Inogen IO-200 Technical Manual
Inogen IO-200 is a portable oxygen concentrator designed to provide supplemental oxygen to patients with respiratory conditions. It features a compact and lightweight design, allowing for easy mobility, and a long battery life, making it perfect for use at home, in the car, or on the go. The device comes with a range of accessories, including a nasal cannula, a carrying bag, and a cart for easy transport.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
▼
Scroll to page 2
of
20
INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL 96-02735-00-01 Revision C CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 1 of 20 1-877-774-9271 This Manual applies to the following Inogen, Inc. products: CONFIDENTIAL and PROPRIETARY AMSR • Inogen One G2 Oxygen Concentrator, model # IO-200 • Inogen Universal Power Supply, model # BA-107 & BA-207 • Battery, model # BA 200 • Battery, model # BA-224 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 2 of 20 1-877-774-9271 TABLE OF CONTENTS INFORMATION FOR PROVIDERS OF THE INOGEN ONE G2 ................................................................................. 4 1. 1.1. CAUTION AND WARNING STATEMENTS ......................................................................................................................... 4 SETTING UP A PATIENT ON INOGEN ONE ............................................................................................................. 6 2. 2.1. RECOMMENDATIONS FOR USE ................................................................................................................................. 6 2.2. SYSTEM COMPONENTS ........................................................................................................................................... 6 2.3. USING THE INOGEN ONE G2 ................................................................................................................................... 7 2.4. SELECTING THE PROPER FLOW SETTING ..................................................................................................................... 7 SERVICING THE INOGEN ONE G2 ........................................................................................................................... 9 3. 3.1. MAINTENANCE BY THE PROVIDER ............................................................................................................................. 9 3.2. MAINTENANCE BY THE PATIENT ............................................................................................................................. 13 3.3. EXPECTED SERVICE REQUIREMENTS ........................................................................................................................ 16 4. INOGEN ONE G2 SYSTEM SPECIFICATIONS ........................................................................................................ 17 5. INOGEN ONE G2 ERROR CODE TABLE .................................................................................................................. 18 6. CONTACTS FOR MORE INFORMATION ................................................................................................................ 20 CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 3 of 20 1-877-774-9271 1. INFORMATION FOR PROVIDERS OF THE INOGEN ONE G2 Thank you for choosing to provide your patients with the Inogen One G2 oxygen concentrator. We are pleased to offer you and your patients one solution for your many oxygen needs. This Technical Manual will familiarize you with provider-specific information regarding the Inogen One G2 Oxygen Concentrator and its accessories. Before reading this Technical Manual, please read and review the Inogen One G2 Patient Manual for description and indications for use of the device. Be sure to thoroughly read all of the information in this manual in its entirety. If you have any additional questions, please see the list of contacts at the end of this Technical Manual. Instructions included in this Technical Manual are intended to help assure that patients are given proper guidance in the use and function of the Inogen One G2 and its accessories. Proper care in relaying this information will not only enhance the user’s experience with the Inogen One G2, but will also protect the patient, prolong the life of the device, and help you avoid unnecessary service calls and complaints from users. 1.1. Caution and Warning Statements You will see Warnings and Cautions throughout this Technical Manual. To ensure effective Oxygen Therapy and proper operation of the Inogen One G2 Oxygen Concentrator, you must observe them carefully. A WARNING indicates that the personal SAFETY of the Patient may be involved. Disregarding a WARNING could result in a significant injury. Be sure that patients WARNING understand all WARNING statements. A CAUTION indicates that a precaution or a service procedure must be followed. CAUTION Disregarding a caution could lead to a minor injury or damage to the equipment. Be sure that patients understand all CAUTION statements. CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 4 of 20 1-877-774-9271 NOTE A NOTE indicates specific information to improve ease of use or maintenance of the equipment. A DESIGN NOTE indicates specific information regarding the design of the Inogen DESIGN One G2 and/or accessories. This information is included in this manual to provide NOTE you with a greater working understanding of the device. This information is not required to operate or maintain the Inogen One G2. The Inogen One G2 User Manual contains additional cautions and warnings. CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 5 of 20 1-877-774-9271 2. SETTING UP A PATIENT ON INOGEN ONE 2.1. Recommendations for Use The Inogen One G2 Oxygen Concentrator is used on a prescriptive basis by patients requiring supplemental oxygen. It supplies a high concentration of oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. You may choose to equip your patients with the Inogen One G2 so it may be used in home, institution, vehicle and various mobile environments. Availability of an alternate source of oxygen is recommended in case of power NOTE outage. Several certifying bodies for Home Health Care Providers require that back-up oxygen be available to the patient. Supplemental oxygen cylinders or extra Inogen One G2 Batteries may satisfy these requirements. Oxygen demand of some patients, particularly those with high breathing rates CAUTION and high flow settings, may exceed the capabilities of the Inogen One G2. Inogen suggests that each patient be titrated to assure that the Inogen One G2 is an appropriate solution for their needs. 2.2. System Components The following are standard components of the Inogen One system: o Inogen One G2 Oxygen Concentrator o 12 cell Battery o AC/DC Power Supply o Cart o Carry Bag o Cannula Patient Set-Up. To properly set up a patient on the Inogen One G2 System, you may need to provide: o Pulse Oximeter (for titration, not included) CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 6 of 20 1-877-774-9271 o Extra cannulas (not included) 2.3. Using the Inogen One G2 To quickly configure the Inogen One G2 for patient use: 1. Slide the battery onto the bottom of the system until the latch clicks into place. 2. Connect AC or DC power cord to the power supply. 3. Connect the power supply to the Inogen One G2. 4. Plug into closest AC or DC outlet. 5. Turn on the Inogen One G2 by pressing the on/off button. 6. Set the Inogen One G2 to correct flow setting prescribed by the physician or clinician by pressing the + (increase flow) or – (decrease flow) button. 7. Attach cannula to the metal hose barb located next to the handle of the Inogen One G2. For further information regarding the use of the Inogen One G2, please consult the User Manual. 2.4. Selecting the Proper Flow Setting 2.4.1. Bolus Volumes Specification All oxygen conserving devices (OCD’s) function differently, and therefore it is prudent to titrate patients for any new conserving device. Delivery timing, bolus volume, and oxygen concentration all contribute to a patient’s fraction of inspired oxygen (FiO2), and therefore to the OCD’s efficacy at maintaining the patient’s blood oxygen saturation. As an oxygen concentrator, the Inogen One G2 does not contain a finite stored volume of oxygen, such as with compressed gas or liquid cryogenic systems. The Inogen One G2 can provide oxygen to the patient as long as a source of electricity is available. However, because the oxygen is being produced as it is used, supply of oxygen is rate- limited. The Inogen One G2 delivers up to 900 ml/min of 90% oxygen. At each flow setting, the Inogen One generates a specific amount of oxygen (180ml per setting), and the on-board OCD attempts to deliver all of this product to the patient. This is equivalent to a conserving ratio of 5.6 at all flow settings and breathing rates. CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 7 of 20 1-877-774-9271 Slower breathing patients will receive larger boluses, and faster breathing patients will receive smaller boluses. This method of bolus volume determination is similar to what is experienced by a DESIGN NOTE patient using a continuous flow concentrator – actual alveolar oxygen inspiration is more closely linked to flow setting than to breathing rate. In general, the Inogen One G2 delivers 12ml per bolus per flow setting at 15 breaths per minute (180ml/min per flow setting). The following table summarizes the bolus volumes delivered by the Inogen One OCD at 20C and Sea Level: Flow setting 1.0 2.0 3.0 4.0 5.0 Flow Rate (ml +/- 10%) 180 180 360 540 720 900 10 BPM (ml +/- 10%) min nominal max 16.2 18.0 19.8 32.4 36.0 39.6 48.6 54.0 59.4 64.8 72.0 79.2 81.0 90.0 99.0 20 BPM (ml +/- 10%) min nominal max 8.1 9.0 11.6 16.2 18.0 23.3 24.3 27.0 34.9 32.4 36.0 46.6 40.5 45.0 58.2 25 BPM (ml +/- 10%) min nominal max 6.5 7.2 7.9 13.0 14.4 15.8 19.4 21.6 23.8 25.9 28.8 31.7 32.4 36.0 39.6 Bolus Volume is tuned to provide the correct bolus volume when delivered DESIGN through a nasal cannula such as the Salter Labs 1600Q. If bolus volumes are NOTE measured without a cannula or with a different type of nasal cannula, bolus volumes will vary from values stated in the table above. 2.4.2. Trigger sensitivity The conserver will trigger when the negative pressure at the internal sensor reaches approximately -0.12 cm H20. This low trigger sensitivity allows for breath detection of very shallow breathing, but may also result in the occasional trigger due to motion of the cannula or motion of the concentrator. 2.4.3. Flow Setting Selection Relative to Physician Prescription Inogen has labeled each of the five settings to provide a guideline for matching the setting of the device to the continuous flow prescription issued by the physician. Actual correlation is dependent upon the patient’s breathing rates, inspiratory tidal CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 8 of 20 1-877-774-9271 volume, and other physiologic factors. Inogen suggests that each patient be titrated (a) while sedentary, (b) while active or ambulatory. 2.4.4. Use with the Inogen Satellite Conserver The Inogen One G2 is not compatible with the Inogen One Satellite Conserver 3. SERVICING THE INOGEN ONE G2 3.1. Maintenance by the Provider 3.1.1. The Inogen One G2 is only intended to be repaired by authorized repair centers. If a repair is required, please contact Inogen to locate your nearest authorized repair center. If you would like to become an authorized repair center, please contact Inogen for access to certification training, service instructions, component part lists and the necessary repair equipment. 3.1.2. Checking Life Clock, Serial #, and SW version To check usage, press and hold the Alarm Bell Button for 5 seconds. The concentrator’s display will show the hour meter, the unit’s serial number, and the software version installed on the device. Do not disassemble the Inogen One G2 or any of the accessories or attempt any maintenance other than tasks described in this Technical Manual unless you have NOTE completed a training course through Inogen. Disassembly of the Inogen One G2 or any of the accessories without proper training certification will void the product warranties. Contact Inogen for information about receiving proper training and certification for service of the Inogen One G2 and accessories. 3.1.3. Suggested Materials for Regular Maintenance To perform regular field maintenance (by a technician) on the Inogen One G2 System, you may need: o Replacement cannulas (RP-128) o Replacement intake screens (RP-200) o Replacement Inogen One G2 Batteries (BA-200 or BA-224) o Hose barb removal tool (Spanner Wrench, RP-102) CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 9 of 20 1-877-774-9271 o Replacement product filters (RP-101) o USB storage drive for data log transfers (such as PNY PN# P-FD2GBATT2-SF) During a normal field maintenance visit, the technician may elect to turn off the concentrator for approximately 30 minutes. If the patient requires oxygen during NOTE this period, Inogen recommends making arrangements to bring an extra oxygen supply (such as a supplemental Inogen One G2 Concentrator). 3.1.4. System Inspection At the start of any maintenance visit: 1. Be sure to ask the patient if they have experienced any difficulties in operating the equipment. 2. Be sure to ask the patient if they have observed any malfunctions or changes in characteristics of the equipment. 3. Visually inspect the device, batteries, and accessories for cracks or other damage. 4. Feel the sides of the device for vibration and listen for unusual noises, rattles, or other signs that the device requires service. Discovery of cracks or other types of external damage may be indicative of other internal damage that may not be visible. If such external damage is discovered, be CAUTION certain to inquire as to how it occurred, and whether any changes in the device have been noticeable since its occurrence. If you have any concern over the safety of the device, arrange for equipment servicing. 3.1.5. Product (Bacterial) Filter Oxygen concentrator safety standards (ISO 8359) require use of a filter in the product line capable of removing particles larger than 10 microns. This filter is intended to protect the user from bacterial infection, and from inhalation of small particles in the product gas flow. The Inogen One G2 includes a product filter, conveniently located behind the removable cannula nozzle fitting. Inogen suggests that this filter be replaced between patients. CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 10 of 20 1-877-774-9271 To replace the Product Filter: 1. Use the Cannula Nozzle Fitting Tool (available from Inogen as RP-102) to access the product filter. The tool has two prongs which mate with two indentations located on the surface of the metal nozzle fitting on the Inogen One G2. 2. Carefully remove the nozzle fitting by unscrewing it in the counter-clockwise direction. 3. The filter, a hard plastic disk with a silicone gasket on its outer edge, will be visible in the recess once the hose barb is removed. 4. Remove the filter, and inspect the recess to make sure it is free of debris. 5. Install a new replacement filter. 6. Carefully screw the nozzle fitting back into the recess (clockwise) until it bottoms out on the filter gasket. Take care to squarely screw the nozzle fitting into the threads, and not to over tighten. Failure to inspect and replace the product filter may result in the filter becoming WARNING clogged or obstructed over time, and in reduced delivery of oxygen to the patient. 3.1.6. Data Logging Operating data from your Inogen One G2 concentrator are periodically recorded to flash memory located within the device. Data is stored for a period of approximately 612 months; new data replaces the oldest data as it is collected. Additionally, the device records errors and system information at the time of the error which is useful in diagnostics. 3.1.6.1. The following information is a list of recorded parameters that can be used to evaluate the device and its usage: CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 11 of 20 1-877-774-9271 • Real Time Clock • Life Clock • Flow Setting • Breaths per Minute • Low, Medium , and High Priority Errors • Ambient Pressure • System Temperature • Battery Status • Charging Status 3.1.7. Data Download To download data from the Inogen One G2, you will need a USB mass storage device. If one of the following drives cannot be obtained, the drive must be less than 2GB in capacity and have a native file format of FAT32. The capacity of the drive and the file system alone do not ensure compatibility with the Inogen One G2. Compatible USB drives include: • PNY Attache 2GB PN: P-FD2GBATT2-SF • SanDisk Cruzer Micro 2GB PN:SDCZ6-2048RB • Kingston Data Traveler G3 2GB PN:DTIG3/2GBZ • Kingston DataTraveler 102 2GB PN:DT102/2GB • Kingston DataTraveler DT101 2GB PN: DT101G2/2GB • Dane-Elec DNL 2GB PN:DNLDAZMP02GCAW • Transcend JetFlash V30 2GB PN:TS2GJFV30 • Patriot Signature 2GB PN: PSF2GUSB To collect data: a) With the Inogen One G2 off, insert the USB mass storage device into the USB port behind the Gross Particle Screen b) Plug in external power to the concentrator c) Press the “alarm bell” and the “light” buttons for 5 seconds CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 12 of 20 1-877-774-9271 d) The administrative mode menu will show normal. Press the “+” or “-“ buttons until the LCD reads “Data Log” e) Press the “alarm bell” button twice and the LCD will read “Transferring” f) When the data is transferred, the LCD will read “Success” g) If the LCD screen reads “Failure” try a different type of USB mass storage device 3.2. Maintenance by the Patient 3.2.1. Cannula Replacement The nasal cannula should be replaced on a regular basis. A single lumen cannula of four to 25 feet in length must be used. Inogen has certified its performance data with the Salter Laboratories 1600Q cannula. CAUTION Do not use total cannula tubing length exceeding 25 feet with the Inogen One G2 unless proper saturation has been verified by a clinician. 3.2.2. Intake Filter Cleaning At the front of the Inogen One G2, ambient air passes through a gross particle screen that removes dust fragments. This particle screen must be cleaned on a weekly basis to ensure adequate air flow through the device. If the particle screen is not cleaned frequently, the life expectancy of the concentrator will be shortened due to higher internal operating temperatures. Particle screens should be cleaned using a mild detergent (e.g., dishwashing solution, such as DawnTM) and water solution; be sure the filter is rinsed in water and allowed to air dry before reuse. Additional replacement filters may be obtained from Inogen (RP200). There are other types of filters inside the Inogen One G2 that provide additional filtration. Maintenance of these filters is not required under normal operating conditions. CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 13 of 20 1-877-774-9271 3.2.3. Surface Cleaning The outside case should be cleaned using a cloth dampened with a solution of mild detergent (e.g., dishwashing detergent, such as DawnTM) and water or disinfecting towelettes (e.g., DisCide Ultra). 3.2.4. Battery Care and Maintenance The Inogen One G2 Lithium Ion Battery requires special care to ensure proper performance and long life. Use only Inogen One G2 Batteries with the Inogen One G2 Oxygen Concentrator. The Inogen One G2 Oxygen Concentrator adjusts its oxygen production rate to match the oxygen demand specified by the user flow setting. When the device DESIGN is used at lower settings, its battery life is extended. Additionally, at lower NOTE flow settings, the concentrator does not generate as much heat and noise, draw as much electric current from external power supplies, and many system components do not wear as quickly. Initial Charging and subsequent charging The Inogen One G2 battery requires an initial charge to full capacity before use. Instruct the patient to avoid running the Inogen One G2 on battery power until this initial charging has been completed. This charge typically takes 2 hours depending on the flow setting. The Inogen One G2 may be run on AC or DC power and used by the patient during this initial charging period Normal Charging The battery is operating properly when a battery icon is displayed on the LCD screen. CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 14 of 20 1-877-774-9271 The Inogen One G2 batteries do not have a “memory” like older NiCad battery packs. The Inogen One G2 batteries can be partially charged and discharged without damaging the battery packs. Effect of Temperature on Battery Performance The Inogen One G2 battery powers the Inogen One G2 Oxygen Concentrator from 2 to 4 hours using the standard 12 cell pack (BA-200) under most environmental conditions. To maintain maximum run-time of the battery, users should avoid running in temperatures less than 40oF (4oC) or higher than 95oF (35oC) for extended periods of time. The number of cycles that the battery will last is highly dependent upon the temperature at which the battery is charged. Inogen recommends that batteries not be charged at room temperatures exceeding 75oF (24oC). The BA-200 and BA-224 will typically achieve 500 charge/discharge cycles while DESIGN NOTE retaining 80% of their original capacity if proper battery care is taken. The BA-200 and BA-224 will typically achieve 500 charge/discharge cycles while retaining 80% of their original capacity if proper battery care is taken. Instruct the user that attempting to charge their battery while attached to a running Inogen One G2 concentrator at elevated temperatures may result in a charging error and termination of the battery charge. In this case, they may either (a) place the concentrator into a cooler environment, (b) charge the battery while the Inogen One G2 is turned off and plugged into an AC power outlet, or (c) remove the battery and charge using the External Battery Charger (optional accessory). Improper charging may also result in decreased run time on a single battery charge and reduced life expectancy of the battery. Battery Time Remaining Clock The Inogen One G2 continuously displays battery time remaining. Explain to the user that this displayed time is only an estimate, and the actual time remaining may vary from this value. CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 15 of 20 1-877-774-9271 To avoid running out of battery power unexpectedly, users should regularly monitor the displayed battery time remaining and/or carry a back-up power supply (extra charged battery or AC Power Supply). Storage Instruct patients to remove the battery from the Inogen One G2 when it is not in use to avoid inadvertent discharge. Leaving a battery attached to an idle Inogen One G2 for prolonged periods will result in battery damage that will severely shorten the expected life of the battery. When the concentrator is off but the battery installed, the battery will continue to DESIGN NOTE provide a small amount of power to the concentrator’s microprocessor. This power draw will empty a full battery in approximately 20 days. Leaving a Battery in the Inogen One G2 while the device is unused will irreversibly damage the battery. After such storage, the battery may not be able to recharge CAUTION or its life cycle and/or capacity will be greatly diminished. A full battery can be damaged in as little as 20 days if left in the concentrator while not plugged in. Instruct users to avoid storing the Inogen One G2 battery in extreme temperatures, below -4oF (-20oC) or above 140oF (60oC), for any amount of time. They should avoid leaving batteries in automobiles, where these temperatures can be regularly exceeded. Storage of the Inogen One G2 battery in a cool, dry location will help to maximize the longevity of the battery. 3.3. Expected Service Requirements The following table is provided as an estimate only and assumes typical environmental conditions for temperature, humidity, and air pollution. Any smoking around the device will severely shorten its life expectancy. Please refer to product warranty coverage terms. CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 16 of 20 1-877-774-9271 Service Item Battery Estimated Service Requirement Frequency 500 full charge/discharge cycles (to approximately 80% capacity) Compressor Function of flow setting and time 17,000 run hours @ flow setting 5 20,000 run hours @ flow setting 2 Oxygen Sensor No Service Requirement Molecular Sieve 36 to 60 months depending on usage Valves 20,000 run hours Intake Filter 20,000 run hours Cooling Fan 20,000 run hours 4. INOGEN ONE G2 SYSTEM SPECIFICATIONS INOGEN ONE G2 CONCENTRATOR Length Cm In 27.2 10.7 27.2 10.7 Width Cm In 9.9 3.9 9.9 3.9 Height Cm In 22.2 8.75 24.1 9.50 Weight: No battery 12 cell battery 27.2 10.7 9.9 3.9 26.0 24 cell battery 27.2 10.7 11.4 4.50 25.4 In Bag 34.3 13.5 20.3 8.00 40.6 On Cart Approximately 7.25 pounds (includes 12 cell battery) Noise: <38 dB on flow setting 2 Warm-Up Time: Approximately 2 minutes Oxygen Concentration: 90 + 6/-3% Flow Control Settings: 1,2,3,4,5 Dimensions: 10.25 10.00 16.00 AC/DC Power Supply Power: CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC AC Input: 100 to 240 VAC, 50 to 60 Hz, 1.0A to 0.5A DC Input: 13.5 to 15 VDC, 10A Output 19 VDC, 90 W INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 17 of 20 1-877-774-9271 Rechargeable Battery: Battery Duration: Battery Charging Time: Voltage: 12.0 to 16.8 VDC, Approximately 2 to 3.5 hours for 12 cell battery Approximately 4 to 7 hours for 24 cell battery Approximately 2.25 to 4.25 hours for 12 cell battery Approximately 4.25 to 8.00 hours for 24 cell battery Environmental Ranges Temperature: 40 – 104oF (4 – 40oC) Intended for Use: Humidity: 0% to 95%, non-condensing Classifications Mode of Operation: Continuous Duty Type of Protection Against Electrical Shock: Class II Degree of Protection Against Electrical Type BF Shock: Not intended for cardiac application Degree of Protection Against Ingress of Concentrator: IP 20 Water: Carry bag: Degree of Safety for Application in Presence of Anesthetic Gases: IP 22 Not suitable for such applications 5. INOGEN ONE G2 ERROR CODE TABLE If more than one alarm is triggered simultaneously, the error code number will be the sum of the error codes given below. Error Code Explanation Possible Causes & Troubleshooting Number 001 Instructions Low voltage on power supply Replace the power supply or remove the battery to see if the concentrator was running on battery power 002 Software error Unplug the external power and remove the battery to clear the error and then restart the concentrator 004 B/C Pressure Error The concentrator was shut down due to a pressure error. Ensure that the CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 18 of 20 1-877-774-9271 battery is sufficiently charged. If the error persists, return the concentrator for service.. 016 System electric current out of There may be a fault with the motor or specifications the battery communication. If the problem occurs at the end of a battery discharge, recharge the battery. If the error is persistent, return the concentrator for service. 128 System Hot Signal or reading out of spec. A sensor has given an out of range Applies to signals from user reading and might need to be serviced interface and other internal if the error persists after restarting the connections. concentrator. temperature out of specifications Remove the concentrator from the System Cold carry bag and restart the concentrator after 10 minutes to cool down. If the error persists, the fan may have failed. If the System Cold message appears, allow the concentrator to warm up for 15 minutes in a room temperature environment and then restart the concentrator. Battery Hot Battery over temperature during The concentrator might need to be Warning discharge moved to a cooler location with Battery Hot improved ventilation Shutdown Oxygen Low The Oxygen purity is < 82% for 30 The sieve beds may require servicing minutes Oxygen Error The Oxygen purity is < 50% for 10 The concentrator has a malfunction minutes such as a leak or a failed valve O2 Sensor The oxygen sensor is returning If the problem persists, return the Failure false readings concentrator for service O2 Delivery A breath is detected, but no bolus Check for kinks or blockages in the Error is delivered tubing between the product manifold CONFIDENTIAL and PROPRIETARY 96-02735-00-01 revC AMSR INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 19 of 20 1-877-774-9271 and the cannula barb. If the problem persists, return the concentrator for service. Battery Error The system is not properly Remove and reinsert the battery to communicating with the battery clear the error. If the error persists, replace the battery Service Needed External Power The compressor has reached The compressor is at its end of life or maximum speed there is a leak in the concentrator The input voltage is less than 17V Remove power to the concentrator and Low remove and reinsert the battery. Try another external power source such as DC power 6. CONTACTS FOR MORE INFORMATION Inogen Website: Inogen : Patient Issues: http://www.inogenone.com Corporate: http://www.inogen.net 326 Bollay Dr Goleta, CA 93117 1-866-765-2800 [email protected] CONFIDENTIAL and PROPRIETARY AMSR 96-02735-00-01 revC INOGEN ONE G2 OXYGEN CONCENTRATOR TECHNICAL MANUAL oxygenconcentratorstore.com Page 20 of 20 1-877-774-9271
advertisement
Key Features
- Lightweight design
- Compact size
- Long-lasting battery
- Adjustable flow settings
- Quiet operation
- Easy to use
- Durable construction
- Portable oxygen concentrator
- Home and mobile use
Frequently Answers and Questions
How long does the battery last?
The battery life is approximately 2 to 4 hours using the 12 cell pack (BA-200) depending on the flow setting and environmental conditions.
How often does the intake filter need to be cleaned?
The intake filter should be cleaned weekly to ensure adequate airflow. It can be cleaned with mild detergent and water and allowed to air dry.
What are the recommended environmental ranges for the Inogen One G2?
The recommended temperature range is 40 to 104 degrees Fahrenheit (4 to 40 degrees Celsius) and the recommended humidity range is 0% to 95% non-condensing.