Caution
Do not mark on the model and other components
with pen or leave any printed materials
contacted on surface. Ink marks on the
product cannot be removed.
PH-10
BMD Chart Phantom UHA
Instruction
Manual
Contents
Introduction
Manufacturer’s Note..................................................P.1
DOs and DON’Ts.........................................................P.1
Before You Start
Set Includes....................................................................P.2
Specifications.................................................................P.2
How to Use
How to use......................................................................P.3
Introduction
Manufacturer’s Note
Manufacturer’s Note
Features
PH-10
BMD Chart Phantom UHA is a Bone Mineral Density (BMD) Chart for microdensitometry (MD) method,
which is widely used to assess bone atrophy in metabolic disorders and skeletal dysplasia. The phantom allows
quantitative assessment of bone mineral using amount of hydroxyapatite in the phantom blocks.
The product is designed for training of healthcare professionals. Usage for any other purpose or the way of use
not described in this manual is strongly discouraged.
Applications
Densitometry
Dos and Don'ts
Handle the phantom and the components with care
Be sure not to drop or add shock to the phantom and its
components, since the product uses special resins and
heavy, such shock may cause damage to the products or
environment.
Do not let ink from pens, oriented materials, this
manual or other sources contact with the product’s
surface, as they cannot be cleaned off.
Storage dark, cool and dry space will help prevent
the product’s colors from fading and product from
deforming.
Never use the organic solvent such as paint thinner
to clean the skin, as this will damage the simulator.
The product may be cleaned with a wet cloth if
necessary, using mildly soapy water or diluted
detergent.
The color of the product’s surface may change
across the ages, this does not affect the quality of
its performance.
1
Before You
Start
Set Includes
Set Includes
Before your first use, ensure that you have all components listed below.
1 chart phantom
30 x 210x 15 mm
Contains 21 pieces of phantom blocks of 30 x 10 x 15 mm
1 storage case
Specifications
Materials
Base materials: Polyurethane resin (Density: 1.06)
Additive: High-purity hydroxyapatite (Density: 2.7)
Each phantom block contains 0 to 400mg/cm3 hydroxyapatite by step of 20mg/cm3.
Compound ratio of Hydroxyapatite in each step (block) *theoretical value
Mg/cm3
20
40
60
80
100
120
140
160
180
200
220
240
260
280
300
320
340
360
380
400
2
How to Use
How to Use
How to Use
1.Take out the phantom blocks from the acrylic case.
It is also possible to radiograph the phantom with the acrylic case, though, be sure to
maintain the same condition for future assessment.
2. Radiograph the phantom together with the right hand of the patient. (see fig.1).
3 Measure the density of the middle point of the second metacarpal on the acquired image
using densitometer.
4. Measure the density of each phantom block and compare the value with that of the
second metacarpal.
5. Following above procedures, assess variation of BMD by days, months, years, or
difference before and after administration of medicines.
X
Measure at the mid point between X and Y
Fig. 1
3
Caution
Do not mark on the model and other components
with pen or leave any printed materials
contacted on surface. Ink marks on the
product cannot be removed.
For inquiries and service, please contact your distributor or KYOTO KAGAKU CO., LTD.
The contents of the instruction manual are subject to change without prior notice.
No part of this instruction manual may be reproduced or transmitted in any form without permission from the manufacturer.
Please contact manufacturer for extra copies of this manual which may contain important updates and revisions.
Please contact manufacturer with any discrepancies in this manual or product feedback. Your cooperation is greatly
appreciated.