iCare DRS Fundus Imaging System User Manual
Icare DRS is a digital fundus camera used to capture high-quality images of the retina without the need for pupil dilation. It features a 45° x 40° field of view and multiple internal fixation targets to capture different retinal fields. The device offers automated image acquisition, manual alignment options, and a convenient user interface for reviewing the captured images. Icare DRS is a valuable tool for eye care professionals in diagnosing and managing various retinal conditions.
PDF
Download
Document
Advertisement
Advertisement
- Operating Manual CenterVue DRS Operating Manual Pag. 1 of 59 MANUAL INFORMATION Date of release: Revision number: Reference software version: Manufacturer: October 30th 2022 29 V2.9 CenterVue S.p.A. Via San Marco 9h, 35129 Padova – Italy Tel. +39 049 501 8399 Fax +39 049 501 8398 The information in this manual is correct at the date of issue of the manual. The device configuration can change as product improvements are incorporated and this manual may not exactly depict your device. Please contact the local distributor if you have any questions about differences. The original language of the DRS Operating Manual is English: in case of conflict of terms, the English version shall prevail. SUMMARY 1. INTRODUCTION ....................................................................................................................................... 4 2. DEVICE...................................................................................................................................................... 5 3. LABELING ................................................................................................................................................. 6 4. SYMBOLS.................................................................................................................................................. 7 4.1 Symbols adopted on the device ............................................................................................................ 7 4.2 Symbols adopted in this manual ............................................................................................................ 7 5. PREPARING THE DRS ............................................................................................................................. 8 6. PREPARING THE PATIENT ..................................................................................................................... 9 7. WARNINGS AND PRECAUTIONS ......................................................................................................... 10 8. NOTES TO OPERATOR ......................................................................................................................... 11 9. PERFORMING THE TEST ...................................................................................................................... 12 9.1 Adding a new patient ........................................................................................................................... 12 9.2 Selecting an existing patient ................................................................................................................ 12 9.3 Starting the image acquisition process ................................................................................................ 13 9.4 Automated acquisition ......................................................................................................................... 15 9.5 Manual alignment ................................................................................................................................ 16 9.6 Anterior Eye ......................................................................................................................................... 17 9.7 Stereo exams ....................................................................................................................................... 18 9.8 Warning messages .............................................................................................................................. 18 10. REVIEWING THE IMAGES ..................................................................................................................... 19 10.1 Patient record screen ........................................................................................................................... 19 10.2 Full image screen ................................................................................................................................ 20 11. REMOTE VIEWER .................................................................................................................................. 23 12. PRINTOUT............................................................................................................................................... 24 12.1 Single image printout ........................................................................................................................... 24 12.2 Dual image printout .............................................................................................................................. 24 13. RETINAL FIELDS .................................................................................................................................... 26 14. BACKUP .................................................................................................................................................. 27 14.1 First-time, total, backup ....................................................................................................................... 28 14.2 Incremental backups ............................................................................................................................ 30 14.3 Removing a backup disk...................................................................................................................... 31 14.4 Partial backups .................................................................................................................................... 31 14.5 Restore ................................................................................................................................................ 32 15. SETTINGS ............................................................................................................................................... 34 CenterVue DRS Operating Manual Pag. 2 of 59 15.1 Fields ................................................................................................................................................... 34 15.2 Exam .................................................................................................................................................... 34 15.3 Network ................................................................................................................................................ 36 15.4 System ................................................................................................................................................. 42 15.5 About.................................................................................................................................................... 46 16. AUTOMATIC SOFTWARE UPDATE ...................................................................................................... 47 17. SYSTEM SHUTDOWN ............................................................................................................................ 47 18. CLEANING............................................................................................................................................... 48 18.1 Front lens ............................................................................................................................................. 48 18.2 Patient cushions .................................................................................................................................. 49 18.3 Touch screen panel ............................................................................................................................. 50 18.4 Plastic shells ........................................................................................................................................ 50 19. MAINTENANCE ....................................................................................................................................... 50 20. ELECTROMAGNETIC COMPATIBILITY ................................................................................................ 51 21. FCC (USA) and IC (Canada) radio certification ....................................................................................... 51 22. TECHNICAL SPECIFICATIONS ............................................................................................................. 52 23. DISPOSAL ............................................................................................................................................... 53 24. TROUBLESHOOTING AND ERROR MESSAGES................................................................................. 54 APPENDIX A - DICOM Statement .................................................................................................................. 56 APPENDIX B – AUTOMATED MOSAIC FUNCTION ..................................................................................... 58 CenterVue DRS Operating Manual Pag. 3 of 59 1. INTRODUCTION The Digital Retinography System (DRS) is a fundus camera for the acquisition of digital images of the retina without the use of a mydriatic agent. In particular, the DRS allows to acquire color images of the retina over a field of view of 45° x 40°, in a fully automatic mode. Seven different retinal fields can be framed using multiple internal fixation targets. The clinical interpretation of the images acquired by the DRS is restricted to licensed eye care practitioners. The process of making a diagnosis using DRS results is the responsibility of the eye care practitioner. A device-specific training is required for any operator to become able to use the system. Federal laws (US) restrict this device to sale by or on the order of a physician or a properly licensed practitioner. CenterVue DRS Operating Manual Pag. 4 of 59 2. DEVICE Forehead rest = applied part Chin rest applied part = Touch screen Operator side Patient side Figure 1 - DRS left side view Power switch Power inlet USB ports Fuses Ethernet port Figure 2 - Detail of the device right side The device is supplied with: 1. 2. 3. 4. 5. 6. 7. 8. 9. CenterVue Front lens cap; Power cord; Spare fuses; Dust cover; USB extension cord; Air blower; Removable forehead rest; Cleaning paper; Cleaning cloth. DRS Operating Manual Pag. 5 of 59 3. LABELING Labelling might be subject to changes depending on local regulatory requirements concerning labelling. CenterVue DRS Operating Manual Pag. 6 of 59 4. SYMBOLS 4.1 Symbols adopted on the device Symbol Explanation Manufacturer Data Manufacturing Date (year/week of production) Serial Number Electronic and electric devices must be recycled Refer to Instruction Manual CE mark: the device complies with the essential requirements of the European Medical Devices Directive 93/42/EC Type B Applied Part Generic Warning 4.2 Symbols adopted in this manual Note Warning CenterVue DRS Operating Manual Pag. 7 of 59 5. PREPARING THE DRS Read carefully and thoroughly paragraph 7 WARNINGS AND PRECAUTIONS before proceeding with first use. The test should be performed in conditions of semi-darkness to facilitate natural dilation of the pupil. To set up your DRS follow these steps: • extract the device from its box; • place the DRS on a suitable electrical table1; • connect the power cord provided with the device to the power inlet (see Figure 2); • optionally connect a compatible printer1 to any of the USB ports (see Figure 2); • attach the silicon forehead rest (included in the tool box) on the forehead rest frame as shown in the pictures below (Figure 3 and Figure 4); Figure 3 - Forehead rest patient side Figure 4 - Forehead rest rear side Time zone configuration The first time the DRS is powered on, the operator is requested to set the time zone. The configuration request (Figure 5) will appear every time the DRS is started, until the time zone is set (see §14.4). Figure 5 – Request to configure time zone Request for camera auto-tuning After a software upgrade to the latest version a message as in Figure 6 may appear. Such message appears only once when a new software is started for the first time: please apply the lens cap and press ok, then wait until the completion message (Figure 7) appears. 1 Not provided with the DRS CenterVue DRS Operating Manual Pag. 8 of 59 Figure 6 – Request to perform camera auto-tuning Figure 7 – Camera auto-tuning completion message 6. PREPARING THE PATIENT This paragraph explains how to prepare a patient for the DRS test. There are no restrictions as to the selection of the patients undergoing DRS testing. The DRS is a non-mydriatic device (minimum pupil diameter 3.8 mm), so there is no need to dilate the patient’s pupil. Patient contacting parts are indicated in Figure 1. Patients wearing Contact Lenses: image quality may be affected by contact lenses. It is recommended to have contact lenses removed before imaging the patient. Before the test inform the patient about the following: 1) the test is non-invasive, in particular the device will never touch your eye and you will only see a flash of light when a photograph is taken; 2) find a comfortable position, keeping the chin and forehead firmly pressed against the rests; 3) at the beginning of each test, the device will move around to find your pupil: this is absolutely normal; 4) when the test starts look straight in front of you and when a small green circle becomes visible anywhere within the field look steadily at it; 5) open your eyes wide so that eyelids do not interfere and try to not blink during the test; 6) do not move during the test; 7) if you follow these instructions the acquisition of a single photo will take less than 30 seconds. CenterVue DRS Operating Manual Pag. 9 of 59 7. WARNINGS AND PRECAUTIONS The following precautions are important to use the device in safety: • Federal laws (US) restrict this device to sale by or on the order of a physician or a properly licensed practitioner. • The clinical interpretation of DRS images is restricted to licensed eye care practitioners. • A device-specific training is required for any operator to become able to use the system. • Do not open the device: this could lead to electric shocks or damage to the system. • Do not use the device in the event that the cover or other parts of the device have been removed. • Only technicians authorized by CenterVue may service the DRS. CenterVue cannot be held responsible for system safety should the DRS be opened, repairs carried out, third parties software be installed, or parts be replaced by unauthorized persons. • Do not expose the device to water: this could lead to fire or electric shock. • Stand clear from moving parts during operation. • The device is supplied with an earth ground by means of a protection conductor contained inside the power supply cable. Before turning on the system, make sure the power supply socket is correctly grounded to avoid the risk of electric shock. • The room where the DRS is operated must respect IEC or ISO safety standards relative to medical use of a room or area. • The DRS must NOT be used in an oxygen rich environment or in presence of flammable anesthetics. • External equipment connected to the DRS within the patient environment must be compliant with IEC 60601-1. Equipment not complying with IEC 60601-1 shall be kept outside the patient environment and complying with IEC 60950-1. Any person who connects external equipment to signal input, signal output or other connectors of the DRS has formed a Medical Electrical System according to the definition of IEC 60601-1 and is therefore responsible for the system to comply with the requirements of IEC 60601-1, § 16. If in doubt, contact the local distributor for any additional information • Data Backup: the DRS contains Patient Health Information. IT IS A RESPONSIBILITY OF THE OPERATOR (END USER) TO KEEP AN UPDATED COPY OF THE DATA GENERATED BY THE DRS, BY PERIODICALLY USING THE BACKUP UTILITY IN ORDER TO BE PROTECTED AGAINST UNPREDICTABLE DATA LOSS. See par.14.4.3 The following precautions are particularly relevant to prevent use errors: • The device must be placed in a room which is not exposed to adverse chemicalphysical conditions, such as the presence of sulfur, salt, dust, direct sunlight, lack of ventilation, high humidity, sudden temperature drops or peaks. The safety and/or effectiveness of the instrument cannot be guaranteed if these conditions are not fulfilled. • The DRS needs to be operated in a semi dark room (except when taking Anterior Eye photos). CenterVue DRS Operating Manual Pag. 10 of 59 • The DRS needs to be operated under the following environmental conditions: Temperature: from 10 °C to 40 °C (50 °F to 104 °F) / Humidity (max): 90% not condensing • The DRS needs to be stored under the following environmental conditions: Temperature: from -10 °C to 60 °C (14 °F to 140 °F) / Humidity (max): 90% not condensing • A not optimal pupil alignment may cause the acquisition of an image of the retina that has a growing light crescent on a board. Such growing light should be considered an artifact and not a feature of the retina under examination. • Do not leave the front lens uncovered while the device is not in use. Note on auto-alignment: The auto-alignment process may fail in some circumstances such as: • long eye-lashes; • presence of make-up; • room too bright; • direct light on a patient’s face; • very big pupil. Under such circumstances the exam should be repeated after removing the cause of the failure. 8. NOTES TO OPERATOR No specific skills are required to operate the DRS. Minimal device-specific training is required to become able to use the device. The operator needs to be acquainted with the following concepts: • pupil: is the opening located in the center of the iris, of variable diameter, which allows light to enter the eyeball. The pupil naturally is open (dilated) and contracts when struck by light. If the pupil is too small the image quality may be impaired. • retina: The inner layer of the eyeball. • fixation / fixating: the ability of a subject to stare at a specific point in space and specifically the DRS internal fixation targets. IT IS A RESPONSIBILITY OF THE OPERATOR TO KEEP AN UPDATED COPY OF THE DATA GENERATED BY THE DRS, BY PERIODICALLY USING THE BACKUP UTILITY, IN ORDER TO BE PROTECTED AGAINST UNPREDICTABLE DATA LOSS. CenterVue DRS Operating Manual Pag. 11 of 59 9. PERFORMING THE TEST This paragraph explains how to operate the DRS to perform a complete test. When the device is turned on and the boot process is over, the Home screen shows up (see Figure 8). Figure 8 - Home screen Several options are available at this point, such as: Function Adding a new patient (see 9.1 below) Selecting an existing patient from the list (see 9.2 below) Icon - Starting image acquisition without identifying the patient (see 9.3 below) Shutting the system down Entering the system settings (see 15) 9.1 Adding a new patient To add a new patient to the list, type the last name (mandatory), first name and date of birth in the corresponding boxes. Optionally select the gender and type a Code of your choice and any patient related notes. Then click on save. 9.2 Selecting an existing patient To select an existing patient from the list, click on the PATIENT LIST button: the complete list of stored patients will show up (see Figure 9). CenterVue DRS Operating Manual Pag. 12 of 59 The following mechanisms are available to find a specific patient: • scroll the list by dragging the screen up or down; • sort the list by one of the available fields (the list can be sorted by last name, date of birth or date of last visit) by clicking on the corresponding arrows ( for descending, for ascending) located at the header of the list (by default patients are sorted by their numeric ID, in ascending order); • search for a specific patient by typing the initials of the last name or first name or code in the search box. Once the desired patient has been identified in the list, click on the corresponding line to select and open the Patient Record screen (see Figure 22). Figure 9 - Patient list screen 9.3 Starting the image acquisition process When a new patient has been created or an existing one selected or none of the above (anonymous), it is possible to start the image acquisition process. Before proceeding, be sure that the patient received complete explanations concerning the test and what she/he is required to do, as explained at par. 6 above. Then choose among the different options / parameters: a. Click OU to acquire images from both eyes (default), OD for the right eye and OS for the left eye; b. Select SINGLE FIELD to acquire one 45° image of the retina (default); c. Select MULTI FIELD to acquire multiple 45° fields. See par. 15.1 for details on how to configure the fields acquired in each of the above two options; d. Select ANTERIOR EYE to acquire the external part of the eye; e. Select STEREO to acquire a stereo pair of the nasal field and produce a stereoscopic view of the optic disc (see details below); f. Position the patient on the chin rest and head rest and use the CHINREST UP and CHINREST DOWN buttons to adjust the chin rest position, if needed. When the AUTOMATIC CHIN REST flag is selected from the settings menu (see par. 15), the chin rest is automatically adapted to the patient head characteristics. Press the START button to start the acquisition procedure. The START button is only enabled if the chin rest is engaged. If the patient is not properly positioned on the chin rest, the START button will not be enabled and it will not be possible to start the acquisition process. CenterVue DRS Operating Manual Pag. 13 of 59 CenterVue DRS Operating Manual Pag. 14 of 59 9.4 Automated acquisition Once the START button is pressed, the DRS will automatically perform the following steps: a. Move the optical head to locate the patient’s pupil (see Figure 10); b. Move the optical head to center the patient’s pupil to the front lens (see Figure 11); the system will also display the pupil current diameter and which eye is being acquired; the red circle in Figure 11 represents the pupil and becomes green when proper centering is achieved. c. Perform auto-focus (see Figure 12); d. Capture an image, by flashing the retina. This step involves also automatic tuning of the flash level; e. Store the image in the local hard drive (see Figure 13); f. In case of acquisition of both eyes and/or multiple fields, the acquired images can be displayed before proceeding to the next one, depending on the settings (see par. 15). At any time during the above sequence it is possible to stop the acquisition process by pressing the CANCEL button. In case the auto-alignment process fails, manual alignment is possible: see par. 9.5 below for details. The DRS stores any images and data on the local hard disk. If the hard disk capacity is reached, no more images can be saved and a warning message is displayed saying “Hard disk full”. Figure 10 - Eye search in progress Figure 11 – Auto-alignment in progress Figure 12 - Auto-focus in progress Figure 13 - Image processing and saving in progress CenterVue DRS Operating Manual Pag. 15 of 59 9.5 Manual alignment If, for some reasons, the automated eye search fails, a message as in Figure 14 shows up. Reasons for failure include: improper patient positioning, patient not fixating steadily, patient moving the head and / or the gaze during eye search. If any of the above situations applies, try to fix the cause (re-position the patient, instruct her/him to steadily fixate the target and/or to not move) before proceeding with manual alignment. Use the UP / DOWN / LEFT / RIGHT buttons to adjust the optical head position and align the patient’s eye to the instrument front lens. As soon as the eye is detected the automated procedure will resume. Figure 14 – Message indicating failure of automatic eye search The result of the acquisition process is one or more retinal images as the one shown in Figure 15. Figure 15 – Example of retinal image captured by the DRS CenterVue DRS Operating Manual Pag. 16 of 59 9.6 Anterior Eye In order to capture a focused image of the anterior part of the eye, the patient’s forehead must be placed on the chinrest at a distance of 3-5 cm (1-2 inches) from the instrument forehead rest. The patient’s chin should be placed near the edge of the chin rest (see Figure 16). The room must be sufficiently lit in order to take an anterior eye image. Once the process is started, the optical head moves in front of the selected eye and performs a rough alignment. When it reaches the optimal position, a live color view of the eye is displayed (Figure 17) so that manual alignment and focus can be performed using the on-screen buttons. Manual adjustment of alignment and focus may be necessary to capture a good quality picture. Click on the live image to toggle a full-screen view (Figure 18). Once satisfied with alignment and focus, press SHOOT to acquire the image. Figure 16 - Anterior eye: correct patient position Figure 17 - Anterior eye acquisition CenterVue Figure 18 - Full screen view of the anterior eye DRS Operating Manual Pag. 17 of 59 9.7 Stereo exams If Stereo is selected before starting the exam, two slightly offset images of the central-nasal field will be captured with automatic alignment and focus. A delay between the shots is applied in order to let the pupil recover. The delay amount can be set in the Settings page (see par. 15). At the end of the acquisition process the resulting two images will appear in the Patient Record screen. See par. 10 for details on how to review stereo pairs. 9.8 Warning messages If the lens cap is placed on the front lens when an exam is started, the message shown in Figure 19 will appear (unless “check lens cap presence” is off, see 15.3.2). Press OK to exit and restart the exam after removing the lens cap. Figure 19 - Message when lens cap is detected If the pupil – measured immediately before the snapshot - is smaller than 3.8 mm, a warning message is shown to inform the operator about the possible degradation of image quality (see Figure 20). Figure 20 - Warning due to small pupil When the auto-focus process is not able to find an optimal focus, a warning message is shown to inform the operator about the possible occurrence of poor focusing conditions (see Figure 21). Figure 21 - Warning due to incorrect focusing CenterVue DRS Operating Manual Pag. 18 of 59 10. REVIEWING THE IMAGES Once the image(s) has (have) been acquired, the system will open the patient’s individual record and present a thumbnail view of all the available images (see Figure 22). 10.1 Patient record screen For each image thumbnail, the following information is displayed: • the eye (OD/OS) and field information, in the top-left corner; • the image number, in the top-right corner; • the acquisition date and time, in the bottom-right corner. Figure 22 - Patient record screen Available functions include: Function Icon Command Point on any thumbnail and drag to the left / right or press the left / right Image scroll buttons located at the bottom on the sides Edit patient Click on this icon to enter the edit mode and correct any mistyped record information or provide additional details Delete patient Click on this icon to permanently delete the current patient and all record related exams. A confirmation is required. Click on this icon to export all images of the selected patient as JPG Batch image files on a USB key (only available if a USB key is plugged in any of the export USB ports) Click on this icon to select one or more images to be retaken. When Retake all images have been selected, click on RETAKE. Images can be images retaken within 5 minutes after they have been captured. Open full Click on the corresponding thumbnail to open the full image screen image screen Open 3D Click on this button to open the 3D viewer. Only available for stereo image view pairs. CenterVue DRS Operating Manual Pag. 19 of 59 Function Icon Command Click on the printer icon to print two images in a single sheet. Icon is enabled when a supported printer is connected. Print Click on this icon to generate and export a dual image printout (available if a USB key is connected, see par. 12.2 for further details). Click on this icon to start generation of a mosaic (a composition) of multiple retinal fields. This icon is only available if the related license is installed. See Appendix B for explanations on how to use this function. Please contact your CenterVue representative for information on how to obtain the license. Export to pdf Automated mosaic 10.2 Full image screen The Full Image screen (see Figure 23) is used to review individual images at full screen. The information available in this screen includes the eye (OD, OS), pupil size at capture, date and time when the image was captured and any image related notes. Figure 23 - Full image screen Functions available in this screen include: Function Icon Command Zoom image - The slider on the right side of the screen allows zooming Pan image - Click on the image and drag it around to frame the desired area Red-free filter CenterVue Click on this icon to display a black and white image representing the green channel of the color image DRS Operating Manual Pag. 20 of 59 Export image to shared folder Click on this icon to export the selected image to a shared folder (see Settings >> Network >> Shared folder) Export image file to USB Click on this icon to export the selected image as a jpg file on any USB key (only available if a USB key is connected) Image print Click on this icon to print the current image (only available if a compatible USB printer is connected) Image delete Click on this icon to permanently delete the selected image Move to next / previous image Click on these icons to move to the next/previous image of the current patient 3D viewer Click on this icon to toggle between normal and 3D viewer for the stereo images. Only available when showing a stereo pair. Swap stereo images Click on this icon to swap right and left image of a stereo pair (“crossed-eye” format). Only available when showing a stereo pair. Add image notes Click on Add Notes to add / edit image specific comments (see Figure 25) Figure 24 – Notes editing screen CenterVue DRS Operating Manual Pag. 21 of 59 Figure 25 - Stereo Images 3D viewer To obtain a three-dimensional view, images must be looked at with appropriate prismatic goggles (see Figure 26). A crossed-eye format is also supported. Figure 26 – Example of prismatic goggles for three-dimensional viewing CenterVue DRS Operating Manual Pag. 22 of 59 11. REMOTE VIEWER There are two ways to review DRS data and images from a remote PC, a tablet or a smart-phone. The first is by connecting to an existing network (LAN) and the second is by creating a private network, even in absence of an existing network infrastructure. The DRS supports both wired and Wi-Fi connections in both the regular and private network modes2. The DRS remote viewer uses standard Internet browsers and does not require installation of any software application on the remote PC. Supported browsers include: Microsoft Internet Explorer (version 8 or later), Mozilla Firefox (version 3 or later), Apple Safari, Google Chrome. Remote visualization through a LAN To configure the DRS for LAN access, see par. 15.3. In order to use the remote viewer with an existing LAN proceed as follows: 1. Connect the DRS to the LAN; 2. Connect the remote PC to the LAN; 3. In case the local area network does not provide automatic IP assignments, configure the DRS as explained at par. 15.3 and configure the remote PC with a different IP; 4. Retrieve the DRS IP, as explained in Settings >> Network (see par. 15.3, Figure 48); 5. Type the DRS IP in the URL box of the remote PC’s Internet browser. Remote visualization through a private network To configure the DRS for a private network access, see par. 15.3. To use the viewer in wireless mode, proceed as follows: 1. Enable the wireless connection of your PC, tablet or smart-phone (some restrictions apply: not all devices support private network); 2. Select the wireless network created by the DRS that will be called “drs-s/n”; 3. Open the browser; 4. Type http://drs in the URL box of the remote PC’s Internet browser. Once the connection is established, you will be required to type a password, which must be configured in the DRS settings when enabling Remote Access (see 15.3). The DRS patient list will then show up on the remote PC, tablet or smart-phone (please consider that not all devices support the private network). Available functions include: • Patient list (same as Figure 9); • Patient record screen (same as Figure 22) • Access to individual images • Possibility to download this Operating Manual in pdf format in About page Figure 27 - DRS patient list in remote viewer 2 some Wi-Fi network cards may fail to establish connection with the DRS when using the private network mode. In such case an external router is required to successfully connect via Wi-Fi CenterVue DRS Operating Manual Pag. 23 of 59 12. PRINTOUT 12.1 Single image printout When a compatible USB printer is connected, any acquired image can be printed by clicking on the printer icon in the Full Image screen. Also, red-free images (see par. 10) can be printed using the same procedure. In the same way a pdf file containing the single printout can be generated if a USB key is connected. 12.2 Dual image printout By clicking on the printer icon in the Patient Record screen it is possible to print two images on the same sheet. Once the button is pressed a dialog box appears prompting for the selection of two images. The list can be scrolled to select the images to be printed by clicking on the printer icon that is placed on the top right corner of each image. Once two images are selected, click the Print button. In the same way, a pdf file containing the dual printout can be generated if a USB key is connected. The single image printout uses portrait format, while the dual image printout uses landscape format. Figure 28 – image selection screen for dual image printout CenterVue DRS Operating Manual Pag. 24 of 59 Figure 29 - Single and dual image printout CenterVue DRS Operating Manual Pag. 25 of 59 13. RETINAL FIELDS The following fields can be acquired (see Figure 30 for fields representation for the right eye): a. b. c. d. e. f. g. CENTRAL: centered on the foveal pit; NASAL: centered 17° nasally to the foveal pit; TEMPORAL: centered 17° temporally to the foveal pit; SUPERIOR-TEMPORAL: centered 12° superiorly and 12° temporally to the foveal pit; CENTRAL-NASAL: centered 5° nasally to the foveal pit; SUPERIOR: centered 17° superiorly to the foveal pit; INFERIOR: centered 17° inferiorly to the foveal pit; f d c a b e optic disk g Figure 30 - DRS retinal fields The field(s) acquired respectively when using the SINGLE FIELD or MULTI FIELD acquisition modes described previously can be configured in the Settings (see par. 15). CenterVue DRS Operating Manual Pag. 26 of 59 14. BACKUP The Backup utility allows to perform the backup of DRS data and can be found in the Settings, under the “System” tab (see Figure 31). The backup utility is based on the concept of incremental backup. The following paragraphs explain: i) how to perform the first-time backup and ii) how to perform any subsequent incremental backups. The backup utility is structured as a wizard and will guide you through the process. • IT IS A RESPONSIBILITY OF THE OPERATOR TO KEEP AN UPDATED BACKUP COPY OF THE DATA GENERATED BY THE DRS, BY PERIODICALLY USING THE BACKUP UTILITY, IN ORDER TO BE PROTECTED AGAINST UNPREDICTABLE FAILURES RESULTING IN DATA LOSS • THE MANUFACTURER CANNOT BE HELD RESPONSIBLE FOR LOSS OF DATA DUE TO UNPREDICTABLE FAILURES • IN CASE OF A FAILURE RESULTING IN DATA LOSS, THE DATA CAN BE RESTORED FROM THE LAST PERFORMED BACKUP • MANUAL MODIFICATION OF BACKUP FILES STORED ON THE BACKUP MEDIA MAY PREVENT THE ABILITY TO RESTORE THE DATA: DO NOT EVER ALTER THOSE FILES Figure 31 - System >> BACKUP tab in Settings when no backup was ever performed CenterVue DRS Operating Manual Pag. 27 of 59 14.1 First-time, total, backup To perform the first-time backup, insert a USB drive in any of the DRS USB ports. The USB drive should preferably be a hard disk and its capacity should exceed 320 GB (although initially the data size is much smaller). The firs-time backup process involves two steps: 1. initializing the USB drive; 2. creating the backup file. • A BACKUP IS ONLY EFFECTIVE WHEN THE “BACKUP COMPLETED” MESSAGE IS DISPLAYED (SEE Figure 35) • IF THE BACKUP PROCESS IS INTERRUPTED USING THE “STOP BACKUP” COMMAND THE BACKUP IS NOT EFFECTIVE, DATA CANNOT BE RESTORED AND THE PROCESS SHALL BE REPEATED • IF DURING THE BACKUP PROCESS AN ERROR MESSAGE IS DISPLAYED (SEE Figure 36) THE RESULTING BACKUP IS NOT EFFECTIVE, DATA CANNOT BE RESTORED AND THE PROCESS SHALL BE REPEATED • A BACKUP CANNOT BE SPLIT INTO DIFFERENT USB DRIVES The duration of the backup process depends on the amount of data stored in the DRS and on the speed of the USB drive used. Typically, first time backups may take longer than subsequent incremental backups. Figure 32 – Backup process, step 1 Figure 33 – Backup process, step 2 CenterVue DRS Operating Manual Pag. 28 of 59 Figure 34 – Backup in progress message Figure 35 – Backup completed message Figure 36 – Error message during backup CenterVue DRS Operating Manual Pag. 29 of 59 14.2 Incremental backups To perform an incremental backup, click on Settings and select the BACKUP tab: information about the USB drive used for previous backups will be displayed (see Figure 37). Retrieve the same USB drive, insert it in any of the DRS USB ports and click on START BACKUP (see Figure 38): information on progress will be displayed. If you use a different USB drive than indicated above you will actually be performing again a first-time, total, backup Figure 37 – System >> BACKUP tab in Settings when a previous backup had already been performed Figure 38 – System >> BACKUP tab in Settings when the USB drive containing the previous backup is connected CenterVue DRS Operating Manual Pag. 30 of 59 14.3 Removing a backup disk If you do not use a backup disk anymore and wish to remove it from the DRS list, click on Settings, select the BACKUP tab and click on the icon corresponding to the disk you want to remove, then click on “REMOVE BACKUP” (see Figure 39). Figure 39 – Removing a backup disk from the list • • • Backup files are not altered by this operation and remain on the disk It is still possible to perform a Restore using the data on the disk It is no longer possible to perform an incremental backup on the disk 14.4 Partial backups Using the backup utility, you can also create a one-time partial backup (see Figure 38): this option allows to create a backup copy of all exams performed after a specified date. Such partial backups can be restored as any other backup. PARTIAL BACKUPS SHALL BE USED ONLY IN THOSE SITUATIONS WHERE NOT ALL PATIENTS / EXAMS ARE OF INTEREST, AS PARTIAL BACKUPS DO NOT ALLOW TO RESTORE EXAMS THAT WERE PERFORMED BEFORE THE SELECTED DATE CenterVue DRS Operating Manual Pag. 31 of 59 14.5 Restore To restore data from an existing backup, click on Settings and select the BACKUP tab: information about the USB drive used for previous backups will be displayed. Retrieve the same USB drive, insert it in any of the DRS USB ports and click on START RESTORE. The restore process involves two steps: 1. selecting the backup file from which data shall be restored (if more than one is available); 2. selecting the restore method (erase & restore or restore new data). • • • • • DO NOT DISCONNECT THE USB DRIVE WHILE THE RESTORE PROCESS IS IN PROGRESS THE “ERASE & RESTORE” METHOD WILL PERMANENTLY ERASE ALL DATA STORED ON THE DRS BEFORE RESTORING THAT PRESENT IN THE BACKUP: USE THIS METHOD WITH CARE THE “RESTORE NEW DATA” METHOD WILL PRESERVE EXISTING DATA AND ADD THE PATIENTS CONTAINED IN THE BACKUP. THIS METHOD MAY RESULT IN DUPLICATED PATIENTS THE RESTORE PROCESS CANNOT BE INTERRUPTED IF THE HARD DISK CAPACITY IS REACHED, NO DATA CAN BE RESTORED AND A WARNING MESSAGE IS DISPLAYED Figure 40 – Restore process, step 1 Figure 41 – Restore process, step 2 CenterVue DRS Operating Manual Pag. 32 of 59 Figure 42 – Restore in progress Figure 43 – Restore completed CenterVue DRS Operating Manual Pag. 33 of 59 15. SETTINGS To access the settings screen, click on the button. In the Settings screen several different tabs as described in the following pages are available. 15.1 Fields This page allows configuring the fields acquired by the SINGLE and MULTI acquisition modes: any 45° field, among the 7 available ones (see par. 13), can be associated to the SINGLE mode, while up to seven 45° fields can be associated to the MULTI mode. Figure 44 - Fields >> SINGLE FIELD tab in Settings 15.2 Exam The Exam tab has three sub-tabs, namely EXAM, SAVING and ADVANCED. 15.2.1 EXAM Use this tab (see Figure 45) to: • set a time delay in seconds (default 0) between subsequent snapshots; • set a time delay between stereo images; • set the fixation target to be permanent or blinking at 3 different speeds; • enable automatic chin rest adjustment; • select UNIFORMITY or CONTRAST image processing strategy (see note below). CenterVue DRS Operating Manual Pag. 34 of 59 Figure 45 - Exam >> EXAM tab in Settings Image Processing Strategy. It is possible to select between two different image processing strategies: Uniformity and Contrast. This setting affects the internal gains and color tunings of the Camera. The final retina image has different color balance and luminosity, but the original image details are the same. UNIFORMITY: with this setting the image appears very uniform in color, with mild shades. Micro aneurisms and macula details may appear more visible with this setting. CONTRAST: with this setting the image will show higher contrast. Details may appear more evident but image has a grainier aspect. 15.2.2 SAVING Use this tab to: • set the name of the Doctor or Clinic that will appear in the header of the printouts (this value is saved in the DICOM tag 0008, 0080 Institution name). • choose to automatically create a folder for each patient name and/or a folder for each exam ID when saving on USB key. This is not a unique ID under all circumstances: several actions (e.g. backup + restore) may alter it. • Enable export of patient and exam data in text format when saving to USB. • enable DICOM export of images from the Remote Viewer. (The DICOM file format contains both the retina image and some corollary field, for example: exam date/time, patient name, type of medical instrument, etc., see Appendix A). CenterVue DRS Operating Manual Pag. 35 of 59 Figure 46 - Exam >> SAVING tab in settings 15.2.3 ADVANCED Use this tab to: • enter the Institution Dept. Name: this value is saved in the DICOM tag 0008,1040; • enter the Study Description: this value is saved in the DICOM tag 0008,1030; • enter the Series Description: this value is saved in the DICOM tag 0008,103e; • find the option to disable the lens cap detection at the start of the exams. • set the threshold of the auto-exposure meter (Exposure Value box). The higher the value specified, the brighter the flash level. The value can be set from 0.5 to 5.0. The “default” button sets the value to the default level (2.0). Figure 47 - Exam >> ADVANCED tab in Settings 15.3 Network Each time the “Network” tab is selected, a network test is performed. Once it completes, the result is displayed on the screen next to the Current network status label. The result can be any of the following: • Internet connection fully available on address xxx.xxx.xxx.xxx: the DRS is fully connected to the Internet. • LAN working on IP xxx.xxx.xxx.xxx: a valid network is set and the local network is ready. • No working connection: the DRS is not connected to any network. CenterVue DRS Operating Manual Pag. 36 of 59 The Network tab has 4 sub-tabs: GENERAL, WIRELESS, SHARED-FOLDER and REMOTE ACCESS. Figure 48 – Network >> GENERAL tab in Settings (Top: valid for units supporting private network via cable and wireless Bottom: valid for units supporting private network via cable only) 15.3.1 GENERAL To enable a wired connection, plug a network cable into the DRS and wait a few seconds. Press “SET NETWORK” to refresh the network configuration. Most networks do not require the operator to set the IP address manually, but in case it is needed, click on the Manual button and enter the IP address (mandatory), netmask (mandatory), gateway (optional) and DNS (optional). Then select “SET NETWORK” To allow the DRS to create a private network click on Private Network button The Private Network allows the use of the DRS remote viewer without using existing networks. CenterVue DRS Operating Manual Pag. 37 of 59 It can be used both in wired and wireless configuration. To go back and use an existing network click on Set Network. If the DRS is plugged into an existing network (where another DHCP server is active), unplug the network cable before enabling the Private Network in order not to interfere with the existing DHCP server. The DRS wireless Private Network is compatible with Microsoft Windows XP, 7, Vista and 8 (up to version 8.0). Android-based devices and Windows 8.1 have known limitations that prevent the correct use with the DRS wireless Private Network. Some DRS units do not support private network via the wireless interface. Figure 49 – Network tab in Settings when a private network has been set-up 15.3.2 WIRELESS To enable the wireless network, select the WIRELESS tab and click the ENABLE button. Make sure that no cable is plugged into the Ethernet port: this could create problems in wireless connection. Press RESCAN to refresh the list of available networks, then select the network you want to connect to. Enter the WiFi password/key and click “Connect” to establish the wireless connection. Figure 50 – Network >> WIRELESS tab in Settings CenterVue DRS Operating Manual Pag. 38 of 59 15.3.3 SHARED FOLDER The DRS can export images into a shared network folder. To enable exporting to shared folder, select the SHARED FOLDER tab and press ENABLE (see Figure 51). The following options are available: • INTERNAL or REMOTE; • AUTOMATIC or MANUAL. Figure 51 – Network >> SHARED FOLDER tab in Settings, when AUTOMATIC mode using a remote folder is selected Figure 52 - Network >> SHARED FOLDER tab in Settings, when MANUAL mode is selected Internal vs Remote If the INTERNAL option is selected, images will be exported in a folder inside the DRS. In this case only jpg format is available. This folder can be reached from a regular PC by searching a network resource named \\DRS-ID\shared (e.g. \\drs-0010\shared). In case of complex networks, it may be necessary to insert the IP address of the DRS instead of its name (e.g. \\192.168.1.10\shared). In this case you might need to seek the support of your local IT staff. If the INTERNAL option is selected, exported files will be only available when the DRS is powered and networked. CenterVue DRS Operating Manual Pag. 39 of 59 If the REMOTE option is selected, images will be exported in a shared network folder. In this case exported image format can be selected (JPEG, DICOM or PDF) and the remote shared folder needs to be configured. Remote folder configuration A server name, shared folder name, username (including domain or workgroup) and password need to be specified (see Figure 53). See par. 24 for possible error conditions. Figure 53 - Network >> SHARED FOLDER tab: remote shared folder configuration Click on CONFIGURE DESTINATION to re-configure the remote shared folder. Click on CHECK DESTINATION to refresh the status of the remote shared folder. If the remote shared folder becomes (temporarily) unavailable/unreachable, newly acquired images cannot be exported and will stay in the export queue, until access is restored. Status icon for automatic export to remote shared folder: Icon Description Folder properly configured. No file in the export queue. Export is initiating. Export in progress. An error has occurred: export cannot be completed, files stay in the export queue. Export completed, no files pending in export queue. CenterVue DRS Operating Manual Pag. 40 of 59 Automatic vs Manual If the AUTOMATIC mode is selected, at the end of each exam, a copy of the image will be automatically saved to the shared network folder. Images in the shared folder can be deleted at any time by the operator or by a program without altering the database and files inside the DRS. Click on REFRESH to update the status of the export queue. Click on EMPTY QUEUE to erase all pending files in the export queue (confirmation required). If the MANUAL mode is selected, a new button will be shown in the upper right corner of the Full image screen: this button allows you to save the exam in the shared folder. Exported file name The naming convention for exported files is as follows: patientID_lastname_firstname_date_time_drsNumber_eye_fieldNumber_examNumber.<ext> 15.3.4 REMOTE ACCESS Use the REMOTE ACCESS sub-tab to: • enable the WebAPI service (see information box below); • enable remote requests over http for the Remote Viewer (see par. 11) and for the WebAPI service; • enable remote requests over https for the WebAPI service; • set a password for the Remote Viewer and WebAPI access to the DRS, to prevent access to the data by unauthorized persons. WebAPIs are Application Program Interfaces for third parties’ software solutions (e.g. EMRs) to connect to the DRS and push/retrieve data. Configuration of a password is mandatory, for security reasons, when any of the Remote Access options is enabled. Passwords must comply with the following rules: at least 8 chars AND at least one lower-case letter AND at least one capital-case letter AND at least one number. Figure 54 – Network >> REMOTE ACCESS tab in Settings CenterVue DRS Operating Manual Pag. 41 of 59 15.4 System The System tab has five sub-tabs: GENERAL, TIME, BACKUP, PRINTERS and SERVICE. 15.4.1 GENERAL Use this tab to: • Enable, disable and configure the screensaver: the delay can be between 5 and 60 min. • If the screensaver is enabled, it can be protected by a password: the operator will also be prompted for the password when the DRS starts; • Enable or disable the “DARK MODE”: a configuration that reduces the emission of light, useful when using the device in a dark room. Screensaver passwords must comply with the following rules: at least 8 chars AND at least one lower-case letter AND at least one capital-case letter AND at least one number. Figure 55 – System >> GENERAL tab in Settings 15.4.2 TIME The system date and time can be set-up: • Automatically from Internet • Manually from the pull-down menus In this submenu, you can also set: the correct time zone, the date and time format Figure 56 - System >> TIME tab in Settings CenterVue DRS Operating Manual Pag. 42 of 59 15.4.3 PRINTERS Use this tab to: • choose among different printer drivers (built-in drivers or generic postscript driver); • adjust image brightness on the printout; • optionally enable the “Automatic fit to page” for printouts; • optionally enable A5 printing format (only for printers that support it). Figure 57 – System >> PRINTERS tab in Settings CenterVue DRS Operating Manual Pag. 43 of 59 15.4.4 SERVICE This tab provides access to various functionalities and tools, in particular: • reset optical head resets the position of the optical head (same procedure performed at device startup); • lens cleaning assistant helps to check the cleaning condition of the front lens. See below for additional information. • service gives access to a special password-protected panel available for authorized personnel only. • open remote assistance opens a remote connection that allows authorized service technicians to connect to the DRS. Once pressed, the disclaimer shown in Figure 59 appears, informing the operator that he/she is authorizing access to all data present on the device, for servicing purposes. Once confirmed, the DRS contacts CenterVue servers and receives a random authorization code. This code must be communicated to the service operator in order to connect. The connection can be closed by clicking the close remote assistance button. • save log files to USB allows to export the log file of your DRS unit onto a USB key, for troubleshooting purposes. • upload log files allows to upload the log file to a Centervue server, for troubleshooting • reset settings erases all device settings to Factory defaults (selected fields, network configuration, shared folder, preferences, etc.). • reset patients erases all stored patient data and images. RESET PATIENTS ERASES THE WHOLE DATABASE CONTENT. ONCE THE OPERATION HAS BEEN CONFIRMED THERE IS NO WAY TO RESTORE THE DATA. USE THIS FUNCTION WITH CAUTION. Figure 58 - System >> SERVICE tab in Settings CenterVue DRS Operating Manual Pag. 44 of 59 Figure 59 - Remote Assistance Lens cleaning assistant Clicking the Lens Cleaning Assistant button opens the box shown in Figure 60. Clicking on Instructions opens the instructions box shown in Figure 61. Clicking on Tools opens a box describing the tools to be used for cleaning the lens. Clicking on SHOOT captures an image which should be mostly black: white particles indicate presence of residual dirt on the outer surface of the lens. Figure 60 – Lens cleaning Assistant information dialog Figure 61 – Lens cleaning instructions CenterVue DRS Operating Manual Pag. 45 of 59 15.5 About This screen shows the DRS serial number, the software and firmware version, the OS version, the Ethernet and WiFi MAC address, the total number of patients and exams stored, the space left on the hard disk, the presence of any optional licenses and the list of compatible printers. It also allows to download this Operating Manual on a USB key, in pdf format. Figure 62 - About tab in Settings CenterVue DRS Operating Manual Pag. 46 of 59 16. AUTOMATIC SOFTWARE UPDATE When a new software update is released by the manufacturer and the DRS is connected to the internet, a new icon in the Home screen top bar, notifies availability of the update. Clicking on the icon, the message shown in Figure 63 will appear. To download and install the new software version click Install, otherwise click Cancel. Figure 63 - Message to confirm the installation of the new software version 17. SYSTEM SHUTDOWN To shut down the system go to the home screen, Patient list or Patient record screen and click on the power off button. Wait for the progress bar to completely roll back. A message on the screen will inform you when it is safe to turn off the main switch. Always put back the cap on the front lens. Do not leave the front lens uncovered while the system is not in use. CenterVue DRS Operating Manual Pag. 47 of 59 18. CLEANING This paragraph explains how to clean the device. The device must be powered off, and the power cord shall be disconnected from mains. Extreme care should be given to always protect the front lens. 18.1 Front lens The front lens should be cleaned by using a small hand pump air blower (e.g. Kaiser Typhoon 6316), to blow away dust. Figure 64 – Hand air blower Only if really needed, for instance due to the presence of a fingerprint, the front lens should be cleaned using photographic cleaning paper (e.g. Edmund Optics NT53-984) and a suitable lens cleaning fluid (e.g. Edmund Optics P57-727) as illustrated in Figure 61. Do not attempt to clean the front lens with a dry cloth. This could lead to scratching its surface. Pass a wet wipe on the front lens with a single circular motion: never reuse it after each pass. Several passes may be needed in order to achieve a good cleaning level. Figure 65 – Cleaning liquid and wipes Figure 66 – Spray the wipes Figure 67 – Clean the lens Use the Lens cleaning assistant tool in order to check the cleaning status of the front lens. Please note that some small white dots will always appear due to dust (see Figure 65), but they do not affect image quality. See Lens cleaning assistant at par.14.4.5, for a detailed assisted procedure. CenterVue DRS Operating Manual Pag. 48 of 59 NO Figure 68 – Sample picture of a very dirty front lens OK Figure 69 – Sample picture of a clean front lens 18.2 Patient cushions The chin rest and the forehead rest are the only parts in direct contact with the patient: they should be wiped with an antiseptic wipe, before each use and allowed to dry prior to reuse. Take care not to sprinkle parts not belonging to the patient rest. If the device is equipped with silicon cushions, it is possible to remove the forehead rest, and rinse it with warm water and a mild hand detergent. The chin rest removable. CenterVue cushion DRS Operating Manual is not Pag. 49 of 59 18.3 Touch screen panel The touch screen panel should be cleaned only with a cloth damped in water. Do not use alcohol or detergents to clean the touch screen, as these may damage the film covering the panel. 18.4 Plastic shells If needed, the external covers of the device can be cleaned by means of a cloth slightly damped in water. Do not use alcohol or detergents to clean the plastic shells, as they might erase labels or indications. 19. MAINTENANCE The only operation the operator can do on the device is the replacement of the fuses. Following is the relevant procedure: • Disconnect the power cord from the mains • Remove the fuse holder with the help of a small screwdriver • Remove and replace the fuses (see section ”Technical specifications”) • Insert the fuse holder • Connect the power cord to the mains. All service interventions other than the above should be carried out by authorized CenterVue personnel. CenterVue recommends requiring a service to the instrument in order to: • check the electrical safety (IEC 60601-1) every year. • check the general functionality every 5 years. Upon request, CenterVue and its authorized personnel can provide specific maintenance contracts. CenterVue DRS Operating Manual Pag. 50 of 59 20. ELECTROMAGNETIC COMPATIBILITY This device is classified in class B according to IEC 60601-1-2. This device has been tested and found to comply with the limits for medical devices contained in IEC 60601-1-2 and Medical Device Directive 93/42/EEC. These limits are intended to provide reasonable protection against harmful interference in a typical medical installation. This instrument generates, uses and can radiate radio frequency energies and, if not installed and used in accordance with these instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. If the system does cause harmful interference to other devices, which can be determined by turning the system off and on, try to eliminate the interference by adopting one or more of the following measures: • reorient and/or relocate the receiving device; • increase the distance between the devices; • connect the system to an outlet on a different circuit than that to which the other devices are connected; • consult the manufacturer or field service technician for help. 21. FCC (USA) and IC (Canada) radio certification The DRS contains a radio module that complies with regulations of the USA and Canada, in particular with part 15 of the FCC rules. Changes or modifications not expressly approved by the party responsible for compliance could void user’s authority to operate the equipment. Operation is subject to the following 2 conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. CenterVue DRS Operating Manual Pag. 51 of 59 22. TECHNICAL SPECIFICATIONS Class and type of applied part I, B (according to IEC 60601-1). IP classification IPX0 (according to IEC 60529, related to the degree of protection provided by the enclosure with respect to harmful penetration of particulate matter or water). Retinal Imaging • Field of view: 45° x 40° • Non-mydriatic operation (3.8 mm minimum pupil size) • Fixation target: internal LEDs • Operating distance: 37 mm • Sensor size: 5 MPixel (2592 x 1944) • Resolution: 48 pixels/deg • Resolution on retina: 15 µm • Pixel pitch on retina: 6 µm Other features: • Patient presence sensor • Motorized chin-rest • Automatic alignment • Auto-focus (adjustment range: from -15D to +15D) • Auto-flash level adjustment • Touch-screen color display • Embedded PC • Ethernet connectivity Dimensions: • Weight: 19 Kg (42 lbs.) • Size: 55 × 55 × 33 cm (22 × 22 × 13 in) Power requirements: • Voltage: 100-240 VAC, 50-60 Hz, fuse 3.15 A (T type) • Power consumption: 100 VA Specifications are subject to change without notice for improvement, as result of ongoing technical development CenterVue DRS Operating Manual Pag. 52 of 59 23. DISPOSAL The DRS is made of different materials, such as plastics, aluminum, electronic parts. In case of instrument disposal, please separate the various materials and follow the laws and regulations regarding disposal or recycling for each material effective in your own country. Separate collection for electrical and electronic equipment The European Directive 2002/96/EC establishes the separate collection for Waste of Electrical and Electronic Equipment (WEEE). The users of Electric and Electronic Equipment (EEE) must not dispose of WEEE as unsorted municipal waste, but have to collect such WEEE separately. The available return and collection system is defined by the local public administration, or in alternative an authorized company can recycle the WEEE. Please refer to public administration about separate collection; if this information is not available, contact the manufacturer of the equipment. Users have a fundamental role in contributing to reuse, recycling and recovery of WEEE. The potentially dangerous substances contained in the WEEE can pollute the environment and produce harmful effects to the human health. Below are a few indications of specific dangers of some substances, which may leach in the environment and in the water system. Lead: damages the nervous system of humans, it affects the endocrine system, the cardiovascular system and kidneys. It accumulates and is very toxic for animals, plants and micro-organisms. Cadmium: accumulates with a half-life of 30 years and can damage the kidneys and cause cancer. Mercury: is easily accumulated in organisms and concentrates through the food chain. It has chronic effects and can cause brain damage. Chromium (Hexavalent): easily absorbed into cells with toxic effects. The results can be allergic reactions, asthma and it is considered to be genotoxic (damages the DNA). Especially dangerous when incinerated. Brominated Flame Retardants: widely used to reduce flammability (e.g. cables, connectors and plastic cases). CenterVue DRS Operating Manual Pag. 53 of 59 24. TROUBLESHOOTING AND ERROR MESSAGES If any problem occurs, check the following table and try to apply the proposed solution. If the problem persists, please contact CenterVue Service representative. Issue / Error message Possible cause The unit does not turn on Main fuses are blown up After a period of functioning, the DRS computer becomes very slow Overheating of the computer After the system booting, the software application sends a connection error saying: “Robot initialization error” or “Unable to connect to control board”. The retinal image is blurred/defocused During the exam, the software application fails the autoalignment. The control board is not powered with 12V, or there is no USB connection between the computer and the control board, or the BIOS of the embedded PC is set not to power control board logic The motor of the focus LED doesn’t move. The motor of the camera lens doesn’t move. One (or both) of the pupil LEDs is damaged or not powered Replace main fuses according to the § “Maintenance” Remove the objects under the DRS that are blocking the cooling air flow Contact an Authorized Service Center Try repeating the exam, possibly on a different eye. If still defocused, contact an Authorized Service Center Contact an Authorized Service Center The mechanical connection of the USK key is not appropriate Use the USB extension cord that comes with the DRS (inside the components box) to connect the USB key. The database got corrupted because of a hard shutdown Click on “proceed with data recovery” The unit underwent a hard shutdown None needed, the units will check data integrity and restart The USB key does not work CenterVue Solution DRS Operating Manual Pag. 54 of 59 Issue / Error message Mosaic does not work with error message “Unable to create mosaic. You need to select at least two images” Mosaic does not work with error message “Unable to create mosaic. You need to include a central field image” Mosaic fails with error message “Mosaic creation failed” “Warning: remote shared folder is not reachable with the current configuration. Do you wish to save it anyway?” Possible cause Solution No valid USB drive is connected Use a different USB drive At least two images must be selected for mosaic generation A central field is missing in the selected images Images are not suitable for stitching The shared folder is currently not reachable or it is incorrectly configured Charge level of the battery of the internal PC is low CenterVue DRS Operating Manual Select at least two images Add a central field to the selection Select (or capture) different images Check that the configured folder is reachable and shared with write access. Check the spelling of the configured parameters. Contact Authorized Service Center Pag. 55 of 59 APPENDIX A - DICOM Statement DRS can export a DICOM-compliant file format. Refer to the table below for the list of supported tags. Tag (0002,0000) (0002,0001) (0002,0002) (0002,0003) (0002,0010) (0002,0012) (0002,0013) (0002,0016) (0008,0016) Description Group Length File Meta Information Version Media Storage SOP Class UID Media Storage SOP Instance UID Transfer Syntax UID Implementation Class UID Implementation Version Name Source Application Entity Title SOP Class UID VR UL OB UI UI UI UI SH AE UI Len 4 2 32 22 20 26 8 10 32 (0008,0018) SOP Instance UID UI 22 (0008,0020) (0008,0021) (0008,0022) (0008,0023) (0008,002A) (0008,0030) (0008,0031) (0008,0032) (0008,0033) (0008,0050) (0008,0060) (0008,0064) (0008,0070) (0008,0090) (0008,0100) (0008,0102) (0008,0104) Study Date Series Date Acquisition Date Image Date Acquisition Datetime Study Time Series Time Acquisition Time Image Time Accession Number Modality Conversion Type Manufacturer Referring Physician's Name Code Value Coding Scheme Designator Code Meaning DA DA DA DA DT TM TM TM TM SH CS CS LO PN SH SH LO 8 8 8 8 14 14 14 14 14 0 2 2 16 0 8 4 14 (0008,1030) Study Description LO 2 (0008,103E) Series Description LO 2 (0010,0010) Patient's Name PN 10 (0010,0020) Patient ID LO 2 (0010,0030) Patient's Birth Date DA 8 (0010,0040) Patient's Sex CS 2 (0010,1000) Other Patient IDs LO 2 (0018,7004) (0020,000D) (0020,000E) Detector Type Study Instance UID Series Instance UID CS UI UI 4 22 22 Value 180 0 1.2.840.10008.5.1.4.1.1.77.1.5.2 2.16.840.1.114496.6.4 1.2.840.10008.1.2.1 1.2.826.0.1.3680043.8.641 CVUEDRS RUBY_DICOM 1.2.840.10008.5.1.4.1.1.77.1.5.2 CenterVue ANSI registration code (unique id number) plus machine id and actual exam id. Example: 2.16.840.1.114496.6.4 Exam date Exam date Exam date Exam date Exam datetime Exam time Exam time Exam time Exam time This tag is always empty OP DI CenterVue S.p.A. This tag is always empty R-1021A SRT Fundus Camera Value entered in Settings | Exam | Advanced | Study Description Value entered in Settings | Exam | Advanced | Series Description Value entered in the patient’s Last Name and First Name fields (last + first) Value entered in the patient’s “code” field Value entered in the patient’s date of birth fields Value entered in the patient’s gender field Value entered in the patient’s “code” field CMOS 2.16.840.1.114496.6.4 2.16.840.1.114496.6.4 DRS Operating Manual Pag. 56 of 59 CenterVue Tag (0020,0010) (0020,0011) (0020,0013) (0020,0020) (0020,0062) (0022,000C) (0028,0002) (0028,0004) (0028,0006) (0028,0010) (0028,0011) (0028,0030) (0028,0100) (0028,0101) (0028,0102) (0028,0103) CenterVue Description Study ID Series Number Instance Number Patient Orientation Image Laterality Horizontal Field of View Samples per Pixel Photometric Interpretation Planar Configuration Rows Columns Pixel Spacing Bits Allocated Bits Stored High Bit Pixel Representation VR SH IS IS CS CS FL US CS US US US DS US US US US Len 0 0 0 2 2 4 2 4 2 2 2 6 2 2 2 2 DRS Operating Manual Value This tag is always empty This tag is always empty This tag is always empty LF R or L (left or right eye) 45.000000 3 RGB 0 1944 2592 0.0064 8 8 7 0 Pag. 57 of 59 APPENDIX B – AUTOMATED MOSAIC FUNCTION A mosaic is the composition of multiple, partially overlapping, fields of the same retina and it is used to obtain a wider view. Mosaic images are permanently stored on the local memory and can be reviewed at any time as individual fields. Typical examples of fields used are: central, temporal and nasal or central, temporal, nasal, superior and inferior. This function is not active by default on the DRS system and may require installation of a dedicated license. Please contact your CenterVue representative for further information on how to activate the mosaic function. To create a mosaic, check that the fields of interest have been captured, then click on the mosaic button in the Patient Record screen to open the image selection screen: the first step requires selection of the eye, the second one requires selection of the fields to be composed. Presence of a central field is always required for mosaic generation. 2 to 6 fields can be used. Only images of the same eye captured on the same day can be used. Only one eye at a time can be selected. Once field selection is completed click on the Create Mosaic button to start mosaic generation or click on Cancel to abort (see Figure 71). Generation of a mosaic image may take from 10 to 90 seconds3, depending on how many fields are used. The DRS cannot be used while mosaic generation is in progress. Figure 70 – Mosaic function, eye selection screen 3 Benchmark refers to devices manufactured starting in August 2015 CenterVue DRS Operating Manual Pag. 58 of 59 Figure 71 – Mosaic function, field selection screen During mosaic generation a dialog box will provide feedback on progress. Click on Cancel to abort a mosaic at any time. Once completed, click on Ok to go back to the Patient Record screen and review the mosaic image (see Figure 72). Images resulting from the mosaic process may contain artifacts (such as duplicated or disconnected vessels) that are generated at the transition between two adjacent fields and that are not present in the original images. Such artifacts can be easily ruled out by comparing the mosaic image with the original single-field images. Figure 72 – Example of a 3-fields mosaic image CenterVue DRS Operating Manual Pag. 59 of 59 ">
Advertisement
Key Features
- Non-Mydriatic
- Automatic Image Acquisition
- Multiple Retinal Fields
- Stereo Imaging
- User-Friendly Interface
- Image Review and Export
- Data Backup
- Software Updates
Frequently asked questions
What is the field of view of the Icare DRS?
The Icare DRS has a field of view of 45° x 40°.
Does the Icare DRS require pupil dilation?
No, the Icare DRS is a non-mydriatic device, so pupil dilation is not required.
How many retinal fields can be captured with the Icare DRS?
The Icare DRS can capture seven different retinal fields.