Siemens | LB 54... | \\Usggswfile02\pll group files\

StreamLAB® Analytical Workcell
System Specifications
Effective: December 2008
Shown with four (4) Dimension® RxL Max® instruments
For further details, see D-01431 Dimension ® RxL Max® STM/RMS Instrument
Specifications. For configurations, contact your Siemens Healthcare Diagnostics Inc.
sales representative.
System Weight and Dimensions
Weight
Input/Output Module
720 lb (327 kg)
Main Track
340 lb (154 kg)
Track Extension
120 lb (54 kg)
Track Spacer
10 lb (5 kg)
Track Turn
227 lb (103 kg)
Tube Sealer Module
200 lb (91 kg)
Optional Centrifuge with Robot
430 lb (195 kg)
Optional Centrifuge UPS
79 lb (36 kg)
Optional Jun-Air Compressor
256 lb (116 kg)
Dimensions
Input/Output Module
70 in. wide x 60 in. high x 35 in. deep
(178 cm w x 152 cm h x 89 cm d)
Main Track
122 in. wide x 40 in. high x 17 in.deep
(310 cm w x 102 cm h x 43 cm d)
Track Extension
67 in. wide x 40 in. high x 17 in. deep
(170 cm w x 102 cm h x 43 cm d)
Track Spacer
18 in. wide x 40 in. high x 17 in. deep
(46 cm w x 102 cm h x 43 cm d)
Track Turn
30.5 in. wide x 40 in. high x 17 in. deep
(77.5 cm w x 102 cm h x 43 cm d)
Tube Sealer Module
33.5 in. wide x 40 in. high x 17 in. deep
(85 cm w x 102 cm h x 43 cm d)
Optional Centrifuge with Robot
23 in. wide x 31 in. high x 29 in. deep
(59 cm w x 79 cm h x 74 cm d)
Optional Centrifuge UPS
17 in. wide x 3.5 in. high x 26 in. deep
(44 cm w x 9 cm h x 66 cm d)
Optional Jun-Air Compressor
26 in. wide x 34 in. high x 26 in. deep
(66 cm w x 86 cm h x 66 cm d)
Electrical Installation Requirements
Current/Operating Power Requirements
Nominal Line
Voltage
vac
Line Voltage
vac
Nominal Line
Frequency
Hz
Maximum Continuous
Current,
AMPS
Power
Consumption
WATTS
Dimension® RxL Max® system w/HM/RMS/STM
115
230
103 to 127
207 to 253
47 to 63
47 to 63
~17
~11
1955
2530
Input/Output Module
Centrifuge1
Jun-Air Compressor2
230
230
230
115
230
230
230
115
47 to 63
47 to 63
47 to 63
47 to 63
~5
~9
~8
~17
1150
2070
1840
1955
1 Refer to Hettich Centrifuge Operator's Guide for additional specifications.
2 Refer to Jun-Air Operator's Guide for additional specifications.
StreamLAB® Analytical Workcell
Clearances (Minimum)
• Overhead: 25 in. (64 cm)
• Doorway opening for installation—
35 in. (89 cm)
Wire Size (North America only)
Average Noise Output
Dimension® RxL Max ® system:
10 AWG minimum
Input/Output Module: <70 dBA at 1 m
while operating
Shield
Not required.
Optional Jun-air Compressor:
<53 dBA at 1 m while operating
Receptacle
Optional Hettich Centrifuge: <62 dBA
at 1m while operating
• Input/Output Module
North America: 220 vac, 60Hz, Single
Phase, 15A
EU : 230 vac, 50Hz, Single Phase, 16A
Input/Output Module: The receptacle
must be accessible to the 9-ft (2.7m)
power cords furnished with the
module.
Compressed Air
• Optional Centrifuge
North America: 220 vac, 60Hz, Single
Phase, 15A
EU : 230 vac, 50Hz, Single Phase, 16A
Leakage Current
A 360° access is needed for service
with a minimum 30-inch (76 cm)
clearance around the workcell.
Recommended Service Outlet
• Optional Jun-Air Compressor
North America: 220 vac, 60Hz, Single
Phase, 15A
EU : 230 vac, 50Hz, Single Phase, 8A
Transient Overvoltage
Installation Category II (branch circuit)
Circuit and Ground
The Input/Output Module should have
a dedicated line with hot, neutral, and
isolated ground in its own conduit.
Connect the conduit from the
distribution panel to the receptacle.
Three-wire distribution to the
receptacle is required for each
instrument. Connect the third (green)
ground wire from the distribution panel
to the receptacle in accordance with
NEC paragraph 250-74, exception 4,
unless local codes prohibit. The
ground wire should not be tied to
grounds from other loads.
Leakage current is less than or equal
to 500 mA. This is measured
according to procedure EN61010-1.
The StreamLAB® Analytical Workcell
requires an external source of
compressed air supplied at 65 NL/min
@ 7 bar (2.3 scfm @ 100 psi.).
Connect the air supply system to the
air inlet fitting of the I/O Module. A
shutoff valve and pressure gauge
should be accessible near the
workcell.
Room Environment
Operating Temperature
Room temperature must be 65–85°F
(17–30°C) with a maximum fluctuation
of 5°F (2.8°C) per hour.
Relative Humidity
Maintain between 20% and 80%
Compressed air requirement with tube
sealer is 105 NL/min @ 7 bar
(3.7 scfm @ 100 psi.)
External Uninterruptible Power Source
(UPS)
3923 BTU/hr Input/Output Module
The Input/Output Module UPS and
Centrifuge Module UPS must be
installed in the same room as their
modules.
6670 BTU/hr Dimension® RxL Max ®
system with HM/RMS/STM
Host Interfacing
Average Thermal Output
7063 BTU/hr Optional Centrifuge
6278 BTU/hr Optional Jun-Air
Compressor
A 25-pin female connector is required
for hookup to the male connector
used for host communications port.
Installation
FCC Compliance
The StreamLAB® Analytical Workcell
will be installed by a qualified
representative of Siemens. The
installation will include checkout of all
aspects necessary to ensure the
equipment is fully operational.
This device complies with Part 15 of
the FCC Rules. Operation is subject
to the following conditions:
Preventive Maintenance Frequency
Four Siemens service preventive
maintenance visits per year for the
StreamLAB® Analytical Workcell.
Multiple pieces of equipment can be
maintained during a single visit.
1. This device may not cause harmful
interference.
2. This device must accept any
interference received, including
interference that may cause undesired operation.
Barcode Scanner
The StreamLAB® Analytical Workcell
barcode reader is a Class II laser and
is a possible eye hazard.
Code Compliance
Validated Connectivity Equipment
Emission Compliance
The StreamLAB® Analytical Workcell
should not be used next to any
Industrial Scientific and Medical (ISM)
equipment that must functionally
produce RF energy (e.g., diathermy
equipment).
• Dimension® RxL Max ® w/HM/RMS/
STM system
• Dimension Vista® System
• Sysmex ® CA-7000 System (May
2006)
• Immulite® 2000 w/SMS
• Immulite® 2500 w/SMS
• ADVIA Centaur® System (outside
U.S. only; September 2006)
• Abbott Architect® Analyzer (outside
U.S. only; June 2007)
Validated Tube Types
Manufacturer
Becton-Dickinson
Tube Type
Hemogard Closures
Size of Tube
13 x 75 mm
13 x 100 mm
16 x 100 mm
Tube Material
Glass or Plastic
Glass or Plastic
Glass or Plastic
Beckton-Dickinson
Sarstedt
Rubber Stopper
S-Monovette®
16 x 100 mm
13 x 90 mm
15.3 x 75 mm
15.3 x 92 mm
13 x 65 mm (capped)*
Glass or Plastic
Greiner
Vacuette®
Terumo
Venosafe®
13 x 75 mm
13 x 100 mm
13 x 75 mm
13 x 100 mm
16 x 100 mm
13 x 75 mm
13 x 100 mm
Any Manufacturer
Uncapped round bottom tube for pour-offs
* For use with Sysmex® CA-7000 Analyzer only.
For additional information or to reach a Siemens
representative, please call 1-800-393-9362.