YSI ProDSS Multiparameter Sampling Instrument
Portable digital sampling system for the measurement of pH, ORP,
dissolved oxygen (optical-based sensor), conductivity, turbidity, temperature, depth and more.
Designed for use in applications such as surface water, groundwater,
coastal waters, and aquaculture, the rugged and reliable ProDSS allows
for measurement of up to 17 parameters. The ProDSS features userreplaceable digital smart sensors that are automatically recognized by
the instrument when connected. A backlit color display, large memory,
convenient calibration procedures, rechargeable battery, and powerful PC
data management program (KorDSS) make the ProDSS user friendly. The
optional GPS function, wide range of sensors, and varying cable lengths
allow for complete customization of the ProDSS. Mil-spec (military spec)
connectors and a waterproof (IP-67), rubber over-molded case ensures
durability to provide years of sampling even in the harshest field conditions.
• Single cable design features universal ports which can accept any 4
sensors plus depth
• Long-life rechargeable lithium-ion battery to power handheld and sensors
• Color display and backlit keypad allow for sampling in all lighting
conditions
• Digital smart sensors are automatically recognized by the instrument and
store calibration data
• User-replaceable cables and sensors provide versatility, reduce down time
and reduce overall cost of ownership
Parameters:
Dissolved Oxygen (Optical)
Turbidity
pH
ORP/Redox
Conductivity
Specific Conductance
Salinity
Total Dissolved Solids (TDS)
Resistivity
Seawater Density
Total Suspended Solids (TSS)
Depth
Ammonium
Ammonia
Chloride
Nitrate
Temperature
• Large memory (> 100,000 data sets) with extensive site list and Data ID tag
capabilities
• KorDSS, a powerful data management software with geo-mapping
capability, is included with the instrument
• USB On-The-Go connector for PC connection, recharging/powering the
ProDSS and connecting directly to a USB stick
• Global Positioning System (GPS) (optional)
• Cable lengths up to 100 meters available with or without a depth sensor
• Rugged, waterproof case (IP-67 rated) with rubber over-mold and metal,
military-spec (MS) cable connectors as well as rugged titanium sensors
• Multiple languages in handheld and KorDSS: English, Spanish, Portuguese,
French, German, Italian, Japanese, Norwegian, and Chinese (simplified and
traditional)
• Warranty: 3-year instrument; 2-year cable assembly and sensors; 1-year pH
and pH/ORP sensor modules; 6-months ammonium, nitrate and chloride
sensor modules.
YSI.com/ProDSS
ProDSS General Specifications
Size
Instrument:
Bulkhead with sensors, without depth (no guard):
Bulkhead with sensors, with depth (no guard):
Bulkhead with guard, without depth:
Bulkhead with guard, with depth:
8.3 cm width x 21.6 cm length x 5.6 cm depth (3.27 in x 8.5 in x 2.21 in)
33.3 cm (13.11 in) length - the length with the ISE sensors is 34.04 cm (13.4 in)
35.84 cm (14.11 in) length - the length with the ISE sensors is 36.58 cm (14.4 in)
42.82 cm (16.86 in) length and 4.75 cm (1.87 in) outer diameter
45.36 cm (17.86 in) length and 4.75 cm (1.87 in) outer diameter
Weight with batteries
567 grams (1.25 lbs)
Power
Rechargeable lithium-ion battery pack provides ~48 hours with the handheld only and ~20 hours with the handheld, cable and four
sensors; battery recharge time is ~9 hours with the AC power adapter
The instrument can also be powered via AC or external power pack through the USB port
Instrument operating temperature
0 to 50 °C (32 to 122 °F)
Instrument storage temperature
0 to 45 °C (32 to 113 °F) with battery installed; 0 to 60 °C (32 to 140 °F) without battery installed
Display
Color, LCD graphic display; 3.9 cm width x 6.5 cm height
USB port
Built-in micro USB On-The-Go port for PC connection, recharging/powering the ProDSS and connecting directly to a USB stick
Cables
Available with or without depth sensor in 1, 4, 10, 20, 30, 40, 50, 60, 70, 80, 90, and 100-meters
Sensor ports
4 universal sensor ports on each cable; can accept any ProDSS sensor
Warranty
3-year instrument; 2-year bulkhead, cable assembly, and sensors; 1-year pH and pH/ORP sensor modules, ODO sensor caps, and Li-ion
battery pack; 6-months ammonium, nitrate, and chloride sensor modules
Memory
> 100000 data sets
Logging modes
Single point or continuous with autostable feature
GLP compliance
Yes; 400 detailed GLP records can be stored and are available to view, download, and print.
Languages
English, Spanish, Portuguese, French, German, Italian, Japanese, Norwegian, Simplified Chinese, Traditional Chinese
Certifications
CEC, CE; RoHS; IP-67; WEEE; FCC; UN Part III, Section 38.3, Test methods for lithium-ion batteries (Class 9)
GPS
Optional internal GPS; coordinates are stored with measurement data and site lists
Sites and data ID
100 user-defined sites and 100 user-defined data ID tags
ProDSS System Specifications (Instrument, Sensor, and Cable)
Sensor/Parameter
Range
Accuracy
Resolution
Units
Temperature
-5 to 70 °C (temperature compensation range
for DO mg/L measurement: -5 to 50 °C )
±0.2 °C
0.1 °C or 0.1 °F (user
selectable)
°C, °F, K
pH
0 to 14 pH units
±0.2 pH units
0.01 pH units
pH, pH mV
ORP
-1999 to 1999 mV
±20 mV
0.1 mV
mV
Dissolved
Oxygen
0 to 500%, 0 to 50 mg/L
0 to 200%: ±1% of reading or 1% saturation, whichever is
greater
200 to 500%: ±8% of reading
0 to 20 mg/L: ±0.1 mg/L or 1% of reading, whichever is greater
20 to 50 mg/L: ±8% of reading
0.01 mg/L and 0.1%, or
0.1 mg/L and 1% (user
selectable)
% saturation, % saturation
local, mg/L, ppm
Barometer
375 to 825 mmHg
±1.5 mmHg from 0 to 50 °C
0.1 mmHg
mmHg, inHg, mbar, psi,
kPa, atm
Conductivity
0 to 200 mS/cm
0 - 100 mS/cm: ±0.5% of reading or .001 mS/cm, whichever
is greater
100 - 200 mS/cm: ±1.0% of reading
0.001, 0.01 or 0.1 µS/
cm (range dependent)
μS/cm, mS/cm
Specific Conductance*
0 to 200 mS/cm
0 - 100 mS/cm: ±0.5% of reading or .001 mS/cm, whichever
is greater
100 - 200 mS/cm: ±1.0% of reading.
0.001, 0.01, 0.1 mS/cm
μS/cm or mS/cm
User selectable reference temperature (15 to 25 °C; default 25
°C) and compensation coefficient (0 to 4%/°C; default 1.91%)
Salinity*
0 to 70 ppt
±1.0% of reading or ±0.1 ppt, whichever is greater
0.01 ppt
ppt or PSU
Total Dissolved Solids
(TDS)*
0 to 100 g/L
Calculated from specific conductance and a user-selectable TDS
multiplier (0.30 to 1.00; default 0.65)
0.001, 0.01, 0.1 g/L
mg/L, g/L, kg/L
Resistivity*
0 to 2 Mohms
±0.1% Full Scale
0.001, 0.01, 0.1 ohms
ohm-cm, kohm-cm,
Mohm-cm
Seawater Density*
0.0 to 50.0 sigma, sigma T
-
0.1 sigma or sigma T
Sigma, Sigma T
Turbidity
0 to 4000 FNU
0 to 999 FNU: 0.3 FNU or ±2% of reading, whichever is greater
1000 to 4000 FNU: ±5% of reading
0.1 FNU
FNU, NTU
Total Suspended Solids
(TSS)*
-
User correlated from turbidity field measurements and lab TSS
measurements from grab samples
0.01, 0.1 mg/L
mg/L
Ammonium**
0 to 200 mg/L NH4-N
±10% of reading or 2 mg/L, whichever is greater
0.01 mg/L
NH4-N mg/L, NH4-N mV
Ammonia*
0 to 200 mg/L NH3-N
-
0.01 mg/L
NH3-N mg/L
Chloride**
0 to 18000 mg/L Cl
±15% of reading or 5 mg/L, whichever is greater
0.01 mg/L
Cl mg/L, Cl mV
Nitrate**
0 to 200 mg/L NO3-N
±10% of reading or 2 mg/L, whichever is greater
0.01 mg/L
NO3-N mg/L, NO3-N mV
Depth
0 to 328 feet (0 to 100 m)
±0.013 ft (±0.004 m) for 1, 4, and 10 m cables
±0.13 ft (±0.04 m) for cables 20 m and longer
0.001 m or 0.01 ft
m, ft
*Derived/calculated parameter **ISEs for freshwater only; 20 meter maximum depth
ProDSS System Specifications (Instrument, Sensor, and Cable)
Sensor/Parameter
Sensor Type/Measurement Method
Calibration
Maximum Depth
Warranty
Temperature
Thermistor, installed on conductivity sensor
Not available
100 m
2 years for conductivity/temperature
sensor
pH
Combination glass bulb electrode,
Ag/AgCl reference electrode with gelled electrolyte
1, 2, or 3 point
100 m
2 years for pH and pH/ORP sensors
ORP
Platinum button with Ag/AgCl reference
1 point
100 m
Dissolved
Oxygen
Optical luminescence - lifetime method
1 or 2 point
100 m
Barometer
-
1 point
-
3 years, integrated into ProDSS handheld
Conductivity
Four nickel electrode cell
1 point
100 m
2 years for conductivity/temperature
sensor
Specific Conductance*
Calculated from conductivity and temperature
1 point
-
-
Salinity*
Calculated from conductivity and temperature
1 point
-
-
Total Dissolved Solids
(TDS)*
Calculated from specific conductance and a userselectable TDS multiplier (0.30 to 1.00; default
0.65)
-
-
-
Resistivity*
Calculated from conductivity and temperature
-
-
-
Seawater Density*
Sigma is calculated from salinity, temperature, and
pressure (depth)
Sigma T is calculated from salinity and temperature
-
-
-
Turbidity
Nephelometric - Optical, 90° scatter
Meets ISO 7027
1, 2, or 3 point
-
2 years for turbidity sensor
Total Suspended Solids
(TSS)*
User correlated from turbidity field measurements
and lab TSS measurements from grab samples
-
-
-
Ammonium**
Ion selective electrode
1, 2, or 3 point
20 m
2 years for ammonium sensor
1 year for pH and pH/ORP sensor
modules
2 years for pH/ORP sensor
1 year for pH/ORP sensor module
2 years for optical DO sensor
1 year for optical DO sensor cap
6 months for ammonium sensor module
Ammonia*
Calculated from ammonium, temperature, salinity,
and pH
-
-
-
Chloride**
Ion selective electrode
1, 2, or 3 point
20 m
2 years for chloride sensor
Nitrate**
Ion selective electrode
1, 2, or 3 point
20 m
2 years for nitrate sensor
Depth
Pressure transducer
1 point
-
2 years, integrated into cable assembly
6 months for chloride sensor module
6 months for nitrate sensor module
*Derived/calculated parameter
**ISEs for freshwater only ; 20 meter maximum depth
ProDSS Order Guide
Step 1: Order Instrument
All instruments include a rechargeable lithium-ion battery (pre-installed), hand strap, USB cable for charging the ProDSS
battery and for connection to a PC, universal AC charger, cable for connection to a USB memory stick, ProDSS quick start
guide, and USB memory stick containing KorDSS software and digital copy of the comprehensive manual.
626870-1: Handheld without GPS
626870-2: Handheld with GPS
Step 2: Order Cable
Cables do not include an integrated temperature sensor. All cables include 4 universal sensor ports and a maintenance kit (3 port plugs, tube of Krytox® lubricant, brush for conductivity probe, syringe for depth sensor, and sensor
installation/removal tool). A 1 lb. weight is included with all cables 10 meters and longer while 10, 20, and 30 meter
cables include a cable management kit.
626909: No depth sensor integrated in 1, 4, 10, 20, 30, 40, 50, 60, 70, 80, 90, or 100 meter cable.
626910: Depth sensor integrated in 1, 4, or 10 meter cable.
626911: Depth sensor integrated in 20, 30, 40, 50, 60, 70, 80, 90, or 100 meter cable.
ProDSS Order Guide continued
Step 3: Order Smart Sensor(s)
A conductivity/temperature sensor must be installed in a ProDSS cable for accurate measurement of all parameters. All
sensors, including conductivity/temperature, must be ordered separately. Please visit ysi.com/ProDSS for replacement
module ordering information.
626900: Optical dissolved oxygen (ODO)
626901: Turbidity
626902: Conductivity and temperature
626903: pH sensor with module
626904: pH/ORP sensor with module
626905: Nitrate sensor with module
626906: Ammonium sensor with module
626907: Chloride sensor with module
Step 4: Order Standards
065270: Conductivity standard, 1000 umhos/cm (quart)
003841: 1 mg/L ammonium standard (500 mL)
065272: Conductivity standard, 10000 umhos/cm (quart)
003842: 10 mg/L ammonium standard (500 mL)
065274: Conductivity standard, 100000 umhos/cm (quart)
003843: 100 mg/L ammonium standard (500 mL)
603824: 2 pints each of 4, 7, and 10 pH buffers
003885: 1 mg/L nitrate standard (500 mL)
061320: Zobell standard for ORP (mV) calibration, 125 mL
003886: 10 mg/L nitrate standard (500 mL)
608000: Turbidity standard, 0 FNU (1 gallon)
003887: 100 mg/L nitrate standard (500 mL)
607200: Turbidity standard, 12.4 FNU (1 gallon)
607300: Turbidity standard, 124 FNU (1 gallon)
607400: Turbidity standard, 1010 FNU (1 gallon)
005580: Confidence Solution® to verify pH, ORP and conductivity data (not for calibration)
Step 5: Order Accessories
626946: Large, hard-sided carrying case
626846: Replacement Li-ion battery pack
603075: Large, soft-sided carrying case
626969*: ProDSS USB flash drive
626945: Small, hard-sided carrying case (1 and 4 meter) 626991*: USB cable for charging and PC connection
599080: Flow cell
626992*: Cable for connection to USB stick
603056: Flow cell mounting spike
626990**: Cable maintenance kit
063507: Tripod
626919**: Probe guard kit
063517: Ultra clamp
599786**: Calibration/storage cup
603070: Shoulder strap
603062***: Cable management kit
626918****: 1 lb weight
603069: Belt clip
605978: 4.9 oz weight
626942: USB car charger
626943: Small external battery pack. Will typically recharge battery to ~50% when completely discharged
626944: Large external battery pack. Will typically recharge battery and charge second battery to ~20%
626940*: AC charger (USA). Includes power supply and USB cable (626991)
626941*: AC charger (International). Includes power supply, USB cable and outlet adapters
*Included with new ProDSS. Only one (1) AC charger is included based on region **Included with all cables
***Included with all 10, 20, and 30 meter cables ****Included with all cables 10 meters and longer
YSI
YSI is a registered trademark. Specifications are subject to change. Please visit YSI.com to verify all specs.
©2015 YSI, a xylem brand
Printed in the USA. W83-03 July, 2015
1725 Brannum Lane, Yellow Springs, OH 45387
Tel +1 937.767.7241 800.897.4151 (US)