ALLPRO IMAGING ScanX DVM, ScanX-12 DVM Digital Imaging System Instruction Manual
Below you will find brief information for Digital Imaging System ScanX DVM, Digital Imaging System ScanX-12 DVM. The ScanX is a digital radiography system that utilizes reusable Phosphor Storage Plates (PSP) in place of X-ray film to produce diagnostic quality intraoral and extraoral digital radiographs. The ScanX produces a digital image by scanning PSPs, which have been exposed to X-rays. The ScanX allows computer storage, processing, retrieval and display of the computed radiographic images utilizing a user supplied software and computer. An additional feature of the ScanX consists of an in-line plate eraser function that removes the latent image from the plate immediately after scanning.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
Part Numbers: B7150 and B7350
Instruction Manual
ScanX-12 DVM
P/N B7350
ScanX DVM
P/N B7150
FOREWORD
ALLPRO Imaging has prepared this document as a guide to the proper use of two ScanX
ScanX ®
®
DVM
Digital Imaging Systems with In-Line Erase, ScanX
.
® -12 DVM and
TABLE OF CONTENTS
Section Page
Congratulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
On-Line Warranty Registration . . . . . . . . . . . . . . . . . . . . . . . . 3
Safety Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Important Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Purpose of this Manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
System Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Unpacking and Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . 9
If You Need Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Computer System Requirements . . . . . . . . . . . . . . . . . . . . . . . 10
Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Site Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
System Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Plate Care and Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Intraoral Imaging Procedures . . . . . . . . . . . . . . . . . . . . . . . . 19
Extraoral Imaging Procedures . . . . . . . . . . . . . . . . . . . . . . . . 22
Powering Down the System . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Scheduled Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 254
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
VISIX Imaging Software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
2
CONGRATULATIONS
Congratulations on your purchase of the ScanX-12 DVM or ScanX DVM Digital
Imaging System with In-Line Erase another innovative imaging product from ALLPRO
Imaging, a leading manufacturer of medical and veterinary equipment since 1962
The ScanX-12 DVM is capable of processing both intraoral and extraoral imaging plates. The ScanX DVM, however, is designed to process intraoral imaging plates only. For details, see the sections UNPACKING THE UNIT and SPECIFICATIONS.
The ScanX-12 DVM and ScanX DVM are hereafter referred to as ScanX in this manual. Each has been designed and manufactured using state-of-the-art technology to give many years of dependable service.
This manual covers the installation, operation and maintenance of the ScanX.
Review and follow the guidelines included in this manual to ensure that your ScanX gives the highest level of service.
WARRANTY
The ScanX is warranted to be free from defects in material and workmanship from the date of installation for a period of one year. The ScanX is designed solely for use in an office environment and this warranty is not applicable to other applications.
If your ScanX cannot be repaired in the field and turns out to be defective due to faulty materials and/or workmanship within the warranty period, ALLPRO Imaging will arrange to replace the unit at its expense within two business days. ALLPRO
Imaging will ship the Customer a replacement, factory refurbished ScanX. The
Customer then returns (at ALLPRO Imaging’s expense) the malfunctioning system back to ALLPRO Imaging in the same shipping container
Upon determination of eligibility by ALLPRO Imaging’s Technical Support personnel, the
Customer must return the malfunctioning ScanX within 15 business days or Customer will be invoiced for the replacement ScanX. Any returned units become the property of ALLPRO Imaging. In case of replacement of the product, the Warranty Period will remain in effect for 6 months or for the remaining period of the original warranty; whichever term is longer.
The ScanX Standard Swap Warranty Coverage applies provided the product is handled properly for its intended use, in accordance with its operating instructions. This warranty does not apply to damage due to shipping, misuse, careless handling or repairs by non-authorized personnel. Warranty is void if installed or serviced by other than authorized service personnel. If unit fails, and the problem is determined to be caused by the lack of recommended scheduled maintenance as outlined in this Instruction Manual (i.e. worn Belt Drives or Brushes) , this failure would not be covered. This warranty does not cover accessories.
The Phosphor Storage Plates (PSPs) are designed for use with the ScanX and will be replaced for a period of 30 days from the date of purchase if defective in manufacturing or packaging.
ON-LINE REGISTRATION
Quickly and easily register your new ScanX on-line. Just have your product model and serial numbers available. Then go to ALLPRO’s web site, www.allproimaging.com
, click the Warranty Registration link and complete the registration form. This on-line registration ensures a record for the warranty period and helps us keep you informed of product updates and other valuable information.
3
SAFETY NOTICE
This equipment has been designed to minimize exposure of personnel to hazards. While the ScanX is designed for safe operation, certain precautions must be observed. Use of the ScanX not in conformance with the instructions specified in this manual may result in permanent failure of the unit.
General Safety Information.
Check with your authorized dealer for packing material requirements if it is necessary to return the product to the manufacturer. Correct packing guarantees optimal safety of the device during transport. Should it become necessary to return the device to the manufacturer during the warranty period, Air Techniques will not accept claims for damage arising from using incorrect packing materials.
Before every use, the operator must check the functional safety and the condition of the device.
The operator must be knowledgeable in the operation of the device.
This device is not to be used in any areas where the atmosphere could cause fire or explosion.
Markings.
The following terms or symbols are used on the equipment or in this manual to denote information of special importance:
CAUTION
CLASS 1 LASER PRODUCT
CLASS 3B LASER
RADIATION WHEN OPEN
AVOID EXPOSURE TO BEAM
DANGER
LASER RADIATION WHEN OPEN
AVOID DIRECT
EXPOSURE TO BEAM
COVER REMOVED MAKES
THIS DEVICE A
CLASS IIIb LASER PRODUCT
The ScanX is a Class I Laser Product [Class 1 Laser Product (IEC)]
This warning label identifies the ScanX as such a product and describes the potential danger to humans in the event the product is opened during service. There is no laser radiation from this product when operated and maintained as instructed.
The Laser Product Accession Number is 0212282-00
Alerts users to important Operating and
Maintenance instructions. Read carefully to avoid any problems.
Warns users that uninsulated voltage within the unit may be of sufficient magnitude to cause electric shock.
LABORATORY
EQUIPMENT
60CB
E234737
Indicates the
ScanX is a UL
Listed product.
2
Indicates item used only once. Discard after use.
ATTENTION USERS:
Manufacturing date code on serial number label is in the format:
MONTH
YYYY.
4
SAFETY NOTICE
Authorized Dealer Service Only.
The interior of the ScanX is only accessible by removing hardware with tools. It should be opened and serviced only by an authorized dealer service technician.
Failure to heed this warning may result in equipment damage or personal injury, and will void any and all warranties. Contact your authorized dealer for service information.
Use of Accessory Equipment.
The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system.
Consideration relating to the choice shall include:
- use of the ACCESSORY in the PATIENT VICINITY
- evidence that the safety certification of the ACCESSORY has been performed
in accordance to the appropriate IEC 60601-1 and/or IEC 60601-1-1
harmonized national standard.
Use of ACCESSORIES or cables other than those specified or provided by Air
Techniques may result in increased EMISSIONS or decreased IMMUNITY of the
EQUIPMENT.
Electrical Safety Notes.
The line cord is the main power disconnect device.
Use only the line cord provided with the unit.
Use only grounded electrical connections.
To avoid risk of electric shock, fire, short-circuit or dangerous emissions, never insert any metallic object into the equipment.
Only use connection cable(s) delivered with the device.
Check the device cables for possible damage before switching on.
Damaged cables, plugs and sockets must be replaced before use.
Never touch open supply outlets and subjects simultaneously.
Do not locate unit where it could be sprayed with water, or in a damp environment.
5
SAFETY NOTICE
Knowledge of Warnings and Cautions.
Users must exercise every precaution to ensure personnel safety, and be familiar with the warnings and cautions presented throughout this manual and summarized below.
In this manual, the following definitions apply for all WARNINGS and CAUTION
Statements:
WARNINGS: Any operation, procedure or practice, which, if not strictly observed, may result in injury or long-term health hazards to personnel.
CAUTIONS: Any operation, procedure or practice, which, if not strictly observed, may result in destruction of equipment or loss of effectiveness or damage to equipment and
Phosphor Storage Plates (PSPs).
DANGER: Opening the ScanX by removing any covers or components makes the equipment into a Class III b Laser Product. [Class 3B Laser Product (IEC)].
WARNINGS -
Only trained professionals should use this device.
Federal law prohibits the sale of this device to individuals other than licensed professionals. Use of this device, other than as described in this manual, may result in injury.
The ScanX contains a laser and is a Class 1 [Class 1 (IEC)]
Laser Product.
Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure. The laser is on only during an active scan.
Only a trained technician from an authorized dealer should remove a cover from the ScanX.
Direct eye contact with the output beam from the laser may cause serious damage and possible blindness.
Do not open the ScanX to maintain it.
The ScanX contains no user serviceable parts. If there is a service problem, contact your authorized dealer.
Operate ScanX in dry environment.
To prevent fire or electrical shock, do not expose this appliance to rain or moisture.
Do not use damaged Phosphor Storage Plates (PSPs).
Damaged PSPs may not provide reliable diagnostic images.
Do not reuse the Barrier Envelopes.
Dispose of used Barrier Envelopes in accordance with all local regulations.
Imaging Plates (PSPs) are toxic.
Never place an intraoral PSP in the mouth without first enclosing it in a sealed barrier envelope. Try to prevent the enveloped PSP from being bitten, chewed or swallowed.
Disposal of PSPs.
Consult with your federal, national, state and local government, for rules and regulations on disposal of PSPs.
6
SAFETY NOTICE
CAUTIONS -
Completely clean and erase PSPs before taking an X-ray exposure.
See the PLATE PREPARATION section of this manual.
Minimize exposing an X-ray exposed PSP to light.
Transfer the PSP into the Inlet slot quickly to minimize exposure to light.
Use care in handling PSPs - Avoid fingerprints and scratching
Refer to the instructions provided with the PSP package for further information on handling.
Use of other manufacturer’s imaging plates
Do not put PSPs designed for drum-type or other scanners in the ScanX. The hooks and/or frames on the ends or around these PSPs, or PSPs of different thickness (especially thicker ones) will damage the ScanX. PSPs from another manufacturer may be used as long as the specifications are the same as those for the ScanX PSPs.
IMPORTANT INFORMATION
General Notes.
All instructions in this manual form an integral part of the unit. They must be kept close to the unit and in readiness whenever required. Precise observance of these instructions is a pre-condition for use of the unit for the intended purpose and for its correct operation. This manual should be passed on to any future purchaser or operator.
Safety of the operator as well as trouble-free operation of the unit are only ensured if use is made of original equipment parts. Moreover, use may only be made of those accessories that are specified in the technical documentation or that have been expressly approved and released by
Air Techniques for the intended purpose. Air Techniques cannot warranty for the safety or proper functioning of this unit in the case where parts or accessories are used that are not supplied by Air Techniques.
There is no guarantee against damage arising where parts or accessories are used that are not supplied by Air Techniques.
Observe the usage and storage conditions.
Appliances which accumulate condensation or become wet through a change of temperature may only be operated after they are fully dry again.
Air Techniques regard themselves as being responsible for the equipment with regard to safety, reliability and proper functioning only if assembly, resetting, changes or modifications and repairs have been carried out by an authorized dealer and if the equipment is used in conformity with the instructions contained in this manual.
The device conforms to the relevant safety standards valid at this time.
Any reprinting of the technical documentation, in whole or in part, is subject to prior written approval by Air Techniques.
7
IMPORTANT INFORMATION
Correct Usage
Operation of the ScanX may only be carried out by suitably qualified personnel.
The ScanX is only to be used in the processing of exposed PSPs.
The ScanX should be used in a room equipped for it.
Room temperature should be in the range 50 to 105°F (10 to 40°C) with relative humidity between 5 and 95%.
If the device is brought into the room of operation from a cooler environment, condensation can build up. Do not connect the device until it has warmed up to room temperature and is absolutely dry.
This room should be free of all possible interferences (e.g. strong magnetic fields), as these could affect the operation.
The ScanX may only be operated together with authorized software such as
Visix or VetRay Imaging Software.
Correct usage includes observing all installation and operating instructions and adherence to the set-up, operation and maintenance instructions.
Any use, above and beyond that described in this manual as correct usage, will invalidate the warranty.
Incorrect Usage
Any use that is not described in this manual as correct usage is considered as incorrect usage. The manufacturer is not to be held liable for any damage caused as a result of incorrect usage. The operator bears all risks.
PURPOSE OF THIS MANUAL
This manual provides the information necessary for the setup, operation and routine care and maintenance of the the ScanX ® -12 DVM (P/N B7350) and the ScanX ®
DVM (P/N B7150) Digital Imaging Systems with In-Line Erase.
This manual is not to be used as a replacement for training in radiography.
For information regarding the computer system and imaging software, refer to the appropriate documentation provided with your computer hardware and software.
SYSTEM DESCRIPTION
The ScanX is a digital radiography system that utilizes reusable Phosphor
Storage Plates (PSP) in place of X-ray film to produce diagnostic quality intraoral and extraoral digital radiographs. The ScanX produces a digital image by scanning
PSPs, which have been exposed to X-rays. The ScanX allows computer storage, processing, retrieval and display of the computed radiographic images utilizing a user supplied software and computer. An additional feature of the ScanX consists of an in-line plate eraser function that removes the latent image from the plate immediately after scanning. This design provides an efficient one-operation scanning and erasing process leaving the user with a PSP ready for the next
X-ray procedure.
8
UNPACKING AND INSPECTION
Unpacking
The ScanX may be shipped in a very special shipping case covered by a doublewall corrugated over-packaging. A deposit fee has been paid for this case, which may be returned to Air Techniques, Inc./ALLPRO Imaging for a full refund of the fee using a Return Authorization (RA) number, or retained, without refund of the fee, to protect the ScanX during transport or future shipping. Retain the over-packaging to return the shipping case.
Included System Components
Each ScanX consists of the indicated main assembly and accessory kit as listed below: (See SPECIFICATIONS for ratings and identification for specific models.)
Description
ScanX DVM,
P/N B7150
B7110V
ScanX-12 DVM,
P/N B7350
B7310V Main ScanX 12 ILE Assembly
Accessory Kit containing:
Power Supply Adapter
Line Cord
Computer Connector Cable, USB
Quick Start Instructions
PSP Protective Cushion Kit
Plate Guides
(Sizes 0, 1, 3, 4 Qty. 1 ea. and Size 2, Qty. 4)
ScanX Cleaning Sheet Sample Kit
Dust Cover
Drivers, Utilities and Instruction Manual Disk
PSP Wipes Sample Kit
B7095
73096
B3554
B7208
-
73444V
B3030
-
B7365
B8912
B7095
73096
B3554
B7208
74030B
73444V
B2030
74508
B7365
B8912
Inspection
Verify that all listed items were received. If any item is missing, notify your authorized dealer. Unpack each component of the ScanX and inspect for physical damage such as scratched panels, damaged connectors, etc. If any damage is noted, notify your authorized dealer immediately so corrective action can be taken.
IF YOU NEED ASSISTANCE
ALLPRO Imaging ScanX systems are designed and manufactured to high standards.
They are easy to install and use and typically deliver high-quality performance. If any difficulties are encountered with this product, please contact ALLPRO Imaging
Technical Support at 1-800-247-8324.
For additional information, please contact your authorized dealer or visit our web site, www.allproimaging.com
Air Techniques, Inc., Allpro Imaging
1295 Walt Whitman Road
Melville, New York 11747 USA
9
COMPUTER SYSTEM REQUIREMENTS
IMPORTANT: To operate the ScanX, it must be connected to a compliant Computer System, and the computer must be loaded with an authorized Imaging Software such as
Visix or VetRay. Neither the Computer System nor the Software is provided by the manufacturer. Contact your dealer for available Computer System and Software options.
Computer System Requirements
The Computer System (computer, monitor, etc.) and any related peripheral or other equipment, supplied by the user, or a third party, must comply with the requirements for Information Technology Equipment (ITE) as specified in
IEC 60950 (EN 60950).
Operating System: Windows XP Professional with Service Pack 3 or later for an Intel 32-bit processor;
Microsoft Windows XP Professional 64-bit Edition with
Service Pack 2 or later for an Intel 64-bit extended (x64) processor;
Microsoft Windows Vista Business, Enterprise, or
Ultimate with Service Pack 1 or later for an Intel 32-bit or an Intel 64-bit extended (x64) processor; or
Microsoft Windows 7 Professional, Enterprise, or
Ultimate for an Intel 32-bit or an Intel 64-bit extended
(x64) processor.
USB Port/Version: USB 2.0 or later
Hard Drive: 200MB available disk space required to start scanning
Image Management Compatible authorized third-party software
Software: (not included with product).
Optical Drive: CD-ROM Capable
Recommended Components
The items listed below are recommended (but not required) computer system components to aid in ScanX operation
CPU/Speed: 3.0 GHz Intel CORE2
System RAM:
Monitor:
2 GB
20-inch SVGA, 1024 x 768 or higher resolution, contrast ratio 450:1,
.22 dot pitch capability.
Video Display Adapter:
Peripherals:
32 MB RAM
Standard Keyboard & Mouse
Backup Device
External Surge Protector
Power supply backup
10
COMPUTER SYSTEM REQUIREMENTS
System Properties.
If unsure of the operating system version installed, check that it meets the necessary requirements by checking the System Properties window.
This is done simply by right clicking the My Computer
icon. Selecting Properties from the menu list displays the System Properties window as shown. The installed operating system version is listed under the General tab.
The System Properties window can also be opened from the Desktop Start button. Just press the Start button and select SettingsControl Panel and then click System icon.
CONTROLS AND INDICATORS
Ready/Standby
Green LED indicator
ERASER
Switch
FLASHING OFF
Blue LED Indicator
READY
Switch
Keypad Control Panel
4 Scanner Track
Indicator LEDs
Keypad Control Panel
READY Switch
Ready/Standby
Green LED indicator
ERASER Switch
FLASHING OFF
Blue LED indicator
Toggles between the Standby and Ready mode as follows:
1. Press to switch from the Standby mode to the Ready mode.
2. Press and hold down for at least 2 seconds to switch to the
Standby mode from the Ready mode.
Illuminates green to indicate that the ScanX is Ready for operation.
When extinguished, it indicates that the ScanX is in the Standby mode.
Toggles between turning the Erase function OFF and ON. This switch has no effect once the plate scanning or erasing starts.
Illuminates steady blue to indicate that the Erase function is ON and
PSPs will be erased after scanning.
Flashes blue to indicate that the Erase function is OFF and PSPs will not be erased after scanning.
4 Scanner Track Indicator LEDs
Illuminate green when the Scanner has been activated, indicating that a PSP can be fed into the ScanX.
Illuminate yellow, indicating the PSP has been sensed and the
Scanner is transporting the PSP.
11
TECHNICAL DATA
Electrical Requirements:
Supply Voltage:
Supply Current:
Line Cord:
Power Supply:
Physical Properties:
ScanX Part Number
Depth (inches)
Width (inches)
Height (inches)
Weight (pounds)
100 to 240VAC +/- 10%, 50/60 Hz
1.2 A Maximum
North American style 10 foot long
Hospital Grade power cord, P/N 73096.
24 Volt Power Supply, P/N B7095 provided
B7110V
15.13
15.0
15.5
43
B7310V
15.4
15.0
27.8
50
Environmental Conditions:
Unit in Operation
Temperature:
Humidity:
Heat emission
Storage and Transport
Temperature:
Humidity:
50°F to 105°F (10°C to 40°C)
5% to 95% RH
<90W
-21°F to 130°F (-29°C to 54°C)
5% - 95% (Non-condensing)
Note: Resolution of the ScanX is dependent on operating mode and specific imaging plate type used.
Resolution Range:
Horizontal:
Vertical:
P/N B7110V
10 to 14 (lp/mm)
P/N B7310V
2.5 to 14 (lp/mm)
10 to 18+ (lp/mm) 2.5 to 18+ (lp/mm)
Compliance Data:
Laser Class:
Laser Product Report
Accession Number:
Installation Category: I
Class I Laser Product (21 CFR 1040.10)
Class 1 Laser Product (IEC 60825-1)
0212282-00
Pollution Degree: 2
User Replacement Items: Power Supply, Line Cord, Plate Guides, PSPs and Computer Connector Cable
IEC 61010-1 Classification:
Class 1, Transportable, Continuous Operation,
Equipment not suitable for use in the presence of flammable anaesthetic mixture(s). Protection against ingress of liquids -Ordinary
Electromagnetic Interference:
Electromagnetic interference between the equipment and other devices can occur. Do not use the equipment in close conjunction with sensitive devices, or devices creating high electromagnetic disturbances.
12
ABBREVIATIONS
Abbreviations used in this manual are summarized below.
A
AC
CD-ROM compact disk, read-only memory
CFR
CPU cm
D
GB
GHz
H
Hz
IEC
IMS ampere(s) alternating current
Code of Federal
Regulations central processing unit
(your computer) centimeter depth gigabyte (2 height
30
Gigahertz (10
10
9
9 bytes)
of Hertz)
Hertz (cycles per second)
International Electro-technical Commission
Image Management
Software
Light emitting diode LED lbs. pounds lp/mm line pair per mm lux a measure of light intensity
MB megabytes
(2 20
10
6 bytes) mm millimeter (103 m)
MONTH YYYY date
Phosphor
PN
PSP
(Month, 4 digit year) a luminescent material part number phosphor storage plate
(imaging plate)
RAM
RH
SVGA
USB
UL
V
W
°C
°F
”, in random access memory relative humidity
Super Video Graphics
Array
Universal Serial Bus
Underwriters
Laboratories
Volts
Watts, width degree Celsius degree Fahrenheit inch
SITE SELECTION
Note: The ScanX is designed to be installed by your authorized dealer. The dealer or user must provide appropriate and compliant computer hardware and software.
The ScanX may be located almost anywhere in the office. Follow these guidelines for optimum performance:
Lighting conditions: Set up the scanner in ordinary room light, however, direct sunlight and light fixture(s) above and near the ScanX producing more than 400 lux of light at the PSP inlet must be avoided.
Provide a stable, flat countertop large enough to hold the scanner, plus provide a working area for resting and opening cassettes.
Locate the computer within 6 feet.
Access to a hospital grade grounded electric Mains AC outlet using the line cord provided, must be within line cord length.
13
SYSTEM SETUP
Note: Authorized Imaging Software, such as Visix or VetRay, supplied by the dealer or other company, must be installed on the computer in order to operate the ScanX.
ScanX Drivers and Utilities Installation
Before connecting the ScanX to your computer or attempting to use it for the first time, run the Setup program on the ScanX Drivers and Utilities Disk included with the ScanX. Normally, this program runs automatically when the CD is inserted into the drive for the first time. If not, run the Setup program located in the root directory of the CD (typically D:\Autorun.exe).
The Setup program guides you through updating the library files on your computer, which must be completed before the ScanX will operate properly.
More information can be found in the Installation Instructions and Notes file on the ScanX Drivers and Utilities Disk included with the ScanX.
ScanX Connection Procedure
Refer to Figure 1 and perform the following procedure to connect the ScanX for operation for the first time.
1. Select a location that meets the SITE SELECTION guidelines.
2. Set up the computer according to the manufacturer's recommendations.
Make sure that the computer meets all requirements listed on page 10.
3. Verify that an authorized Imaging Software is installed properly on the computer.
4. Verify that the supplied ScanX Drivers and Utilities Disk which contains the
USB drivers was properly installed per instructions above.
5. Connect the high speed USB cable between the USB Type B connector located on the ScanX rear panel and the USB Type A connector located on the computer.
Note: Connect the 24V Power Supply Adapter to the ScanX prior to plugging the line cord into the Mains outlet.
6. Connect the 24V Power Supply Adapter Output Connection Cable to the
ScanX DC Power Input Jack.
7. Connect the line cord between the Mains outlet and the 24V Power Supply
Adapter.
8. Switch the scanner from standby to ON by pressing the membrane READY switch ( ) located on the Keypad Control Panel on the top of the scanner.
Verify that both the green and blue LED indicators above the READY and
ERASER switches illuminate.
9. With both the ScanX and computer turned on, Windows detects the ScanX as a new USB Device. Windows should automatically find the drivers installed from the ScanX Drivers and Utilities Disk.
14
SYSTEM SETUP
Line Cord
The Line Cord is the Mains disconnect device.
Green LED
Indicator
Blue LED
Indicator
Mains
Outlet
Power
Supply
Power Supply
Line Cord
Computer Connector
Cable
READY
Switch
Press HERE to turn ON or OFF
ERASER
Switch
TO
POWER
SUPPLY
TO
COMPUTER
USB PORT
Connections for ScanX
P/Ns B7300 and B7350
Connections for ScanX
P/N B7150
Figure 1. Typical ScanX Installation
15
PLATE CARE & PREPARATION
Prior to performing the Intraoral and
Extraoral imaging procedures provided on the following pages, the user must be familiar with the care, handling and preparation of the PSP in order to ensure successful image scanning.
Figure 2 shows the configuration of a typical Intraoral Size #2 PSP and Figure
3 shows a typical Extraoral plate.
Printed side
Tube or Sensitive side
Figure 2. Intraoral PSP Configuration
Black Side or Back of PSP
White or Sensitive Side or Front of PSP
Figure 3. Typical Extraoral Plate Configuration
Handle PSPs with Care.
Avoid scratching or soiling PSPs.
Do not bend PSPs or apply unnecessary pressure.
Do not store PSPs in a hot or moist area.
Protect the PSPs from direct sunlight and ultraviolet rays.
Pick up the PSPs using two fingers around the edges to avoid unnecessary contact with the plates.
CAUTION: Always use a Barrier Envelope for each Intraoral plate. Exercise care in handling the plate so as not to scratch the sensitive surface or nick the edges.
Plate Protection
When storing or transferring PSPs use a Plate Transfer Box for Intraoral PSPs or an X-ray Cassette for Extraoral PSPs.
Plate Transfer Box. Place the Intraoral PSP covered with a Barrier Envelope into the Plate Transfer Box with the sensitive (front) side of the PSP facing down.
Note: Cassettes must not contain intensifying screens when using PSPs.
X-ray Cassette. Place the Extraoral PSP into the appropriate X-ray Cassette with the sensitive (front) side of the PSP towards the Tube-side of the cassette and close cassette.
16
PLATE CARE & PREPARATION
IMPORTANT: PSPs must always be erased prior to use.
Note: Use PSPs within 24 hours of last erasure. Repeat erasing process if PSPs have been stored longer than 24 hours.
Erase the PSP
Each Intraoral and Extraoral PSP should be used (i.e. X-ray exposed and scanned) within 24 hours of erasure since natural radiation will add noise to the PSP. Erase
PSPs by simply using the ScanX In-Line Erase Feature. This can be accomplished using one of two methods as follows:
Note: Both erasing methods will result in an erased PSP suitable for reuse. The user will not observe any difference in ScanX operation when using either method.
Method #1
Perform the Imaging Procedures for Intraoral PSPs on page 20 or the procedures for Extraoral PSPs on page 22. Except when performing step 4 of either Activate Scanner procedure, select the Erase option from the installed authorized imaging software to activate the ScanX. This method does not scan the plate and no image will be acquired.
Method #2.
Perform the normal Imaging Procedures for Intraoral PSPs on page 20 or the procedures for Extraoral PSPs on page 22. This method scans the plate and the imaging software may acquire a “junk image” (scanned latent plate image) that should be subsequently deleted from the imaging software.
IMPORTANT: Wipe the phosphor side of the plate only making sure not to scratch the plates.
Cleaning Phosphor Storage Plates
For the best images, PSPs should be handled carefully and kept clean. Use the following steps to clean plates:
1. Use lint-free, 100% cotton gauze (not cotton balls). Gently wipe the cotton gauze over the dry Plate surface. Wipe using an even side to side motion and then in a circular motion.
2. To clean any remaining stains wipe using an even side to side motion per one of the following options:
Option 1 – Use Air Techniques PSP wipes (P/N B8910).
Option 2 - 100% cotton gauze dampen with anhydrous ethanol or anhydrous
isopropyl alcohol.
3. Completely dry the surface by wiping with another piece of lint-free cotton gauze. Make sure that the PSP is completely dry before use.
IMPORTANT: Do not soak overnight.
Disinfecting the Phosphor Storage Plates
There is no reason to routinely disinfect the PSPs unless contamination is suspected.
If a PSP has touched a contaminated surface, it may be immersed briefly in a cold sterilant (such as a 2% Gluteraldehyde solution) according to sterilant manufacturers directions. Do not immerse the plate if there is any evidence of deep scratches in the surface of the plate or nicks in the edges of the plate. After disinfection, clean and dry the plate using the instructions above.
Disposal of Phosphor Storage Plates
Consult with your federal, national, state and local government, for rules and regulations on disposal of Phosphor Storage Plates.
17
PLATE CARE & PREPARATION
Barrier Envelope Handling
Always guard against contamination by using standard infectious control procedures when handling individual barrier envelopes. It is best to discard suspected contaminated envelopes since no cleaning or disinfection procedure exist or are required for barrier envelopes.
2 The Barrier Envelope must be used only once and disposed of properly in accordance with local code.
Preparing Intraoral Plates for Patient Use
Insert the erased PSP into the Barrier Envelope so the printed side of the PSP is visible through the transparent side of the envelope. Peel off the adhesive strip and seal the envelope. See Figures 4a - 4c below.
Figure 4a
Insert the PSP.
Figure 4b
Peel off the adhesive strip.
Figure 4c
Seal the Barrier Envelope.
18
INTRAORAL IMAGING PROCEDURES
Note: 1. The orientation letter “a”, printed on the PSP, may be used for reference
as you would use the dot on an intraoral X-ray film. In addition, a backwards “a” a
2. If using holders with alligator clips, it is advisable to file down the points to avoid puncturing the Barrier Envelope. It is important to take care not to puncture the
Barrier Envelope or damage the PSP.
Take an X-ray Image
Put an image on the PSP by performing the following procedure.
1. Place the erased intraoral PSP in the sealed Barrier Envelope into the subject's mouth exactly as you would use X-ray film. Make sure the opaque side of the Barrier Envelope is facing the tubehead.
2. Take the exposure. The X-ray dose may typically be reduced by 80 - 85% of that required for D-speed intraoral film (depending on X-ray system used; the actual X-ray dose should be determined through experimentation).
3. Wearing gloves, remove the exposed PSP from the subject's mouth and place to the side making sure the sensitive side of the barrier envelope is facing away from any light source.
4. Repeat steps 1 through 3 as necessary to complete the subject's X-ray series. When all necessay plates in the X-ray session have been exposed, prepare each plate by performing the procedure below.
Preparing the Exposed Plate for Scanner Processing
1. Disinfect the Barrier Envelope (with plate still inside) and your gloves by washing with disinfectant hand soap and water. Dry the Barrier Envelope thoroughly.
2. Remove gloves and wash any powder from hands. Powder on a PSP will degrade the image, and an accumulation of powder in the scanner will lead to degradation of scanner performance.
IMPORTANT: Be sure that the sensitive side of the PSP is facing down when it lands inside the box (See Figure 5). If it is not sensitive side down,
TURN IT OVER IMMEDIATELY. Failure to do so may result in erasure of the PSP.
3. Open the Plate Transfer Box with the lid away from you. Then, remove the exposed PSP from its washed and dried envelope as follows: a. Hold the enveloped exposed PSP, with the printed side facing up, parallel to and about an inch above the open
Plate Transfer Box.
b. Tear the envelope lengthwise starting at the notch to eject the PSP into the open Transfer Box. See Figure 5.
Figure 5
Ejecting a Plate into the
Plate Transfer Box
4. Bring the Plate Transfer Box with all the exposed PSPs properly prepared to the ScanX. The PSP is now ready to be scanned to read the image from the PS by performing the Intraoral Imaging Procedures.
19
INTRAORAL IMAGING PROCEDURES
Configure the Intraoral Plate Guides. If the desired Plate Guides are not in place, install these Guides at this time. Do not operate the scanner without a full complement of four Guides in place. Any combination of Guides may be used. See Figure 6.
Figure 6. ScanX with a full Complement of Plate Guides in Place.
IMPORTANT: Transfer the PSP from the Plate Transfer Box to the Plate Guide slot quickly. Always keep the sensitive side away from any light source to minimize image loss.
Activate Scanner
Activate the ScanX by performing the following procedures.
1. Make sure the ScanX and computer components are correctly connected as shown in Figure 1.
2. Switch the scanner from Standby to ON by pressing the membrane READY switch ( ) located on the Keypad Control Panel on the top of the scanner.
3. Verify that both the green and blue LED indicators above the READY and
ERASER switches, respectively, illuminate. (Default mode has eraser ON.)
4. Use the user-supplied authorized Imaging
Software to activate the Scanner and to select the desired image type and resolution.
5. Verify that the four Scanner track indicators illuminate green when the Scanner has been activated, indicating that PSPs can be fed into the corresponding Plate Guide.
Scanning and Erasing Plates
Scan and erase an intraoral PSP in one operation by performing the following procedures.
1. Carefully open the Plate Transfer Box.
2. Grasp a PSP by long edges between your thumb and index finger. See Figure 7.
3. With printed side facing you, carefully and quickly insert the PSP into the corresponding
Plate Guide slot as far as possible.
4. Immediately press the PSP all the way down with your fingertip (from position Figure
8 to position Figure 9) until the scanner's transport mechanism takes over and the PSP moves down on its own.
5. Verify that the track light illuminates yellow, indicating the Plate Guide slot is in use, the PSP has been sensed and the Scanner is transporting the PSP.
20
Figure 7. PSP Pick-up.
Figure 8
Feeding Intraoral Plate.
Figure 9.
Fully Inserted Plate.
INTRAORAL IMAGING PROCEDURES
Scanning and Erasing Plates (continued)
Plate Guide at a time as long as the corresponding track indicator light is illuminated green. The next PSP may be fed into a Plate Guide slot only after the corresponding track indicator LED changes from yellow to green.
6. Observe that a red glow eminates from the scanner exit slot.
7. Repeat steps 2 through 6 to process additional PSPs as necessary. Another
PSP may be fed into any Plate Guide slot as long as the track indicator light illuminates green.
8. Observe that each scanned and erased PSP drops into the receiving tray at the bottom of the scanner. Since the ScanX default operation is with the erase mode enabled (blue LED indicator above the ERASER switch is steadily illuminated), each PSP is erased and ready for reuse.
9. Observe that all transport indicators illuminate green and the red glow from the exit slot extinguish after the last PSP drops to the tray.
10. Retrieve the processed (scanned and erased) PSPs for reuse or storage.
Make sure not to scratch the sensitive surface or nick the edges when removing from the scanner outlet.
11. View and save each scanned image using the user-supplied authorized
Imaging Software.
IMPORTANT: PSPs will not be erased after scanning when operating the ScanX with the eraser disabled. PSPs must always be erased prior to exposure to X-rays for new images.
Scanning Plates without Erasing
The ScanX can be operated with the in-line eraser feature turned off. When the eraser mode is disabled, the ScanX scans the same as when the eraser is enabled except that the PSPs are not erased after scanning. Scan an intraoral PSP without erasing the image by performing the following procedures.
1. Activate the scanner by performing the procedures on previous page.
GREEN LED
Indicator
BLUE LED
Indicator eraser mode enabled. This must be disabled prior to scanning to prevent erasing of the scanned PSP.
2. Disable the eraser mode of operation by pressing the membrane ERASER switch located on the
Keypad Control Panel. See Figure 10.
3. Verify that the blue LED indicator above the
ERASER switch blinks blue to indicate that the Erase function is OFF. PSPs will not be erased after scanning.
READY
Switch
ERASER
Switch
Figure 10.
ScanX Keypad Control Panel
4. Insert the exposed PSPs to be scanned into the ScanX Plate Guides by performing the Scanning and Erasing Plates procedures provided on the previous page.
5. The scanned PSPs still contain latent images that require erasure. Make sure to erase each PSP prior to reuse for new images.
21
EXTRAORAL IMAGING PROCEDURES
Note: Cassettes must not contain intensifying screens when using PSPs.
Take an X-ray Image
Put an image on the extraoral PSP by performing the following procedure.
1. Place the erased PSP into the appropriate X-ray cassette, with the Tube side
(sensitive side) of PSP towards Tube side of cassette.
2. Load cassette into the exposure device as previously done for film, and take the exposure following the X-ray device manufacturer’s instructions for
PSPs.
3. Bring the cassette containing the exposed extraoral PSP to the ScanX. The
PSP is now ready to be scanned.
Activate Scanner
Activate the ScanX by performing the following procedures.
1. Make sure the ScanX and computer components are correctly connected as shown in Figure 1.
2. Switch the scanner from Standby to ON by pressing the membrane READY switch ( ) located on the Keypad Control Panel on the top of the scanner.
3. Verify that both the green and blue LED indicators above the READY and
ERASER switches, respectively, illuminate. (Default has eraser mode ON)
Note: Change in motor noise pitch levels is normal when switching to different resolution setting selections.
4. Use the user-supplied authorized Imaging Software to activate the Scanner and to select the desired image type and resolution.
5. Verify that the four Scanner track indicators illuminate green when the
Scanner has been activated, indicating a PSP can be fed into the ScanX.
Note: 1. Leave the four Plate Guides in place when scanning extraoral PSPs.
2. Only one extraoral exposed PSP can be fed into the ScanX at a time. The next PSP may be fed only after all four track indicator light LEDs change from yellow to green.
Scanning and Erasing Plates
Scan and erase an extraoral PSP in one operation by performing the following procedures.
1. Orient the cassette so that the Tube side is facing down and the hinge is away from you.
2. Open the cassette and grasp the PSP by its ends with your finger tips, and quickly (minimizing exposure to ambient light) move it to the ScanX, with Tube side of PSP towards
ScanX.
3. Position the extraoral plate behind the four
Plate Guides against the curved surface and gently slide the plate down (See Figure 11) until the transport mechanism takes over and the plate moves on its own.
4. At this point, all four track lights will turn yellow, indicating the PSP has been sensed and the
Scanner is transporting the PSP.
Figure 11.
Feeding an Extra Oral
Plate into the ScanX
22
EXTRAORAL IMAGING PROCEDURES
Scanning and Erasing Plates (continued)
5. Observe that a red glow eminates from the scanner exit slot.
6. Repeat steps 1 through 5 to process additional PSPs as necessary. Another
PSP may be fed into the ScanX when all four track indicator lights illuminate green.
7. Observe that the scanned PSP exits through the scanner arch. Since the
ScanX default operation is with the erase mode enabled (blue LED indicator above the ERASER switch is illuminated), the PSP is erased and ready for reuse.
8. Observe that all transport indicators illuminate green and the red glow from the exit slot extinguishes after the last PSP drops to the tray.
9. Retrieve the processed (scanned and erased) PSP for reuse or storage in an X-ray cassatte. Make sure not to scratch the sensitive surface or nick the edges when removing from the scanner outlet.
10. View and save the image using an user-supplied authorized Imaging
Software.
IMPORTANT: PSPs will not be erased after scanning when operating the ScanX with the eraser disabled. PSPs must always be erased prior to exposure to X-rays for new images.
Scanning Plates without Erasing
The ScanX can be operated with the in-line eraser feature turned off. When the eraser mode is disabled, the ScanX scans the same as when the eraser is enabled except that the PSPs are not erased after scanning. Scan an extraoral PSP without erasing the image by performing the following procedures.
1. Activate the scanner by performing the procedures on previous page.
Note: Upon activation, the ScanX defaults with the eraser mode enabled.
This must be disabled prior to scanning to prevent erasing of the scanned
PSP.
2. Disable the eraser mode of operation by pressing the membrane ERASER switch located on the Keypad Control Panel. See Figure 10.
3. Verify that the blue LED indicator above the ERASER switch blinks blue to indicate that the Erase function is OFF. The extraoral PSP will not be erased after scanning.
4. Insert the exposed PSP to be scanned into the ScanX by performing the Scanning and Erasing Plates procedures provided on the previous page.
5. Retrieve the processed (scanned only) PSP and store in an X-ray cassatte.
The scanned PSPs still contain latent images that require erasure. Make sure to erase each PSP prior to reuse for new images.
23
POWERING DOWN THE SYSTEM
IMPORTANT: Never power down the system during a scanning session.
The ScanX is designed to be left on continuously during the active day. At the end of the day, or whenever desired, power down the system simply by pressing and holding the membrane READY switch ( ) on the Keypad Control Panel for approximately two seconds, until the green LED above the READY switch extinguishes.
MAINTENANCE
Maintenance Procedures
The ScanX is designed for many years of trouble-free operation. Maintenance as described herein is minimal.
IMPORTANT: Do not spray solvents or liquid directly on the scanner.
Cleaning the ScanX
Turn off the ScanX disconnect the line cord from the Mains wall outlet and disconnect the computer connection cable from the ScanX before cleaning. Wipe the outside surfaces with a soft paper towel dampened with a disinfectant solution or non-abrasive household cleaner. Be careful not to allow solvents TO RUN
OR DRIP into the ScanX. This could cause damage to the ScanX. Allow to air dry before plugging in or turning back on.
Note: ScanX Cleaning Sheets cannot be used on the ScanX
(P/N B7100) because of height limitations .
®
Intraoral
Cleaning the Plate Transport
Over time, small debris and dust may accumulate in the plate transport mecha-
®
To ensure optimal performance of the ScanX (P/N B7350), the plate transport should be cleaned once per week using a new ScanX Cleaning Sheet each time.
Sample sheets are included with the ScanX and additional sheets can be purchased from your dealer. Refer to ACCESSORIES page 27.
Disinfecting the ScanX
No disinfection is required for the ScanX.
Cleaning the Plate Guides
Wash the Plate Guides with a non-abrasive cleaner and water. Dry thoroughly.
IMPORTANT: Plate Guides must not be put in an autoclave.
Note: Always use standard infection control procedures when handling devices that contact the patient.
Disinfecting the Plate Guides
There is no reason to routinely disinfect the Plate Guide unless contamination is suspected. If a Plate Guide has touched a contaminated surface, it may be immersed briefly in a cold sterilant according to sterilant manufacturers directions.
After disinfection, clean and dry the Plate Guide using the instructions above.
24
MAINTENANCE
Phosphor Storage Plates (PSPs).
PSP's are subject to "wear" on the black side during normal handling and use. They can appear scratched, while the sensitive side (blue or white) remains relatively
"smooth". This scratched look on the black side has absolutely no effect on the quality of the image and should be expected under normal conditions. If the phosphor side is scratched make sure the plates are being handled properly and not being dragged from the ScanX tray area or other surfaces that could cause scratching of the plate. Make sure to review the Plate Care and Preparation information provided on page 16 of this manual.
SCHEDULED MAINTENANCE
Like all precision products, the ScanX requires a certain amount of care on a regularly scheduled basis. A well-organized maintenance program aids dependable equipment operation and reduces problems to a minimum. Routine checks help to detect general overall wear, and replacement of parts can often be made before a problem occurs. Adherence to the maintenance schedule will ensure that the ScanX Digital Imaging System will continue performing at its best with uninterrupted service.
Understanding this, we have established two basic maintenance kits that will help insure continuous operation of the ScanX Digital Imaging System. The kits and their associated parts number along with the recommended performance schedule are listed below.
IMPORTANT: All service requiring access to the interior of the ScanX must be performed only by an authorized dealer service technician with the proper training.
Service Requirement
Replace transport drive belts and springs
Replace complete transport belt drive belt assembly
Schedule
2 years
4 years
Kit Part No.
B7920
B7434
25
TROUBLESHOOTING
Trouble
1. No power/
No green light on membrane switch
Possible Cause Corrective Action
• Not plugged in.
• No power at Mains
Outlet
• Defective power supply
• Check the line cord connection is firmly plugged in.
• Make sure outlet is grounded and has power.
• Call your authorized Air
Techniques dealer.
• Defective light or circuitry.
• Call your authorized Air
Techniques dealer.
2. Green, Blue or
Yellow indicator light does not work.
3. Image
Management
Software does not recognize the ScanX when selected.
• Inadequate Computer
System.
• The ScanX has not been turned on.
• Verify Computer System requirements (Page 10).
• The computer connection cable is loose or defective.
• The computer does not recognize that the ScanX is connected.
• ScanX hardware problem.
4. Plate does not scan properly.
• The PSP was not pushed far enough into the ScanX.
5. No image appears after scanning.
• The PSP fed backwards
(printed side towards
ScanX).
Note:
was disabled during scanning.
• Make sure that the READY switch is set to ON and the green indicator light is illuminated.
• Reconnect the cable. Check for tightness. Replace if necessary.
• Verify that the Setup program was correctly
(Page 14).
installed
• Call your authorized Air
Techniques dealer.
• Fully feed the PSP into the Plate
Guide.
• If Eraser mode was disabled during scanning, quickly refeed the plate with the printed side out.
• If Eraser mode was enabled during scanning, retake image.
Important:
Do not allow the PSP to be exposed to light between taking an X-ray and scanning with the ScanX.
• The PSP was erased prior to scanning.
• Hardware failure.
• X-ray source failed.
• Feed the PSPs into the scanner immediately and quickly after removal from the Plate Transfer
Box or the cassette.
• Call your authorized Air
Techniques dealer.
• Call your X-ray service dealer.
26
TROUBLESHOOTING
Trouble
6.
Image is too dark.
7 Intraoral image appears skewed on monitor.
8 Extraoral image appears skewed on monitor.
9 Image contains ghost images or shadows.
10 Image shows artifacts.
11 Intraoral PSP does not drop into the receiving tray
Possible Cause Corrective Action
• PSP has been overexposed • Use software to adjust brightness. If this is not possible, retake image with proper
(lower) exposure and a newly erased PSP.
• PSP was fed skewed either behind or without a Plate
Guide in place.
• Plate Guide used was not the proper size.
• Verify that Plate Guide is in place and PSP is fed into slot or if a
Size 4 PSP, it is fed in straight behind a Guide.
• Check PSP size and ensure that the correct Plate Guide is in place.
• PSP was fed skewed.
• When inserting PSP into feed slot, be sure to “feel” for resistance of light seal brush, align
PSP, and then push down uniformly on top edge of PSP.
• PSP was not completely erased prior to use.
• Imaging Plate was exposed with the back of the IP facing the tubehead.
• PSP has been stored too long in barrier envelope or cassette.
• Partial erasure of the image due to exposure to light during handling of the PSP.
• The PSP surface is not clean and has dirt, stains or scratches on it.
• ScanX plate transport path may contain debris.
• PSP may have clung to the bottom of transport arch due to static electricity.
• Make sure ScanX is operating with eraser turned on (blue LED indicator above the ERASER switch is illuminated steadily).
• Make sure the plates are inserted properly into the barrier envelope or cassette with the proper orientation to the X-ray source.
• Do not store PSPs in barrier envelopes or cassettes for more than 24 hours.
• Do not leave exposed PSPs in well lit areas. Even in barrier envelopes, some light may penetrate and partially erase the PSP. Transfer PSPs from their protective barriers to the ScanX within one hour of exposure.
• Clean the PSP. If the PSP is scratched or stained, do not reuse.
• Clean transport path using a
ScanX Cleaning Sheet.
• Look below transport arch and gently touch the PSP to allow it to drop.
27
ACCESSORIES
The following lists the ordering number and description for accessory components available to maintain the ScanX to meet your professional needs. Contact your authorized dealer for information.
Description
ScanX Cleaning Sheets* for B7150
ScanX Cleaning Sheets* for B7350
Quantity
Box of 12
Box of 25
Box of 12
Box of 25
1 ScanX Dust Cover Kit
Intraoral Plate Guides
Size 0
Size 1
Size 2
Size 3
Intraoral Phosphor Storage Plates
Size 0
Size 1
Size 2
Size 3
Size 4
Intraoral Barrier Envelopes
Size 0
Size 1
Size 2
Size 3
Size 4
Medical Phosphor Storage Plates
8 x 10 inches
10 x 12 inches
Plate Protectors
8 x 10 inches
10 x 12 inches
Cassettes
8 x 10 inches
10 x 12 inches
Plate Transfer Box
Power Supply
Line Cord
6-Ft Computer Connector Cable, USB
15-Ft Computer Connector Cable, USB
PSP Wipes
1
1
1
1
2 per package
2 per package
4 per package
2 per package
1 per package
100 per package
100 per package
300 per package
100 per package
50 per package
1 per package
1 per package
1 per package
1 per package
1 per package
1 per package
1
1
1
1
1
Box of 25
* One re-usable Release Liner is included in each box of ScanX Cleaning Sheets .
Part No.
B3010
B3020
B2010
B2020
74050
73566-0
73566-1
73566-2
73566-3
73445-0
73445-1
73445-2
73445-3
73445-4
73248-0
73248-1
73248-2
73248-3
73248-4
73578-8M
73578-10
B2160-8M
B2160-10
74190-8
74190-10
74570
B7095
73096
B3554
B5026
B8910
28
VISIX IMAGING SOFTWARE
Consider VISIX, our comprehensive digital imaging software application, providing seamless integration with our ScanX PSP digital imaging systems and Camera product family.
Key Features.
Easy to learn.
Simple to use.
Acquire, mount, view and store images with just a few mouse clicks.
Customize images to personal preferences with our simple Set-Up 1-2-3 tool.
Exposure - Our Stop Light Exposure System is a unique indicator to ensure optimum data capture. It helps you acquire great images by monitoring
X-ray exposure.
Viewing - VISIX provides automatic image mounting that is completely customizable.
Flexibility - VISIX includes a built-in bridging solution to over 35 Practice
Management Software titles.
VISIX is the superb software complement for ScanX. Contact your authorized dealer for further information.
29
NOTES
30
31
NOTES
ALLPRO Imaging is a leading manufacturer of imaging systems for the medical, veterinary and security markets. Thousands of
ALLPRO Imaging products have been sold globally, installed and serviced through our network of authorized dealers.
100 Plus
2010
Polaris
J
ScanX
®
12
ScanX
®
12 DVM
ScanX
®
Discover
ScanX
®
Duo
ScanX
®
Fit
ScanX
®
Quantum
ScanX
®
Scout
ScanX
®
Trek
Visix
J
1295 Walt Whitman Road, Melville, NY 11747-3062
Website: www.allproimaging.com
Toll Free: 1-800-247-8324
© Copyright ALLPRO Imaging, 2006 • P/N B7157, Rev. D - December 2010
advertisement
Key Features
- Digital Radiography
- Reusable Phosphor Storage Plates
- Inline Erase
- Intraoral & Extraoral Imaging
- Computer Storage & Retrieval
- High-Resolution Imaging
Frequently Answers and Questions
What kind of computer system do I need to use with the ScanX?
How do I clean the Phosphor Storage Plates (PSPs)?
How do I erase the PSPs?
What is the proper way to handle and store the PSPs?
Related manuals
advertisement
Table of contents
- 1 TOP
- 2 FOREWORD
- 2 TABLE OF CONTENTS
- 3 CONGRATULATIONS
- 3 WARRANTY
- 3 ON-LINE REGISTRATION
- 4 SAFETY NOTICE
- 7 IMPORTANT INFORMATION
- 8 PURPOSE OF THIS MANUAL
- 8 SYSTEM DESCRIPTION
- 9 UNPACKING AND INSPECTION
- 9 IF YOU NEED ASSISTANCE
- 10 COMPUTER SYSTEM REQUIREMENTS
- 11 CONTROLS AND INDICATORS
- 12 TECHNICAL DATA
- 13 ABBREVIATIONS
- 13 SITE SELECTION
- 14 SYSTEM SETUP
- 16 PLATE CARE & PREPARATION
- 19 INTRAORAL IMAGING PROCEDURES
- 22 EXTRAORAL IMAGING PROCEDURES
- 24 POWERING DOWN THE SYSTEM
- 24 MAINTENANCE
- 25 SCHEDULED MAINTENANCE
- 26 TROUBLESHOOTING
- 28 ACCESSORIES
- 29 VISIX IMAGING SOFTWARE