The Pfeiffer ACM 2000 is a multi-channel controller designed to work with various pressure gauges, including Pirani, combined Pirani/Cold cathode, combined Crystal/Hot cathode, and capacitive gauges. This controller provides power, control, and pressure display for one, two, or three gauges simultaneously. It can be controlled by an operator via the front panel or externally by a computer via the built-in RS232 link. The ACM 2000 can monitor pressure in your system, providing valuable insights for process control and vacuum management. This controller can help you ensure the integrity and reliability of your vacuum systems.
PDF
Download
Document
Advertisement
Advertisement
--Réalisation/Publication: Alcatel Vacuum Technology France - User’s Manual - Ed 03 - Date : 09/2010 - P/N : 113162 *113162* ACM 2000 Multi-channel Controller User’s Manual Alcatel Vacuum Technology, as part of the Alcatel-Lucent Group, has been supplying vacuum pumps, helium and hydrogen leak detection systems, plasma sensors, vacuum measurement for several years. Thanks to its complete range of products, the company has become an essential player in multiple applications : instrumentation, Research & Development, industry and semiconductors. Alcatel Vacuum Technology has launched Adixen, its new brand name, in recognition of the company’s international standing in vacuum position. With both ISO 9001 and 14001 certifications, the French company is an acknowlegded expert in service and support, and Adixen products have the highest quality and environmental standards. Gb 03420 - Edition 01 - August 2007 With 45 years of experience, AVT today has a worldwide presence, through its international network that includes a whole host of experienced subsidiaries, distributors and agents. The first step was the founding of Alcatel Vacuum Products (Hingham- MA) in the United States, thirty years ago, reinforced today by 2 others US subsidiaries in Fremont (CA) and Tempe (AZ). In Europe, AVTF-France headquarters and its subsidiaries, Alcatel Hochvakuumtechnik (Germany), Alcatel Vacuum Technology UK (Scotland), Alcatel Vacuum Technology Benelux (Netherlands), Alcatel Vacuum Systems (Italy) and more recently Adixen Sensistor AB in Sweden (in 2007) form the foundation for the European partner network. In Asia, our presence started in 1993 with Alcatel Vacuum Technology (Japan), and has been strengthened with Alcatel Vacuum Technology Korea (in 1995), Alcatel Vacuum Technology Taiwan (in 2001), Alcatel Vacuum Technology Singapore, Alcatel Vacuum Technology Shanghai (China) (in 2004) This organization is rounded off by more than 40 representatives based in a variety of continents. Thus, whatever the circumstances, the users of Adixen products can always rely on quick support of our specialists in Vacuum Technology. ACM 2000 Multi-channel controller Welcome Dear customer, You have just acquired an Adixen ACM 2000 controller. We thank you for your purchase and are proud to be able to count you among our customers. GB 03423 - Edition 03 - October 09 Alcatel Vacuum Technology has acquired a vast wealth of experience in the design of controllers over the years. To guarantee optimum performance and full satisfaction in using this equipment, please take the necessary time to become familiar with this manual, especially the section covering installation and commissioning, before installing or operating this controller. The ACM 2000 is a multi-channel controller designed to work with the following gauges: - PIRANI gauge: AP 2004 - APN 2004 - Combined PIRANI/Cold cathode gauge: ACC 2009 - Combined CRYSTAL/Hot cathode gauge: AHC 2010 - Capacitive gauges: ASD 200X, ARD 200X series Alcatel Vacuum Technology France - User's Manual ACM 2000 1/2 ACM 2000 Multi-channel controller Copyright/Intellectual property: The use of Adixen products are subject to copyright and intellectual property rights in force in any jurisdiction. All rights reserved, including copying this document in whole or any part without prior written authorization from Alcatel Vacuum Technology France. Specifications and information are subject to change without notice by Alcatel Vacuum Technology France. 2/2 Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03423 - Edition 03 - October 09 This product meets all the essential requirements contained in European directives and standards, as listed in the Declaration of Conformity shown in Appendix 9 of this manual. These Directives are amended by Directive 93/68/EEC (CE mark). General contents MANUAL REFERENCE: 113162 EDITION: 03 - September 2010 User’s Manual ACM 2000 Controller English original version Chapter A Commissioning - General safety recommendations - Unpacking and storing the controller - Installation - Quick commissioning procedure Chapter B Product description - Overview of multi-channel controller - Navigation the controls - Display modes - Configuring the start-up display - Configuring the graph display Chapter C Settings - Overview - Configuring the gauges - Set point management - General settings for the ACM 2000 Chapter D Test mode - Access to test mode - Analog/Digital converter test - Controller display test - Memory tests - Controller interface tests Chapter E RS232 mode GB 03422 - Edition 02 - September 07 - Introduction - List of commands - Detailed description of RS232 command Chapter F Appendices Appendix 1: Technical characteristics Appendix 2: Dimensions Appendix 3: Rack Installation Appendix 4: Cable and connector details Appendix 5: Factory settings Appendix 6: Error messages Appendix 7: Accessories Appendix 8: Conversion of measuring units Appendix 9: Declaration of conformity Appendix 10: Detailed example of navigation Appendix 11: Storing and Retrieving Pressure Data Alcatel Vacuum Technology France - User's Manual ACM 2000 1/1 A Commissioning ACM 2000 User’s Manual Detailed contents General safety recommendations 2/9 Unpacking and storing the controller 5/9 - Unpacking - Storing the controller - Decontamination - Product recycling Installation 6/9 - Assembly, installation - Table mounting - Rack mounting - Power connections Quick commissioning procedure 7/9 GB 03424 - Edition 03 - October 09 - Bringing the ACM 2000 on line - Start-up screen - Measurement screen and display - Warning or error message - Alternative displays - Shutting down Alcatel Vacuum Technology France - User's Manual ACM 2000 1/9 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices General safety rules This controller is designed to be completely safe to use. Nonetheless, there are certain risks that cannot be eliminated. Such risks are indicated throughout this manual by the following warning messages: Warning symbols used in this manual This type of warning is used to indicate a potential risk that can cause significant damage to equipment and/or installations if instructions are not followed. This type of warning is used to indicate a potential risk that can cause minor injury if instructions are not followed. This type of warning is used to indicate a potential risk that can cause serious injury or death if instructions are not followed. GB 03424 - Edition 03 - October 09 This type of warning is used to indicate an imminent risk that can cause serious injury or death if instructions are not followed. 2/9 Alcatel Vacuum Technology France - User's Manual ACM 2000 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices General safety rules Safety instructions for the installation Before switching on the power to this controller, the user must be familiar with this manual and must follow the safety instructions herein. • The controllers must be connected to an electrical installation fitted with a type TT earthed socket in compliance with ruling 88-1056 of 14th November, 1988 (for French regulation). • Our products are designed to meet all CEE regulations in effect. Any modifications by the user are liable to break compliance with regulations and jeopardise the EMC (Electromagnetic Compatibility) performance and product safety. The manufacturer renounces all responsibility for the consequences of any such intervention. The EMC performance of this product is obtained only if the installation is carried out in compliance with the applicable EMC rules. In particular, in environments liable to emit strong EM disturbances, you must - use shielded cables and connections for the interfaces - protect the power line by shielding from the power source to within 3 metres of the power input to the product. GB 03424 - Edition 03 - October 09 The enclosure is designed to ensure normal safety conditions in its normal operating environment (in a rack). When used on a table, or when handling the enclosure make sure no objects are introduced into the ventilation holes and that these holes are not blocked. When switching off equipment containing condensers charged at more than 60 VDC or 25 VAC, take all necessary precautions to protect against contact with the pins on the connectors (single phase motors, equipment with mains filter, frequency inverter, monitor, etc.) Alcatel Vacuum Technology France - User's Manual ACM 2000 3/9 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices General safety rules This product’s safety and performance can not be guaranteed unless it is used in compliance with this manual. This symbol on the controller body indicates that certain internal components are under voltage and can cause electric shock if touched. This symbol recommends that the controller be disconnected from the mains before any work is carried out on it or that the circuit breaker be correctly identified and locked. This symbol on the controller body indicates a potential risk that could cause significant damage to the equipment and/or installations if instructions are not followed. Responsibility and guarantee 4/9 Jobs described in this document may be carried out only by staff possessing the appropriate technical training and necessary experience or having been trained to this end by the product owner. The manufacturer’s responsibility is lifted and the guarantee voided if the owner or a third party • fails to observe the instructions in this document, • uses this product in a way that is not consistent with the manufacturer’s intended use, • modifies this product in any way whatsoever, • uses the product with accessories not listed in the product documentation. Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03424 - Edition 03 - October 09 Qualification of personnel A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Unpacking Unpacking and storing the controller Take care in unpacking the equipment and keep its all of the packaging. Make sure there is no shipping damage. If any such damage is noted, take the necessary measures with the carrier and notify the manufacturer if necessary. In all cases we advise you to keep the packaging (recyclable material) in case the controller needs to be shipped again, or must be stored for a long time. This product was manufactured in a clean environment and packaged carefully to maintain this cleanliness. We advise you not to unpack the controller until it is at its place of installation. Remove the equipment manually from its package (weight 2.1 kg). The electrical cables (power cable and gauge connection cables) are shipped in a separate package (see appendix 7). Storing the controller The controller can be stored in its original packaging at temperature of –20°C to +60°C. GB 03424 - Edition 03 - October 09 Decontamination and product recycling In compliance with Directive 2002/96/CE concerning the handling of electrical and electronic equipment waste and Directive 2002/9/CE concerning restrictions on hazardous substances, Adixen products that have reached the end of their service life must be returned to the manufacturer for decontamination and reuse. The manufacturer’s obligation to recover such equipment applies only to complete items of equipment that have been neither modified nor retrofitted and have used only spare parts from Alcatel Vacuum Technology sold by Alcatel Vacuum Technology and including all their assemblies and sub-assemblies. This obligation does not include the cost of transporting the product to a reprocessing centre, nor the cost of the service which will be invoiced on to the customer. For all equipment returned to an Alcatel service center, the safety questionnaire at the end of manual (or available online at www. adixen.com) must be filled in and sent to the service center in advance. Alcatel Vacuum Technology France - User's Manual ACM 2000 5/9 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Assembly, installation Installation The controller can be placed on a table or mounted in a 19" rack. Table mounting Ventilation holes Locate the appliance in a way that ensures good air circulation and avoids direct exposure to the sun. Rack mounting See appendix 3 Electric shock hazard. Before connecting the gauges, set the power switch to the «0» (off) position. Always connect the gauges before switching on power to the controller. The supply voltage to the unit must be between 100 and 240 VAC. Check the supply voltage and always used adequate cabling. Failure to follow these instructions can lead to electric shock. 6/9 Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03424 - Edition 03 - October 09 Controller fan Internal components can deteriorate through overheating if there is inadequate ventilation. DO not block the ventilation holes. A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Bringing the ACM2000 on line Quick commissioning procedure 1 - Place the ACM 2000 controller on a table 2 - Connect the vacuum gauges to the connections at the back of the controller. 3 - Connect the mains power cable Products which are not properly connected to ground are a danger to life in the event of failure. The controller frame must be connected to earth and an earth wire is provided for this purposes in the 3-pin power cable (PE). Always connect the power cable to a socket with a ground. 4 - Set the power switch to the «1» (ON) position. 4 2 GB 03424 - Edition 03 - October 09 3 1 Alcatel Vacuum Technology France - User's Manual ACM 2000 7/9 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Start-up screen Displaying the start-up screen Quick commissioning procedure The start up screen appears for 5 seconds and shows the controller type and the software version number. Type of controller 㨇X.XX ? Software version number Measurement board version After start-up, the ACM 2000 • performs automatic testing • identifies what gauges are connected to it • continues with the same parameters as it had when power was last turned off • switches to «Measuring» mode and to the sensor that was active before power was last turned off • adapts any parameters where necessary (if the gauges have changed) The measurement is available if: - The gauges are properly connected - The pressure lies within the measuring range of the connected gauge Measurement readout Channel No. 00:00:00 (*) Exponent Type of gauge _____ Name (User settable parameter) _____ AD _____ 0970 Unit Function bar (*) See chapter B/display description E.g., on channel "2", the "AHC" type gauge measures a pressure of 6.6 x 10-7 mbar. 8/9 Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03424 - Edition 03 - October 09 Measurement screen and display A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Warning or error messages Quick commissioning procedure The following warning message comes up when no gauge is connected. E.g.: no gauge connected to channel 3. _____ An error message is displayed when an error is detected - see Appendix 6 for the error message list. E.g.: Error 01 on channel 2. _ _AD ___ Alternative displays 2 different measuring screens can be displayed in addition to that shown above. The alternatives are: Graph display Analog display 00:00:00 (*) 00:00:00 (*) Pa CH1 CH1 C-ION CH2 PIR CH3 CCPIR 1E+ 0 -1 -7 - -5 2H 2L 1 1H -1 1 1L 2 1H 2 1L TIME/DIV - - -5 - CH3 - - 1H -3 -2 1 + +1 +2 1L +3 1H -3 -2 1 + +1 +2 +3 -7 - -5 - -3 -2 1 + +1 +2 +3 mbar 30min/DIV SCALE -7 - 1L 1H 2H 1L 2L -1 1E- 1 GB 03424 - Edition 03 - October 09 2H CH2 5E- 1 DISP - - L CK 2L DISP (*) See chapter B/display description You can cycle from screen to screen by pressing the button below DISP . Shutting down Switch off the ACM 2000 by setting the power switch to the «O» position (or via the distribution box if rack mounted). Before switching the ACM 2000 on again, wait at least 10 seconds to allow it to reset. Alcatel Vacuum Technology France - User's Manual ACM 2000 9/9 Product description B ACM 2000 User’s Manual Detailed contents Overview of the multi-channel controller 2/9 - Controller concept - Rear view of the controller - Control panel - Description of display Navigation controls 5/9 - Buttons - Viewing the selection - Function bar Display modes 7/9 - Digital readout - Graph display - Analog display 8/9 Configuring the graph display 9/9 GB 03425 - Edition 03 - October 09 Configuring the start-up display Alcatel Vacuum Technology France - User's Manual ACM 2000 1/9 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Controller concept Overview of the multi-channel controller The ACM 2000 provides power, control, and pressure display for one, two, or three gauges simultaniously. It can be controlled by an operator via the front panel, or externally by a computer via the builtin RS232 link. ACM 2000 Ver 3-X.XX GB 03425 - Edition 03 - October 09 㨇X.XX ? Technical and dimensional characteristics 2/9 Appendices 1 and 2 Alcatel Vacuum Technology France - User's Manual ACM 2000 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Overview of the multi-channel controller Rear view of the controller RS232 connector (see Section E and Appendix 4) Power indicator light Power switch REMOTE connectors (see appendix 4) Power connection Ground connection (if required) Connectors for gauges (see appendix 4) Rear face GB 03425 - Edition 03 - October 09 Control panel The ACM 2000 is operated from the front face of the enclosure which includes touch buttons and a display. Settings button Display ACM 2000 Ver 3-X.XX 㨇X.XX ? Measuring channel select button Navigation and function buttons Front face Alcatel Vacuum Technology France - User's Manual ACM 2000 3/9 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Overview of the multi-channel controller Description of display Channel Type of sensor (*) Name (user settable) Measured pressure or No Gauge message (if no gauge) or Error message: (Errxx) Pressure signal flashes if over range. ”Lock” displayed when settings are locked 00:00:00 (*) Exponent of pressure reading, e.g 10-2 _____ (*) AP ACC AHC AD = AP 2004 - APN 2004 = ACC 2009 = AHC 2010 = ASD 200X - ARD 200X Display of set point status for each channel _____ AD _____ 00:00:00 (*) 0970 Status of specific functions concerning the gauge connected to this channel Units: mbar, Pa, Torr Function bar (*)This data displays from software version: 3-1.10 [1.10] GB 03425 - Edition 03 - October 09 This is the screen shown when the ”Digital display” mode is set. See page 8, section B for all available display modes. 4/9 Alcatel Vacuum Technology France - User's Manual ACM 2000 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Navigation controls You can navigate through the controller settings using the buttons and the function bar. Only the buttons are described below. Refer to appendix 10 if you would like a detailed example of how to navigate. Buttons Access to PARAMETER SETTING MODE. Back to previous menu up to DISPLAY screen. 0,5 s Access to TEST MODE. 3s CHANNEL selection (measuring channel) when 1, 2, or 3 is displayed Button function varies and is defined on display function bar. These buttons are used to navigate the menus, select variables, change the variables and validate the selections or changes Examples of displays a CHANNEL selected a MENU selected _____ _____ 0970 GB 03425 - Edition 03 - October 09 a Parameter selected non-active menu item a parameter value selected See the ”display modes” paragraph on page 7 for a full preview of the displays. Alcatel Vacuum Technology France - User's Manual ACM 2000 5/9 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Navigation controls Function bar DIGITAL readout GRAPH display TIME/DIV DISP FUNCTIONS linked with the gauge for the selected channel Change DISPLAY screen 2-column MENU Confirm selected MENU item Select MENU item ANALOG display 30min/DIV SCALE Choose SCALE Change VALUE of scale Change DISPLAY screen DISP Change DISPLAY screen Single-column MENU or Modification of parameter value Confirm selected MENU OR Confirm modification of parameter VALUE Select MENU or PARAMETER OR Modify VALUE of parameter GB 03425 - Edition 03 - October 09 See the ”display modes” paragraph on page 7 for a full preview of the displays. 6/9 Alcatel Vacuum Technology France - User's Manual ACM 2000 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Digital readout Display modes By default, the measurement screen will be in digital readout mode. 00:00:00 (*) _____ _____ AD _____ 0970 You can change the display mode by pressing the ”DISP” button to switch to a graph display. Graph display 00:00:00 (*) mbar CH1 C-ION CH2 PIR CH3 CCPIR 1E+ 0 Graph of the pressure signals over time 5E- 1 1 1H 1 1L Set points shown 2 1H 2 1L 1E- 1 30min/DIV TIME/DIV SCALE GB 03425 - Edition 03 - October 09 DISP Display settings bar You can change the display mode by pressing the ”DISP” button to switch to an analog display. Analog display 00:00:00 (*) CH1 -1 - - L CK 2L 2H -7 - -5 - CH2 -1 - - -7 - 1L Set points shown 1H -3 -2 1 + +1 +2 2H 2L 1L -5 - -3 -2 1 + +3 1H +1 +2 +3 Bargraphs showing the measurements of the different channels 1H 2H 1L 2L CH3 -1 - - -7 - -5 - -3 -2 1 + +1 +2 +3 mbar DISP You can change the display mode by pressing the ”DISP” button to switch to digital readout. (*) See chapter B / display description Alcatel Vacuum Technology France - User's Manual ACM 2000 7/9 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Configuring the start-up display The power on display mode is factory set to the digital readout. This can be changed by the user. Modification of the start-up display: 00:00:00 (*) _____ _____ AD _____ 0970 (*) See chapter B / display description Access: SET/PARAMETER SETTING MODE/Other/Start-up Display Select The controller offers a choice of 3 display modes: Digital, Graph or Analog. Start-up Display Select Digital Graph Analog Digital The modification will take effect the next time the controller is turned on. 8/9 Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03425 - Edition 03 - October 09 (*) See chapter B / display description A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Configuring the graph display The function bar on the ”Graph” display allows the user to change the time scale and maximum and minimum pressures displayed. Adjustment of the time scale by the ”TIME DIV” buttons Adjustment of the pressure scale by the buttons: - MAX RNG: maximum value of pressure axis - MIN RNG: minimum value of pressure axis Choose which parameter to modify by pressing the ”scale” button until the corresponding function is shown. It will cycle through ”TIME DIV”, ”MAX RNG” and ”MIN RNG”. TIME DIV Time axis Time range displayed full scale 10 180 sec. s/DIV 1 min/DIV 18 min 10 min/DIV 180 min 30 min/DIV 540 min 1 h/DIV 18 hr In the event of a power failure, the controller can not memorize the measurements taken previously. Recording will begin again at zero when the controller is restarted. GB 03425 - Edition 03 - October 09 MAX RNG Unit Adjustment range Pa 1E+6~1E-7 Torr 1E+4~1E-9 Mbar 1E+4~1E-9 Unit Adjustment range Pa 1E+5~1E-8 Torr 1E+3~1E-10 Mbar 1E+3~1E-10 MIN RNG Alcatel Vacuum Technology France - User's Manual ACM 2000 9/9 C Settings ACM 2000 User’s Manual Detailed contents Overview 2/16 - Access to settings - List of available settings Configuring the gauges 4/16 General settings - Channel name - Filtering - “Gas” correction factor - Analog gauge output Specific settings - AP 2004 - APN 2004 Pirani gauge - ACC 2009 Combined PIRANI/Cold cathode gauge - AHC 2010 Combined CRYSTAL/Hot cathode gauge - ASD 200X/ARD 200X Capacitance gauge Set point management 11/16 - View by display type - Adjustment by gauge type General settings for the ACM 2000 13/16 GB 03426 - Edition 03 - October 09 - RS232 settings - Locking the settings - Settings for measurements - Restoring factory default settings - Available remote Inputs/Outputs Alcatel Vacuum Technology France - User's Manual ACM 2000 1/16 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Overview This mode gives access to the user settable functions for - the controller - the gauges - measurements. Set point , button to bring for configuring the set points Measurement for configuring the correction of measurements Digital Display for setting up the digital display Graph Display for setting up the graph display Other 2/16 On the measurement display screen, press the up the following screen: configuration of general settings Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03426 - Edition 03 - October 09 Access to settings A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Overview List of available settings Gauge settings Setting AP 2004 ACC 2009 AHC 2010, ASD 200X APN 2004 Combined ARD 200X Combined Pirani Capacitance Pirani/ Cold Crystal/ Gauge Cathode Hot Cathode Menu 1. Set points 1.1: Set points • • • • • • • • • • • • • • • • 1.2: Check gauge low level • 2. Measurement 2.1: Gas correction factor 2.2: Filtering 2-3: Full scale • • 3. Digital display 3-1: Digit 3-2: Unit Settings ACM2000 controller • Menu 4. Graph display GB 03426 - Edition 03 - October 09 4-1: Set point selection 4-2: Channel selection 5. Sensor Control • • NOT USED 6. Other 6-1: Baud-Rate speed 6-2: Lock setting buttons 6-3: Factory default settings 6-4: Input channel name 6-5: RS232 adjustment 6-6: Select start up display Alcatel Vacuum Technology France - User's Manual ACM 2000 • • • • • • 3/16 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Channel name Configuring the gauges General settings The channel name can be found only on the digital screen. It allows the channels to be given custom names. 1 AP ----- Channel number Type of gauge User selectable channel name Access to this setting Access : SET/PARAMETER SETTING MODE/Other/Channel Name Input Example: 1 Name : ABCDE AP ABCDE GB 03426 - Edition 03 - October 09 Factory default setting The space reserved for the name is empty ( _ _ _ _ _ ). 4/16 Alcatel Vacuum Technology France - User's Manual ACM 2000 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Filtering Configuring the gauges General settings This function smooths the measurement information with a certain frequency response time: Slow/Normal/Fast. This can help with unstable processes or fluctuating measurement signals. Normal filter p GB 03426 - Edition 03 - October 09 Choose the “standard” setting to obtain a good compromise between a fast signal response and a noisy signal, on both the display and the set point functions. t Slow filter Choose “slow” to remove noise from the displayed signal, and to eliminate premature tripping of set points caused by signal noise. The controller will then respond more slowly to variations in the measurement. p Fast filter p Choose “fast” if the read out and set points must respond quickly to measurement fluctuations. The controller will then be more sensitive to measurement disturbances. t t Access to this setting. Access: SET/PARAMETER SETTING MODE/Measurement/Filter Setting Factory default setting This function is set to “NORM”. Alcatel Vacuum Technology France - User's Manual ACM 2000 5/16 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Gas correction factor Configuring the gauges General settings This correction factor corrects the measurements from a gauge for gases other than nitrogen (N2). The correction factor has an influence on - the read out - the switching thresholds of the set points. This function is available for all gauges except capacitive ones. Access to this setting Access : SET/PARAMETER SETTING MODE/Measurement/GAS Setting Factory default setting The calibration factor is factory set to a value of 1. A 0-10 V analog output is available on the Remote connector for each measuring channel. Its value is identical to the analog signal delivered by the gauge. See the user manual supplied with the gauge for the voltage to pressure conversion functions and graphs. See Appendix 4: Wiring up the Remote gauge connector. GB 03426 - Edition 03 - October 09 Analog gauge output 6/16 Alcatel Vacuum Technology France - User's Manual ACM 2000 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Pirani gauge (AP 2004 - APN 2004) Configuring the gauges Specific settings No furthur settings for this gauge. Combined PIRANI/ Cold cathode gauge (ACC 2009) Switching function (UR control). Description This setting allows the set point contacts for this channel to be held open or closed when the gauge indicates “UR” (Under Range = pressure below the gauge’s measuring capabilities). There are several possible causes of this “UR” indication: - the pressure really is low - the cold cathode gauge is not primed - the gauge is faulty. ON setting Choose ON if you consider it abnormal for the gauge to have gone UR. Previously activated set points will then be desactivated (detection of UR status). OFF setting Choose OFF if you consider it normal for the gauge to have gone UR. The set point contacts will then remain closed. Access to this setting Access : SET/PARAMETER SETTING MODE/Set point/Under-Range Control GB 03426 - Edition 03 - October 09 Factory default setting The switching function is OFF by default. Alcatel Vacuum Technology France - User's Manual ACM 2000 7/16 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Configuring the gauges Specific settings Combined CRYSTAL/ Hot cathode gauge (AHC 2010) Degassing function Description Deposits on the electrode system of an AHC 2010 combined CRYSTAL/Hot Cathode Gauge can effect the measured values or make them unstable. Degassing involves heating the electron collector grid to 70°C by electronic bombardment, thus cleaning the electrode system. Degassing normally lasts 3 minutes but can be interrupted prematurely. The function is permitted for pressures < 5x10-3 mbar (pressure advised < 1x10-5).. 00:00:00 (*) _____ _____ AD _____ Activating the degassing function: –> via the controller 1 - Select the channel to which the gauge is connected by pressing the DEGAS 0970 button. HV-DSBL DEGAS 2 - Press the . button. While degassing is in progress (approx. 3 minutes), - “Degas” is displayed on the measuring screen for the channel in question - ignore the displayed pressure values which will be wrong during this operation. Hot cathode deactivation function 00:00:00 (*) 2 HV-DSB 0970 HV-DSBL DEGAS 8/16 This function deactivates the hot cathode gauge. Only pressure values measured with the CRYSTAL gauge appear. –> via the controller 1 - Select the channel to which the gauge is connected by pressing the 10 AD _____ Description Activing the HV-DSBL function: _____ _____ is active. button. 2 - Press the button. HV-DSB comes up on the screen. –> via the Remote connector (HV-DSBL wiring - see Appendix 4) To reactivate measurement, press again. Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03426 - Edition 03 - October 09 –> via the Remote connector (DEGAS wiring - see Appendix 4) Note: degassing is impossible when - degassing has just stopped after 3 minutes - the pressure is higher than 5 x 10-3 mbar A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Configuring the gauges Specific settings Capacitive gauge (ASD 200X-ARD 200X) Digit function Description This function shows or hides the 5th digit in the displayed pressure. Activing the DIGIT function. - Select the channel to which the gauge is connected by pressing the button. DIGIT - Press the AD _____ function. 09704 DIGIT ZERO DP-POS Factory default setting The switching function deactivated by default and the display set to 4 digits. Zero function Description GB 03426 - Edition 03 - October 09 This function allows you to assign a measured pressure value as a datum point. The values displayed will then be negative when the pressure is less than the “ZERO” value entered. Remark: this is only possible in the voltage range of –0.2V to +0.2V delivered on the gauge’s analog output. Activing the ZERO function: - Select the channel to which the gauge is connected by pressing the button. Press the appear: AD _____ ZERO button for 2 seconds and the following display will 0000 + 0 ZERO Zero is displayed DIGIT ZERO DP-POS Alcatel Vacuum Technology France - User's Manual ACM 2000 9/16 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Configuring the gauges Specific settings DP-POS function Description This function lets you modify the read out of the displayed pressure value by changing the position of the decimal point. The exponent changes to suit. Accessing this function - Select the channel to which the gauge is connected by pressing the button. - Press the DP-POS button. 00:00:00 (*) _____ _____ AD _____ HV-DSB 0970 + 3 0970 DIGIT ZERO DP-POS AD _____ 0970 + 2 0970 + 3 DIGIT ZERO DP-POS AD 0970 + 2 0970 + 1 _____ DIGIT ZERO DP-POS AD 0970 + 1 _____ Setting «Full scale» ZERO DP-POS Description This setting specifies the top value in the measuring range depending on the gauge used. Full Scale Select Adjusting this setting: CH Access : SET/PARAMER SETTING MODE/Measurement/Full Scale Select 1 Factory default setting: 1333 mbar 2 The «Full scale» value is factory set to the highest possible value (1333 mbar). 133.3 mbar 13.33 mbar 1.333 mbar 䂥 3 䂥 10/16 0.333 mbar 1333 mbar 0.133 mbar 䂯 E NTE R Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03426 - Edition 03 - October 09 DIGIT A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Set point management View by display type Depending on what type of display is selected, the set points are represented in different ways: Digital readout 00:00:00 (*) A white dot is displayed when the set point (SP) is at the ON position. _____ _____ AD _____ Graph display 0970 No dot The set points (SP1, SP2) are at the OFF position. 00:00:00 (*) mbar CH1 C-ION CH2 PIR CH3 CCPIR 1E+ 0 5E- 1 1 1H 1 1H 1 1L 2 1H 2 1L 1 1L 1E- 1 30min/DIV TIME/DIV SCALE DISP Analog display 00:00:00 (*) GB 03426 - Edition 03 - October 09 CH1 -1 - - L CK 2L 2H 1L -7 - -5 - -3 -2 1 + 2H 2L 1L CH2 2L 2H 1L 1H -1 - - -7 - -5 - CH3 -7 - -5 2H 2L -3 -2 1 + +1 1H +1 +2 +3 1H -3 -2 1 + +1 +2 +3 1H 2H 1L 2L -1 - - The set points are represented by dotted lines the same colour as their channel. They are displayed only if they are activated. The set point is ON when the threshold label is shown in red. The set point is OFF when the threshold label is shown in white. -7 - -5 - -3 -2 1 + +1 +2 +3 The high and low thresholds are shown on the bargraph as long as set points are activated in the Graph Display menu1. The set point is ON when the threshold label is shown in red. The set point is OFF when the threshold label is shown in white. mbar DISP 1L Alcatel Vacuum Technology France - User's Manual ACM 2000 11/16 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Set point management Adjustment by gauge type The set points, 2 per channel, are used to control external devices using the “remote” connector. The outputs on the remote connector depend on the setpoint levels programmed and the pressure signal. Each set point has 2 thresholds: - 1 low threshold: SP-L - 1 high threshold: SP-H The set point opens as the pressure signal goes below the low threshold and remains open until the pressure signal crosses the high threshold. SP- H SP- L Time Set Point output Closed Open Closed Open Accessing set point adjustment: Access: SET/PARAMETER SETTING MODE/Set-point select/Graph Display AP 2004 Pirani APN 2004 Pirani AHC 2010 Crystal Hot cathode ACC2009 Cold cathode ASD 200x capacitive ARD 200x capacitive 12/16 Pa 1.00E+5 Torr 750 mbar Pa Initial setting 5.00E-2 5.00E-2 3.75E-4 3.75E-4 1000 5.00E-4 5.00E-4 1.00E+5 5.00E-8 5.00E-8 ≥(SP-H)≥(SP-L)≥ ≥(SP-H)≥(SP-L)≥ Torr 750 3.75E-10 3.75E-10 mbar 1000 5.00E-10 5.00E-10 Pa 1.00E+5 1.00E-7 1.00E-7 Torr 750 7.50E-10 7.50E-10 mbar 1000 1.00E-9 1.00E-9 Pa FSx100 (FS×100)/1000 (FS×100)/1000 Torr FSx0.75 (0.75xFS)/1000 (0.75xFS)/1000 mbar FS FS/1000 FS/1000 ≥(SP-H)≥(SP-L)≥ ≥(SP-H)≥(SP-L)≥ Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03426 - Edition 03 - October 09 Pressure adjustment range for set points and initial setting for each type of gauge Gauge Units min setting max setting ≥(SP-H)≥(SP-L)≥ A B C D E F General settings for the ACM 2000 controller Commissioning Product description Settings Test mode RS232 mode Appendices RS232 settings Communication speed Description This setting is used to adjust the communication speed of the RS232 serial interface. The speed is expressed in baud units. Accessing this setting: Access : SET/PARAMETER SETTING MODE/Other/Baud-Rate Setting Modification of this setting does not require the controller to be restarted. Factory default setting: The communication speed is factory set to 9600 bauds. RS-232 Setting Selecting the end of line characters for the RS232 transmissions CR (carriage return) = return to start of line only (with a risk of overwriting the previous data sent) CR LF (carriage return + line Feed) = advance to start of next line (with no chance of overwriting the previously sent data). Accessing this setting Access : SET/PARAMETER SETTING MODE/Other/RS232C Setting Factory default setting: GB 03426 - Edition 03 - October 09 The controller is factory set to the CR position. Locking the settings Description This setting stops anyone modifying the configuration. LOCK is displayed when the LOCK setting is ON. “Err05” will appear anytime anyone tries to modify a setting. See appendix 6: error messages). To unlock the settings, set parameter lock to "OFF". Accessing this setting: Access : SET/PARAMETER SETTING MODE/Other/Parameter Lock Setting Modification of this setting does not require the controller to be restarted. Factory default setting: By default the controller is not locked: LOCK is factory set to OFF. Alcatel Vacuum Technology France - User's Manual ACM 2000 13/16 A B C D E F General settings for the ACM 2000 controller Commissioning Product description Settings Test mode RS232 mode Appendices Settings for measurements Units Description This setting lets you choose between the following units for displaying pressure: - mbar - Torr - Pascal The pressure units are shown in the bottom right of the screen. mbar Accessing this setting Access : SET/PARAMETER SETTING MODE/Digital Display/Unit Setting Changing this setting does not require the ACM2000 controller to be restarted. Factory default setting The default pressure unit is factory set to mbar. Digit Description This setting lets you chose between 2 or 3 digit display for values. This setting concerns all gauges except the capacitive gauge. Access : SET/PARAMETER SETTING MODE/Digital Display/Digit Setting Changing this setting does not require the ACM2000 controller to be restarted. Factory default setting The measured pressure display is factory set to 2 digits. 14/16 Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03426 - Edition 03 - October 09 Accessing this setting: A B C D E F General settings for the ACM 2000 controller Commissioning Product description Settings Test mode RS232 mode Appendices Restoring factory settings Description This setting causes the ACM 2000 controller to reset all settings to their default values (factory default setting). Accessing this setting: Access : SET/PARAMETER SETTING MODE/Other/Factory Setting - Select ON and press . Factory Setting OFF ON Setting 䂥 䂯 ENTER - the screen automatically switches to the OTHER menu. Factory default settings. See appendix 5. GB 03426 - Edition 03 - October 09 Available Remote Inputs/Output Analog recording output A 0-10 V no configurable analog output is available at each Remote connector (pins 2 & 3) corresponding to that measuring channel. Its value is independent of the type of gauge connected to the channel. The signal can be used by applying the following formula: output (V) = 1/2 (Log Pdisp + 12) or Pdisp = 10(2 x output voltage(V) – 12) (Pdisp : value of displayed pressure) See Appendix 4 for wiring of the Remote connector. A voltage of 10V, 9.5V, or 0.5V is considered an error. If the internal fan stops, the output recording displays 0.5V. See list of error messages in Appendix 6. Alcatel Vacuum Technology France - User's Manual ACM 2000 15/16 A B C D E F General settings for the ACM 2000 controller Commissioning Product description Settings Test mode RS232 mode Appendices Available Remote Inputs/Outputs (continued) Remote control The connected gauge can be controlled from the Remote connector. The criteria controlled depend on the gauge type. Table of remote control functions Gauge Gauge input “function1” Gauge input “function2” Pirani AP 2004 - APN 2004 not used not used CRYSTAL/hot cathode AHC2010 HV-DSBL DEGAS Cold cathode/Pirani ACC2009 not used not used Capacitive ASD200X/ARD200X no used not used GB 03426 - Edition 03 - October 09 See Appendix 4 for wiring of the Remote connector. 16/16 Alcatel Vacuum Technology France - User's Manual ACM 2000 Test mode D ACM 2000 User’s Manual Detailed contents Access to test mode 2/6 Analog/Digital converter test 3/6 - Gauge input voltage - Gauge identification resistance Controller display test 4/6 Memory tests 4/6 - RAM and EEPROM tests Controller interface tests 5/6 GB 03427 - Edition 02 - September 07 - Verification of recorder output voltage - Input/output relay switching test - Testing the front panel buttons - Testing the RS232 link Alcatel Vacuum Technology France - User's Manual ACM 2000 1/6 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Access to test mode You can access the “test mode” by pressing the least 3 seconds. button for at 3s The “TEST MODE” screen comes up. Test controller display Test Analog/Digital converter (A/D) Test memories Test controller interface (Button, Input/Output, RS232) buttons to select which test to perform and press the method). button to confirm your choice (see navigation GB 03427 - Edition 02 - September 07 Use the 2/6 Alcatel Vacuum Technology France - User's Manual ACM 2000 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Gauge input voltage Analog/Digital converter test Description Use this test to check the input voltage from the gauges. Accessing the menu Access : SET/TEST MODE/A/D Test/Gauge Input - Press button to confirm. - The controller displays the input voltage for the connected gauges. When there is no gauge connected, the controller displays [-00.000] on the channel in question. Leaving this menu Press the Gauge identification resistance button to get back to the previous screen. Description Use this test to check the identification resistance of the gauges. Accessing the menu Access : SET/TEST MODE/A/D Test/ID Resistance button to confirm. - Press - The controller displays the value of the identification resistance. This value depends on the type of gauge (see appendix 4). GB 03427 - Edition 02 - September 07 Leaving this menu Press the button to get back to the previous screen. Alcatel Vacuum Technology France - User's Manual ACM 2000 3/6 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Controller display test Description Use this test to visually check the display quality. A blank screen is presented when this test is activated. Any black squares correspond to dead pixels. Accessing this test Access : SET/TEST MODE/DISPLAY TEST - Press button to confirm. Leaving this test Press the button to get back to the previous screen. Memory tests RAM and EEPROM tests Description Use this test to check: - “Live” memory (RAM) - “Program” memory (EEPROM) Access : SET/TEST MODE/Memory Test/RAM test Access : SET/TEST MODE/Memory Test/EEPROM Test Press to confirm. The program starts up automatically and the screen displays “RUN” throughout the test. “PASS” is displayed to indicated the test has been passed. “Err” is displayed to indicate the test has failed. Leaving this test Press the 4/6 button to get back to the previous screen. Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03427 - Edition 02 - September 07 Accessing these test A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Controller interface tests Description Use these tests to check - the output voltage of each channel for recording - the communication functions of the input/output relays for each channel. - proper operation of the front panel buttons - the RS232 protocol. Verification of recorder output voltage Accessing this test Access : SET/TEST MODE/Interface Test/D/A Test Connect a voltage measuring device to the recorder output. Use the button to position the cursor on the desired channel - Press button to confirm.. The voltage displayed is 10.0 (Volts) (max value). - Use the 0.1V). buttons to select a different output voltage (10V to Check that the measured voltage is the same as that displayed on the controller. Leaving this test Press GB 03427 - Edition 02 - September 07 Input/output relay switching test to get back to the previous screen. Accessing this test Access : SET/TEST MODE/Interface Test/I/O Test The relays can be switched manually, independently of the pressure. It is therefore a good idea to disconnect the cables attached to the Remote connectors. -- Use the press button to position the cursor on the desired channel and button to confirm. As soon as you press the position. button, the SP1 relay switches to the “ON” buttons to change the output relay states and use Use the an appropriate instrument to check switching of the relays (ON/OFF). Leaving this test Press to get back to the previous screen. Alcatel Vacuum Technology France - User's Manual ACM 2000 5/6 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Testing the front panel buttons Controller interface tests Accessing this test Access : SET/TEST MODE/Interface Test/Key Test - Press button to confirm. Note: the test becomes active as soon as you press the button. The screen displays the name of the “F4” button. Press the various front panel buttons and check that the corresponding names come up on the screen. Ex: —> Screen display «CH» Leaving this test To leave this test and get back to the previous screen, press and hold down the Testing the RS232 link button. Use this test to check the data transfer to and from a computer. Accessing this test - Press button to confirm. The screen displays “COMM” while data is being transmitted from the computer to the controller. The transmitted character is resent from the controller to the computer. Leaving this test Press the 6/6 button to get back to the previous screen. Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03427 - Edition 02 - September 07 Access : SET/TEST MODE/Interface Test/RS232C Test E RS232 mode ACM 2000 User’s Manual Detailed contents Introduction 2/9 - RS232 interface - Location of the RS232 interface - Configuring the RS232 protocol - Symbols used - Transmission syntax List of commands 3/9 - Interpreting replies - List of commands 4/9 GB 03428 - Edition 03 - October 09 Detailed description of RS232 commands Alcatel Vacuum Technology France - User's Manual ACM 2000 1/9 Commissioning Product description Settings Test mode RS232 mode Appendices Introduction RS232 interface Location of the RS232 interface Configuration of RS232 protocol Symbols used Transmission syntax The RS232 interface is used to operate the controller from a computer via an RS232 cable link (see appendix 4) The male sub-D 9 pin connector is located on the rear panel of the controller (see chapter B “Overview of multi-channel controller”). See the “General settings for the ACM 2000 controller” paragraph in the “Configuration” chapter. Data is transmitted in the following format: - 9600 bauds - 8 data bits - 1 stop bit - no parity bit. The following symbols are used in this chapter: < > Definition [ ] Option The transmission syntax is as follows: <STX> <COMMAND> [,PARAMETER1] … [, PARAMETERn] <EOM> Control characters (Start of text) (always ‘$’) Reply Used when parameters are required, see “Detailed description of RS232 commands: Syntax of commands for setting parameters”, pages 4 to 9. <STX> <DATA> <EOM> [LF] Data, see “Detailed description of RS232 commands: Queries”, pages 4 to 9. 2/9 End of message characters (CR: Carriage Return) Added if configured at the controller). Carriage return with line feed (LF) Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03428 - Edition 03 - October 09 A B C D E F A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices List of commands Interpretation of replies Reply Statuts Meaning OK No error data received correctly ERR_00001 Operation error unauthorised request ERR_00010 Command error command wording is incorrect ERR_00100 Parameter error parameter value is incorrect ERR_01000 Transmission error a transmission error has been encountered ERR_10000 Hardware error The controller hardware is faulty List of commands GB 03428 - Edition 03 - October 09 Command Function Query Set Type BAU Set transmission speed Controller CON Request continuous reception of pressure value Pressure CPF Set confirmation of power cut Controller DGS Request degassing function for AHC gauges DGT Set number of digits displayed Channel Controller AHC ERR Status of current errors FDS Disable filament of AHC gauges FLT Set filtering Channel FSR Set full scale for AD gauges Channel GAS Adjust gas correction factor Channel LOC Lock settings Controller OFS Request ‘zero’ function for AD gauges Channel PRA Query pressure value for all channels Pressure PRD Query pressure value for a channel Pressure AHC SP1 Adjust set point ‘SP1’ Channel SP2 Adjust set point ‘SP2’ Channel SPS Query set points status Channel TAD Test Analogique/Digital converters Test TDI Test display Test TEE Test EEPROM Test TID Query type of gauge connected TIO Test set point relays Test TKY Test panel buttons Test TRA Test RAM Test Channel TRS Test RS232 link UNI Set pressure measurement units Controller VER Query controller software version Controller Alcatel Vacuum Technology France - User's Manual ACM 2000 Test 3/9 Detailed description of RS232 commands Commissioning Product description Settings Test mode RS232 mode Appendices BAU Syntax Reply Description Set Command Query A B C D E F $BAU,? $a Sets or quieries RS232 Transmission speed a= transmission speed 0 -> 9600 bauds $BAU,a CON $CON,a - 1 -> 19200 bauds 2 -> 38400 bauds $b,c,d,e,f,g Request continuous reception of pressure value a = reception interval 0 -> 100 ms 1-> 1s 2 -> 1min b, d, f= status of gauge (b for channel 1, d for channel 2, f for channel 3) 0 -> OK A -> OK with ‘FAN WARNING’ alarm 1-> Ur B -> Ur with ‘FAN WARNING’ alarm 2 -> Or C -> Or with ‘FAN WARNING’ alarm 3 -> Err 03 or Err 04 or Err C# or FAN STOP 4 -> not used 5 -> No Gauge 6 -> Id Err 7 -> Err Hi or Err Lo or Err 06 or Err 07 c, e, g= measured pressure value (c for channel 1, e for channel 2, g for channel 3) for gauges AP, ACC, AHC: x.xxE ± xx for gauge AD: ± x.xxxxE ± xx Press any button to stop transmission. CPF $CPF,? $a Set confirmation of power cut a= memorise power supply problem 0 -> OFF DGS $CPF,a $OK $DGS,a,? $a 1 -> ON Request degassing function for AHC gauges a= channel 1, 2, or 3 $DGS,a,b $OK b= degassing status for AHC gauge 0 -> OFF 1 -> ON 4/9 Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03428 - Edition 03 - October 09 no gauge: 0.00E+00 Detailed description of RS232 commands Commissioning Product description Settings Test mode RS232 mode Appendices DGT Syntax Reply Description Set Command Query A B C D E F $DGT,? $b $DGT,a $OK $ERR $ERR_a Set number of digits displayed a= number of digits 2 or 3 ERR Status of current errors a= error code meaning 00000 -> no error 00001 -> operation error 00010 -> command error 00100 -> parameter error 01000 -> syntax error 10000 -> hardware error FDS $FDS,a,? $b Disable AHC filament a = channel 1, 2 or 3 $FDS,a,b $OK b = status of disable function 0 -> OFF (not disabled) 1 -> ON (disabled) FLT $FLT,? $a Set filtering a= channel 1, 2, or 3 $FLT,a,b $OK b= filter 0 -> SLOW 1-> NORMAL GB 03428 - Edition 03 - October 09 2 -> FAST FSR $FSR,a,? $b $FSR,a,b $OK Set full scale for AD gauges a= channel 1, 2, or 3 b= full scale for AD gauge 0 -> 1333 mbar 1 -> 133.3 mbar 2 -> 13.33 mbar 3 -> not used 4 -> 1.333 mbar 5 -> not used 6 -> (0.333 mbar) 7 -> not used 8 -> 0.133 mbar Alcatel Vacuum Technology France - User's Manual ACM 2000 5/9 Detailed description of RS232 commands Commissioning Product description Settings Test mode RS232 mode Appendices GAS Syntax Reply Description Set Command Query A B C D E F $GAS,a,? $b Adjust gas correction factor a= channel 1, 2, or 3 LOC $GAS,a,b $OK $LOC,? $a b= adjustment of value to suit gas, from 0.10 to 9.99 Lock settings a= status of settings lock 0 -> OFF OFS $LOC,a $OK $OFS,a,? $b $OFS,a,b $OK 1 -> ON Request ‘zero’ function for AD gauges a= channel 1, 2, or 3 b= zero status of AD gauge 0 -> OFF 1 -> ON PRA $PRA $a,b,c,d,e,f Query pressure value for a channel a= channel 1, 2, or 3 b= gauge status 0 -> OK A -> OK with ‘FAN WARNING’ alarm 1-> Ur B -> Ur with ‘FAN WARNING’ alarm 2 -> Or C -> Or with ‘FAN WARNING’ alarm 4 -> not used 5 -> No Gauge 6 -> Id Err 7 -> Err Hi or Err Lo or Err 06 or Err 07 c= measured pressure value for gauges AP, ACC, AHC: x.xxE ± xx for gauge AD: ± x.xxxxE ± xx no gauge: 0.00E+00 6/9 Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03428 - Edition 03 - October 09 3 -> Err 03 or Err 04 or Err C# or FAN STOP Detailed description of RS232 commands Commissioning Product description Settings Test mode RS232 mode Appendices PRD Syntax Reply Description Set Command Query A B C D E F $PRD,a $b,c Query pressure value for a channel a= channel 1, 2, or 3 b= gauge status 0 -> OK A -> OK with ‘FAN WARNING’ alarm 1-> Ur B -> Ur with ‘FAN WARNING’ alarm 2 -> Or C -> Or with ‘FAN WARNING’ alarm 3 -> Err 03 or Err 04 or Err C# or FAN STOP 4 -> not used 5 -> No Gauge 6 -> Id Err 7 -> Err Hi or Err Lo or Err 06 or Err 07 c= measured pressure value for gauges AP, ACC, AHC: x.xxE ± xx for gauge AD: ± x.xxxxE ± xx no gauge: 0.00E+00 SP1 $SP1,a,? $b,c Adjust set point ‘SP1’ a= channel 1, 2, or 3 b= set point low threshold SP L GB 03428 - Edition 03 - October 09 c= set point high threshold SP H SP2 $SP1,a,b,c $OK $SP2,a,? $b,c x.xxE ± xx SP H >= SP L Adjust set point ‘SP2’ a= channel 1, 2, or 3 $SP2,a,b,c $OK b= set point low threshold SP2L c= set point high threshold SP2H x.xxE ± xx SP2H >= SP2L Alcatel Vacuum Technology France - User's Manual ACM 2000 7/9 Detailed description of RS232 commands Commissioning Product description Settings Test mode RS232 mode Appendices SPS Syntax Reply Description Set Command Query A B C D E F $SPS,a $b,c Query set points status a= channel 1, 2, or 3 b= status of set point SP1 c= status of set point SP2 0 -> OFF 1 -> ON TAD $TAD,a $b,c Test Analogique/Digital converters a= channel 1, 2, or 3 b= value of analogue signal delivered by the gauge ± xx.xxxV -00.000V if no gauge is connected c= value of the gauge identification resistance xx.xE+3 ohms 00.0E+0 ohms if no gauge is connected TDI TEE $TDI,? $a $TDI,a $OK $TEE $a Test display a= screen test status 0 -> OFF 1 -> ON Test EEPROM a= result of EEPROM test PASS -> OK ERR_ -> error $TID,a $b Query type of gauge connected a= channel 1, 2, or 3 b= gauge connected AP, ACC, AHC, AD, or NoGau TIO $TIO,a,? $TIO,a,b $b $OK Test set point relays a= channel 1, 2, or 3 b= state of set point relays 0 -> initial state 8/9 1 -> set points and 2 OFF 2 -> set points and 2 ON Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03428 - Edition 03 - October 09 TID Detailed description of RS232 commands Commissioning Product description Settings Test mode RS232 mode Appendices TKY Syntax Reply Description Set Command Query A B C D E F $TKY $a Test panel buttons a= status of command buttons x***** -> F1 button *x**** -> F2 button **x*** -> F3 button ***x** -> F4 button ****x* -> CH button *****x -> SET button 0 for OFF, TRA $TRA $a for ON Test RAM a= result of RAM test PASS -> OK ERR_ -> error TRS $TRS $OK UNI $UNI,? $a $UNI,a $OK Test RS232 link End of test when ‘ETX’ ( ‘Ctrl C’ ) is transmitted. Set pressure measurement units = pressure measurement units 0 -> Pa 1 -> Torr 2 -> mbar VER $VER $a Query controller software version a= software version GB 03428 - Edition 03 - October 09 3-x.xx Alcatel Vacuum Technology France - User's Manual ACM 2000 9/9 GB 03429 - Edition 03 - July 2010 F Appendices ACM 2000 User’s Manual Detailed contents Appendix 1: Technical characteristics 2/22 Appendix 2: Dimensions 4/22 Appendix 3: Rack Installation 5/22 Appendix 4: Cable and connector details 6/22 Appendix 5: Factory settings 9/22 Appendix 6: Error messages 10/22 Appendix 7: Accessories 12/22 Appendix 8: Conversion of measuring units 13/22 Appendix 9: Declaration of conformity 14/22 Appendix 10: Detailed example of navigation 16/22 Appendix 11: Storing and Retrieving Pressure Data 17/22 Alcatel Vacuum Technology France - User's Manual ACM 2000 1/22 Commissioning Product description Settings Test mode RS232 mode Appendices General specifications Environment Gauges connected Sockets Commands Display Measurement Appendix 1 Technical characteristics Voltage 100 - 240 VAC Frequency 50/60 Hz Power requirement < 100 VA Weight 2.1 kg Overvoltage category II Protection class I Storage temperature –20 ...+60 °C Running temperature +5 ... +50 °C Relative humidity < 70 % Use Inside use only, max altitude 2000 metres Pollution level II Protection IP 30 Number 3 Usable types Pirani: AP 2004 - APN 2004 CRYSTAL/Hot cathode: AHC2010 Pirani / Cold cathode: ACC2009 Capacitive: ASD200X/ARD200X Gauges D-Sub 9P (female) x3 Remote D-Sub 5P (female) x3 RS232 D-Sub 9P (male) Front panel buttons 6 buttons Rear panel connectors see “Overview of multi-channel controller” paragraph in this manual Screen LCD TFT colour Refresh time 100 ms Measuring range specific to gauge Measurement error amplifier: 0.02 % FS Filter time constant offset : 0.05 % FS slow: t = 750 ms (fq = 0.2 Hz) normal: t = 150 ms (fq = 1 Hz) fast: Gauge power supply Gauge commands t = 20 ms (fq = 8 Hz) Pressure mbar, Pa, or Torr Zero function Capacitive gauges only Voltage 24 V CC ± 5 % Max power requirement (per channel) 13 W Current protection 900 mA by resettable fuse Number 2 (on gauge connector) ON/OFF signal ON: +24 V, OFF: 0 V continued on page 3/5 2/22 Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03429 - Edition 03 - July 2010 A B C D E F A B C D E F Appendix 1 Technical characteristics Commissioning Product description Settings Test mode RS232 mode Appendices Remote control power supply Voltage /Max current 24 V DC ± 5 % ; 1 00 mA (on Remote connector) Current protection 300 mA by resettable fuse Set Point output Number 2 (on REMOTE connector) Type of contact Dry contact Allowable load 125 V AC, 0,3 A / 30 V DC, 1 A Mechanical service life 5 x 107 switching cycles Electrical service life 1 x 105 switching cycles (at max load) Number 2 (on REMOTE connector) Photocoupler input 24 V DC Number 1 (on REMOTE connector) Voltage range - 0,5…10.5 V Accuracy ± 50 mV (difference from displayed value) Impedance 200 Ω Number 1 (on REMOTE connector) Voltage range 0…10 V Precision ± 20 mV Impedance 200 Ω Standard RS232 (RS232 connector) Protocol 1 start bit, 8 data bits, 1 stop bit, no parity bit RS232 TDX, RXD and GND Speed 9600 / 19200 / 38400 Bauds CE mark Low voltage directive: EN61010-1 Remote control input Analog output Recording output Interface (RS232) Standard EN61326 (Class A emission) GB 03429 - Edition 02 - July 2010 EMC directive: Alcatel Vacuum Technology France - User's Manual ACM 2000 3/22 Commissioning Product description Settings Test mode RS232 mode Appendices Appendix 2 Dimensions 103 mm (4.05 inch) inch) 110 mm (4.33 inch) 212 .34 mm (8 128,5 mm (5.05 inch) 4 x Ø 3,5 m m 91,4 mm (3.59 inc h) 106,3 mm Ø 0.137 inch (4.18 inc h) Front panel thickness: 2,5 mm 4/22 Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03429 - Edition 03 - July 2010 A B C D E F A B C D E F Appendix 3 Rack Installation Commissioning Product description Settings Test mode RS232 mode Appendices To optimize space the ACM 2000 controller can be rack mounted. When doing so we recommend you to supply the power through a breaker panel. (4.82 inch) 122.5 mm 112 mm (4.4 inch) 105 mm (4.13 inch) 91,4 mm (3.59 inch) GB 03429 - Edition 02 - July 2010 4 M3 holes Ø 0,137 inch If the controller is rack mounted, the controller temperature will rise and this could cause damage. To avoid this leave 20 mm free space around the controller. In addition, a fan must be installed in the rack to avoid the ambient temperature rising above 50°C. Alcatel Vacuum Technology France - User's Manual ACM 2000 5/22 Commissioning Product description Settings Test mode RS232 mode Appendices Appendix 4 Cable and connector details This controller has dry contact outputs, it is the customer’s responsibility to use these outputs in compliance with all appropriate safety standards. This chapter gives details of the specifications for each connector and the various cable connections. «CH» connector and gauge cables Before connecting or disconnecting the gauge cables, always turn the controller power supply off first to avoid causing damage to the equipment. ACM2000 CH connector D-sub9P (Female) Gauge cable Gauge cable D-sub9P (Male) 1 6 1 7 3 2 3 5 9 8 4 9 5 2 7 7 8 4 1 6 6 2 7 8 D-sub9P (Male) 1 6 2 9 Gauge Gauge cable D-sub9P (Female) 3 4 5 3 8 9 4 5 N.B. the “CH” connectors are wired up identically for each measuring channel. Table of gauge identification resistance values Pin No. Signal name Description 1 Measured signal input (+) Analog signal (+) 0–10 V delivered by the gauge 2 Not used Not used 3 Gauge “function” output “HV-DSBL” function for AHC2010 gauge 4 Power supply output (+) 24V DC (+) for power supply to gauge 5 Identification input Gauge “function 2” output Identification de la gauge (see table below) “DEGAS” function for AHC2010 gauge 6 Not used Not used 7 Not used Not used 8 Measured signal input (-) Analog signal (-) 0-10 V delivered by the gauge 9 Power supply output 24 V DC (0V) for power supply to gauge Gauge Resistance value Pirani AP 2004 - APN 2004 13 KΩ CRYSTAL/cathode chaude AHC2010 4,7 KΩ Cathode froide/Pirani ACC2009 15 KΩ (CC mode) 18 KΩ (Pirani mode) Capacitive ASD200X/ARD200X 6/22 7,5 KΩ Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03429 - Edition 03 - July 2010 A B C D E F A B C D E F Appendix 4 Cable and connector details Commissioning Product description Settings Test mode RS232 mode Appendices “Remote” connector ACM2000 “Remote” connector D-sub15P (Female) 1 9 +24VDC PS2911-1(K) 2 10 3 22 k 11 PS2911-1(K) 4 22 k 12 5 13 SP2 14 SP1 15 6 7 8 GB 03429 - Edition 02 - July 2010 N.B. the “Remote” connectors are wired up identically for each measuring channel. Pin No. Signal name Description 1 Measured signal output (+) Analog signal (+) 0–10 volts delivered by the gauge 2 Output signal (-) 0V of measured signal and recording signal 3 Recording signal output (+) Analog signal (+) 0–10 volts delivered by the controller 4 Gauge “function 1” (+) input 24 VDC to activate the control 5 Gauge “function 1” (-) input 0 V to deactivate the control 6 Set point (NO) 7 Set point common 8 Set point (NC) 9 24 V DC power supply output (+) 10 24 V DC power supply output (-) 11 Input (+) “function 2” gauge 24 VDC to activate the control 12 Input (–) “function 2” gauge 0 V to deactivate the control 13 Set point 14 Set point 2 common 15 Set point 2 (NC) Set point dry contact 24 VDC power supply, max 100 mA 2 (NO) Set point 2 dry contact NO : Normally open NC: Normally closed The 24 V DC power supply output is limited to 100 mA. Moreover, it is not isolated from the internal circuit. When the power supply source for the controller is used for the “Remote” or “Set-Point” control, isolation of the photocoupler and relays is no longer guaranteed. To guarantee complete isolation, therefore, use an external power supply. Before connecting or disconnecting the «Remote» cable, always turn the controller power supply off first to avoid causing damage to the equipment. Alcatel Vacuum Technology France - User's Manual ACM 2000 7/22 Commissioning Product description Settings Test mode RS232 mode Appendices Appendix 4 Cable and connector details Remote connector (continued) Table of Remote control functions Gauge Gauge “function 1” input Gauge “function 2” input Pirani AP 2004 - APN 2004 not used not used CRYSTAL/hot cathode AHC2010 HV-DSBL DEGAS Cold cathode/Pirani ACC2009 not used not used Capacitive ASD200X/ARD200X not used not used In compliance with electromagnetic compatibility standards, use shielded cables for connecting the interfaces. Connect the screening to the chassis earth of the other equipment. Without this, “noise” can be generated which can cause damage to equipment. Comply with the cut off rating for the set point outputs: 125 V AC - 0.3 A or 30 V DC - 1 A. Overvoltages and overcurrents can cause internal electrical damage. Description and connection of RS232 cables ACM2000 RS232 connector D-sub9P (Male) RS232 cable D-sub9P (Female) D-sub9P to PC Pin No. Signal name Description 2 RXD Data reception 3 TXD Data transmission 5 GND Chassis earth If the pin number is not described, it is because it is not used. The RS232 specifications are described in chapter E “RS232 Mode”. 8/22 Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03429 - Edition 03 - July 2010 A B C D E F A B C D E F Appendix 5 Factory settings Commissioning Product description Settings Test mode RS232 mode Appendices The values indicated in the table below correspond to factory default settings and can be restored at any time. You can use the «User value» column to note your own specific settings. Default value Under-Range control OFF User value Gas setting 1 Filter setting NORM Full scale setting 1333 mbar Digit setting 2 digit Unit setting mbar Set point select OFF Chanel select ON Baud rate setting 9600 bps Parameter lock setting OFF RS232 setting CR Start-up display select Digital Set-Point see “Settings by gauge type” table Chapter C GB 03429 - Edition 02 - July 2010 Setting Alcatel Vacuum Technology France - User's Manual ACM 2000 9/22 Appendix 6 Error messages Commissioning Product description Settings Test mode RS232 mode Appendices Fault Controller Gauge Led Voltage & display Definition Possible fault Led OFF No display Power supply board Led ON No display Display or display control P Q 1- Check power supply 2- Contact customer Service for repair R Err 01 Action Eprom fault Channel control board Button activation other than «DISP» or «CH» «Err C#» running Contact customer Service for repair 1- Press ENTER for 3 seconds «Setting» appears in flashing then, «SET» is displayed. Switch controller off, then switch on 2- Contact customer Service for repair Err 02 display 2 seconds S Err 03 A/D CPU converter fault -Measurement stopped- Main board Correct «Err C#» 1- Switch controller off, then switch on 2- Contact customer Service for repair T Err 04 A/D converter fault -Measurement stopped- Channel control board 1- Switch controller off, then switch on 2- Contact customer Service for repair Err 05 display 2 seconds Modification of a parameter Function «Parameter Lock Setting» (ON) running Deactivate the function «Parameter Lock Setting» (OFF) Err 06 Gauge fault: Eprom or converter A/D -Measurement stopped- Gauge error See gauges: Fault catalog Err 07 AHC gauge fault: Filament B-A or electronic -Measurement stopped- Gauge error See gauges: Fault catalog Err 08 display 2 seconds Starting of the AHC “degas” function -Function not started- Conditions required not present Check the required conditions Err 09 display 2 seconds Starting of the A/D “zero” function -Function not started- Conditions required not present Check the required conditions Err Hi Gauge fault: signal 0-10V too high -Measurement stopped- Gauge error See gauges: Fault catalog E F A Led ON 10/22 Led OFF Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03429 - Edition 03 - July 2010 A B C D E F A B C D E F Appendix 6 Error messages Commissioning Product description Settings Test mode RS232 mode Appendices Fault Controller Led Voltage & display Definition Possible fault Action Or Pressure over the maximum readable pressure of the gauge (Over range) Gauge message See gauges: Fault catalog Ur Pressure below the maximum readable pressure of the gauge (Under range) Gauge message See gauges: Fault catalog Err Lo Gauge fault: signal 0-10V too low -Measurement stopped- Gauge error See gauges: Fault catalog Err Id Identification error of the gauge The gauge is not recognized Check the gauge Check the cable No gauge connected No gauge connected Check the gauge Check the cable Gauge B D G No GAUGE U Err C# Internal communication on channel 1,2 or 3 Main board 1- Switch controller off, then switch on 2- Contact customer Service for repair V Fan WARNING Poor functioning fan Fan 1- Check the cleanliness of inlet and fan 2- Contact customer Service for repair W Fan stopped Fan Contact customer Service for repair Led OFF GB 03429 - Edition 02 - July 2010 Led ON Fan STOP Alcatel Vacuum Technology France - User's Manual ACM 2000 11/22 12/16 (green) (green) Alcatel Vacuum Technology France - User's Manual ACM 2000 ON - O - I Light indicators : - H (red) V 0,2 V - - Flashing - - V < -0,5 V (red) (red) V < -0,2 V - - (red) 0,1 ≤ T < 0,3 V (red) V 0,1 V - - (red) 0,2 ≤ V < 10,5 or (green) Voltage ≥10,5 (red) - ASD 200x ARD 200x (red) 0,3 V < 0,5 V (green) 0,5 V < 0,774 V GB 03429 - Edition 03 - October 09 OFF - (red) V 0,2 V - (green) 0,774 ≤ V<10,061 V or 10,061≤ V < 10,3 V (green) (red) V 10,3 V AHC 2010 unpredictable message -30.67 Err Lo Err 07 Err 06 Ur 09677 -20.67 ASD-ARD AP/APN-ACC et AHC 1.2 +2 Or Err Hi Controller message electronic «zéro» drift or contaminated electronic B-A broken filament or contaminated electronic ASD-ARD (red) : pressure over the full scale range or temperature out range ASD-ARD (green): if value measurement is not correct = gauge not «zeroed» or «zero» drift or contamination AP/APN, ACC, AHC: if correct ATM/ VAC adjustment is not possible or in case of erratic measurement = Sensor contaminated AP/APN, ACC: broken filament of Pirani AHC: crystal sensor contaminated or defect Possible defect replace the gauge 1- «zéro» adjustment 2- replacement of the gauge replace electronic 1- check B-A filament 2- run «degas» function 3- replace sensor head replace electronic ASD-ARD (green): «zéro» adjustment ASD-ARD (red): none 2- replace sensor head 1- AP/APN, ACC, AHC: check sensor head AHC: start «degas» 2- replace sensor head 1- AP/APN, ACC: check Pirani filament AHC: check crystal sensor Action Commissioning Product description Settings Test mode RS232 mode Appendices G - E F (red) 0,2 V < 0,5 V (red) 0,2 V < 0,5 V D 1,8 V < 8,645 V (green) 0,5 V < 2,199 V 2,199 V < 8,582 V 8,645≤ V < 9,5 V (green) (green) 8,582 ≤V < 9,5 V (red) V 9,5 V ACC 2009 (red) V 9,5 V AP/APN 2004 Signalisation des jauges : Led Power & Voltage (V) (green) 0,5 V< 2,199 V C B A Fault Gauges: Fault catalog A B C D E F Appendix 6 Error messages A B C D E F Appendix 7 Accessories Commissioning Product description Settings Test mode RS232 mode Appendices Power sockets and cables 1a 1b 1c Item Description P.N 1a mains cable - Europe - 2m 103566 1b mains cable - US+Japan - 2m 103567 1c mains cable - UK - 2m 104411 1d mains cable - Switzerland - 2m 103718 1e mains cable - Italy - 2 m 104758 1d 1e GB 03429 - Edition 02 - July 2010 Gauge cables Adapter kit series 1000/series 2000 measuring range Alcatel Vacuum Technology France - User's Manual ACM 2000 Length P.N 5m 112752 10 m 112753 20 m 112754 Description P.N Gauge adapter 113152 Controller adapter 113153 13/22 Commissioning Product description Settings Test mode RS232 mode Appendices Appendix 8 Conversion of measuring units Weight kg oz 35,274 kg 1 2,205 lb 0,454 1 31,081x10-3 16 slug 14,594 32,174 1 514,785 oz 28,349x10-3 62,5x10-3 1,943x10-3 1 N/m2, Pa bar mbar Torr at N/m2, Pa 1 10x10-6 10x10-3 7,5x10-3 9,869x10-6 bar 100x103 1 103 750,062 0,987 mbar 100 10-3 1 750,062x10-3 0,987x10-3 Torr 133,322 1,333x10-3 1,333 1 1,316x10-3 at 101,325x103 1,013 1,013x103 760 1 mbar Pascal Torr mmH2O psi mbar 1 100 750,062x10-3 10,2 14,504x10-3 Pascal 10x10-3 1 7,5x10-3 0,102 0,145x10-3 Torr 1,333 133,322 1 13,595 19,337x10-3 mmH2O 9,81x10-2 9,81 7,356x10-2 1 1,422x10-3 68,948 6,895x103 51,715 703 1 psi Length mm m inch ft mm 1 10-3 39,37x10-3 3,281x10-3 m 103 1 39,37 3,281 inch 25,4 25,4x10-3 1 8,333x10-2 ft 304,8 0,305 12 1 Temperature Kelvin (K) Celsius (°C) Fahrenheit (°F) Kelvin (K) 1 °C+273,15 (°F+459,67)x5/9 Celsius (°C) K-273,15 1 5/9x°F-17,778 9/5x(°C+17,778) 1 Fahrenheit (°F) 9/5xK-459,67 14/22 slug 68,522x10-3 Pressure Vacuum technique pressure units lb Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03429 - Edition 03 - July 2010 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Appendix 9 Declaration of conformity GB 03429 - Edition 02 - July 2010 A B C D E F Alcatel Vacuum Technology France - User's Manual ACM 2000 15/22 Commissioning Product description Settings Test mode RS232 mode Appendices 1 Appendix 10 Detailed example of navigation (Adjusting a set point) 2 Change to settings mode (*) see capter/ display description Menu choice (1) + (1) and confirm (2) 3 Sub-menu choice (1) + (1) and confirm (2) 00:00:00 (*) Set-Point _____ Set-Point Pressure Under-Range Control _____ _____ mbar 1 1 4 5 Choose channel (1) Choose setting (2) + (2) and confirm (2) Set-Point Pressure CH 1 1 2 Change value of setting (1) + (1) 6 Set-Point Pressure mbar 1H 1L 2H 2L 1 2 CH 3.75E-4 3.75E-4 1.23E-2 1.23E-2 1 2 Confirm value of setting (1) Set-Point Pressure mbar 1H 1L 2H 2L 2 CH 4.62E-1 3.75E-4 1.23E-2 1.23E-2 mbar 1H 1L 2H 2L 4.62E-1 3.75E-4 1.23E-2 1.23E-2 1 2 2 7 3 3 2 Leave setting 1 8 Set-Point Pressure CH 1 1 Leave sub-menu 9 Leave settings mode Set-Point mbar 1H 1L 2H 2L Set-Point Pressure 3.75E-4 3.75E-4 1.23E-2 1.23E-2 Under-Range Control 10 16/22 1 1 Back to initial screen Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03429 - Edition 03 - July 2010 A B C D E F A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Appendix 11 Storing and Retrieving Pressure Data (Unit with V1.11 firmware or upper only) ACM 2000 gauge controllers with firmware V 1.11 or higher offer the possibility to store up to 24 hours of pressure data in the instruments memory. One data record of the pressures measured on each channel is stored per second. The data can be retrieved via the RS232 interface of the instrument. Recording data By pushing the “CH” key for two seconds data recording can be started and stopped. This recording function is only available while the GRAPH screen is displayed. The maximum recording time is 24 hours. In case the recording time exceeds 24 hours, the oldest data is overwritten. Therefore the stored data is the latest data stored before the recording is stopped. Recording time is displayed. Its color will be red in case the recording time exceeds 24 hours. GB 03429 - Edition 02 - July 2010 “REC” is displayed during recording. “COMM” is displayed during data read out via RS232. While data is retrieved “COMM” appears on the screen) data recording cannot be started. Push “CH” for 2 s to start or stop recording. Alcatel Vacuum Technology France - User's Manual ACM 2000 17/22 Commissioning Product description Settings Test mode RS232 mode Appendices Annexe 11 Storing and Retrieving Pressure Data (Unit with V1.11 firmware or upper only) “REC” and timer are also displayed on the ANALOG and DIGITAL screens during recording. However, the start and stop operations can only be activated on the GRAPH screen. 18/22 Retrieving Data The stored data can be retrieved via the RS232 interface of the instrument when the recording has been stopped. The data is send as “comma separated values” using the “1K-XMODEM” protocol. This protocol can be handled by many communication programs, for example the HyperTerminal program available on standard WINDOWS® installations. An example of retrieving data with HyperTerminal is given hereafter. Details of the commands, 1KXMODEM protocol and data structure of csv-file follow in the next chapter. Retrieving Data with Connect the ACM 2000 with an appropriate cable to a personal HyperTerminall computer (see appendix 4, page 8). 18/22 Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03429 - Edition 03 - July 2010 A B C D E F A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Annexe 11 Storing and Retrieving Pressure Data (Unit with V1.11 firmware or upper only) Retrieving Data with Open Start /All Programs /Accessories /Communication / HyperTerminall (Cdt.) HyperTerminal. Select the correct settings for the serial communication (see chapter E, page 2). Start and Stop data acquisition with the ACM 2000 controller as described above. The “REC” info must not show on the screen and the timer should display a value other than “00:00:00”. - type “REC,0CR” on the HyperTerminal (CR = carriage return). - start the x-modem data acquisition - in the menu choose “Transfer/Receive data…”. - enter a file name and choose “1K XMODEM” from the dropdown list as transmission protocol - and push “Receive“. The ACM 2000 will immediately start data transmission. A status screen is displayed during data sending. When it disappears, data transmission is finished. Open the csv-file with a spread sheet program to see and process the data. GB 03429 - Edition 02 - July 2010 Details of commands and Data transmission Procedure to read out the recorded data: • send command «REC» via RS232 <$> <REC> <,a> <CR> [LF] Argument a Meaning 0 all channels 1 select channel 1 2 select channel 2 3 select channel 3 During pressure data transmission “REC” command cannot be accepted and the ‘illegal operation’ error will be returned. The addition of LF code for communication of “REC” command and recorded data follows the setting in “PARAMETER SETTING MODE“. • send code (“C” 0x43) to request the start of the 1K-XMODEM protocol. • receive the data in blocks of 1029 bytes. Alcatel Vacuum Technology France - User's Manual ACM 2000 19/22 Commissioning Product description Settings Test mode RS232 mode Appendices Annexe 11 Storing and Retrieving Pressure Data (Unit with V1.11 firmware or upper only) Details of commands and Data transmission Cdt.) • transmission is based on the 1K XMODEM protocol • Time out period from sending “REC“ command to sending start code “C” is 60 sec. • in case ACM 2000 does not receive ACK (0x06) for 5 sec or longer after sending a packet or EOT, it also returns to normal command mode. • in case ACM 2000 sends a packet and receives NAK (0x06), it resents the same packet. • in case ACM 2000 receives CAN (0x18), it discontinues communication and returns to normal command mode. • if file size is not a multiple of 1024, parts without data in the final packet are filled with SUB (0x1A). • packet size is a 1029 byte record and it is sent with the format below: 20/22 Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03429 - Edition 03 - July 2010 A B C D E F A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Annexe 11 Storing and Retrieving Pressure Data (Unit with V1.11 firmware or upper only) Details of commands and Data transmission Cdt.) STX BC BC 1024 data bytes STX header byte (0x02) to start the block BC block count from 0x00 to 0xFF BC complement of block count (0x00~0xFF) CRCH high-order 8 bits of cyclic redundance check code CRCL low-order 8 bits of cyclic redundance check code Details of Data format (total 13 bytes) (total 10 bytes) CRCL Exclusing header, CRC part and SUB code The format of the recorded pressure data depends on the type of gauge(s) connected to the controller. - data format for capacitance gauges: b,±#.####E±## - data format for non-capacitance gauges: b,#.##E±## Reponse b GB 03429 - Edition 02 - July 2010 CRCH Meaning 0 [A] measured value is OK [with ‘fan stopped’ warning active] 1 [B] underrange [with ‘fan stopped’ warning active] 2 [C] overrange [with ‘fan stopped’ warning active] 3 controller failure (Err 03, Err 04, Err C#, FAN STOP) 4 off (ion gauge only) 5 no gauge 6 identification resistance error (Err ID) 7 gauge error (Err Hi, Err Lo, Err 10, Err 06, Err 07) The received data are preceded by a header with information on the measuring channels: Alcatel Vacuum Technology France - User's Manual ACM 2000 21/22 A B C D E F Commissioning Product description Settings Test mode RS232 mode Appendices Annexe 11 Storing and Retrieving Pressure Data (Unit with V1.11 firmware or upper only) Details of Data format - the argument of the REC command is “0” (all channels) CH1,NAME1,CH2,NAME2,CH3,NAME3<CR>[LF] (total 30~31byte) - the argument of the REC command is “1” (channel one) CH1,NAME1<CR>[LF] (total 10~11byte) - the argument of the REC command is “2” (channel two) CH2,NAME2<CR>[LF] (total 10~11byte) - the argument of the REC command is “3” (channel three) CH3,NAME3<CR>[LF] (total 10~11byte) NAME1, NAME2, and NAME3 are the channel names as entered by the user in PARAMETER SETTING MODE (see chapter C, page 4, Channel Name Input). Example for Communication Data CH1 capacitance gauge (pressure value measured: 1.2345E+03 Pa, NAME setting: CAP) CH2 Pirani gauge (pressure value measured: 6.78E+04 Pa, NAME setting: PIR) CH3 not connected (NAME setting: NGAUG) Example for data recorded for three seconds under the conditions above (communication codes are not described). Data received (excluding header, CRC part, and SUB codes): CH1,CAP,CH2,PIR,CH3,NGAUG <CR>[LF] 0,+1.2345E+03,0,6.78E+04,5,0.00E+00<CR>[LF] 0,+1.2345E+03,0,6.78E+04,5,0.00E+00<CR>[LF] 0,+1.2345E+03,0,6.78E+04,5,0.00E+00<CR>[LF] The same data imported into a spread sheet program and presented in a formatted table (the “Time“ column was added manually for convenience): Heure CH1 22/22 CAP CH2 PIR CH3 NGAUG 0 0 1.2345E+03 0 6.78E+04 5 0.00E+00 1 0 1.2345E+03 0 6.78E+04 5 0.00E+00 2 0 1.2345E+03 0 6.78E+04 5 0.00E+00 Alcatel Vacuum Technology France - User's Manual ACM 2000 GB 03429 - Edition 03 - July 2010 Command sent: $REC,0<CR>[LF] CHINA JAPAN SWEDEN Alcatel Vacuum Technology, Shanghai N°82 Lane 887 Zuchongzhi Road Zhangjiang High-Tech Park, Shanghai 201203 China Tel. (86) 21 5027 0628 Fax. (86) 21 3895 3815 Alcatel-Lucent Japan Ltd. 1-9-4 Kita Shin-Yokohama Kohoku-ku Yokohama, Kanagawa 223-0059 Japan Tel. (81) 3 6431 7130 Fax. (81) 45 544 0049 Adixen Sensistor AB. Box 76 SE-581 02 Linköping Sweden Tel. 46 (0)13 35 59 00 Fax. 46 (0)13 35 59 01 GERMANY KOREA TAIWAN Alcatel Hochvakuumtechnik GmbH Am Kreuzeck 10 - Postfach 1151 97877 Wertheim Germany Tel. (49) 9342 9610 0 Fax. (49) 9342 9610 30 Alcatel Vacuum Technology Korea 315-2 Banwol-dong, Hawsung-si, 445-330 Kyungki-do, South Korea Tel. (82) 031-206-6277 Fax. (82) 031-204-6279 Alcatel Vacuum Technology Taïwan No. 169-3, Sec.1, Kang-Leh Rd Song-Lin Village, Hsin-Feng 304 Hsin-Chu County Taïwan - R.O.C. Tel. (886) 3 559 9230 Fax.(886) 3559 9231 INDIA NETHERLANDS UNITED KINGDOM Alcatel Vacuum Technology India Deepak 812, 8th Floor, Park Centra, Sector-30, Gurgaon - Haryana - 122 001 INDIA Tel. (91) 124-4737777 Fax. (91) 124-4737799 ALCATEL Vacuum Technology Benelux Landzichtweg 60 NL 4105 DP Culemborg The Netherlands Tel. (31) 345 478 400 Fax. (31) 345 531 076 Alcatel Vacuum Technology UK Ltd 8 Bain Square - Kirkton Campus Livingston - West Lothian EH54 7DQ - Scotland United Kingdom Tel. (44) 1 506 418 000 Fax. (44) 1 506 418 002 ITALY SINGAPORE USA Alcatel Vacuum Systems Via Trento, 30 20059 Vimercate (Mi) Italy Tel. (39) 0396 86 38 55 Fax. (39) 039 66 71 25 Alcatel-Lucent Singapore 49 Jalan Pemimpin #01-02 APS Industrial Building 577203 Singapore Tel. (65) 6254 0828 Fax. (65) 6254 7018 Alcatel Vacuum Products 67, Sharp Street Hingham - MA 02043 USA Tel. (1) 781 331 4200 Fax. (1) 781 331 4230 Alcatel Vacuum Technology France - 98, avenue de Brogny - BP 2069 - 74009 Annecy cedex - FRANCE Tel. (33) 4 50 65 77 77 - Fax. (33) 4 50 65 77 89 Web site: www.adixen.com ">

Public link updated
The public link to your chat has been updated.
Advertisement
Key features
- Multi-channel control
- Gauge compatibility
- Pressure display and monitoring
- RS232 communication
- Set point management
- Error message display
- Display modes (digital, graph, analog)
- Safety features
- User-configurable settings
- Remote control capability
Frequently asked questions
The ACM 2000 is compatible with the following gauges: PIRANI gauge: AP 2004 - APN 2004, Combined PIRANI/Cold cathode gauge: ACC 2009, Combined CRYSTAL/Hot cathode gauge: AHC 2010, Capacitive gauges: ASD 200X, ARD 200X series.
You can change the display mode between digital, graph, and analog by using the "DISP" button on the controller's front panel.
To set up a set point, you need to access the "SET" menu on the controller. Follow the instructions in the manual to navigate the settings and configure the desired set point parameters for each gauge.
The ACM 2000 has a built-in RS232 interface. You can connect the controller to a computer using a standard RS232 cable. Refer to the manual for detailed instructions on configuring the RS232 connection.
The ACM 2000 is designed for safe use. The manual outlines important safety recommendations and instructions for using the device. These instructions include guidelines for connecting the controller to electrical power, handling the enclosure, and ensuring adequate ventilation.