Chauvin Arnoux CA 6116N - CA 6117 User's manual

Chauvin Arnoux CA 6116N - CA 6117 User's manual

The CHAUVIN ARNOUX CA 6116N - CA 6117 installation testers are portable devices with a color graphic display. They're designed for checking electrical installations, whether new, existing, or in need of troubleshooting. These devices can test various aspects of an installation, including voltage, resistance, insulation resistance, and earth resistance, as well as performing RCD testing.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

CHAUVIN ARNOUX CA 6116N - CA 6117 Manual | Manualzz
EN - User’s manual
C.A 6116N
C.A 6117
Installation testers
Thank you for purchasing a C.A 6116N or C.A 6117 installation tester.
To obtain the best service from your unit:
■ read this user manual carefully,
■ comply with the precautions for use.
WARNING, risk of DANGER! The operator must refer to these instructions whenever this danger symbol appears.
Useful information or tip.
Current clamp.
USB socket.
Auxiliary rod.
The voltage on the terminals must not exceed 550 V.
Equipment protected by double insulation.
The product is declared recyclable following a life cycle analysis in accordance with standard ISO 14040.
Chauvin Arnoux has adopted an Eco-Design approach in order to design this appliance. Analysis of the complete
lifecycle has enabled us to control and optimize the effects of the product on the environment. In particular this appliance exceeds regulation requirements with respect to recycling and reuse.
The CE marking indicates compliance with the European Low Voltage Directive (2014/35/EU), Electromagnetic
Compatibility Directive (2014/30/EU), and Restriction of Hazardous Substances Directive (RoHS, 2011/65/EU and
2015/863/EU).
The UKCA marking certifies that the product is compliant with the requirements that apply in the United Kingdom, in
particular as regards Low-Voltage Safety, Electromagnetic Compatibility, and the Restriction of Hazardous Substances.
The rubbish bin with a line through it indicates that, in the European Union, the product must undergo selective disposal
in compliance with Directive WEEE 2012/19/EU. This equipment must not be treated as household waste.
PRECAUTIONS FOR USE
This instrument is compliant with safety standard IEC/EN 61010-2-034 and the leads are compliant with IEC/EN 61010-031, for
voltages of 600V category III or 300V in category IV (under shelter).
Failure to observe the safety instructions may result in electric shock, fire, explosion, and destruction of the instrument and of the
installations.
■ Do not exceed the maximum rated voltage and current and the measurement category.
■ Never exceed the protection limits indicated in the specifications.
■ Comply with the conditions of use, namely the temperature, the humidity, the altitude, the degree of pollution, and the place of use.
■ Do not use the device or its accessories if they seem damaged.
■ Do not use the device if the battery compartment cover is missing or incorrectly installed.
■ To recharge the battery, use only the mains adapter unit provided with the device.
■ To replace the battery, disconnect everything connected to the device and set the switch to OFF.
■ Do not use a battery with a damaged jacket.
■ Use connection accessories of which the overvoltage category and service voltage are greater than or equal to those of the
measuring device (600 V Cat. III or 300 V Cat. IV).
■ Troubleshooting and metrological checks must be done only by accredited skilled personnel.
■ Wear the appropriate protective gear.
2
CONTENTS
1. FIRST START-UP..................................................................................................................................................................5
1.1. Unpacking.....................................................................................................................................................................5
1.2. Accessories..................................................................................................................................................................6
1.3. Replacement parts.......................................................................................................................................................6
1.4. Charging the battery.....................................................................................................................................................7
1.5. Carrying the device.......................................................................................................................................................7
1.6. Use on a desktop..........................................................................................................................................................8
1.7. Brightness of the display..............................................................................................................................................8
1.8. Choice of language.......................................................................................................................................................9
2. PRESENTATION OF THE DEVICES.................................................................................................................................10
2.1. Functions of the devices ............................................................................................................................................ 11
2.2. Keypad....................................................................................................................................................................... 11
2.3. Display unit.................................................................................................................................................................12
2.4. USB port.....................................................................................................................................................................12
3. USE.......................................................................................................................................................................................13
3.1. General.......................................................................................................................................................................13
3.2. Voltage measurement.................................................................................................................................................13
3.3. Resistance and continuity measurement....................................................................................................................15
3.4. Insulation resistance measurement............................................................................................................................19
3.5. 3P earth resistance measurement..............................................................................................................................22
3.6. Loop impedance measurement (ZS)...........................................................................................................................26
3.7. Earth measurement on live circuit (Za, Ra)..................................................................................................................30
3.8. Selective earth measurement on live circuit...............................................................................................................35
3.9. Measurement of the line impedance (Zi)....................................................................................................................38
3.10. Measurement of the voltage drop in the cables (ΔV)...............................................................................................42
3.11. Test of residual current device..................................................................................................................................45
3.12. Current and leakage current measurement..............................................................................................................53
3.13. Direction of phase rotation........................................................................................................................................55
3.14. Power measurement................................................................................................................................................57
3.15. Harmonics................................................................................................................................................................60
3.16. Compensation for the resistance of the measurement leads................................................................................... 63
3.17. Adjustment of the alarm threshold............................................................................................................................65
4. ERROR INDICATION..........................................................................................................................................................66
4.1. No connection.............................................................................................................................................................67
4.2. Out of measurement range.........................................................................................................................................67
4.3. Presence of dangerous voltage..................................................................................................................................67
4.4. Invalid measurement..................................................................................................................................................67
4.5. Device too hot.............................................................................................................................................................67
4.6. Check of internal protection devices...........................................................................................................................68
5. SET-UP.................................................................................................................................................................................69
6. MEMORY FUNCTION.........................................................................................................................................................72
6.1. Choice of mode..........................................................................................................................................................72
6.2. Tree mode..................................................................................................................................................................73
6.3. Table mode.................................................................................................................................................................78
7. DATA EXPORT SOFTWARE..............................................................................................................................................83
7.1. Functions....................................................................................................................................................................83
7.2. Obtain the ICT software..............................................................................................................................................83
7.3. Installing ICT ..............................................................................................................................................................83
8. TECHNICAL CHARACTERISTICS...................................................................................................................................84
8.1. General reference conditions.....................................................................................................................................84
8.2. Electrical characteristics.............................................................................................................................................84
8.3. Variations in the range of use.....................................................................................................................................95
8.4. Intrinsic uncertainty and operating uncertainty...........................................................................................................98
8.5. Power supply..............................................................................................................................................................98
8.6. Environmental conditions.........................................................................................................................................100
8.7. Mechanical characteristics ......................................................................................................................................100
8.8. Conformity to international standards.......................................................................................................................100
8.9. Electromagnetic compatibility (EMC)........................................................................................................................100
9. DEFINITIONS OF SYMBOLS..........................................................................................................................................101
10. MAINTENANCE...............................................................................................................................................................103
10.1. Cleaning.................................................................................................................................................................103
10.2. Replacing the battery..............................................................................................................................................103
10.3. Resetting the device...............................................................................................................................................104
10.4. Updating of the internal software............................................................................................................................104
11. WARRANTY ....................................................................................................................................................................105
12. APPENDIX.......................................................................................................................................................................106
12.1. Table of fuses managed by the C.A 6117...............................................................................................................106
3
Definition of measurement categories
■ Measurement category IV (CAT IV) corresponds to measurements taken at the source of low-voltage installations.
Example: power feeds, meters and protection devices.
■ Measurement category III (CAT III) corresponds to measurements on building installations.
Example: distribution panel, circuit-breakers, stationary machines or fixed industrial devices.
■ Measurement category II (CAT II) corresponds to measurements taken on circuits directly connected to low-voltage installations.
Example: power supply to domestic appliances and portable tools.
Example to identify the locations of measurement categories
5
7
6
7
8
11
10
888
16
7
L1
(L2, L3)
4
888
3
9
15
N
1
12
2
13
CAT IV
14
CAT III
1
Low-voltage supply source over or under ground
9
2
Service fuse
10 Light switch
3
Tariff meter
11
4
Mains circuit breaker or isolator switch *
12 Junction box
5
Photovoltaic panel
13 Socket wiring
6
DC to AC inverter
14 Socket outlets
7
Circuit breaker or isolator switch
15 Plug-in lamps
8
Generation meter
16 Household appliances, portable tools
CAT II
Distribution board
Lighting
* : The mains circuit breaker or isolator switch may be installed by the service provider. If not, the demarcation point between CAT
IV and CAT III is the first isolating switch in the distribution board.
4
1. FIRST START-UP
1.1. UNPACKING
①
③
④
②
⑦
⑨
⑥
⑤
⑫
⑪
⑧
⑩
⑬
⑭
⑮
⑯
①
②
③
④
⑤
⑥
⑦
⑧
⑨
⑩
⑪
⑫
⑬
⑭
⑮
⑯
One C.A 6116N or C.A 6117.
One mains charger with cable for the battery.
One USB A/B cord.
One tripod cable with mains plug (adapted to the country of sale.
One tripod cable, 3 safety leads.
Three probe tips (red, blue, and green).
Three crocodile clips (red, blue, and green).
Two elbowed-straight safety leads (red and black).
One hand strap.
One remote probe.
One carrying bag.
One multilingual quick start guide.
One multilingual safety sheet.
One battery data sheet.
One test report.
One 4-point hands-free strap.
5
1.2. ACCESSORIES
15m earth kit (red/blue/green)
3P earth kit (50 m)
3P earth kit (100 m)
1P earth kit (30 m, black)
C177A clamp (200 A)
MN77 clamp (20 A)
Continuity pole
Lithium Ion pack charging stand
Dataview Software
1.3. REPLACEMENT PARTS
Battery pack Lithium Ion
USB-A USB-B Cord
Mains power unit/type 2 charger
2P mains cable Euro
2P mains cable GB
2P mains cable US
Screen protection film
4-point hands-free strap
No. 22 carrying bag
Remote probe
Black prod for remote control probe
Tripod cable, Euro plug
Tripod cable, GB plug
Tripod cable, IT plug
Tripod cable, CH plug
Tripod cable, US plug
Measuring cable-3 safety leads (red, blue and green)
Measuring cable-3 safety leads (red, blue and green) CH
3 4mm-dia. probe tips (red, blue and green)
3 crocodile clips (red, blue and green)
2 elbowed-straight safety leads (red and black) 3m long
Hand strap
For accessories and spare parts, visit our website:
www.chauvin-arnoux.com
6
1.4. CHARGING THE BATTERY
Before the first use, start by fully charging the battery. The charging must be done between 0 and 45°C.
> 90 VAC
< 264 VAC
50 / 60 Hz
Battery
loading...
The indicator of the device
lights.
Remove the cover from the mains connector on
the device.
Loading
completed.
Charging time:
approximately 5h.
When charging is complete,
the indicator goes off. The
animation on the screen
will continue as long as the
charger is connected.
Set the switch to OFF, but charging is possible when the device is not off.
1.5. CARRYING THE DEVICE
The 4-point hands-free strap will let you use the device while leaving your
hands free. Snap the four fasteners of the strap onto the four lugs on the
device.
Pass the strap around your neck.
Adjust the length of the strap,
then the tilt of the device.
7
To withdraw the strap, slide a flat screwdriver under the tab of the fastener to lift it, then slide the fastener down.
1.6. USE ON A DESKTOP
For use on a desktop, have the device rest on the fasteners of the hand strap and on the housing.
This lets the display unit be read directly.
1.7. BRIGHTNESS OF THE DISPLAY
To adjust the brightness of the display, press the

+

or
key and one of the arrow keys simultaneously.

sustained
press
8
1.8. CHOICE OF LANGUAGE
Before using the device, first choose the language in which you want the device to display messages.
Set the switch to SET-UP.
Use the directional keypad to select the languages icon:
SET UP
OK
OFF

OK
Press the OK key to validate your
choice.
Select your language, from among those proposed, using the ▲▼ keys and validate by pressing the OK key again.
You can download other languages from our site’s support space (see §10.4).
9
2. PRESENTATION OF THE DEVICES
TEST button to start
the measurements.
Switch for selection
of the measurement
function or SETUP.

Connection terminals.
TEST
Four function
keys.
/
SET UP
OFF
Indicator light.
OK
Stud for fixing on the
4-point hands-free
strap.
Brightness adjustment key.
Help key.
Directional keypad:
four navigation keys
and one validation
key.
Fasteners for the hand
strap, also used to tilt
the device.
USB port for data
transfer to a PC.
Battery charging
connector.
10
2.1. FUNCTIONS OF THE DEVICES
C.A 6116N and C.A 6117 installation testers are portable measuring devices with a colour graphic display. They are powered by a
rechargeable battery with a built-in charger and external power supply unit.
These instruments are intended to check the safety of electrical installations. It can be used to test a new installation before it is
powered up, to check an existing installation, whether in operation or not, or to diagnose a malfunction in an installation.
Measurement
functions
■
■
■
■
■
■
■
■
■
■
■
■
■
■
■
voltage measurement
continuity and resistance measurement
insulation resistance measurement
earth resistance measurement (with 3 rods)
loop impedance measurement (Zs)
earth resistance on live circuit measurement (with an auxiliary probe)
selective earth resistance measurement (with an auxiliary probe and an optional current clamp)
calculation of the short-circuit current and of the fault voltages
line impedance measurement (Zi)
measurement of the voltage drop in the cables (for the C.A 6117 only)
test of type AC, A , F, B, B+ and EV residual current devices, in ramp mode, in pulse mode, or in
non-tripping mode (types B, B+ and EV with the C.A 6117 only)
current measurement (with an optional current clamp)
detection of direction of phase rotation
active power and power factor measurement (single-phase or balanced three-phase network) with
display of the voltage and/or current curves
voltage and current harmonic analysis (with an optional clamp)
Controls
one thirteen-position switch, one five-key navigator, one keypad with four function keys, one contextsensitive help key, one brightness key, and one TEST button.
Display
5.7” (115 x 86mm) colour graphic display unit, 1/4 VGA (320 x 240 points).
The only difference between the C.A 6116N and the C.A 6117 is that the C.A 6117 can test type B RCDs.
2.2. KEYPAD
The actions of the 4 function keys are indicated on the display unit by adjacent icons. They depend on the context.
The help key can be used in all functions. The help function is context-sensitive: it depends on the function.
The
key is used to adjust the brightness of the display.
The directional keypad comprises four navigation keys and one validation key.
11
2.3. DISPLAY UNIT
③
②
①
50 . 0
11-21-2013 10:47
④
Ω
⑤
50 . 1 Hz
L
6 mA
⑥
L-PE
230.3 V
L-N
230.4 V
⑦
⑧
0.8 V
N-PE
6%
⑨
LOOP ZS
⑩
①
②
③
④
⑤
⑥
⑪
⑦
⑧
⑨
⑩
⑪
Top strip
Date and time
Alarm threshold
Frequency measured
Condition of the battery
Position of the phase on the socket outlet
Display of measurement results
Bottom strip
Name of function
Information about the measurement in progress
Icons representing the functions of the keys
2.4. USB PORT
The USB port of the device is used to transfer the stored data to a PC (see §7). This operation requires the prior installation of a
specific peripheral driver and other software.
The USB port can also be used to update the device’s internal software (see §10.4).
The USB cord and the associated software are supplied with the device.
12
3. USE
3.1. GENERAL
When it leaves the plant, the device is configured so that it can be used without changing the parameters. For most measurements, simply select the measurement function by turning the switch and press the TEST button.
However, you can also parameterize:
■ the measurements, using the function keys,
■ or the device itself, using SET-UP.
The device is not designed to operate when the charger is connected. The measurements must be made using battery
power.
3.1.1. CONFIGURATION
When configuring the measurements, you can always choose between:
■ validating by pressing the OK key,
■ or exiting without saving by pressing the
key.
3.1.2. HELP
In addition to an intuitive interface, the instrument provides complete help in use and analyses and appraisals. Three types of help
function are available:
■ Help before the measurement can be accessed using the
key. It indicates the connections to be made for each function
and important recommendations.
■ Error messages appear, as soon as the TEST button is pressed, to report connection errors, measurement parameterizing
errors, out-of-range values, defective installations tested, etc.
■ Help associated with the error messages. Messages containing the
nate the error found.
icon invite you to look up the help for ways to elimi-
3.1.3. REFERENCE POTENTIAL
The user is assumed to be at the reference earth potential. He/she must therefore not be insulated from earth: must not
wear insulating shoes or insulating gloves and must not use a plastic object to press the TEST button.
3.2. VOLTAGE MEASUREMENT
Whichever function is chosen, except for SET-UP, the device always starts by measuring the voltage present on its terminals.
3.2.1. DESCRIPTION OF THE MEASUREMENT PRINCIPLE
The device separates the alternating voltage from the direct voltage and compares the amplitudes to decide whether the signal
is AC or DC. In the case of an AC signal, the frequency is measured and the device calculates the RMS value of the AC part and
displays it. In the case of a DC signal, the device does not measure its frequency, but calculates its mean value and displays it.
For measurements made at the mains voltage, the device checks that the connection is correct and displays the position of the
phase on the socket outlet. It also checks the presence of a protective conductor on the PE terminal by means of the contact the
user makes with his/her finger by touching the TEST button.
13
3.2.2. MAKING A MEASUREMENT
Connect the lead to the device to be tested. As soon as the device is powered up, it measures the voltages present on its terminals
and displays them, whatever the setting of the switch.
600 V CAT III
/
In the ZS (RA/SEL.) and RCD, settings, the device also indicates the position of the phase on the display unit using the
The mains socket outlet of the measuring cable is marked with a white reference spot.
■
: the phase is on the right-hand contact of the mains plug when the white spot is up.
■
: the phase is on the left-hand contact of the mains plug when the white spot is up.
■
symbol.
: the device cannot determine where the position of the phase, probably because the PE is not connected or the L and PE
conductors are interchanged.
The L symbol is displayed as soon as the voltage is high enough (> UL programmable in SET-UP). The terminal identified
as L is the one that has the highest voltage with respect to PE.
3.2.3. ERROR INDICATION
The only errors reported in voltage measurement are values outside the voltage measurement range. These errors are reported
in clear language on screen.
14
3.3. RESISTANCE AND CONTINUITY MEASUREMENT
3.3.1. DESCRIPTION OF THE MEASUREMENT PRINCIPLE
For continuity measurements, the device generates a DC current of 200 or 12 mA, at the user’s discretion, between the Ω and
COM terminals. It then measures the voltage present between these two terminals and from it deduces the value of R = V/I.
For resistance measurements (current chosen = kΩ), the device generates a DC voltage between the Ω and COM terminals. It
then measures the current between these two terminals and from it deduces the value of R = V/I.
In the case of a measurement at high current (200 mA), at the end of one second, the device reverses the direction of the current
and makes another measurement for one second. The result displayed is the mean of these two measurements. It is possible to
make measurements with either the positive or the negative polarity of the current disabled.
For measurements at low current (12 mA or kΩ), the polarity is positive only.
3.3.2. MAKING A MEASUREMENT
To comply with standard IEC-61557, the measurements must be made at 200 mA. The reversal of the current serves to compensate
for any residual electromotive forces and, more important, to check that the continuity is in fact duplex.
When you make continuity measurements that are not contractual, prefer a current of 12 mA. Even though the results cannot be
regarded as those of a normative test, this significantly increases the life of the device between charges and forestalls untimely
tripping of the installations if there is a connection error.
The permanent mode is used to chain measurements without having to press the TEST button each time.
If the object to be measured is inductive, it is better to switch to pulse mode at 200 mA and make a measurement at positive polarity, then a measurement at negative polarity, manually, in order to leave time for the measurement to settle.
The alarm, if activated, serves to inform the user, by an audible signal, that the measurement is below threshold, making it unnecessary to look at the display unit to check this point.
Set the switch to Ω
.
Use the leads to connect the device to be tested between the Ω and COM terminals of the device. The object to be tested must not be live.
SET UP
R
600 V CAT III
/
OFF
3.3.3. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can configure it by modifying the parameters displayed:
200mA
Choice of measurement current: kW, 12 mA or 200 mA.
■ The high current (200 mA) can be used only to measure low resistances, up to 40 Ω.
■ The low current (12 mA) is used to make measurements up to 400 Ω.
■ The choice kΩ is used to make resistance measurements up to 400 kΩ.
To correct for the resistance of the measurement leads (leads and probe tips or crocodile clips), for measurements
at 12 and 200 mA (see §3.16).
Pressing the TEST button starts only one measurement (pulse mode).
Pressing the TEST button starts the continuous measurement (permanent mode). To
stop it, you must press the TEST button again.
15
Automatic reversal of polarity for a measurement at 200 mA.
Measurement at positive polarity only.
Measurement at negative polarity only.
To activate the alarm.
To deactivate the alarm.
Ω 002.00
kΩ
6%
To set the alarm threshold (see §3.17). The default threshold is 2Ω.
Before the measurement: to display the measurements already recorded.
After the measurement: to record it.
The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
Once the parameters have been defined, you can start the measurement.

If you selected the pulse mode, press the TEST button once and the measurement stops automatically when it is over.
If you selected the permanent mode, press the TEST button once to start the measurement and a second time to
stop it, or else press the record key
6%
directly.
TEST
3.3.4. READING OF THE RESULT
■ In the case of a 200 mA current:
Value of the alarm threshold.
11-22-2013 10:47
2.00 Ω
- - .- Hz
Measurement result:
R = (R+) + (R-)
2
200 mA
Measurement current.
0.83 Ω
I
207.4mA
R+
0.59
Ω
R-
1.08
Ω
Measurement with a positive current (R+).
Measurement with a negative current (R-).
Case where the measurement is
below the alarm threshold.
1%
CONTINUITY
Measurement with reversal of polarity.
Compensation for the resistance of
the measurement leads is activated.
Permanent mode.
16

To see the next display page.
22/07/2014 10:47
2.00 Ω
- - .- Hz
200 mA
UΩ
L - PE
L-N
N - PE
0.0 V
0 . 00 V
X
V
X
V
External voltage present on the
terminals just before the start of the
measurement.
1%
CIAGLOSC
■ In the case of a 12 mA current, there is no current reversal.
11-22-2013 10:47
2.00 Ω
- - .- Hz
Value of the alarm threshold.
12 mA
1 8.4
I
Ω
12.3mA
Measurement result.
Current measurement.
Case where the measurement is
above the alarm threshold.
1%
CONTINUITY
The polarity of the current is positive.
Compensation for the resistance of
the measurement leads is activated.
Pulse mode.
17
■ In the case of a resistance measurement (kΩ), there is no current reversal and no compensation for the measurement leads.
11-22-2013 10:47
- - .- Hz
kΩ
1 . 58 k Ω
Measurement result.
Case where the measurement is
below the alarm threshold.
1%
RESISTANCE
Permanent mode.
3.3.5. ERROR INDICATION
The commonest error in the case of a continuity or resistance measurement is the presence of a voltage on the terminals. An error
message is displayed if a voltage greater than 0.5 VRMS is detected and you press the TEST button.
In this case, the measurement is not enabled. Eliminate the cause of the interference voltage and start the measurement over.
Another possible error is measurement of an overly inductive load that prevents the measurement current from stabilizing. In this
case, start the measurement in permanent mode with only one polarity and wait for the measurement to stabilize.

For help with connections or any other information, use the help function.
18
3.4. INSULATION RESISTANCE MEASUREMENT
3.4.1. DESCRIPTION OF THE MEASUREMENT PRINCIPLE
The device generates a DC test voltage between the COM and MΩ terminals. The value of this voltage depends on the resistance
to be measured: it is greater than or equal to UN when R ≥ RN = UN /1 mA, and less otherwise. The device measures the voltage
and current present between the two terminals and from them deduces the value of R = V / I.
The COM terminal is the voltage reference point. The MΩ terminal therefore provides a negative voltage.
3.4.2. MAKING A MEASUREMENT
The alarm, if activated, serves to inform the user, by an audible signal, that the measurement is below threshold, making it unnecessary to look at the display unit to check this point.
Use the leads to connect the device to be tested between the COM and MΩ
terminals of the device. The object to be tested must not be live.
Set the switch to MΩ.
SET UP
R
600 V CAT III
/
OFF
To avoid leakage during the insulation measurement, which would
throw off the measurement, do not use the measuring cable when you
make this type of measurement, but two simple leads.
Generally, an insulation measurement on an installation is made between the interconnected phase(s) and neutral, on the one
hand, and earth, on the other.
L1
L2
L3
N
PE
600 V CAT III
/
If the insulation is not sufficient, you must then make the measurement between each of the pairs to locate the fault. It is for this
reason possible to index the recorded value with one of the following values:
L-N, L-PE, N-PE, L1-PE, L2-PE, L3-PE, L1-N, L2-N, L3-N, L1-L2, L2-L3 or L1-L3
The remoted TEST button of the optional remote control probe makes it easier to trigger the measurement. To use the remote
control probe, refer to its user’s manual.
Operation only
with
C.A 6113/16/17
L1
L2
L3
PE
19
3.4.3. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can configure it by modifying the parameters displayed:
To choose the nominal test voltage UN: 50, 100, 250, 500 or 1000 V.
To activate the alarm.
To deactivate the alarm.
k Ω 0500.0
MΩ
To set the alarm threshold (see §3.17). As default, the threshold is set to
R (kΩ) = UN / 1 mA.
Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
6%

Once the parameters have been defined, you can start the measurement.
Keep the TEST button pressed until the measurement is stable. The measurement stops when the TEST button
is released.
TEST
Before disconnecting the leads or starting another measurement, wait a few seconds for the UN voltage to be zero.
3.4.4. READING OF THE RESULT
Value of the alarm threshold.
11-22-2013 10:47
500 kΩ
- - .- Hz
The bargraph provides a rapid quantitative indication of the insulation.
3 1 . 0 6 MΩ
2%
Measurement result.
577
V
The test voltage UN is present and
dangerous.
7
s
Duration of the measurement.
Case where the measurement is
above the alarm threshold.
Press TEST until the measurement
is stable
INSULATION
To change display pages.
20

To see the next display page.
500 kΩ
11-22-2013 10:47
UMΩ
L - PE
L-N
N - PE
2%
- - .- Hz
0.3 V
X
X
X
External voltage present on the
terminals just before the start of the
measurement.
V
V
V
Press TEST until the measurement
is stable
To change display pages.
INSULATION
3.4.5. ERROR INDICATION
The commonest error in the case of an insulation measurement is the presence of a voltage on the terminals. If it is greater than
10 V (the exact value depends on UN, see § 8.2.5), the insulation measurement is not enabled. Eliminate the voltage and start the
measurement over.
The measurement may be unstable, probably because of an overly capacitive load or an insulation fault. In this case, read the
measurement on the bargraph.

For help with connections or any other information, use the help function.
21
3.5. 3P EARTH RESISTANCE MEASUREMENT
This function is the only one that can measure an earth resistance when the electrical installation to be tested is not live (new installation, for example). It uses two auxiliary rods, with the third rod being constituted by the earth electrode to be tested (whence
the name “3P”).
It can be used on an existing electrical installation, but the power must be cut off (main RCD). In all cases (new or existing installation), the earthing strip of the installation must be open during the measurement.
It is possible to make a rapid measurement and measure only RE or else to make a more detailed measurement by also measuring
the resistances of the rods.
3.5.1. DESCRIPTION OF THE MEASUREMENT PRINCIPLE
The device generates between the H and E terminals a square wave at a frequency of 128 Hz and an amplitude of 35 V. It measures the resulting current, IHE , along with the voltage present between the S and E terminals, USE. It then calculates the value of
RE = USE/IHE.
To measure the resistances of the RS and RH rods, the device internally reverses the E and S terminals and makes a measurement.
It then does likewise with the E and H terminals.
3.5.2. MAKING A MEASUREMENT
There are several measurement methods. We recommend the «62%» method.
Set the switch to RE 3P.
SET UP
Plant the H and S rods in line with the earth electrode. The distance between the S rod and
the earth electrode must be approximately 62% of the distance between the H rod and the
earth electrode.
In order to avoid electromagnetic interference, we recommend paying out the full length of
the cables, placing them as far apart as possible, and not making loops.
OFF
H
earth
strap
S
d
62% d
600 V CAT III
/
Connect the cables to the H and S terminals. Power down the installation and disconnect the earth strap. Then connect the E
terminal to the earth electrode to be checked.
The alarm, if activated, serves to inform the user, by an audible signal, that the measurement is above threshold, making it unnecessary to look at the display unit to check this point.
3.5.3. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can configure it by modifying the parameters displayed:
Choice of type of measurement: rapid, to measure RE only (icon crossed out), or detailed, to measure also rod
resistances RS and RH. This last case is useful if the ground is dry, making the resistance of the rods high.
To compensate for the resistance of the lead connected to the E terminal, for measurements of low values (see §3.16).
22
To activate the alarm.
To deactivate the alarm.
Ω
kΩ
050.00
To set the alarm threshold (see §3.17). As default, the threshold is set to 50Ω.
Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
6%
If the measurement must be made in a damp environment, remember to change the value of maximum contact voltage UL
in SET-UP (see §5) and set it to 25 V.

Press the TEST button to start the measurement. The measurement stops automatically.
This symbol invites you to wait while the measurement is in progress.
TEST
Do not forget to reconnect the earth strap at the end of the measurement before powering the installation back up.
3.5.4. READING OF THE RESULT
In the case of a detailed measurement:
11-22-2013 10:47
50.0 kΩ
- - .- Hz
Value of the alarm threshold.
RE
32.08
Ω
Measurement result.
Rs
1.58 kΩ
Resistance of the S rod.
Rh
1.32 kΩ
Resistance of the H rod.
Case where the measurement is
below the alarm threshold.
3%
To see the voltages before the beginning of the test.
EARTH 3P
Compensation for the resistance of
the measurement leads is activated.
23
3.5.5. VALIDATION OF THE MEASUREMENT
To validate your measurement, move the S rod towards the H rod by 10% of d and make another measurement. Then move the
S rod, again by 10% of d, but towards the earth electrode.
S
H
52% d
62% d
72% d
600 V CAT III
/
d
The 3 measurement results must be the same to within a few percent. If this is the case, the measurement is valid. If not, it is
because the S rod is in the zone of influence of the earth electrode.
If the resistivity of the ground is homogeneous, it is necessary to increase distance d and repeat the measurements. If the resistivity
of the ground is inhomogeneous, the measurement point must be moved either towards the H rod or towards the earth terminal
until the measurement is valid.
3.5.6. POSITIONING OF THE AUXILIARY RODS
To make sure that your earth measurements are not distorted by interference, we recommend repeating the measurement with the
auxiliary rods placed at a different distance and in another direction (for example rotated 90° from the first alignment).
H
H
S
d2
S
E
d1
If you find the same values, your measurement is reliable. If the measured values are substantially different, it is probable that they
were influenced by earth currents or a groundwater artery. It may be useful to drive the rods deeper.
If the in-line configuration is not possible, you can plant the rods in a triangle. To validate the measurement, move the S rod on
either side of the line HE.
S
E
H
S
Avoid routing the connecting cables of the earth rods near or parallel to other cables (transmission or power supply), metal pipes,
rails, or fences, this in order to avoid the risk of cross-talk with the measurement current.
24
3.5.7. ERROR INDICATION
The commonest errors in the case of an earth measurement are the presence of an interference voltage or rod resistances that
are too high.
If the device detects:
■ a rod resistance greater than 15 kΩ,
■ a voltage greater than 25 V on H or on S when the TEST button is pressed.
In these two cases, the earth measurement is not enabled. Move the rods and start the measurement over.
To reduce the resistance of the rods RH (RS), you can add one or more rods, two metres apart, in the H (S) branch of the circuit.
You can also drive them deeper and pack the earth around them, or wet it with a little water.

For help with connections or any other information, use the help function.
25
3.6. LOOP IMPEDANCE MEASUREMENT (Z S)
In a TN or TT type installation, the loop impedance measurement is used to calculate the short-circuit current and to size the protections of the installation (fuses or RCDs), especially their breaking capacity.
In a TT type installation, the loop impedance measurement makes it easy to determine the earth resistance without planting any
rods and without cutting off power to the installation. The result obtained, ZS, is the loop impedance of the installation between the
L and PE conductors. It is barely greater than the earth resistance.
From this value and the conventional touch voltage limit (UL), it is then possible to choose the rated differential operating current
of the RCD: I∆N < UL / ZS.
This measurement cannot be made in an IT type installation because of the high earthing impedance of the supply transformer,
which may even be completely isolated from earth.
3.6.1. DESCRIPTION OF THE MEASUREMENT PRINCIPLE
The device starts by generating pulses having a duration of 1.1 ms and an amplitude of at most 7 A between the L and N terminals.
This first measurement is used to determine ZL.
It then applies a low current, 6, 9 or 12 mA at the user’s discretion, between the L and PE terminals. This low current serves to
avoid tripping residual current devices of which the nominal current is greater than or equal to 30 mA. This second measurement
is used to determine ZPE.
The device then calculates loop resistance ZS = ZL-PE = ZL+ZPE , and short-circuit current Ik = ULPE/ZS.
The value of Ik serves to check the proper sizing of the protections of the installation (fuses or RCDs).
For greater accuracy, it is possible to measure ZS with a high current (TRIP mode), but this measurement may trip the RCD of the
installation.
3.6.2. MAKING A MEASUREMENT
Set the switch to ZS (RA/SEL.).
Connect the measuring cable to the device, then to the socket outlet of the installation to be tested.
At the time of connection, the device first checks that the voltages present on
its terminals are correct, then determines the position of the phase (L) and of
the neutral (N) with respect to the protective conductor (PE) and displays it. If
necessary, it then automatically switches the L and N terminals so that the loop
measurement can be made without modifying the connections of the device.
If possible, first disconnect all loads from the network on which you make
SET UP
the loop measurement.
It is possible to eliminate this step if you use a measurement current of 6 mA, which
allows a leakage current of up to 9 mA for an installation protected by a 30 mA residual current device.
Case of a TT installation
OFF
Case of a TN installation
RL
L
RL
L
RN
N
RN
N
RE
PE
PE
Rb
Rb
Ra
600 V CAT III
600 V CAT III
/
/
In trip mode, it is not necessary to connect the N terminal.
26
For a more accurate measurement, you can choose a high current (TRIP mode), but the RCD that protects the installation may trip.
The alarm, if activated, serves to inform the user, by an audible signal, that the measurement is above threshold, making it unnecessary to look at the display unit to check this point.
The signal can be smoothed to produce a mean of several values. But the measurement then takes longer.
3.6.3. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can configure it by modifying the parameters displayed:
6 mA
Choice of measurement current in non-tripping mode: 6, 9, 12 mA
or TRIP mode to use a high current that will give a more accurate measurement.
To compensate for the resistance of the measurement leads, for measurements of low values (see §3.16).
To activate or deactivate the smoothing of the signal.
(Ik)
The device proposes choosing the voltage for the Ik calculation from among the following values:
■ ULN (the measured voltage value),
■ the voltage of the old standard (for example 220 V),
■ the voltage of the current standard (for example 230 V).
Depending on the voltage ULN measured, the device proposes the following choices:
■ if 170<ULN<270 V: ULN, 220 V, or 230 V.
■ if 90<ULN<150 V: ULN, 110 V or 127 V.
■ if 300<ULN<500 V: ULN, 380 V or 400 V.
To deactivate the alarm.
Z-R
To activate the alarm on ZLPE (in TRIP mode) or on RLPE (in non-tripping mode).
Ω
kΩ
Ik
050.00
To activate the alarm on Ik.
A 010.00
kA
Isc
To set the alarm threshold (see §3.17). As default,
the threshold is set to 50 Ω.
To set the alarm threshold (see §3.17). As default,
the threshold is set to 10 kA.
To activate the alarm on Isc, a current that may help in choosing a fuse (for the C.A 6117
only).
To enter the table of fuses.
27
You can then choose:
■ The delay (the duration of application of IN before the fuse blows): 0.1s, 0.2s, 0.4s,
5s, or 35ms.
■ The type of fuse: gG, LSB, LSC or LSD.
■ The rated current IN: any standardized value between 2 and 1000A.
The choices available depend on the choices already made. As does the value of Isc.
6%

Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
Press the TEST button to start the measurement. The measurement stops automatically.
When the TEST button is pressed, the device checks that the contact voltage is less than UL. If not, it does not make
the loop impedance measurement.
This symbol invites you to wait while the measurement is in progress.
TEST
28
3.6.4. READING OF THE RESULT
■ In the case of a non-tripping measurement, with smoothing:
Value of the alarm threshold.
11-22-2013 10:47
50 .0 Ω
50 . 1 Hz
L
Value of the short-circuit current.
Value of the impedance.
6 mA
Ik
Value of the resistance.
1 5 2.0 A
Z s
1.52
Ω
R s
1.36
Ω
L s
2.2 mH
Value of the inductance.
Case in which the measurement is
below the alarm threshold.
To change display pages.
4%
Value of the reference voltage for the
calculation of Ik.
LOOP Zs
Programmed maximum contact
voltage.
Compensation for the resistance of
the measurement leads is activated.
■ In the case of a measurement with tripping (TRIP) and without smoothing:
11-22-2013 10:47
10.0
50 . 1 Hz
Ω
L
Ik
1 1.8 A
Z s
19.31
Ω
R s
19.08
Ω
L s
Value of the short-circuit current.
Value of the impedance.
Value of the resistance.
Case where the measurement is
above the alarm threshold.
9.6 mH
Value of the inductance.
4%
LOOP Zs
3.6.5. ERROR INDICATION
See §3.8.5.
29
3.7. EARTH MEASUREMENT ON LIVE CIRCUIT (Z A, RA)
This function is used to make an earth resistance measurement in a place where it is impossible to make a 3P earth measurement
or to disconnect the earth connection strap, often the case in an urban environment.
This measurement is made without disconnecting the earth, with only one additional rod, saving time with respect to a conventional
earth measurement with two auxiliary rods.
In the case of a TT type installation, this measurement is a very simple way to measure the earth of frame grounds.
In the case of a TN type installation, to determine the value of each of the earths put in parallel, it is necessary to perform a selective earth measurement on live circuit using a current clamp (see §3.8). Without this clamp, what you find is the value of the global
earth connected to the network, which is rather meaningless.
It is then more useful to measure the loop impedance to size the fuses and RCDs, and to measure the fault voltage to check the
protection of persons.
3.7.1. DESCRIPTION OF THE MEASUREMENT PRINCIPLE
The device starts by making a loop measurement ZS (see §3.6) with a low current or a high current, at the user’s discretion. It
then measures the potential between the PE conductor and the auxiliary rod and from it deduces RA = UPI-PE/I, I being the current
chosen by the user.
For greater accuracy, it is possible to make the measurement with a high current (TRIP mode), but this measurement may trip the
RCD of the installation.
3.7.2. MAKING A MEASUREMENT
Set the switch to ZS (RA/SEL.).
Connect the measuring cable to the device, then to the socket outlet of the installation to be tested.
At the time of connection, the device detects the positions of the phase (L)
and of neutral (N) with respect to the protective conductor (PE) and displays
them. If necessary, it then automatically switches the L and N terminals so
that the loop measurement can be made without modifying the connections
of the terminals of the device.
SET UP
If possible, first disconnect all loads from the network on which you make the earth
measurement on line circuit.
It is possible to eliminate this step if you use a measurement current of 6 mA, which
allows a leakage current of up to 9 mA for an installation protected by a 30 mA residual
current device.
OFF
Plant the auxiliary rod at a distance of more than 25 metres from the earth electrode and
(RA SEL) terminal of the device. The
symbol is then displayed.
connect it to the
Case of a TT installation
Case of a TN installation
RL
L
RL
L
RN
N
RN
N
RE
PE
PE
Rb
Rb
Ra
> 25 m
600 V CAT III
PE
Ra
> 25 m
/
30
600 V CAT III
/
To make this measurement, you can choose:
■ either a low current which avoids any untimely tripping out of the installation but gives only the earth resistance (RA).
■ or a high current (TRIP mode), which yields a more accurate earth impedance (ZA) with good measurement stability and can
also be used to calculate the short-circuit fault voltage, UFK, in accordance with standard SEV 3569.
The alarm, if activated, serves to inform the user, by an audible signal, that the measurement is above threshold, making it unnecessary to look at the display unit to check this point.
The signal can be smoothed to produce a mean of values. But the measurement then takes longer.
3.7.3. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can configure it by modifying the parameters displayed:
6 mA
Choice of measurement current: 6 (default), 9, 12 mA,
or TRIP to use a high current that will yield a more accurate measurement.
To compensate for the resistance of the measurement leads, for measurements of low values (see §3.16).
To activate or deactivate the smoothing of the signal.
(Ik)
The device proposes choosing the voltage for the Ik calculation from among the following values:
■ ULN (the measured voltage value),
■ the voltage of the old standard (for example 220 V),
■ the voltage of the current standard (for example 230 V).
Depending on the voltage ULN measured, the device proposes the following choices:
■ if 170<ULN<270 V: ULN, 220 V or 230 V.
■ if 90<ULN<150 V: ULN, 110 V or 127 V.
■ if 300<ULN<500 V: ULN, 380 V or 400 V.
To deactivate the alarm.
Z-R
To activate the alarm on ZA (in TRIP mode) or on RA (in non-tripping mode).
Ω
kΩ
Ik
050.00
To activate the alarm on Ik (in TRIP mode only).
A 010.00
kA
6%
To set the alarm threshold (see §3.17). As default,
the threshold is set to 50 Ω.
To set the alarm threshold (see §3.17). As default,
the threshold is set to 10 kA.
Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
31

Press the TEST button to start the measurement. The measurement stops automatically.
This symbol invites you to wait while the measurement is in progress.
TEST
3.7.4. READING OF THE RESULT
■ In the case of a measurement with a high current (TRIP mode), without smoothing:
Value of the alarm threshold.
11-25-2013 10:47
50 . 0
Ω
50 . 1 Hz
L
UFk
I K
U FK
468
A
0.6 V
Value of the short-circuit current.
Earth electrode fault voltage in the
event of a short-circuit.
Case where the measurement is
above the alarm threshold.
To change display pages.
Value of the reference voltage for the
calculation of Ik.
6%
The rod is connected.
EARTH 1P (Ra)
Programmed maximum contact
voltage.
Compensation for the resistance of
the measurement leads is activated.
UFk is calculated only in earth measurement on live circuit with a high current (TRIP mode). UFk = Ik x ZA.
32

To see the next display page.
11-25-2013 10:47
50 . 0
Ω
50 . 1 Hz
L
UFk
ZA
2 5.1 0 Ω
R a
24.8 Ω
La
5.6
Value of the impedance.
Value of the resistance.
mH
Value of the inductance.
To change display pages.
6%
EARTH 1P (Ra)
The third page displays the values of ZS, RS, LS. The fourth page displays the voltages ULN, ULPE, UNPE and on the rod (
the measurement.
) before
■ In the case of a measurement with a low current and smoothing, the first display screen is the following:
Value of the alarm threshold.
11-25-2013 10:47
50 . 0
Ω
50 . 1 Hz
L
Measurement result.
6 mA
RA
2 5.1 0 Ω
Case where the measurement is
below the alarm threshold.
To change display pages.
Value of the reference voltage for the
calculation of Ik.
6%
The rod is connected.
EARTH 1P (Ra)
Programmed maximum contact
voltage.
Compensation for the resistance of
the measurement leads is activated.
33
3.7.5. VALIDATION OF THE MEASUREMENT
Move the rod ± 10% of the distance from the earth electrode and make two more measurements. The 3 measurement results must
be the same to within a few percent. In this case the measurement is valid.
If this is not the case, this means that the rod is in the zone of influence of the earth electrode. You must then move the rod away
from the earth electrode and redo the measurements.
3.7.6. ERROR INDICATION
See §3.8.5.
34
3.8. SELECTIVE EARTH MEASUREMENT ON LIVE CIRCUIT
This function is used to make an earth measurement and to select one earth from among others, in parallel, and measure it. It
requires the use of an optional current clamp.
3.8.1. DESCRIPTION OF THE MEASUREMENT PRINCIPLE
The device starts by making a loop measurement ZS between L and PE (see §3.6) with a high current, and therefore with a risk of
tripping out the installation. This high current must be used to ensure that the current flowing in the clamp is large enough to be
measured. The device then measures the current flowing in the circuit surrounded by the clamp. Finally, it measures the potential
of the PE conductor with respect to the auxiliary rod and from it deduces RASEL = UPI-PE / ISEL, ISEL being the current measured by
the clamp.
3.8.2. MAKING A MEASUREMENT
Connect the measuring cable to the device, then to the socket outlet of the installation to be tested.
Set the switch to ZS (RA/SEL.).
At the time of connection, the device detects the positions of the phase (L)
and of neutral (N) with respect to the protective conductor (PE) and displays
them. If necessary, it then automatically switches the L and N terminals so
that the measurement can be made without modifying the connections of the
terminals of the device.
SET UP
Plant the auxiliary rod at a distance of more than 25 metres from the earth electrode
(RA SEL) terminal of the device. The
symbol is then disand connect it to the
played.
Connect the clamp to the device; the
symbol is displayed. Then place it on the
earth circuit to be measured.
OFF
Case of a TN installation
RL
L
RN
N
RE
PE
PE
Ra1
Rb
Ra2
Ra3
> 25 m
600 V CAT III
/
Case of a TT installation
RL
L
RN
N
PE
Rb
Ra1
Ra2
Ra3
> 25 m
600 V CAT III
/
For a more accurate measurement, you can choose a high current (TRIP mode), but the RCD that protects the installation may trip.
The alarm, if activated, serves to inform the user, by an audible signal, that the measurement is above threshold, making it unnecessary to look at the display unit to check this point.
The signal can be smoothed to produce a mean of several values. But the measurement then takes longer.
35
In the selective earth measurement on live circuit, it is essential to do a compensation of the measurement leads and to
redo it if it has not been done recently or if you have changed leads.
3.8.3. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can configure it by modifying the parameters displayed:
The measurement current must be a high current (TRIP mode).
To compensate for the resistance of the measurement leads (see §3.16). It is essential for the selective earth measurement on live circuit.
To activate or deactivate the smoothing of the signal.
(Ik)
The device proposes choosing the voltage for the Ik calculation from among the following values:
■ ULN (the measured voltage value),
■ the voltage of the old standard (for example 220 V),
■ the voltage of the current standard (for example 230 V).
Depending on the voltage ULN measured, the device proposes the following choices:
■ if 170<ULN<270 V: ULN, 220 V or 230 V.
■ if 90<ULN<150 V: ULN, 110 V or 127 V.
■ if 300<ULN<500 V: ULN, 380 V or 400 V.
To deactivate the alarm.
Z-R
To activate the alarm on RASEL.
Ω
kΩ
Ik
050.00
To activate the alarm on Ik (in TRIP mode only).
A 010.00
kA
6%

TEST
To set the alarm threshold (see §3.17). As default,
the threshold is set to 50 Ω.
To set the alarm threshold (see §3.17). As default,
the threshold is set to 10 kA.
Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
Press the TEST button to start the measurement. The measurement stops automatically.
This symbol invites you to wait while the measurement is in progress.
36
3.8.4. READING OF THE RESULT
Value of the alarm threshold.
11-25-2013 10:47
50 . 0
50 . 1 Hz
Ω
Measurement result.
Value of the current measured by
the clamp.
Value of the impedance.
R Asel
7%
3 8.4 2 Ω
I sel
163.5mA
Z a
3.840
Ω
R a
3.838
Ω
L a
Value of the resistance.
Value of the inductance.
Case where the measurement is
above the alarm threshold.
To change display pages.
2.6mH
Value of the reference voltage for the
calculation of Ik.
EARTH Ra Sel.
The rod is connected.
Programmed maximum contact
voltage.
Compensation for the resistance of
the measurement leads is activated.
The clamp is connected.
The second page is used to see the value of short-circuit current Ik, f loop impedance ZS, of loop resistance RS and of loop inductance LS.
The third page is used to see the value of the voltages ULN, ULPE, UNPE and on the rod (
) before the measurement.
3.8.5. ERROR INDICATION (LOOP, EARTH ON LIVE CIRCUIT, AND SELECTIVE EARTH ON LIVE CIRCUIT)
The commonest errors in the case of a loop impedance measurement or earth measurement on live circuit are:
■ A connection error.
■ An earth rod resistance that is too high (>15 kΩ): reduce it by packing the earth around the rod and moistening it.
■ A voltage on the protective conductor that is too high.
■ A voltage on the rod that is too high: move the rod out of the influence of the earth electrode.
■ Tripping in the non-tripping mode: reduce the test current.
■ A current measured by the clamp in selective earth on live circuit that is too low: the measurement is not possible.
The user may have picked up a charge of static electricity, for example by walking on a carpet. In this case, when he/she
presses the TEST button, the device displays the error message «earth potential too high». The user must then be discharged by touching an earth before making the measurement.

For help with connections or any other information, use the help function.
37
3.9. MEASUREMENT OF THE LINE IMPEDANCE (Zi)
The loop impedance measurement Zi (L-N, L1-L2, or L2-L3 or L1-L3) is used to calculate the short-circuit current and size the protections of the installation (fuse or RCD), whatever type of neutral the installation uses.
3.9.1. DESCRIPTION OF THE MEASUREMENT PRINCIPLE
The device generates pulses having a duration of 1.1 ms and an amplitude of at most 7 A between the L and N terminals. It then
measures the voltages UL and UN and from them deduces Zi.
The device then calculates the short-circuit current Ik = ULN/Zi the value of which serves to check the proper sizing of the protections of the installation.
3.9.2. MAKING A MEASUREMENT
Connect the measuring cable to the device, then to the socket outlet of the installation to be tested.
Set the switch to Zi.
At the time of connection, the device first checks that the voltages present on
its terminals are correct, then determines the position of the phase (L) and of
the neutral (N) with respect to the protective conductor (PE) and displays it.
If necessary, it then automatically switches the L and N terminals so that the
line impedance measurement can be made without modifying the connections
of the terminals of the device.
SET UP
If you use the measuring cable that is terminated by three leads, you can connect the
PE lead (green) to the N lead (blue). Otherwise, the device cannot display the position
of the phase. But this does not prevent making the measurement.
OFF
Case of a TT installation
Case of a TN installation
RL
L
RL
L
RN
N
RN
N
RE
PE
PE
Rb
Rb
Ra
600 V CAT III
600 V CAT III
/
/
Case of an IT installation
RL
L
RN
N
PE
Z
Ra
600 V CAT III
/
The alarm, if activated, serves to inform the user, by an audible signal, that the measurement is above threshold, making it unnecessary to look at the display unit to check this point.
The signal can be smoothed to produce a mean of values. But the measurement then takes longer.
38
3.9.3. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can configure it by modifying the parameters displayed:
To select measurement of Zi (line impedance measurement) or of ∆V (measurement of the voltage drop in the cables,
for the C.A. 6117 only). Here, you must select Zi.
To compensate for the resistance of the measurement leads, for measurements of low values (see §3.16).
To activate or deactivate the smoothing of the signal.
(Ik)
The device proposes choosing the voltage for the Ik calculation from among the following values:
■ ULN (the measured voltage value),
■ the voltage of the old standard (for example 220 V),
■ the voltage of the current standard (for example 230 V).
Depending on the voltage ULN measured, the device proposes the following choices:
■ if 170<ULN<270 V: ULN, 220 V, or 230 V.
■ if 90<ULN<150 V: ULN, 110 V or 127 V.
■ if 300<ULN<500 V: ULN, 380 V or 400 V.
To deactivate the alarm.
Z-R
To activate the alarm on Zi.
Ω
kΩ
Ik
050.00
To activate the alarm on Ik.
A 010.00
kA
ISC
To set the alarm threshold (see § 3.17). As default,
the threshold is set to 50 Ω.
To set the alarm threshold (see §3.17). As default,
the threshold is set to 10 kA.
To activate the alarm on Isc, a current that may help in choosing a fuse (for the C.A 6117
only).
To enter the table of fuses.
39
You can then choose:
■ The delay (the duration of application of IN before the fuse blows): 0.1s, 0.2s, 0.4s,
5s, or 35ms.
■ The type of fuse: gG, LSB, LSC or LSD.
■ The rated current IN: any standardized value between 2 and 1000A.
The choices available depend on the choices already made. As does the value of Isc.
Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
6%

Press the TEST button to start the measurement. The measurement stops automatically.
When the TEST button is pressed, the device checks that the contact voltage is less than UL. If not, it does not make
the loop impedance measurement.
This symbol invites you to wait while the measurement is in progress.
TEST
If Ik is less than Isc, the fuse is not suited to the installation it protects and must be replaced.
40
3.9.4. READING OF THE RESULT
Value of the alarm threshold.
11-25-2013 10:47
50 . 0
Ω
50 . 1 Hz
L
Value of the short-circuit current.
Value of the impedance.
Ik
1316 A
Z i
0.29
Ω
R i
0.15
Ω
L i
0.8 mH
Value of the resistance.
Value of the inductance.
Case where the measurement is
below the alarm threshold.
To change display pages.
Value of the reference voltage for the
calculation of Ik.
5%
Programmed maximum contact
voltage.
LOOP Zi
Compensation for the resistance of
the measurement leads is activated.
3.9.5. ERROR INDICATION
See §3.8.5.
41
3.10. MEASUREMENT OF THE VOLTAGE DROP IN THE CABLES (ΔV)
For the C.A. 6117 only. The voltage drop in the cables is measured to check that the cross section of the cables is sufficient for the
installation. A voltage drop that is too large (> 5%) means that the cross section of the cables is too small.
This measurement can be made whatever the type of neutral used in the installation.
3.10.1. DESCRIPTION OF THE MEASUREMENT PRINCIPLE
The device makes a first measurement of Zi at a reference point, then a second measurement of Zi at the measurement point.
The voltage drop is then calculated: ∆V = 100 (Zi - Zi ref ) x IN / UREF. IN is the rated current of the fuse that protects the installation.
The result is expressed in %.
3.10.2. MAKING A MEASUREMENT
You must make two measurements.
For the first, connect the tripod cable/3 safety leads to the instrument. Operate just
after the fuse that protects the installation. Connect the L cord (red) to the phase and
the N cord (blue) to the neutral. Connect the PE cord (green) to the N cord (blue).
Set the switch to Zi.
SET UP
At the time of connection, the device first checks that the voltages present on
its terminals are correct, then determines the position of the phase (L) and of
the neutral (N) with respect to the protective conductor (PE) and displays it.
If necessary, it then automatically switches the L and N terminals so that the line
impedance measurement can be made without modifying the connections of the
terminals of the device.
OFF
RL
L
F
N
RN
PE
Rb
Ra
600 V CAT III
/
You can make the first measurement as many times as necessary. When you are satisfied with it, you can enter it as reference by
pressing the
key. You can also start from a zero reference by pressing the
key without first making a measurement. When
.
a reference has been entered, the key becomes
For the second measurement, connect the tripod cable to the instrument and to one of the outlets of the installation.
RL
L
F
N
RN
PE
Rb
Ra
600 V CAT III
/
Here again, you can make as many measurements as necessary, always with the same first measurement as reference. And you
can record the result each time.
42
The alarm, if activated, serves to inform the user, by an audible signal, that the measurement is above threshold, making it unnecessary to look at the display unit to check this point.
The signal can be smoothed to produce a mean of values. But the measurement then takes longer.
For this measurement, it is not necessary to connect the PE terminal.
3.10.3. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can configure it by modifying the parameters displayed:
To select the measurement of Zi (line measurement impedance) or of ∆V (measurement of the voltage drop in the
cables). Here, you must select ∆V.
Indicates whether a first measurement has already been entered as reference. If the symbol is greyed out, this is
not the case. Otherwise, the reference value is indicated.
2A
(Ik)
Can be used to specify the characteristics of the fuse by entering the table of fuses.
■ Choice of delay (the duration of application of IN before the fuse blows): 0,1 s, 0,2 s, 0,4 s, 5 s
and 35 ms.
■ Choice of type of fuse: gG, LSB, LSC or LSD.
■ Choice of rated current IN: any standardized value between 2 and 1000 A.
The choices available depend on the choices already made. As does the value of Isc.
The device proposes choosing the voltage for the Ik calculation from among the following values:
■ ULN (the measured voltage value),
■ the voltage of the old standard (for example 220 V),
■ the voltage of the current standard (for example 230 V).
Depending on the voltage ULN measured, the device proposes the following choices:
■ if 170<ULN<270 V: ULN, 220 V, or 230 V.
■ if 90<ULN<150 V: ULN, 110 V or 127 V.
■ if 300<ULN<500 V: ULN, 380 V or 400 V.
To deactivate the alarm.
∆V
To activate the alarm on ∆V.
%
To adjust the alarm threshold (see § 3.17). The default
threshold is 5%.
Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
6%

TEST
5.00
The percentage indicates the quantity of memory already used.
Press the TEST button to start the measurement. The measurement stops automatically.
When the TEST button is pressed, the device checks that the contact voltage is less than UL. If not, it does not make
the loop impedance measurement.
This symbol invites you to wait while the measurement is in progress.
If Ik is greater than Isc, the fuse is unsuited to the installation it protects and must be replaced.
43
3.10.4. READING OF THE RESULT
After the first measurement:
50 . 1 Hz
01/17/2017 10:47
L
Value of the short-circuit current.
Value of the reference impedance.
Ik
Value of the resistance.
1316 A
Z ref
0.836 Ω
R i
0.154 Ω
L i
0.8 mH
Value of the inductance.
key to enter the measPress the
urement as reference.
Value of the reference voltage for the
calculation of Ik.
5%
LOOP Dv.
Programmed maximum contact
voltage.
Compensation for the resistance of
the measurement leads is activated.
Change the connection as explained above and press the TEST button again to make the second measurement.
After the second measurement:
50 . 1 Hz
01/17/2017 10:47
L
Result of the calculation of ∆V.
Value of the reference impedance.
V
0.836
-0.33 %
Z ref
0.836
Ω
Z l ine
0.788
Ω
I n
16.00
A
5%
LOOP Dv.
3.10.5. ERROR INDICATION
See §3.8.5.
44
Value of the 2nd impedance.
Value of the rated current of the fuse.
3.11. TEST OF RESIDUAL CURRENT DEVICE
The device can be used to perform three types of test on RCDs:
■ a tripping test in ramp mode,
■ a tripping test in pulse mode,
■ a non-tripping test.
The test in ramp mode serves to determine the exact value of the tripping current of the RCD.
The test in pulse mode serves to determine the tripping time of the RCD.
The non-tripping test serves to check that the RCD does not trip at a current of 0.5 I∆N. For the test to be valid, the leakage current
must be negligible with respect to 0.5 I∆N and, to ensure this, all loads connected to the installation protected by the RCD that is
being tested must be disconnected.
3.11.1. DESCRIPTION OF THE MEASUREMENT PRINCIPLE
For each of the three types of test, the device starts by checking that the RCD can be tested without compromising the user’s safety,
i.e. without causing the fault voltage, UF , to exceed 50 V (or 25 V or 65 V according to what is defined in the SET-UP for UL ). The
device therefore starts by generating a low current (<0.3 I∆N) in order to measure ZS, as it would for a loop impedance measurement.
It then calculates UF = ZS x I∆N (or UF = ZS x 2 I∆N or UF = ZS x 5 I∆N depending on the type of test requested), which will be the maximum voltage produced during the test. If this voltage is greater than UL, the device does not perform the test. The user can then
reduce the measurement current (to 0.2 I∆N) so that the test current combined with the leakage current present in the installation
will not lead to a voltage greater than UL.
For a more accurate measurement of the fault voltage, we recommend planting an auxiliary rod, as for earth measurements on
live circuits. The device then measures RA and calculates UF = RA x I∆N (or UF = RA x 2 I∆N or UF = ZS x 5 I∆N depending on the type
of test requested).
Once this first part of the measurement has been made, the device goes on to the second part, which depends on the type of test.
■ For the ramp mode test, the device generates a sinusoidal current of which the amplitude increases gradually, in plateaus, from
0.3 to 1.06 I∆N between the L and PE terminals for type AC, A and F RCDs and from 0,2 to 2,2 I∆N for type B, B+ and EV RCDs
(for the C.A 6117 only). When the RCD opens the circuit, the device displays the exact value of the tripping current and the
tripping time. This time is an indication and may differ from the trip time in pulse mode, which is closer to the operational reality.
■ For the pulse mode test, the device generates a sinusoidal current at the mains frequency, having an amplitude of I∆N, 2 I∆N or
5 I∆N between the L and PE terminals for type AC, A and F RCDs and 2 I∆N or 4 I∆N for type B, B+ and EV RCDs (for the C.A
6117 only), lasting at most 500 ms. And it measures the time the RCD takes to open the circuit. This time must be less than
500 ms.
■ For the non-tripping test, the device generates a current of 0.5 I∆N for one or two seconds, depending on what the user has
programmed. Normally, the tripping must not trip.
In the ramp and pulse mode tests, if the RCD does not trip, the device sends a current pulse between the L and N terminals. If the
RCD trips, it is because it was incorrectly installed (N and PE reversed).
3.11.2. PERFORMING A TEST IN RAMP MODE
Set the switch to RCD
SET UP
.
Connect the measuring cable to the device, then to a socket outlet included in the circuit
protected by the RCD to be tested.
At the time of connection, the device detects the positions of the phase (L) and
of neutral (N) with respect to the protective conductor (PE) and displays them. If
necessary, it then automatically switches the L and N terminals so that the test
can be done without modifying the connections of the terminals.
OFF
45
RCD
RL
L
N
RN
PE
Rb
Ra
600 V CAT III
/
If possible, first disconnect all loads from the network on which you test the RCD. This prevents interference with the test
by any leakage currents due to these loads.
If you have a current clamp, you can measure the leakage current (see §3.12) at the RCD and so make allowance for it during the test.
To make a more accurate measurement of the fault voltage, plant the auxiliary rod at a distance of more than 25 metres
symbol is then displayed.
(RA SEL) terminal of the device. The
from the earth electrode and connect it to the
RCD
RL
L
N
RN
PE
Rb
Ra
600 V CAT III
> 25 m
/
Particular case:
To test a residual current device located downstream of another residual current device having a smaller nominal current, you
must use the measuring cable terminated by 3 leads and make the connections shown opposite (upstream-downstream method).
RCD
RL
RCD
L
N
RN
PE
Rb
30 mA
blue
300 mA
green
Ra
600 V CAT III
/
46
red
3.11.3. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can configure it by modifying the parameters displayed:
30 mA
■ Choice of the nominal current of the residual current device I∆N: VAR. (variable: the user programs a value between 6 and 999mA for types AC, A, and F, or a value between 6 and 499 mA for types B, B+, and EV), 6 mA,
10 mA, 30 mA, 100 mA, 300 mA, 500 mA, 650 mA or 1000 mA (except 1000 A for types B, B+, and EV residual
current devices).
Type EV residual current devices must be tested with DC at 6 mA.
■ Choice of type of residual current device: STD (standard), S or G (the S type is tested with a current of 2 I∆N as
default).
■ Choice of the form of the test signal:
signal that starts with a positive alternation (type AC RCDs),
signal that starts with a negative alternation (type AC RCDs),
signal containing only positive alternations (type A or F RCDs),
signal containing only negative alternations (type A or F RCDs),
continuous positive DC signal (type B, B+ or EV RCDs),
continuous negative DC (type B, B+ or EV RCDs).
To restore the factory adjustment parameters: I∆N = 30 mA, STD and signal
0.3 I∆N
types.
To perform a prior check of voltage UF, choose a test current: 0.2, 0.3, 0.4, or 0.5 I∆N .
For type EV residual current devices, or for a quicker measurement, eliminate the preliminary check
of voltage UF by selecting: --x--.
To activate or deactivate the audible voltage alarm (the threshold being equal to UL).
This function makes it possible to locate, on the distribution panel, using the audible signal, the RCD
protecting a remote current socket outlet (typical case of a panel at a distance from the socket outlet)
without being in the immediate vicinity of the device.
6%
Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.

Press the TEST button to start the measurement. The measurement stops automatically.
In the case of type S or G circuit-breakers, the device counts 30 seconds between the prior test of UF and the test of
the RCD itself, in order to allow its demagnetization. This wait can be cut short by pressing the TEST button again.
TEST
This symbol invites you to wait while the measurement is in progress.
47
3.11.4. READING OF THE RESULT
50 . 1 Hz
11-25-2013 10:47
L
30 mA
UF
1.0 7 3 V
I a
2 2.3 m A
Ta
1 3.8 m s
UF = ZS x IA or RA x IA.
Tripping current.
Tripping time.
The measurement results are correct.
To change display pages.
8%
RCD : Ia
Type of signal.
Type of RCD.
Programmed maximum contact
voltage.
48
3.11.5. MAKING A TEST IN PULSE MODE
Set the switch to RCD
Connect the measuring cable to the device, then to a socket outlet included in the circuit
protected by the circuit-breaker to be tested.
.
At the time of connection, the device detects the positions of the phase (L) and
of neutral (N) with respect to the protective conductor (PE) and displays them. If
necessary, it then automatically switches the L and N terminals so that the test
can be made without modifying the connections of the terminals of the device.
SET UP
RCD
RL
L
OFF
N
RN
PE
Rb
Ra
600 V CAT III
/
For a more accurate measurement of the fault voltage, plant the auxiliary rod at a distance of more than 25 metres from
symbol is then displayed.
(RA SEL) terminal of the device. The
the earth electrode and connect it to the
RCD
RL
L
N
RN
PE
Rb
Ra
600 V CAT III
> 25 m
/
Particular case:
To test a residual current device located downstream of another residual current device having a smaller nominal current, you
must use the measuring cable terminated by 3 leads and make the connections shown opposite (upstream-downstream method).
RCD
RL
RCD
L
N
RN
PE
Rb
30 mA
blue
300 mA
green
red
Ra
600 V CAT III
/
If it is active, the alarm on the tripping time informs the user by an audible signal, that the measurement is outside the range limits,
so there is no need to look at the display unit.
A type S RCD is normally tested at 2 I∆N.
The tests at 0.5 I∆N are made with the
waveform.
49
3.11.6. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can configure it by modifying the parameters displayed:
30 mA
■ Choice of the nominal current of the residual current device I∆N: VAR. (variable: the user programs a value between 6 and 999 mA for types AC, A, and F, or a value between 6 and 499 mA for types B, B+, and EV), 6 mA,
10 mA, 30 mA, 100 mA, 300 mA, 500 mA, 650 mA or 1000 mA (except 1000 A for types B, B+, and EV residual
current devices).
Type EV residual current devices must be tested with DC at 6 mA.
■ Choice of type of residual current device: STD (standard), S or G (the S type is tested with a current of 2 I∆N as
default)
■ Choice of pulse current in multiples of I∆N : x1, x2, x4, x5, x0,5/1s, x0,5/2s or UF . The 2 values at 0.5 I∆N are used
to perform a non-tripping test.
■ Choice of the form of the test signal:
signal that starts with a positive alternation (type AC RCDs),
signal that starts with a negative alternation (type AC RCDs),
signal containing only positive alternations (type A or F RCDs),
signal containing only negative alternations (type A or F RCDs),
continuous positive DC (type B, B+ or EV RCDs, current x2 or x4),
continuous negative DC (type B, B+ or EV RCDs, current x2 or x4).
Depending on the type of fuse and the form of the test signal, only some values of the pulse current are possible.
To restore the factory adjustment parameters: I∆N = 30 mA, STD type RCD, pulse current = I∆N and
signal.
0.3 I∆N
To perform a prior check of voltage UF, choose a test current: 0.2, 0.3, 0.4, or 0.5 I∆N.
For type EV residual current devices, or for a quicker measurement, eliminate the preliminary check
of voltage UF by selecting: --x--.
To deactivate the alarm.
TAmin
To program an alarm on the minimum tripping time.
TAmax
To program an alarm on the maximum tripping time.
TAmin/TAmax
To program an alarm on the minimum tripping time and on the maximum
tripping time (see §3.17).
The default TA min is 0 ms.
The default TA max is 300 ms for standard differentials (STD) and 500 ms for selective differentials
(S or G).
50
6%

Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
Press the TEST button to start the measurement. The measurement stops automatically.
In the case of type S or G RCD, the device counts 30 seconds between the prior test of UF and the test of the RCD
itself, in order to allow its demagnetization. This wait can be cut short by pressing the TEST button again.
This symbol invites you to wait while the measurement is in progress.
TEST
3.11.7. READING OF THE RESULT
■ In the case of a test in pulse mode with tripping:
Value of the alarm threshold.
200 ms
11-25-2013 10:47
50 . 1 Hz
L
30 mA
UF
1.1 4 6 V
Ta
UF = ZS x IA or RA x IA.
for type AC and A RCDs. UF is not
calculated for type B RCDs.
Trip time.
1 7 1.6 m s
Case where: TAmin < TA < TAmax.
To change display pages.
8%
RCD : Ta TRIP
Type of signal.
x2
Type of RCD.
Programmed maximum contact
voltage.
Pulse current as a multiple of I∆N.
51
■ In the case of a non-tripping test in pulse mode:
11-25-2013 10:47
50 . 1 Hz
L
30 mA
UF = ZS x IA or RA x IA.
UF
0.1 4 6 V
Ta
The RCD did not trip out during the
duration of application of the current
of 0.5 I∆N.
> 1.0 0 s
To change display pages.
8%
RCD : Ta NO TRIP
Type of signal.
x0.5/1
Type of RCD.
Programmed maximum contact
voltage.
Non-tripping test lasting one second.
3.11.8. ERROR INDICATION
The commonest errors in the case of a test of a residual current device are:
■ The RCD did not trip out during the test. Now, to ensure the safety of users, a RCD must trip within 300 ms, or 200 ms for a
type S. Check the wiring of the RCD. If it is OK, the RCD itself must be declared defective and replaced.
■ The RCD trips out when it should not. The leakage currents are probably too high. First disconnect all loads from the network
on which you are performing the test. Then perform a second test with the current reduced (in UF check) as far as possible. If
the problem persists, the RCD must be declared defective.

For help with connections or any other information, use the help function.
52
3.12. CURRENT AND LEAKAGE CURRENT MEASUREMENT
This measurement requires the use of a specific optional current clamp.
It can measure very low currents (of the order of a few mA) like fault currents or leakage currents, and high currents (of the order
of a few hundred Amperes).
3.12.1. DESCRIPTION OF THE MEASUREMENT PRINCIPLE
The specific clamps operate on the current transformer principle: the primary is constituted by the conductor in which the current is
to be measured, while the secondary is constituted by the internal winding of the clamp. This winding is itself closed through a resistance having a very low value, located in the device. The voltage across the terminals of this resistance is measured by the device.
Two of the four points of connection of the clamp serve to identify the type of clamp (x 1,000 or x 10,000) and the other two to measure
the current. Knowing the ratio of the clamp, the device displays a direct reading of the current.
3.12.2. MAKING A MEASUREMENT
Set the switch to
.
Connect the clamp to the
terminal on the device. The
symbol is then displayed.
Actuate the trigger to open the clamp and encircle the conductor to be measured.
Release the trigger.
SET UP
I
OFF
600 V CAT III
/
The current measurement can be made on different conductors of an installation. This is why it has been made possible to index
the value recorded with one of the following values:
1, 2, 3, N, PE, or 3L (sum of the phase currents or phase and neutral currents, to measure the leakage current).
3.12.3. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can program an alarm:
To deactivate the alarm.
To activate the alarm.
m A 200.0
A
6%

To set the alarm threshold (see §3.17). As default, the threshold is set to 200 A.
Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
Press the TEST button once to start the measurement and a second time to stop it.
TEST
53
3.12.4. READING OF THE RESULT
010 . 0 A
11-25-2013 10:47
50 . 1 Hz
Value of the alarm threshold.
1 9 7.3 m A
Measurement result.
Case where the measurement is
below the alarm threshold.
9%
CURRENT
The clamp is connected.
3.12.5. ERROR INDICATION
The commonest errors in the case of a current measurement are:
■ The clamp is not connected.
■ The current measured by the clamp is too low. Use a clamp having a lower ratio or pass the conductor through the clamp
several times to increase the measured current.
Here, the conductor passes through the clamp 4 times. You will
have to divide the measured current by 4 to know the true value of I.
■ The frequency is too unstable for the measurement. In this
I
red
600 V CAT III
green
/
case connect the corresponding mains voltage between L and PE. The device will
then synchronize to the frequency of the voltage and will be able to measure the
current at this same frequency.
blue
I
600 V CAT III
/
■ The current measured by the clamp is too high. Use a clamp having a higher ratio.

For help with connections or any other information, use the help function.
54
3.13. DIRECTION OF PHASE ROTATION
This measurement is made on a three-phase network. It is used to check the phase order of the network.
3.13.1. DESCRIPTION OF THE MEASUREMENT PRINCIPLE
The device checks that the three signals are at the same frequency, then compares the phases to determine their order (direct or
reverse direction).
3.13.2. MAKING A MEASUREMENT
Set the switch to
.
Connect the measuring cable terminated by 3 leads to the device and to each of the phases:
the red to L1, the blue to L2, and the green to L3.
SET UP
L1
L2
L3
red
blue
green
N
PE
OFF
600 V CAT III
/
There are no parameters to program before starting the measurement.

Press the TEST button once to start the measurement and a second time to stop it.
TEST
3.13.3. READING OF THE RESULT
11-25-2013 10:47
50 . 1 Hz
The + sign indicates a forward
direction and the - sign a reverse
direction.
U12: 379 V
U23: 380 V
U31: 378 V
10 %
PHASE SEQUENCE
55
Voltages between the phases.
6%
Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
3.13.4. ERROR INDICATION
The commonest errors in the case of a test of direction of phase rotation are:
■ One of the three voltages is outside the measurement range (connection error).
■ The frequency is outside the measurement range.

For help with connections or any other information, use the help function.
56
3.14. POWER MEASUREMENT
This measurement requires the use of the optional specific C177A current clamp. It can be made on a single-phase network or on
a three-phase network that is balanced in voltage and in current.
3.14.1. DESCRIPTION OF THE MEASUREMENT PRINCIPLE
For a single-phase network, the device measures the voltage between the L and PE terminals, then multiplies it by the current measured by the clamp.
For a three-phase network balanced in voltage and in current, the device measures one of the three phase-to-phase voltages, multiplies it by the current of the third phase, then multiplies the result by √3. Example: P3Φ = U12 x I3 x √3
3.14.2. MAKING A MEASUREMENT
Set the switch to W.
In the case of a single-phase network, connect the measuring cable terminated by 3 leads
to the device and to a socket outlet of the installation to be tested, using the red and blue
leads. With the clamp, surround either the phase conductor, to obtain the total power, or the
conductor of one of the loads, to obtain the partial power.
SET UP
L
red
N
blue
green
PE
OFF
600 V CAT III
/
In the case of a three-phase network balanced in voltage and in current, connect the measuring cable terminated by 3 leads to the
device and to two of the three voltages U12, U23 or U31 using the red and blue leads. Then, with the clamp, surround the conductor
of the third phase I3 (for U12), I1 (for U23) or I2 (for U31).
red
L1
L2
L3
blue
green
600 V CAT III
/
The power measurement can be made on different phases of an installation. This is why it has been made possible to index the
recorded power value with one of the following values: 1, 2, or 3 (single-phase measurements on a three-phase network).
3.14.3. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can configure it by modifying the parameters displayed:
Choice of type of network: single-phase or balanced three-phase.
6%

Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
Press the TEST button once to start the measurement and a second time to stop it.
TEST
57
3.14.4. READING OF THE RESULT
50 . 1 Hz
11-25-2013 10:47
+
64
U
232.5
I
2 7 8 . 1 mA
PF
+
W
Measurement result.
The + sign indicates power consumed. The - sign indicates power
returned.
Voltage between the L and PE terminals.
V
Current measured by the clamp.
0.87
Power factor.
The + sign indicates that the load
is resistive or inductive. The - sign
indicates that the load is capacitive.
12 %
POWER
The clamp is connected.
If the phase of the current with respect to the voltage is not correct, turn the clamp around, with the help of the arrow marked on
the jaws, in order to reverse the phase by 180°.
Press this function key to display the voltage and current curves, as on an oscilloscope. If the clamp is not connected,
only the voltage curve is displayed. The current curve cannot be displayed alone.
The representation of the curves is normalized:
■ in amplitude, the curves are automatically adjusted to fill the screen.
■ on the time scale, approximately one period is shown.
11-25-2013 10:47
50 . 1 Hz
URMS = 232.5 V
UMAX = 328.8 V
IRMS = 278.1 mA
IMAX = 393.3 mA
Numerical values.
Voltage curve (blue).
Current curve (red).
POWER
To move the key if it masks part of the curves.
58
3.14.5. POWER FACTOR
In the case of sinusoidal signals, the sign of cos j indicates whether the measurement is being made on a generator (cos j < 0) or
on a receiver (cos φ > 0). The power factor, PF, can be regarded as equivalent to cos φ but generalized to non-sinusoidal signals,
which is often the case with currents.
However, on the instrument, the sign of the PF is treated conventionally, meaning that it indicates only the phase advance or delay
(inductive or capacitive load) and not whether a receiver or a generator is involved.
The phase angle is counted algebraically. It represents the angular difference of the voltage vector with respect to the current
vector, taken as reference.
i(t)
i
ϕ
time
V(t)
ϕ
+
V
Indications provided by the instrument
Mean power 1
Sign of the PF
Phase[V(t);i(t)]
Type of equipment
Reactive component
-180° < φ < -90°
Generator
inductive
Negative
Positive
- 90° < φ < 0°
Receiver
capacitive
Positive
Negative
0° < φ < +90°
Receiver
inductive
Positive
Positive
+90° < φ < +180°
Generator
capacitive
Negative
Negative
1: under the receiver convention.
3.14.6. ERROR INDICATION
The commonest errors in the case of a power measurement are:
■ The voltage is outside the measurement range.
■ The frequency is outside the measurement range.
■ The current is too low to be measured.
■ The power measured is negative. Check that the clamp is correctly placed on the cable (look at the direction of the arrow). If
it is it means that, what you are measuring is power returned (from receiver to generator).

For help with connections or any other information, use the help function.
59
3.15. HARMONICS
This function is used to display the harmonic analysis of a voltage or current of which the signal is steady-state or quasi-steadystate. It is used to establish a first diagnostic of the harmonic pollution of an installation.
The current analysis requires the use of the C177A current clamp (optional).
3.15.1. DESCRIPTION OF THE MEASUREMENT PRINCIPLE
The device measures the voltage and, if the clamp is connected, the current. Then, depending on what the user has chosen (FFT U
or FFT I), it performs an FFT limited to the first 50 harmonics either of the voltage or of the current. Harmonic 0 (the DC component)
is not displayed.
3.15.2. MAKING A MEASUREMENT
Set the switch to
Connect the measuring cable terminated by 3 leads to the device and to a socket outlet of the
installation to be tested, using the red and blue leads.
Or connect the C177A clamp to the device and encircle the phase.
.
SET UP
L
red
N
blue
green
PE
OFF
600 V CAT III
/
3.15.3. CONFIGURING THE MEASUREMENT
Before starting the measurement, you can configure it by modifying the parameters displayed:
To choose to perform an FFT on the voltage (U) or on the current (I).
To choose the display format for the FFT:
LIN
LOG
H_RMS
linear scale,
logarithmic scale,
result in the form of an alphanumeric list.
F
Choice of calculation of the level of distortion with respect to the fundamental (THD-F) or of the distortion factor with
respect to the RMS amplitude (THD-R or DF).
6%
Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.

Press the TEST button once to start the measurement and a second time to stop it.
TEST
60
3.15.4. READING OF THE RESULT
50 . 0 Hz
01-23-2013 10:47
THDF =
2.8 %
ULPE = 225.9 V
Display of the THD-F and of the RMS
voltage.
100
log
10
1
Representation of the harmonics.
F
225.3 V
100.0 %
H1
13 %
Indication of the name of the selected
spike and of its amplitude.
HARMONICS U
The frequency and amplitude of the selected harmonic (in green) are indicated at the bottom of the graph. To select another harmonic, use the ◄ ► keys. The device then shifts from the fundamental (H1) to harmonic H2, then to harmonics (H3, H4, ..., H25).
And on the next page it sweeps the harmonics from H26 to H50.
Frequency F1 is displayed on the top strip of the display unit.
The frequency of harmonic Hn is n x F1.
The display in list form gives the following screen:
50 . 0 Hz
11-25-2013 10:47
H%RMS
F
T H D _ F :
2 . 5 %
U _ R M S :
2 2 6 . 2 V
H 0 1 :
2 2 5 . 8 V
1 0 0 . 0 %
H 0 2 :
0 . 4 V
0 . 2 %
H 0 3 :
2 . 1 V
0 . 9 %
H 0 4 :
0 . 1 V
0 . 1 %
H 0 5 :
3 . 7 V
1 . 7 %
H 0 6 :
0 . 1 V
0 . 1 %
Display of the THD-F and of the
voltage.
For each harmonic, its amplitude and
the percentage with respect to the
amplitude of the fundamental (H01)
are specified.
To see the rest of the display of the
measurement.
13 %
HARMONICS U
You must scroll through 6 other screens using the ► key to display the values of all 50 harmonics.
61
3.15.5. ERROR INDICATION
The commonest errors in the case of an analysis of a signal into harmonics are:
■ The voltage is outside the measurement range.
■ The frequency is outside the measurement range.
■ The current is too low to be measured.
■ The signal is not steady-state.

For help with connections or any other information, use the help function.
62
3.16. COMPENSATION FOR THE RESISTANCE OF THE MEASUREMENT LEADS
Compensation for the resistance of the measurement leads serves to neutralize their values and obtain a more accurate measurement when the resistance to be measured is low. The cords are already compensated in the plant; you must perform a new
compensation if you use cords other than those provided.
The device measures the resistance of the accessories (leads, probe tips, crocodile clips, etc.) and subtracts this value from the
measurements before displaying them.
Compensation for the resistance of the measurement leads is possible in continuity, 3P earth, and loop tests. It is different for each
of these functions. It must be renewed at each change of accessories.
Press the
key to enter the function.
The current value(s) of the compensation is(are) displayed at top right. A value of zero indicates that no compensation has
been determined. The R∆
0 symbol, present on the bottom strip of the display unit, reminds you that the resistance of the
leads is compensated.
3.16.1. IN CONTINUITY
R∆:
0.052
Ω
Connect the two leads that you are going to use for the measurement to
the Ω and COM terminals, short-circuit them, then press the TEST button.
The device measures the resistance of the leads and displays it. Press OK
to use this value or
to keep the old value.
3.16.2. IN 3P EARTH
R∆:
0.750
Ω
Connect the lead that you are going to use to connect the E terminal to the
earth between the H and E terminals, then press the TEST button.
The device measures the lead and displays its value. Press OK to use this
to keep the old value.
value or
3.16.3. IN LOOP (ZS OR Zi)
R∆E :
R∆N :
R∆L :
0.127
0.105
0.114
Ω
Ω
Ω
Connect the three leads that you are going to use for the measurement to
the L, N, and PE terminals, short-circuit them, then press the TEST button.
The device measures each of the three leads and displays their values.
Press OK to use this value or
63
to keep the old values.
3.16.4. ELIMINATING THE COMPENSATION
Proceed as for compensation, but rather than short-circuiting the leads, leave them disconnected. Then press the TEST button.
The device removes the compensation,
RΔL = RΔN = RΔE = 0,030 Ω.
R∆
then returns to voltage measurement. The 0 symbol disappears from the display unit and the
icon is crossed out.
3.16.5. ERROR
■ If the resistance of the measurement leads is too high (>2.5 Ω per lead), compensation is impossible. Check the connections
and any junctions and leads that might be open-circuit.
■ If, during a continuity, 3P earth, or loop impedance measurement, you obtain a negative measurement result, you must have
changed the accessories without redoing the compensation. In this case, perform a compensation with the accessories you
are now using.
64
3.17. ADJUSTMENT OF THE ALARM THRESHOLD
The device makes an audible signal and the indicator flashes:
■ in continuity, resistance and insulation measurement, if the measurement is below threshold;
■ for earth and loop measurements and measurements of the voltage drop in the cables, if the measurement is above threshold;
■ for short-circuit current measurements, if the measurement is below threshold;
■ in test of residual current device, if the measurement is not between the two thresholds (Tmin and Tmax).
In continuity measurement, the audible signal is used to validate the measurement.
In all the others functions, it reports an error.
The alarm threshold is adjusted in essentially the same way for all measurements.
Start by entering the alarm function by pressing the
or
key.
If the alarm is not active, press the▼
key to activate it.
Ω 002.00
kΩ
Using the ► key, move the cursor to the units.
Ω 002.00
kΩ
Using the ▲ ▼ keys, choose the unit of the alarm threshold you want to set: Ω or kΩ. Depending
on the function chosen, MΩ, mA, A, kA, and ms are also possible.
Ω 002.00
kΩ
Using the ► key, move the cursor to the value of the threshold.
Ω 042.00
kΩ
Using the ▲ ▼ keys, modify the selected digit. Then move the cursor to the next digit to modify
it, and so on.

OK
To validate the modified threshold, press the OK key.
To abort without saving, press the
key or turn the switch.
65
4. ERROR INDICATION
Generally, errors are reported in clear language on screen.
Example of error screen:

Press the OK key to erase the message.
OK

Or press the help key for help in solving your problem.
The following screen is then displayed.
Press the OK key or the
key to exit from the help function.
66
4.1. NO CONNECTION
One or more terminals are not connected.
4.2. OUT OF MEASUREMENT RANGE
>40.0Ω
< 5.0 V
}
The value is outside the measurement range of the device. The minimum and maximum
values depend on the function.
4.3. PRESENCE OF DANGEROUS VOLTAGE
The voltage is regarded as dangerous from 25, 50, or 65V, depending on the value of UL programmed in SET-UP.
For measurements made without voltage (continuity, insulation, and 3P earth), if the device detects a voltage, it
disables starting of the measurement by the pressing of the TEST button and displays an explanatory error message.
For measurements that are made on live circuits, the device detects the absence of voltage, the absence of a protective conductor, a frequency or voltage outside the measurement range. When the TEST button is pressed, the device then disables starting
of the measurement by the pressing of the TEST button and displays an explanatory error message
4.4. INVALID MEASUREMENT
If the device detects an error in the measurement configuration or in the connection, it displays this symbol and a
corresponding error message.
4.5. DEVICE TOO HOT
The internal temperature of the device is too high. Wait for
the device to cool off before making another measurement.
This case concerns essentially the test of residual current
devices.
67
4.6. CHECK OF INTERNAL PROTECTION DEVICES
The device includes two internal protection devices that cannot be reset and that the user cannot replace. These devices act only
under extreme conditions (e.g. a lightning strike).
To check the condition of these protections:
Disconnect the input terminals.
Set the switch to ZS (RA/SEL.).
SET UP
OFF
600 V CAT III
/
If the internal protection devices are intact, the display should indicate:
50 .0 Ω
11-25-2013 10:47
6 mA
50 . 1 Hz
L-PE
V
L-N
V
N-PE
V
6%
LOOP Zs
If UL-PE does not display – x --, the
protection in the L terminal has been
activated.
11-25-2013 10:47
6 mA
50 .0 Ω
50 . 1 Hz
L-PE
0.0 V
L-N
N-PE
If UN-PE does not display – x --, the
protection in the N terminal has been
activated.
11-25-2013 10:47
50 .0 Ω
6 mA
50 . 1 Hz
L-PE
V
V
L-N
V
V
N-PE
Case where both protections have
been activated.
11-25-2013 10:47
6 mA
0.0 V
6%
6%
6%
LOOP Zs
LOOP Zs
LOOP Zs
In these last three cases, the device must be sent in for repair.
68
50 .0 Ω
50 . 1 Hz
L-PE
0.0 V
L-N
0.0 V
N-PE
0.0 V
5. SET-UP
Set the switch to SET-UP.
SET UP
OFF
OK
Use the directional keypad to select an icon, select a field, and modify it.
This key is used to exit from the current screen without saving.
Used to display all parameters of the device:
■ the software version (internal to the device),
■ the hardware version (of the internal boards and components of the device),
■ the date format,
■ the time format,
■ activation of the audible signal,
■ the serial number,
next page
■ the duration of operation of the device before automatic switching off,
■ the language.
To set the date and time and choose the display format.
To activate or deactivate the audible signal.
To set the contact voltage to 25 V, 50 V (default), or 65 V.
■ 50 V is the standard voltage (default).
■ 25 V should be used for measurements in a damp environment.
■ 65 V is the default voltage in some countries (Austria, for example).
69
Adjustment of the time to automatic switching off of the device: 5 min (default), 10 min, 30 min, or ∞ (permanent operation).
Used to access the memory to:
■ read the measurements already made,
■ or prepare a tree before a measurement campaign.
See storage in §6.
To erase all of the memory in tree mode.
The device requests confirmation before erasing all memory then formatting in tree mode.
To erase all of the memory in table mode.
The device requests confirmation before erasing all memory then formatting in table mode.
To return to the factory configuration (compensation for resistance of measurement leads and all adjustable
parameters in the various measurements). The device requests confirmation before executing.
The default configuration of the device is as follows:
General configuration
■ Audible signal: activated
■ UL = 50 V
■ Duration of lighting of the backlighting: 2 min.
■ Duration of operation of the device before automatic switching off: 5 min.
■ Date and time format: DD/MM/YYYY and 24 h.
■ Language: English.
The memorization is not affected by the return to the plant configuration.
Resistance and continuity measurement
■ Measurement mode: permanent.
■ Measurement current: 200 mA.
■ Polarity of the current: duplex
■ Compensation of the measurement leads: 60 mΩ.
■ Alarm activated.
■ Alarm threshold: 2 Ω.
Insulation measurement
■ Test voltage: 500 V.
■ Alarm activated.
■ Alarm threshold: 1 MΩ.
3P earth resistance measurement
■ Simple measurement (no measurement of the rods).
■ Compensation of the measurement lead RE = 30 mΩ.
■ Alarm activated.
■ Alarm threshold: 50 Ω.
Measurement of loop impedance (ZS), of earth on live circuit, and of selective earth resistance on
live circuit
■ Measurement current: 6 mA.
■ Compensation of the cords: 30 mΩ, 30 mΩ, 30 mΩ respectively for R∆L, R∆N, R∆PE (measuring cable with
mains plug).
■ UREF = UMEAS.
■ Alarm deactivated.
■ No smoothing of the measurement.
Line impedance measurement (Zi)
■ Compensation of the leads: 30 mΩ, 30 mΩ respectively for R∆L, R∆N (measuring cable cord with mains plug).
■ UREF = UMEAS.
■ Alarm deactivated.
■ No smoothing of the measurement.
70
∆V)
Measurement of the voltage drop in the cables (∆
■ Alarm activated.
■ Alarm threshold: 5%.
Test of RCD
■ Nominal range I∆N = 30 mA.
■ Type of circuit-breaker: Standard (STD).
■ Test waveform: sinusoidal signal that begins with a positive half-wave.
■ Test current for determination of UF = 0.3 I∆N.
■ Alarm deactivated.
■ Audible RCD identification function: deactivated.
Current and leakage current measurement
■ Alarm deactivated.
Direction of phase rotation
■ No configuration.
Power measurement
■ Single-phase network.
Harmonics
No default configuration. Each time the device is started up, the configuration is:
■ Voltage harmonics.
■ Display in bar-chart form with linear ordinates.
■ Calculation of the total distortion referred to the fundamental (THD-F).
To choose the language.
71
6. MEMORY FUNCTION
6.1. CHOICE OF MODE
The memory can be used in 2 different modes:
■ Tree mode
■ Table mode
6.1.1. TREE MODE
In tree mode, the measurements are organized as follows:
SITE 1
ROOM 1
OBJECT 1
OBJECT 2
ROOM 2
OBJECT 1
SITE 2
ROOM 1
...
Each OBJECT can contain 9 tests of each type (insulation, earth measurement, RCD test, etc.).
6.1.2. TABLE MODE
In table mode, the measurements are organized as follows:
000
001
002
003
...
The objects are numbered from 000 to 999 and each object can contain 130 tests.
6.1.3. CHANGING MODES
As default, the memory operates in tree mode. To change to table mode, erase the memory and format it in SET-UP (see § 5),
using the icon:
Changing modes requires completely erasing the memory. Take care to back up your measurements before this operation.
To return to tree mode, erase the memory and format it in SET-UP (see § 5), using the icon:
Each time you erase the memory, use the icon that matches the mode chosen, tree or table.
72
6.2. TREE MODE
6.2.1. ORGANIZATION OF THE MEMORY AND NAVIGATION
The device has 1000 memory locations to record measurements. They are organized in a tree on three levels, as follows:
SITE 1
ROOM 1
OBJECT 1

OBJECT 2

ROOM 2
OBJECT 1

SITE 2
ROOM 1
...
Navigation in the tree is done using the directional keypad. The titles of the SITES, ROOMS, and OBJECTS can be parameterized
by the user.
If a SITE or ROOM is preceded by the sign, it means that this level has sub-levels that can be expanded using the ► key or the
OK key. The
sign is then replaced by the
sign.
To compress the tree (change from the
sign to the
sign), use the ◄ or OK key.
Measurements are always recorded on an OBJECT. In the OBJECT, measurements are classified by TYPE OF TEST (continuity,
insulation, loop, etc.). Each OBJECT can contain up to nine TESTS belonging to the same TYPE OF TEST. Each TEST corresponds to one measurement.
To see the tests contained in an OBJECT, go to the OBJECT and press the OK key.
A status symbol displayed to the right of the OBJECTS, of the TYPES OF TEST, and of the TEST indicates:
 that the OBJECT has not yet been tested,
R that all TESTS of the OBJECT are OK,
T that at least one TEST of the OBJECT is not OK.
6.2.2. ENTERING THE STORAGE FUNCTION
When a measurement is over, the device proposes recording it by displaying the recording icon (arrow pointing in) at bottom left
of the measurement results:
The percentage indicates the level of occupancy of the memory.
If you want to record the measurement you have just made, press the key corresponding to the record icon.
For a measurement to be «recordable», the TEST button must have been pressed. It is not possible to record voltage
measurements alone.
The device displays the following message:
73
The following screen then appears:
500 kΩ
11-25-2013 10:47
- - .- Hz
Position in the tree.
Site1
To create a new SITE.
To create a ROOM in a SITE or an
OBJECT in a ROOM.
To delete an element.
To exit from the memory function.
INSULATION
6.2.3. CREATE A TREE
As default, the device proposes the beginning of a tree (SITE1, ROOM1, OBJECT1). If you do not want to create a tree, this lets
you record all of your measurements in OBJECT1 or change to table mode.
To expand the tree, use the ► key or the OK key.
To create a new SITE, press the
key. The name entry screen is displayed.
500 kΩ
11-25-2013 10:47
- - .- Hz
Erase a letter.
Path in the tree.
/m e m
Change between
upper- and lowercase.
Name of the new
SITE.
Site
1
Save and exit.
Alphanumeric
keyboard.
Exit without saving.
INSULATION
You can then rename the SITE. Start by erasing the existing text. Then move about on the keyboard using the directional keypad
( ▲▼◄ ► ) and validate each letter by pressing the OK key.
A sustained press on one of the ▲▼◄ ► keys speeds up the scrolling
74
To add a ROOM to a SITE, place the cursor on the chosen SITE and press the
key. Give the ROOM a name and validate it.
Then press the key again to create an OBJECT in the ROOM. This results in the following tree:
500 kΩ
11-25-2013 10:47
Site1
House
Room1
- - .- Hz
Object1
Entry
Switchboard
INSULATION
To avoid losing time when you are making the measurements, you can prepare your tree in advance.
6.2.4. RECORD THE MEASUREMENT
To record the measurement, place the cursor on the desired OBJECT and press the OK key.
For insulation, loop impedance, line impedance, current, and power measurements and the harmonic analysis, the device proposes
indexing your measurement, because several measurements are possible.
01/24/2017 10:47
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
500 kΩ
- - .- Hz
LN - PE
L-N
L - PE
N - PE
L1 - PE
L2 - PE
L3 - PE
L1 - N
L2 - N
L3 - N
L1 - L2
L2 - L3
L1 - L3
INSULATION
Using the ▲▼ arrows, select the type of insulation measurement you have just made and validate by pressing the OK key.
75
You can in this way make several insulation measurements on the electrical panel. And then move on to another type of measurement, still on the electrical panel, for example a loop impedance measurement.
100
11/24/2017 10:47
LOOP
Zs
LOOP
LOOP
LOOP
LOOP
L - PE
L1 - PE
L2 - PE
L3 - PE
Ω
50 . 1 Hz
As in the case of insulation, you can
index the measurement.
LOOP Zs
6.2.5. READ THE RECORDS
You can read the measurement made by pressing the 6 % key (arrow pointing out). The device then displays the tree again. The
last OBJECT on which a measurement has been recorded is selected.
To change levels in the tree, use the ◄ and ► keys.
To move on the same level (from SITE to SITE, ROOM to ROOM, or OBJECT to OBJECT), use the ▲▼ keys.
To see all of the measurements made on the selected OBJECT, press the OK key.
11-25-2013 10:47
500 kΩ
- - .- Hz
House / Entry / Switchboard
INSULATION
Zs (LOOP)
Insulation resistance measurements
have been made on this OBJECT
and at least one of them is not OK.
Loop impedance measurements
have been made on this OBJECT
and they are all OK.
INSULATION
76
Press the OK key to expand a TYPE OF TEST.
500 kΩ
11-25-2013 10:47
- - .- Hz
House / Entry / Switchboard
INSULATION
INSULAT.
1. 2. INSULAT.
Zs (LOOP)
1. LOOP
2. LOOP
Path in the tree.
L1-PE
L1-N
L1-PE
L2-PE
List of tests performed in the Panel
OBJECT.
INSULATION

Press the OK key again to see the recorded measurement.
OK
500 kΩ
11-25-2013 10:47
- - .- Hz
The top and bottom strips of the
display unit are displayed in reverse
video to clearly differentiate a measurement just made from a reading in
memory.
3 1 . 0 6 MΩ
7
s
2%
INSULAT. L - PE
Press the
Indexed measurement.
key to return to the tree.
77
6.2.6. ERASURE
You can erase a SITE, a ROOM, an OBJECT or a record either when creating the tree or while reading in memory.
Move the cursor onto the element to be erased using the keys of the directional keypad ( ▲▼◄ ► ).
500 kΩ
11-25-2013 10:47
Site1
House
Room1
Entry
Kitchen
- - .- Hz
Object1
Switchboard
Socket1
Socket1
Socket2
Socket3
INSULATION
Press the
key to erase ROOM1. The device asks you to confirm by pressing the OK key or abort by pressing the
key.
If the number of measurements recorded is large, the erasure may last several minutes.
6.2.7. ERRORS
The commonest errors during storage are the following:
■ The name given already exists. Change the name or index it (ROOM1, ROOM2, etc.)
■ The memory is full. You must eliminate at least one OBJECT to be able to record your new measurement.
■ It is not possible to record a measurement in a SITE or a ROOM. You must create an OBJECT in a ROOM or access an
existing OBJECT and record the measurement there.
6.3. TABLE MODE
6.3.1. RECORDING A MEASUREMENT
When a measurement is over, the device proposes recording it, by displaying the record icon (arrow pointing in) at bottom left of
the measurement results:
The percentage indicates the level of occupancy of the memory.
If you want to record the measurement you have just made, press the key next to the record icon.
For a measurement to be «recordable», the TEST button must have been pressed. It is not possible to record voltage
measurements alone.
78
The following screen then appears:
01/17/2017 10:47
500 kΩ
- - .- Hz
000
001
002
003
004
005
006
007
008
009
010
011
012
013
014
015
To move up 100 objects.
To move down 100 objects.
To delete a recording.
To exit from the memory function.
INSULATION
Use the ▲▼ arrows to select the object in which you want to record your measurement and validate by pressing the OK key.
For insulation, loop impedance, line impedance, current, and power measurements and the harmonic analysis, the device proposes
indexing your measurement, because several measurements are possible.
01/24/2017 10:47
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
INSULAT.
500 kΩ
- - .- Hz
LN - PE
L-N
L - PE
N - PE
L1 - PE
L2 - PE
L3 - PE
L1 - N
L2 - N
L3 - N
L1 - L2
L2 - L3
L1 - L3
INSULATION
Using the ▲▼ arrows, select the type of insulation measurement you have just made and validate by pressing the OK key.
You can in this way make several insulation measurements on the electrical panel. And then move on to another type of measurement, still on the electrical panel, for example a loop impedance measurement.
79
6.3.2. READ THE RECORDS
You can read the measurement made by pressing the 6 % key (arrow pointing out). The device then displays the list of objects,
going to the last object in which a measurement was recorded.
500 kΩ
01/17/2017 10:47
- - .- Hz
000
001
002
003
004
005
006
007
008
009
010
011
012
016
017
015
INSULATION
A status symbol displayed to the right of an object indicates:
 that the object does not contain any tests,
 that all tests in the object are OK,
 that at least one test in the object is not OK.
To see all of the measurements made on the selected OBJECT, press the OK key.
01/17/2017 10:47
500 kΩ
- - .- Hz
Insulation resistance measurements
have been made on this object and
at least one of them is not OK.
===> 001
INSULATION
Zs (LOOP)
Loop impedance measurements
have been made on this object and
they are all OK.
INSULATION
80
Press the OK key to expand one type of test.
500 kΩ
01/17/2017 10:47
===> 001
INSULATION
- - .- Hz
Object number.
Zs (LOOP)
INSULAT.
INSULAT.
L1-PE
L1-N
LOOP
LOOP
L1-PE
L2-PE
List of tests performed in the object.
INSULATION
Use the ( ▲▼◄ ► ) keys to select a measurement. Press the OK key again to see the recorded measurement.
500 kΩ
11-25-2013 10:47
- - .- Hz
The top and bottom strips of the
display unit are displayed in reverse
video to clearly differentiate a measurement just made from a reading in
memory.
3 1 . 0 6 MΩ
7
s
2%
Indexed measurement.
INSULAT. L - PE
Press the
key to return to the previous screen.
6.3.3. EFFACEMENT
To erase an object or a recording, select it using the keys of the directional keypad ( ▲▼◄ ► ).
Press the
key. The device asks you to confirm by pressing the OK key or abort by pressing the
If the number of measurements recorded is large, the erasure may last several minutes.
81
key.
6.3.4. ERRORS
When the memory is full, you can no longer record measurements. You must then delete at least one object to be able to record
your new measurement.
82
7. DATA EXPORT SOFTWARE
7.1. FUNCTIONS
The data export software ICT (Installation Controller Transfer) is used to:
■ configure the parameters of the measurements,
■ prepare the tree in memory,
■ export the recorded measurements in an Excel file.
The Dataview software (in option) is used to recover the measurements from the Excel file and present them in the form of a report
conforming to the standard in your country.
7.2. OBTAIN THE ICT SOFTWARE
You can download the latest version from our web site:
www.chauvin-arnoux.com
Go to the Support tab, then Download our software.
Then search on the name of your instrument.
Download the software
7.3. INSTALLING ICT
To install it, run the set-up.exe file, then follow the instructions on screen.
You must have administrator privileges on your PC to install the ICT software.
Do not connect the instrument to the PC until the software and the driver have been installed.
Then, turn the device on by turning the switch to any setting.
Connect the device to the PC using the USB cord provided
with the instrument and removing the cover that protects
the USB port of the device.
SET UP
OFF
When the device is in communication with a PC, it does
nothing else and its keys are inactive. It then displays the
following message:
The data rate is 115,200 Bauds.
To use ICT, refer to the help functions of the software.
Once the USB cord has been disconnected, the device restarts after a few seconds.
83
8. TECHNICAL CHARACTERISTICS
8.1. GENERAL REFERENCE CONDITIONS
Quantity of influence
Reference values
Temperature
20 ± 3 °C
Relative humidity
45 to 55 % HR
Supply voltage
10.6 ± 0.2 V
Frequency
DC and 45 to 65 Hz
Electric field
< 1 V/m
Magnetic field
< 40 A/m
Supply
on battery (mains not connected)
The intrinsic uncertainty is the error defined under the reference conditions.
The operating uncertainty includes the intrinsic uncertainty plus the effects of variation of the quantities of influence (supply voltage, temperature, interference, etc.) as defined in standard IEC-61557.
The device is not designed to operate when the charger is connected. The measurements must be made using the battery.
8.2. ELECTRICAL CHARACTERISTICS
8.2.1. VOLTAGE MEASUREMENTS
Particular reference conditions:
Peak factor = 1.414 in AC (sinusoidal signal)
AC component <0.1% in DC measurement
DC component <0.1% in AC measurement
Voltage measurements (L, N, PE)
Resolution
0.2 - 399.9 V
2.0 - 399.9 V
0.1 V
Intrinsic uncertainty
± (1.5 % + 2 ct)
Measurement range (AC or DC)
400 - 550 V
1V
± (1.5 % + 1 ct)
Input impedance
and PE
270 kΩ between terminals L, N,
530 kΩ between terminals L and N
Frequency of use
DC and 15.8 to 450 Hz
Ω, PE)
Voltage measurements in insulation measurement (MΩ
Measurement range (AC or DC)
Resolution
Intrinsic uncertainty
5,0 - 399,9 V
400 - 550 V
0,1 V
1V
± (3,7 % + 2 pt)
± (3,7 % + 1 pt)
Input impedance
145 kΩ
Frequency of use
DC and 15.8 to 65 Hz
Contact voltage measurements
Measurement range (AC)
2.0 - 100.0 V
Intrinsic uncertainty
± (15% + 2 ct)
Input impedance
6 MΩ
Frequency of use
15.8 … 65 Hz
This voltage is displayed only if it exceeds UL (25 V, 50 V or 65 V).
84
Measurements of potential of the voltage probe
The characteristics are the same as in the voltage measurements except that the input impedance is 200 kΩ.
This voltage must normally be between 0 and UL.
8.2.2. FREQUENCY MEASUREMENTS
Particular reference conditions:
Voltage
≥2V
Voltage
≥ 20 V for the MΩ voltage input
or current
≥ 30 mA for the MN77 clamp,
≥ 50 mA for the C177A clamp.
Beyond these values, the frequency is indeterminate (display of - - - - ).
Measurement range
15.8 - 399.9 Hz
Resolution
400.0 - 499.9 Hz
0.1 Hz
1 Hz
Intrinsic uncertainty
± (0.1 % + 1 ct)
8.2.3. CONTINUITY MEASUREMENTS
Particular reference conditions:
Resistance of the leads: zero or compensated.
Inductance of the leads: zero.
External voltage on the terminals: zero.
Inductance in series with the resistance: zero.
Compensation of the leads up to 5 Ω.
The maximum acceptable superposed external AC voltage is 0.5 VRMS in sine wave.
200 mA current
Measurement range
0.00 - 39.99 Ω
Resolution
0.01 Ω
Measurement current
≥ 200 mA
Intrinsic uncertainty
± (1.5% + 2 ct)
Operating uncertainty
± (8.5% + 2 ct)
No-load voltage
9.5 V ± 10%
Maximum inductance in series
40 mH
12 mA current
Measurement range
Resolution
0.00 - 39.99 Ω
40.0 - 399.9 Ω
0.01 Ω
0.1 Ω
Measurement current
12 mA
Intrinsic uncertainty
± (1.5% + 5 ct)
Operating uncertainty
± (8.5% + 5 ct)
No-load voltage
9.5 V ± 10%
Maximum inductance in series
40 mH
85
8.2.4. RESISTANCE MEASUREMENTS
Particular reference conditions:
External voltage on the terminals: zero.
Inductance in series with the resistance: zero.
Measurement range
0.001 - 3.999 kΩ
4.00 - 39.99 kΩ
40.0 - 399.9 kΩ
1Ω
10 Ω
100 Ω
Resolution
Measurement current
≤ 22 µA
≤ 22 µA
≤ 17 µA
Intrinsic uncertainty
± (1.5% + 5 ct)
± (1.5% + 2 ct)
± (1.5% + 2 ct)
No-load voltage
3.1 V ± 10%
8.2.5. INSULATION RESISTANCE MEASUREMENTS
Particular reference conditions:
Capacitance in parallel: zero.
Maximum acceptable external AC voltage during the measurement: zero.
Frequency of external voltages: DC and 15.8 ... 65 Hz.
The frequency is guaranteed only for a voltage ≥ 20 V .
Maximum no-load voltage
No-load voltage (50 V range)
Nominal current
Short-circuit current
External AC voltage on the terminals
Intrinsic uncertainty on the measurement of the test voltage
Measurement range at 50 V
1.254 x UN (for UN ≥ 100 V)
48 V ≤ U ≤ 70 V)
≥ 1 mA
≤ 3 mA
zero
± (2.5% + 3 ct)
0.01 - 7.99 MΩ
8.00 - 39.99 MΩ
Measurement range at 100 V
0.01 - 3.99 MΩ
4.00 - 39,99 MΩ
Measurement range at 250 V
0.01 - 1.99 MΩ
2.00 - 39.99 MΩ
Measurement range at 500 V
0.01 - 0.99 MΩ
1.00 - 39.99 MΩ
Measurement range at 1000 V
0.01 - 0.49 MΩ
0.50 - 39.99 MΩ
10 kΩ
10 kΩ
Resolution
Intrinsic uncertainty
± (5% + 3 ct)
Uncertainty of operation
± (12% + 3 ct)
± (2% + 2 ct)
± (10% + 2 ct)
Typical test voltage vs load curve
The voltage developed as a function of the resistance measured has the following form:
I = 1 mA
UN
RN = UN / 1 mA
The maximum capacitance between the terminals is 3 μF.
86
40.0 - 399.9 MΩ
400 - 1999 MΩ
100 kΩ
1 MΩ
± (2% + 2 ct)
50V range:
Value for guidance
Other ranges:
± (2% + 2 ct)
± (10% + 2 ct)
50V range:
Value for guidance
Other ranges:
± (10% + 2 ct)
Typical measurement settling time as a function of the elements tested
These values include influences due to the capacitive component of the load, to the automatic range system, and to the regulation
of the test voltage.
Test voltage
50 V - 250 V
250 V - 500 V - 1000 V
Load
Non-capacitive
With 100 nF
With 1 µF
10 MΩ
1s
1000 MΩ
1s
10 MΩ
1s
2s
12 s
1000 MΩ
1s
4s
30 s
50 V
100 V
250 V
500 V
1000 V
0,25 s x C
0,5 s x C
1sxC
2sxC
4sxC
-
Typical discharge time of a capacitive element to reach 25 V
Test voltage
Discharge time (C in µF)
8.2.6. 3P EARTH RESISTANCE MEASUREMENTS
Particular reference conditions:
Resistance of the E lead: zero or compensated.
Interference voltages: zero.
Inductance in series with the resistance: zero.
(RH + RS ) / RE < 300 and RE < 100 x RH with RH and RS ≤ 15,00 kΩ.
Compensation of the lead RE up to 2.5 Ω.
Measurement range
0.50 - 39.99 Ω
40.0 - 399.9 Ω
400 - 3999 Ω
0.01 Ω
0.1 Ω
1Ω
Resolution
0.20 - 15.00 kΩ 1
10 Ω
Intrinsic uncertainty
± (2% + 10 ct)
± (2% + 2 ct)
± (10% + 2 ct)
Operating uncertainty
Typical peak-to-peak measurement current 2
Measurement frequency
± (9% + 20 ct)
± (9% + 5 ct)
-
4.3 mA
4.2 mA
3.5 mA
128 Hz
No-load voltage
38.5 V peak-to-peak
1: the 40 kΩ display range is used only for measurements of the RH and RS rods.
2: current at mid-range with RH = 1000 Ω.
Maximum acceptable interference voltage:
25 V on H from 50 to 500 Hz.
25 V on S from 50 to 500 Hz.
Accuracy on the measurement of the interference voltages:
Characteristics the same as for the voltage measurements in §8.2.1.
8.2.7. EARTH MEASUREMENTS ON LIVE CIRCUITS
Particular reference conditions:
Voltage of the installation: 90 to 500 V.
Stability of the voltage source: < 0.05%.
Frequency of the installation: 15.8 to 17.5 Hz and 45 to 65 Hz.
Resistance of the leads: zero or compensated.
Impedance of the inductive part: < 0.1 x the resistive part of the impedance measured
Contact voltage (potential of the protective conductor with respect to the local earth): < 5 V.
Resistance of the voltage measurement probe: ≤ 15 kΩ.
Potential of the voltage probe with respect to the PE: ≤ UL.
Residual leakage current of the installation: zero.
Compensation of the lead RE up to 2.5 Ω.
87
-
Characteristics in trip mode:
Measurement range
0.100 - 0.500 Ω
Resolution
0.510 - 3.999 Ω
4.00 - 39.99 Ω
40.0 - 399.9 Ω
0.01 Ω
0.1 Ω
0.001 Ω
Intrinsic uncertainty on the impedance measurement
± (10% + 20 ct)
± (5% + 20 ct)
Peak measurement current between 90 and 270 V
2,45 - 7,57 A
2,27 - 7,55 A
1,36 - 7,02 A
0,274 - 4,20 A
Peak measurement current between 270 and 550 V
4,48 - 6,66 A
4,3 - 6,66 A
3,05 - 6,39 A
0,78 - 4,53 A
Intrinsic uncertainty on the resistive part
± (10% + 20 ct)
± (5% + 20 ct)
± (5% + 2 ct)
Intrinsic uncertainty on the inductive part 3
± (10% + 2 ct)
± (10% + 2 ct)
-
Operating uncertainty on the impedance measurement
± (17% + 20 ct)
± (12% + 20 ct)
± (12% + 2 ct)
Frequency of operation
± (5% + 2 ct)
15.8 to 17.5 and 45 to 65 Hz
3: when the impedance is >15 Ω, the instrument displays 0.0 mH for the inductive part.
The duration of the measurement depends on the voltage of the installation, on the measured impedance value, and on the activation of the smoothing filter (SMOOTH).
If smoothing is activated (SMOOTH mode), the instability of the intrinsic uncertainty is then halved (for example: ±5 digits becomes
±2.5 digits).
Maximum acceptable resistance of the voltage probe: 15 kΩ.
Intrinsic uncertainty on the probe resistance measurement: ± (10% + 5 digits), resolution 0.1 kΩ.
Maximum acceptable inductance for the measurement: 15 mH, resolution 0.1 mH.
Calculation of the fault voltage if there is a short-circuit, UFk:
Calculation range
0.2 - 399.9 V
400 - 550 V
0.1 V
1V
Resolution
= √ (Intrinsic uncertainty on the voltage measurement if UMEAS is used)²
Intrinsic uncertainty
+ (Intrinsic uncertainty on the loop measurement)²
Frequency of operation
15,8 to 17,5 and 45 to 65 Hz
Characteristics in non-tripping mode:
Measurement range
0.20 - 0.99 Ω
Resolution
RMS measurement current
Intrinsic uncertainty on the impedance measurement 4
Intrinsic uncertainty on the resistive part
Intrinsic uncertainty on the inductive part
Operating uncertainty on the
impedance measurement
1.00 - 1.99 Ω
2.00 - 39.99 Ω
40.0 - 399.9 Ω
400 - 3999 Ω
0.1 Ω
1Ω
0.01 Ω
choice of 6. 9. or 12 mA
± (15% + 3 ct)
± (15% + 3 ct)
± (10% + 3 ct)
± (5% + 2 ct)
± (15% + 3 ct)
± (15% + 3 ct)
± (10% + 3 ct)
± (5% + 2 ct)
± (10% + 3 ct)
± (10% + 3 ct)
± (10% + 3 ct)
± (5% + 2 ct)
± (20% + 3 ct)
± (20% + 3 ct)
± (12% + 3 ct)
-
-
4: There is no measurement of the inductive in L-PE loop part with a low current.
The intrinsic uncertainty is defined for 0.1 ≤ RL / RN ≤ 10 with RL and RN ≥ 1 Ω.
The duration of the measurement depends on the voltage of the installation, on the measured impedance value, and on the activation of the smoothing filter (SMOOTH).
If smoothing is activated (SMOOTH mode), the instability of the intrinsic uncertainty is then halved (for example: ±5 digits becomes
±2.5 digits) and the duration of the measurement is of the order of 30 s.
Maximum acceptable resistance of the voltage probe: 15 kΩ.
Intrinsic uncertainty on the probe resistance measurement: ± (10% + 5 digits), resolution 0.1 kΩ.
Maximum acceptable inductance for the measurement: 13,17 mH with R < 0,50 Ω.
88
Characteristics in selective mode:
Measurement range
Resolution
Intrinsic uncertainty on the resistance measurement 5
0.50 - 39.99 Ω
40.0 - 399.9 Ω
0.01 Ω
0.1 Ω
± (10% + 10 ct)
5: there is no measurement of the inductive part in selective mode.
The duration of the measurement depends on the voltage of the installation, on the measured impedance value, and on the activation of the smoothing filter (SMOOTH).
Maximum acceptable resistance of the voltage probe: 15 kΩ.
Accuracy on the probe resistance measurement: ± (10% + 5 digits), resolution 0.1 kΩ.
The measurement current corresponds to the test current indicated in the table of characteristics in tripping mode divided by the
ratio RSEL/RA avec RSEL/RA ≤ 100. Beyond this, the maximum current, 20 mA peak, is reached.
8.2.8. LOOP IMPEDANCE MEASUREMENTS
Particular reference conditions:
Voltage of the installation: 90 to 500 V.
Stability of the voltage source: <0.05%.
Frequency of the installation: 15.8 to 17.5 Hz and 45 to 65 Hz.
Resistance of the leads: zero or compensated.
Contact voltage (potential of the protective conductor with respect to the local earth): < 5 V.
Residual leakage current of the installation: zero.
Compensation of the leads up to 5 Ω.
Characteristics in 3-wire mode with tripping:
See § 8.2.7
Characteristics in 3-wire mode without tripping:
See § 8.2.7
Characteristics of the short-circuit current calculation:
Calculation formula : Ik = UREF / ZS
Calculation range
Resolution
Intrinsic uncertainty
Operating uncertainty
0.1 - 399.9 A
400 - 3999 A
4.00 - 6.00 kA
0.1 A
1A
10 A
= √ (Intrinsic uncertainty on the voltage measurement if UMEAS is used)²
+ (Intrinsic uncertainty on the loop measurement)²
= √ (Operating uncertainty on the voltage measurement if UMEAS is used)²
+ (Operating uncertainty on the loop measurement)²
8.2.9. LINE IMPEDANCE MEASUREMENTS
Particular reference conditions:
Voltage of the installation: 90 to 500 V.
Stability of the voltage source: <0.05%.
Frequency of the installation: 15.8 to 17.5 Hz and 45 to 65 Hz.
Resistance of the leads: zero or compensated.
Impedance of the inductive part: < 0.1 x the resistive part of the impedance measured
Compensation of the leads up to 5 Ω.
Characteristics in 2-wire mode (power-level current):
See § 8.2.7
89
8.2.10. VOLTAGE DROP IN THE CABLES
Particular reference conditions:
Voltage of the installation: 90 to 500 V.
Stability of the voltage source: <0.05%.
Frequency of the installation: 15.8 to 17.5 Hz and 45 to 65 Hz.
Resistance of the leads: zero or compensated.
Impedance of the inductive part: < 0.1 x the resistive part of the impedance measured
Compensation of the leads up to 5 Ω.
The voltage drop is a calculated value.
Calculation formula: ∆V = 100 (Zi - Zi ref ) x IN / UREF
Calculation range
-40% to +40%
Resolution
0,01%
8.2.11. TEST OF RESIDUAL CURRENT DEVICE
Particular reference conditions:
Voltage of the installation: 90 to 500 V.
Frequency of the installation: 15.8 to 17.5 Hz and 45 to 65 Hz.
Contact voltage (potential of the protective conductor with respect to the local earth): <5 V.
Resistance of the voltage probe (if used): < 100 Ω.
Potential of the voltage measurement (if used) with respect to the PE: <5 V..
Residual leakage current of the installation: zero.
Limitation of the ranges accessible as a function of the voltage for type AC, A and F RCDs
I∆N
6 mA
10 mA
30 mA
100 mA
300 mA
500 mA
650 mA
1000 mA
Variable
6 - 999 mA
90 - 280 V







≥ 100 V
I∆N ≤ 950 mA
280 - 550 V








I∆N ≤ 500 mA
Limitation of the test current as a function of the nature of the test signal for type AC, A and F RCDs
Depending on the range I∆N chosen and the nature of the test signal, some test modes will be unavailable. This test of coherence
is performed when the test of RCDs is started.
Wave
or
I
6 mA
10 mA
30 mA
100 mA
300 mA
500 mA
650 mA
1000 mA
Variable
6 - 999 mA
Ramp









I∆N pulse









2 x I∆N pulse








I∆N ≤ 500 mA
5 x I∆N pulse








I∆N ≤ 200 mA
I
6 mA
10 mA
30 mA
100 mA
300 mA
500 mA
650 mA
1000 mA
Variable
6 - 999 mA
Ramp








I∆N ≤ 500 mA
Wave
or
I∆N pulse








I∆N ≤ 500 mA
2 x I∆N pulse








I∆N ≤ 250 mA
5 x I∆N pulse








I∆N ≤ 100 mA
90
Characteristics in pulse mode for type AC, A and F RCDs:
6 mA - 10 mA - 30 mA - 100 mA - 300 mA - 500 mA - 650 mA - 1000 mA
Variable (6 to 999 mA) 6
Range I∆N
Nature of the test
Determination of UF
0.2 x I∆N … 0.5 x I∆N
Test current
Intrinsic uncertainty on the test
+0 -7%
current
± 2 mA
Maximum duration of application
from 32 to 72
of the test current (STD)
periods
Maximum duration of application
from 32 to 72
of the test current (S or G)
periods
9
Non-tripping test
Tripping test
Tripping test
Tripping test
0.5 x I∆N
+0 -7%
± 2 mA
I∆N
-0 +7%
± 2 mA
2 x I∆N
-0 +7%
± 2 mA
5 x I∆N
-0 +7%
± 2 mA
1000 or 2000 ms 8
300 ms
150 ms
40 ms
1000 or 2000 ms 8
500 ms
200 ms
150 ms
7
6: the upper limit of the variable range (999 mA) depends on the nature of the test performed and on the type of test current (half
or full wave).
7: this current can be adjusted in steps of 0.1 I∆N and must not be less than 4 mA. As default, this current is 0.3 I∆N.
8: to be chosen when configuring the measurement.
9: when the signal contains only positive or only negative alternations, I∆N is multiplied by 1.4.
Characteristics in ramp mode for type AC, A and F RCDs:
6 mA - 10 mA - 30 mA - 100 mA - 300 mA - 500 mA - 650 mA - 1000 mA
Variable (6 to 999 mA) 9
Range I∆N
Nature of the test
Determination of UF
Tripping test
Test current
Intrinsic uncertainty on the test
current
Maximum duration of application
of the test current
0.2 x I∆N … 0.5 x I∆N
0.9573 x I∆N x k / 28 11
12
10
+0 -7% ± 2 mA
-0 +7% ± 2 mA
4600 ms to 50 and 60 Hz
4140 ms to 16.6 Hz
-0 +7% + 3.3 % I∆N ± 2 mA
Resolution de 0.1 mA up to 400 mA
and 1 mA thereafter
from 32 to 72 periods
Intrinsic uncertainty on the indication of the tripping current
-
9: the upper limit of the variable range (999 mA) depends on the nature of the test performed and on the type of test current (half
or full wave).
10: this current can be adjusted in steps of 0.1 I∆N and must not be less than 4 mA. As default, this current is 0.3 I∆N.
11: k is between 9 and 31. The waveform so generated goes from 0.3 I∆N to 1.06 I∆N in 22 steps of 3.3% I∆N each having a duration
of 200 ms (180 ms at 16.66Hz).
12: when the signal contains only positive or only negative alternations, I∆N is multiplied by 1.4.
Characteristics of the trip time (TA) for type AC, A and F RCDs:
Pulse mode
Ramp mode
Measurement range (STD)
5.0 - 299.9 ms
-
10.0 - 200.0 ms
Measurement range (S or G)
5.0 - 399.9 ms
400 - 500 ms
10.0 - 200.0 ms
1 ms
0.1 ms
Resolution
0.1 ms
Intrinsic uncertainty
± 2 ms
± 2 ms
Operating uncertainty
± 3 ms
± 3 ms
Characteristics of the fault voltage calculation (UF) for type AC, A and F RCDs:
Measurement range
Resolution
Intrinsic uncertainty
5.0 - 70.0 V
0.1 V
± (10% + 10 ct)
Calculation formula:
UF = ZLPE x I∆N or ZA x I∆N or RA x I∆N or ZLPE x 2I∆N if the test is at 2I∆N
91
Limitation of the ranges available as a function of the voltage for type B, B+ and EV RCDs
I∆N
6 mA
10 mA
30 mA
100 mA
300 mA
500 mA
90 - 280 V






280 - 550 V






Variable
6 - 499 mA


Limitation of the test current as a function of the nature of the test signal for type B, B+ and EV RCDs
Depending on the range I∆N chosen and the nature of the test signal, some test modes will be unavailable. This test of coherence
is performed when the test of RCDs is started.
Wave
or
I
6 mA
10 mA
30 mA
100 mA
300 mA
500 mA
Variable
6 - 499 mA
Ramp







2 x I∆N pulse





I∆N ≤ 250 mA
4 x I∆N pulse







I∆N ≤ 125 mA
Characteristics in pulse mode for type B, B+ and EV RCDs:
Range I∆N
6 mA - 10 mA - 30 mA - 100 mA - 300 mA - 500 mA
Nature of the test
Tripping test
Tripping test
Test current
Intrinsic uncertainty on the test
current
Maximum duration of application
of the test current
2.2 x 2 I∆N
-0 + 3.5%
± 2 mA
2.4 x 4 I∆N
-0 + 3.5%
± 2 mA
300 ms
150 ms
12: this current can be adjusted in steps of 0.1 I∆N and must not be less than 10 mA. As default, this current is 0.2 I∆N.
Characteristics in ramp mode for type B, B+ and EV RCDs:
Range I∆N
6 mA - 10 mA - 30 mA - 100 mA - 300 mA - 500 mA
Nature of the test
Tripping test
Test current
Intrinsic uncertainty on the test
current
Maximum duration of application
of the test current
0.2 to 2.2 x I∆N
-0 +7% ± 2 mA
6000 ms
(-0 +7% + 3.3 % I∆N ± 2 mA
Resolution de 0.1 mA up to 400 mA
and 1 mA thereafter
Intrinsic uncertainty on the indication of the tripping current
Characteristics of the trip time (TA) for type B, B+ and EV RCDs:
Pulse mode
Measurement range
Resolution
5.0 - 399.9 ms
400 - 500 ms
0.1 ms
1 ms
Intrinsic uncertainty
± 2 ms
Operating uncertainty
± 3 ms
The tripping time is not displayed.
92
8.2.12. CURRENT MEASUREMENT
Particular reference conditions:
Peak factor = 1,414
DC component< 0.1 %.
Frequency: 15.8 450 Hz.
For the measurement of ISEL, the intrinsic uncertainty is increased by 5%.
Characteristics with the MN77 clamp:
Transformation ratio: 1000 / 1
Measurement range
5.0 - 399.9 mA
0.400 - 3.999 A
4.00 - 19.99 A
0.1 mA
1 mA
10 mA
± (2% + 5 ct)
± (1.5% + 2 ct)
± (1.2% + 2 ct)
Resolution
Intrinsic uncertainty
When a voltage is connected between the L and PE terminals, the device synchronizes to the frequency of this voltage, allowing
current measurements from 1 mA.
Characteristics with the C177A clamp:
Transformation ratio: 10 000 / 1
Measurement range
Resolution
Intrinsic uncertainty
5.0 - 399.9 mA
0.400 - 3.999 A
4.00 - 39.99 A
40.0 - 199.9 A
0.1 mA
1 mA
10 mA
100 mA
± (2% + 5 ct)
± (1.5% + 2 ct)
± (1% + 2 ct)
± (1.2% + 2 ct)
When a voltage is connected between the L and PE terminals, the device synchronizes to the frequency of this voltage, allowing
current measurements from 5 mA.
In selective current measurements, the intrinsic error of the clamps is increased by 5 %.
8.2.13. DIRECTION OF PHASE ROTATION
Particular reference conditions:
Three-phase network.
Voltage of the installation: 20 to 500 V.
Frequency: 15.8 to 17. 5 Hz and 45 to 65 Hz.
Acceptable level of amplitude unbalance: 20%.
Acceptable level of phase unbalance: 10%.
Acceptable level of harmonics (voltage): 10%.
Characteristics:
The phase order is «positive» if rotation L1-L2-L3 is anticlockwise.
The phase order is «negative» if rotation L1-L2-L3 is clockwise.
The three voltages are measured (see the characteristics in §8.2.1) and indicated as U12, U23 and U31.
93
8.2.14. POWER MEASUREMENTS (WITH C177A CLAMP)
Particular reference conditions:
Sinusoidal voltage and current signals: cosφ = 1.
Voltage ≥ 10 V.
Current ≥ 0.1 A.
Frequency: 15.8 to 17.5 Hz and 45 to 65 Hz.
No DC component.
Measurement range
1W
10 W
40.0 - 110.0 kW 13
40.0 - 330.0 kW
100 W
± (2% + 5 ct)
± (2% + 2 ct)
± (2% + 2 ct)
5 - 3999 W
Resolution
Intrinsic uncertainty
4.00 - 39.99 kW
13: full scale is 110 kW (550V x 200A) in single-phase and 330 kW in three-phase.
8.2.15. POWER FACTOR
Particular reference conditions:
Voltage of the installation: 10 to 500 V.
Current: 0.1 to 200 A.
Measurement range
(±) 0.2 - 0.49
Resolution
Intrinsic uncertainty
(±) 0.50 - 1.00
0.01
± (2% + 2 ct)
± (1% + 2 ct)
If the power is zero, the power factor is indeterminate.
The sign of the power factor depends on whether the phase of the voltage leads or lags that of the current. This can be used to
determine whether the load is inductive (+ sign) or capacitive (- sign).
8.2.16. HARMONICS
Particular reference conditions:
Signal without inter-harmonics, of which the fundamental is stronger than the other harmonic components and than the DC
component.
Frequency of the fundamental: 16.66 Hz, 50 Hz, or 60 Hz ± 0.05 Hz.
Peak factor of the signal ≤ 4.
Characteristics:
Characteristics of voltage display
Characteristics of current display
Stability of the current and voltage display
10 to 500 V, the display range being determined by the value of the
strongest harmonic component.
1 to 200 A, the display range being determined by the value of the
strongest harmonic component.
± 2 ct
Domain of use
Harmonics of orders 1 to 50
Measurement range for the harmonic factor
0.2 - 399.9 %
Detection threshold for the harmonic factor
0.1 %
Measurement range in THD-F and THD-R
0.2 - 100 %
Resolution for the harmonic factor, THD-F and THD-R
0.1%
Factor > 10% and order < 13: 5 ct
Factor <10% and order < 13: 10 ct
Factor > 10% and order > 13: 10 ct
Factor > 10% and order > 13: 15 ct
10 ct
Intrinsic uncertainty on the RMS value and the harmonic
factor
Intrinsic uncertainty on the THD-F and THD-R
94
Method and definitions:
Determination of harmonics: Cooley-Tukey FFT algorithm on 16 bits
Sampling frequency: 256 times the frequency of the fundamental
Filtering window: rectangular, 4 periods
THD-F: Total distortion referred to the fundamental of the signal.
n=50
√Σ
THD-F =
n=2
H n²
H1
THD-R: Total distortion referred to the RMS value of the signal (also called DF: distortion factor).
n=50
√Σ
THD-R =
n=2
H n²
R[RMS]
8.3. VARIATIONS IN THE RANGE OF USE
8.3.1. VOLTAGE MEASUREMENT
Quantities of influence
Temperature
Limits of the range of use
-10 … + 55 °C
Variation of the measurement
Typical
Maximum
1 %/10 °C ± 1 ct
2 %/10 °C + 2 ct
Relative humidity
10 … 85 % RH at 45°C
2%
3 % + 2 ct
Supply voltage
8.4 … 12.7 V
0.1% or 1 ct
0.5% + 2 ct
Frequency (except in MΩ setting)
15.8 … 450 Hz
0,5%
4,5 % + 1 ct
Frequency (in MΩ setting)
15.8 … 65 Hz
4%
1% + 1 ct
0 … 500 Vac
50 dB
40 dB
Series mode rejection in AC
50/60Hz series mode rejection in DC
Common mode rejection in 50/60Hz AC
8.3.2. INSULATION MEASUREMENT
Quantities of influence
Temperature
Limits of the range of use
-10 … + 55 °C
Variation of the measurement
Typical
Maximum
1 %/10 °C ± 1 ct
2 %/10 °C + 2 ct
Relative humidity
10 … 85 % RH at 45°C
2%
3 % + 2 ct
Supply voltage
8.4 … 12.7 V
0.25% or 2 ct
2% + 2 ct
1%
5% + 2 ct
1%
1%
1% + 1 ct
10% + 5 ct
Ranges 50 V and 100 V
R ≤ 100 MΩ : 2 V
R > 100 MΩ : 0,7 V
50/60Hz AC voltage superposed on the
test voltage (UN)
Ranges 250 V and 500 V
R ≤ 100 MΩ : 6 V
R > 100 MΩ : 2 V
Ranges 500 V and 1000 V
R ≤ 100 MΩ : 10 V
R > 100 MΩ : 3 V
Capacitance in parallel on the resistance
to be measured
0 … 5 µF @ 1 mA
0 … 2 µF @ 2000 MΩ
95
8.3.3. RESISTANCE AND CONTINUITY MEASUREMENT
Quantities of influence
Temperature
Relative humidity
Supply voltage
50/60Hz AC voltage superposed on the
test voltage
Limits of the range of use
Variation of the measurement
Typical
Maximum
1 %/10 °C ± 1 ct
2 %/10 °C + 2 ct
10 … 85 % RH at 45°C
2%
3 % + 2 ct
8.4 … 12.7 V
0.25% or 1 ct
1% + 2 ct
0.5 Vac
0,5%
1% + 2 ct
-10 … + 55 °C
8.3.4. 3P EARTH MEASUREMENT
Quantities of influence
Temperature
Relative humidity
Supply voltage
Voltage in series in the voltage measurement loop (S-E)
Fundamental = 16.6/50/60Hz + odd harmonics
Voltage in series in the current injection
loop (H-E)
Fundamental = 16.6/50/60Hz + odd harmonics
Current loop rod resistance (RH)
Voltage loop rod resistance (RS)
Limits of the range of use
Variation of the measurement
Typical
Maximum
-10 … + 55 °C
1 %/10 °C ± 1 ct
2 %/10 °C + 2 ct
10 … 85 % RH at 45°C
2%
3 % + 2 ct
8.4 … 12.7 V
0.25% or 1 ct
1% + 2 ct
15 V (RE ≤ 40 Ω)
25 V (RE > 40 Ω)
0.5% or 10 ct
2% + 50 ct
2% + 2 ct
15 V (RE ≤ 40 Ω)
25 V (RE > 40 Ω)
0.5% or 10 ct
2% + 50 ct
2% + 2 ct
0 to 15 kΩ
0.3%
1% + 2 ct
0 to 15 kΩ
0.3%
1% + 2 ct
8.3.5. CURRENT MEASUREMENT
Quantities of influence
Temperature
Limits of the range of use
-10 … + 55 °C
Variation of the measurement
Typical
Maximum
1 %/10 °C ± 1 ct
2 %/10 °C + 2 ct
Relative humidity
10 … 85 % RH at 45°C
2%
3 % + 2 ct
Supply voltage
8.4 … 12.7 V
15.8 … 45 Hz
45 … 450 Hz
0 … 500 Vac
0.1% or 2 ct
1%
0.5%
50 dB
0.5% + 2 ct
1% + 1 ct
1.5% + 1 ct
40 dB
Frequency
50/60Hz series mode rejection in AC
96
8.3.6. EARTH MEASUREMENT ON LIVE CIRCUIT, LOOP AND SELECTIVE EARTH
Quantities of influence
Limits of the range of use
Temperature
-10 … + 55 °C
Relative humidity
10 … 85 % RH at 45°C
Supply voltage
8.4 … 12.7 V
99 to 101% of the nominal
Network frequency of the installation tested
frequency
85 to 110% of the nominal
Network voltage of the installation tested
voltage
Phase difference between the internal load
0 …20°
and the measured impedance or inductor 0 … 400 mH
ance of the measured impedance or L/R
or 0 … 500 ms
ratio of the measured impedance
Resistance in series with the voltage probe
(earth on live circuit only)
0 … 15 kΩ
Contact voltage (UC)
0 … 50 V
Variation of the measurement
Typical
Maximum
1 %/10 °C ± 1 ct
2 %/10 °C + 2 ct
2%
3 % + 2 ct
0.5% or 2 ct
2% + 2 ct
0.1% or 1 ct
0.1% + 1 ct
0.1% or 1 ct
0.1% + 1 ct
1%/10°
1%/10°
Negligible
(taken into account in the
intrinsic uncertainty)
Negligible
(taken into account in the
intrinsic uncertainty)
Negligible
(taken into account in the
intrinsic uncertainty)
Negligible
(taken into account in the
intrinsic uncertainty)
8.3.7. TEST OF RESIDUAL CURRENT DEVICE
Quantities of influence
Limits of the range of use
Temperature
Relative humidity
Variation of the measurement
Typical
Maximum
-10 … + 55 °C
1 %/10 °C ± 1 ct
2 %/10 °C + 2 ct
10 … 85 % RH at 45°C
2%
3 % + 2 ct
0.1% or 1 ct
0.5% + 2 ct
0.1% or 1 ct
0.1% + 1 ct
0.1% or 1 ct
0.1% + 1 ct
Supply voltage
8.4 … 12.7 V
99 to 101% of the nominal
Network frequency of the installation tested
frequency
85 to 110% of the nominal
Network voltage of the installation tested
voltage
8.3.8. DIRECTION OF PHASE ROTATION
No quantity of influence
8.3.9. POWER
Quantities of influence
Temperature
Relative humidity
Limits of the range of use
-10 … + 55 °C
10 … 85 % RH at 45°C
Supply voltage
8.4 … 12.7 V
Network frequency of the installation
99 to 101% of the nominal
tested
frequency
Network voltage of the installation 85 to 110% of the nominal volttested
age
0.50 … 1.00 at 45…65 Hz
0.20 … 0.49 at 45…65 Hz
Power factor
0.50 … 1.00 at 15.8…17.5 Hz
0.20 … 0.49 at 15.8…17.5 Hz
Variation of the measurement
Typical
Maximum
1 %/10 °C ± 1 ct
2 %/10 °C + 2 ct
2%
3 % + 2 ct
0.1% or 1 ct
0.5% + 2 ct
0.1% or 1 ct
0.1% + 1 ct
0.1% or 1 ct
0.1% + 1 ct
0.5%
1.5%
2%
4%
1% + 2 ct
3% + 2 ct
2.5% + 2 ct
5% + 2 ct
8.3.10. VOLTAGE AND CURRENT HARMONICS
The quantities of influence and the associated variations are the same as for voltage measurements and current measurements,
respectively.
97
8.4. INTRINSIC UNCERTAINTY AND OPERATING UNCERTAINTY
The installation testers comply with standard IEC-61557, which requires that the operating uncertainty, called B, be less than 30%.
■ In insulation, B = ± ( |A| + 1,15 √ E1² + E2² + E3² )
with
A = intrinsic uncertainty
E1 = influence of the reference position ± 90°.
E2 = influence of the supply voltage within the limits indicated by the manufacturer
E3 = influence of the temperature between 0 and 35°C.
■ In continuity measurement, B = ± ( |A| + 1,15 √ E1² + E2² + E3² )
■ In loop measurement, B = ± ( |A| + 1,15 √ E1² + E2² + E3² + E6² + E7² + E8² )
with
E6 = influence of the phase angle from 0 to 18°.
E7 = influence of the network frequency from 99 to 101% of the nominal frequency.
E8 = influence of the network voltage from 85 to 110% of the nominal voltage.
■ In earth measurement, B = ± ( |A| + 1,15 √ E1² + E2² + E3² + E4² + E5² + E7² + E8² )
with
E4 = influence of the interference voltage in series mode (3 V at 16.6, 50, 60, and 400 Hz)
E5 = influence of the resistance of the rods from 0 to 100 x RA but ≤ 50 kΩ.
■ In test of residual current device, B = ± ( |A| + 1,15 √ E1² + E2² + E3² + E5² + E8² )
with
E5 = influence of the resistance of the probes within the limits indicated by the manufacturer.
8.5. POWER SUPPLY
The device is powered by a10.8 V 5.8 Ah rechargeable Lithium-ion battery pack.
Battery mass: about 365 g including 5.04 g of lithium.
8.5.1. LITHIUM-ION TECHNOLOGY
The Li-ion technology has many advantages:
■ long life between charges with limited bulk and weight,
■ no memory effect: you can recharge your battery even if it is not fully discharged, without reducing its capacity,
■ very low self discharge,
■ the possibility of recharging your battery rapidly,
■ protection of the environment through the absence of polluting materials such as lead and cadmium.
8.5.2. BATTERY CHARGE
The instrument is not designed to operate when the charger is connected. The measurements must be made using battery
power.
The battery charger of the device is in two distinct parts: an external power supply and a charger built into the device.
The charger manages the charging current, the battery voltage, and the internal temperature of the battery simultaneously. This
optimizes charging while guaranteeing a long battery life.
The day before you use your device, check its charge condition. If the battery level indicator shows less than three bars, charge
the device overnight (see §1.4).
The charging time is approximately 5 h.
In order to extend the life of your battery:
■
Use only the charger supplied with your device. The use of another charger may prove dangerous!
■ Charge your device only between 0 and 45°C.
■ Observe the conditions of use and storage stated in this data sheet.
Following prolonged storage, the battery may be fully discharged. In this case, the first charge may take longer.
98
Set the switch to OFF; charging is possible when the device is not off, but will take longer.
8.5.3. OPTIMIZE BATTERY CHARGING
During charging, the temperature of the battery rises, especially towards the end. A safety device, built into the battery, checks constantly that the battery temperature does not exceed an acceptable maximum. If this maximum is exceeded, the charger switches
off automatically, even if charging is not complete.
The battery is at the bottom of the device, and the evacuation of the heat can be facilitated by placing the device upright while
charging. The battery temperature is then lower and it will be charged more fully.
This precaution is especially important when the air temperature is high (in summer).
8.5.4. LIFE BETWEEN CHARGES
The mean battery life depends on the type of measurement and on how the device is used. Approximately:
■ 12 h if the automatic switching off function is deactivated,
■ 24 h if the automatic switching off function is activated,
How long your device can operate when the battery is fully charged depends on several factors:
■ The consumption of the device, which depends on the measurements you make,
■ The capacity of the battery. It is greatest when the battery is new, and declines as the battery ages.
Here are a few ways to extend battery life between charges:
■ Use the back-lighting only when it is strictly necessary,
■ Set the brightness of the display to the lowest level at which you can still read the display unit,
■ Program the shortest time to automatic switching off you are comfortable with (see SET-UP, § 5),
■ Use the pulse mode for continuity measurements at 200 mA,
■ If the continuity measurement at 200 mA is made in permanent mode, do not let the measurement leads touch each other when
you are not making a measurement,
■ When making insulation measurements at high test voltages, stop pressing the TEST button when the measurement is over.
Typical life between charges of the device:
Function
Device off
Device on standby
At 50%
brightness
At 100%
brightness
Number of measurements per hour
> 3 months 14
> 3 months 14
-
Conditions
> 10 days
> 10 days
-
Voltage / Current / Power /
Harmonics
81 h
57 h
-
A
Continuity at 200 mA
20 h
16 h
120
B
Continuity at 12 mA
23 h
18 h
120
B
Insulation
22 h
17 h
120
B
Earth, 3P
25 h
18 h
30
C
Loop / RCD
22 h
18 h
300
D
Loop / RCD (smooth)
20 h
16 h
20
E
Earth, 1P / Selective earth
22 h
18 h
300
D
Earth, 1P / Selective earth (smooth)
22 h
18 h
20
E
14: If the device is to be left unused for more than 2 months, remove the battery. To keep it fully charged, recharge it every 4 to 6
months.
A: With automatic switching off after 10 minutes, one measurement every 30 minutes, 7 hours a day.
B: With one 5-second measurement every 25 seconds and a programmed automatic switching-off time.
C: With 5 consecutive 10-second measurements every 10 minutes and a programmed automatic switching-off time.
D: With 5 consecutive 5-second measurements every minute and a programmed automatic switching-off time.
E: With 5 consecutive 30-second measurements every 3 minutes and a programmed automatic switching-off time.
99
8.5.5. END OF BATTERY LIFE
The internal resistance of a battery at the end of its life is high. The result is an abnormally short charging time.
After a full charge, the device indicates “charging over”, but as soon as the charger is disconnected, the display unit loses its
contrast and goes off, meaning that the battery no longer holds a charge.
8.6. ENVIRONMENTAL CONDITIONS
Indoor and outdoor use.
Specified operating range 15
Range for recharging the battery
Range in storage (without battery)
Altitude
Pollution degree
-20 to 60°C and 10% to 85% RH
10 to 45°C
-40°C to +70°C and 10% to 90% RH
<2,000m
2
15: This range corresponds to the range of the operating uncertainty defined by standard IEC-61557. When the device is used
outside this range, it is necessary to add 1.5%/10°C and 1.5% between 75 and 85% RH to the operating uncertainty.
8.7. MECHANICAL CHARACTERISTICS
Dimensions (L x D x H)
Weight
280 x 190 x 128 mm
approximately 2.2 kg
Protection class
IP 53 per IEC 60529 if the cover of the USB port is closed, IP 51 if it is open.
IK 04 per IEC 62262
Free fall test
Per IEC/EN 61010-2-034
8.8. CONFORMITY TO INTERNATIONAL STANDARDS
The instrument is compliant with safety standard IEC/EN 61010-2-034 and the leads are compliant with IEC/EN 61010-031, for
voltages of 600V category III or 300V in category IV (under shelter).
Assigned characteristics: measurement category III, 600V with respect to earth (or 300V in CAT IV under shelter), 550V in differential between the terminals, and 300V, CAT II on the charger input.
The device is in conformity with IEC-61557 parts 1, 2, 3, 4, 5, 6, 7 and 10.
8.9. ELECTROMAGNETIC COMPATIBILITY (EMC)
The device is in conformity with standard IEC/EN 61326-1.
100
9. DEFINITIONS OF SYMBOLS
Here is a list of the symbols used in this document and on the display unit of the device.
3P
3-point earth resistance measurement with 2 auxiliary rods.
AC
AC (Alternating Current) signal.
DC
DC (Direct Current) signal.
DF
Distortion Factor = THD-R.
E
E terminal (earth electrode, measurement current return terminal).
FFT
harmonic analysis of a signal (Fast Fourier Transform).
G
selective residual current device, specific to Austria.
H
H terminal (measurement current injection terminal in 3P earth measurement).
Hz
Hertz: indicates the frequency of the signal.
I
current.
I1
current in phase 1 of a three-phase network.
I3
current in phase 3 of a three-phase network.
Ia
RCD tripping current of the residual current device.
Ik
short-circuit current between the L and N, L and PE, N and PE, or L and L terminals.
IN
rated current of the fuse.
IT
Type of link to earth defined in standard IEC-60364-6.
I2
I∆N
current in phase 2 of a three-phase network.
assigned operating current of the RCD to be tested.
Isc
current the fuse must withstand before blowing. It depends on the type of fuse, on IN, and on its delay.
ISEL
current flowing in the earthing resistance to be measured in selective earth measurement on live circuit.
L
L terminal (phase).
Li
inductance in the L-N or L-L loop.
LS
inductance in the L-PE loop.
N
N terminal (neutral).
φ
phase difference of the current with respect to the voltage.
P
active power, P = U . I . PF.
PE
PE terminal (protective conductor).
PF
power factor (cosφ for sinusoidal signal).
PIT
Permanent Insulation Tester.
R
mean resistance calculated from R+ and R-.
R+
resistance measured with a positive current flowing from terminal Ω to terminal COM.
R-
resistance measured with a negative current flowing from terminal Ω to terminal COM.
R±
resistance measured alternately with a positive current, then a negative current.
R∆
resistance of the accessories subtracted from the measurement (compensation of the measurement leads).
RCD
acronym designating a Residual Current Device.
RA
earth resistance in earth measurement on live circuit.
RE
earth resistance connected to the E terminal.
RL-N
resistance in the L-N loop.
RASEL
selective earth resistance in selective earth measurement on live circuit.
RH
resistance of the rod connected to the H terminal.
RL-PE
resistance in the L-PE loop.
RMS
Root Mean Square: root-mean-square value of the signal, the square root of the mean of the squares of the signals.
RN-PE
resistance in the N-PE loop.
RPI
resistance of the auxiliary rod in earth measurement on live circuit.
RN
nominal resistance in insulation measurement RN = UN/1mA.
RPE
resistance of protective conductor PE.
RS
resistance of the rod connected to the S terminal.
S
terminal S (acquisition of measurement potential for the earth resistance calculation).
S
selective residual current device.
101
STD
standard type differential.
TA
effective trip time of the residual current device.
THD-F
level of harmonic distortion referred to the fundamental.
THD-R
level of harmonic distortion referred to the RMS value of the signal.
TN
type of link to earth defined in standard IEC-60364-6.
TT
type of link to earth defined in standard IEC-60364-6.
U12
voltage between phases 1 and 2 of a three-phase network.
U23
voltage between phases 2 and 3 of a three-phase network.
UC
contact voltage between conducting parts when they are touched simultaneously by a person or an animal (IEC61557).
UF
fault voltage appearing during a fault condition between accessible conducting parts (and/or external conducting
parts) and the reference frame ground (IEC-61557).
U31
UFk
UH-E
UL
UL-N
UL-PE
UN
voltage between phases 3 and 1 of a three-phase network.
fault voltage, in the event of a short-circuit, according to Swiss standard SEV 3569.
UFk = Ik x ZA = UREF x ZA/ZS.
voltage measured between terminals H and E.
conventional maximum contact voltage (IEC-61557).
voltage measured between the L and N terminals.
voltage measured between the L and PE terminals.
nominal test voltage in insulation measurement, generated between the MΩ and COM terminals.
UN-PE
voltage measured between the N and PE terminals.
UREF
reference voltage for calculation of the short-circuit current.
UPE
voltage between the PE conductor and the local earth measured when the user presses the TEST button.
US-E
voltage measured between the S and E terminals.
ZA
earth impedance in earth measurement on live circuit.
Zi
impedance in the loop between the phase and the neutral or between two phases (line loop impedance).
ZL-PE
impedance in the L-PE loop.
ZS
ZL-N
impedance in the loop between the phase and the protective conductor.
impedance in the L-N loop.
102
10. MAINTENANCE
Except for the battery, the instrument contains no parts that can be replaced by personnel who have not been specially
trained and accredited. Any unauthorized repair or replacement of a part by an “equivalent” may gravely impair safety.
10.1. CLEANING
Disconnect anything connected to the device and set the switch to OFF.
Use a soft cloth, dampened with soapy water. Rinse with a damp cloth and dry rapidly with a dry cloth or forced air. Do not use
alcohol, solvents, or hydrocarbons.
10.2. REPLACING THE BATTERY
The battery of this device is specific: it has precisely matched protection and safety elements. Replacement of the battery by a
model other than the one specified may result in damage to equipment or bodily injury by explosion or fire.
To keep the device safe, replace the battery only with the original model. Do not use a battery with a damaged jacket.
Replacement procedure:
1. Disconnect anything connected to the device and set the switch to OFF.
MADE IN FRANCE
TO AVOID ELECTRICAL SHOCK
DISCONNECT LEADS, PROBES AND POWER SUPPLY
BEFORE REMOVING LI-ION MODULE
ONLY REPLACE WITH
LI-ION MODULE
2. Turn the device over and insert a screwdriver into the hole in the battery pack.
TO AVOID ELECTRICAL SHOCK
DISCONNECT LEADS, PROBES AND POWER SUPPLY
BEFORE REMOVING LI-ION MODULE
ONLY REPLACE WITH
LI-ION MODULE
3. Then push the screwdriver towards the rear and the battery slides out of its compartment.
Spent batteries must not be treated as ordinary household waste. Take them to the appropriate recycling collection
point.
The internal clock of the instrument continues to run for at least 60 minutes with the battery out.
4. Insert the new battery pack in its compartment and press it home.
103
10.3. RESETTING THE DEVICE
If the device crashes, it can be reset, like a PC.
Set the switch to ZS (RA/SEL.).
Press the 3 keys indicated below simultaneously.
SET UP
/
SET UP
OFF
OFF
OK
10.4. UPDATING OF THE INTERNAL SOFTWARE
With a view to providing, at all times, the best possible service in terms of performance and technical upgrades, Chauvin Arnoux
invites you to update the embedded software of the device by downloading the new version, available free of charge on our web site.
See you on our site:
www.chauvin-arnoux.com
In Support, click on Download our software and enter the name of the instrument.
Connect the device to your PC using the USB cord provided.
The update of the embedded software depends on its compatibility with the hardware version of the device. This version is, indicated in SET-UP (see § 5).
The update of the embedded software overwrites the whole configuration. As a precaution save the data to a PC before
updating the embedded software.
104
11. WARRANTY
Except as otherwise stated, our warranty is valid for 24 months starting from the date on which the equipment was sold. The
extract from our General Terms of Sale is available on our website.
www.chauvin-arnoux.com/en/general-terms-of-sale
The warranty does not apply in the following cases:
■ Inappropriate use of the equipment or use with incompatible equipment;
■ Modifications made to the equipment without the explicit permission of the manufacturer’s technical staff;
■ Work done on the device by a person not approved by the manufacturer;
■ Adaptation to a particular application not anticipated in the definition of the equipment or not indicated in the user’s manual;
■ Damage caused by shocks, falls, or floods.
105
12. APPENDIX
12.1. TABLE OF FUSES MANAGED BY THE C.A 6117
Per standard EN 60227-1 § 5.6.3
DIN gG per standards IEC 60269-1, IEC 60269-2, and DIN VDE 0636-1/2
Iks: break-induced current for a specified time (opening time indicated for each table)
Opening time = 5 s
Nominal current
IN (A)
DIN gG/gL fuse
Iks max (A)
RCD LS-B
Iks max (A)
RCD LS-C
Iks max (A)
RCD LS-D
Iks max (A)
2
6
10
20
20
4
19
20
40
40
28
30
60
60
50
80
100
6
Slow-blow fuse
Iks max (A)
21
8
10
35
38
47
55
65
90
100
60
65
80
100
110
13
16
20
75
85
100
150
150
25
100
110
125
170
170
32
150
150
160
220
220
35
150
173
175
228
228
40
160
190
200
250
250
50
220
250
250
300
300
63
280
320
315
500
500
80
380
425
400
500
520
100
480
580
500
600
650
625
750
820
125
715
160
950
200
1250
250
1650
315
2200
400
2840
500
3800
630
5100
800
7000
1000
9500
1250
106
Opening time = 400 ms
Nominal current
IN (A)
Slow-blow fuse
Iks max (A)
DIN gG/gL fuse
Iks max (A)
RCD LS-B
Iks max (A)
RCD LS-C
Iks max (A)
RCD LS-D
Iks max (A)
2
6
10
20
20
4
19
20
40
40
34
46
30
60
120
55
81
50
100
200
100
65
130
260
6
8
10
13
16
80
107
80
160
320
20
120
146
100
200
400
25
160
180
125
250
500
32
240
272
160
320
640
35
240
309
160
320
640
40
280
319
200
400
800
50
350
464
250
500
1000
63
510
545
315
630
1260
80
837
100
1018
125
1455
160
1678
200
2530
250
2918
315
4096
400
5451
500
7516
630
9371
800
107
Opening time = 200 ms
Nominal current
IN (A)
Slow-blow fuse
Iks max (A)
DIN gG/gL fuse
Iks max (A)
RCD LS-B
Iks max (A)
RCD LS-C
Iks max (A)
RCD LS-D
Iks max (A)
2
19
20
4
39
40
6
57
30
60
120
10
97
50
100
200
13
118
65
130
260
16
126
80
160
320
20
171
100
200
400
25
215
125
250
500
32
308
160
320
640
35
374
175
350
700
40
381
200
400
800
50
545
250
500
1000
63
663
315
630
1260
8
80
965
400
800
1600
100
1195
500
1000
2000
125
1708
625
1250
2500
160
2042
200
2971
250
3615
315
4985
400
6633
500
8825
630
108
Opening time = 100 ms
Nominal current
IN (A)
Slow-blow fuse
Iks max (A)
DIN gG/gL fuse
Iks max (A)
2
0
4
47
6
72
RCD LS-B
Iks max (A)
RCD LS-C
Iks max (A)
RCD LS-D
Iks max (A)
30
60
120
8
92
10
110
50
100
200
13
140,4
65
130
260
16
150
20
80
160
320
100
200
400
25
260
125
250
500
32
350
160
320
640
35
453,2
175
350
700
40
450
200
400
800
50
610
250
500
1000
63
820
315
630
1260
80
1100
400
800
1600
100
1450
500
1000
2000
125
1910
625
1250
2500
160
2590
200
3420
250
4500
315
6000
400
8060
500
109
Opening time = 35 ms
Nominal current
IN (A)
Slow-blow fuse
Iks max (A)
DIN gG/gL fuse
Iks max (A)
RCD LS-B
Iks max (A)
RCD LS-C
Iks max (A)
RCD LS-D
Iks max (A)
103
30
60
120
10
166
50
100
200
13
193
65
130
260
16
207
80
160
320
20
277
100
200
400
25
361
125
250
500
32
539
160
320
640
35
618
175
350
700
40
694
200
400
800
50
919
250
500
1000
63
1 217
315
630
1260
2
4
6
8
80
1 567
400
800
1600
100
2 075
500
1000
2000
125
2 826
625
1250
2500
160
3 538
200
4 556
250
6 032
315
7 767
400
110
111
© Chauvin Arnoux - All rights reserved and reproduction prohibited
694202D02 - Ed. 1 - 08/2024
FRANCE
Chauvin Arnoux
12-16 rue Sarah Bernhardt
92600 Asnières-sur-Seine
Tél : +33 1 44 85 44 85
Fax : +33 1 46 27 73 89
[email protected]
www.chauvin-arnoux.com
INTERNATIONAL
Chauvin Arnoux
Tél : +33 1 44 85 44 38
Fax : +33 1 46 27 95 69
Our international contacts
www.chauvin-arnoux.com/contacts

advertisement

Key Features

  • Portable
  • Color graphic display
  • Voltage, resistance, insulation, earth resistance
  • RCD testing
  • Safety testing
  • Data logging
  • USB connectivity
  • Intuitive interface

Frequently Answers and Questions

What types of electrical installations can I test with the CA 6116N - CA 6117?
These devices are designed to check electrical installations of various types, including new installations before they are powered up, existing installations, and troubleshooting malfunctioning installations.
What types of measurements can I perform with the CA 6116N - CA 6117?
You can perform a wide range of measurements, including voltage, resistance, insulation resistance, earth resistance, loop impedance, and RCD testing.
Can I connect the CA 6116N - CA 6117 to a computer?
Yes, the devices have a USB port for data transfer to a PC. You can use this to download data and store it for analysis.

Related manuals

Download PDF

advertisement