TSI VOC probe specs
VOC_Probes:US_5001365_RevB
E N E R G Y
1/9/2012
A N D
3:53 PM
Page 1
C O M F O R T
Ventilation Test Instruments
Volatile Organic Compounds (VOC)
Indoor Air Quality Probes
TSI VOC probes are pre-calibrated, plug-and-play accessory
probes for multi-purpose meters Models 7575 Q-TRAK monitor
and 9565 VELOCICALC meter. When combined with these meters,
long-term data analysis can be performed and reported which is
useful for monitoring or investigating the indoor environment.
Applications
Features and Benefits
IAQ investigations
Industrial hygiene surveys
Building commissioning
Track down emissions to their source
Pre-calibrated plug-and-play accessory probes for TSI Models
7575 Q-TRAK™ monitor and 9565 VELOCICALC® meter
Compact design with ergonomic handle
Data Collection and Reporting
Four versions available with multiple measurement capability
Expanded data logging
capacity and the inclusion of
TRAKPRO Data Analysis
Software provides the
capabilities to work more
effectively and efficiently. The Q-TRAK monitor can store up to
38.9 days of data collected at one-minute log intervals. The
stored data can be recalled, reviewed on screen, and downloaded
for easy reporting. TRAKPRO software helps you to generate
professional graphs for your reports.
– Model 984 - Low concentration (ppb) VOC and temperature
– Model 985 - High concentration (ppm) VOC and temperature
– Model 986 - Low concentration (ppb) VOC, temperature,
CO2 and humidity
– Model 987 - High concentration (ppm) VOC, temperature,
CO2 and humidity
Calculates VOC exposure in mass concentration
– Requires knowledge of VOC being measured
– Enter specific response factor into meter
Two-year factory warranty
Send only the probe back for factory calibration
VOC_Probes:US_5001365_RevB
E N E R G Y
1/9/2012
A N D
3:53 PM
Page 2
C O M F O R T
Ventilation Test Instruments
Field Service
TSI VOC probes are competitively priced to others in the marketplace
but feature a lower cost of ownership due its compact design and
construction which is geared for field service capability.
Field calibration
– Temperature
– Relative humidity
Specifications
– Volatile organic compounds (VOC)
VOC Probes for Multi-Purpose Meters
Models 984, 985, 986, and 987
– Carbon dioxide (CO2)
Replaceable sensors
Sensor maintenance
Model 984 Low Concentration (ppb) VOC and Temperature
Range
10 to 20,000 ppb, 14 to 140°F (-10 to 60°C)
Accuracy
±1.0°F (±0.5°C)1
Resolution
Up to 10 ppb, 0.1°F (0.1°C)
– Lamp cleaning
– Lamp replacement
Model 985 High Concentration (ppm) VOC and Temperature
Range
1 to 2,000 ppm, 14 to 140°F (-10 to 60°C)
Accuracy
±1.0°F (±0.5°C)1
Resolution
Up to 10 ppm, 0.1°F (0.1°C)
Instrumentation
The 7575 Q-TRAK monitor or 9565 VELOCICALC meter feature a menudriven user interface for easy operation. VOC probes are also
compatible with AIRFLOW™ Instruments TA460 series and TSI’s
discontinued 7565 and 9555 series, if instrument firmware is at
revision 2.10 or higher.
Model 986 Low Concentration (ppb) VOC, Temperature, CO2, and Humidity
Range
10 to 20,000 ppb VOC, 0 to 5,000 ppm CO2
14 to 140°F (-10 to 60°C), 5 to 95% RH
Accuracy
±3% of reading or 50 ppm CO2, whichever is greater
±1.0°F (±0.5°C)1, ±3% RH2
Resolution
Up to 10 ppb VOC, 0.1 ppm CO2, 0.1°F (0.1°C), 0.1% RH
On-screen prompts and step-by-step instructions guide the user
through operation and field calibration. These instruments also feature
an ergonomic, over molded case design and a keypad lockout to
prevent tampering during unattended use.
Model 987 High Concentration (ppm) VOC, Temperature, CO2, and Humidity
Range
1 to 2,000 ppm VOC, 0 to 5,000 ppm CO2
14 to 140°F (-10 to 60°C), 5 to 95% RH
Accuracy
±3% of reading or 50 ppm CO2, whichever is greater
±1.0°F (±0.5°C)1, ±3% RH2
Resolution
Up to 10 ppm VOC, 0.1 ppm CO2, 0.1°F (0.1°C), 0.1% RH
Display up to five measurements simultaneously
Log multiple parameters to investigate trends
Calculate dew point, wet bulb and percent outside air (VOC
models 986 and 987)
Store up to 38.9 days of data collected at one-minute
log intervals
User selectable logging intervals and start/stop times
Internal barometric pressure sensor
Probe Dimensions
Length
Base Diameter
Tip Diameter
Note: The 984 and 986 probes are designed to measure ppb concentrations of VOCs. The
10 to 20,000 ppb range corresponds to 0.01 to 20 ppm.
1
2
Download data to TRAKPRO™ data analysis software
– Report generation
Specifications are subject to change without notice.
.
– Instrument programming
USA
UK
France
Germany
India
China
Singapore
Accuracy with instrument case at 77°F (25°C), add uncertainty of 0.05°F/°F (0.03°C/°C) for
change in instrument temperature.
Accuracy with probe at 77°F (25°C). Add uncertainty of 0.1% RH/°F (0.2% RH/°C) for change in
probe temperature. Includes 1% hysteresis.
Q-TRAK, TRAKPRO, and VELOCICALC are trademarks of TSI Incorporated.
– Graph creation
TSI Incorporated
- 500 Cardigan Road, Shoreview, MN 55126-3996 USA
Tel: +1 800 874 2811
Tel: +44 149 4 459200
Tel: +33 491 11 87 64
Tel: +49 241 523030
Tel: +91 80 41132470
Tel: +86 10 8251 6588
Tel: +65 6595 6388
E-mail: answers@tsi.com Website: www.tsi.com
E-mail: tsiuk@tsi.com
Website: www.tsiinc.co.uk
E-mail: tsifrance@tsi.com Website: www.tsiinc.fr
E-mail: tsigmbh@tsi.com Website: www.tsiinc.de
E-mail: tsi-india@tsi.com
E-mail: tsibeijing@tsi.com
E-mail: tsi-singapore@tsi.com
Contact your local TSI Distributor or visit our website www.tsi.com for more detailed specifications.
P/N 5001365 Rev B
7.0 in. (17.8 cm)
0.75 in. (1.9 cm)
1.0 in. (2.54 cm)