M-System PTAF PT Transmitter Instruction manual

M-System PTAF PT Transmitter Instruction manual

Below you will find brief information for PT Transmitter PTAF. This document describes the necessary points of caution when using this product, including installation, connection, and basic maintenance procedures. It also covers component identification, terminal connections, adjustment procedures, checking points, and environmental considerations.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

PT Transmitter PTAF Instruction Manual | Manualzz
PTAF
INSTRUCTION MANUAL
PT TRANSMITTER
PTAF
MODEL
(high speed response)
■ FRONT PANEL CONFIGURATIONS
BEFORE USE ....
Thank you for choosing M-System. Before use, check the
package you received as below.
If you have any problems or questions with the product,
please contact M-System's Sales Office or representatives.
Zero Adj.
■ PACKAGE INCLUDES:
Signal conditioner (body + base socket) ................. (1)
Span Adj.
■ MODEL NO.
Check that model No. described on specification label is
exactly what you ordered.
Front Cover (open)
■ INSTRUCTION MANUAL
This manual describes necessary points of caution when you
use this product, installation, connection and basic maintenance procedures.
COMPONENT IDENTIFICATION
Body
Base Socket
Connection
Diagram Label
Front Cover
INSTALLATION
Detach the yellow clamps located at the top and bottom of the
unit for separating the body from the base socket.
■ DIN RAIL MOUNTING
Set the base socket so that its
DIN rail adaptor is at the bottom. Position the upper hook
at the rear side of base socket
on the DIN rail and push in
the lower. When removing the
socket, push down the DIN
rail adaptor utilizing a screwdriver (–) and pull.
Clamp
(top & bottom)
DIN Rail
35mm wide
Spring Loaded
DIN Rail Adaptor
■ WALL MOUNTING
Refer to the drawings in the Shape and size of the base socket
are slightly different with various
following page.
socket types.
Specification
Label
■ HOW TO OPEN THE FRONT COVER:
Position your finger on the hook at the top of front cover and
pull.
TERMINAL CONNECTIONS
Connect the unit as in the diagram below or refer to the
connection diagram label on top of the unit.
SOURCE
1 2
FUSE U PT u
V
LOAD
v
3
1 +
4
2 –
OUTPUT
7 U
POWER
The shape of base socket may be different
for some models.
8 V
P. 1 / 2
EM-1914
PTAF
CHECKING
POINTS OF CAUTION
1) Terminal wiring: Check that all cables are correctly
connected according to the connection diagram.
2) Power input voltage: Check voltage across the terminal 7
–␣ 8 with a multimeter.
3) Input: Check that the input signal is within 0 – 100% of
the full-scale.
4) Output: Check that the load resistance meets the described specifications.
■ GENERAL
• Before you remove the unit from its base socket or mount it,
turn off the power supply and input signal for safety.
ADJUSTMENT PROCEDURE
This unit is calibrated at the factory to meet the ordered
specifications, therefore you usually do not need any calibration.
For matching the signal to a receiving instrument or in case
of regular calibration, adjust the output as explained in the
following.
■ HOW TO CALIBRATE THE OUTPUT SIGNAL
Use a signal source and measuring instruments of sufficient
accuracy level. Turn the power supply on and warm up for
more than 10 minutes.
1) ZERO: Apply 10% input and adjust output to 10%.
2) SPAN: Apply 100% input and adjust output to 100%.
3) Check ZERO adjustment again with 10% input.
4) When ZERO value is changed, repeat the above procedure
1) – 3).
■ WIRING
• Do not install cables (power supply, input and output) close
to noise sources (relay drive cable, high frequency line, etc.).
• Do not bind these cables together with those in which noises
are present. Do not install them in the same duct.
■ AND ....
• The unit is designed to function as soon as power is
supplied, however, a warm up for 10 minutes is required for
satisfying complete performance described in the data sheet.
MAINTENANCE
Regular calibration procedure is explained below:
■ CALIBRATION
Warm up the unit for at least 10 minutes. Apply 10%, 25%,
50%, 75% and 100% input signal. Check that the output
signal for the respective input signal remains within accuracy described in the data sheet. When the output is out of
tolerance, recalibrate the unit according to the "ADJUSTMENT PROCEDURE" explained earlier.
mm (inch)
CLAMP
(top & bottom)
7.8 (.31)
20
(.79)
EXTERNAL DIMENSIONS
■ ENVIRONMENT
• When heavy dust or metal particles are present in the air,
install the unit inside proper housing and ventilate it.
• Do not install the unit where it is subjected to continuous
vibration. Do not apply physical impact to the unit.
• Environmental temperature must be within -5 to +60°C (23
to 140°F) with relative humidity within 30 to 90% RH in order
to ensure adequate life span and operation.
107 (4.21)
127 (5)
5
4
3
7
8
1
2
2–4.5 (.18) dia.
MTG HOLE
15 (.59) deep
8–M3.5
SCREW
50 (1.97)
6
80 (3.15)
80 (3.15)
35.4 (1.39)
DIN RAIL
35mm wide
40 (1.57)
[3.3 (.13)]
50 (1.97)
•When mounting, no extra space is needed between units.
M-SYSTEM WARRANTY
M-System warrants such new M-System product which it manufactures to be free from defects in materials and workmanship during the 36-month period following the date that such
product was originally purchased if such product has been used under normal operating conditions and properly maintained, M-System's sole liability, and purchaser's exclusive remedies,
under this warranty are, at M-System's option, the repair, replacement or refund of the purchase price of any M-System product which is defective under the terms of this warranty. To
submit a claim under this warranty, the purchaser must return, at its expense, the defective M-System product to the below address together with a copy of its original sales invoice.
THIS IS THE ONLY WARRANTY APPLICABLE TO M-SYSTEM PRODUCT AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. M-SYSTEM SHALL HAVE NO LIABILITY FOR CONSEQUENTIAL, INCIDENTAL OR
SPECIAL DAMAGES OF ANY KIND WHATSOEVER.
M-System Co., Ltd., 5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN, Phone: (06) 6659-8201, Fax: (06) 6659-8510, E-mail: [email protected]
P. 2 / 2
EM-1914

advertisement

Key Features

  • High speed response
  • DIN rail mounting
  • Wall mounting
  • Zero and Span adjustment
  • Calibration
  • Environmental protection
  • 36-month warranty

Frequently Answers and Questions

How do I install the PT Transmitter?
The PT Transmitter can be installed on a DIN rail or on a wall. Instructions for both installation methods are provided in the manual.
How do I calibrate the PT Transmitter?
The PT Transmitter is calibrated at the factory, but you can adjust the output signal for matching it to a receiving instrument or for regular calibration. The manual provides detailed instructions on how to calibrate the transmitter.
What are the environmental conditions for using the PT Transmitter?
The PT Transmitter can operate in an environment with temperatures ranging from -5 to +60°C (23 to 140°F) and relative humidity within 30 to 90% RH. It is recommended to install the unit inside proper housing and ventilate it when heavy dust or metal particles are present in the air.
What is the warranty on the PT Transmitter?
M-System provides a 36-month warranty on the PT Transmitter. This warranty covers any defects in materials and workmanship that occur under normal operating conditions and proper maintenance.
What are the terminal connections for the PT Transmitter?
The manual includes a connection diagram that shows the terminal connections for the power supply, input signal, output signal, and fuse.

Related manuals