- No category
advertisement
5522A Multi-Product
Calibrator
Extended specifications
General Specifications
The following tables list the 5522A specifications. All specifications are valid after allowing a warm-up period of
30 minutes, or twice the time the 5522A has been turned off. (For example, if the 5522A has been turned off for
5 minutes, the warm-up period is 10 minutes.)
All specifications apply for the temperature and time period indicated. For temperatures outside of tcal 5 C (tcal is the ambient temperature when the 5522A was calibrated), the temperature coefficient as stated in the General Specifications must be applied.
The specifications also assume the Calibrator is zeroed every seven days or whenever the ambient temperature changes more than 5 C. The tightest ohms specifications are maintained with a zero cal every 12 hours within 1 C of use.
Also see additional specifications later in this chapter for information on extended specifications for ac voltage and current.
Warmup Time .....................................................Twice the time since last warmed up, to a maximum of 30 minutes.
Settling Time ......................................................Less than 5 seconds for all functions and ranges except as noted.
Standard Interfaces ............................................IEEE-488 (GPIB), RS-232
Temperature
Operating .........................................................0 C to 50 C
Calibration (tcal) ..............................................15 C to 35 C
Storage ............................................................-20 to +70 C; The DC current ranges 0 to 1.09999 A and 1.1 A to
2.99999 A are sensitive to storage temperatures above 50 C. If the 5522A is stored above 50 C for greater than 30 minutes, these ranges must be re-calibrated. Otherwise, the 90 day and 1 year uncertainties of these ranges double.
Temperature Coefficient ....................................Temperature coefficient for temperatures outside tcal 5 C is
10 % of the stated specification per C.
Relative Humidity
Operating .........................................................<80 % to 30 C, <70 % to 40 C, <40 % to 50 C
Storage ............................................................<95 %, non-condensing. After long periods of storage at high humidity, a drying-out period (with power on) of at least one week may be required.
Altitude
Operating .........................................................3,050 m (10,000 ft) maximum at 120 V line operation
2000 m (6,500 ft) maximum at 120 V line operation
Non-operating .................................................12,200 m (40,000 ft) maximum
Safety ..................................................................IEC 61010-1: Overvoltage CATII, Pollution Degree 2
Output Terminal Electrical Overload Protection Provides reverse-power protection, immediate output disconnection, and/or fuse protection on the output terminals for all functions. This protection is for applied external voltages up to 300 V peak.
Analog Low Isolation .........................................20 V normal operation, 400 V peak transient
Electromagnetic Environment ...........................IEC 61326-1: Controlled
Electromagnetic Compatibility ..........................If used in areas with Electromagnetic fields of 1 to 3 V/m from
0.08-1GHz, resistance outputs have a floor adder of 0.508
Performance not specified above 3 V/m. This instrument may be susceptible to electro-static discharge (ESD) to the binding posts.
Good static awareness practices should be followed when handling this and other pieces of electronic equipment.
Additionally this instrument may be susceptible to electrical fast transients on the mains terminals. If any disturbances in operation are observed, it is recommended that the rear panel chassis ground terminal be connected to a known good earth ground with a low inductance ground strap. Note that a mains power outlet while providing a suitable ground for protection against electric shock hazard may not provide an adequate ground to properly drain away conducted rf disturbances and may in fact be the source of the disturbance. This instrument was certified for
EMC performance with data I/O cables not in excess of 3m.
2 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
Line Power ......................................................... Line Voltage (selectable): 100 V, 120 V, 220 V, 240 V
Line Frequency: 47 Hz to 63 Hz
Line Voltage Variation: 10 % about line voltage setting
For optimal performance at full dual outputs (e.g. 1000 V, 20 A) choose a line voltage setting that is 7.5 % from nominal.
Power Consumption .......................................... 600 VA
Dimensions (HxWxL) ........................................... 17.8 cm x 43.2 cm x 47.3 cm (7 in x 17 in x 18.6 in) Standard rack width and rack increment, plus 1.5 cm (0.6 in) for feet on bottom of unit.
Weight (without options) .................................... 22 kg (49 lb)
Absolute Uncertainty Definition ....................... The 5522A specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. You do not need to add anything to determine the total specification of the 5522A for the temperature range indicated.
Specification Confidence Level ........................ 99 %
Detailed Specifications
DC Voltage
Range
0 to 329.9999 mV
0 to 3.299999 V
0 to 32.99999 V
30 to 329.9999 V
100 to 1020.000 V
0 to 329.9999 mV
0.33 to 3.299999 V
3.3 to 7 V
Absolute Uncertainty, tcal 5 C
(ppm of output +V)
90 days
15 + 1
9 + 2
10 + 20
15 + 150
15 + 1500
300 + 350
300 + 350
Auxiliary Output (dual output mode only)
[2]
300 + 350
1 year
20 + 1
11 + 2
12 + 20
18 + 150
18 + 1500
400 + 350
400 + 350
400 + 350
Stability
24 hours, 1 C
(ppm of output
+V)
3 + 1
2 + 1.5
2 + 15
2.5 + 100
3 + 300
30 + 100
30 + 100
30 + 100
Resolution V Max Burden
0.1
1
10
100
1000
1
10
100
TC Simulate and Measure in Linear 10 V/C and 1 mV/C modes
[3]
40 + 3 50 + 3 5 + 2 0.1
65
10 mA
10 mA
5 mA
5 mA
5 mA
5 mA
5 mA
10
[1]
0 to 329.9999 mV
[1] Remote sensing is not provided. Output resistance is <5 m for outputs 0.33 V. The AUX output has an output resistance of <1 . TC simulation has an output impedance of 10 1 .
[2] Two channels of dc voltage output are provided.
[3] TC simulating and measuring are not specified for operation in electromagnetic fields above 0.4 v/m.
3 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
Range
0 to 329.9999 mV
0 to 3.299999 V
0 to 32.99999 V
30 to 329.9999 V
100 to 1020.000 V
0 to 329.9999 mV
0.33 to 3.299999 V
3.3 to 7 V
[1] Two channels of dc voltage output are provided.
Noise
Bandwidth 0.1 Hz to 10 Hz p-p
(ppm of output + floor)
0 + 1 V
0 + 10 V
0 + 100 V
10 + 1 mV
10 + 5 mV
Auxiliary Output (dual output mode only)
[1]
0 + 5 V
0 + 20 V
0 + 100 V
Bandwidth 10 Hz to 10 kHz rms
6 V
60 V
600 V
20 mV
20 mV
20 V
200 V
1000 V
DC Current
Range
0 to 329.999 A
0 to 3.29999 mA
0 to 32.9999 mA
0 to 329.999 mA
0 to 1.09999 A
1.1 to 2.99999 A
0 to 10.9999 A
(20 A Range)
11 to 20.5 A
[1]
Absolute Uncertainty, tcal 5 C
(ppm of output +A)
90 days
120 + 0.02
80 + 0.05
80 + 0.25
80 + 2.5
160 + 40
300 + 40
380 + 500
1 year
150 + 0.02
100 + 0.05
100 + 0.25
100 + 2.5
200 + 40
380 + 40
500 + 500
Resolution
1 nA
0.01 A
0.1 A
1 A
10 A
10 A
100 A
Max Compliance
10
10
7
7
6
6
4
Voltage V
Max Inductive
Load mH
400
800 + 750
[2]
1000 + 750
[2]
100 A 4
[1] Duty Cycle: Currents <11 A may be provided continuously. For currents >11 A, see Figure 1. The current may be provided
Formula 60-T-I minutes any 60 minute period where T is the temperature in C (room temperature is about 23 C) and I is the output current in amperes. For example, 17 A, at 23 C could be provided for 60-23-17 = 20 minutes each hour. When the 5522A is outputting currents between 5 and 11 amps for long periods, the internal self-heating reduces the duty cycle. Under those conditions, the allowable "on" time indicated by the formula and Figure 1 is achieved only after the
5522A is outputting currents <5 A for the "off" period first.
[2] Floor specification is 1500 A within 30 seconds of selecting operate. For operating times >30 seconds, the floor specification is 750 A.
Range
0 to 329.999 A
0 to 3.29999 mA
0 to 32.9999 mA
0 to 329.999 mA
0 to 2.99999 A
0 to 20.5 A
Bandwidth 0.1 Hz to 10 Hz p-p
2 nA
20 nA
200 nA
2000 nA
20 A
200 A
Noise
Bandwidth 10 Hz to 10 kHz rms
20 nA
200 nA
2.0 A
20 A
1 mA
10 mA
4 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
50
45
40
35
30
25
20
15
10
5
0
11 12
40 °C
30 °C
20 °C
10 °C
Ambient
0 °C
13 14 15 16
Current (Amps)
17 18
Figure 1. Allowable Duration of Current >11 A
19 20
40%
30%
20%
80%
70%
60%
50%
10%
0%
5 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
Resistance
Range
[1]
Absolute Uncertainty, tcal 5 C (ppm of output +floor) ppm of output
90 days 1 year
Floor ()
[2]
Time and temp since ohms zero cal
12 hrs 1 C 7 days 5 C
Resolution
Allowable Current
0 to
10.9999
11 to
32.9999
33 to
109.9999
110 to
329.9999
330 to
1.099999 k
1.1 to
3.299999 k
3.3 to
10.99999 k
11 to
32.99999 k
33 to
109.9999 k
110 to
329.99999 k
35 40 0.001 0. 01
25
22
22
22
22
22
22
22
25
30
28
28
28
28
28
28
28
32
0.0015
0.0014
0.002
0.002
0.02
0.02
0.2
0.2
2
0.015
0.015
0.02
0.02
0.2
0.1
1
1
10
0.0001
0.0001
0.0001
0.0001
0.001
1 mA to 125 mA
1 mA to 125 mA
1 mA to 70 mA
1 mA to 40 mA
1 mA to 18 mA
1.099999 M
1.1 to
3.299999 M
3.3 to
10.99999 M
11 to
32.99999 M
33 to
109.9999 M
110 to
329.9999 M
330 to
1100 M
25
40
110
200
400
2500
12000
32
60
130
250
500
3000
15000
[1] Continuously variable from 0 �to 1.1 G.
2
30
50
2500
3000
100000
500000
10
150
250
2500
3000
100000
500000
1
10
10
100
1000
10000
250 nA to 5 A
250 nA to 1.8 A
25 nA to 500 nA
25 nA to 180 nA
2.5 nA to 50 nA
1 nA to 13 nA
[2] Applies for 4-WIRE compensation only. For 2-WIRE and 2-WIRE COMP, add an additional amount to the floor specification as calculated by: (5 V divided by the stimulus current in amps). For example, in 2-WIRE mode, at 1 k the floor specification within 12 hours of an ohms zero cal for a measurement current of 1 mA is: 0.002 + (5 V / 1 mA) = (0.002 + 0.005) =
0.007 .
[3] For currents lower than shown, the floor adder increases by Floor(new) = Floor(old) x Imin/Iactual. For example, a 50 A stimulus measuring 100 has a floor specification of: 0.0014 x 1 mA/50 A = 0.028 assuming an ohms zero calibration within 12 hours.
6 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
AC Voltage (Sine Wave)
Range
1.0 mV to
32.999 mV
33 mV to
329.999 mV
0.33 V to
3.29999 V
3.3 V to
32.9999 V
33 V to
329.999 V
330 V to
1020 V
Frequency
10 Hz to 45 Hz
45 Hz to 10 kHz
10 kHz to 20 kHz
20 kHz to 50 kHz
600 + 6
120 + 6
160 + 6
800 + 6
50 kHz to 100 kHz 3000 + 12
100 kHz to 6000 + 50
500 kHz
10 Hz to 45 Hz
45 Hz to 10 kHz
10 kHz to 20 kHz
20 kHz to 50 kHz
50 kHz to 100 kHz
100 kHz to
500 kHz
250 + 8
140 + 8
150 + 8
300 + 8
600 + 32
1600 + 70
10 Hz to 45 Hz
45 Hz to 10 kHz
10 kHz to 20 kHz
20 kHz to 50 kHz
250 + 50
140 + 60
160 + 60
250 + 50
50 kHz to 100 kHz 550 + 125
100 kHz to 2000 + 600
500 kHz
10 Hz to 45 Hz
45 Hz to 10 kHz
10 kHz to 20 kHz
20 kHz to 50 kHz
50 kHz to 100 kHz 750 + 1600
45 Hz to 1 kHz
1 kHz to 10 kHz
Absolute Uncertainty, tcal 5 C
(ppm of output + V)
90 days
250 + 650
125 + 600
220 + 600
300 + 600
150 + 2000
160 + 6000
10 kHz to 20 kHz 220 + 6000
20 kHz to 50 kHz 240 + 6000
1 year
Normal Output
800 + 6
150 + 6
200 + 6
1000 + 6
3500 + 12
8000 + 50
300 + 8
145 + 8
160 + 8
350 + 8
800 + 32
2000 + 70
300 + 50
150 + 60
190 + 60
300 + 50
700 + 125
2400 + 600
300 + 650
150 + 600
240 + 600
350 + 600
900 + 1600
190 + 2000
200 + 6000
250 + 6000
300 + 6000
50 kHz to 100 kHz 1600 + 50000 2000 + 50000
45 Hz to 1 kHz
1 kHz to 5 kHz
5 kHz to 10 kHz
250 + 10000 300 + 10000
200 + 10000 250 + 10000
250 + 10000 300 + 10000
Resolution
1 V
1 V
10 V
100 V
1 mV
10 mV
Max
Burden
65
65
10 mA
10 mA except
20 mA for
2 mA, except
6 mA for
45 Hz to
65 Hz
Max Distortion and
Noise
10 Hz to 5 MHz
Bandwidth
(% of output + floor)
0.15 + 90 V
0.035 + 90 V
0.06 + 90 V
0.15 + 90 V
0.25 + 90 V
0.3 + 90 V
0.8 + 10 mV
1.0 + 10 mV
[1] Max Distortion for 100 kHz to 200 kHz. For 200 kHz to 500 kHz, the maximum distortion is 0.9 % of output + floor as shown.
[1]
0.15 + 90 V
0.035 + 90 V
0.06 + 90 V
0.15 + 90 V
0.20 + 90 V
0.20 + 90 V
[1]
0.15 + 200 V
0.035 + 200 V
0.06 + 200 V
0.15 + 200 V
0.20 + 200 V
0.20 + 200 V
0.15 + 2 mV
0.035 + 2 mV
0.08 + 2 mV
0.2 + 2 mV
0.5 + 2 mV
0.15 + 10 mV
0.05 +10 mV
0.6 + 10 mV
0.15 + 30 mV
0.07 + 30 mV
0.07 + 30 mV
[1]
Note
Remote sensing is not provided. Output resistance is <5 m for outputs 0.33 V. The AUX output resistance is <1 . The maximum load capacitance is 500 pF, subject to the maximum burden current limits
7 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
AC Voltage (Sine Wave) (cont.)
Range Frequency
[1]
Absolute Uncertainty, tcal 5 C
(% of output + V)
90 days 1 year
Resolution
Max
Burden
Max Distortion and
Noise
10 Hz to 5 MHz
Bandwidth
(% of output + floor)
10 mV to
329.999 mV
0.33 V to
3.29999 V
3.3 V to 5 V
10 Hz to 20 Hz
20 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 5 kHz
5 kHz to 10 kHz
10 kHz to 30 kHz
10 Hz to 20 Hz
20 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 5 kHz
10 Hz to 20 Hz
20 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 5 kHz
5 kHz to 10 kHz
0.15 + 370
0.08 + 370
0.08 + 370
0.15 + 450
0.3 + 450
4.0 + 900
0.15 + 450
0.08 + 450
0.07 + 450
0.15 + 1400
5 kHz to 10 kHz 0.3 + 1400
10 kHz to 30 kHz 4.0 + 2800
0.15 + 450
0.08 + 450
0.07 + 450
0.15 + 1400
0.3 + 1400
AUX Output
0.2 + 370
0.1 + 370
0.1 + 370
0.2 + 450
0.4 + 450
5.0 + 900
0.2 + 450
0.1 + 450
0.09 + 450
0.2 + 1400
0.4 + 1400
5.0 + 2800
0.2 + 450
0.1 + 450
0.09 + 450
0.2 + 1400
0.4 + 1400
1 V
10 V
100 V
5 mA
5 mA
5 mA
[1] There are two channels of voltage output. The maximum frequency of the dual output is 30 kHz.
Note
Remote sensing is not provided. Output resistance is <5 m for outputs 0.33 V. The AUX output resistance is <1 . The maximum load capacitance is 500 pF, subject to the maximum burden current limits
0.2 + 200 V
0.06 + 200 V
0.08 + 200 V
0.3 + 200 V
0.6 + 200 V
1 + 200 V
0.2 + 200 V
0.06 + 200 V
0.08 + 200 V
0.3 + 200 V
0.6 + 200 V
1 + 200 V
0.2 + 200 V
0.06 + 200 V
0.08 + 200 V
0.3 + +200 V
0.6 + 200 V
8 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
AC Current (Sine Wave)
Range Frequency
Absolute Uncertainty, tcal 5 C
(% of output + A)
Compliance adder (A/V)
Max Distortion
& Noise 10 Hz to 100 kHz BW
(% of output + floor)
Max
Inductive
Load H
29.00 to
329.99 A
0.33 to
3.29999 mA
3.3 to
32.9999 mA
33 to
329.999 mA
10 to 20 Hz
20 to 45 Hz
45 Hz to 1 kHz
1 to 5 kHz
5 to 10 kHz
10 to 30 kHz
10 to 20 Hz
20 to 45 Hz
45 Hz to 1 kHz
1 to 5 kHz
5 to 10 kHz
10 to 30 kHz
10 to 20 Hz
20 to 45 Hz
45 Hz to 1 kHz
1 to 5 kHz
5 to 10 kHz
10 to 30 kHz
10 to 20 Hz
20 to 45 Hz
45 Hz to 1 kHz
1 to 5 kHz
5 to 10 kHz
10 to 30 kHz
10 to 45 Hz
45 Hz to 1 kHz
1 to 5 kHz
90 days
0.16 + 0.1
0.12 + 0.1
0.1 + 0.1
0.25 + 0.15
0.6 + 0.2
1.2 + 0.4
0.16 + 0.15
0.1 + 0.15
0.08 + 0.15
0.16 + 0.2
0.4 + 0.3
0.8 + 0.6
0.15 + 2
0.075 + 2
0.035 + 2
0.065 + 2
0.16 + 3
0.32 + 4
0.15 + 20
0.075 + 20
0.035 + 20
0.08 + 50
0.16 + 100
0.32 + 200
0.15 + 100
0.036 + 100
0.5 + 1000
1 year
LCOMP Off
0.2 + 0.1
0.15 + 0.1
0.125 + 0.1
0.3 + 0.15
0.8 + 0.2
1.6 + 0.4
0.2 + 0.15
0.125 + 0.15
0.1 + 0.15
0.2 + 0.2
0.5 + 0.3
1.0 + 0.6
0.18 + 2
0.09 + 2
0.04 + 2
0.08 + 2
0.2 + 3
0.4 + 4
0.18 + 20
0.09 + 20
0.04 + 20
0.10 + 50
0.2 + 100
0.4 + 200
0.18 + 100
0.05 + 100
0.6 + 1000
0.05
0.05
0.05
1.5
1.5
10
0.05
0.05
0.05
1.5
1.5
10
0.05
0.05
0.05
1.5
1.5
10
0.05
0.05
0.05
1.5
1.5
10
0.15 + 0.5 A
0.1 + 0.5 A
0.05 + 0.5 A
0.5 + 0.5 A
1.0 + 0.5 A
1.2 + 0.5 A
0.15 + 1.5 A
0.06 + 1.5 A
0.02 + 1.5 A
0.5 + 1.5 A
1.0 + 1.5 A
1.2 + 0.5 A
0.15 + 5 A
0.05 + 5 A
0.07 + 5 A
0.3 + 5 A
0.7 + 5 A
1.0 + 0.5 A
0.15 + 50 A
0.05 + 50 A
0.02 + 50 A
0.03 + 50 A
0.1 + 50 A
0.6 + 50 A
0.2 + 500 A
0.07 + 500 A
1 + 500 A
200
200
50
50
0.33 to
1.09999 A
[2]
2.5
5 to 10 kHz
10 to 45 Hz
45 Hz to 1 kHz
2.0 + 5000
0.15 + 100
0.05 + 100
2.5 + 5000
0.18 + 100
0.06 + 100
[3]
2 + 500 A
0.2 + 500 A
0.07 + 500 A
1.1 to
2.99999 A
1 to 5 kHz
5 to 10 kHz
0.5 + 1000
2.0 + 5000
0.6 + 1000
2.5 + 5000
[2]
[3]
1 + 500 A
2 + 500 A
2.5
3 to
10.9999 A
11 to
20.5 A
[1]
45 to 100 Hz
100 Hz to 1 kHz
1 to 5 kHz
45 to 100 Hz
100 Hz to 1 kHz
1 to 5 kHz
0.05 + 2000 0.06 + 2000
0.08 + 2000 0.10 + 2000
2.5 + 2000
0.1 + 5000
3.0 + 2000
0.12 + 5000
0.13 + 5000 0.15 + 5000
2.5 + 5000 3.0 + 5000
0.2 + 3 mA
0.1 + 3 mA
0.8 + 3 mA
0.2 + 3 mA
0.1 + 3 mA
0.8 + 3 mA
1
1
[1] Duty Cycle: Currents <11 A may be provided continuously. For currents >11 A, see Figure 1. The current may be provided 60-T-I minutes any 60 minute period where T is the temperature in C (room temperature is about 23 C) and I is the output current in Amps. For example, 17 A, at 23 C could be provided for 60-23-17 = 20 minutes each hour.
When the 5522A is outputting currents between 5 and 11 amps for long periods, the internal self-heating reduces the duty cycle. Under those conditions, the allowable "on" time indicated by the formula and Figure 1 is achieved only after the 5522A is outputting currents <5 A for the "off" period first.
[2] For compliance voltages greater than 1 V, add 1 mA/V to the floor specification from 1 to 5 kHz.
[3] For compliance voltages greater than 1 V, add 5 mA/V to the floor specification from 5 to 10 kHz.
9 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
AC Current (Sine Wave) (cont.)
Range
29.00 to
329.99 A
0.33 to
3.29999 mA
3.3 to
32.9999 mA
33 to
329.999 mA
0.33 to
2.99999 A
3 to 20.5 A
[1]
Frequency
10 to 100 Hz
100 Hz to 1 kHz
10 to 100 Hz
100 Hz to 1 kHz
10 to 100 Hz
100 Hz to 1 kHz
10 to 100 Hz
100 Hz to 1 kHz
10 to 100 Hz
100 to 440 Hz
45 to 100 Hz
100 to 440 Hz
Absolute Uncertainty,
(% of output + A)
90 days tcal 5 C
1 year
LCOMP On
0.2 + 0.2
0.5 + 0.5
0.2 + 0.3
0.5 + 0.8
0.07 + 4
0.18 + 10
0.07 + 40
0.18 + 100
0.1 + 200
0.25 + 1000
0.1 + 2000
0.8 + 5000
[2]
[3]
0.25 + 0.2
0.6 + 0.5
0.25 + 0.3
0.6 + 0.8
0.08 + 4
0.2 + 10
0.08 + 40
0.2 + 100
0.12 + 200
0.3 + 1000
0.12 + 2000
[2]
1.0 + 5000
[3]
Max Distortion &
Noise 10 Hz to
100 kHz BW
(% of output + floor)
0.1 + 1.0 A
0.05 + 1.0 A
0.15 + 1.5 A
0.06 + 1.5 A
0.15 + 5 A
0.05 + 5 A
0.15 + 50 A
0.05 + 50 A
0.2 + 500 A
0.25 + 500 A
0.1 + 0 A
0.5 + 0 A
Max
Inductive
Load H
400
400
[4]
[1] Duty Cycle: Currents <11 A may be provided continuously. For currents >11 A, see Figure 1. The current may be provided Formula 60-T-I minutes any 60 minute period where T is the temperature in C (room temperature is about
23 C) and I is the output current in Amps. For example, 17 A, at 23 C could be provided for 60-23-17 = 20 minutes each hour. When the 5522A is outputting currents between 5 and 11 amps for long periods, the internal self-heating reduces the duty cycle. Under those conditions, the allowable "on" time indicated by the formula and Figure 1 is achieved only after the 5522A is outputting currents <5 A for the "off" period first.
[2] For currents >11 A, Floor specification is 4000 A within 30 seconds of selecting operate. For operating times
>30 seconds, the floor specification is 2000 A.
[3] For currents >11 A, Floor specification is 10000 A within 30 seconds of selecting operate. For operating times
>30 seconds, the floor specification is 5000 A.
[4] Subject to compliance voltages limits.
Range
0.029 to 0.32999 mA
0.33 to 3.29999 mA
3.3 to 32.9999 mA
33 to 329.999 mA
0.33 to 2.99999 A
3 to 20.5 A
Resolution A
0.01
0.01
0.1
1
10
100
[1] Subject to specification adder for compliance voltages greater than 1 V rms.
Max Compliance Voltage V rms
[1]
7
7
5
5
4
3
10 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
Capacitance
Range
Absolute Uncertainty, tcal 5 C
(% of output + floor)
[1] [2] [3]
Allowed Frequency or
Charge-Discharge Rate
90 days 1 year
Resolution
Min and Max to
Meet
Specification
Typical Max for
<0.5 % Error
Typical Max for
<1 % Error
220.0 to
399.9 pF
0.4 to
1.0999 nF
1.1 to
3.2999 nF
0.38 + 10 pF 0.5 + 10 pF
0.38 + 0.01 nF 0.5 + 0.01 nF
0.38 + 0.01 nF 0.5 + 0.01 nF
0.1 pF
0.1 pF
0.1 pF
10 Hz to 10 kHz
10 Hz to 10 kHz
10 Hz to 3 kHz
20 kHz
30 kHz
30 kHz
40 kHz
50 kHz
50 kHz
10.9999 nF
11 to
32.9999 nF
33 to
109.999 nF
110 to
329.999 nF
0.33 to
1.09999 F
1.1 to
3.29999 F
3.3 to
10.9999 F
11 to
32.9999 F
33 to
109.999 F
110 to
329.999 F
0.33 to
1.09999 mF
1.1 to
3.29999 mF
3.3 to
10.9999 mF
11 to
32.9999 mF
33 to
110 mF
0.19 + 0.01 nF 0.25 + 0.01 nF
0.19 + 0.1 nF 0.25 + 0.1 nF
0.19 + 0.1 nF 0.25 + 0.1 nF
0.19 + 0.3 nF 0.25 + 0.3 nF
0.19 + 1 nF
0.19 + 3 nF
0.19 + 10 nF
0.30 + 30 nF
0.34 + 100 nF 0.45 + 100 nF
0.34 + 300 nF 0.45 + 300 nF
0.34 + 1 F
0.34 + 3 F
0.34 + 10 F
0.7 + 30 F
1.0 + 100 F
0.25 + 1 nF
0.25 + 3 nF
0.25 + 10 nF
0.40 + 30 nF
0.45 + 1 F
0.45 + 3 F
0.45 + 10 F
0.75 + 30 F
0.1 pF
0.1 pF
1 pF
1 pF
10 pF
10 pF
100 pF
100 pF
1 nF
1 nF
10 nF
10 nF
100 nF
100 nF
1.1 + 100 F 10
[1] The output is continuously variable from 220 pF to 110 mF.
10 Hz to 1 kHz
10 Hz to 1 kHz
10 Hz to 1 kHz
10 Hz to 1 kHz
10 to 600 Hz
10 to 300 Hz
10 to 150 Hz
10 to 120 Hz
10 to 80 Hz
0 to 50 Hz
0 to 20 Hz
0 to 6 Hz
0 to 2 Hz
0 to 0.6 Hz
0 to 0.2 Hz
20 kHz
8 kHz
4 kHz
2.5 kHz
1.5 kHz
800 Hz
450 Hz
250 Hz
150 Hz
80 Hz
45 Hz
30 Hz
15 Hz
7.5 Hz
3 Hz
25 kHz
10 kHz
6 kHz
3.5 kHz
2 kHz
1 kHz
650 Hz
350 Hz
200 Hz
120 Hz
65 Hz
40 Hz
20 Hz
10 Hz
5 Hz
[2] Specifications apply to both dc charge/discharge capacitance meters and ac RCL meters. The maximum allowable peak voltage is 3 V. The maximum allowable peak current is 150 mA, with an rms limitation of 30 mA below 1.1 F and 100 mA for 1.1 F and above.
[3] The maximum lead resistance for no additional error in 2-wire COMP mode is 10 .
11 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
Temperature Calibration (Thermocouple)
TC
Type
B
C
E
J
K
Range
C
[2]
600 to 800
800 to 1000
1000 to 1550
1550 to 1820
0 to 150
150 to 650
650 to 1000
1000 to 1800
1800 to 2316
-250 to -100
-100 to -25
-25 to 350
350 to 650
650 to 1000
-210 to -100
-100 to -30
-30 to 150
150 to 760
760 to 1200
-200 to -100
-100 to -25
-25 to 120
120 to 1000
1000 to 1372
Absolute Uncertainty
Source/Measure tcal 5 C
C
[3]
90 days
0.42
0.34
0.30
0.26
0.23
0.19
0.23
0.38
0.63
0.38
0.12
0.10
0.12
0.16
0.20
0.12
0.10
0.13
0.18
0.25
0.14
0.12
0.19
0.30
1 year
0.44
0.34
0.30
0.33
0.30
0.26
0.31
0.50
0.84
0.50
0.16
0.14
0.16
0.21
0.27
0.16
0.14
0.17
0.23
0.33
0.18
0.16
0.26
0.40
TC
Type
L
N
R
S
T
U
Range
C
[2]
-200 to -100
-100 to 800
800 to 900
-200 to -100
-100 to -25
-25 to 120
120 to 410
410 to 1300
0 to 250
250 to 400
400 to 1000
1000 to 1767
0 to 250
250 to 1000
1000 to 1400
1400 to 1767
-250 to -150
-150 to 0
0 to 120
120 to 400
-200 to 0
0 to 600
[1] Temperature standard ITS-90 or IPTS-68 is selectable.
TC simulating and measuring are not specified for operation in electromagnetic fields above 0.4 V/m.
0.34
0.48
0.18
0.12
0.10
0.56
0.27
0.21
0.48
0.28
0.26
0.30
0.47
0.30
0.28
Absolute Uncertainty
Source/Measure tcal 5 C
C
[3]
90 days
0.37
0.26
0.17
0.30
0.17
0.15
0.14
0.27
0.57
0.35
0.33
0.40
0.47
0.36
0.37
1 year
0.37
0.26
0.17
0.40
0.22
0.19
0.18
0.46
0.63
0.24
0.16
0.14
0.56
0.27
[2] Resolution is 0.01 C
[3] Does not include thermocouple error
12 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
Temperature Calibration (RTD)
RTD Type
Range
C
[1]
Pt 385,
100
Pt 3926,
100
Pt 385,
200
-200 to -80
-80 to 0
0 to 100
100 to 300
300 to 400
400 to 630
630 to 800
-200 to -80
-80 to 0
0 to 100
100 to 300
300 to 400
400 to 630
-200 to -190
-190 to -80
-80 to 0
0 to 100
100 to 260
260 to 300
300 to 400
400 to 600
600 to 630
-200 to -80
-80 to 0
0 to 100
100 to 260
260 to 300
300 to 400
400 to 600
600 to 630
[1] Resolution is 0.003 C
Absolute Uncertainty
0.04
0.05
0.07
0.08
0.09
0.10
0.25
0.04
tcal 5 C
90 days
0.04
0.05
0.07
0.08
0.09
0.10
0.21
0.05
0.06
0.06
0.07
0.08
0.08
0.21
0.03
0.03
0.04
0.04
0.11
0.12
0.12
0.14
C
1 year
0.05
0.05
0.07
0.09
0.10
0.12
0.23
0.05
0.05
0.07
0.09
0.10
0.12
0.25
0.04
0.05
0.06
0.07
0.08
0.09
0.10
0.23
0.04
0.04
0.04
0.05
0.12
0.13
0.14
0.16
RTD Type
Pt 385,
500
Pt 385,
1000
PtNi 385,
120
(Ni120)
Cu 427
10
Range
C
[1]
-200 to -80
-80 to 0
0 to 100
100 to 260
260 to 300
300 to 400
400 to 600
600 to 630
-200 to -80
-80 to 0
0 to 100
100 to 260
260 to 300
300 to 400
400 to 600
600 to 630
-80 to 0
0 to 100
100 to 260
-100 to 260
Absolute Uncertainty
tcal 5 C
C
90 days
0.03
0.04
0.05
0.06
0.07
0.07
0.08
0.09
0.03
0.03
0.03
0.04
0.05
0.05
0.06
0.22
0.06
0.07
0.13
0.3
0.11
0.03
0.03
0.04
0.05
0.06
0.07
0.07
0.23
0.08
0.08
0.14
1 year
0.04
0.05
0.05
0.06
0.08
0.08
0.09
0.3
[2] Applies for COMP OFF (to the 5522A Calibrator front panel NORMAL terminals) and 2-wire and 4-wire compensation.
[3] Based on MINCO Application Aid No. 18
13 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
DC Power Specification Summary
90 days
1 year
Voltage Range
33 mV to 1020 V
33 mV to 1020 V
0.33 to
329.99 mA
3 to
20.5 A
Absolute Uncertainty, tcal 5 C, (% of watts output)
0.021
0.023
Current Range
0.33 to
2.9999 A
0.019
[2]
0.022
[2]
[1]
0.06
0.07
[2]
[2]
[1] To determine dc power uncertainty with more precision, see the individual “DC Voltage Specifications,” “DC Current
Specifications,” and “Calculating Power Uncertainty.”
[2] Add 0.02 % unless a settling time of 30 seconds is allowed for output currents >10 A or for currents on the highest two current ranges within 30 seconds of an output current >10 A.
AC Power (45 Hz to 65 Hz) Specification Summary, PF=1
90 days
1 year
90 days
1 year
Voltage Range
33 to 329.999 mV
330 mV to 1020 V
33 to 329.999 mV
330 mV to 1020 V
Voltage Range
33 to 329.999 mV
330 mV to 1020 V
33 to 329.999 mV
330 mV to 1020 V
3.3 to
8.999 mA
9 to
Current Range
33 to
32.999 mA 89.99 mA
90 to 329.99 mA
Absolute Uncertainty, tcal 5 C, (% of watts output)
0.13
0.11
0.14
0.12
0.09
0.07
0.10
0.08
Current Range
[2]
0.13
0.11
0.14
0.12
0.33 to
0.8999 A
0.9 to
2.1999 A
2.2 to
4.4999 A
[1]
0.09
0.07
0.10
0.08
Absolute Uncertainty, tcal 5 C, (% of watts output)
0.12
0.10
0.13
0.11
0.10
0.08
0.11
0.09
0.12
0.11
0.13
0.12
4.5 to
20.5 A
[1]
0.10
0.09
0.11
0.10
[1] To determine ac power uncertainty with more precision, see the individual “AC Voltage Specifications” and “AC Current
Specifications” and “Calculating Power Uncertainty.”
[2] Add 0.02 % unless a settling time of 30 seconds is allowed for output currents >10 A or for currents on the highest two current ranges within 30 seconds of an output current >10 A.
Power and Dual Output Limit Specifications
Frequency dc
10 to 45 Hz
45 to 65 Hz
65 to 500 Hz
65 to 500 Hz
500 Hz to 1 kHz
1 to 5 kHz
5 to 10 kHz
10 to 30 kHz
Notes
Voltages
(NORMAL)
0 to 1020 V
33 mV to 32.9999 V
33 mV to 1020 V
330 mV to 1020 V
3.3 to 1020 V
330 mV to 1020 V
3.3 to 500 V
3.3 to 250 V
3.3 V to 250 V
Currents
0 to 20.5 A
3.3 mA to 2.99999 A
3.3 mA to 20.5 A
33 mA to 2.99999 A
33 mA to 20.5 A
33 mA to 20.5 A
33 mA to 2.99999 A
33 to 329.99 mA
33 mA to 329.99 mA
Voltages
(AUX)
0 to 7 V
10 mV to 5 V
10 mV to 5 V
100 mV to 5 V
100 mV to 5 V
100 mV to 5 V
100 mV to 5 V
1 to 5 V
1 V to 3.29999 V
Power Factor
(PF)
0 to 1
0 to 1
0 to 1
0 to 1
0 to 1
0 to 1
0 to 1
0 to 1
The range of voltages and currents shown in “DC Voltage Specifications,” “DC Current Specifications,” “AC Voltage (Sine Wave)
Specifications,” and “AC Current (Sine Wave) Specifications” are available in the power and dual output modes (except minimum current for ac power is 0.33 mA). However, only those limits shown in this table are specified. See “Calculating
Power Uncertainty” to determine the uncertainty at these points.
The phase adjustment range for dual ac outputs is 0 to 179.99 . The phase resolution for dual ac outputs is 0.01 degree.
14 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
Phase
10 to
65 Hz
0.10
Note
65 to
500 Hz
0.25
1-Year Absolute Uncertainty, tcal 5 C, ( )
500 Hz to
1 kHz
0.5
1 to
5 kHz
See Power and Dual Output Limit Specifications for applicable outputs.
2.5
5 to
10 kHz
5
10 to
30 kHz
10
Phase ()
Watts
0
Phase ()
VARs
90
10 80
30 60
50 40
70 20
80 10
90 0
PF
1.000
0.985
0.940
0.866
0.766
0.643
0.500
0.342
10 to
65 Hz
0.00 %
0.03 %
0.06 %
0.10 %
0.15 %
0.21 %
0.30 %
0.48 %
0.174 0.99 %
0.000 ––
Power Uncertainty Adder due to Phase Error
65 to
500 Hz
0.00 %
0.08 %
0.16 %
0.25 %
0.37 %
0.52 %
0.76 %
1.20 %
2.48 %
––
500 Hz to
1 kHz
0.00 %
0.16 %
0.32 %
0.51 %
0.74 %
1.04 %
1.52 %
2.40 %
4.95 %
––
1 to
5 kHz
0.10 %
0.86 %
1.68 %
2.61 %
3.76 %
5.29 %
7.65 %
12.08 %
24.83 %
––
5 to
10 kHz
0.38 %
1.92 %
3.55 %
5.41 %
7.69 %
10.77 %
15.48 %
24.33 %
49.81 %
––
10 to
30 kHz
1.52 %
4.58 %
7.84 %
11.54 %
16.09 %
22.21 %
31.60 %
49.23 %
100.00 %
––
To calculate exact ac Watts power adders due to phase uncertainty for values not shown, use the following formula:
Adder
100 ( 1
Cos
(
Cos
)
)
For example: At 60 Hz, for a PF of .9205 ( = 23) and a phase uncertainty of adder is:
= 0.10, the ac Watts power
Adder
100 ( 1
Cos
(23
Cos
.
10 )
Calculating Power Uncertainty
) 0 .
074 %
Overall uncertainty for power output in Watts (or VARs) is based on the root sum square (rss) of the individual uncertainties in percent for the selected voltage, current, and power factor parameters:
Watts uncertainty
U power
U
2 voltage
U
2 current
U
2
PFadder
VARs uncertainty
U
VARs
U
2 voltage
U
2 current
U
2
VARsadder
Because there are an infinite number of combinations, you should calculate the actual ac power uncertainty for your selected parameters. The method of calculation is best shown in the following examples (using 1 year specifications):
Example 1 Output: 100 V, 1 A, 60 Hz, Power Factor = 1.0 (=0).
Voltage Uncertainty Uncertainty for 100 V at 60 Hz is 190 ppm + 2 mV, totaling:
100 V x 190 x 10
-6
= 19 mV added to 2 mV = 21 mV. Expressed in percent:
21 mV/100 V x 100 = 0.021 % (see “AC Voltage (Sine Wave) Specifications”).
Current Uncertainty Uncertainty for 1 A is 0.05 % �100 A, totaling:
1 A x 0.0005 = 500 A added to 100 A = 0.6 mA. Expressed in percent:
0. 6 mA/1 A x 100 = 0.06 % (see “AC Current (Sine Waves) Specifications”).
PF Adder Watts Adder for PF = 1 (=0) at 60 Hz is 0 % (see “Phase Specifications”).
Total Watts Output Uncertainty = U power
0 .
021
2
0 .
06 2 0 2 0 .
064 %
Example 2 Output: 100 V, 1 A, 400 Hz, Power Factor = 0.5 (=60)
Voltage Uncertainty Uncertainty for 100 V at 400 Hz is, 190 ppm + 2 mV, totaling:
100 V x 190 x 10
-6
= 19 mV added to 2 mV = 21 mV. Expressed in percent:
21 mV/100 V x 100 = 0.021 % (see “AC Voltage (Sine Wave) Specifications”).
Current Uncertainty Uncertainty for 1 A is 0.05 % � 100 A, totaling:
1 A x 0.0005 = 500 A added to 100 A = 0.6 mA. Expressed in percent:
0.6 mA/1 A x 100 = 0.06 % (see “AC Current (Sine Waves) Specifications”).
PF Adder Watts Adder for PF = 0.5 (=60) at 400 Hz is 0.76 % (see “Phase Specifications”).
15 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
Total Watts Output Uncertainty =
U power
0 .
021
2
0 .
06 2 0 .
76 2 0 .
76 %
VARs When the Power Factor approaches 0.0, the Watts output uncertainty becomes unrealistic because the dominant characteristic is the VARs (volts-amps-reactive) output. In these cases, calculate the Total VARs Output Uncertainty, as shown in example 3:
Example 3 Output: 100 V, 1 A, 60 Hz, Power Factor = 0.174 (=80)
Voltage Uncertainty Uncertainty for 100 V at 60 Hz is, 190 ppm + 2 mV, totaling:
100 V x 190 x 10
-6
= 19 mV added to 2 mV = 21 mV. Expressed in percent:
21 mV/100 V x 100 = 0.021 % (see “AC Voltage (Sine Wave) Specifications”).
Current Uncertainty Uncertainty for 1 A is 0.05 % � 100 A, totaling:
1 A x 0.0005 = 500 A added to 100 A = 0.6 mA. Expressed in percent:
0.6 mA/1 A x 100 = 0.06 % (see “AC Current (Sine Waves) Specifications”).
VARs Adder VARs Adder for =80 at 60 Hz is 0.03 % (see “Phase Specifications”).
Total VARS Output Uncertainty =
U
VARs
= 0 .
021
2
0 .
06 2 0 .
03 2 0 .
070 %
Additional Specifications
The following paragraphs provide additional specifications for the 5522A Calibrator ac voltage and ac current functions.
These specifications are valid after allowing a warm-up period of 30 minutes, or twice the time the 5522A has been turned off. All extended range specifications are based on performing the internal zero-cal function at weekly intervals, or when the ambient temperature changes by more than 5 C.
Frequency
Frequency Range Resolution
1-Year Absolute Uncertainty, tcal 5 C
Jitter
0.01 to 119.99 Hz
120.0 to 1199.9 Hz
1.200 to 11.999 kHz
12.00 to 119.99 kHz
120.0 to 1199.9 kHz
1.200 to 2.000 MHz
0.01 Hz
0.1 Hz
1.0 Hz
10 Hz
100 Hz
1 kHz
2.5 ppm +5 Hz
[1]
100 ns
[1] With REF CLK set to ext, the frequency uncertainty of the 5522A is the uncertainty of the external 10 MHz clock 5 Hz.
The amplitude of the 10 MHz external reference clock signal should be between 1 V and 5 V p-p.
Harmonics (2
nd
to 50
th
)
Fundamental
Frequency
[1]
10 to 45 Hz
45 to 65 Hz
65 to 500 Hz
500 Hz to 5 kHz
5 to 10 kHz
Voltages
NORMAL Terminals
33 mV to 32.9999 V
33 mV to 1020 V
33 mV to 1020 V
330 mV to 1020 V
3.3 to 1020 V
Currents
3.3 mA to 2.99999 A
3.3 mA to 20.5 A
33 mA to 20.5 A
33 mA to 20.5 A
33 to
329.9999 mA
33 to
329.9999 mA
Voltages
AUX Terminals
10 mV to 5 V
10 mV to 5 V
100 mV to 5 V
100 mV to 5 V
100 mV to 5 V
Amplitude
Uncertainty
Same % of output as the equivalent single output, but twice the floor adder.
10 to 30 kHz 3.3 to 1020 V
100 mV to
3.29999 V
[1] The maximum frequency of the harmonic output is 30 kHz (10 kHz for 3.3 to 5 V on the Aux terminals). For example, if the fundamental output is 5 kHz, the maximum selection is the 6th harmonic (30 kHz). All harmonic frequencies (2nd to
50th) are available for fundamental outputs between 10 Hz and 600 Hz (200 Hz for 3.3 to 5 V on the Aux terminals).
16 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
Phase Uncertainty .............................................. Phase uncertainty for harmonic outputs is 1 degree or the phase uncertainty shown in “Phase Specifications” for the particular output, whichever is greater. For example, the phase uncertainty of a 400 Hz fundamental output and 10 kHz harmonic output is 5
(from “Phase Specifications”). Another example, the phase uncertainty of a 50 Hz fundamental output and a 400 Hz harmonic output is 1 degree.
Example of determining Amplitude Uncertainty in a Dual Output Harmonic Mode
What are the amplitude uncertainties for the following dual outputs?
NORMAL (Fundamental) Output:
100 V, 100 Hz ............................................. From “AC Voltage (Sine Wave) 90 Day Specifications” the single output specification for 100 V, 100 Hz, is 0.015 % + 2 mV. For the dual output in this example, the specification is 0.015 % +4 mV as the 0.015 % is the same, and the floor is twice the value (2 x
2 mV).
AUX (50th Harmonic) Output:
100 mV, 5 kHz ............................................ From “AC Voltage (Sine Wave) 90 Day Specifications” the auxiliary output specification for 100 mV, 5 kHz, is 0.15 % + 450 mV. For the dual output in this example, the specification is 0.15 %
900 mV as the 0.15 % is the same, and the floor is twice the value
(2 x 450 mV).
AC Voltage (Sine Wave) Extended Bandwidth
Range Frequency
1-Year Absolute
Uncertainty tcal 5 C
Normal Channel (Single Output Mode)
1.0 to 33 mV
34 to 330 mV
0.4 to 33 V
0.01 to 9.99 Hz
(5.0 % of output
+0.5 % of range)
0.3 to 3.3 V
10 to 330 mV
0.4 to 5 V
500.1 kHz to 1 MHz
1.001 to 2 MHz
-10 dB at 1 MHz, typical
-31 dB at 2 MHz, typical
Auxiliary Output (Dual Output Mode)
0.01 to 9.99 Hz
(5.0 % of output
+0.5 % of range)
Max Voltage
Resolution
Two digits, e.g., 25 mV
Three digits
Two digits
Two digits
Three digits
Two digits
17 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
AC Voltage (Non-Sine Wave)
Triangle Wave &
Truncated Sine
Range, p-p
[1]
2.9 to 92.999 mV
93 to 929.999 mV
0.93 to 9.29999 V
9.3 to 93 V
29 to 929.999 mV
0.93 to 9.29999 V
9.3 to 14.0000 V
Frequency
Normal Channel (Single Output Mode)
0.01 to 10 Hz
10 to 45 Hz
45 Hz to 1 kHz
1 to 20 kHz
1-Year Absolute Uncertainty, tcal 5 C,
(% of output + % of range)
[2]
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
0.5 + 0.25
5.0 + 0.5
20 to 100 kHz
0.01 to 10 Hz
[3]
10 to 45 Hz
45 Hz to 1 kHz
1 to 20 kHz
20 to 100 kHz
0.01 to 10 Hz
[3]
10 to 45 Hz
45 Hz to 1 kHz
1 to 20 kHz
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
0.5 + 0.25
5.0 + 0.5
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
0.5 + 0.25
5.0 + 0.5
20 to 100 kHz
0.01 to 10 Hz
10 to 45 Hz
45 Hz to 1 kHz
1 to 20 kHz
[3]
20 to 100 kHz
[3]
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
0.5 + 0.25
5.0 + 0.5
Auxiliary Output (Dual Output Mode)
0.01 to 10 Hz
10 to 45 Hz
45 Hz to 1 kHz
1 to 10 kHz
0.01 to 10 Hz
10 to 45 Hz
45 Hz to 1 kHz
1 to 10 kHz
0.01 to 10 Hz
10 to 45 Hz
45 Hz to 1 kHz
1 to 10 kHz
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
5.0 + 0.5
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
5.0 + 0.5
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
5.0 + 0.5
Max Voltage
Resolution
Two digits on each range
Six digits on each range
Two digits on each range
Six digits on each range
Two digits on each range
Six digits on each range
Two digits on each range
Six digits on each range
Two digits on each range
Six digits on each range
Two digits on each range
Six digits on each range
Two digits on each range
Six digits on each range
[1] To convert p-p to rms for triangle wave, multiply the p-p value by 0.2886751. To convert p-p to rms for truncated sine wave, multiply the p-p value by 0.2165063.
[2] Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.
[3] Uncertainty for Truncated Sine outputs is typical over this frequency band.
18 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
AC Voltage (Non-Sine Wave) (cont.)
Square Wave
Range
(p-p)
2.9 to 65.999 mV
66 to 659.999 mV
0.66 to 6.59999 V
6.6 to 66.0000 V
29 to 659.999 mV
0.66 to 6.59999 V
6.6 to 14.0000 V
Frequency
1-Year Absolute Uncertainty, tcal 5 C,
(% of output + % of range)
[2]
0.01 to 10 Hz
Normal Channel (Single Output Mode)
5.0 + 0.5
0.25 + 0.5
10 to 45 Hz
45 Hz to 1 kHz
1 to 20 kHz
20 to 100 kHz
0.01 to 10 Hz
10 to 45 Hz
45 Hz to 1 kHz
1 to 20 kHz
20 to 100 kHz
0.01 to 10 Hz
10 to 45 Hz
45 Hz to 1 kHz
1 to 20 kHz
20 to 100 kHz
0.01 to 10 Hz
10 to 45 Hz
45 Hz to 1 kHz
1 to 20 kHz
20 to 100 kHz
0.01 to 10 Hz
10 to 45 Hz
45 Hz to 1 kHz
1 to 10 kHz
[3]
0.01 to 10 Hz
10 to 45 Hz
45 Hz to 1 kHz
1 to 10 kHz
[3]
0.01 to 10 Hz
10 to 45 Hz
45 Hz to 1 kHz
1 to 10 kHz
[3]
0.25 + 0.25
0.5 + 0.25
5.0 + 0.5
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
0.5 + 0.25
5.0 + 0.5
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
0.5 + 0.25
5.0 + 0.5
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
0.5 + 0.25
5.0 + 0.5
Auxiliary Output (Dual Output Mode)
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
5.0 + 0.5
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
5.0 + 0.5
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
5.0 + 0.5
[1] To convert p-p to rms for square wave, multiply the p-p value by 0.5.
[2] Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.
[3] Limited to 1 kHz for Auxiliary outputs 6.6 V p-p.
Max Voltage
Resolution
Two digits on each range
Six digits on each range
Two digits on each range
Six digits on each range
Two digits on each range
Six digits on each range
Two digits on each range
Six digits on each range
Two digits on each range
Six digits on each range
Two digits on each range
Six digits on each range
Two digits on each range
Six digits on each range
19 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
AC Voltage, DC Offset
Range
[1]
(Normal Channel)
3.3 to 32.999 mV
33 to 329.999 mV
0.33 to 3.29999 V
3.3 to 32.9999 V
Offset Range
[2]
Max Peak
Signal
1-Year Absolute Uncertainty, tcal 5 C
[3]
(% of dc output + floor)
Sine Waves (rms)
0 to 50 mV
0 to 500 mV
0 to 5 V
0 to 50 V
80 mV
800 mV
8 V
55 V
Triangle Waves and Truncated Sine Waves (p-p)
0.1 + 33 V
0.1 + 330 V
0.1 + 3300 V
0.1 + 33 mV
9.3 to 92.999 mV
93 to 929.999 mV
0.93 to 9.29999 V
9.3 to 93.0000 V
6.6 to 65.999 mV
66 to 659.999 mV
0.66 to 6.59999 V
6.6 to 66.0000 V
0 to 50 mV
0 to 500 mV
0 to 5 V
0 to 50 V
0 to 500 mV
0 to 5 V
0 to 50 V
80 mV
800 mV
8 V
55 V
Square Waves (p-p)
0 to 50 mV 80 mV
800 mV
8 V
55 V
0.1 + 93 V
0.1 + 930 V
0.1 + 9300 V
0.1 + 93 mV
0.1 + 66 V
0.1 + 660 V
0.1 + 6600 V
0.1 + 66 mV
[1] Offsets are not allowed on ranges above the highest range shown above.
[2] The maximum offset value is determined by the difference between the peak value of the selected voltage output and the allowable maximum peak signal. For example, a 10 V p-p square wave output has a peak value of 5 V, allowing a maximum offset up to 50 V to not exceed the 55 V maximum peak signal. The maximum offset values shown above are for the minimum outputs in each range.
[3] For frequencies 0.01 to 10 Hz, and 500 kHz to 2 MHz, the offset uncertainty is 5 % of output, 1 % of the offset range.
AC Voltage, Square Wave Characteristics
1 kHz
Typical
Settling Time @
1 kHz Typical
<10 s to 1 % of final value
@ 1 kHz
Typical
Duty Cycle Range Duty Cycle Uncertainty
<1 s <2 %
1 % to 99 % <3.3 V p-p.
0,01 Hz to 100 kHz
(0.02 % of period + 100 ns), 50 % duty cycle
(0.05 % of period + 100 ns), other duty cycles from 10 % to 90 %
AC Voltage, Triangle Wave Characteristics (typical)
Linearity to 1 kHz
0.3 % of p-p value, from 10 % to 90 % point
Aberrations
<1 % of p-p value, with amplitude >50 % of range
20 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
AC Current (Non-Sine Wave)
Triangle Wave &
Truncated Sine Wave
Range p-p
Frequency
10 to 45 Hz
0.047 to
0.92999 mA
[1]
45 Hz to 1 kHz
1 to 10 kHz
0.93 to
9.29999 mA
9.3 to
92.9999 mA
[1]
[1]
10 to 45 Hz
45 Hz to 1 kHz
1 to 10 kHz
10 to 45 Hz
45 Hz to 1 kHz
1 to 10 kHz
10 to 45 Hz
45 Hz to 1 kHz
93 to
929.999 mA
[1]
0.93 to
8.49999 A
[2]
1 to 10 kHz
10 to 45 Hz
45 Hz to 1 kHz
1 to 10 kHz
8.5 to 57 A
[2]
45 to 500 Hz
500 Hz to 1 kHz
[1] Frequency limited to 1 kHz with LCOMP on.
[2] Frequency limited to 440 Hz with LCOMP on.
1-Year Absolute Uncertainty tcal 5 C
(% of output + % of range)
0.25 + 0.5
0.25 + 0.25
10 + 2
0.25 + 0.5
0.25 + 0.25
10 + 2
0.25 + 0.5
0.25 + 0.25
10 + 2
0.25 + 0.5
0.25 + 0.5
10 + 2
0.5 + 1.0
0.5 + 0.5
10 + 2
0.5 + 0.5
1.0 + 1.0
Max Current
Resolution
Six digits
Six digits
Six digits
Six digits
Six digits
21 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
AC Current (Non-Sine Wave) (cont.)
Square Wave
Range p-p
Frequency
1-Year Absolute Uncertainty tcal 5 C
(% of output + % of range)
10 to 45 Hz 0.25 + 0.5
0.047 to
0.65999 mA
[1]
45 Hz to 1 kHz 0.25 + 0.25
1 to 10 kHz 10 + 2
0.66 to
6.59999 mA
6.6 to
65.9999 mA
[1]
[1]
10 to 45 Hz
45 Hz to 1 kHz
1 to 10 kHz
10 to 45 Hz
45 Hz to 1 kHz
1 to 10 kHz
66 to
659.999 mA
[1]
10 to 45 Hz
45 Hz to 1 kHz
1 to 10 kHz
10 to 45 Hz
0.66 to
5.99999 A
[2]
45 Hz to 1 kHz
1 to 10 kHz
45 to 500 Hz
6 to 41 A
[2]
500 Hz to 1 kHz
[1] Frequency limited to 1 kHz with LCOMP on.
[2] Frequency limited to 440 Hz with LCOMP on.
0.25 + 0.5
0.25 + 0.25
10 + 2
0.25 + 0.5
0.25 + 0.25
10 + 2
0.25 + 0.5
0.25 + 0.5
10 + 2
0.5 + 1.0
0.5 + 0.5
10 + 2
0.5 + 0.5
1.0 + 1.0
Max Current
Resolution
Six digits
Six digits
Six digits
Six digits
AC Current, Square Wave Characteristics (typical)
Range
I <6 A @ 400 Hz
LCOMP off
3 A & 20 A Ranges on
Risetime Settling Time
25 s 40 s to 1 % of final value
Overshoot
<10 % for <1 V Compliance
<10 % for <1 V Compliance
AC Current, Triangle Wave Characteristics (typical)
Linearity to 400 Hz
0.3 % of p-p value, from 10 % to 90 % point
Aberrations
<1 % of p-p value, with amplitude >50 % of range
22 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
Ordering Information
Model
5522A
5522A/6
Multi-Product Calibrator
Multi-Product Calibrator with 600
MHz Oscilloscope Calibration Option
5522A/1GHZ
5522A-PQ
Multi-Product Calibrator with 1100
MHz Oscilloscope Calibration Option
Multi-Product Calibrator with Power
Quality Option
5522A-PQ/6 Multi-Product Calibrator with Power
Quality and 600 MHz Oscilloscope
Calibration Options
5522A-PQ/1GHZ Multi-Product Calibrator with Power
Quality and 1100 MHz Oscilloscope
Calibration Options
Accessories
5522A/
CARRYCASE
55XX/CASE
Rugged Carrying Case with removable front/back panels
Transit Case with Wheels
5520A-525A/ Thermocouple and Test
LEADS Lead Set
5500A/COIL
5500A/HNDL
50-Turn Coil
Side Handle
Y5537 Rack Mount Kit
Fluke-700Pxx Pressure Module Series
Software
MET/CAL MET/CAL Plus Calibration
Management Software
Other calibration solutions
Fluke Calibration provides the broadest range of calibrators and standards, software, service, support and training in electrical, temperature, pressure, RF and flow calibration.
Visit www.Fluke.com/FlukeCal for more information about
Fluke Calibration products and services.
Fluke Calibration.
Precision, performance, confidence.
™
Fluke Calibration
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (877) 355-3225 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.flukecal.com
©2011-2013 Fluke Calibration.
Specifications subject to change without notice.
Printed in U.S.A. 11/2013 3931295D D-EN-N
Pub-ID 11721-eng Rev. 01
Modification of this document is not permitted without written permission from Fluke Corporation.
23 Fluke Calibration 5522A Multi-Product Calibrator Extended Specifications
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project