SITRANS T - Temperature Transmitters Brochure


Add to my manuals
12 Pages

advertisement

SITRANS T - Temperature Transmitters Brochure | Manualzz

siemens.com/sitranst

Combining perfect senses and intelligence for every degree and condition

SITRANS T: highly accurate temperature sensors and transmitters with unique communication capabilities

Answers for industry.

Meeting all process industry‘s demands:

The SITRANS T family for temperature measurement

Whether it is about faster processes, higher product quality or quality control, optimized plant availability and safety, more efficient workflow or lower costs – process industries need to master a variety of demands for competitive advantage. Devices for measuring temperature are one but essential part to achieve these goals. With the product family SITRANS T

Siemens offers you a complete spectrum of temperature sensors and transmitters including comprehensive services – from a single source.

Your application requires individual components or complete solutions? No problem with SITRANS T. The product family delivers highly accurate temperature sensors and transmitters for every application in the process industry – even under extreme conditions: general purpose, intrinsically safe or explosion-proof and with globally recognized certificates. They are as easily integrated in different control systems as configurated. They prove highest flexibility as they are available as a complete measuring point with a widespread portfolio or as transmitter for head-, rail- or field-mount installation. Furthermore, they convince with communications capability and offer unique features as diagnostics displays. And to make the portfolio perfect,

Siemens supports you with application and process knowhow from worldwide presales and technical support to on-site servicing.

2

It’s all about communication – easy integration and configuration of SITRANS T

With SITRANS T you benefit from continuous ease of use and highest transparency – thanks to powerful software and a whole range of communications capabilities that allow the integration of field devices into the overall plant.

Most of our temperature transmitters use a communication based on industrial standards like HART, PROFIBUS PA or FOUNDATION Fieldbus. The EDD (Enhanced Device

Description) technology is used to integrate these devices in different process control systems like SIMATIC PCS 7.

SIMATIC PDM (Process Device Manager) is a consistent, vendor-independent tool for operation, setup, maintenance and diagnostics of intelligent field devices. It uses exactly these EDDs to operate devices with HART and

PROFIBUS PA communication. HART means here: wired

HART and WirelessHART which is part of HART V7.

So the new WirelessHART temperature transmitter can also be operated with SIMATIC PDM, based on a modern

EDD. Most devices are also integrated in third-party tools like HART Communicator or AMS.

For FOUNDATION Fieldbus, AMS or the F Communicator offer easy integration. For 4 - 20 mA devices without

HART a special tool SIPROM T is available for point-topoint communications. This tool reflects the advanced operating philosophy and look and feel of SIMATIC PDM, so that it is also familiar to the PDM user.

3

Precision where the temperature takes place:

SITRANS TS temperature sensors

SITRANS TS temperature sensors are the perfect base for accurate and secure temperature measurement. They support a very broad scope of different applications in the process industries. Equipped with our head-mount transmitters they provide easy system integration. With the optional display a comfortable read out of the process variable supports local operation and monitoring.

Benefits at a glance

• Broad portfolio support of all main applications

• Secure use in intrinsic safe and Ex d instrumented plants

• Modular system for easy configuration

• Support with 4 - 20 mA, HART, PROFIBUS PA and

FOUNDATION Fieldbus

4

Intelligence on-site:

SITRANS T temperature transmitters

SITRANS T temperature transmitters are the first choice to bring intelligence in your temperature sensor. The status of your measurement is easy to monitor remotely or locally. Unique support of maintenance such as two-color diagnostics LED and test pins let you detect with one glance suspect measurements and with one touch you measure the current loop without interruption of the measuring.

Benefits at a glance

• Support of various designs: head-, rail-, field-mount

• Easy integration in your system with 4 - 20 mA, HART,

PROFIBUS PA, FOUNDATION Fieldbus as well as

WirelessHART

• Extended diagnostics

• Maintenance benefits

5

6

The complete range

SITRANS TS temperature sensors

SITRANS TS100

This cable temperature sensor product series comes with a direct mounted cable. As a basic or mineral-insulated version a wide field of application is supported. The installation is easy and flexible by using compression or soldering fittings. With the optional adapter surface measurement is simple to apply. The intrinsic safe version has the approval for operating even in zone 0 without an additional protection tube. In such application the excellent response time of the sensor will be an outstanding benefit.

SITRANS TS500

The industry temperature sensor series supports a wide field of measurements, from simple applications up to solutions for harsh environments. Designed as a modular system of tubular or barstock thermowell, extension, connection head and optional transmitter and display, the customers profit from the use of standard components for individual applications. Intrinsic safe versions are available as well as Ex d.

SITRANS TS200

The compact temperature sensor series adds to the excellent benefits of our SITRANS TS100. Instead of the flexible cable, it comes with a fixed connection M12, Lemo etc.

SITRANS T temperature sensors – special for food and pharma

Our food and pharma temperature sensor product series is featured with a wide range of appropriate process connections – the classical method. With the clamp-on temperature sensor Siemens strikes a new path. Comparable with built-in measurement regarding response time and accuracy the advantages especially at small pipe diameters are obviously. No welding and welding validation, no process disturbance, easy pigging, easy dismantling for recalibration.

SITRANS T temperature transmitters

SITRANS TH

Weak sensor signal, expensive compensation cables, various I/O for each sensor type? The use of our SITRANS TH for mounting in the connection head of the temperature sensor makes you forget that troubles. The sensor signal is amplified and transmitted EMI-resistant to the control system. With the full range of outputs 4 - 20 mA, HART,

PROFIBUS PA, FOUNDATION Fieldbus you will profit from easy integration in your process control system. The galvanic insulation and the fault detection provide a proper operation. Further benefits are the outstanding maintenance options: extended diagnostics and unique diag- nostics displays that show the device status at a glance.

The technician simply connects an ammeter via the test sockets and can then read the output current without opening the measuring circuit. International engineering is no problem with the broad range of approvals e.g. for

Europe, USA, Canada and Brazil. Functional safety is certified for SIL 2/3 applications.

SITRANS TW

Looking for a Swiss army knife in temperature measurement, this real universal device is your choice. Multi- voltages power supply, 4 - 20 mA, 2 - 10 V, HART, relay output, individual curves are only an extract of the features of that device.

SITRANS TF

You have extreme environmental conditions? Temperature sensor mounted in an inconvenient location? With the rug- ged-built field device SITRANS TF, available in die-cast alu- minum or durable stainless steel housing, also harsh con- ditions can be coped. Mount your transmitter near to the sensor, but out of a hot zone of heat or vibration. Direct on the plant floor easy read out of the metering is your benefit.

SITRANS TR

You want to use the excellent features of the SITRANS TH family but prefer a centralized access to all connected measurement points in junction boxes or in the control room close to the process? SITRANS TR family, the twowire rail-mount transmitters, support this strategy in a perfect way.

SITRANS TF280

The SITRANS TF280 is a WirelessHART temperature transmitter that provides all measured process values as well as diagnostics information, parameters and functions via radio. The device is powered by an internal battery and designed for ultra-low power consumption. Its compact and rugged design makes it especially suitable for direct mounting on tanks and pipes in remote parts of plants, and on moving or rotating equipment for process monitoring or asset management applications.

7

SITRANS TS at a glance

8

Type SITRANS TS Insert SITRANS TS100 SITRANS TS200 SITRANS TS300 SITRANS TS500 SITRANS TS500 SITRANS TS500

Measuring inser spares

Mineral-insulated execution (MIC)

Temperature Sensors in cable version

Mineralinsulated version

Temperature Sensors in compact version

Mineral-insulated version

Temperature Sensors for

Food&Beverage/Pharma

In-pipe or clamp-on

Temperature Sensors for installation in existing thermowells

Suitable for thermowells according to DIN 43772 as well as

ASME B40.9-2001

Temperature Sensors with tubular thermowell for low to middle process load

Thermowell Form 2 or Form 3

(tapered) according to DIN 43772 and Form 2N; with thread, flange or without process connection

Temperature Sensors with barstock thermowell for high process load

Thermowell according to

DIN 43772 Form 4 for weld-in or Form 4F with flange

Application

Process connection

Certificates

Output

Wetted material

Spares Plant and machinery construction, bearing temperature, surface measurement

Compression or soldering fittings:

G [1/4, 1/2]“; 1/2“NPT; M8x1;

M18x1.5

Surface mounting adapter for installation on flat surfaces pipes

Plant and machinery construction, bearing temperature, surface measurement

Compression or soldering fittings:

G [1/4, 1/2]“; 1/2“NPT; M8x1;

M18x1.5

Surface mounting adapter for installation on flat surfaces pipes

For advanced hygienic requirements

In-pipe:

Clamp-flange; DIN 11851;

Varivent; BioControl; Neumo;

Ingold; spherical-welding-sleeve

Clamp-on: collar 4…57 mm strap up to 200 mm

Europe+IEC EX:

• intrinsic safety “ia”, “ic”

• for flameproof enclosure “d”; for dust protection by enclosure “t”

Europe+IEC EX:

• intrinsic safety “ia”, “ic”

Europe+IEC EX:

• intrinsic safety “ia”, “ic”

Direct sensor signal

4...20 mA (TH100/TH200)

HART (TH300)

PA (TH400)

FF (TH400)

SS similar 1.4404 (RTD),

2.4816 (Thermocouple)

(SS sim. 316L, Alloy 600)

Direct sensor signal

SS similar 1.4404 (RTD),

2.4816 (Thermocouple)

(SS sim. 316L, Alloy 600)

Direct sensor signal

SS similar 1.4404 (RTD),

2.4816 (Thermocouple)

(SS sim. 316L, Alloy 600)

Direct sensor signal

4 - 20 mA (TH100/TH200)

HART (TH300)

PA (TH400)

FF (TH400)

In-pipe:

1.4404/316L

Clamp-on: no wetted parts

Vessel and pipes

Connection to thermowell:

M18x1.5; G1/2“, 1/2“NPT

Europe+IEC EX:

• intrinsic safety “ia”, “ic”

• flameproof enclosure “d”;

dust protection by enclosure “t”

• non-sparking “n”

Direct sensor signal

4 - 20 mA (TH100/TH200)

HART (TH300)

PA (TH400)

FF (TH400) no wetted parts

Technical data

Temperature limits*

Minimum response time t

0,5

Ingress Protection

Pt100 Basic: –30...+400 °C

Pt100 Extend: –196...+600 °C

Thermocouple: –196...+1100 °C

(depends on type)

2...6s

IP54

Pt100 Basic: –30...+400 °C

Pt100 Extend: –196...+600 °C

Thermocouple: –196...+1100 °C

(depends on type)

2...6s

IP54

* Combination of loads (temperature, flow, vibration, pressure) lower this values partly extensive.

Further temperature limits are a result of the used thermowell materials

(e.g. 1.4571 / 316Ti resistant to compression stress until 450...550 °C, material limit 800 °C.)

Pt100 Basic: –30...+400 °C

Pt100 Extend: –196...+600 °C

Thermocouple: –196...+1100 °C

(depends on type)

2...6s

IP54

(some connetors lower)

In-pipe:

–30…+300 °C

Clamp-on:

–20…+160 °C

5 s

IP65 (IP54 at some head types)

Pt100 Basic: –30...+400 °C

Pt100 Extend: –196...+600 °C

Thermocouple: –196...+1100 °C

(depends on type)

Depends on type of thermowell

IP65 (IP54 at some head types)

Vessel and pipes Vessel and pipes

Compression fitting G1/2“; 1/2“NPT; welded thread

G1/2“; G1“; 1/2“NPT; welded flange

DN25PN40; 1RF150; 1.5RF150;

1.5RF300

Form 4 for weld-in

Form 4F with flange:

DN25PN40; 1RF150;

1RF300; 1.5RF150; 1.5RF300

Europe+IEC EX:

• intrinsic safety “ia”, “ic”

• flameproof enclosure “d”;

dust protection by enclosure “t”

• non-sparking “n”

Direct sensor signal

4 - 20 mA (TH100/TH200)

HART (TH300)

PA (TH400)

FF (TH400)

1.4404; 1.4571

(316L; 316TI)

Europe+IEC EX:

• intrinsic safety “ia”, “ic”

• flameproof enclosure “d”;

dust protection by enclosure “t”

• non-sparking “n”

Direct sensor signal

4 - 20 mA (TH100/TH200)

HART (TH300)

PA (TH400)

FF (TH400)

Form 4F: 1.4404; 1.4571

(316L; 316TI)

Form 4 additionally 1.7335; 1.5415

(A 182 F11; A 204 Gr.A)

Pt100 Basic: –30...+400 °C

Pt100 Extend: –196...+600 °C

Thermocouple: –196...+1100 °C

(depends on type)

7...45s

IP65 (IP54 at some head types)

Pt100 Basic: –30...+400 °C

Pt100 Extend: –196...+600 °C

Thermocouple: –196...+1100 °C

(depends on type)

20...45s

IP65 (IP54 at some head types)

SITRANS TS at a glance

Type SITRANS TS Insert SITRANS TS100 SITRANS TS200 SITRANS TS300 SITRANS TS500 SITRANS TS500 SITRANS TS500

Measuring inser spares

Mineral-insulated execution (MIC)

Temperature Sensors in cable version

Mineralinsulated version

Temperature Sensors in compact version

Mineral-insulated version

Temperature Sensors for

Food&Beverage/Pharma

In-pipe or clamp-on

Temperature Sensors for installation in existing thermowells

Suitable for thermowells according to DIN 43772 as well as

ASME B40.9-2001

Temperature Sensors with tubular thermowell for low to middle process load

Thermowell Form 2 or Form 3

(tapered) according to DIN 43772 and Form 2N; with thread, flange or without process connection

Temperature Sensors with barstock thermowell for high process load

Thermowell according to

DIN 43772 Form 4 for weld-in or Form 4F with flange

Application

Process connection

Certificates

Output

Wetted material

Spares Plant and machinery construction, bearing temperature, surface measurement

Compression or soldering fittings:

G [1/4, 1/2]“; 1/2“NPT; M8x1;

M18x1.5

Surface mounting adapter for installation on flat surfaces pipes

Plant and machinery construction, bearing temperature, surface measurement

Compression or soldering fittings:

G [1/4, 1/2]“; 1/2“NPT; M8x1;

M18x1.5

Surface mounting adapter for installation on flat surfaces pipes

For advanced hygienic requirements

In-pipe:

Clamp-flange; DIN 11851;

Varivent; BioControl; Neumo;

Ingold; spherical-welding-sleeve

Clamp-on: collar 4…57 mm strap up to 200 mm

Europe+IEC EX:

• intrinsic safety “ia”, “ic”

• for flameproof enclosure “d”; for dust protection by enclosure “t”

Europe+IEC EX:

• intrinsic safety “ia”, “ic”

Europe+IEC EX:

• intrinsic safety “ia”, “ic”

Direct sensor signal

4...20 mA (TH100/TH200)

HART (TH300)

PA (TH400)

FF (TH400)

SS similar 1.4404 (RTD),

2.4816 (Thermocouple)

(SS sim. 316L, Alloy 600)

Direct sensor signal

SS similar 1.4404 (RTD),

2.4816 (Thermocouple)

(SS sim. 316L, Alloy 600)

Direct sensor signal

SS similar 1.4404 (RTD),

2.4816 (Thermocouple)

(SS sim. 316L, Alloy 600)

Direct sensor signal

4 - 20 mA (TH100/TH200)

HART (TH300)

PA (TH400)

FF (TH400)

In-pipe:

1.4404/316L

Clamp-on: no wetted parts

Vessel and pipes

Connection to thermowell:

M18x1.5; G1/2“, 1/2“NPT

Europe+IEC EX:

• intrinsic safety “ia”, “ic”

• flameproof enclosure “d”;

dust protection by enclosure “t”

• non-sparking “n”

Direct sensor signal

4 - 20 mA (TH100/TH200)

HART (TH300)

PA (TH400)

FF (TH400) no wetted parts

Technical data

Temperature limits*

Minimum response time t

0,5

Ingress Protection

Pt100 Basic: –30...+400 °C

Pt100 Extend: –196...+600 °C

Thermocouple: –196...+1100 °C

(depends on type)

2...6s

IP54

Pt100 Basic: –30...+400 °C

Pt100 Extend: –196...+600 °C

Thermocouple: –196...+1100 °C

(depends on type)

2...6s

IP54

* Combination of loads (temperature, flow, vibration, pressure) lower this values partly extensive.

Further temperature limits are a result of the used thermowell materials

(e.g. 1.4571 / 316Ti resistant to compression stress until 450...550 °C, material limit 800 °C.)

Pt100 Basic: –30...+400 °C

Pt100 Extend: –196...+600 °C

Thermocouple: –196...+1100 °C

(depends on type)

2...6s

IP54

(some connetors lower)

In-pipe:

–30…+300 °C

Clamp-on:

–20…+160 °C

5 s

IP65 (IP54 at some head types)

9

Pt100 Basic: –30...+400 °C

Pt100 Extend: –196...+600 °C

Thermocouple: –196...+1100 °C

(depends on type)

Depends on type of thermowell

IP65 (IP54 at some head types)

Vessel and pipes Vessel and pipes

Compression fitting G1/2“; 1/2“NPT; welded thread

G1/2“; G1“; 1/2“NPT; welded flange

DN25PN40; 1RF150; 1.5RF150;

1.5RF300

Form 4 for weld-in

Form 4F with flange:

DN25PN40; 1RF150;

1RF300; 1.5RF150; 1.5RF300

Europe+IEC EX:

• intrinsic safety “ia”, “ic”

• flameproof enclosure “d”;

dust protection by enclosure “t”

• non-sparking “n”

Direct sensor signal

4 - 20 mA (TH100/TH200)

HART (TH300)

PA (TH400)

FF (TH400)

1.4404; 1.4571

(316L; 316TI)

Europe+IEC EX:

• intrinsic safety “ia”, “ic”

• flameproof enclosure “d”;

dust protection by enclosure “t”

• non-sparking “n”

Direct sensor signal

4 - 20 mA (TH100/TH200)

HART (TH300)

PA (TH400)

FF (TH400)

Form 4F: 1.4404; 1.4571

(316L; 316TI)

Form 4 additionally 1.7335; 1.5415

(A 182 F11; A 204 Gr.A)

Pt100 Basic: –30...+400 °C

Pt100 Extend: –196...+600 °C

Thermocouple: –196...+1100 °C

(depends on type)

7...45s

IP65 (IP54 at some head types)

Pt100 Basic: –30...+400 °C

Pt100 Extend: –196...+600 °C

Thermocouple: –196...+1100 °C

(depends on type)

20...45s

IP65 (IP54 at some head types)

SITRANS TH, TR, TW and TF at a glance

Installation In the connection head

Two-wire

On DIN rail

Two-wire Four-wire

Field device

Two-wire Wireless

Type

Input

(connectable sensors)

Output

Local display

SITRANS TH100

Pt100 resistance thermometers

4 - 20 mA

N/A

SITRANS TH200

Resistance thermometers

Thermocouples

Resistance-type sensors

DC sources

4 - 20 mA

N/A

SITRANS TH300

Resistance thermometers

Thermocouples

Resistance-type sensors

DC sources

4 - 20 mA, HART

N/A

SITRANS TH400

Resistance thermometers

Thermocouples

Resistance-type sensors

DC sources

PROFIBUS PA version

FOUNDATION Fieldbus version

N/A

Power supply

Housing material

8.5 - 36 V DC

(30 V for Ex)

Molded plastic, embedded electronics

–40...+85 °C

11 - 35 V DC

(30 V for Ex)

Molded plastic, embedded electronics

–40...+85 °C

11 - 35 V DC

(30 V for Ex)

Molded plastic, embedded electronics

–40...+85 °C

9 - 32 V DC

(30 V for Ex and

17.5 V for FISCO)

Molded plastic,

embedded electronics

–40...+85 °C

Ambient temperature

Degree of protection

Certificates

IP40 IP40 IP40 IP40

Operator input

SIMATIC PDM

Handheld 375

AMS

SIPROM T and special modem

Local configuration using

3 push buttons

Europe (ATEX): Ex ia, ib, ic,

Ex n

USA (cFMus): IS, NI

Canada (cFMus): IS, NI other certificates:

GOST,NEPSI,PESO

Europe (ATEX): Ex ia, ib, ic,

Ex n

USA (cFMus): IS, NI

Canada (cFMus): IS, NI,

NiFW, DIP

SIL 2 und SIL2/3 other certificates:

GOST,NEPSI,PESO,EXPLABS

Europe (ATEX): Ex ia, ib, ic

Ex n

USA (cFMus): IS, NI

Canada (cFMus): IS, NI,

NiFW, DIP

SIL 2 und SIL2/3 other certificates:

GOST,NEPSI,PESO,EXPLABS

Europe (ATEX): Ex ia, Ex n

Ex ib, Ex n

USA (cFMus): IS, NI

Canada (cFMus): IS, NI other certificates:

GOST, PESO

PROFIBUS PA

FOUNDATION Fieldbus

FF version

FF version

10

SITRANS TR200

Resistance thermometers

Thermocouples

Resistance-type sensors

DC sources

4 - 20 mA

SITRANS TR300

Resistance thermometers

Thermocouples

Resistance-type sensors

DC sources

4 - 20 mA, HART

SITRANS TW

Resistance thermometers

Thermocouples

Resistance-type sensors

DC sources

0/4 - 20 mA, HART

0/2 - 10 V, HART

SITRANS TF

Resistance thermometers

Thermocouples

Resistance-type sensors

DC sources

Current sources

4 - 20 mA

4 - 20 mA, HART

PROFIBUS PA

FOUNDATION Fieldbus

LCD (with version

4 - 20 mA and 4 - 20 mA,

HART)

SITRANS TF280

Pt100 resistance thermometer

WirelessHART back-lighted LCD

11 - 35 V DC

(30 V for Ex ia)

Molded plastic

–40 - +85 °C

11 - 35 V DC

(30 V for Ex ia)

Molded plastic

–40 - +85 °C

115/230 V UC or 24 V UC

Molded plastic

–25 - +70 °C

DC 11/13,5 - 35V

(30 V for Ex ia and

17.5 V for FISCO)

Varnished die-cast aluminum or stainless steel

–40 - +85 °C

Lithium thionyl battery

D-cell 3.6 V

Varnished die-cast aluminum

–40 - +80 °C

IP20 IP20 IP20 IP67

Europe (ATEX): Ex ia or

Ex ib, Ex n

SIL 2 und SIL2/3 other certificates:

NEPSI

Europe (ATEX): Ex ia or

Ex ib, Ex n

SIL 2 und SIL2/3 other certificates:

NEPSI

Europe (ATEX): Ex (ia) or

Ex (ib)

SIL 1 other certificates:

GOST

Europe (ATEX): Ex ia,

Ex d, Ex n

USA: XP/DIP/NI/S

SIL 2 und SIL2/3 (4...20 mA/HART) other certificates:

GOST,INMETRO,

NEPSI,KOSHA

IP65

R&TTE, FCC

HART / PROFIBUS PA/

FOUNDATION Fieldbus

HART / FOUNDATION

Fieldbus

HART / FOUNDATION

Fieldbus

4...20 mA

SITRANS TH, TR, TW and TF at a glance

Installation In the connection head

Two-wire

On DIN rail

Two-wire Four-wire

Field device

Two-wire Wireless

Type

Input

(connectable sensors)

Output

Local display

SITRANS TH100

Pt100 resistance thermometers

4 - 20 mA

N/A

SITRANS TH200

Resistance thermometers

Thermocouples

Resistance-type sensors

DC sources

4 - 20 mA

N/A

SITRANS TH300

Resistance thermometers

Thermocouples

Resistance-type sensors

DC sources

4 - 20 mA, HART

N/A

SITRANS TH400

Resistance thermometers

Thermocouples

Resistance-type sensors

DC sources

PROFIBUS PA version

FOUNDATION Fieldbus version

N/A

Power supply

Housing material

8.5 - 36 V DC

(30 V for Ex)

Molded plastic, embedded electronics

–40...+85 °C

11 - 35 V DC

(30 V for Ex)

Molded plastic, embedded electronics

–40...+85 °C

11 - 35 V DC

(30 V for Ex)

Molded plastic, embedded electronics

–40...+85 °C

9 - 32 V DC

(30 V for Ex and

17.5 V for FISCO)

Molded plastic,

embedded electronics

–40...+85 °C

Ambient temperature

Degree of protection

Certificates

IP40 IP40 IP40 IP40

Operator input

SIMATIC PDM

Handheld 375

AMS

SIPROM T and special modem

Local configuration using

3 push buttons

Europe (ATEX): Ex ia, ib, ic,

Ex n

USA (cFMus): IS, NI

Canada (cFMus): IS, NI other certificates:

GOST,NEPSI,PESO

Europe (ATEX): Ex ia, ib, ic,

Ex n

USA (cFMus): IS, NI

Canada (cFMus): IS, NI,

NiFW, DIP

SIL 2 und SIL2/3 other certificates:

GOST,NEPSI,PESO,EXPLABS

Europe (ATEX): Ex ia, ib, ic

Ex n

USA (cFMus): IS, NI

Canada (cFMus): IS, NI,

NiFW, DIP

SIL 2 und SIL2/3 other certificates:

GOST,NEPSI,PESO,EXPLABS

Europe (ATEX): Ex ia, Ex n

Ex ib, Ex n

USA (cFMus): IS, NI

Canada (cFMus): IS, NI other certificates:

GOST, PESO

PROFIBUS PA

FOUNDATION Fieldbus

FF version

FF version

SITRANS TR200

Resistance thermometers

Thermocouples

Resistance-type sensors

DC sources

4 - 20 mA

SITRANS TR300

Resistance thermometers

Thermocouples

Resistance-type sensors

DC sources

4 - 20 mA, HART

SITRANS TW

Resistance thermometers

Thermocouples

Resistance-type sensors

DC sources

0/4 - 20 mA, HART

0/2 - 10 V, HART

SITRANS TF

Resistance thermometers

Thermocouples

Resistance-type sensors

DC sources

Current sources

4 - 20 mA

4 - 20 mA, HART

PROFIBUS PA

FOUNDATION Fieldbus

LCD (with version

4 - 20 mA and 4 - 20 mA,

HART)

SITRANS TF280

Pt100 resistance thermometer

WirelessHART back-lighted LCD

11 - 35 V DC

(30 V for Ex ia)

Molded plastic

–40 - +85 °C

11 - 35 V DC

(30 V for Ex ia)

Molded plastic

–40 - +85 °C

115/230 V UC or 24 V UC

Molded plastic

–25 - +70 °C

DC 11/13,5 - 35V

(30 V for Ex ia and

17.5 V for FISCO)

Varnished die-cast aluminum or stainless steel

–40 - +85 °C

Lithium thionyl battery

D-cell 3.6 V

Varnished die-cast aluminum

–40 - +80 °C

IP20 IP20 IP20 IP67

Europe (ATEX): Ex ia or

Ex ib, Ex n

SIL 2 und SIL2/3 other certificates:

NEPSI

Europe (ATEX): Ex ia or

Ex ib, Ex n

SIL 2 und SIL2/3 other certificates:

NEPSI

Europe (ATEX): Ex (ia) or

Ex (ib)

SIL 1 other certificates:

GOST

Europe (ATEX): Ex ia,

Ex d, Ex n

USA: XP/DIP/NI/S

SIL 2 und SIL2/3 (4...20 mA/HART) other certificates:

GOST,INMETRO,

NEPSI,KOSHA

IP65

R&TTE, FCC

HART / PROFIBUS PA/

FOUNDATION Fieldbus

HART / FOUNDATION

Fieldbus

HART / FOUNDATION

Fieldbus

4...20 mA

11

More information:

www.siemens.com/sitranst www.siemens.com/processinstrumentation

Siemens AG

Industry Sector

Sensors and Communication

76181 KARLSRUHE

GERMANY

Subject to change without prior notice 09/12

Order No.: E20001-A930-P710-V1-7600

DISPO 27900

21/36212 MI.SC.PS.XX01.52.2.06

WS 09125.0

Printed in Germany

© Siemens AG 2012

The information provided in this brochure contains merely general descriptions or characteristics of performance which in actual case of use do not always apply as described or which may change as a result of further development of the products. An obligation to provide the respective characteristics shall only exist if expressly agreed in the terms of contract.

All product designations may be trademarks or product names of Siemens AG or supplier companies whose use by third parties for their own purposes could violate the rights of the owners.

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

advertisement