Midmark ClassicSeries Wet-Ring Vacuums Maintenance Guide

Midmark ClassicSeries Wet-Ring Vacuums Maintenance Guide

Below you will find brief information for Wet-Ring Vacuums ClassicSeries Wet-Ring Vacuums. This maintenance guide provides a comprehensive schedule for daily, and annual service to ensure years of reliable operation. It outlines the steps for cleaning and replacing essential components such as vacuum piping, filters, and strainers. This guide also includes instructions on how to operate the water by-pass valve for troubleshooting clogs in the recycler.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Wet-Ring Vacuums ClassicSeries Wet-Ring Vacuums Maintenance Guide | Manualzz

Vacuum Level Adjustment

Procedure

Note: Vacuum Relief Valve is factory set at 10” Hg.

Perform “Vacuum Level Adjustment” Procedure if adjustment is needed. Never adjust the VRV above

12” Hg on the vacuum gauge.

Vacuum Relief Valve Filter

Single Model

Increase “Hg

Decrease “Hg

AA186500

A. Rotate barrel clockwise to increase

“Hg.

B. Rotate barrel counterclockwise to decrease “Hg.

VRV Location on

Twin Model

Pull filter out of Vacuum Relief

Valve end cap to clean or replace.

Twin Model

AA199800

LIMITED WARRANTY

SCOPE OF WARRANTY

Midmark Corporation (“Midmark”) warrants to the original purchaser its new ClassicSeries™ products and components (except for components not warranted under “Exclusions”) manufactured by Midmark to be free from defects in material and workmanship under normal use and service. Midmark’s obligation under this warranty is limited to the repair or replacement, at

Midmark’s option, of the parts or the products the defects of which are reported to Midmark within the applicable warranty period and which, upon examination by Midmark, prove to be defective.

APPLICABLE WARRANTY PERIOD

The applicable warranty period, measured from the date of installation for the original user, shall be:

• ClassicSeries™ Wet Ring Vacuums: Five (5) years or 10,000 usage hours (whichever comes first)

EXCLUSIONS

This warranty does not cover and Midmark shall not be liable for the following: (1) repairs and replacements because of misuse, abuse, negligence, alteration, accident, freight damage, or tampering; (2) products which are not installed, used, and properly cleaned as required in the Midmark “Installation” manual and or

“maintenance” guide for this applicable product; (3) products considered to be of a consumable nature; (4) accessories or parts not manufactured by Midmark; (5) charges by anyone for adjustments, repairs, replacement parts, installation, or other work performed upon or in connection with such products which is not expressly authorized in writing in advance by Midmark.

EXCLUSIVE REMEDY

Midmark’s only obligation under this warranty is the repair or replacement of defective parts. Midmark shall not be liable for any direct, special, indirect, incidental, exemplary, or consequential damages or delay, including, but not limited to, damages for loss of profits or loss of use.

NO AUTHORIZATION

No person or firm is authorized to create for Midmark any other obligation or liability in connection with the products.

THIS WARRANTY IS MIDMARK’S ONLY WARRANTY AND

IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR

IMPLIED. MIDMARK MAKES NO IMPLIED WARRANTIES

OF ANY KIND INCLUDING ANY WARRANTIES OF

MERCHANTABILITY OR FITNESS FOR ANY PARTICU-

LAR PURPOSE. THIS WARRANTY IS LIMITED TO THE

REPAIR OR REPLACEMENT OF DEFECTIVE PARTS.

AA199500

ClassicSeries™

Wet-Ring Vacuums

Maintenance Guide

To ensure years of reliable service, follow the maintenance schedule outlined in this guide.

If service is required, contact your authorized

Midmark dealer.

Dealer (name / phone):

__________________________________

__________________________________

__________________________________

Model Number:

Serial Number:

__________________

__________________

Date of Purchase: _________________

Date:

Annual Maintenance Log

Services Performed: Technician:

_____ _______________________ __________

_____

_____

_____

_____

_____

_____

_____

_____

_____

_______________________

_______________________

_______________________

_______________________

_______________________

_______________________

_______________________

_______________________

_______________________

© Midmark Corporation 2007

10546500 Rev. E (2/12)

__________

__________

__________

__________

__________

__________

__________

__________

__________

Daily

A. Clean Vacuum Piping System.

Clean piping system with a non-foaming vacuum system cleaner. Midmark recommends

Precision Clense™, Order No. 77000570.

B. Clean Vacuum Inlet Strainer.

Turn vacuum “OFF”. Remove, replace and or clean filter bowl, screen and gasket. Reassemble. Turn vacuum “ON”.

Note - Recycler Models Only

If you have low to no suction due to a clog by residue building up over time, there is a feature on vacuum to keep it running until a service technician can troubleshoot the problem. Turn the water by-pass valve to the OPEN position. This will by-pass the recycler and may keep the pump running.

Water

Inlet

Strainer

Water

By-Pass

Valve In

Open

Position

Recycler

Models

Non-Recycler

Models

Annually

A. Clean and or Replace Vacuum Relief

Filter.

Pull filter out of end cap. Blow off filter and or replace if damaged. Replace filter in VRV end cap.

B. Clean Water Inlet Strainer.

Turn vacuum “OFF”. Remove cap on “Y” strainer housing and pull out the wire intake strainer.

Rinse with water and blow off. Replace strainer and cap. Turn vacuum “ON”.

C. Clean / Dust Vacuum Pump(s).

Wipe with a clean cloth to remove dust & debris.

CLOSED Position

Shown

For Recycler To Be

Activated

During Normal

Operation The Water

By-Pass Valve Should

Be In the CLOSED postion.

Vacuum

Inlet

Strainer

AA199700

Midmark Corporation

60 Vista Drive

Versailles, Ohio 45380

1-800- Midmark (643-6275) www.midmark.com

Recycler

Models

Maintenance Kits

Vacuum Relief Valve Filter

Order #: 62935601

Includes:

• Filter

Vacuum Inlet Strainer

Order #:

Single Models – 62937301

Twin Models - 77005155

Includes:

• Filter bowl, screen and o-ring.

Dispos-A-Bowl

Order #:

Single Models – 77005167

Twin Models - 77005168

Includes:

• Container, lid and red safety tape.

Accessories

Exhaust Silencer

Order #: 77000040

Includes:

• Exhaust Silencer

ADP

Mineral Reducer Kit

Order #: AVA8200

Includes:

• Installation instructions, housing,

cartridge, all necessary mounting

and installation hardware.

Order #: PFM8200

Includes:

• Cartridge for Mineral Reducer.

advertisement

Key Features

  • Vacuum relief valve filter
  • Vacuum inlet strainer
  • Water inlet strainer
  • Vacuum pump cleaning
  • Water by-pass valve
  • Maintenance log
  • Daily cleaning
  • Annual service

Frequently Answers and Questions

How often should I clean the vacuum piping system?
The vacuum piping system should be cleaned daily with a non-foaming vacuum system cleaner.
What is the purpose of the water bypass valve?
The water bypass valve is used to bypass the recycler in case of a clog, allowing the vacuum pump to continue running until a technician can address the issue.
When should I perform annual maintenance on the vacuum?
Annual maintenance should be performed on the vacuum to ensure optimal performance. This includes cleaning and replacing components like the vacuum relief filter, vacuum inlet strainer, and water inlet strainer.