E+E Elektronik EE364 T63xx, Tx3xx Moisture in Oil Transmitter Quick Guide

E+E Elektronik EE364 T63xx, Tx3xx Moisture in Oil Transmitter Quick Guide

The EE364 T63xx and Tx3xx are moisture in oil transmitters designed for measuring the water activity (Aw) and the water content (Xk/Xm) in oil. These transmitters are suitable for use in a variety of applications, including power transformers, oil-filled electrical equipment, and hydraulic systems. The EE364 is easy to install and configure, and it comes with a variety of features that make it a reliable and accurate choice for your moisture in oil monitoring needs.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

EE364 T63xx, Tx3xx Moisture in Oil Transmitter Quick Guide | Manualzz

QUICK GUIDE

www.epluse.com

WIRING / ANSCHLUSS

EE364-T63xx

View on sensor-plug

Blick auf Sensorstecker

Output 1: 4-20 mA

8

4

Output 2: 4-20 mA 5

6

3

2

10 *) ...28V DC *) 10V+0.02*R

L 8

Output 1: 4-20 mA

4

Output 2: 4-20 mA 5 mA mA

EE364 - Moisture in Oil Transmitter

6

-

+

3

2

RS485 A

RS485 B

10 *) ...28V DC *) 10V+0.02*R

L mA

-

+ mA

RS485 A

RS485 B

4

5

3

1 2

EE364-Tx3xx

5

View on sensor-plug

Blick auf Sensorstecker

4

3

1

10 *) ...28V DC *) 10V+0.02*R

L

-

+

RS485 A

RS485 B

Note:

To satisfy the EMC Directive 2004/108/EC a shielded cable must be used.

1

10 *) ...28V DC *) 10V+0.02*R

L

-

+

Accessory HA0103xx - connecting cable for EE364-T63xx

Zubehör HA0103xx - Anschlusskabel für EE364-T63xx

RS485 A

Anmerkung:

Zur Erfüllung der EMV Richtlinie 2004/108/EG ist ein geschirmtes Kabel zu verwenden.

Accessory HA0108xx - connecting cable for EE364-Tx3xx

Zubehör HA0108xx - Anschlusskabel für EE364-Tx3xx

5 RS485 B

6

7

4

5

1

2

3

8

WH NC (not connected)

BN RS485 B

GN RS485 A

YE

GY

PK

BU

RD

Out1 4-20mA

Out2 4-20mA

GND

NC (not connected)

V+

1

2

3

4

5

BN V+

WH NC (not connected)

BU GND

BK

GY

RS485 A

RS485 B

MODBUS MAP

Instructions for Modbus-Protocol-Setup please see

Application Note AN0103 (www.epluse.com/EE364).

• The EE364 can be used in Modbus networks with max. 32 units.

• The bus termination shall be realized with 120 Ohm resistor (not included in the scope of supply)

• For proper function the power supply must be strong enough to ensure supply voltage within the specified range (see technical data) at any time and at all devices in the bus. This is particularly relevant when using long and thin cables which can cause high voltage drop.

Grundlagen zur Modbus-Protokoll-Einstellungen finden Sie in der Application Note AN0103 (www.epluse.com/EE364).

• Der EE364 ist für Modbus Netzwerke mit max. 32

Busteilnehmern geeignet.

• Der Abschluss der Busleitung (bus termination) muss mit einem 120 Ohm Widerstand (nicht im Lieferumfang enthalten) realisiert werden.

• Für eine einwandfreie Funktion muss das Netzteil stark genug sein, um die Versorgungsspannung innerhalb des angegebenen Bereichs (siehe technische Daten) jederzeit und auf allen Geräten im Bus zu gewährleisten. Dies gilt insbesondere bei der Verwendung von langen und dünnen Kabel, die einen hohen

Spannungsabfall verursachen können.

Factory setting:

Slave-ID 243; Baudrate 9600, Parity even, Stopbit 1

Werkseinstellung:

Slave-ID 243; Baudrate 9600, Parity even, Stopbit 1

32Bit FLOAT:

Coil / Register

Numbers

30052

30054

30026

30224

30226

60101

60103

Data-

Addresses

0x33

0x35

0x19

0xDF

0xE1

0x64

0x66

Parameter name water activity Aw water content Xk/Xm temperature Tx

A parameter (read)

B parameter (read)

A parameter (write)

B parameter (write)

16Bit Data:

Coil / Register

Numbers

60001

60002

Data-

Addresses

0x00

0x01

Parameter name

Slave-ID

RS485 Setting

SETUP AND ADJUSTMENT / SETUP UND JUSTAGE

The EE364 transmitter is ready to use and does not require any confi guration by the user. The factory setup of EE364 corresponds to the type number ordered. If needed, the user can change the factory setup or make an adjustment by using the optional

USB-Modbus converter (E+E part no. HA011013) and the E+E

Product Configuration Software (EE-PCS). The E+E Product

Configuration Software (EE-PCS) is free and can be downloaded from www.epluse.com/configurator .

Der EE364 ist sofort einsatzbereit und erfordert keine

Konfi guration durch den Kunden. Die Werkseinstellung des EE364 entspricht der bestellten Typenbezeichung.

Bei Bedarf kann die

Werkseinstellung mit dem optionalen USB-Modbus Konverter

(E+E Artikel Nr. HA011013) und der E+E Product Confi guration

Software (EE-PCS) geändert werden.

Die E+E Product Configuration Software (EE-PCS) steht unter www.epluse.com/configurator gratis zum Download bereit.

INSTALLATION / EINBAU

OK

DIMENSIONS / ABMESSUNGEN

mm (inch)

140 (5.5)

ISO

92 (3.6)

34 (1.36)

Ø12 (0.5)

140 (5.5) 

NOT OK

92 (3.6)

G1/2“

NPT

51 (2)

AD27

77 (3)

89 (3.5)

Ø30 (1.2)

Ø12 (0.5)

G1/2“

AD27

77 (3)

51 (2) 89 (3.5)

SCOPE OF SUPPLY / LIEFERUMFANG

- EE364 Transmitter according to Ordering Guide

- M12x1 Attachable mating connector - 8 pin, straight

- User manual - Quick Guide

- Inspection certifi cate according to DIN EN10204 - 3.1

74 (2.9)

Ø30 (1.2)

E+E Elektronik Ges.m.b.H.

Langwiesen 7 • A-4209 Engerwitzdorf

Tel: +43 7235 605-0 • Fax: +43 7235 605-8 [email protected] • www.epluse.com

LG Linz Fn 165761 t • UID-Nr. ATU44043101

Place of Jurisdiction: A-4020 Linz • DVR0962759

1/2“NPT AD22

74 (2.9)

- EE364 Messumformer lt. Bestellinformation

- Konfektionierbarer Gegenstecker M12x1, 8 pol., gerade

- Bedienungsanleitung - Quick Guide

- Abnahmeprüfzeugnis gemäß DIN EN10204-3.1

BA_ee364 // v3.3 // Technische Änderungen vorbehalten // 302724

advertisement

Key Features

  • Measures water activity and water content in oil
  • 4-20mA analog output and RS485 digital communication
  • Suitable for use in a variety of applications
  • Easy to install and configure
  • Reliable and accurate
  • Shielded cable required for EMC compliance
  • MODBUS protocol support for network integration
  • Optional USB-Modbus converter for configuration

Questions & Answers

2024-08-05

H H N

How do I connect the EE364 transmitter to a Modbus network?
[Answer based on document]

Frequently Answers and Questions

What is the maximum number of units that can be used in a Modbus network?
The EE364 can be used in Modbus networks with max. 32 units.
What type of cable must be used to satisfy the EMC Directive?
To satisfy the EMC Directive 2004/108/EC a shielded cable must be used.
What is the factory setting for the EE364?
Factory setting: Slave-ID 243; Baudrate 9600, Parity even, Stopbit 1
Can the factory settings be changed?
Yes, the factory setup can be changed by using the optional USB-Modbus converter (E+E part no. HA011013) and the E+E Product Configuration Software (EE-PCS).

Related manuals