ICS Electronics 8065 Ethernet-to-GPIB Controller Application Bulletin

ICS Electronics 8065 Ethernet-to-GPIB Controller Application Bulletin

Below you will find brief information for Ethernet-to-GPIB Controller 8065. This document explains how to program the 8065 device using the VXI-11 RPC protocol. It also includes some code examples for a simple program that reads the IDN message from a device on the instrument address.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Ethernet-to-GPIB Controller 8065 Application Bulletin | Manualzz
ICS
AB80-3
ICS
ELECTRONICS
division of Systems West Inc.
APPLICATION
BULLETIN
VXI-11 RPC Programming Guide for the 8065
An Introduction to RPC Programming
INTRODUCTION
channel then allows the client to request of the server the execution
of channel-specific set of remote procedure calls. RPC calls are the
requests to the server to execute one of the functions.
ICS's new 8065 Ethernet-to-GPIB Controller is a VXI-11 compatible device. The 8065 can be programmed with calls to a VXI-11
compliant VISA1, with Agilent's SICL library or with VXI-11 RPC
commands. This Application Note describes what RPC commands
are and shows how RPC calls can be used with ICS's 8065 to control
GPIB instruments.
The programmer should have a RPC package for his system that
includes a protocol converter (rpcgen). Because the VXI-11 Specification contains an RPCL (RPC Language) description of the protocol
in Section C, it is possible to use the protocol generator (rpcgen)
to generate client and server stubs for C language programs which
makes application development much easier.
VXI-11 INTRODUCTION
The VXI-11 Specification specifies a protocol for communication
with (test or measurement) devices over a network (LAN) via TCP/IP.
This protocol uses the ONC/RPC (Open Network Computing/Remote
Procedure Call) standard which is based on TCP/IP and is described
in the ‘VXI-11: TCP/IP Instrument Protocol Specification’.
For more information on RPC, the readers may want to look at the
references at the end of this Application Note.
CHANNELS
The VXI-11 standard states that a network instrument host has to
implement a certain set of RPC functions on three different channels. These are the:
• Core Channel: send commands from the Network interface client
to the network interface server,
• Abort Channel: abort a previously sent (core channel) command,
• Intr Channel: send an interrupt from the network interface (8065
as a client) to the Application (as a server).
VXI-11 is the overall VXI-11 document and describes the network
protocol. There are three sub-specifications. VXI-11.1 is for a VXI
chassis and is not applicable to the 8065. The VXI-11.2 Specification defines how the gateway, like the 8065, controls the GPIB bus
and performs bus specific functions. The VXI-11.3 Specification
describes instrument specific functions including data transfer to
and from an instrument.
ICS's 8065 has both VXI-11.2 and VXI-11.3 capability so it can
be used just as you would any 488.2 Controller. Generating IFCs
and performing 488.2 protocols such as FindLstn are no problem
for the 8065. Competitive units with only VXI-11.3 capability are
limited to just device specific functions.
CORE CHANNEL
The core channel is used to establish links to the 8065 and to the
devices on the bus. The links can be created and destroyed. The
create_link call takes an identifying string as parameter and returns,
if successful, a handle (of type Device_Link), which is to be used
in following calls like device_read or device_write. All addressing
of the devices is hidden behind this handle. Additionally one may
call create_link with an identification string of the network interface
server itself (no special device). The resulting link can then be used
to perform GPIB bus operations or modify the server's operation.
RPC
RPC stands for Remote Procedure Calls. An RPC server, like the
8065, implements a desired set of RPC protocols to carry out specific
functions. In RPC language they are called programs and these can
exist in different versions. If a client and a server agree on a certain
program in a certain version, they as a result establish a channel. This
Notes: 1. Some VISA libraries like those from National Instruments do not provide
interface level commands and limit the 8065 to just VXI-11.3 functions.
1
09-06-11
In the special situation of the GPIB server (8065), the identifying
string is “gpib0” for the server and “gpib0,<n>” for devices, where
<n> is the GPIB address of the device. Some calls like device_docmd
only work with the server link, and not with device links. Other
commands like device_read and device_write are intended for
device links.
an SRQ handler or RPC server, and uses the create_intr_chan call
to ask the 8065 to create an intr channel.
To generate the interrupt, the client enables the device to generate
the SRQ and gives it an id key. This also sets the 8065 to Auto
Serial Poll. When the enabled instrument generates a SRQ, the
8065 serial polls the enabled device(s) and passes a message with
the id key of the SRQ generator to the RPC service function in the
PC (Application). The RPC service function receives but does not
reply to the interrupt message. Instead a flag or other signal should
be set so that the main routine in the client application will read the
status byte from the instrument who generated the SRQ and service
the instrument.
Since the network interface server resides in a network, it is possible for more than one client to establish a connection to the server.
To avoid concurrent access to devices, links can be locked by the
client. The device_lock and device_unlock calls are used for this
purpose. Because a one client might not be aware of the existence
of another client, it is possible to specify a lock_timeout in the calls
so the calls wait a certain number of milliseconds for an existing
lock to be released.
For more details on creating and using the intr channel, see Application Note AB80-4 'SRQ Handling with a VXI-11 Interrupt Channel
on ICS’s Model 8065 Ethernet-to-GPIB Controller'
FLAGS FIELD
SIMPLE PROGRAM STEPS
The Flags field is only used during actual Read/Write operations. Of
the three fields, one is used for writing and one is used for reading,
with the third flag (WaitLock) used for both operations.
The following are the basic steps for writing a simple program that
reads the IDN message from an instrument at address 28.
The End flag is used when writing to an instrument. In the GPIB
world, a write operation normally has the last character sent with
the EOI signal asserted to indicate that the current character is the
last character of the transmission. It is possible to write multiple
blocks of data as a single write, if the End flag is zero on all but the
last block. The last block of a write should always have the End flag
set. If the device wants a terminating character, such as a linefeed
(0x0A), it can be added as the last character in the message.
1. Initialize the RPC layer.
2. Using RPC, send the following VXI-11 commands.
a) create_link to “gpib0,28”
b) device_write “*IDN?” with the End flag set to one
c) device_read with TermChrSet set to zero
d) destroy_link
3. Close the RPC layer
USING THE GENERATED PROTOCOL
The TermChrSet is used when reading from an instrument. Most
instruments use a terminating character (normally a linefeed, 0x0A)
at the end of the data message to terminate the message. (Binary
data is terminated by asserting EOI on the last character of the
message) If TermChrSet is set to one, then the character will cause
the 8065 to terminate the read and return all prior read data to the
client. If it is set to zero, then only an EOI will be able to terminate
a read. Note that the read length being met will always cause the
read to terminate.
The following examples are for a UNIX operating system but apply
to any operating system with an RPC package. While the examples
are believed to be correct, the user assumes all responsibility for
their operation in his system.
The VXI-11 RPCL protocol description consists of two files:
vxi11core.rpcl (in paragraph C.1) containing core and abort channel
and vxi11intr.rpcl (in paragraph C.2) containing the intr channel.
The WaitLock tells the 8065 to wait for up to lock_timeout amount
of time if another client has the instrument locked. The simply ignore the WaitLock flag while you’re learning how to communicate
with the 8065.
The rpcgen utility program creates a number of C language files
from the RPCL protocol description. The files include:
• a header file xxx.h, containing the type definitions, the function
numbers and the prototypes for the generated functions
• a C source file xxx_clnt.c containing client side functions
• a C source file xxx_svc.c containing server side code and
• a C source file xxx_xdr.c containing data type conversion functions for the data types declared in xxx.rpcl.
ABORT AND INTERRUPT CHANNEL
The intr and abort channels both support only one call (device_intr_srq
and device_abort). With the device_abort call the client can abort a
previously given (core channel) call that is still in progress. Some
GPIB operations are completed immediately and the device_abort
call will have no affect. An useful example is cancelling a read
from a device that has no data.
Not all of the generated files are needed unless you want to implement the complete protocol. The abort and intr channel are optional
in a network interface client (Application). Note that the 8065
includes a complete core, abort and intr channel implementation
for a network server.
The intr channel is used to implement GPIB service requests (SRQ).
The intr channel effectively reverses the rolls of the network interface
server and client. Here the 8065 becomes a client and the application
becomes the server. To setup the intr channel the application supplies
To illustrate how rpcgen transforms an RPCL declaration into C
source code, consider the following simple example:
2
EXAMPLE RPC -> C
/* Do the call; rpcClient is the RPC CLIENT
handle */
resp = simple_func_1(&parms, rpcClient);
This is the RPCL declaration of a function taking two integers as
parameter and returning two floats together with the type declaration
of its parameters and return values (file simple.rpcl):
/* Evaluate result */
if (resp == NULL)
{
clnt_perror(rpcClient, “<name of the
server>”);
return;
}
/* resp now points to a (static) structure
holding the
result values
*/
struct Parms
{
int p1;
int p2;
};
struct Resp
{
float a;
float b;
};
program SIMPLE
{
version SIMPLE_VERSION
{
Resp simple_func (Parms) = 1;
} = 1;
} = 0x0607AF;
Server Code
The server side only has to define the function:
Resp *simple_func_1(Parms *parms)
{
...
}
This is transformed by rpcgen to the C declarations (simple.h):
The server routine automatically calls simple_func_1 as soon as the
RPC request arrives from a client.
#include <rpc/types.h>
struct Parms {
int p1;
int p2;
};
typedef struct Parms Parms;
bool_t xdr_Parms();
UNDER UNIX
The protocol generator, rpcgen, generates source code that is directly
usable under the operating system. While this example is for UNIX,
the example applies to any operating system with an RPC package.
The basic steps are:
struct Resp {
float a;
float b;
};
typedef struct Resp Resp;
bool_t xdr_Resp();
#define
#define
#define
extern
Include Files
The following header files must be included:
SIMPLE ((u_long)0x444)
SIMPLE_VERSION ((u_long)1)
simple_func ((u_long)1)
Resp *simple_func_1();
#include <rpc/rpc.h>
#include “vxi11core.h”
#include “vxi11.h”
Open RPC Connection
The last declared function is the one the user is actually going to
call. NOTE: the generated function prototypes do not contain parameter declarations. If you want to be sure, read the corresponding
xxx_clnt.c file. The general rule is: first argument is a pointer to the
(user allocated) parameter; second argument is the CLIENT handle
as obtained from clnt_create (declared in rpc/rpc.h).
To start, we must open an RPC connection. The client handle rpcClient is used in all subsequent RPCs.
CLIENT
static char
Client Code
/* open rpc connection */
rpcClient = clnt_create(svName,DEVICE_CORE,
DEVICE_CORE_VERSION, “tcp”);
if (rpcClient == NULL)
{
clnt_pcreateerror(svName);
return 1;
}
The client side call of the remote procedure could look like this:
Parms parms;
Resp *resp;
/* Fill parms structure */
parms.p1 = 1;
parms.p2 = 2;
ICS Electronics
*rpcClient;
*svName = “box10.acc”;
7034 Commerce Circle, Pleasanton, CA 94588
3
Phone: (925) 416-1000
Fax: (925) 416-0105
Create Device Link
PORTING TO OTHER SYSTEMS
Before sending any data to our device we must create a device link.
Here the device is at a GPIB primary address of 28. Use 'gpib0,28'
as the device.
The source code generated by rpcgen uses some static variables
(e.g. for the returned structures) and is therefore not reentrant. The
user has to take this into account when writing his application. Of
interest are the following generated files:
/* fill Create_LinkParms structure */
crlp.clientId = rpcClient;
crlp.lockDevice = 0;
crlp.lock_timeout = 10000;
crlp.device = “gpib0,28”;
• vxi11core_clnt.c: client side functions for the core and abort
channel
• vxi11intr_svc.c: server side functions for the intr channel
Be especially aware of this limitation when the RPC code is called
by a higher level library or application with multi-treaded capability.
One such example is VxWorks.
/* call create_link on instrument server */
crlr = create_link_1(&crlp, rpcClient);
if (crlr == NULL)
{
clnt_perror(rpcClient, svName);
return 1;
};
SuSE EXAMPLE
The following is a complete Linux example for SuSE 10.0. The
example creates links to an ICS 8065 then to a HP437B Power
Meter at GPIB address 12.
Example: Send String
To send a string (cl) to our device we do the following:
Device_WriteParms
Device_WriteResp
#include “vxi11core.h”
#include <stdio.h>
dwrp;
*dwrr;
#define WAITLOCK_FLAG 1
#define WRITE_END_CHAR 8
#define READ_END_CHAR 0x80
#define VXI_ENDW (WAITLOCK_FLAG | WRITE_END_
CHAR)
dwrp.lid = crlr->lid;
dwrp.io_timeout = IO_TIMEOUT;
dwrp.lock_timeout = LOCK_TIMEOUT;
dwrp.flags = VXI_ENDW;
dwrp.data.data_len = strlen(dval);
dwrp.data.data_val = dval;
CLIENT *rpcClient;
Create_LinkParms crlp;
Create_LinkResp *crlr;
dwrr = device_write_1(&dwrp, cl);
...
/* error test here like above */.
Device_WriteParms dwrp;
Device_WriteResp *dwrr;
Note that most GPIB devices use the linefeed terminating character
to terminate a ASCII string. Newer IEEE-488.2 devices can also
sense the EOI line when it is asserted on the last character of the
string. Use the End flag when writing. Set the End Flag to 1. Use
TermChrSet to set the terminating character. Specify the TermChr
within the device_read command.
Device_ReadParms drdp;
Device_ReadResp *drdr;
Device_Error *derr;
Destroy Device Link
static char *svName = “192.168.46.105”;
If the link to the device is no longer needed destroy it:
int main(){
char sendMessage[1024];
char responseMessage[2048];
rpcClient = clnt_create(svName, DEVICE_CORE, DEVICE_
CORE_VERSION, “tcp”);
if (rpcClient == NULL){
printf(“Error creating client\n”);
clnt_pcreateerror(svName);
return 1;
}
printf(“Hello %s\n”, svName);
crlp.clientId = rpcClient;
crlp.lockDevice = 0;
crlp.lock_timeout = 10000;
crlp.device = “gpib0,12”;
Device_Error *derr;
derr = destroy_link_1(&(crlr->lid), rpcClient);
Close RPC Connection
At the and of the program the client handle must be destroyed.
clnt_destroy(rpcClient);
4
SUMMARY
crlr = create_link_1(&crlp, rpcClient);
if (crlr == NULL){
clnt_perror(rpcClient, svName);
return 1;
}
This application note has described RPC calls and shown how the
calls are generated by the rpcgen utility. A brief examples are provided so the reader can better understand how RPC calls are used
with the 8065 to control GPIB devices.
printf(“Link created to %s\n”, crlp.device);
This application note is only intended to introduce the reader to the
RPC concept and guidelines for programming with the 8065. The
following references are recommended reading before attempting
RPC coding:
dwrp.lid = crlr->lid;
dwrp.io_timeout = 1000;
dwrp.lock_timeout = 10000;
dwrp.flags = VXI_ENDW;
sprintf(sendMessage, “*IDN?\n”);
dwrp.data.data_len = strlen(sendMessage);
dwrp.data.data_val = sendMessage;
1. "Cisco IOS for S/390 RPC/XDR Programmer’s Reference".
http://www.cisco.com/univercd/cc/td/doc/product/software/ioss390/
ios390rp/
dwrr = device_write_1(&dwrp, rpcClient);
if (dwrr == NULL){
clnt_perror(rpcClient, svName);
return 1;
}
printf (“device_write returns error code %d\n”, dwrr->error);
2. Halfway down, “Managing the Server”
http://publib16.boulder.ibm.com/pseries/en_US/aixprggd/progcomc/rpc_ref.htm
3. Builds and registers a simple RPC service.
http://publib16.boulder.ibm.com/pseries/en_US/aixprggd/progcomc/rpc_lowest_ex.htm#a287x92028
drdp.lid = crlr->lid;
drdp.io_timeout = 1000;
drdp.lock_timeout = 10000;
drdp.flags = VXI_ENDW;
drdp.termChar = ‘\n’;
drdp.requestSize = 1024;
4. Similar to the second URL, but with more examples.
http://www.unet.univie.ac.at/aix/aixprggd/progcomc/rpc_ref.htm
The reader is urged to use Google to search for an RPC example or
programming information for his system since RPC implementations vary from system to system and an example that works on one
system may not work on another system.
drdr = device_read_1(&drdp, rpcClient);
if (drdr == NULL){
clnt_perror(rpcClient, svName);
return 1;
}
printf(“device read returns error %d, reason %d.\n”, drdr->error,
drdr->reason);
printf(“response = %s\n”, drdr->data.data_val);
DISCLAIMER
While the examples herein are correct to the best of our knowledge,
the user assumes all responsibility for their operation in his system.
The authors are not responsible for the examples' operation.
ACKNOWLEDGEMENTS
The author wishes to thank B. Franksen at Bessy GmbH in Germany for his paper 'VXI-11 and HP E2050' Many of his ideas are
presented in this Application Note.
derr = destroy_link_1(&(crlr->lid), rpcClient);
clnt_destroy(rpcClient);
Additional thanks to Bob Clemmons at Aeroflex for his patience
and SuSE example.
}
DEBUGGING HINTS
Lost device LIDs, lost data or overwritten LIDs may be caused by
not saving the LID or data before calling the next RPC function.
Some RPC implementations return a pointer to a LID or data buffer when executing a create_link or read call. The user must read
and save the values in his own variables to keep them from being
overwritten by a subsequent RPC call.
ICS Electronics
7034 Commerce Circle, Pleasanton, CA 94588
5
Phone: (925) 416-1000
Fax: (925) 416-0105

advertisement

Key Features

  • VXI-11 compliant
  • Ethernet-to-GPIB interface
  • RPC programming support
  • GPIB bus control
  • Code examples provided

Frequently Answers and Questions

How can I send a string to a device using the 8065?
To send a string to a device, you would need to create a device link, fill the Device_WriteParms structure with the string data and the End flag set to 1, and then call the device_write_1 function.
What is the purpose of the TermChrSet flag?
The TermChrSet flag is used when reading data from a device. It determines if the read operation will terminate on encountering a specific terminating character (TermChr) or only on receiving the EOI signal. If TermChrSet is set to 1, the read will terminate upon encountering the specified TermChr, otherwise it'll only terminate upon receiving the EOI signal.
How do I handle SRQs (Service Requests) with the 8065?
The 8065 supports SRQ handling using an interrupt channel. You can set up an SRQ handler in your application and use the create_intr_chan call to create an interrupt channel on the 8065. The 8065 will then automatically pass SRQ messages to your SRQ handler.