Mercoid 3200G Pressure Transmitter Datasheet

Mercoid 3200G Pressure Transmitter Datasheet
110_Layout 2 7/12/13 2:04 PM Page 110
PRESSURE
Series
3200G
Explosion-proof Pressure Transmitter
®
HART , Push Button Configuration, Rangeability (100:1)
4-1/4 [108]
2-3/4
[70]
Single Pressure Transmitters
3-1/2 [89]
1-1/2
[38]
6-1/4
[159]
1/2 NPT
The Mercoid® Series 3200G Smart Pressure Transmitter is a microprocessor-based
high performance transmitter, which has flexible pressure calibration, push button
configuration, and programmable using HART® Communication. The Series 3200G is
capable of being configured with the zero and span buttons, a field calibrator is not required
for configuration. The transmitter software compensates for thermal effects, improving
performance. EEPROM stores configuration settings and stores sensor correction
coefficients in the event of shutdowns or power loss. The Series 3200G is FM approved for
use in hazardous (Classified) locations. The 100:1 rangeability allows the smart transmitter
to be configured to fit any application.
FEATURES
• Completely configurable using zero/span buttons (No calibrator required)
• Rangeability (100:1)
• High accuracy (±0.075%)
• Automatic sensor temperature compensation
• Fail-mode process function
Model
3200G-1-FM-1-1
3200G-2-FM-1-1
3200G-3-FM-1-1
3200G-4-FM-1-1
3200G-1-FM-1-1-LCD
3200G-2-FM-1-1-LCD
3200G-3-FM-1-1-LCD
3200G-4-FM-1-1-LCD
Range
psi (kPa)
-14.5 to 21 (-100 to 150)
-14.5 to 217 (-100 to 1500)
0 to 725 (0 to 5000)
0 to 3600 (0 to 25000)
-14.5 to 21 (-100 to 150)
-14.5 to 217 (-100 to 1500)
0 to 725 (0 to 5000)
0 to 3600 (0 to 25000)
Calibrated Span
(Min. to Max.) psi (kPa)
0.22 to 21 (1.5 to 150)
2 to 217 (15 to 1500)
7.25 to 725 (50 to 5000)
36 to 3600 (250 to 25000)
0.22 to 21 (1.5 to 150)
2 to 217 (15 to 1500)
7.25 to 725 (50 to 5000)
36 to 3600 (250 to 25000)
SPECIFICATIONS
Service: Compatible gases, steam, liquids or vapors.
Wetted Materials: 316L SS.
Accuracy: ±0.075% FS (@ 20°C).
Rangeability: 100:1 turn down.
Stability: ±0.125% FSO/yr.
Temperature Limits:
Process: -40 to 248°F (-40 to 120°C);
Ambient: Without LCD -40 to 185°F (-40 to 85°C);
With LCD -22 to 176°F (-30 to 80°C).
Thermal Effect: ±0.125% span/32°C.
Power Requirements: 11.9 to 45 VDC.
Output Signal: 4 to 20 mA / HART® Communication.
Response Time: 0.12 seconds.
Damping Time: 0.25 to 60 seconds.
Loop Resistance:
Operation: 0 to 1500 Ω;
HART® Communication: 250 to 500 Ω.
Electrical Connection: Two 1/2˝ female NPT conduit, screw terminal.
Process Connections: 1/2˝ female NPT.
Display: Optional 5 digit LCD.
Enclosure Rating: NEMA 4X (IP66) and explosion proof for Class I,
Div I Groups A, B, C and D.
Weight: 5.5 lb (2.5 kg).
Agency Approvals: ATEX, CE, FM.
Max. Pressure
psi (bar)
58 (4)
580 (40)
2000 (138)
10000 (690)
58 (4)
580 (40)
2000 (138)
10000 (690)
LCD Display
No
No
No
No
Yes
Yes
Yes
Yes
Contact factory for custom calibration.
ACCESSORIES
A-630, Stainless steel angle type bracket with SS bolts
A-631, Stainless steel flat type bracket with SS bolts
BBV-0N, 2-valve block manifold
DevCom2000, HART® Communication Protocol Software
HART® is a registered trademark of Hart Communication Foundation.
110
DWYER INSTRUMENTS, INC. | www.dwyer-inst.com
111_Layout 2 7/12/13 2:04 PM Page 111
3200G-2-FM-3-1-LESS2A105S2-05-LCD
3200G Explosion-Proof Pressure Transmitter
-14.5 to 21 psig (factory set 0 to 21 psig)
-14.5 to 217 psig (factory set 0 to 217 psig)
0 to 725 psig
0 to 3600 psig
0 to 8500 psig
FM approved
ATEX approved
NEMA 4X (IP66) [Only available with 316 SS housing]
1/2˝ female NPT
Diaphragm seal
1/2˝ female NPT
1 extended diaphragm seal direct mount
1 extended diaphragm seal capillary type high
1 flush diaphragm seal direct mount
1 flush diaphragm seal capillary type
2˝ (50 mm) 316L SS
3˝ (80 mm) 316L SS
ANSI class 150#
ANSI class 300#
DIN PN 10/16
DIN PN 25/40
JIS 10 K
JIS 20 K
No extension [standard for flush mount]
2˝ extension
4˝ extension
6˝ extension
316L SS diaphragm
PTFE and 316L SS diaphragm
Hastelloy C-276 diaphragm
Tantallum diaphragm
Silicon oil (-40 to 400°F)
0 to 20 feet
5 digit LCD
316 SS housing (Only available with WP approval)
NIST calibration
Custom calibration
LCD
00
05
10
15
S
P
H
T
Fill Fluid
Capillary Length
Options
2
XX
LCD
SSH
NIST
CC
Single Pressure Transmitters
Diaphragm Material
05 S 2 05
PRESSURE
3200G 2 FM
1 LES S2 A1
3200G
3
1
2
3
4
5
Approval
FM
ATEX
WP
Process Connection
1
3
Electrical Connection
1
Diaphragm Seal Type
LED
LES
LFD
LFS
Mounting Flange
S2
S3
Mounting Flange
A1
Rating
A2
D1
D2
J1
J2
Extension Length
Example
Series
Range
Custom Calibration Values
in H2O, ft H2O, mm H2O, in Hg, psig, g/cm2, kg/cm2, Pa, kPa, bar, mbar, Torr, Atm, mm Hg
Primary Units
Upper Range Limit 20 mA value
Lower Range Limit 4 mA value
0 to 60 seconds
Damping Time
Primary unit, %, mA, rotate
Display Mode
Model
DevCom2000
®
HART Communication Protocol Software
Includes USB HART Modem
DevCom2000 Software
HART Field Device
USB HART Modem
Windows®-Based PC
The Model DevCom2000 HART Communicator Software turns your PC into a fullfeatured HART communicator. Now it is possible to configure transmitters and control
valves at the desktop or in the field. DevCom2000 uses device descriptions (DDs) to retrieve
data that is stored in the memory of smart field devices. This software is a simple, reliable
and secure method to add new measurement values to control systems without the need of
additional wires. This software eliminates the need to purchase and maintain a separate
handheld HART communicator.
®
®
®
FEATURES
• Complete DD library
• Includes USB HART modem
• USB 1.1 and 2.0 compatible
• Self powered modem
SPECIFICATIONS
HART Communicator Software
DD Library: Included.
Generic DD: Included.
Operating System: Windows NT ,
Windows 2000, WindowsXP ,
Windows Vista (32/64), Windows 7
(32/64).
®
®
®
®
®
®
USB HART MODEM
Material: High strength ABS plastic.
Temperature Limits: 0 to 50°C (32 to
122°F).
Storage Temperature: -40 to 85°C
(-40 to 185°F).
Humidity: 0 to 99% (non-condensing).
HART Cable Length: 4’ (1.2 m).
Model DevCom2000, HART Communicator Software
®
HART Cable Connectors: Minigrabber.
USB Cable Length: 18˝ (0.5 m).
USB Cable Connector: USB Type A.
USB: USB 1.1, USB 2.0.
Power: USB port provides power to
unit.
Current Draw: 20 mA.
Output: 600 mVpp.
Leakage: < 10 uA.
Isolation Voltage: 1500 VDC.
HART : HART 4, HART 5, HART 6,
HART 7, HART Physical Layer Spec
HCF_SPEC-54.
Weight: 3 oz (85 g).
Agency Approval: CE.
®
®
®
®
®
®
®
®
Windows , Windows Vista , Windows NT® , Windows XP®, are registered trademarks of Microsoft Corporation
®
®