Keysight Noise Figure Measurement Capability Upgrade Kit

Keysight Noise Figure Measurement Capability Upgrade Kit
Keysight Noise Figure Measurement
Capability Upgrade Kit
To Upgrade PNA-X N5241A Option 419 or Option 423, N5242A
Option 419 or Option 423, N5249A Option 419 or Option 423,
N5241A-H85, N5242A-H85, or N5249A-H85 to include Option
029
Upgrade Kit Order Numbers: N5241AU-929, N5242AU-929, and
N5249AU-929
Keysight Kit Number: N5242-60106
Installation Note
Notices
© Keysight Technologies, Inc.
2009-2016
No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Keysight Technologies, Inc. as
governed by United States and
international copyright laws.
Trademark Acknowledgments
Manual Part Number
N5242-90010
Edition
Edition 1, March 21, 2016
Printed in USA/Malaysia
Published by:
Keysight Technologies
1400 Fountaingrove Parkway
Santa Rosa, CA 95403
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES,
EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED
WARRANTIES OF
MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
KEYSIGHT SHALL NOT BE LIABLE
FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE
FURNISHING, USE, OR
PERFORMANCE OF THIS
DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH
THESE TERMS, THE WARRANTY
TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance
with the terms of such license.
U.S. Government Rights
The Software is “commercial
computer software,” as defined
by Federal Acquisition Regulation
(“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and
Department of Defense FAR
Supplement (“DFARS”) 227.7202,
the U.S. government acquires
commercial computer software
under the same terms by which
the software is customarily
provided to the public.
Accordingly, Keysight provides
the Software to U.S. government
customers under its standard
commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of
which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA
represents the exclusive authority
by which the U.S. government
may use, modify, distribute, or
disclose the Software. The EULA
and the license set forth therein,
does not require or permit,
among other things, that
Keysight: (1) Furnish technical
information related to
commercial computer software
or commercial computer
software documentation that is
not customarily provided to the
public; or (2) Relinquish to, or
otherwise provide, the
government rights in excess of
these rights customarily provided
to the public to use, modify,
reproduce, release, perform,
display, or disclose commercial
computer software or
commercial computer software
documentation. No additional
government requirements
beyond those set forth in the
EULA shall apply, except to the
extent that those terms, rights, or
licenses are explicitly required
from all providers of commercial
computer software pursuant to
the FAR and the DFARS and are
set forth specifically in writing
elsewhere in the EULA. Keysight
shall be under no obligation to
update, revise or otherwise
modify the Software. With
respect to any technical data as
defined by FAR 2.101, pursuant
to FAR 12.211 and 27.404.2 and
DFARS 227.7102, the U.S.
government acquires no greater
than Limited Rights as defined in
FAR 27.401 or DFAR 227.7103-5
(c), as applicable in any technical
data.
Safety Notices
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in damage to the
product or loss of important data. Do
not proceed beyond a CAUTION
notice until the indicated conditions
are fully understood and met.
A WARNING notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated
conditions are fully understood and
met.
Keysight Noise Figure Measurement Capability Upgrade Kit
Upgrade Kit Order Number: N5241AU-929, N5242AU-929, and
N5249AU-929
Installation Note
Description of the Upgrade
This kit may contain either a N5242-60098 or a N5245-60124. Either board
will work in your PNA. Refer to Table 1 on page 8.
This upgrade adds noise figure measurement capability to your Option 419 or
423 4-port analyzer by adding Option 029 which includes:
— a noise down converter and noise receiver
— a bypass switch in ports 1 and 2
3
-
Description of the Upgrade
Getting Assistance from Keysight
Getting Assistance from Keysight
By internet or phone, get assistance with all your test and measurement needs.
Contacting Keysight
Assistance with test and measurements needs and information on finding a
local Keysight office are available on the Web at:
http://www.keysight.com/find/assist
If you do not have access to the Internet, please contact your Keysight field
engineer.
In any correspondence or telephone conversation, refer to the Keysight
product by its model number and full serial number. With this information,
the Keysight representative can determine whether your product is still
within its warranty period.
If You Have Problems With the Upgrade Kit Contents
Keysight stands behind the quality of the upgrade kit contents. If you have
problems with any item in the kit, email Keysight Component Test Division
(CTD) Support at support_ctd-soco@keysight.com, or telephone the CTD
Hotline at (707) 577-6802 and leave a voice mail message. Please include
details of the problem.
4
Installation Note N5242-90010
Description of the Upgrade
Getting Prepared
Getting Prepared
The PNA contains extremely sensitive components that can be ruined if
mishandled. Follow instructions carefully when making cable connections,
especially wire harness connections.
The person performing the work accepts responsibility for the full cost of
the repair or replacement of damaged components.
To successfully install this upgrade kit, you will need the following:
— A license key - refer to “License Key Redemption” below.
— A PDF copy or a paper copy of the PNA Service Guide - refer to
“Downloading the Online PNA Service Guide” below.
— An ESD-safe work area - refer to “Protecting Your Workspace from
Electrostatic Discharge” below.
— Correct tools - refer to “Tools Required for the Installation” on page 7.
— Enough time - refer to “About Installing the Upgrade” on page 7.
— Test equipment for the post-upgrade adjustments. To view the equipment
list, click the Chapter 3 bookmark “Tests and Adjustments” in the PDF
Service Guide1.
License Key Redemption
The enclosed Option Entitlement Certificate is a receipt, verifying that you
have purchased a licensed option for the PNA of your choice. You must
now use a Keysight Web page to request a license key for the instrument
that will receive the option.
To enable the option product, you must request a license key from:
http://www.keysight.com/find/softwarelicense. To complete the request, you
will need to gather the following information:
— From the certificate
-
Order number
-
Certificate number
— From your instrument
-
Model number
-
Serial number
-
Host ID
1. See “Downloading
Installation Note N5242-90010
the Online PNA Service Guide” on page 6.
5
-
Description of the Upgrade
Getting Prepared
The instrument information is available on the network analyzer – on the
analyzer’s Help menu, click About Network Analyzer.
If you provide an email address, Keysight will promptly email your license key.
Otherwise, you will your receive your license key via postal mail.
Downloading the Online PNA Service Guide
To view the online Service Guide for your PNA model number, use the following
steps:
1. Go to www.keysight.com.
2. In the Search box, enter the model number of the analyzer (Ex: N5242A)
and click Search.
3. Click Technical Support > Manuals.
4. Click Service Manual.
5. Click the service guide title to download the PDF file.
6. When the PDF of the Service Guide is displayed, scroll through the
Contents section bookmarks to locate the information needed.
Protecting Your Workspace from Electrostatic Discharge
For information, click on the Chapter 1 bookmark, “Electrostatic Discharge
Protection” in the PDF Service Guide1.
ESD Equipment Required for the Installation
Description
Keysight Part Number
ESD grounding wrist strap
9300-1367
5-ft grounding cord for wrist strap
9300-0980
2 x 4 ft conductive table mat and 15-ft grounding wire
9300-0797
ESD heel strap (for use with conductive floors)
9300-1308
1. See “Downloading
6
the Online PNA Service Guide” on page 6.
Installation Note N5242-90010
Description of the Upgrade
Getting Prepared
Tools Required for the Installation
Description
Qty
Part Number
T-10 TORX driver (set to 9 in-lbs)
1
N/A
T-20 TORX driver (set to 21 in-lbs)
1
N/A
5/16-in torque wrench (set to 10 in-lbs)
1
N/A
5/16-in torque wrench (set to 21 in-lbs)
1
N/A
1-in torque wrench (set to 72 in-lbs)
1
N/A
Use a 5/16-in torque wrench set to 10 in-lbs on all cable connections
except the front and rear panel bulkhead connectors. Torque these
connections to 21 in-lb.
About Installing the Upgrade
Products affected
N5241A, N5242A, and N5249A Option 219 or 224, and N5241A-H85,
N5242A-H85, and N5249A-H85
Installation to be performed by
Keysight service center or personnel qualified by Keysight
Estimated installation time
5.0 hours
Estimated adjustment time
0.5 hours
Estimated full instrument calibration time
4.5 hours
Installation Note N5242-90010
7
-
Description of the Upgrade
Items Included in the Upgrade Kit
Items Included in the Upgrade Kit
Check the contents of your kit against the following list. If any part is missing or
damaged, contact Keysight Technologies. Refer to “Getting Assistance from
Keysight” on page 4.
Table 1
Contents of Upgrade Kit N5242-90010
Ref
Desig.
Description
Installation note (this document)
Qty
1
Part Number
N5242-90010
Accessory Items
Note to Installer
The following two items are used when performing vector-corrected noise figure measurements and are not used in the
upgrade. They are packaged separately in the kit and must remain with the analyzer.
Female-to-female adapter used to connect an ECal module
1
85052-60013
RF cable used to connect an ECal module
1
N5242-20169
Was
N5242-20137
Assemblies
A7
Noise receiver board
1
N5242-60098
A16
Power supply
1
0950-4934
A29,
A32
Test port couplers
2
A38,
A41
Bias tee (includes wire harness)
2
5067-4865
A52
Bypass switch, port 1
1
N1811-60030
A53
Bypass switch, port 2
1
N1811-60028
A54
Bridge
1
5087-7794
A55
Noise downconverter
1
5087-7767
Gap pad for test port coupler
2
E4403-20033
Vibe mount for test port coupler
2
0460-2725
Bumper for test port coupler
1
0403-0285
Switch/attenuator bracket
2
N5242-00031
Bias tee bracket
2
N5242-00032
5087-7813
Was
5087-7710
Hardware/Miscellaneous
8
Installation Note N5242-90010
Description of the Upgrade
Items Included in the Upgrade Kit
Table 1
Contents of Upgrade Kit N5242-90010
Ref
Desig.
Description
Qty
Part Number
Machine screw, M3.0 x 18, pan head (to attach A52 and A53 to switch brackets)
4
0515-0666
Machine screw, M3.0 x 6, pan head (to attach bias tee brackets)
4
0515-0430
Machine screw, M3.0 x 20, pan head (to attach A54 to side frame)
1
0515-1410
Hex nut with lock washer (for screw 0515-1410; to attach A54 to side frame)
1
0535-0031
Machine screw, M3.0 x 10, pan head (to attach A55 to side frame)
3
0515-0374
Lower front panel overlay, Option 029 without H85
1
N5242-80012
Lower front panel overlay, Option 029 with H85
1
N5242-80019
Bulkhead connector assembly for test set front plate
13
1250-3805
Cables
W14
RF cable, A29 port 1 coupler to front-panel Port 1 CPLR ARM
1
N5222-20030
W24
RF cable, front panel port 2 CPLR THRU to A32 port 2 coupler
1
N5222-20053
W26
RF cable, A32 port 2 coupler to front-panel Port 2 CPLR ARM
1
N5222-20034
W74
RF cable, A38 port 1 bias tee to A29 port 1 coupler
1
N5222-20012
W85
RF cable, front panel port 2 CPLR THRU to A41 port 2 bias tee
1
N5222-20035
W86
RF cable, A41 port 2 bias tee to A32 port 2 coupler
1
N5222-20013
W121
RF cable, A25 port 1 bridge to A34 port 1 source attenuator
1
N5242-20273
W122
RF cable, A34 port 1 source attenuator to A52 port 1 bypass switch
1
N5242-20298
W123
RF cable, A52 port 1 bypass switch to front-panel Port 1 SOURCE OUT
1
N5242-20297
W124
RF cable, front-panel Port 1 CPLR THRU to A52 port 1 bypass switch
1
N5242-20295
W125
RF cable, A52 port 1 bypass switch to A38 port 1 bias tee
1
N5242-20296
W126
RF cable, A28 port 2 bridge to A37 port 2 source attenuator
1
N5242-20272
W127
RF cable, A37 port 2 source attenuator to A53 port 2 bypass switch
1
N5242-20292
W128
RF cable, A53 port 2 bypass switch to front-panel Port 2 SOURCE OUT
1
N5242-20303
W129
RF cable, A53 port 2 bypass switch to A54 port 2 bridge
1
N5242-20293
W130
RF cable, A52 port 2 bypass switch to A54 port 2 bridge
1
N5242-20302
W131
RF cable (male-to-male adapter), A54 port 2 bridge to A55 noise down converter
1
1250-3576
W132
RF cable, front-panel Port 1 RCVR A IN to A42 port 1 receiver attenuator
1
N5242-20277
W133
RF cable, A42 port 1 receiver attenuator to A23 mixer brick (A)
1
N5242-20275
Installation Note N5242-90010
9
-
Description of the Upgrade
Items Included in the Upgrade Kit
Table 1
Ref
Desig.
Contents of Upgrade Kit N5242-90010
Description
Qty
Part Number
W134
RF cable, A45 port 2 receiver attenuator to A23 mixer brick (B)
1
N5242-20276
W135
RF cable, front-panel Port 2 RCVR B IN to A45 port 2 receiver attenuator
1
N5242-20278
W136
RF cable, A33 reference mixer switch to A23 mixer brick (R1)
1
N5242-20274
W137
RF cable, A28 port 2 bridge to front-panel REF 2 SOURCE OUT
1
N5242-20279
W138
RF cable, front-panel REF 2 RCVR R2 IN to A23 mixer brick (R2)
1
N5242-20280
W140
RF cable, A24 mixer brick to A55 noise down converter
1
N5242-20294
W141
RF cable, A55 noise downconverter to A7 noise receiver board LO
1
N5242-20299
W142
RF cable, A55 noise downconverter J4 to A7 noise receiver board P2
1
N5242-60041
W143
RF cable, A55 noise downconverter to A7 noise receiver board RF
1
N5242-20300
W144
RF cable, A52 port 1 noise bypass switch to A29 port 1 coupler
1
N5242-20304
W153
RF cable, front panel port 2 CPLR THRU to A32 port 2 coupler
1
N5242-20310
Ribbon cable, A19 test set motherboard J548 to A55 noise downconverter J1
1
N5242-60033
Extra quantities of items such as protective plastic caps, screws, cable ties,
and cable clamps may be included in this upgrade kit. It is normal for some
of these items to remain unused after the upgrade is completed.
10
Installation Note N5242-90010
Description of the Upgrade
Installation Procedure for the Upgrade
Installation Procedure for the Upgrade
The network analyzer must be in proper working condition prior to installing
this option. Any necessary repairs must be made before proceeding with this
installation.
This installation requires the removal of the analyzer’s protective
outer covers. The analyzer must be powered down and
disconnected from the mains supply before performing this
procedure.
Overview of the Installation Procedure
“Step 1. Obtain a Keyword and Verify the Information.”
“Step 2. Remove the Outer Cover.”
“Step 3. Remove the Inner Cover.”
“Step 4. Remove the Front Panel Assembly.”
“Step 5. Remove the Existing Test Set Cables.”
“Step 6. Remove the Existing Attenuators, Brackets, and Bias Tees (If
Present).”
“Step 7. Assemble the Old Attenuators and the New A52 and A53 Bypass
Switches onto the New Brackets.”
“Step 8. Install the New Brackets, with the Old Attenuators and the New A52
and A53 Bypass Switches, into the Analyzer.”
“Step 9. Assemble the A29 and A32 Test Port Coupler Assemblies (For
Analyzers with Serial Numbers Prefixed MY/SG/US5310 and Below).”
“Step 10. Install the New (3.5 mm) A29 and A32 Test Port Couplers onto the
Front Plate (For Analyzers with Serial Numbers Prefixed MY/SG/US5310 and
Below).”
“Step 11. Assemble and Install the A54 Bridge and A55 Noise Down
converter.”
“Step 12. Install the A7 Noise Receiver Board.”
“Step 13. Install the Bulkhead Connectors in the Test Set Front Plate (For
Analyzers with Serial Numbers Prefixed MY/SG/US5310 and Below).”
“Step 14.Install the New Test Set Cables.”
“Step 15. Replace the A16 Power Supply (If Necessary).”
“Step 16. Replace the lower front panel overlay.”
“Step 17.Reinstall the Front Panel Assembly and Front Panel Jumpers.”
“Step 18. Position the Cables and Wires to Prevent Pinching.”
Installation Note N5242-90010
11
-
Description of the Upgrade
Installation Procedure for the Upgrade
“Step 19. Reinstall the Inner Cover.”
“Step 20. Reinstall the Outer Cover.”
“Step 21. Enable Option 029.”
“Step 22. Verify Option 028 is Not Installed.”
“Step 23.Perform Post-Upgrade Adjustments and Calibration.”
“Step 24. Prepare the PNA for the User.”
Step 1.Obtain a Keyword and Verify the Information
Follow the instructions on the Option Entitlement Certificate supplied to obtain
a license key for installation of this upgrade. Refer to “License Key
Redemption” on page 5.
Verify that the model number, serial number, and option number information
on the license key match those of the instrument on which this upgrade will be
installed.
If the model number, serial number, or option number do not match those on
your license key, you will not be able to install the option. If this is the case,
contact Keysight for assistance before beginning the installation of this
upgrade. Refer to “Contacting Keysight” on page 4.
Once the license key has been received and the information verified, you can
proceed with the installation at step 2.
Step 2. Remove the Outer Cover
This procedure is best performed with the analyzer resting on its front
handles in the vertical position. Do not place the analyzer on its front panel
without the handles. This will damage the front panel assemblies.
Refer to Figure 1 for this step of the procedure.
1. Disconnect the power cord (if it has not already been disconnected).
2. Remove the strap handles (item ①) by loosening the screws (item ➁), with
a T-20 TORX driver, on both ends until the handle is free of the analyzer.
3. Remove the foot locks (item ③) from the four bottom feet (item ④) and
then remove the four bottom feet from the outer cover.
4. Remove the four rear panel feet (item ⑤) by removing the center screws
(item ⑥) with a T-20 TORX driver.
5. Slide the outer cover toward the rear of the analyzer and remove it.
12
Installation Note N5242-90010
Description of the Upgrade
Installation Procedure for the Upgrade
Figure 1
Outer Cover Removal
Step 3.Remove the Inner Cover
Refer to Figure 2 for this step of the procedure.
1. With a T-10 TORX driver, remove the 12 pan head screws (item ①).
2. With a T-10 TORX driver, remove the 9 flat head screws (item ➁).
3. Lift off the cover.
Installation Note N5242-90010
13
-
Description of the Upgrade
Installation Procedure for the Upgrade
Figure 2
Inner Cover Removal
Step 4.Remove the Front Panel Assembly
Refer to Figure 3 for this step of the procedure.
1. With a 5/16-in wrench, remove all front panel jumpers (item ①).
2. With a T-10 TORX driver, remove the screws (item ➁) from the sides of the
frame.
Before removing the front panel from the analyzer, lift and support the
front of the analyzer chassis.
3. Slide the front panel over the test port connectors.
4. Disconnect the front panel interface ribbon cable (item ③). The front panel
is now free from the analyzer.
14
Installation Note N5242-90010
Description of the Upgrade
Installation Procedure for the Upgrade
Figure 3
Front Panel Assembly Removal
Step 5.Remove the Existing Test Set Cables
Be careful not to damage the center pins of the semirigid cables. Some
flexing of the cables may be necessary but do not over-bend them.
Leave the gray flexible cables, the wire harnesses, and the ribbon cables
connected where possible. Any that are removed should be labeled for
reconnection later.
Refer to Figure 4 for this step of the procedure. Although only Option 419 is
shown in the illustration, Option 423 is similar in appearance.
1. Place the analyzer bottom-side up on a flat surface.
2. Remove the following cables in the order listed. Discard these cables; they
will not be reused.
For analyzers with serial numbers prefixed MY/SG/US5310 and below:
— W26
Installation Note N5242-90010
A32 port 2 coupler to front-panel Port 2 CPLR ARM
15
-
Description of the Upgrade
Installation Procedure for the Upgrade
For all analyzer serial numbers:
— W40
Front-panel REF 2 RCVR R2 IN to A23 mixer brick (R2)
— W94
A45 port 2 receiver attenuator to A23 mixer brick (B)
— W88
A42 port 1 receiver attenuator to A23 mixer brick (A)
— W37
A33 reference mixer switch to A23 mixer brick (R1)
— W71
A25 port 1 bridge to A34 port 1 source attenuator
— W25
A28 port 2 bridge to front-panel REF 2 SOURCE OUT
— W83
A28 port 2 bridge to A37 port 2 source attenuator
— W72
A34 port 1 source attenuator to front-panel Port 1 SOURCE OUT
— W14
A29 port 1 coupler to front-panel Port 1 CPLR ARM
If bias tee option is installed (NOT N5241/42A-H85):
For analyzers with serial numbers prefixed MY/SG/US5310 and below:
— W86
A41 port 2 bias tee to A32 port 2 coupler
— W74
A38 port 1 bias tee to A29 port 1 coupler
— W85
A41 port 2 bias tee to front panel port 2 CPLR THRU
For all analyzer serial numbers:
— W73
Front-panel Port 1 CPLR THRU to A38 port 1 bias tee
If bias tee option is not installed (N5241/42A-H85):
For analyzers with serial numbers prefixed MY/SG/US5310 and below:
16
— W12
Front panel port 1 CPLR THRU to A29 port 1 coupler
— W24
Front panel port 2 CPLR THRU to A32 port 2 coupler
Installation Note N5242-90010
Description of the Upgrade
Installation Procedure for the Upgrade
For all analyzer serial numbers:
— W87
Front-panel Port 1 RCVR A IN to A42 port 2 receiver attenuator
— W84
A37 port 2 source attenuator to front-panel Port 2 SOURCE OUT
— W93
Front-panel Port 2 RCVR B IN to A45 port 2 receiver attenuator
3. Disconnect the ribbon cable from the port 1 and port 2 step attenuator
connectors. It is not necessary to disconnect these ribbon cables from the
A19 test set motherboard; connector designations are given here for
reference only.
a. Item ① connects between the A42 port 1 receiver attenuator and A19
test set motherboard connector J205.
b. Item ➁ connects between the A34 port 1 source attenuator and A19
test set motherboard connector J201.
c. Item ③ connects between the A45 port 2 receiver attenuator and A19
test set motherboard connector J208.
d. Item ④ connects between the A37 port 2 source attenuator and A19
test set motherboard connector J204.
4. Remove the termination (item ⑤) from the A24 mixer brick. This
termination will not be reinstalled.
Installation Note N5242-90010
17
-
Description of the Upgrade
Installation Procedure for the Upgrade
Figure 4
18
Existing Test Set Cables Removal
Installation Note N5242-90010
Description of the Upgrade
Installation Procedure for the Upgrade
Step 6.Remove the Existing Attenuators, Brackets, and Bias Tees (If
Present)
Refer to Figure 5 for this step of the procedure. Although only Option 419 is
shown in the illustration, Option 423 is similar in appearance.
1. Place the analyzer bottom-side up on a flat surface.
If bias tee option is installed (NOT N5241/42A-H85):
For analyzers with serial numbers prefixed MY/SG/US5321 and above:
2. Remove two screws (item ①) from the A38 port 1 bias tee and the A41
port 2 bias tee and lift them out of the PNA, and set them aside for
reinstallation later.
For analyzers with serial numbers prefixed MY/SG/US5310 and below:
3. Remove two screws (item ①) from the A38 port 1 bias tee and the A41
port 2 bias tee and lift them out of the PNA. These old bias tees have 2.4
mm connectors and will be replaced later with bias tees having 3.5 mm
connectors.
For all analyzers:
4. Remove three screws (item ➁) from each attenuator bracket.
5. Remove the attenuator brackets, with the attenuators still attached, from
the analyzer.
6. If it is necessary to remove any other cables from the analyzer to facilitate
removal of the attenuators and bias tees brackets, be sure to note their
locations and set them aside for reinstallation later.
Installation Note N5242-90010
19
-
Description of the Upgrade
Installation Procedure for the Upgrade
Figure 5
Existing Attenuators, Brackets, and Bias Tees Removal
Step 7.Assemble the Old Attenuators and the New A52 and A53
Bypass Switches onto the New Brackets
Refer to Figure 6 for this step of the procedure. New parts are listed in Table 1
on page 8.
1. Position each new bypass switch on a new switch bracket as shown. Make
sure that the switches are oriented as shown; they are not oriented the
same for both ports. The A52 switch, with the longest cable, is for port 1.
2. Secure each switch to its bracket using two screws (item ①) for each.
3. Remove the source and receiver attenuators from the old bracket and,
using the same screws (item ➁), attach them to the new brackets. Make
sure that they are oriented as shown; they are not oriented the same for
both ports.
4. Secure the bias tee brackets to the attenuator brackets using two screws
(item ③) for each.
20
Installation Note N5242-90010
Description of the Upgrade
Installation Procedure for the Upgrade
Figure 6
Attenuators and Bypass Switches Assembly
Installation Note N5242-90010
21
-
Description of the Upgrade
Installation Procedure for the Upgrade
Step 8.Install the New Brackets, with the Old Attenuators and the
New A52 and A53 Bypass Switches, into the Analyzer
Refer to Figure 7 and Figure 8 for this step of the procedure. Although only
Option 419 is shown in Figure 7, Option 423 is similar in appearance. New
parts are listed in Table 1 on page 8.
1. Position the attenuator brackets, with the attenuators and bypass
switches attached, in the analyzer as shown. Make sure that the brackets
are placed as shown; the attenuators and bypass switches are not oriented
the same for both ports.
2. Secure the attenuator brackets to the analyzer test set deck using three
screws (item ➁) for each.
If bias tee option is installed (NOT N5241/42A-H85):
Installing these bias tees backwards will cause damage to the analyzer
source modules.
For analyzers with serial numbers prefixed MY/SG/US5321 and above:
3. Position the existing bias tees, configured as shown in Figure 8, on the new
bias tee brackets and secure each of them with the original two mounting
screws (item ①).
4. Reconnect existing cables W74, W85, W86, and new cable W125 to the
bias tees.
For analyzers with serial numbers prefixed MY/SG/US5310 and below:
5. Position the new bias tees, configured as shown in Figure 8, on the new
bias tee brackets and secure each of them with the original two mounting
screws (item ①).
6. Connect new cables W85 and W125 to the bias tees. New cables W74 and
W86 will be installed later.
For all analyzers:
7. Reinstall other cables that were removed to facilitate removal of the old
attenuator brackets.
22
Installation Note N5242-90010
Description of the Upgrade
Installation Procedure for the Upgrade
Figure 7
New Brackets Installation into the Analyzer
Figure 8
A23 and A24 Mixer Brick Assembly
Installation Note N5242-90010
23
-
Description of the Upgrade
Installation Procedure for the Upgrade
Step 9. Assemble the A29 and A32 Test Port Coupler Assemblies (For
Analyzers with Serial Numbers Prefixed MY/SG/US5310 and Below)
1. Remove the A29 test port 1 coupler and A32 test port 2 coupler from the
PNA. For instructions, click the Chapter 7 bookmark, “Removing and
Replacing the A29 - A32 Test Port Couplers” in the PDF Service Guide1.
2. Discard the test port couplers you just removed from the PNA. These old
couplers have 2.4 mm connectors and must be replaced with the new
couplers (3.5 mm connectors) included in the kit.
3. Follow the two instructions shown in Figure 9. New parts are listed in
Table 1 on page 8 of this document.
If bias tee option is installed (NOT N5241/42A-H85):
4. Connect new cables W14, W26, W74 and W86.
If bias tee option is not installed (N5241/42A-H85):
Option 419/H85:
5. Connect new cables W14, W24, W26, and W144.
Option 423/H85:
6. Connect new cables W14, W26, W144, and W153.
1. See “Downloading
24
the Online PNA Service Guide” on page 6.
Installation Note N5242-90010
Description of the Upgrade
Installation Procedure for the Upgrade
Figure 9
A29 and A32 Test Port Coupler Assembly
Installation Note N5242-90010
25
-
Description of the Upgrade
Installation Procedure for the Upgrade
Step 10. Install the New (3.5 mm) A29 and A32 Test Port Couplers
onto the Front Plate (For Analyzers with Serial Numbers Prefixed
MY/SG/US5310 and Below)
1. Install the two new test port couplers onto the front plate.
2. Reinstall the two mounting nuts, but hand tighten only. Do not torque.
They will be fully tightened later.
26
Installation Note N5242-90010
Description of the Upgrade
Installation Procedure for the Upgrade
Step 11.Assemble and Install the A54 Bridge and A55 Noise Down
converter
Refer to Figure 10 for this step of the procedure. New parts are listed in Table
1 on page 8.
1. Connect the A54 bridge to the A55 noise downconverter, using cable
adapter W131, as shown. Place these assemblies on a flat surface and
torque each W131 connector to 10 in-lbs while holding the A54 and A55
assemblies flat.
2. Connect the ribbon cable (item ①) to the A55 noise downconverter
connector as shown.
3. Place the combined A54/W131/A55 assembly into the analyzer as shown.
4. Secure the A55 noise downconverter to the side frame using three screws
(item ➁), as shown.
5. Secure the A54 bridge to the side frame using one screw (item ③) and one
hex nut with lock washer (item ④), as shown. Install the screw from the
outside of the side frame, through the top hole in the A54 bridge, and
place the nut/washer on the inside. Hold the hex nut with a wrench while
tightening the screw.
Figure 10
A54 Bridge and A55 Noise Down Converter Installation
Installation Note N5242-90010
27
-
Description of the Upgrade
Installation Procedure for the Upgrade
Step 12.Install the A7 Noise Receiver Board
Refer to Figure 11 for this part of this step of the procedure. Although only
Option 419 is shown in the illustration, Option 423 is similar in appearance.
New parts are listed in Table 1 on page 8.
1. Place the analyzer top-side up on a flat surface.
2. Insert the A7 noise receiver board in the analyzer as shown. Make sure it is
fully seated in the motherboard connector.
28
Installation Note N5242-90010
Description of the Upgrade
Installation Procedure for the Upgrade
Figure 11
Noise Receiver Board Installation
Installation Note N5242-90010
29
-
Description of the Upgrade
Installation Procedure for the Upgrade
Step 13.Install the Bulkhead Connectors in the Test Set Front Plate
(For Analyzers with Serial Numbers Prefixed MY/SG/US5310 and
Below)
If your analyzer’s serial number is MY/SG/US5321 and above, ignore this step.
Refer to Figure 12 for this procedure. Some bulkhead connectors may already
be installed on your analyzer’s front plate. New parts are listed in Table 1 on
page 8.
1. From the back side of the test set front plate, insert a bulkhead connector
into a hole in the plate.
2. Install 1x washer and 1x nut. Hand tighten nut and ensure bulkhead
connector hexagon nut, on the back side of test set front plate, is aligned
to the test set subpanel hexagon indent.
3. Repeat previous two steps for the remaining bulkhead connectors.
4. Torque nuts, on the front side of test set front plate, to 21 in-lbs.
5. Torque test ports 1 and 2 mounting nuts to 72 in-lbs.
Figure 12
30
Bulkhead Connectors Installation
Installation Note N5242-90010
Description of the Upgrade
Installation Procedure for the Upgrade
Step 14.Install the New Test Set Cables
Follow instructions carefully when making cable connections, especially
wire harness connections. Incorrect connections can destroy components,
resulting in additional customer costs.
Be careful not to damage the center pins of the semirigid cables. Some
flexing of the cables may be necessary but do not over-bend them.
Use a 5/16-in torque wrench set to 10 in-lbs on all cable connections
except the front and rear panel bulkhead connectors. On these, use a 9
mm nutsetter or open end torque wrench set to 21 in-lb.
Refer to Figure 13 and Figure 14 for this step of the procedure. Although only
Option 419 is shown in the illustrations, Option 423 is similar in appearance.
New parts are listed in Table 1 on page 8.
1. Connect the following wire harness and ribbon cables:
—
—
—
—
—
—
—
①—A42 port 1 receiver attenuator to A19 test set motherboard J205
➁—A34 port 1 source attenuator to A19 test set motherboard J201
③—A45 port 2 receiver attenuator to A19 test set motherboard J208
④—A37 port 2 source attenuator to A19 test set motherboard J204
⑤—A53 port 2 bridge to A55 noise downconverter J11
⑥—A52 port 1 bridge to A55 noise downconverter J10
⑦—A55 noise down converter J1 to A19 test set motherboard J548
2. Install the following semirigid cables in the order listed. Use a 5/16-in
torque wrench set to 10 in-lbs to tighten all cable connectors.
— W135 (N5242-20278) Front-panel Port 2 RCVR B IN to A45 port 2
receiver attenuator
— W128 (N5242-20303) A53 port 2 bypass switch to front-panel Port 2
SOURCE OUT
— W127 (N5242-20292) A37 port 2 source attenuator to A53 port 2
bypass switch
— W130 (N5242-20302) A53 port 2 bypass switch to A54 port 2 bridge
— W129 (N5242-20293) A53 port 2 bypass switch to A54 port 2 bridge
If bias tee option is not installed (N5241/42A-H85):
— W132 (N5242-20277) Front-panel Port 1 RCVR A IN to A42 port 1
receiver attenuator
— W124 (N5242-20295) Front-panel Port 1 CPLR THRU to A52 port 1
bypass switch
Installation Note N5242-90010
31
-
Description of the Upgrade
Installation Procedure for the Upgrade
— W123 (N5242-20297) A52 port 1 bypass switch to front-panel Port 1
SOURCE OUT
— W122 (N5242-20298) A34 port 1 source attenuator to A52 port 1
bypass switch
— W126 (N5242-20272) A28 port 2 bridge to A37 port 2 source
attenuator
— W137 (N5242-20279) A28 port 2 bridge to front-panel REF 2 SOURCE
OUT
— W121 (N5242-20273) A25 port 1 bridge to A34 port 1 source
attenuator
— W136 (N5242-20274) A33 reference mixer switch to A23 mixer brick
(R1)
— W133 (N5242-20275) A42 port 1 receiver attenuator to A23 mixer
brick (A)
— W134 (N5242-20276) A45 port 2 receiver attenuator to A23 mixer
brick (B)
— W138 (N5242-20280) Front-panel REF 2 RCVR R2 IN to A23 mixer
brick (R2)
— W140 (N5242-20294) A24 mixer brick to A55 noise down converter
3. Position the analyzer as shown in Figure 14 on page 34 (fans facing
upwards) and loosely install the following cables. Route each of the cables
through the opening in the test set deck to the top side of the analyzer.
The other ends will be connected in the next step.
— W143 (N5242-20300) Semirigid cable, A55 noise down converter to
A7 noise receiver board RF
— W141 (N5242-20299) Semirigid cable, A55 noise down converter to
A7 noise receiver board LO
— W142 (N5242-60041) Flexible cable, A55 noise down converter J4 to
A7 noise receiver board P2
32
Installation Note N5242-90010
Description of the Upgrade
Installation Procedure for the Upgrade
Figure 13
New Test Set Cable Installation, Part 1
Installation Note N5242-90010
33
-
Description of the Upgrade
Installation Procedure for the Upgrade
Refer to Figure 14 for this part of this step of the procedure. Although only
Option 419 is shown in the illustration, Option 423 is similar in appearance.
New parts are listed in Table 1 on page 8.
4. The analyzer should be positioned on its left side (fans facing upwards) as
shown.
5. Connect semirigid cables W141 and W143 as indicated. Torque
connectors to 10 in-lbs.
6. Connect flexible cable W142 as indicated.
7. Go back and torque the connectors on the other ends of W141 and W143
to 10 in-lbs.
Figure 14
34
New Test Set Cable Installation, Part 2
Installation Note N5242-90010
Description of the Upgrade
Installation Procedure for the Upgrade
Step 15.Replace the A16 Power Supply (If Necessary)
Refer to Figure 15 for this step of the procedure. The analyzer model shown in
the illustration is for reference only and may not be the same as your analyzer.
New parts are listed in Table 1 on page 8.
1. Check the part number of the power supply currently installed in your
analyzer. If the correct power supply is installed, you need not replace it.
a. The part number should be as shown in Table 1 on page 8.
b. If the part number IS as shown, proceed to “Step 16. Replace the lower
front panel overlay”.
c. If the part number IS NOT as show, continue with this step.
2. Remove the three flat head screws (item ①) from the power supply
bracket.
3. Remove the seven pan head screws (item ➁) from the power supply rear
panel.
4. Slide the A16 power supply assembly out the rear of the analyzer.
5. Slide the new A16 power supply assembly into position in the analyzer.
6. Reinstall the seven pan head screws (item ➁) in the power supply rear
panel.
7. Reinstall the three flat head screws (item ①) in the power supply bracket.
Installation Note N5242-90010
35
-
Description of the Upgrade
Installation Procedure for the Upgrade
Figure 15
36
Power Supply Replacement
Installation Note N5242-90010
Description of the Upgrade
Installation Procedure for the Upgrade
Step 16.Replace the lower front panel overlay
Refer to Figure 15 for this step of the procedure. New parts are listed in Table 1
on page 8.
1. From the back side of the front panel, use a blunt object in the cutouts in
the lower front dress panel to push on the old overlay (item ①) and
separate it from the front dress panel.
2. From the front side of the front panel, pull off the overlay completely and
discard it.
3. Remove any adhesive remaining on the front panel.
4. Remove the protective backing from the new front panel overlay (item ➁).
5. Starting from either side, loosely place the overlay in the recess on the
lower front panel, ensuring that it fits tightly against the edges of the
recess.
6. Once the overlay is in place, press it firmly onto the frame to secure it.
Installation Note N5242-90010
37
-
Description of the Upgrade
Installation Procedure for the Upgrade
Figure 16
Lower Front Panel Overlay Replacement
Step 17.Reinstall the Front Panel Assembly and Front Panel Jumpers
Before installing the front panel assembly onto the analyzer, lift and
support the front of the analyzer chassis.
Refer to Figure 17 for this step of the procedure. New parts are listed in Table 1
on page 8.
1. Make sure all of the hex nuts on the front-panel cable connectors have
been tightened using a 5/16-in torque wrench set to 21-in lbs.
2. Reconnect the ribbon cable (item ③) to the A1 front panel interface board.
38
Installation Note N5242-90010
Description of the Upgrade
Installation Procedure for the Upgrade
3. Slide the front panel over the front-panel connectors.
4. With a T-10 TORX driver, reinstall the 12 screws (item ➁) in the sides of
the frame.
5. Reinstall the semirigid jumpers (item ①) on the front panel, and tighten
each of the connectors to 10-in lbs.
Figure 17
Front Panel Assembly Reinstallation
Step 18.Position the Cables and Wires to Prevent Pinching
On the top side of the PNA, carefully position the grey flex cables so they can’t
be pinched between the covers and the rails.
On the bottom side of the PNA, carefully fold or push down the ribbon cables
and wires so they can’t be pinched between the hardware and the outer cover.
Ribbon cables and wires must never be positioned on top of hardware.
Step 19. Reinstall the Inner Cover
Refer to Figure 18 for this step of the procedure.
1. Position the inner cover on the analyzer.
2. With a T-10 TORX driver, install the 12 pan head screws (item ①).
Installation Note N5242-90010
39
-
Description of the Upgrade
Installation Procedure for the Upgrade
3. With a T-10 TORX driver, install the 9 flat head screws (item ➁).
Figure 18
Inner Cover Reinstallation
Step 20.Reinstall the Outer Cover
This procedure is best performed with the analyzer resting on its front
handles in the vertical position. Do not place the analyzer on its front panel
without the handles. This will damage the front panel assemblies
Refer to Figure 19 for this step of the procedure.
1. Slide the outer cover over the analyzer frame.
2. Install the four rear panel feet (item ⑤) by installing the center screws
(item ⑥) with a T-20 TORX driver.
3. Install the four bottom feet (item ④) onto the bottom of the outer cover
then install the foot locks (item ③).
4. Install the strap handles (item ①) by tightening the screws (item ➁) on
both ends of each strap handle with a T-20 TORX driver.
40
Installation Note N5242-90010
Description of the Upgrade
Installation Procedure for the Upgrade
Figure 19
Outer Cover Reinstallation
Step 21.Enable Option 029
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must be running.
— A keyboard must be connected to the network analyzer.
Option Enable Proced ure
1. To start the option enable utility, press UTILITY System , then Service ,
then Option Enable . An option enable dialog box will appear.
2. Click the arrow in the Select Desired Option box. A list of available options
will appear.
3. In the Select Desired Option list, click 029 - Full Noise Measurements.
4. Using the keyboard, enter the license key in the box provided. The license
key is printed on the license message you received from Keysight. Enter
this key exactly as it is printed on the message.
5. Click Enable.
Installation Note N5242-90010
41
-
Description of the Upgrade
Installation Procedure for the Upgrade
6. Click Yes in answer to the displayed question in the Restart Analyzer? box.
7. When the installation is complete, click Exit.
Option Verification Procedure
Once the analyzer has restarted and the Network Analyzer program is again
running:
1. On the analyzer’s Help menu, click About Network Analyzer.
2. Verify that “029” is listed after “Options:” in the display. Click OK.
If Option 029 has not been enabled, perform the “Procedure
Requirements” on page 41 again. If the option is still not enabled, contact
Keysight Technologies. Refer to “Getting Assistance from Keysight” on
page 4.
Step 22. Verify Option 028 is Not Installed
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must be running.
— A keyboard and mouse must be connected to the network analyzer.
Verify Option 028 is Not Installed
— On the toolbar at the top of the PNA display screen, click Help.
— On the drop-down list, click About Network Analyzer.
— In the About Network Analyzer window, under Options, verify that “028” is
not listed.
— If it is listed, continue with the instructions in this step to remove
Option 028.
— If it is not listed, skip the remainder of this step and go to the next
one.
Option 028 Removal Procedure
1. To start the option enable utility, press UTILITY System , then Service ,
then Option Enable . An option enable dialog box will appear.
2. Click the arrow in the Select Desired Option box. A list of available options
will appear.
3. In the Select Desired Option list, click 028.
4. Click Remove.
5. Click OK to confirm that you want to remove the license for the selected
option.
42
Installation Note N5242-90010
Description of the Upgrade
Installation Procedure for the Upgrade
6. Click Yes in answer to the displayed question in the Restart Analyzer? box.
Step 23.Perform Post-Upgrade Adjustments and Calibration
Ad justments
The following adjustments must be made due to the hardware changes of the
analyzer.
— Default EE adjustment - select the LO Drive-NF adjustment and either
adjust or initialize the values.
— Source adjustment
— Receiver adjustment
— Noise figure adjustment
These adjustments are described in the PNA Service Guide and in the PNA
on-line HELP. A list of equipment required to perform these adjustments is
also found in the service guide.
To view this service guide information, click the Chapter 3 bookmark “Tests
and Adjustments” in the PDF Service Guide1.
After the specified adjustments have been performed, the analyzer should
operate and phase lock over its entire frequency range.
Operator’s Check
Perform the Operator’s Check to check the basic functionality of the analyzer.
For instructions, click the Chapter 3 bookmark “Tests and Adjustments” in the
PDF Service Guide1.
If you experience difficulty with the basic functioning of the analyzer, contact
Keysight. Refer to “Contacting Keysight” on page 4.
Calibration
Although the analyzer functions, its performance relative to its specifications
has not been verified. It is recommended that a full instrument calibration be
performed using the analyzer’s internal performance test software. To view
information on the performance test software, click the Chapter 3 bookmark
“Tests and Adjustments” in the PDF Service Guide1.
Step 24.Prepare the PNA for the User
1. If necessary, reinstall front jumper cables.
2. Install the cable guards, pushing them over the front jumper cables until
the cushioning material touches the front panel of the PNA.
3. Install the dust caps on the test ports.
1. See “Downloading
Installation Note N5242-90010
the Online PNA Service Guide” on page 6.
43
-
Description of the Upgrade
Installation Procedure for the Upgrade
4. Clean the analyzer, as needed, using a damp cloth.
44
Installation Note N5242-90010
This information is subject to change without notice.