GE
Sensing
Features
• Sources and reads mA, mV, V, ohms, frequency and
pulses
• Convenient one-handed operation
• Simulates and reads 8 RTDs and 12 thermocouples
• Plug/play connector for IDOS Universal Pressure
Modules
• 32 pressure modules from 10 in H20 to 10,000 psi
(25 mbar to 700 bar)
Applications
• Simultaneous dual reading capability
• Test and maintenance
• Automatic switch test and pressure leak test
• Transmitter calibration
• 1000 point data storage with real time clock
• Loop set-up and diagnostics
• 24 V loop power supply
• Switch, trip and alarm verification
• HART® loop resistor
The DPI 800 Series is a complete range of advanced, robust
and simple to use hand-held instruments. Highly cost
effective, these tools are ideal for test/calibration of many
popular process parameters. Advanced features and
technical innovations address more applications in less time
and deliver results you can rely on.
• Large backlit display
• Robust and weatherproof
• Compact, simple to use, easy to carry
DPI 880
Druck Multi-function
Calibrator
DPI 880 is a Druck product.
Druck has joined other
GE high-technology sensing
businesses under a new
name—GE Sensing.
g
IDOS
GE
Sensing
The DPI 880 Multi-function Calibrator is an ultra compact
and simple to use tool for testing, configuring and
calibrating virtually all process parameters.
It measures, sources and simulates mA, mV, V, RTDs,
thermocouples, ohms, frequency, pulses and pressure,
captures switch values and provides 24 V loop power.
C/J error due to 54°F (30°C) transient change
4.5°F (2.5°C)
DPI 880
Leading competitor
Error
DPI 880
Specifications
Unique thermocouple cold junction compensation
Virtually eliminates errors caused by changing
environmental conditions on test instruments used
outside.
32°F (0°C)
0
Simultaneous dual readings
Reads both input and output parameters simplifying
calibrations and system diagnostics. Calibration values
are captured on one screen and adjustments are seen in
real time, for example, when making zero and span
adjustments.
mA
mA (24V)
V
Switch test
IDOS Pressure
Leak test
mA
Measure or source
V mV Hz RTD TC
IDOS
V Pressure
Programmable step and ramp outputs
Quickly step through calibrations with the %Step output
or make zero and full scale adjustments using Span
Check. The Ramp output is ideal for delicate analogue
meters, rate of change indicators and checking valve
travel and slew rates.
Adjustable nudge output
Provides small incremental output changes for accurately
setting valve positioners, switches, trips, and alarms.
Automatic switch test
Captures open/closed trip values providing a fast and
highly accurate "safety system" check.
HART resistor
Can be switched into the loop when required for a HART
digital communicator avoiding the inconvenience of
carrying a 250 Ω resistor.
24V loop power supply
Energises transmitters and control loops.
Temperature
Measures or simulates RTD or thermocouple sensors and
is the ideal tool for checking probes, transmitters,
process loops, indicators and controllers. Use with a
temperature probe to provide a versatile thermometer.
5
10
15
tmins
Automatic detection of 2, 3 and 4 wire RTDs
Quickly detect damaged sensors and faulty wiring that
could otherwise be missed or cause system inaccuracy.
Thermometer with wide probe compatibility
Compatibility with 8 RTDs and 12 thermocouples allows
probe selection for any thermometer application, e.g.
general purpose, high temperature, hygienic and
aggressive media.
Frequency
Measures or sources Hz, kHz, CPM and CPH providing a
highly accurate calibration standard and versatile test
tool for process technicians and electronic engineers.
Dedicated features facilitate test and maintenance of
electronic circuits, frequency/flow meters, batch
counters, tachometers, and motion pickups.
Automatic trigger
Detects the best value regardless of waveform or
amplitude.
Pressure
Intelligent Digital Output Sensor (IDOS) Pressure Modules
are available from 10 in H2O to 10,000 psi (25 mbar to
700 bar). IDOS modules are Plug & Play requiring no
instrument calibration or set-up to provide a fully
featured pressure calibrator.
Standard and Premier accuracy
Standard accuracy from 0.05% FS includes operation
over 32°F to 122°F (0°C to 50°C), one year stability and
calibration uncertainty. The Premier range provides
laboratory grade precision to 0.01% FS.
Total flexibility
IDOS modules can be used with any compatible
instrument carrying the IDOS logo; for example,
the DPI 800 series and DPI 150.
GE
Sensing
Electrical specification
Measure accuracy
0 to 24.000 mA
0 to 55.000 mA
Type
Standard
*Accuracy
Range
Source accuracy
K
IEC 584
1.1°F (0.6°C)
-454°F to 2502°F (-270°C to 1372°C)
0.02% rdg + 2 counts
J
IEC 584
0.9°F (0.5°C)
-346°F to 2192°F (-210°C to 1200°C)
T
IEC 584
0.6°F (0.3°C)
-454°F to 752°F (-270°C to 400°C)
0.02% rdg + 3 counts
0 to 120.00 mV
0.02% rdg + 2 counts
0 to 12.000 V
0.02% rdg + 2 counts
0 to 30.000 V
0.03% rdg + 2 counts
B
IEC 584
1.8°F (1.0°C)
32°F to 3308°F (0°C to 1820°C)
R
IEC 584
1.8°F (1.0°C)
-58°F to 3214°F (-50°C to 1768°C)
S
IEC 584
2.5°F (1.4°C)
-58°F to 3214°F (-50°C to 1768°C)
E
IEC 584
0.7°F (0.4°C)
-454°F to 1832°F (-270°C to 1000°C)
0 TO 4000.0 Ω
0.1 to 1.3 Ω
N
IEC 584
1.1°F (0.6°C)
-454°F to 2372°F (-270°C to 1300°C)
Switch detection
Open and closed, 2 mA current
L
DIN 43710
0.6°F (0.3°C)
-328°F to 1652°F (-200°C to 900°C)
Loop power
24 V +/-10% (35 mA maximum)
U
DIN 43710
0.6°F (0.3°C)
-328°F to 1112°F (-200°C to 600°C)
HART mA loop resistor
250 Ω (menu selection)
C
1.8°F (1.0°C)
32°F to 4199°F (0°C to 2315°C)
D
1.8°F (1.0°C)
32°F to 4514°F (0°C to 2490°C)
mV
0.2%rdg + 0.01%FS
-10 to 75 mV
Frequency specification
Measure accuracy
Source accuracy
0 to 999.999 Hz
0.003% rdg + 2 counts
0.003% rdg + 0.0023 Hz
0 to 50.0000 kHz
0.003% rdg + 2 counts
0.003% rdg + 0.0336 Hz
IDOS Universal Pressure Modules
0 to 999999 cpm
0.003% rdg + 2 counts
0.003% rdg + 0.138 cpm
Pressure Range
0 to 59999 cpm
0 to 999999 cph
*Mid range figure quoted
Cold Junction Error 0.4°F (0.2°C) maximum for 86°F (30°C) change in ambient temperature.
G
A
0.003% rdg + 2 counts
0 to 99999 cph
Media
_
+
*Accuracy %FS
S
P
±10 in H2O (25 mbar)
0.1
0.03
0.075
0.03
0.1
N/A
0.05
0.01
0.075
N/A
0.05
0.01
0.075
N/A
0.003% rdg + 0.5 cph
Output waveform
Square wave (zero crossing)
±1, 3, 5, or 10 psi
(70, 200, 350, or 700 m bar)
Voltage input
30 V maximum
5 psi (350 mbar)
Trigger level
0 to 12 V resolution 0.1V
Output amplitude
0 to 12 VDC +/- 1% (10 mA maximum)
0 to 12 VAC pk-pk +/- 5% (10 mA maximum)
-15 to 15 or 30 psi
(-1 to 1 or 2 bar)
Temperature specification
Measure and
Simulate
Standard
*Accuracy
Range
Pt 50 (385)
IEC 751
0.9°F (0.5°C)
-328°F to 1562°F
(-200°C to 850°C)
Pt 100 (385)
IEC 751
0.45°F (0.25°C)
-328°F to 1562°F
(-200°C to 850°C)
Pt 200 (385)
IEC 751
1.08°F (0.6°C)
-328°F to 1562°F
(-200°C to 850°C)
Pt 500 (385)
IEC 751
0.72°F (0.4°C)
-328°F to 1562°F
(-200°C to 850°C)
Pt 1000 (385)
IEC 751
0.36°F (0.2°C)
-328°F to 752°F
(-200°C to 400°C)
D 100 (392)
JIS 1604-1989 0.45°F (0.25°C)
-328°F to 1202°F
(-200°C to 650°C)
Ni 100
DIN 43760
0.36°F (0.2°C)
-76°F to 482°F
(-60°C to 250°C)
Ni 120
MINCO 7-120
0.36°F (0.2°C)
-112°F to 500°F
(-80°C to 260°C)
0 to 4000
0.1 to 1.3 Ω
Ohms
G/D
30 psi (2 bar)
-15 to 50, 100, 150 or
300 psi (-1 to 3.5, 7 10
or 20 bar)
100, 300 psi (7, 20 bar)
* Mid range figure quoted
Excitation: 0.2 to 0.5 mA measure, 0.05 to 3 mA simulate
Pulse excitation currents minimum duration 10 ms
500, 1000, 1500, 2000
or 3000 psi
(35, 70, 100, 135, 200 bar)
0.05
0.01
5000 or 10,000 psi
(350 or 700 bar)
Sealed gauge
0.05
N/A
G = gauge, A = absolute, G/D = gauge/differential; calibrated referenced to
atmosphere maximum line pressure 30 psi (2 bar).. Stainless steel,
compatibility Non-corrosive gas/fluid and Non-corrosive gas. (N/A = not
available). Accuracy assumes regular zero correction.
*IDOS UPM-S Standard Accuracy
Total accuracy over 32°F to 122°F (0°C to 50°C) and one
year stability.
*IDOS UPM-P Premier Accuracy
Precision over 65°F to 82°F (18°C to 28°C).
Option A) Negative calibration for Premier ranges
Pressure Connections
G 1/8 female or 1/8 NPT female
Please refer to IDOS UPM data sheet for full specification
GE
Sensing
General specifications
Electrical connection
4mm sockets and mini-jack thermocouple socket
Calibrated temperature
50°F to 86°F (10°C to 30°C) unless otherwise stated
Operating Temperature
14°F to 122°F (-10°C to 50°C) unless otherwise stated
Temperature coefficient 14°F to 50°F, 86°F to 122°F
0.0017%FS/°F (-10°C to 10°C, 30°C to 50°C 0.003%FS/°C)
For ohms 14°F to 50°F, 86°F to 122°F 0.0028%FS/°F
(-10°C to 10°C, 30°C to 50°C 0.005%FS/°C)
IO800E
Data logging upgrade and RS232 lead
Log data periodically (1 second to 23 hours 59 minutes
59 seconds) or manually by key press. Review data
onscreen or upload to a PC via the RS232 interface.
No software purchase is necessary as standard
Microsoft® applications provide data transfer
(HyperTerminal) and analysis (Excel). Alternatively, print
directly to a compatible serial printer. Real time clock with
date. Memory: 1000 single or 750 dual reading screens
with date and time. Header tags: 6 user characters to
identify groups of readings. RS232: 19.2 k baud, 8 data bits,
1 stop bit, no parity, Xon/Xoff. Data output: comma
separated ASCII.
Storage Temperature
-4°F to 158°F (-20°C to 70°C)
Ordering Information
Humidity
0% to 90% non-condensing, Def Stan 66-31, 8.6 Cat III
For DPI 880
Please state the model number DPI 880 and accessories
as separate items.
Shock and Vibration
BS EN61010:2001, Def Stan 66-31, 8.4 Cat III
EMC
BS EN61326-1:1998 + A2:2001
Safety
Electrical BS EN61010:2001. Pressure Equipment
Directive (PED), Class SEP. CE marked
Display
Graphic LCD with backlight.
Each unit is supplied with batteries, calibration certificate,
user guide and a set of electrical test leads.
For IDOS UPM
Please state the model number IDOS UPM S for the
standard accuracy version or IDOS UPM P for the premier
accuracy version followed by the range, G/D, G or A and G
1/8 female or 1/8 NPT female.
Each unit is supplied with calibration certificate and user guide.
Supporting Services (order as separate items)
Size (l x w x h) and Weight
7.1 in x 3.3 in x 2 in (180 mm x 85 mm x 50 mm),
15 oz (425 g)
Related Products
Batteries
3 AA alkaline, >50 hours measure, >10 hours 24V source
GE is a world leader in the design and manufacture of
pressure, temperature and electrical field calibrators,
laboratory/workshop calibration equipment and pressure
sensors.
Accessories
IO800A
Soft fabric carrying case with accessory pocket
IO800B
Belt clip, wrist strap/hanging loop and bench stand
IO800C
NiMh batteries with charger, batteries charged externally
g
Supporting Services
Our highly trained staff can support you, no matter where
you are in the world. We can provide training, nationally
accredited calibration - both initially and at periodic
intervals - extended warranty terms, maintenance and
even rental of portable or laboratory calibrators. Further
details can be found in
www.gesensing.com/productservices/service.htm
©2008 GE. All rights reserved.
920-139B
All specifications are subject to change for product improvement without notice.
®
GE is a registered trademark of General Electric Co. Other company or product
names mentioned in this document may be trademarks or registered trademarks
of their respective companies, which are not affiliated with GE.
www.gesensing.com
Download PDF