Mark Of Fitness MF 46 User manual
Mark Of Fitness MF 46 is an easy-to-use home blood pressure monitor that helps you conveniently track and maintain your heart health. Equipped with oscillometric technology, it accurately measures systolic and diastolic blood pressure, and heart rate within an upper arm circumference range of 8.7 to 12.6 inches.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
MARK OF FITNESS FIVE YEAR WARRANTY
-U.S. Customers Only-
Mark of Fitness, Inc., warrants this product to be free of defects in material and workmanship for a period of five years, except cuff which has one year warranty.
Should any defects be discovered within five years of purchase, Mark of Fitness, Inc. will repair the defective product or parts at no charge for labor or materials provided:
1.
Said unit is returned postage paid to the factory for servicing, marked Attn: Service Department.
2.
3.
Proof of date of purchase is included with the product.
A letter accompanies the product and indicates the complaint or defect to be remedied.
4.
A check or money order in the amount of $7.95 is included to cover the shipping and handling costs.
In the event the defect may be remedied without returning the unit, contact the firm by phone or mail first, and every effort will be made to supply the replacement parts quickly and at no charge.
This warranty does not apply to any unit that has been tampered with, abused, accidentally damaged or to damage incurred through improper packing and mishandling in transit to or from the factory.
This warranty gives you specific legal rights, and you may have other rights which vary from state to state.
Model MF-46
CONTENTS Page
BASIC PRODUCT INFORMATION . . . . . . . . . . . 9
PRECAUTIONS FOR USE . . . . . . . . . . . . . . . . 10
TIPS ON TAKING YOUR BLOOD PRESSURE . . 10
INSTALLING BATTERIES AND SETTING THE
CLOCK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
CORRECT MEASURING POSTURE . . . . . . . . . 12
Page
WRAPPING THE CUFF . . . . . . . . . . . . . . . . . . 12
MEASUREMENT PROCEDURES . . . . . . . . . . . 13
MEMORY FUNCTIONS . . . . . . . . . . . . . . . . . . 13
DELETING STORED DATA . . . . . . . . . . . . . . . . 14
CARE AND MAINTENANCE . . . . . . . . . . . . . . 14
ERROR DISPLAYS AND TROUBLESHOOTING . . 15
INSTRUCTIONS
E
NGLISH
INSTRUCTIONS
E
NGLISH
Mark of Fitness, Inc.
621 Shrewsbury Avenue
Shrewsbury, New Jersey 07702
Phone (732) 842-7200 Fax (732) 842-0558
Customer Service (888) 343-1558 e-mail [email protected]
web site www.markoffitness.com
A11XXXX-1A
INSTRUCTIONS
E
NGLISH
MANUAL DE INSTRUÇOES
P
ORTUGUÊS
1
10
11
7
5
3
4
2
6
12
9
8
13 14
PARTS NAMES AND PRODUCT COMPONENTS
1
5
DISPLAY (SYS:systolic/DIA:diastolic/PUL:pulse)
AIR CONNECTOR 6 AC ADAPTER JACK
10 AIR HOSE 11 TUBE PLUG
2
7
ON/OFF BUTTON
CUFF HOLDER
12 CARRYING BAG
NOM DES PIÈCES ET COMPOSANTES DU PRODUIT
3
8
MEMORY BUTTON M1
CUFF
13 4 X AA BATTERY
4
9
MEMORY BUTTON M2
CUFF RING
14 INSTRUCTION MANUAL
NOMBRE DE LAS PIEZAS Y COMPONENTES DEL PRODUCTO
NOMES DAS PEÇAS E COMPONENTES DO PRODUTO
- 2 -
10
fig.1
INSTALLING BATTERIES AND SETTING THE CLOCK p.11
1
-1
1
-2
3
7
YEAR
ANNEE
AÑO
ANO
5
8
HOUR
HEURE
HORAS
HORA
- 3 -
4
2
1
2
6
11
DAY
JOUR
DIA
DIA
9
MONTH
MOIS
MES
MÊS
12
13
MINUTE
MINUTE
MINUTOS
MINUTO fig.2
CORRECT MEASURING POSTURE p.12
POSTURA CORRECTA DE MEDICIÓN p.X
14
POSTURE DE MESURE p.X
POSTURA CORRECTA PARA MEDIÇÃO p.X
fig.3
WRAPPING THE CUFF p.12
COLOCACIÓN DEL BRAZALETE p.X
1
2
ENROULEMENT DU BRASSARD p.X
COLOCAÇÃO DA BANDA p.X
1 inch
- 4 -
3
fig.4
MEASUREMENT PROCEDURES p.13
CÓMO EFECTUAR LA MEDICIÓN p.X
1 2
11
5
MÉTHODE DE MESURE p.X
PROCEDIMENTOS DE MEDIÇÃO p.X
3
SYS mmHg
DIA mmHg
PUL
/min
MEMORY BANK NUMBER
NUMÉRO INUTILISÉ DE LA
MÉMOIRE
NÚMERO DE BANCO DE MEMORIA
NÚMERO DO BANCO DE MEMÓRIA
- 5 sfhdfrejrgyk; dfhfgjtkdfghhg!
4
fig.5
MEMORY FUNCTIONS p.13
FUNCIONES DE MEMORIA p.X
1
2
latest data
Dernières données
FONCTIONNEMENT DE LA MÉMOIRE p.X
FUNÇÕES DA MEMÓRIA p.X
Average
SYS mmHg
DIA mmHg
PUL
/min moyenne
Media médio
MEMOR A NUMBER
NO. DES DONNÉES DE LA MÉMOIRE number of total stored data
HEURE
HORA
TEMPO
DATE
FECHA
DATA
3
4
2nd latest data
Avant-dernières données
2º medición más reciente
Penúltimos dados
Oldest data
Données les plus anciennes
Dato más antiguo
Dados mais antigos
30 30
- 6 fig.6
DELETING STORED DATA p.14
1
2
1
2
- 7 -
BASIC PRODUCT INFORMATION
This manual is intended to assist the user in the safe and efficient operation of the Model MF-46 Digital Blood Pressure Monitor. The product must be used in accordance with the procedures contained in this manual and must not be used for purposes other than those described herein. It is important to read and understand the entire manual. In particular, please read carefully and become familiar with the section entitled “TIPS ON
TAKING YOUR BLOOD PRESSURE”.
pressure against the cuff produced by the expansion and contraction of the arteries in the arm in response to each heart beat. The amplitude of each pressure wave is measured, converted to millimeters of mercury, and displayed on the LCD as a digital value. Memory circuit stores the 30 most recent measurement results for comparison.
The user is cautioned that the system may fail to yield specified measurement accuracy if operated or stored in temperature or humidity conditions outside the limits stated in the specifications section of this manual. Also, an inaccurate reading may result if this instrument is used on a child's arm.
Consult your physician if you wish to take a child's blood pressure.
Accuracy was evaluated during clinical studies in which results with the MF-
46 system were compared with simultaneous auscultatory measurements obtained with a standard sphygmomanometer. In these studies, the onset of the fourth (or fifth) Korotkoff sound was taken as the diastolic pressure for the purpose of determining overall efficacy. A copy of this study is available from the manufacturer on request. To insure that accuracy is maintained, compare readings obtained with the MF-46 system with values measured by a trained observer using a manual sphygmomanometer at least every
6 months.
Indications for Use
The MF-46 system is intended for the non-invasive measurement of systolic and diastolic blood pressure and determination of heart rate in adult patients, i.e., age 15 and above. The product is recommended for use by patients with labile blood pressure or known hypertension in a home care environment as an adjunct to medical management. The cuff will accommodate an upper arm circumference range of approximately 8.7 to 12.6 inches (22 to 32 centimeters). Pressure is measured over a range of 0 to 300 mmHg and heart rate over a range of 40 to 160 beats/minute.
Blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the cuff stethoscope auscultation method, within the limits prescribed by the American National
Standard, Electronic or Automated Sphygmomanometers.
The MF-46 system has the ability to function according to its specifications in the presence of common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation.
Method of Measurement
This product employs the oscillometric method for measurement of blood pressure and heart rate. The cuff is connected to the main unit and wrapped around the arm. Circuits within the monitor sense the small oscillations in
CAUTION!
Do not use cuffs other than the original cuff included with this product.
SPECIFICATIONS
Model
Operating Principle
Indicator
: MF-46
: Oscillometric method
: 13 digits LCD display
Pressure indicating Range : 0 to 300 mmHg (Cuff pressure)
Measuring Range
Accuracy
Inflation
: 50 to 250 mmHg (systolic)
40 to 180 mmHg (diastolic)
40 to 160 pulse/min. (pulse rate)
:
±
3 mmHg (Cuff pressure)
±
5% of reading (pulse rate)
: Automatic inflation
Deflation : Electric control valve
Exhaust
Power Supply
: Automatic exhaust valve
: 4 pcs. 1.5 volt "AA" (R6P/LR6) type batteries or AC adapter
Power Consumption : 4W (max.)
Memory : 30 measurement results x 2, computation of the average of stored Data
Operating environment : +50˚F to+104˚F, 15 to 90% relative humidity
Storage environment : -4˚F to+122˚F, 15 to 95% relative humidity
Cuff
Main Unit
: Coverage arm circumference ; 8.7 to 12.6 in.
[22 to 32 cm]
: Weight ; Approx. 13.4 oz. [370 g] (without batteries)
Size ; 6.4 x 7 x 3.2 in [163 x 180 x 83 mm]
(W x D x H) when CUFF HOLDER attached
Specifications are subject to change without notice due to improvements in performance.
- 9 -
PRECAUTIONS FOR USE
Do not use this instrument without consultation with your doctor if you are under dialysis therapy or on anticoagulants, antiplatelets or steroids. Use of this instrument under such conditions could cause internal bleeding.
To avoid any possibility of accidental strangulation, keep this unit away from children and
For specific information on your blood pressure, contact your physician. Never make any judgment on measurement results on your own.
Use of this device in areas near portable phones, microwave ovens or other devices with strong electromagnetic field may cause malfunctions.
TIPS ON TAKING YOUR BLOOD PRESSURE
Blood pressure is a measurement of the force exerted by the heart in pumping the blood through the arteries and the resistance by the veins in this flow.
Blood pressure changes easily as influenced by mental and physical factors and is never constant.
Causes for Changes in Blood Pressure
• Body movement
• Drinking Alcohol
• Conversation
• Smoking
• Mental Tension • Emotions
• Recent Urination or Bowel Movement
• Meals
• Temperature • Changes in the surroundings such as movement or noise, etc.
In general, blood pressure is highest during the working hours and gradually decreases during the afternoon and evening hours. It is low during sleep and slowly increases at a relatively fast rate after arising from bed.
Blood pressure measured at home tends to be lower than when measured in a hospital, clinic or doctor's office.
When blood pressure is measured in a hospital, it could be higher than when measured at home. This is because you are tense at the hospital and relaxed at home. It is important to know your stable normal blood pressure at home.
Let a qualified physician interpret your blood pressure readings.
Depending on your age, weight and general condition, blood pressure can be slightly different.
Consult with your doctor on determining what blood pressure is normal for you.
Blood pressure classification by WHO (1999)
Blood pressure changes shown below.
Upper arm blood pressure date.
200
180
160
140
120
100
80
60
40
20
12 13 14 15 16 17 18 19 20 21 22 23 24 1 2 3 4 5 6 7 8 9 10 11
Time
180~ (severe)
160~179 (moderate)
140~159 (mild)
130~139 High Normal
120~129 Normal
~120
SYS mmHg
DIA
Systolic
Diastolic
Hypertension
Optimal
80
80
84
85
89
90 100 110
99 109
Rest for approximately five minutes and take your blood pressure while relaxed in a quiet environment.
- 10 -
TIPS ON TAKING YOUR BLOOD PRESSURE
14
Take blood pressure measurement with the correct posture and do not move nor speak during measurement.
Avoid exercise, eating, drinking alcohol, smoking and other activities that affect your blood pressure right before a measurement.
Take your blood pressure every day at the same time.
The ambient temperature sould be approximately 68˚F (20˚C) when you take your blood pressure.
INSTALLING BATTERIES AND SETTING THE CLOCK
see fig.1
page 3
1
2
3
4
5
Turn the unit over and remove CUFF HOLDER from the unit.
7
CAUTION! Do not use rechargeable batteries.
Open the battery cover.
Never forcefully remove CUFF HOLDER or the battery cover.
Install four “AA” type batteries into the compartment as indicated. Make sure that the polarities correspond to the (+) and (-) marks.
Close the battery compartment cover.
Year flashes on the display.
Clock can be set only after the batteries are reinstalled. Adjust the clock so that measurement results are stored with correct date and time.
battery replacement indication
Replace all the batteries when the battery replacement indication appears on the display or nothing is displayed
Use alkaline batteries to increase battery life. Use of regular Carbon-
Zinc batteries will necessitate more frequent replacement. The enclosed batteries are for monitoring, and their life may be shorter than that of commercial batteries.
Since neither the device nor batteries are household waste, follow your local recycling rules and dispose of them at appropriate collection sites.
Operation with an AC adapter
6
7
8 to next step.
Month flashes.
Adjust month with M1 and M2 buttons.
Press ON/OFF BUTTON and day flashes.
the right side of the unit. Do not use any adapters other than the one specified below.
Specifications of the AC adapter to be selected are:
Output voltage : 6 VDC
±
5%
Input voltage range : Standard voltage
±
15%
Maximum electric load : 600mA or higher
- 11 -
INSTALLING BATTERIES AND SETTING THE CLOCK
9
10
Adjust day with M1 and M2 buttons.
Press ON/OFF BUTTON and hour flashes.
The clock operates in 24-hour mode.
Output plug polarity : Center Negative
Output plug shape : External diameter
: Internal diameter
: Length
5.5ø
±
0.1 mm
2.1ø
±
0.1 mm
9.5
±
0.3 mm
An approved adapter is the RadioShack PN 273-1761, with adaptaplug "M" PN 273-1716.
11 Adjust with M1 and M2 buttons.
12 Press ON/OFF BUTTON and minute flashes.
13 Adjust with M1 and M2 buttons.
14 Press ON/OFF BUTTON, the clock is set and the unit is turned off.
The clock is displayed while the unit is turned off.
Remove the batteries when operating the unit with AC adapter for extended period of time.
When batteries are removed or AC adapter is unplugged, the clock returns to the default value.
CORRECT MEASURING POSTURE
see fig.2
page 4
Sit at the table and let the table support your arm as you take the blood pressure measurement.
Make sure that the measurement location on the upper arm is at approximately the same height as the heart, and that the forearm is extended naturally on the table and does not move.
You may lie on your back and take the measurement. Look at the ceiling, stay calm, and do not move your neck or body during the measurement. Again, make sure that the measurement location on the arm is at approximately the same height as the heart.
WRAPPING THE CUFF
see fig.3
page 4
1
2
With the tube of the cuff positioned in the direction of the fingertips, place it on your left arm.
Wrap the cuff around the arm with the edge of the cuff approximately 1 inch (2.5 cm) above the elbow.
3 Press the surface of the cuff to make sure that it is attached securely.
When wrapping the cuff, wrap it loosely enough around the arm so that two fingers can be placed between the cuff and the arm. If the cuff is wrapped more tightly or loosely than this, inaccurate blood pressure readings may result.
If you are wearing a shirt that might restrict circulation in your upper arm or you roll your sleeve up over the upper arm, the blood flow will be restricted, preventing accurate measurement.
- 12 -
MEASUREMENT PROCEDURES
see fig.4
page 5
14
Take deep breaths and relax.
Do not move, chat or strain your arm or hand during measurement.
1
2
3
Beep sounds and the monitor automatically starts inflation.
The inflation stops as beep sounds and the displayed value (pressure) decreases.
The pulse is indicated by heart mark " " with synchronizing beep.
Long beep sounds as the measurement completes and maximum and minimum blood pressures and pulse rate are displayed.
Deflation mark appears and the unit automatically exhausts the air from the cuff.
Press either one of MEMORY BUTTONs,
4 the selected bank.
The selected memory bank number is shown on the display. The result is stored in the bank which is selected when the unit is turned off. When a measurement resulted in an error, it will not be stored.
Press ON/OFF BUTTON to turn off the power.
If you forget to turn off the unit, it will be automatically turned off after 3 minutes.
Stopping Measurement
Press ON/OFF BUTTON 2 and the monitor will stop inflation, discharge air rapidly, and then turn off.
Do not take repeated measurements. blood can be forced into tissues beyond the cuff, which would result in incorrect measurements. Let your arm rest for at least 5 minutes.
Automatic Repressurization
If the pressure applied is judged insufficient in the early phase of measurement or if movement of arm or hand occurs during measurement, the unit will inflate again to a level about 40 mmHg higher. Automatic Repressurization is repeated until a measurement is made. However, this does not indicate a problem.
To manually control pressurization, hold down ON/OFF
BUTTON immediately after the start of inflation and release it when the pressure reaches the level 50 mmHg higher than the expected maximum systolic pressure. If the pressure value has exceeded 180 mmHg, the inflation will stop when the key is released. Pressure can be increased to approximately 300 mmHg.
MEMORY FUNCTIONS
see fig.5
page 6
Measured result is stored in either M1 or M2. Each of two memory banks can hold up to 30 results and their average. When the number of measurements exceeds 30, the oldest data will be deleted to record new data. You may decide which bank to store your measurement results to avoid data mixture with someone else's, or may use two banks to save data measured in the morning and evening separately.
Recalling measured data
1
- 13 -
MEMORY FUNCTIONS
The avarage of the stored result in the selected bank is displayed.
The latest result stored is displayed when there is only a single result and the clock display remains when there are no results stored.
The indication at the top of the display alternately changes from memory data number to date and to time.
The memory data is displayed for approximately 30 seconds. Approximately 30 seconds after MEMORY BUTTON is released, the apparatus will turn off.
2 Press MEMORY BUTTON to move to next data.
The memory data number 1 is the latest among the stored data in the selected memory bank.
3 Every depression of the key switches among the memory data.
As MEMORY BUTTON is pressed, the memory data number increases; the bigger the number, the older the result.
4
DELETING STORED DATA
see fig.6
page 7
Deleting a single data
1 Show the data to be deleted following the steps in MEMORY FUNCTIONS.
2 Press and hold down MEMORY BUTTON stored.
The displayed data starts flashing.
Hold down the button until the display disappears.
Deleting all data in a bank
1 4 and show the average of the bank to be deleted.
2 Press and hold down selected MEMORY
BUTTON.
The displayed data starts flashing.
Hold down the button until the display disappears.
CARE AND MAINTENANCE
GASOLINE
PAINT
THINNER
Because the unit includes precision parts, care should be taken to avoid extreme temperature variations, humidity, shock, dust, and direct sunlight. Do not drop or strike the unit. Make sure not to expose the unit to moisture. This unit is not water resistant.
Use only a soft, dry cloth to clean the unit. Do not use gasoline, paint thinner, or other strong solvents on the unit. Since the cuff may absorb perspiration and other fluids, inspect it for stain and discoloration after each use. When cleaning the cuff, do not machine wash or scrub it. Use a synthetic detergent and gently
When storing the unit, do not place heavy objects on it and do not coil AIR HOSE too tightly. When the unit has been stored at a temperature below the freezing point, keep it for at least 1 hour in a warm place before using it. Remove the batteries if the instrument is to be stored for an extended period of time. Keep
- 14 -
CARE AND MAINTENANCE
14 the batteries out of reach of children.
We suggest that you have your blood pressure monitor checked every 2 years. This operation may only be performed by the manufacturer or by firms authorized by the manufacturer.
Do not inflate the cuff when it is not wrapped around your arm.
Do not disassemble or modify the unit.
ERROR DISPLAYS AND TROUBLESHOOTING
ERROR SYMOL / SYMPTOM
OVER-PRESSURIZATION
CAUSE / CHECK POINT
The cuff was inflated to the maximum pressure because of movement of body etc.
REMEDY
Do not move during measurement.
MEASUREMENT ERROR Measurement could not be made because of moving or talking during measurement.
Remain still and quiet during measurement.
INFLATION ERROR securely inserted.
DEFLATION ERROR Movement or talking was detected during measurement.
Turn off the power and make another measurement while being quiet and still.
IMPROPER OPERATION during battery replacement.
once and press it again to start a measurement.
No display when ON/OFF BUTTON
2 is pressed.
Have the batteries polarities been positioned incorrectly?
Have the batteries run down?
Are the battery terminals clean?
Reinsert the batteries in the correct position.
Replace them with new ones.
Clean the battery terminals with a dry cloth.
- 15 -
ERROR DISPLAYS AND TROUBLESHOOTING
ERROR SYMOL / SYMPTOM
No display when MEMORY BUTTON
CAUSE / CHECK POINT
Is there any measured result stored in the memory?
Are you pressing the right MEMORY BUTTON?
Have the batteries run down?
REMEDY
Nothing is displayed when there is no measured data stored.
Check the memory bank number.
Replace them with new ones.
Cannot make great number of measurements.
Are you using alkaline batteries?
Cannot complete the measurement.
Have the batteries run down?
Use alkaline batteries to increase battery life.
Replace all four batteries with new ones.
Blood pressure is different each time. The reading is extremely low
(or high).
Are you measuring with correct posture?
Blood pressure readings constantly vary with time of measurement and nervous condition.
Measure with the correct posture.
See TIPS ON TAKING YOUR BLOOD PRESSURE.
Pulse rate is too low (or too high).
Did you move during the measurement?
Did you make measurement right after exercise?
Remain still during the measurement.
Measure again after resting for more than 5 minutes.
The power is automatically turned off.
Have you left the instrument untouched after the measurement?
This does not indicate a problem. The instrument automatically turns off 3 minutes after a measurement or 30 seconds after memory display.
If you cannot get correct measurement with the methods above, do not tamper with the internal mechanism. Contact your dealer. If the unit malfunctions, please return it to the dealer or an authorized service representatives for service according to the warranty. Service data are provided to qualified dealers by the manufacturer.
- 16 - 17 -