Viewstation OR Operators Manual
Operating Instructions
Datascope
Viewstation
0070-01-0663-01_revB.indd 1
™
4/6/10 4:24:13 PM
Operating Instructions
Datascope
Viewstation
™
Masimo SET® is a U.S. registered trademark of Masimo Corp.
Nellcor® is U.S. registered trademark of Tyco International Ltd.
Panorama™ is a U.S. trademark of Mindray DS USA Inc.
Spectrum OR™ is a U.S. trademark of Mindray DS USA Inc.
Viewstation OR™ is a U.S. trademark of Mindray DS USA Inc.
Copyright © Mindray DS USA Inc., 2007. All rights reserved. Contents of this publication may not be reproduced in any
form without permission of Mindray DS USA Inc. Due to the continuous evolution of our products, all capabilities
documented in this manual may not be currently available.
Viewstation OR™ Operating Instructions
0070-10-0663-01
Table of Contents
Foreword ....................................................................................................................................................... v
Warnings, Precautions and Notes..................................................................................................................... v
Indications for Use .......................................................................................................................................... viii
Unpacking Information .................................................................................................................................... viii
Technical Support ........................................................................................................................................... viii
Symbols and Descriptions ................................................................................................................................ ix
System Overview ............................................................................................................. 1 - 1
The Viewstation OR......................................................................................................................................... 1 - 2
Basic Configuration......................................................................................................................................... 1 - 2
Supported Device ........................................................................................................................................... 1 - 4
Spectrum OR™ ........................................................................................................................................ 1 - 4
Viewstation OR Main Screen ............................................................................................. 2 - 1
The Main Screen ............................................................................................................................................ 2 - 1
Waveform Parameter Tiles........................................................................................................................ 2 - 2
Waveform Area .............................................................................................................................. 2 - 3
Waveform Digital Area .................................................................................................................... 2 - 4
Digital Parameter Tiles.............................................................................................................................. 2 - 5
Menu Bar ............................................................................................................................................... 2 - 7
System Status/Demographics Area ............................................................................................................ 2 - 8
Date and Time ........................................................................................................................................ 2 - 8
Display States ......................................................................................................................................... 2 - 9
Patient Setup Functions ..................................................................................................... 3 - 1
Display Tab.................................................................................................................................................... 3 - 2
Accessing the Display Tab ........................................................................................................................ 3 - 2
Components of the Display Tab ................................................................................................................. 3 - 2
Configuring the Waveform and Digital Parameter Areas .............................................................................. 3 - 4
Saving a Display Configuration................................................................................................................. 3 - 6
Loading a New Display Configuration ............................................................................................... 3 - 7
Removing a Display Configuration .................................................................................................... 3 - 7
Editing a Display Name ................................................................................................................... 3 - 7
Sidebar Buttons ....................................................................................................................................... 3 - 9
Troubleshooting....................................................................................................................................... 3 - 9
Wave Gain Tab (Patient) ................................................................................................................................. 3 - 10
Accessing the Wave Gain Tab (Patient)...................................................................................................... 3 - 10
Configuring Wave Gain Settings............................................................................................................... 3 - 10
Sidebar Buttons ....................................................................................................................................... 3 - 11
Troubleshooting....................................................................................................................................... 3 - 11
System Setup Functions..................................................................................................... 4 - 1
Parameter Color Tab ....................................................................................................................................... 4 - 2
Accessing the Parameter Color Tab ........................................................................................................... 4 - 2
Selecting a Color for Parameter Data ......................................................................................................... 4 - 2
Sidebar Button - Normal Screen ................................................................................................................ 4 - 3
Troubleshooting....................................................................................................................................... 4 - 3
Equipment Setup Tab....................................................................................................................................... 4 - 3
Accessing the Equipment Setup Tab ........................................................................................................... 4 - 4
Connecting Equipment ............................................................................................................................. 4 - 4
Sidebar Button - Normal Screen ................................................................................................................ 4 - 4
Troubleshooting....................................................................................................................................... 4 - 4
Wave Gain Tab (System)................................................................................................................................. 4 - 5
Accessing the Wave Gain Tab.................................................................................................................. 4 - 5
Changing the System Default Wave Gain Settings ....................................................................................... 4 - 5
Sidebar Button - Normal Screen ................................................................................................................ 4 - 5
Viewstation OR™ Operating Instructions
0070-10-0663-01
i
Table of Contents
Touch Screen Tab ........................................................................................................................................... 4 - 6
Accessing the Touch Screen Tab................................................................................................................ 4 - 6
Turning the Touch Screen On and Off ................................................................................................ 4 - 6
Recalibrating the Touch Screen ......................................................................................................... 4 - 7
Sidebar Button - Normal Screen ................................................................................................................ 4 - 8
Troubleshooting....................................................................................................................................... 4 - 8
User Maintenance ............................................................................................................. 5 - 1
Care and Cleaning of the Touch Screen ............................................................................................................ 5 - 2
Parameter Specifications................................................................................................... 6 - 1
Heart Rate (HR)............................................................................................................................................... 6 - 2
Respiration (Resp) ........................................................................................................................................... 6 - 2
NIBP ............................................................................................................................................................. 6 - 3
SpO2 ............................................................................................................................................................ 6 - 4
IBP ................................................................................................................................................................ 6 - 4
CO2.............................................................................................................................................................. 6 - 5
Temperature (T1, T2, and T Blood) .................................................................................................................... 6 - 5
Cardiac Output .............................................................................................................................................. 6 - 6
Cardiac Index ................................................................................................................................................ 6 - 6
Pulmonary Artery Wedge Pressure (PAWP) ........................................................................................................ 6 - 6
Invalid Data Display........................................................................................................................................ 6 - 7
Performance Specifications ............................................................................................... 7 - 1
ECG.............................................................................................................................................................. 7 - 2
ECG Derived Heart Rate Meter ................................................................................................................. 7 - 2
ECG Respiration...................................................................................................................................... 7 - 2
NIBP ............................................................................................................................................................. 7 - 3
Systolic Pressure Measurement .................................................................................................................. 7 - 3
Diastolic Pressure Measurement................................................................................................................. 7 - 3
Pulse Rate ............................................................................................................................................... 7 - 3
IBP ................................................................................................................................................................ 7 - 3
IBP Heart Rate Meter................................................................................................................................ 7 - 3
Temperature ................................................................................................................................................... 7 - 4
SpO2 ............................................................................................................................................................ 7 - 4
CO2.............................................................................................................................................................. 7 - 4
Viewstation OR Computer ................................................................................................................................ 7 - 5
Real Time Clock ...................................................................................................................................... 7 - 5
Power Supply.......................................................................................................................................... 7 - 5
Operating Temperature ............................................................................................................................ 7 - 5
Operating Humidity ................................................................................................................................. 7 - 5
Operating Altitude................................................................................................................................... 7 - 5
Storage Temperature................................................................................................................................ 7 - 6
Storage Humidity..................................................................................................................................... 7 - 6
Storage Altitude ...................................................................................................................................... 7 - 6
Safety Standards ..................................................................................................................................... 7 - 6
Electromagnetic Compatibility ................................................................................................................... 7 - 7
Physical Characteristics ............................................................................................................................ 7 - 10
I/O Control Devices........................................................................................................................................ 7 - 11
Displays......................................................................................................................................................... 7 - 11
Mindray DS Supplied ELO-Compatible Touch Screen Display........................................................................ 7 - 11
Customer Supplied Compatible Displays .................................................................................................... 7 - 11
Warranty....................................................................................................................................................... 7 - 12
Phone Numbers and How To Get Assistance...................................................................................................... 7 - 14
Mindray DS’s Responsibility ............................................................................................................................. 7 - 14
ii
0070-10-0663-01
Viewstation OR™ Operating Instructions
Table of Contents
Glossary........................................................................................................................... 8 - 1
Terms, Acronyms, and Abbreviations ................................................................................................................ 8 - 2
Viewstation OR™ Operating Instructions
0070-10-0663-01
iii
Table of Contents
This page intentionally left blank.
iv
0070-10-0663-01
Viewstation OR™ Operating Instructions
Foreword
Introduction
Foreword
The Viewstation OR Operating Instructions are intended to provide the information required
to operate the Viewstation OR. For additional information and technical assistance, please
contact a Mindray DS Patient Monitoring Representative.
A general understanding of the features and functions of the Viewstation OR and its
components (shown in FIGURE 1-1) is a prerequisite for the proper use of this equipment.
Therefore, do not operate this equipment before reading these instructions thoroughly,
including all appropriate warnings and cautions.
CAUTION:
US Federal law restricts the sale of this device to, by or on
the order of a physician.
NOTE:
Figures in this manual are provided for reference purposes
only. Screens will likely differ based on the monitoring
device configuration and selected parameter configurations
of the Viewstation OR.
Warnings, Precautions and Notes
Please read and adhere to all of the warnings and precautions listed throughout this manual.
A WARNING is provided to alert the user to potentially serious outcomes (death, injury or
serious adverse events) to the patient or the user.
A CAUTION is provided to alert the user that special care should be taken for the safe and
effective use of the device. They will include actions to be taken to avoid effects on patients
or users that will not be potentially life threatening or result in serious injury, but about which
the user should be aware.
A NOTE is provided when additional general information is available.
Viewstation OR™ Operating Instructions
0070-10-0663-01
v
Introduction
Warnings, Precautions and Notes
Warnings
WARNING: Only qualified and trained personnel or Mindray DS Service
personnel should attempt to service Mindray DS equipment.
Service is defined as any activity requiring the cover to be
removed for internal adjustments, parts replacements,
repairs or software upgrades of any kind to insure
compatibility.
WARNING: To insure compatibility with the operating system and
applications software, use only Mindray DS USA Inc.
supplied and/or approved components to repair any part of
the Viewstation OR. Use of unauthorized software, devices,
accessories, or cables other than those approved by
Mindray DS may render the application unsuitable for
medical patient monitoring. It may also result in increased
electromagnetic emissions or decreased Immunity of the
system.
WARNING: Route cables neatly. Ensure cables are not in the way of
patient or hospital personnel.
WARNING: All of the equipment on the Viewstation OR network must
utilize the hospital emergency power system. Failure to do
so will result in loss of monitoring during extended periods
of power failure.
WARNING: Do not put MPSO (Multiple Portable Socket Outlets, i.e.,
multiple outlet extension cords) used with the
Viewstation OR equipment on the floor. Connect only
approved accessories and components to the same MPSO as
the Viewstation OR equipment. Do not overload MPSOs. Use
only MPSOs that comply with the requirements of IEC
60601-1-1.
WARNING: PACEMAKER PATIENTS - Indication of the heart-rate may be
adversely affected by cardiac pacemaker pulses or by
cardiac arrhythmias. Keep pacemaker patients under close
surveillance. Refer to "ECG" on page 7-2 for disclosure of
the pacer rejection capability.
WARNING: PACEMAKER PATIENTS - Rate meters may continue to count
the pacemaker rate during occurrences of cardiac arrest or
some arrhythmias. Do not rely entirely upon rate meter
alarms. Keep pacemaker patients under close surveillance.
WARNING: When connecting any Spectrum OR monitor, verify its
location and associated device ID and bed number.
vi
0070-10-0663-01
Viewstation OR™ Operating Instructions
Warnings, Precautions and Notes
Introduction
Cautions
CAUTION:
For proper operation do not obstruct the fan air holes.
CAUTION:
For proper operation use only hospital grade power cords.
CAUTION:
For proper operation, do not use the Viewstation OR
equipment with a frayed or damaged power cord.
CAUTION:
For proper operation never place fluids on top of this
equipment. In case of accidental spillage, wipe clean
immediately and have the system serviced to ensure no
hazard exists.
CAUTION:
For proper monitoring of patient information, do not block
the operators' clear view of the Viewstation OR data
display(s).
CAUTION:
Changes or modifications not expressly approved by
Mindray DS USA Inc. could void the user's warranty.
CAUTION:
To avoid potential system issues, deactivate the touch screen
before cleaning.
CAUTION:
DO NOT spray glass cleaner directly on a display as it could
possibly leak inside a non-sealed unit and cause damage.
CAUTION:
Some pacemakers may contain a respiratory sensor that
may produce artifact on an ECG waveform.
CAUTION:
The Viewstation OR must be installed outside the patient’s
vicinity.
CAUTION:
The Viewstation OR CPU should not be used within 25 cm of
the point where leakage or discharge of a flammable
anesthetic mixture with oxygen or nitrous oxide may occur.
CAUTION:
Do not use the Viewstation OR if the case has been dropped
or damaged.
Viewstation OR™ Operating Instructions
0070-10-0663-01
vii
Introduction
Indications for Use
Indications for Use
The indications for use for the Viewstation OR include viewing real time patient clinical data.
The clinical data displayed by the Viewstation OR is obtained from a Spectrum OR™ patient
monitor and includes: ECG waveforms, Invasive and Non-Invasive Blood Pressures, Blood
Oxygenation (SpO2), Heart Rate, Respiration Rate, PVC/min, Temperature, CO2 inspired
and end tidal, CO/CI, CCO/CCI, SvO2 and PAWP for a patient.
The Viewstation OR is intended for use in a fixed, non-portable, permanent location as a
secondary independent remote display in the Operating Room department. The
Viewstation OR does not display or annunciate physiological or technical alarms originating
from the Spectrum OR monitor. The Viewstation OR is not intended to be directly connected
to the patient at any time or installed within a six foot radius of the patient.
Unpacking Information
The Viewstation OR must be installed by a certified Mindray DS Service Technician or
authorized distributor. Please contact your authorized Mindray DS Service Representative for
additional information.
Technical Support
For technical assistance please contact your area or regional Mindray DS Patient Monitoring
Representative.
Additional information relating to the servicing and repair of the Viewstation OR is provided
in the Viewstation OR Service Manual (P/N 0070-00-0665).
viii
0070-10-0663-01
Viewstation OR™ Operating Instructions
Symbols and Descriptions
Introduction
Symbols and Descriptions
SYMBOL
DESCRIPTION
DESCRIPTION
Attention, Consult Accompanying
Documents / Refer to Manual
Monitor
Alternating Current (AC)
Touch Screen
Direct Current (DC)
Earth Ground
Patient
Network
Keyboard
Viewstation OR™ Operating Instructions
SYMBOL
Panorama Patient Network
Ethernet
Mouse
Mark of compliance with U.S.
and Canadian safety standards
requirement.
Input/Output
Do not dispose of this product as
unsorted municipal waste.
0070-10-0663-01
ix
Introduction
Symbols and Descriptions
This page intentionally left blank.
x
0070-10-0663-01
Viewstation OR™ Operating Instructions
1.0
System Overview
This section provides a high level view of the authorized components for the Viewstation OR.
• The Viewstation OR
• Basic Configuration
• Supported Device
Viewstation OR™ Operating Instructions
0070-10-0663-01
1-1
The Viewstation OR
1.1
System Overview
The Viewstation OR
Viewstation OR enables real-time clinical data to be presented on an independent, remote,
secondary display for surgeons and perfusionists.
The Viewstation OR CPU receives data from a Spectrum OR™ and transmits it to the
independent display which can be configured to meet the needs of various clinicians. (See
section 1.2, Basic Configuration.)
While accessing the tabs that are associated with the menu bar buttons, the Viewstation OR
display automatically re-configures as a split-screen. The top half of the screen displays
patient data while the bottom half of the screen is converted to a work space for
simultaneously entering and modifying display configurations.
1.2
Basic Configuration
As depicted in FIGURE 1-1, the basic configuration of the Viewstation OR is:
• Spectrum OR Bedside Monitor
• Viewstation OR computer (CPU)
• Independent Display
• Touch Pad
INDEPENDENT
DISPLAY
TOUCH PAD
SPECTRUM OR
MONITOR
CPU
FIGURE 1-1 Basic Components of the Viewstation OR
This configuration can be varied to include two identically configured displays per
Viewstation OR computer and may also access the Spectrum OR monitor through an Ethernet
network switch (see FIGURE 1-2).
1-2
0070-10-0663-01
Viewstation OR™ Operating Instructions
System Overview
Basic Configuration
SPECTRUM OR
MONITOR
ELAN
ETHERNET
8 PORT
SWITCH
CPU
CPU
TO PANORAMA
OR GATEWAY
VIDEO
SPLITTER
CPU
VGA
TOUCH SCREEN
VGA
EXTERNAL
DISPLAY
TOUCH PAD
EXTERNAL
DISPLAY
EXTERNAL
DISPLAY
TOUCH PAD
FIGURE 1-2 Example Multiple Display Configuration of the Viewstation OR
Viewstation OR™ Operating Instructions
0070-10-0663-01
1-3
Supported Device
System Overview
1.3
Supported Device
1.3.1
Spectrum OR™
All Viewstation OR data is sourced from the Spectrum OR, a vital signs bedside monitor
designed specifically for the operating room environment. The following table lists the
Spectrum OR parameters that are supported by the Viewstation OR.
SUPPORTED PARAMETERS
HR
PVC/min
Resp
NIBP (S/D/M)
SpO2
IBP1 (S/D/M)
IBP2 (S/D/M)
IBP3 (S/D/M)
IBP4 (S/D/M)
I/E CO2
T1
T2
T Blood
CO
CI
CCO
CCI
SvO2
PAWP
NOTE:
1-4
The Viewstation OR does not display or annunciate
physiological or technical alarms. All physiological alarms
originate in the Spectrum OR.
0070-10-0663-01
Viewstation OR™ Operating Instructions
2.0
Viewstation OR Main Screen
This section provides an overview of the Viewstation OR main screen.
2.1
The Main Screen
The Main Screen for the Viewstation OR is used to display patient data and to navigate the
system.
The Main Screen includes the following elements:
•
Waveform Parameter Tiles
•
Digital Parameter Tiles
•
Menu Bar
•
System Status/Demographics Area
•
Date and Time
•
Display States
Viewstation OR™ Operating Instructions
0070-10-0663-01
2-1
The Main Screen
2.1.1
Viewstation OR Main Screen
Waveform Parameter Tiles
Waveform parameter tiles are the core elements of the Viewstation OR system (see
FIGURE 2-1). These tiles display patient data obtained from the Spectrum OR™. Three to
eight waveform parameter tiles may be displayed at once.
Each waveform parameter tile is divided into two sections:
•
Waveform Area
•
Waveform Digital Area
FIGURE 2-1 Example Waveform Parameter Tile
Configuring Waveform Parameter Tiles
1. Select the desired waveform parameter tile. The waveform configuration dialog box
opens (see the examples in FIGURE 2-2 and FIGURE 2-3). The configuration options
include Sensor Color, Wave Gain (all waveforms except SpO2), Change To and
Pacer Enhancement (ECG1 waveform only). The Pacer Enhancement factory default
setting is OFF.
FIGURE 2-2 Example ECG Waveform Configuration Dialog
FIGURE 2-3 Example IBP Waveform Configuration Dialog
2-2
NOTE:
The Change To choice control is based on the sensors utilized
by the Spectrum OR. When viewing the top waveform area,
the Change To choices are limited to ECG waveforms only.
NOTE:
The Change To choice control is hidden in ECG waveforms
when a 3-wire ECG lead set is utilized by the Spectrum OR.
0070-10-0663-01
Viewstation OR™ Operating Instructions
Viewstation OR Main Screen
The Main Screen
2. Select the Done or Cancel button.
• The Done button accepts the changes.
• The Cancel button rejects the changes.
2.1.1.1
NOTE:
Changes made in the Waveform Configuration Dialog box
apply to both the Waveform Area and the Waveform Digital
Area.
NOTE:
Changes made in the Waveform Configuration Dialog box
will be reflected in the Display Tab, and vice versa.
Waveform Area
The waveform area displays waveform data parameters. The following is a list of the
available waveform areas.
FIGURE 2-4 Example ECG Waveform Area
•
The ECG1 waveform is always located in the top waveform area. Based on the ECG
lead set that is in use, other ECG waveforms can be displayed in other waveform areas.
A scale bar is viewable and wave gain settings are displayed. Filters set at the
Spectrum OR™ can be viewed in this area, including Monitor bandwidth (“MON”),
Extended bandwidth (“EXT”) and ST bandwidth (“ST”).
•
The SpO2 waveform area displays an SpO2 waveform.
•
The IBP1 through IBP4 waveforms include a scale bar and wave gain settings. When
an IBP area has been labeled at the Spectrum OR, the label will also display in the
Viewstation OR. Wave gain settings include 0 – 20, 0 – 40, 0 – 80, 60 – 140,
0 – 160, 0 – 225, 0 – 320, and (-10) – 10 mmHg.
•
The Respiration waveform area includes a scale and wave gain settings. The wave
gain choices include 1, 2, 3, 4, and 5.
•
The CO2 waveform area displays the wave gain scale with the associated unit of
measure (mmHg, %, or kPa) as set at the Spectrum OR.
Waveforms are shown in a configured color and wave gain. If individual configurations are
not set, the system default settings are used. A waveform can be displayed in more than one
tile.
Viewstation OR™ Operating Instructions
0070-10-0663-01
2-3
The Main Screen
2.1.1.2
Viewstation OR Main Screen
Waveform Digital Area
Waveform digital areas display waveform numeric data. Below is a list of the available
waveform digital areas.
FIGURE 2-5 Example Resp Waveform Digital Area
•
The Heart Rate waveform digital area is always located adjacent to the ECG1
waveform. It displays the heart rate numeric value, ECG lead indicator, Beats Per Minute
(BPM) and Pacer Enhancement icon (if enabled). (The Pacer Enhancement icon is only
available in the Heart Rate tile.)
FIGURE 2-6 Pacer Icon
•
The ECG Lead waveform digital areas are located adjacent to ECG waveforms that are
not in the ECG1 position. They display the ECG lead indicator.
•
The SpO2 waveform digital area displays SpO2 numeric values and the % unit of
measure.
•
The IBP1 through IBP4 waveform digital areas display Systolic, Diastolic and Mean
blood pressure numeric values. When an IBP area has been labeled at the
Spectrum OR, it will also display in the Viewstation OR. Labels include ART, ICP, CVP,
LA, LV, RA, PA and UA.
The IBP1 through IBP4 waveform digital areas can also be viewed with Pulsatile and
Non-Pulsatile layouts. A Pulsatile layout configures Systolic and Diastolic as the primary
parameters and Mean as the secondary parameter. A Non-Pulsatile layout configures
Mean as the primary parameter and Systolic and Diastolic as the secondary
parameters.
•
The Resp waveform digital area displays the Resp numeric value and the RPM unit of
measure.
•
The CO2 waveform digital area displays Inspired and End-Tidal numeric values for CO2
Its unit of measure (mmHg, %, or kPa) is set at the Spectrum OR.
Digital data is shown in the color assigned to the parameter. If individual configurations are
not set, the system default settings are used.
2-4
0070-10-0663-01
Viewstation OR™ Operating Instructions
Viewstation OR Main Screen
2.1.2
The Main Screen
Digital Parameter Tiles
The digital parameter tiles display up to six parameters, based on the sensors available for
the attached Spectrum OR™. Below is a list of available digital parameter tiles.
FIGURE 2-7 Example Digital Parameter Tiles
•
The PVC/min digital parameter tile displays the PVC/min numeric value.
•
The Temperature digital parameter tile (T1, T2, Tb) displays the temperature reading
and its unit of measure (Fahrenheit or Celsius) as set at the Spectrum OR.
•
The CO/CI digital parameter tile displays the Cardiac Output/Cardiac Input numeric
values in LPM/LPM/m2 units of measure.
•
The PAWP digital parameter tile displays the PAWP numeric value in mmHg units of
measure and the time the most recent measurement was taken.
•
The CCO/CCI digital parameter tile displays the CCO/CCI numeric values in
LPM/LPM/m2 units of measure and the time the most recent measurement was taken.
•
The SvO2 digital parameter tile displays the SvO2 numeric value in % units of measure.
•
The NIBP digital parameter tile displays the measures for Systolic, Diastolic, Mean,
Elapsed Time and the time the most recent measurement was taken. It is fixed in the
bottom right corner of the screen and only its display color can be modified.
NOTE:
The CO, PAWP and NIBP parameters have a “Timeout”
setting in the Spectrum OR monitor. If this setting is
exceeded for any of these parameters, the numeric data in
the associated tile will be replaced with dashes and the
“time stamp” will be removed.
Configuring Digital Parameter Tiles
1. Select the desired digital parameter tile. The digital configuration dialog box opens (see
the examples in FIGURE 2-8 and FIGURE 2-9). The configuration options include
Sensor Color and Change To.
FIGURE 2-8 Example CO/CI Digital Configuration Dialog
Viewstation OR™ Operating Instructions
0070-10-0663-01
2-5
The Main Screen
Viewstation OR Main Screen
FIGURE 2-9 Example PVC/min Digital Configuration Dialog
NOTE:
The Change To choice control is based on the sensors utilized
by the Spectrum OR™.
2. Select the Done or Cancel button.
• The Done button accepts the changes.
• The Cancel button rejects the changes.
NOTE:
2-6
Changes made in the Digital Configuration Dialog box will
be reflected in the Display Tab, and vice versa.
0070-10-0663-01
Viewstation OR™ Operating Instructions
Viewstation OR Main Screen
2.1.3
The Main Screen
Menu Bar
The menu bar provides navigation for the Viewstation OR and is fixed at the bottom of the
screen. The table below provides a list of the functions and tabs associated with each menu
button.
FIGURE 2-10 Menu Bar
BUTTON
PURPOSE
Lead
Cycles through each of the waveforms in the ECG ordered set for
ECG1. Not available if using a 3-lead cable. Has no associated tabs.
Patient Setup
Provides access to the following tabs:
System Setup
Viewstation OR™ Operating Instructions
•
The Display tab is used to select the parameter information
displayed on the main screen.
•
The Wave Gain (patient) tab is used to adjust the wave gain
used for displayed waveforms.
Provides access to the following tabs:
•
The Parameter Color tab is used to create the system default
parameter color assignment.
•
The Equipment Setup tab is used to connect and disconnect
the Spectrum OR™ that is displayed at the Viewstation OR.
•
The Touch Screen tab is used to turn the touch screen on and
off and to recalibrate the touch screen display.
•
The Wave Gain (system) tab is used to adjust the system
default wave gain setting for displaying waveforms.
0070-10-0663-01
2-7
The Main Screen
2.1.4
Viewstation OR Main Screen
System Status/Demographics Area
The Viewstation OR displays system status and demographics information at the bottom of
the screen, between the menu bar and the date and time. System events indicate system-wide
issues and are listed based on the system violation. System events are resolved when the
system-wide condition no longer exists. The patient’s bed number and the touch screen status
are also displayed in this area.
NOTE:
If there are multiple messages to display, they will scroll in
the System Status area.
The table below provides the system event messages and bed ID information that may be
shown in the System Status/Demographics area.
MESSAGE/ISSUE*
REASON
System Service required Contact Mindray DS or
Authorized Mindray DS
Distributor (100)
This message will be displayed after the Viewstation OR is
automatically restarted because a system issue was detected.
System Service required Contact Mindray DS or
Authorized Mindray DS
Distributor (300)
The Viewstation OR has detected a hard drive error. Please notify
Mindray DS Service personnel when this system event message is
displayed.
NOTE:
Touch Screen Off
The touch screen functionality is off.
Bed number
The bed number is updated when the bed number is changed on
the Spectrum OR™.
*
2.1.5
It is important to notify Mindray DS Service
personnel when this system event message is
displayed. This message indicates that while
data was being written to or read from the
system hard drive, an error was detected.
Until service personnel correct this issue,
patient data may be lost (for example,
demographic information, alarm
configurations, and historical patient data).
Real time parameter monitoring and alarm
checking will not be affected by this message.
This message will be displayed until reset by
service personnel.
Messages are shown in all bold text.
Date and Time
The date and time, which are provided in the bottom right corner of the Viewstation OR
display, are formatted during software installation.
FIGURE 2-11 Date and Time
2-8
0070-10-0663-01
Viewstation OR™ Operating Instructions
Viewstation OR Main Screen
2.1.6
The Main Screen
Display States
The Viewstation OR displays full-screen messages when it has left active display mode. The
messages and possible resolutions are described below.
MESSAGE
REASON
RESOLUTION
Standby
The Spectrum OR is in a standby
state.
Remove the Spectrum OR from the
standby state. Refer to the Spectrum OR
Operating Instructions Manual.
Discharged
The Spectrum OR is in a
discharged state.
Remove the Spectrum OR from the
discharged state. Refer to the Spectrum
OR Operating Instructions Manual.
Communications
Lost
Communication between the
Viewstation OR computer and the
Spectrum OR has been lost.
Reestablish communication with the
Spectrum OR.
Waiting for
Patient
The Viewstation OR computer has
been rebooted.
Wait for the Viewstation OR computer
to reboot.
The Spectrum OR is not yet
connected to the Panorama
Patient Monitoring Network.
Connect the Spectrum OR to the
Panorama Patient Monitoring Network.
Viewstation OR™ Operating Instructions
0070-10-0663-01
2-9
The Main Screen
Viewstation OR Main Screen
This page intentionally left blank.
2 - 10
0070-10-0663-01
Viewstation OR™ Operating Instructions
3.0
Patient Setup Functions
This section describes patient setup at the Viewstation OR.
• Display Tab
• Wave Gain Tab (Patient)
Viewstation OR™ Operating Instructions
0070-10-0663-01
3-1
Display Tab
3.1
Patient Setup Functions
Display Tab
The Display tab is used to select the waveform parameters that will appear in the main
screen for the active session.
FIGURE 3-1 Example Display Tab
3.1.1
Accessing the Display Tab
• From the menu bar, select the Patient Setup button. The tabs associated with patient
setup are displayed and the Display tab is enabled. The top four Waveform Parameter
tiles and all of the Digital Parameter tiles remain visible.
3.1.2
Components of the Display Tab
The Waveform Parameter tiles are represented by long rectangular buttons (see FIGURE 3-2).
The Digital Parameter tiles are represented by short rectangular buttons (see FIGURE 3-3).
Each button is labeled with its assigned parameter and is presented in a layout that simulates
the main screen. The active display configuration name is listed in the black text box. An
asterisk appears to the left of the display configuration name when unsaved changes have
been made.
FIGURE 3-2 Example Waveform Parameter Tile Buttons
FIGURE 3-3 Example Digital Parameter Tile Buttons
3-2
0070-10-0663-01
Viewstation OR™ Operating Instructions
Patient Setup Functions
Display Tab
Display Configuration Label Dialog
Up to 10 pre-configured displays are available on the Viewstation OR. The Display
Configuration Label dialog lists all of the saved Viewstation OR display configuration labels.
The following factory default settings and their associated parameters may be altered as
needed during initial setup:
• Cardiac
• ECG II waveform parameter tile
• ECG V waveform parameter tile
• SpO2 waveform parameter tile
• IBP1 waveform parameter tile
• IBP2 waveform parameter tile
• PVC/min digital parameter tile
• CO/CI digital parameter tile
• PAWP digital parameter tile
• ALL ECG
• ECG II waveform parameter tile
• ECG V waveform parameter tile
• ECG I waveform parameter tile
• ECG III waveform parameter tile
• ECG aVR waveform parameter tile
• ECG aVL waveform parameter tile
• ECG aVF waveform parameter tile
• Resp
• ECG II waveform parameter tile
• SpO2 waveform parameter tile
• Resp waveform parameter tile
• PVC/min digital parameter tile
• GI
• ECG II waveform parameter tile
• ECG V waveform parameter tile
• SpO2 waveform parameter tile
• IBP I waveform parameter tile
• PVC/min digital parameter tile
Viewstation OR™ Operating Instructions
0070-10-0663-01
3-3
Display Tab
Patient Setup Functions
• Ortho
• ECG II waveform parameter tile
• SpO2 waveform parameter tile
• Resp waveform parameter tile
• PVC/min digital parameter tile
• Neuro
• ECG II waveform parameter tile
• SpO2 waveform parameter tile
• Resp waveform parameter tile
• PVC/min digital parameter tile
NOTE:
The available ECG parameter tiles are based on the ECG
lead set and sensors that are being utilized.
See “Loading a New Display Configuration,” “Removing a Display Configuration,” and
“Editing a Display Name” for more information on the Display Configuration Label dialog.
3.1.3
Configuring the Waveform and Digital Parameter Areas
1. Select the waveform parameter tile button or digital parameter tile button that
corresponds to the tile to which a new parameter will be assigned. The Parameter list
box opens (see FIGURE 3-4). This list box is not presented in the Display tab until a
parameter tile button is selected.
FIGURE 3-4 Example Parameter List Box
2. From the Parameter list box, select the new parameter to be assigned to the tile and
choose the Select button.
NOTE:
An asterisk appears next to the configuration name,
indicating that unsaved changes have been made to that
configuration.
3. Repeat steps 1 and 2 for all tiles to which a new parameter will be assigned.
3-4
0070-10-0663-01
Viewstation OR™ Operating Instructions
Patient Setup Functions
Display Tab
4. When all new parameters have been assigned, select the Save sidebar button and
proceed as described in section 3.1.4. To exit the Display tab without saving any
changes, select the Normal Screen sidebar button. The dialog box shown in
FIGURE 3-5 will open with the message, “Exiting the tab will not save the configuration.
Are you sure you want to exit this tab without saving?” Select Yes to exit the Display
tab without saving any changes. Select No to close the dialog box and return to the
Display tab.
FIGURE 3-5 Exiting without saving
Viewstation OR™ Operating Instructions
0070-10-0663-01
3-5
Display Tab
3.1.4
Patient Setup Functions
Saving a Display Configuration
To save a new display or overwrite an existing display with newly assigned parameters,
proceed as follows:
1. Select the Save sidebar button.
• If fewer than 10 configurations have been saved, the Overwrite/New message
dialog box (shown in FIGURE 3-6) will open. Proceed to step 2.
FIGURE 3-6 Overwrite/New Message Dialog Box
• If 10 configurations have already been saved, a dialog box will open with the
message, “You have reached the maximum number of saved configurations. Do you
want to overwrite an existing display?” (see FIGURE 3-7). Select Yes to open the
Display Config Label dialog box (see FIGURE 3-8). Proceed to the first bullet under
step 2 for further directions. Select No to close the dialog box and return to the
Display tab without making changes.
FIGURE 3-7 “Maximum Configurations Reached” Message Dialog Box
2. To overwrite an existing display, select the Overwrite button. To save as a new
display, select the New button.
• Selecting the Overwrite button opens the Display Config Label dialog box (see
FIGURE 3-8). Select the name of the configuration to be overwritten and choose the
Select button. (Or choose the Cancel button to return to the Display tab without
making changes.)
• Selecting the New button opens the Configuration Label keyboard (see
FIGURE 3-9). Enter the new configuration name and select the Enter button. (Or select
the Close button to return to the Display tab without saving the new configuration.)
• Select the Cancel button to return to the Display tab without making changes.
3-6
0070-10-0663-01
Viewstation OR™ Operating Instructions
Patient Setup Functions
3.1.4.1
Display Tab
Loading a New Display Configuration
1. From the Display tab, select the Load sidebar button. The Display Config Label
dialog box opens.
FIGURE 3-8 Example Display Config Label Dialog Box
2. Select the desired display and choose the Select button, or select the Cancel button to
return to the Display tab without loading the display.
3.1.4.2
Removing a Display Configuration
1. From the Display tab, select the Remove sidebar button. The Display Config Label
dialog box is opened.
2. Select the desired display and choose the Select button, or select the Cancel button to
return to the Display tab without removing the display.
3.1.4.3
NOTE:
An active display cannot be removed.
NOTE:
The Remove button is disabled if there is only one saved
display.
Editing a Display Name
1. From the Display tab, select the Edit Labels sidebar button. The Display Config
Label dialog box is opened.
2. Select the desired display and choose the Select button. The Configuration Label
keyboard (see FIGURE 3-9) is opened. (Select the Cancel button to return to the
Display tab without editing the display.)
3. Edit the label name and select the Enter button or select the Cancel button to return to
the Display tab without editing the display name.
Viewstation OR™ Operating Instructions
0070-10-0663-01
3-7
Display Tab
Patient Setup Functions
Configuration Label Keyboard
The Configuration Label keyboard (shown in FIGURE 3-9) allows users to enter and edit
alphanumeric information. This keyboard is similar to a standard computer keyboard.
NOTE:
Tab controls are disabled when the Configuration Label
keyboard is displayed.
FIGURE 3-9 Configuration Label Keyboard
The Configuration Label Keyboard includes:
3-8
•
The name of the selected field
•
An Edit box to edit and enter the label
•
A Close button to discard changes
•
An Enter button to accept changes
•
A Backspace (<==) button to clear a single character in the edit box
•
A Caps Lock button to enable or disable caps lock (A Caps Lock indicator is shown
when Caps Lock is enabled.)
0070-10-0663-01
Viewstation OR™ Operating Instructions
Patient Setup Functions
3.1.5
Display Tab
Sidebar Buttons
The following sidebar buttons are used to expand the functionality of this tab.
Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, and
close all tabs, message boxes, and dialog boxes.
Save
The Save sidebar button saves changes made.
Load
The Load sidebar button was previously described in "Loading a New Display
Configuration" on page 3-7.
Remove
The Remove sidebar button was previously described in "Removing a Display
Configuration" on page 3-7.
Edit Labels
The Edit Labels sidebar button was previously described in "Editing a Display Name" on
page 3-7.
3.1.6
Troubleshooting
This section lists some of the potential issues that may occur while using the Display tab.
ISSUE
REASON
SOLUTION
A waveform is not displayed.
The waveform has not been
configured or the parameter is
not being utilized.
Configure the waveform area
and save the configuration.
A digital parameter is not
displayed.
The digital parameter has not
been configured or the
parameter is not being utilized.
Configure the digital
parameter area and save the
configuration.
A display configuration cannot
be removed.
The configuration is the current
display.
Close the display and then
remove the configuration.
The configuration is not
displayed.
The configuration has not been
saved.
Save the configuration.
Viewstation OR™ Operating Instructions
0070-10-0663-01
3-9
Wave Gain Tab (Patient)
3.2
Patient Setup Functions
Wave Gain Tab (Patient)
The Wave Gain tab (shown in FIGURE 3-10) is used to change the wave gain settings for
the patient’s available waveforms. The layout, functionality, and choice controls available in
this tab will change based on the Spectrum OR™ configurations.
3.2.1
Accessing the Wave Gain Tab (Patient)
1. From the menu bar, select the Patient Setup menu button. The tabs associated with
patient setup are displayed.
2. Select the Wave Gain tab.
FIGURE 3-10 Example Wave Gain Tab
3.2.2
Configuring Wave Gain Settings
Viewstation OR waveform tiles will initially display the default settings as described in "Wave
Gain Tab (System)" on page 4-5. Configurations made in Patient Setup mode only affect
the active session.
1. In the Wave Gain tab, select the ECG mm/mV button until the desired wave gain
displays. The settings available for the ECG wave gain include 2.5, 5.0, 10.0, 20.0,
and 30.0 mm/mV.
2. Select the IBP1 mmHg through IBP4 mmHg buttons until the desired wave gain
displays. The settings available for the IBP wave gain include 0 – 20, 0 – 40, 0 – 80,
60 – 140, 0 – 160, 0 – 225, 0 – 320, and (-10) – 10 mmHg.
NOTE:
If the IBP button(s) have been labeled at the Spectrum OR
monitor, the same label(s) will be displayed on the
individual wave gain button(s).
3. Select the Resp Gain button until the desired wave gain is displayed. The settings
available for the resp gain wave gain include 1, 2, 3, 4, and 5.
4. Select the CO2 (%, kPa, mmHg) button until the desired wave gain is displayed.
• The CO2 % wave gain settings include 5.0, 7.5, and 10.0 %.
• The CO2 kPa wave gain settings include 5.0, 7.5, and 10.0 kPa.
• The CO2 mmHg wave gain settings include 40, 60, and 100 mmHg.
3 - 10
0070-10-0663-01
Viewstation OR™ Operating Instructions
Patient Setup Functions
Wave Gain Tab (Patient)
5. Select the Done button to accept the wave gain settings shown in the Wave Gain tab
or select the Restore Previous Settings sidebar button to reset the wave gains to
their previous settings.
3.2.3
Sidebar Buttons
The following sidebar buttons are used to expand the functionality of the Wave Gain tab.
Restore Previous Settings
Select the Restore Previous Settings sidebar button to return patient display settings to
those in use when the tab was opened.
Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, and
close all tabs, message boxes, and dialog boxes.
3.2.4
Troubleshooting
This section lists some of the potential issues that may occur while using the patient Wave
Gain tab.
ISSUE
REASON
SOLUTION
The wave gain settings did not
change
The Done button was not
selected after the wave gain
settings were changed.
Adjust the wave gain settings
and select the Done button
when the settings are
completed.
Viewstation OR™ Operating Instructions
0070-10-0663-01
3 - 11
Wave Gain Tab (Patient)
Patient Setup Functions
This page intentionally left blank.
3 - 12
0070-10-0663-01
Viewstation OR™ Operating Instructions
4.0
System Setup Functions
This section outlines the tabs associated with the Viewstation OR default configuration
settings.
• Parameter Color Tab
• Equipment Setup Tab
• Wave Gain Tab (System)
• Touch Screen Tab
Viewstation OR™ Operating Instructions
0070-10-0663-01
4-1
Parameter Color Tab
4.1
System Setup Functions
Parameter Color Tab
The Parameter Color tab (shown in FIGURE 4-1) selects the system default colors for the
parameters which will be displayed in the parameter tiles. Individual parameter colors may
also be adjusted via the Sensor Color configuration option in any waveform configuration
dialog box (see the examples in FIGURE 2-2 and FIGURE 2-3).
4.1.1
Accessing the Parameter Color Tab
1. From the menu bar, select the System Setup button. The tabs associated with the
System Setup button are displayed. By default, the Parameter Color tab is
displayed first.
FIGURE 4-1 Example Parameter Color Tab
4.1.2
Selecting a Color for Parameter Data
This section outlines assigning a color to a parameter in the Parameter Color tab (see
FIGURE 4-2).
1. Select a color tile button. A checkmark indicates that the colored tile is selected.
FIGURE 4-2 Example of a selected color tile
4-2
0070-10-0663-01
Viewstation OR™ Operating Instructions
System Setup Functions
Equipment Setup Tab
2. Select a parameter in the Parameter list box. When a parameter is selected, the
associated label changes from a black background with white text to a white
background with black text.
3. Choose the Select button to accept the parameter selection. The color is assigned to the
selected parameter and the color icon is shown in the list box to the left of the parameter
name.
4.1.3
Sidebar Button - Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, and
close all tabs, message boxes, and dialog boxes.
4.1.4
Troubleshooting
This section lists some of the potential issues that may occur while using the Parameter
Color tab.
4.2
ISSUE
REASON
SOLUTION
A parameter color is
different than expected.
The Sensor Color configuration option
in a configuration dialog box was
changed. When the color is changed in
either a configuration dialog box or
from the Parameter Color tab, the
new color becomes the default.
Choose a new color as
desired. This color
becomes the default.
Equipment Setup Tab
The Equipment Setup tab connects and disconnects the Spectrum OR™ from the
Viewstation OR computer. The Device ID text box lists the device ID of the Spectrum OR that
is currently connected to the Viewstation OR.
FIGURE 4-3 Example Equipment Setup Tab
Viewstation OR™ Operating Instructions
0070-10-0663-01
4-3
Equipment Setup Tab
4.2.1
System Setup Functions
Accessing the Equipment Setup Tab
1. From the menu bar, select the System Setup button.
2. Select the Equipment Setup tab.
4.2.2
Connecting Equipment
WARNING: When connecting any Spectrum OR monitor, verify its
location and associated device ID and bed number.
1. From the Equipment Setup tab, disconnect any active equipment by selecting the
Disconnect button.
2. Select the Connect button. The Device ID dialog box opens, listing all Spectrum OR™
bedside monitors available on the network.
3. Select the desired Spectrum OR to be connected.
4. Select the Done or Cancel button.
• The Done button accepts the change.
• The Cancel button rejects the change and closes the Device ID dialog box.
4.2.3
Sidebar Button - Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, and
close all tabs, message boxes, and dialog boxes.
4.2.4
Troubleshooting
This section lists some of the potential issues that may occur while using the Equipment
Setup tab.
4-4
ISSUE
REASON
SOLUTION
The data that is being displayed at the
Viewstation OR does not correspond
to the data that is being displayed at
the Spectrum OR monitor.
A different Spectrum OR
monitor than the one that is
desired is being displayed
at the Viewstation OR.
Connect the desired
Spectrum OR monitor to
the Viewstation OR.
Data is not displayed.
A Spectrum OR monitor is
not connected to the
Viewstation OR.
Connect the desired
Spectrum OR monitor to
the Viewstation OR.
There are no Spectrum OR
monitors on the network.
Add Spectrum OR
monitors to the network
and connect the desired
Spectrum OR monitor to
the Viewstation OR.
0070-10-0663-01
Viewstation OR™ Operating Instructions
System Setup Functions
4.3
Wave Gain Tab (System)
Wave Gain Tab (System)
The Wave Gain tab (shown in FIGURE 4-4) is used to select the default wave gain settings.
For information regarding individual wave gain settings, refer to the "Wave Gain Tab
(Patient)" on page 3-10.
NOTE:
4.3.1
The changes made in the Wave Gain tab (System) will apply
to newly admitted Spectrum OR monitors.
Accessing the Wave Gain Tab
1. From the menu bar, select the System Setup button.
2. Select the Wave Gain tab.
FIGURE 4-4 Example Wave Gain Tab (System)
4.3.2
Changing the System Default Wave Gain Settings
This section outlines entering and modifying the system default wave gain settings.
1. Select the ECG mm/mV button until the desired ECG wave gain is displayed.
• The settings available for the ECG wave gain include 2.5, 5.0, 10.0, 20.0, and 30.0
mm/mV. The factory default setting is 10.0 mm/mV.
2. Select the IBP1 mmHg through IBP4 mmHg buttons until the desired IBP wave gains
are displayed. There are four IBP buttons.
• The settings available for the IBP wave gains include 0 – 20 mmHg, 0 – 40 mmHg,
0 – 80 mmHg, 60 – 140 mmHg, 0 – 160 mmHg, 0 – 225 mmHg, 0 – 320 mmHg
and (-10) – 10 mmHg. The factory default setting is 0 – 320 mmHg.
3. Select the Resp Gain button until the desired Resp Gain wave gain is displayed.
• The settings available for the Resp Gain wave gain include 1, 2, 3, 4, and 5. The
factory default setting is 1.
4. Select the CO2 (%, kPa, mmHg) button until the desired wave gain is displayed.
• CO2 % wave gain settings include 5.0, 7.5, and 10.0 %.
• CO2 kPa wave gain settings include 5.0, 7.5, and 10.0 kPa.
• CO2 mmHg wave gain settings include 40, 60, and 100 mmHg.
4.3.3
Sidebar Button - Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, and
close all tabs, message boxes, and dialog boxes.
Viewstation OR™ Operating Instructions
0070-10-0663-01
4-5
Touch Screen Tab
4.4
System Setup Functions
Touch Screen Tab
A Viewstation OR configuration can include a touch screen display. A touch pad or mouse
can also be used for point-and-click system navigation. This section provides an overview of
how to turn the touch screen on and off and how to recalibrate it.
FIGURE 4-5 Touch Screen Tab
4.4.1
Accessing the Touch Screen Tab
1. From the menu bar, select the System Setup button.
2. Select the Touch Screen tab.
4.4.1.1
Turning the Touch Screen On and Off
The touch screen functionality must be turned off before cleaning (see "Care and Cleaning of
the Touch Screen" on page 5-2). There is one button for turning the touch screen on or off.
The touch screen must be enabled to exit the Touch Screen tab.
• When the Touch Screen On button displays, the touch screen is active.
• When the Touch Screen Off button displays, the touch screen is inactive.
FIGURE 4-6 Touch Screen On Button
NOTE:
4-6
FIGURE 4-7 Touch Screen Off Button
When the Touch Screen Off button displays, the message
“Touch Screen OFF” will display in the System Status/
Demographics area.
0070-10-0663-01
Viewstation OR™ Operating Instructions
System Setup Functions
4.4.1.2
Touch Screen Tab
Recalibrating the Touch Screen
The touch screen should be recalibrated whenever the touch interface becomes difficult to
maneuver.
1. From the Touch Screen tab, select Recalibrate Touch Screen. The recalibration
process is initiated (see FIGURE 4-8).
NOTE:
The text in the Recalibrate Touch Screen display is always
displayed in the English language.
NOTE:
During the recalibration process, the parameter tiles, menu
bar, and sidebar buttons will be replaced by the
recalibration screen.
FIGURE 4-8 Touch Screen Recalibration
2. Touch the red targets on the screen as instructed. Three calibration targets are displayed.
3. The message Touch the screen. Does the cursor follow your finger? is
displayed (see FIGURE 4-9). Touch all four corners and all four sides of the screen.
FIGURE 4-9 Check Calibration Screen
Viewstation OR™ Operating Instructions
0070-10-0663-01
4-7
Touch Screen Tab
System Setup Functions
The
and
buttons appear.
• Select the
button if the targets moved as they were selected. Once selected, the
screen is recalibrated.
• Select the
button if the targets did not move as they were selected. Once selected,
the recalibration process begins again.
The message box will time out in thirty seconds and the recalibration of the touch screen will
automatically be accepted.
4.4.2
Sidebar Button - Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, and
close all tabs, message boxes, and dialog boxes.
NOTE:
4.4.3
The Normal Screen sidebar button does not display while
the touch screen is recalibrating.
Troubleshooting
This section lists some of the potential issues that may occur while recalibrating the Touch
Screen tab.
ISSUE
REASON
SOLUTION
Recalibration settings not
accepted
The Recalibration screen timed
out because the first target was
not touched within 30 seconds.
Try to recalibrate again and
touch the first target before the
30 second time out.
Recalibration steps are
repeated
The
error.
Recalibrate again and if the
targets moved as they were
selected, select the
button.
button was selected in
Contact your local Mindray DS Service Representative if recalibration of the touch screen
does not improve the performance of the touch screen display.
4-8
0070-10-0663-01
Viewstation OR™ Operating Instructions
5.0
User Maintenance
This section of the manual outlines proper maintenance of the Viewstation OR touch screen.
Viewstation OR™ Operating Instructions
0070-10-0663-01
5-1
Care and Cleaning of the Touch Screen
5.1
User Maintenance
Care and Cleaning of the Touch Screen
For the best performance, it is recommended that the Viewstation OR touch screen be kept
clean.
NOTE:
The touch screen must be turned off prior to cleaning. (See
"Turning the Touch Screen On and Off" on page 4-6.)
Observe the following precautions when cleaning the touch screen:
CAUTION:
To avoid potential system issues, deactivate the touch screen
before cleaning.
CAUTION:
DO NOT spray glass cleaner directly on a display as it could
possibly leak inside a non-sealed unit and cause damage.
• Always remember to use a cloth or towel to apply glass cleaner to the touch screen.
• Any standard glass cleaner can be used to clean the touch screen. DO NOT use abrasive
cleaning materials to clean a touch screen. DO NOT use alcohol or solvents containing
chlorinated hydrocarbons.
• Remove fingerprints and stains by using a liquid lens cleaner and a soft cloth.
• Use a fine soft-hair brush to carefully brush away dust and dirt particles.
NOTE:
5-2
Remember to turn the touch screen back on after cleaning.
(See "Turning the Touch Screen On and Off" on page 4-6.)
0070-10-0663-01
Viewstation OR™ Operating Instructions
6.0
Parameter Specifications
This section is for reference only to list the allowable range specifications for all of the
parameters measured by the Spectrum OR™.
Viewstation OR™ Operating Instructions
0070-10-0663-01
6-1
Heart Rate (HR)
6.1
Parameter Specifications
Heart Rate (HR)
The Heart Rate parameter obtains data from ECG and IBP sources.
Unit of Measure:
Beats Per Minute (bpm)
Resolution:
1 bpm
Heart Rate Source: ECG
PATIENT SIZE
MINIMUM VALUE (bpm)
MAXIMUM VALUE (bpm)
Neonate
30
350
Pediatric
30
300
Adult
30
300
The Viewstation OR displays ECG waveforms when using 3-wire and 5-wire lead cables. The
ECG leads available at the Viewstation OR are dependent on the lead cable connected to
the patient.
LEAD
WAVEFORM LEADS AVAILABLE
3-wire lead set
I, II, III
5-wire lead set
I, II, III, aVR, aVL, aVF, V
Heart Rate Source: IBP
PATIENT SIZE
6.2
MINIMUM VALUE (bpm)
MAXIMUM VALUE (bpm)
Neonate
30
350
Pediatric
30
300
Adult
30
300
Respiration (Resp)
The Respiration parameter obtains data from ECG and CO2 sources.
Unit of Measure:
Respirations Per Minute (rpm)
Resolution:
1 rpm
Respiration Source: ECG Thoracic Impedance
6-2
PATIENT SIZE
MINIMUM VALUE (rpm)
MAXIMUM VALUE (rpm)
Neonate
4
199
Pediatric
4
199
Adult
4
199
0070-10-0663-01
Viewstation OR™ Operating Instructions
Parameter Specifications
NIBP
Respiration Source: CO2
PATIENT SIZE
MINIMUM VALUE (rpm)
MAXIMUM VALUE (rpm)
Neonate
0
150
Pediatric
0
150
Adult
0
150
The Viewstation OR supports respiration waveforms.
6.3
NIBP
The NIBP parameter includes measurements for the Systolic (Sys), Diastolic (Dia), Mean and
Elapsed Time (ET).
Unit of Measure:
mmHg
Resolution:
1 mmHg
ET
Unit of Measure:
minute
Resolution:
1 minute
NIBP (Sys)
PATIENT SIZE
MINIMUM VALUE (mmHg)
MAXIMUM VALUE (mmHg)
Neonate
45
120
Pediatric
55
160
Adult
55
235
PATIENT SIZE
MINIMUM VALUE (mmHg)
MAXIMUM VALUE (mmHg)
Neonate
20
100
Pediatric
30
150
Adult
30
200
NIBP (Dia)
NIBP (Mean)
PATIENT SIZE
MINIMUM VALUE (mmHg)
MAXIMUM VALUE (mmHg)
Neonate
greater than the NIBP (Dia) value
less than the NIBP (Sys) value
Pediatric
greater than the NIBP (Dia) value
less than the NIBP (Sys) value
Adult
greater than the NIBP (Dia) value
less than the NIBP (Sys) value
Viewstation OR™ Operating Instructions
0070-10-0663-01
6-3
SpO2
Parameter Specifications
NIBP (ET)
6.4
PATIENT SIZE
MINIMUM VALUE (MINUTES)
MAXIMUM VALUE (MINUTES)
Neonate
0
999
Pediatric
0
999
Adult
0
999
SpO2
The SpO2 parameter includes measurements from Masimo SET® and Nellcor® SpO2
sensors.
Unit of Measure:
%
Resolution:
1%
PATIENT SIZE
MINIMUM VALUE (%)
MAXIMUM VALUE (%)
Neonate
---
100
Pediatric
---
100
Adult
---
100
The Viewstation OR supports SpO2 waveforms.
6.5
IBP
The Viewstation OR supports up to four invasive blood pressure parameters. The
measurements shown for each pressure include Systolic (Sys), Diastolic (Dia) and Mean.
NOTE:
IBP parameter labels include IBP, Art, UA, LV, PA, CVP,
ICP, LA, and RA. The label used is based on the label
selected at the Spectrum OR monitor.
Resolution:
1 mmHg
IBP (Sys)
PATIENT SIZE
MINIMUM VALUE (mmHg)
MAXIMUM VALUE (mmHg)
Neonate
-30
300
Pediatric
-30
300
Adult
-30
300
MINIMUM VALUE (mmHg)
MAXIMUM VALUE (mmHg)
IBP (Dia)
PATIENT SIZE
6-4
Neonate
-30
300
Pediatric
-30
300
Adult
-30
300
0070-10-0663-01
Viewstation OR™ Operating Instructions
Parameter Specifications
CO2
IBP Mean
PATIENT SIZE
MINIMUM VALUE (mmHg)
MAXIMUM VALUE (mmHg)
Neonate
-30
300
Pediatric
-30
300
Adult
-30
300
The Viewstation OR supports IBP waveforms. The Viewstation OR will not display an IBP
waveform if the IBP cable is not connected to the system.
6.6
CO2
The Viewstation OR supports the mmHg, kPa and % units of measure for CO2.
NOTE:
6.7
The unit of measure shown for CO2 is based on the selection
made at the Spectrum OR monitor.
Units of Measure:
mmHg, %, kPa
Resolution:
1 mmHg, 1 %, 1 kPa
Temperature (T1, T2, and T Blood)
The Viewstation OR supports four temperature measurements.
Units of Measure:
F° and C°
Resolution:
0.1 F° and 0.1 C°
T1 and T2
PATIENT SIZE
MINIMUM VALUE (F°)
MAXIMUM VALUE (F°)
Neonate
59
113
Pediatric
59
113
Adult
59
113
PATIENT SIZE
MINIMUM VALUE (C°)
MAXIMUM VALUE (C°)
Neonate
15
45
Pediatric
15
45
Adult
15
45
T Blood Values
PATIENT SIZE
MINIMUM VALUE (F°)
MAXIMUM VALUE (F°)
Neonate
Not Applicable
Not Applicable
Pediatric
63.7
109.1
Adult
63.7
109.1
Viewstation OR™ Operating Instructions
0070-10-0663-01
6-5
Cardiac Output
6.8
6.9
6.10
6-6
Parameter Specifications
Cardiac Output
Unit of Measure:
lpm
Resolution:
0.1 lpm
PATIENT SIZE
MINIMUM VALUE (LPM)
MAXIMUM VALUE (LPM)
Pediatric
0.2
20.0
Adult
0.2
20.0
Cardiac Index
Unit of Measure:
lpm/m2
Resolution:
0.1 lpm/m2
PATIENT SIZE
MINIMUM VALUE (LPM/M2)
MAXIMUM VALUE (LPM/M2)
Pediatric
0.1
20.0
Adult
0.1
20.0
Pulmonary Artery Wedge Pressure (PAWP)
Unit of Measure:
mmHg
Resolution:
1 mmHg
PATIENT SIZE
MINIMUM VALUE (mmHg)
MAXIMUM VALUE (mmHg)
Pediatric
-30
300
Adult
-30
300
0070-10-0663-01
Viewstation OR™ Operating Instructions
Parameter Specifications
6.11
Invalid Data Display
Invalid Data Display
The Viewstation OR uses dashes (---), Xs (XX) or a straight line (waveform) to show invalid
data.
Dashes (---) Digital Data
Digital data is replaced with dashes (---) for the following reasons:
• Data is above the maximum value
• Data is below the minimum value
• Parameter is not available
• Lead or a sensor is off
• Sensor is not zeroed
• Parameter is zeroing
• Measurement has not been taken
• Measurement is being taken
• Measurement has timed out
• Measurement is being retried
• Parameter is turned off
Xs (XX) Digital Data
Digital data is replaced with Xs (XX) for the following reasons:
• IBP parameter is not zeroed
• NIBP parameter failed to obtain a reading
• CO2 not calibrated
Straight Line (Waveforms)
Waveform data is replaced with a continuous straight line for the following reasons:
• ECG leads are disconnected
• ECG cable is not connected to the system
• Respiration parameter is disabled at the Spectrum OR monitor and the source is ECG
• SPO2 cable is connected and not in use
• IBP cable is connected but not zeroed or is zeroing
• CO2 filter line is not in use or is zeroing
Viewstation OR™ Operating Instructions
0070-10-0663-01
6-7
Invalid Data Display
Parameter Specifications
This page intentionally left blank.
6-8
0070-10-0663-01
Viewstation OR™ Operating Instructions
7.0
Performance Specifications
This section outlines performance specifications for the Viewstation OR.
• ECG
• ECG Derived Heart Rate Meter
• Viewstation OR Computer
• I/O Control Devices
• Displays
Viewstation OR™ Operating Instructions
0070-10-0663-01
7-1
ECG
Performance Specifications
7.1
ECG
Three Lead Displayable Leads:
I, II, III (one vector at a time)
Five Lead Displayable Leads:
I, II, III, aVR, aVL, aVF, V
Aspect Ratio
0.4 ± 0.08 sec/mV (met by 25 mm/s)
ANSI/AAMI EC13-2002, 4.2.9.7 (b)
Channel Width
40 mm. min
ANSI/AAMI EC13-2002, 4.2.9.7 (a)
Pacer Rejection
The heart rate meter rejects pulses of amplitude signals
from ±2mV to ±700mV and duration 0.1ms to 2ms with
no tail according to ANSI/AAMI EC13-2002 Section
4.1.4.1 (No tails).
Pacemaker Pulse
The pacer detector’s slew rate is 6.024V/s
Rejection Capability
7.1.1
ECG Derived Heart Rate Meter
Range:
30 to 300 bpm Adult/Pediatric ANSI/AAMI
EC13–2002, 4.2.7
30 to 350 bpm Neonatal ANSI/AAMI
EC13–2002, 4.2.7
Resolution:
7.1.2
7-2
1 bpm
ECG Respiration
Range:
4 to 199 breaths per minute
Resp. Scale:
1, 2, 3, 4 or 5 with standard ECG cable
0070-10-0663-01
Viewstation OR™ Operating Instructions
Performance Specifications
NIBP
7.2
NIBP
7.2.1
Systolic Pressure Measurement
Range:
55 to 235 mmHg in Adult mode
55 to 160 mmHg in Pediatric mode
45 to 120 mmHg in Neonatal mode
Resolution:
7.2.2
1 mmHg
Diastolic Pressure Measurement
Range:
30 to 200 mmHg in Adult mode
30 to 150 mmHg in Pediatric mode
20 to 100 mmHg in Neonatal mode
Resolution:
7.2.3
1 mmHg
Pulse Rate
Range:
35 to 245 bpm, for Adults/Pediatric
70 to 245 bpm, for Neonate
Resolution:
7.3
1 bpm
IBP
-30 to 300 mmHg after zeroing at the Spectrum OR™
Range:
7.3.1
IBP Heart Rate Meter
Range:
30 to 300 bpm Adult/Pediatric
30 to 350 bpm Neonatal
Resolution:
Viewstation OR™ Operating Instructions
1 bpm
0070-10-0663-01
7-3
Temperature
7.4
Performance Specifications
Temperature
Scale:
Range:
Selectable Celsius or Fahrenheit
15 °C to 45 °C
59 °F to 113 °F
Resolution:
0.1 °C
0.1 °F
7.5
7.6
7-4
SpO2
SpO2 Range:
70% to 100%
SpO2 Pulse Rate Range:
20 to 249 bpm
CO2
Range:
0 - 13%s
Respiration Rate Range:
0 - 150 breaths/minute
0070-10-0663-01
Viewstation OR™ Operating Instructions
Performance Specifications
7.7
Viewstation OR Computer
Viewstation OR Computer
There is one basic hardware platform for the Viewstation OR computer system. It contains the
following:
• 1 GB of RAM
• 80 GB system hard drive, minimum
• Four channel serial interface for touch screen inputs and future features
• 2 on-board Ethernet connectors
• CD-R/W
• 3.5” Floppy Drive
7.7.1
Real Time Clock
This is used for various time-related functions in Viewstation OR.
7.7.2
Power Supply
Voltage
• 100-240 VAC (±10%)
Frequency
• 60/50 Hz (±3Hz)
Viewstation OR Power Consumption Maximum
• 115 VAC @ 1.4 A + 20%
• 230 VAC @ 0.6 A + 20%
7.7.3
Operating Temperature
• +5 to +40 º C
7.7.4
Operating Humidity
• 20% to 80% Relative Humidity Maximum, non-condensing, maximum wet bulb 29 º C
7.7.5
Operating Altitude
• 1060 to 700 hPa (-1250 to 9889 feet ASL) (-380 - 3014 m) (795 to 525 mmHg)
Viewstation OR™ Operating Instructions
0070-10-0663-01
7-5
Viewstation OR Computer
7.7.6
Performance Specifications
Storage Temperature
• -20 to +60 º C
7.7.7
Storage Humidity
• 5% to 80% Relative Humidity Maximum, non-condensing, maximum wet bulb 35 º C
7.7.8
Storage Altitude
• 1060 to 700 hPa (-1250 to 9889 feet ASL) (-380 - 3014 m) (795 to 525 mm Hg)
7.7.9
Safety Standards
The Viewstation OR complies with relevant portions of the following standards:
1. UL 60950-1:2003
Safety of Information Technology Equipment
2. CSA-C22.2 60950-1-03
Safety of Information Technology Equipment
3. IEC 60601-1-2:2001
Electromagnetic compatibility
4. ANSI/AAMI EC 13:2002,
4.2.9.7 (a) (b), 4.1.4.1
Cardiac Monitors, heart rate meters, and alarms
5. ISTA Test Procedure 1A
Shipping Integrity Test
SAFETY DESIGNATIONS
SEE SECTION 7.7.10 (TABLE 7-1, TABLE 7-2, TABLE 7-3, TABLE 7-4)
7-6
Mode of Operation
Continuous
Protection Against Hazards of Explosion
Non Protected (ordinary)
Protection Against Ingress of Liquids
IPX0 (ordinary, non-protected)
Degree of Mobility
Stationary, Intra Hospital
0070-10-0663-01
Viewstation OR™ Operating Instructions
Performance Specifications
7.7.10
Viewstation OR Computer
Electromagnetic Compatibility
The Viewstation OR meets the requirements of IEC 60601-1-2:2001.
NOTE:
The Viewstation OR needs special precautions regarding
Electro Magnetic Compatibility (EMC) and needs to be
installed and put into service according to the EMC
information provided in this section.
NOTE:
Portable and mobile RF communications equipment can
affect the Viewstation OR. Refer to Table 7-1 on page 7-7,
Table 7-2 on page 7-8, and Table 7-4 on page 7-10.
TABLE 7-1
GUIDANCE AND MINDRAY DS USA INC. DECLARATION - ELECTROMAGNETIC
EMISSIONS
The Viewstation OR is intended for use in the electromagnetic environment specified below. The
customer or the user of the Viewstation OR should assure that it is used in such an environment.
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
EMISSIONS TEST
COMPLIANCE
RF emissions
CISPR 11
Group 1
The Viewstation OR uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A
Harmonic emissions
IEC 61000-3-2 *
Class A
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3 *
Complies
The Viewstation OR is suitable for use in
all establishments other than domestic
and those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes
*
Viewstation OR and Panorama Server only
Viewstation OR™ Operating Instructions
0070-10-0663-01
7-7
Viewstation OR Computer
Performance Specifications
TABLE 7-2
GUIDANCE AND MINDRAY DS USA INC. DECLARATION - ELECTROMAGNETIC
IMMUNITY
The Viewstation OR is intended for use in the electromagnetic environment specified below. The
customer or the user of the Viewstation OR should assure that it is used in such an environment.
IMMUNITY
TEST
IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines ±1
kV for input/
output lines
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Mains power quality should be typical
of a commercial or hospital
environment.
Surge IEC
61000-4-5
±1 kV
differential
mode ±2 kV
common mode
± 1 kV
differential mode
± 2 kV common
mode
Mains power quality should be typical
of a commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines IEC
61000-4-11
<5% UT (>95%
dip in UT) for
0,5 cycle
<5% UT (>95%
dip in UT) for 0,5
cycle
40% UT (60%
dip in UT) for
5 cycles
40% UT (60%
dip in UT) for
5 cycles
Mains power quality should be typical
of a commercial or hospital
environment. If the user requires
continued operation during power
mains interruptions, it is recommended
that the device be powered from an
uninterruptible power supply or a
battery.
70% UT (30%
dip in UT) for
25 cycles
70% UT (30%
dip in UT) for
25 cycles
< 5% UT (>95%
dip in UT) for 5
sec
< 5% UT (>95%
dip in UT) for 5
sec
3 A/m
3 A/m
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
UT
7-8
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
is the a.c. mains voltage prior to application of the test level.
0070-10-0663-01
Viewstation OR™ Operating Instructions
Performance Specifications
Viewstation OR Computer
TABLE 7-3
GUIDANCE AND MINDRAY DS USA INC. DECLARATION - ELECTROMAGNETIC
IMMUNITY
The Viewstation OR is intended for use in the electromagnetic environment specified below. The
customer or the user of the Viewstation OR should assure that it is used in such an environment.
IMMUNITY
TEST
IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC ENVIRONMENT GUIDANCE
Portable and mobile RF communications
equipment should be used no closer to any part
of the Telepack, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
Conducted
RF IEC
61000-4-6
3 Vrms 150
kHz to 80
MHz
3 Vrms
d = 1.2 x
Radiated RF
IEC
61000-4-3
3 V/m 80
MHz to 2.5
GHz
3 V/m
d = 1.2 x P 80 MHz to 800 MHz
P
d = 2.3 x P 80 MHz to 2.5 GHz
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey*
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
NOTE:
At 80 MHz and 800 MHz, the higher frequency range
applies.
NOTE:
These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
*
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment, due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength where the Telepack is
used exceeds the applicable RF compliance level, the Telepack should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the Telepack.
NOTE:
Viewstation OR™ Operating Instructions
Viewstation OR equipment should not be used in the
presence of high energy amplitude modulated (AM)
signals that are operating within the audible range. These
signals are not normally found in the hospital
environment.
0070-10-0663-01
7-9
Viewstation OR Computer
Performance Specifications
TABLE 7-4
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND THE VIEWSTATION OR AND TELEMETRY
SERVER
The Viewstation OR is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of thee Viewstation OR can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Viewstation OR as recommended below, according
to the maximum output power of the communications equipment.
RATED MAXIMUM
OUTPUT POWER OF
TRANSMITTER W (WATTS)
SEPARATION DISTANCE ACCORDING TO FREQUENCY
OF TRANSMITTER M (METERS)
150 KHZ TO
80 MHZ
80 MHZ TO
800 MHZ
800 MHZ TO
2.5 GHZ
D = 1.2 x
D = 1.2 x
D = 2.3 x
P
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
P
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
7.7.11
NOTE:
At 80 MHz and 800 MHz, the separation distance for the
higher frequency range applies.
NOTE:
These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Physical Characteristics
Maximum Size
• Depth: 17.5" (445.5 mm)
• Height: 17.0" (431.8 mm)
• Width: 7.5" (190.5 mm)
Maximum Weight
• 30.0 lbs (13.6 kg)
7 - 10
0070-10-0663-01
Viewstation OR™ Operating Instructions
Performance Specifications
7.8
I/O Control Devices
I/O Control Devices
The Viewstation OR can be used with a mouse, touch pad and touch screen. These I/O
control devices are used to make selections from various menus and screens. The mouse and
touch pad are supported for system diagnostics and user interface, operating independent of
the touch screen.
NOTE:
The mouse and touch pad connect to the PS/2 port on the
CPU. Since there is only one PS/2 port, these I/O control
devices cannot be used simultaneously.
NOTE:
Touch screen capability is dependent on the chosen display
as detailed in the next section.
7.9
Displays
7.9.1
Mindray DS Supplied ELO-Compatible Touch Screen Display
• A Surface Acoustical Wave (SAW) touch screen is supported as the standard input device
on the 19” display.
• The touch screen operates independent of a mouse or touch pad interface.
• The actual touch screen and its controller are integrated into the display.
• The touch screen communication is accepted by Viewstation OR via a serial port.
• Use of the touch screen does not preclude the use of a mouse or touch pad.
7.9.2
Customer Supplied Compatible Displays
• Contact Mindray DS Customer Support for requirements.
Viewstation OR™ Operating Instructions
0070-10-0663-01
7 - 11
Warranty
7.10
Performance Specifications
Warranty
Domestic Warranty
Mindray DS USA Inc. warrants that its products will be free from defects in workmanship and
materials for a period of one (1) year from the date of purchase. Disposable or one-time use
products are warranted to be free from defects in workmanship and materials up to a date
one year from the date of purchase or the date of first use, whichever is sooner. This warranty
does not cover consumable items such as, but not limited to, batteries, external cables,
sensors, cuffs, hoses, or mounts.
Mindray DS USA Inc. shall not be liable for any incidental, special, or consequential loss,
damage, or expense directly or indirectly arising from the use of its products, liability under
this warranty and the buyer’s exclusive remedy under this warranty is limited to servicing or
replacing at Mindray DS’s option at the factory or at an authorized Mindray DS Distributor,
any product which shall under normal use and service appear to the Company to have been
defective in material or workmanship.
No agent, employee, or representative of Mindray DS USA Inc. has any authority to bind
Mindray DS USA Inc. to any affirmation, representation, or warranty concerning its products,
and any affirmation, representation or warranty made by any agent, employee, or
representative shall not be enforceable by buyer.
This warranty is expressly in lieu of any other express or implied warranties, including any
implied warranty or merchantability or fitness, and of any other obligation on the part of the
seller.
Damage to any product or parts through misuse, neglect, accident, or by affixing any
non-standard accessory attachments or by any customer modification voids this warranty.
Mindray DS USA Inc. makes no warranty whatever in regard to trade accessories, such
being subject to the warranty of their respective manufacturers.
A condition of this warranty is that this equipment or any accessories which are claimed to
be defective be returned when authorized by Mindray DS, freight prepaid to Mindray DS
USA Inc., Mahwah, New Jersey 07430. Mindray DS USA Inc. shall not have any
responsibility in the event of loss or damage in transit.
Calibration may be performed without the need to disassemble the instrument. It is the
responsibility of the purchaser to perform calibration as necessary, in accordance with the
instructions provided in this manual.
International Warranty
Mindray DS USA Inc. warrants that its products will be free from defects in workmanship and
materials for a period of two (2) years from the date of purchase. Disposable or one-time use
products are warranted to be free from defects in workmanship and materials up to a date
one year from the date of purchase or the date of first use, whichever is sooner. This warranty
does not cover consumable items such as, but not limited to, batteries, external cables,
sensors, cuffs, hoses, or mounts.
7 - 12
0070-10-0663-01
Viewstation OR™ Operating Instructions
Performance Specifications
Warranty
Mindray DS USA Inc. shall not be liable for any incidental, special, or consequential loss,
damage, or expense directly or indirectly arising from the use of its products, liability under
this warranty and the buyer’s exclusive remedy under this warranty is limited to servicing or
replacing at Mindray DS’s option at the factory or at an authorized Mindray DS Distributor,
any product which shall under normal use and service appear to the Company to have been
defective in material or workmanship.
No agent, employee, or representative of Mindray DS USA Inc. has any authority to bind
Mindray DS USA Inc. to any affirmation, representation, or warranty concerning its products,
and any affirmation, representation or warranty made by any agent, employee, or
representative shall not be enforceable by buyer.
This warranty is expressly in lieu of any other express or implied warranties, including any
implied warranty or merchantability or fitness, and of any other obligation on the part of the
seller.
Damage to any product or parts through misuse, neglect, accident, or by affixing any
non-standard accessory attachments or by any customer modification voids this warranty.
Mindray DS USA Inc. makes no warranty whatever in regard to trade accessories, such
being subject to the warranty of their respective manufacturers.
A condition of this warranty is that this equipment or any accessories which are claimed to
be defective be returned when authorized by Mindray DS, freight prepaid to Mindray DS
USA Inc., Mahwah, New Jersey 07430. Mindray DS USA Inc. shall not have any
responsibility in the event of loss or damage in transit.
Calibration may be performed without the need to disassemble the instrument. It is the
responsibility of the purchaser to perform calibration as necessary, in accordance with the
instructions provided in this manual.
Viewstation OR™ Operating Instructions
0070-10-0663-01
7 - 13
Phone Numbers and How To Get Assistance
7.11
Performance Specifications
Phone Numbers and How To Get Assistance
Mindray DS maintains a network of service representatives and factory-trained distributors.
Prior to requesting service, perform a complete operational check of the instrument to verify
proper control settings. If operational problems continue to exist, contact the Mindray DS
Service Department at (800) 288-2121 or (201) 995-8116 for assistance in determining the
nearest field service location.
Please include the instrument model number, the serial number, and a description of the
problem with all requests for service.
Any questions regarding the warranty should be directed to the nearest Mindray DS location.
A list of international offices, along with their phone numbers, is provided at the end of this
manual.
7.12
Mindray DS’s Responsibility
Mindray DS is responsible for the effects on safety, reliability and performance of the
equipment only if:
a. assembly operations, extensions, readjustments, modifications or repairs are carried
out by persons authorized by Mindray DS; and
b. the electrical installation of the relevant room complies with the appropriate
requirements; and
c. the equipment is used in accordance with the instructions for use
7 - 14
0070-10-0663-01
Viewstation OR™ Operating Instructions
8.0
Glossary
This section provides a glossary of terminology used throughout the Viewstation OR
Operating Instructions.
• Terms, Acronyms, and Abbreviations
Viewstation OR™ Operating Instructions
0070-10-0663-01
8-1
Terms, Acronyms, and Abbreviations
8.1
Glossary
Terms, Acronyms, and Abbreviations
TERMS
DEFINITIONS
Bedside Monitor
A monitoring device that has its own user interface and does
not require the Viewstation OR to be its primary monitoring
source
CCI
Continuous Cardiac Index
CCO
Continuous Cardiac Output
CI
Cardiac Index
CO
Cardiac Output
Heart Rate
Number of heartbeats detected during a one minute time
period
Invasive Blood
Method of obtaining blood pressure using an internal pressure
Pressure (IBP)
sensor
Monitoring Device
The instrumentation connected to the patient. The monitoring
device acquires the patient data and then transmits it onto the
network.
Non-Invasive Blood
Method of obtaining blood pressure using an external cuff
Pressure (NIBP)
8-2
Panorama Patient
Supports hardwired communication of the real-time patient
Network
data collected by monitoring devices attached to the patient
PAWP
Pulmonary Artery Wedge Pressure
Respiration Rate
Number of breaths during a one minute time period
SpO2
Level of Oxygen Saturation in the blood
SvO2
Mixed Venous Oxygen Saturation
0070-10-0663-01
Viewstation OR™ Operating Instructions
Glossary
Terms, Acronyms, and Abbreviations
ACRONYMS/
ABBREVIATIONS
DEFINITIONS
ART
Arterial Line
BPM
Beats per minute
CO2
Carbon Dioxide
CVP
Central Venous Pressure
ECG
Electrocardiogram
EMC
Electromagnetic Compatibility
ET
End-Tidal when referencing CO2 parameters
Elapsed Time when referencing NIBP
HR
Heart Rate
IBP
Invasive Blood Pressure
ICP
Intracranial Pressure
IEC
International Electro-Technical Commission
INSP
Inspired
kPa
Kilopascals Per Atmosphere
LA
Left Arm when referring to ECG Leads (AHA)
LV
Left Ventricle
MDD
Medical Device Directive
mmHg
Millimeters of Mercury
NIBP
Non-Invasive Blood Pressure
PA
Pulmonary Artery
Pleth
Plethysmograph
PVC
Premature Ventricular Contraction
RA
Right Atrium when referring to IBP
Right Arm when referring to ECG Leads (AHA)
RESP
Respiration Rate
RL
Right Leg when referring to ECG Leads (AHA)
RPM
Respirations per minute
SAW
Surface Acoustical Wave
Tb
Temperature of the Blood
UA
Umbilical Artery
V
Chest Lead (AHA)
Viewstation OR™ Operating Instructions
0070-10-0663-01
8-3
0070-10-0663-01
Rev C
September 06, 2010
Mindray DS USA, Inc. • 800 MacArthur Boulevard • Mahwah, NJ 07430 • USA •
Dom. Customer Service: 1.800.288.2121 • Intl. Customer Service: +1.201.995.8000 •
Dom. Fax: 1.800.926.4275 • Intl. Fax: +1.201.995.8680 • www.mindray.com
Mindray Medical Netherlands B.V.• P.O. Box 26 • 3870 CA Hoevelaken • The Netherlands •
Tel: +31 33 25 44 911 • Fax: +31 33 25 37 621
Mindray (UK) Limited • 3 Percy Road • St. John’s Park • Huntingdon • Cambridgeshire PE29 6SZ •
United Kingdom • Tel: 01480 416840 • Fax: 01480 436588
Mindray Medical France SARL • Europarc Créteil •123, Chemin des Bassins •
94035 Créteil Cedex • France • Tel: (0)1.45.13.91.50 • Fax: (0)1.45.13.91.51
Mindray Medical German GmbH • Zwischen den Bächen 4 • 64625 Bensheim • Germany •
Tel: +49.6251.17524-0 • Fax: +49.6251.17524-20
Mindray Medical International Ltd. • 2813 Office Tower, Convention Plaza • No 1 Harbour Road •
Wanchai • Hong Kong • Tel: +852 2793 5596 • Fax: +852 2344 8824
0002-08-8918
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertisement