Thermo Fisher Scientific STARA2130 Orion™ Star AA Battery Powered Dissolved Oxygen or RDO Benchtop Meter Use and Care Manual

Add to my manuals
110 Pages

advertisement

Thermo Fisher Scientific STARA2130 Orion™ Star AA Battery Powered Dissolved Oxygen or RDO Benchtop Meter Use and Care Manual | Manualzz

Thermo Scientific

Orion Star A210 Series

Electrochemistry Benchtop Meters

68X000441

Revision A

June 2015

Important Note

Please read this user guide thoroughly before using your meter. Any use outside of these instructions may invalidate your warranty and cause permanent damage to the meter.

Contact Information

For assistance with Thermo Scientific

Orion

products, contact Technical Support by email at [email protected]

or by phone – within the United States call 1-800-225-1480 and outside the United States call +1-978-232-6000 or fax +1-978-232-6031.

For additional product information, contact your local authorized dealer, local Thermo Scientific

Orion technical sales representative or contact us using the Water and Laboratory Products

(WLP) information on the page back of this manual.

Applications and Technical Resources

Visit www.thermoscientific.com/water to view Thermo Scientific Orion products and download product literature, user guides and manuals, software updates, and the latest application and technical resources.

Table of Contents

Chapter 1 ....................................................................................................................... 6

Meter Introduction ........................................................................................................ 6

Meter Overview ............................................................................................................................ 6

Packing List ................................................................................................................................. 7

Intended Use ............................................................................................................................... 7

Chapter 2 ....................................................................................................................... 8

Meter Basics ................................................................................................................. 8

Using the Universal Power Adapter ............................................................................................. 8

Installing Batteries (Optional Power Source) ............................................................................... 9

Attaching the Electrode Stand and Holder ................................................................................. 10

Using the Electrode Holder .............................................................................................. 11

Meter Connections ..................................................................................................................... 12

Meter Keypad ............................................................................................................................ 14

Function Keys .................................................................................................................. 15

Meter Display ............................................................................................................................. 16

Measurement Display Icons ............................................................................................. 18 pH Electrode Condition Icon ............................................................................................. 18

Meter Models and Measurement Capabilities ............................................................................ 19

Meter Maintenance .................................................................................................................... 19

Chapter 3 ..................................................................................................................... 20

Meter Setup Menus ..................................................................................................... 20

Main Setup Menu ....................................................................................................................... 20

General Setup Menu Navigation ...................................................................................... 20

Channel-Specific Method, Mode and Temperature Menus........................................................ 24

Method Menu ................................................................................................................... 25

Mode and Settings Menu ................................................................................................. 28

Temperature Menu ........................................................................................................... 41

Instrument Settings Setup Menu................................................................................................ 45

Log View Menu .......................................................................................................................... 46

Data Log ........................................................................................................................... 46

Calibration Log ................................................................................................................. 47

Diagnostics Menu ...................................................................................................................... 48

Meter Self Test Procedure ............................................................................................... 48

Electrode Stability Test Procedure ................................................................................... 49

Chapter 4 ..................................................................................................................... 50

Using the pH or pH/ISE Channel ............................................................................... 50

Meter and Electrode Preparation ............................................................................................... 50

pH Calibration Procedure .......................................................................................................... 51

pH Calibration Editing....................................................................................................... 52

ORP Calibration Procedure (Relative mV Mode) ....................................................................... 53

ORP Calibration Procedure (E

H

Units) ....................................................................................... 54

ISE Calibration Procedure ......................................................................................................... 55

ISE Calibration Editing ..................................................................................................... 56

Measurement Procedure ........................................................................................................... 57

Chapter 5 ..................................................................................................................... 58

Using the Conductivity Channel ............................................................................... 58

Meter and Sensor Preparation ................................................................................................... 58

Conductivity Calibration Procedure ............................................................................................ 59

Conductivity Calibration Editing ........................................................................................ 60

Alternative Certified Cell Constant Entry Calibration Procedure ...................................... 60

Conductivity Standards vs. Temperature Table ......................................................................... 61

Conductivity Meter Verification Procedure ................................................................................. 62

Measurement Procedure ........................................................................................................... 63

Chapter 6 ..................................................................................................................... 64

Using the DO/RDO Channel ....................................................................................... 64

Meter and Sensor Preparation ................................................................................................... 64

Dissolved Oxygen Calibration Procedure .................................................................................. 65

Air (Water-saturated Air) Calibration ................................................................................ 66

Water (Air-saturated Water) Calibration ........................................................................... 66

Manual (Winkler) Calibration ............................................................................................ 67

Set Zero Calibration ......................................................................................................... 68

Measurement Procedure ........................................................................................................... 69

Chapter 7 ..................................................................................................................... 70

Data Transfer and Software Updates ........................................................................ 70

Data Storage and Transfer Settings .......................................................................................... 70

Measurement Read Type Setting ..................................................................................... 70

Data Log Setting .............................................................................................................. 71

Export Data Settings ........................................................................................................ 72

Printer Compatibility and Requirements .................................................................................... 74

Computer Compatibility and Requirements ............................................................................... 75

Using the USB-to-Serial Computer Cable ........................................................................ 76

Using the USB Computer Cable ....................................................................................... 78

Orion Star Com Communication Software ....................................................................... 81

Meter Interfacing with Other Computer Programs ............................................................ 82

Star A200-A300 Meter Remote Control Protocols ............................................................ 83

Meter Software Upgrade Procedure .......................................................................................... 90

Chapter 8 ..................................................................................................................... 94

Customer Services ..................................................................................................... 94

Troubleshooting Tips ................................................................................................................. 95

Meter Factory Reset Procedure ....................................................................................... 96

Meter User Reset Procedure ............................................................................................ 96

Notice of Compliance ................................................................................................................. 97

WEEE Compliance .................................................................................................................... 97

Declaration of Conformity .......................................................................................................... 98

Meter Specifications .................................................................................................................. 99

Ordering Information ................................................................................................................ 106

Meter Accessories, Electrodes and Solutions ................................................................ 108

C H A P T E R 1

1

Meter Introduction

Meter Overview

Thermo Scientific

Orion Star

A210 series benchtop meters feature an informative, easy to read graphic display with backlight and onscreen calibration and setup instructions for intuitive, user-friendly operation with minimum training. The comprehensive keypad with menu-specific function keys and useful shortcut keys allows fast and efficient meter navigation and control.

The multilanguage user interface allows customization of the meter in a variety of local languages and new languages can be added through software updates.

Measurements can be taken quickly and reliably with the onscreen reading stability indicator and selectable read modes – Auto-Read, timed or continuous with hold function. The data log collects up to 2000 measurement sets with time and date stamp and the non-volatile meter memory preserves all data, even with the loss of power. Use Thermo Scientific

Orion

Star

Com

software to facilitate data transfer from the meter to a computer via USB or RS232, export data to an Excel spreadsheet or comma separated value file (.csv) and print data to a network or local printer.

Utilize the Orion Star stirrer probe for meter-powered and meter-controlled sample stirring with five selectable stirring speeds. Use the stirrer probe with the included electrode stand to simplify placement into and out of solutions and eliminate the need for magnetic stir bars and plates.

The IP54-rated dust and splash resistant meter housing offers the flexibility to place the meter where it is needed – on the bench or wall-mounted to save space. For added site adaptability, power the meter using the included universal power adapter or optional four AA batteries. Six

Orion Star A210 series benchtop meters are available to meet your exact measurement needs.

6 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 1

|

Meter Introduction

Orion Star A211 pH Benchtop Meter

Measure pH, mV, relative mV or ORP with temperature

Orion Star A212 Conductivity Meter

Measure conductivity, TDS, salinity or resistivity with temperature

Orion Star A213 RDO/DO Meter

Measure dissolved oxygen as % saturation or concentration with temperature using either

RDO

®

optical or polarographic DO sensors

Orion Star A214 pH/ISE Meter

Measure ion concentration using an ion selective electrode (ISE), pH, mV, relative mV or ORP with temperature

Orion Star A215 pH/Conductivity Meter

Measure pH, mV, relative mV or ORP with temperature on channel one and measure conductivity, TDS, salinity or resistivity with temperature on channel two

Orion Star A216 pH/RDO/DO Meter

Measure pH, mV, relative mV or ORP with temperature on channel one and measure dissolved oxygen as % saturation or concentration with temperature on channel two

Packing List

Orion Star A210 series benchtop meters and meter kits include the following items:

·

Meter-attachable electrode stand and holder

· Universal power adapter

·

Literature CD

· Printed quick start guide

·

Computer interface cable

· Meter test certificate

Each Orion Star A210 series meter is fully tested and certified by Thermo Fisher Scientific and a certificate of calibration is included with each meter. For specific meter and kit contents, refer

to the Ordering Information

section. Visit our website at www.thermoscientific.com/OrionMeters to download the complimentary Orion Star Com data transfer computer software and Star

A200/A300 series USB driver.

Intended Use

Please read this reference guide thoroughly. Any use outside of these instructions may invalidate the meter warranty and cause permanent damage to the meter.

Orion Star A210 Series Benchtop Meter User Manual |

7

C H A P T E R 2

2

Meter Basics

Using the Universal Power Adapter

A universal power adapter (Catalog Number 1010003) with US, EU, UK and China plug plates is included with the Orion Star A210 series benchtop meters. This universal power adapter is specifically for use with Star A210 series benchtop meters. Use of other power adapters can damage the meter and void the warranty.

1. Select the appropriate plug plate for the power outlet that will be used.

2. Remove the clear plastic cover from the groove on the back of the power adapter.

3. Slide the appropriate plug plate into the groove on the back of the power adapter.

4. Connect the assembled power adapter to a power outlet and the meter input labeled Power.

A surge protector or uninterrupted power supply (UPS) is also recommended.

8 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 2

|

Meter Basics

Installing Batteries (Optional Power Source)

1. Select four new AA alkaline batteries.

2. Confirm that the meter is powered off.

3. Turn the meter over, with the meter display facing down, on a clean dry surface.

4. Remove the battery compartment cover – push down on the battery compartment tab and lift the battery compartment cover up.

5. Install the new batteries in the battery compartment, orientating the batteries as shown in the battery compartment housing.

6. Replace the battery compartment cover and turn the meter over, with the meter display facing up.

Orion Star A210 Series Benchtop Meter User Manual |

9

Chapter 2

|

Meter Basics

Attaching the Electrode Stand and Holder

The electrode stand can be attached to either side of the meter and up to two stands can be attached to each meter. A weighted base (Catalog Number STARA-HB) is also available to support the stand without attachment to the meter.

1. Open the box containing the electrode stand. The box will include a base plate, electrode arm, pivot pin, electrode holder, ATC spacer and RDO clip.

2. Turn the meter over, with the meter display facing down, on a clean dry surface.

3. Identify the side of the meter that the stand will be installed on and remove the screw between the circles on that side of the meter.

4. Align the base plate of the stand with the circles on the meter.

5. Replace the screw from step 3 to attach the base plate to the meter.

6. Turn the meter over, with the meter display facing up.

7. Insert the electrode arm into the metal post on the base plate.

8. Connect the electrode holder to the electrode arm using the pivot pin.

10 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Chapter 2

|

Meter Basics

Using the Electrode Holder

Place electrodes in the stand for easy movement in and out of containers during calibration, sample measurement and storage. Recommended electrode positions are shown below.

15 mm holder for conductivity and dissolved oxygen sensors

Holder with 20° angle for gas sensing ion selective electrodes

Stirrer probe holder

12 mm flexible holder for pH, ORP and ion selective electrodes

12 mm holder with optional ATC probe insert

Optional RDO optical dissolved oxygen sensor holder

Thermo Scientific Orion Star A210 Series Benchtop Meter User Manual |

11

Chapter 2

|

Meter Basics

Meter Connections

The following diagram depicts all the possible meter connections available on the Orion Star

A210 series benchtop meters. Some Star A210 series benchtop meters will have fewer connections, depending on the meter’s measurement capabilities.

Connector Function

Power

Stirrer

USB

RS232

BNC

Ref.

ATC/CON

Ground

Connect the universal power adapter (included with meter) to power the meter

Connect an Orion Star stirrer probe (Catalog Number

096019) for meter-controlled and meter-powered sample stirring with five selectable speeds

Connect the USB cable for bi-directional data transfer and communication via USB to a printer or computer

Connect the RS232 cable for bi-directional data transfer and communication via RS232 to a printer or computer

Connect a pH electrode, ORP/redox electrode or ion selective electrode (ISE) with BNC connector

Connect a half-cell reference electrode with standard 2.5 mm pin-tip connector

Connect an ATC temperature probe with 8 pin MiniDIN connector

Connect a conductivity sensor or ATC temperature probe with 8 pin MiniDIN connector

Applicable Meter Models

All Star A210 series meters

All Star A210 series meters

All Star A210 series meters

All Star A210 series meters

Star A211, Star A214,

Star A215, Star A216 meters

Star A211, Star A214,

Star A215, Star A216 meters

Star A211, Star A214,

Star A216 meters

Star A212, Star A215 meters

Use to reduce interference generated by other equipment All Star A210 series meters

DO/RDO

Connect a dissolved oxygen sensor (either RDO optical or polarographic) with 9 pin MiniDIN connector

Star A213, Star A216 meters

Orion Star A210 series benchtop meters are compatible with the same electrodes and sensors used with Thermo Scientific

Orion

Versa Star

meters and legacy Thermo Scientific

Orion

Star

and Star Plus meters. This includes the Thermo Scientific

Orion

ROSS Ultra

Triode

™ pH/ATC electrodes, Thermo Scientific

Orion

ROSS

pH electrodes and Thermo Scientific

Orion

DuraProbe

conductivity sensors.

12 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 2

|

Meter Basics

Orion Star A211 pH Meter Connections

Orion Star A212 Conductivity Meter Connections

Orion Star A213 RDO/DO Meter Connections

Orion Star A214 pH/ISE Meter Connections

Orion Star A215 pH/Conductivity Meter Connections

Orion Star A216 pH/RDO/DO Meter Connections

Orion Star A210 Series Benchtop Meter User Manual |

13

Chapter 2

|

Meter Basics

Meter Keypad

The Orion Star A210 series benchtop meter keypad includes menu-specific function keys that update on the display for fast and efficient meter operation and shortcut keys for easy navigation to the measurement, setup, data log and calibration log menus.

Key Icon and Name Function f1, f2, f3 power measure (esc)

Press the f1, f2 and f3 function keys to perform the action shown above each key on the display.

Press the power key to turn on the meter.

When the meter is powered on, press and release the power key to switch the display backlight on and off.

Press and hold the power key for about three seconds to turn off the meter.

Press the measure (esc) key in the Auto-Read measurement mode to start a new measurement.

Press the measure (esc) key to escape the current mode or menu and return to the measurement mode.

Press the log view key to access the data log and calibration log from the measurement mode.

log view

Press the stirrer key to turn the stirrer probe on and off in the continuous or timed measurement mode and the calibration edit mode.

stirrer

14 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 2

|

Meter Basics

Key Icon and Name Function

setup / up arrow (p)

Press the setup key to enter the setup menu from the measurement mode.

Press the up arrow (p) key to scroll up through a list of items.

mode / right arrow (u)

Press the mode key to change the measurement mode of the displayed channel.

Press the right arrow (u) key to scroll right through a list of items.

log/print / down arrow

(q)

Press the log/print key to manually log and/or print a measurement, depending on the selected measurement mode and data output settings.

Press the down arrow (q) key to scroll down through a list of items.

In the continuous measurement mode, press the hold key to lock (freeze) the current displayed measurement and press the hold key again to release

(unfreeze) the measurement.

Press the left arrow (t) key to scroll left through a list of items.

Hold / left arrow (t)

Function Keys

The following image shows the f1, f2 and f3 keys with their corresponding actions above each function key on the meter display. Pressing the f1 (cal) key will prompt the meter to enter the calibration mode. Pressing the f2 (sample ID) key will prompt the meter to enter the sample ID setup mode. Pressing the f3 (setup) key will prompt the meter to enter the main setup menu.

Orion Star A210 Series Benchtop Meter User Manual |

15

Chapter 2

|

Meter Basics

Temperature measurement and source

Main measurement mode

Main measurement value

Calibration points from most recent active calibration

Temperature measurement and source

Main measurement mode

Main measurement value

Calibration points from most recent active calibration

Temperature measurement and source

Main measurement mode

Main measurement value

Barometric pressure

Meter Display

Measurement Display Examples

The following displays are examples only. Actual meter displays will vary based on the selected meter setup parameters, active calibration data, etc.

Orion Star A211 pH Measurement Display

Date and time

Method

Main measurement units mV corresponding to main measurement

Orion Star A212 Conductivity Meter Display

Date and time

Method

Main measurement units

Temperature compensation setting

Orion Star A213 RDO/DO Meter Display

Date and time

Method

Main measurement units

Salinity value

16 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Temperature measurement and source

Main measurement mode

Main measurement value

Channel 1 main measurement mode

Channel 1 temperature measurement and source

Channel 2 main measurement mode

Channel 2 temperature measurement and source

Channel 1 main measurement mode

Channel 1 temperature measurement and source

Channel 2 main measurement mode

Channel 2 temperature measurement and source

Orion Star A214 pH/ISE Meter Display

Date and time

Method

Main measurement units

Chapter 2

|

Meter Basics mV corresponding to main measurement

Orion Star A215 pH/Conductivity Meter Display

Display each channel separately or both channels simultaneously using the f3 (channel) key.

Date and time

Channel 1 main measurement value

Channel 1 main measurement units

Channel 2 main measurement value

Channel 2 main measurement units

Orion Star A216 pH/RDO/DO Meter Display

Display each channel separately or both channels simultaneously using the f3 (channel) key.

Date and time

Channel 1 main measurement value

Channel 1 main measurement units

Channel 2 main measurement value

Channel 2 main measurement units

Thermo Scientific Orion Star A210 Series Benchtop Meter User Manual |

17

Chapter 2

|

Meter Basics

Measurement Display Icons

Display Icon Description

Shown when the meter is running on AC power

Shown when the meter is running on battery power

Indicates when data is exported to a computer or printer

Indicates when a measurement is recorded in the data log

Shown when an alarm is set and the alarm is triggered

Indicates the RS232 port is selected for printer or computer interfacing

Indicates the USB port is selected for printer or computer interfacing

Displays the active temperature measurement and indicates the source as an

ATC temperature sensor (ATC) or manually entered temperature value (MAN)

Shown when the hold key is pressed and displayed measurements are frozen, press the hold key a second time to release the hold function

Indicates a calibration has been successfully completed, flashes when the calibration alarm is set and the alarm is triggered

Indicates the pH electrode condition as good (two bars), fair (one bar) or bad

(slash through it), based on the last saved calibration and electrode stability

Indicates the active measurement method number, M100 and M200 are the default methods and indicate a password-protected method is not in use

Stability indicator will flash stabilizing while the measurement is changing and show ready when the measurement is stable

When the read type is set to Auto-Read, the icon will flash while the reading is stabilizing and remain solid when the reading is stable and locked on the display

Indicates the sample ID function is active, the operator assigned number is shown to the right of the icon

Indicates the user ID function is active, the operator assigned name is shown to the right of the icon

pH Electrode Condition Icon

Orion Star A211 pH meters, Orion Star A214 pH/ISE meters, Orion Star A215 pH/conductivity meters and Orion Star A216 pH/RDO/DO meters include a pH electrode condition icon. In the measurement mode, the pH electrode condition icon indicates the performance of the pH electrode, based on the last saved calibration and electrode measurement stability.

Icon pH Electrode Status

Electrode condition is good. The electrode slope is in the range of 95.1% to 104.9%.

Electrode condition is fair. The electrode slope is 85.1% to 95% or 105% to 114.9%.

Electrode condition is bad. The electrode slope is 85% or lower or 115% or higher. Consult the pH electrode manual for instructions on how to clean, condition and troubleshoot the electrode.

Note: This is a general indicator of overall electrode status – always refer to the pH electrode user manual for specific information on the recommended slope range for each pH electrode.

18 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 2

|

Meter Basics

Meter Models and Measurement Capabilities

The table below shows the available models of Orion Star A210 series benchtop meters and their available measurement modes. All measurements include temperature.

Meter Model

Star A211 pH Meter

Star A212

Conductivity Meter

Star A213

RDO/DO Meter

Star A214 pH/ISE Meter

Star A215 pH/Conductivity

Meter

Star A216 pH/RDO/DO Meter

pH mV

RmV

ORP

ISE pH mV

RmV

ORP pH mV

RmV

ORP

Channel 1

Measurement Modes

pH mV

RmV

ORP

Conductivity

TDS

Salinity

Resistivity

DO as % Saturation

DO as mg/L

Channel 2

Measurement Modes Temperature Modes

n/a

Automatic

Manual n/a

Automatic

Manual n/a n/a

Conductivity

TDS

Salinity

Resistivity

DO as % Saturation

DO as mg/L

Automatic

Automatic

Manual

Automatic

Manual

Automatic

Manual (Ch 1 only)

Meter Maintenance

·

For routine meter maintenance, dust and wipe with a damp cloth. If necessary, warm water or mild water-based detergent can be used.

· Meter maintenance can be performed on a daily, weekly or monthly basis, as required by the operating environment.

·

Remove any spilled substances immediately from the meter or module using the proper cleaning procedure for that spill type.

Orion Star A210 Series Benchtop Meter User Manual |

19

C H A P T E R 3

3

Meter Setup Menus

Main Setup Menu

The main setup menu of the Orion Star A210 series meters contains menus for measurement settings, instrument settings, calibration and data logs and meter diagnostics in one location.

General Setup Menu Navigation

1. In the measurement mode, press the setup key to access the main setup menu.

2. Press the p, q, t or u key to scroll through and highlight a setup menu icon and press the f3 (select) key to access the submenus for the selected menu.

3. Press the p or q key to scroll and highlight a submenu option and press the f3 (select) key to access the parameters for the selected submenu.

4. Perform the appropriate actions to set the parameters and settings in the selected menu. a. To select a value from a list of options, press the p or q key to highlight the desired value and then press the f3 (select) key to set the value. b. To enter a numeric value, use the popup number entry screen. i. Press the f3 (edit) key to open the number entry screen.

0 1 2 . ii. Press the p, q, t or u key to highlight a number, decimal point or negative sign and then press the f3

3 4 5 6

7 8 9 –

(enter) key to select the highlighted item. Repeat until the

desired value is shown at the top of the screen. backspace

back done enter

iii. Press the f2 (done) key to save the value and exit the number entry screen.

5. Press the f1 (back) key to navigate out of a menu and press the measure (esc) key to return to the measurement mode at any time.

20 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Chapter 3

|

Meter Setup Menus

Main Setup Menu Icons and Descriptions

Icon Description

Use the pH Channel menu to customize measurement, calibration and alarm settings for pH, mV, RmV (relative mV), ORP and temperature

Use the pH/ISE Channel menu to customize measurement, calibration and alarm settings for pH, mV,

RmV (relative mV), ORP, ISE and temperature

Use the COND Channel menu to customize measurement, calibration and alarm settings for conductivity, salinity, TDS, resistivity and temperature

Use the DO/RDO Channel menu to customize measurement, calibration and alarm settings for dissolved oxygen and temperature

Use the Instrument Settings menu to update meter settings for data transfer, data log, date and time, language, sounds, stirrer speed, display contrast, autoshutoff, user ID and sample ID

Access the View Log menu to view, export or clear saved data in the data log and view or print the ten most recent calibrations per channel in the calibration log

Access the Diagnostics menu to initiate a meter reset, perform a meter self test, test electrode stability or view the meter serial number and software revision

Applicable Meter Models

Star A211 pH meter

Star A215 pH/conductivity meter

Star A216 pH/RDO/DO meter

Star A214 pH/ISE meter

Star A212 conductivity meter

Star A215 pH/conductivity meter

Star A213 RDO/DO meter

Star A216 pH/RDO/DO meter

All Star A210 series meters

All Star A210 series meters

All Star A210 series meters

Example Main Setup Menu Displays

Star A211 pH Meter Setup Menu Star A214 pH/ISE Meter Setup Menu

Thermo Scientific

Star A215 pH/Conductivity Meter Setup Menu Orion Star A216 pH/RDO/DO Meter Setup Menu

Orion Star A210 Series Benchtop Meter User Manual |

21

Chapter 3

|

Meter Setup Menus

Main Setup Menu Flow Chart

Main Setup

Menu

Channel* Settings View Log Diagnostics

Method

Mode and

Settings

Export Data

Data Log

Data Log

Factory

Reset

Calibration

Log

User Reset

Temperature Date / Time Self Test

Stability Test

About Meter

Language

Key Press

Beep

Alarm Beep

Stirrer

Contrast

Auto Shut

Off

User ID

Sample ID

* Refer to next figure for a detailed list of the measurement, calibration and alarm settings within each channel-specific Mode and Settings menu.

22 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 3

|

Meter Setup Menus

Channel-specific Mode and Settings Menu Flow Chart

Main Setup

Menu pH Channel

Mode and

Settings

Measure

Mode

Read Type pH/ISE

Channel

Mode and

Settings

Measure

Mode

Read Type

COND

Channel

Mode and

Settings

Measure

Mode

Read Type

DO/RDO

Channel

Mode and

Settings

Measure

Mode

Measure

Unit

Resolution

Buffer

Group

Stability

Averaging

Limit Alarm

Cal Due

Alarm

Resolution

Buffer

Group

Measure

Unit

Blank

Correct

Stability

Averaging

Limit Alarm

Cal Due

Alarm

Low Lv

Stability

Cell K

Type

(Salinity)

TDS Factor

Ref. Temp.

Temp.

Comp.

Temp.

Coeff.

Stability

Resolution

Read Type

Baro

Pressure

Salinity

Correct

Stability

Averaging

Averaging

Limit Alarm

Cal Due

Alarm

RDO Cap

Life

Limit Alarm

Cal Due

Alarm

Cell Type

Orion Star A210 Series Benchtop Meter User Manual |

23

Chapter 3

|

Meter Setup Menus

Channel-Specific Method, Mode and

Temperature Menus

Within the pH Channel, pH/ISE Channel, COND Channel and DO/RDO Channel menus are

Method, Mode and Settings and Temperature submenus, which can be used to customize the measurement, calibration and alarm settings for each channel. The channel menus displayed by each meter will depend on the meter model and its measuring capabilities.

Meter Model

Orion Star A211 pH meter

Orion Star A212 conductivity meter

Orion Star A213 RDO/DO meter

Orion Star A214 pH/ISE meter

Orion Star A215 pH/conductivity meter

Orion Star A216 meters pH/RDO/DO meter

Available Channel-specific Menus

pH Channel

COND Channel

RDO/DO Channel pH/ISE Channel pH Channel

COND Channel pH Channel

RDO/DO Channel

·

Method – Use the Method setup menu to create, load, copy, edit or delete password

protected methods.

·

Mode and Settings – Select the Mode and Settings menu to review and update the

measurement, calibration and alarm settings for the selected channel.

·

Temperature – Use the Temperature menu to manually enter a sample temperature value,

set the temperature units as °C or °F, perform a temperature calibration on an ATC probe or conductivity / dissolved oxygen sensor with built-in temperature and set temperature input source for meters with dual temperature sources.

24 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 3

|

Meter Setup Menus

Method Menu

Save up to ten channel-specific methods in the Method menu for fast and easy recollection of custom channel-specific measurement, calibration and alarm settings.

The default method (M100 or M200) is updated whenever the channel-specific measurement, calibration and alarm settings are changed in the Mode and Settings menu and the default method is not password protected.

The custom methods (M101-M110 or M201-M210) can be protected using a three to eight digit password. When a protected method is created, loaded and active in the measurement mode and a calibration is performed, that calibration will be saved to the method so every time the method is loaded, the corresponding calibration will also be loaded.

Protected methods are helpful when using two or more electrodes on one channel. For example, a pH electrode and an ion selective electrode (ISE) used on the same BNC input of the Star A214 pH/ISE meter or a low range conductivity sensor and standard range conductivity sensor used on the same 8 pin MiniDIN input of the Star A212 conductivity meter.

Creating a New Method using the Current Meter Settings

1. In the measurement mode, press the setup key.

2. Press the p, q, t or u key to highlight pH Channel, pH/ISE Channel, COND Channel or

DO/RDO Channel and press the f3 (select) key.

3. Press the p or q key to highlight Method and press the f3 (select) key to access the methods list.

4. Press the p or q key to highlight Current Settings and press the f3 (save) key.

5. Press the p or q key to highlight an open method (M101-M110 or M201-M210) to save the current channel-specific settings to and press the f2 (accept) key. a. Open methods have no date, time or mode shown in the methods list.

6. Use the popup number entry screen to create a method-specific password (3 to 8 digits). a. Press the f3 (edit) key to access the number entry screen. b. Press the p, q, t or u key to highlight a number, press the

f3 (enter) key to select the number and repeat until the desired

password is shown at the top of the number entry screen. c. Press the f2 (done) key to accept the password and exit the number entry screen.

7. Press the f2 (accept) key to save the entered password.

0 1 2 .

3 4 5 6

7 8 9 – backspace

back done enter

8. The meter will return to the methods list and the selected method number will have the date, time and mode shown in the methods list.

Orion Star A210 Series Benchtop Meter User Manual |

25

Chapter 3

|

Meter Setup Menus

Loading a Method

Use the load function to activate a protected method for use in the measurement mode. If no protected method has been created, the open method is active. A password is not required to load a protected method.

1. In the measurement mode, press the setup key.

2. Press the p, q, t or u key to highlight pH Channel, pH/ISE Channel, COND Channel or

DO/RDO Channel and press the f3 (select) key.

3. Press the p or q key to highlight Method and press the f3 (select) key to access the methods list.

4. Press the p or q key to highlight a protected method to be loaded in the measurement mode and press the f2 (load) key. a. Protected methods have the date, time and mode shown in the methods list.

5. The meter will automatically proceed to the measurement mode.

Copying a Method

Use the copy function to save the channel-specific measurement, calibration and alarm settings of an existing method to a new method, so the original method settings are preserved and the new method settings can be modified. When using the copy function to create a new method, a new password must be created for the new method to be saved.

1. In the measurement mode, press the setup key.

2. Press the p, q, t or u key to highlight pH Channel, pH/ISE Channel, COND Channel or

DO/RDO Channel and press the f3 (select) key.

3. Press the p or q key to highlight Method and press the f3 (select) key to access the methods list.

4. Press the p or q key to highlight a protected method to be copied to an open protected method and press the f3 (options) key. a. Protected methods have the date, time and mode shown in the methods list.

5. Use the popup number entry screen to enter the specific password for the selected method. a. Press the f3 (edit) key to access the number entry screen. b. Press the p, q, t or u key to highlight a number, press the f3 (enter) key to select the number and repeat until the correct password is shown at the top of the screen. c. Press the f2 (done) key to accept the password and exit the number entry screen.

6. Press the f2 (accept) key to submit the password for the selected method.

7. Press the p or q key to highlight Copy and press the f2 (accept) key.

8. Press the p or q key to highlight an open method and press the f2 (accept) key. a. Open methods have no date, time or mode shown in the methods list.

9. Use the popup number entry screen to create a password for the new method.

26 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 3

|

Meter Setup Menus a. Press the f3 (edit) key to access the number entry screen. b. Press the p, q, t or u key to highlight a number, press the f3 (enter) key to select the number and repeat until the desired password is shown at the top of the screen. c. Press the f2 (done) key to accept the password and exit the number entry screen.

10. Press the f2 (accept) key to submit the new password for the new method.

11. The meter will return to the methods list and the new method number will have the date, time and mode shown in the methods list.

Editing or Deleting a Method

Use the edit function to modify the channel-specific measurement, calibration and alarm settings of an existing protected method. Use the delete function to convert an existing protected method to an open method.

1. In the measurement mode, press the setup key.

2. Press the p, q, t or u key to highlight pH Channel, pH/ISE Channel, COND Channel or

DO/RDO Channel and press the f3 (select) key.

3. Press the p or q key to highlight Method and press the f3 (select) key to access the methods list.

4. Press the p or q key to highlight a protected method and press the f3 (options) key. a. Protected methods have the date, time and mode shown in the methods list.

5. Use the popup number entry screen to enter the specific password for the selected method. a. Press the f3 (edit) key to access the number entry screen. b. Press the p, q, t or u key to highlight a number, press the f3 (enter) key to select the number and repeat until the correct password is shown at the top of the screen. c. Press the f2 (done) key to accept the password and exit the number entry screen.

6. Press the f2 (accept) key to submit the password for the selected method.

7. Press the p or q key to highlight Edit or Delete and press the f2 (accept) key. a. If Edit is selected: i. Perform the appropriate actions to set the channel-specific measurement, calibration and alarm settings within the selected method. ii. Once the required settings have been edited, press the f1 (back) key to return to the methods list. b. If Delete is selected: i. Press the f2 (yes) key to confirm the deletion of the selected method. ii. The method will be deleted and the meter will automatically proceed to the measurement mode.

Orion Star A210 Series Benchtop Meter User Manual |

27

Chapter 3

|

Meter Setup Menus

Mode and Settings Menu

General Mode and Settings Options

Read Types

The read type determines how the meter displays, logs and exports measurements. Ensure the appropriate data export and data log settings are enabled in the Instrument Settings menu before logging or exporting measurements.

·

Auto-Read:

o

Press the measure (esc) key to start a measurement. The AR icon blinks as the changing value stabilizes. When stable, the AR icon stops blinking and the measurement is locked on the display until the measure (esc) key is pressed again. o

When enabled, the stable measurement is automatically saved to the data log and exported to a printer or computer.

·

Continuous:

o

Measurement values are continuously updated on the display and the Stabilizing or

Ready icon indicates the measurement stability status. o

When enabled, press the log/print key to save a measurement to the data log and export it to a printer or computer.

·

Timed:

o

Measurement values are continuously updated on the display and the Stabilizing or

Ready icon indicates the measurement stability status. o

When enabled, measurements are saved to the data log and exported to a printer or computer at the selected time interval of 3 seconds (00:00:03) to 24 hours

(24:00:00).

Stability Setting Information

The stability setting determines when a measurement is recognized as stable by the meter, using an algorithm that evaluates measurement fluctuations. When Smart Stability is selected, the measurement resolution is utilized in conjunction with the algorithm. Approximate mV per minute values, estimated with ideal measurement conditions, are shown below. Values are theoretical only and actual values will vary based on individual measurement conditions.

Smart Stability Setting

0.1 resolution or 1 significant digit

0.01 resolution or 2 significant digits

0.001 resolution or 3 significant digits

mV per Minute Stability Setting

7.6

2.9

1.0

Fast Stability

Medium Stability

Slow Stability

mV per Minute

7.6

2.9

1.0

28 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Chapter 3

|

Meter Setup Menus

Averaging Setting

The averaging setting allows the operator to select Automatic Smart averaging for faster measurement stability or Off for no measurement averaging once the stability criteria is met.

Limit Alarm Setting

The limit alarm setting allows measurements to be evaluated using set high limit and/or low limit values. If a measurement goes above the high limit value or below the low limit value, the alarm is triggered. For example, when the high limit is set to 8.50 pH, the alarm will activate when a measurement of 8.51 pH or higher is read in the measurement mode. Turn on the alarm beep setting in the Instrument Settings menu to enable an audible beep when the alarm is triggered.

Calibration Due Alarm Setting

The calibration due alarm allows the operator to set a custom calibration interval in hours. If a calibration is not performed within the specified calibration interval, the alarm is triggered. Turn on the alarm beep setting in the Instrument Settings menu to enable an audible beep when the alarm is triggered.

Thermo Scientific Orion Star A210 Series Benchtop Meter User Manual |

29

Chapter 3

|

Meter Setup Menus

Mode and Settings Menu for pH Channel

Use the Mode and Settings menu to customize the measurement, calibration and alarm settings for the pH channel of Orion Star A211 pH meters, Orion Star A215 pH/conductivity meters and

Orion Star A216 pH/RDO/DO meters. Default meter settings are bold.

Parameter

Measure Mode

Read Type

Resolution

(pH only)

Buffer Group

(pH only)

Stability

Averaging

Limit Alarm

Cal Due Alarm

(pH, RmV and

ORP only)

Settings

·

pH

· mV

·

RmV

· ORP

·

Auto

· Continuous

·

1 Decimal Place

2 Decimal Places

3 Decimal Places

·

·

·

Smart Stability

· Fast

·

Medium

·

Slow

· Off

·

Automatic Smart

·

Limit Alarm

Off On

·

Alarm Settings

High High/Low Low

·

High Limit

·

Low Limit

·

·

Timed 00:00:03

USA

DIN

Cal Due Alarm

On Off

Cal Due Limit

12 Hrs

Details

Set the displayed measurement mode, the selected mode determines which additional settings are shown and the type of calibration the meter will perform

Set the read type to determine how the meter takes a measurement and when the measurement is sent to the data log and/or external device

0.1

0.01

0.001

Set the resolution of the pH measurement value

Set the buffer group for automatic buffer recognition during pH calibrations

USA: pH 1.68, 4.01, 7.00, 10.01 and 12.46 buffers

DIN: pH 1.68, 4.01, 6.86 and 9.18 buffers

Set when a measurement is recognized as stable,

Smart Stability compensates for measurement conditions and optimizes the meter response time

Set averaging for faster measurement stability,

Automatic Smart compensates for measurement conditions and optimizes the meter response time

Set the limit alarm off or on, when alarm is on – set the alarm setting as high, high/low or low and then enter the appropriate measurement limit values

The limit alarm is triggered when the measurement goes above the high limit or below the low limit

Set the calibration due alarm off or on, when the cal due alarm is on – enter the calibration time interval in hours (1 to 9998 hours)

The cal due alarm is triggered after the entered time has elapsed without a calibration being performed

30 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 3

|

Meter Setup Menus

Buffer Group Setting

Orion Star A211 pH meters, Orion Star A215 pH/conductivity meters and Orion Star A216 pH/RDO/DO meters are capable of automatically recognizing pH buffer values within the selected buffer set during a pH calibration. During the pH calibration, the meter uses the selected pH buffer set and the raw mV reading of the pH electrode in the buffer to recognize and display the buffer value at the measured temperature. The raw mV reading of the pH electrode in the buffer must be within one pH unit (approximately ±59 mV) of the buffer’s theoretical mV value for the meter to recognize the buffer.

pH Buffer

1.68

4.01

7.00

10.01

12.46

USA Buffer Set mV range

+255 mV to +374 mV

+117 mV to +236 mV

-59 mV to +59 mV

-237 mV to -119 mV

-382 mV to -264 mV

pH Buffer

1.68

4.01

6.86

9.18

DIN Buffer Set mV range

+255 mV to +374 mV

+117 mV to +236 mV

-51 mV to +67 mV

-189 mV to -70 mV

Testing a pH Electrode for Automatic Buffer Recognition

Use the following procedure to verify that the raw mV reading of the pH electrode is within one pH unit (±59 mV) from the theoretical mV reading of the pH buffer, and therefore verify that the pH electrode in use is capable of performing automatic buffer recognition.

1. Prepare the pH electrode according to the electrode manual. Set the meter measurement mode to mV.

2. Rinse the pH electrode with distilled water, blot it dry with a lint-free tissue and place the pH electrode into a pH 4.01 buffer at approximately 25 °C.

3. Wait for the measurement to stabilize and record the mV value of the pH 4.01 buffer when the measurement is stable.

4. Remove the pH electrode from the pH 4.01 buffer.

5. Rinse the pH electrode with distilled water, blot it dry with a lint-free tissue and place the pH electrode into a pH 7.00 buffer at approximately 25 °C.

6. Wait for the measurement to stabilize and record the mV value of the pH 7.00 buffer when the measurement is stable.

7. The mV reading of the pH electrode in pH 4 buffer should be +117 to +236 mV and in pH 7 buffer should be -59 to +59 mV. If the mV readings are in the correct ranges, the pH electrode is capable of performing automatic buffer recognition. If the mV readings are not in the correct ranges, manually enter the pH buffer values during a pH calibration.

Orion Star A210 Series Benchtop Meter User Manual |

31

Chapter 3

|

Meter Setup Menus

Mode and Settings Menu for pH/ISE Channel

Use the Mode and Settings menu to customize the measurement, calibration and alarm settings for the pH/ISE channel of Orion Star A214 pH/ISE meters. Default meter settings are bold.

Parameter

Measure Mode

Read Type

Resolution

(pH only)

Buffer Group

(pH only)

Resolution

(ISE only)

Measure Unit

(ISE only)

Blank Correct

(ISE only)

Stability

Averaging

Limit Alarm

Cal Due Alarm

(pH, RmV, ORP and ISE only)

Settings

·

pH

· mV

· RmV

·

·

ORP

ISE

·

Auto

·

Continuous

· Timed 00:00:03

1 Decimal Place

2 Decimal Places

3 Decimal Places

Details

Set the displayed measurement mode, the selected mode determines which additional settings are shown and the type of calibration the meter will perform

Set the read type to determine how the meter takes a measurement and when the measurement is sent to the data log and/or external device

·

·

USA

DIN

0.1

0.01

0.001

Set the resolution of the pH measurement value

Set the buffer group for automatic buffer recognition during pH calibrations

USA: pH 1.68, 4.01, 7.00, 10.01 and 12.46 buffers

DIN: pH 1.68, 4.01, 6.86 and 9.18 buffers

1 significant digit

2 significant digits

3 significant digits

· ppm

·

M

· mg/L

·

·

·

Percentage (%)

ppb

None

Set the resolution of the ion concentration (ISE) measurement value

Set the units displayed with the ion concentration

(ISE) measurement value

·

·

Yes

No

Set the automatic blank correction function off or on for ISE measurements, when the function is on – an algorithm is used to compensate for the non-linear response of the electrode in low level standards and samples for improved measurement stability

·

Smart Stability

·

Fast

·

Medium

· Slow

Set when a measurement is recognized as stable,

Smart Stability compensates for measurement conditions and optimizes the meter response time

·

·

Off

Automatic Smart

Set averaging for faster measurement stability,

Automatic Smart compensates for measurement conditions and optimizes the meter response time

·

Limit Alarm

Off On

· Alarm Settings

High High/Low Low

· High Limit

·

Low Limit

Set the limit alarm off or on, when alarm is on – set the alarm setting as high, high/low or low and then enter the appropriate measurement limit values

The limit alarm is triggered when the measurement goes above the high limit or below the low limit

·

·

Cal Due Alarm

On Off

Cal Due Limit

12 Hrs

Set the calibration due alarm off or on, when the cal due alarm is on – enter the calibration time interval in hours (1 to 9998 hours)

The cal due alarm is triggered after the entered time has elapsed without a calibration being performed

Set the low level stability function off or on for ISE calibrations, when the function is on – the minimum stabilization time is increased (~3-5 minutes) to improve accuracy in low level calibration standards

32 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Chapter 3

|

Meter Setup Menus

Buffer Group Setting

Orion Star A214 pH/ISE meters are capable of automatically recognizing pH buffer values within the selected buffer set during a pH calibration. During the pH calibration, the meter uses the selected pH buffer set and the raw mV reading of the pH electrode in the buffer to recognize and display the buffer value at the measured temperature. The raw mV reading of the pH electrode in the buffer must be within one pH unit (approximately ±59 mV) of the buffer’s theoretical mV value for the meter to recognize the buffer.

Blank Correct Setting

Orion Star A214 pH/ISE meters offer the option to use blank (non-linear) correction for ISE measurements when a multi-point calibration is performed. The automatic blank correction feature uses an algorithm to compensate for the non-linearity of an ion selective electrode in low level standards and samples.

With the blank correction feature enabled, the meter decides whether blank correction is the best measurement strategy by analyzing the electrode response during a multi-point calibration.

A separate blank does not have to be run. Graphically, blank correction is equivalent to drawing a smooth curve through the lowest three points of the multi-point calibration and extrapolating to zero concentration based on the assumption of Nernstian electrode behavior.

Multi-point calibrations at the lower limit of detection are desirable when the response of an ion selective electrode is non-linear and cannot be characterized with a one or two point calibration.

This is usually seen as a low electrode slope. Generally, the electrode is behaving in a

Nernstian manner but the effect of a blank is being observed. See the figure below.

Thermo Scientific

This blank may be a true reagent blank, traces of analate ion in the reagents or it may be the

“mud” value of the electrode. It could also be an interference in the reagents that becomes apparent at low levels of analate ion, or it could be any combination of these effects. The expanded version of the Nernst equation traditionally used for blank correction is as follows:

E = E

0

+ S * log (C + b) where b is the blank

Orion Star A210 Series Benchtop Meter User Manual |

33

Chapter 3

|

Meter Setup Menus

In a multi-point calibration, a set of equations is generated and the relationship between them evaluated. For example, the equations generated in a three point calibration would be:

E

1

= E

0

+ S * log (C

1

+ b)

E

2

= E

0

+ S * log (C

2

+ b)

E

3

= E

0

+ S * log (C

3

+ b)

The meter evaluates the relationships between the three potentials E

1

, E

2

and E

3

and the three concentrations C

1

, C

2

and C

3

. If the relationships indicate that blank correction is desired, a blank will automatically be calculated and the non-linearity will be corrected for in a Nernstian manner. If the appropriate conditions are not met, the blank is set at zero and each segment of the multi-point calibration is treated independently.

When all three of the following conditions are met, blank correction is invoked.

1. The concentration of the first standard is zero, or the slope of the electrode between the first and second standards is less than the slope between the second and third.

2. Potential differences between points are significant. For example, E

3

- E

1

> 10 mV

3. The blank correction algorithm converges at reasonable blank and slope values. Conditions in steps 1 and 2 prevent failure to converge in most situations. However, the slope will attain any value necessary for convergence to a calculated blank value of 3 x C

3

.

When blank correction is implemented, the slope value for the electrode may be outside the range of values normally considered acceptable during an ordinary calibration. For best results, calibration standards should be close in range to the expected sample concentrations and should bracket the expected sample concentration. When conditions 1, 2 and 3 are not met, the calibration data is handled by the multi-point calibration method described earlier.

In calibrations with more than three points, a combination of methods is utilized. Automatic blank correction is used if the lowest three points satisfy the criteria and multi-point calibration is used for the other points. The slope calculated in the automatic blank correction algorithm and the slopes for each additional segment are used to calculate the average slope.

Low Level Stability Setting

Orion Star A214 pH/ISE meters offer the option to use the low level stability feature when calibrating ion selective electrodes. The low level stability feature improves the accuracy of low concentration ISE measurements by adjusting the timing for calibration points of low level standards, allowing a longer stabilization time for the electrode in the calibration standards. The stabilization time is typically extended to about three to five minutes per calibration point, but will vary based on the actual readings from the electrode during calibration.

34 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 3

|

Meter Setup Menus

Mode and Settings Menu for COND Channel

Use the Mode and Settings menu to customize the measurement, calibration and alarm settings for the conductivity channel of Orion Star A212 conductivity meters and Orion Star A215 pH/conductivity meters. Default meter settings are bold.

Parameter

Measure Mode

Read Type

Settings

·

Conductivity

· Salinity

·

TDS

·

Resistivity

·

Auto

·

Continuous

·

Timed 00:00:03

Cell K

Details

Set the displayed measurement mode, the selected mode determines which additional settings are shown and the type of calibration the meter will perform

Set the read type to determine how the meter takes a measurement and when the measurement is sent to the data log and/or external device

Cell K

Type

(Salinity only)

TDS Factor

(TDS only)

·

·

Practical Salinity

Sea Water

TDS Factor

Set the type of salinity measurement as practical salinity

(psu) or natural sea water (ppt)

Ref. Temp.

Temp. Comp.

· 5 °C

·

10°C

·

15°C

·

20°C

·

25°C

·

Off

·

Linear

· nLFn

· nLFu

·

EP

Set the reference temperature for temperature compensated measurements, readings are adjusted to the selected reference temperature when the temperature compensation setting is active

Set the temperature compensation type as linear, nLFn

(non linear natural water), nLFu (non linear ultra pure water), EP (compensation off with alarm when values are outside EP requirements for ultra pure water) or off

Temp. Coeff. Temp. Coefficient 2.10 Enter the temperature coefficient used with the linear

Stability

·

Smart Stability

·

Fast

· Medium

·

Slow

Set when a measurement is recognized as stable, Smart

Stability compensates for measurement conditions and optimizes the meter response time

Averaging

·

·

Off

Automatic Smart

Set averaging for faster measurement stability, Automatic

Smart compensates for measurement conditions and optimizes the meter response time

Limit Alarm

·

Limit Alarm

Off On

· Alarm Settings

High High/Low Low

·

High Limit

·

Low Limit

Set the limit alarm off or on, when alarm is on – set the alarm setting as high, high/low or low and then enter the appropriate measurement limit values

The limit alarm is triggered when the measurement goes above the high limit or below the low limit

Cal Due Alarm

(Conductivity only)

Cell Type

·

·

Cal Due Limit

12 Hrs

·

·

Cal Due Alarm

On Off

Standard

USP

Set the calibration due alarm off or on, when on – enter the calibration interval in hours (1 to 9998 hrs)

The calibration due alarm is triggered after the set time has elapsed without performing a calibration

Set the type of conductivity senor as standard (most sensors) or USP (2-cell sensors in ultra pure water)

Orion Star A210 Series Benchtop Meter User Manual |

35

Chapter 3

|

Meter Setup Menus

Cell K Setting

Orion Star A212 conductivity meters and Orion Star A215 pH/conductivity meters are capable of automatically recognizing Thermo Scientific Orion 100 μS conductivity standard, 1413 μS conductivity standard and 12.9 mS conductivity standard when the nominal cell constant (K) value of the conductivity sensor is entered in the setup menu.

Catalog Number

011008

011007

01100710

011006

01100610

Description

Orion 100 μS conductivity standard, 5 x 60 mL

Orion 1413 μS conductivity standard, 5 x 60 mL

Orion 1413 μS conductivity standard, 10 individual use pouches

Orion 12.9 mS conductivity standard, 5 x 60 mL

Orion 12.9 mS conductivity standard, 10 individual use pouches

The nominal cell constant (K) values for Thermo Scientific Orion conductivity sensors compatible with Star A210 series conductivity meters are listed below.

Catalog

Number

Description Measurement Range

Nominal Cell

Constant

1 μS/cm to 200 mS/cm 0.475 cm

-1

1 μS/cm to 200 mS/cm 0.475 cm

-1

1 μS/cm to 200 mS/cm 0.475 cm

-1

1 μS/cm to 200 mS/cm 0.475 cm

-1

10 μS/cm to 200 mS/cm 0.55 cm

-1

10 μS/cm to 200 mS/cm 0.55 cm

-1

0.01 μS/cm to 300 μS/cm 0.1 cm

-1

10 μS/cm to 200 mS/cm 1.0 cm

-1

1 μS/cm to 20 mS/cm 1.0 cm

-1

10 μS/cm to 2000 mS/cm 10 cm

-1

36 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 3

|

Meter Setup Menus

Reference Temperature, Temperature Compensation and Coefficient Settings

Orion Star A212 conductivity meters and Orion Star A215 pH/conductivity meters offer several options for temperature compensated conductivity measurements. Temperature has a large effect on the conductivity value of a solution. The temperature compensation feature allows the meter to use the sample conductivity and temperature readings to calculate and display the conductivity that the sample would be expected to have at a selected reference temperature, such as 25 °C. To accurately use the temperature compensation feature, select the proper type of temperature compensation for the samples to be measured; set the required reference temperature as 5 °C, 10 °C, 15 °C, 20 °C or 25 °C; and use a conductivity sensor with built-in

ATC temperature probe. The options for temperature compensation are:

·

Linear – uses the temperature coefficient value to apply a constant percent correction factor to every degree change in temperature. A few common linear coefficient values are:

Solution (25 °C to 50 °C)

Sugar Syrup

Ultra Pure Water

98% Sulfuric Acid

Salt (NaCl)

Dilute Ammonia

5% NaOH

10% HCl

5% Sulfuric Acid

Temperature Coefficient (% / °C)

5.64

4.55

2.84

2.12

1.88

1.72

1.32

0.96

· nLFn (non-linear ultra pure non-degassed water) – applies a non-constant correction factor to pure water samples that have a varying response to changes in temperature. This mode is for temperature compensated readings of low conductivity waters that are in equilibrium with the carbon dioxide in air, such as pure waters reading near 1 μS/cm at 25 °C and natural waters having a composition comparable to natural ground, well or surface waters.

· nLFu (non-linear ultra pure degassed water) – applies a non-constant correction factor to pure water samples that have a varying response to changes in temperature. This mode is for temperature compensated readings of ultra pure water that contains no air and no carbon dioxide, such as ultra pure water (18 megohms resistance or higher) directly from the source without aeration.

· EP – no temperature correction is applied (temperature compensation turned off) and a warning is displayed if the measured conductivity value is over the EP requirement for pure water at the measured sample temperature.

·

Off - no temperature correction is applied and the actual conductivity value is displayed at the measured sample temperature.

Orion Star A210 Series Benchtop Meter User Manual |

37

Chapter 3

|

Meter Setup Menus

Salinity Type Setting

Orion Star A212 conductivity meters and Orion Star A215 pH/conductivity meters offer two salinity types: practical salinity and natural sea water. Practical salinity measurements are based on relating the sample measurement to the reading of a standard potassium chloride

(KCl) solution at 15 °C. Orion Star A210 series conductivity meters will automatically temperature compensate and report the expected practical salinity result as practical salinity units (psu) at 15 °C when using a conductivity sensor with built-in temperature. Natural sea water measurements use a historic convention known as UNESCO 1966 and Orion Star A210 series conductivity meters will report the expected natural sea water result as parts per thousand (ppt).

TDS Factor Setting

Orion Star A212 conductivity meters and Orion Star A215 pH/conductivity meters measure TDS as the total amount of dissolved inorganic material in a solution. The dissolved inorganic material carries a current that is measured by the conductivity sensor. Since there is a direct relationship between conductivity and TDS, conductivity readings are used to estimate the presence of inorganic material using the TDS factor entered in the setup menu.

The standard method of determining TDS involves evaporating a sample to dryness at 180 °C and weighing the residue. The TDS factor is calculated by taking the residue weight and dividing it by the sample conductivity. Subsequent conductivity readings are multiplied by the

TDS factor to determine the TDS value of the sample.

Cell Type Setting

Orion Star A212 conductivity meters and Orion Star A215 pH/conductivity meters accept 2-cell and 4-cell conductivity sensors and allows the type of conductivity sensor to be set as Standard

(most conductivity sensors) or USP (for the ultra pure water conductivity sensor, Catalog

Number 013016MD, when disabling temperature compensation).

38 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 3

|

Meter Setup Menus

Mode and Settings Menu for DO/RDO Channel

Use the Mode and Settings menu to customize the measurement, calibration and alarm settings for the dissolved oxygen channel of Orion Star A213 RDO/DO meters and Orion Star A216 pH/RDO/DO meters. Default meter settings are bold.

Parameter

Measure Mode

Measure Unit

Resolution

(% saturation)

Resolution

(mg/L)

Read Type

Baro Press

Salinity Correct

Stability

Averaging

Limit Alarm

Cal Due Alarm

RDO Cap Life

(RDO only)

Settings

·

·

·

·

·

·

·

·

DO – Polarographic

RDO

Auto mg/L

% sat

1 Decimal Place

2 Decimal Places

1 Decimal Place

2 Decimal Places

·

Auto

· Continuous

· Timed 00:00:03

Auto

Manual 760.0mmHg

Manual 0.0

1

0.1

0.1

0.01

Details

Set the type of dissolved oxygen sensor connected to the meter as DO polarographic or RDO optical or select Auto and the meter will automatically detect which type of dissolved oxygen sensor is connected and update the displayed measurement mode

Set the units displayed with the dissolved oxygen measurement value

Set the resolution of the dissolved oxygen measurement value as percent saturation

Set the resolution of the dissolved oxygen measurement value as mg/L

Set the read type to determine how the meter takes a measurement and when the measurement is sent to the data log and/or external device

Set the automatic barometric pressure compensation source as the internal barometer (Auto) or a manually entered barometric pressure value (Manual)

Enter the salinity value of samples for automatic salinity correction of dissolved oxygen measurements

·

Smart Stability

·

Fast

·

Medium

· Slow

Set when a measurement is recognized as stable,

Smart Stability compensates for measurement conditions and optimizes the meter response time

·

·

· Limit Alarm

Off On

·

Alarm Settings

High High/Low Low

· High Limit

·

Low Limit

·

·

Off

Automatic Smart

Cal Due Alarm

On Off

Cal Due Limit

12 Hrs

RDO Cap Life

Serial Number

0.0

0

Set averaging for faster measurement stability,

Automatic Smart compensates for measurement conditions and optimizes the meter response time

Set the limit alarm off or on, when the limit alarm is on – select the alarm settings as high, high/low or low and then enter the measurement limit values

The limit alarm is triggered when a measurement goes above the entered high limit or below the entered low limit

Set the calibration due alarm off or on, when the cal due alarm is on – enter the calibration time interval in hours (1 to 9998 hours)

The cal due alarm is triggered after the entered time has elapsed without a calibration being performed

View the remaining RDO optical DO sensor cap life in days and the serial number

Orion Star A210 Series Benchtop Meter User Manual |

39

Chapter 3

|

Meter Setup Menus

Barometric Pressure Compensation Setting

Orion Star A213 RDO/DO meters and Orion Star A216 pH/RDO/DO meters have an internal barometer that can be used for automatic pressure compensated dissolved oxygen readings.

Manual entry of the barometric pressure can be used for automatic pressure compensated dissolved oxygen readings, which is useful when dissolved oxygen is measured with a submerged sensor or in a pressurized vessel. The pressure value must be entered as mm Hg.

1 mm Hg = 0.03937 inch Hg = 1.3332 hPa (mBar) = 0.01934 PSI

Salinity Correction Setting

Orion Star A213 RDO/DO meters and Orion Star A216 pH/RDO/DO meters are capable of performing automatic salinity correction for dissolved oxygen readings (measured as mg/L) when the salinity value of the sample is manually entered as ppt (parts per thousand) in the setup menu.

13

14

15

10

11

12

16

17

18

19

Conductivity at 20 °C

(mS/cm)

5

8

9

6

7

8

9

10

6

7

8

10

11

12

13

Salinity

Correction

Value (ppt)

3

5

6

4

4

28

29

30

25

26

27

31

32

33

34

Conductivity at 20 °C

(mS/cm)

20

21

22

23

24

46

48

50

40

42

44

52

54

56

Conductivity at 20 °C

(mS/cm)

35

36

37

38

39

19

20

21

17

18

18

22

22

23

24

Salinity

Correction

Value (ppt)

13

14

15

15

16

Calculated from the International Oceanographic Tables, Vol. 1, National Institute of

Oceanography of Great Britain, Womley, Godaming, Surrey, England and Unesco, Paris 1971

33

35

37

29

30

32

38

40

42

Salinity

Correction

Value (ppt)

25

25

26

27

28

40 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 3

|

Meter Setup Menus

Temperature Menu

Orion Star A210 series benchtop meters offer a Temperature menu used to customize temperature settings for each channel by manually entering a temperature value, selecting the temperature units as °C or °F and performing a temperature calibration on an ATC probe, conductivity sensor or dissolved oxygen sensor with built-in temperature. Orion Star A216 pH/RDO/DO meters also have the option to use one temperature input source for both measurement channels.

Meter Model

Star A211 pH Meter

Star A212

Conductivity Meter

Star A213 RDO/DO

Meter

Star A214 pH/ISE

Meter

Star A215 pH/

Conductivity Meter

Channel

pH Channel

COND

Channel

DO/RDO

Channel

Temperature Menu

·

Manual Temp Value

·

Temperature Unit

·

Temperature Calibration

·

Manual Temp Value

·

Temperature Unit

·

Temperature Calibration

·

Temperature Unit

· Temperature Calibration

(polarographic DO sensor only)

pH/ISE

Channel pH Channel

COND

Channel

·

Manual Temp Value

·

Temperature Unit

·

Temperature Calibration

·

Manual Temp Value

·

Temperature Unit

·

Temperature Calibration

· Manual Temp Value

· Temperature Unit

· Temperature Calibration

· Manual Temp Value

· Temperature Unit

Star A216 pH/RDO/DO

Meter pH Channel

·

Temperature Calibration

DO/RDO

Channel

· Temperature Input

·

Temperature Unit

· Temperature Calibration

(polarographic DO sensor only)

Menu Options

Enter temperature value

·

Celsius

·

ATC

·

Fahrenheit

Enter temperature value

·

Celsius

·

Fahrenheit

·

ATC

·

Celsius

·

Solution Temperature

· Membrane Temperature

· Sol & Mem Temp

Enter temperature value

·

Celsius

·

Fahrenheit

·

ATC

Enter temperature value

·

Celsius

·

Fahrenheit

·

ATC

Enter temperature value

· Celsius

· ATC

· Fahrenheit

Enter temperature value

· Celsius · Fahrenheit

·

ATC

·

Solution Temperature

·

Membrane Temperature

· Sol & Mem Temperature

· ATC, Sol & Mem Temp

· ATC

· DO Probe

·

Celsius

· Manual

·

Fahrenheit

·

Solution Temperature

·

Membrane Temperature

·

Sol & Mem Temp

Orion Star A210 Series Benchtop Meter User Manual |

41

Chapter 3

|

Meter Setup Menus

Manual Temperature Value

1. In the measurement mode, press the setup key.

2. Press the p, q, t or u key to highlight pH Channel, pH/ISE Channel or COND Channel and press the f3 (select) key.

3. Press the p or q key to highlight Temperature and press the f3 (select) key.

4. Press the p or q key to highlight Manual Temp Value and press the f3 (select) key.

5. Press the f3 (edit) key access the popup number entry screen and enter the sample temperature value. a. Press the p, q, t or u key to highlight a number, decimal point or negative sign and then press the f3 (enter) key to select the highlighted item. Repeat until the sample temperature value is shown at the top of the screen. b. Press the f2 (done) key to save the value and exit the number entry screen.

0 1 2 .

3 4 5 6

7 8 9 – backspace

back done enter

6. Press the f1 (back) key to return to the main temperature menu and press the measure

(esc) key to return to the measurement mode.

Temperature Unit

1. In the measurement mode, press the setup key.

2. Press the p, q, t or u key to highlight pH Channel, pH/ISE Channel, COND Channel or

DO/RDO Channel and press the f3 (select) key.

3. Press the p or q key to highlight Temperature and press the f3 (select) key.

4. Press the p or q key to highlight Temperature Unit and press the f3 (select) key.

5. Press the p or q key to highlight Celsius or Fahrenheit and press the f3 (select) key.

6. Press the measure (esc) key to return to the measurement mode.

42 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 3

|

Meter Setup Menus

Temperature Calibration

When using Orion Star A210 series benchtop meters, a temperature calibration can be performed on an ATC probe, conductivity sensor with built-in temperature or polarographic dissolved oxygen sensor with built-in temperature when the probe or sensor is connected to the meter. Use the temperature calibration function only when necessary, since the meter has a relative temperature accuracy of ±0.1°C and ATC probes have varying accuracies, usually

±0.5°C to ±2°C. Since the temperature offset calculated during the calibration is applied to all future temperature measurements, recalibrate the temperature if a different ATC probe is used.

1. In the measurement mode, press the setup key.

2. Press the p, q, t or u key to highlight pH Channel, pH/ISE Channel, COND Channel or

DO/RDO Channel and press the f3 (select) key.

3. Press the p or q key to highlight Temperature and press the f3 (select) key.

4. Press the p or q key to highlight Temperature Calibration and press the f3 (select) key.

5. Press the p or q key to highlight ATC and press the f3 (select) key. a. If using the Star A213 meter, press the p or q key to highlight Solution Temperature,

Membrane Temperature or Sol & Mem Temperature and press the f3 (select) key. b. If using the Star A216 meter, press the p or q key to highlight ATC, Solution

Temperature, Membrane Temperature, Sol & Mem Temperature or ATC, Sol & Mem

Temp and press the f3 (select) key.

6. Place the ATC probe and/or sensor into a solution with a known, stable temperature. Two

NIST-traceable thermometers should be used to measure and verify the temperature.

7. Wait for the temperature value on the meter to stabilize and then press the f3 (edit) key to access the popup number entry screen and enter the temperature value. a. Press the p, q, t or u key to highlight a number, decimal point or negative sign and then press the f3 (enter) key to select the highlighted item. Repeat until the desired temperature value is shown at the top of the screen. b. Press the f2 (done) key to exit the number entry screen.

8. Press the f2 (accept) key to save the entered temperature value.

9. Press the f1 (meas) key to return to the measurement mode.

Temperature Calibration for Meters with Multiple Temperature Inputs

When a polarographic dissolved oxygen sensor with two temperature inputs is connected with an Orion Star A213 RDO/DO meter or Orion Star A216 pH/RDO/DO meter, the solution temperature input and membrane temperature input can be calibrated separately or together.

When an ATC probe and polarographic dissolved oxygen sensor are both connected to the

Orion Star A216 pH/RDO/DO meter, the temperature calibration for both inputs can also be performed together.

Orion Star A210 Series Benchtop Meter User Manual |

43

Chapter 3

|

Meter Setup Menus

Temperature Input

When using the Orion Star A216 pH/RDO/DO meter, a dissolved oxygen sensor with built-in temperature can be set as the temperature input source for all measurements. For example, when using a pH electrode without an ATC probe, the dissolved oxygen sensor with built-in temperature can be used as the temperature source for pH measurements, as long as both the pH electrode and dissolved oxygen sensor are placed in the same solution.

1. In the measurement mode, press the setup key.

2. Press the p, q, t or u key to highlight pH Channel and press the f3 (select) key.

3. Press the p or q key to highlight Temperature and press the f3 (select) key.

4. Press the p or q key to highlight Temperature Input and press the f3 (select) key.

5. Press the p or q key to highlight ATC, DO Probe or Manual and press the f3 (select) key.

6. Press the measure (esc) key to return to the measurement mode.

44 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 3

|

Meter Setup Menus

Instrument Settings Setup Menu

Use the Instrument Settings menu to update meter settings for data transfer, data log, date and time, language, sounds, stirrer speed, display contrast, auto-shutoff, user ID and sample ID.

Instrument

Setting

Export Data

Data Log

Date and Time

Language

Options

·

Printing

Off On

· Comm Setup

RS232 USB

·

Data Format

Printer PC (CSV)

· Comm Config

1200 9600

2400 19200

4800 38400

·

Off

·

On

·

Date

DD/MM/YY

MM/DD/YY

Date: 01/01/12

· Time

12 Hour Clock

Time: 02:30 PM

24 Hour Clock

Time: 14:30

·

English

· Español

· Deutsch

·

Français

·

Italiano

· 中文

(Chinese)

Details

Access the data export menu to set data transfer settings for exporting measurement and calibration information to a printer or computer

Turn printing on to allow data transfer from the meter to an external device, select the data transfer format as printer (standard text) or PC (comma delimited text) and set the meter baud rate to match the printer or computer baud rate

The meter will automatically detect when an RS232 or

USB cable is connected to a meter port

Turn data logging on to save measurement data to the data log according to the selected read type

Use the date setting to select the date format as day/month/year (DD/MM/YY) or month/day/year

(MM/DD/YY) and then enter the month, day and year values

Use the time setting to set the time format as 12 hour clock or 24 hour clock and then enter the hour and minute values

Set the language to be used for the meter interface, additional languages (including Portuguese and

Korean) are available with software updates on our website at www.thermoscientific.com/OrionMeters

Alarm Beep

Stirrer

Contrast

Auto Shut Off

User ID

Sample ID

·

Off

· On

·

Off

·

On

ABCDE

·

Off

·

Manual

·

Auto Increment

Turn the audible beep on or off each time a meter key is pressed

Turn the audible beep on or off when an alarm is triggered

Adjust the stirrer speed setting from one (slowest) to five (fastest) when using an Orion Star stirrer probe

Adjust the display contrast higher or lower to improve display viewing in different light conditions

Turn on function to automatically power off the meter when no keys are pressed for 20 minutes

Enter a user ID value (up to six characters) using the alphanumeric entry screen

Set sample ID as off, manual (up to six alphanumeric characters) or auto incremental (up to six numbers that are automatically increased for each measurement)

Orion Star A210 Series Benchtop Meter User Manual |

45

Chapter 3

|

Meter Setup Menus

Log View Menu

Use the Log View menu to access the data log and calibration log. All information in the data and calibration logs is protected by the meter’s nonvolatile memory, so stored information is saved in the meter even when it is not connected to a power supply.

Data Log

Orion Star A210 series benchtop meters offer a 2000 point data log. Each point includes the actively displayed measurements, depending on the meter model and display configuration, with date and time stamp. When the data log function is turned on, the read type selected for each displayed channel (Auto-Read, Continuous or Timed) determines how the point is saved to the data log.

1. In the measurement mode, press the setup key.

2. Press the p, q, t or u key to highlight View Log and press the f3 (select) key.

3. Press the p or q key to highlight Data Log and press the f2 (select) key.

4. The meter will display a list of the data log points. The list shows the sequential number of the points and the data and time the point was saved.

5. To view measurement information for an individual point: a. Press the p or q key to highlight a point and then press the f2 (select) key. b. Press the f1 (back) key to return to the list of data log points.

6. To export the data log to a printer or computer: a. Press the f3 (options) key, press the p or q key to highlight Log Export and then press the f2 (accept) key. b. Press the p or q key to highlight Today, Last 10, Range or All and then press the f2

(accept) key.

i. If Range is selected, press the p or q key to highlight the number of the first

(from) and last (to) points to be exported, use the popup number entry screen to edit the numbers and press the f2 (accept) key to save the entered range. c. Press the f2 (yes) key to confirm that the meter is interfaced with a printer or computer.

The meter will automatically return to the data log list once the export is completed.

7. To clear the data log memory: a. Press the f3 (options) key, press the p or q key to highlight Log Clear and then press the f2 (accept) key. b. Press the f2 (yes) key to confirm erasing the entire data log from the meter’s memory. c. Press the f3 (edit) key, use the popup number entry screen to enter the default meter password of 111111 and then press the f2 (done) key. d. Press the f2 (accept) key and wait for the data log to be cleared.

46 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Exported Calibration Log Example,

Printer Data Format:

Chapter 3

|

Meter Setup Menus

Calibration Log

Orion Star A210 series meters save up to ten of the most recent calibrations per channel.

1. In the measurement mode, press the setup key.

2. Press the p, q, t or u key to highlight View Log and press the f3 (select) key.

3. Press the p or q key to highlight Calibration Log and press the f2 (select) key.

4. Press the p or q key to highlight pH–Channel, pH/ISE–Channel, Conductivity–Channel or

DO/RDO–Channel and press the f2 (select) key. a. If pH–Channel is selected, press the p or q key to highlight pH, RmV or ORP and press the f2 (select) key. b. If pH/ISE–Channel is selected, press the p or q key to highlight pH, RmV, ORP or

ISE and press the f2 (select) key. c. If Conductivity–Channel is selected, press the p or q key to highlight Conductivity,

Resistivity, TDS or Salinity and press the f2 (select) key. d. If DO/RDO–Channel is selected, press the p or q key to highlight DO or RDO and press the f2 (select) key.

5. The meter will display a list of calibrations for the selected channel and calibration mode.

The list chows the sequential number of the calibration log and the data and time the calibration was completed.

6. To view detailed information for an individual calibration, press the p or q key to highlight a calibration and then press the f2 (select) key. a. To export the calibration to a printer or computer, press the f2 (print) key.

Thermo Scientific Orion Star A210 Series Benchtop Meter User Manual |

47

Chapter 3

|

Meter Setup Menus

Diagnostics Menu

Use the Diagnostics Menu to initiate a meter reset, test electrode stability, perform a meter self test or view meter and module serial numbers and software revisions.

· Factory Reset – All meter settings are reset to factory defaults. Data log, calibration log and methods are deleted. Use the default meter password of 111111 to complete the reset.

· User Reset – All channel-specific measurement, calibration and alarm settings as well as all instrument settings are reset to factory defaults and methods are erased. Data log and calibration log are retained.

·

Self Test – Tests the meter accuracy. A self test is also performed automatically by the meter at each power up.

·

Stability Test – Tests the stability of an electrode by measuring the raw input reading per minute drift and noise.

·

About Meter – Displays the meter model, serial number and software revision.

Meter Self Test Procedure

1. In the measurement mode, press the setup key.

2. Press the p, q, t or u key to highlight Diagnostics and press the f3 (select) key.

3. Press the p or q key to highlight Self Test and press the f2 (select) key.

4. Disconnect all electrodes and sensors from the meter, attach the BNC shorting cap to the

BNC input if applicable and press the f2 (yes) key.

5. Press the f2 (yes) key to start the self test.

6. The meter will perform the self test. Once Self Test Passed is displayed, press the f1 (esc) key and the meter will automatically proceed to the measurement mode.

Meter Verification Procedure (pH and pH/ISE Channels Only)

1. After completing the self test, the meter will return to the measurement mode. Keep all of the electrodes and sensors disconnected from the meter and keep the BNC shorting cap attached to the BNC input on the meter.

2. Set the measurement mode to pH. The meter should read a steady 7.000 pH ±0.002 pH. a. If the meter does not read a steady 7.000 ±0.002, perform a one point pH calibration with the BNC shorting cap attached to the BNC input. Set the pH value to 7.000 and the slope value to 100.0. Refer to the pH Calibration section for detailed instructions.

3. Set the measurement mode to mV. The meter should read a steady 0.0 mV ±0.2 mV. a. If the meter does not read a steady 0.0 mV ±0.2 mV with the BNC shorting cap attached to the BNC input, contact Technical Support.

48 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 3

|

Meter Setup Menus

Electrode Stability Test Procedure

1. In the measurement mode, press the setup key.

2. Press the p, q, t or u key to highlight Diagnostics and press the f3 (select) key.

3. Press the p or q key to highlight Stability Test and press the f2 (select) key.

4. Press the p or q key to highlight pH–Channel, pH/ISE–Channel, Conductivity–Channel or

DO/RDO–Channel and press the f2 (next) key.

5. Make sure the electrode or sensor to be tested is connected to the channel selected in the previous step and press the f2 (yes) key.

6. Place the electrode or sensor in an appropriate solution with a stable temperature and press the f2 (yes) key. a. Thermo Scientific Orion pH 7.00 buffer (Catalog Number 910107, 910760 or 910710) is recommended for testing a pH electrode. b. Thermo Scientific Orion ORP standard (Catalog Number 967901 or 967961) is recommended for testing an ORP electrode. c. A calibration standard with a moderate concentration of the ion of interest is recommended for testing an ion selective electrode (ISE). d. Thermo Scientific Orion 1413 µS conductivity standard (Catalog Number 011007 or

01100710) is recommended for testing most conductivity sensors. Thermo Scientific

Orion 100 µS conductivity standard (Catalog Number 011008) is recommended for testing ultra pure water conductivity sensors. e. A prepared water-saturated air calibration sleeve is recommended for testing RDO optical and polarographic dissolved oxygen sensors.

7. The meter will perform the stability test and show the raw input reading per minute drift and noise on the display.

8. The meter will display Stability Test Passed or Stability Test Failed. Press the f1 (esc) key to return to the measurement mode.

Orion Star A210 Series Benchtop Meter User Manual |

49

C H A P T E R 4

4

Using the pH or pH/ISE Channel

Use the Orion Star A211 pH meter, Orion Star A215 pH/conductivity meter or Orion Star A216 pH/RDO/DO meter to measure pH, mV, relative mV, ORP and temperature. Use the Orion Star

A214 pH/ISE meter to measure pH, ion concentration using an ISE (ion selective electrode), mV, relative mV, ORP and temperature. The following chapter provides instructions on connecting electrodes, calibrating the system and measuring these parameters.

Meter and Electrode Preparation

1. Familiarize yourself with general meter operation and prepare the meter and accessories

according to the instructions in Chapter 2 Meter Basics .

a. It is recommended that electrodes be placed in the meter-attached electrode stand for easy movement in and out of containers during calibration, measurement and storage.

2. Refer to the instructions in Chapter 3 Meter Setup Menus to customize measurement,

calibration and alarm settings within the pH Channel or pH/ISE Channel setup menu and data transfer, data log, date and time, language, sounds, stirrer speed, display contrast, auto-shutoff, user ID and sample ID settings within the Instrument Settings setup menu.

3. Connect all necessary electrodes, probes and sensors to the appropriate meter inputs. a. Connect a pH, ORP or ion selective electrode to the BNC meter input. b. Connect an ATC temperature probe to the 8 pin MiniDIN meter input. c. Connect a reference half-cell electrode to the pin tip meter input (for use when a halfcell sensing electrode with BNC connector is used). d. Connect the stirrer probe (Catalog Number 096019) to the Stirrer meter input.

4. Prepare all connected electrodes for use as instructed in the electrode manuals.

50 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 4

|

Using the pH or pH/ISE Channel

pH Calibration Procedure

Orion Star A211 pH meters, Orion Star A214 pH/ISE meters, Orion Star A215 pH/conductivity meters and Orion Star A216 pH/RDO/DO meters can perform a pH calibration using one to five pH buffers with easy to follow onscreen instructions to assist with the calibration process.

Always use fresh pH buffers and select buffers that bracket the sample pH and are one to three pH units apart. Prepare the pH electrode according to the instructions in the electrode manual.

Connect all electrodes to be used to the appropriate meter inputs.

Make sure the meter measurement mode is set to pH. For Star A215 and Star A216 meters, press the f3 (channel) key until the pH channel is shown in the measurement mode.

5. Press the f1 (cal) key to start the calibration. a. If more than one channel is displayed in the measurement mode, press the p or q key to highlight pH-Channel and press the f2 (select) key.

6. Rinse the pH electrode and any other electrodes in use with distilled water, blot gently with a lint-free tissue to remove excess water and place into the pH buffer.

7. When the electrode and buffer are ready, press the f3 (start) key. a. If using a stirrer probe, the stirrer probe will start stirring when the f3 (start) key is pressed and stop stirring when the reading stabilizes.

8. Wait for the pH value on the meter to stabilize and perform one of the following actions: a. Press the f2 (accept) key to accept the pH buffer value determined by the automatic buffer recognition function. or b. Press the f3 (edit) key to access the popup number entry screen and manually enter the pH buffer value. i. Press the p, q, t or u key to highlight a number or decimal point, press the

f3 (enter) key to select the highlighted item and repeat until the desired value is

shown at the top of the screen. ii. Press the f2 (done) key to save the value and exit the number entry screen. iii. Press the f2 (accept) key to accept the manually entered value.

9. Press the f2 (next) key to proceed to the next pH buffer and repeat steps 2 through 4 or press the f3 (cal done) key to save and end the calibration. If five buffers are used, the calibration will save and end once the fifth value is accepted. a. When performing a one point calibration, press the f2 (accept) key to accept the displayed slope value or press the f3 (edit) key, use the popup number entry screen to enter a new slope value, press the f2 (done) key and then press the f2 (accept) key.

10. The meter will display the calibration summary including the slope and export the data to the calibration log. Press the f1 (meas) key to proceed to the measurement mode or press the

f2 (print) key to export the calibration data to a printer or computer.

Orion Star A210 Series Benchtop Meter User Manual |

51

Chapter 4

|

Using the pH or pH/ISE Channel

pH Calibration Editing

When the calibration summary is displayed, the calibration editing option can be selected to fix individual points without having to perform a full recalibration.

1. In the calibration summary display (after step 5 in the pH Calibration Procedure), press the

f3 (cal edit) key.

2. Press the p or q key to highlight a calibration point to edit and press the f2 (select) key.

3. Press the p or q key to highlight Remeasure, Edit or Delete and press the f2 (select) key. a. If Remeasure is selected, repeat steps 2 through 4 in the pH Calibration Procedure. b. If Edit is selected, press the f3 (edit) key, use the popup number entry screen to enter a new pH buffer value, press the f2 (done) key and then press the f2 (accept) key.

Select another calibration point to edit or press the f1 (back) key. c. If Delete is selected, the calibration point will be deleted. If deleting the point changes it to a one point calibration, enter the slope using the popup number entry screen.

4. The meter will display an updated calibration summary and export the data to the calibration log. Press the f1 (meas) key to proceed to the measurement mode or press the f2 (print) key to export the calibration data to a printer or computer.

52 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 4

|

Using the pH or pH/ISE Channel

ORP Calibration Procedure (Relative mV Mode)

Orion Star A211 pH meters, Orion Star A214 pH/ISE meters, Orion Star A215 pH/conductivity meters and Orion Star A216 pH/RDO/DO meters can perform an ORP calibration in the relative mV (RmV) mode using one standard with easy to follow onscreen instructions to assist with the calibration process.

Use the Thermo Scientific Orion ORP standard (Catalog Number 967901 or 967961) or other

ORP standard as desired. Prepare the electrode according to the instructions in the electrode manual. Connect all electrodes to be used to the appropriate meter inputs.

Make sure the meter measurement mode is set to RmV. For Star A215 and Star A216 meters, press the f3 (channel) key until the RmV channel is shown in the measurement mode.

1. Press the f1 (cal) key to start the calibration. a. If more than one channel is displayed in the measurement mode, press the p or q key to highlight pH-Channel and press the f2 (select) key.

2. Rinse the electrode and any other electrodes in use with distilled water, blot gently with a lint-free tissue to remove excess water and place into the standard.

3. When the electrode and standard are ready, press the f3 (start) key. a. If using a stirrer probe, the stirrer probe will start stirring when the f3 (start) key is pressed and stop stirring when the reading stabilizes.

4. Wait for the mV value on the meter to stabilize (indicated by the flashing stabilizing or solid

ready icon) and perform one of the following actions: a. Press the f2 (accept) key to accept the mV value. or b. Press the f3 (edit) key to access the popup number entry screen and manually enter the mV value expected at the measured temperature. i. Press the p, q, t or u key to highlight a number, decimal point or negative sign; press the f3 (enter) key to select the highlighted item and repeat until the desired value is shown at the top of the screen. ii. Press the f2 (done) key to save the value and exit the number entry screen. iii. Press the f2 (accept) key to accept the manually entered value.

5. The meter will display the calibration summary including the mV offset and export the data to the calibration log. Press the f1 (meas) key to proceed to the measurement mode or press the f2 (print) key to export the calibration data to a printer or computer.

Orion Star A210 Series Benchtop Meter User Manual |

53

Chapter 4

|

Using the pH or pH/ISE Channel

ORP Calibration Procedure (E

H

Units)

Orion Star A211 pH meters, Orion Star A214 pH/ISE meters, Orion Star A215 pH/conductivity meters and Orion Star A216 pH/RDO/DO meters can perform a one point automatic ORP calibration to the E

H

value (relative to the Standard Hydrogen Electrode, 420 mV at 25 °C) when using the following items:

· Thermo Scientific Orion ORP standard (Catalog Number 967901 or 967961)

·

Thermo Scientific Orion ORP electrode (Catalog Number 9678BNWP, 9778BNWP,

9180BNMD or 9179BNMD)

·

Thermo Scientific Orion 4 M KCl filling solution (Catalog Number 900011) or 4 M KCl gel

(in the 9179BNMD electrode)

Orion ORP standard is nonhazardous, stable and will not change over time, so it is an ideal standard for any ORP measurement system. Always use fresh ORP standard. Prepare the

ORP electrode according to the instructions in the electrode manual. Connect all electrodes to be used to the appropriate meter inputs.

Make sure the meter measurement mode is set to ORP. For Star A215 and Star A216 meters, press the f3 (channel) key until the ORP channel is shown in the measurement mode.

1. Press the f1 (cal) key to start the calibration. a. If more than one channel is displayed in the measurement mode, press the p or q key to highlight pH-Channel and press the f2 (select) key.

2. Rinse the electrode and any other electrodes in use with distilled water, blot gently with a lint-free tissue to remove excess water and place into the standard.

3. When the electrode and standard are ready, press the f3 (start) key. a. If using a stirrer probe, the stirrer probe will start stirring when the f3 (start) key is pressed and stop stirring when the reading stabilizes.

4. Wait for the mV value on the meter to stabilize (indicated by the flashing stabilizing or solid

ready icon) and press the f2 (accept) key once the mV value is stable.

5. The meter will display the calibration summary including the mV offset and export the data to the calibration log. Press the f1 (meas) key to proceed to the measurement mode or press the f2 (print) key to export the calibration data to a printer or computer.

54 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 4

|

Using the pH or pH/ISE Channel

ISE Calibration Procedure

Orion Star A214 pH/ISE meters can perform an ion concentration calibration using one to five standards with easy to follow onscreen instructions to assist with the calibration process.

Always use fresh standards and select standards that bracket the sample concentration and are one decade (ten times) apart in concentration. To accurately prepare calibration standards from a stock solution, serial dilution is recommended using calibrated pipettes. If ISA will be added to samples, the same ISA must be added to all standards prior to calibration to ensure a consistent dilution factor. If more than one standard is used for calibration, start with the lowest concentration standard and work up to the highest concentration standard last.

Prepare the ion selective electrode according to the instructions in the electrode manual.

Connect all electrodes to be used to the appropriate meter inputs. Make sure the meter measurement mode is set to ISE.

1. Press the f1 (cal) key to start the calibration.

2. Rinse the ion selective electrode and any other electrodes in use with distilled water, blot gently with a lint-free tissue to remove excess water and place into the standard.

3. When the electrode and standard are ready, press the f3 (start) key. a. If using a stirrer probe, the stirrer probe will start stirring when the f3 (start) key is pressed and stop stirring when the reading stabilizes.

4. Wait for the value on the meter to stabilize and perform one of the following actions: a. Press the f2 (accept) key to accept the displayed concentration value. or b. Press the f3 (edit) key to access the popup number entry screen and manually enter the concentration value. i. Press the p, q, t or u key to highlight a number or decimal point, press the

f3 (enter) key to select the highlighted item and repeat until the desired value is

shown at the top of the screen. ii. Press the f2 (done) key to save the value and exit the number entry screen. iii. Press the f2 (accept) key to accept the manually entered value.

5. Press the f2 (next) key to proceed to the next standard and repeat steps 2 through 4 or press the f3 (cal done) key to save and end the calibration. If five standards are used, the calibration will save and end once the fifth value is accepted. a. When performing a one point calibration, press the f2 (accept) key to accept the displayed slope value or press the f3 (edit) key, use the popup number entry screen to enter a new slope value, press the f2 (done) key and then press the f2 (accept) key.

6. The meter will display the calibration summary including the slope and export the data to the calibration log. Press the f1 (meas) key to proceed to the measurement mode or press the

f2 (print) key to export the calibration data to a printer or computer.

Orion Star A210 Series Benchtop Meter User Manual |

55

Chapter 4

|

Using the pH or pH/ISE Channel

ISE Calibration Editing

When the calibration summary is displayed, the calibration editing option can be selected to fix individual points without having to perform a full recalibration.

1. In the calibration summary display (after step 5 in the ISE Calibration Procedure), press the

f3 (cal edit) key.

2. Press the p or q key to highlight a calibration point to edit and press the f2 (select) key.

3. Press the p or q key to highlight Remeasure, Edit or Delete and press the f2 (select) key. a. If Remeasure is selected, repeat steps 2 through 4 in the ISE Calibration Procedure. b. If Edit is selected, press the f3 (edit) key, use the popup number entry screen to enter a new standard value, press the f2 (done) key and then press the f2 (accept) key.

Select another calibration point to edit or press the f1 (back) key. c. If Delete is selected, the calibration point will be deleted. If deleting the point changes it to a one point calibration, enter the slope using the popup number entry screen.

4. The meter will display an updated calibration summary and export the data to the calibration log. Press the f1 (meas) key to proceed to the measurement mode or press the f2 (print) key to export the calibration data to a printer or computer.

56 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 4

|

Using the pH or pH/ISE Channel

Measurement Procedure

Prepare the electrodes according to the instructions in the electrode manuals. Connect all electrodes to be used to the appropriate meter inputs. Make sure the electrodes are recently calibrated and working correctly.

Ensure the meter measurement mode is set to the desired measurement parameter. For Orion

Star A215 and Orion Star A216 meters, press the f3 (channel) key until the desired channel or channels are shown in the measurement mode.

1. Rinse the electrodes with distilled water or appropriate solution, blot gently with a lint-free tissue to remove excess water and place into the sample.

2. Start the measurement and wait for the reading to stabilize or reach the predefined time. a. Auto-Read: Press the measure (esc) key to start the measurement. If using a stirrer probe, the stirrer probe will start stirring when the measure (esc) key is pressed and stop stirring when the measurement stabilizes. b. Continuous: Measurements will start immediately when in the measurement mode. If using a stirrer probe, press the stirrer key to start and stop stirring. c. Timed: Measurements will start immediately when in the measurement mode. If using a stirrer probe, press the stirrer key to start and stop stirring.

3. Once the measurement is stable or reaches the set time, record all applicable parameters. a. Auto-Read: Once the measurement is stable, it will be locked and held on the display and the AR icon will remain solid. If the data log function is enabled, the measurement will be exported to the data log. b. Continuous: The flashing Stabilizing icon will update to the solid Ready icon when the measurement is stable. If the data log function is enabled, press the log/print key to export the measurement to the data log. c. Timed: Measurements will be recorded at the predefined time interval. If the data log function is enabled, the < icon will be shown when each time interval has elapsed and the measurement is exported to the data log.

4. Remove the electrodes from the sample, rinse with distilled water or appropriate solution, blot gently with a lint-free tissue to remove excess water and place into the next sample.

5. Repeat steps 2 through 4 for all samples. When all samples have been measured, store the electrodes according to the instructions in the electrode manuals.

Orion Star A210 Series Benchtop Meter User Manual |

57

C H A P T E R 5

5

Using the Conductivity Channel

Use the Orion Star A212 conductivity meter or Orion Star A215 pH/conductivity meter to measure conductivity, salinity, TDS, resistivity and temperature. The following chapter provides instructions on connecting sensors, calibrating the system and measuring these parameters.

Meter and Sensor Preparation

1. Familiarize yourself with general meter operation and prepare the meter and accessories

according to the instructions in Chapter 2 Meter Basics .

a. It is recommended that sensors be placed in the meter-attached electrode stand for easy movement in and out of containers during calibration, measurement and storage.

2. Refer to the instructions in Chapter 3 Meter Setup Menus to customize measurement,

calibration and alarm settings within the COND Channel setup menu and data transfer, data log, date and time, language, sounds, stirrer speed, display contrast, auto-shutoff, user ID and sample ID settings within the Instrument Settings setup menu.

3. Connect all necessary electrodes, probes and sensors to the appropriate meter inputs. a. Connect a conductivity sensor to the 8 pin MiniDIN meter input. If the conductivity sensor has built-in temperature, the temperature will also be measured when the conductivity sensor is connected with the meter. b. Connect the stirrer probe (Catalog Number 096019) to the Stirrer meter input.

4. Prepare all connected sensors for use as instructed in the sensor manuals.

58 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 5

|

Using the Conductivity Channel

Conductivity Calibration Procedure

Orion Star A212 conductivity meters and Orion Star A215 pH/conductivity meters can perform a conductivity calibration using one to five conductivity standards with easy to follow onscreen instructions to assist with the calibration process or by the manual entry of the certified conductivity cell constant (K) value.

Always use fresh standards and select standards that are near the expected sample conductivity. Prepare the conductivity sensor according to the instructions in the sensor manual.

Connect all sensors to be used to the appropriate meter inputs.

Make sure the meter measurement mode is set to conductivity (Cond). For Star A215 meters, press the f3 (channel) key until the conductivity channel is shown in the measurement mode.

Note: For an automatic calibration, enter the nominal cell constant of the conductivity sensor in the channel-specific setup menu and use Orion 100 μS, 1413 μS and/or 12.9 mS standards.

1. Press the f1 (cal) key to start the calibration. a. If more than one channel is displayed in the measurement mode, press the p or q key to highlight Conductivity-Channel and press the f2 (select) key.

2. Rinse the conductivity sensor and any other electrodes in use with distilled water, blot gently with a lint-free tissue to remove excess water and place into the conductivity standard.

3. When the sensor and standard are ready, press the f3 (start) key. a. If using a stirrer probe, the stirrer probe will start stirring when the f3 (start) key is pressed and stop stirring when the reading stabilizes.

4. Wait for the value on the meter to stabilize and perform one of the following actions: a. Press the f2 (accept) key to accept the conductivity value determined by automatic standard recognition function (Orion 100 μS, 1413 μS and 12.9 mS standards only). or b. Press the f3 (edit) key to access the popup number entry screen and manually enter the conductivity standard value expected at the measured temperature. i. Press the p, q, t or u key to highlight a number or decimal point, press the

f3 (enter) key to select the highlighted item and repeat until the desired value is

shown at the top of the screen. ii. Press the f2 (done) key to save the value and exit the number entry screen. iii. Press the f2 (accept) key to accept the manually entered value.

5. Press the f2 (next) key to proceed to the next standard and repeat steps 2 through 4 or press the f3 (cal done) key to save and end the calibration. If five standard are used, the calibration will save and end once the fifth value is accepted.

6. The meter will display the calibration summary and export the data to the calibration log.

Press the f1 (meas) key to proceed to the measurement mode or press the f2 (print) key to export the calibration data to a printer or computer.

Orion Star A210 Series Benchtop Meter User Manual |

59

Chapter 5

|

Using the Conductivity Channel

Conductivity Calibration Editing

When the calibration summary is displayed, the calibration editing option can be selected to fix individual points without having to perform a full recalibration.

1. In the calibration summary display (after step 5 in the Conductivity Calibration Procedure), press the f3 (cal edit) key.

2. Press the p or q key to highlight a calibration point to edit and press the f2 (select) key.

3. Press the p or q key to highlight Remeasure, Edit or Delete and press the f2 (select) key. a. If Remeasure is selected, repeat steps 2 through 4 in the Calibration Procedure. b. If Edit is selected, press the f3 (edit) key, use the popup number entry screen to enter a new conductivity standard value, press the f2 (done) key and then press the f2

(accept) key. Select another calibration point to edit or press the f1 (back) key.

c. If Delete is selected, the calibration point will be deleted.

4. The meter will display an updated calibration summary and export the data to the calibration log. Press the f1 (meas) key to proceed to the measurement mode or press the f2 (print) key to export the calibration data to a printer or computer.

Alternative Certified Cell Constant Entry Calibration Procedure

When using the cell constant entry calibration procedure, enter the certified cell constant (K) value that is printed on the conductivity sensor cable or certificate of calibration.

1. Press the f1 (cal) key to start the calibration. a. If more than one channel is displayed in the measurement mode, press the p or q key to highlight Conductivity-Channel and press the f2 (select) key.

2. Press the f2 (cell K) key.

3. Wait for the conductivity value on the meter to stabilize and then press the f3 (edit) key to access the popup number entry screen and enter the certified cell constant value. a. Press the p, q, t or u key to highlight a number or decimal point, press the f3

(enter) key to select the highlighted item and repeat until the desired value is shown at

the top of the screen. b. Press the f2 (done) key to save the value and exit the number entry screen. c. Press the f2 (accept) key to accept the entered cell constant value.

4. Press the f2 (accept) key to save and end the calibration and export the data to the calibration log. The meter will automatically proceed to the measurement mode.

5. The conductivity measurement system is now calibrated. Verify the calibration by reading a known conductivity standard or according to your SOP.

60 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 5

|

Using the Conductivity Channel

Conductivity Standards vs. Temperature Table

130.4

132.5

134.6

136.7

138.9

141.0

143.2

145.4

113.9

115.9

117.9

120.0

122.0

124.1

126.2

128.3

147.6

149.8

152.0

154.2

156.4

158.7

160.9

163.2

165.4

98.08

100.0

102.0

103.9

105.9

107.9

109.9

111.9

82.98

84.83

86.69

88.56

90.45

92.34

94.24

96.15

Orion 111.9mS

Standard (mS/cm)

65.10

66.84

68.59

70.35

72.12

73.91

75.70

77.50

79.32

81.15

38

39

40

41

34

35

36

37

30

31

32

33

26

27

28

29

46

47

48

49

42

43

44

45

50

22

23

24

25

18

19

20

21

14

15

16

17

10

11

12

13

Temperature

(°C)

0

1

8

9

6

7

4

5

2

3

15.10

15.35

15.61

15.86

16.12

16.37

16.63

16.89

13.10

13.35

13.59

13.84

14.09

14.34

14.59

14.85

17.15

17.40

17.66

17.92

18.18

18.44

18.70

18.96

19.22

11.16

11.40

11.64

11.88

12.12

12.36

12.61

12.85

9.309

9.535

9.763

9.993

10.22

10.46

10.69

10.93

Orion 12.9mS

Standard (mS/cm)

7.135

7.344

7.555

7.768

7.983

8.200

8.418

8.638

8.860

9.084

1665

1693

1722

1751

1780

1808

1837

1866

1441

1468

1496

1524

1552

1580

1608

1636

1896

1925

1954

1983

2013

2042

2071

2101

2130

1224

1251

1277

1304

1331

1358

1386

1413

1017

1043

1068

1094

1119

1145

1171

1198

Orion 1413μS

Standard (μS/cm)

776

799

918

943

968

992

822

846

870

894

127

129

131

134

119

121

123

125

110

112

114

117

102

104

106

108

145

147

149

151

136

138

140

142

154

94

96

98

100

87

88

90

92

79

81

83

85

72

73

75

77

Orion 100μS

Standard (μS/cm)

54

56

64

66

68

70

58

59

61

63

185

188

191

194

173

176

179

182

161

164

167

170

150

153

156

159

209

212

215

219

197

200

203

206

222

138

141

144

147

127

130

133

136

116

119

122

125

106

108

111

114

Orion 147μS

Standard (μS/cm)

81

83

96

98

101

103

86

88

91

93

Orion Star A210 Series Benchtop Meter User Manual |

61

Chapter 5

|

Using the Conductivity Channel

Conductivity Meter Verification Procedure

Use the Orion Star series conductivity meter verification kit, Catalog Number 1010001, to verify the accuracy of the conductivity measurements on Orion Star A212 conductivity meters and

Orion Star A215 pH/conductivity meters. The measured conductivity value of each resistor should fall within the relative accuracy of the resistor (±0.1% of the actual resistor conductance) plus the relative accuracy of the meter (±0.5% of reading ±1 digit for readings greater than 3

μS/cm or ±0.5% of reading ±0.01μS/cm for readings less than or equal to 3 μS/cm).

1. Ensure that the conductivity resistor kit is certified and has a valid calibration date.

2. In the measurement mode, press the f1 (cal) key. a. If more than one channel is displayed in the measurement mode, press the p or q key to highlight Conductivity-Channel and press the f2 (select) key.

3. Press the f2 (cell K) key.

4. Press the f3 (edit) key to access the popup number entry screen and enter a cell constant value of 1.0000. a. Press the p, q, t or u key to highlight a number or decimal point, press the f3

(enter) key to select the highlighted item and repeat until the desired value is shown at

the top of the screen. b. Press the f2 (done) key to save the value and exit the number entry screen. c. Press the f2 (accept) key to accept the entered cell constant value.

5. The meter will automatically proceed to the measurement mode and Cell Constant: 1.0000 will be shown on the display.

6. Connect a resistor to the meter, press the measure (esc) key to start the measurement, wait for the measurement value to stabilize and then record the displayed conductivity value of the resistor.

7. Repeat step 6 for all six resistors in the kit.

Resistor Number Nominal Resistance Nominal Conductance Nominal Acceptance Range

1010001-A

1010001-B

1010001-C

1010001-D

1010001-E

1010001-F

1000 KΩ

100 KΩ

10 KΩ

1 KΩ

100 Ω

10 Ω

1 μS

10 μS

100 μS

1000 μS

10 mS

100 mS

0.984 to 1.016 μS

9.930 to 10.07 μS

99.30 to 100.7 μS

993.0 to 1007 μS

9.930 to 10.07 mS

99.30 to 100.7 mS

Note: These are the nominal values for the resistors only. The actual resistance and conductance is printed on each resistor and the actual acceptance range must be calculated from the actual value on each resistor.

62 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 5

|

Using the Conductivity Channel

Measurement Procedure

Prepare the sensors according to the instructions in the sensors manuals. Connect all electrodes, sensors and probes to be used to the appropriate meter inputs. Make sure the sensors are recently calibrated and working correctly.

Ensure the meter measurement mode is set to the desired measurement parameter. For Star

A215 meters, press the f3 (channel) key until the desired channel or channels are shown in the measurement mode.

1. Rinse the sensors with distilled water or appropriate solution, blot gently with a lint-free tissue to remove excess water and place into the sample.

2. Start the measurement and wait for the reading to stabilize or reach the predefined time. a. Auto-Read: Press the measure (esc) key to start the measurement. If using a stirrer probe, the stirrer probe will start stirring when the measure (esc) key is pressed and stop stirring when the measurement stabilizes. b. Continuous: Measurements will start immediately when in the measurement mode. If using a stirrer probe, press the stirrer key to start and stop stirring. c. Timed: Measurements will start immediately when in the measurement mode. If using a stirrer probe, press the stirrer key to start and stop stirring.

3. Once the measurement is stable or reaches the set time, record all applicable parameters. a. Auto-Read: Once the measurement is stable, it will be locked and held on the display and the AR icon will remain solid. If the data log function is enabled, the measurement will be exported to the data log. b. Continuous: The flashing Stabilizing icon will update to the solid Ready icon when the measurement is stable. If the data log function is enabled, press the log/print key to export the measurement to the data log. c. Timed: Measurements will be recorded at the predefined time interval. If the data log function is enabled, the < icon will be shown when each time interval has elapsed and the measurement is exported to the data log.

4. Remove the sensors from the sample, rinse with distilled water or appropriate solution, blot gently with a lint-free tissue to remove excess water and place into the next sample.

5. Repeat steps 2 through 4 for all samples. When all samples have been measured, store the sensors according to the instructions in the sensor manuals.

Orion Star A210 Series Benchtop Meter User Manual |

63

C H A P T E R 6

6

Using the DO/RDO Channel

Use the Orion Star A213 RDO/DO meter or Orion Star A216 pH/RDO/DO meter to measure dissolved oxygen as percent saturation or mg/L and temperature using either Orion RDO optical or polarographic dissolved oxygen sensors. The following chapter provides instructions on connecting sensors, calibrating the system and measuring these parameters.

Meter and Sensor Preparation

1. Familiarize yourself with general meter operation and prepare the meter and accessories

according to the instructions in Chapter 2 Meter Basics .

a. It is recommended that sensors be placed in the meter-attached electrode stand for easy movement in and out of containers during calibration, measurement and storage.

2. Refer to the instructions in Chapter 3 Meter Setup Menus to customize measurement,

calibration and alarm settings within the DO/RDO Channel setup menu and data transfer, data log, date and time, language, sounds, stirrer speed, display contrast, auto-shutoff, user

ID and sample ID settings within the Instrument Settings setup menu.

3. Connect all necessary electrodes, probes and sensors to the appropriate meter inputs. a. Connect an RDO optical or polarographic dissolved oxygen sensor to the 9 pin MiniDIN meter input. The meter will automatically detect which dissolved oxygen sensor is connected. The temperature will also be measured when the dissolved oxygen sensor is connected with the meter. b. Connect the stirrer probe (Catalog Number 096019) to the Stirrer meter input.

4. Prepare all connected sensors for use as instructed in the sensor manuals.

64 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 6

|

Using the DO/RDO Channel

Dissolved Oxygen Calibration Procedure

Note: A polarographic dissolved oxygen sensor must be polarized prior to use. The polarographic dissolved oxygen sensor is continuously polarized when connected to the meter.

If the polarographic dissolved oxygen sensor is new, has been serviced or has not been connected to the meter – connect the sensor to the meter, power on the meter and wait 30 minutes for the sensor to polarize.

Orion Star A213 RDO/DO meters and Orion Star A216 pH/RDO/DO meters can perform the following calibrations:

·

Air (water-saturated air) – This is the simplest and most accurate method and uses the calibration sleeve included with most dissolved oxygen sensors. For best accuracy, the calibration temperature should match the expected sample temperature. Moisten the sponge in the calibration sleeve with distilled water and insert the sensor into the sleeve.

Alternatively, a BOD bottle can be used with just enough distilled water to cover the bottom without touching the sensor.

·

Water (air-saturated water) – This method uses water that is 100% saturated with air.

Bubble air into a water sample for an extended period, preferably overnight.

·

Manual (Winkler) – This method uses a water sample with a known concentration of dissolved oxygen and is typically used to calibrate the DO sensor to the value achieved by a Winkler titration. Due to possible titration errors, this method is inherently less accurate.

· Set Zero – This method uses an oxygen free solution to add a zero point to an existing

Water Saturated Air or Air Saturated Water calibration. This calibration is generally recommended when measurements are below 10% saturation or 1 mg/L.

Prepare the dissolved oxygen sensor according to the instructions in the sensor manual.

Connect all sensors to be used to the appropriate meter inputs.

Make sure the meter measurement mode is set to dissolved oxygen (DO or RDO). For Star

A216 meters, press the f3 (channel) key until the dissolved oxygen channel is shown in the measurement mode.

Orion Star A210 Series Benchtop Meter User Manual |

65

Chapter 6

|

Using the DO/RDO Channel

Air (Water-saturated Air) Calibration

1. Press the f1 (cal) key to start the calibration. a. If more than one channel is displayed in the measurement mode, press the p or q key to highlight DO/RDO-Channel and press the f2 (select) key.

2. Press the p or q key to highlight Air and press the f3 (select) key.

3. Prepare the DO sensor and calibration sleeve and allow them to reach equilibrium.

4. When the DO sensor and calibration sleeve are ready, press the f3 (start) key.

5. Wait for the dissolved oxygen value on the meter to stabilize. a. With a polarographic DO sensor, 102.3% will be displayed when the reading stabilizes. b. With an RDO optical DO sensor, 100.0% will be displayed when the reading stabilizes.

6. Press the f2 (cal done) key to save and end the calibration.

7. The meter will display the calibration summary and export the data to the calibration log.

Press the f1 (meas) key to proceed to the measurement mode or press the f2 (print) key to export the calibration data to a printer or computer.

Water (Air-saturated Water) Calibration

1. Press the f1 (cal) key to start the calibration. a. If more than one channel is displayed in the measurement mode, press the p or q key to highlight DO/RDO-Channel and press the f2 (select) key.

2. Press the p or q key to highlight Water and press the f3 (select) key.

3. Prepare the DO sensor and 100% air-saturated water and allow them to reach equilibrium. a. Use plastic paraffin film to seal the open area between the DO sensor and vessel containing the 100% air-saturated water.

4. When the DO sensor and calibration apparatus are ready, press the f3 (start) key. a. If using a stirrer probe, the stirrer probe will start stirring when the f3 (start) key is pressed and stop stirring when the reading stabilizes.

5. Wait for the dissolved oxygen value on the meter to stabilize. a. 100.0% will be displayed when the reading stabilizes.

6. Press the f2 (cal done) key to save and end the calibration.

7. The meter will display the calibration summary and export the data to the calibration log.

Press the f1 (meas) key to proceed to the measurement mode or press the f2 (print) key to export the calibration data to a printer or computer.

66 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 6

|

Using the DO/RDO Channel

Manual (Winkler) Calibration

1. Press the f1 (cal) key to start the calibration. a. If more than one channel is displayed in the measurement mode, press the p or q key to highlight DO/RDO-Channel and press the f2 (select) key.

2. Press the p or q key to highlight Manual and press the f3 (select) key.

3. Prepare the DO sensor and calibration solution and allow them to reach equilibrium. a. Use plastic paraffin film to seal the open area between the DO sensor and vessel containing the calibration solution.

4. When the DO sensor and calibration apparatus are ready, press the f3 (start) key. a. If using a stirrer probe, the stirrer probe will start stirring when the f3 (start) key is pressed and stop stirring when the reading stabilizes.

5. Wait for the dissolved oxygen value on the meter to stabilize and perform one of the following actions: a. Press the f2 (accept) key to accept the displayed dissolved oxygen value.

Or b. Press the f3 (edit) key to access the popup number entry screen and manually enter the dissolved oxygen value. i. Press the p, q, t or u key to highlight a number or decimal point, press the

f3 (enter) key to select the highlighted item and repeat until the desired value is

shown at the top of the screen. ii. Press the f2 (done) key to save the value and exit the number entry screen.

6. Press the f2 (cal done) key to save and end the calibration.

7. The meter will display the calibration summary and export the data to the calibration log.

Press the f1 (meas) key to proceed to the measurement mode or press the f2 (print) key to export the calibration data to a printer or computer.

Orion Star A210 Series Benchtop Meter User Manual |

67

Chapter 6

|

Using the DO/RDO Channel

Set Zero Calibration

An Air (water-saturated air) calibration or Water (air-saturated water) calibration must be completed before a Set Zero calibration can be performed.

Prepare a sodium sulfite solution by dissolving about 15.0 grams of Na

2

SO

3

in about 250 mL of distilled water. Transfer the solution to a BOD bottle or flask and use plastic paraffin film to seal the bottle. A small amount of cobalt salt can be added to the sodium sulfite solution to act as an indicator and change color when the sodium sulfite solution no longer has zero oxygen content.

1. Press the f1 (cal) key to start the calibration. a. If more than one channel is displayed in the measurement mode, press the p or q key to highlight DO/RDO-Channel and press the f2 (select) key.

2. Press the p or q key to highlight Set Zero and press the f3 (select) key.

3. Prepare the DO sensor and zero oxygen solution and allow them to reach equilibrium. a. Use plastic paraffin film to seal the open area between the DO sensor and vessel containing the calibration solution.

4. When the DO sensor and calibration apparatus are ready, press the f3 (start) key. a. If using a stirrer probe, the stirrer probe will start stirring when the f3 (start) key is pressed and stop stirring when the reading stabilizes.

5. Wait for the dissolved oxygen value on the meter to stabilize. a. 0.0% will be displayed for P2 when the reading stabilizes.

6. Press the f2 (cal done) key to save and end the calibration.

7. The meter will display the calibration summary and export the data to the calibration log.

Press the f1 (meas) key to proceed to the measurement mode or press the f2 (print) key to export the calibration data to a printer or computer.

68 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 6

|

Using the DO/RDO Channel

Measurement Procedure

Prepare the sensors according to the instructions in the sensors manuals. Connect all electrodes, sensors and probes to be used to the appropriate meter inputs. Make sure the sensors are recently calibrated and working correctly.

Ensure the meter measurement mode is set to the desired measurement parameter. For Star

A216 meters, press the f3 (channel) key until the desired channel or channels are shown in the measurement mode.

1. Rinse the sensors with distilled water or appropriate solution, blot gently with a lint-free tissue to remove excess water and place into the sample.

2. Start the measurement and wait for the reading to stabilize or reach the predefined time. a. Auto-Read: Press the measure (esc) key to start the measurement. If using a stirrer probe, the stirrer probe will start stirring when the measure (esc) key is pressed and stop stirring when the measurement stabilizes. b. Continuous: Measurements will start immediately when in the measurement mode. If using a stirrer probe, press the stirrer key to start and stop stirring. c. Timed: Measurements will start immediately when in the measurement mode. If using a stirrer probe, press the stirrer key to start and stop stirring.

3. Once the measurement is stable or reaches the set time, record all applicable parameters. a. Auto-Read: Once the measurement is stable, it will be locked and held on the display and the AR icon will remain solid. If the data log function is enabled, the measurement will be exported to the data log. b. Continuous: The flashing Stabilizing icon will update to the solid Ready icon when the measurement is stable. If the data log function is enabled, press the log/print key to export the measurement to the data log. c. Timed: Measurements will be recorded at the predefined time interval. If the data log function is enabled, the < icon will be shown when each time interval has elapsed and the measurement is exported to the data log.

4. Remove the sensors from the sample, rinse with distilled water or appropriate solution, blot gently with a lint-free tissue to remove excess water and place into the next sample.

5. Repeat steps 2 through 4 for all samples. When all samples have been measured, store the sensors according to the instructions in the sensor manuals.

Orion Star A210 Series Benchtop Meter User Manual |

69

C H A P T E R 7

7

Data Transfer and Software Updates

Data Storage and Transfer Settings

Meter settings that affect data storage and transfer include the measurement read type, export data settings and data log setting. The date and time settings should also be reviewed and update as needed.

Measurement Read Type Setting

The selected read type determines when the meter sends measurements to the data log, if the data log is enabled in the setup menu, and when the meter exports measurements to a printer or computer, if a printer or computer is properly connected to the meter and export data settings are enabled in the setup menu. Select from Auto-Read, continuous and timed read types.

Auto-Read

Press the measure (esc) key to start a measurement. The AR icon blinks as the changing value stabilizes. When the measurement is stable, the AR icon stops blinking and the measurement is locked on the display until the measure (esc) key is pressed again. When the data log and export data settings are enabled, the stable measurement is automatically saved to the data log and exported to a printer or computer.

Continuous

Measurement values are continuously updated on the display and the Stabilizing or Ready icon indicates the measurement stability status. When the data log and export data settings are enabled, press the log/print key to save a measurement to the data log and export it to a printer or computer.

70 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 7

|

Data Transfer and Software Updates

Timed

Measurement values are continuously updated on the display and the Stabilizing or Ready icon indicates the measurement stability status. When the data log and export data settings are enabled, measurements are saved to the data log and exported to a printer or computer at the selected time intervals from 3 seconds (00:00:03) to 24 hours (24:00:00).

To Set Measurement Read Type:

1. In the measurement mode, press the setup key to access the main setup menu.

2. Press the p, q, t or u key to highlight pH Channel, pH/ISE Channel, COND Channel or

DO/RDO Channel and press the f3 (select) key.

3. Press the p or q key to highlight Mode and Settings and press the f3 (select) key.

4. Press the p or q key to highlight Read Type and press the f3 (select) key.

5. Press the p or q key to highlight Auto, Continuous or Timed and press the f3 (select) key to set the desired measurement mode. a. If Timed is selected, press the t or u key to highlight the hours (00:00:00), minutes

(00:00:00) or seconds (00:00:00). b. Press the f3 (edit) key to access the popup number entry screen. c. Press the p, q, t or u key to highlight a number, press the f3 (enter) key to select the number and repeat until a two digit value is shown at the top of the screen. d. Press the f2 (done) key to save the value and exit the number entry screen. e. Repeat until the correct time interval is entered for the hours, minutes and seconds.

Time intervals from 3 seconds (00:00:03) to 24 hours (24:00:00) can be entered.

6. Press the f1 (back) key to navigate out of the menu and then press the measure (esc) key to return to the measurement mode.

Data Log Setting

To save up to 2000 measurement sets to the meter data log, turn the data log setting on in the setup menu.

1. In the measurement mode, press the setup key to access the main setup menu.

2. Press the p, q, t or u key to highlight Settings and press the f3 (select) key.

3. Press the p or q key to highlight Data Log and press the f3 (select) key.

4. Press the p or q key to highlight On and press the f3 (select) key.

5. Press the f1 (back) key to navigate out of the menu and press the measure (esc) key to return to the measurement mode.

Orion Star A210 Series Benchtop Meter User Manual |

71

Chapter 7

|

Data Transfer and Software Updates

Export Data Settings

To transfer measurements to a computer or printer, turn the export data function on in the setup menu and update export settings to match the external device settings, including the baud rate.

The default meter communication connection is RS232 and the meter will automatically detect and adjust the connection setting when an active USB cable is connected with the meter.

1. In the measurement mode, press the setup key to access the main setup menu.

2. Press the p, q, t or u key to highlight Settings and press the f3 (select) key.

3. Press the p or q key to highlight Export Data and press the f3 (select) key.

4. Press the p or q key to highlight Printing and press the f3 (select) key. Press the p or q

key to highlight On and press the f3 (select) key.

5. Press the p or q key to highlight Data Format and press the f3 (select) key. Press the p or q key to highlight Printer or PC (CSV) and press the f3 (select) key.

6. Press the p or q key to highlight Comm Config and press the f3 (select) key. Press the p

or q key to highlight 1200, 2400, 4800, 9600, 19200 or 38400 as the meter baud rate and press the f3 (select) key.

7. Press the f1 (back) key to navigate out of the menu and press the measure (esc) key to return to the measurement mode.

Data Transfer Examples

5 Point pH Calibration, CSV Format, HyperTerminal View: pH Measurement, CSV Format, HyperTerminal View:

72 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

5 Point pH Calibration, Printer Format:

Chapter 7

|

Data Transfer and Software Updates

pH Measurement, Printer Format:

Thermo Scientific Orion Star A210 Series Benchtop Meter User Manual |

73

Chapter 7

|

Data Transfer and Software Updates

Printer Compatibility and Requirements

Orion Star A210 series benchtop meters can print directly to the Orion Star series inkjet printer,

Catalog Number 1010006. Measurement and calibration data sent from the meter to the Orion

Star series inkjet printer is automatically formatted to fit the paper width when the printer data format is selected within the Export Data setting. The Orion Star series inkjet printer has a 9600 baud rate and it is packaged with the RS232 printer cable required to interface an Orion Star

A210 series meter to this printer.

The baud rate of Orion Star A210 series meters can be set to 1200, 2400, 4800, 9600, 19200 or 38400 for communication to different printers. The fixed meter settings are:

Number of Data Bits: 8

Stop Bits:

1

Parity:

Flow Control:

None

XON/XOFF

The meter will send measurements and calibration data to the printer if the printer option is turned on in the setup menu. Alternatively, measurements and calibration data can be sent to the data log and calibration log. From the data log and calibration log, the operator can chose to print a single data log point, a range of data log points or the entire data log.

74 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 7

|

Data Transfer and Software Updates

Computer Compatibility and Requirements

The Orion Star A210 series meters can send measurements and calibration data to a computer in a comma delimited format that is easy to parse in computer programs like Excel.

The baud rate of Orion Star A210 series meters can be set to 1200, 2400, 4800, 9600, 19200 or 38400. The fixed meter settings are:

Number of Data Bits: 8

Stop Bits:

Parity:

Flow Control:

1

None

XON/XOFF

The meter will send measurements and calibration data to the computer if the printer option is turned on in the setup menu. Alternatively, measurements and calibration data can be sent to the data log and calibration log. From the data log and calibration log, the operator can chose to print a single data log point, a range of data log points or the entire data log.

The computer must be up-to-date with the latest Windows updates installed and have .NET framework version 4.0 with the latest updates.

All computer screen savers and power management settings should be disabled when the meter is interfaced with the computer. The computer should not go into sleep/standby mode while the meter is transferring data. Laptop computers should be plugged into a power supply, running on battery power is not recommended.

Three computer cables can be used to interface the Orion Star A210 series benchtop meters with a computer – the Orion Star series USB-to-Serial computer cable (Cat. No. 1010005),

Orion Star series RS232 computer cable (Cat. No. 1010053) or a standard Mini B USB to USB computer cable.

Computer Cable

USB-to-Serial Computer Cable,

Cat. No. 1010005

RS232 Computer Cable,

Cat. No. 1010053

USB Computer Cable (standard

Mini B USB to USB cable)

Meter

Connection

RS232

RS232

USB (Mini B)

Computer

Connection

USB

RS232

USB

Driver Required

Yes

No

Yes

Orion Star A210 Series Benchtop Meter User Manual |

75

Chapter 7

|

Data Transfer and Software Updates

Using the USB-to-Serial Computer Cable

Turn on the power to the meter and connect the USB-to-Serial computer cable first to the meter and then to the computer. The USB-to-Serial driver must be installed when the meter will be interfaced with a computer using the USB-to-Serial computer cable. The USB-to-Serial driver is compatible with Microsoft

®

Windows

®

XP, Windows Vista

®

, Windows 7 and Windows 8 operating systems.

1. Connect the meter to a wall outlet using the universal power adapter (catalog number

1010003) or install four AA batteries and power on the meter.

2. Connect the USB-to-Serial computer cable first to the meter’s RS232 input and then to the computer’s USB input.

3. After the USB-to-Serial computer cable is connected to the computer, the computer will automatically identify the device and install the driver. Once the device driver is done installing, the USB-to-Serial computer cable can be used to transfer data from the meter to the computer using Orion Star Com computer software, HyperTerminal or similar program and send remote commands from the computer to the meter.

4. Record the COM port location of the USB-to-Serial cable, for example COM8. To view the

COM port location, open the computer’s Device Manager tool and expand the Ports option. a. For Windows 7: Click the Start button and then click Control Panel. In the Control Panel window, click Device Manager if viewing by large icons or small icons or click System and Security and then click Device Manager if viewing by category. In the Device

Manager window, expand the Ports option. b. For Windows XP: Click the Start button and then click Control Panel. In the Control

Panel window, click the Performance and Maintenance link and then click the System icon or for Control Panel Classic View, just double-click the System icon. In the System

Properties window, click the Hardware tab and then click the Device Manager button. In the Device Manager window, expand the Ports option.

76 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Chapter 7

|

Data Transfer and Software Updates

Thermo Scientific

5. The meter is now able to transfer data to the computer using the Virtual COM port and

Orion Star Com computer software, HyperTerminal or similar program.

Note: If the computer does not automatically install the device driver, open the computer’s

Device Manager tool, expand the Other Deceives option, double click the device and manually install the driver. The drivers for some RS232 to USB to adapters may need to be downloaded from the manufacturer’s website. For example, when using a Tripp Lite adapter, go to http://www.tripplite.com/support/downloads/ , enter the model number of the adapter (i.e. U209-

000-R) and select the appropriate driver for the computer operating system.

Orion Star A210 Series Benchtop Meter User Manual |

77

Chapter 7

|

Data Transfer and Software Updates

Using the USB Computer Cable

The Orion Star A200-A300 USB driver must be installed when the meter will be interfaced with a computer using the USB computer cable. The Orion Star A200-A300 USB driver is compatible with Microsoft

®

Windows

®

XP, Windows Vista

®

, Windows 7 and Windows 8 operating systems. When using a USB computer cable, always turn on the power to the meter and connect the USB computer cable first to the meter and then to the computer.

1. Go to www.thermoscientific.com/OrionMeters , download the Orion Star A200-A300 USB driver folder and unzip/extract the files to the computer’s desktop.

2. Connect the power adapter to the meter and power on the meter.

3. Connect the USB cable first to the Mini B USB input on the meter and then to the USB input on a computer.

4. Once the USB cable is connected to the computer, the computer will try to search for the device driver software and then a message will appear that the device driver software was not successfully installed. Open the computer’s Device Manager tool. a. For Windows 7: Click the Start button and then click Control Panel. In the Control

Panel window, click Device Manager if viewing by large icons or small icons or click

System and Security and then click Device Manager if viewing by category. In the

Device Manager window, expand the Ports option. b. For Windows XP: Click the Start button and then click Control Panel. In the Control

Panel window, click the Performance and Maintenance link and then click the System icon or for Control Panel Classic View, just double-click the System icon. In the

System Properties window, click the Hardware tab and then click the Device Manager button. In the Device Manager window, expand the Ports option.

5. Locate and double-click on the Virtual COM Port (CDC) device in the Device Manager tool.

78 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Chapter 7

|

Data Transfer and Software Updates

6. The Virtual COM Port (CDC) properties window will open. Click the Update Driver button.

7. Click the “Browse my computer for driver software” option.

Thermo Scientific Orion Star A210 Series Benchtop Meter User Manual |

79

Chapter 7

|

Data Transfer and Software Updates

8. Click the Browse button, set the computer desktop as the location and click the Next button.

9. Wait while the driver software is installed and accept any warning messages.

10. Once the installation is complete, record the COM port location and click the Close button.

11. The meter is now able to transfer data to the computer using the Virtual COM port and

Orion Star Com computer software, HyperTerminal or similar program.

80 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Chapter 7

|

Data Transfer and Software Updates

Orion Star Com Communication Software

The Orion Star Com communication software is compatible with Orion Star A210 series benchtop meters and a free download is available at www.thermoscientific.com/OrionMeters .

The Star Com software facilitates the transfer of calibration and measurement data from the meter to a computer and then allows data to be exported to an Excel (.xls) or Comma-

Separated Values (.csv) file. Transferred data can also be printed from the computer.

Orion Star A210 series benchtop meters must have software revision 2.59 or higher to use the

Star Com software. The Orion Star A200/A300 series USB driver software must be installed on the computer if the USB port will be used to connect the meter with a computer.

Thermo Scientific Orion Star A210 Series Benchtop Meter User Manual |

81

Chapter 7

|

Data Transfer and Software Updates

Meter Interfacing with Other Computer Programs

If using the meter USB port to connect with the computer, the Orion Star A200/A300 series

USB driver software (virtual COM port) must be installed on the computer and the meter should be updated to software revision 3.04 or higher.

1. Power on the meter and connect the meter to the computer using the appropriate computer interface cable.

2. Open a computer communication program that uses COM ports. Programs such as

Terminal and HyperTerminal apply.

3. Select the correct Virtual COM Port and connect. The following examples are from

HyperTerminal:

4. Verify that the meter export data setting is on. Match the baud rate set in the meter to the computer setting. Selecting a higher baud rate will allow for faster data transfer.

5. Data will transfer from the meter to the computer, depending on the meter read type and command settings.

Note: If the meter is powered off for any reason (auto-shut off, loss of power, etc.) the connection will need to be reestablished by repeating the steps above.

82 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 7

|

Data Transfer and Software Updates

Star A200-A300 Meter Remote Control Protocols

Rules for Remote Control Usage

1. CR (Carriage Return, ASCII 13) is used to terminate a command. Whenever this character is received, the internal buffer will be processed.

2. All characters except for NL (New Line, ASCII 10) are significant. The NL character will be ignored. The meter is not case sensitive, though all lower case characters are converted to upper case internally.

3. Only one command can be executed at a time. No new command can issued until the current command is finished being processed. When the current command is finished, it will issue the “>” prompt to the user, indicating a new command can be entered.

4. Empty commands (i.e. just a CR) will be ignored and a new prompt will be issued.

5. Turn on the Export Data and Data Log meter settings. The default meter communication settings are:

Bits per second: 9600

Data bits: 8

Parity: None

Stop bits:

Flow control:

1

None

Remote Control Engine

The remote control engine receives input from the serial/USB port and processes it as needed.

Commands sent to the remote control interface will be in the form of “OPCODE

<OPERAND(s)>

CR

”. Line feeds will be ignored. New commands cannot be issued until the previous command has been completed and a prompt is given. A prompt is shown as the greater than symbol (“>“‘) followed by a space.

Orion Star A210 Series Benchtop Meter User Manual |

83

Chapter 7

|

Data Transfer and Software Updates

Star A200-A300 Meter Remote Control Commands

Press the Esc key on the computer keypad to stop a command being executed.

Command

GETMEAS

GETCAL

GETLOG

SYSTEM

SETRTC

Command Details

GETMEAS <CR>

Prints the current channel measurement immediately.

GETMEAS Data Count <CR>

Prints the current channel measurement for a set number of times.

Example: GETMEAS 2

GETMEASTIMED CH_Channel Combination, Time Interval <CR>

Set the channel(s) to be measured and the time interval for the measurement, measurements are sent to the meter data log

Channel Combination: Channel numbers to be measured.

Time Interval: The measurement retrieving time interval in seconds.

Example: GETMEASTIMED CH_12 5 <CR>

STOP <CR>

Command exits Timed measurement.

GETCAL MODE <CR>

Prints all the channel mode calibration data. If no calibration is available, will return

“>” to receive next command.

Example: GETCAL MODE <CR>

GETCAL MODE <CR>

Print calibration data for specific MODE as:

PH COND DO

RMV

ORP

RES

SALT

RDO

ISE TDS

Example: GETCAL PH <CR>

GETLOG <CR>

Prints all the logged measurement data. If no data is logged, will return “>” to receive next command.

Example: GETLOG <CR>

GETLOG START, END <CR>

Prints the data in a specified range. If no data is available, will return “>” to receive next command.

Example: GETLOG 10 100 <CR>

SYSTEM <CR>

Prints the Meter Model, Serial Number, Software Version

Example: 329, 12345, 2.53

SETRTC YY MM DD HH MM SS <CR>

Set the date and time (in 24 hour format) for the meter

Example: SETRTC 13 07 15 08 30 00

84 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 7

|

Data Transfer and Software Updates

Command

SETMODE

GETMODE

SETCSV

SETKEYLOCK

Command Details

SETMODE MODE <CR>

Set the current channel measurement MODE as:

PH

RMV

ORP

COND

RES

SALT

DOSAT

DOCON

ISE TDS

Example: SETMODE PH <CR>

GETMODE CHANNEL <CR>

Print the channel measurement mode

Example: GETMODE 1

SETCSV <CR>

Set printing format to CSV

SETKEYLOCK NUMBER <CR>

To disable the key pad: Number = 0

To enable the key pad Number = 1

Meter Measurement Data from GETMEAS Command

Measure

Mode

PC (CSV) Format

pH

Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,

Channel, Mode, pH Value, pH Unit, mV Value, mV Unit, Temperature Value, Temperature

Unit, Slope Value, Slope Unit, Method #, Log # mV

Example:

A211 pH, X01036, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, pH, 7.000, pH, 0.0, mV, 25.0, C, 98.1,%, M100, #1 <CR>

Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,

Channel, Mode, mV Value, mV Unit, Temperature Value, Temperature Unit, Method #,

Log #

RmV

Example:

A211 pH, X01036, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, mV, 0.0, mV, 25.0,

C, M100, #2 <CR>

Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,

Channel, Mode, Relative mV Value, Relative mV Unit, Offset mV Value, Offset mV Unit,

Temperature Value, Temperature Unit, Method #, Log #

ORP

Example:

A211 pH, X01036, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, RmV, 0.0, RmV,

0.0, mV, 25.0, C, M100, #3 <CR>

Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,

Channel, Mode, ORP Value, ORP Unit, mV Value, mV Unit, Temperature Value,

Temperature Unit, Method #, Log #

Example:

A211 pH, X01036, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, ORP, 0.0, mV, 0.0, mV, 25.0, C, M100, #4 <CR>

Orion Star A210 Series Benchtop Meter User Manual |

85

Chapter 7

|

Data Transfer and Software Updates

Measure

Mode

PC (CSV) Format

ISE

Conductivity

Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,

Channel, Mode, ISE Value, ISE Unit, mV Value, mV Unit, Temperature Value,

Temperature Unit, Slope Value, Slope Unit, Method #, Log #

Example:

A214 pH/ISE, X01037, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, ISE, 1, ppb, 0.0, mV, 25.0, C, 59.2, mV/dec, M100, #1 <CR>

Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,

Channel, Mode, Conductivity Value, Conductivity Unit, Conductance Value, Conductance

Unit, Temperature Value, Temperature Unit, Temperature Coefficient Value, Temperature

Coefficient Unit, Reference Temperature Value, Reference Temperature Unit, Cell

Constant Value, Cell Constant Unit, Method #, Log #

TDS

Salinity

Resistivity

Example:

A212 Cond, X01038, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, COND, 936.41, uS/cm, 1977.59, uS, 25.0, C, 2.1,%/C, 25.0, C, 0.4750, /cm, M100, #1 <CR>

Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,

Channel, Mode, TDS Value, TDS Unit, Conductance Value, Conductance Unit,

Temperature Value, Temperature Unit, Temperature Coefficient Value, Temperature

Coefficient Unit, Reference Temperature Value, Reference Temperature Unit, Cell

Constant Value, Cell Constant Unit, Method #, Log #

Example:

A212 Cond, X01038, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, TDS, 460, ppm,

1975, uS, 25.0, C, 2.1,%/C, 25.0, C, 0.4750, /cm, M100, #2 <CR>

Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,

Channel, Mode, Salinity Value, Salinity Unit, Conductance Value, Conductance Unit,

Temperature Value, Temperature Unit, Salinity Type, Reference Temperature Value,

Reference Temperature Unit, Cell Constant Value, Cell Constant Unit, Method #, Log #

Example:

A212 Cond, X01038, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, SALT, 0.5112, psu, 1973.120, uS, 25.0, C, Practical Salinity, 15.0, C, 0.4750, /cm, M100, #3 <CR>

Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,

Channel, Mode, Resistivity Value, Resistivity Unit, Conductance Value, Conductance Unit,

Temperature Value, Temperature Unit, Temperature Coefficient Value, Temperature

Coefficient Unit, Reference Temperature Value, Reference Temperature Unit, Cell

Constant Value, Cell Constant Unit, Method #, Log #

DO

% Saturation

Example:

A212 Cond, X01038, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, RES, 1068, Ohmcm, 1982, uS, 25.0, C, 2.1,%/C, 25.0, C, 0.4750, /cm, M100, #4 <CR>

Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,

Channel, Mode, Percent Saturation Value, Percent Saturation Unit, Concentration Value,

Concentration Unit, Current Value, Current Unit, Solution Temperature Value, Solution

Temperature Unit, Membrane Temperature Value, Membrane Temperature Unit,

Barometric Pressure Value, Barometric Pressure Unit, Salinity Correction Value, Salinity

Correction Unit, Slope Value, Slope Unit, Method #, Log #

Example:

A213 DO/RDO, X01039, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, DO, 0.3,%

Sat, 0.03, mg/L, 3.4, nA, 24.5, C, 24.6, C, 761.0, mmHg, 0.0, ppt, 11.800, Na/%Sat, M100,

#1 <CR>

86 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 7

|

Data Transfer and Software Updates

Measure

Mode

PC (CSV) Format

DO mg/L

Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,

Channel, Mode, Concentration Value, Concentration Unit, Percent Saturation Value,

Percent Saturation Unit, Current Value, Current Unit, Solution Temperature Value,

Solution Temperature Unit, Membrane Temperature Value, Membrane Temperature Unit,

Barometric Pressure Value, Barometric Pressure Unit, Salinity Correction Value, Salinity

Correction Unit, Slope Value, Slope Unit, Method #, Log #

RDO

% Saturation

Example:

A213 DO/RDO, X01039, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, DO, 0.03, mg/L, 0.3,% sat, 3.4, nA, 24.5, C, 24.6, C, 761.0, mmHg, 0.0, ppt, 11.800, Na/%Sat, M100,

#2 <CR>

Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,

Channel, Mode, Percent Saturation Value, Percent Saturation Unit, Concentration Value,

Concentration Unit, Partial Pressure Value, Partial Pressure Unit, Solution Temperature

Value, Solution Temperature Unit, Barometric Pressure Value, Barometric Pressure Unit,

Salinity Correction Value, Salinity Correction Unit, Slope Value, Slope Unit, Method #, Log

#

RDO mg/L

Example:

A213 DO/RDO, X01039, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, RDO,

100.0,% Sat, 8.40, mg/L, 138.5, Torr, 24.0, C, 761.0, mmHg, 0.0, ppt, 1.383, Torr/%Sat,

M100, #3 <CR>

Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,

Channel, Mode, Concentration Value, Concentration Unit, Percent Saturation Value,

Percent Saturation Unit, Partial Pressure Value, Partial Pressure Unit, Solution

Temperature Value, Solution Temperature Unit, Barometric Pressure Value, Barometric

Pressure Unit, Salinity Correction Value, Salinity Correction Unit, Slope Value, Slope Unit,

Method #, Log #

Example:

A213 DO/RDO, X01039, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, RDO, 8.40, mg/L, 100.0,% sat, 138.5, Torr, 24.0, C, 761.0, mmHg, 0.0, ppt, 1.383, Torr/%Sat, M100,

#4 <CR>

Meter Model, Serial Number, Software Revision, User ID, Date & Time, Sample ID,

(followed by measurement data for each channel, as shown above)

Multi

Channel

Example:

A215 pH/Cond, X01040, 3.04, ABCDE, 01/03/15 16:05:41, SAMPLE, CH-1, pH, 7.000, pH, 0.0, mV, 25.0, C, 98.1,%, M100, CH-2, COND, 936.41, uS/cm, 1977.59, uS, 25.0, C,

2.1,%/C, 25.0, C, 0.4750, /cm, M200, #1 <CR>

Orion Star A210 Series Benchtop Meter User Manual |

87

Chapter 7

|

Data Transfer and Software Updates

Meter Calibration Data from GETCAL Command

Cal Mode pH

RmV

ORP

ISE

Conductivity

PC (CSV) Format

Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &

Time, Total Calibration Points,

repeated for each calibration point: Calibration Point, pH Value, pH Unit, mV Value, mV Unit, Temperature Value, Temperature Unit,

repeated for each point-to-point segment: Slope Value, Slope Unit,

repeated for each point-to-point segment: Offset Value, Offset Unit,

Average Slope Value, Average Slope Unit, Calibration Number

Example (5 point calibration):

A211 pH, X01036, 3.04, CH-1, pH, 01/03/15 16:05:41, 5, 1, 2.00, pH, 290.5, mV, 25.0,

C, 2, 4.01, pH, 174.3, mV, 25.0, C, 3, 7.00, pH, 0.0, mV, 25.0, C, 4, 10.01, pH, -173.9, mV, 25.0, C, 5, 12.00, pH, -289.7, mV, 25.0, C, 97.7,%, 98.5,%, 97.7,%, 98.4,%, 1.4, mV, 0.0, mV, 0.0, mV, 1.3, mV, 98.1,%, #1 <CR>

Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &

Time, Raw mV Value, Raw mV Unit, Relative mV Offset Value, Relative mV Offset Unit,

Reference mV Value, Reference mV Unit, Calibration Number

Example:

A211 pH, X01036, 3.04, CH-1, RmV, 01/03/15 16:05:41, 0.0, mV, 0.0, mV, 0.0, mV, #1

<CR>

Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &

Time, Raw mV Value, Raw mV Unit, Offset Value, Offset Unit, Temperature Value,

Temperature Unit, Calibration Number

Example:

A211 pH, X01036, 3.04, CH-1, ORP, 01/03/15 16:05:41, 176.9, mV, 242.6, mV, 25.0, C,

#1 <CR>

Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &

Time, Total Calibration Points,

repeated for each calibration point: Calibration Point, Concentration Value,

Concentration Unit, mV Value, mV Unit, Temperature Value, Temperature Unit,

repeated for each point-to-point segment: Slope Value, Slope Unit,

repeated for each point-to-point segment: Offset Value, Offset Unit,

Average Slope Value, Average Slope Unit, Blank Value, Blank Unit, Calibration Number

Example (5 point calibration):

A214 pH/ISE, X01036, 3.04, CH-1, ISE, 01/03/15 16:05:41, 5, 1, -1, ppb, 0.0, mV, 25.0,

C, 2, 2, ppb, 19.0, mV, 25.0, C, 3, 300, ppb, 150.1, mV, 25.0, C, 4, -2000, ppb, 200.1, mV, 25.0, C, 5, 4000, ppb, 210.1, mV, 25.0, C, 63.1, mV/dec, 60.2, mV/dec, 60.7, mV/dec, 33.2, mV/dec, 0.0, mV, 0.9mV, 0.0mV, -0.2mV, 90.4, mV, 54.3, mV/dec,

0.0000, ppb, #1 <CR>

Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &

Time, Total Calibration Points,

repeated for each calibration point: Calibration Point, Conductivity Value,

Conductivity Unit, Conductance Value, Conductance Unit, Temperature Value,

Temperature Unit (MTC/ATC), Calibration Type, Calibration Factor,

Average Cell Constant (K), Calibration Number

Example (5 point calibration):

A212 Cond, X01038, 3.04, CH-1, Cond, 01/03/15 16:05:41, 5, 1, 52.8, uS/cm, 111.1, uS,

25.0, C, Manual, 0.4750, 2, 68.6, uS/cm, 143.1, uS, 25.0, C, Manual, 0.4956, 3, 100, uS/cm, 200.2, uS, 25.0, C, Manual, 0.5511, 4, 158.4, uS/cm, 333.4, uS, 25.0, C, Manual,

0.4375, 5, 475, uS/cm, 1000, uS, 25.0, C, Manual, 0.4749, 0.4868 #1 <CR>

88 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 7

|

Data Transfer and Software Updates

Cal Mode

DO

(Air or Water

Calibration )

DO

(Manual

Calibration )

RDO

(Air or Water

Calibration)

RDO

(Manual

Calibration)

PC (CSV) Format

Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &

Time, Calibration Point, Calibration Type, Calibration Saturation Value, Calibration

Saturation Unit, Set Zero Calibration Current Value, Set Zero Calibration Current Unit,

Calibration Current Value, Calibration Current Unit, Solution Temperature Value,

Solution Temperature Unit, Membrane Temperature Value, Membrane Temperature

Unit, Calibration Pressure Value, Calibration Pressure Unit, Salinity Value, Salinity Unit,

Calibration Slope Value, Calibration Slope Unit, Calibration Point, Calibration Number

Example:

A213 DO/RDO, X01039, 3.04, CH-1, DO, 01/03/15 16:05:41, 1, Auto- Air, 102.3,%, 0.0, nA, 405.1, nA, 5.0, C, 5.0, C, 736.1, mmHg, 0.0, ppt, 4.1, nA/%sat, 1, #1 <CR>

Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &

Time, Calibration Point, Calibration Type, Calibration Concentration Value, Calibration

Concentration Unit, Set Zero Calibration Current Value, Set Zero Calibration Current

Unit, Calibration Current Value, Calibration Current Unit, Solution Temperature Value,

Solution Temperature Unit, Membrane Temperature Value, Membrane Temperature

Unit, Calibration Pressure Value, Calibration Pressure Unit, Salinity Value, Salinity Unit,

Calibration Slope Value, Calibration Slope Unit, Calibration Point, Calibration Number

Example:

A213 DO/RDO, X01039, 3.04, CH-1, DO, 01/03/15 16:05:41, 1, Manual, 12.62, mg/l, 0.0, nA, 405.1, nA, 5.0, C, 5.0, C, 736.1, mmHg, 0.0, ppt, 4.1, nA/%sat, 1, #2 <CR>

Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &

Time, Probe Serial Number, Calibration Point, Calibration Type, Calibration Saturation

Value, Calibration Saturation Unit, Set Zero Partial Pressure Value, Set Zero Partial

Pressure Unit, Calibration Partial Pressure Value, Calibration Partial Pressure Unit,

Solution Temperature Value, Solution Temperature Unit, Barometric Pressure Value,

Barometric Pressure Unit, Salinity Value, Salinity Unit, Calibration Slope Value,

Calibration Slope Unit, Calibration Point, Calibration Number

Example:

A213 DO/RDO, X01039, 3.04, CH-1, RDO, 01/03/15 16:05:41, 123454, 1, Auto–Air,

100.0,%, 8.2, Torr, 149.6, Torr, 24.3, C(ATC), 749.7, mmHg, 0.1, ppt, 1.52, Torr/%sat, 1,

#1 <CR>

Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &

Time, Probe Serial Number, Calibration Point, Calibration Type, Calibration

Concentration Value, Calibration Concentration Unit, Set Zero Partial Pressure Value,

Set Zero Partial Pressure Unit, Calibration Partial Pressure Value, Calibration Partial

Pressure Unit, Solution Temperature Value, Solution Temperature Unit, Barometric

Pressure Value, Barometric Pressure Unit, Salinity Value, Salinity Unit, Calibration Slope

Value, Calibration Slope Unit, Calibration Point, Calibration Number

Example:

A213 DO/RDO, X01039, 3.04, CH-1, RDO, 12/16/12, 09:21:00, 123454, 1, Manual, 6.9, mg/l, 8.2, Torr, 149.6, Torr, 24.3, C(ATC), 749.7, mmHg, 0.1, ppt, 1.52, Torr/%sat, 1, #1

<CR>

Orion Star A210 Series Benchtop Meter User Manual |

89

Chapter 7

|

Data Transfer and Software Updates

Meter Software Upgrade Procedure

The software upgrade program for Orion Star A200 series and Orion A300 series meters has been tested with Microsoft Windows 7, XP and Vista operating systems only. We are always working to update our programs; however, the software update program has not been tested with any other operating systems at this time.

All computer screen savers and power management settings should be disabled when the meter is interfaced with the computer. The computer must not go into sleep/standby mode while the software update is performed. Many laptop computers go into sleep/standby mode when the lid is closed, so the lid should remain open throughout the software update. Laptop computers must also be plugged into a power supply when running the software update, performing the software update while the computer is on battery power is not recommended.

Note: Back up any stored meter data before upgrading the meter software.

1. Go to www.thermoscientific.com/OrionMeters , download the latest software for Orion Star

A200-A300 series meters and unzip/extract the software files to the computer’s desktop. a. There are two versions of the Orion Star A200-A300 series meter software update for

Windows 32-bit operating systems and Windows 64-bit operating systems. To determine which to use with your computer: i. For Windows 7: Click the Start button and then click Control Panel. In the

Control Panel window, click System if viewing by large icons or small icons or click System and Security and then click System if viewing by category. In the

System window, next to System Type, either 32-bit or 64-bit will be shown. ii. For Windows XP: Click the Start button and then click Control Panel. In the

Control Panel window, click the Performance and Maintenance link and then click the System icon or for Control Panel Classic View, just double-click the

System icon. In the System Properties window, click the General tab and under

System, if "x64 Edition" is shown, use the 64-bit version and if "x64 Edition" isn’t shown, use the 32-bit version.

2. Uninstall any previous versions of the Orion Star A200-A300 series meter software update using the Computer’s Programs and Features tool.

90 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 7

|

Data Transfer and Software Updates

3. Make sure the computer interface cable is connected with the meter and computer and fully

setup using the instructions in the Using the USB-to-Serial Computer Cable

or Using the

USB Computer Cable section.

4. Power on the meter. To ensure uninterrupted power through the update process, use the power adapter with benchtop meters and fully charged batteries or power adapter with portable meters.

5. Open the folder containing the unzipped/extracted software files and double click the setup.exe file to install the setup program onto the computer.

6. Follow all prompts to install the setup program onto the computer. If any security warning messages appear, select the run software option.

Orion Star A210 Series Benchtop Meter User Manual |

91

Chapter 7

|

Data Transfer and Software Updates

7. After successful installation of the setup program, a new StarA200_300SoftwareUpgrade icon will appear on the computer’s desktop. Double click the icon to launch the program.

8. Follow all prompts, including reviewing and accepting the licensing agreement, to start the software update. If any security warning messages appear, select the run software option.

9. Set the interface parameters for running the software update program. a. When using the USB-to-Serial computer cable, select Serial as the connection method, enter the COM port location (viewable in Device Manager under Ports) and enter the current meter baud rate (default setting is 9600). b. When using the USB computer cable, the software update will detect the USB connection (USB driver must be installed prior to launching the software update).

92 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Chapter 7

|

Data Transfer and Software Updates

10. Click the Upgrade Firmware button. Wait while the update is installed on the meter. a. Warning: The software update must not be interrupted once the Update Firmware

button is clicked. Interrupting the software update may corrupt the meter and cause it to

not power on correctly.

11. The update will take approximately 15 minutes. Once the software update is complete, the dialog box will show “Done!”

Thermo Scientific

12. Close the software update and detach the cable from the meter and computer.

Note: If the following message is shown and the USB cable is being used, turn the power to the meter off by unplugging the power adapter and/or remove batteries, disconnect the USB cable, close the software update

and retry this procedure using the RS232 cable.

Orion Star A210 Series Benchtop Meter User Manual |

93

C H A P T E R 8

8

Customer Services

For any questions or if you require assistance, contact our Technical Support Specialists:

·

Email [email protected]

·

Within the United States, call 1-800-225-1480

·

Outside the United States, call +1-978-232-6000 or fax +1-978-232-6031

For additional product information, contact your local authorized dealer, local Thermo Scientific

Orion technical sales representative or contact us using the Water and Laboratory Products

(WLP) information on the page back of this manual.

Visit www.thermoscientific.com/water to view Thermo Scientific Orion products and download product literature, user guides and manuals, software updates, and additional application and technical resources.

For the most current warranty information, refer to the Thermo Scientific Orion warranty card included on the Thermo Scientific Orion Star A210 Series Meter literature CD and available online at www.thermoscientific.com/water .

94 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 8

|

Customer Services

Troubleshooting Tips

If an issue is encountered while using the Orion Star A210 Series meter, ensure the correct power adapter is being used and try power cycling the meter: disconnect the power adapter from the meter, wait 15 seconds and reconnect the power adapter to the meter.

Meter Issue

The measurement value is flashing 9999 and displaying Over Range or

Under Range

The meter locks up while connected to a computer via the USB cable

The meter keypad is unresponsive when using the Star Com computer software

The measurement value freezes and will not change

Recommended Action

The measurement is outside the allowable measurement range. Verify that the correct channel is shown on the meter display. Make sure the electrode or sensor is fully connected with the meter.

Disconnect the USB cable from the meter and computer and then disconnect and reconnect the power adapter from the meter. Reestablish the meter and computer USB connection.

The keypad lock option is enabled in the Star Com computer program. To disable the keypad lock: in the Star Com program click the Settings icon, uncheck the box next to Keypad Lock and press the Save icon.

The meter display goes blank, shows random lines or intermittently freezes

The meter does not automatically recognize the pH buffer during calibration

The meter does not recognize the conductivity standard during calibration

The meter does not recognize an RDO optical or polarographic dissolved oxygen sensor

The read type is set to Auto-Read mode (AR icon shown on the display).

Press the measure (esc) key to take a new measurement or use the setup menu to change the read type to continuous.

Ensure the correct power supply for the Star A series meter is being used.

This power supply is different from the one supplied with Thermo Scientific

Orion Versa Star meters. The use of a surge protector or uninterrupted power supply (UPS) is also recommended. Perform a factor reset on the meter.

Verify the correct buffer set was selected in the setup menu. The meter uses raw mV readings to recognize the buffer. As the electrode ages or becomes dirty, its mV readings will drift. Check the buffers and clean the electrode according to the instructions in the electrode manual.

Verify that the correct nominal cell constant is entered in the setup menu for the conductivity sensor being calibrated. The cell constant is typically printed on the cable of the conductivity sensor. Verify that the conductivity standard is one that can be automatically recognized by the meter.

Recalibrate the conductivity sensor using new conductivity standard.

Make sure the dissolved oxygen sensor is fully connected with the meter and verify the correct channel is shown on the meter display. Wait about 15 seconds after connecting a dissolved oxygen sensor for the meter to recognize the DO sensor type. Press the measure (esc) key to initiate a new measurement and update the meter display.

Orion Star A210 Series Benchtop Meter User Manual |

95

Chapter 8

|

Customer Services

Meter Factory Reset Procedure

Warning: Performing a factory reset will erase the meter’s calibration log and data log and reset all meter setup parameters to the factory default settings.

1. In the measurement mode, press the setup key.

2. Press the p, q, t or u key to highlight Diagnostics and press the f3 (select) key.

3. Press the p or q key to highlight Factory Reset and press the f2 (select) key.

4. Press the f2 (yes) key to start the factory reset procedure.

5. Press the f3 (edit) key to access the popup number entry screen and enter the default meter password of 111111. a. Press the p, q, t or u key to highlight the number 1 and press the f3 (enter) key six times until 111111 is shown at the top of the screen. b. Press the f2 (done) key to save the value and exit the number entry screen.

6. Press the f2 (accept) key to initiate the factory reset.

7. Wait while the factory reset is performed. Once completed, the meter will go through a power cycle and then proceed to the measurement mode. All meter settings will be reset to the factory default settings and the calibration log and data log will be erased.

Meter User Reset Procedure

1. In the measurement mode, press the setup key.

2. Press the p, q, t or u key to highlight Diagnostics and press the f3 (select) key.

3. Press the p or q key to highlight User Reset and press the f2 (select) key.

4. Press the f2 (yes) key to initiate the user reset procedure.

5. Wait while the user reset is performed. Once completed, the meter will proceed to the measurement mode. All meter settings will be reset to the factory default settings; however the calibration log and data log will be retained.

96 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 8

|

Customer Services

Notice of Compliance

This equipment generates, uses, and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual, may cause interference to radio communications. It has been tested and found to comply with the limits for a Class A computing device pursuant to Subpart J of Part 15 of FCC Rules, which are designed to provide reasonable protection against such interference when operated in a commercial environment.

Operation of this equipment in a residential area is likely to cause interference in which case the user, at his own expense, will be required to take whatever measures may be required to correct the interference.

“This digital apparatus does not exceed the (Class A) limits for radio noise emissions from digital apparatus set out in the Radio Interference Regulations of the Canadian Department of

Communications.”

“Le present appareil numerique n’ emet pas de bruits radioelectriques depassant les limites applicables aux appareils numeriques (de la class A) prescrites dans le Reglement sur le brouillage radioelectrique edicte par le ministere des Communications du Canada.”

WEEE Compliance

This product is required to comply with the European Union’s Waste Electrical & Electronic

Equipment (WEEE) Directive 2002/96/EC. It is marked with the symbol shown here.

Further information on compliance with these directives, the recyclers in your country, and information on Thermo Scientific Orion products that may assist the detection of substances subject to the RoHS Directive are available by contact us using the Water and Laboratory

Products (WLP) information on the back page of this manual.

Orion Star A210 Series Benchtop Meter User Manual |

97

Chapter 8

|

Customer Services

Declaration of Conformity

Manufacturer: Thermo Fisher Scientific Inc

Address: Ayer Rajah Crescent

Blk 55 #04-16/24

Singapore 139949

Singapore

Hereby declares that the following products:

Thermo Scientific Orion Star A210 Series Benchtop Meters are rated 100 to 240 VAC,

50/60 Hz, 0.5A

Benchtop Meter Models:

Orion Star A211 pH Meter

Orion Star A212 Conductivity Meter

Orion Star A213 RDO/DO Meter

Orion Star A214 pH/ISE Meter

Orion Star A215 pH/Conductivity Meter

Orion Star A216 pH/RDO/DO Meter

Equipment Class:

Measurement, control and laboratory

Orion Star A-series meters are EMC Class A

Conforms with the following directives and standards:

EN61326-1:2013 Electromagnetic Compatibility (EMC Directive)

EN61010-1:2010

UL61010-1:2012

Electrical equipment for measurement, control and laboratory use -

EMC requirements

Safety Standards

Safety requirements for electrical equipment for measurement,

CAN/CSA C22.2 No. 61010-1:2012 control and laboratory use - general requirements.

Cheow Kwang Chan

QA/Regulatory Manager

Place and Date of Issue:

16 February 2015

Singapore

98 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 8

|

Customer Services

Meter Specifications

Orion Star A210 Series Benchtop Meter Operating Conditions

Operating Ambient Temperature

5 to 45 °C

Operating Relative Humidity

5 to 85%, non-condensing

Storage Temperature

Storage Relative Humidity

-20 to 60 °C

5 to 85%, non-condensing

Pollution

Overvoltage

Meter Weight

Meter Dimensions (H x W x D)

Regulatory and Safety

Degree 2

Category II

0.9 kg (2 lb)

11 cm x 18 cm x 24 cm (4.33” x 7.09” x 9.45”)

Power Rating

CE, TUV 3-1, FCC Class A

DC input: 9VDC, 1A

Battery: 4 x AA

Shock: drop test in packaging per ISTA #1A

Shock and Vibration

Enclosure (designed to meet)

Warranty

Vibration: shipping/handling per ISTA #1A

IP54

3 years

Universal Power Adapter Operating Conditions

Operating Ambient Temperature

0 to 50 °C

Operating Relative Humidity

Storage Temperature

Storage Relative Humidity

Pollution

Overvoltage

0 to 90%, non-condensing

-20 to 75 °C

0 to 90%, non-condensing

Degree 2

Category II

Orion Star A210 Series Benchtop Meter Specifications

Measurement Channels

1 to 2

Display

Measurement Modes

Data Log Memory

Data Log Functions

Data Log Transfer

Data Log Editing

Graphic LCD

Auto-Read, Continuous, Timed

2000 data points with time and date stamp

Automatic logging with Auto-Read and Timed measure modes; manual logging with Continuous measure mode

Transfer single, range or all data points to printer or computer

Delete all data points

Orion Star A210 Series Benchtop Meter User Manual |

99

Chapter 8

|

Customer Services

Calibration Log

Computer Software

Methods

Alarms

Sample ID

User ID

Communication Ports

Stirrer Probe Ports

Power Source

Orion Star A210 Series Benchtop Meter Specifications

10 calibrations per channel with time and date stamp

Orion Star Com data transfer software available free on website

10 per channel with password protection

Limit alarm, calibration alarm

Manual entry, auto-incremental or off

Multilanguage User Interface

Manual entry or off

RS232, USB

1

Universal AC power adapter, 90-260 VAC, 50-60 Hz or optional 4 AA batteries (average 800 hours of battery life)

English, Spanish, French, Italian, German and Chinese

Portuguese and Korean available with software update on web

Orion Star A211 pH Benchtop Meter Specifications

Measurement Channels

Range

1 – pH, mV, relative mV (RmV) or ORP with temperature

- 2.000 to 20.000

Resolution

0.1 / 0.01 / 0.001

Relative Accuracy ±0.002

pH Calibration Points

Up to 5

Calibration Editing Yes

Input Impedance

> 10^

12

ohms

Electrode Status

Range

Resolution

Onscreen indicator for good, fair or bad electrode status

±2000.0 mV

0.1 mV

mV / RmV

Relative Accuracy ±0.2 mV or ±0.05% of reading, whichever is greater

Relative mV Mode

Yes

E

H

ORP Mode

Range

Yes

-5 to 105 °C, 22 to 221 °F

Temperature

Resolution

0.1 °C, 0.1 °F

Relative Accuracy ±0.1 °C

Offset Calibration

1 point

Sensor

Inputs

Source Options

BNC

Pin Tip

8 pin MiniDIN

Manual or automatic with ATC probe pH or ORP electrode

Reference electrode

ATC temperature probe

100 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 8

|

Customer Services

Orion Star A212 Conductivity Benchtop Meter Specifications

Measurement Channels

Range

Resolution

1 – conductivity, salinity, TDS or resistivity with temperature

0.001 μS to 3000 mS

0.001 μS minimum, auto ranging, up to 4 significant digits

Conductivity

Salinity

Reference

Temperature

Temperature

Compensation

Cell Constants

5 °C, 10 °C, 15 °C, 20 °C, 25 °C

Linear (0 to 10.0%/ºC), nonlinear nLFn, nonlinear nLFu, EP

(USP), off

0.001 to 199.9 cm

-1

Calibration Points

Up to 5

Calibration Editing Yes

Range

0.06 to 80.00 psu, 0.05 to 42.00 ppt

Resolution

0.01 psu or 0.01 ppt minimum, auto ranging

Relative Accuracy 0.5% of reading ±1 digit

Type

Practical salinity (psu) or natural sea water (ppt)

Range

Resolution

0.001 to 200.0 ppm

0.001 ppm minimum, auto ranging, up to 4 significant digits

TDS

Resistivity

Relative Accuracy 0.5% of reading ±1 digit

TDS Factor

Linear (0.02 to 9.99)

Range

2 Ω to 100 MΩ

Resolution

1 Ω or 0.1 MΩ, auto ranging

Relative Accuracy 0.5% of reading ±1 digit

Range

-5 to 105 °C, 22 to 221 °F

Temperature

Resolution

0.1 °C, 0.1 °F

Relative Accuracy ±0.1 °C

Offset Calibration

1 point

Source Options

Manual or automatic with built-in temperature sensor

Sensor Input 8 pin MiniDIN

Conductivity sensor with built-in temperature

Orion Star A210 Series Benchtop Meter User Manual |

101

Chapter 8

|

Customer Services

Orion Star A213 RDO/DO Benchtop Meter Specifications

Measurement Channels

Polarographic

1 – dissolved oxygen as % saturation or mg/L with temperature

Concentration % Saturation

Dissolved

Oxygen

Range

Resolution

RDO Optical

Range

Resolution

Relative Accuracy

Barometric Pressure Correction

Salinity Correction Factor

Calibration Types

Compatible Sensors

Range

Resolution

0 to 90 mg/L

0.01 / 0.1 mg/L

0 to 600 %

0.1 / 1 %

±2 % saturation or ±2% of reading, whichever is greater

% Saturation Concentration

0 to 50 mg/L 0 to 500 %

0.01 / 0.1 mg/L 0.1 / 1 %

±0.1 mg/L up to 8 mg/L,

±0.2 mg/L 8 to 20 mg/L,

±10% of reading up to 50 mg/L

±2 % saturation ≤ 200 % saturation, ±10 % saturation

> 200 % saturation

400 to 850 mmHg, automatic using built-in barometer (±6 mmHg) or manual entry

0.0 to 45.0 ppt, automatic using manual entry of sample salinity

Water-saturated air, air-saturated water, manual (Winkler), zero point

Polarographic, RDO optical

0 to 50 °C, 32 to 122 °F

0.1 °C, 0.1 °F

Temperature

Relative Accuracy ±0.1 °C

Offset Calibration

1 point

Source Options

Sensor Input 9 pin MiniDIN

Automatic with built-in temperature sensor

Dissolved oxygen sensor with built-in temperature

102 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 8

|

Customer Services

Orion Star A214 pH/ISE Benchtop Meter Specifications

Measurement Channels

Range

Resolution

1 – pH, mV, RmV, ORP or ion concentration with temperature

- 2.000 to 20.000

0.1 / 0.01 / 0.001

pH

Relative Accuracy ±0.002

Calibration Points

Up to 5

Calibration Editing Yes

ISE (Ion

Concentration)

Input Impedance

Electrode Status

Range

Resolution

> 10^

12

ohms

Onscreen indicator for good, fair or bad electrode status

0.0001 to 19900

0.0001 minimum, 1 to 3 significant digits (user selectable)

Relative Accuracy ±0.2 mV or ±0.05% of reading, whichever is greater

Units

ppm, Molar, mg/L,%, ppb, none

Calibration Points

Up to 5

mV / RmV

Temperature

Sensor Inputs

Calibration Editing Yes

Advanced Features

Segmented (point-to-point) slope, non-linear selectable autoblank, low concentration range stability

Range

±2000.0 mV

Resolution

0.1 mV

Relative Accuracy ±0.2 mV or ±0.05% of reading, whichever is greater

Relative mV Mode

Yes

E

H

ORP Mode

Range

Yes

-5 to 105 °C, 22 to 221 °F

Resolution

0.1 °C, 0.1 °F

Relative Accuracy ±0.1 °C

Offset Calibration

1 point

Source Options

Manual or automatic with ATC probe

BNC

pH electrode, ORP electrode or ion selective electrode (ISE)

Pin Tip

8 pin MiniDIN

Reference electrode

ATC temperature probe

Orion Star A210 Series Benchtop Meter User Manual |

103

Chapter 8

|

Customer Services

Orion Star A215 pH/Conductivity Benchtop Meter Specifications

Measurement Channels

2 – Ch. 1: pH, mV, relative mV (RmV) or ORP with temperature

Ch. 2: conductivity, salinity, TDS or resistivity with temperature

Range

Resolution

- 2.000 to 20.000

0.1 / 0.01 / 0.001

Relative Accuracy ±0.002

pH Calibration Points

Up to 5

Calibration Editing Yes

Input Impedance

Electrode Status

> 10^

12

ohms

Onscreen indicator for good, fair or bad electrode status

mV / RmV

Conductivity

Salinity

Range

Resolution

±2000.0 mV

0.1 mV

Relative Accuracy ±0.2 mV or ±0.05% of reading, whichever is greater

Relative mV Mode

Yes

E

H

ORP Mode

Range

Yes

0.001 μS to 3000 mS

Resolution

0.001 μS minimum, auto ranging, up to 4 significant digits

Relative Accuracy 0.5% of reading ±1 digit > 3 μS, 0.5% of reading ±0.01 μS ≤ 3 μS

Reference Temp.

5 °C, 10 °C, 15 °C, 20 °C, 25 °C

Temp. Comp.

Linear , nonlinear nLFn, nonlinear nLFu, EP (USP), off

Calibration Points

Up to 5 with calibration editing option

Range

0.06 to 80.00 psu, 0.05 to 42.00 ppt

Resolution

0.01 psu or 0.01 ppt minimum, auto ranging

TDS

Resistivity

Relative Accuracy 0.5% of reading ±1 digit

Type

Practical salinity (psu) or natural sea water (ppt)

Range

Resolution

0.001 to 200.0 ppm

0.001 ppm minimum, auto ranging, up to 4 significant digits

Relative Accuracy 0.5% of reading ±1 digit

TDS Factor

Linear (0.02 to 9.99)

Range

Resolution

2 Ω to 100 MΩ

1 Ω or 0.1 MΩ, auto ranging

Temperature

Sensor

Inputs

Relative Accuracy 0.5% of reading ±1 digit

Range

-5 to 105 °C, 22 to 221 °F

Resolution

0.1 °C, 0.1 °F

Relative Accuracy ±0.1 °C

Offset Calibration

1 point

Source Options

Manual or automatic with ATC probe or built-in temp. sensor

BNC

Pin Tip

8 pin MiniDIN

pH or ORP electrode

Reference electrode

Conductivity sensor with built-in temperature or ATC probe

104 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 8

|

Customer Services

Orion Star A216 pH/RDO/DO Benchtop Meter Specifications

Measurement Channels

2 – Ch. 1: pH, mV, relative mV (RmV) or ORP with temperature

Ch. 2: dissolved oxygen as% saturation or mg/L with temperature

Range

Resolution

- 2.000 to 20.000

0.1 / 0.01 / 0.001

pH

Relative Accuracy ±0.002

Calibration Points

Up to 5

Calibration Editing Yes

Input Impedance

> 10^

12

ohms

mV / RmV

Dissolved

Oxygen

Temperature

Sensor

Inputs

Electrode Status

Range

Onscreen indicator for good, fair or bad electrode status

±2000.0 mV

Resolution

0.1 mV

Relative Accuracy ±0.2 mV or ±0.05% of reading, whichever is greater

Relative mV Mode

Yes

E

H

ORP Mode

Yes

Polarographic Concentration

Range

Resolution

0 to 90 mg/L

0.01 / 0.1 mg/L

% Saturation

0 to 600 %

0.1 / 1 %

±2 % saturation or ±2% of reading, whichever is greater

% Saturation

RDO Optical

Concentration

Range

Resolution

Relative Accuracy

0 to 50 mg/L

0.01 / 0.1 mg/L

±0.1 mg/L up to 8 mg/L,

±0.2 mg/L 8 to 20 mg/L,

±10% of reading up to 50 mg/L

400 to 850 mmHg, automatic using built-in barometer (±6 mmHg) or manual entry

0 to 500 %

0.1 / 1 %

±2 % saturation ≤ 200 % saturation, ±10 % saturation

> 200 % saturation

Barometric

Pressure Correction

Salinity Correction 0.0 to 45.0 ppt, automatic using manual entry of sample salinity

Calibration Types

Water-saturated air, air-saturated water, manual, zero point

Compatible Sensors Polarographic, RDO optical

Range

0 to 50 °C, 32 to 122 °F

Resolution

0.1 °C, 0.1 °F

Relative Accuracy ±0.1 °C

Offset Calibration

1 point

Source Options

BNC

Pin Tip

8 pin MiniDIN

9 pin MiniDIN

Manual or automatic with ATC probe or built-in temp. sensor pH or ORP electrode

Reference electrode

ATC temperature probe

Dissolved oxygen sensor with built-in temperature

Note: All specifications are subject to change without notice.

Orion Star A210 Series Benchtop Meter User Manual |

105

Chapter 8

|

Customer Services

Ordering Information

Cat. No. Description

STARA2115

STARA2116

STARA2117

Star A211 pH meter kit with electrode stand, universal power adapter, literature CD, printed quick start guide, computer interface cable and meter test certificate

·

8302BNUMD ROSS Ultra Triode refillable glass-body pH/ATC electrode

· 810199 ROSS pH buffer and storage solution kit

Star A211 pH meter kit with electrode stand, universal power adapter, literature CD, printed quick start guide, computer interface cable and meter test certificate

·

8157BNUMD ROSS Ultra Triode refillable epoxy-body pH/ATC electrode

· 810199 ROSS pH buffer and storage solution kit

Star A211 pH meter kit with electrode stand, universal power adapter, literature CD, printed quick start guide, computer interface cable and meter test certificate

·

9157BNMD Standard Triode refillable epoxy-body pH/ATC electrode

· 910199 pH buffer and storage solution kit

STARA2125

STARA2126

Star A212 conductivity meter kit with electrode stand, universal power adapter, literature

CD, printed quick start guide, computer interface cable and meter test certificate

·

013005MD DuraProbe 4-cell (K=0.475) epoxy-body conductivity sensor

·

011007 Orion 1413 μS conductivity standard, 5 x 60 mL

Star A212 conductivity meter kit with electrode stand, universal power adapter, literature

CD, printed quick start guide, computer interface cable and meter test certificate

·

013016MD Pure water 2-cell (K=0.1) conductivity sensor with detachable flow cell

·

011008 Orion 100 μS conductivity standard, 5 x 60 mL

STARA2135

STARA2136

Star A213 RDO/DO meter kit with electrode stand, universal power adapter, literature

CD, printed quick start guide, computer interface cable and meter test certificate

·

083005MD Polarographic DO sensor with calibration sleeve

· 080513 DO sensor maintenance kit

·

BOD adapter, funnel and stirrer

Star A213 RDO/DO meter kit with electrode stand, universal power adapter, literature

CD, printed quick start guide, computer interface cable and meter test certificate

· 086030MD Polarographic Auto-Stir BOD/DO sensor

·

080513 DO sensor maintenance kit

·

080514 DO sensor electrolyte solution

STARA2145

Star A214 pH/ISE meter kit with electrode stand, universal power adapter, literature CD, printed quick start guide, computer interface cable and meter test certificate

·

8102BNUWP ROSS Ultra refillable glass-body pH electrode

· 927007MD Orion stainless steel ATC temperature probe

·

096019 Orion Star stirrer probe

·

810199 ROSS pH buffer and storage solution kit

106 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 8

|

Customer Services

Cat. No.

STARA2146

STARA2147

STARA2148

Description

Star A214 pH/ISE meter kit with electrode stand, universal power adapter, literature CD, printed quick start guide, computer interface cable and meter test certificate

· 8102BNUWP ROSS Ultra refillable glass-body pH electrode

· 9512HPBNWP Orion high-performance ammonia ion selective electrode (ISE)

·

927007MD Orion stainless steel ATC temperature probe

·

096019 Orion Star stirrer probe

·

951007 Orion 1000 ppm ammonia standard, 475 mL

· 951210 Orion low-level ammonia ISA solution, 475 mL

·

951213 Orion ammonia electrode storage solution, 475 mL

Star A214 pH/ISE meter kit with electrode stand, universal power adapter, literature CD, printed quick start guide, computer interface cable and meter test certificate

· 8102BNUWP ROSS Ultra refillable glass-body pH electrode

·

9609BNWP Orion fluoride ion selective electrode (ISE)

·

927007MD Orion stainless steel ATC temperature probe

·

096019 Orion Star stirrer probe

· 040906 Orion 1 ppm with TISAB II fluoride standard, 475 mL

· 040907 Orion 2 ppm with TISAB II fluoride standard, 475 mL

·

040908 Orion 10 ppm with TISAB II fluoride standard, 475 mL

·

940909 Orion TISAB II solution, 1 gallon

Star A214 pH/ISE meter kit with electrode stand, universal power adapter, literature CD, printed quick start guide, computer interface cable and meter test certificate

·

8102BNUWP ROSS Ultra refillable glass-body pH electrode

·

8611BNWP ROSS sodium ion selective electrode (ISE) with standards and ISA

·

927007MD Orion stainless steel ATC temperature probe

·

096019 Orion Star stirrer probe

STARA2155

Star A215 pH/conductivity meter kit with electrode stand, universal power adapter, literature CD, printed quick start guide, computer interface cable and meter test certificate

·

8157BNUMD ROSS Ultra Triode refillable epoxy-body pH/ATC electrode

·

013005MD DuraProbe 4-cell (K=0.475) epoxy-body conductivity sensor

· 810199 ROSS pH buffer and storage solution kit

· 011007 Orion 1413 μS conductivity standard, 5 x 60 mL

STARA2165

Star A216 pH/RDO/DO meter kit with electrode stand, universal power adapter, literature

CD, printed quick start guide, computer interface cable and meter test certificate

·

8157BNUMD ROSS Ultra Triode refillable epoxy-body pH/ATC electrode

·

083005MD Polarographic DO sensor with calibration sleeve

· 810199 ROSS pH buffer and storage solution kit

·

080513 DO sensor maintenance kit

·

BOD adapter, funnel and stirrer

Orion Star A210 Series Benchtop Meter User Manual |

107

Chapter 8

|

Customer Services

Meter Accessories, Electrodes and Solutions

Cat. No. Description

STARA-BEA

STARA-HB

810017

1010003

1010053

1010005

1010006

096019

Star A and V

ERSA

S

TAR

benchtop meter-attachable electrode stand, includes electrode arm, holder and meter bracket

Freestanding weighted base for use with Star A and V

ERSA

S

TAR

electrode stand

Storage sleeve and base for 12mm diameter electrodes

Universal power adapter for Star A series meters

Star series RS232 computer cable

Star series RS232 to USB cable and adapter

Star series inkjet printer, 110V/220V, with RS232 printer cable

Orion Star stirrer probe, pin tip connector

927007MD

927005MD

928007MD

Orion ATC temperature probe with stainless steel body, MiniDIN connector

Orion ATC temperature probe with epoxy body, MiniDIN connector

Orion micro ATC temperature probe with stainless steel tip, MiniDIN connector

8102BNUWP ROSS Ultra glass-body refillable pH electrode, BNC connector

8156BNUWP ROSS Ultra epoxy-body refillable pH electrode, BNC connector

8172BNWP

8165BNWP

ROSS Sure-Flow glass-body refillable pH electrode, BNC connector

ROSS Sure-Flow epoxy-body refillable pH electrode, BNC connector

910001

910168

910104

910105

910686

910107

910918

910110

910112

8302BNUMD ROSS Ultra Triode glass-body refillable pH/ATC electrode, BNC & MiniDIN connectors

8157BNUMD ROSS Ultra Triode epoxy-body refillable pH/ATC electrode, BNC & MiniDIN connectors

8107BNUMD ROSS Ultra Triode epoxy-body gel-filled pH/ATC electrode, BNC & MiniDIN connectors

8135BNUWP ROSS Ultra epoxy-body refillable pH electrode with flat surface bulb, BNC connector

8163BNWP ROSS spear tip glass-body refillable pH electrode, BNC connector

8103BNUWP ROSS Ultra semi-micro glass-body refillable pH electrode, BNC connector

8115BNUWP ROSS Ultra semi-micro epoxy-body refillable pH electrode, BNC connector

8220BNWP ROSS micro glass-body refillable pH electrode, BNC connector

810199

810001

ROSS All-in-One pH buffer and storage solution kit, includes pH 4, 7, 10 buffers; ROSS storage solution; pH electrode cleaning solution; pH electrode storage bottle

ROSS pH electrode storage solution, 475 mL

Standard pH electrode storage solution, 475 mL

Orion pH 1.68 buffer, 475 mL

Orion pH 4.01 buffer, 475 mL

Orion pH 5.00 buffer, 475 mL

Orion pH 6.86 buffer, 475 mL

Orion pH 7.00 buffer, 475 mL

Orion pH 9.18 buffer, 475 mL

Orion pH 10.01 buffer, 475 mL

Orion pH 12.46 buffer, 475 mL

108 | Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific

Thermo Scientific

Chapter 8

|

Customer Services

040907

040908

9707BNWP

920707

930711

930710

8611BNWP

841108

841111

013005MD

013016MD

018020MD

011008

011007

011006

1010001

083005MD

086030MD

087010MD

Cat. No.

9678BNWP

9180BNMD

Description

Orion Sure-Flow ORP epoxy-body refillable electrode, BNC connector

Orion Triode ORP/ATC epoxy-body refillable electrode, BNC & MiniDIN connectors

967901

967961

Orion ORP standard solution, 475 mL

Orion ORP standard solution, 5 x 60 mL

9512HPBNWP Orion high-performance ammonia ion selective electrode, BNC connector

9512BNWP

951007

951211

951210

9609BNWP

940907

940909

040906

Orion standard ammonia ion selective electrode, BNC connector

Orion 1000 ppm ammonia standard, 475 mL

Orion ammonia ionic strength adjuster (ISA) with pH-indicating blue dye, 475 mL

Orion low level ammonia ISA with pH-indicating blue dye, 475 mL

Orion fluoride ion selective electrode, BNC connector

Orion 100 ppm fluoride standard, 475 mL

Orion TISAB II total ionic strength adjustment buffer for fluoride analysis, 3.8 L

Orion 1 ppm fluoride standard premixed with TISAB II, 475 mL

Orion 2 ppm fluoride standard premixed with TISAB II, 475 mL

Orion 10 ppm fluoride standard premixed with TISAB II, 475 mL

Orion nitrate ion selective electrode, BNC connector

Orion 1000 ppm nitrate standard, 475 mL

Orion nitrate ionic strength adjuster (ISA), 475 mL

Orion nitrate interference suppressor solution, 475 mL

ROSS sodium ion selective electrode, BNC connector

Orion 1000 ppm sodium standard, 475 mL

Orion sodium ionic strength adjuster (ISA), 475 mL

Orion DuraProbe 4-cell (K=0.475)conductivity/temperature sensor, MiniDIN connector

Orion pure water 2-cell (K=0.1) conductivity/temperature sensor, MiniDIN connector

Orion high range 2-cell conductivity sensor (K=10), MiniDIN connector

Orion 100 μS/cm conductivity standard, 5 x 60 mL

Orion 1413 μS/cm conductivity standard,, 5 x 60 mL

Orion 12.9 mS/cm conductivity standard, 5 x 60 mL

Orion conductivity verification resistor kit

Orion polarographic DO sensor with calibration sleeve, MiniDIN connector

Orion Auto-Stir BOD polarographic sensor with calibration sleeve, MiniDIN connector

RDO optical DO sensor with optical cap, cal sleeve and guard, MiniDIN connector

Visit www.thermoscientific.com/water for a complete listing of all available Thermo Scientific

Orion meters, electrodes, solutions and accessories.

Orion Star A210 Series Benchtop Meter User Manual |

109

thermoscientific.com/water

© 2015 Thermo Fisher Scientific Inc. All rights reserved. RDO is a registered trademark of In-Situ Inc. Microsoft, Windows and Microsoft Vista are registered trademarks of Microsoft Corporation. All other trademarks are the property of Thermo Fisher Scientific Inc. & its subsidiaries.

Water and Lab Products

North America

Toll Free: 1-800-225-1480

Tel: 1-978-232-6000

[email protected]

Germany

Tel: (49) 6184-90-6000 [email protected]

China

Tel: (86) 21-68654588 [email protected]

India

Tel: (91) 22-4157-8800 [email protected]

Singapore

Tel: (65) 6778-6876 [email protected]

Japan

Tel: (81) 045-453-9175 [email protected]

Australia

Tel: (613) 9757-4300

In Australia (1300) 735-295

[email protected]

68X000441 0615 RevA

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement

Table of contents