05-437 PRIMALERT ® 35 Area Radiation Monitor

190N Portable Neutron Survey Meter

ASM990 Advanced Survey Meter (ASM)

Radiation

Safety

Product Catalog

2007/2008

Fluke Biomedical.

The accuracy you need from a company you can trust.

1060AM-V Vehicle Mounted Radiation Detection System 451P Pressurized µR Ion Chamber Survey Meter

2

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

Radiation Safety

Product Catalog

2007/2008

Providing solutions, not just products

Today, biomeds, physicists, RSO’s, other medical personnel must meet increasing regulatory pressures, higher quality standards, and rapid technological growth, while performing their work faster and more efficiently than ever. Fluke Biomedical provides a diverse range of software and hardware tools to meet today’s challenges.

Service

Fluke Biomedical is dedicated to providing the best service within the healthcare industry. Equipped with the best- credentialed facilities, onsite experts, and full asset-management capabilities, Fluke Biomedical’s service team is always on call to take care of its customers. Fluke Biomedical’s world-class staff leads the industry in post- and pre-sale support, including helping customers choose the best products and accessories for their needs, technical support, product calibration, and repairs.

Regulatory compliance

Fluke Biomedical is a major force in the industry and, like its parent company Fluke, is a leader in quality. Fluke Biomedical operates to the most rigorous standards in the industry, including compliance with ISO 9001:2000, ISO 13485:2003,

FDA/QSR, and NRC/Part 50, Appendix B/Part 21 and adheres to

ISO 17025:2005, ANSI Z540, Mammography MQSA and CNSC.

Many of the Fluke Biomedical products are CE-marked and CSAcertified. In addition, the Global Calibration Laboratory holds its

NVLAP certification and is traceable to both the NIST & PTB.

Legacy

You may be familiar with some of our legacy brand names, including:

Victoreen

®

Nuclear Associates

Keithley

Metron

DNI Nevada

Bio-Tek Instruments

Fluke Biomedical has taken the best elements and products of these former brands and have incorporated them into the Fluke

Biomedical culture and product line available today.

Our newest catalog

Our Radiation Safety catalog contains a variety of survey meters and probes, area monitors, and other monitoring accessories that can help Radiation Safety Officers (RSOs), Health Physicists,

Emergency Responders and other radiation-minded professionals manage diagnostic imaging QA, regulatory compliance and radiation emergencies.

If you are interested in receiving catalogs or information about any of Fluke Biomedical’s other product-lines, please visit www.flukebiomedical.com/catalogs.

Catalogs are also available for the following product lines:

Biomedical Test

Diagnostic Imaging QA

Radiation Oncology QA

Service

Key features

Range: 10

-1

to 10

7

mR/h

Energy response: 10 % from

Fluke Biomedical leads

Life expectancy: up to 10 7 facture of biomedical test and simulation products,

No external power is needed

Three twisted-pair cables automated performance testing and documentation

Environmentally hardened for most trusted and accurate

Seismically tested with high

Microprocessor controlled tions for regulatory compliance.

About Fluke

Corporation

Fluke Biomedical is a division of Fluke Corporation.

Fluke Corporation is the world leader in the manufacture, distribution, and service of electronic test tools and software and is a wholly owned subsidiary of Danaher

Corporation (NYSE:DHR).

Elso Philips Service, Trenčín

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

Radiation Safety

Product Catalog

Postioning statement or for statement

Ion Chamber Survey Meters

451B Ion Chamber Survey Meter with Beta Slide .......................................................... 4-5

451P Pressurized μR Ion Chamber Survey Meter ............................................................6-7

451EXL 451 Assistant for Excel ...............................................................................................8

440RF/D Low Energy RF Shielded Survey Meter ................................................................9

Survey Meters and Probes

ASM-990 Series Advanced Survey Meter (ASM) ....................................................... 10-12

489-110D GM Pancake Probe ................................................................................................13

Geiger-Mueller and Scintillation Probe Selection Guide ......................................14-16

190N Portable Neutron Survey Meter ...................................................................................17

Area Monitors

190F Area Monitor/Frisker Count Rate Meter .....................................................................18

1060AM-V Radiation Detection System ....................................................................... 19-20

1060AM Digital Smart Detector Area Monitor ....................................................................21

05-437 PRIMALERT

®

35 Area Radiation Monitor ............................................................ 22

05-443/444 PRIMALERT Digital Area Monitors ............................................................... 23

05-450 PRIMALERT Digital Doorway Monitor .................................................................. 24

Radiation Protection

RS-500 The Lawrence Livermore Realistic Phantom .......................................................25

RS-550 The Fission-Product Phantom .................................................................................26

Fluke Biomedical

6045 Cochran Road

Cleveland, OH 44139-3303 U.S.A.

Fluke Biomedical Europe AS

Vegamot 8, N-7048 Trondheim,

Norway

For more information, contact us:

In the U.S.A. (800) 850-4608 or

Fax (440) 349-2307

In Europe/M-East/Africa

+47 73954700 or

Fax +47 73954701

From other countries

+1 (440) 248-9300 or

Fax +1 (440) 349-2307

Email: [email protected]

Web access: www.flukebiomedical.com

Service Center

1-800-850-4608 ext. 2564

Email: [email protected]

Repair/Calibration instructions:

Request a SRA (Service Return

Authorization) by calling

440-498-2564 or emailing [email protected]. Shipping instructions will be provided when you receive your SRA.

Elso Philips Service, Trenčín

4

0.8

0.7

0.6

0.5

0.4

0.3

0.2

1 0

1.2

1.1

1

0.9

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

451B

Ion Chamber Survey

Meter with Beta Slide

The auto-ranging 451B measures radiation rate and accumulated dose from beta, gamma and x-ray radiation sources. The 451B’s site surveying capabilities make it well-suited for a wide range of end users, including: police and fire departments, x-ray manufacturers, government agencies, state inspectors, emergency response and

HAZMAT teams, nuclear medicine labs, hospital radiation safety officers, and nuclear power workers.

The ion chamber detector allows for a fast response time to radiation from leakage, scatter beams and pinholes. Additionally, the low noise chamber bias supply provides for fast background settling time. A sliding beta shield serves as an equilibrium thickness for photon measurements and enables beta discrimination.

The digital display features an analog bar graph,

2.5 digit digital readout, low battery and freeze

(“peak hold”) mode indicators, and an automatic backlight function. User controls consist of an

ON/OFF button and a MODE button. The case is constructed of lightweight, high strength materials and is sealed against moisture. The RS-232 interface can be connected directly to a computer for use with the

Excel add-in for Windows

®

(451EXL), enhancing the functionality of the instrument. This software allows for data retrieval, user parameter selection and provides a virtual instrument display with audible

(requires sound card) and visual alarm indication.

Key features

High sensitivity measurement of rate and dose simultaneously, with the capability to record peak rate

Auto-ranging and autozeroing

RS-232 communications interface with optional

Windows-based Excel add-in for data logging

Ergonomic, anti-fatigue handle with replaceable grip, wrist strap and tripod mount

Programmable flashing LCD display and audible alarm

Easily-accessible battery door

(operated by two 9-volt alkaline batteries) on the outside of the bottom case

Available with dose equivalent energy response (SI units)

451B typical energy dependence

1 0 0

Effective (keV)

1 0 0 0

Side

Slide - Closed

Slide - Open

1 0 0 0 0

Elso Philips Service, Trenčín

Ion Chamber Survey Meters

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

451B

Ion Chamber Survey

Meter with Beta Slide

Specifications

Radiation detected Alpha above 7.5 MeV, Beta above 100 keV, and Gamma above

7 keV

Operating ranges

Accuracy

0 to 5 mR/h or 0 to 50 μSv/h

0 to 50 mR/h or 0 to 500 μSv/h

0 to 500 mR/h or 0 to 5 mSv/h

0 to 5 R/h or 0 to 50 mSv/h

0 to 50 R/h or 0 to 500 mSv/h

Within 10 % of reading between 10 % and 100 % of full scale indication on any range, exclusive of energy response.

Calibration source is

137

Cs.

Detector

Chamber

Chamber wall

Chamber window

Beta slide

451B-DE-SI

Controls

Automatic features

Response time

349 cc volume air ionization

246 mg/cm

2

thick phenolic

6.6 mg/cm

2

mylar, protected by steel mesh, 46 cm

2

detection area

440 mg/cm

2

In order to achieve energy response consistent with measurements of H*(10) as required by ICR4-47, aluminum has been added to the back wall, 38 % of the side wall area, and to the beta slide. With the Beta Shield open, the 451B can measure skin dose at 10*(0.07), and Deep Dose H*(10) with

Beta Shield closed.

ON/OFF and MODE

Auto-zeroing, auto-ranging, and auto-backlight

Range

0 to 5 mR/h (0 to 50 μSv/h)

Response

8 seconds

0 to 50 mR/h (0 to 500 μSv/h) 2.5 seconds

0 to 500 mR/h (0 to 5 mSv/h) 2 seconds

0 to 5 R/h (0 to 50 mSv/h) 2 seconds

0 to 50 R/h (0 to 500 mSv/h) 2 seconds

Display LCD analog/digital with backlight

Analog

Digital

100 element bar graph 6.4 cm long. Bar graph is divided into 5 major segments, each labeled with the appropriate value for the range of the instrument.

2.5 digit display is followed by a significant zero digit depending on the operating range of the instrument. The units of measurement are indicated on the display at all times. Digits are 6.4 mm (0.25 in) high. Low battery and freeze indicators are also provided on the display.

Modes

Integrate mode

Freeze mode

Operates continuously 30 seconds after the instrument has been turned on. Integration is performed even if the instrument is displaying in mR/h or R/h.

Will place a tick mark on the bar graph display to hold on the peak displayed value. The unit will continue to read and display current radiation values.

Environmental

Power requirements

Warm-up time

Temperature range

Relative humidity

Geotropism

Dimensions (WxDxH)

Weight

Two 9 V alkaline, 200 hours operation

One minute

-20 °C to 70 °C (-4 °F to 158 °F)

0 to 100 %, @ 60 °C

Less than 1 %

10 cm x 20 cm x 15 cm (4 in x 8 in x 6 in)

1.11 kg (2.5 lb)

Optional accessories

451EXL 451 Assistant for Excel, includes RS-232 interface cable

190HPS Single Unit Carrying

Case

450UCS Check Source,

238

Uranium, 0.064 μCi, impregnated 2 x 2 in yellow card

Ordering information

451B-RYR Ion Chamber Survey

Meter with Beta Slide and standard chamber

451B-DE-SI-RYR Ion Chamber

Survey Meter with Beta Slide and dose equivalent chamber

Elso Philips Service, Trenčín

Ion Chamber Survey Meters

5

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

451P

Pressurized µR Ion Chamber

Survey Meter

The auto-ranging 451P features a pressurized ion chamber, providing enhanced sensitivity

(μR resolution) and improved energy response to measure radiation rate and dose from x-ray and gamma sources.

Originally designed to measure leakage and scatter around diagnostic x-ray and radiation therapy suites, the 451P’s site surveying capabilities make it well-suited for a wide range of end users, including: x-ray manufacturers, government agencies, state inspectors, biomedical technicians, and maintenance technicians for airport baggage scanners.

The ion chamber detector allows for a fast response time to radiation from leakage, scatter beams and pinholes. Additionally, the low noise chamber bias supply provides for fast background settling time.

The digital display features an analog bar graph, 2.5 digit digital readout, low battery and freeze (“peak hold”) mode indicators, and an automatic backlight function. User controls consist of an ON/OFF button and a MODE button. The case is constructed of lightweight, high strength materials and is sealed against moisture. The RS-232 interface can be connected directly to a computer for use with the Excel add-in for Windows (451EXL), enhancing the functionality of the instrument. This software allows for data retrieval, user parameter selection and provides a virtual instrument display with audible (requires sound card) and visual alarm indication.

Key features

High sensitivity μR meas- urements of rate and dose simultaneously, with the capability to record peak rate

Ergonomic, anti-fatigue handle with replaceable grip, wrist strap and tripod mount

Programmable flashing LCD display and audible alarm

Easily-accessible battery door (operated by two 9-volt alkaline batteries) on the outside of the bottom case

RS-232 communications interface with optional

Windows-based Excel add-in for data logging

Available with dose equivalent energy response (SI units)

6

1.2

1.1

1

0.9

0.8

0.7

0.6

0.5

0.4

0.3

0.2

1 0

Ion Chamber Survey Meters

451P typical energy dependence

1 0 0

Effective (keV)

1 0 0 0

Side

Face

Front

1 0 0 0 0

Typical energy dependence

16

Nitrogen gamma rays are

110 % to 120 % of indicated readings as determined at the

University of Lowell

Elso Philips Service, Trenčín

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

451P

Pressurized µR Ion Chamber

Survey Meter

Specifications

Radiation detected

Operating ranges

Beta above 1 MeV, Gamma and x-rays above 25 keV

Accuracy

Detector

Chamber

Controls

Automatic features

Response time

Analog response time from

10 % to 90 % of reading for a full scale step increase is dependent on operating range.

Response time for a step increase in radiation exposure rate from background:

230 cc volume pressurized air ionization chamber to 8 atmospheres or 125 psi

ON/OFF and MODE

Auto-zeroing, auto-ranging, and auto-backlight

Step increase, background to

400 μR/h

4 mR/h

10 mR/h

Time to reach 90 % of final value

4.8 sec

3.3 sec

4.3 sec

This table shows time measured from 10 % to 90 % of final value for a step increase or decrease in exposure rate such that a range change does not occur.

These values are the response times for the various ranges:

Range

0 to 500 μR/h (5 μSv/h)

0 to 5 mR/h (50 μSv/h)

0 to 50 mR/h (500 μSv/h)

10 % to 90 %

5 sec

2 sec

1.8 sec

0 to 500 mR/h (5 mSv/h)

0 to 5 R/h (50 mSv/h)

1.8 sec

1.8 sec

Analog/Digital display LCD with backlight

Analog 100 element bar graph 6.4 cm (2.5 in) long. Bar graph is divided into five major segments, each labeled with the appropriate value for the range of the instrument.

Digital 2.5 digit display is followed by a significant zero digit depending on the operating range of the instrument. The units of measurement are indicated on the display at all times. Digits are 0.25 inches

(6.4 mm) high. Low battery and freeze indicators are also provided on the display.

Modes

Integrate mode

40 mR/h

100 mR/h

1 R/h

4 R/h

4.5 sec

2.7 sec

2 sec

2.7 sec

Freeze mode

Operates continuously 30 seconds after the instrument has been turned on. Integration is performed even if the instrument is displaying in mR/h or R/h.

Will place a tick mark on the bar graph display to hold on the peak displayed value. The unit will continue to read and display current radiation values.

Environmental

Power requirements

Warm-up time

0 to 500 μR/h or 0 to 5 μSv/h

0 to 5 mR/h or 0 to 50 μSv/h

0 to 50 mR/h or 0 to 500 μSv/h

0 to 500 R/h or 0 to 5 mSv/h

0 to 5 R/h or 0 to 50 mSv/h

Within 10 % of reading between 10 % and 100 % of full scale indication on any range, exclusive of energy response. Calibration source is

137

Cs

Temperature range

Relative humidity

Geotropism

Dimensions (WxDxH)

Weight

Two 9 V alkaline, 200 hours operation

Less than two minutes for initial operation when the instrument is in equilibrium with ambient temperature.

-20 °C to 50 °C (-4 °F to 122 °F)

0 to 100 %

Negligible

10 cm x 20 cm x 15 cm (4 in x 8 in x 6 in)

1.07 kg (2.4 lb)

Optional accessories

451EXL 451 Assistant for Excel, includes RS-232 interface cable

190HPS Single Unit Carrying

Case

62-103 Check Source,

137

Cs,

10 μCi. Flat disc, 1 inch diameter

Ordering information

451P-RYR Pressurized μR Ion

Chamber Survey Meter with standard chamber

451P-DE-SI-RYR Pressurized μR

Ion Chamber Survey Meter with dose equivalent chamber

Note: Due to the pressurized ion chamber, the 451P is considered U.S.

Department of Transportation (DOT)

“Dangerous Goods” and must be shipped via IAW DOT special permit

DOT-SP 13187.

Elso Philips Service, Trenčín

Ion Chamber Survey Meters

7

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

451EXL

Assistant for Excel

The 451EXL provides remote control for many of the 451B and 451P functions via a

Microsoft

®

Excel-based user interface, including real-time data logging with user defined alarm parameters, upload of the internal data log into Excel worksheet, real-time virtual instrument display, and accumulated dose measurement over a user defined integration period. This information management software is ideal for the facility Radiation Safety Officer or anyone responsible for maintaining a permanent record of spills and accidents for adherence to state and NRC requirements.

The 451EXL’s data logging function automatically records realtime measured data into an Excel worksheet. The 451 Assistant provides user configurable audible and visual alarms for the real time logged data, including the color-coding of each data entry for quick identification for radiation levels and alarm acknowledgment status. This 451EXL information management software program is ideal for the facility radiation safety officer or anyone responsible for maintaining a permanent record of spills and accidents for adherence to state and NRC requirements.

Key features

Real time data logging and uploading of 451 internal data log into protected Excel worksheet

Virtual instrument display with user-defined audible and visual alarm indication

Compatible with Windows

®

2000 and above, and Excel

97, 2000

Package includes manual, diskette set, and 25 ft RS-232 cable, Model 1020039000

System requirements

Windows 2000 and above

Microsoft Excel 97 or 2000

One serial port (COM1 through

COM4)

Ordering information

451EXL 451 Assistant for Excel

Elso Philips Service, Trenčín

8

Ion Chamber Survey Meters

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

440RF/D

Low Energy RF Shielded

Survey Meter

The 440RF/D is a highly sensitive, low energy, RF shielded survey meter suited for fast, accurate measurements of background and other low radiation levels. In particular, the 440RF/D is used to measure radiation exposure in the color television industry (cathode-ray tube leakage), radar and transmission towers where RF may be present, and surveying applications in high RF fields.

Approved by the Electronic Industries Alliance, the

440RF/D is designed to meet the radiation sensitivity and measurement requirements for television receivers set forth by the US Department of Health and Human Services.

Entitled “Performance Standard for Televisions Receivers”

(21 CFR 1020.10), this standard requires that “radiation exposure rates produced by a television receiver shall not exceed 0.5 milliroentgens per hour at a distance of five centimeters from any point on the external surface of the receiver.”

Key features

Meets US Dept. HHS radiation survey requirements for Television Receiver and Cabinet

X-ray Systems requirements

(21 CFR 1020.10 and 21 CFR

1020.40, respectively)

Insensitive to 10 mW/cm

2

RF fields

Resolves 0.02 mR/h from

15 kV x-rays

Measures low energy radiation exposure down to 12 keV

Auto-zeroing

Batteries accessible from outside instrument

Specifications

Radiation detected Beta above 150 keV; gamma and x-ray above 12 keV

RF response No response in RF fields up to 10 mW/cm2

Accuracy Within 10 % of reading between 10 % and 100 % of full scale indication on any range, exclusive of energy response. Calibration at 21 keV x-ray

(27.5 KVCP, 0.9 mmA/HVL).

Geotropism Within 2 % of full scale in any orientation

Temperature range 20 °C to 40 °C (68 °F to 104 °F)

Relative humidity

Pressure dependence

0 to 95 %, non-condensing

Pressure transducer and temperature sensor automatically apply standard air density correction factors from 70 to 106 kPa to the unsealed ion chamber

Initial stabilization Approximately 30 seconds

Detector Internal ionization chamber 3.56 cm diameter by 5.87 cm long, cross sectional area 10 cm

2

volume, with 1.5 mg/cm

2

aluminized mylar window and an external magnesium window 13 mg/cm

2

thick. Center of ion chamber volume is 5 cm from the plane determined by the tips of three plastic bumpers.

Display 9.5 cm (3.7 in) meter scale, marked 0 to 3 and 0 to 10

Power requirements Five 9 V batteries; 200 hours operation. Three in parallel configuration for electronic supplies and 2 in series configuration for—18 V chamber bias.

Controls

Zero adjust

Single rotary switch and spring-loaded check source switch

Auto-zeroing

Check source

Housing material

Built-in operational uranium check source

All metal, splash-proof

Dimensions (WxDxH) 12.7 cm x 20.3 cm x 27.6 cm (5 in x 8 in x 10.8 in)

Weight 3.2 kg (6.8 lb)

Shipping vol/wt 0.071 m

3

(2.5 cu ft) 6.05 kg (13.3 lb)

1.2

1.0

0.8

15 kVcp

27.5 kVcp

42.5 kVcp

50 kVcp

250 kVcp

100 kVcp

200 kVcp

150 kVcp

137 Cs

60 Co

Typical energy dependence

X-ray and gamma ray: Within

10 % from 12.5 keV to 42 keV.

Maximum response peak of

+40 % at 100 keV. Within

10 % at

137

Cs and

60

Co with added equilibrium wall.

0.6

0.4

0.2

with added equilibrium wall material in front of window

1 0 0 1 0 0 0

Operating ranges

440RF/D exposure rate in five overlapping ranges: 0 to 1, 0 to 3, 0 to

10, 0 to 30, and 0 to 100 mR/h

440RF/D-SI dose equivalent rate in five overlapping ranges: 0 to 10,

0 to 30, 0 to 100, 0 to 300, and

0 to 1000 μSv/h

Response time 90 % of final indication in:

Range Response

7 sec 0 to 1 mR/h

(0 to 10 μSv/h)

0 to 3 mR/h

(0 to 30 μSv/h)

7 sec

5 sec 0 to 10 mR/h

(0 to 100 μSv/h)

0 to 30 mR/h

(0 to 300 μSv/h)

0 to 100 mR/h

(0 to 1000 μSv/h)

5 sec

5 sec

Ordering information

440RF/D Low Energy RF

Shielded Survey Meter

Elso Philips Service, Trenčín

Ion Chamber Survey Meters

9

10

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

ASM-990 Series

Advanced Survey Meter (ASM)

The ASM-990 series can detect alpha, beta, gamma, or x-ray radiation within an operating range of 1 μR/hr to 1 R/hr (1 to 5,000,000 CPM), depending on the selected probe (Geiger-

Mueller, neutron, proportional counter, scintillation). With the proper probe combination, this meter can be used as a general survey meter, an area monitor, a wipe test counter, and a contamination monitor.

Designed to meet the high technology requirements of health physics, medical physics, and nondestructive testing applications, the ASM-990 Series is well-suited for a wide range of end users, including: Radiation safety officers (RSO), nuclear medicine laboratories, diagnostic x-ray and hospital emergency room technicians, environmental health physicists, and emergency responders.

The units, with purchased probe, are shipped calibrated, ready-to-use, and include a MHV connector to ensure compatibility with all Fluke Biomedical probes. The 992 includes a fully calibrated internal energy compensated

1 R/hr GM detector. The 993 features a fully calibrated internal pancake detector as well as an internal energy compensated 1 R/hr GM detector.

Key features

Auto-scaling measurement of rate and dose simultaneously, with the capability to record peak rate

Up to five different probes can be calibrated with one unit

Data logging survey mode feature allows user to store up to 5 separate survey sequences

Saved data can be uploaded to a PC via included Infrared

Data (IrDA) transmitter

Easy-to-use multifunction keypad for intuitive menu navigation

Backlit analog/digital LCD display with full-range audio output capability

Barcode scanner (optional)

Auto Power-Down feature extends battery life

Specifications

ASM-990 and ASM-992

Operating modes

Operating rate ranges

(dependent on selected probe)

Complementary units in the integrate mode with the integrated time value in seconds

Accuracy

Detector

Temperature range

Relative humidity

Warm up time

Check source

Power requirements

Housing material

Display

Rate

Integrate

µR/hr

μrem/hr

μSv/hr

CPM

DPM 99mTc

Bq

125

I

μCi

67

Ga

µR

μrem

μSv

C (counts)

D (distintigrations)

Scaler (dual option): “Based On

Measurement” or “Based On Time” mR/hr mrem/hr mSv/hr

CPS

DPS 131I kBq mCi mR mrem mSv kC

123

18

F

I kD 99mTc

Timed Peak Hold

Data Logging

R/hr rem/hr

Sv/hr

MBq

201

Tl

Ci

57

Co

R rem

Sv

MC

MC 131I

Within 10 % of reading between 10 % to 100 % of full scale indication on any range, exclusive of typical energy dependence. Accuracy is probe dependent.

Accepts GM detectors and scintillation probes operating at high voltages between 500 and 1300 volts

-10 °C to +50 °C (14 °F to 122 °F)

0 to 95 %, non-condensing

5 second diagnostic check

Natural uranium, mounted on the case

Two “D” cells, 150 hours operation, automatically indicates when battery is low

Proprietary polycarbonate, splash-proof case

Liquid crystal display, 5.6 cm x 5.6 cm (2.2 in x 2.2 in)

Elso Philips Service, Trenčín

Survey Meters and Probes

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

ASM-990 Series

Advanced Survey Meter (ASM)

Data logging modes

The ASM-990 series’ Log Data feature can easily be accessed via the Setup Sub-menu. The unit can log/save a maximum of 500 data points in any of three separate modes (Manual and Survey modes can utilize the optional barcode scanner).

Manual: Individual rate data points can be saved by pressing the Start/Stop/Rst/

Save button.

Timed: A data point will automatically be saved at user selectable time intervals in the range of 1 to 255 seconds.

Survey: Programmed sequences can be accessed via the menu system. Pressing the Start/Stop/

Rst/Save button saves the current reading and displays the next survey location.

Programming of survey sequences, as well as retrieval of logged data, is accomplished via the built-in IrDA port.

Label names up to 20 characters can be programmed into the unit to identify the individual survey locations

Probe connector: The unit is available with a MHV connector. The unit can be used with multiple probes (5 total) by selecting the appropriate probe from the main menu. All calibration data for each probe is stored in the unit’s EEPROM.

Specifications

ASM-992 and ASM-993

Range

Radiation detected

Accuracy

Weight (without probe)

Dimensions (WxDxH)

0.1 mR/hr to 1 R/hr

Gamma above 60 keV

± 10 % of reading between 10 % and 100 % of full scale on any range, exclusive of energy dependence

ASM 990, 992: 0.95 kg (2.1 lb)

ASM 993: 1.09 kg (2.4 lb)

10.47 cm x 27.71 cm x 6.35 cm

(4.125 in x 10.91 in x 2.5 in)

Typical energy dependence

Internal energy compensated 1 R GM detector

1.4000

1.2000

1.0000

0.8000

0.6000

0.4000

0.2000

0.0000

1 0 1 0 0 1 0 0 0

Energy (keV)

1 0 0 0 0

Elso Philips Service, Trenčín

Survey Meters and Probes

11

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

ASM-990 Series

Advanced Survey Meter (ASM)

12

Specifications

ASM-993

Radiation detected

Range

Window

Typical background

Protective screen

Accuracy

Efficiency

The internal pancake detector efficiency is shown below. In a recent performance check, the numbers shown represent typical results obtained:

Note: The efficiency formula used to calculate the %Efficiency is:

Eff. % = (CPM x 100)/DPM

Alpha above 3.5 MeV, beta above 35 keV and gamma above 6 keV

Background to 80 mR/hr

15 cm

2

(1.75 in Ø) mica, 1.4 to 2.0 mg/cm

2

30 CPM

Stainless steel, hexagonal pattern providing 86 % open area

± 10 % of reading between 10 % and 100 % of full scale on any range, exclusive of energy dependence (protective cover open)

Isotope

14

C

%Efficiency

5 %

99

Tc

137 Cs

90 Sr

36

241

129

Cl

230

I

Am

Th

12 %

24 %

59 %

26 %

8 %

2 %

15 %

239 Pu 12 %

Typical energy dependence

Internal pancake detector

5

4.5

4

3.5

3

2.5

2

1.5

1

0.5

0

10 100

125

Energy (keV)

1000

137

Cs

60

Co

Available model(s)

ASM-990

Series

990

990BC

992

992BC

993

993BC

Advanced survey meter

Barcode reader

Internal energy compensated

1 R/hr GM detector

Internal pancake detector

10000

Survey Meters and Probes

Optional accessories

990-IR-USB USB Port IrDA

Adapter

990CC Carrying Case

990WM Wall Mounting Bracket

990PH Probe Holder for

489-110D

990UPH Universal Probe Holder

990SH Soft-Sided Holster

990SA Shoulder Strap Assembly

Note: The shoulder strap assembly is only available for the ASM-993 and must be ordered with the instrument and factory installed.

Note: The ASM-990 series, with the customer selected probe is calibrated to NIST standards. The ASM-990 series with GM probe standard calibration is in R, Sv, and rems.

Scintillation detectors are calibrated in counts. Radionuclide specific efficiency calibrations are available upon request. For probe selection and calibration services, see next page.

Ordering information

990 Advanced Survey Meter

990BC Advanced Survey Meter with barcode reader

992 Advanced Survey Meter with an internal 1 R GM detector

992BC Advanced Survey Meter with an internal 1 R GM detector and barcode reader

993 Advanced Survey Meter with an internal 1 R GM detector and internal pancake detector

993BC Advanced Survey Meter with an internal 1 R GM detector, internal pancake detector, and barcode reader

Elso Philips Service, Trenčín

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

489-110D

GM Pancake Probe

Key features

All purpose GM Pancake

Probe detects alpha, beta, gamma, and x-ray radiations

High detection efficiency

Lightweight, ergonomic design

Detachable probe cable

BNC or MHV connector

Easy to decontaminate

Specifications

Detector

Radiation detected

Operating voltage

Window

Typical background

Sensitivity

Protective screen

Housing material

Cable

Dimensions

Designed for use in conjunction with the ASM-990 Series and other standard GM survey meters, the 489-110D can detect alpha, beta, and gamma radiation. It is configured for operating convenience in table-top and floor surveys as well as surveys of personnel and equipment. Prime applications for this probe include nuclear medicine counter tops and frisker stations, leakage detection for low energy diagnostic x-ray machines, geological and environmental surveys or any place where there exists the suspicion that some form of radiation is present, especially emergency response teams.

For storage and carrying ease, the probe fits into the standard handle clip on a survey meter.

The GM probe comes in two configurations: 489-110D with an ABS plastic housing, MHV connector, and foam grip, or

489-110E with a BNC connector. This selection of connectors provides the ability to attach the GM probe to most GM survey meters on the market. Replacement foam grip handles are available: 489-130-44. This same probe design is available in a rugged metal housing as 489-110C.

Weight (pancake probe only)

Typical energy dependence

5

4.5

4

3.5

3

2.5

2

1.5

1

0.5

0

10

125

GM Pancake Probe typical energy dependence

Efficiency

100 1000

137

Cs

Effective (keV)

60

Co

489-110D GM Pancake Probe efficiency is shown below. In a recent performance check, the numbers shown represent typical results obtained:

Isotope

14

99

C

Tc

137 Cs

90

Sr

36

241

Cl

Am

129

I

230

Th

%Efficiency

5

12

24

59

26

8

2

15

239

Pu 12

Note: The efficiency formula used to calculate the % Efficiency is: Eff. % =

(CPM x 100)/DPM

Halogen-quenched “Pancake” GM tube

Alpha above 3.5 MeV, beta above 35 keV and gamma above 6 keV

900 V; compatible with all GM survey meters

15 cm

2

(1.75 in Ø) mica, 1.4 to 2 mg/cm

2

thick

30 CPM

3500 CPM/mR/hr

Stainless steel, hexagonal pattern providing 86 % open area

Replacement parts

489-130-44 Foam Grip Handle

ABS plastic housing and foam grip handle

Shielded cord; approximately 4.5 ft long MHV coaxial connector or

BNC connector

Detector housing (WxDxH): 6.36 cm x 2.2 cm x 10.8 cm

(2.50 in x 0.875 in x 4.25 in)

Handle (excluding connector): 2.5 cm Ø x 16.5 cm dia.

(1 in Ø x 6.25 in dia.)

0.28 kg (0.625 lb)

Ordering information

489-110D GM Pancake Probe with ABS plastic housing, MHV connector, and foam grip handle

489-110E GM Pancake Probe with ABS plastic housing, BNC connector, and foam grip handle

489-110C GM Pancake Probe with metal housing, MHV connector, and foam grip handle

Elso Philips Service, Trenčín

10000

Survey Meters and Probes

13

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

489-XXX, 90-12,

425-XXX, 190-100BGS

Geiger-Mueller and Scintillation Probe

Selection Guide

489-110D GM Pancake Probe

Alpha above 3.5 MeV

Beta above 35 keV

Gamma and x-ray > 6 keV

To 80 mR/hr (800 μSv/hr)

489-50 Gamma Scintillation Probe

Gamma and x-ray > 60 keV

1 in x 1 in, 1.5 in x 1.5 in and 2 in x 2 in NaI (Tl) detectors available

Key features

Rugged and reliable designs

GM probes are available in pancake style, energy compensated and with beta discrimination

Scintillators are test selected and optically coupled to photomultiplier tubes

489-200 Scintillation Pancake Probe

Beta above 100 keV

Gamma and x-ray > 25 keV

NaI (Tl) rectangular

90-12 Energy Compensated GM Probe

Beta above 200 keV

Gamma and x-ray > 12 keV

Up to 1 R/hr (10 mSv/hr)

489-35 Thin End Window GM Probe

Alpha above 4 MeV

Beta above 70 keV

Gamma and x-ray > 6 keV

Up to 80 mR/hr (800 μSv/hr)

489-60 Alpha Scintillation Probe

Alpha above 4 MeV

1.5 in Ø ZnS (Ag)

425-110 Low Energy Gamma Scintillation Probe

Gamma and x-ray > 10 keV

NaI (Tl) 1 mm thick

491-40 Utility 1 R/hr GM Probe

Beta above 200 keV

Gamma and x-ray > 12 keV

Up to 1 R/hr (10 mSv/hr)

425-200 Alpha/Beta Scintillation Probe

Alpha above 350 keV

Beta above 14 keV

Plastic scintillator

190-100BGS 100 cm 2

Scintillation Probe

Beta/Gamma

Beta

90

Sr efficiency 65.0 %

126 cm

2

active area

Plastic scintillator

GM probes for qualitative radiation detection—the

Geiger-Mueller (GM) detectors fulfill a wide variety of radiation measurement needs for alpha, beta, gamma and x-ray sources. Some probes have are provided with a 360° shield to permit discrimination between penetrating and non-penetrating radiation.

The GM detectors have a field-proven design to ensure dependable performance, reliability and ruggedness. Standard MHV type connectors readily allow interchange of all detector probes. The life expectancy of the counters ranges from

108 total counts to unlimited life depending on the type of quench gas utilized.

Enhanced sensitivity to lowlevel alpha, beta, gamma and x-ray radiation is achieved when using the unique

498-110D Pancake GM Probe.

The following probe selection guide lists various probes and suggested applications. Applications include nuclear medicine counter top surveys, leakage detection from diagnostic x-ray and linear accelerators, geological surveys, scrap metal yards and unknown wells.

Elso Philips Service, Trenčín

14

Survey Meters and Probes

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

489-XXX, 90-12,

425-XXX, 190-100BGS

Geiger-Mueller and Scintillation Probe

Selection Guide

Scintillation probes for quantitative radiation assessment (counts/minute) –

The scintillation detectors are optically coupled to photomultiplier tubes, which are then both magnetically shielded with mu-metal and specially shock-mounted to provide trouble-free performance. The entire detector, crystal and photomultiplier, are secured in a sturdy cylindrical aluminum housing. Where appropriate, a thin window has been utilized to provide alpha or low energy gamma response.

Applications outlined in this guide include nuclear medicine labs, HAZMAT spills, radiation safety office surveys, industrial hygiene, industrial x-ray manufacturing, and geological surveys.

Specifications

Scintillation probes

Model

Type

489-50

NaI (Tl) Sodium

Iodide 1 x 1, scintillator optically coupled to PMT

489-55

NaI (Tl) Sodium

Iodide 1.5 x 1.5, scintillator optically coupled to

PMT

489-120

NaI (Tl) Sodium

Iodide 2 x 2, scintillator optically coupled to

PMT

489-60

ZnS (Ag)

Alpha, scintillator optically coupled to PMT

425-110

NaI (Tl)

Thin Scintillator for Low Energy

Gamma, scintillator optically coupled to

PMT

Gamma and x-ray above 10 keV

425-200

NE 102A Plastic

Scintillator

Flashlight Probe, scintillator optically coupled to

PMT

489-200

NaI (Tl)

Pancake, scintillator optically coupled to PMT

Radiation detected

Applications

Gamma and x-ray above 60 keV

Nuclear medicine

Industrial hygiene

Gamma and x-ray above 60 keV

Industrial x-ray manufacturing

Geological surveys

Radiation safety office

Gamma and x-ray above 60 keV

Nuclear

medicine seed

finder

Alpha above

4 MeV

Alpha detection

Uranium,

Plutonium

HAZMAT

RSO

Primary probe

for nuclear

medicine

Low energy

x-ray

manufacturing

Industrial

hygiene

Alpha above

350 keV, beta above 14 keV

Alpha, beta

counting of

filter paper

HAZMAT

spills

Nuclear

medicine

missing sources

137

Cs 2 pts/scale to 100 mR/hr

137

Cs 2 pts/scale to 10 mR/hr

Sensitivity to

241

Am

± 10 %

Sensitivity to

129

I

Sensitivity to

99

Tc,

137

Cs,

14

C

90

Sr,

137 Cs 13 % 137 Cs 26 % 239

241

Pu 13 %

Am 8 %

90 Sr 22 %

36 Cl 8 %

241 Am 8 %

133 Ba 34 %

90 Sr 7 %

99 Tc 3 %

137 Cs 5 %

14 C 1 %

0 to 95 %

-40 °C to + 50 °C (-40 °F to +120 °F), maximum temperature increase of 20 °F/hr

0.68 kg (1.5 lb) 0.91 kg (2.0 lb) 0.68 kg (1.5 lb) 0.68 kg (1.5 lb) 0.35 kg (0.78 lb)

Gamma and x-ray above 25 keV, beta above

100 keV

Beta, gamma

frisker for

nuclear

medicine is 10

times more

sensitive than

GM probe

Environmental

surveys

3000 Typical background (CPM)

Nominal sensitivity

Wall material

Window

Sensitive area

Crystal dim.

Probe dia.

1750 5000 6000 20 200 38

160,000 CPM/ mR/hr 137

Cs

0.04 in Al,

1 mm thick

108 mg/cm

2

Al

5 cm

2

2.5 cm x 2.5 cm

(1 in x 1 in)

350,000 CPM/ mR/hr

137

Cs

0.04 in Al,

1 mm thick

108 mg/cm

2

Al

11.4 cm

2

3.8 cm x 3.8 cm

(1.5 in x 1.5 in)

700,000 CPM/ mR/hr

137

Cs

0.04 in Al,

1 mm thick

108 mg/cm

2

Al

20 cm

2

5.1 cm x 5.1 cm

(2 in x 2 in)

300,000 CPM/μCi

241

Am

0.04 in Al,

1 mm thick

3 mg/cm

2

Al Mylar

11.4 cm

2

3.8 cm Ø

(1.5 in Ø)

3,000,000 CPM/

μCi

129 l

0.04 in Al,

1 mm thick

8 mg/cm

2

Al

5 cm

2

2.5 cm Ø

(1 in Ø)

0.0012 CPM/DPM/

100 cm

63

0.04 in Al,

1 mm thick

0.25 mg/cm

Plastic

20.3 cm

2

2

Ni

2

650 CPM/μR/hr

137

Cs

0.04 in Al,

1 mm thick

130 mg/cm

2

Al

59.2 cm

2

5.1 cm (2 in) 5.1 cm (2 in) 5.7 cm (2.25 in) 5.1 cm (2 in) 5.1 cm (2 in)

5.1 cm Ø

(2 in Ø)

5.1 cm x 5.1 cm x

1.3 cm (2 in x 2 in x 0.5 in)

6.7 cm (2.625 in) 5.7 cm x 1.8 cm

(2.25 in x 0.69 in)

22.2 cm (8.75 in) 23.2 cm (9.125 in) 24.5 cm (9.625 in) 18.4 cm (7.25 in) 20.6 cm (8.125 in) 20.3 cm (8 in)

122 cm (48 in) 122 cm (48 in) 122 cm (48 in) 122 cm (48 in)

900 V

122 cm (48 in) 122 cm (48 in)

28 cm (11 in)

122 cm (48 in)

Probe length

Cable length

Operating voltage

Calibration

137

Cs 2 pts/scale to 10 mR/hr

Cal. tolerance

Efficiency 137 Cs 6 %

57 Co 9 %

133 Ba 6 %

60 Co 2 %

Humidity range

Operating temp

Weight (approx.) 0.68 kg (1.5 lb)

Sensitivity to

99m

90

60

Tc

SR 5 %

137

133

Cs 11 %

Ba 34 %

Co 16 %

0.35 kg (0.78 lb)

Elso Philips Service, Trenčín

Survey Meters and Probes

15

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

489-XXX, 90-12,

425-XXX, 190-100BGS

Geiger-Mueller and Scintillation Probe

Selection Guide

16

Specifications

Geiger-Mueller probes

Model

Type

Radiation detected

Applications

Typical background (shielded)

Maximum exposure rate

Nominal sensitivity to 1 mR/hr of 60

Co

Replacement GM tube part number

Wall material

80 mR/h

(800 μSv/hr)

3500 CPM

P-115

Wall thickness

Active length

Quenching gas

Diameter of probe

Length of probe

Cable length

Weight (approx.)

Operating voltage

Humidity range

Operating temperature range

Pressure range

489-110C/D/E*

Pancake alpha, beta, gamma, and x-ray with thin pancake window

Alpha above

3.5 MeV, beta above

35 keV, gamma and x-ray above 6 keV

All purpose

sensitive alpha,

beta, and gamma

and x-ray probe

Nuclear medicine

counter tops

Detects leakage

from diagnostic

x-ray machines,

especially

mammography

Geological surveys

Scrap metal yards

HAZMAT

30 CPM

90-12

Energy compensated beta, gamma, and x-ray with 360° linear movement shield for beta discrimination

Beta above 200 keV and gamma above

12 keV

Energy

compensated

to eliminate low

energy over

response

Convenient size

to fit in small

spaces around

linear

accelerators

X-ray tube

manufacturers

15 CPM

Stainless steel with mica window

1.5 to 2.0 mg/cm

2

38 mm (1.5 in Ø)

Neon and halogen

68 mm (2.6875 in)

248 mm (9.75 in)

122 cm (48 in)

0.45 kg (1.0 lb)

900 V

0 to 95 %

-56 °C to +85 °C

(-65 °F to +185 °F)

To 5 psig

1 R/h

(10 mSv/hr)

720 CPM

35-166

Stainless steel

40 to 60 mg/cm

2

19.1 mm (0.75 in)

Neon and halogen

35 mm (1.375 in)

170 mm (6.7 in)

122 cm (48 in)

0.26 kg (0.59 lb)

900 V

0 to 95 %

-56 °C to +85 °C

(-65 °F to +185 °F)

To 15 psig

489-35

Alpha, beta, gamma, and x-ray with

0.875 inch thin end window

Alpha above 4 MeV, beta above 70 keV, and gamma and x-ray above 6 keV

Ultra sensitive

alpha, beta,

gamma probe

with directional

focus

Nuclear medicine

Emergency

response

50 CPM

80 mR/h

(800 μSv/hr)

3900 CPM

489-76

493-50

Beta, gamma, and x-ray with sliding

360° metal shield for beta discrimination

Gamma above

12 keV and beta above 200 keV

Rugged probe

with beta

discrimination

Scrap metal yards

Rugged to drop

down wells

Nuclear medicine

15 CPM

1 R/h

(10 mSv/hr)

720 CPM

35-166

Stainless steel with mica window

1.4 to 2.0 mg/cm

2

Stainless steel

40 to 60 mg/cm

2

102 mm (4 in)

Neon and halogen

19.1 mm (0.75 in)

Neon and halogen

33.4 mm (1.3125 in) 32 mm (1.25 in)

191 mm (7.5 in)

122 cm (48 in)

0.45 kg (1.0 lb)

900 V

0 to 95 %

-56 °C to +85 °C

(-65 °F to +185 °F)

To 5 psig

84 mm (3.3125 in)

122 cm (48 in)

0.45 kg (1.0 lb)

900 V

0 to 95 %

-56 °C to +85 °C

(-65 °F to +185 °F)

To 15 psig

491-40

Beta, gamma, and x-ray with sliding

360° metal shield for

Beta discrimination

491-30

Beta, gamma, and x-ray with sliding

360° metal shield for beta discrimination

Gamma above

12 keV and beta above 200 keV

15 CPM

1 R/h

(10 mSv/hr)

720 CPM

35-166

Stainless steel

40 to 60 mg/cm

0.45 kg (1.0 lb)

900 V

0 to 95 %

To 15 psig

2

19.1 mm (0.75 in)

Neon and halogen

30 mm (1.1875 in)

136 mm (5.375 in)

122 cm (48 in)

-56 °C to +85 °C

(-65 °F to +185 °F)

Gamma above

12 keV and beta above 200 keV

Beta, gamma

probe is more

sensitive than

491-40 or 493-50,

but has max. rate

of 100 mR/hr

20 CPM

100 mR/h

(1 mSv/hr)

2200 CPM

35-150

Stainless steel

30 to 40 mg/cm

2

57.2 mm (2.25 in)

Neon and halogen

30 mm (1.1875 in)

136 mm (5.375 in)

122 cm (48 in)

0.45 kg (1.0 lb)

900 V

0 to 95 %

-56 °C to +85 °C

(-65 °F to +185 °F)

To 15 psig

Ordering information

489-110D GM Pancake Probe

489-50 Gamma Scintillation Probe

489-200 Scintillation Pancake Probe

90-12 Energy Compensated GM Probe

489-35 Thin End Window GM Probe

489-60 Alpha Scintillation Probe

425-110 Low Energy Gamma

Scintillation Probe

491-40 Utility 1 R/hr GM Probe

425-200 Alpha/Beta Scintillation

Probe

190-100BGS 100 cm 2

Scintillation Probe

Beta/Gamma

Elso Philips Service, Trenčín

Survey Meters and Probes

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

190N

Portable Neutron

Survey Meter

The self-contained 190N measures mRem in accordance with the classical Anderson and Braun design. The neutron probe can be attached to either a 190 Survey Meter or a 190F Frisker for continuous neutron surveys or area monitoring.

This product has all the salient features of an auto-scaling digital survey meter including data logging. Using the

190-lA Infrared Communicator, manual data logging or automatic preset time data logging is accessible for data handling. Neutron Probe, RP-N, can be interfaced to the 190F Frisker, with ac power for continuous monitoring.

Key features

Auto-scaling measurement of rate and dose (integrate mode includes dose and time accumulation)

True Rem readings recorded across a wider (lower and higher) rate range

Data logging with the 190-1A

Infrared Communicator to a PC

Ergonomic, portable design: adjustable shoulder strap and rugged handle with padded grip

Flexible detector assembly,

190 can be removed for remote readings

Available in SI units

Specifications

Readout

Alarm

Logging of data

Detector assembly,

RP-N

Programmable features of a standard 190 Survey Meter. Refer to the 190 data sheet for complete details.

Audio and visual setpoint can be programmed into the 190N via the 190-1A Infrared Communicator

The 190-1A Infrared Communicator interfaced to a personal computer can be used to set up data logging

The detector assembly is a polyethylene cylinder, 9.5 in L x 8.5 in dia., containing a BF3 proportional counter and neutron energy compensating materials. It is based upon the standard reliable Anderson and Braun design for neutron energy response. The handle is padded for ease of gripping. An adjustable shoulder strap is provided.

BF

3

operating characteristics

The BF

3

proportional counter operates at 1150 V. Active length is 5.08 cm (2 in). Fill gas is enriched BF expectancy is greater than 10

10

counts.

3

, 96 %

Boron 10. Gas pressure is 20 cm Hg. Resolving time is 1 microsecond, plateau slope is 2 % per 100 V and tube life

Nominal 2000 counts per mRem Typical neutron sensitivity

Range Rate: 0 μRem/h to 75 Rem/h

0 μSv/R to 0.75 Sv/h

0 CPM to 2.5 x 10

6

CPM

0 CPS to 41,660 CPS

Integrate: 0 μRem to 1000 Rem

0 μSv to 10 Sv

0 to 10 9 counts

No response in 137 Cs gamma radiation in fields up to 500 R/h Gamma sensitivity/ rejection

Accuracy

Dimensions

Miscellaneous

Weight

10 % of theoretical ICRP dose rate

31.75 Ø x 26 cm dia. (12.50 Ø x 10.25 in dia.)

Detector assembly cable length: 1.37 m (4.5 ft). An optional 9.14 m (30 ft) cable is available.

9.52 kg (21 lb) (total 190 + detector assembly)

Directionality Less than 20 % in three orthogonal directions

Temperature range 190 operating range: -10 °C to +40 °C (14 °F to 104 °F)

Detector assembly operating range: -80 °C to +80 °C (-112 °F to 176 °F)

Power requirements Four 9 V alkaline batteries supplied, 100 hours operation

Calibration 190N is calibrated against a NIST traceable “Tissue Equivalent Proportional Counter” and uses Radium/Beryllium neutrons at a distance of 100 cm

Typical energy dependence

10E1

10E0

10E-1

190N Sensitivity

ICRP Response

20 CPS/2.5 MREM/HR

10E-2

10E-3

10E-2 10E-1 10E0 10E1 10E2 10E3 10E4 10E5

190N Typical Energy Response

10E6 10E7 10E8

Survey Meters and Probes

Ordering information

190N Portable Neutron Survey

Meter

190N-SI Portable Neutron Survey

Meter, SI Unit

Elso Philips Service, Trenčín

17

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

190F

Specifications

Area Monitor/Frisker Count

Rate Meter

The easy-to-use, auto-ranging 190F is compatible with GM detectors, neutron probes, proportional counters, and scintillation probes operating from 300 volts to 1,300 volts. Depending on probe selection, the

190F detects alpha, beta, gamma, x-ray or neutron radiation within an operating range of 1 μR/h to 1 R/h (1 CPM to 1,000,000

CPM). The unit is available with either an

MHV or a BNC connector.

Visual indication of selected parameters, as well as measured values, are displayed on the analog/digital display.

The 190F Area Monitor/Frisker Count Rate

Meter, with purchased probe, is shipped calibrated and ready-to-use.

Accuracy

Detector

Adapter module

Log

Power requirements

Batteries

Within 10 % of reading between 10 % to 100 % of full scale indication on any range, exclusive of energy dependence. Accuracy is probe dependent.

Accepts GM detectors, neutron probes, scintillation probes, and proportional counters operating at high voltages between 300 V and 1300 V.

Contains calibration data and high voltage settings for a specified probe.

The module is available with an MHV or a BNC connector. Specify the type of connector with order.

Note: Additional adapter modules can be purchased for use with multiple probes: Specify 190060 for MHV adapter module and 190070 for BNC adapter module. By using multiple replaceable probe adaptor modules, each module can be assigned to a specific probe. The module’s

EEPROM stores the calibration factors for a specific probe. When plugged into a 190F Area Monitor and Count Rate Meter, it automatically sets the high voltage and activates the calibration data set for the specific probe. By using modules married to specific probes, the user has the convenience of using only one 190F with multiple probes for survey work.

Logs 211 data points and sequentially labels data points. (Data retrieval requires the 190-1A Infrared Communicator). With the communicator, alphanumerics up to 16 characters can be programmed into the 190F to name the locations of individual data points to be collected. The location name is displayed when the Log button is pressed. Press the Log button again, and the data point is stored.

9 V dc regulated power converter

Three 9 V batteries, 150 hours operation, automatically indicates when battery is low

Warm up time

Check source

Temperature range

Relative humidity

Housing material

15 second diagnostic check

Natural uranium, mounted on the case

-10 °C to +60 °C (14 °F to 140°F)

0 to 95 %, non-condensing

Molded ABS plastic, splash-proof case. Probe fits into side-mounted ABS plastic probe holder with Velcro

®

straps.

Dimensions (WxDxH) 9.2 cm x 23.4 cm x 5 cm (3.75 in x 9.2 in x 2.1 in)

Weight (without probe) 0.70 kg (1.56 lb)

Key features

Auto-scaling measurement of rate and dose simultaneously

Adjustable Alarm

Backlit analog/digital LCD display with bar graph and operational units

Interchangeable probe adapter module

Data logging

Operating ranges (dependent on selected probe)

Toggles and selects rate units:

μR/hr

CPM

μSv/hr

DPM mR/hr

CPS mSv/hr

Bq/cm 2

R/hr

μCi/cm 2 and the complementary units in the integrate mode:

R μR

CTS

μSv

Bq mR

D mSv

μCi with the integrated time value in seconds

Optional accessories

190-1A Infrared Communicator

Additional features can be activated, such as Log Mode,

Alarm Setpoint, Energy Specific

Calibrations, and default setting changes. Features and pushbuttons can also be locked-out to set up the 190F in a user defined mode of operation.

Note: The 190F Area Monitor and

Count Rate Meter, with the customer selected probe is calibrated to NIST standards. The 190F and probe is calibrated in mR/h or μSv/h units as a standard. The end user may calibrate in additional radiation units using the 190-1A Infrared

Communicator.

Ordering information

190F Area Monitor/Frisker Count

Rate Meter

Elso Philips Service, Trenčín

Area Monitors

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

1060AM-V

Radiation Detection System

The 1060AM-V is a vehiclemounted radiation detector that can alert first responders to the presence of gamma radiation.

This product is designed to warn a response team prior to leaving the truck and entering a field of radiation so that the truck can assess an incident scene with a minimum of exposure.

It is available in environmental, low, medium, and high range versions employing internal Geiger-Mueller (GM) detectors. Circuitry that detects pulse pileup conditions in a high radiation field is employed to prevent erroneous readings.

WIN1060 PC software provides the ability to display multiple channels, to maintain both alarm and measurement history, and to access system configuration options. The 1060DA remote display, consisting of a visual alarm indicator and a 100 decibel audible alarm along with a self scaling digital readout, is also available.

Key features

Available in 4 operating ranges: environmental, low, medium and high

Self contained, reliable, continuous gamma radiation monitors

Cab view digital display with alarming light and siren

Wide range of applications in

NEMA 4 Enclosure

Optional computer display software, WIN1060 Windows

® for cab or remote viewing

EMI shielded watertight

NEMA ® enclosure that is CE marked

Specifications

Operating range Environmental range: 1 μR/hr to 1000 μR/hr (.01 μSv/hr to

10 μSv/hr)

Low range: 0.01 mR/hr to 1 R/hr (0.1 μSv/hr to 10 mSv/hr)

Medium range: 0.1 mR/hr to 10 R/hr (1 μSv/hr to 100 mSv/hr)

Radiation detected

High range: 1 mR/hr to 100 R/hr (10 μSv/hr to 1 Sv/hr)

Gamma rays

Typical energy dependence ± 15 % from 100 keV to 1.5 MeV

High voltage Regulated 500 V dc to 2500 V dc, < 1 mV ripple, digitally controlled with 1 V resolution, 500 microamperes at 1400 V dc

Input circuitry High and low discriminator setpoints. Jam detection (anti-jam).

Power requirements

Enclosure

12 V dc @ 500 mA

A plastic rectangular housing, NEMA 4 (IP65) type for outdoor or indoor applications

9.65 cm x 28.26 cm x 10.16 cm (3.8 in x 11.125 in x 4 in) Dimensions (LxWxH)

Mounting hole pattern

User interface

7.46 cm x 23.81 cm (2.9375 in x 9.375 in)

RS485 supporting multi-drop applications for communications with IBM

®

compatible personal computer running WIN1060 applications software

Temperature range

Operating limits

Relative humidity

Shock and vibration

Operating system

-20 °C to 50 °C (-4 °F to 122 °F)

-47 °C to 110 °C (-57 °F to 230 °F)

5 % to 95 %, non-condensing

Mechanical shock and vibration, 15 Grms, 10 to 2,000 Hz and SAE

J1455 cab mounted and mid-frame spectrum

Real-time, interrupt driven, embedded system

Elso Philips Service, Trenčín

Area Monitors

19

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

1060AM-V

Radiation Detection System

Specifications

1060DA Remote Display

Range

Operator interface

Radiation values display

Type

Line 1

Line 2

5 decades, based on 1060AM-V

Digital, microprocessor controlled, and intelligent display

PLED, 16 x 2 matrix, 2 lines, 16 characters per line

Displays radiation value and engineering units

Displays alarm messages (HIGH ALARM, OVERRANGE,

JAM, FAIL)

Luminance

Display size (WxH)

Character size

Color

Value

Units

High-alarm indicators

Visual

Audible

350 cd/m 2

6.5 cm x 1.6 cm (2.56 in x 0.63 in)

0.5 cm (0.20 in), typical (0.105 in x 0.197 in)

Yellow/Green

Up to five digits with floating decimal point

Auto ranging, mR/hr; SI

Alarm silence

Red, High intensity LED, 0.78 in Ø default setpoint is 2.0 mR/hr

103 ± 5 db(A) siren at 15 V dc, 60.96 cm (24 in), 25 °C, 2,900

± 250 Hz operating frequency, stainless steel diaphragm

Single pole, push to make, momentary push button switch silences audible alarm. Also initiates a self-test routine when held for three (3) seconds.

Enclosure

Enclosure Ingress protection IP41, NEMA1 up to NEMA 4 (IP65)

Enclosure dimensions (LxWxH) 19 cm x 5.7 cm x 3.2 cm (7.5 in x 2.24 in x 1.26 in), approx.

Overall dimensions (LxWxH)

Weight

Mounting

22.6 cm x 8.7 cm x 4.7 cm (8.9 in x 3.42 in x 1.88 in)

.39 kg (0.85 lb), approx.

Four adjustable foot mounting brackets provided

Included accessories

1060AM-V-LR Vehicle area monitor low range

1060AM-NM3-DC Power cable assembly, 25 ft

1060AM-NM3-IO I/O cable assembly, 25 ft

1060AM-NM3-PC RS485 to USB cable, 25 ft

941060WN WIN1060 application software

Ordering information

1060AM-V Vehicle-Mounted

Radiation Detection System

Elso Philips Service, Trenčín

20

Area Monitors

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

1060AM

Digital Smart Detector

Area Monitor

The versatile 1060AM is designed for reliable, continuous area monitoring for gamma or x-ray sources in a medical facilities, or any other facility with radioisotope sources. Employing internal

Geiger-Mueller (GM) detectors, it is available in environmental, low, medium, and high range versions, with an optional MHV interface that can accommodate a wide variety of external

GM probes.

The 1060AM is suitable for stand-alone operation or in a network environment employing multiple channels, communicating via an RS485 interface to a main computer system. The optional

WIN1060 PC software provides the ability to display multiple channels, to maintain both alarm and measurement history, and to access to system configuration options. An optional remote display, consisting of a visual alarm indicator and a logarithmic meter corresponding to the detector range, is available. In addition, the 1060AM provides an EMI shielded watertight National

Electrical Manufacturers’ Association (NEMA

®

) enclosure that is CE marked. The 1060AM provides two RS485 connectors to simplify connections between multiple units.

Specifications

Key features

Available in 4 operating ranges: environmental, low, medium and high

Optional MHV interface for external probes

Wide range of applications in

NEMA 4 Enclosure

RS485 interface for multi-drop applications

Optional WIN1060 Windows

® software monitors up to 30 channels

Optional remote display with alarm indicator

All versions available in SI units

Optional accessories

941060WN WIN1060

Applications Software

90-177 Converter RS-232/

RS-485 power cube, cable (US)

90-178 Converter RS-232/

RS-485 power cube, cable

(Europe)

90-179 Converter RS-232/

RS-485 power cube, cable

(Australia)

90-180 Converter RS-232/

RS-485 power cube, cable (UK)

External Probes (consult factory)

Custom configurations available

Operating range Environmental range: 1 μR/hr to 100 μR/hr (0.01 μSv/hr to

10 μSv/hr)

Low range: 0.01 mR/hr to 1 R/hr (0.1 μSv/hr to 10 mSv/hr)

Medium range: 0.1 mR/hr to 10 R/hr (1 μSv/hr to 100 mSv/hr)

Radiation detected

High range: 1 mR/hr to 100 R/hr (10 μSv/hr to 1 Sv/hr)

Gamma rays

Typical energy dependence ± 15 % from 100 keV to 1.5 MeV

High voltage Regulated 500 V dc to 2500 V dc, < 1 mV ripple, digitally controlled with 1 V resolution, 500 microamperes at 1400 V

Input circuitry

Power requirements

Enclosures

Dimensions (LxWxH)

Mounting hole pattern

User interface

High and low discriminator setpoints. Jam detection (anti-jam).

12 V dc @ 500 mA power converter

A plastic rectangular housing, NEMA 4 type for outdoor or indoor applications

9.65 cm x 28.26 cm x 10.16 cm (3.8 in x 11.125 inx 4.0 in)

7.46 cm x 23.81 cm (2.9375 in x 9.375 in)

RS-485 supporting multi-drop applications for communications with IBM ® compatible personal computer running WIN1060 applications software

Temperature range

Relative humidity

Shock and vibration

Operating system

0 °C to 50 °C (32 °F to 122 °F)

5 % to 95 %, non-condensing

Mechanical shock and vibration specifications are per ANSI

N42.17A, Section 8.4 and 8.5

Real-time, interrupt driven, embedded system

Ordering information

1060AM-NM-ER NEMA

Enclosure, environmental range

1060DS-ER (-SI) Remote Display

1060AM-NM-LR NEMA

Enclosure, low range

1060DS-LR (-SI) Remote Display

1060AM-NM-MR NEMA

Enclosure, medium range

1060DS-mR (-SI) Remote Display

1060AM-NM-HR NEMA

Enclosure, high range

1060DS-HR (-SI) Remote Display

1060MHV-NM External probe 1

1

Consult factory for external probe

options

Add -SI for SI units

Elso Philips Service, Trenčín

Area Monitors

21

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

05-437

PRIMALERT

®

35 Area

Radiation Monitor

The PRIMALERT 35 contains an energy-compensated GM detector and has six range indicators

(1, 2, 4, 8, 16, and 32 mR/hr) that can clearly display an increase or decrease in radiation levels.

The visible and audible alarms can be set at any of the six levels by a front-panel, screwdriveradjustable control. When each preset radiation level is exceeded, personnel are alerted by bright flashing red lights (visible over a 180° field) and a loud intermittent audio signal. The alarms stop automatically when the radiation level falls below each of the preset values. This permits instant radiation-level recognition not readily distinguishable on meter-type instruments.

Fail-safe operation is assured by a light which continuously indicates background radiation and provides visual proof that the unit is functioning. The monitor will not jam or show false readings in high radiation fields. A mounting bracket and a

110 V ac adapter/power converter are also included.

Key features

Provides continuous visual indication of radiation levels and produces audible and/ or visual alarms at any of six programmable radiation levels

Assures reliable, continuous monitoring wherever radioactive materials are present

Displays the radiation level in bright color-coded lights

Optional Primalarm Remote

Alarm, which functions up to

100 feet from monitor

Specifications

Power requirements

Dimensions (WxHxT)

Weight

105 V to 125 V/60 Hz/8 W

9 cm x 15 cm x 4 cm (3.5 in x 6 in x 1.5 in)

0.9 kg (2 lb)

Optional accessories

62-103 Check Source, 137

Cs,

10 μCi. Flat disc, 1 in diameter

Included accessories

AC adapters, specify with order

14-314 110 V ac 12 V dc

500 mA (USA, Japan)

14-400 230 V ac 12 V dc

500 mA (Europe)

14-417 230 V ac 12 V dc

580 mA (UK)

14-436 230 V ac 12 V dc

580 mA (Australia)

Ordering information

05-437 Primalert 35 Area

Radiation Monitor

Elso Philips Service, Trenčín

22

Area Monitors

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

05-443 and 05-444

PRIMALERT

®

Digital

Area Monitors

The PRIMALERT Digital Area Monitors are designed for a wide range of gamma radiation area monitoring applications. Two self-contained configurations are available, each with an internal energy compensated

GM detector (detection range in parentheses): 05-443 (0.1 mR/h to 1 R/h) and

05-444 (1 mR/h to 4 R/h). Both models are ac powered with an internal battery backup, have user settable low and high alarms, and are available with an optional remote alarm for added security.

The versatile PRIMALERT Digital Area Monitors can be used in industrial applications, medical settings, or wherever there is a need to warn personnel of increasing radiation levels and/or to limit the accumulated exposure of personnel to gamma radiation.

Key features

Simple installation and setup

(calibration controls easily accessed through front panel)

Anti-jam circuitry prevents erroneous readings at tube saturation

LED digital display with

Detector Fail indicator

Programmable low and high alarm indicators, with an optional remote alarm available

Data output/RS-232

Specifications

Indicated use Radiation area monitoring

Internal GM detector range 05-443: 0.1 mR/hr to 1 R/hr; 05-444: 1 mR/hr to 4 R/hr

Display

Display units

Linearity

Response

Status (green light)

Low alarm

High alarm

Detector fail

4 digit LED display with 2 cm (0.8 in) character height; Display range: 000.0 to 9999

Can be made to display in μR/hr, mR/hr, R/h, μSv/h, mSv/h, Sv/h, cpm, cps and others

Reading within ± 10 % of true value with detector connected

Typically 3 seconds from 10 % to 90 % of final reading

Indicates the instrument is functioning properly

Indicated by a yellow light and slow beep (1 per sec) audible tone (can be set at any point from 0.0 to 9999)

Indicated by a red light and fast beep (4 per sec) audible tone (can be set at any point from 0.0 to 9999)

Red light and audible tone; > 68 dB at 2 ft indicates detector overload, no count from detector, or instrument failure

Low battery (yellow)

Calibration controls

High voltage

Threshold

Indicates < 2 hours of battery power remaining

Accessible from front of instrument (protective cover provided)

Adjustable from 200 V to 2500 V

Adjustable from 2 mV to 100 mV

Dead time

Overload

Overrange

Data output

Adjustable to compensate for dead time of the detector and electronics (can be read on the display)

Senses detector saturation (indicated by display reading “-OL”)

Indicates the radiation field being measured has exceeded the counting range of the instrument

(indicated by display reading “----”)

9 pin connector providing 5 decade log output, RS-232 output, signal ground connection, FAIL and Alarm signals (current sink), and direct connection to battery and ground

Power requirements

Battery life

Battery charger

Housing material

Temperature range

Dimensions (WxDxH)

Weight

95 V ac to 135 V ac (178 V ac to 240 V ac available), 50 Hz to 60 Hz single phase (< 100 mA), 6 V sealed lead acid rechargeable battery (built-in)

Typically 48 hours in non-alarm condition, 12 hours in alarm condition

Battery is continuously trickle charged when instrument is connected to line power and turned on

Aluminum housing with white polyurethane enamel paint

-20 °C to 50 °C (-4 °F to 122 °F). May be certified for operation from -40 °C to 65 °C (-40 °F to 150 °F).

24.6 cm x 6.4 cm x 18.7 cm (9.7 in x 2.5 in x 7.4 in)

2.3 kg (6.5 lb)

Note: audible indicators can be configured as a single beep if desired.

Ordering information

05-443 PRIMALERT Digital Area

Monitor with internal energy compensated 0.1 mR/hr to

1 R/hr GM detector

05-443-2200 PRIMALERT

Digital Area Monitor with internal energy compensated

1 μSv to 10 mSv/hr GM detector,

220 V ac operation

05-444 PRIMALERT Digital Area

Monitor with internal energy

05-444-2200 PRIMALERT

Digital Area Monitor with internal energy compensated

10 μSv to 40 mSv/hr GM detector, 220 V ac operation

05-446 Remote Display

Elso Philips Service, Trenčín

Area Monitors

23

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

05-450

PRIMALERT

®

Digital

Doorway Monitor

The highly sensitive 05-450 is designed to detect low levels of gamma radiation that pass through an entryway. Common installations of the 05-450 include hospital entrances, emergency rooms, laundry rooms, nuclear medicine labs and procedure rooms, waste disposal chutes and any other area of the hospital where radiation contamination could be a concern.

The system consists of a digital monitor, two shielded NaI (Tl) scintillation detectors with NEMA enclosures, associated cabling and a 10 μCi

137

Cs check source. The system is AC powered with internal battery backup and user selectable alarm settings.

Key features

Dual detectors—highly sensitive lead shielded NaI

(Tl) scintillators

Configuration with NEMA enclosures

Fast response time with LED digital display

Audio and visual alarms

Battery backup

24

Specifications

Detectors

Connectors

Sensitivity

Two 3 in Ø x 1 in thick (7.6 cm x 2.5 cm) shielded NaI (Tl) scintillation detectors with up to 200 ft cables

(NEMA 4x enclosures included)

BNC (others available on request)

Detects an unshielded 40 μCi 137Cs source at 10 ft and unshielded 10 μCi 137Cs source at 5 ft from the detector

Check source

Display

Display units

Linearity

Response

Status

Low alarm

High alarm

0.875 in Ø 10 μCi 137 Cs check source

4 digit LED display with 2 cm (0.8 in) character height

Can be made to display in μR/hr, mR/hr, R/hr, μSv/h, mSv/h, Sv/h, μrem/hr, mrem/hr, rem/hr, cpm, cps and others

Reading within ± 10 % of true value with detector connected

Typically 3 seconds from 10 % to 90 % of final reading

(green light) Indicates the instrument is functioning properly

Indicated by a yellow light and slow beep (1 per sec) audible tone (can be set at any point from 0.0 to 9999)

Indicated by a red light and fast beep (4 per sec) audible tone (can be set at any point from 0.0 to 9999)

Note: Audible indicators can be configured as a single beep if desired.

Detector fail

Low battery

High voltage

Threshold

Dead time

Overload

Overrange

Data output

Indicates overload, no count from detector, or instrument failure (red light and audible tone; > 68 dB at 2 ft)

Yellow light indicates < 2 hours of battery power remaining

Adjustable from 200 V to 2500 V

Adjustable from 2 mV to 100 mV

Adjustable to compensate for dead time of the detector and electronics (can be read on the display)

Senses detector saturation (indicated by display reading “-OL”)

Radiation field being measured exceeds the counting range of the instrument (indicated by display reading “----”)

9 pin connector providing 5 decade log output, RS-232 output, signal ground connection, FAIL and Alarm signals (current sink), and direct connection to battery and ground

Power requirements 95 V ac to 135 V ac (178 V ac to 240 V ac available), 50 Hz to 60 Hz single phase (< 100 mA), 6 V sealed lead acid rechargeable battery (built-in)

Battery life Typically 48 hours in non-alarm condition, 12 hours in alarm condition

Battery charger Battery is continuously trickle charged when instrument is connected to line power and turned on

Battery dependence

< 3 % change in readings to battery endpoint

Temperature range -20 °C to 50 °C (-4 °F to 122 °F). May be certified for operation from -40 °C to 65 °C (-40 °F to 150 °F)

Dimensioins

(WxDxH)

Electronics: 24.6 cm x 6.4 cm x 18.7 cm (9.7 in x 2.5 in x 7.4 in)

Detectors: 43.2 cm x 21.6 cm x 33 cm (17 in x 8.5 in x 13 in)

Weight Electronics: 2.3 kg (6.5 lb)

Detectors: 14.5 kg (32 lb)

Ordering information

05-450 PRIMALERT Digital

Doorway Monitor, Sv/hr, 220 V power

05-450-2200 PRIMALERT Digital

Doorway Monitor, Sv/hr, 220 V power

Elso Philips Service, Trenčín

Area Monitors

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

RS-500 Series

The Lawrence Livermore

Realistic Phantom

The Lawrence Livermore

Realistic Phantom

The Lawrence Livermore Realistic Phantom was developed under the direction of the US

Department of Energy, primarily as a reference standard for the in-vivo counting of emissions from low-energy transuranic nuclides. The organs of primary interest—the lungs, liver and lymph nodes—may be either radioactive or inert. These organs are accommodated in a thorax generally similar to average adult males.

The Livermore Phantom contains a synthetic bone skeleton molded within a soft-tissueequivalent material. The organs are located in an internal cavity with a separate torso cover that closes the phantom. Anterior sections of ribs and the sternum are molded into this cover.

Soft-tissue-equivalent blocks are used to position the organs and fill significant air spaces, providing continuity of the soft-tissues throughout the phantom. The phantom is shipped assembled with inert organs. Any or all of these organs can be replaced with radioactive organs, which are shipped in separate packages.

Organs

All organs are made of soft-tissue-equivalent material. Lungs are available with the nuclides uniformly dispersed throughout the lung molding material. Optional organs (developed by Fluke Biomedical and not part of the original DOE specifications) are inert, but have holes in 2 cm x 2 cm grids. These can be filled with either inert plugs or with active capsules to establish any desired distribution within the lungs.

The liver has the same loading flexibility.

However, another option is available, designed for use with solutions having relatively short half-lives. This liver is a hollow shell with a fill/drain port.

If the liver is not required, it may be replaced with the Abdominal Contents, which are also available with a hole matrix or active.

Lymph nodes are available as hollow shells or as active packets. Two are 8 mm in diameter by 17 mm long. A third is 13 mm in diameter by 25 mm long. They are positioned in recessed molds in a lymph node block.

Key features

A worldwide reference standard for in-vivo counting

Fission-product version available for nuclear power plant applications

Meets the standards of IRCU

Report No. 44

Inert and radioactive organs available

Capsules

Liver with hole matrix

Lungs

Elso Philips Service, Trenčín

25

Radiation Protection

25

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

RS-500 Series

The Lawrence Livermore

Realistic Phantom

Chest overlay plates

Transuranic emissions from individuals with varying amounts of muscle and/ or adipose tissue may be so attenuated as to be undetectable. Chest overlay plates were developed to ensure the validity of in-vivo counting of such individuals.

There are three sets of plates, each in four graded thicknesses. One set is equivalent to 87 % adipose and 13 % muscle, another is equivalent to 50 % adipose and 50 % muscle, and the third is equivalent to 100 % muscle. Plates of each material and/or each thickness are available separately.

Targets

Three sets of concentric circles are drawn in permanent black ink over the torso cover and over each chest overlay plate. They range from 1 to 5 inch in diameter in 1 inch increments, with a 5.5 inch diameter outer circle. One set of circles is placed over each lung region, with the outer circle tangent to the inferior aspect of the clavicle and protruding over the sternum. The third set is placed over the liver.

A 2 cm

2

grid is projected in red ink over most of the chest surface by the same technique that is used for the circles. All targets are consistent among all phantoms.

Organs

The organs of the Lawrence

Livermore Realistic Phantom fit exactly into the Fission-

Product Phantom. However, the internal construction has been modified to accommodate additional organs: thyroid, kidneys, stomach, spleen, pancreas, and small and large intestines.

The Fission-Product Phantom system is shipped with inert organs. Any or all organs can be replaced by radioactive organs, which are shipped in separate packages.

The same choices of nuclide loadings are available for active organs as for the liver of the

Livermore phantom (uniformly dispersed in the molding materials, hollow shells or with hole grids).

Chest overlay plates

These are fitted specifically to the Fission-Product Phantom and are not interchangeable with the Livermore phantom.

The Fission-Product Phantom

The Fission-Product Phantom is partially modeled after the Lawrence Livermore Realistic

Phantom. It was developed to meet the needs of nuclear power plants and other facilities where more organs are needed than are provided by the Lawrence Livermore Realistic Phantom.

This phantom may be used on a couch, in a chair or standing. It includes the head and neck, the complete torso and stub legs articulated at the hips (which are also needed for the sitting position). The same access to the torso cavity is provided as in the Livermore phantom.

The materials of construction are multienergetic formulations. The skeleton has been extended to include the head and full torso.

The “Superhuman Skeleton,” which is multienergetic, is used. Materials closely meet the standard of the International Commission on

Radiation Units and Measurements (ICRU) Report No. 44 (Tissue Substitutes in

Radiation Dosimetry and Measurement, 1989.)

26

Radiation Protection

Elso Philips Service, Trenčín

26

www.elso.sk/Fluke-Biomedical riešenia na presné meranie

RS-500 Series

The Lawrence Livermore

Realistic Phantom

Ordering information

Livermore Realistic Phantom System

RS–500 Livermore Realistic Phantom System, includes torso and gridded torso cover, inert heart, lungs, lymph nodes, liver and liver envelope or abdominal contents. All blocks needed for organ location and soft–tissue continuity within torso. Documentation. Permanent shipping and storage case.

RS–501

RS–502

RS–504

Torso and gridded torso cover

Lungs – foam, inert (pair)

Heart – inert

RS–506

RS–507*

RS–510

Lymph Nodes – solid, inert (set of 3)

Liver and Liver Envelope – solid, inert

Abdominal contents – solid, inert

Organ options: To replace the above supplied inert organs, or if ordered separately.

Hole matrix organs for loading with active capsules

RS–502H Lungs – solid, inert, w/hole matrix and inert plugs

RS–508H

RS–510H

Liver and Liver Envelope – solid, inert, w/hole matrix and inert plugs

Abdominal contents – solid, inert, w/hole matrix and inert plugs

Hollow organs

RS–506S

RS–519S

Lymph Nodes – hollow, shell (set of 3)

Liver – hollow, shell

Active organs

RS–503A**

RS–506A**

Lungs – foam, active (pair)

Lymph Nodes – solid, active (set of 3)

RS–508A**

RS–511A**

Liver – solid, active

Abdominal Contents – solid, active

**Price for active organs refers to single nuclide with single isotopic concentration.

Capsules***

RS–520

RS–521

Capsule, active (advise isotope and concentration)

Capsule with cap – to be filled by user (price per capsule)

***Set–up and capsule loading price to be quoted.

Chest Overlay Plates

RS–513–1–A Chest Overlay Plate 1 mm to 6 mm plate thickness.

Matl A – 87 % fat and 13 % muscle.

RS–513–1–B Chest Overlay Plate 1 mm to 6 mm plate thickness.

Matl B – 50 % fat and 50 % muscle.

RS–513–1–C Chest Overlay Plate, 1 mm to 6 mm plate thickness.

Matl C – 100 % muscle.

RS–513–2–A Chest Overlay Plate, 2 mm to 11 mm plate thickness. Matl A – 87 % fat and 13 % muscle.

RS–513–2–B Chest Overlay Plate, 2 mm to 11 mm plate thickness. Matl B – 50 % fat and 50 % muscle.

RS–513–2–C Chest Overlay Plate, 2 mm to 11 mm plate thickness. Matl C – 100 % muscle.

RS–513–3–A

Chest Overlay Plate, 3 mm to 16 mm plate thickness. Matl A – 87 % fat and 13 % muscle.

RS–513–3–B Chest Overlay Plate, 3 mm to 16 mm plate thickness. Matl B – 50 % fat and 50 % muscle.

RS–513–3–C Chest Overlay Plate, 3 mm to 16 mm plate thickness. Matl C – 100 % muscle.

RS–513–4–A Chest Overlay Plate, 4 mm to 24 mm plate thickness. Matl A – 87 % fat and 13 % muscle.

RS–513–4–B Chest Overlay Plate, 4 mm to 24 mm plate thickness. Matl B – 50 % fat and 50 % muscle.

RS–513–4–C Chest Overlay Plate, 4 mm to 24 mm plate thickness. Matl C – 100 % muscle.

RS–514–A

RS–514–B

RS–514–C

Chest Overlay Plates, Matl A – 87 % fat & 13 % muscle.

One of each plate size 6, 11, 16 and 24 mm.

Chest Overlay Plates, Matl B – 50 % fat and 50 % muscle. One of each plate size 6, 11, 16 and 24 mm.

Chest Overlay Plates, Matl C – 100 % muscle.

One of each plate size 6, 11, 16 and 24 mm.

Radiation Protection

Fission–Product Phantom

RS–550 Fission–Product Phantom, includes head, neck, neck cover plate for thyroid access compete torso and gridded torso cover, inert heart, lungs, lymph nodes, thyroid, liver, kidneys, stomach, pancreas, spleen, small and large intestines. Stub legs.

RS–512

Head, neck, neck cover plate for thyroid access, gridded torso cover.

RS–524

RS–504

Stub legs (pair) with pivot bolt

Heart – inert

RS–549

RS–502

RS–506

RS–543S

RS–507

RS–530

Auxiliary cover under gridded cover

Lungs – foam, inert (pair)

Lymph Nodes – solid, inert (set of 3)

Thyroid – hollow, shell inert

Liver only – solid, inert

Kidneys – solid, inert

RS–533

RS–536

RS–539

Stomach – solid, inert

Pancreas – solid, inert

Spleen – solid, inert

RS–544

RS–546

Small intestine – solid, inert

Large intestine – solid, inert

Organ options: To replace the above supplied inert organs, or if ordered separately.

Hole matrix organs for loading with active capsules

RS–502H Lungs – solid, inert, w/hole matrix and inert plugs

RS–507H

RS–530H

RS–533H

Liver and Liver Envelope – solid, inert, w/hole matrix and inert plugs.

Kidneys – solid, inert, (pair) w/hole matrix and inert plugs.

Stomach – solid, inert, w/hole matrix and inert plugs

RS–536H

RS–539H

RS–544H

Pancreas – solid, inert, w/hole matrix and inert plugs

Spleen – solid, inert w/hole matrix and inert plugs

Small intestine – solid w/hole matrix and inert plugs

RS–546H Large intestine – solid w/hole matrix and inert plugs

Note: All organs with hole hole matrices are supplied with full complement of inert plugs with 20 % excess.

Hollow organs

RS–506S

RS–519S

RS–541S

RS–543S

RS–532S

Lymph Nodes – hollow, shell (set of 3)

Liver – hollow, shell

Spleen – hollow, inert

Thyroid – hollow, shell inert

Kidney – hollow, shell

Stomach – hollow, shell

Pancreas – hollow shell, inert

RS–535S

RS–538S

Active organs

RS–503A**

RS–506A**

RS–543A

RS–508A**

RS–531A

RS–534A

Lungs – foam, active (pair)

Lymph Nodes – solid, active (set of 3)

Thyroid – hollow, active

Liver – solid, active

Kidneys – solid, active

Stomach – solid, active

RS–537A

RS–540A

Pancreas – solid, active

Spleen – solid, active

Please specify isotope with quantity per organ, lymph node or capsule.

Delivery may be delayed on certain isotopes, some of which are available only at intervals during the year.

Note: all organs with hole matrices are supplied with full complement of inert plugs with 20% excess.

Note: Lungs are not available as hollow, shell or with hollow matrix.

** Price for active organs refers to single nuclide with single isotopic concentration.

Prices for multiple concentrations and mixed nuclides will be quoted upon request.

27

Publications

The following Fluke Biomedical catalogs are also available

Fluke Biomedical Diagnostic Imaging QA

The Diagnostic Imaging QA catalog is a comprehensive source book of solutions for the Imaging QA Technologist, Physicist, Biomedical/

Clinical Engineer, or Service Engineer. The catalog contains information about the test devices, phantoms, and accessories needed to manage diagnostic imaging QA and maintain regulatorycompliance.

Order the Diagnostic Imaging QA catalog online by visiting www.flukebiomedical.com/catalogs

Fluke Biomedical Radiation Oncology QA

The Radiation Oncology QA catalog provides a full range of QA solutions for the Radiation Oncology Physicist, Therapist, and

Dosimetrist. The catalog contains information about the linear accelerator QA instruments, radiation oncology chambers, phantoms, and accessories needed to manage radiation oncology QA and maintain a safe, regulatory-compliant facility.

Order the Radiation Oncology QA catalog online by visiting www.flukebiomedical.com/catalogs

Fluke Biomedical Biomedical Test

The Biomedical Test catalog emphasizes the complete line of biomedical test and simulation products for Biomedical/Clinical

Engineers and Technicians. The catalog contains information about

Fluke Biomedical’s test and simulation products, including standalone electrical safety testers, patient simulators, and performance analyzers, as well as fully integrated and automated performancetesting and documentation systems.

Order the Biomedical Test catalog online by visiting www.

flukebiomedical.com/catalogs

Your authorized Fluke Biomedical distributor

riešenia na presné meranie™

Elso Philips Service; tel: +421 32 6582410 email: [email protected]

; web: www.elso.sk

Fluke Biomedical.

The accuracy you need from a company you can trust.

Fluke Biomedical

6045 Cochran Road

Cleveland, OH 44139-3303 U.S.A.

Fluke Biomedical Europe AS

Vegamot 8, N-7048 Trondheim, Norway

For more information, contact us:

In the U.S.A. (800) 850-4608 or

Fax (440) 349-2307

In Europe/M-East/Africa +47 73954700 or

Fax +47 73954701

From other countries +1 (440) 248-9300 or

Fax +1 (440) 349-2307

Email: [email protected]

Web access: www.flukebiomedical.com

©2007 Fluke Biomedical. All OEM company trademarks are implied. Specifications subject to change without notice. Printed in U.S.A.

9/2007 3038334 C-EN Rev A

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals