PXI-9820 M

PXI-9820 M
PXI-9820
2-CH, 65MS/s, 14-Bit Digitizer with SDRAM
User's Guide
Recycle Paper
©Copyright 2004 ADLINK Technology Inc.
All Rights Reserved.
Manual Rev. 1.00 : September 21, 2004
Part No: 50-17015-100
The information in this document is subject to change without prior notice in
order to improve reliability, design, and function and does not represent a
commitment on the part of the manufacturer.
In no event will the manufacturer be liable for direct, indirect, special, incidental, or consequential damages arising out of the use or inability to use
the product or documentation, even if advised of the possibility of such
damages.
This document contains proprietary information protected by copyright. All
rights are reserved. No part of this manual may be reproduced by any
mechanical, electronic, or other means in any form without prior written
permission of the manufacturer.
Trademarks
PXI is registered trademarks of PXI Systems Alliance. Other product
names mentioned herein are used for identification purposes only and may
be trademarks and/or registered trademarks of their respective companies.
Getting Service from ADLINK
Customer Satisfaction is top priority for ADLINK TECHNOLOGY INC. If
you need any help or service, please contact us.
ADLINK TECHNOLOGY INC.
Web Site
http://www.adlinktech.com
Sales & Service
Service@adlinktech.com
TEL
Address
+886-2-82265877
FAX
+886-2-82265717
9F, No. 166, Jian Yi Road, Chungho City, Taipei, 235 Taiwan
Please email or FAX your detailed information for prompt, satisfactory,
and consistent service.
Detailed Company Information
Company/Organization
Contact Person
E-mail Address
Address
Country
TEL
FAX
Web Site
Questions
Product Model
Environment
Detail Description
Suggestions for ADLINK
OS:
Computer Brand:
M/B:
Chipset:
Video Card:
NIC:
Other:
CPU:
BIOS:
Table of Contents
Chapter 1 Introduction ................................................. 1
1.1 Features ....................................................................................2
1.2 Applications ...............................................................................2
1.3 Functional Block Diagram .........................................................3
1.4 Specifications ............................................................................6
Chapter 2 Installation ................................................. 11
2.1 Contents of Package ...............................................................11
2.2 Unpacking................................................................................12
2.3 Mechanical Drawing ................................................................13
2.4 Analog Input Impedance Setting .............................................14
2.5 Installation ...............................................................................15
2.6 Software Support.....................................................................17
Chapter 3 Operation Theory ...................................... 19
3.1 Block Diagram .........................................................................19
3.2 Analog Input Signal Source Control ........................................20
3.3 A/D Sampling rate and Timebase Control...............................20
3.4 Trigger Sources .......................................................................24
3.5 Trigger Modes .........................................................................27
3.6 Trigger Signals Exporting ........................................................31
3.7 Data Transfers.........................................................................32
3.8 AI Data Format ........................................................................33
3.9 Synchronizing Multiple Devices...............................................33
3.10 Auto-calibration .....................................................................38
Warranty Policy........................................................... 39
Table of Contents • i
How to Use This Guide
This user’s guide is designed to help users understand and operate the
PXI-9820. It is divided into three chapters:
Chapter 1
Introduction
Gives an overview of product features, applications, specifications and a functional block diagram.
Chapter 2
Installation
Describes how to install the PXI-9820.
Chapter 3
Operation Theory
Describes how to operate the PXI-9820, including the control and setting of signal, timebase, and trigger sources;
trigger modes; data transfers; synchronizing multiple modules; and auto-calibration.
How to Use This Guide • iii
1
Introduction
The ADLINK PXI-9820 is a 65MS/s, high-resolution PXI digitizer with deep
SODIMM SDRAM memory that features flexible input configurations, including programmable input ranges and user-selectable input impedance.
With deep onboard acquisition memory, the PXI-9820 is not limited by the
PCI’s 132MB/s bandwidth, and can record waveforms for extended periods
of time. The PXI-9820 is ideal for high-speed waveform capturing such as
radar, ultrasound, software radio, and other signal digitizing applications
needing large amounts of memory for data storage. PXI-9820 uses the PXI
trigger bus to send and receive trigger and timebase signals to and from
other devices. In additional, the PXI-9820 can phase-lock the reference
clock from the PXI backplane or from an external connector for better
synchronization capabilities.
This chapter gives an overview of the PXI-9820 high speed digitizer, and
covers the following topics:
Features
Applications
Functional Block Diagram
Specifications
Introduction • 1
1.1 Features
1.2
•
PXI specifications Rev. 2.1 compliant
•
3U Eurocard form factor, CompactPCI compliant (PICMG 2.0 R3.0)
•
14-bit A/D resolution
•
Up to 60MS/s sampling rate per channel with internal timebase
•
Up to 65MS/s sampling rate per channel with external timebase
•
Up to 130MS/s sampling rate using “ping pong” mode for single-channel acquisition
•
2-CH simultaneous-sampled single-ended analog inputs
•
Programmable input ranges of ±1V and ±5V
•
User-selectable input impedance of 50Ω or high input impedance
•
>30MHz -3dB bandwidth
•
Up to 512MB on-board SODIMM SDRAM
•
2-CH synchronous digital inputs
•
Scatter-gather DMA data transfers
•
Analog and digital triggering
•
Fully auto calibration
•
Multiple modules synchronization through PXI trigger bus
•
On-board phase-lock loop circuit
Applications
•
Software Radio / Wireless Communication
•
Radar / Sonar / Lidar
•
Ultrasound
•
Imaging
•
Military / Laboratory / Research
2 • Introduction
1.3 Functional Block Diagram
The following topics overview the PXI-9820 main features as shown in the
functional block diagram below.
Figure 1.1 PXI-9820 Functional Block Diagram
1.3.1 Analog Input
The PXI-9820 features two analog input channels. Signal bandwidth of
each channel exceeds 30MHz, satisfying the Nyquist sampling theory. The
input ranges are programmable as either ±5V or ±1V. The 14-bit A/D
resolution makes the PXI-9820 ideal both for time-domain and frequency-domain applications.
1.3.2 Acquisition System
The PXI-9820 uses a pair of 65MS/s, 14-bit pipeline ADCs to digitize input
signals. It provides an internal 60MHz timebase for data acquisition. The
maximum real-time sampling rate is 60MS/s with internal timebase and up
to 65MS/s with external timebase. By using the “ping pong” mode, the
sampling rate is up to 120MS/s with internal timebase or 130MS/s with
external timebase.
1.3.3 Acquisition Memory
The PXI-9820 supports SODIMM SDRAM ranging from 128MB to 512MB.
Digitized data is stored in the on-board SDRAM before being transferred to
host memory. The PXI-9820 uses scatter-gather bus mastering DMA to
move data to the host memory. If the data throughput from the PXI-9820 is
less than the available PCI bandwidth, the PXI-9820 will bypass the
Introduction • 3
SDRAM and use the on-board 3k-sample FIFO to achieve real-time
transfer directly to the host memory.
1.3.4 Triggering
The PXI-9820 features flexible triggering functions, such as analog and
digital triggering. The analog trigger features programmable trigger
thresholds on rising or falling edges of both input channels. The 5V/TTL
digital trigger comes from the external SMB connector, PXI trigger bus or
PXI_STAR for synchronizing multiple devices.
Post-trigger, pre-trigger, delay-trigger and middle-trigger modes are
available to acquire data around the trigger event. The PXI-9820 also
features repeated trigger acquisition to acquire data in multiple segments
coming with successive trigger events at extremely short rearming intervals.
Figure 1.2 PXI-9820 Trigger Modes
1.3.5 Multiple-Instrument Synchronization
To achieve multiple-instrument synchronization, PXI-9820 routes timing
and trigger signals between one or more PXI-9820 devices by the PXI
trigger bus. Users can assign timebase or trigger signals to a specific PXI
trigger bus[0:7] via software, providing better flexibility for multiple devices
synchronization. The PXI-9820 also adopts PXI_STAR signal for better
synchronization performance. Both trigger signals and timebase clock can
be transferred to 13 peripheral slots through the PXI_STAR signal on slot 2.
The reference clock is used in the PXI-9820 phase-lock loop (PLL) circuit
to synchronize the timebase clock to the reference. The PXI-9820 can
4 • Introduction
accept a reference clock the CLK IN on the front panel as well as from
PXI_CLK10. The reference clock should be a fixed and stable 10MHz
clock.
Figure 1.3 PXI-9820 Timebase Control
1.3.6 Synchronous Digital Input
The PXI-9820 features two high-speed digital input lines. User can clock in
digital data synchronous to the analog input timebase clock. Viewing
analog and digital correlated operations is effortless.
Figure 1.4 PXI-9820 Synchronous Digital Inputs
1.3.7 Auto Calibration
The auto-calibration function of the PXI-9820 is performed with trim DACs
to calibrate offset and gain errors of the analog input channels. Once the
Introduction • 5
calibration process is complete, the calibration constant will be stored in
EEPROM. These values are loaded and used as needed by the board.
Because all the calibration is conducted automatically by software commands, users do not have to adjust trimpots to calibrate the boards.
Figure 1.5 PXI-9820 Auto Calibration
1.4 Specifications
Analog Input
•
•
•
•
•
•
•
•
•
Number of channels: 2 simultaneous-sampled single-ended
Resolution: 14 bits
Max sampling rate:
60MS/s per channel with internal timebase
65MS/s per channel with external timebase
120MS/s using “ping pong” mode on CH0 with internal timebase
130MS/s using “ping pong” mode on CH0 with external timebase
On-board memory size:
SODIMM SDRAM: 128MB standard, up to 512MB
FIFO buffer: 3056 samples
Bandwidth (-3 dB): 30MHz minimum, 42MHz typical
Passband flatness (typical, referenced at 250 kHz):
±0.4dB, DC to 15 MHz
±1dB, 15 to 25 MHz
Input signal ranges:
±5V, ±1V (software programmable)
Input coupling: DC
Overvoltage protection:
6 • Introduction
Range
±5V
±1V
•
•
Overvoltage protection
±14V
±5V
Input impedance:
50Ω (default), 1.5MΩ (soldering selectable)
System Noise: (typical)
Range
Noise (LSBrms)
1.21
±5V
1.68
±1V
•
•
•
Crosstalk: < - 80dB, DC to 1MHz
Total Harmonic Distortion (THD)*: -75 dB
Signal-to-noise ratio (SNR)*:
Range
SNR (dB)
69
±5V
63.5
±1V
•
Spurious-free dynamic range (SFDR)*: 76.9dB
*Measured using 200kHz sine wave input with amplitude of 95% of full scale at
60MS/s
Synchronous Digital Input
•
•
•
•
•
Number of channels: 2
Compatibility: 5V/TTL
Digital logic levels
Input high voltage: 0V to 0.8V
Input low voltage: 2.0V to 5V
Maximum input overload: -0.5V to +7V
Hysteresis voltage: 0.7V minimum
Timebase System
•
Sources:
- Internal, on-board 60MHz:
24
Internal sample clock frequency: 60 MHz / n, 1≦ n ≦2 -1
Timebase accuracy: ±25 ppm
Jitter: < 6ps
- External, CLK IN sine wave:
Connector: front panel SMB connector (CLK IN)
Impedance: 50Ω
Coupling: AC
Input amplitude: 1Vpp to 2Vpp
Overvoltage protection: 2.5Vpp
Introduction • 7
Frequency range:
Ping-pong mode: 25MHz - 65MHz
Others: 500kHz - 65MHz
- External, PXI Trigger Bus[0:7]:
Source selection: from one of PXI Trigger bus[0:7], software selectable
Compatibility: 5V/TTL
Maximum input overload: -0.5V to +5.5V
Frequency range:
Ping-pong mode: 25MHz - 65MHz
Others: 500kHz - 65MHz
- External, PXI_STAR
Source selection: software selectable
Compatibility: 5V/TTL
Maximum input overload: -0.5V to +7.0V
Frequency range:
Ping-pong mode: 25MHz - 65MHz
Others: 500kHz - 65MHz
•
Timebase clock exporting
Destinations: PXI_STAR[0:12], PXI Trigger Bus[0:7]
PLL Reference Clock
•
Sources: PXI 10MHz reference (backplane, PXI_CLK10), external
10MHz reference (front panel SMB, CLK IN)
•
Frequency requirement: 10MHz ± 50ppm
- External 10MHz reference
Connector: front panel SMB connector (CLK IN)
Impedance: 50Ω
Coupling: AC
Input amplitude: 1Vpp to 2Vpp
Overvoltage protection: 2.5Vpp
- PXI 10MHz
Connector: from PXI backplane
Compatibility: 5V/TTL
Maximum input overload: -0.5V to +5.5V
Triggering
•
Sources: software, analog trigger, external digital trigger (TRG IO),
PXI_STAR, PXI Trigger Bus[0:7]
- Analog triggering
Sources: CH0 and CH1
8 • Introduction
Slope: rising/falling
Coupling: DC
Trigger sensitivity: 256 steps in full-scale voltage range
Hysteresis: 1.5% of the full range
Offset error: 1.25% of the full range
- Digital triggering
Connectors: front panel SMB connector (TRG IO)
Slope: rising/falling
Compatibility: 5V/TTL
Minimum pulse width: 10ns
- PXI_STAR
Slope: rising/falling
Compatibility: 5V/TTL
Minimum pulse width: 10ns
- PXI Trigger Bus[0:7]
Source: software selectable from one of PXI trigger bus[0:7]
Compatibility: 5V/TTL
Minimum pulse width: 10ns
•
•
•
•
•
Modes: pre-trigger, middle-trigger, post-trigger, delay-trigger
Repeated trigger rearming interval: 2 cycles of timebase
Pre-trigger depth: 128MB to 512MB, depending on memory size
Post-trigger depth: 128MB to 512MB, depending on memory size
Trigger Exporting
Destinations: Front panel SMB connector (TRG IO), PXI Star
Trigger[0:12], PXI Trigger Bus[0:7]
Calibration
•
•
Recommended warm-up time: 15 minutes
On-board calibration reference:
Level: 5.000V
Temperature coefficient: ±2ppm/°C
Long-term stability: 6ppm/1000Hr
General Specifications
•
•
•
Dimensions: (not including connectors)
Single 3U PXI module, 100mm by 160mm
I/O connector:
BNC x 2 for analog inputs(CH0 & CH1)
SMB x 4 for external timebase (CLK IN), external digital trigger
(TRG IO) and synchronous digital input 0 & 1(SDI0, SDI1)
PCI signaling environment:
Supports a 32-bit 3.3V or 5V PCI bus
Introduction • 9
•
•
•
Operating environment:
Ambient temperature: 0 to 50°C
Relative humidity: 5% to 95% non-condensing
Storage environment :
Ambient temperature: -20 to 80°C
Relative humidity: 5% to 95% non-condensing
Power requirement: (typical)
Power Rail
5V
12V
3.3V
10 • Introduction
Current (mA)
900
305
360 (with 128MB onboard SDRAM memory)
500 (with 512MB onboard SDRAM memory)
2
Installation
This chapter describes how to install the PXI-9820. Please read the contents
of the package and unpacking information carefully before you install the
module.
2.1
Contents of Package
Check the shipping carton for any visible damage. If the shipping carton and
contents are damaged, notify the dealer for a replacement. Retain the shipping carton and packing materials for inspection by the dealer. Remember to
obtain authorization before returning any products to ADLINK.
Check the following in the package. If there is any missing item, contact your
dealer:
•
PXI-9820 2CH, 65MS/s, 14-Bit Digitizer with SDRAM
•
This User’s Guide
•
Software Installation Guide
•
ADLINK All-In-One CD
Installation • 11
2.2
Unpacking
Your PXI-9820 module contains electro-static sensitive components that can
be easily be damaged by static electricity.
Therefore, the module should be handled on a grounded anti-static mat. The
operator should be wearing an anti-static wristband, grounded at the same
point as the anti-static mat.
After opening the shipping carton, extract the module and place it only on a
grounded anti-static surface with component side up.
Again, inspect the module for damage. Press down on all the socketed IC's to
make sure that they are properly seated. Do this only with the module place
on a firm flat surface.
Note: DO NOT APPLY POWER TO THE MODULE IF IT HAS BEEN
DAMAGED.
12 • Installation
2.3
Mechanical Drawing
CLK IN
TRG IO
SO-DIMM
SDI0
SDI1
CH0
CH1
Figure 2.1 PXI-9820 Mechanical Drawing
The ADLINK PXI-9820 module is packaged in a Eurocard form factor compliant with PXI Specifications measuring 160mm in length and 100mm in
height (not including connectors). The connector types and functions are
described as follows.
CLK IN:
The SMB connector is a 50Ω, AC-coupled external timebase
input or external 10MHz reference clock input.
TRG IO:
The SMB connector is for external digital trigger input or output.
SDI0:
The SMB connector is for synchronous digital input channel 0.
SDI1:
The SMB connector is for synchronous digital input channel 1.
CH0:
The BNC connector is for attaching the analog input signal to
measure on channel 0.
CH1:
The BNC connector is for attaching the analog input signal to
measure on channel1.
SO-DIMM:
The SO-DIMM connector is for plugging the 144-pin SDRAM
SODIMM.
Installation • 13
2.4 Analog Input Impedance Setting
CH0 and CH1 input impedance can be set to 50Ω or 1.5MΩ by soldering gap
switches J6 and J7 on the backside of the PXI-9820. The location of J6, J7
and the corresponded input impedance setting are shown in Fig. 2.2 and
Table 2.1. The default setting is 50Ω input impedance.
J6
Open
Close
(Default)
CH0 Input Impedance
High (1.5MΩ)
Low (50Ω)
J7
CH1 Input Impedance
Open
Close
(Default)
High (1.5MΩ)
Low (50Ω)
Table 2.1: Location of solder switches
Figure 2.2: Location of solder switches
NOTE:
If the high input impedance 1.5MΩ is selected, the output impedance of the
signal sources should be kept low to avoid of the offset voltage caused by the
input bias current, which is 2μA min. and 25μA max.
14 • Installation
2.5 Installation
Windows will find the new PXI-9820 module automatically. If this is the first
time the PXI-9820 is running on your Windows system, please refer to the
following step-by-step installation procedures.
2.5.1 Software Installation
For Windows 98/2000/XP users:
1.
Insert the ADLINK ALL-In-One CD. Under X:\Software\WD-DASK\W98nt2k\DISK1, you will find SETUP.EXE (X is
the drive letter of your CDROM).
2.
Execute SETUP.EXE and follow the installation shield to complete
WD-DASK installation.
3.
When you complete the software installation, turn off your system and
follow the hardware installation guide to insert PXI-9820 into your
system.
2.5.2 Hardware Installation/Removal Procedures
The PXI connectors are rigid and require careful handling when inserted and
removed. Improper handling of modules can easily damage the backplane.
To insert the ADLINK PXI-9820 module into a PXI chassis, please follow the
procedures below.
1.
Turn off your system
2.
Align the module’s edge with the card guide in the PXI chassis.
3.
Slide the module into the chassis, until resistance is felt from the PXI
connector.
4.
Push the ejector upwards and fully insert the module into the chassis
5.
Once inserted, a “click” can be heard from the ejector latch.
6.
Tighten the screw on the front panel.
7.
Power on
Installation • 15
To remove the module, reverse steps 2 through 6 above.
2.5.3 Installation check
When you complete both software and hardware installations, you can check
the list in device manager to confirm whether PXI-9820 is working properly. In
the device manager, look for a device named “PCI9820 Device” under NuDAQ Boards. If it is found, installation was successful.
16 • Installation
2.6 Software Support
ADLINK provides versatile software drivers and packages for users’ differing
approaches to building up a system. ADLINK not only provides programming
libraries such as DLLs for most Windows based systems, but also drivers for
®
other software packages such as LabVIEW .
All software options are included in the ADLINK All-In-One CD. Non-free
software drivers are protected with licensing codes. Without the software
code, you can install and run the evaluation version for two hours for
trial/demonstration purposes. Please contact ADLINK dealers to purchase
the formal license.
2.6.1
Programming Library
For customers who are writing their own programs, we provide function libraries for many different operating systems, including:
WD-DASK: Includes device drivers and DLLs for Windows 98, Windows
NT, and Windows 2000. DLL is a binary compatible across Windows 98,
Windows NT, and Windows 2000. All applications developed with
WD-DASK are compatible across Windows 98, Windows NT, and Windows 2000. The developing environment can be VB, VC++, Delphi, BC5,
or any Windows programming language that allows calls to a DLL. The
function reference manual of WD-DASK are in the All-In-One CD
(\Manual_PDF\Software\WD-DASK).
WD-DASK/X: Includes device drivers and shared libraries for Linux. The
developing environment can be Gnu C/C++ or any programming language that allows linking to a shared library. The function reference
manual
of
WD-DASK/X
are
in
the
All-In-One
CD
(\Manual_PDF\Software\WD-DASK).
2.6.2
DAQ-LVIEW PnP: LabVIEW® Driver
DAQ-LVIEW PnP contains the VIs, which are used to interface with National
®
Instrument’s LabVIEW software package. DAQ-LVIEW PnP supports
®
Windows 98/NT/2000. The LabVIEW driver is shipped free with the device.
Users can install and use them without a license. For detailed information
about DAQ-LVIEW PnP, please refer to the user’s guide and function reference
manual
in
the
All-In-One
CD
(\Manual_PDF\Software\DAQ-LVIEW PnP).
Installation • 17
2.6.3
WD-OCX: ActiveX Controls
We suggest customers who are familiar with ActiveX controls and VB/VC++
programming use WD-OCX ActiveX control component libraries for developing applications. WD-OCX is designed for Windows 98/NT/2000. For more
detailed information about WD-OCX, please refer to the function reference
manual in the All-In-One CD (\Manual_PDF\Software\WD-OCX)
In addition, ADLINK supplies ActiveX control software DAQBench.
DAQBench is a collection of ActiveX controls for measurement or automation
applications. With DAQBench, users can easily develop custom interfaces to
display data, analyze acquired data or data received from other sources, or
integrate with popular applications or other data sources. For more detailed
information about DAQBench, please refer to the All-In-One CD
(\Manual_PDF\Software\DAQBench)
Users can obtain a 4-hour evaluation version of DAQBench for free from the
All-In-One CD. Please contact ADLINK or an ADLINK dealer to purchase the
software license.
18 • Installation
3
Operation Theory
The operation theory of the PXI-9820 is described in this chapter, including
the control and setting of signal sources, timebase sources, trigger sources,
trigger modes, data transfers, synchronizing multiple modules, and
auto-calibration.
3.1
Block Diagram
Figure 3.1 PXI-9820 Block Diagram
Operation Theory • 19
3.2 Analog Input Signal Source Control
Number of Channels
The PXI-9820 provides two simultaneously sampled analog input channels in
SE (single ended) connection. Each channel can be enabled individually.
Signal Range and Input impedance
The available signal input ranges are ±5V or ±1V, which can be set by
software. All signals are DC-coupled. The input impedance for high-speed
applications should also be considered. The selectable input impedance
values are 50Ω and 1.5MΩ. Please refer to section 2.4 for setting impedance.
3.3
A/D Sampling rate and Timebase Control
The PXI-9820 supports six timebase sources for analog input conversion:
•
Internal 60MHz (Internal VCXO free run mode)
•
Internal 60MHz (timebase clock locked to PXI_CLK10)
•
Internal 60MHz (timebase clock locked to external reference 10MHz)
•
External sine wave
•
PXI Trigger Bus[0:7]
•
PXI_STAR
The following diagram shows the timebase clock architecture of the
PXI-9820.
Figure 3.2 PXI-9820 Timebase Clock Architecture
20 • Operation Theory
Once choosing the timebase source, users can set a 24-bit counter to divide
the timebase to get the needed sampling rate. The following formula determines the ADC sampling frequency:
Sampling Rate = Timebase Frequency / ADC Clock Divisor
where the ADC Clock Divisor = 1,2,3,4,5… 224-1(maximum)
3.3.1
Internal Timebase Clock Source
The PXI-9820 has an onboard Voltage Control Crystal Oscillator (VCXO)
running at 60MHz. When selects internal timebase clock, you can have
free-run mode (none phase-locked) or PLL mode. In free-run mode, the
60MHz timebase clock is coming from the VCXO which is controlled by an
internal trim DAC. In PLL mode, the onboard 60MHz timebase phase locked
to the reference 10MHz clock. The reference 10MHz clock comes from PXI
backplane (PXI_CLK10) or front panel SMB connector (CLK IN).
3.3.2
External Timebase Clock Source
When users need a specific timebase in some applications that the onboard
60MHz timebase is not achievable, a timebase clock from an external device
can replace onboard timebase clock. In addition, external timebases also
provides a method to synchronize PXI-9820 to other measurement modules
by distributing/receiving a common clock to/from multiple modules. The
PXI-9820 can receive an external timebase clock from the front panel connector (CLK IN), PXI Star trigger or one of the PXI Trigger bus line.
Users can supply the timebase from external SMB connector CLK IN, which
should be a sine wave signal. This signal is AC coupled with 50Ω input impedance and the valid input level is from 1 to 2 volts peak-to -peak. Note that
the external clock must be continuous for correct ADC operation because of
the pipeline architecture of the ADC.
PXI-9820 can receive the timebase clock via one of the trigger bus lines by
software selection. The eight PXI trigger bus lines (PXI_TRIG[0:7]) provide
intermodule synchronization and communication. Note that this function is
only available when the PXI-9820 is in a PXI system. It’s not supported when
PXI-9820 is in a CompactPCI system.
When PXI-9820 is plugged into a generic peripheral slot, it can receive
timebase clock from PXI_STAR. The PXI_STAR signal comes from star
trigger controller are matched in propagation delay within 1ns and the delay
from star trigger slot to a peripheral slot is less than 5ns. According to these
hardware features, PXI-9820 can achieve very good synchronization performance when using PXI_STAR as timebase clock source. Note that this
Operation Theory • 21
function is only available when the PXI-9820 is in a PXI system. It’s not
supported when PXI-9820 is in a CompactPCI system.
3.3.3
PLL Reference Clock Source
The PXI-9820 equipped with a phase-locked loop circuit to synchronize the
timebase clock to reference clock. The PXI-9820 can receive a reference
clock from the front panel connector (CLK IN) as well as from PXI_CLK10.
The reference clock from front panel connector should be a stable 10MHz
clock. The input of CLK IN is AC coupled with 50Ω input impedance and the
valid input level is from 1 to 2 volts peak-to -peak. Note that the reference
clock must be continuous for correct PLL operation. Once the reference clock
disconnected, the timebase clock will still exist but not phase lock to reference clock anymore. When using external timebase clock, it’s not valid to
lock to reference clock.
3.3.4
Timebase Clock Exporting
The PXI-9820 can export timebase clock to the following two destinations:
PXI_STAR or one of PXI Trigger bus lines. When PXI-9820 plugged into star
trigger controller slot of the PXI backplane, it can export timebase clock to the
other 13 generic peripheral slots. The PXI-9820 can also export timebase
clock onto one of the eight PXI trigger bus lines. By software programming,
user can pick up a trigger line to transmit timebase clock. These two features
are very useful when synchronize to multiple measurement modules.
3.3.5
Timebase 50% duty cycle restore function
Because the on-board sample-and-hold circuit is in hold mode when the
timebase is high and in track mode when timebase is low, the PXI-9820 can
get the best dynamic response performance for high-frequency analog inputs
when the timebase has a 50% duty cycle. However, it is often difficult to
maintain a 50% duty cycle; especially when driving the timebase with a single-ended or sinewave input. To ease the constraint of providing an accurate
50% timebase, the PXI-9820 has an optional duty cycle restore function to
restore the timebase duty cycle to 50%, independent of the clock input duty
cycle. Low jitter on the rising edge (sampling edge) of the timebase is preserved while the falling edge is interpolated.
It may be desirable to disable the duty cycle restore function when users use
the external timebase input whose frequency is varied. Once the timebase
frequency is changed, over 100 clock cycles may be required for the duty
cycle restore circuit to settle to the new speed. Duty cycle restore function is
software programmable when configuring your PXI-9820.
22 • Operation Theory
3.3.6
130MS/s Sampling using Ping-Pong Mode
The PXI-9820 uses two A/D converters, each running at 60MS/s, to provide a
dual-channel simultaneous real-time sampling rate of 60MS/s. (65MS/s with
external timebase)
For the single-channel acquisition, the two ADCs can be clocked in a
“ping-pong” mode to achieve up to 120MS/s sampling (130MS/s with external
timebase). Figure 3.3 illustrates the operation. Note that only CH0 can be
applied to ping-pong mode. The onboard auto-calibration circuitry allows the
two channels to be matched in order to reduce the image signal.
Amplifier
circuitry
CH0
Input
clk
Amplifier
circuitry
CH1
Input
Timebase with 180 degree
phase difference
ADC
ADC
clk
Figure 3.3 “Ping-pong” mode operations
3.3.7 Synchronous Digital Input
The PXI-9820 contains two synchronous digital input channels, SDI0 and
SDI1. These two digital input lines can be sampled synchronously with
timebase clock for mixed signal applications. Thus the data transfer can up to
60Mbit/s when using internal 60MHz timebase clock. These two digital input
lines are combined with analog input channels, respectively. Please refer to
Figure 3.4 for the data format of ADC and SDI. When the analog input
channel of PXI-9820 is configured as “ping-pong” mode, only SDI0 is enabled.
Operation Theory • 23
Figure 3.4 Synchronous digital input operations
3.4
Trigger Sources
In addition to the internal software trigger, the PXI-9820 also supports external analog triggers, external digital triggers, PXI_STAR triggers and PXI
Trigger Bus[0:7]. Users can configure the trigger source by software. Refer to
Figure 3.5 for PXI-9820 trigger architecture.
Figure 3.5 PXI-9820 Trigger Architecture
3.4.1
Software Trigger
This trigger mode does not need any external trigger source. The trigger
asserts right after executing specified function calls to begin the operation.
24 • Operation Theory
3.4.2
External Analog Trigger
Users can choose either CH0 or CH1 as the trigger signal while using external analog trigger source. The trigger level can be set by software with
8-bit resolution. Please refer to Table 3.1 for the ideal transfer characteristic.
Trigger Level digital setting
Trigger voltage
Trigger voltage
(±5V range)
(±1V range)
0xFF
4.96V
0.992V
0xFE
4.92V
0.984V
---
---
---
0x81
0.04V
0.008V
0x80
0
0
0x7F
-0.04V
-0.008V
---
---
---
0x01
-4.96V
-0.992V
Table 3.1 Analog trigger ideal transfer characteristic
The trigger conditions for analog triggers are illustrated in Figure 3.6 and
described as follows:
• Positive-slope trigger - The trigger event occurs when the trigger
signal (analog input signal) changes from a voltage that is lower than the
specified trigger level to a voltage that is higher than the specified trigger
level.
• Negative-slope trigger - The trigger event occurs when the trigger
signal (analog input signal) changes from a voltage that is higher then the
specified trigger level to a voltage that is lower than the specified trigger level.
Trigger
level
Positive-slope trigger event
occurs
Negative-slope trigger event
occurs
Figure 3.6 Analog trigger conditions
Operation Theory • 25
3.4.3
External Digital Trigger
An external digital trigger occurs when a TTL rising edge or a falling edge is
detected at the SMB connector TRG IO on the front panel, as illustrated in
Figure 3.7. The trigger polarity can be selected by software. Note that the
signal level of the external digital trigger signal should be TTL-compatible,
and the minimum pulse width is 10ns.
Figure 3.7 External digital trigger input
3.4.4 PXI_STAR Trigger
When users select PXI_STAR as trigger source, the PXI-9820 can accept a
TTL-compatible digital signal as a trigger signal. The trigger occurs when a
rising edge or falling edge is detected at PXI_STAR. User can use software
to configure the trigger polarity. The minimum pulse width requirement of this
digital trigger signal is 10ns.
3.4.5 Triggers from PXI Trigger Bus[0:7]
The PXI-9820 utilizes PXI Trigger Bus[0:7] as System Synchronization Interface (SSI). Using the interconnected bus provided by PXI Trigger Bus,
user can easily synchronize multiple modules. When configured as input, the
PXI-9820 can accept three different SSI signals, SSI_TRIG1, SSI_TRIG2
and SSI_START_OP (for more detail about these signals, please refer to
section 3.9). Each signal can be routed from one of the PXI Trigger Bus[0:7]
by software programming.
26 • Operation Theory
3.5
Trigger Modes
The PXI-9820 provides 5 trigger sources (internal software trigger, external
analog trigger, external digital trigger, PXI_Star trigger, and PXI Trigger
Bus[0:7] signals). Users must select one of them as the source of the trigger
event. A trigger event occurs when the specified condition is detected on the
selected trigger source (For example, a rising edge on the external digital
trigger input).
There are 4 trigger modes (pre-trigger, post-trigger, middle-trigger, and delay-trigger) working with the 5 trigger sources to initiate different data acquisition timing when a trigger event occurs. They are described as follows.
3.5.1
Post-trigger Acquisition
Use post-trigger acquisition when you want to collect data after the trigger
event, as illustrated in Figure 3.8.
Operation
start
Acquisition
stop
Trigger
event
Acquired & stored
data with the specified amount
Time
Figure 3.8 Post trigger
3.5.2
Pre-trigger Acquisition
Use pre-trigger acquisition to collect data before the trigger event. The
quisition starts once specified function calls are executed to begin
pre-trigger operation, and it stops when the trigger event occurs.
If the trigger event occurs after the specified amount of data has been
quired, the system only stores the data before the trigger event with
specified amount, as illustrated in Figure 3.9.
acthe
acthe
Operation Theory • 27
Operation
start
Trigger
event
Acquired data
Acquired & stored
data with the specified amount
Time
Figure 3.9 Pre trigger (the trigger event occurs after the specified amount of
data has been acquired)
However, if the trigger event occurs before the specified amount of data has
been acquired, the system can either stop the acquisition immediately (which
implies the stored data will be less then the amount you specified) or ignore
the trigger signal until the specified amount of data has been acquired (which
assures the user can get the specified amount of data). These can be set by
software and are illustrated in Figure 3.10 and Figure 3.11.
Operation
Trigger
start
event
Acquired
&
stored data
Specified
amount of data
Time
Figure 3.10 Pre trigger (The trigger signal is accepted anytime after the operation starts)
28 • Operation Theory
Operation
start
Trigger signals that occur
before the specified amount of
data has been acquired
(shadow area) will be ignored.
Trigger
event
Time
Specified
amount of data
Acquired & stored
data with the specified amount
Figure 3.11 Pre trigger (The trigger signal will be ignored until the specified
amount of data is acquired)
3.5.3
Middle-trigger Acquisition
Use middle-trigger acquisition when you want to collect data before and after
the trigger event. The amount of stored data before and after the trigger can
be set individually (M and N), as illustrated in Figure 3.12.
Operation
start
Trigger
event
Acquired & stored
data with the specified amount (M)
Acquired & stored
data with the specified amount (N)
Time
Figure 3.12 Middle trigger
Operation Theory • 29
Like pre-trigger mode, the stored data may be less than the amount specified
if the trigger event occurs before the specified amount of data (M) has been
acquired. Users can also set by program to ignore trigger signals until the
specified amount of data (M) has been acquired.
3.5.4
Delay-trigger Acquisition
Use delay trigger acquisition to delay the data collection after the trigger
event, as illustrated in Figure 3.13. The delay time is specified by a 32-bit
counter value so that the maximum delay time is the period of timebase *
32
(2 –1), while the minimum delay time is the period of timebase.
Operation
start
Trigger
event
Acquisition
start
Delay
Acquisition
stop
Acquired & stored
data with the specified amount
Time
Figure 3.13 Delay trigger
3.5.5
Post-trigger or Delay-trigger Acquisition with Re-trigger
Use post-trigger or delay-trigger acquisition with re-trigger function to collect
data after several trigger events, as illustrated in Figure 3.14. Users can
program the number of triggers then the PXI-9820 will acquire an additional
record each time a trigger is accepted until all the requested records have
been stored in memory. After the initial setup, the process does not require
software intervention.
30 • Operation Theory
Operation
start
Trigger
event
Trigger
event
Acquired
Data
Acquired
Data
Figure 3.14 Post-trigger with re-trigger
Time
3.6 Trigger Signals Exporting
The PXI-9820 can export trigger signals to the following connectors: TRG IO,
PXI_STAR or PXI Trigger Bus[0:7].
The TRG IO on the front panel can also be programmed to output the trigger
signal when the trigger source is from software trigger, analog trigger,
PXI_STAR trigger, or PXI Trigger Bus[0:7]. The timing characteristic is in
Figure 3.15.
Tw
TRG IO
Tw = 2 - 3 TIMEBASE clocks
Figure 3.15 TRG IO output signal timing
When the PXI-9820 is plugged into the star trigger controller slot, i.e. slot 2, in
a PXI system, it can export trigger signals to other 13 generic peripheral slots
(PXI_STAR0 ~ PXI_STAR12) through the PXI star trigger pin. The trigger
signal source can be from software trigger, external digital trigger signal from
TRG IO, analog trigger and PXI Trigger Bus[0:7]. The output trigger signal
timing characteristic is identical to the TRG IO which is shown in figure 3.15.
The PXI-9820 utilize PXI Trigger Bus[0:7] as System Synchronize Interface.
When configured as output, the PXI-9820 can output 3 different trigger signals, SSI_TRIG1, SSI_TRIG2 and SSI_START_OP. User can route these
signals to any one of PXI Trigger Bus[0:7] by software programming.
Operation Theory • 31
3.7
Data Transfers
Since the maximum data throughput on the PXI-9820 (60MS/s * 2 channels *
2 Bytes/channel = 240MB/s) is much higher than the 32bit/33MHz PCI-bus
bandwidth, samples are acquired into the onboard SDRAM memory before
being transferred to the host computer. Since the number of stored samples
per acquisition is limited by the amount of on-board memory, the PXI-9820
supports different sizes of SODIMM SDRAM ranging from 128MB to 512MB
in order to meet application requirements.
Once all the data has been stored in the on-board memory, the data will be
transferred to the host computer’s memory through bus-mastering DMA.
In a multi-user or multi-tasking OS, like Microsoft Windows, Linux, and so on,
it is difficult to allocate a large continuous memory block to do the DMA
transfer. Therefore, the PXI-9820 provides the function of scatter /gather
DMA to link the non-continuous memory blocks into a linked list so that users
can transfer very large amounts of data without being limited by the fragment
of small size memory, as illustrated in Figure 3.16.
Figure 3.16 Scatter/gather DMA for data transfer
If the data throughput from the PXI-9820 is less than the available PCI
bandwidth (For example: 20MS/s * 2 channels * 2 Bytes/channel = 80MB/s),
the PXI-9820 also features on-board 3k-sample FIFO to achieve real-time
transfer bypassing the SDRAM, directly to host memory.
32 • Operation Theory
3.8
AI Data Format
Table 3.2 illustrates the ideal transfer characteristics of various input ranges
of the PXI-9820. Bit13-0 is the acquired 14-bit A/D data with binary coding
format while bit14 is the out-of-range (OTR) indicator (logic “1” means
out-of-range). As described in section 3.3.7, bit 15 is the synchronous digital
input data. SDI0 is within the data format of analog channel 0 while SDI1 is
within the data format of analog channel 1.
Description
Analog Input Voltage
Full-scale Range
±5V
±1V
Least significant bit
0.61mV
0.122mV
> = FSR
Digital code
>= 5V
>= 1V
7FFF
FSR-1LSB
4.99939V
0.999878V
3FFF
Midscale +1LSB
0.61mV
0.122mV
2001
Midscale
0V
0V
2000
Midscale –1LSB
-0.61mV
-0.122mV
1FFF
-FSR
-5V
-1V
0000
< -FSR
< -5V
< -1V
4000
Table 3.2 Analog input voltage and the output digital code (Note that bit14 is
the out-of-range indicator)
3.9
Synchronizing Multiple Devices
The eight interconnected lines on PXI backplane named as PXI Trigger
Bus[0:7] provide a flexible interface for multiple modules synchronization.
The PXI-9820 utilizes the PXI Trigger Bus[0:7] as the System Synchronization Interface (SSI). By providing flexible routing of timebase clock and trigger
signals onto PXI Trigger Bus, the PXI-9820 makes the synchronization between multiple modules easy and simple.
The bi-directional SSI I/Os provide a flexible connection between modules,
which allows one SSI master PXI-9820 to output the SSI signals to other
slaves PXI-9820s to receive the signals. Table 3.3 lists the summary of SSI
timing signals and the functionalities. Figure 3.17shows the architecture of
SSI. Note that it’s not allowed to route different signals onto the same trigger
bus line.
Operation Theory • 33
SSI timing signal Functionality
SSI_TIMEBASE
Input/Output 60MHz timebase signal through SSI
SSI_TRIG1
Input/Output the trigger signal through SSI
SSI_TRIG2
Input/Output the clocked trigger signal through SSI
SSI_START_OP
Input/Output the acquisition start signal in pre-trigger or middle-trigger mode
Table 3.3 Summary of SSI timing signals and the corresponding function
Figure 3.17 SSI Architecture
3.9.1
SSI_TIMEBASE
As an output, the SSI_TIMEBASE signal outputs the onboard 60MHz LVTTL
timebase through PXI Trigger Bus.
As an input, the PXI-9820 accepts the SSI_TIMEBASE signal to be the
source of timebase.
34 • Operation Theory
3.9.2
SSI_TRIG1
As an output, the SSI_TRIG1 signal reflects the trigger event signal in an
acquisition sequence. Users can use the function SSI_SourceConn() to
output the SSI_TRIG1 signal.
As an input, the PXI-9820 accepts the SSI_TRIG1 signal to be the trigger
event source. The signal is configured in the rising edge-detection mode.
When selecting the trigger sources of the PXI-9820, Users can select
TRSRC_SSI_1 to set SSI_TRIG1 as the source of trigger event.
Figure 3.18 and Figure 3.19 show the input and output timing requirements.
Tw
SSI_TRIG1
Tw = 2 - 3 TIMEBASE clocks
Figure 3.18 SSI_TRIG1 output signal timing
Tw
SSI_TRIG1
T w = 20 ns minimum
Figure 3.19 SSI_TRIG1 input signal timing
3.9.3
SSI_TRIG2 and SSI_START_OP
As an output, the SSI_TRIG2 signal is a clocked SSI_TRIG1 signal by
TIMEBASE, as illustrated in Figure 3.20.
Operation Theory • 35
SSI_TRIG1
TIMEBASE
Tw
SSI_TRIG2
T w = 2 TIMEBASE clocks
Figure 3.20 SSI_TRIG2 output signal timing
As an input, the PXI-9820 accepts the SSI_TRIG2 signal to be the source of
a one-clock delayed trigger event. The controller on the PXI-9820 will then
compensate the one-clock delay if using SSI_TRIG2 as the source of trigger
event. The signal is configured in the rising edge-detection mode.
Tw
SSI_TRIG2
T w = 20 ns minimum
Figure 3.21 SSI_TRIG2 input signal timing
As an output, the SSI_START_OP signal reflects the operation start signal in
a pre-trigger or middle-trigger acquisition sequence. Please refer to Figure
3.9 - Figure 3.12 for the relationship between the operation start signal and
the acquisition sequence.
As an input, the PXI-9820 accepts the SSI_START_OP signal to be the operation start signal in a pre-trigger or middle-trigger acquisition sequence.
The signal is configured in the rising edge-detection mode. Figure 3.22 and
Figure 3.23 show the SSI_START_OP signal input and output timing requirements.
For enabling output operations, users can use the function SSI_SourceConn()
to output the SSI_TRIG2 and SSI_START_OP signals.
For the input operations, users can select TRSRC_SSI_2 to set SSI_TRIG2
and SSI_START_OP as the source of the trigger event and operation start
signal.
36 • Operation Theory
Tw
SSI_START_OP
T w = 2 TIMEBASE clocks
Figure 3.22 SSI_START_OP output signal timing
Tw
SSI_START_OP
T w = 20 ns minimum
Figure 3.23 SSI_START_OP input signal timing
3.9.4
Comparing the different trigger sources from SSI
When selecting TRSRC_SSI_1 as the trigger source input, the signal
SSI_TRIG1 reflects the trigger event signal in an acquisition sequence.
However, when synchronizing multiple PXI-9820 devices, each PXI-9820
may recognize the trigger signal with one-clock time difference because the
signal is not related to the timebase.
There is another phenomenon if using TRSRC_SSI_2 in pre-trigger and
middle-trigger mode. The operation start signal is generated by a software
command so multiple PXI-9820 devices don’t start the data acquisition simultaneously, which may result in the fact that the amount of stored samples
are different if the trigger event occurs before the specified amount of data
has been acquired.
When selecting TRSRC_SSI_2 as the trigger source input, SSI_TRIG2 and
SSI_START_OP can achieve better synchronization between multiple
PXI-9820 devices. A clocked SSI_TRIG2 can guarantee all PXI-9820 devices recognize the trigger event at the same clock edge if they use the same
timebase. In pre-trigger and middle-trigger mode, SSI_START_OP guarantees all the PXI-9820 devices start the data acquisition at the same time.
Operation Theory • 37
3.10 Auto-calibration
By using the auto-calibration feature of the PXI-9820, the calibration software
can measure and correct offset and gain errors without any external signal
connections, reference voltages, or measurement devices.
After the auto-calibration procedure finishes, the calibration constants can be
saved into the EEPROM. In addition to the default bank of factory calibration
constants, there are three extra user-modifiable banks in the EERPOM for
users to store three sets of calibration constants according to different environments and re-load the calibration constants when necessary.
Because of the fact that errors in measurements will vary with time and
temperature, it is recommended that users re-calibrate the PXI-9820 when
the module is installed in a new environment.
Note: Before auto-calibration procedure starts, please warm up the module
for at least 15 minutes.
38 • Operation Theory
Warranty Policy
Thank you for choosing ADLINK. To understand your rights and enjoy all the
after-sales services we offer, please read the following carefully.
1.
Before using ADLINK’s products please read the user manual and follow
the instructions exactly. When sending in damaged products for repair,
please attach an RMA application form which can be downloaded from:
http://rma.adlinktech.com/policy/.
2.
All ADLINK products come with a limited two-year warranty, one year for
products bought in China.
• The warranty period starts on the day the product is shipped from
ADLINK’s factory.
• Peripherals and third-party products not manufactured by ADLINK will
be covered by the original manufacturers' warranty.
• For products containing storage devices (hard drives, flash cards, etc.),
please back up your data before sending them for repair. ADLINK is not
responsible for any loss of data.
• Please ensure the use of properly licensed software with our systems.
ADLINK does not condone the use of pirated software and will not
service systems using such software. ADLINK will not be held legally
responsible for products shipped with unlicensed software installed by
the user.
• For general repairs, please do not include peripheral accessories. If
peripherals need to be included, be certain to specify which items you
sent on the RMA Request & Confirmation Form. ADLINK is not responsible for items not listed on the RMA Request & Confirmation
Form.
3.
Our repair service is not covered by ADLINK's guarantee in the following
situations:
• Damage caused by not following instructions in the User's Manual.
• Damage caused by carelessness on the user's part during product
transportation.
• Damage caused by fire, earthquakes, floods, lightening, pollution, other
acts of God, and/or incorrect usage of voltage transformers.
• Damage caused by inappropriate storage environments such as with
high temperatures, high humidity, or volatile chemicals.
• Damage caused by leakage of battery fluid during or after change of
batteries by customer/user.
• Damage from improper repair by unauthorized ADLINK technicians.
Warranty • 39
•
•
•
4.
Products with altered and/or damaged serial numbers are not entitled to
our service.
This warranty is not transferable or extendible.
Other categories not protected under our warranty.
Customers are responsible for all fees necessary to transport damaged
products to ADLINK.
For further questions, please e-mail our FAE staff: service@adlinktech.com
40 • Operation Theory