isolated 64 ch 20 MS/s

isolated 64 ch 20 MS/s
MEMORY HiCORDER MR8847A
Max.
20 MS/s
high-speed sampling
All analog channels
isolated
Max. 16 channels
5 new modules
Logic channels
Max.
64 ch
16 channels standard
Support for a total of 13 units
NEW
ARBITRARY WAVEFORM GENERATOR UNIT
HIGH VOLTAGE UNIT
DIGITAL VOLTMETER UNIT
WAVEFORM GENERATOR UNIT
PULSE GENERATOR UNIT
For on-site work and R&D testing
Global Standard Recorder
High-voltage 1000 V direct input measurement
HIGH-VOLTAGE UNIT
Max. 1 MS/s high-speed sampling, 16-bit resolution measurement
Generate and record in a single unit
ARBITRARY WAVEFORM GENERATOR UNIT
Reproduce and output problematic waveform measurements
No amp needed; max. 15 V output
2
Testing
Vibration testing at the product development stage
A high-spec, high-quality
versatile measuring device
20 MS/sec sampling speed
Isolated input for all channels
Perform multi-channel, high-speed sampling at 20 M samples/
C onnections bet ween analog
sec (time axis resolution: 50 nsec) for all channels at the same
input channels, and between the
time.
input channel and the main unit,
are isolated by isolation elements.
So potential differences can be
measured without any concerns,
just like with an oscilloscope.
Isolation element
1 MS/s
20 MS/s
A4 size built-in printer
Print large, high-definition hard copies for easy on-site checking.
High-speed sampling allows you to measure the rising edge
Paper is easy to replace by inserting a new roll, rolling out the
of pulses and detect anomaly operations and instantaneous
paper slightly, and then closing the cover.
waveforms that occur suddenly with high precision.
Observe the rising edge of pulses
Input amp with integrated A/D converter
Simply open the cover, insert the new paper, and then close the cover.
3
Development
Surveys
Generating + monitoring
Power supply surveys for power equipment
5 new modules added
NEW
Hioki has added new high-performance
ARBITRARY WAVEFORM GENERATOR UNIT U8793 ►
modules in response to overwhelming
HIGH VOLTAGE UNIT U8974 ►
demand.
The Memory HiCorder now supports
a wide variety of measurements with a
total of 13 plug-in modules.
64 logic input channels
+10 analog channels
DIGITAL VOLTMETER UNIT MR8990 ►
WAVEFORM GENERATOR UNIT MR8790 ►
PULSE GENERATOR UNIT MR8791 ►
Large 512 MW capacity
Hioki has developed an internal storage
FPGA for super-high-speed access. Used
The MR8847A has 16 built-in logic input channels.
in combination with large capacity high-
Add 3 logic input units to record a total of 64 channels
speed memory, this enables many hours of
at once. You can also display the waveforms for
high-speed sampling to be recorded.
all channels on a single screen—ideal for timing
measurements.
Up to 10 channels of analog waveforms can be
recorded at the same time for efficiency.
(MR8847-53 only)
Input
Input
NEW
A/D
conversion
Isolated
A/D
conversion
Isolated
Memory
Write
SSD 128 GB storage media
The new internal SSD unit (available as an
additional option) has 128 GB of capacity,
allowing large amounts of data to be stored.
Durable design, with resistance to
dropping up to 50 cm
T h e M R 8 8 47A i s r e s i s t a n t t o s t r o n g
mechanical shock and vibration, such as
short drops.
The durable design has been tested to
withstand vertical drops of up to 50 cm.
Measure and display multiple relays at the same time
* Tested based on in-house conditions. A dropped unit is not guaranteed to be
free of damage or trouble.
Screen display
Print output
4
NEW
HIGH VOLTAGE UNIT U8974
Directly input high voltage
without a differential probe
2 channels, banana input terminal
Both channels support 1000 V input
1000 V DC, 700 V AC high-voltage direct input
S inc e yo u c a n dire ctly inpu t up to
1000 V DC and 700 V AC, a differential
probe is no longer necessary.
Maximum rated voltage to ground is
1000 V for CAT III and 600 V for CAT IV
environments.
Battery
Load
Inverter
Motor
No differential probe needed
Direct voltage measurement
Global power supply line
measurement
Ideal for primary and secondary measurements of
UPS power supplies and commercial power supply
transformers, and for recording the primary and
secondary waveforms of inverters. It can also be used
to measure high-voltage power supply lines, such as
380 V and 480 V systems used in many countries.
Supports high voltage systems around the world
Applicable to a variety of
characteristics tests
Maximum 1 MS/s high-speed sampling and 16-bit
resolution allow the MR8847A to be used for interruption
testing and switch testing.
The voltage of each battery cell can be input separately.
This uses 1000 V DC input, which can withstand even if
high voltage is applied when a cell shorts.
The digital voltmeter unit, which allows input up to 500 V
DC, is suitable for the testing of individual battery cells.
Battery
Battery package
Transformer Dump Tests
Application of each unit allows
analysis of the correlation
between voltage before and
af ter the interruption of a
generator, RPM fluctuation
rate, governor servo motor
operation conditions, and
suppression machine switch
timing.
Interchannel isolation allows for safe circuit
connections. Simultaneous high-speed sampling
can record waveforms before and after the dump.
Input large numbers of control and circuit signals.
Recommended units
HIGH VOLTAGE UNIT
U8974
CURRENT UNIT
8971
LOGIC UNIT
8973
● Maximum 1 MS/s high-speed sampling and 16-bit resolution in the high-voltage unit allow
the MR8847A to be used for interruption and switch testing.
5
NEW
ARBITRARY WAVEFORM
GENERATOR UNIT U8793
Generate and record
in a single unit
2 channels, SMB terminals
2 types of output cables (sold separately) ►
Output and record results seamlessly
Just one MEMORY HiCORDER gives
y o u a f u n c t i o n g e n e ra to r m o d e,
arbitrary waveform generator mode,
and waveform measurement mode.
This makes it easy to obser ve
waveforms while varying test
conditions, such as changing the
signal's amplitude and frequency and
programming various waveforms to
output in order.
Generate
Record
Standalone testing
Output recorded waveforms
without modification
For example, you could output actual waveforms recorded
from a car without modification, and then use them for
standalone testing. You can also generate isolated output
of up to 15 V without a generator or amplifier, which is
traditionally necessary in order to generate output while
varying the signal's amplitude and frequency.
Measurement
on actual car
Reproducibility testing
Output waveform example
Waveform Maker Software included
After you install the included SF8000 Waveform Maker
software on your computer, you can create waveforms
easily by either entering them directly or by entering the
functions behind them. You can also quickly add noise
and multiply waveforms.
Standalone
testing
Waveform Maker SF8000
for processing
as needed
Process actual waveforms for
reproducibility testing
Process and calculate signals recorded with the
M E M O R Y H i C O R D E R a n d o u t pu t t h e a r bi t ra r y
waveforms that you create.
Max. 15 V output
Sine wave
Triangular
wave
Square
wave
Pulse wave
Ramp up
wave
Ramp down
wave
Sweep
output
Program
output
Program and generate connected waveforms
Anomaly Simulation
Reproduce and output the observed waveforms
without modification. When resolving problems
observed during research or development, you can
reproduce such problems for efficient testing.
Recommended units
Record anomalous
waveforms
ARBITRARY WAVEFORM
GENERATOR UNIT
U8793
ANALOG UNIT
8966
HIGH RESOLUTION
UNIT
8968
●
Max. 15 V output + amplifier
Reproduce and output
anomalous waveforms
Create power supply waveforms such as power supply dips, instantaneous
interruptions, and voltage fluctuations for immunity tests to regulate malfunctions in
equipment caused by power supply harmonics to perform evaluation testing.
6
The right unit for your
measurement needs
Inverter / UPS Test
Operation testing and evaluation during load
fluctuation
● Confirmation of UPS switching
●
Recommended
units
ANALOG UNIT 8966
LOGIC UNIT 8973
CURRENT UNIT 8971
Perfect for inverter and UPS evaluation / start-up tests.
Record using both logic (control signals) and analog
(primary/secondary voltage or current for a UPS or
inverter).
UPS
Inverter
Power Monitor and Logger
Torque
sensor
Identify power fluctuations when power supply is
turned ON/OFF and during load fluctuations
● Long-term fluctuations in power
●
Recommended
units
Inverter
ANALOG UNIT 8966
Pulse
Encoder
Load
Motor
Battery
HIGH RESOLUTION UNIT 8968
FREQ UNIT 8970
Load the analog output for the rms (instant power
/ voltage / current, etc.) calculated by the power
analyzer, or import the waveform output from the power
analyzer to observe data for long-term tests or irregular
waveforms.
D/A output
Power Analyzer (PW6001)
Control Simulation
Generate simulated output of each type of sensor
signal
● Fluctuating simulated output for 12 V DC car batteries
●
Recommended
units
ARBITRARY WAVEFORM GENERATOR UNIT U8793
WAVEFORM GENERATOR UNIT MR8490
PULSE GENERATOR UNIT MR8791
Use actual waveforms to perform testing on control
boards, such as for engine control, airbags, brake
systems, power steering, and active suspension. This
allows efficient simulation of actual waveforms obtained
from cars.
Generation
13 units
to choose
from
Voltage
ARBITRARY WAVEFORM HIGH VOLTAGE UNIT
GENERATOR UNIT
U8974
U8793
NEW
No. of channels: 2
Arbitrary waveform output
Output frequency range
10m Hz to 100 kHz
●
Max. output: 15 V
●
NEW
Measurement resolution: 16-bit
1/1600 of measurement range
High voltage
Commercial power supply
(primary/secondary)
●
Power equipment characteristics testing
Perfect for control testing of automobiles, high speed trains, and traditional trains
DC voltage
DIGITAL VOLTMETER
UNIT
MR8990
Generation
WAVEFORM
GENERATOR UNIT
MR8790
NEW
NEW
Measurement resolution: 24-bit No. of channels: 4
1/50 000 of measurement range Waveform output
●
●
●
●
●
Multi-channel
Minute sensor voltage
EV battery voltage
●
●
DC output: -10 V to 10 V
Sine wave output
10 mHz to 20 kHz
Pulse
PULSE GENERATOR
UNIT
MR8791
NEW
No. of channels: 8
Pulse output
●
●
Pulse output
0.1 Hz to 20 kHz
Pattern output
Voltage
ANALOG UNIT
8966
Measurement resolution: 12-bit
20 MS/s high-speed sampling
Various amps
Transducers
Sensors
●
Industrial meters
●
●
●
7
MEMORY HiCORDER MR8847A
Vibration / Endurance Tests
Analyze the relationship between engine control
and vibration
● Confirm equipment durability
●
Recommended
units
ARBITRARY WAVEFORM GENERATOR UNIT U8793
HIGH RESOLUTION UNIT 8968
STRAIN UNIT 8969
512 MW of high-capacity memory makes it easy to
observe vibration waveforms for many hours while
performing high-speed sampling. This feature is
perfect for detecting waveform peaks.
Observe minor vibrations with
high precision
Vibration testing equipment
Replace multiple DMMs with a single unit
Save space by replacing multiple desktop DMM units with a single MEMORY HiCORDER.
This eliminates the need to control multiple units and simplifies your system.
Recommended
units
DIGITAL VOLTMETER UNIT
MR8990
Install up to 8 DVM Units to
expand up to 16 channels
2 channels, banana input terminal
High precision, high resolution
DIGITAL VOLTMETER UNIT MR8990
NEW
Fine precision and resolution
Proprietary specifications for DC voltage
measurements
Measure minute fluctuations in sensor output for
automobiles or voltage fluctuations in batteries with
high precision and at high resolution. The maximum
voltage that you can input is 500 V DC. Another
feature is high input resistance.
Measurement range
Effective input
range
(Guaranteed
measurement
accuracy range)
Measurement accuracy
Max.
resolution
5 mV/div
(f.s. = 100 mV)
-120 mV to 120 mV
50 mV/div
(f.s. = 1000 mV)
-1200 mV to 1200 mV
1 μV
-12 V to 12 V
10 μV
500 mV/div (f.s. = 10 V)
0.1 μV
5 V/div
(f.s. = 100 V)
-120 V to 120 V
100 μV
50 V/div
(f.s. = 1000 V)
- 500 V to 500 V
1 mV
Input
resistance
100 MΩ
or more
NPLC:
less than 1
NPLC:
1 or more
±0.01% rdg.
±0.015% f.s.
±0.01% rdg.
±0.01% f.s.
10 MΩ
±5%
±0.01% rdg.
±0.0025% f.s.
±0.025% rdg.
±0.0025% f.s.
● 6.5-digit display (Resolution: 0.1 μV), 24-bit high resolution
Temperature
Voltage
Distortion
Frequency, RPM
Current
Voltage
TEMP UNIT
8967
HIGH RESOLUTION
UNIT
8968
STRAIN UNIT
8969
FREQ UNIT
8970
CURRENT UNIT
8971
Measurement resolution: 16-bit
1/1000 of measurement range
Measurement resolution: 16-bit
1/1600 of measurement range
Measurement resolution: 16-bit
1/1250 of measurement range
Measurement resolution: 16-bit
1/2000 of measurement range
Measurement resolution: 12-bit Measurement resolution: 12-bit
Clamp sensor direct connection RMS measurement
●
Thermocouple
K, J, E, T, N, R, S, B, W
Supply voltage
Primary / secondary inverter
voltage
●
Motor voltage, etc.
Strain gauge converter
Dynamic strain ● Vibration
Pressure ● Acceleration
●
Weight, etc.
●
●
●
●
●
●
●
Encoder
Rotating pulse
●
●
●
Supply current
Inverter current
Motor current, etc.
DC/RMS UNIT
8972
Supply voltage
Primary / secondary inverter
voltage
●
Motor voltage, etc.
●
●
Contact
LOGIC UNIT
8973
No. of channels: 16
Observation of control signal
Voltage / non-voltage
contacts
Relay signals
●
AC / DC signals
●
●
8
Full range of supporting functions
On-site assistance
Help function
Understand operation methods without even reading
the instruction manual using the built-in Help function.
Place the cursor on a field in the settings and press
the HELP button to view a detailed description of that
setting.
Press the HELP button.
A detailed description of the setting is displayed.
Master triggers
Set triggers while viewing
waveforms
Set input triggers while checking waveforms. You
can also display the settings screen separately as a
floating screen.
Trigger functions for m onitoring all
measurement channels
- Level trigger for comparing a single voltage value
- Window trigger for comparing 2 voltage values
- Voltage drop trigger for detecting voltage drops in
commercial power lines
- Period trigger for monitoring periods
- Glitch trigger for detecting anomalies in pulses
- Pattern trigger for comparisons when the logic
signal is ON/OFF
Acquiring data with triggers, and
post-acquisition searching
The MR8847A includes a search function for finding
abnormal waveforms within all of the acquired data.
You can use this function to search for anomalies after
data has been acquired, when it is too difficult to set
triggers because it is not possible to predict what
types of anomalies might be observed.
Adjust levels while displaying waveforms
Set the number of events for each source
* Only for level and glitch triggers
Set trigger conditions in a variety of combinations.
Label each channel
Comment entry function
Set comments for each channel and display them on
the screen, even when observing multiple channels,
making identification easy.
W hen printing, you can also print the channel
comments.
Input comments directly on the unit or by using a USB
keyboard.
Detect instantaneous outages
Setting screen for number of events
9
MEMORY HiCORDER MR8847A
Enlarge waveforms
Zoom function
Display time axis reduced waveforms at the top of
the screen, and time axis enlarged waveforms at the
bottom of the screen. You can use the scroll function
to display the entire waveform while also observing
specific parts.
Collapse waveform
−
Check the entire waveform.
Expand waveform
+
Enlarge/shrink along the time/vertical axes.
Enlarge to observe waveform details
Scan and clip
AB cursor function
Apply the Zoom function to set point A and point B for
the area you want to clip.
Scan
Scan data at the cursor and the waveform's cross
point.
Extract
Specify the segment to save as binary or CSV data.
Conveniently manage scanned data on your computer
PC operations
Connect to L AN for HT TP/F TP
server functions
Use the HT TP function to operate the MEMORY
HiCORDER with a browser on a PC connected via
LAN. You can also use the FTP function to acquire
data from the internal memory or from storage media
inserted in the MEMORY HiCORDER.
You can even acquire data from the internal memory
or from storage media connected to the MEMORY
HiCORDER via USB.
LAN connection
Hub
Operate from computer
Acquire data
10
Record the data you need
Simultaneous recording on storage media
Memory functions
Recording
method
ampling is done at the set period,
S
and all data is recorded.
● Automatic data saving on SSD / CF card or USB memory stick
●During high-speed sampling, data is written to internal memory
Record all
sampling data
first and later saved on other media
● During low-speed sampling, data is written to internal memory
while also saved on other media
Sampling period
Input
● Highly effective for long-term recording
Input
A/D conversion
Maximum Recording Time to internal memory (excerpt)
MR8847-51
MR8847-52
MR8847-53
Maximum recording length
fluctuates depending on
number of channels used.
16 analog channels +
16 internal logic channels
16 analog channels +
16 internal logic channels
16 analog channels +
16 internal logic channels
Sampling
period
40 000 divisions
160 000 divisions
320 000 div
(64 MW)
Time axis
(256 MW)
(512 MW)
Save in real time
CF card
SSD
USB memory stick
Save on devices and media
5 μs/div
50 ns
0.2 s
0.8 s
1.6 s
10 μs/div
100 ns
0.4 s
1.6 s
3.2 s
100 μs/div
1 μs
4s
16 s
32 s
1 ms/div
10 μs
40 s
2 min 40 s
5 min 20 s
100 ms/div
1 ms
1 h 06 min 40 s
4 h 26 min 40 s
8 h 53 min 20 s
1 s/div
10 ms
11 h 06 min 40 s
1 d 20 h 26 min 40 s
3 d 16 h 53 min 20 s
1 min/div
600 ms
27 d 18 h 40 min 00 s
5 min/div
Store on internal memory
111 d 02 h 40 min 00 s 222 d 05 h 20 min 00 s
3.0 s 138 d 21 h 20 min 00 s 555 d 13 h 20 min 00 s 1111 d 02 h 40 min 00 s
-Caution: Available recording duration is determined by internal RAM capacity,
not by external media.
- Caution: Although USB memory sticks enable automatic data saving, for more
reliable data protection, we recommend use of HIOKI CF cards, which are
guaranteed to work with the instrument.
- Note: Table shows maximum values at arbitrary recording length settings.
- Note: Saving to media in near real-time is possible at sampling speeds of
100 ms/div (1 msec sampling) or slower.
Recording and analysis software
WAVE PROCESSOR 9335
LAN COMMUNICATOR 9333
(Software sold separately)
(Software sold separately)
● Waveform display, calculations
● Auto-save waveform data to PC
● Print function
● Remote control via LAN
connection
● Save data in CSV format and
transfer to spreadsheet programs
9335 Brief Specifications
9333 Brief Specifications
Supported
Supported
OS
Windows 8/7 (32/64-bit), Vista (32-bit), XP
OS
- Auto-saves waveform data to PC, Remote control of Memory HiCorder (by
- Display functions: Waveform display, X-Y display, Cursor function, etc.
sending key codes and receiving images on screen), print report, print
- F ile loading: Readable data formats (.MEM, .REC, .RMS, .POW) / Maximum
Functions
files from the Memory HiCorder (binary only)
size may be limited depending on the computer configuration)
etc.
Printing
images from the screen, receive waveform data in same format as waveform
loadable file size: Maximum file size that can be saved by a given device (file
- Data conversion: Conversion to CSV format, Batch conversion of multiple files,
- Print function: Printing image file output (expanded META type, ".EMF")
- Print formatting: 1 up, 2-to-16 up, 2-to-16 rows, X-Y 1-to-4 up, preview, hard
copy
Windows 8/7 (32/64-bit), Vista (32-bit), XP (9333 ver.1.09 or later)
Functions
- Waveform data acquisition: Accept auto-saves from the Memory HiCorder,
same format as auto-save files of Memory HiCorder (binary only), print
automatically with a Memory HiCorder from a PC. The Memory HiCorder's print
key launches printouts on the PC
- Waveform viewer: Simple display of waveform files, conversion to CSV format,
etc.
11
MEMORY HiCORDER MR8847A
Chart recording without missing transient events
Recorder functions
Recording Sampling is done at the set period,
method and data other than the maximum and minimum values is thinned out for recording.
● High-speed sampling ensures that transient events are captured
also with slow recording
● Data compression achieved by recording maximum/minimum value pairs
● Max. 833-day (1 hr/div) long-term recording even for 64 MW model
● Continuous recording until paper runs out for chart output
Sampling
period
To internal memory
20000 divisions
100 ms/div
*Calculated as 30 m = 2,970 divisions
*Changing paper enables semi-permanent
continuation of recording.
Display only
200 ms/div
1 h 6 min 40 s
Display only
500 ms/div
2 h 46 min 40 s
24 min 45 s
5 h 33 min 20 s
49 min 30 s
5 s/div 1 ms, 10 ms,
10 s/div 100 ms
30 s/div
* Limited by
50 s/div combination of
1 min/div
1/100 on time
axis and time axis
2 min/div setting for memory
5 min/div recording
11 h 6 min 40 s
1 h 39 min 00 s
1 d 3 h 46 min 40 s
4 h 7 min 30 s
2 d 7 h 33 min 20 s
8 h 15 min 00 s
6 d 22 h 40 min 00 s
24 h 45 min 00 s
11 d 13 h 46 min 40 s
1 d 17 h 15 min 00 s
23 d 3 h 33 min 20 s
3 d 10 h 30 min 00 s
100 s/div selections under
13 d 21 h 20 min 00 s
2 d 1 h 30 min 00 s
27 d 18 h 40 min 00 s
4 d 3 h 00 min 00 s
69 d 10 h 40 min 00 s
10 d 7 h 30 min 00 s
10 min/div
138 d 21 h 20 min 00 s
20 d 15 h 00 min 00 s
30 min/div
416 d 16 h 00 min 00 s
61 d 21 h 00 min 00 s
1 hr/div
833 d 8 h 00 min 00 s
123 d 18 h 00 min 00 s
HMR Terminal
(iPad App for MEMORY HiCORDER)
Store on internal memory
Chart recording
Continuous (approx. recording time with
30 m paper roll)
33 min 20 s
1 s/div 1 μs, 10 μs,
2 s/div 100 μs,
Screen
display
Input
A/D conversion
Maximum Recording Time with the Recorder function
REC time axis
Input
before data culling
Waveform Viewer Wv
● Easy waveform operation on iPad
● Fingertip operation of Max. 32 channels
of waveform data
● O p e r a t e M E M O R Y H i C O R D E R v i a
network, change settings, and monitor
waveforms during measurement
* New function in Ver 2.0
● Check waveforms with binar y
data on a PC
● Save data in C SV format and
transfer to spreadsheet programs
environment
Supported
OS
iOS on the iPad (Apple Inc.)
Sampling period
Notes
- When opening data created with the Recorder function on a computer, the
maximum and minimum data pairs are lined up in a time series.
- Length of printer paper roll is 30 meters. Paper can be changed during
operation without stopping the recording process.
- With settings between 100 ms and 200 ms/div on the time axis, continuous
recording is not possible if printer is ON.
- The table shows values for the MR8847-51 (64 M-words memory capacity).
Model MR8847-52 (256 MW) can record four times and Model MR8847-53
(512 MW) eight times as much. At ''Continuous'' setting in recording length,
total recording time cannot be increased.
(Bundled software)
Operating
Minimum value
High-speed sampling
Free app (exclusively for iPad) downloadable from the App Store
HMR Terminal Brief Specifications
Maximum value
Reduced saving allows
long-term recording
even during high-speed
sampling
Waveform Viewer (Wv) Brief Specifications
Supported
OS
Windows 8/7 (32/64-bit), Vista (32-bit), XP, 2000
iOS
- Data acquisition: Send to iPad via FTP using a WiFi router, or load to iPad via
iTunes (PC app)
Functions
- Intuitively operate waveform level searches, maximum / minimum / average
values, zero position adjustment, and more at your fingertips
- Waveform monitor
- Meter setting
* Logic waveforms and computational waveforms are not supported.
Functions
- Simple display of waveform files
- Convert binary data files to text format, CSV, etc.
- Scroll function, enlarge/reduce display, jump to cursor/trigger position, etc.
12
Definitive analysis of important data
Calculate parameter values
from measured waveforms
T h e M R 8 8 4 7A c a n p e r f o r m 2 4 c a l c u l a t i o n s ,
including RMS, peak value, and maximum value,
from measured waveforms. It can also per form
time difference measurements, phase difference
measurements, histogram measurements for HIGH
level and LOW level, and statistical processing.
Calculation results are displayed together on the
waveform observation screen.
Process waveforms with
formulas
If you know the required formulas, you can also
per form complicated calculations. By entering
formulas, you can perform a variety of calculations
even after measurements are complete. For example,
you can make the settings shown on the right to find
the RMS value from a measured waveform.
RMS=
1
n
n
i=1
Measured
waveform
di 2
Calculated
waveform
RMS = RMS value
n = Data number
di = Data for channel i
Results are the same
for both numerical
calculations
and waveform processing
Enter formula
FFT analysis function
Specify analysis
points
T he M R 8 8 47A can per form one - signal F F T for
analyzing frequency components, two-signal FFT for
analyzing transfer functions, and octave analysis for
acoustics.
Memory
waveform
FF T calculations from memory
waveforms
When performing FFT analysis of data measured with
the memory function, you can use the jog shuttle to
specify analysis points while also viewing the calculation
results at the same time. You can also display both the
raw data measured with the memory function and the
calculation results for storage waveforms at the same
time, which improves operability during analysis by
displaying spectrum waveforms while checking the
results of window functions.
Running spectrum display
Display the calculation source (memory waveform) and FF T
calculation results at the same time
Change the number of calculation points after measurement
1000 points
Changed to 10000 points
Scaling by "dB"
Before scaling
Display the spectrum as it changes over time in 3D
After scaling
13
MEMORY HiCORDER MR8847A
X-Y RECORDER
Now even easier to use with independent pen up/
down control. Saving data in chronological order
allows records to be saved as digital data, rather than
paper hardcopies that need to be stored.
Pen up/down control
Pen up/down during X-Y recording is controlled
independently. Press the function button or use an
external control terminal (EXT. IN 1, 2, 3) for external
control.
Replaces mechanical pen recorders
Use pen up/down control to record only the required
data. T his allows you to reduce the amount of
unnecessary data that is recorded, and lower the
running cost for paper.
Pen up/down while recording X-Y waveforms
EXT. IN 1 / EXT. IN 2 / EXT. IN 3
Control terminals
Determine waveform quality
Use the waveform judgment function, which monitors
whether a waveform extends beyond the given area,
to easily determine the quality of signal waveforms
that are normally difficult to judge.
For time axis ranges that are slower than 100 msec/
div, you can even make judgments while loading
waveforms. This allows you to take the appropriate
action the moment a poor waveform is detected on
the production line. You can stop the line as soon as
an abnormality is detected.
Judge FFT analysis waveforms
Judge FFT analysis waveforms in the same way.
Judge waveform quality by area
Judge X-Y waveforms
In addition to time axis signals, the MR8847A also
has a waveform judgment function for X-Y waveforms
built in. Use this to detect:
- Displacement and pressure of presses
- Pressure and flow rate of pumps
The X-Y waveforms of the above and other data can
be tested automatically based on area judgment.
Judge FFT analysis waveforms and X-Y waveforms by area
Judgment: Poor
14
Product Specifications
Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Measurement
functions
MEMORY (high-speed recording), RECORDER (real-time recording)
X-Y RECORDER, FFT
Number of input
units
[8 analog input modules]: 16 analog channels + 16 logic channels
(built-in)
[5 analog input modules + 3 logic input modules]: 10 analog
channels + 64 logic channels (16 built-in channels + 48 channels in
logic input modules)
* For analog units, channels are isolated from each other and from
frame GND. For logic units and internal standard logic terminals,
all channels have a common ground.
Max.
sampling speed
20 MS/second (50 ns period, all channels simultaneously)
External sampling (10 MS/second, 100 ns period)
Memory capacity
MR8847-51: Total 64 M-words (Memory expansion: none)
32 MW/ch (using 2 Analog channels), to 4 MW/ch (using 16 Analog
channels)
MR8847-52: Total 256 M-words (Memory expansion: none)
128 MW/ch (using 2 Analog channels), to 16 MW/ch (using 16 Analog
channels)
MR8847-53: Total 512 M-words (Memory expansion: none)
256 MW/ch (using 2 Analog channels), to 32 MW/ch (using
16 Analog channels)
Removable storage
CF card slot (standard) × 1 (up to 2GB, FAT, or FAT-32 format), SSD
(128 GB, optional), USB memory stick (USB 2.0)
Backup function
Clock and parameter setting backup: at least 10 years,
Waveform backup function: none
(At 25°C/ 77°F)
MEMORY (High-speed recording)
Time axis
5 µs to 5 min/div (100 samples/div) 26 ranges, External sampling
(100 samples/div, or free setting), Time axis zoom: x2 to x10 in
3 stages, compression: 1/2 to 1/200 000 in 16 stages
Sampling period
1/100 of time axis range (minimum 50 ns period)
Recording length
MR8847-51: 16 ch mode: 25 to 20 000 div, 2 ch mode: 25 to 200 000 div
(built-in presets) or arbitrary setting in 1-div steps (max. 320 000 div)
MR8847-52: 16 ch mode: 25 to 100 000 div, 2 ch mode: 25 to 1 000 000
div (built-in presets) or arbitrary setting in 1-div steps (max. 1 280 000 div)
MR8847-53: 16 ch mode: 25 to 200 000 div, 2 ch mode: 25 to 2 000 000
div (built-in presets) or arbitrary setting in 1-div steps (max. 2 560 000 div)
Pre-trigger
Record data from before the trigger point at 0 to +100% or -95% of the
recording length in 15 stages, or in 1 div step settings
Numerical
calculations
- Simultaneous calculation for up to 16 selected channels
Average value, effective (rms) value, peak to peak value, maximum
value, time to maximum value, minimum value, time to minimum
value, period, frequency, rise time, fall time, standard deviation,
area value, X-Y area value, specified level time, specified time level,
pulse width, duty ratio, pulse count, four arithmetic operations, time
difference, phase difference, high-level and low-level
- Calculation result evaluation output: GO/NG (with open-collector
5 V output)
- Automatic saving of calculation results
Waveform
processing
- For up to 16 freely selectable channels, the following functions can be
performed (results are automatically stored):
Automatic saving of four arithmetic operations, absolute value,
exponentiation, common logarithm, square root, moving average,
differentiation (primary, secondary), integration (primary,
secondary), parallel displacement along time axis, trigonometric
functions, reverse trigonometric functions, calculation results
Control terminals
External trigger input, Trigger output, External sampling input, Two
external outputs (GO, NG), Three external inputs (START, STOP,
PRINT)
External interface
LAN: 100BASE-TX (FTP server, HTTP server)
USB: USB2.0 compliant, series A receptacle ×1, series B receptacle
×1, (File transfer internal drive/CF card to PC, or remote control
from PC)
Memory
segmentation
- Max. 1024 blocks, sequential storage, multi-block storage
Operation: -10°C to 40°C (14°F to 104°F), 20% to 80% RH
With printer and/or SSD in use: 0°C to 40°C (32°F to 104°F), 20% to
80% RH
Storage: -20°C to 50°C (-4°F to 122°F), 90% RH or less
Other
- No logging
- X-Y waveform synthesis (1-screen, 4-screens)
- Overlay (always overlay when started/overlay only required
waveforms)
- Automatic/ Manual/ A-B cursor range printing/ Report printing
Environmental
conditions
(no condensation)
RECORDER (Real-time recording)
Compliance standard
Safety: EN61010
EMC: EN61326, EN61000-3-2, EN61000-3-3
Power supply
100 to 240 V AC, 50/60 Hz
10 to 28 V DC (use the DC POWER UNIT 9784: Factory installation only)
Time axis
10 ms to 1 hour/div, 19 ranges, time axis resolution 100 points/div * Out of data
acquired at selected sampling rate, only maximum and minimum value data
determined using 100 points/div units are stored.
Time axis compression selectable in 14 steps, from ×1/2 to ×1/50 000
Power consumption
130 VA max. (Printer not used), 220 VA max. (Printer used)
Sampling period
1/10/100 μs, 1/10/100 ms (selectable from 1/100 or less of time axis)
Dimensions and
mass
Approx. 351 mm (13.82 in) W × 261 mm (10.28 in) H × 140 mm
(5.51 in) D, 7.6 kg (268.1 oz) (main unit only)
Instruction Manual ×1, Measurement Guide ×1, Application Disk
(Waveform Maker Software SF8000, Wave Viewer Wv, Communication
Commands table) ×1, Power cord ×1, Input cord label ×1, USB cable ×1,
Printer paper ×1, Roll paper attachment ×2, Ferrite clamp ×1
Real-time printing
Accessories
Supported
* Real-time printing is possible at time axis settings slower than 500 ms/div
* Delayed print is performed when recording length is not set to "Continuous"
and time axis setting is 10 ms to 200 ms/div
* When recording length is set to "Continuous" and time axis setting is 10 ms
to 200 ms/div, manual printing can be performed after measurement stop
Features
Printer paper one-touch loading, high-speed thermal printing
Recording length
Recording Paper
216 mm (8.50 in) × 30 m (98.43 ft), thermal paper roll (use 9231 paper)
Waveform section recording width: 200 mm (7.87 in) 20 division full
scale, 1 div = 10 mm (0.39 in) 80 dots
MR8847-51: Built-in presets of 25 to 20 000 div, or "Continuous" or
arbitrary setting in 1-div steps (max. 20 000 div)
MR8847-52: Built-in presets of 25 to 50 000 div, or "Continuous" or
arbitrary setting in 1-div steps (max. 80 000 div)
MR8847-53: Built-in presets of 25 to 100 000 div, or "Continuous" or
arbitrary setting in 1-div steps (max. 160 000 div)
Recording speed
Max. 50 mm (1.97 in)/sec
Paper feed density
10 dots/mm
Display section
Internal printer
Additional recording Supported (recording is resumed without overwriting previous data)
Waveform memory
10.4 inch SVGA-TFT color LCD (800 × 600 dots)
(Time axis 25 div × Voltage axis 20 div, X-Y waveform
20 div × 20 div)
MR8847-51: Store data for most recent 20 000 div in memory
MR8847-52: Store data for most recent 80 000 div in memory
MR8847-53: Store data for most recent 160 000 div in memory
* Backward scrolling and re-printing available
Auto saving
Data are automatically saved on CF card, USB memory stick or internal
drive after measurement stops.
Display languages
English, Japanese, Korean, Chinese
Other
- No logging
- Manual/ A-B cursor range printing/ Report printing
Waveform display
zoom/compression
Time axis: x10 to x2 (zoom at MEMORY function only), x1, x1/2 to
x1/20 000
Voltage axis: x100 to x2, x1, x1/2 to x1/10
Variable display
Upper/Lower limit set, display/div set
Scaling
10:1 to 1000:1, automatic scaling for various probes
Manual scaling (conversion ratio setting, 2-point setting, unit setting)
Display
X-Y RECORDER (X-Y real-time recording)
Sampling period
1/10/100 ms (dot), 10/100 ms (line)
Recording length
Continuous
Screen, Printing
Split screen (1 or 4), Manual printing only
Number of X-Y
1 to 8 phenomena
X-Y channel setting
Any 8 channels out of 16 can be selected for X axis and Y axis
respectively
Comment entry
Alphanumeric input (title, analog and logic channels), Simple input,
history input, phrase input
Logic waveform
Display point move 1% step, Line width 3 types
Display partition
Max. 16 graphs
X-Y axis resolution
25 dots/div (screen), horizontal 80 dots/div × vertical 80 dots/div
(printer)
Monitor functions
- Level monitor
- Numerical value (sampling 10 kS/s fixed, refresh rate 0.5 s)
Waveform memory
Sampling data for last 4 000 000 points are stored in memory
Pen up/down
Simultaneous for all phenomena
External pen control
Possible via external input connector (simultaneous up/down for all
phenomena)
Other display
functions
- Waveform inversion (positive/negative)
- Cursor measurement (A, B, 2-cursor, for all channels)
- Vernier function (amplitude fine adjustment)
- Zoom function (horizontal screen division, zoomed waveform shown
in lower section)
- 16 selectable colors for waveform display
- Zero position shift in 1% steps for analog waveform
- Global zero adjust for all channels and all ranges
15
MEMORY HiCORDER MR8847A
Trigger functions
Other
Trigger mode
MEMORY (high-speed recording), FFT: Single, Repeat, Auto
RECORDER (real-time recording): Single, Repeat
Trigger source
CH1 to CH16 (analog), Standard Logic 16ch + Logic Unit (Max.
3 units 48 channels), External (a rise of 2.5V or terminal short
circuit), Timer, Manual (either ON or OFF for each source), Logical
AND/OR of sources
Trigger types
- Level: Triggering occurs when preset voltage level is crossed
(upwards or downwards)
- Voltage drop: Triggering occurs when voltage drops below peak
voltage setting (for 50/60 Hz commercial power supply only)
- Window: Triggering occurs when window defined by upper and
lower limit is entered or exited
- Period: Rising edge or falling edge cycle of preset voltage value
is measured and triggering occurs when defined cycle range is
exceeded
- Glitch: Triggering occurs when pulse width from rising or falling
edge of preset voltage value is under run
- Event setting: Event count is performed for each source, and
triggering occurs when a preset count is exceeded
- Logic: 1, 0, or ×, Pattern setting
- Area comparison with reference waveform area for time domain
waveform, X-Y waveform, or FFT analysis waveform
Waveform judgment
- Parameter calculated value comparison with reference value
function
(In MEMORY or
- Output: GO/NG decision, Open-collector 5V,
FFT function)
*100 msec/div (1 msec sampling) and thereafter allows for evaluation in almost
real-time.
FFT function
Analysis mode
Storage waveform, Linear spectrum, RMS spectrum, Power
spectrum, Density of power spectrum, Cross power spectrum,
Auto-correlation function, Histogram, Transfer function,
Cross-correlation function, Impulse response, Coherence function,
1/1 Octave analysis, 1/3 Octave analysis, LPC analysis, Phase
spectrum
Analysis channels
Selectable from all analog input channels
133 mHz to 8 MHz, External
(resolution 1/400, 1/800, 1/2000, 1/4000)
Level setting
resolution
0.1% of full scale (full scale = 20 divisions)
Frequency range
Trigger filter
Selectable 0.1 div to 10.0 div, or OFF (high-speed recording)
ON (10 ms fixed) or OFF (at RECORDER function)
Number of sampling
1000, 2000, 5000, 10 000 points
points
Trigger output
Open collector (5 voltage output, active Low)
At Level setting: pulse width (Sampling period × data number after
trigger)
At Pulse setting: pulse width (2 ms)
Window functions
Rectangular, Hanning, Hamming, Blackman, Blackman-Harris,
Flat-top, Exponential
Display format
Single, Dual, Nyquist, Running spectrum
Other functions
Trigger priority (OFF/ON), Pre-trigger function for capturing
data from before / after trigger event (at MEMORY function),
Level display during trigger standby, Start and stop trigger (at
RECORDER function), Trigger search function
Averaging function
Time axis / frequency axis simple averaging, Exponential
averaging, Peak hold (frequency axis), Averaging times: 2 times to
10000 times
Print functions
Same as the MEMORY function (partial print not available)
- Measurement Indices (Input units sold separately)
- Maximum Internal Memory Recording Time (MEMORY Function)
MR8847-51 (64 MW)
Maximum recording length
increases depending on
number of channels used
Time axis
5 μs/div
Sampling
period
50 ns
Analog 16 ch
+ internal Logic 16 ch
40 000 divisions
Analog 2 ch
+ internal Logic 16 ch
320 000 div
0.2 s
1.6 s
MR8847-52 (256 MW)
Analog 16 ch
+ internal Logic 16 ch
160 000 divisions
Analog 2 ch
+ internal Logic 16 ch
1 280 000 divisions
0.8 s
MR8847-53 (512 MW)
Analog 16 ch
+ internal Logic 16 ch
320 000 div
Measurement
targets
12.8 s
10 μs/div
100 ns
0.4 s
3.2 s
1.6 s
12.8 s
3.2 s
25.6 s
20 μs/div
200 ns
0.8 s
6.4 s
3.2 s
25.6 s
6.4 s
51.2 s
50 μs/div
500 ns
2s
16 s
8s
1 min 04 s
16 s
2 min 08 s
100 μs/div
1 μs
4s
32 s
16 s
2 min 08 s
32 s
4 min 16 s
200 μs/div
2 μs
8s
1 min 04 s
32 s
4 min 16 s
1 min 04 s
8 min 32 s
500 μs/div
5 μs
20 s
2 min 40 s
1 min 20 s
10 min 40 s
2 min 40 s
21 min 20 s
1 ms/div
10 μs
40 s
5 min 20 s
2 min 40 s
21 min 20 s
5 min 20 s
42 min 40 s
2 ms/div
20 μs
1 min 20 s
10 min 40 s
5 min 20 s
42 min 40 s
10 min 40 s
1 h 25 min 20 s
5 ms/div
50 μs
3 min 20 s
26 min 40 s
13 min 20 s
1 h 46 min 40 s
26 min 40 s
3 h 33 min 20 s
10 ms/div
100 μs
6 min 40 s
53 min 20 s
26 min 40 s
3 h 33 min 20 s
53 min 20 s
7 h 06 min 40 s
20 ms/div
200 μs
13 min 20 s
1 h 46 min 40 s
53 min 20 s
7 h 06 min 40 s
1 h 46 min 40 s
14 h 13 min 20 s
50 ms/div
500 μs
33 min 20 s
4 h 26 min 40 s
2 h 13 min 20 s
17 h 46 min 40 s
4 h 26 min 40 s
35 h 33 min 20 s
100 ms/div
1 ms
1 h 06 min 40 s
8 h 53 min 20 s
4 h 26 min 40 s
1 d 11 h 33 min 20 s
8 h 53 min 20 s
2 d 23 h 06 min 40 s
200 ms/div
2 ms
2 h 13 min 20 s
17 h 46 min 40 s
8 h 53 min 20 s
2 d 23 h 06 min 40 s
17 h 46 min 40 s
5 d 22 h 13 min 20 s
500 ms/div
5 ms
5 h 33 min 20 s
1 d 20 h 26 min 40 s
22 h 13 min 20 s
7 d 09 h 46 min 40 s
44 h 26 min 40 s
14 d 19 h 33 min 20 s
1 s/div
10 ms
11 h 06 min 40 s
3 d 16 h 53 min 20 s
1 d 20 h 26 min 40 s
14 d 19 h 33 min 20 s
3 d 16 h 53 min 20 s
29 d 15 h 06 min 40 s
2 s/div
20 ms
22 h 13 min 20 s
7 d 09 h 46 min 40 s
3 d 16 h 53 min 20 s
29 d 15 h 06 min 40 s
7 d 09 h 46 min 40 s
59 d 06 h 13 min 20 s
9 d 06 h 13 min 20 s
Max.
resolution
100 mV f.s. to 400 V f.s.
50 µV
HIGH
RESOLUTION
UNIT 8968
100 mV f.s. to 400 V f.s.
3.125 µV
DC/RMS UNIT
8972
100 mV f.s. to 400 V f.s.
50 µV
HIGH-VOLTAGE
UNIT U8974
4 V f.s. to 1000 V f.s.
0.125 mV
Current
CURRENT UNIT
8971 + optional
current sensor
20 A f.s. or larger
When driving current sensors
with separate power supply,
measurement can be conducted
with voltage input units.
1 mA or
larger
RMS AC
voltage
DC/RMS UNIT
8972
100 mV f.s. to 400 V f.s.
50 µV
Temperature
(Thermocouple
input)
TEMP UNIT 8967
200°C (392°F) f.s. to
2000°C (3632°F) f.s.
0.01°C
Note: Upper and lower
(0.02°F)
limit values depend on the
thermocouple
Frequency,
RPM
FREQ UNIT 8970
20 Hz f.s. to 100 kHz f.s.
2 (kr/min) f.s. to 2000 (kr/min) f.s.
2 mHz
0.2 (r/min)
Power supply
frequency
FREQ UNIT 8970
40 to 60 Hz, 50 to 70 Hz,
390 to 410 Hz
0.01 Hz
Integration
count
FREQ UNIT 8970
40 k-counts f.s. to 20
M-counts f.s.
1 count
2 560 000 divisions
1.6 s
Display range
ANALOG UNIT
8966
Analog 2 ch
+ internal Logic 16 ch
6.4 s
With use
input unit
Voltage
5 s/div
50 ms
2 d 07 h 33 min 20 s
18 d 12 h 26 min 40 s
74 d 01 h 46 min 40 s
18 d 12 h 26 min 40 s 148 d 03 h 33 min 20 s
10 s/div
100 ms
4 d 15 h 06 min 40 s
37 d 00 h 53 min 20 s
18 d 12 h 06 min 40 s 148 d 03 h 33 min 20 s
37 d 00 h 53 min 20 s 296 d 07 h 06 min 40 s
30 s/div
300 ms
13 d 21 h 20 min 00 s
111 d 02 h 40 min 00 s
55 d 13 h 20 min 00 s 444 d 10 h 40 min 00 s
111 d 02 h 40 min 00 s 888 d 21 h 20 min 00 s
Pulse duty
ratio
FREQ UNIT 8970
100% f.s.
0.01%
Pulse width
FREQ UNIT 8970
0.01 s f.s. to 2 s f.s.
1 µs
Vibration
stress
STRAIN UNIT 8969 400 µe f.s. to 20000 µe f.s. 0.016 µe
Relay contacts,
voltage on/off
LOGIC UNIT 8973
50 s/div
500 ms
23 d 03 h 33 min 20 s 185 d 04 h 26 min 40 s
92 d 14 h 13 min 20 s
740 d 17 h 46 min 40 s
185 d 04 h 26 min 40 s
"H"
1 min/div
600 ms
27 d 18 h 40 min 00 s 222 d 05 h 20 min 00 s
111 d 02 h 40 min 00 s 888 d 21 h 20 min 00 s
222 d 05 h 20 min 00 s
"H"
100 s/div
1.0 s
46 d 07 h 06 min 40 s 370 d 08 h 53 min 20 s
185 d 04 h 26 min 40 s
"H"
370 d 08 h 53 min 20 s
"H"
2 min/div
1.2 s
55 d 13 h 20 min 00 s 444 d 10 h 40 min 00 s
222 d 05 h 20 min 00 s
"H"
444 d 10 h 40 min 00 s
"H"
5 min/div
3.0 s 138 d 21 h 20 min 00 s
555 d 13 h 20 min 00 s
"H"
"H"
"H"
"H"
Notes
- The above table shows maximum values at arbitrary recording length settings.
- Saving to media in near real-time is possible at sampling speeds of 100 ms/div (1 msec sampling) or slower.
- Operation cannot be guaranteed for extended recording periods one year or longer. The above table represents
theoretical values.
—
—
Notes
- Each unit has two input channels, except Logic Unit.
- Besides logic units (16 channels), the MR8847A series comes
standard with 16 logic inputs integrated in the device.
16
Optional Specifications (sold separately)
Dimensions and mass: approx. 106 mm (4.17 in) W ×
19.8 mm (0.78 in) H × 196.5 mm (7.74 in) D, approx. 250 g (8.8 oz)
Accessories: None
ANALOG UNIT 8966
(Accuracy at 23 ±5°C/73 ±9°F, 20 to 80% rh after 30 minutes of warm-up time and zero
adjustment; Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Measurement functions
Number of channels: 2, for voltage measurement
Isolated BNC connector (input impedance 1 MΩ, input capacitance 30 pF),
Max. rated voltage to ground: 300 V AC or DC (with input isolated from the main unit,
Input terminals
the maximum voltage that can be applied between input channel and chassis, and between
input channels without damage)
5 mV to 20 V/div, 12 ranges, full scale: 20 div,
AC voltage for possible measurement/display using the memory function: 280 V rms,
Measurement range
Low-pass filter: 5/50/500 Hz, 5 k/50 k/500 kHz
Measurement resolution 1/100 of range (using 12-bit A/D conversion)
20 MS/s (simultaneous sampling in 2 channels)
Maximum sampling rate
Measurement accuracy ±0.5% of full scale (with filter 5 Hz, zero position accuracy included)
Frequency
DC to 5 MHz -3 dB, (with AC coupling: 7 Hz to 5 MHz -3 dB)
characteristics
AC/DC/GND
Input coupling
400 V DC (maximum voltage that can be applied between input connectors without
Maximum input voltage
damage)
Dimensions and mass: approx. 106 mm (4.17 in) W × 19.8 mm
(0.78 in) H × 204.5 mm (8.05 in) D, approx. 240 g (8.5 oz)
Accessories: Ferrite clamp × 2
TEMP UNIT 8967
Measurement functions
Input terminals
Temperature measurement
range
Note: Upper and lower limit values
depend on the thermocouple
Thermocouple range
(JIS C 1602-1995)
(ASTM E-988-96)
Data refresh rate
Measurement accuracy
(Accuracy at 23 ±5°C/73 ±9°F, 20 to 80% rh after 30 minutes of warm-up time and zero
adjustment; Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Number of channels: 2, for temperature measurement with thermocouple (voltage
measurement not available)
Thermocouple input: plug-in connector, Recommended wire diameter: single-wire, 0.14 to 1.5
mm 2, braided wire 0.14 to 1.0 mm 2 (conductor wire diameter min. 0.18 mm), AWG 26 to 16
Input impedance: min. 5 M Ω (with line fault detection ON/OFF),
Max. rated voltage to ground: 300 V AC or DC (with input isolated from the main unit, the
maximum voltage that can be applied between input channel and chassis, and between input
channels without damage)
10°C (50°F)/div (-100°C to 200°C (-148°F to 392°F)), 50°C (122°F)/div (-200°C to 1000°C
(-328°F to 1832°F)), 100°C (212°F)/div (-200°C to 2000°C (-328°F to 3632°F)), 3 ranges,
full scale: 20 div,
Measurement resolution: 1/1000 of measurement range (using 16-bit A/D conversion)
K: -200°C to 1350°C (-328°F to 2462°F), J: -200°C to 1100°C (-328°F to 2012°F), E: -200°C
to 800°C (-328°F to 1472°F), T: -200°C to 400°C (-328°F to 752°F), N: -200°C to 1300°C
(-328°F to 2372°F), R: 0°C to 1700°C (32°F to 3092°F), S: 0°C to 1700°C (32°F to 3092°F),
B: 400°C to 1800°C (752°F to 3272°F), W (WRe5-26): 0°C to 2000°C (32°F to 3632°F),
Reference junction compensation: internal/ external (switchable), Line fault detection ON/
OFF possible
3 methods, Fast: 1.2 ms (digital filter OFF), Normal: 100 ms (digital filter 50/60 Hz), Slow:
500 ms (digital filter 10 Hz)
Thermocouple K, J, E, T, N: ±0.1% of full scale ±1°C (±1.8°F) (±0.1% of full scale ±2°C
(±3.6°F) at -200°C to 0°C (-328°F to 32°F)),
Thermocouple R, S, B, W: ±0.1% of full scale ±3.5°C (±6.3°F) (at 0°C (32°F) to less than 400°C
(752°F); However, no accuracy guarantee of less than 400°C (752°F) for B), ±0.1% f.s. ±3°C
(±5.4°F) (at 400°C (752°F) or more)
Reference junction compensation accuracy: ±1.5°C (±2.7°F) (added to measurement accuracy
with internal reference junction compensation)
Dimensions and mass: approx. 106 mm (4.17 in) W × 19.8 mm
(0.78 in) H × 196.5 mm (7.74 in) D, approx. 250 g (8.8 oz)
Accessories: None
HIGH RESOLUTION UNIT 8968
(Accuracy at 23 ±5°C/73 ±9°F, 20 to 80% rh after 30 minutes of warm-up time and zero
adjustment; Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Measurement functions
Number of channels: 2, for voltage measurement
Isolated BNC connector (input impedance 1 MΩ, input capacitance 30 pF),
Max. rated voltage to ground: 300 V AC, DC (with input isolated from the unit,
Input terminals
the maximum voltage that can be applied between input channel and chassis and between
input channels without damage)
5 mV to 20 V/div, 12 ranges, full scale: 20 div,
AC voltage for possible measurement/display using the memory function: 280 V rms,
Measurement range
Low-pass filter: 5/50/500 Hz, 5k/50k Hz
Integrated filter for suppressing aliasing distortion caused by FFT processing (automatic
Anti-aliasing filter
cutoff frequency setting/OFF)
Measurement resolution 1/1600 of measurement range (using 16-bit A/D conversion)
Maximum sampling rate
Measurement accuracy
Frequency
characteristics
Input coupling
Maximum input voltage
1 MS/s (simultaneous sampling in 2 channels)
±0.3% of full scale (with filter 5 Hz, zero position accuracy included)
DC to 100 kHz -3 dB (with AC coupling: 7 Hz to 100 kHz -3 dB)
AC/DC/GND
400 V DC (maximum voltage that can be applied between input connectors without
damage)
Dimensions and mass: approx. 106 mm (4.17 in) W × 19.8 mm
(0.78 in) H × 196.5 mm (7.74 in) D, approx. 220 g (7.8 oz)
Accessories: Conversion cable 9769 × 2 (cable length 50 cm/1.64 ft)
STRAIN UNIT 8969
(Accuracy at 23 ±5°C/73 ±9°F, 20 to 80% rh after 30 minutes of warm-up time and autobalance; Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Number of channels: 2, for distortion measurement (electronic auto-balancing, balance
adjustment range
within ±10 000 µe or less)
Weidmuller SL 3.5/7/90G
(via Conversion Cable 9769, TAJIMI PRC03-12A10-7M10.5)
Max. rated voltage to ground: 33 V rms or 70 V DC (with input isolated from the unit,
Input terminals
the maximum voltage that can be applied between input channel and chassis and between
input channels without damage)
Strain gauge converter, Bridge impedance: 120 Ω to 1 kΩ, Bridge voltage: 2 V ±0.05 V,
Suitable transducer
Gauge rate: 2.0
20 µe to 1000 µe/div, 6 ranges, full scale: 20 div,
Measurement range
Low-pass filter: 5/10/100 Hz, 1 kHz
Measurement resolution 1/1250 of measurement range (using 16-bit A/D conversion)
Dimensions and mass: approx. 106 mm (4.17 in) W × 19.8 mm
(0.78 in) H × 196.5 mm (7.74 in) D, approx. 250 g (8.8 oz)
Accessories: None
FREQ UNIT 8970
Measurement functions
Input terminals
Isolated BNC connector (input impedance 1 MΩ, input capacitance 30 pF),
Max. rated voltage to ground: 300 V AC or DC (with input isolated from the main unit,
the maximum voltage that can be applied between input channel and chassis, and between
input channels without damage)
Frequency mode
Range: Between DC to 100 kHz (minimum pulse width 2 μs), 1 Hz/div to 5 kHz/div (full
scale = 20 div), 8 settings
Accuracy: ±0.1% f.s. (exclude 5 kHz/div), ±0.7% f.s. (at 5 kHz/div)
Rotation mode
Range: Bet ween 0 to 2 million rotations/minute (minimu m pulse width 2 μs),
100 (r/min)/div to 100 k (r/min)/div (full scale = 20 div), 7 settings
Accuracy: ±0.1% f.s. (excluding 100 k (r/min)/div), ±0.7% f.s. (at 100 k (r/min)/div)
Power frequency mode
Range: 50 Hz (40 to 60 Hz), 60 Hz (50 to 70 Hz), 400 Hz (390 to 410 Hz) (full scale =
20 div), 3 settings
Accuracy: ±0.03 Hz (50, 60 Hz), ±0.1 Hz (400 Hz range)
Integration mode
Range: 2 k counts/div to 1 M counts/div, 6 settings
Accuracy: ±range/2000
Duty ratio mode
Range: Between 10 Hz to 100 kHz (minimum pulse width 2 μs), 5%/div (full scale =
20 div)
Accuracy: ±1% (10 Hz to 10 kHz), ±4% (10 kHz to 100 kHz)
Pulse width mode
Range: Between 2 μs to 2 sec, 500 μs/div to 100 ms/dv (full scale = 20 div), Accuracy:
±0.1% f.s.
Measurement resolution
1/2000 of range (Integration mode), 1/500 of range (exclude integration, power frequency
mode), 1/100 of range (power frequency mode)
Input voltage range and
threshold level
±10 V to ±400 V, 6 settings, selectable threshold level at each range
Other functions
Slope, Level, Hold, Smoothing, Low-pass filter, Switchable DC/AC input coupling,
Frequency dividing, Integration over-range keep/return
Dimensions and mass: approx. 106 mm (4.17 in) W × 19.8 mm (0.78 in)
H × 196.5 mm (7.74 in) D, approx. 250 g (8.8 oz)
Accessories: CONVERSION CABLE 9318 × 2
(To connect the current sensor to the 8971)
CURRENT UNIT 8971
200 kS/s (simultaneous sampling across 2 channels)
±(0.5% f.s. +4 µe) (5 Hz filter ON)
DC to 20 kHz +1/-3 dB
(Accuracy at 23 ±5°C/73 ±9°F, 20 to 80% rh after 30 minutes of warm-up time and zero
adjustment; Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Measurement functions
Number of channels: 2, Current measurement with optional current sensor,
Note: Maximum 4 units connectable to main unit
Input terminals
Sensor connector (input impedance 1 MΩ, exclusive connector for current sensor via
conversion cable the 9318, common GND with recorder)
Compatible current
sensors
CT6863, CT6862, 9709, CT6841, CT6843, CT6844, CT6845, 9272-10 (To connect the 8971
via conversion cable the 9318)
Measurement range
Using 9272-10 (20 A), CT6841: 100 mA to 5 A/div (f.s. = 20 div, 6 settings)
Using CT6862: 200 mA to 10 A/div (f.s. = 20 div, 6 settings)
Using 9272-10 (200 A), CT6843, CT6863: 1 A to 50 A/div
(f.s. = 20 div, 6 settings)
Using CT6844, CT6845, 9709: 2 A to 100 A/div (f.s. = 20 div, 6 settings)
Measurement accuracy
(with 5 Hz filter ON)
Note: Add the accuracy
and attributes of the
current sensor being used.
±0.65% f.s.
RMS amplitude accuracy: ±1% f.s. (DC, 30 Hz to 1 kHz), ±3% f.s. (1 kHz to 10 kHz)
RMS response time: 100 ms (rise time from 0 to 90% of full scale),
Crest factor: 2
Frequency characteristics: DC to 100 kHz, ±3 dB (with AC coupling: 7 Hz to 100 kHz)
Measurement resolution 1/100 of range (using 12-bit A/D conversion)
Maximum sampling rate
1 MS/s (simultaneous sampling in 2 channels)
Other functions
Input coupling: AC/DC/GND, Low-pass filter: 5, 50, 500, 5 k, 50 kHz
Dimensions and mass: approx. 106 mm (4.17 in) W × 19.8 mm
(0.78 in) H × 196.5 mm (7.74 in) D, approx. 250 g (8.8 oz)
Accessories: None
DC/RMS UNIT 8972
(Accuracy at 23 ±5°C/73 ±9°F, 20 to 80% rh after 30 minutes of warm-up time and zero
adjustment; Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Measurement functions
Number of channels: 2, for voltage measurement, DC/RMS selectable
Input terminals
Isolated BNC connector (input impedance 1 MΩ, input capacitance 30 pF),
Max. rated voltage to ground: 300 V AC or DC (with input isolated from the main unit,
the maximum voltage that can be applied between input channel and chassis, and between
input channels without damage)
Measurement range
5 mV to 20 V/div, 12 ranges, full scale: 20 div,
AC voltage for possible measurement/display using the memory function: 280 V rms,
Low-pass filter: 5/50/500 Hz, 5 k/100 kHz
Measurement resolution 1/100 of range (using 12-bit A/D conversion)
Maximum sampling rate
1 MS/s (simultaneous sampling in 2 channels)
Measurement accuracy
±0.5% of full scale (with filter 5 Hz, zero position accuracy included)
RMS measurement
RMS amplitude accuracy: ±1% f.s. (DC, 30 Hz to 1 kHz), ±3% of full scale (1 kHz to
100 kHz)
Response time: SLOW 5 s (rise time from 0 to 90% of full scale), MID 800 ms (rise time
from 0 to 90% of full scale), FAST 100 ms (rise time from 0 to 90% of full scale),
Crest factor: 2
Frequency
characteristics
DC to 400 kHz -3 dB, (with AC coupling: 7 Hz to 400 kHz -3 dB)
Input coupling
AC/DC/GND
Maximum input voltage
400 V DC (maximum voltage that can be applied between input connectors without
damage)
Measurement functions
Maximum sampling rate
Measurement accuracy
After auto-balancing
Frequency
characteristics
(Accuracy at 23 ±5°C/73 ±9°F, 20 to 80% rh after 30 minutes of warm-up time;
Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Number of channels: 2, for voltage input based frequency measurement, rotation, power
frequency, integration, pulse duty ratio, pulse width
Dimensions and mass: approx. 106 mm (4.17 in) W × 19.8 mm
(0.78 in) H × 196.5 mm (7.74 in) D, approx. 190 g (6.7 oz)
Accessories: None
LOGIC UNIT 8973
Measurement functions Number of channels: 16 channels (4 ch/1 probe connector × 4 connectors)
Input terminals
Mini DIN connector (for HIOKI logic probes only),
Compatible logic probes: 9320-01, 9327, MR9321-01
17
MEMORY HiCORDER MR8847A
NEW
Dimensions and mass: approx. 106 mm (4.17 in) W × 19.8 mm
(0.78 in) H × 196.5 mm (7.74 in) D, approx. 260 g (9.2 oz)
Accessories: None
DIGITAL VOLTMETER UNIT MR8990
(Accuracy at 23 ±5°C/73 ±9°F, 20 to 80% rh after 30 minutes of warm-up time and calibration,
Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Measurement functions
Number of channels: 2, for DC voltage measurement
Input terminals
Banana input connectors (Input resistance: 100 MΩ or higher with 100 mV f.s. to 10 V f.s.
range, otherwise 10 MΩ)
Max. rated voltage to ground: 300 V AC or DC (with input isolated from the main unit,
the maximum voltage that can be applied between input channel and chassis, and between
input channels without damage)
Measurement range
100 mV f.s. (5 mV/div) to 1000 V f.s. (50 V/div), 5 ranges, full scale: 20 div
Measurement resolution
1/50 000 of measurement range (using 24 bit ΔΣ modulation A/D)
Integration time
20 ms ×NPLC (during 50 Hz), 16.67 ms ×NPLC (during 60 Hz)
Response time
2 ms +2× integration time or less (rise - f.s. → + f.s., fall + f.s. → - f.s.)
Basic measurement
accuracy
±0.01% rdg. ±0.0025% f.s. (at range of 1000 mV f.s.)
Maximum input voltage
500 V DC (maximum voltage that can be applied between input connectors without damage)
Note: Cannot use with 8847 or MR8847
NEW
Dimensions and mass: approx. 106 mm (4.17 in) W × 19.8 mm
(0.78 in) H × 196.5 mm (7.74 in) D, approx. 230 g (8.1 oz)
Accessories: None
HIGH-VOLTAGE UNIT U8974
(Accuracy at 23 ±5°C/73 ±9°F, 20 to 80% rh after 30 minutes of warm-up time and zero
adjustment; Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Measurement functions
Number of channels: 2, for voltage measurement, DC/RMS selectable
Maximum rated voltage to ground: 1000 V AC or DC (CAT III), 600 V AC or DC
(CAT IV)
Input terminals
Banana input terminal (Input impedance: 4 MΩ, Input capacitance: 5 pF)
Measurement range
200 mV, 500 mV, 1, 2, 5, 10, 20, 50 V/div (DC mode)
500 mV, 1, 2, 5, 10, 20, 50 V/div (RMS mode)
Measurement
resolution
1/1600 of measurement range (using 16-bit A/D conversion)
Maximum sampling rate
1 MS/s
Measurement accuracy
±0.25% f.s. (with filter 5 Hz, zero position accuracy included)
RMS measurement
RMS accuracy: ±1.5% f.s. (DC, 30 Hz to 1 kHz), ±3% f.s. (1 kHz to 100 kHz)
Response time: High speed 150 ms, Medium speed 500 ms, Low speed 2.5 s
Frequency
characteristics
DC to 100 kHz -3 dB
Input coupling
DC / GND
Maximum input voltage
1000 V DC, 700 V AC
NEW
Dimensions and mass: approx. 106 mm (4.17 in) W × 19.8 mm
(0.78 in) H × 196.5 mm (7.74 in) D, approx. 250 g (8.8 oz)
Accessories: None
ARBITRARY WAVEFORM GENERATOR UNIT U8793
Output terminal
Output voltage range
Max. output current
FG function
Arbitrary waveform
generator mode
Sweep function
Program function
Other
(Accuracy at 23 ±5°C/73 ±9°F, 80% rh or less after 30 minutes or more of warm-up time;
Power supply frequency range of installed MEMORY HiCORDER at 50 Hz/60 Hz ±2 Hz;
Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Number of channels: 2, SMB terminal (Output impedance: 1 Ω or less)
Max. rated voltage to ground: 33 V rms AC or 70 V DC
-10 V to 15 V (Amplitude setting range: 0 V to 20 V p-p, Setting resolution: 1 mV)
10 mA (Allowable load resistance: 1.5 kΩ or more)
DC, Sine wave, Square wave, Pulse wave, Triangular wave, Ramp wave, Output
frequency: 0 Hz to 100 kHz
Waveforms measured by MR8847A, etc., generated by Hioki Model 7075 or SF8000, CSV waveforms
D/A refresh rate: 2 MHz (using 16-bit D/A)
Frequency, Amplitude, Offset, Duty (Pulse only)
Max. 128 steps (Number of loops for each step, Number of total loops)
Self-test function (Voltage), External input/output control
Note: Cannot use with 8847 or MR8847
NEW
Dimensions and mass: approx. 106 mm (4.17 in) W × 19.8 mm
(0.78 in) H × 196.5 mm (7.74 in) D, approx. 230 g (8.1 oz)
Accessories: None
WAVEFORM GENERATOR UNIT MR8790
(Accuracy at 23 ±5°C/73 ±9°F, 80% rh after 30 minutes of warm-up time;
Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Output terminal
Number of channels: 4, SMB terminal (Output impedance: 1 Ω or less)
Max. rated voltage to ground: 33 V rms AC or 70 V DC
Output voltage range
-10 V to 10 V (Amplitude setting range: 0 V to 20 V p-p, Setting resolution: 1 mV)
Max. output current
5 mA
Output function
DC, Sine wave (Output frequency range: 0 Hz to 20 kHz)
Accuracy
Amplitude accuracy: ±0.25% of setting ±2 mV p-p (1 Hz to 10 kHz)
Offset accuracy: ±3 mV
DC output accuracy: ±0.6 mV
Other
Self-test function (Voltage, Current)
Cable length and mass: Input side: 70 cm (2.30 ft), Output side: 1.5 m
(4.92 ft), Approx. 170 g (6.0 oz)
DIFFERENTIAL PROBE P9000
Measurement modes
Division ratio
DC output accuracy
Effective value
measurement accuracy
Input resistance/
capacity
Maximum input voltage
Maximum rated voltage to
ground
Operating temperature
range
Power supply
Accessories
(Accuracy guaranteed for 1 year,
Post-adjustment accuracy guaranteed for 1 year)
P9000-01: For waveform monitor output, Frequency characteristics: DC to 100 kHz
-3 dB
P9000-02: Switches between waveform monitor output/AC effective value output
Wave mode frequency characteristics: DC to 100 kHz -3 dB,
RMS mode frequency characteristics: 30 Hz to 10 kHz, Response time: Rise
300 ms, Fall 600 ms
Switches between 1000:1, 100:1
±0.5% f.s. (f.s. = 1.0 V, division ratio 1000:1), (f.s. = 3.5 V, division ratio 100:1)
±1% f.s. (30 Hz to less than 1 kHz, sine wave), ±3% f.s. (1 kHz to 10 kHz, sine wave)
H-L: 10.5 MΩ, 5 pF or less (At 100 kHz)
1000 V AC, DC
1000 V AC, DC (CAT III)
-40°C to 80°C (-40°F to 176°F)
(1) AC adapter Z1008 (100 to 240 V AC, 50/60 Hz), 6 VA (including AC adapter),
0.9 VA (main unit only)
(2) USB bus power (5 V DC, USB micro-B connector), 0.8 VA
(3) External power source 2.7 V to 15 V DC, 1 VA
Instruction manual ×1, Alligator clip ×2, Carrying case ×1
Cable length and mass: Main unit cable 1.3 m (4.27 ft), input section
cable 46 cm (1.51 ft), approx. 350 g (12.3 oz)
DIFFERENTIAL PROBE 9322
(Accuracy guaranteed for 1 year)
Functions
For high-voltage f loating measurement, power line surge noise detection, RMS
rectified output measurement
DC mode
For waveform monitor output, Frequency characteristics: DC to 10 MHz (±3 dB),
Amplitude accuracy: ±1% of full scale (at max. 1000 V DC), ±3% of full scale (at
max. 2000 V DC) (full scale: 2000 V DC)
AC mode
For detection of power line surge noise, Frequency characteristics: 1 kHz to 10 MHz
±3 dB
RMS mode
DC/AC voltage RMS output detection, Frequency characteristics: DC, 40 Hz to 100
kHz, Response speed: 200 ms or less (400 V AC), Accuracy: ±1% of full scale (DC,
40 Hz to 1 kHz), ±4% of full scale (1 kHz to 100 kHz) (full scale: 1000 V AC)
Input
Input type: balanced differential input, Input impedance/capacitance: H-L 9 MΩ/
10 pF, H/L-unit 4.5 MΩ/20 pF,
Max. rated voltage to ground: when using grabber clip 1500 V AC/DC (CAT II),
600 V AC/DC (CAT III), when using alligator clip: 1000 V AC/DC (CAT II), 600 V
AC/DC (CAT III)
Maximum input voltage
2000 V DC, 1000 V AC (CAT II), 600 V AC/DC (CAT III)
Output
Voltage divider for 1/1000 of input, BNC connectors (output switchable for 3 modes
DC, AC, RMS)
Power supply
Any of the following: (1) AC Adapte 9418-15, (2) Power Cord 9248 with Probe Power
Unit 9687, (3) Power Cord 9324 + Conversion Cable 9323 with HiCORDER
logic terminal, (4) Power Cord 9325 with F/V Unit 8940
Cable length and mass: Main unit cable 1.5 m (4.92 ft), input section
cable 30 cm (0.98 ft), approx. 150 g (5.3 oz)
Note: The unit-side plug of the 9320-01 and 9327 is different from the 9320.
LOGIC PROBE 9320 -01/9327
Functions
Detection of voltage signal or relay contact signal for High/Low state recording
Input
4 channels (common ground between unit and channels), digital/contact input,
switchable (contact input can detect open-collector signals)
Input resistance: 1 MΩ (with digital input, 0 to +5 V)
500 kΩ or more (with digital input, +5 to +50 V)
Pull-up resistance: 2 kΩ (contact input: internally pulled up to +5 V)
Digital input threshold
1.4 V/ 2.5 V/ 4.0 V
Contact input
detection resistance
1.4 V: 1.5 kΩ or higher (open) and 500 Ω or lower (short)
2.5 V: 3.5 kΩ or higher (open) and 1.5 kΩ or lower (short)
4.0 V: 25 kΩ or higher (open) and 8 kΩ or lower (short)
Response speed
9320-01: 500 ns or lower, 9327: detectable pulse width 100 ns or higher
Maximum input voltage
0 to +50 V DC (the maximum voltage that can be applied across input pins without
damage)
Note: Cannot use with 8847 or MR8847
NEW
Dimensions and mass: approx. 106 mm (4.17 in) W × 19.8 mm
(0.78 in) H × 196.5 mm (7.74 in) D, approx. 230 g (8.1 oz)
Accessories: None
PULSE GENERATOR UNIT MR8791
Output terminal
Output mode 1
Output mode 2
Other
(Accuracy at 23 ±5°C/73 ±9°F, 80% rh or less with no condensation;
Accuracy guaranteed for 1 year)
Number of channels: 8, Connector: D-sub, half-pitch, 50-pin
Max. rated voltage to ground: 33 V rms AC or 70 V DC (between unit and output channels)
Logic output/Open collector output
Pattern output: Read frequency: 0 Hz to 120 kHz, 2048 logic patterns
Pulse output: Frequency 0 Hz to 20 kHz, Duty 0.1% to 99.9%
Logic output: Output voltage level: 0 V to 5 V
(H level: 3.8 V or more, L level: 0.8 V or less)
Open collector output: Absolute maximum rated voltage for collector/emitter: 50 V
Overcurrent protection: 100 mA
Self-test function
Note: Cannot use with 8847 or MR8847
Cable length and mass: Main unit cable 1.5 m (4.92 ft), input section
cable 1 m (3.28 ft), approx. 320 g (11.3 oz)
Note: The unit-side plug of the MR9321-01 is different from
the MR9321.
LOGIC PROBE MR9321-01
Detection of AC or DC relay drive signal for High/Low state recording
Can also be used for power line interruption detection
4 channels (isolated between unit and channels), HIGH/LOW range switching
Input
Input resistance: 100 kΩ or higher (HIGH range), 30 kΩ or higher (LOW range)
170 to 250 V AC, ±DC 70 to 250 V (HIGH range)
Output (H) detection
60 to 150 V AC, ±DC 20 to 150 V (LOW range)
0 to 30 V AC, ±DC 0 to 43 V (HIGH range)
Output (L) detection
0 to 10 V AC, ±DC 0 to 15 V (LOW range)
Rising edge 1 ms max., falling edge 3 ms max. (with HIGH range at 200 V DC, LOW
Response time
range at 100 V DC)
250 V rms (HIGH range), 150 V rms (LOW range) (the maximum voltage that can be
Maximum input voltage
applied across input pins without damage)
Functions
18
System Chart of Options
Model : MEMORY HiCORDER MR8847A
Model Name
(Order Code)
MR8847-51
MR8847-52
MR8847-53
Input modules
(Note)
* Input cords not included. Please purchase them separately.
* When using 9709 with Current Unit 8971, a total of 7 current probes can be used.
ANALOG UNIT 8966
2 ch, Voltage input, DC to 5 MHz bandwidth
(64MW memory, main unit only)
(256MW memory, main unit only)
TEMP UNIT 8967
(512MW memory, main unit only)
*Cannot operate alone, You must install other options
2 ch, thermocouple temperature input
NEW
HIGH RESOLUTION UNIT 8968
2 ch, voltage input, DC to 100 kHz bandwidth
STRAIN UNIT 8969
2 ch, strain gauge type converter amp
Conversion Cable 9769
(For and bundled with the strain unit)
FREQ UNIT 8970
2 ch, for measurement of frequency, RPM, pulse, etc.
CURRENT UNIT 8971
2 ch, for measuring current using dedicated current sensors,
bundled two Conversion cable 9318
* Max. up to four modules of the Current unit 8971
Note: Main unit MR8847A cannot operate alone.
You must install one or more optional input modules in the unit.
Power supply unit
DC/RMS UNIT 8972
* Manufacturing option that must be specified upon
order.
2 ch, voltage/DC to 400 kHz, RMS rectifier, DC and 30 to 100
kHz bandwidth
DC POWER UNIT 9784
Dimensions: 290 mm (11.42 in) W × 29 mm (1.14
in) H × 219.5 mm (8.64 in) D
Mass: 1.2 kg (42.3 oz)
Specified upon factory shipping, built-in on
back of unit, 10 V to 28 V DC
Storage Devices
and Media
LOGIC UNIT 8973
4 terminals, 16 ch
Max. 3 modules can be installed in the MR8847A
NEW
DIGITAL VOLTMETER UNIT MR8990
NEW
HIGH-VOLTAGE UNIT U8974
* SSD is a built-in option that must be specified upon
order. * The CF card includes a PC card adapter.
NEW
SSD UNIT U8331
Specified upon order; built-in type, 128 GB
2ch, high-precision DC V, 0.1 μV resolution, maximum sampling
rate 500 times/s
2ch, voltage input, max. 1000 V DC and 700 V AC
* CF Card Precaution
Use only CF Cards sold by HIOKI. Compatibility and performance are not
guaranteed for CF cards made by other manufacturers. You may be unable to read
from or save data to such cards.
PC CARD 2G 9830
Output modules
WAVEFORM GENERATOR UNIT MR8790
NEW
PULSE GENERATOR UNIT MR8791
NEW
ARBITRARY WAVEFORM GENERATOR UNIT U8793
(2 GB)
PC CARD 1G 9729
(1 GB)
* Input cords not included. Please purchase separately.
NEW
PC CARD 512M 9728
4ch, DC Output: ±10 V,
Sine wave output: 10 mHz to 20 kHz
8ch, Pulse output: 0.1 Hz to 20 kHz, Pattern output
(512 MB)
PC Software
2ch, 10 mHz to 100 kHz function generator, arbitrary waveform
generator with 2 MHz D/A refresh rate, -10 V to 15 V output
WAVE PROCESSOR 9335
Convert data, print and display waveforms
LAN COMMUNICATOR 9333
• Waveform data collect function
• Remote control with the PC
iPad App for MEMORY HiCORDER HMR
Terminal
Download from the App Store (exclusively for
Apple iPad)
LAN CABLE 9642
Output cable
* Please contact your local HIOKI distributor for
connectors that support Model MR8791.
NEW
CONNECTION CABLE L9795-01
NEW
CONNECTION CABLE L9795-02
Maximum rated voltage to ground: 33 V AC
rms or 70 V DC, SMB terminal - alligator
clip, Cord length: 1.5 m (4.92 ft)
Maximum rated voltage to ground: 33 V
AC rms or 70 V DC, SMB terminal - BNC
terminal, Cord length: 1.5 m (4.92 ft)
Straight Ethernet cable, supplied with straight
to cross conversion cable, 5 m (16.41 ft) length
Logic signal measurement
Printer options
RECORDING PAPER 9231
A4 width 216 mm (8.50 in) × 30 m (98.43 ft),
6 rolls/set
Case
LOGIC PROBE 9327
4-channel type, for voltage/contact signal
ON/OFF detection (response pulse width
100 ns or more, miniature terminal type)
LOGIC PROBE MR9321-01
4 isolated channels, ON/OFF detection of
AC/DC voltage (miniature terminal type)
LOGIC PROBE 9320-01
4-channel type, for voltage/contact signal
ON/OFF detection (response pulse width
500 ns or more, miniature terminal type)
CARRYING CASE 9783
Hard trunk type to protect unit during transport
CONVERSION CABLE 9323
* Used for connecting the 9320/9321/MR9321 and
the 9324 to the Memory HiCorder with small logic
terminal models * This cable is not required for the
small-terminal types 9327, 9320-01, 9321-01 and
MR9321-01.
19
MEMORY HiCORDER MR8847A
* You can connect up to 4 Current Unit 8971 to the Memory HiCorder main unit, allowing up to 8 current sensors to be used.
* There is no limit if you connect a current sensor to the voltage input analog unit.
INPUT CORD (A)
* Voltage is limited to the specifications of the
input modules in use
CONNECTION CORD L9790
Flexible φ 4.1 mm (0.16 in) thin dia. cable allowing
for up to 600 V input, 1.8 m (5.91 ft) length * The
end clip is sold separately.
ALLIGATOR CLIP L9790-01
Red/black set attaches to the ends of the cables
L9790
GRABBER CLIP 9790-02
Red/black set attaches to the ends of the cables
L9790 * When this clip is attached to the end of
the L9790, input is limited to CAT II 300 V. Red/
black set.
CONTACT PIN 9790-03
Red/black set attaches to the ends of the cables
L9790
INPUT CORD (B)
* Voltage is limited to the specifications of the
input modules in use
CONNECTION CORD L9198
φ 5.0 mm (0.20 in) dia., cable allowing for up to
300 V input, 1.7 m (5.58 ft) length, small alligator
clip
CONNECTION CORD L9197
φ 5.0 mm (0.20 in) dia., cable allowing for up to
600 V input, 1.8 m (5.91 ft) length, detachable
large alligator clips are bundled
GRABBER CLIP 9243
Attaches to the tip of the banana plug cable, CAT
III 1000 V, 196 mm (7.72 in) length
INPUT CORD (C)
* Voltage is limited to the specifications of the
input modules in use
10:1 PROBE 9665
Note: This probe does not expand the maximum rated
voltage above ground of an isolated input. Max. rated
voltage to ground is same as for input module, max. input
voltage 1 kV rms (up to 500 kHz), 1.5 m (4.92 ft) length
Up to 200 A (High precision)
High-Precision pull-through current sensors, observe waveforms
from DC to distorted AC
AC/DC CURRENT SENSOR CT6862, 1 MHz band width, 50 A
AC/DC CURRENT SENSOR CT6863, 500 kHz band width,
200 A
Observe waveforms from DC to distorted AC
AC/DC CURRENT PROBE CT6841, 1 MHz band width, 20 A
AC/DC CURRENT PROBE CT6843, 500 kHz band width, 200 A
Observe waveforms of distorted AC (cannot for DC)
CLAMP ON SENSOR 9272-10, 100 kHz band width, 200 A
Up to 1000 A (High precision)
High-Precision pull-through current sensors, observe waveforms
from DC to distorted AC
AC/DC CURRENT SENSOR 9709, 100 kHz band width, 500 A
Observe waveforms from DC to distorted AC
AC/DC CURRENT PROBE CT6844, 200 kHz band width,
500 A
AC/DC CURRENT PROBE CT6845, 100 kHz band width,
500 A
AC/DC CURRENT PROBE CT6846, 20 kHz band width,
1000 A
DC to 100 MHz wide band response, mA-class current up to 30
Arms
CLAMP ON PROBE 3274
DC to 10 MHz wideband response, mA-class current up to 150
Arms
CLAMP ON PROBE 3275
DC to 2 MHz wideband response, mA-class current up to 500
Arms
Power supply * Necessary for use high speed current probes
POWER SUPPLY 3272
For the 3270 series, single sensor connectable
(2 units possible depending on conditions)
POWER SUPPLY 3269
For the 3270 series, connect up to four sensors
100 A to 2000 A (Medium speed) A
AC/DC CURRENT SENSOR CT7631, (Auto zero
CT7731)
DC, 1 Hz to 10 kHz (-3dB), 100 A, 1 mV/A output
* For measuring high voltage. *Voltage to ground is within this
product’s specifications, and is not affected by connected
input modules. Separate power source is also required.
AC/DC CURRENT SENSOR CT7636, (Auto zero
CT7736)
(Wave Only) For Memory HiCorder,
1 kV AC, DC, Frequency band: 100 kHz
AC/DC CURRENT SENSOR CT7642, (Auto zero
CT7742)
DIFFERENTIAL PROBE P9000-02
(Switch between Wave/RMS) For Memory
HiCorder, 1 kV AC, DC, Frequency band: 100
kHz
AC ADAPTER Z1008
100 to 240 V AC
DC, 1 Hz to 10 kHz (-3dB), 600 A, 1 mV/A output
Custom cable
*For P9000. Inquire with your
local Hioki distributor.
(1) Bus powered USB cable
(2) USB(A)- Micro B cable
(3) 3-prong cable
DC to 15 kHz (-3dB), 2000 A, Output 0.2 V/f.s. Set model with the CT6590
500 A to 5000 A *For commercial power lines, 50/60 Hz
CLAMP ON PROBE 9018-50
Other options for Input
CLAMP ON PROBE 9132-50
CONVERSION ADAPTER 9199
10 Hz to 20 kHz, 5000 A/ 500 A AC, 500 mV/f.s. output, φ 100 to 254 mm
(3.94 to 10.00 in), 3 loop diameters
CONNECTION CABLE L4940
EXTENSION CABLE L4931
Receiving side banana terminal,
output BNC terminal
ALLIGATOR CLIP L4935
Metallic BNC at both ends, for
metallic BNC terminals, 1.5 m (4.92
ft) , not CE marked
BUS BAR CLIP L4936
Attach to the tip of banana plug cables, CAT III 600 V
CONNECTION CORD 9165
CONVERSIONCABLE 9318
For connecting CT6841/43 and
similar probes to 8971/40/51.
LEAK CURRENT *For commercial power lines, 50/60 Hz
CLAMP ON LEAK HiTESTER 3283
10 mA range/10 μA resolution to 200 A range, with monitor/analog
output 1 V f.s.
OUTPUT CORD 9094
3.5 mm (0.14 in) dia. mini plug to banana, 1.5 m (4.92 ft) length
AC ADAPTER 9445-02
For USA, 100 to 240 V AC, 9 V/ 1 A
GRABBER CLIP 9243
AC ADAPTER 9445-03
Attach to the tip of banana plug cables, red/black set,
full length: 196mm (7.72 in), CAT III 1000 V
Cable length: 70 cm, tips interchangeable with
a pin test lead or alligator clip, maximum input
voltage: CAT IV 600 V, CAT III 1000 V
AC FLEXIBLE CURRENT SENSOR CT9667-01/-02/-03
Receiving side banana, output BNC terminal
Attach to the tip of banana plug cables,CAT III 1000 V
TEST LEAD L2200
Frequency characteristics: 40 Hz to 1 kHz, 20 to 1000 A AC range, output
0.2 V AC f.s.
CONVERSION ADAPTER 9199
MAGNETIC ADAPTER L4937
* For the MR8990 *Voltage is limited to the
specifications of the input modules in use
Good phase characteristics, Frequency characteristics: 40 Hz to 3 kHz, 10
to 500 A AC range, output 0.2 V AC f.s.
CONNECTION CORD L9217
Cord has insulated BNC connectors
at both ends, signal output use, 1.6 m
(5.25 ft) length
Attach to the tip of banana plug cables, CAT IV 600 V,
CAT III 1000 V
100 A to 2000 A (Medium speed) B
CLAMP ON AC/DC SENSOR CT9691-90
CLAMP ON AC/DC SENSOR CT9693-90
100 to 240 V AC
Extend the length of banana plug cables, Cable length:
1.5 m (4.92 ft)
Power supply for the CT7600/7700 series single drive, Measure,
Display, Signal output function
DC to 20 kHz (-3dB), 200 A, Output 0.2 V/f.s. Set model with the CT6590
AC ADAPTER 9418-15
Banana plug - banana plug, Cord length: 1.5 m (4.92 ft)
DISPLAY UNIT CM7290
CLAMP ON AC/DC SENSOR CT9692-90
1 kV AC, 2 kV DC, Frequency band: 10 MHz
* Voltage input via banana terminals limited by the
voltage specifications of the respective input unit.
DC, 1 Hz to 10 kHz (-3dB), 2000 A, 1 mV/A output
DC to 10 kHz (-3dB), 100 A, Output 0.1 V/f.s. Set model with the CT6590
* For measuring high voltage. *Voltage to ground is within this
product’s specifications, and is not affected by connected
input modules. Separate power source is also required.
DIFFERENTIAL PROBE 9322
INPUT CORD (G)
Cord has insulated BNC connectors at both ends, signal output
use, 1.6 m (5.25 ft) length
CLAMP ON PROBE 3276
DIFFERENTIAL PROBE P9000-01
INPUT CORD (F)
CONNECTION CORD L9217
DC to 50 MHz wide band response, mA-class current up to 30
Arms
Note: This probe does not expand the maximum rated
voltage above ground of an isolated input. Max. rated
voltage to ground is same as for input module, max. input
voltage 5 kV peak (up to 1 MHz), 1.5 m (4.92 ft) length
INPUT CORD (E)
Power supply for the Current Sensor, used alone
10 mA class to 500 A (High speed)
CLAMP ON PROBE 3273-50
100:1 PROBE 9666
INPUT CORD (D)
Power supply * Not necessary when using Current Unit 8971.
SENSOR UNIT 9555-10
For EU 100 to 240 V AC, 9 V/ 1 A
Temperature sensor
THERMOCOUPLE
*For reference only. Please purchase
locally.
Example sets
8 ch
waveform
output
8 ch
voltage
measurement
Anomaly simulation testing
Output measured anomalous waveforms and processed arbitrary waveforms at max. 15 V, and record the results
without modification.
Set example
MEMORY HiCORDER
ARBITRARY WAVEFORM
GENERATOR UNIT
ANALOG UNIT
CONNECTION CABLE
CONNECTION CORD
Generate
Record
Standalone testing
Reproduce anomalous waveforms
Record results while testing
voltage
measurement
3 ch
current
measurement
1 unit
4
8966
L9795-01
L9198
4
8
8
Arbitrary waveforms generated for 8 units,
max. 16 channels
Isolated output for all channels
Output both measured anomalous waveforms
and waveforms that you created yourself for
testing. You can also measure the results at the
same time.
3 ch
MR8847-51
U8793
Channels can be expanded without connecting additional
recorders. Isolation between the Memory HiCorder and each
channel and between the channels allows connections with
devices with different potentials.
High-voltage direct input measurement
Direct input is also possible without a differential probe for high voltage of 1000 V DC and 700 V AC.
Set example
Battery
Load
Inverter
MEMORY HiCORDER
HIGH-VOLTAGE UNIT
CURRENT UNIT
CLAMP ON SENSOR
CONNECTION CABLE
ALLIGATOR CLIP
Motor
MR8847-51
U8974
8971
9272-10
L4940
L4935
No DIFFERENTIAL PROBE
needed for direct high-voltage
measurements
Perform direct measurement of up to 1000 V DC
and 700 V AC for high-voltage power equipment
as well as 380 V and 480 V systems used globally.
Multi-type
Video
multirecording channel
measurement
1 unit
2
2
3
3
3
Simultaneous
measurement
of all waveforms
Battery
Test ECUs, inverters and motors.
Simultaneous measurement with
high-speed camera recording
* Please contact your local Hioki
distributor for more information about
the use of high-speed cameras.
Synchronize high-speed video with multi-channel signals for recording.
Visualize prototype evaluations and
problem analyses together with
measurement data.
Easily visualize the relationship between various
factors through the simultaneous measurement of data
such as multi-system voltage, current and vibration
together with high-speed camera recording.
Set example
MEMORY HiCORDER
ANALOG UNIT
CONNECTION CORD
High-speed camera
MR8847-51
8966
9197
—
1 unit
1
1
1 unit
Note: Company names and Product names appearing in this catalog are trademarks or registered trademarks of various companies.
HIOKI (Shanghai) SALES & TRADING CO., LTD.
TEL +86-21-63910090 FAX +86-21-63910360
http://www.hioki.cn / E-mail: [email protected]
DISTRIBUTED BY
HIOKI INDIA PRIVATE LIMITED
TEL +91-124-6590210
HEADQUARTERS
E-mail: [email protected]
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 FAX +81-268-28-0568
HIOKI SINGAPORE PTE. LTD.
http://www.hioki.com / E-mail: [email protected] TEL +65-6634-7677 FAX +65-6634-7477
E-mail: [email protected]
HIOKI USA CORPORATION
HIOKI KOREA CO., LTD.
TEL +1-609-409-9109 FAX +1-609-409-9108
TEL +82-2-2183-8847 FAX +82-2-2183-3360
http://www.hiokiusa.com / E-mail: [email protected] E-mail: [email protected]
All information correct as of Feb. 12, 2016. All specifications are subject to change without notice.
MR8847A_E2-62M
Printed in Japan