PXIe VNA Startup Guide

PXIe VNA Startup Guide
Keysight M937xA PXIe
Vector Network Analyzer
Modules
M9370A, 300 kHz to 4 GHz
M9371A, 300 kHz to 6.5 GHz
M9372A, 300 kHz to 9 GHz
M9373A, 300 kHz to 14 GHz
M9374A, 300 kHz to 20 GHz
M9375A, 300 kHz to 26.5 GHz
Startup Guide
Notices
© Keysight Technologies,
Inc. 2014-2016
No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Keysight Technologies, Inc. as
governed by United States and
international copyright laws.
Trademark Acknowledgments
Manual Part Number
M9370-90001
Edition
Edition 2, April 2016
Printed in USA/Malaysia
Published by:
Keysight Technologies
1400 Fountaingrove Parkway
Santa Rosa, CA 95403
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES,
EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED
WARRANTIES OF
MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
KEYSIGHT SHALL NOT BE LIABLE
FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE
FURNISHING, USE, OR
PERFORMANCE OF THIS
DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH
THESE TERMS, THE WARRANTY
TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance
with the terms of such license.
U.S. Government Rights
The Software is “commercial
computer software,” as defined
by Federal Acquisition Regulation
(“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and
Department of Defense FAR
Supplement (“DFARS”) 227.7202,
the U.S. government acquires
commercial computer software
under the same terms by which
the software is customarily
provided to the public.
Accordingly, Keysight provides
the Software to U.S. government
customers under its standard
commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of
which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA
represents the exclusive authority
by which the U.S. government
may use, modify, distribute, or
disclose the Software. The EULA
and the license set forth therein,
does not require or permit,
among other things, that
Keysight: (1) Furnish technical
information related to
commercial computer software
or commercial computer
software documentation that is
not customarily provided to the
public; or (2) Relinquish to, or
otherwise provide, the
government rights in excess of
these rights customarily provided
to the public to use, modify,
reproduce, release, perform,
display, or disclose commercial
computer software or
commercial computer software
documentation. No additional
government requirements
beyond those set forth in the
EULA shall apply, except to the
extent that those terms, rights, or
licenses are explicitly required
from all providers of commercial
computer software pursuant to
the FAR and the DFARS and are
set forth specifically in writing
elsewhere in the EULA. Keysight
shall be under no obligation to
update, revise or otherwise
modify the Software. With
respect to any technical data as
defined by FAR 2.101, pursuant
to FAR 12.211 and 27.404.2 and
DFARS 227.7102, the U.S.
government acquires no greater
than Limited Rights as defined in
FAR 27.401 or DFAR 227.7103-5
(c), as applicable in any technical
data.
Safety Notices
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in damage to the
product or loss of important data. Do
not proceed beyond a CAUTION
notice until the indicated conditions
are fully understood and met.
A WARNING notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated
conditions are fully understood and
met.
Contents
Table of Contents
1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Getting Started
Introduction
Related Documentation
STEP 1. Unpack and Inspect the Network Analyzer Modules
5
5
5
Inspect for Damage
6
How to Return an Instrument for Service
7
STEP 2. Check the Shipment
7
STEP 3. Install the Software
8
System Requirements
8
Hardware Requirements
8
Power Up the Controller
8
Software Installation Overview
9
STEP 4. Install the Network Analyzer Modules
11
Recommended Best Practices to Ensure Proper and Safe Module Operating Conditions
11
Tools Required for the Installation Procedure
12
Installation Procedure
12
Front Panel Features
19
STEP 5. Verify Operation of the Network Analyzer Modules
2
3
20
The Operator’s Check
20
If the Operator’s Check Fails
21
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Getting Help with Your Network Analyzer Module
Help System
23
N937xA on the Internet
23
Contacting Keysight
24
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety and Regulatory Information
Maintenance
25
Shipment for Service
25
Safety Symbols
26
General Safety Considerations
27
Before Applying Power
27
Servicing
27
Operating Conditions
27
Keysight M937xA Startup Guide
3
Contents
Regulatory Information
28
Instructions for Use
28
EMC
28
Safety
28
Instrument Markings
29
Declaration of Conformity
29
Warranty
4
30
Keysight M937xA Startup Guide
Keysight PXIe Vector Network Analyzers
M937xA
Startup Guide
1
Getting Started
Introduction
The scope of this Startup Guide is to detail the processes of receiving and
installing the network analyzer module and cables that compose the Keysight
M937xA PXIe Vector Network Analyzer module. Additionally, installing the
required software and verifying network analyzer module operation is
documented. If you have any questions after reviewing this information, please
contact your local Keysight Technologies Inc. representative or contact us
through our support Website at
http://www.keysight.com/techicalsupport/main.
Related Documentation
You can find all M937xA PXI network analyzer module documentation and
other resources at: http://na.support.keysight.com/pxi/help
or at the respective product pages on www.keysight.com (go to Document
Library > Manuals).
STEP 1. Unpack and Inspect the Network Analyzer Modules
The network analyzer modules are shipped in materials which prevent
damage from static. The modules should only be removed from the
packaging in an anti-static area ensuring that correct anti-static
precautions are taken. Store all modules in anti-static envelopes when not
in use.
Electrostatic discharge (ESD) can damage or destroy electronic components.
Use a static-safe work station to perform all work on electronic assemblies.
Figure 1-1 on page 6 shows a static-safe work station using two types of ESD
protection: conductive table-mat and wrist-strap combination, and conductive
floor-mat and heel-strap combination. Both types, when used together,
provide a significant level of ESD protection. Of the two, only the table-mat and
5
Getting Started
STEP 1. Unpack and Inspect the Network Analyzer Modules
wrist-strap combination provides adequate ESD protection when used alone.
To ensure user safety, the static-safe accessories must provide at least 1
meg-ohm of isolation from ground.
Do NOT use these techniques for a static-safe work station when
working on circuitry with a voltage potential greater than 500 volts.
Figure 1-1
ESD protection setup
Unpack the network analyzer modules from the shipping containers. Keep the
black and red plastic port protectors and the shipping containers for possible
reuse.
Inspect for Damage
After unpacking a network analyzer module, inspect it for any shipping
damage. Report any damage to the shipping agent immediately, as such
damage is not covered by the warranty.
To avoid damage when handling the network analyzer modules, do not
touch exposed connector pins.
6
Keysight Technologies Startup Guide
Getting Started
STEP 2. Check the Shipment
How to Return an Instrument for Service
Should it become necessary to return a network analyzer module for repair or
service, follow the steps below:
1. Review the warranty information shipped with your product.
2. Contact Keysight to obtain a Return Material Authorization (RMA) and
return address. For assistance finding Keysight contact information, go to
www.keysight.com/find/assist (worldwide contact information for repair
and service).
3. Write the following information on a tag and attach it to the network
analyzer module:
— Name and address of owner. A P.O. box is not acceptable as a return
address.
— Network analyzer module serial numbers. The serial number label is
located on the side panel of the module. The serial number can also
be read from the Soft Front Panel (SFP) interface, after the hardware
is installed.
— Description of the service required or the failure.
4. On the shipping label, write ATTENTION REPAIR DEPARTMENT and the
RMA number.
5. Ship the analyzer module using the original packaging materials. Shipping
the analyzer module in anything other than the original packaging may
result in non-warranted damage.
In your correspondence, refer to the network analyzer module by serial
number and by model number.
STEP 2. Check the Shipment
Use the Contents List in the shipping container to verify the completeness of
your shipment. If not complete, refer to “Contacting Keysight” on page 24.
Keysight Technologies Startup Guide
7
Getting Started
STEP 3. Install the Software
STEP 3. Install the Software
System Requirements
Topic
Windows 7 and Vista Requirements
Operating system
Windows 7 (32 bit and 64 bit)
Windows Vista, SP 1 and SP2 (32-bit and 64-bit)
Processor speed
1.5 GHz dual core (x86 or x64) minimum
(2.4 GHz recommended)
Available memory
1 GB minimum (8 GB recommended)
Available disk space
1.5 GB available hard disk space
Hardware Requirements
Topic
Requirements
Chassis
PXIe or PXI-H chassis slot
Controllers
PXI or PXI Express embedded controller or remote controller
(external PC connected to the chassis by a PCI-to-PXI
interface) is required.
Embedded controller
Keysight M903xA or an equivalent embedded controller that
meets the following requirements:
— A PXIe system controller (PXI-1 embedded
controllers are not compatible).
— Utilizes a 2x8, 4x4, or 1x8 PXIe system slot link
configuration.
— Runs one of the operating systems listed in
System Requirements (above).
Remote controller
(For Keysight M9018A chassis use only) A PC running one of
the operating systems listed in System Requirements
and a Keysight M9021A Cable Interface x8 with one of the
following PC interface options:
— Keysight M9045B PCIe ExpressCard Adaptor x1,
with cable (for a laptop PC)
— Keysight M9048B PCIe Desktop Adaptor x8, with
cable (for a desktop PC)
Power Up the Controller
8
Keysight Technologies Startup Guide
Getting Started
STEP 3. Install the Software
If you are using a remote controller, complete the following steps:
1. Install the cable interface between the remote controller and the chassis.
If you are using an Keysight M9021A cable interface, refer to the Keysight
M9021A documentation for further details.
2. Power up the chassis.
3. Power up the PC.
If you are using a remote controller and you have installed the interface
cable, you must power up the chassis BEFORE you power up the PC.
When you power down your system, shut down the PC BEFORE you power
down the chassis.
If you are using an embedded controller, complete the following steps:
1. Install the embedded controller module into a compatible chassis. The
Keysight M9018A PXIe chassis is recommended. Refer to the Keysight
M9018A documentation for further details.
2. Connect peripherals (mouse, keyboard, monitor).
3. Power up the chassis.
Software Installation Overview
The software installation includes the following items:
— Keysight IO Libraries Suite (IOLS), which includes the Keysight Connection
Expert. Download the latest version at www.keysight.com/find/iosuite. If
you are unable to access the Web, you can purchase Option 2CD which
includes an IO Library Suite CD. This software must be installed first.
Version 16.3.6603.3 (or newer) of the Keysight IO Libraries Suite is
required.
— Network analyzer module software, which includes the soft front panel
(SFP) software, device drivers (IVI-C and IVI-COM, and LabVIEW G), and
documentation for the M937xA PXIe Vector Network Analyzer modules.
Download the latest version at
http://www.keysight.com/techicalsupport/main. Enter the model for your
instrument and click Find. If you are unable to access the Web, you can
purchase Option 2CD which includes a software CD.
Keysight Technologies Startup Guide
9
Getting Started
STEP 3. Install the Software
Software Installation Procedure
1. Install the Keysight IO Libraries Suite by downloading it from
www.keysight.com/find/iosuite (or the IO Libraries Suite CD, if you
purchased Option 2CD). Follow the installer prompts to install the IO
libraries.
2. Install the network analyzer module software by downloading it from
http://www.keysight.com/techicalsupport/main (or the software CD, if
you purchased Option 2CD). Enter the model for your instrument and click
Find. Follow the installer prompts.
3. Complete the installation:
If Using an Embedded Controller:
a. After the InstallShield Wizard has completed, click Finish.
b. Restart the embedded controller PC, using Start > Restart,
and wait for the system to restart.
If Using a Remote Controller:
a. After the InstallShield Wizard has completed, click Finish.
b. Shut down the remote controller PC, using Start > Shut
down.
c. Power down the chassis.
d. Power up the chassis.
e. Power up the remote controller PC.
Check for software updates at:
http://www.keysight.com/techicalsupport/main. Enter the model for your
instrument and click Find. Follow the installer prompts.
10
Keysight Technologies Startup Guide
Getting Started
STEP 4. Install the Network Analyzer Modules
STEP 4. Install the Network Analyzer Modules
Please review all safety information located in Chapter 3 before
proceeding with installing and using the network analyzer module.
These modules can be used in a chassis with a PXIe, or PXI-H chassis slots.
Recommended Best Practices to Ensure Proper and Safe Module
Operating Conditions
— Ensure proper chassis air flow is maintained.
— The use of a Keysight M9018A chassis is recommended for optimum
temperature control. If a different chassis is used, make sure it provides
thermal protection if fans become inoperable or forced air cooling is
obstructed.
— If using the Keysight M9018A PCIe Cable Interface, download the
Keysight Multiple PXIe and AXIe Chassis System Configuration Tool.
Save the file to your local hard drive; do not try to open the file from
the Internet location.
— Use slot blockers and EMC filler panels in empty module slots to ensure
proper operating temperatures. Keysight chassis and slot blockers optimize
module temperature performance and reliability of test.
— Set chassis fans to high or auto. Do not disable fans.
— Position chassis to allow plenty of space around chassis air intake and fan
exhaust.
— At ambient temperatures above 45° C (113° F) set the chassis fan to high.
Figure 1-2
M937xA chassis air flow
Keysight Technologies Startup Guide
11
Getting Started
STEP 4. Install the Network Analyzer Modules
Tools Required for the Installation Procedure
— Phillips #1 torque driver, set to 4 in-lbs (0.45 N.m) - not supplied.
— Socket adapter (part number 5023-1450) - supplied.
— Torque driver (for use with socket adapter), set to 8 in-lb (0.91 N.m) - not
supplied.
— Cable removal tool (part number 5002-3361) - supplied.
Installation Procedure
PXI hardware does not support “hot-swap” (changing network analyzer
modules while power is applied to the chassis) capabilities. Before
installing or removing a module to/from the chassis, power-off the chassis
to prevent damage to the module.
The network analyzer module can be installed in any PXIe or hybrid PXI
slot marked with a peripheral slot compatibility image (solid black
circle fro PXIe, or solid black circle with the letter “H” for hybrid.
1. Make sure that the line cord is plugged in to establish earth ground and
that the chassis power switch is Off (see callouts 1 and 3 in Figure 1-3 on
page 13).
2. Wear a grounded wrist strap for this procedure (see callout 2 in Figure 1-3
on page 13).
3. If the chassis has multiple fan speed settings, ensure that the fan switch is
set to AUTO.
4. Position the chassis to provide ample space between the chassis fan
intake and exhaust vents. Blockage by walls or obstructions affects the air
flow needed for cooling. (Refer to the chassis documentation for more
information about cooling.)
5. Before inserting the network analyzer module into the chassis, back-out
the top mounting screw (see callout 7 in Figure 1-3 on page 13) on the
module to ensure that there is no interference between the screws and the
mounting rails.
6. Holding the module by the injector/ejector handle, slide it into a slot, as
shown in callouts 4 – 6 in Figure 1-3 on page 13.
a. Install the module into the slot of the chassis by placing the module
card edges into the front module guides (top and bottom).
12
Keysight Technologies Startup Guide
Getting Started
STEP 4. Install the Network Analyzer Modules
b. Hold down the silver locking tab while pushing down the
injector/ejector handle to the unlatched (downward) position, and
slide the module to the rear of the chassis.
c. Slide the module completely into the chassis. When you begin to
feel resistance, pull up on the injector/ejector handle to fully inject
the module into the chassis. The injector/ejector handle should lock
into place. (To remove the module from the chassis, you must first
press the injector/ejector handle to unlock it.)
Figure 1-3 Installing the Network Analyzer Module
7. Repeat for other modules, if present.
8. Tighten the top mounting screw (see callout 7 in Figure 1-3) and the
bottom mounting screw (located below the injector/ejector handle) on the
module(s) using a Phillips #1 torque drive set to 4 in-lbs (0.45 N.m).
9. Install EMC filler panels and slot blockers into empty module slots to
assure proper air circulation and operating temperatures.
10.If using only one M937xA module in the chassis, leave the 50 ohm loads
attached as shown in Figure 1-4 on page 14.
Keysight Technologies Startup Guide
13
Getting Started
STEP 4. Install the Network Analyzer Modules
Figure 1-4
Single M937xA Network Analyzer Module Configuration
In the following instructions, eight network analyzer modules are
configured. Use these instructions as a general guide if installing a
different number of modules.
11.As seen in Figure 1-5 on page 15, each module is shipped with two 50
ohm loads attached to each module. To enable connections between
module pairs, some of these loads must be removed. To remove the loads,
use the socket adapter, seen in Figure 1-6 on page 15, attached to the 8
in-lb (0.91 N.m) torque driver.
a. On the module in slot 1, remove the 50 ohm load from the LO Out
port. Leave the load installed on the LO In port.
b. On the module in slot 8, remove the 50 ohm load from the LO In
port. Leave the load installed on the LO Out port.
c. On each of the remaining modules, remove both 50 ohm loads.
d. Save the 50 ohm loads that were just removed for possible reuse in
another module configuration.
14
Keysight Technologies Startup Guide
Getting Started
STEP 4. Install the Network Analyzer Modules
Figure 1-5 50 Ohm Loads Installed on the Network Analyzer Modules
Figure 1-6 Socket Adapter for 50 Ohm Loads (part number 5023-1450)
12.Install a semi-rigid cable (M9370-20018), seen in “Installing Semi-Rigid
Cable M9370-20018” on page 16, between the Trig Out port and the Trig
In port on each of the module pairs (1 & 2, 2 & 3, 3 & 4, 4 & 5, 5 & 6, 6 & 7,
and 7 & 8).
Keysight Technologies Startup Guide
15
Getting Started
STEP 4. Install the Network Analyzer Modules
Figure 1-7
Installing Semi-Rigid Cable M9370-20018
Use the supplied cable removal tool (5002-3361) to remove the
M9370-20018 cables if necessary. See Figure 1-8 on page 16.
Figure 1-8
16
Cable Removal Tool (part number 5002-3361)
Keysight Technologies Startup Guide
Getting Started
STEP 4. Install the Network Analyzer Modules
13.Install a semi-rigid cable (M9370-20006), seen in Figure 1-9, between the
LO Out port and the LO In port on each of the module pairs (1 & 2, 2 & 3, 3
& 4, 4 & 5, 5 & 6, 6 & 7, and 7 & 8), hand tightening the connectors.
Figure 1-9
Installing Semi-Rigid Cable M9370-20006
14.Tighten the semi-rigid cable using the socket adapter, seen in Figure 1-6
on page 15, attached to an 8 in-lb (0.91 N.m) torque driver.
15.Install a semi-rigid cable (M9370-20019), seen in Figure 1-10, between
the Ref Out port and the Ref In port on each of the module pairs (1 & 2, 2 &
3, 3 & 4, 4 & 5, 5 & 6, 6 & 7, and 7 & 8).
Keysight Technologies Startup Guide
17
Getting Started
STEP 4. Install the Network Analyzer Modules
Figure 1-10
Installing Semi-Rigid Cable M9370-20019
Use the supplied cable removal tool (5002-3361) to remove the
M9370-20019 cables if necessary. See Figure 1-8 on page 16.
You have now completed the hardware installation procedure.
18
Keysight Technologies Startup Guide
Getting Started
STEP 4. Install the Network Analyzer Modules
Front Panel Features
Front Panel Photo
Feature
Description
Status LED
OFF: Power supply off. Safe to remove the network analyzer module from
the chassis.
AMBER: Power supply on, but no firmware is in control.
GREEN: Firmware is in control; instrument is probably idle.
FLASHING GREEN: A sweep or measurement has been triggered.
RED: Last sweep had an error.
Port 1
APC 3.5 mm female connector. Sources stimulus signals for the DUT and
receives response signals from the DUT. Frequency range is determined
by the network analyzer module model.
Port 2
LO In
SMA female connector. If the network analyzer module is not configured
with another module, the connector is terminated with a 50 ohm load. If
the module is configured with a module in an adjacent slot, the connector
receives the LO signal from the configured module.
LO Out
SMA female connector. If the network analyzer module is not configured
with another module, the connector is terminated with a 50 ohm load. If
the analyzer module is configured with a module in an adjacent slot, the
connector provides the LO signal to the configured module.
Ref In
SMB female connector. A 10 MHz external reference signal supplied to
this port can be used as the instrument frequency reference instead of the
internal frequency reference. To activate this feature on the soft front
panel, select Utility > System > Configure > External Reference.
Ref Out
SMB female connector. Outputs a 10 MHz frequency reference signal for
use by other test equipment.
Trig In
SMB female connector. Inputs a signal to trigger the network analyzer
module.
Trig Out
SMB female connector. Outputs a signal either before or after a
measurement.
Trig Ready
SMB female connector. Outputs a “READY’ signal to other devices.
Keysight Technologies Startup Guide
19
Getting Started
STEP 5. Verify Operation of the Network Analyzer Modules
STEP 5. Verify Operation of the Network Analyzer Modules
The Operator’s Check
Allow the network analyzer modules to warm up for at least 15 minutes
before performing the Operator’s Check.
The operator’s check is a software driven test that checks the basic operation
of the assemblies in all of the measurement port signal paths. By performing
the operator’s check, the following are determined:
— attenuation ranges of all installed attenuators
— calibration of the receivers
— frequency response of the receivers
— phase lock
— noise floor and trace noise
— various voltages throughout the network analyzer module
Accessories Used in the Operator’s Check
Male short(s) or open(s) from a 3.5 mm calibration kit.
Performing the Operator’s Check
Perform the Operator’s Check on each network analyzer module in your
configuration.
1. On your PC screen, click the Network Analyzer icon.
2. Select a network analyzer module in your configuration.
3. On the soft front panel, select Utility > System > Service > Operator’s
Check.
4. In the PXI Operator’s Check dialog box (refer to Figure 1-11 on page 21),
under Configure, select either Prompt for attachment of Short/Open, to
pause at each step in the process to allow moving the short/open to the
appropriate port, or Shorts/Opens are attached to ALL ports, to run
through the test without stopping. Shorts and opens can be mixed on the
test ports.
5. Select Begin.
6. If shorts and opens are not connected to all ports, you will be prompted to
connect them as they are needed.
7. The result of the operator’s check will be shown as a PASS or FAIL next to
each test (refer to Figure 1-11 on page 21.
20
Keysight Technologies Startup Guide
Getting Started
STEP 5. Verify Operation of the Network Analyzer Modules
Figure 1-11
Operator’s Check Dialog Box
If the Operator’s Check Fails
1. Clean the test ports, and the open(s) or short(s). Torque to specification.
Repeat the check.
2. If the check still fails, return the network analyzer module to Keysight. See
“How to Return an Instrument for Service” on page 7.
Keysight Technologies Startup Guide
21
Getting Started
STEP 5. Verify Operation of the Network Analyzer Modules
22
Keysight Technologies Startup Guide
Keysight PXIe Vector Network Analyzers
M937xA
Startup Guide
2
Getting Help with Your Network Analyzer Module
Help System
Use the Help System to quickly reference programming and user
documentation
To access Help:
— From the analyzer module software
— On your PC screen, click the Network Analyzer icon.
— After making your analyzer module configuration settings, click Help.
— Online, go to http://na.support.keysight.com/pxi/help.
— If you purchased Option 2CD, which includes a software CD (CD part
number M9370-90003), load it, and click Soft Front Panel Help.
N937xA on the Internet
The M937xA PXI Network Analyzer System Web page is another resource for
help with the analyzer module. You can find all M937xA PXI network analyzer
module documentation and other resources at:
http://na.support.keysight.com/pxi/help.
Check for software updates at:
http://www.keysight.com/techicalsupport/main. Enter the model for your
instrument and click Find
23
Getting Help with Your Network Analyzer Module
Contacting Keysight
Contacting Keysight
Assistance with test and measurements needs and information on finding a
local Keysight office are available on the Web at:
www.keysight.com/find/assist.
If you do not have access to the Internet, please contact your Keysight field
engineer.
In any correspondence or telephone conversation, refer to the Keysight
product by its model number and full serial number. With this information,
the Keysight representative can determine whether your product is still
within its warranty period..
24
Keysight Technologies Startup Guide
Keysight PXIe Vector Network Analyzers
M937xA
Startup Guide
3
Safety and Regulatory Information
Maintenance
To remove dirt or dust from the external case of the network analyzer module,
clean the case using a dry or slightly-dampened cloth only.
To prevent electrical shock, remove analyzer module from chassis
slot for cleaning. Use a dry cloth or one slightly dampened with
water to clean the external case parts. Do not attempt to clean
internally.
Shipment for Service
Contact Keysight Technologies for instructions on where to ship the analyzer
module for service. Refer to “Contacting Keysight” on page 24.
Ship the analyzer module using the original packaging materials. Shipping the
analyzer module in anything other than the original packaging may result in
non-warranted damage.
25
Safety and Regulatory Information
Safety Symbols
Safety Symbols
The following safety symbols are used throughout this manual. Familiarize
yourself with each of the symbols and its meaning before operating this
instrument.
Denotes a hazard. It calls attention to a procedure that, if not correctly
performed or adhered to, would result in damage to or destruction of the
product. Do not proceed beyond a caution note until the indicated
conditions are fully understood and met.
Denotes a hazard. It calls attention to a procedure which, if not
correctly performed or adhered to, could result in injury or loss of
life. Do not proceed beyond a warning note until the indicated
conditions are fully understood and met.
26
Keysight Technologies Startup Guide
Safety and Regulatory Information
General Safety Considerations
General Safety Considerations
Before Applying Power
If this product is not used as specified, the protection provided by
the equipment could be impaired. This product must be used in a
normal condition (in which all means for protection are intact) only.
The measuring terminals on this instrument are designed to be used with
external signals described in Measurement Category I, but NOT with
external signals described in Categories II, III, and IV. The input of this
instrument cannot be connected to the mains.
Servicing
These servicing instructions are for use by qualified personnel only.
To avoid electrical shock, do not perform any servicing unless you
are qualified to do so.
Operating Conditions
Operator is responsible for maintaining safe operating conditions. To
ensure safe operating conditions, modules should not be operated
beyond the full temperature range documented in the M937xA Data
Sheet (http://literature.cdn.keysight.com/litweb/pdf/M9370-90002.pdf ).
Not maintaining safe operating conditions can result in shorter
module lifespans, improper module performance, and user safety
issues. If modules are allowed to exceed the specified full
temperature range, module surface temperatures can cause
discomfort or burns if touched. If a module exceeds the full
temperature range, always allow it to cool before touching it or
removing it from the chassis.
Keysight Technologies Startup Guide
27
Safety and Regulatory Information
Regulatory Information
Regulatory Information
This section contains information that is required by various government
regulatory agencies.
Instructions for Use
This product has been designed and tested in accordance with accepted
industry standards, and has been supplied in a safe condition. The
documentation contains information and warnings that must be followed by
the user to ensure safe operation and to maintain the product in a safe
condition.
EMC
Complies with the essential requirements of the European EMC Directive as
well as current editions of the following standards (dates and editions are cited
in the Declaration of Conformity):
— IEC/EN 61326-1
— CISPR Pub 11 Group 1, class A
— AS/NZS CISPR 11
— ICES/NMB-001
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB-001 du Canada.
South Korean Class A EMC declaration: This equipment is Class A suitable for
professional use and is for use in electromagnetic environments outside of the
home.
A 급 기기 ( 업무용 방송통신기자재 ) 이 기기는 업무용 (A 급 ) 전자파적합기
기로서 판 매자 또는 사용자는 이 점을 주 의하시기 바라 며 , 가정외의 지역
에서 사용하는 것을 목적으 로 합니
Safety
Acoustic statement (European Machinery Directive)
— Acoustic noise emission
— LpA <70 dB
— Operator position
— Normal operation mode per ISO 7779
28
Keysight Technologies Startup Guide
Safety and Regulatory Information
Regulatory Information
Instrument Markings
Some instrument markings may not appear on your analyzer.
The instruction documentation symbol. The product is marked with this symbol when
it is necessary for the user to refer to the instructions in the documentation.
The AC symbol indicates the required nature of the line module input power.
This symbol indicates separate collection for electrical and electronic equipment,
mandated under EU law as of August 13, 2005. All electric and electronic equipment
are required to be separated from normal waste for disposal (Reference WEEE
Directive, 2002/96/EC).
This symbol is used to identify a terminal which is internally connected to the product
frame or chassis.
The CE mark is a registered trademark of the European Community. (If accompanied
by a year, it is when the design was proven.)
The CSA mark is a registered trademark of the CSA International.
This mark designates the product is an Industrial Scientific and Medical Group 1
Class A product (reference CISPR 11, Clause 5).
ICES/NMB-001
This is a marking to indicate product compliance with the Canadian
Interference-Causing Equipment Standard (ICES-001).
Direct Current.
The RCM mark is a registered trademark of the Australian Communications and
Media Authority.
Indicates the time period during which no hazardous or toxic substance elements are
expected to leak or deteriorate during normal use. Forty years is the expected useful
life of the product.
This symbol on all primary or secondary packaging indicates compliance to China
standard GB 18455-2001.
South Korean Certification (KC) mark; includes the marking's identifier code which
follows the format: MSIP-REM-YYY-ZZZZZZZZZZZZZZ or
KCC-REM-YYY-ZZZZZZZZZZZZ.
Declaration of Conformity
A declaration of conformity is available upon request, or a copy is available on
the Keysight Technologies Web site at
Keysight Technologies Startup Guide
29
Safety and Regulatory Information
Warranty
http://www.keysight.com/go/conformity or by contacting Keysight - see
“Contacting Keysight” on page 24.
Warranty
The actual warranty on your instrument depends on the date it was ordered as
well as whether or not any warranty options were purchased at that time. To
determine the exact warranty on your instrument, contact Keysight
Technologies with the model and serial number of your instrument. Refer to
“Contacting Keysight” on page 24.
Keysight Technologies does not warrant third-party system-level (combination
of chassis, controllers, modules, etc.) performance, safety, or regulatory
compliance, unless specifically stated.
30
Keysight Technologies Startup Guide
This information is subject to change
without notice.
© Keysight Technologies 2014-2016
Edition 2, April 2016
www.keysight.com