Agilent Technologies | 34411A | Agilent 34410A and 34411A Multimeters Setting the Standard for

Agilent 34410A and 34411A Multimeters
Setting the Standard for Next Generation
Benchtop and System Testing
Product Overview
Fully compliant to
LXI Class C specification
Agilent 34410A 61/2-Digit
High-Performance DMM
• 10,000 readings/s @
5 1/2 -digits direct to PC
• 1,000 readings/s @
6 1/2 -digits direct to PC
• 30 PPM 1-year Basic DC accuracy
• LAN, USB & GPIB standard
• DCV, ACV, DCI, ACI, 2-wire and
4-wire Resistance, Frequency,
Period, Continuity, and Diode Test
• Capacitance & Temperature
measurements
• Expanded measurement ranges
• Data logger with 50 k reading
non-volatile memory
Agilent 34411A 61/2-Digit
Enhanced-Performance DMM
All the features of the 34410A, plus:
• 50,000 readings/s @
41/2 -digits direct to PC
• 1 million reading memory
• Analog level triggering
• Programmable Pre/Post triggering
The Best Just Got Better
The Agilent 34410A and
34411A 61/2-Digit DMMs represent
the latest generation of multimeters
from Agilent Technologies. Building
on the phenomenal success of the
industry-standard Agilent 34401A,
these new meters offer improved
accuracy, expanded measurement
capability, dramatically improved
measurement speed and throughput,
and modern computer interfaces
including LAN and USB. The dual
display offers both dual measurement capabilities and ease of use
when setting up and configuring
the DMM. Improvements have been
made in every facet of the 34401A
to make the best even better,
whether you use it on the bench
or in a system.
Dramatic Speed Improvements
Whether it’s raw reading speed
or fast system throughput, the
34410A sets a new benchmark in
performance. Using a new A/D
technology, the 34410A achieves
an impressive 10,000 readings a
second at 51/2-digits, and can stream
readings to your computer at this
same speed! Triggering is fast and
precise, with both trigger latency
and trigger jitter less than 1 µs,
while bus query response is less
than 500 µs. ACV measurements are
faster as well thanks to a digital
measurement technique that additionally improves accuracy at high
and low frequencies. For even
greater reading speeds, select the
34411A, which achieves 50,000
readings a second at 41/2-digits.
Enhanced Measurement Performance
The 34410A and 34411A offer
Temperature and Capacitance
capabilities, in addition to those
measurements you have come to
expect, such as DCV, ACV, DCI,
ACI, 2-wire and 4-wire Resistance,
Frequency, Period, Continuity and
Diode Test. You also get Offset
Compensated Ohms, allowing you to
accurately measure resistance in the
presence of voltages. Measurement
ranges have been expanded as well;
for example, DC and AC Current
Ranges now go down to 100 µ A,
resulting in 100 pA resolution.
Real-time math and statistics
are included, and a peak-detect
capability allows you to capture
peaks as short as 20 µs.
Agilent Technologies
Even Greater Performance
with the 34411A
The 34411A has all the features
of the 34410A, plus additional
performance that makes it even
more powerful. With the ability to
make 50,000 readings per second at
41/2 -digits, Analog Level Triggering,
programmable Pre- and Post- Trigger
and 1 million readings of volatile
memory in addition to 50,000
reading of non-volatile memory,
you now have the ability to capture
low-frequency waveforms, characterize device performance and
transfer results for analysis on
your computer.
Data Logger Function
A front panel data logger function
allows you to set the meter up to
make unattended, paced measurements over a fixed time or number
of events, then pull up the results
later for review or transfer to a
computer for analysis. Set the
meter up to take measurements
every 10 seconds for an hour, go
have lunch, and check the results
upon your return. The contextual
front panel sequences make setup
and read back a breeze.
Improved Ease of Use
From the inclusion of a second
display to configuring setups for
each of the measurement functions,
these new DMMs offer significant
improvements in usability. Simple
things are simple; more complicated
setups are easier than ever. There
is even a new probe set designed to
more easily probe today’s fine-pitch
components. Finally, there is a
built-in Graphical Web Interface
that allows you to interactively
control the DMM without the
hassle of programming at all!
Modern I/O for Improved Connectivity
When connecting to a computer,
select LAN, USB or GPIB interfaces; all three are standard on the
34410A and 34411A. Concerned
about the viability of your existing
software programs? These new DMMs
respond to Standard Commands
for Programmable Instrumentation
(SCPI), and there is even a 34401A
Emulation Mode to ensure the
easiest upgrade possible. Agilent’s
I/O Library Suite ships with the
34410A and 34411A to help you
quickly establish an error-free
connection between your PC and
instrument. It provides robust
instrument control and works
with the software development
environment you choose.
Product Reference CD-ROM with
soft documentation and software:
LXI - Class C
LAN Extensions for Instruments (LXI)
provides the next generation I/O
for system applications requiring
the highest throughput. Transfer
rates of over 250,000 readings/s are
attainable ensuring even the most
data intensive measurements are
fast, without the overhead cost
of an instrument cardcage. Both
the 34410A and the 34411A are
LXI - Class C compliant.
Bench Dimensions:
Built To Last
Our new DMMs were designed
to high standards of ruggedness
and reliability. From the robust
package with its shock absorbing
bumpers to careful selection of
components and conservative
circuit design, these meters are
built to last. Calculated Mean
Time Between Failure (MTBF) is in
excess of 100,000 hours. Backed by
a 1-year warranty and a worldwide
network of service centers, you can
buy with confidence.
•
•
•
•
•
•
•
Optional Printed Documentation:
• Quick Start Tutorial
• User’s Guide
• Service Guide
261.2 mm
103.8 mm
303.2 mm
System Dimensions:
Go to the Web
For the latest information on
these or other Agilent DMMs,
go to www.agilent.com/find/dmm
Accessories Included:
• Test Lead Kit with probes
and SMT attachments.
• Test report, power cord,
and USB interface cable.
Programmer’s Reference Help
Quick Start Tutorial
User’s Guide
Service Guide
Programming Examples
IntuiLink for Multimeters
LabVIEW and IVI-COM drivers
212.8 mm
88.3 mm
272.3 mm
2
Accuracy Specifications ± (% of reading + % of range)1
Function
DC Voltage
True RMS
AC Voltage5
Resistance6
DC Current
True RMS
AC Current 7
Frequency
or Period
Capacitance8
Temperature9
RTD
Thermistor
Continuity
Diode Test10
Range 3
100.0000 mV
1.000000 V
10.00000 V
100.0000 V
1000.000 V 4
100.0000 mV
to 750.000 V
100.0000 Ω
1.000000 kΩ
10.00000 kΩ
100.0000 kΩ
1.000000 MΩ
10.00000 MΩ
100.0000 MΩ
1.000000 GΩ
100.0000 µA
1.000000 mA
10.00000 mA
100.0000 mA
1.000000 A
3.000000 A
100.0000 µA to
3.00000 A
100 mV to
750 V
1.0000 nF
10.000 nF
100.00 nF
1.0000 µF
10.000 µF
-200°C to 600°C
-80°C to 150°C
1000.0 Ω
1.0000 V
Frequency,
Test Current or
Burden Voltage
3 Hz – 5 Hz
5 Hz – 10 Hz
10 Hz – 20 kHz
20 kHz – 50 kHz
50 kHz – 100 kHz
100 kHz – 300 kHz
1 mA
1 mA
100 µA
10 µA
5 µA
500 nA
500 nA || 10 MΩ
500 nA || 10 MΩ
< 0.03 V
< 0.3 V
< 0.03 V
< 0.3 V
< 0.8 V
< 2.0 V
3 Hz – 5 kHz
5 kHz – 10 kHz
3 Hz – 5 Hz
5 Hz – 10 Hz
10 Hz – 40 Hz
40 Hz – 300 kHz
500 nA
1 µA
10 µA
10 µA
100 µA
1 mA
1 mA
24 Hour 2
Tcal ±1°C
90 Day
Tcal ±5°C
1 Year
Tcal ±5°C
0.0030 + 0.0030
0.0020 + 0.0006
0.0015 + 0.0004
0.0020 + 0.0006
0.0020 + 0.0006
0.50 + 0.02
0.10 + 0.02
0.02 + 0.02
0.05 + 0.04
0.20 + 0.08
1.00 + 0.50
0.0030 + 0.0030
0.0020 + 0.0005
0.0020 + 0.0005
0.0020 + 0.0005
0.0020 + 0.0010
0.0100 + 0.0010
0.200 + 0.001
2.000 + 0.001
0.010 + 0.020
0.007 + 0.006
0.007 + 0.020
0.010 + 0.004
0.050 + 0.006
0.100 + 0.020
0.10 + 0.04
0.20 + 0.04
0.070 + 0.000
0.040 + 0.000
0.020 + 0.000
0.005 + 0.000
0.50 + 0.50
0.40 + 0.10
0.40 + 0.10
0.40 + 0.10
0.40 + 0.10
0.0040 + 0.0035
0.0030 + 0.0007
0.0020 + 0.0005
0.0035 + 0.0006
0.0035 + 0.0006
0.50 + 0.03
0.10 + 0.03
0.05 + 0.03
0.09 + 0.05
0.30 + 0.08
1.20 + 0.50
0.008 + 0.004
0.007 + 0.001
0.007 + 0.001
0.007 + 0.001
0.010 + 0.001
0.030 + 0.001
0.600 + 0.001
6.000 + 0.001
0.040 + 0.025
0.030 + 0.006
0.030 + 0.020
0.030 + 0.005
0.080 + 0.010
0.120 + 0.020
0.10 + 0.04
0.20 + 0.04
0.070 + 0.000
0.040 + 0.000
0.020 + 0.000
0.006 + 0.000
0.50 + 0.50
0.40 + 0.10
0.40 + 0.10
0.40 + 0.10
0.40 + 0.10
0.0050 + 0.0035
0.0035 + 0.0007
0.0030 + 0.0005
0.0040 + 0.0006
0.0040 + 0.0006
0.50 + 0.03
0.10 + 0.03
0.06 + 0.03
0.10 + 0.05
0.40 + 0.08
1.20 + 0.50
0.010 + 0.004
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.012 + 0.001
0.040 + 0.001
0.800 + 0.001
8.000 + 0.001
0.050 + 0.025
0.050 + 0.006
0.050 + 0.020
0.050 + 0.005
0.100 + 0.010
0.150 + 0.020
0.10 + 0.04
0.20 + 0.04
0.070 + 0.000
0.040 + 0.000
0.020 + 0.000
0.007 + 0.000
0.50 + 0.50
0.40 + 0.10
0.40 + 0.10
0.40 + 0.10
0.40 + 0.10
0.06°C
0.08°C
0.002 + 0.010
0.002 + 0.010
0.06°C
0.08°C
0.008 + 0.020
0.008 + 0.020
0.06°C
0.08°C
0.010 + 0.020
0.010 + 0.020
1
2
3
4
5
Specifications are for 90 minute warm-up and 100 PLC..
Relative to calibration standards..
20% overrange on all ranges, except DCV 1000 V, ACV 750 V, DCI and ACI 3 A ranges.
For each additional volt over ± 500 V add 0.02 mV of error.
Specifications are for sinewave input > 0.3% of range and > 1 mVrms. Add 30 uV error for frequencies below 1 kHz.
750 VAC range limited to 8 x 10 7 Volts-Hz. For each additional volt over 300 Vrms add 0.7 mVrms of error.
6 Specifications are for 4-wire resistance measurements, or 2-wire using Math Null.
Without Math Null, add 0.2 Ω additional error in 2-wire resistance measurements.
7 Specifications are for sinewave input > 1% of range and > 10 µArms. Frequencies > 5 kHz are typical
for 1 A and 3 A ranges.
8 Specifications are for 1-hour warm-up using Math Null. Additional errors may occur for non-film capacitors.
9 For total measurement accuracy, add temperature probe error.
10 Accuracy specifications are for the voltage measured at the input terminals only. 1 mA test current is typical.
Variation in the current source will create some variation in the voltage drop across a diode junction.
3
Temperature Coefficient/°C
0°C to (Tcal -5°C)
(Tcal +5°C) to 55°C
0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.010 + 0.003
0.008 + 0.003
0.005 + 0.003
0.010 + 0.005
0.020 + 0.008
0.120 + 0.020
0.0006 + 0.0005
0.0006 + 0.0001
0.0006 + 0.0001
0.0006 + 0.0001
0.0010 + 0.0002
0.0030 + 0.0004
0.1000 + 0.0001
1.0000 + 0.0001
0.0020 + 0.0030
0.0020 + 0.0005
0.0020 + 0.0020
0.0020 + 0.0005
0.0050 + 0.0010
0.0050 + 0.0020
0.015 + 0.006
0.030 + 0.006
0.005 + 0.000
0.005 + 0.000
0.001 + 0.000
0.001 + 0.000
0.05 + 0.05
0.05 + 0.01
0.01 + 0.01
0.01 + 0.01
0.01 + 0.01
0.003°C
0.002°C
0.0010 + 0.0020
0.0010 + 0.0020
A-to-D Converter Noise Performance
1
2
3
4
5
Resolution
(ppm of range)1
Normal Mode
Rejection (dB)2
Readings/Second4
0.0015
0.002 5
0.006
0.02
0.06
0.2
1
2
10
100
30
15
6
3
1.5
0.7
0.3
0.2
0.1
0.03
0
0
0
0
0
0
55
110 3
110 3
110 3
50,000
25,000
10,000
3,000
1,000
300
60 (50)
30 (25)
6 (5)
0.6 (0.5)
Resolution is defined as the typical DCV 10 V range RMS noise.
Auto-zero on for NPLC ≥ 1. See manual for additional noise characteristics.
Normal mode rejection for power line frequency ± 0.1%.
For power-line frequency ± 1% 75 dB and for ± 3% 55 dB.
Maximum rate with auto-zero off for 60 Hz and (50 Hz) operation.
Only available for the 34411A.
100
RMS Noise (ppm of range)
Integration Time
(NPLC)
10
1
0.1
0.01
0.001
0.01
0.1
1
10
Integration Time (NPLC)
System Reading and Throughput Rates
DMM memory to PC (Maximum reading rate out of memory)1
System Reading Architecture
Drawing – Path B
Reading
Format
ASCII
32-bit Binary
64-bit Binary
GPIB
Readings/s
USB 2.0
Readings/s
4,000
89,000
47,000
8,500
265,000
154,000
LAN (VXI-11)
Readings/s
LAN (Sockets)
Readings/s
7,000
110,000
60,000
8,500
270,000
160,000
Direct I/O Measurements (Single reading – measure and I/O time)1
Drawing – Path C
Function
DCV/2-wire
Resistance
ACV/
Frequency
1
Maximum Reading
Rate into Memory
or to Direct I/O
Resolution
(NPLC)
GPIB
ms
USB 2.0
ms
LAN
(VXI-11)
ms
LAN
(Sockets)
ms
0.006 (0.001)
2.6
2.9
4.6
3.2
10,000 (50,000)
Fast Filter
1 ms Gate
10.0
10.0
10.0
10.0
500
(Readings/s)
Drawing – Path A or C
1/2 scale input signal, immediate trigger, trigger delay 0, auto-zero off, auto-range off, no math, null off,
60 Hz line frequency. Specifications are for 34410A or (34411A). See manual for performance on other functions.
System Performance
Function
1
Change (ms)
Range
Change (ms)2
LAN/GPIB
Autorange (ms)3
Maximum
External
Trigger Rate
Maximum Internal
Trigger Rate4
DCV/2-wire
Resistance
22
3.9/2.6
7.5
5,000/s
10,000/s
(50,000/s)
ACV/
Frequency
37
6.5/6.4
19
500/s
500/s
1
Time to change from 2-wire Resistance to this specified function,
or DCV to 2-wire Resistance using the SCPI “FUNC” command.
2
Time to change from one range to the next higher range, ≤10 V, ≤10 MΩ.
3
Time to automatically change one range and be ready for the new measurement, ≤ 10 V, ≤ 10 MΩ.
4
Specifications are for 34410A or (34411A).
4
100
Measurement Characteristics
DC Voltage
True RMS AC Current
Additional 34411A Specifications
Measurement Method:
Continuously integrating
multi-slope IV A/D converter
Measurement Method:
AC-coupled True RMS measurement.
Directly coupled to the fuse and shunt.
Digital sampling with anti-alias filter.
Resolution: See table on page 4
Linearity:
(10 V range)
0.0002% of reading
+ 0.0001% of range
Input Resistance:
0.1 V, 1 V, 10 V
Ranges
100 V, 1000 V
Ranges
10 MΩ or > 10 GΩ
(Selectable)
10 MΩ ±1%
(Fixed)
Overall Bandwidth, DCV & DCI:
15 kHz typical @ 20 µs aperture (-3 dB)
Triggering: Pre/Post, Int/Ext, Pos/Neg
Current Shunt:
200 Ω for 100 µA, 1 mA
2 Ω for 10 mA, 100 mA
0.1 Ω for 1 A, 3 A
Timebase Resolution: 19.9524 µs 0.01% accuracy
Trigger Jitter:
2 µs (p-p), 20 µs (p-p) when pre-triggered
Input Bias Current: < 50 pA at 25°C
Maximum Input:
The peak value of the DC + AC current must
be <300% of range. The RMS current must be
<3 A including the DC current content.
Input Protection: 1000 V
Input Protection: 3 A, 250 V fuse
DC CMRR:
1
140 dB
Frequency and Period
True RMS AC Voltage
Measurement Method:
AC-coupled True RMS measurement.
Digital sampling with anti-alias filter.
Crest Factor:
No additional error for crest factors < 10.
Limited by peak input and 300 kHz bandwidth.
Peak Input:
300% of range or 1100 V
70 dB
Input Impedance:
1 MΩ ± 2% in parallel with < 150 pF
Input Protection: 750 Vrms all ranges
2
Function DCV
Range
Spur-Free
1V
-75 dB
10 V 1
-70 dB
100 V
-75 dB
1
10 V range: 2 V (p-p) <signal < 16 V (p-p)
SNDR
60 dB
60 dB
60 dB
Triggering and Memory
Reading Hold Sensitivity: 1% of reading
Samples per Trigger:
1 to 50,000
(34410A)
1 to 1,000,000 (34411A)
Trigger Delay: 0 to 3600 s (20 µs step size)
External Trigger:
Programmable edge, Low-power TTL compatible
Delay: < 1 µs Max rate: 5,000/s
Jitter: < 1 µs
Min Pulsewidth: 1 µs
Capacitance
Overload Ranging:
Will select higher range if peak input
overload is detected during auto range.
Overload is reported in manual ranging.
AC CMR:
Measurement Method:
Reciprocal-counting technique. AC-coupled input
using the AC voltage measurement function.
Spurious-Free Dynamic Range
& Signal to Noise Distortion Ratio
Measurement Method:
Current input with measurement of
resulting ramp.
Voltmeter Complete: 3 V Logic output,
2 µs pulse with programmable edge
Connection Type: 2-wire
Nonvolatile Memory: 50,000 readings
Maximum Input: 400 Vdc, 1100 Vpk
Temperature
Input Impedance:
1 MΩ ± 2% in parallel with < 150 pF
Thermistor:
2.2 kΩ, 5 kΩ, and 10 kΩ
Input Protection: 750 Vrms all ranges
RTD: a = 0.00385
Ro from 49 Ω to 2.1 kΩ
Volatile Memory:
50,000 readings (34410A)
1,000,000 readings (34411A)
Sample Timer:
Range: 0 to 3600 s (20 µs step sizes)
Jitter:
< 100 ns
Resistance
Continuity/Diode Test
Offset Compensation:
Selectable on the 100 Ω, 1 kΩ,
and 10 kΩ ranges
Continuity Threshold: Fixed at 10 Ω
Power Line Frequency:
45 Hz to 66 Hz and 360 Hz to 440 Hz,
Automatically sensed at power-on
Max. Lead Resistance (4-wire):
10% of range per lead for 100 Ω, 1 kΩ.
1 kΩ per lead on all other ranges
Operating Characteristics
Maximum readings/second
Power Consumption: 25 VA peak (16 W average)
Operating Environment: Full accuracy for
0°C to 55°C, 80% R.H. at 40°C non-condensing
Input Protection:
1000 V on all ranges
Function3
DCV
2-wire Ω
DCI
ACV
ACI
Frequency
Period
DC Current
Current Shunt:
200 Ω for 100 µA, 1 mA
2 Ω for 10 mA, 100 mA
0.1 Ω for 1 A, 3 A
Input Protection:
Digits
4.5
50 k4
25 k4
3k
500
500
450
450
5.5
10 k
3k
60(50)
500
150
90
90
3 A, 250 V fuse
For 1 kΩ unbalanced in LO lead, ± 500 V peak maximum
2
For 1 kΩ unbalanced in LO lead and <60 Hz, ± 500 V peak maximum
3
Maximum rate for DCV, DCI, and resistance functions
(using zero settling delay, autozero off, manual range)
34411A only
Power Supply:
100 V/120 V/220 V/240 V ±10%
Response Time:
300 samples/sec with audible tone
1
4
General Specifications
Measurement Method:
Selectable 2-wire or 4-wire.
Current source referenced to LO input.
6.5
1k
60(50)
60(50)
150
150
10
10
Storage Temperature: -40°C to 70°C
Weight: 3.72 kg (8.2 lbs)
Safety: IEC 61010-1, EN 61010-1, UL 61010-1,
CAN/CSA-C22.2 No. 61010-1, Refer to
Declarations of Conformity for current
revisions. Measurement CAT II 300 V,
CAT I 1000 V. Pollution Degree 2
EMC: IEC 61326, EN 61326, CISPR 11, ICES-001,
AS/NZS 2064.1, Refer to Declaration of
Conformity for current revisions.
Vibration & Shock: MIL-T-28800E,
Type III, Class 5 (Sine Only)
LXI Compliance: LXI Class C, ver. 1.0
5
Warranty: 1 year
Ordering Information
Agilent 34410A and
34411A Multimeters
Accessories included
Test Lead Kit with probes and
SMT attachments, test report,
power cord, and USB interface cable.
Product Reference CD-ROM with
soft documentation and software:
•
•
•
•
•
•
•
Programmer’s Reference Help
Quick Start Tutorial
User’s Guide
Service Guide
Programming Examples
IntuiLink for Multimeters
LabVIEW and IVI-COM drivers
Options
Opt. 1CM
Rack Mount Kit
Remove all doubt
Our repair and calibration services will get
your equipment back to you, performing
like new, when promised. You will get
full value out of your Agilent equipment
throughout its lifetime. Your equipment
will be serviced by Agilent-trained technicians using the latest factory calibration
procedures, automated repair diagnostics
and genuine parts. You will always have the
utmost confidence in your measurements.
Agilent offers a wide range of additional
expert test and measurement services for
your equipment, including initial start-up
assistance onsite education and training,
as well as design, system integration, and
project management.
For more information on repair and
calibration services, go to
Opt. ABA
Printed Manual Set (English)
www.agilent.com/find/removealldoubt
Opt. 0B0
Delete printed Manual Set (full
documentation included on CD ROM)
Opt. A6J
ANSI Z540 compliant calibration
Agilent Accessories
11059A Kelvin Probe set
11060A Surface Mount Device
(SMD) test probes
11062A Kelvin clip set
34131A Hard transit case
34162A Accessory pouch
34171A Input terminal connector
(sold in pairs)
34172A Input calibration short
(sold in pairs)
34330A 30 A current shunt
E2308A 5 kΩ thermistor probe
* For racking two side-by-side,
order both items below:
Lock link kit (P/N 5061-9694)
Flange kit (P/N 5063-9212)
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the products
and applications you select.
www.agilent.com
For more information on Agilent
Technologies’ products, applications
or services, please contact your local
Agilent office. The complete list is
available at:
www.agilent.com/find/contactus
Phone or Fax
United States:
(tel) 800 829 4444
(fax) 800 829 4433
Canada:
(tel) 877 894 4414
(fax) 800 746 4866
China:
(tel) 800 810 0189
(fax) 800 820 2816
Europe:
(tel) 31 20 547 2111
Japan:
(tel) (81) 426 56 7832
(fax) (81) 426 56 7840
Korea:
(tel) (080) 769 0800
(fax) (080) 769 0900
Agilent Direct
www.agilent.com/find/quick
Quickly choose and use your test
equipment solutions with confidence.
Agilent
Open
www.agilent.com/find/open
Agilent Open simplifies the process of
connecting and programming test systems
to help engineers design, validate and
manufacture electronic products. Agilent
offers open connectivity for a broad range
of system-ready instruments, open industry
software, PC-standard I/O and global support, which are combined to more easily
integrate test system development.
Latin America:
(tel) (305) 269 7500
Taiwan:
(tel) 0800 047 866
(fax) 0800 286 331
Other Asia Pacific Countries:
(tel) (65) 6375 8100
(fax) (65) 6755 0042
Email: tm_ap@agilent.com
Revised: 09/14/06
Product specifications and descriptions in this
document subject to change without notice.
© Agilent Technologies, Inc. 2006
Printed in USA, October 5, 2006
5989-3738EN
is the US registered trademark of
the LXI Consortium.
Agilent Technologies
Download PDF