REVISEDdpg bulletin 5/2/05 11:57 AM Page 1
Bulletin A-34
Series DPG-100 Digital Pressure Gage
Specifications - Installation and Operating Instructions
3
[76.20]
1-5/8
[40.39]
3-5/8
[91.39]
The Series DPG-100 Digital Pressure Gage has a precise +/-
0.25% full scale accuracy. The 4 digit digital display will reduce the potential for errors in readings by eliminating parallax error commonly produced with analog gages.
The DPG-100 is packaged in a durable extruded aluminum case designed to meet NEMA 4/4X. The unit is battery powered and has an auto-shut off to conserve battery life. A four button key pad allows easy access to features without the need to work through complex menus or difficult key combinations. These features include backlight, peak and valley, auto zero and conversion of the pressure units. Readings may be converted to various engineering units. See reverse page for available units.
DPG Quick Start Button Operation
For complete detailed instructions on the four button operation see the Operation Chart provided on back.
On/Backlight: Push to turn on; Press again for backlight. Press and hold for 5 seconds to turn unit off.
Zero: Zeros the display - Push & hold for 2 seconds to zero the display. If after pressing zero the display does not re-zero, the unit is no longer meeting specified accuracy and should be sent back to the factory for calibration. Do not use the zero button when pressure is applied.
Units: Scroll (see next page for Model/Range Chart).
P/V: Peak recorded reading retrieved by initial push of button; push again retrieves the valley or lowest value recorded.
INSTALLATION
When installing gage always use 1˝ hex at the base of the housing to tighten the gage to a mating fitting. Do not apply wrench to housing.
MAINTENANCE
A “LOW BAT” descriptor indication will appear on the display when batteries need to be replaced.
Battery Replacement: Remove the three screws on backplate. Lift backplate off by hand. Replace batteries per polarity indicators. To reassemble, ensure O-Ring seats into housing evenly and press until seated completely. Replace screws.
WARNING: Substitution of components may impair intrinsic safety. Batteries must be changed in a non-hazardous location only.
DO NOT mix batteries from different manufacturers. DO NOT mix old batteries with new batteries.
SPECIFICATIONS
Service: Liquids and combustible compatible gases (for
FM listings see agency approvals below).
Wetted Materials: Type 316L
SS.
Housing Materials: Black
Polycarbonate front & back cover, anodized aluminum extruded housing with recessed grooves, Polycarbonate Overlay,
Buna-N O-Rings, 316L SS
Sensor Construction.
Accuracy: 0.25% F.S. +/- 1 least significant digit 32 to 130°F
(0 to 55°C). (Includes linearity, hysteresis, repeatability) ±2%
F.S. @ 10°F. ±5% F.S. @ 0°F.
Pressure Limit: 2x pressure range for models ≤1000 psi;
5000 psi for 3000 psi range;
7500 psi for 5000 psi range.
Enclosure Rating: Designed to meet NEMA 4/4X (IP65).
Temperature Limits: 0 to
130°F (-18 to 55°C).
Thermal Effect: Between 70 to
130°F is 0.016%/F.
Between 32 to 70°F is
0.026%/F. Between 10 to 32°F
1/4 NPT is 0.09%/F. Between 0 to 10°F is 0.50%/F.
Size: 3.00˝ OD x 1.90 Deep
(max).
Process Connection: 1/4˝ male
NPT.
Weight: 8.84 oz (275 g).
Display: 4 digit (.425 H x .234
W digits).
Sample Rate: 2 sec.
Power Requirements: Two
AAA batteries.
Battery Life: 2000 hours typical; Low battery indicator.
Auto Shut-Off:
Gage: 60 minute auto shutoff. Auto shut-off may be disengaged.
Backlight: 2 minute auto shut-off.
Agency Approvals: FM approved to be intrinsically safe for Class I, Division I, Groups A,
B, C and D, for ranges 0-15 to
0-3000 psi.
CE EMC Directive:
EN 61000-4-2
EN 61000-4-3 & ENV50204
EN 61000-6-2
EN 61000-6-4
EN 55011.
WARNING: DO NOT pressurize DPG over maximum allowable pressure limits. Extreme damage can occur if limits are exceeded
(see Pressure Limit under Specifications). An overflow “OFL” designator will flash to acknowledge that the gage’s pressure range has been exceeded.
PANEL MOUNT OPTION
Drill a 3-1/4˝ (82.55) diameter hole in a panel, locate three mounting holes for a #6 screw 120° apart on a 3-58˝ (92.08) bolt circle.
(3) MOUNTING HOLES FOR #6 FLAT HEAD SCREW
EQUALLY SPACED ON A 3-5/8 (92.08) B.C.
4-1/16
(103.19)
3-1/4 (82.55)
HOLE IN PANEL
DWYER INSTRUMENTS, INC.
P.O. Box 373 • Michigan City, IN 46361-0373, U.S.A.
Phone: 219/879-8000
Fax: 219/872-9057
5/64
(1.98)
9/32 (7.14) MAX
PANEL THICKNESS www.dwyer-inst.com
e-mail: [email protected]
PAGE 1 OF 4
PAGE 1 OF 4
PAGE 1 OF 4
PAGE 1 OF 4
REVISEDdpg bulletin 5/2/05 11:57 AM Page 2
Bulletin A-34
Page 2
DPG-100 DISPLAY
LOW BATTERY INDICATION
UNITS
AUTO SHUT-OFF
INDICATOR
PRESSURE READING
DPG BUTTON OPERATION
Button Operation
Turn Unit On
Turn Backlight On (Auto Shut-Off: 2 Min)
Turn Backlight Off plus
Turn Unit Off
Zero Gage
Unit Selection
View Peak
View Valley
Reset Peak & Valley
Exit Peak & Valley
Disengage Auto Shut-Off
Action
Press and hold ON button
While gage is on press ON button
While backlight is on press ON button
While gage is on, press and hold ON button 5 seconds for unit to turn off.
With no pressure applied to the gage, press and hold ZERO button until display flashes
If gage will not zero, then gage is no longer meeting specified accuracy and should be sent back to the factory for recalibration
NOTE: Gage is not to be zeroed while pressure is applied
Press UNIT button to change units on gage (Not all units are available for each range, please refer to model/range chart located below for available units)
Press P/V button to read peak (highest pressure) measurement
Press P/V button again to read valley (lowest pressure) measurement
Press and hold P/V button to reset peak and valley
Press ON button to exit peak/valley
Auto shut-off will turn off an idle gage after 60 minutes. To disengage auto shut-off, press and hold ON button, then press and hold P/V button. A decimal will blink in the lower left hand corner of the display to indicate the auto-shut off is turned off. To turn on auto shut-off, repeat above procedure until decimal point stops blinking
MODEL/RANGE CHART
Model
Number
DPG-100*
DPG-102
DPG-103
DPG-104
DPG-105
DPG-106
DPG-107
DPG-108
DPG-109
DPG-110
DPG-111*
Range psi
-14.70-0
15.00
30.00
50.00
100.0
200.0
300.0
500.0
1000
3000
5000 kg/cm2
-1.033
1.055
2.109
3.515
7.03
14.06
21.09
35.15
70.3
210.9
351.5
bar
-1.013
1.034
2.069
3.448
6.895
13.79
20.69
34.48
68.98
206.9
344.8
in Hg
-29.93
30.54
61.08
101.8
203.6
407.2
610.8
1018
2036
6108
Pressure Ranges ft wc
-33.94
kPa
34.61
69.21
115.4
230.7
461.4
692.1
1154
2307
6921
-101.4
103.4
206.9
344.8
689.5
1379
2069
3448
6895 oz/in 2
-235.2
240
480
800
1600
3200
4800
8000 in wc
-407.3
415.2
830.4
1384
2768
5536
8304 mbar
-1013
1034
2069
3448
6895 cm wc
-1034
1055
2109
3515
7031 mm Hg
-760.7
775.7
1551
2586
5172
Compound Ranges available: DPG-120* Range: 30˝ Hg-0-15 psi; DPG-121* Range: 30˝ Hg-0-30 psi; DPG-122* Range: 30˝ Hg-0-45 psi;
DPG-123*: Range 30˝ Hg-0-60 psi; DPG-124*: 30˝ Hg-0-100 psi.
* Models DPG-100, DPG-111, DPG-120, DPG-121, DPG-122, DPG-123 and DPG-124 are not FM approved.
©Copyright 2005 Dwyer Instruments, Inc.
Printed in U.S.A. 5/05 FR# 18-443202-04 Rev. 8