Swift™ LT for Her
nasal PIllOWs sYsTEM
User Guide
English
Swift™ LT for Her
nasal PIllOWs sYsTEM
Thank you for choosing the Swift LT.
Rx Only
Intended Use
The Swift LT channels airflow noninvasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bilevel system.
The Swift LT is:
• to be used by adult patients (> 66 lb/30 kg) for whom positive airway
• pressure has been prescribed intended for single-patient re-use in the home environment and multipatient re-use in the hospital/institutional environment.
Using the Swift LT for Her
When using the Swift LT with ResMed flow generators that have mask setting options, if available select ‘SWIFT’; otherwise select ‘MIRAGE’ as the mask option.
Notes:
•
The Swift LT is not compatible for use with ResMed AutoSet CS™ 2 and
•
•
VPAP™ Adapt SV devices.
SmartStop may not operate effectively when using Swift LT with some flow generators that have this feature.
If you experience nasal dryness or irritation, use of a humidifier is recommended.
For a full list of compatible devices for this mask, see the Mask/Device
Compatibility List on
www.resmed.com
on the
Products
page under
Service
& Support.
If you do not have internet access, please contact your ResMed representative.
1
Parts of the Swift LT for Her & Ordering Information
Complete system
(includes XS*, S* and
M* pillows)**
60588
Frame system
(pillows, frame, short tube assembly)
XS*
60585
S*
60582
M*
60583
L*
60584
Frame assembly
(frame, short tube assembly)
60581
Plastic tube retainer
60598
Top buckle
60597 (1)
60596 (10)
Back buckle
60740 (1)
60741 (10)
Pillows
60574 (XS*)
60571 (S*)
60572 (M*)
Frame
60575
Elbow
Soft wraps
60599
Swivel
Short tube assembly
(short tube, elbow and swivel)
60577 (1)
60742 (10)
* XS Extra Small, S Small, M Medium, L Large
** Note: Large pillows are available separately 60573 (L)
2
Headgear assembly
(headgear, soft wraps and buckles)
60595
Frame assembly
(frame, short tube assembly)
60581
Fitting your Swift LT for Her
B
A
1
(A)
Hold the pillows at your nose;
(B)
pull the headgear over your head.
2
Adjust top straps (overtightening may cause excess pressure on your nose).
3
Adjust each end of the backstrap (overtightening may cause excess pressure on upper lip).
Headgear assembly
(headgear, soft wraps and buckles)
60595
4
Rotate frame to adjust pillow angle to ensure it sits comfortably.
5
Connect the short tube to the flow generator’s air tubing.
Using the Plastic Tube Retainer (Optional)
6
The Swift LT is now fitted and ready for use. Headgear should sit midway between ear and eye.
1
Push tube retainer into buckle. Clip either short tube or air tubing to tube retainer.
2
When using the tube retainer, the tubing can be positioned to either side of your face or centered.
3
3
If the mask pulls on your nose, increase length of tube between mask and tube retainer.
Disassembling your Swift LT for Her
1
Detach air tubing from short tube.
2
Remove headgear from frame.
3
Remove pillows from frame.
4
Remove elbow from frame.
Cleaning your Swift LT for Her in the home
Notes:
•
The nasal pillows system, soft wraps and headgear should be cleaned by
•
handwashing.
The soft wraps should be removed from the headgear before
•
handwashing.
The remainder of the headgear may be washed without being disassembled.
4
WARNING
Do not use aromatic-based solutions or scented oils (eg, eucalyptus or essential oils), bleach, alcohol or products that smell strongly (eg, citrus) to clean any of the system’s components. Residual vapours from these solutions can be inhaled if not rinsed thoroughly. They may also damage the system, causing cracks.
CAUTION
If any visible deterioration of a system component is apparent (cracking, crazing, tears etc.), the component should be discarded and replaced.
Daily/After Each Use
Handwash the separated mask components (excluding headgear) by gently rubbing in warm (approx. 86°F/30°C) water using mild soap for one minute.
Rinse all components well with drinking quality water and allow them to air dry out of direct sunlight.
Weekly
Handwash the headgear, soft wraps and tube retainer in warm (approx.
86°F/30°C) water using mild soap. Rinse well and allow them to air dry out of direct sunlight.
Reprocessing your Swift LT for Her between Patients
The Nasal Pillows System should be reprocessed when used between patients. Cleaning, disinfection and sterilization instructions are available from the ResMed website, www.resmed.com/masks/sterilization/americas.
If you do not have Internet access, please contact your ResMed representative.
Storage
Ensure that your Swift LT is thoroughly clean and dry before storing it for any length of time. Store the mask in a cool dry place out of direct sunlight.
Disposal
The Swift LT does not contain any hazardous substances and may be disposed of with your normal household refuse.
5
Limited Warranty
ResMed warrants that your ResMed system, including its components: frame, pillow sleeve, headgear, tubing and other accessories (“Product”) shall be free from defects in material and workmanship for a period of ninety (90) days from the date of purchase by the initial consumer.
For further details refer to the Warranty.
6
Reassembling your Swift LT for Her
1
Insert elbow into frame.
2
Align pillows with frame.
3
Push pillows firmly into frame. Ensure top and bottom retention tabs slot into place.
4
5
Check positioning markers on pillows and headgear are on the same side.
Slide left and right headgear stability arms onto frame.
Ensure inner wall of pillows is in correct position.
If not, squeeze pillow sides to release inner walls.
Reassembling the headgear
A
A
B
Lay out the headgear as shown, writing facing upwards.
B
(A)
Thread the top straps through the top buckle, then
(B)
back over to secure.
7
(A)
Thread both backstraps through the back buckle
(Velcro™ facing outwards), then
(B)
fold back over to secure.
Troubleshooting
Possible reason Solution
Pillows won’t seal properly, are uncomfortable or cause red marks.
Pillows may have been fitted incorrectly, incorrectly adjusted, or the headgear is too tight.
Inner wall position of one/ both pillows is incorrect.
Pillows are wrong size.
Pillows may be dirty.
Carefully follow instructions in “Fitting your Swift LT”.
Check headgear is not over-tightened.
Check ResMed logo on top of pillows is facing outwards.
Squeeze pillow sides to release inner wall. The correct position in shown in “Reassembling your
Swift LT”.
Talk to your clinician.
Clean pillows according to instructions in “Cleaning your Swift LT in the Home”.
Nasal pillows system is too noisy
System is assembled incorrectly.
Reassemble the Swift LT. Make sure the pillows are fully inserted into the frame.
Before using your Swift LT for Her
•
•
•
•
•
•
•
•
•
•
WARNINGS
The vent holes must be kept clear.
The mask should not be used unless the CPAP system is turned on and operating properly.
Follow all precautions when using supplemental oxygen.
Oxygen flow must be turned off when the flow generator is not operating, so that unused oxygen does not accumulate within the flow generator enclosure and create a risk of fire.
At a fixed flow rate of supplemental oxygen flow, the inhaled oxygen concentration varies, depending on the pressure settings, patient breathing pattern, mask, point of application and leak rate.
The technical specifications of the mask are provided for your clinician to check that they are compatible with the flow generator. If used outside specification or if used with incompatible devices, the seal and comfort of the mask may not be effective, optimum therapy may not be achieved, and leak, or variation in the rate of leak, may affect the flow generator function.
Discontinue using the Swift LT if you have ANY adverse reaction to the use of the mask, and consult your physician or sleep therapist.
As with all masks, some rebreathing may occur at low CPAP pressures.
Refer to your flow generator manual for details on settings and operational information.
Remove all packaging before using the mask.
Troubleshooting
Possible reason Solution
Pillows won’t seal properly, are uncomfortable or cause red marks.
Pillows may have been fitted incorrectly, incorrectly adjusted, or the headgear is too tight.
Inner wall position of one/ both pillows is incorrect.
Pillows are wrong size.
Pillows may be dirty.
Carefully follow instructions in “Fitting your Swift LT”.
Check headgear is not over-tightened.
Check ResMed logo on top of pillows is facing outwards.
Squeeze pillow sides to release inner wall. The correct position in shown in “Reassembling your
Swift LT”.
Talk to your clinician.
Clean pillows according to instructions in “Cleaning your Swift LT in the Home”.
Nasal pillows system is too noisy
System is assembled incorrectly.
Reassemble the Swift LT. Make sure the pillows are fully inserted into the frame.
Before using your Swift LT for Her
•
•
•
•
•
•
•
•
•
•
WARNINGS
The vent holes must be kept clear.
The mask should not be used unless the CPAP system is turned on and operating properly.
Follow all precautions when using supplemental oxygen.
Oxygen flow must be turned off when the flow generator is not operating, so that unused oxygen does not accumulate within the flow generator enclosure and create a risk of fire.
At a fixed flow rate of supplemental oxygen flow, the inhaled oxygen concentration varies, depending on the pressure settings, patient breathing pattern, mask, point of application and leak rate.
The technical specifications of the mask are provided for your clinician to check that they are compatible with the flow generator. If used outside specification or if used with incompatible devices, the seal and comfort of the mask may not be effective, optimum therapy may not be achieved, and leak, or variation in the rate of leak, may affect the flow generator function.
Discontinue using the Swift LT if you have ANY adverse reaction to the use of the mask, and consult your physician or sleep therapist.
As with all masks, some rebreathing may occur at low CPAP pressures.
Refer to your flow generator manual for details on settings and operational information.
Remove all packaging before using the mask.
Technical specifications
Note:
The mask system does not contain latex, PVC or DEHP materials.
Pressure–Flow
Curve
Note
: The mask contains passive venting to protect against rebreathing. As a result of manufacturing variations, the vent flow rate may vary.
50
Pressure
(cm H
2
O)
Flow
(L/min)
40 4 21
30
8 28
20
10
12 33
0
4 6 8 10 12 14 16 18 20
16 38
Mask Pressure (cm H
2
O)
20 42
Dead Space
Information
Therapy
Pressure
Resistance
Physical dead space is the empty volume of the mask to the end of the swivel. Using the large cushions this is 94 mL.
4 to 20 cm H
2
O.
Sound
Environmental
Conditions
Drop in pressure measured (nominal) at 50 L/min: 0.54 cm H
2
O and at 100 L/min: 1.95 cm H
2
O.
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871. The A-weighted sound power level of the mask is 25 dBA, with uncertainty 3 dBA. The
A-weighted sound pressure level of the mask at a distance of
1 m is 17 dBA, with uncertainty 3 dBA.
Operating temperature: +41°F to 104°F (+5°C to +40°C)
Operating humidity: up to 95% non-condensing
Storage and transport temperature: -4°F to 140°F (-20°C to
+60°C)
Storage and transport humidity: up to 95% non-condensing.
380 mm (H) x 57 mm (W) x 62 mm (D). Mask fully assembled with short tube assembly – no headgear.
Gross
Dimensions
Note:
The manufacturer reserves the right to change these specifications without notice.
Symbols
Caution, consult accompanying documents;
LOT
Lot number;
REF
Part number; Temperature limitation; Humidity limitation; Does not contain latex; Indicates a Warning or Caution and alerts you to a possible injury or explains special measures for the safe and effective use of the device.
9
ResMed Ltd
(Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia
ResMed Corp
(US Designated Agent) 14040 Danielson Street Poway CA 92064-6857 USA
ResMed (UK) Ltd
(EU Authorized Representative) 96 Milton Park Abingdon Oxfordshire OX14
4RY UK
ResMed Offices
Australia, Austria, Belgium, Brazil, China, Finland, France, Germany, Greece,
Hong Kong, India, Ireland, Italy, Japan, Malaysia, Mexico, Netherlands, New Zealand, Norway,
Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, UK, USA (see www.resmed.com
for contact details).
Swift LT Nasal Pillows System
Protected by patents: AU 785376, DE 20321468.4, DE 20321469.2, DE 20321470.6, DE
20321471.4, DE 20321472.2, EP 1314445, NZ 562417, NZ 562418, NZ 562419, NZ 553756, NZ
553757, NZ 553822, NZ 553824, NZ 553825, US 6581594, US 6823865, US 7159587, US 7318437.
Other patents pending.
Protected by design registrations: AU 301943, CN 200430102469.3, JP 1243026, NZ 407161,
US D552731. Others designs pending.
Swift, AutoSet CS and VPAP are trademarks of ResMed Ltd and are registered in U.S. Patent and
Trademark Office.
Velcro is a registered trademark of Velcro Industries B.V.
© 2008 ResMed Ltd.
1
3
608241-Eng/2
08 11
SWIFT LT for Her
USER
AMER MULTI
6 0 8 2 4 1
Global leaders in sleep and respiratory medicine
www.resmed.com
4
2
5
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project