Planar S5048 Programming Manual

Planar S5048 Programming Manual
PLANAR 304/1
PLANAR 804/1
S5048, S7530
Network Analyzers
Programming Manual
COM/DCOM
Fourth Edition
2013
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
TABLE OF CONTENTS
1
Introduction........................................................................................................................................ 8
2
Scope of Manual ................................................................................................................................ 8
3
Registering COM Server ................................................................................................................... 8
4
COM Technology Overview.............................................................................................................. 9
5
Automation Server ............................................................................................................................. 9
6
Automation Controllers ..................................................................................................................... 9
7
Local and Remote Server ................................................................................................................. 10
8
Structure of COM Objects ............................................................................................................... 12
9
Accessing the Application Object .................................................................................................... 13
10 Object Methods ................................................................................................................................ 15
11 Object Properties ............................................................................................................................. 15
12 Error Handling ................................................................................................................................. 16
13 COM Automation Data Types ......................................................................................................... 18
14 Measurement Data Arrays ............................................................................................................... 19
15 COM Server Commands .................................................................................................................. 20
NAME ................................................................................................................................................... 20
Ready..................................................................................................................................................... 21
SCPI.ABORt ......................................................................................................................................... 22
SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt). USER.FILename ........................ 23
SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).STATe ......................................... 25
SCPI.CALCulate(Ch).FSIMulator.SENDed.PMCircuit.PORT(Pt). USER.FILename ......................... 26
SCPI.CALCulate(Ch).FSIMulator.SENDed.PMCircuit.PORT(Pt).STATe ......................................... 27
SCPI.CALCulate(Ch).FSIMulator.SENDed. ZCONversion.PORT(Pt).Z0.R ...................................... 28
SCPI.CALCulate(Ch).FSIMulator.SENDed.ZCONversion.STATe ..................................................... 29
SCPI.CALCulate(Ch).PARameter.COUNt ........................................................................................... 30
SCPI.CALCulate(Ch).PARameter(Tr).DEFine..................................................................................... 31
SCPI.CALCulate(Ch).PARameter(Tr).SELect ..................................................................................... 33
SCPI.CALCulate(Ch).PARameter(Tr).SPORt ...................................................................................... 34
SCPI.CALCulate(Ch).SELected.CONVersion.FUNCtion .................................................................... 35
SCPI.CALCulate(Ch).SELected.CONVersion.STATe ......................................................................... 36
SCPI.CALCulate(Ch).SELected.CORRection.EDELay.TIME ............................................................ 37
SCPI.CALCulate(Ch).SELected.CORRection.OFFSet.PHASe............................................................ 38
SCPI.CALCulate(Ch).SELected.DATA.FDATa .................................................................................. 39
SCPI.CALCulate(Ch).SELected.DATA.FMEMory ............................................................................. 40
SCPI.CALCulate(Ch).SELected.DATA.SDATa .................................................................................. 41
SCPI.CALCulate(Ch).SELected.DATA.SMEMory ............................................................................. 42
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.CENTer .......................................................... 43
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.SHAPe ............................................................ 44
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.SPAN.............................................................. 45
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STARt ............................................................ 46
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STATe ............................................................ 47
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STOP .............................................................. 48
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.TYPE .............................................................. 49
SCPI.CALCulate(Ch).SELected.FORMat ............................................................................................ 50
SCPI.CALCulate(Ch).SELected.FUNCtion.DATA.............................................................................. 51
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.COUPle ............................................................ 52
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STARt .............................................................. 53
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STATe .............................................................. 54
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STOP ................................................................ 55
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute ......................................................................... 56
SCPI.CALCulate(Ch).SELected.FUNCtion.PEXCursion..................................................................... 57
SCPI.CALCulate(Ch).SELected.FUNCtion.POINts ............................................................................. 58
SCPI.CALCulate(Ch).SELected.FUNCtion.PPOLarity ........................................................................ 59
SCPI.CALCulate(Ch).SELected.FUNCtion.TARGet ........................................................................... 60
SCPI.CALCulate(Ch).SELected.FUNCtion.TTRansition..................................................................... 61
SCPI.CALCulate(Ch).SELected.FUNCtion.TYPE ............................................................................... 62
SCPI.CALCulate(Ch).SELected.LIMit.DATA ..................................................................................... 63
SCPI.CALCulate(Ch).SELected.LIMit.DISPlay.STATe ...................................................................... 64
SCPI.CALCulate(Ch).SELected.LIMit.FAIL ....................................................................................... 65
2
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.AMPLitude................................................................ 66
SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.MARKer.................................................................... 67
SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.STIMulus .................................................................. 68
SCPI.CALCulate(Ch).SELected.LIMit.REPort.ALL ........................................................................... 69
SCPI.CALCulate(Ch).SELected.LIMit.REPort.DATA ........................................................................ 70
SCPI.CALCulate(Ch).SELected.LIMit.REPort.POINts ....................................................................... 71
SCPI.CALCulate(Ch).SELected.LIMit.STATe .................................................................................... 72
SCPI.CALCulate(Ch).SELected.MARKer(Mk).ACTivate ................................................................... 73
SCPI.CALCulate(Ch).SELected.MARKer(Mk).BWIDth.DATA ......................................................... 74
SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.REFerence .......................................................... 75
SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.STATe ................................................................ 76
SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.THReshold ......................................................... 77
SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.TYPE .................................................................. 78
SCPI.CALCulate(Ch).SELected.MARKer.COUPle ............................................................................. 79
SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.COUPle ............................................ 80
SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STARt .............................................. 81
SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STATe .............................................. 82
SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STOP ................................................ 83
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.EXECute .................................................. 84
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.PEXCursion ............................................. 85
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.PPOLarity ................................................ 86
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TARGet ................................................... 87
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TRACking ............................................... 88
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TTRansition ............................................. 89
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TYPE ....................................................... 90
SCPI.CALCulate(Ch).SELected.MARKer.REFerence.STATe ............................................................ 91
SCPI.CALCulate(Ch).SELected.MARKer(Mk)._SET.......................................................................... 92
SCPI.CALCulate(Ch).SELected.MARKer(Mk).STATe ....................................................................... 93
SCPI.CALCulate(Ch).SELected.MARKer(Mk).X ................................................................................ 94
SCPI.CALCulate(Ch).SELected.MARKer(Mk).Y ................................................................................ 95
SCPI.CALCulate(Ch).SELected.MARKer.COUNt .............................................................................. 96
SCPI.CALCulate(Ch).SELected.MATH.FUNCtion ............................................................................. 97
SCPI.CALCulate(Ch).SELected.MATH.MEMorize ............................................................................ 98
SCPI.CALCulate(Ch).SELected.MSTatistics.DATA ........................................................................... 99
SCPI.CALCulate(Ch).SELected.MSTatistics.DOMain.MARKer.STARt .......................................... 100
SCPI.CALCulate(Ch).SELected.MSTatistics.DOMain.MARKer.STOP ........................................... 101
SCPI.CALCulate(Ch).SELected.MSTatistics.DOMain.STATe ......................................................... 102
SCPI.CALCulate(Ch).SELected.MSTatistics.STATe......................................................................... 103
SCPI.CALCulate(Ch).SELected. RLIMit.DATA ............................................................................... 104
SCPI.CALCulate(Ch).SELected.RLIMit.DISPlay.LINE .................................................................... 105
SCPI.CALCulate(Ch).SELected.RLIMit.DISPlay.SELect ................................................................. 106
SCPI.CALCulate(Ch).SELected.RLIMit.DISPlay.VALue ................................................................. 107
SCPI.CALCulate(Ch).SELected.RLIMit.FAIL .................................................................................. 108
SCPI.CALCulate(Ch).SELected.RLIMit.REPort.DATA.................................................................... 109
SCPI.CALCulate(Ch).SELected.RLIMit.STATe................................................................................ 110
SCPI.CALCulate(Ch).SELected.SMOothing.APERture .................................................................... 111
SCPI.CALCulate(Ch).SELected.SMOothing.STATe ......................................................................... 112
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.CENTer ........................................................... 113
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.IMPulse.WIDTh .............................................. 114
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.KBESsel .......................................................... 115
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.LPFRequency .................................................. 116
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.SPAN .............................................................. 117
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STARt ............................................................. 118
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STATe ............................................................. 119
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STEP.RTIMe ................................................... 120
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STIMulus......................................................... 121
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STOP ............................................................... 122
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.TYPE ............................................................... 123
SCPI.CALCulate(Ch).TRACe(Tr).DATA.FDATa ............................................................................. 124
SCPI.CALCulate(Ch).TRACe(Tr).DATA.FMEMory ........................................................................ 125
SCPI.CALCulate(Ch). TRACe(Tr).DATA.SDATa ............................................................................ 126
3
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch). TRACe(Tr).DATA.SMEMory ....................................................................... 127
SCPI.DISPlay.COLor.BACK .............................................................................................................. 128
SCPI.DISPlay.COLor.GRATicule ...................................................................................................... 129
SCPI.DISPlay.COLor.RESet............................................................................................................... 130
SCPI.DISPlay.COLor.TRACe(Tr).DATA .......................................................................................... 131
SCPI.DISPlay.COLor.TRACe(Tr).MEMory ...................................................................................... 132
SCPI.DISPlay.ENABle ....................................................................................................................... 133
SCPI.DISPlay.FSIGn .......................................................................................................................... 134
SCPI.DISPlay.IMAGe......................................................................................................................... 135
SCPI.DISPlay.SPLit ............................................................................................................................ 136
SCPI.DISPlay._UPDate.IMMediate.................................................................................................... 137
SCPI.DISPlay.WINDow(Ch).ACTivate ............................................................................................. 137
SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.ALIGn.TYPE .................................................. 138
SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.SINGle.STATe................................................ 139
SCPI.DISPlay.WINDow(Ch).SPLit .................................................................................................... 140
SCPI.DISPlay.WINDow(Ch).TITLe.DATA....................................................................................... 141
SCPI.DISPlay.WINDow(Ch).TITLe.STATe ...................................................................................... 142
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).ANNotation.MARKer. POSition.X ................................. 143
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).ANNotation.MARKer. POSition.Y ................................. 144
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).MEMory. STATe ............................................................. 145
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).STATe.............................................................................. 146
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.AUTO ............................................................. 147
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.PDIVision........................................................ 148
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.RLEVel ........................................................... 149
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.RPOSition ....................................................... 150
SCPI.DISPlay.WINDow(Ch).X.SPACing .......................................................................................... 151
SCPI.DISPlay.WINDow(Ch).Y.SCALe.DIVisions ............................................................................ 152
SCPI.HCOPy.ABORt .......................................................................................................................... 153
SCPI.HCOPy.DATE.STAMp ............................................................................................................. 153
SCPI.HCOPy.IMAGe ......................................................................................................................... 154
SCPI.HCOPy.IMMediate .................................................................................................................... 155
SCPI.HCOPy.PAINt ........................................................................................................................... 156
SCPI.IEEE4882.CLS........................................................................................................................... 157
SCPI.IEEE4882.IDN ........................................................................................................................... 157
SCPI.IEEE4882.RST........................................................................................................................... 158
SCPI.IEEE4882.TRG .......................................................................................................................... 158
SCPI.IEEE4882.WAI .......................................................................................................................... 159
SCPI.INITiate(Ch).CONTinuous ........................................................................................................ 160
SCPI.INITiate(Ch).IMMediate............................................................................................................ 161
SCPI.MMEMory.COPY(Src, Dst) ...................................................................................................... 162
SCPI.MMEMory.DELete(File) ........................................................................................................... 162
SCPI.MMEMory.LOAD.CHANnel.STATe ....................................................................................... 163
SCPI.MMEMory.LOAD.CKIT(Ck) .................................................................................................... 164
SCPI.MMEMory.LOAD.LIMit........................................................................................................... 165
SCPI.MMEMory.LOAD.PLOSs(Pt)................................................................................................... 166
SCPI.MMEMory.LOAD.RLIMit ........................................................................................................ 167
SCPI.MMEMory.LOAD.SEGMent .................................................................................................... 168
SCPI.MMEMory.LOAD.STATe ........................................................................................................ 169
SCPI.MMEMory.MDIRectory ............................................................................................................ 170
SCPI.MMEMory.STORe.CHANnel.CLEar ....................................................................................... 171
SCPI.MMEMory.STORe.CHANnel.STATe ...................................................................................... 171
SCPI.MMEMory.STORe.CKIT(Ck) ................................................................................................... 172
SCPI.MMEMory.STORe.FDATa ....................................................................................................... 173
SCPI.MMEMory.STORe.IMAGe ....................................................................................................... 174
SCPI.MMEMory.STORe.LIMit .......................................................................................................... 175
SCPI.MMEMory.STORe.PLOSs(Pt) .................................................................................................. 176
SCPI.MMEMory.STORe.RLIMit ....................................................................................................... 177
SCPI.MMEMory.STORe.SEGMent ................................................................................................... 178
SCPI.MMEMory.STORe.SNP.DATA ................................................................................................ 179
SCPI.MMEMory.STORe.SNP.FORMat ............................................................................................. 180
SCPI.MMEMory.STORe.SNP.TYPE.S1P.......................................................................................... 181
4
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.STORe.SNP.TYPE.S2P.......................................................................................... 182
SCPI.MMEMory.STORe.STATe ....................................................................................................... 183
SCPI.MMEMory.STORe.STYPe ........................................................................................................ 184
SCPI.OUTPut.STATe ......................................................................................................................... 185
SCPI.SENSe(Ch).AVERage.CLEar .................................................................................................... 186
SCPI.SENSe(Ch).AVERage.COUNt .................................................................................................. 187
SCPI.SENSe(Ch).AVERage.STATe................................................................................................... 188
SCPI.SENSe(Ch).BANDwidth.RESolution ........................................................................................ 189
SCPI.SENSe(Ch).CORRection.CLEar................................................................................................ 190
SCPI.SENSe(Ch).CORRection.COEFficient.DATA(Str, Pt_r, Pt_s) ................................................ 191
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.ERESponse .................................................. 192
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.RESPonse.OPEN......................................... 193
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.RESPonse.SHORt ....................................... 194
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.RESPonse.THRU ........................................ 195
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.SOLT1 ......................................................... 196
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.SOLT2 ......................................................... 197
SCPI.SENSe(Ch).CORRection.COEFficient.SAVE .......................................................................... 198
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.ISOLation .......................................................... 199
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.LOAD................................................................ 200
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.OPEN ................................................................ 201
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire. SHORt .............................................................. 202
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.THRU ................................................................ 203
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.LABel ..................................................................... 204
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.ORDer.LOAD(Pt) .................................................. 205
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.ORDer.OPEN(Pt) ................................................... 206
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.ORDer.SHORt(Pt).................................................. 207
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.ORDer.THRU(Pt_m, Pt_n) .................................... 208
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.RESet ...................................................................... 209
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.SELect .................................................................... 209
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).ARBitrary ............................................ 210
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C0 ........................................................ 212
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C1 ........................................................ 213
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C2 ........................................................ 214
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C3 ........................................................ 215
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).DELay ................................................. 216
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L0 ........................................................ 217
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L1 ........................................................ 218
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L2 ........................................................ 219
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L3 ........................................................ 220
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LABel .................................................. 221
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LOSS ................................................... 222
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).TYPE ................................................... 223
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).Z0 ........................................................ 224
SCPI.SENSe(Ch).CORRection.COLLect.CLEar................................................................................ 225
SCPI.SENSe(Ch).CORRection.COLLect.DATA.ISOLation(Pt_r, Pt_s)........................................... 226
SCPI.SENSe(Ch).CORRection.COLLect.DATA.LOAD(Pt) ............................................................. 227
SCPI.SENSe(Ch).CORRection.COLLect.DATA.OPEN(Pt).............................................................. 228
SCPI.SENSe(Ch).CORRection.COLLect.DATA.SHORt(Pt) ............................................................ 229
SCPI.SENSe(Ch).CORRection.COLLect.DATA.THRU. MATCh(Pt_r,Pt_s) .................................. 230
SCPI.SENSe(Ch).CORRection.COLLect.DATA.THRU. TRANsmission(Pt_r, Pt_s) ...................... 231
SCPI.SENSe(Ch).CORRection.COLLect.ECAL.INFormation .......................................................... 232
SCPI.SENSe(Ch).CORRection.COLLect.ECAL.ORIenation.STATe................................................ 233
SCPI.SENSe(Ch).CORRection.COLLect.ECAL.PATH(Pt) .............................................................. 234
SCPI.SENSe(Ch).CORRection.COLLect.ECAL.SOLT1 ................................................................... 235
SCPI.SENSe(Ch).CORRection.COLLect.ECAL.SOLT2 ................................................................... 235
SCPI.SENSe(Ch).CORRection.COLLect.ECAL.UCHar ................................................................... 236
SCPI.SENSe(Ch).CORRection.COLLect.ECAL.UTHRu.STATe ..................................................... 237
SCPI.SENSe(Ch).CORRection.COLLect.METHod.ERESponse ....................................................... 238
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.OPEN .............................................. 239
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.SHORt ............................................ 240
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.THRU ............................................. 241
5
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.METHod.SOLT1 .............................................................. 242
SCPI.SENSe(Ch).CORRection.COLLect.METHod.SOLT2 .............................................................. 243
SCPI.SENSe(Ch).CORRection.COLLect.METHod.TYPE ................................................................ 244
SCPI.SENSe(Ch).CORRection.COLLect.SAVE ................................................................................ 245
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).FREQuency(Ls) ............................................ 246
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).INCLude(Ls).STATe .................................... 247
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).LDC .............................................................. 248
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).LOSS(Ls) ...................................................... 249
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).TIME ............................................................ 250
SCPI.SENSe(Ch).CORRection.EXTension.STATe ........................................................................... 251
SCPI.SENSe.CORRection.IMPedance.INPut.MAGNitude ................................................................ 252
SCPI.SENSe(Ch).CORRection.OFFSet.CLEar .................................................................................. 253
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.LOAD .................................................. 254
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.OPEN ................................................... 255
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.SHORt.................................................. 256
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.THRU .................................................. 257
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.PMETer................................................ 258
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.CLEar .................................................................. 259
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.METHod.SMIX2 ................................................. 260
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.SAVE .................................................................. 261
SCPI.SENSe(Ch).CORRection.RECeiver(Pt).COLLect.ACQuire ..................................................... 262
SCPI.SENSe(Ch).CORRection.RECeiver(Pt).STATe........................................................................ 263
SCPI.SENSe(Ch).CORRection.STATe .............................................................................................. 264
SCPI.SENSe(Ch).CORRection.TYPE(Tr) .......................................................................................... 265
SCPI.SENSe(Ch).FREQuency.CENTer.............................................................................................. 266
SCPI.SENSe(Ch).FREQuency.CW ..................................................................................................... 267
SCPI.SENSe(Ch).FREQuency.DATA ................................................................................................ 268
SCPI.SENSe(Ch).FREQuency.SPAN ................................................................................................. 269
SCPI.SENSe(Ch).FREQuency.STARt ................................................................................................ 270
SCPI.SENSe(Ch).FREQuency.STOP ................................................................................................. 271
SCPI.SENSe(Ch).OFFSet.ADJust.CONTinuous.PERiod ................................................................... 272
SCPI.SENSe(Ch).OFFSet.ADJust.CONTinuous.STATe ................................................................... 273
SCPI.SENSe(Ch).OFFSet.ADJust.EXECute ...................................................................................... 274
SCPI.SENSe(Ch).OFFSet.ADJust.PORT ........................................................................................... 275
SCPI.SENSe(Ch).OFFSet.ADJust.STATe .......................................................................................... 276
SCPI.SENSe(Ch).OFFSet.ADJust.Value ............................................................................................ 277
SCPI.SENSe(Ch).OFFSet.PORT(Pt).FREQuency.DATA ................................................................. 278
SCPI.SENSe(Ch).OFFSet.PORT(Pt).FREQuency.DIVisor ............................................................... 279
SCPI.SENSe(Ch).OFFSet.PORT(Pt).FREQuency. MULTiplier ........................................................ 280
SCPI.SENSe(Ch).OFFSet.PORT(Pt).FREQuency. OFFSet ............................................................... 281
SCPI.SENSe(Ch).OFFSet.PORT(Pt).FREQuency. STARt ................................................................ 282
SCPI.SENSe(Ch).OFFSet.PORT(Pt).FREQuency. STOP .................................................................. 283
SCPI.SENSe(Ch).OFFSet.STATe....................................................................................................... 284
SCPI.SENSe(Ch).ROSCillator.SOURce ............................................................................................. 285
SCPI.SENSe(Ch).SEGMent.DATA .................................................................................................... 286
SCPI.SENSe(Ch).SWEep.POINt.TIME ............................................................................................. 287
SCPI.SENSe(Ch).SWEep.POINts....................................................................................................... 288
SCPI.SENSe(Ch).SWEep.TYPE......................................................................................................... 289
SCPI.SERVice.CHANnel.ACTive ...................................................................................................... 290
SCPI.SERVice.CHANnel.COUNt ...................................................................................................... 290
SCPI.SERVice.CHANnel(Ch).TRACe.ACTive ................................................................................. 291
SCPI.SERVice.CHANnel.TRACe.COUNt ......................................................................................... 291
SCPI.SERVice.PORT.COUNt ............................................................................................................ 292
SCPI.SERVice.SWEep.FREQency.MAXimum ................................................................................. 292
SCPI.SERVice.SWEep.FREQency.MINimum ................................................................................... 293
SCPI.SERVice.SWEep.POINts........................................................................................................... 293
SCPI.SOURce(Ch).POWer.CENTer................................................................................................... 294
SCPI.SOURce(Ch).POWer.LEVel.IMMediate.AMPLitude ............................................................... 295
SCPI.SOURce(Ch).POWer.LEVel.SLOPe.DATA ............................................................................. 296
SCPI.SOURce(Ch).POWer.LEVel.SLOPe.STATe ............................................................................ 297
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.COLLect.ACQuire ........................................... 298
6
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.COLLect.TABLe. LOSS.DATA ..................... 299
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.COLLect.TABLe. LOSS.STATe ..................... 300
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.DATA .............................................................. 301
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.STATe.............................................................. 302
SCPI.SOURce(Ch).POWer.SPAN ...................................................................................................... 303
SCPI.SOURce(Ch).POWer.STARt ..................................................................................................... 304
SCPI.SOURce(Ch).POWer.STOP ...................................................................................................... 305
SCPI.STATus.OPERation.CONDition ............................................................................................... 306
SCPI.STATus.OPERation.ENABle .................................................................................................... 306
SCPI.STATus.OPERation.EVENt ...................................................................................................... 307
SCPI.STATus.OPERation.NTRansition ............................................................................................. 307
SCPI.STATus.OPERation.PTRansition .............................................................................................. 308
SCPI.STATus.PRESet......................................................................................................................... 308
SCPI.STATus.QUEStionable.CONDition .......................................................................................... 309
SCPI.STATus.QUEStionable.ENABle ............................................................................................... 309
SCPI.STATus.QUEStionable.EVENt ................................................................................................. 310
SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).CONDition ....................................................... 310
SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).ENABle ............................................................ 311
SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).EVENt .............................................................. 311
SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).NTRansition ..................................................... 312
SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).PTRansition ...................................................... 313
SCPI.STATus.QUEStionable.LIMit.CONDition ................................................................................ 313
SCPI.STATus.QUEStionable.LIMit.ENABle ..................................................................................... 314
SCPI.STATus.QUEStionable.LIMit.EVENt ....................................................................................... 314
SCPI.STATus.QUEStionable.LIMit.NTRansition .............................................................................. 315
SCPI.STATus.QUEStionable.LIMit.PTRansition .............................................................................. 316
SCPI.STATus.QUEStionable.NTRansition ........................................................................................ 317
SCPI.STATus.QUEStionable.PTRansition ......................................................................................... 318
SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).CONDition ..................................................... 318
SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).ENABle.......................................................... 319
SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).EVENt............................................................ 319
SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).NTRansition ................................................... 320
SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).PTRansition ................................................... 321
SCPI.STATus.QUEStionable.RLIMit.CONDition ............................................................................. 321
SCPI.STATus.QUEStionable.RLIMit.ENABle .................................................................................. 322
SCPI.STATus.QUEStionable.RLIMit.EVENt .................................................................................... 322
SCPI.STATus.QUEStionable.RLIMit.NTRansition ........................................................................... 323
SCPI.STATus.QUEStionable.RLIMit.PTRansition ............................................................................ 324
SCPI.SYSTem.BEEPer.COMPlete.IMMediate .................................................................................. 325
SCPI.SYSTem.BEEPer.COMPlete.STATe ........................................................................................ 325
SCPI.SYSTem.BEEPer.WARNing.IMMediate .................................................................................. 326
SCPI.SYSTem.BEEPer.WARNing.STATe ........................................................................................ 326
SCPI.SYSTem.COMMunicate.ECAL.TEMPerature.SENSor ............................................................ 327
SCPI.SYSTem.COMMunicate.ECAL.IMPedance(Pt) ....................................................................... 328
SCPI.SYSTem.COMMunicate.ECAL.THRU(Pt1, Pt2) ..................................................................... 328
SCPI.SYSTem.CORRection.STATe................................................................................................... 329
SCPI.SYSTem.DATE ......................................................................................................................... 330
SCPI.SYSTem.LOCal ......................................................................................................................... 330
SCPI.SYSTem.PRESet ....................................................................................................................... 332
SCPI.SYSTem.REMote ...................................................................................................................... 332
SCPI.SYSTem.RWLock ..................................................................................................................... 333
SCPI.SYSTem.TEMPerature.SENSor(Idx) ........................................................................................ 333
SCPI.SYSTem.TIME .......................................................................................................................... 334
SCPI.TRIGger.SEQuence.IMMediate ................................................................................................ 335
SCPI.TRIGger.SEQuence.SINGle ...................................................................................................... 336
SCPI.TRIGger.SEQuence.SOURce .................................................................................................... 337
Appendix 1. IEE488.2 Status Reporting System ........................................................................ 338
Appendix 2. Error Codes .................................................................................................................. 344
Appendix 3. Programming Examples ............................................................................................ 345
7
PLANAR 304/1, 804/1, S5048, S7530
1
COM/DCOM Programming manual
Introduction
This Manual contains information on Network Analyzer remote control and its data
communication by means of user programs written with COM/DCOM technology.
COM technology is used when a user program runs together with an external
measurement instrument program on one PC. DCOM technology is used when a user
program runs on a PC connected with the measurement instrument by LAN.
Methods and techniques for writing of user programs are same for the both
technologies. The only difference between the technologies is that the DCOM
technology requires additional LAN setting performed by the LAN administrator.
Before reading this Manual, familiarize yourself with Analyzer Operating Manual.
2
Scope of Manual
This Manual directly applies to the PLANAR 304/1, PLANAR 804/1, S5048, and
S7530 models of Network Analyzer.
Throughout this manual, the terms Planar, Planar304, and Planar804 will be used
interchangeably to refer to the executable module name and to the COM object
name.
Except where specifically noted, all commands and descriptions apply across all
models covered by this document.
3
Registering COM Server
COM server is registered automatically when application Planar304.exe
(Planar804.exe, S5048.exe, S7530.exe; hereafter Planar.exe) starts up. Administrative
rights is required to register COM server. Run application as administrator at least once
before using COM server functionality.
8
PLANAR 304/1, 804/1, S5048, S7530
4
COM/DCOM Programming manual
COM Technology Overview
COM stands for Component Object Model. This programming technology was
developed by Microsoft for two purposes:
5

the model provides the specification for interaction of binary
modules created in different programming languages;

the model defines the interfacing between a client application and a
server application running either on the same PC or on two different
PCs. In the latter case, the technology has DCOM abbreviation –
Distributed COM.
Automation Server
The Planar.exe application contains a built-in COM server that enables other programs
to access its functionality. The Planar.exe application COM server was developed in
conformity with the COM automation specification. COM automation is a technology
allowing control over the COM server by the programs written in both traditional
compiling programming languages and interpreting programming languages, such as
VBScript. This enables the server applications to make their functionality accessible to
many more clients.
6
Automation Controllers
Automation controllers are client programs, which use internal functionality of COM
servers. Automation controller programs are developed by users for writing their own
add-ons for the system.
User programs can be written in different languages:

programming languages with built-in COM support, such as Visual
Basic®, Delphi, Java;

universal programming languages, such as C, C++;

Microsoft Excel and Word office applications as they include builtin programming language Visual Basic for Applications®;

program generators, such as National Instruments LabVIEW®,
Agilent VEE®, MathWorks MATLAB®.
Examples represented in this Manual are written in Visual Basic (VB). Appendix 3
contains examples written in VB, C++, and LabView languages.
9
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
Examples\COM\VBA folder contains source codes for examples written in Visual
Basic for Applications® (Microsoft Excel files).
Examples\COM\CPP folder contains source codes for the C++ language examples.
Examples\COM\LabView folder contains source codes for the LabView language
examples.
7
Local and Remote Server
Planar.exe application can function either as a local server or as a remote server of
COM automation.
Local server runs on the same PC with the automation controller and each of the
programs is executed as an individual application in a separate window. COM
technology is used in this case (Figure 1).
Remote server and the automation controller run on different PCs connected by LAN.
DCOM (Distributed COM) technology is used in this case (Figure 2). When using
DCOM it is necessary to configure the local network by means of DCOM Windows
tools.
The same automation controller is used for the both COM and DCOM technology.
Some changes to the user program may be required in operators, which establish
connection with the server. Moreover, DCOM technology requires additional settings
of the LAN performed by the LAN administrator.
10
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
User
Program
USB
COM
PLANAR.exe
Figure 1. COM technology
DCOM
PLANAR.exe
USB
User
Program
Ethernet
Figure 2. DCOM technology
11
PLANAR 304/1, 804/1, S5048, S7530
8
COM/DCOM Programming manual
Structure of COM Objects
COM server contains several objects, which provide different functionality of the
server. The COM objects of the Planar.exe application are organized in a hierarchical
structure. Figure 3 shows the main COM objects, which comprise the first three levels
of the hierarchical structure of the Planar.exe application COM objects. COM objects
provide various methods and properties, which allow access to the server functions;
besides, they allow access to the objects of the lower levels, which are not shown in
Figure 3.
Application
SCPI
CALCulate
DISPlay
IEEE4882
INITiate
MMEMory
SENSe
STATus
SOURce
SERVice
HCOPy
OUTPut
TRIGer
SYSTem
Figure 3. The structure of COM objects
The Object Application of the Planar.exe application is in the top of the hierarchy.
Access to the lower level objects is implemented via higher level objects.
12
PLANAR 304/1, 804/1, S5048, S7530
Note
COM/DCOM Programming manual
The hierarchy of COM objects and their names are taken from
the command system used for control over the Analyzer via
GPIB interface. GPIB communication protocol complies with
the IEE488.2 standard. In accordance with this standard the
SCPI programming language is used. Operators in SCPI have
hierarchical chain structure, for example:
CALCulate:PARameter:DEFine S11
The same command in COM written in VB programming
language is as follows:
app.SCPI.CALCulate.PARameter.DEFine = "S11"
9
Accessing the Application Object
To establish connection with the COM server application, create an object reference in
the client program. In COM programming the object reference needs to be acquired
preliminarily, to be used later to access the object functionality. To define an object in
Visual Basic perform the following:
1) Declare a variable as an object.
2) Assign the object to this variable.
To declare a variable, use Dim operator or other declaration statement (Public, Private
or Static). The variables used for references should of the types Variant, Object, or a
type of a specific object. For example, the following three operators declare app
variable:
Dim app
Dim app as Object
Dim app as Planar304.Application
Use Set operator and CreateObject (ObjectName, HostName) function to assign a
specific object to a variable.
ObjectName
Automation object name is always equal to “Planar304.Applcation”
(“Planar804.Applcation”, “S5048.Application”, “S7530.Application”)
HostName
Network name of the PC hosting the COM server. This parameter is not
specified in case of a local server.
13
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
For example, the following operators create Application object and assign it to app
variable:
Set app = CreateObject("Planar304.Applcation")
Set app = CreateObject("Planar304.Applcation", "Analyzer_Name")
Note
The first form of the operator is used to create the reference to
the local COM server, the second one is used to create the
reference to the remote DCOM server.
To allow access to the objects of a lower level of the hierarchy, these objects are
specified after the reference to the higher level object and separated from it by a dot.
For example:
Dim SystObj
Set SystObj = app.SCPI.SYSTem
COM objects can have indices. For example, CALCulate, INITiate, SENSe, SOURce
objects represent various aspects of the 16 measurement channels of the Analyzer.
Therefore, it is necessary to write the channel index from 1 to 16 to acquire the data of
these objects. For example:
Set SensObj1 = app.SCPI.SENSe(1)
Set SensObj2 = app.SCPI.SENSe(2)
Visual Basic allows omitting of such indices; in this case the indices are considered as
equal to 1. For example, the following VB operators are equivalent:
Set SensObj = app.SCPI.SENSe(1)
Set SensObj = app.SCPI.SENSe
14
PLANAR 304/1, 804/1, S5048, S7530
10
COM/DCOM Programming manual
Object Methods
Objects have methods. Methods are actions that can be applied to objects. The object
methods are specified after the object name and separated from it by a dot.
The following example shows the PRESet method of SYSTem object. This method
performs setting of the Analyzer to the preset condition:
app.SCPI.SYSTem.PRESet
11
Object Properties
Along with methods, objects have properties. Properties are object characteristics that
can be set or read out. The object properties are specified after the object name and
separated from it by a dot.
To modify an object characteristic, write the value of the corresponding property. To
define an object characteristic, read out the value of its property. The following
example show the setting of the POINts property of SWEep object, i.e. the number of
sweep points:
app.SCPI.SENSe.SWEp.POINts = 201
Note
Some object properties cannot be written, and some object
properties cannot be read. In such cases, the properties are
indicated as “read only” or “write only”.
15
PLANAR 304/1, 804/1, S5048, S7530
12
COM/DCOM Programming manual
Error Handling
You can use different approaches to error handling in VB program:

check the value of Err.Number variable after execution of VB
operator, which contains the call to Planar304 object;

use On Error goto VB operator.
These approaches are represented in the examples below. The following operator
causes an error in VB program as "S13" value of the DEFine property is incorrect.
app.SCPI.PARameter.DEFine = "S13"
In the first example, the value of the Err.Number variable is checked after execution of
the VB operator, which contains the call to Planar304 object. On Error Resume Next
directive instructs VB not to interrupt the program execution when the error is detected
but to pass control to the next operator in natural order.
Dim app
Public Sub HandleError1()
Set app = CreateObject("Planar304.Application")
On Error Resume Next
app.SCPI.PARameter.DEFine = "S13"
If Err.Number <> 0 Then
Msg = "Error # " & Str(Err.Number) & " was generated by " &_
Err.Source & Chr(13) & Err.Description
MsgBox Msg,,"Error"
End If
...
End Sub
16
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
In the second example, On Error GoTo ErrHandler directive instructs VB to interrupt
the program execution when the error is detected and to pass control to ErrHandler
label.
Dim app
Public Sub HandleError2()
Set app = CreateObject("Planar804.Applcation")
On Error GoTo ErrHandler
app.SCPI.PARameter.DEFine = "S13"
...
Exit Sub
ErrHandler:
Msg = "Error # " & Str(Err.Number) & " was generated by " &_
Err.Source & Chr(13) & Err.Description
MsgBox Msg,,"Error"
End Sub
17
PLANAR 304/1, 804/1, S5048, S7530
13
COM/DCOM Programming manual
COM Automation Data Types
In COM automation, there are the following data types, which can be used for client-toserver communication:
Long
32-bit signed integer, value range from –2147483648 to 2147483647
Double
64-bit double-precision floating point,
value range from
–1.79769313486232E308 to –4.94065645841247E–324 for negative
values, and from 4.94065645841247E–324 to 1.79769313486232E308
for positive values
Boolean
16-bit integer, two values 0 – false, 1 – true
String
Variable-length string
Variant
Can be either a value of arbitrary type or an array of values of arbitrary
type. In this case, the term “arbitrary type” means any allowed type of
COM automation. A variable contains information about its type and
array size (if it is an array). It is used for communication of data arrays
between a client and a server.
18
PLANAR 304/1, 804/1, S5048, S7530
14
COM/DCOM Programming manual
Measurement Data Arrays
Measurement data can be either complex values or real values. This depends on the
format selected by the user. For example, the data is real in logarithmic magnitude
format and the data is complex in polar format.
The measurement data is transferred in a Variant type variable, which represents an
array of Double type. To transfer one complex measurement, two adjacent array cells
are used. To transfer one real measurement two adjacent array cells are used as well but
the second cell is always equal to 0. Thus, measurement data array size is a double
number of the measurement points.
Measurement 1
Measurement 2
Real
Real
Imag
Imag
Measurement N
…
Real
Imag
Figure 4. Array of complex measurements
Measurement 1
Measurement 2
Value
Value
0
0
Measurement N
…
Value
0
Figure 5. Array of real measurements
19
PLANAR 304/1, 804/1, S5048, S7530
15
COM/DCOM Programming manual
COM Server Commands
NAME
Object Type
Property (read only)
Data Type
String
Description
Instrument information string. String format: manufacturer, model, serial
number, number of firmware version and number of software version.
Range
up to 40 characters
Dim ID As String
Syntax
ID = app.NAME
Equivalent Softkeys
None
20
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
Ready
Object Type
Property (read only)
Data Type
Boolean
Description
Ready state of the instrument. Reads out the True value after successful
completion of the boot process (about 10 sec). The VNA must be connected to
PC by a USB cable.
Dim State as Boolean
Syntax
State = app.Ready
Equivalent Softkeys
None
21
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.ABORt
Object Type
Method
Description
Aborts the sweep and switches the triggers of all the channels to Hold. The
channels in Continuous trigger mode switch to waiting for a trigger. If the
trigger source is set to Internal, the channel from waiting for a trigger turns to a
new sweep.
Syntax
app.SCPI.ABORt
Equivalent Softkeys
Stimulus > Trigger > Restart
22
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).
USER.FILename
Object Type
Property (read/write)
Data Type
String
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
De-embedding function file name (*.s2p). The file contains the circuit S–
parameters in Touchstone format.
Range
up to 256 characters
Preset Value
""
Dim File As String
File =
Syntax
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).USER.FILename
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).USER.FILename =
"network.s2p"
Notes
If the full path to the file is not specified, the \FixtureSim subdirectory of the
main directory will be searched for the file.
Equivalent Softkeys
Analysis > Fixture Simulator > De–Embedding > S–parameters File
23
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
Table 1. Ch: Channel Number
Data Type
Long
Description
Channel number.
Range
from 1 to 16
Out of Range
An error occurs. Error code: 201.
Notes
If the channel number is not specified, it is taken as equal to 1.
Table 2. Pt: Port Number
Data Type
Long
Description
Port number.
Range
from 1 to 2
Out of Range
An error occurs. Error code: 114.
Notes
If the port number is not specified, it is taken as equal to 1.
24
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).STATe
Object Type
Property (read/write)
Data Type
Boolean
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
The ON/OFF state of the e-embedding function.
Allowable Values
True: De-embedding function ON
False: De-embedding function OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).STATe
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).STATe = True
Equivalent Softkeys
Analysis > Fixture Simulator > De–Embedding > Port n
25
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).FSIMulator.SENDed.PMCircuit.PORT(Pt).
USER.FILename
Object Type
Property (read/write)
Data Type
String
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
Embedding function file name (*.s2p). The file contains the circuit S–
parameters in Touchstone format.
Range
up to 256 characters
Preset Value
""
Dim File As String
File =
Syntax
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.PMCircuit.PORT(Pt).USER.FILename
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.PMCircuit.PORT(Pt).USER.FILename =
"network.s2p"
Notes
If the full path to the file is not specified, the \FixtureSim subdirectory of the
main directory will be searched for the file.
Equivalent Softkeys
Analysis > Fixture Simulator > Embedding > Port n
26
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).FSIMulator.SENDed.PMCircuit.PORT(Pt).STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
The ON/OFF state of the embedding function.
Allowable Values
True: Embedding function ON
False: Embedding function OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).STATe
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).STATe = True
Equivalent Softkeys
Analysis > Fixture Simulator > Embedding > Port n
27
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).FSIMulator.SENDed. ZCONversion.PORT(Pt).Z0.R
Object Type
Property (read/write)
Data Type
Double
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
The value of the impedance for port impedance conversion function.
Range
from 1e–6 to 1e6
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
50
Unit
Ω (Ohm)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).FSIMulator.SENDed.ZCONversion.PORT(Pt).Z0.R
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.ZCONversion.PORT(Pt).Z0.R = 75
Equivalent Softkeys
Analysis > Fixture Simulator > Port Z Conversion > Port n Z0
28
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).FSIMulator.SENDed.ZCONversion.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the port impedance conversion function.
Allowable Values
True: Port Z conversion function ON
False: Port Z conversion function OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).FSIMulator.SENDed.ZCONversion.STATe
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.ZCONversion.STATe = True
Equivalent Softkeys
Analysis > Fixture Simulator > Port Z Conversion > Port Z Conversion
29
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).PARameter.COUNt
Object Type
Property (read/write)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The number of traces in the channel.
Range
from 1 to 16
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1
Dim TraceNum As Long
Syntax
TraceNum = app.SCPI.CALCulate(Ch).PARameter.COUNt
app.SCPI.CALCulate(Ch).PARameter.COUNt = 2
Equivalent Softkeys
Display > Num of Traces
30
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).PARameter(Tr).DEFine
Object Type
Property (read/write)
Data Type
String
Target
Trace Tr of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Tr:
trace number 1–16 (see Table 3 on page 32)
Description
The measurement parameter of the trace.
Allowable Values
Out of Range
"S11"
: S11 parameter
"S21"
: S21 parameter
"S12"
: S12 parameter
"S22"
: S22 parameter
"A"
: Receiver A
"B"
: Receiver B
"R1"
: Receiver R1
"R2"
: Receiver R2
An error occurs. Error code: 208.
Depends on the trace number.
Preset Value
Tr 1, Tr 5, Tr 9, Tr 13:
Tr 2, Tr 6, Tr 10, Tr 14:
Tr 3, Tr 7, Tr 11, Tr 15:
Tr 4, Tr 8, Tr 12, Tr 16:
"S11"
"S21"
"S12"
"S22"
Dim Meas As String
Syntax
Meas = app.SCPI.CALCulate(Ch).PARameter(Tr).DEFine
app.SCPI.CALCulate(Ch).PARameter(Tr).DEFine = "S11"
Measurement > Measurement > S11 | S21 | S12 | S22
Equivalent Softkeys
Measurement > Measurement > Absolute > { A(1) | B(1) | R1(1) | A(2) | B(2) | R2(2) }
31
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
Table 3. Tr: Trace Number
Data Type
Long
Description
Trace number
Range
from 1 to 16
Out of Range
An error occurs. Error code: 202.
Notes
If the trace number is not specified, it is taken as equal to 1.
32
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).PARameter(Tr).SELect
Object Type
Method
Target
Trace Tr of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Tr:
trace number 1–16 (see Table 3 on page 32)
Description
Sets the active channel and trace.
Syntax
app.SCPI.CALCulate(Ch).PARameter(Tr).SELect
Notes
If the channel number is greater than the number of the channels displayed, an
error occurs and the command is ignored. If the trace number is greater than the
number of the traces displayed in the channel, an error occurs and the
command is ignored.
Equivalent Softkeys
Display > Active Trace/Channel > Active Channel
Display > Active Trace/Channel > Active Trace
33
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).PARameter(Tr).SPORt
Object Type
Property (read/write)
Data Type
Long
Target
Trace Tr of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Tr:
trace number 1–16 (see Table 3 on page 32)
Description
The number of the source port for absolute measurements.
Range
from 1 to 2
Out of Range
An error occurs. Error code: 208.
Preset Value
1
Dim StimPort As Long
Syntax
StimPort = app.SCPI.CALCulate(Ch).PARameter(Tr).SPORt
app.SCPI.CALCulate(Ch).PARameter(Tr).SPORt = 1
Equivalent Softkeys
Measurement > Measurement > Absolute > A(1) | B(1) | R1(1) | A(2) | B(2) | R2(2)
34
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.CONVersion.FUNCtion
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch: channel number 1–16 (see Table 1 on page 24)
Description
The S–parameter conversion function type.
Range
"ZREFlection"
: Reflection equivalent impedance
"ZTRansmit"
: Transmission equivalent impedance
"YREFlection"
: Reflection equivalent admittance
"YTRansmit"
: Transmission equivalent admittance
"INVersion"
: Inverse S–parameter
"ZTSHunt"
: Shunt equivalent impedance
"YTSHunt"
: Shunt equivalent admittance
"CONJugation"
: S–parameter conjugate
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
An error occurs. Error code: 217.
Preset Value
"ZREF"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.CONVersion.FUNCtion
app.SCPI.CALCulate(Ch).SELected.CONVersion.FUNCtion = "ZTR"
Equivalent Softkeys
Analysis > Conversion > Zr | Zt | Yr | Yt | 1/S | Z Trans–Shunt |
Y Trans–Shunt | Conjugation
35
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.CONVersion.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the S-parameter conversion function.
Allowable Values
True: S–parameter conversion function ON
False: S–parameter conversion function OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.CONVersion.STATe
app.SCPI.CALCulate(Ch).SELected.CONVersion.STATe = true
Equivalent Softkeys
Analysis > Conversion > Conversion
36
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.CORRection.EDELay.TIME
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The value of the electrical delay.
Range
from –10 to 10
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.CORRection.EDELay.TIME
app.SCPI.CALCulate(Ch).SELected.CORRection.EDELay.TIME = 1e–9
Equivalent Softkeys
Scale > Electrical Delay
37
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.CORRection.OFFSet.PHASe
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The value of the phase offset.
Range
from –360 to 360
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
° (degree)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.CORRection.OFFSet.PHASe
app.SCPI.CALCulate(Ch).SELected.CORRection.OFFSet.PHASe = 360
Equivalent Softkeys
Scale > Phase Offset
38
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.DATA.FDATa
Object Type
Property (read only)
Data Type
Variant (array of double)
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The formatted data array. The array elements contain measurements in the
current format, for example, in logarithmic magnitude format (Log Mag). Also,
see section “Measurement Data Arrays” on page 19.
The array size is 2N, where N is the number of measurement points.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
real number in rectangular format, real part in polar and
Smith chart formats;
Data(2n–1)
0 in rectangular format, imaginary part in polar and Smith
chart formats.
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch).SELected.DATA.FDATa
Equivalent Softkeys
None
39
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.DATA.FMEMory
Object Type
Property (read only)
Data Type
Variant (array of double)
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The formatted memory array. The array elements contain saved measurements
in the current format, for example, in logarithmic magnitude format (Log Mag).
Also, see section “Measurement Data Arrays” on page 19.
The array size is 2N, where N is the number of measurement points.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
real number in rectangular format, real part in polar and
Smith chart formats;
Data(2n–1)
0 in rectangular format, imaginary part in polar and Smith
chart formats.
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch).SELected.DATA.FMEMory
Equivalent Softkeys
None
40
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.DATA.SDATa
Object Type
Property (read only)
Data Type
Variant (array of double)
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The corrected data array. The corrected measurements are complex numbers.
Also, see section “Measurement Data Arrays” on page 19.
The array size is 2N, where N is the number of measurement points.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
the real part of corrected measurement;
Data(2n–1)
the imaginary part of corrected measurement.
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch).SELected.DATA.SDATa
Equivalent Softkeys
None
41
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.DATA.SMEMory
Object Type
Property (read only)
Data Type
Variant (array of double)
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The corrected memory array. The corrected measurements are complex
numbers. Also, see section “Measurement Data Arrays” on page 19.
The array size is 2N, where N is the number of measurement points.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
the real part of corrected measurement memory;
Data(2n–1)
the imaginary part of corrected measurement memory.
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch).SELected.DATA.SMEMory
Equivalent Softkeys
None
42
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.CENTer
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The gate center value of the gating function.
Range
Varies depending on the frequency span and the number of points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.CENTer
app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.CENTer = 1e–8
Equivalent Softkeys
Analysis > Gating > Center
43
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.SHAPe
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The gate shape of the gating function.
Range
"MAXimum"
: Maximum shape
"WIDE"
: Wide shape
"NORMal"
: Normal shape
"MINimum"
: Minimum shape
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
An error occurs. Error code: 218.
Preset Value
"NORM"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.SHAPe
app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.SHAPe = "MAX"
Equivalent Softkeys
Analysis > Gating > Shape > Maximum | Wide | Normal | Minimum
44
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.SPAN
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The gate span value of the gating function.
Range
Varies depending on the frequency span and the number of points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
2e–8
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.SPAN
app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.SPAN = 1e–8
Equivalent Softkeys
Analysis > Gating > Span
45
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STARt
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The gate start value of the gating function.
Range
Varies depending on the frequency span and the number of points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
–1e–8
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STARt
app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STARt = 1e–7
Equivalent Softkeys
Analysis > Gating > Start
46
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the gating function.
Allowable Values
True: Gating function ON
False: Gating function OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STATe
app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STATe = Status
Equivalent Softkeys
Analysis > Gating > Gating
47
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STOP
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The gate stop value of the gating function.
Range
Varies depending on the frequency span and the number of points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1e–8
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STOP
app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STOP = 1e–7
Equivalent Softkeys
Analysis > Gating > Stop
48
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.TYPE
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The gate type of the gating function.
Range
"BPASs"
: Bandpass type
"NOTCh"
: Notch type
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
An error occurs. Error code: 219.
Preset Value
"BPAS"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.TYPE
app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.TYPE = "bpas"
Equivalent Softkeys
Analysis > Gating > Type
49
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FORMat
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Data format.
Range
"MLOGarithmic"
"PHASe"
"GDELay"
"SLINear"
"SLOGarithmic"
"SCOMplex"
"SMITh"
"SADMittance"
"PLINear"
"PLOGarithmic"
"POLar"
"MLINear"
"SWR"
"REAL"
"IMAGinary"
"UPHase"
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
An error occurs. Error code: 209.
Preset Value
"MLOG"
: Logarithmic magnitude
: Phase
: Group delay time
: Smith chart format (Lin)
: Smith chart format (Log)
: Smith chart format (Real/Imag)
: Smith chart format (R + jX)
: Smith chart format (G + jB)
: Polar format (Lin)
: Polar format (Log)
: Polar format (Real/Imag)
: Linear magnitude
: Voltage standing wave ratio
: Real part
: Imaginary part
: Expanded phase
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.FORMat
app.SCPI.CALCulate(Ch).SELected.FORMat = "PHAS"
Format > Log Mag | Phase | Group Delay | Lin Mag | SWR | Real | Imag | Phase >
180
Equivalent Softkeys
Format > Smith > Log/Phase | Lin/Phase | Real/Imag | R+jX | G+jB
Format > Polar > Log/Phase | Ling/Phase | Real/Imag
50
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.DATA
Object Type
Property (read only)
Data Type
Variant (array of double)
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The data array of analysis executed by the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
The array size is 2N, where N is the number of points defined by the
SCPI.CALCulate(Ch).SELected.FUNCtion.POINts property.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
the response value in the n–th measurement point;
Data(2n–1)
the stimulus value in the n–th measurement point. Always
set to 0 for the analysis of mean value, standard deviation,
and peak–to–peak value.
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch).SELected.FUNCtion.DATA
Equivalent Softkeys
None
51
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.COUPle
Object Type
Property (read/write)
Data Type
Boolean
Target
All traces of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF coupling state of the analysis range for the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
Allowable Values
True: Coupling state ON
False: Coupling state OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.COUPle
app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.COUPle = Status
Equivalent Softkeys
None
52
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STARt
Object Type
Property (read/write)
Data Type
Double
Target
All traces of channel Ch (if the coupling is set to OFF by the
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.COUPle property), the active
trace of channel Ch (if otherwise),
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The start value of the analysis range set by the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
Range
From the stimulus current start value to the stimulus current stop value.
Out of Range
No limitation
Preset Value
0
Unit
Hz (Hertz) | s (second) | dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STARt
app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STARt = 1e9
Equivalent Softkeys
None
53
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
All traces of channel Ch (if the coupling is set to OFF by the
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.COUPle property), the active
trace of channel Ch (if otherwise),
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the arbitrary range when executing the analysis by the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
Allowable Values
True: Arbitrary range ON
False: Arbitrary range OFF (entire sweep range)
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STATe
app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STATe = true
Equivalent Softkeys
None
54
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STOP
Object Type
Property (read/write)
Data Type
Double
Target
All traces of channel Ch (if the coupling is set to OFF by the
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.COUPle property), the active
trace of channel Ch (if otherwise),
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The stop value of the analysis range set by the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
Range
From the stimulus current start value to the stimulus current stop value.
Out of Range
No limitation
Preset Value
0
Unit
Hz (Hertz) | s (second) | dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STOP
app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STOP = 2e9
Equivalent Softkeys
None
55
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute
Object Type
Method
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Executes the analysis specified with the
Description
property.
The analysis result can then be read out with the
SCPI.CALCulate(Ch).SELected.FUNCtion.DATA property.
SCPI.CALCulate(Ch).SELected.FUNCtion.TYPE
Syntax
app.SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute
Equivalent Softkeys
None
56
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.PEXCursion
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The lower limit for the peak excursion value when executing the peak search
by the SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
Range
Varies depending on the trace format.
Out of Range
No limitation
Preset Value
3
Unit
dB (decibel) | ° (degree) | s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.FUNCtion.PEXCursion
app.SCPI.CALCulate(Ch).SELected.FUNCtion.PEXCursion = 1.5
Equivalent Softkeys
None
57
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.POINts
Object Type
Property (read only)
Data Type
Long
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The number of points (data pairs) of the analysis result by the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
Always equal to 1, when the search is executed for the maximum, minimum,
mean, standard deviation, peak, and peak–to–peak values. The actual number
of points is read out, when the search is executed for all peaks or all targets.
Dim Value As Long
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.FUNCtion.POINts
Equivalent Softkeys
None
58
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.PPOLarity
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Range
The
polarity selection
when
performing
the
peak
search
by
the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
"POSitive"
: Positive peaks
"NEGative"
: Negative peaks
"BOTH"
: Both positive peaks and negative peaks
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
The value is ignored.
Preset Value
"POS"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.FUNCtion.PPOLarity
app.SCPI.CALCulate(Ch).SELected.FUNCtion.PPOLarity = "NEG"
Equivalent Softkeys
None
59
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.TARGet
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The target level when performing the search for the trace and the target level
crosspoints by the SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
Range
Varies depending on the trace format.
Out of Range
No limitation
Preset Value
0
Unit
dB (decibel) | ° (degree) | s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.FUNCtion.TARGet
app.SCPI.CALCulate(Ch).SELected.FUNCtion.TARGet = –10
Equivalent Softkeys
None
60
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.TTRansition
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The transition type selection when performing the search for the trace and the
target level crosspoints by the SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute
method.
Range
"POSitive"
: Positive peaks
"NEGative"
: Negative peaks
"BOTH"
: Both positive peaks and negative peaks
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
The value is ignored.
Preset Value
"POS"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.FUNCtion.TTRansition
app.SCPI.CALCulate(Ch).SELected.FUNCtion.TTRansition = "both"
Equivalent Softkeys
None
61
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.TYPE
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Range
The
selection
of
the
type
of
analysis
executed
by
the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
"PTPeak"
: Peak–to–peak (difference between the maximum value and
the minimum value)
"STDEV"
: Standard deviation
"MEAN"
: Mean value
"MAXimum"
: Maximum value
"MINimum"
: Minimum value
"PEAK"
: Search for the peak
"APEak"
: Search for all the peaks
"ATARget"
: Search for all targets
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
The value is ignored.
Preset Value
"PTP"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.FUNCtion.TYPE
app.SCPI.CALCulate(Ch).SELected.FUNCtion.TYPE = "STDEV"
Equivalent Softkeys
None
62
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.DATA
Object Type
Property (read/write)
Data Type
Variant (array of double)
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The data array, which is the limit line in the limit test function. The array size is
1 + 5N, where N is the number of measuring points.
For the n–th point, where n from 1 to N:
Data(0)
The number of limit line segments N is from 0 to 100.
Setting 0 clears the limit line;
Data(5n–4)
type of the n–th limit line segment;
0: OFF
Description
1: Upper limit
2: Lower limit
Notes
Data(5n–3)
the stimulus value in the start point of the n–th segment;
Data(5n–2)
the stimulus value in the end point of the n–th segment;
Data(5n–1)
the response value in the start point of the n–th segment;
Data(5n–0)
the response value in the end point of the n–th segment.
If the array size is not 1 + 5N, where N is Data(0), an error occurs (error code
214). If Data(5n – 4) is less than 0 or more than 2, an error occurs (error code
214). When Data(5n–3), Data(5n–2), Data(5n–1) and Data(5n–0) elements are
out of allowable range, the value is set to the limit, which is closer to the
specified value.
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch).SELected.LIMit.DATA
app.SCPI.CALCulate(Ch).SELected.LIMit.DATA = Array(1,2,800,900,–10,–10)
Equivalent Softkeys
Analysis > Limit Test > Edit Limit Line
63
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.DISPlay.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the limit line display of the limit test function.
Allowable Values
True: Limit line display ON
False: Limit line display OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.LIMit.DISPlay.STATe
app.SCPI.CALCulate(Ch).SELected.LIMit.DISPlay.STATe = true
Equivalent Softkeys
Analysis > Limit Test > Limit Line
64
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.FAIL
Object Type
Property (read only)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The limit test result.
Allowable Values
True: Fail
False: Pass
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.LIMit.FAIL
Equivalent Softkeys
None
65
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.AMPLitude
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The value of the limit line offset along Y–axis.
Range
Varies depending on the trace format.
Out of Range
No limitation
Preset Value
0
Unit
dB (decibel) | ° (degree) | s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.AMPLitude
app.SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.AMPLitude = –10
Equivalent Softkeys
Analysis > Limit Test > Limit Line Offsets > Response Offset
66
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.MARKer
Object Type
Method
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Sets the value of the limit line offset along Y–axis to the active marker value.
Syntax
app.SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.MARKer
Equivalent Softkeys
Analysis > Limit Test > Limit Line Offsets > Marker > Response Ofs
67
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.STIMulus
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The value of the limit line offset along X–axis.
Range
From the stimulus current start value to the stimulus current stop value.
Out of Range
No limitation
Preset Value
0
Unit
Hz (Hertz) | s (second) | dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.STIMulus
app.SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.STIMulus = 1e6
Equivalent Softkeys
Analysis > Limit Test > Limit Lines Offsets > Stimulus Offset
68
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.REPort.ALL
Object Type
Property (read only)
Data Type
Variant (array of double)
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The data array, which is the limit test results. The array size is 4N, where N is
the number of measurement points.
For the n–th point, where n from 1 to N:
Data(4n–3)
the stimulus value in the n–th point
Data(4n–2) the limit test result in the n–th point
–1: No limit
0: Fail
1: Pass
Description
Data(4n–1)
the upper limit value in the n–th point (0 – if there is no
limit)
Data(4n–0)
the lower limit value in the n–th point (0 – if there is no
limit)
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch).SELected.LIMit.REPort.ALL
Equivalent Softkeys
None
69
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.REPort.DATA
Object Type
Property (read only)
Data Type
Variant (array of double)
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The data array, which is the stimulus values at all the measurement points that
failed the limit test. The array size is defined by the
SCPI.CALCulate(Ch).SELected.LIMit.REPort.POINts property.
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch).SELected.LIMit.REPort.DATA
Equivalent Softkeys
None
70
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.REPort.POINts
Object Type
Property (read only)
Data Type
Long
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The number of the measurement points that failed the limit test.
Description
The array of stimulus values of the points can be read out by the
SCPI.CALCulate(Ch).SELected.LIMit.REPort.DATA property.
Dim Cnt As Long
Syntax
Cnt = app.SCPI.CALCulate(Ch).SELected.LIMit.REPort.POINts
Equivalent Softkeys
None
71
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the limit test function.
Allowable Values
True: Limit test function ON
False: Limit test function OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.LIMit.STATe
app.SCPI.CALCulate(Ch).SELected.LIMit.STATe = true
Equivalent Softkeys
Analysis > Limit Test > Limit Test
72
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).ACTivate
Object Type
Method
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Mk: marker number 1–15, or reference marker number 16 (see Table
4 on page 73)
Sets the active marker. If a marker is OFF this function will turn it ON.
Description
Syntax
Turning ON a marker with the number from 1 to 15 will turn ON all the
markers of smaller numbers. Turning ON the reference marker with number 16
does not turn ON the markers with the numbers from 1 to 15, but switches
these markers to the relative measurement mode.
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).ACTivate
Markers > Select > Marker n
Equivalent Softkeys
Markers > Reference Marker
Table 4. Mk: Marker Number
Data Type
Long
Description
Marker number. Numbers from 1 to 15 are for regular markers, number 16 is
for the reference marker.
Range
from 1 to 16
Out of Range
An error occurs. Error code: 203.
Notes
If the marker number is not specified, it is taken as equal to 1.
73
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).BWIDth.DATA
Object Type
Property (read only)
Data Type
Variant (array of double)
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Mk: marker number 1–15, or reference marker number 16 (see Table
4 on page 73)
The bandwidth search result. The bandwidth search can be performed relatively
to the marker Mk, or relatively to the absolute maximum value of the trace (in
this case the marker number is ignored), what is set by the
SCPI.CALCulate(Ch).SELected.MARKer(Mk).BWIDth.REFerence property.
Description
Notes
The array contains 4 elements:
Data(0)
Bandwidth;
Data(1)
Center frequency;
Data(2)
Q value;
Data(3)
Loss.
If the bandwidth search is impossible, all the read out values are 0. If the search
is performed relatively to a maker, which is OFF, an error occurs (error code
204).
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).BWIDth.DATA
Equivalent Softkeys
None
74
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.REFerence
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The selection of the reference point for the bandwidth search function:
reference marker or absolute maximum value of the trace.
"MARKer"
Range
: Bandwidth search relative to the reference marker
"MAXimum" : Bandwidth search relative to the absolute maximum of the
trace
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
The value is ignored.
Preset Value
"MAX"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.REFerence
app.SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.REFerence = "marker"
Equivalent Softkeys
Markers > Marker Math > Bandwidth Search > Search Ref To
75
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the bandwidth search function.
Allowable Values
True: Bandwidth search function ON
False: Bandwidth search function OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.STATe
app.SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.STATe = true
Equivalent Softkeys
Markers > Marker Math > Bandwidth Search > Bandwidth Search
76
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.THReshold
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The bandwidth definition value.
Range
Varies depending on the trace format.
Out of Range
No limitation
Preset Value
–3
Unit
dB (decibel) | ° (degree) | s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).BWIDth.THReshold
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).BWIDth.THReshold = –6.0
Equivalent Softkeys
Markers > Marker Math > Bandwidth Search > Bandwidth Value
77
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.TYPE
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The type of the bandwidth search function.
Range
"BPASs"
: Bandpass
"NOTCh"
: Notch
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
The value is ignored.
Preset Value
"BPAS"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.TYPE
app.SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.TYPE = "notc"
Equivalent Softkeys
Markers > Marker Math > Bandwidth Search > Type
78
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.COUPle
Object Type
Property (read/write)
Data Type
Boolean
Target
All traces of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the marker coupling function.
Allowable Values
True: Marker coupling ON
False: Marker coupling OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.MARKer.COUPle
app.SCPI.CALCulate(Ch).SELected.MARKer.COUPle = false
Equivalent Softkeys
Marker > Properties > Marker Couple
79
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.COUPle
Object Type
Property (read/write)
Data Type
Boolean
Target
All traces of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the marker search range coupling for different traces.
Allowable Values
True: Marker search range coupling ON
False: Marker search range coupling OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.COUPle
app.SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.COUPle = false
Equivalent Softkeys
Markers > Marker Search > Couple
80
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STARt
Object Type
Property (read/write)
Data Type
Double
Target
All traces of channel Ch (if the marker search range coupling is set to OFF by
the SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.COUPle property), the active
trace of channel Ch (if otherwise),
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The start value of the marker search range.
Range
From the stimulus current start value to the stimulus current stop value.
Out of Range
No limitation
Preset Value
3e5
Unit
Hz (Hertz) | s (second) | dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STARt
app.SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STARt = 1e6
Equivalent Softkeys
Markers > Marker Search > Search Start
81
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
All traces of channel Ch (if the marker search range coupling is set to OFF by
the SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.COUPle property),
the active trace of channel Ch (if otherwise),
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the arbitrary range when executing the marker search.
Allowable Values
True: Marker search range ON
False: Marker search range OFF (entire sweep range)
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STATe
app.SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STATe = true
Equivalent Softkeys
Markers > Marker Search > Search Range
82
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STOP
Object Type
Property (read/write)
Data Type
Double
Target
All traces of channel Ch (if the marker search range coupling is set to OFF by
the SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.COUPle property),
the active trace of channel Ch (if otherwise),
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The stop value of the marker search range.
Range
From the stimulus current start value to the stimulus current stop value.
Out of Range
No limitation
Preset Value
3.2e9 (PLANAR 304/1)
8.0e9 (PLANAR 804/1)
4.8e9 (S5048)
3.0e9 (S7530)
Unit
Hz (Hertz) | s (second) | dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STOP
app.SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STOP = 1e6
Equivalent Softkeys
Markers > Marker Search > Search Stop
83
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.EXECute
Object Type
Method
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Mk: marker number 1–15, or reference marker number 16 (see Table
4 on page 73)
Description
Executes the marker search according to the specified criterion. The type of the
marker search is set by the
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TYPE property.
Syntax
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.EXECute
Markers > Marker Search > Maximum | Minimum
Equivalent Softkeys
Markers > Marker Search > Peak > Search Peak | Search Max Peak | Search
Peak Left | Search Peak Right
Markers > Marker Search > Target > Search Target | Search Target Left | Search
Target Right
84
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.PEXCursion
Object Type
Property (read/write)
Data Type
Double
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Mk: marker number 1–15, or reference marker number 16 (see Table
4 on page 73)
Description
The peak excursion value, when the marker search for peak is performed by the
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.EXECute method.
Range
Varies depending on the trace format.
Out of Range
No limitation
Preset Value
1
Unit
dB (decibel) | ° (degree) | s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.PEXCursion
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.PEXCursion = 3.0
Equivalent Softkeys
Markers > Marker Search > Peak > Peak Excursion
85
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.PPOLarity
Object Type
Property (read/write)
Data Type
String
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Mk: marker number 1–15, or reference marker number 16 (see Table
4 on page 73)
Description
The peak polarity selection, when the marker search for peak is performed by
the SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.EXECute method.
Range
"POSitive"
: Positive polarity
"NEGative"
: Negative polarity
"BOTH"
: Both positive polarity and negative polarity
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
The value is ignored.
Preset Value
"POS"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.PPOLarity
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.PPOLarity = "neg"
Equivalent Softkeys
Markers > Marker Search > Peak > Peak Polarity > Positive | Negative | Both
86
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TARGet
Object Type
Property (read/write)
Data Type
Double
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Mk: marker number 1–15, or reference marker number 16 (see Table
4 on page 73)
Description
The target value, when the marker search for target is performed by the
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.EXECute method.
Range
Varies depending on the trace format.
Out of Range
No limitation
Preset Value
0
Unit
dB (decibel) | ° (degree) | s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TARGet
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TARGet = –10
Equivalent Softkeys
Markers > Marker Search > Target > Target Value
87
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TRACking
Object Type
Property (read/write)
Data Type
Boolean
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Mk: marker number 1–15, or reference marker number 16 (see Table
4 on page 73)
Description
The ON/OFF state of the marker search tracking function.
Allowable Values
True: Marker search tracking ON
False: Marker search tracking OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TRACking
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TRACking = true
Equivalent Softkeys
Markers > Marker Search > Tracking
88
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TTRansition
Object Type
Property (read/write)
Data Type
String
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Mk: marker number 1–15, or reference marker number 16 (see Table
4 on page 73)
Description
The selection of the type of the target transition, when the marker search for
transition is performed by the
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.EXECute method.
Range
"POSitive"
: Positive target transition
"NEGative"
: Negative target transition
"BOTH"
: Both positive target transition and negative target transition
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
The value is ignored.
Preset Value
"POS"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TTRansition
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TTRansition = "neg"
Equivalent Softkeys
Marker > Marker Search > Target > Target Transition
89
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TYPE
Object Type
Property (read/write)
Data Type
String
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Mk: marker number 1–15, or reference marker number 16 (see Table
4 on page 73)
Description
The selection of the type of the marker search, which is performed by the
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.EXECute method.
Range
"MAXimum"
"MINimum"
"PEAK"
"LPEak"
"RPEak"
"TARGet"
"LTARget"
"RTARget"
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
The value is ignored.
Preset Value
"MAX"
:
:
:
:
:
:
:
:
Maximum value search
Minimum value search
Peak search
Peak search to the left from the marker
Peak search to the right from the marker
Target search
Target search to the left from the marker
Target search to the right from the marker
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TYPE
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TYPE = "MIN"
Markers > Marker Search > Maximum | Minimum
Equivalent Softkeys
Markers > Marker Search > Peak > Search Peak | Search Max Peak | Search Peak
Left | Search Peak Right
Markers > Marker Search > Target > Search Target | Search Target Left | Search
Target Right
90
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.REFerence.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the reference marker. When the reference marker is
turned ON, all the values of the other markers turn to relative values.
Allowable Values
True: Reference marker ON
False: Reference marker OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.MARKer.REFerence.STATe
app.SCPI.CALCulate(Ch).SELected.MARKer.REFerence.STATe = true
Equivalent Softkeys
Markers > Reference Marker
91
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk)._SET
Object Type
Property (write only)
Data Type
String
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Mk: marker number 1–15, or reference marker number 16 (see Table
4 on page 73)
Description
The setting of the value of the specified item to the value of the position of the
marker.
"STARt" : Sweep start value set to the stimulus value of the marker position.
"STOP"
Range
: Sweep stop value set to the stimulus value of the marker position.
"CENTer" : Sweep center value set to the stimulus value of the marker
position.
"RLEVel" : Reference value set to the response value of the marker position.
"DELay" : Delay value set to the response value of the marker position.
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
The value is ignored.
Syntax
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk)._SET = "STOP"
Equivalent Softkeys
Markers > Marker Functions > Marker–>Start | Marker–>Stop |
Marker –>Center | Marker–>Ref Value | Marker–>Delay
92
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Mk:
marker number 1–15, or reference marker number 16 (see Table
4 on page 73)
Description
The ON/OFF state of a marker. Turning ON a marker with the number from 1
to 15 will turn ON all the markers of smaller numbers. Turning OFF a marker
with the number from 1 to 15 will turn OFF all the markers of greater numbers
(except for the reference marker). Turning ON/OFF the reference marker with
number 16 does not turn ON/OFF the markers with the numbers from 1 to 15,
but switches these markers to the relative measurement mode.
Allowable Values
True: Marker ON
False: Marker OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).STATe
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).STATe = true
Markers > Add Marker | Remove Marker
Equivalent Softkeys
Markers > Reference Marker
93
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).X
Object Type
Property (read/write)
Data Type
Double
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Mk: marker number 1–15, or reference marker number 16 (see Table
4 on page 73)
Description
The stimulus value of the marker.
Range
From the stimulus current start value to the stimulus current stop value.
Out of Value
Sets the value of the limit, which is closer to the specified value.
Preset Value
Stimulus center value
Unit
Hz (Hertz) | s (second) | dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).X
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).X = 1e9
Equivalent Softkeys
Markers > Edit Stimulus
94
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).Y
Object Type
Property (read only)
Data Type
Variant (array of double)
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Mk: marker number 1–15, or reference marker number 16 (see Table
4 on page 73)
The response value of the marker. If the reference marker is turned ON, the
values of the markers from 1 to 15 are read out as relative values to the
reference marker.
Description
The array includes 2 elements:
Data(0)
real number in rectangular format, real part in polar and
Smith chart formats;
Data(1)
0 in rectangular format, imaginary part in polar and Smith
chart formats.
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).Y
Equivalent Softkeys
None
95
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.COUNt
Object Type
Property (read/write)
Data Type
Long
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The number of the turned ON markers.
Range
from 0 to 16
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Dim MarkerCnt As Long
Syntax
MarkerCnt = app.SCPI.CALCulate(Ch).SELected.MARKer.COUNt
app.SCPI.CALCulate(Ch).SELected.MARKer.COUNt = 5
Equivalent Softkeys
None
96
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MATH.FUNCtion
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The selection of the math operation between the measurement data and the
memory trace data. The math result replaces the data trace. If the data trace is
not saved, the command is ignored.
"DIVide"
: Division Data / Mem.
"MULTiply" : Multiplication Data x Mem.
Range
"ADD"
: Addition Data + Mem.
"SUBTract"
: Subtraction Data – Mem.
"NORMal"
: No math
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
An error occurs. Error code 210.
Preset Value
"NORM"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.MATH.FUNCtion
app.SCPI.CALCulate(Ch).SELected.MATH.FUNCtion= "DIV"
Equivalent Softkeys
Display > Data Math > Data/Mem | Data*Mem | Data+Mem | Data–Mem | OFF
97
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MATH.MEMorize
Object Type
Method
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Saves the measurement data to the memory trace. Automatically turns on the
display of the memory trace.
Syntax
app.SCPI.CALCulate(Ch).SELected.MATH.MEMorize
Equivalent Softkeys
Display > Data–>Memory
98
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MSTatistics.DATA
Object Type
Property (read only)
Data Type
Variant (array of double)
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The math statistics data array. The statistics function is applied either over the
whole range (for all the trace), or within the range specified by the
SCPI.CALCulate(Ch).SELected.MSTatistics.DOMain.STATe property (the range
limits are determined by two markers).
Description
The array includes 3 elements:
Data(0)
Mean value;
Data(1)
Standard deviation;
Data(2)
Peak–to–peak (difference between the maximum value
and the minimum value).
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch).SELected.MSTatistics.DATA
Equivalent Softkeys
None
99
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MSTatistics.DOMain.MARKer.STARt
Object Type
Property (read/write)
Data Type
Long
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The number of the marker, which specifies the start frequency of the math
statistics range.
Range
from 1 to 16
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1
Dim MkrNum As Long
Syntax
MkrNum = app.SCPI.CALCulate(Ch).SELected. MSTatistics.DOMain.MARKer.STARt
app.SCPI.CALCulate(Ch).SELected. MSTatistics.DOMain.MARKer.STARt = 3
Equivalent Softkeys
Markers > Marker Math > Statistics > Statistics Start
100
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MSTatistics.DOMain.MARKer.STOP
Object Type
Property (read/write)
Data Type
Long
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The number of the marker, which specifies the stop frequency of the math
statistics range.
Range
from 1 to 16
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
2
Dim MarkerNum As Long
MarkerNum =
Syntax
app.SCPI.CALCulate(Ch).SELected. MSTatistics.DOMain.MARKer.STOP
app.SCPI.CALCulate(Ch).SELected. MSTatistics.DOMain.MARKer.STOP = 4
Equivalent Softkeys
Markers > Marker Math > Statistics > Statistics Stop
101
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MSTatistics.DOMain.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the math statistics range.
Allowable Values
True: Statistics range ON
False: Statistics range OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected. MSTatistics.DOMain.STATe
app.SCPI.CALCulate(Ch).SELected. MSTatistics.DOMain.STATe = true
Equivalent Softkeys
Markers > Marker Math > Statistics > Statistics Range
102
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MSTatistics.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the math statistics display.
Allowable Values
True: Statistics display ON
False: Statistics display OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.MSTatistics.STATe
app.SCPI.CALCulate(Ch).SELected.MSTatistics.STATe = true
Equivalent Softkeys
Markers > Marker Math > Statistics > Statistics
103
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected. RLIMit.DATA
Object Type
Property (read/write)
Data Type
Variant (array of double)
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The data array, which is the limit line for the ripple limit function. The array
size is 1 + 4N, where N is the number of limit line segments.
For the n–th point, where n from 1 to N:
Data(0)
Description
Notes
Data(4n–3)
the number of limit line segments N is the integer from 0
to 12. Setting 0 clears the limit line;
type of the n–th limit line segment;
0: Off
1: On
Data(4n–2)
the stimulus value in the beginning point of the n–th
segment;
Data(4n–2)
the stimulus value in the end point of the n–th segment;
Data(4n–0)
the ripple limit value of the n–th segment.
If the array size is not 1 + 4N, where N is Data(0), an error occurs (error code
214). If Data(4n – 3) is less than 0 or more than 1, an error occurs (error code
214). When Data(4n–2), Data(4n–1), and Data(4n–0) elements are out of
allowable range, the value is set to the limit, which is closer to the specified
value.
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch).SELected.RLIMit.DATA
app.SCPI.CALCulate(Ch).SELected.RLIMit.DATA = Array(1,1,800,900,10)
Equivalent Softkeys
Analysis > Ripple Limit > Edit Ripple Limit
104
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.RLIMit.DISPlay.LINE
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the ripple limit line display.
Allowable Values
True: Ripple limit line ON
False: Ripple limit line OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.RLIMit.DISPlay.LINE
app.SCPI.CALCulate(Ch).SELected.RLIMit.DISPlay.LINE = true
Equivalent Softkeys
Analysis > Ripple Limit > Ripple Limit
105
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.RLIMit.DISPlay.SELect
Object Type
Property (read/write)
Data Type
Long
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The number of the ripple limit test band selected for the ripple value display.
Range
from 1 to 12
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1
Dim Value As Long
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.RLIMit.DISPlay.SELect
app.SCPI.CALCulate(Ch).SELected.RLIMit.DISPlay.SELect = 2
Equivalent Softkeys
Analysis > Ripple Limit > Ripple Value Band
106
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.RLIMit.DISPlay.VALue
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The selection of the display type of the ripple value in the specified band.
"OFF"
: Ripple value display OFF
"ABSolute"
: Absolute value
"MARgin"
: Margin (difference between the ripple limit and the absolute
value)
Range
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
The value is ignored.
Preset Value
"OFF"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.RLIMit.DISPlay.VALue
app.SCPI.CALCulate(Ch).SELected.RLIMit.DISPlay.VALue = "ABS"
Equivalent Softkeys
Analysis > Ripple Limit > Ripple Value
107
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.RLIMit.FAIL
Object Type
Property (read only)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Ripple limit test result.
Allowable Values
True: Fail
False: Pass
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.RLIMit.FAIL
Equivalent Softkeys
None
108
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.RLIMit.REPort.DATA
Object Type
Property (read only)
Data Type
Variant (array of double)
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The data array, which is the ripple limit test results. The array size is 1+3N,
where N is the number of ripple limit bands.
For the n–th point, where n from 1 to N:
Description
Data(0)
N total number of the bands;
Data(3n–2)
n number of the band;
Data(3n–1)
Ripple value in the n–th band;
Data(3n–0)
Ripple limit test result in the n–th band:
0: Pass
1: Fail
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch).SELected.RLIMit.REPort.DATA
Equivalent Softkeys
None
109
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.RLIMit.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the ripple limit test.
Allowable Values
True: Ripple limit test ON
False: Ripple limit test OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.RLIMit.STATe
app.SCPI.CALCulate(Ch).SELected.RLIMit.STATe = true
Equivalent Softkeys
Analysis > Ripple Limit > Ripple Test
110
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.SMOothing.APERture
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The smoothing aperture for the smoothing function.
Range
from 0.01 to 20
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1
Unit
%
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.SMOothing.APERture
app.SCPI.CALCulate(Ch).SELected.SMOothing.APERture = 1.5
Equivalent Softkeys
Average > Smo Aperture
111
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.SMOothing.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the trace smoothing function.
Allowable Values
True: Trace smoothing ON
False: Trace smoothing OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.SMOothing.STATe
app.SCPI.CALCulate(Ch).SELected.SMOothing.STATe = true
Equivalent Softkeys
Average > Smoothing
112
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.CENTer
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Еhe time domain center value, when the time domain transformation function
is turned ON.
Range
Varies depending on the specified frequency range and the number of points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.CENTer
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.CENTer = 1e–8
Equivalent Softkeys
Analysis > Time Domain > Center
113
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.IMPulse.WIDTh
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The impulse width (time domain transformation resolution), coupled with the
Kaiser–Bessel window shape β parameter. The impulse width setting changes
the β parameter, and setting of β parameter changes the impulse width.
Range
Varies depending on the specified frequency range and the number of points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
None
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.IMPulse.WIDTh
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.IMPulse.WIDTh = 1e–8
Equivalent Softkeys
Analysis > Time Domain > Window > Impulse Width (when the transformation type is
set to Bandpass or Lowpass Impulse)
114
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.KBESsel
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The β parameter, which controls the Kaiser–Bessel window shape, when
performing time domain transformation.
Range
from 0 to 13
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
6
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.KBESsel
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.KBESsel = 13
Equivalent Softkeys
Analysis > Time Domain > Window > Kaiser Beta
115
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.LPFRequency
Object Type
Method
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Changes the frequency range to match with the lowpass type of the time
domain transformation function.
Syntax
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.LPFRequency
Equivalent Softkeys
Analysis > Time Domain > Set Frequency Low Pass
116
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.SPAN
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The time domain span value, when the time domain transformation function is
turned ON.
Range
Varies depending on the specified frequency range and the number of points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
2e–8
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.SPAN
app. SCPI.CALCulate(Ch).SELected.TRANsform.TIME.SPAN = 1e–8
Equivalent Softkeys
Analysis > Time Domain > Span
117
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STARt
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The start value used for the transformation function of the time domain
function.
Range
Varies depending on the specified frequency range and the number of points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
–1e–8
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STARt
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STARt = 1e–8
Equivalent Softkeys
Analysis > Time Domain > Start
118
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the time domain transformation function.
Allowable Values
True: Time domain transformation ON
False: Time domain transformation OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STATe
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STATe = true
Equivalent Softkeys
Analysis > Time Domain > Time Domain
119
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STEP.RTIMe
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The rise time of the step signal (time domain transformation resolution),
coupled with the Kaiser–Bessel window shape β parameter. The impulse width
setting changes the β parameter, and setting of β parameter changes the impulse
width.
Range
Varies depending on the specified frequency range and the number of points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
None
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.IMPulse.WIDTh
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.IMPulse.WIDTh = 1e–8
Equivalent Softkeys
Analysis > Time Domain > Window > Impulse Width (when the transformation type is
set to Lowpass Step)
120
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STIMulus
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The selection of the stimulus type for the time domain transformation function:
impulse or step.
Range
"IMPulse"
Impulse
"STEP"
Step
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
The value is ignored.
Preset Value
"IMP"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STIMulus
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STIMulus = "STEP"
Equivalent Softkeys
Analysis > Time Domain > Type > Bandpass | Lowpass Step | Lowpass Impulse
121
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STOP
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The time domain stop value, when the time domain transformation function is
turned ON.
Range
Varies depending on the specified frequency range and the number of points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1e–8
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STOP
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STOP = 2e–8
Equivalent Softkeys
Analysis > Time Domain > Stop
122
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.TYPE
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The selection of the transformation type for the time domain transformation
function: bandpass response or direct current circuit.
Range
"BPASs"
Bandpass
"LPASs"
Lowpass
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
The value is ignored.
Preset Value
"BPASs"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.TYPE
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.TYPE = "STEP"
Equivalent Softkeys
Analysis > Time Domain > Type > Bandpass | Lowpass Step | Lowpass Impulse
123
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).TRACe(Tr).DATA.FDATa
Object Type
Property (read only)
Data Type
Variant (array of double)
Target
The specified trace Tr of channel Ch,
Tr:
trace number 1–16 (see Table 3 on page 32)
Ch:
channel number 1–16 (see Table 1 on page 24)
The formatted data array. The array elements contain measurements in the
current format, for example, in logarithmic magnitude format (Log Mag). Also,
see section “Measurement Data Arrays” on page 19.
The array size is 2N, where N is the number of measurement points.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
real number in rectangular format, real part in polar and
Smith chart formats;
Data(2n–1)
0 in rectangular format, imaginary part in polar and Smith
chart formats.
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch).Trace(Tr).DATA.FDATa
Equivalent Softkeys
None
124
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch).TRACe(Tr).DATA.FMEMory
Object Type
Property (read only)
Data Type
Variant (array of double)
Target
The specified trace Tr of channel Ch,
Tr:
trace number 1–16 (see Table 3 on page 32)
Ch:
channel number 1–16 (see Table 1 on page 24)
The formatted memory array. The array elements contain saved measurements
in the current format, for example, in logarithmic magnitude format (Log Mag).
Also, see section “Measurement Data Arrays” on page 19.
The array size is 2N, where N is the number of measurement points.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
real number in rectangular format, real part in polar and
Smith chart formats;
Data(2n–1)
0 in rectangular format, imaginary part in polar and Smith
chart formats.
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch). Trace(Tr).DATA.FMEMory
Equivalent Softkeys
None
125
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch). TRACe(Tr).DATA.SDATa
Object Type
Property (read only)
Data Type
Variant (array of double)
Target
The specified trace Tr of channel Ch,
Tr:
trace number 1–16 (see Table 3 on page 32)
Ch:
channel number 1–16 (see Table 1 on page 24)
The corrected data array. The corrected measurements are complex numbers.
Also, see section “Measurement Data Arrays” on page 19.
The array size is 2N, where N is the number of measurement points.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
the real part of corrected measurement;
Data(2n–1)
the imaginary part of corrected measurement.
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch). Trace(Tr).DATA.SDATa
Equivalent Softkeys
None
126
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.CALCulate(Ch). TRACe(Tr).DATA.SMEMory
Object Type
Property (read only)
Data Type
Variant (array of double)
Target
The specified trace Tr of channel Ch,
Tr:
trace number 1–16 (see Table 1 on page 32)
Ch:
channel number 1–16 (see Table 1 on page 24)
The corrected memory array. The corrected measurements are complex
numbers. Also, see section “Measurement Data Arrays” on page 19.
The array size is 2N, where N is the number of measurement points.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
the real part of corrected measurement memory;
Data(2n–1)
the imaginary part of corrected measurement memory.
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch). Trace(Tr).DATA.SMEMory
Equivalent Softkeys
None
127
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.COLor.BACK
Object Type
Property (read/write)
Data Type
Variant (array of long)
Target
Instrument
The background color for trace display.
The array contains 3 elements:
Description
Data(0)
Red value R;
Data(1)
Green value G;
Data(2)
Blue value B.
Range
For all the array elements from 0 to 255.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0, 0, 0
Dim Data As Variant
Syntax
Data = app.SCPI.DISPlay.COLor.BACK
app.SCPI.DISPlay.COLor.BACK = Array(0, 0, 0)
Equivalent Softkeys
Display > Properties > Color > Background > Red | Green | Blue
128
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.COLor.GRATicule
Object Type
Property (read/write)
Data Type
Variant (array of long)
Target
Instrument
The grid and the graticule label color for trace display.
The array contains 3 elements:
Description
Data(0)
Red value R;
Data(1)
Green value G;
Data(2)
Blue value B.
Range
For all array elements from 0 to 255.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Data(0)
160;
Data(1)
160;
Data(2)
164.
Dim Data As Variant
Syntax
Data = app.SCPI.DISPlay.COLor.GRATicule
app.SCPI.DISPlay.COLor. GRATicule = Array(128, 128, 128)
Equivalent Softkeys
Display > Properties > Color > Grid > Red | Green | Blue
129
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.COLor.RESet
Object Type
Method
Target
Instrument
Description
Restores the display settings to the default values.
Syntax
app.SCPI.DISPlay.COLor.RESet
Equivalent Softkeys
Display > Properties > Set Defaults
130
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.COLor.TRACe(Tr).DATA
Object Type
Property (read/write)
Data Type
Variant (array of long)
Target
Trace number Tr in all channels,
Tr:
trace number 1–16 (see Table 3 on page 32)
The data trace color.
The array contains 3 elements:
Description
Data(0)
Red value R;
Data(1)
Green value G;
Data(2)
Blue value B.
Range
For all array elements from 0 to 255.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the trace number.
Dim Data As Variant
Syntax
Data = app.SCPI.DISPlay.COLor.TRACe(Tr).DATA
app.SCPI.DISPlay.COLor.TRACe(Tr).DATA = Array(255, 255, 0)
Equivalent Softkeys
Display > Properties > Color > Data Trace > Red | Green | Blue
131
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.COLor.TRACe(Tr).MEMory
Object Type
Property (read/write)
Data Type
Variant (array of long)
Target
Trace number Tr in all channels
Tr:
trace number 1–16 (see Table 3 on page 32)
The memory trace color.
The array contains 3 elements:
Description
Data(0)
Red value R;
Data(1)
Green value G;
Data(2)
Blue value B.
Range
For all array elements from 0 to 255.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the trace number.
Dim Data As Variant
Syntax
Data = app.SCPI.DISPlay.COLor.TRACe(Tr).MEMory
app.SCPI.DISPlay.COLor.TRACe(Tr). MEMory = Array(255, 255, 0)
Equivalent Softkeys
Display > Properties > Color > Data Trace > Red | Green | Blue
132
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.ENABle
Object Type
Property (read/write)
Data Type
Boolean
Target
Instrument
Description
The ON/OFF state of the display update function.
Allowable Values
True: Display update ON
False: Display update OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI.DISPlay.ENABle
app.SCPI.DISPlay.ENABle = true
Equivalent Softkeys
Display > Update
133
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.FSIGn
Object Type
Property (read/write)
Data Type
Boolean
Target
Instrument
Description
The ON/OFF state of the Fail sign display, when performing limit test or ripple
limit test.
Allowable Values
True: Fail sign display ON
False: Fail sign display OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.DISPlay.FSIGn
app.SCPI.DISPlay.FSIGn = true
Analysis > Limit Test > Fail Sign
Equivalent Softkeys
Analysis > Ripple Limit > Fail Sign
134
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.IMAGe
Object Type
Property (read/write)
Data Type
String
Target
Instrument
Description
The inverted color display of the data traces.
Range
"NORMal"
: Normal display
"INVert"
: Inverted color display
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
The value is ignored.
Preset Value
"NORM"
Dim Param As String
Syntax
Param = app.SCPI.DISPlay.IMAGe
app.SCPI.DISPlay.IMAGe = "INV"
Equivalent Softkeys
Display > Properties > Invert Color
135
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.SPLit
Object Type
Property (read/write)
Data Type
Long
Target
Instrument
Description
The number and layout of the channel windows on the screen. The channel
window layout is in Table 5 below.
Range
from 1 to 16
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1
Dim Value As Long
Syntax
Value = app.SCPI.DISPlay.SPLit
app.SCPI.DISPlay.SPLit = 2
Equivalent Softkeys
Display > Allocate channels
Table 5. Channel Window Layout on the Screen
1:
2:
3:
4:
5:
6:
7:
8:
9:
10:
11:
12:
13:
14:
15:
16:
136
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay._UPDate.IMMediate
Object Type
Method
Target
Instrument
Description
Updates the display once, when the display update is set to OFF
(SCPI.DISPlay.ENABle property is set to False).
Syntax
app.SCPI.DISPlay._UPDate.IMMediate
Equivalent Softkeys
None
SCPI.DISPlay.WINDow(Ch).ACTivate
Object Type
Method
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Sets the active channel.
Notes
The channel window must be displayed. At attempt to set to the active channel
the channel, which is not displayed, an error occurs.
Syntax
app.SCPI.DISPlay.WINDow(Ch).ACTivate
Equivalent Softkeys
Display > Active Trace / Channel > Active Channel
137
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.ALIGn.TYPE
Object Type
Property (read/write)
Data Type
String
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The alignment mode of the marker display position of each trace, when the
only
active
trace
display
feature
is
turned
OFF
(SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.SINGle.STATe property is set to
False).
Range
"VERTical"
: Vertical alignment
"HORizontal"
: Horizontal alignment
"NONE"
: No alignment
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
The value is ignored.
Preset Value
"NONE"
Dim Param As String
Syntax
Param = app.SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.ALIGn.TYPE
app.SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.ALIGn.TYPE = "VERT"
Equivalent Softkeys
Markers > Properties > Align > Vertical | Horizontal | OFF
138
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.SINGle.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the marker display for the active trace only.
Allowable Values
True: Only active trace markers display ON
False: Only active trace markers display OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.SINGle.STATe
app.SCPI SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.SINGle.STATe = true
Equivalent Softkeys
Markers > Properties > Active Only
139
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).SPLit
Object Type
Property (read/write)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The code of the graph layout of channels. The channel window layout is in
Table 5 on page 132.
Range
from 1 to 16
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1
Notes
This property does not define the number of traces in the channel window, the
number of traces is defined by the SCPI.CALCulate(Ch).PARameter.COUNt
property.
Dim Value As Long
Syntax
Value = app.SCPI.DISPlay.WINDow(Ch).SPLit
app.SCPI.DISPlay.WINDow(Ch).SPLit = 2
Equivalent Softkeys
Display > Allocate Traces
140
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TITLe.DATA
Object Type
Property (read/write)
Data Type
String
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The channel title label.
Range
up to 254 characters
Preset Value
""
Dim Text As String
Syntax
Text = app.SCPI.DISPlay.WINDow(Ch).TITLe.DATA
app.SCPI.DISPlay.WINDow(Ch).TITLe.DATA = "Example1"
Equivalent Softkeys
Display > Edit Title Label
141
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TITLe.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the title label display.
Allowable Values
True: Title label display ON
False: Title label display OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.DISPlay.WINDow(Ch).TITLe.STATe
app.SCPI.DISPlay.WINDow(Ch).TITLe.STATe = true
Equivalent Softkeys
Display > Title Label
142
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).ANNotation.MARKer.
POSition.X
Object Type
Property (read/write)
Data Type
Double
Target
Trace Tr of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Tr:
trace number 1–16 (see Table 3 on page 32)
Description
The display position of the marker value on the X–axis by a percentage of the
display width.
Range
from 0 to 100
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
%
Dim Value As Double
Syntax
Value = app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).ANNotation.MARKer.POSition.X
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).ANNotation.MARKer.POSition.X = 50
Equivalent Softkeys
Markers > Properties > Data X Position
143
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).ANNotation.MARKer.
POSition.Y
Object Type
Property (read/write)
Data Type
Double
Target
Trace Tr of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Tr:
trace number 1–16 (see Table 3 on page 32)
Description
The display position of the marker value on the Y–axis by a percentage of the
display height.
Range
from 0 to 100
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
%
Dim Value As Double
Syntax
Value = app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).ANNotation.MARKer.POSition.Y
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).ANNotation.MARKer.POSition.Y = 50
Equivalent Softkeys
Markers > Properties > Data Y Position
144
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).MEMory. STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Trace Tr of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Tr:
trace number 1–16 (see Table 3 on page 32)
Description
The ON/OFF state of the memory trace display.
Allowable Values
True: Memory trace display ON
False: Memory trace display OFF
Preset Value
False
Notes
If the memory is empty, an error occurs and the object is ignored.
Dim Status As Boolean
Syntax
Status = app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).MEMory.STATe
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).MEMory.STATe = true
Equivalent Softkeys
Display > Display > Memory | Data & Memory (ON)
Display > Display > Data | OFF (OFF)
145
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Trace Tr of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Tr:
trace number 1–16 (see Table 3 on page 32)
Description
The ON/OFF state of the data trace display.
Allowable Values
True: Data trace display ON
False: Data trace display OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).STATe
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).STATe = false
Equivalent Softkeys
Display > Display > Data | Data & Memory (ON)
Display > Display > Memory | OFF (OFF)
146
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.AUTO
Object Type
Method
Target
Trace Tr of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Tr:
trace number 1–16 (see Table 3 on page 32)
Description
Executes the auto scale function for the trace.
Syntax
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.AUTO
Equivalent Softkeys
Scale > Auto Scale
147
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.PDIVision
Object Type
Property (read/write)
Data Type
Double
Target
Trace Tr of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Tr:
trace number 1–16 (see Table 3 on page 32)
Description
The trace scale. Sets the scale per division, when the data format is the
rectangular format. Sets the full scale value, when the data format is the Smith
chart format or the polar format.
Range
from 10E–18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the format.
Logarithmic Magnitude: 10 dB/Div
Phase: 40 °/Div
Expand Phase: 100 °/Div
Group Delay: 10e–9 s/Div
Smith Chart, Polar, SWR: 1 /Div
Linear Magnitude: 0.1 /Div
Real part, Imaginary part: 0.2 /Div
Unit
dB/Div (decibel per division), °/Div (degree per division), s/Div (second per
division)
Dim Value As Double
Syntax
Value = app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.PDIVision
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.PDIVision = 20
Equivalent Softkeys
Scale > Scale
148
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.RLEVel
Object Type
Property (read/write)
Data Type
Double
Target
Trace Tr of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Tr:
trace number 1–16 (see Table 3 on page 32)
Description
The value of the reference line (response value on the reference line). For the
rectangular format only.
Range
from –1E–18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0 (except for SWR: 1)
Unit
dB (decibel) | ° (degree) | s (second)
Dim Value As Double
Syntax
Value = app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.RLEVel
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.RLEVel = 10
Equivalent Softkeys
Scale > Ref Value
149
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.RPOSition
Object Type
Property (read/write)
Data Type
Long
Target
Trace Tr of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Tr:
trace number 1–16 (see Table 3 on page 32)
Description
The position of the reference line. For the rectangular format only.
Range
From
0
to
the
number
of
the
scale divisions (set
SCPI.DISPlay.WINDow(Ch).Y.SCALe.DIVisions property, 10 by default).
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
5 (except for SWR: 0)
by
Dim Value As Long
Syntax
Value = app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.RPOSition
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.RPOSition = 10
Equivalent Softkeys
Scale > Ref Position
150
the
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).X.SPACing
Object Type
Property (read/write)
Data Type
String
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24).
Description
The selection of the display method of the graph horizontal axis for the
segment sweep.
Range
"LINear"
: Frequency base (linear frequency axis)
"OBASe"
: Order base (linear axis of the point numbers)
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
The value is ignored.
Preset Value
"LIN"
Dim Param As String
Syntax
Param = app.SCPI.DISPlay.WINDow(Ch).X.SPACing
app.SCPI.DISPlay.WINDow(Ch).X.SPACing = "OBAS"
Equivalent Softkeys
Stimulus > Segment Table > Segment Display
151
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).Y.SCALe.DIVisions
Object Type
Property (read/write)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24).
Description
The number of the vertical scale divisions. For the rectangular format only.
Range
from 4 to 30
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
10
Resolution
2
Dim Value As Long
Syntax
Value = app.SCPI.DISPlay.WINDow(Ch).Y.SCALe.DIVisions
app.SCPI.DISPlay.WINDow(Ch).Y.SCALe.DIVisions = 12
Equivalent Softkeys
Scale > Divisions
152
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.HCOPy.ABORt
Object Type
Method
Target
Instrument
Description
Aborts the printout.
Syntax
app.SCPI.HCOPy.ABORt
Equivalent Softkeys
None
SCPI.HCOPy.DATE.STAMp
Object Type
Property (read/write)
Data Type
Boolean
Target
Instrument
Description
The ON/OFF state of the current date and time printout in the upper right
corner.
Allowable Values
True: Date & time printout ON
False: Date & time printout OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI.HCOPy.DATE.STAMp
app.SCPI.HCOPy.DATE.STAMp = False
Equivalent Softkeys
System > Print > Print Date & Time
153
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.HCOPy.IMAGe
Object Type
Property (read/write)
Data Type
String
Target
Instrument
Description
The inverted color image printout.
Range
"NORMal"
: Normal printout
"INVert"
: Inverted color printout
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
The value is ignored.
Preset Value
"NORM"
Dim Param As String
Syntax
Param = app.SCPI.HCOPy.IMAGe
app.SCPI.HCOPy.IMAGe = "INV"
Equivalent Softkeys
System > Print > Invert Image
154
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.HCOPy.IMMediate
Object Type
Method
Target
Instrument
Description
Prints out the image displayed on the screen without previewing.
Syntax
app.SCPI.HCOPy.IMMediate
Equivalent Softkeys
System > Print > Print Embedded
155
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.HCOPy.PAINt
Object Type
Property (read/write)
Data Type
String
Target
Instrument
Description
The color chart for the image printout.
Range
"COLor"
: Color printout
"GRAY"
: Grayscale printout
"BW"
: Black&white printout
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
The value is ignored.
Preset Value
"BW"
Dim Param As String
Syntax
Param = app.SCPI.HCOPy.PAINt
app.SCPI.HCOPy.PAINt = "COL"
Equivalent Softkeys
System > Print > Print Color
156
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.IEEE4882.CLS
Object Type
Method
Target
Status Reporting System
Description
Clears the following:
• Error Queue
• Status Byte Register
• Standard Event Status Register
• Operation Status Event Register
• Questionable Status Event Register
• Questionable Limit Status Event Register
• Questionable Limit Channel Status Event Register
Syntax
app.SCPI.IEEE4882.CLS
Equivalent Softkeys
None
SCPI.IEEE4882.IDN
Object Type
Property (read only)
Data Type
String
Target
Instrument
Description
The instrument information string. The string format: ''{manufacturer},
{model}, {serial number}, {software version/firmware version}''.
Range
up to 40 characters
Dim ID As String
Syntax
ID = app.NAME
Equivalent Softkeys
None
157
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.IEEE4882.RST
Object Type
Method
Target
Instrument
Description
Restores the default settings of the instrument. There is difference from
presetting the instrument with the SCPI.SYSTem.PRESet method – in this case
the trigger mode is set to Hold.
Syntax
app.SCPI.IEEE4882.RST
Equivalent Softkeys
None
SCPI.IEEE4882.TRG
Object Type
Method
Target
Instrument
Description
If the trigger source is set to GPIB/LAN (SCPI.TRIGger.SEQuence.SOURce
property is set to ''BUS'' ), triggers a sweep.
If the trigger source is not set to the bus (SCPI.TRIGger.SEQuence.SOURce
property is not set to ''BUS'' ) or the instrument is not waiting for a trigger, the
method is ignored.
Syntax
app.SCPI.IEEE4882.TRG
Related Commands
SCPI.TRIGger.SEQuence.SOURce
SCPI.INITiate(Ch).CONTinuous
SCPI.INITiate(Ch).IMMediate
Equivalent Softkeys
None
158
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.IEEE4882.WAI
Object Type
Method
Target
Instrument
Description
Waits for the execution of all commands sent before this command.
Syntax
app.SCPI.IEEE4882.WAI
Equivalent Softkeys
None
159
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.INITiate(Ch).CONTinuous
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the continuous trigger initiation mode. If the continuous
trigger initiation mode is set to OFF, the channel turns to the hold state.
Allowable Values
True: Continuous trigger initiation mode ON
False: Continuous trigger initiation mode OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI.INITiate(Ch).CONTinuous
app.SCPI.INITiate(Ch).CONTinuous = False
Notes
The sweep start in continuous trigger initiation mode depends on the trigger
source. If the trigger is set to internal, the sweeps will go immediately one after
another. If the trigger is set otherwise, the sweep will start when the trigger
signal is received.
Equivalent Softkeys
Stimulus > Trigger > Continuous
Stimulus > Trigger > Hold
160
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.INITiate(Ch).IMMediate
Object Type
Method
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Sets the channel to the single trigger mode. Before this method is called, the
channel must be in hold state, otherwise an error occurs (error code 213) and
the method is ignored.
On completion of the sweep, the channel goes back into the hold state.
The method returns control before the end of the sweep.
Syntax
app.SCPI.INITiate(Ch).IMMediate
Notes
The sweep start in the single trigger mode depends on the trigger source. If the
trigger is set to internal, the sweep will start immediately after the method is
called. If the trigger is set otherwise, the sweep will start when the trigger
signal is received.
Equivalent Softkeys
Stimulus > Trigger > Single
161
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.COPY(Src, Dst)
Object Type
Method
Target
Instrument
Description
Copies a file.
Syntax
app.SCPI.MMEMory.COPY(Src, Dst)
Src – Source file name. String data type.
Parameter
Dst – Destination file name. String data type.
Equivalent Softkeys
None
SCPI.MMEMory.DELete(File)
Object Type
Method
Target
Instrument
Description
Deletes a file.
Syntax
app.SCPI.MMEMory.DELete(File)
Parameter
File – File name. String data type.
Equivalent Softkeys
None
162
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.LOAD.CHANnel.STATe
Object Type
Property (write only)
Data Type
String
Target
Active channel
Description
Recalls the instrument state for the active channel, saved in one of the four
memory registers by the SCPI.MMEMory.STORe.CHANnel.STATe property.
Range
"A"
: Recall from register A
"B"
: Recall from register B
"C"
: Recall from register C
"D"
: Recall from register D
Out of Range
The value is ignored.
Syntax
app.SCPI.MMEMory.LOAD.CHANnel.STATe = "A"
Equivalent Softkeys
Save/Recall > Recall Channel > State A | B | C | D
163
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.LOAD.CKIT(Ck)
Object Type
Property (write only)
Data Type
String
Target
Calibration kit Ck,
Ck:
calibration kit number 1–11 (see Table 6 on page 164)
Description
Recalls the definition file for the calibration kit. The file must be saved by the
SCPI.MMEMory.STORe.CKIT(Ck) property.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.LOAD.CKIT(Ck) = File
Notes
If the full path to the file is not specified, the \CalKit subdirectory of the main
directory will be searched for the file. The calibration kit definition file has
*.ckd extension by default.
Equivalent Softkeys
None
Table 6. Ck: Calibration Kit Number
Data Type
Long
Description
Calibration kit number.
Range
from 1 to 11
Out of Range
An error occurs. Error code: 114.
Notes
If the calibration kit number is not specified, it is taken as equal to 1.
164
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.LOAD.LIMit
Object Type
Property (write only)
Data Type
String
Target
Active trace of the active channel.
Description
Recalls the specified limit table file. The file must be saved by the
SCPI.MMEMory.STORe.LIMit property.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.LOAD.LIMit = File
Notes
If the full path to the file is not specified, the \Limit subdirectory of the main
directory will be searched for the file. The limit table files have *.lim extension
by default.
Equivalent Softkeys
Analysis > Limit Test > Edit Limit Line > Restore Limit Table
165
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.LOAD.PLOSs(Pt)
Object Type
Property (write only)
Data Type
String
Target
Port Pt of the active channel,
Pt:
port number 1–2 (see Table 2 on page 24).
Description
Recalls the specified loss compensation table file. The file must be saved by the
SCPI.MMEMory.STORe.PLOSs(Pt) property.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.LOAD.PLOSs(Pt) = File
Notes
If the full path to the file is not specified, the \CalKit subdirectory of the main
directory will be searched for the file. The loss compensation file has *.lct
extension by default.
Equivalent Softkeys
Calibration > Power Calibration > Loss Compen > Import Loss Table
166
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.LOAD.RLIMit
Object Type
Property (write only)
Data Type
String
Target
Active trace of the active channel.
Description
Recalls the ripple limit table file. The file must be saved by the
SCPI.MMEMory.STORe.RLIMit property.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.LOAD.RLIMit = File
Notes
If the full path to the file is not specified, the \Limit subdirectory of the main
directory will be searched for the file. The ripple limit files have *.rlm
extension by default.
Equivalent Softkeys
Analysis > Ripple Limit > Edit Ripple Limit > Restore Ripple Limit Table
167
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.LOAD.SEGMent
Object Type
Property (write only)
Data Type
String
Target
Active channel
Description
Recalls the segment table file. The file must be saved by the
SCPI.MMEMory.STORe.SEGMent property.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.LOAD.SEGMent = File
Notes
If the full path to the file is not specified, the \Segment subdirectory of the main
directory will be searched for the file. The segment files have *.seg extension
by default.
Equivalent Softkeys
Stimulus > Segment Table > Recall…
168
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.LOAD.STATe
Object Type
Property (write only)
Data Type
String
Target
Instrument
Description
Recalls the specified instrument state file. The file must be saved by the
SCPI.MMEMory.STORe.STATe property.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.LOAD.STATe = File
Notes
If the full path to the file is not specified, the \State subdirectory of the main
directory will be searched for the file. The instrument state files have *.sta
extension by default.
Equivalent Softkeys
Save/Recall > Recall State > State…
169
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.MDIRectory
Object Type
Property (write only)
Data Type
String
Target
Instrument
Description
Creates a new directory (folder). Contains the full path to the folder being
created.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.MDIRectory = Path
Equivalent Softkeys
None
170
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.STORe.CHANnel.CLEar
Object Type
Method
Target
Instrument
Description
Clears the memory of the channel state saved by the
SCPI.MMEMory.STORe.CHANnel.STATe property.
Syntax
app.SCPI.MMEMory.STORe.CHANnel.CLEar
Equivalent Softkeys
Save/Recall > Save Channel > Clear States
SCPI.MMEMory.STORe.CHANnel.STATe
Object Type
Property (write only)
Data Type
String
Target
Active channel
Description
Saves the instrument state of the items set for the active channel into one of the
four memory registers.
Range
"A"
: Save to register A
"B"
: Save to register B
"C"
: Save to register C
"D"
: Save to register D
Out of Range
The value is ignored.
Syntax
app.SCPI.MMEMory.STORe.CHANnel.STATe = "A"
Equivalent Softkeys
Save/Recall > Save Channel > State A | B | C | D
171
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.STORe.CKIT(Ck)
Object Type
Property (write only)
Data Type
String
Target
Calibration kit Ck,
Ck:
calibration kit number 1–11 (see Table 6 on page 164)
Description
Saves the definition file for the calibration kit parameters.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.CKIT(Ck) = File
Notes
If the full path to the file is not specified, the file will be saved to the \CalKit
subdirectory of the main directory. The calibration kit definition file has *.ckd
extension by default.
Equivalent Softkeys
None
172
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.STORe.FDATa
Object Type
Property (write only)
Data Type
String
Target
Active trace of the active channel
Description
Saves the CSV formatted data into a file.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.FDATa = File
Notes
If the full path to the file is not specified, the file will be saved to the \CSV
subdirectory of the main directory. The files have *.csv extension by default.
Equivalent Softkeys
Save/Recall > Save Trace Data
173
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.STORe.IMAGe
Object Type
Property (write only)
Data Type
String
Target
Instrument
Description
Saves the display image in BMP or PNG format into a file.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.IMAGe = File
Notes
If the full path to the file is not specified, the file will be saved to the \Image
subdirectory of the main directory. If the file has *.png extension, the file has
PNG format, in all the other cases the file has BMP format.
Equivalent Softkeys
System > Print > Print Windows > Save as…
174
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.STORe.LIMit
Object Type
Property (write only)
Data Type
String
Target
Active trace of the active channel
Description
Saves the limit table into a file with the specified name.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.LIMit = File
Notes
If the full path to the file is not specified, the file will be saved to the \Limit
subdirectory of the main directory. The files have *.lim extension by default.
Equivalent Softkeys
Analysis > Limit Test > Edit Limit Line > Save Limit Table
175
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.STORe.PLOSs(Pt)
Object Type
Property (write only)
Data Type
String
Target
Port Pt of the active channel,
Pt:
port number 1–2 (see Table 2 on page 24)
Description
Saves the loss compensation table into a file with the specified name.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.PLOSs(Pt) = File
Notes
If the full path to the file is not specified, the file will be saved to the \CalKit
subdirectory of the main directory. The loss compensation files have *.lct
extension by default.
Equivalent Softkeys
Calibration > Power Calibration > Loss Compen > Export Loss Table
176
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.STORe.RLIMit
Object Type
Property (write only)
Data Type
String
Target
Active trace of the active channel
Description
Saves the ripple limit table into a file with the specified name.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.RLIMit = File
Notes
If the full path to the file is not specified, the file will be saved to the \Limit
subdirectory of the main directory. The ripple limit files have *.rlm extension
by default.
Equivalent Softkeys
Analysis > Ripple Limit > Edit Ripple Limit > Save Ripple Limit Table
177
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.STORe.SEGMent
Object Type
Property (write only)
Data Type
String
Target
Active channel
Description
Saves the segment table in a file with the specified name.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.SEGMent = File
Notes
If the full path to the file is not specified, the file will be saved to the \Segment
subdirectory of the main directory. The segment files have *.seg extension by
default.
Equivalent Softkeys
Stimulus > Segment Table > Save…
178
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.STORe.SNP.DATA
Object Type
Property (write only)
Data Type
String
Target
Active channel
Description
Saves the measured S–parameters of the active channel into a Touchstone file
with the specified name. The file type (1–port or 2–port) is set by the
SCPI.MMEMory.STORe.SNP.TYPE.S1P and SCPI.MMEMory.STORe.SNP.TYPE.S2P
properties. 1–port type file saves one reflection parameter: S11 or S22. 2–port
type file saves all the four parameters: S11, S21, S12, S22.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.SNP.DATA = File
Notes
If the full path to the file is not specified, the file will be saved to the
\FixtureSim subdirectory of the main directory. The 1–port measurement files
have *.slp extension; the 2–port measurement files have *.s2p extension.
Equivalent Softkeys
Save/Recall > Save Data to Touchstone File > Save File…
179
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.STORe.SNP.FORMat
Object Type
Property (read/write)
Data Type
String
Target
Active channel
Description
The data format for the S–parameters saving by the
SCPI.MMEMory.STORe.SNP.DATA property.
Range
" MA"
: Logarithmic Magnitude / Angle format
" DB"
: Linear Magnitude / Angle format
" RI"
: Real part /Imaginary part format
Out of Range
The value is ignored.
Preset Value
"RI"
Dim Param As String
Syntax
Param = app.SCPI.MMEMory.STORe.SNP.FORMat
app.SCPI.MMEMory.STORe.SNP.FORMat = "DB"
Equivalent Softkeys
Save/Recall > Save Data to Touchstone File > Format
180
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.STORe.SNP.TYPE.S1P
Object Type
Property (read/write)
Data Type
Long
Target
Active channel
Description
Selects 1-port measurements and sets port numbers (selects S11 or S22
parameter) when saving the measured S-parameters by the
SCPI.MMEMory.STORe.SNP.DATA property.
Range
from 1 to 2
Out of Range
An error occurs. Error code 222.
Preset Value
1
Dim Value As Long
Syntax
Value = app.SCPI.MMEMory.STORe.SNP.TYPE.S1P
app.SCPI.MMEMory.STORe.SNP.TYPE.S1P = 2
Save/Recall > Save Data to Touchstone File > Type > 1–Port (s1p)
Equivalent Softkeys
Save/Recall > Save Data to Touchstone File > Select Port
181
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.STORe.SNP.TYPE.S2P
Object Type
Property (read/write)
Data Type
Variant (array of long)
Target
Active channel
Selects 2-port measurements and sets port numbers when saving the measured
S-parameters by the SCPI.MMEMory.STORe.SNP.DATA property.
Description
The array contains 2 elements:
Data(0)
First port number (reads out 1, ignored when written);
Data(1)
Second port number (reads out 2, ignored when written).
Dim Data As Variant
Syntax
Data = app.SCPI.MMEMory.STORe.SNP.TYPE.S2P
app.SCPI.MMEMory.STORe.SNP.TYPE.S2P = Array(1, 2)
Equivalent Softkeys
Save/Recall > Save Data to Touchstone File > Type > 2–Port (s2p)
182
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.STORe.STATe
Object Type
Property (write only)
Data Type
String
Target
Instrument
Description
Saves the instrument state into a file with the specified name.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.STATe = File
Notes
If the full path to the file is not specified, the file will be saved to the \State
subdirectory of the main directory. The state files have *.sta extension by
default.
Equivalent Softkeys
Save/Recall > Save State > State…
183
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.MMEMory.STORe.STYPe
Object Type
Property (read/write)
Data Type
String
Target
Instrument
Description
Selects the type of the instrument or channel state saving by the
SCPI.MMEMory.STORe.STATe
or
SCPI.MMEMory.STORe.CHANnel.STATe
property.
Range
"STATe"
: Measurement conditions
"CSTate"
: Measurement conditions and calibration tables
"DSTate"
: Measurement conditions and data traces
"CDSTate" : Measurement conditions, calibration tables and data traces
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
An error occurs. Error code 205.
Preset Value
"CST"
Dim Param As String
Syntax
Param = app.SCPI.MMEMory.STORe.STYPe
app.SCPI.MMEMory.STORe.STYPe = "STATe"
Equivalent Softkeys
Save/Recall > Save Type
184
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.OUTPut.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Instrument
Description
Sets the ON/OFF state of the stimulus signal output. Measurements cannot be
performed when the stimulus signal output is set to OFF.
Allowable Values
True: Stimulus signal output ON
False: Stimulus signal output OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI.OUTPut.STATe
app.SCPI.OUTPut.STATe = False
Equivalent Softkeys
Stimulus > Power > RF Out
185
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).AVERage.CLEar
Object Type
Method
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Resets the averaging data count to 0. Restarts the averaging process.
Syntax
app.SCPI.SENSe(Ch).AVERage.CLEar
Equivalent Softkeys
None
186
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).AVERage.COUNt
Object Type
Property (read/write)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The averaging factor, when the averaging function is set to ON by the
SCPI.SENSe(Ch).AVERage.STATe property.
Range
from 1 to 999
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
10
Dim Value As Long
Syntax
Value = app.SCPI.SENSe(Ch).AVERage.COUNt
app.SCPI.SENSe(Ch).AVERage.COUNt = 2
Equivalent Softkeys
Average > Avg Factor
187
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).AVERage.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the averaging function.
Allowable Values
True: Averaging ON
False: Averaging OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.SENSe(Ch).AVERage.STATe
app.SCPI.SENSe(Ch).AVERage.STATe = False
Equivalent Softkeys
Average > Averaging
188
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).BANDwidth.RESolution
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The IF bandwidth value.
Range
from 1 to 30000
Resolution
In steps of 1, 1.5, 2, 3, 5, 7.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
10000
Unit
Hz (Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).BANDwidth.RESolution
app.SCPI.SENSe(Ch).BANDwidth.RESolution = 100
Equivalent Softkeys
Average > IF Bandwidth
189
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.CLEar
Object Type
Method
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Clears the calibration coefficient table.
Syntax
app.SCPI.SENSe(Ch).CORRection.CLEar
Equivalent Softkeys
None
190
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COEFficient.DATA(Str, Pt_r, Pt_s)
Object Type
Property (read/write)
Data Type
Variant (array of double)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The calibration coefficient data array set by the type of the corrected error Str,
the number of the receiver port Pt_r and the number of the source port Pt_s,
Str
: error type (see below)
Pt_r : the number of the receiver port 1–2 (see Table 2 on page 24)
Pt_s : the number of the source port 1–2 (see Table 2 on page 24)
The array size is 2N, where N is the number of measurement points.
For the n–th point, where n from 1 to N:
Parameter
Data(2n–2)
real part of the calibration coefficients
Data(2n–1)
imaginary part of the calibration coefficients
String Str – corrected error type:
"ES": Source match
"ER": Reflection tracking
"ED": Directivity
"EL": Load match
"ET": Transmission tracking
"EX": Isolation
When ES, ER, or ED is used, the numbers of the ports Pt_r and Pt_s must be
the same. When EL, ET, or EX is used, the numbers of the ports Pt_r and Pt_s
must be different.
Dim Data As Variant
Syntax
Data = app.SCPI.SENSe(Ch).CORRection.COEFficient.DATA (Str, Pt_r, Pt_s)
app.SCPI.SENSe(Ch).CORRection.COEFficient.DATA (Str, Pt_r, Pt_s) = Data
Notes
The written calibration coefficients become effective only after the
SCPI.SENSe(Ch).CORRection.COEFficient.SAVE method is invoked.
Equivalent Softkeys
None
191
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.ERESponse
Object Type
Property (write only)
Data Type
Variant (array of long)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Selects the port numbers and sets the 1–path 2–port calibration type, when the
written
calibration
coefficients
are
made
effective
by
the
SCPI.SENSe(Ch).CORRection.COEFficient.SAVE method.
Description
The array contains 2 elements:
Data(0)
the number of the receiver port;
Data(1)
the number of the source port.
Range
Port number is 1 or 2. Array elements can not contain the same port numbers.
Out of Range
If an incorrect port number is specified, an error occurs (error code: 222). If the
same port numbers are specified, an error occurs (error code: 220).
Dim Ports As Variant
Syntax
Ports = Array(2, 1)
app.SCPI.SENSe(Ch).CORRection.COEFficient.METHod.ERESponse = Ports
Equivalent Softkeys
None
192
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.RESPonse.OPEN
Object Type
Property (write only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Selects the port and sets the response calibration (Open) type, when the written
calibration
coefficients
are
made
effective
by
the
SCPI.SENSe(Ch).CORRection.COEFficient.SAVE method.
Range
from 1 to 2
Out of Range
An error occurs. Error code: 222.
Dim Port As Long
Syntax
Port = 1
app.SCPI.SENSe(Ch).CORRection.COEFficient.METHod.RESPonse.OPEN = Port
Equivalent Softkeys
None
193
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.RESPonse.SHORt
Object Type
Property (write only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Selects the port and sets the response calibration (Short) type, when the written
calibration
coefficients
are
made
effective
by
the
SCPI.SENSe(Ch).CORRection.COEFficient.SAVE method.
Range
from 1 to 2
Out of Range
An error occurs. Error code: 222.
Dim Port As Long
Syntax
Port = 1
app.SCPI.SENSe(Ch).CORRection.COEFficient.METHod.RESPonse.SHORt = Port
Equivalent Softkeys
None
194
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.RESPonse.THRU
Object Type
Property (write only)
Data Type
Variant (array of long)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Selects the ports and sets the response calibration (Thru) type, when the
written
calibration
coefficients
are
made
effective
by
the
SCPI.SENSe(Ch).CORRection.COEFficient.SAVE method.
Description
The array contains 2 elements:
Data(0)
the number of the receiver port;
Data(1)
the number of the source port.
Range
Port number is 1 or 2. The array elements can not contain the same port
numbers.
Out of Range
If an incorrect port number is specified, an error occurs (error code: 222). If the
same port numbers are specified, an error occurs (error code: 220).
Dim Ports As Variant
Syntax
Ports = Array(2, 1)
app.SCPI.SENSe(Ch).CORRection.COEFficient.METHod.RESPonse.THRU = Ports
Equivalent Softkeys
None
195
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.SOLT1
Object Type
Property (write only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Selects the port and sets the full 1–port calibration type, when the written
calibration
coefficients
are
made
effective
by
the
SCPI.SENSe(Ch).CORRection.COEFficient.SAVE method.
Range
from 1 to 2
Out of Range
An error occurs. Error code: 222.
Dim Port As Long
Syntax
Port = 1
app.SCPI.SENSe(Ch).CORRection.COEFficient.METHod.SOLT1= Port
Equivalent Softkeys
None
196
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.SOLT2
Object Type
Property (write only)
Data Type
Variant (array of long)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Selects the ports and sets the full 2–port calibration type, when the written
calibration
coefficients
are
made
effective
by
the
SCPI.SENSe(Ch).CORRection.COEFficient.SAVE method.
Description
The array contains 2 elements:
Data(0)
the number of the receiver port;
Data(1)
the number of the source port.
Range
Port number is 1 or 2. The array elements can not contain the same port
numbers.
Out of Range
If an incorrect port number is specified, an error occurs (error code: 222). If the
same port numbers are specified, an error occurs (error code: 220).
Dim Ports As Variant
Syntax
Ports = Array(2, 1)
app.SCPI.SENSe(Ch).CORRection.COEFficient.METHod.SOLT2 = Ports
Equivalent Softkeys
None
197
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COEFficient.SAVE
Object Type
Method
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Enables the written calibration coefficients depending on the selected
calibration type.
Description
On completion of the method the written calibration coefficients are cleared,
the error correction automatically turns ON.
At the attempt to execute this method before all the needed calibration
coefficients are written, an error occurs and the method is ignored.
Syntax
app.SCPI.SENSe(Ch).CORRection.COEFficient.SAVE
Calibration type selection:
Related Commands
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.ERESponse
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.RESPonse.OPEN
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.RESPonse.SHORt
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.RESPonse.THRU
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.SOLT1
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.SOLT2
Calibration coefficient writing:
SCPI.SENSe(Ch).CORRection.COEFficient.DATA (Str, Pt_r, Pt_s)
Equivalent Softkeys
None
198
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.ISOLation
Object Type
Property (write only)
Data Type
Variant (array of long)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Measures the isolation calibration data between the source port and the receiver
port.
Description
The array contains 2 elements:
Data(0)
the number of the receiver port;
Data(1)
the number of the source port.
Range
Port number is 1 or 2. The array elements can not contain the same port
numbers.
Out of Range
If an incorrect port number is specified, an error occurs (error code: 222). If the
same port numbers are specified, an error occurs (error code: 220).
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.ISOLation = Array(1, 2)
Notes
The property writing starts the measurement for the channel independently of
the trigger initiation and trigger source settings. The function of the property
writing waits for the completion of the measurement.
Calibration > Calibrate > Response (Thru) > Isolation (Optional)
Equivalent Softkeys
Calibration > Calibrate > One Path 2–Port Cal > Isolation (Optional)
Calibration > Calibrate > Full 2–Port Cal > Isolation (Optional)
199
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.LOAD
Object Type
Property (write only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Measures the calibration data of the load standard for the specified port.
Range
Port number is 1 or 2.
Out of Range
An error occurs (error code: 222).
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.LOAD = 1
Notes
The property writing starts the measurement for the channel independently of
the trigger initiation and trigger source settings. The function of the property
writing waits for the completion of the measurement.
Calibration > Calibrate > Response (Open) > Load (Optional)
Calibration > Calibrate > Response (Short) > Load (Optional)
Equivalent Softkeys
Calibration > Calibrate > Full 1–Port Cal > Load
Calibration > Calibrate > One Path 2–Port Cal > Load
Calibration > Calibrate > Full 2–Port Cal > Port n Load
200
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.OPEN
Object Type
Property (write only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Measures the calibration data of the open standard for the specified port.
Range
Port number is 1 or 2.
Out of Range
An error occurs (error code: 222).
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.OPEN= 1
Notes
The property writing starts the measurement for the channel independently of
the trigger initiation and trigger source settings. The function of the property
writing waits for the completion of the measurement.
Calibration > Calibrate > Response (Open) > Open
Calibration > Calibrate > Full 1–Port Cal > Open
Equivalent Softkeys
Calibration > Calibrate > One Path 2–Port Cal > Open
Calibration > Calibrate > Full 2–Port Cal > Port n Open
201
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire. SHORt
Object Type
Property (write only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Measures the calibration data of the short standard for the specified port.
Range
Port number is 1 or 2.
Out of Range
An error occurs (error code: 222).
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.SHORt = 1
Notes
The property writing starts the measurement for the channel independently of
the trigger initiation and trigger source settings. The function of the property
writing waits for the completion of the measurement.
Calibration > Calibrate > Response (Short) > Short
Calibration > Calibrate > Full 1–Port Cal > Short
Equivalent Softkeys
Calibration > Calibrate > One Path 2–Port Cal > Short
Calibration > Calibrate > Full 2–Port Cal > Port n Short
202
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.THRU
Object Type
Property (write only)
Data Type
Variant (array of long)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Measures the calibration data of the thru standard between the source port and
the receiver port.
Description
The array contains 2 elements:
Data(0)
the number of the receiver port;
Data(1)
the number of the source port.
Range
Port number is 1 or 2. The array elements can not contain the same port
numbers.
Out of Range
If an incorrect port number is specified, an error occurs (error code: 222). If the
same port numbers are specified, an error occurs (error code: 220).
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.THRU= Array(1, 2)
Notes
The property writing starts the measurement for the channel independently of
the trigger initiation and trigger source settings. The function of the property
writing waits for the completion of the measurement.
Calibration > Calibrate > Response (Thru) > Thru
Equivalent Softkeys
Calibration > Calibrate > One Path 2–Port Cal > Thru
Calibration > Calibrate > Full 2–Port Cal > Port 1–2 Thru
203
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.LABel
Object Type
Property (read/write)
Data Type
String
Target
Calibration kit, selected for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The calibration kit label.
Range
up to 254 characters
Varies depending on the number of the calibration kit.
1: 85032B/E
2: 05CK10A-150
3: 85033D/E
Preset Value
4: 03CK10A-150
5: 85036B/E
6: 88050Q
7–11: None
Dim Lab As String
Syntax
Lab = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.LABel
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.LABel = "User1"
Equivalent Softkeys
Calibration > Edit Cal Kit > Label
204
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.ORDer.LOAD(Pt)
Object Type
Property (read/write)
Data Type
Long
Target
Calibration kit, selected for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The number of the calibration standard of the load type, used for the
measurement of the specified port Pt,
Pt:
port number 1–2 (see Table 1 on page 24)
Range
From 1 to the number of standards in the calibration kit.
Out of Range
If the specified standard number is greater than the number of standards in the
kit, an error occurs (error code: 222). If the specified standard number is not the
load standard number, an error occurs (error code: 220).
Dim Num As Long
Syntax
Num = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.ORDer.LOAD(Pt)
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.ORDer.LOAD(Pt) = 1
Equivalent Softkeys
Calibration > Edit Cal Kit > Load n > Connection
205
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.ORDer.OPEN(Pt)
Object Type
Property (read/write)
Data Type
Long
Target
Calibration kit, selected for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The number of the calibration standard of the open type, used for the
measurement of the specified port Pt,
Pt:
port number 1–2 (see Table 1 on page 24)
Range
From 1 to the number of standards in the calibration kit.
Out of Range
If the specified standard number is greater than the number of standards in the
kit, an error occurs (error code: 222). If the specified standard number is not the
open standard number, an error occurs (error code: 220).
Dim Num As Long
Syntax
Num = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.ORDer.OPEN(Pt)
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.ORDer.OPEN(Pt) = 1
Equivalent Softkeys
Calibration > Edit Cal Kit > Open n > Connection
206
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.ORDer.SHORt(Pt)
Object Type
Property (read/write)
Data Type
Long
Target
Calibration kit, selected for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The number of the calibration standard of the short type, used for the
measurement of the specified port Pt,
Pt:
port number 1–2 (see Table 1 on page 24)
Range
From 1 to the number of standards in the calibration kit.
Out of Range
If the specified standard number is greater than the number of standards in the
kit, an error occurs (error code: 222). If the specified standard number is not the
short standard number, an error occurs (error code: 220).
Dim Num As Long
Syntax
Num = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.ORDer.SHORt(Pt)
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.ORDer.SHORt(Pt) = 1
Equivalent Softkeys
Calibration > Edit Cal Kit > Short n > Connection
207
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.ORDer.THRU(Pt_m, Pt_n)
Object Type
Property (read/write)
Data Type
Long
Target
Calibration kit, selected for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The number of the calibration standard of the thru type, used for the
measurement between the Pt_m and Pt_n ports,
Pt_m: port number 1–2 (see Table 2 on page 24)
Pt_n: port number 1–2 (see Table 2 on page 24)
Range
From 1 to the number of standards in the calibration kit.
Out of Range
If the specified standard number is greater than the number of standards in the
kit, an error occurs (error code: 222). If the specified standard number is not
the thru standard number, an error occurs (error code: 220).
Dim Num As Long
Syntax
Num = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.ORDer.THRU(1, 2)
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.ORDer.THRU(1, 2) = 1
Equivalent Softkeys
Calibration > Edit Cal Kit > Thru n > Connection
208
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.RESet
Object Type
Method
Target
Calibration kit, selected for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Resets the calibration kit to the factory settings.
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.RESet
Equivalent Softkeys
Calibration > Restore Cal Kit
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.SELect
Object Type
Property (read/write)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The selected calibration kit for the channel.
Range
from 1 to 11
Out of Range
An error occurs. Error code: 222.
Preset Value
1
Dim Value As Long
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.SELect
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.SELect = 3
Equivalent Softkeys
Calibration > Cal Kit > Cal Kit n
209
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).ARBitrary
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Std: standard number (see Table 7 on page 211)
Description
The value of the arbitrary impedance for the load standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
50 or 75, depending on the selected calibration kit.
Unit
Ω (Ohm)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).ARBitrary
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).ARBitrary = 100
Equivalent Softkeys
Calibration > Edit Cal Kit > Load n > Arb. Impedance
210
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
Table 7. Std: Calibration Standard Number
Data Type
Long
Description
The number of the standard.
Range
Varies depending on the number of the standards in the calibration kit.
Out of Range
If the specified standard number is greater than the number of standards in the
kit, an error occurs (error code: 222).
Notes
If the standard number is not specified, it is taken as equal to 1.
211
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C0
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Std: standard number (see Table 7 on page 211)
Description
The C0 value of the open calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
1E–15 F (Farad)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C0
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C0 = 100
Equivalent Softkeys
–15
Calibration > Edit Cal Kit > Open n > C0 10
F
212
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C1
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Std: standard number (see Table 7 on page 211)
Description
The C1 value of the open calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
1E–27 F/Hz (Farad/Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C1
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C1 = 100
Equivalent Softkeys
Calibration > Edit Cal Kit > Open n > C1 10
–27
F/Hz
213
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C2
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Std:
standard number (see Table 7 on page 211)
Description
The C2 value of the open calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
1E–36 F/Hz2 (Farad/Hertz2)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C2
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C2 = 100
Equivalent Softkeys
Calibration > Edit Cal Kit > Open n > C2 10
–36
F/Hz
2
214
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C3
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Std: standard number (see Table 7 on page 211)
Description
The C3 value of the open calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
1E–45 F/Hz3 (Farad/Hertz3)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C3
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C3 = 100
Equivalent Softkeys
Calibration > Edit Cal Kit > Open n > C3 10
–45
F/Hz
3
215
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).DELay
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Std: standard number (see Table 7 on page 211)
Description
The offset delay value of the calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).DELay
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).DELay = 93E–12
Equivalent Softkeys
Calibration > Edit Cal Kit > Open | Short | Load | Thru > Offset Delay
216
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L0
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Std: standard number (see Table 7 on page 211)
Description
The L0 value of the short calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
1E–12 H (Henry)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L0
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L0 = 100
Equivalent Softkeys
Calibration > Edit Cal Kit > Short n > L0 10
–12
H
217
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L1
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Std: standard number (see Table 7 on page 211)
Description
The L1 value of the short calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
1E–24 H/Hz (Henry/Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L1
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L1 = 100
Equivalent Softkeys
Calibration > Edit Cal Kit > Short n > L1 10
–24
H/Hz
218
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L2
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Std: standard number (see Table 7 on page 211)
Description
The L2 value of the short calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
1E–33 H/Hz2 (Henry/Hertz2)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L2
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L2 = 100
Equivalent Softkeys
Calibration > Edit Cal Kit > Short n > L2 10
–33
H/Hz
2
219
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L3
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Std: standard number (see Table 7 on page 211)
Description
The L3 value of the short calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
1E–42 H/Hz3 (Henry/Hertz3)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L3
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L3 = 100
Equivalent Softkeys
Calibration > Edit Cal Kit > Short n > L3 10
–42
H/Hz
3
220
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LABel
Object Type
Property (read/write)
Data Type
String
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Std: standard number (see Table 7 on page 211)
Description
The label of the calibration standard.
Range
up to 254 characters
Preset Value
Varies depending on the selected calibration kit and the standard.
Dim Lab As String
Syntax
Lab = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LABel
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LABel = "Open"
Equivalent Softkeys
Calibration > Edit Cal Kit > Open | Short | Load | Thru > Label
221
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LOSS
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Std: standard number (see Table 7 on page 211)
Description
The offset loss value of the calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
Ω/s (Ohm/second)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LOSS
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LOSS = 700E6
Equivalent Softkeys
Calibration > Edit Cal Kit > Open | Short | Load | Thru > Offset Loss
222
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).TYPE
Object Type
Property (read/write)
Data Type
String
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Std: standard number (see Table 7 on page 211)
Description
The type of the calibration standard.
Range
"OPEN"
: Open
"SHORt"
: Short
"LOAD"
: Load
"THRU"
: Thru
"NONE"
: Not defined
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
An error occurs. Error code: 216.
Preset Value
Varies depending on the selected calibration kit and the standard.
Dim Param As String
Syntax
Param = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).TYPE
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).TYPE = "OPEN"
Equivalent Softkeys
Calibration > Edit Cal Kit > Open | Short | Load | Thru > STD Type
223
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).Z0
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Std: standard number (see Table 7 on page 211)
Description
The offset Z0 value of the calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
50 or 75, depending on the selected calibration kit.
Unit
Ω (Ohm)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).Z0
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).Z0 = 50
Equivalent Softkeys
Calibration > Edit Cal Kit > Open | Short | Load | Thru > Offset Z0
224
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CLEar
Object Type
Method
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Clears the measurement values of the calibration standards.
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.CLEar
Equivalent Softkeys
Calibration > Calibrate > Response (Open) | Response (Short) | Response (Thru) |
One Path 2–Port Cal | Full 1–Port Cal | Full 2–Port Cal > Cancel > OK
225
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.DATA.ISOLation(Pt_r, Pt_s)
Object Type
Property (read/write)
Data Type
Variant (array of double)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The array of the isolation calibration measurements performed between the
receiver port Pt_r and the source port Pt_s,
Pt_r:
the number of the receiver port 1–2 (see Table 2 on page 24)
Pt_s:
the number of the source port 1–2 (see Table 2 on page 24)
Description
The array elements are complex numbers. Also see section “Measurement
Data Arrays” on page 19.
The array size is 2N, where N is the number of measurement points.
For the n–th point, where n from 1 to N:
Data(2n–2)
real part of the measurement;
Data(2n–1)
imaginary part of the measurement.
Dim Data As Variant
Syntax
Data = app.SCPI.SENSe(Ch).CORRection.COLLect.DATA.ISOLation(Pt_r, Pt_s)
app.SCPI.SENSe(Ch).CORRection.COLLect.DATA.ISOLation(Pt_r, Pt_s) = Data
Related Commands
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.ISOLation
Equivalent Softkeys
None
226
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.DATA.LOAD(Pt)
Object Type
Property (read/write)
Data Type
Variant (array of double)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The array of the load calibration standard measurements for the port Pt,
Pt: port number 1–2 (see Table 2 on page 24)
The array elements are complex numbers. Also see section “Measurement
Data Arrays” on page 19.
Description
The array size is 2N, where N is the number of measurement points.
For the n–th point, where n from 1 to N:
Data(2n–2)
real part of the measurement;
Data(2n–1)
imaginary part of the measurement.
Dim Data As Variant
Syntax
Data = app.SCPI.SENSe(Ch).CORRection.COLLect.DATA.LOAD(Pt)
app.SCPI.SENSe(Ch).CORRection.COLLect.DATA.LOAD(Pt) = Data
Related Commands
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.LOAD
Equivalent Softkeys
None
227
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.DATA.OPEN(Pt)
Object Type
Property (read/write)
Data Type
Variant (array of double)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The array of the open calibration standard measurements for the port Pt,
Pt:
port number 1–2 (see Table 2 on page 24)
The array elements are complex numbers. Also see section “Measurement
Data Arrays” on page 19.
Description
The array size is 2N, where N is the number of measurement points.
For the n–th point, where n from 1 to N:
Data(2n–2)
real part of the measurement;
Data(2n–1)
imaginary part of the measurement.
Dim Data As Variant
Syntax
Data = app.SCPI.SENSe(Ch).CORRection.COLLect.DATA.OPEN(Pt)
app.SCPI.SENSe(Ch).CORRection.COLLect.DATA.OPEN(Pt) = Data
Related Commands
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.OPEN
Equivalent Softkeys
None
228
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.DATA.SHORt(Pt)
Object Type
Property (read/write)
Data Type
Variant (array of double)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The array of the short calibration standard measurements for the port Pt,
Pt:
port number 1–2 (see Table 2 on page 24)
The array elements are complex numbers. Also see section “Measurement
Data Arrays” on page 19.
Description
The array size is 2N, where N is the number of measurement points.
For the n–th point, where n from 1 to N:
Data(2n–2)
real part of the measurement;
Data(2n–1)
imaginary part of the measurement.
Dim Data As Variant
Syntax
Data = app.SCPI.SENSe(Ch).CORRection.COLLect.DATA.SHORt(Pt)
app.SCPI.SENSe(Ch).CORRection.COLLect.DATA.SHORt(Pt) = Data
Related Commands
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.SHORt
Equivalent Softkeys
None
229
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.DATA.THRU.
MATCh(Pt_r,Pt_s)
Object Type
Property (read/write)
Data Type
Variant (array of double)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The array of the reflection measurements of the thru standard connected
between the receiver port Pt_r and the source port Pt_s,
Pt_r: the number of the receiver port 1–2 (see Table 2 on page 24)
Pt_s: the number of the source port 1–2 (see Table 2 on page 24)
Description
The array elements are complex numbers. Also see section “Measurement
Data Arrays” on page 19.
The array size is 2N, where N is the number of measurement points.
For the n–th point, where n from 1 to N:
Data(2n–2)
real part of the measurement;
Data(2n–1)
imaginary part of the measurement.
Dim Data As Variant
Syntax
Data = app.SCPI.SENSe(Ch).CORRection.COLLect.DATA.THRU.MATCh(Pt_r, Pt_s)
app.SCPI.SENSe(Ch).CORRection.COLLect.DATA.THRU.MATCh(Pt_r, Pt_s) = Data
Related Commands
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.THRU
Equivalent Softkeys
None
230
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.DATA.THRU.
TRANsmission(Pt_r, Pt_s)
Object Type
Property (read/write)
Data Type
Variant (array of double)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The array of the transmission measurements performed between the receiver
port Pt_r and the source port Pt_s, using the thru standard,
Pt_r: the number of the receiver port 1–2 (see Table 2 on page 24)
Pt_s: the number of the source port 1–2 (see Table 2 on page 24)
Description
The array elements are complex numbers. Also see section “Measurement
Data Arrays” on page 19.
The array size is 2N, where N is the number of measurement points.
For the n–th point, where n from 1 to N:
Data(2n–2)
real part of the measurement;
Data(2n–1)
imaginary part of the measurement.
Dim Data As Variant
Syntax
Data = app.SCPI.SENSe(Ch).CORRection.COLLect.DATA.THRU.TRANsmission(Pt_r,
Pt_s)
app.SCPI.SENSe(Ch).CORRection.COLLect.DATA.THRU.TRANsmission(Pt_r, Pt_s) =
Data
Related Commands
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.THRU
Equivalent Softkeys
None
231
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.ECAL.INFormation
Object Type
Property (read only)
Data Type
String
Target
AutoCal module (Ch – arbitrary number 1 to 16)
Description
Gets information about the AutoCal Module connected to the Network
Analyzer in a string with comma separated fields.
Autocal Module Information:
Model Name,
Serial Number,
Current Temperature of AutoCal Module,
Selected Characterization Information:
Characterization Name,
Characterization Date and Time,
Min Frequency,
Max Frequency,
Number of Points,
Characterization Temperature,
PortA Connector,
PortB Connector,
PortA Adapter,
PortB Adapter,
Analyzer,
Location,
Operator.
Dim ID As String
Syntax
ID = app.SCPI.SENSe(Ch).CORRection.COLLect.ECAL.INFormation
Equivalent Softkeys
Calibration > AutoCal > Characterization Info…
232
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.ECAL.ORIenation.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
AutoCal (Ch – arbitrary number 1 to 16)
Description
The ON/OFF state of the Auto-Orientation function used when executing
AutoCal.
Allowable Values
True: Auto-Orientation function ON
False: Auto-Orientation function OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status =
app.SCPI.SENSe(Ch).CORRection.COLLect.ECAL.ORIentation.STATe
app.SCPI.SENSe(Ch).CORRection.COLLect.ECAL.ORIentation.STATe =
False
Equivalent Softkeys
Calibration > AutoCal > Orientation > Auto-Orientation
233
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.ECAL.PATH(Pt)
Object Type
Property (read/write)
Data Type
Long
Target
AutoCal (Ch – arbitrary number 1 to 16)
Description
Sets or reads out the AutoCal module port number which is connected to a
selected Network Analyzer port (Pt).
Allowable Values
1: Port A of AutoCal Module
2: Port B of AutoCal Module
3: Port C of AutoCal Module (4 port AutoCal module only)
4: Port D of AutoCal Module (4 port AutoCal module only)
Dim Value As Long
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.ECAL.PATH(Pt)
app.SCPI.SENSe(Ch).CORRection.COLLect.ECAL.PATH(Pt) = 2
Equivalent Softkeys
Calibration > AutoCal > Orientation > Port 1 | Port 2
234
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.ECAL.SOLT1
Object Type
Property (write only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Executes 1-port calibration of the specified port of selected channel (Ch) using
the AutoCal module.
Dim Port As Long
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.ECAL.SOLT1 = Port
Equivalent Softkeys
Calibration > AutoCal > 1-Port AutoCal > Port 1 | Port 2
SCPI.SENSe(Ch).CORRection.COLLect.ECAL.SOLT2
Object Type
Property (write only)
Data Type
Variant (array of long)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Executes full 2-port calibration between the specified 2 ports of selected
channel (Ch) using the AutoCal module.
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.ECAL.SOLT2 = Array(2, 1)
Equivalent Softkeys
Calibration > AutoCal > 2-Port AutoCal
235
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.ECAL.UCHar
Object Type
Property (read/write)
Data Type
String
Target
AutoCal (Ch – arbitrary number 1 to 16)
Description
Sets or reads out the Characteristic used when executing AutoCal (factory or
user characterization)..
Range
Preset Value
"CHAR0"
: Factory characterization
"CHAR1"
: User characterization 1
"CHAR2"
: User characterization 2
"CHAR3"
: User characterization 3
"CHAR4"
: User characterization 4
"CHAR5"
: User characterization 5
CHAR0
Dim Param As String
Syntax
Param = app.SCPI.SENSe(Ch).CORRection.COLLect.ECAL.UCHar
app.SCPI.SENSe(Ch).CORRection.COLLect.ECAL.UCHar = "CHAR0"
Equivalent Softkeys
Calibration > AutoCal > Characterization
236
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.ECAL.UTHRu.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
AutoCal (Ch – arbitrary number 1 to 16)
Description
The ON/OFF state of the Unknown Thru feature used when executing AutoCal.
Allowable Values
True: Unknown Thru feature ON
False: Unknown Thru feature OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.SENSe(Ch).CORRection.COLLect.ECAL.UTHRu.STATe
app.SCPI.SENSe(Ch).CORRection.COLLect.ECAL.UTHRu.STATe = False
Note
PLANAR 804/1, S5048, and S7530 only
Equivalent Softkeys
Calibration > AutoCal > Unkn Thru
237
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.METHod.ERESponse
Object Type
Property (write only)
Data Type
Variant (array of long)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Selects the ports and sets the one path 2–port calibration type for the
calculation of the calibration coefficients on completion of the calibration
executed by the SCPI.SENSe(Ch).CORRection.COLLect.SAVE method.
Description
The array contains 2 elements:
Data(0)
the number of the receiver port;
Data(1)
the number of the source port.
Range
Port number is 1 or 2. The array elements can not contain the same port
numbers.
Out of Range
If an incorrect port number is specified, an error occurs (error code: 222). If the
same port numbers are specified, an error occurs (error code: 220).
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.ERESponse = Array(2, 1)
Equivalent Softkeys
Calibration > Calibrate > One Path 2–Port Cal > Select Port
238
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.OPEN
Object Type
Property (write only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Selects the port and sets the response calibration (Open) type for the
calculation of the calibration coefficients on completion of the calibration
executed by the SCPI.SENSe(Ch).CORRection.COLLect.SAVE method.
Range
from 1 to 2
Out of Range
An error occurs. Error code: 222.
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.OPEN = 1
Equivalent Softkeys
Calibration > Calibrate > Response (Open) > Select Port
239
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.SHORt
Object Type
Property (write only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Selects the port and sets the response calibration (Short) type for the
calculation of the calibration coefficients on completion of the calibration
executed by the SCPI.SENSe(Ch).CORRection.COLLect.SAVE method.
Range
from 1 to 2
Out of Range
An error occurs. Error code: 222.
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.SHORt = 1
Equivalent Softkeys
Calibration > Calibrate > Response (Short) > Select Port
240
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.THRU
Object Type
Property (write only)
Data Type
Variant (array of long)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Selects the ports and sets the response calibration (Thru) type for the
calculation of the calibration coefficients on completion of the calibration
executed by the SCPI.SENSe(Ch).CORRection.COLLect.SAVE method.
Description
The array contains 2 elements:
Data(0)
the number of the receiver port;
Data(1)
the number of the source port.
Range
Port number is 1 or 2. Array elements can not contain the same port numbers.
Out of Range
If an incorrect port number is specified, an error occurs (error code: 222). If the
same port numbers are specified, an error occurs (error code: 220).
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.THRU = Array(2, 1)
Equivalent Softkeys
Calibration > Calibrate > Response (Thru) > Select Port
241
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.METHod.SOLT1
Object Type
Property (write only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Selects the port and sets the full 1–port calibration type for the calculation of
the calibration coefficients on completion of the calibration executed by the
SCPI.SENSe(Ch).CORRection.COLLect.SAVE method.
Range
from 1 to 2
Out of Range
An error occurs. Error code: 222.
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.SOLT1 = 1
Equivalent Softkeys
Calibration > Calibrate > Full 1–Port Cal > Select Port
242
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.METHod.SOLT2
Object Type
Property (write only)
Data Type
Variant (array of long)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Selects the port and sets the full 2–port calibration type for the calculation of
the calibration coefficients on completion of the calibration executed by the
SCPI.SENSe(Ch).CORRection.COLLect.SAVE method.
Description
The array contains 2 elements:
Data(0)
the number of the receiver port;
Data(1)
the number of the source port.
Range
Port number is 1 or 2. Array elements can not contain the same port numbers.
Out of Range
If an incorrect port number is specified, an error occurs (error code: 222). If the
same port numbers are specified, an error occurs (error code: 220).
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.SOLT2 = Array(2, 1)
Equivalent Softkeys
Calibration > Calibrate > Full 2–Port Cal
243
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.METHod.TYPE
Object Type
Property (read only)
Data Type
String
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The calibration type selected for calculating of the calibration coefficients on
completion
of
the
calibration
executed
by
the
SCPI.SENSe(Ch).CORRection.COLLect.SAVE method.
Range
"RESPO"
: Response (Open)
"RESPS"
: Response (Short)
"RESPT"
: Response (Thru)
"SOLT1"
: Full 1–port calibration
"SOLT2"
: Full 2–port calibration
"1PATH"
: One path 2–port calibration
"NONE"
: Not defined
Dim Param As String
Syntax
Param = app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.TYPE
Equivalent Softkeys
None
244
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.SAVE
Object Type
Method
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Calculates the calibration coefficients from the calibration standards
measurements depending on the selected calibration type.
Description
On completion of the method, all the calibration standards measurements are
cleared and the error correction automatically turns ON.
At the attempt to execute this method before all the needed standards are
measured, an error occurs and the method is ignored.
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.SAVE
Calibration type selection:
Related Commands
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.OPEN
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.SHORt
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.THRU
SCPI.SENSe(Ch).CORRection.COLLect.METHod.ERESponse
SCPI.SENSe(Ch).CORRection.COLLect.METHod.SOLT1
SCPI.SENSe(Ch).CORRection.COLLect.METHod.SOLT2
Calibration standards measurement:
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.ISOLation
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.LOAD
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.OPEN
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.SHORt
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.THRU
Equivalent Softkeys
Calibration > Calibrate > Response (Open) | Response (Short) | Response (Thru) |
One Path 2–Port Cal | Full 1–Port Cal | Full 2–Port Cal > Apply
245
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).FREQuency(Ls)
Object Type
Property (read/write)
Data Type
Double
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
The value of the frequency at the point number Ls for calculation of the loss for
the port extension function,
Ls:
point number 1–2 (see Table 8 on page 246)
Range
from 3E5 to 3.2E9 (PLANAR 304/1)
from 3E5 to 8.0E9 (PLANAR 804/1)
from 2E4 to 4.8E9 (S5048)
from 2E4 to 3.0E9 (S7530)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1E9
Unit
Hz (Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).FREQuency(Ls)
app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).FREQuency(Ls) = 100Е6
Equivalent Softkeys
Calibration > Port Extensions > Loss > Freq1 | Freq2
Table 8. Ls: Point Number
Data Type
Long
Description
The point number for the loss setting in the port extension function.
Range
1 or 2
Out of Range
An error occurs. Error code: 222.
246
PLANAR 304/1, 804/1, S5048, S7530
Notes
COM/DCOM Programming manual
If the point number is not specified, it is taken as equal to 1.
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).INCLude(Ls).STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
The ON/OFF state of the loss compensation at the point number Ls to calculate
the loss for the port extension function,
Ls:
point number 1–2 (see Table 8 on page 246)
Allowable Values
True: Loss compensation ON
False: Loss compensation OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).INCLude(Ls).STATe
app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).INCLude(Ls).STATe = True
Equivalent Softkeys
Calibration > Port Extensions > Loss > Loss1 | Loss2
247
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).LDC
Object Type
Property (read/write)
Data Type
Double
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
The loss value at DC for the loss calculation of the port extension function.
Range
from –200 to 200
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
dB (decibel)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).LDC
app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).LDC = 10
Equivalent Softkeys
Calibration > Port Extensions > Loss > Loss at DC
248
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).LOSS(Ls)
Object Type
Property (read/write)
Data Type
Double
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
The loss value at the point number Ls to calculate the loss for the port
extension function,
Ls:
point number 1–2 (see Table 7 on page 236)
Range
from –200 to 200
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
dB (decibel)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).LOSS(Ls)
app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).LOSS(Ls) = 10
Equivalent Softkeys
Calibration > Port Extensions > Loss > Loss1 | Loss2
249
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).TIME
Object Type
Property (read/write)
Data Type
Double
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
The electrical delay value for the port extension function.
Range
from –10 to 10
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).TIME
app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).TIME = 10
Equivalent Softkeys
Calibration > Port Extensions > Extension Port1 | Extension Port2
250
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.EXTension.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the port extension function.
Allowable Values
True: Port extension function ON
False: Port extension function OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.SENSe(Ch).CORRection.EXTension.STATe
app.SCPI.SENSe(Ch).CORRection.EXTension.STATe = True
Equivalent Softkeys
Calibration > Port Extensions > Extension
251
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe.CORRection.IMPedance.INPut.MAGNitude
Object Type
Property (read/write)
Data Type
Double
Target
Instrument
Description
The system impedance Z0.
Range
from 0.001 to 1000
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
50
Unit
Ω (Ohm)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe.CORRection.IMPedance.INPut.MAGNitude
app.SCPI.SENSe.CORRection.IMPedance.INPut.MAGNitude = 75
Equivalent Softkeys
Calibration > System Z0
252
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.OFFSet.CLEar
Object Type
Method
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Clears the scalar mixer calibration coefficient table.
Syntax
app.SCPI.SENSe(Ch).CORRection.OFFSet.CLEar
Equivalent Softkeys
None
253
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.LOAD
Object Type
Property (write only)
Data Type
Variant (array of long)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Measures the calibration data of the load standard of the specified port when
the frequency offset feature is on (scalar mixer cal).
Description
The array contains 2 elements:
Port(0)
Measurement port number;
Port(1)
Frequency port number.
Range
Port number is 1 or 2. The array elements can contain the same port numbers.
Out of Range
If an incorrect port number is specified, an error occurs (error code: 222).
Syntax
app.SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.LOAD = Array(1, 2)
Notes
The property writing starts the measurement for the channel independently of
the trigger initiation and trigger source settings. The function of the property
writing waits for the completion of the measurement.
Equivalent Softkeys
Calibration > Mixer/Converter Calibration > Scalar Mixer Calibration > Reflection
Port n > Port n @Freq m Load
254
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.OPEN
Object Type
Property (write only)
Data Type
Variant (array of long)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Measures the calibration data of the open standard of the specified port when
the frequency offset feature is on (scalar mixer cal).
Description
The array contains 2 elements:
Port(0)
Measurement port number;
Port(1)
Frequency port number.
Range
Port number is 1 or 2. The array elements can contain the same port numbers.
Out of Range
If an incorrect port number is specified, an error occurs (error code: 222).
Syntax
app.SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.OPEN = Array(1, 2)
Notes
The property writing starts the measurement for the channel independently of
the trigger initiation and trigger source settings. The function of the property
writing waits for the completion of the measurement.
Equivalent Softkeys
Calibration > Mixer/Converter Calibration > Scalar Mixer Calibration > Reflection
Port n > Port n @Freq m Open
255
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.SHORt
Object Type
Property (write only)
Data Type
Variant (array of long)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Measures the calibration data of the short standard of the specified port when
the frequency offset feature is on (scalar mixer cal).
Description
The array contains 2 elements:
Port(0)
Measurement port number;
Port(1)
Frequency port number.
Range
Port number is 1 or 2. The array elements can contain the same port numbers.
Out of Range
If an incorrect port number is specified, an error occurs (error code: 222).
Syntax
app.SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.SHORt = Array(1, 2)
Notes
The property writing starts the measurement for the channel independently of
the trigger initiation and trigger source settings. The function of the property
writing waits for the completion of the measurement.
Equivalent Softkeys
Calibration > Mixer/Converter Calibration > Scalar Mixer Calibration > Reflection
Port n > Port n @Freq m Short
256
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.THRU
Object Type
Property (write only)
Data Type
Variant (array of long)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Measures the calibration data of the thru standard of the specified port when
the frequency offset feature is on (scalar mixer cal).
Description
The array contains 2 elements:
Port(0)
Response port number;
Port(1)
Stimulus port number.
Range
Port number is 1 or 2. The array elements can not contain the same port
numbers.
Out of Range
If an incorrect port number is specified, an error occurs (error code: 222). If the
same port numbers are specified, an error occurs (error code: 220).
Syntax
app.SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.THRU = Array(1, 2)
Notes
The property writing starts the measurement for the channel independently of
the trigger initiation and trigger source settings. The function of the property
writing waits for the completion of the measurement.
Equivalent Softkeys
Calibration > Mixer/Converter Calibration > Scalar Mixer Calibration > Reflection
Port n > Port n @Freq m Thru
257
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.PMETer
Object Type
Property (write only)
Data Type
Variant (array of long)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Measures the scalar-mixer calibration data using the power meter when the
frequency offset feature is ON.
The array contains 3 elements:
Description
Port(0)
Measurement port number;
Port(1)
Frequency port number.
Port(2)
Reserved.
Range
Port number is 1 or 2. The array elements can contain the same port numbers.
Out of Range
If an incorrect port number is specified, an error occurs (error code: 222).
Syntax
app.SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.PMETer = Array(1, 2, 0)
Notes
The property writing starts the measurement for the channel independently of
the trigger initiation and trigger source settings. The function of the property
writing waits for the completion of the measurement.
Equivalent Softkeys
Calibration > Mixer/Converter Calibration > Scalar Mixer Calibration > Power >
Port n @Freq m
258
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.CLEar
Object Type
Method
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Clears the calibration measurement data when the frequency offset feature is
ON (scalar mixer calibration).
Syntax
app.SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.CLEar
Equivalent Softkeys
Calibration > Mixer/Converter Calibration > Scalar Mixer Calibration > Cancel
259
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.METHod.SMIX2
Object Type
Property (write only)
Data Type
Variant (array of long)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Selects the ports and sets the scalar mixer calibration type when the frequency
offset feature is on for the calculation of the calibration coefficients on
completion
of
the
calibration
executed
by
the
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.SAVE method.
The array contains 2 elements:
Port(0)
Port number 1;
Port(1)
Port number 2.
Range
Port number is 1 or 2. The array elements can not contain the same port
numbers.
Out of Range
If an incorrect port number is specified, an error occurs (error code: 222). If the
same port numbers are specified, an error occurs (error code: 220).
Syntax
app.SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.METHod.SMIX2 = Array(2, 1)
Equivalent Softkeys
Calibration > Mixer/Converter Calibration > Scalar Mixer Calibration
260
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.SAVE
Object Type
Method
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Calculates the calibration coefficient for the selected calibration type (scalar
mixer calibration only) from the calibration data measured with the frequency
offset feature is ON.
Description
If this command is executed before all necessary calibration data for
calculating the calibration coefficient is measured, an error occurs when
executed.
Syntax
app.SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.SAVE
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.METHod.SMIX2
Related Commands
Equivalent Softkeys
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.LOAD
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.OPEN
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.SHORt
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.THRU
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.PMETer
Calibration > Mixer/Converter Calibration > Scalar Mixer Calibration > Apply
261
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.RECeiver(Pt).COLLect.ACQuire
Object Type
Property (write only)
Data Type
Long
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
The property writing executes the receiver calibration for the specified port Pt,
using the specified source port Src to the right of the equal sign.
Range
The number of the source port from 1 to 2.
Out of Range
Error occurs. Error code: 222.
Syntax
app.SCPI.SENSe(Ch).CORRection.RECeiver(Pt).COLLect.ACQuire = Src
Notes
The property writing triggers the channel sweep independently of its current
mode (continuous, single, hold) and trigger source setting. The function of the
property writing waits for the completion of the calibration.
Equivalent Softkeys
Calibration > Receiver Calibration > Take Cal Sweep
262
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.RECeiver(Pt).STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
The ON/OFF state of the receiver correction for the port Pt.
Allowable Values
True: Receiver correction ON
False: Receiver correction OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.SENSe(Ch).CORRection.RECeiver(Pt).STATe
app.SCPI.SENSe(Ch).CORRection.RECeiver(Pt).STATe = True
Equivalent Softkeys
Calibration > Receiver Calibration > Correction
263
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the error correction.
Allowable Values
True: Error correction ON
False: Error correction OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.SENSe(Ch).CORRection.STATe
app.SCPI.SENSe(Ch).CORRection.STATe = True
Equivalent Softkeys
Calibration > Correction
264
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.TYPE(Tr)
Object Type
Property (read only)
Data Type
Variant (Variant array)
Target
Trace Tr of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Tr:
trace number 1–16 (see Table 3 on page 32)
The information about the applied calibration type and the port numbers for the
specified trace.
The array contains 3 elements:
Description
Data(0)
calibration type (see below);
Data(1)
the number of the receiver port to be calibrated;
Data(2)
the number of the source port to be calibrated.
Calibration type in the element Data(0):
Range
"RESPO"
: Response (Open)
"RESPS"
: Response (Short)
"RESPT"
: Response (Thru)
"SOLT1"
: Full 1–port calibration
"SOLT2"
: Full 2–port calibration
"1PATH"
: One path 2–port calibration
"NONE"
: Not defined
Dim CalInfo As Variant
Syntax
CalInfo = app.SCPI.SENSe(Ch).CORRection.TYPE(Tr)
Equivalent Softkeys
None
265
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).FREQuency.CENTer
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The stimulus center value of the sweep range for linear or logarithmic sweep
types.
Range
from 3E5 to 3.2E9 (PLANAR 304/1)
from 3E5 to 8.0E9 (PLANAR 804/1)
from 2E4 to 4.8E9 (S5048)
from 2E4 to 3.0E9 (S7530)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1.60015E9 (PLANAR 304/1)
4.00015E9 (PLANAR 804/1)
2.40001E9 (S5048)
1.50001E9 (S7530)
Unit
Hz (Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).FREQuency.CENTer
app.SCPI.SENSe(Ch).FREQuency.CENTer = 1Е9
Equivalent Softkeys
Stimulus > Center
266
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).FREQuency.CW
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The value of the fixed frequency for the power sweep.
Range
from 3E5 to 3.2E9 (PLANAR 304/1)
from 3E5 to 8.0E9 (PLANAR 804/1)
from 2E4 to 4.8E9 (S5048)
from 2E4 to 3.0E9 (S7530)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
3E5
Unit
Hz (Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).FREQuency.CW
app.SCPI.SENSe(Ch).FREQuency.CW = 1Е9
Equivalent Softkeys
Stimulus > Power > CW Freq
267
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).FREQuency.DATA
Object Type
Property (read only)
Data Type
Variant (array of double)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The array of the measurement points frequency for linear, logarithmic or
segment sweep type.
Description
The array size is N, where N is the number of measurement points.
For the n–th point, where n from 1 to N:
Data(n–1)
Syntax
the frequency value at the n–th measurement point.
Dim Data As Variant
Data = app. SCPI.SENSe(Ch).FREQuency.DATA
Equivalent Softkeys
None
268
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).FREQuency.SPAN
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The stimulus span value of the sweep range for linear or logarithmic sweep
types.
Range
from 0 to 3.1997E9 (PLANAR 304/1)
from 0 to 7.9997E9 (PLANAR 804/1)
from 0 to 4.79998E9 (S5048)
from 0 to 2.99998E9 (S7530)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
3.1997E9 (PLANAR 304/1)
7.9997E9 (PLANAR 804/1)
4.79998E9 (S5048)
2.99998E9 (S7530)
Unit
Hz (Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).FREQuency.SPAN
app.SCPI.SENSe(Ch).FREQuency.SPAN = 2Е9
Equivalent Softkeys
Stimulus > Span
269
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).FREQuency.STARt
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The stimulus start value of the sweep range for linear or logarithmic sweep
types.
Range
from 3E5 to 3.2E9 (PLANAR 304/1)
from 3E5 to 8.0E9 (PLANAR 804/1)
from 2E4 to 4.8E9 (S5048)
from 2E4 to 3.0E9 (S7530)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
3E5
Unit
Hz (Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).FREQuency.STARt
app.SCPI.SENSe(Ch).FREQuency.STARt = 1Е6
Equivalent Softkeys
Stimulus > Start
270
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).FREQuency.STOP
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The stimulus stop value of the sweep range for linear or logarithmic sweep
types.
Range
from 3E5 to 3.2E9 (PLANAR 304/1)
from 3E5 to 8.0E9 (PLANAR 804/1)
from 2E4 to 4.8E9 (S5048)
from 2E4 to 3.0E9 (S7530)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
3.2E9 (PLANAR 304/1)
8.0E9 (PLANAR 804/1)
4.8E9 (S5048)
3.0E9 (S7530)
Unit
Hz (Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).FREQuency.STOP
app.SCPI.SENSe(Ch).FREQuency.STOP = 1Е6
Equivalent Softkeys
Stimulus > Stop
271
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).OFFSet.ADJust.CONTinuous.PERiod
Object Type
Property (read/write)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Selects the period for continuous offset adjust feature:
1 – 3 sec;
Description
2 – 10 sec;
3 – 30 sec.
Range
from 1 to 3
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
2
Unit
n/a
Dim Value As Long
Syntax
Value = app.SCPI.SENSe(Ch).OFFSet.ADJust.PERiod
app.SCPI.SENSe(Ch).OFFSet.ADJust.PERiod = 1
Equivalent Softkeys
Stimulus > Frequency Offset > Offset Adjust > Adjust Period
272
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).OFFSet.ADJust.CONTinuous.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the continuous frequency offset adjust feature.
Allowable Values
True: continuous offset adjust feature is ON
False: continuous offset adjust feature is OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.SENSe(Ch).OFFSet.ADJust.CONTinuous.STATe
app.SCPI.SENSe(Ch).OFFSet.ADJust.CONTinuous.STATe = True
Equivalent Softkeys
Stimulus > Frequency Offset > Offset Adjust > Continuous Adjust
273
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).OFFSet.ADJust.EXECute
Object Type
Method
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Executes the offset adjust when the frequency offset feature is ON.
Syntax
app.SCPI.SENSe(Ch).OFFset.ADJust.EXECute
Equivalent Softkeys
Stimulus > Frequency Offset > Offset Adjust > Adjust Immediate
274
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).OFFSet.ADJust.PORT
Object Type
Property (read/write)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Selects the port number for frequency offset adjust feature.
Range
from 1 to 2
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1
Unit
n/a
Dim Value As Long
Syntax
Value = app.SCPI.SENSe(Ch).OFFset.ADJust.PORT
app.SCPI.SENSe(Ch).OFFset.ADJust.PORT = 2
Equivalent Softkeys
Stimulus > Frequency Offset > Offset Adjust > Select Port
275
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).OFFSet.ADJust.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the frequency offset adjust feature.
Allowable Values
True: offset adjust feature is ON
False: offset adjust feature is OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.SENSe(Ch).OFFSet.ADJust.STATe
app.SCPI.SENSe(Ch).OFFSet.ADJust.STATe = True
Equivalent Softkeys
Stimulus > Frequency Offset > Offset Adjust > Offset Adjust
276
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).OFFSet.ADJust.Value
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The frequency offset adjust value of the offset adjust feature.
Range
from –5E5 to 5E5
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
Hz
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).OFFSet.ADJust.VALue
app.SCPI.SENSe(Ch).OFFSet.ADJust.VALue = 1000
Equivalent Softkeys
Stimulus > Frequency Offset > Offset Adjust > Adjust Value
277
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).OFFSet.PORT(Pt).FREQuency.DATA
Object Type
Property (read only)
Data Type
Variant (array of double)
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Reads the frequency data for the selected port Pt when the frequency offset
feature is ON, for the selected channel Ch.
Description
The array size is N, where N is the number of measurement points.
For the n–th point, where n from 1 to N:
Data(n–1)
Syntax
the frequency value at the n–th measurement point.
Dim Data As Variant
Data = app.SCPI.SENSe(Ch).OFFSet.PORT(Pt).FREQuency.DATA
Equivalent Softkeys
None
278
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).OFFSet.PORT(Pt).FREQuency.DIVisor
Object Type
Property (read/write)
Data Type
Double
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
The basic frequency range divisor when the frequency offset feature is ON.
Range
from 1 to 100
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1
Unit
n/a
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).OFFset.PORT(Pt).FREQuency.DIVisor
app.SCPI.SENSe(Ch).OFFset.PORT(Pt).FREQuency.DIVisor = 2
Equivalent Softkeys
Stimulus > Frequency Offset > Port n > Divider
279
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).OFFSet.PORT(Pt).FREQuency. MULTiplier
Object Type
Property (read/write)
Data Type
Double
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
The basic range frequency multiplier when the frequency offset feature is ON.
Range
from –100 to 100
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1
Unit
n/a
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).OFFset.PORT(Pt).FREQuency.MULTiplier
app.SCPI.SENSe(Ch).OFFset.PORT(Pt).FREQuency.MULTiplier = 2
Equivalent Softkeys
Stimulus > Frequency Offset > Port n > Multiplier
280
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).OFFSet.PORT(Pt).FREQuency. OFFSet
Object Type
Property (read/write)
Data Type
Double
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
The basic frequency range offset when the frequency offset feature is ON.
Range
from –1e–12 to 1e12
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
Hz
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).OFFset.PORT(Pt).FREQuency.OFFSet
app.SCPI.SENSe(Ch).OFFset.PORT(Pt).FREQuency.OFFSet = 1e9
Equivalent Softkeys
Stimulus > Frequency Offset > Port n > Offset
281
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).OFFSet.PORT(Pt).FREQuency. STARt
Object Type
Property (read only)
Data Type
Double
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
The frequency sweep start when the frequency offset feature is ON.
Unit
Hz
Syntax
Dim Value As Double
Value = app.SCPI.SENSe(Ch).OFFset.PORT(Pt).FREQuency.STARt
Equivalent Softkeys
Stimulus > Frequency Offset > Port n > Start
282
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).OFFSet.PORT(Pt).FREQuency. STOP
Object Type
Property (read only)
Data Type
Double
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
The frequency sweep stop when the frequency offset feature is ON.
Unit
Hz
Syntax
Dim Value As Double
Value = app.SCPI.SENSe(Ch).OFFset.PORT(Pt).FREQuency.STOP
Equivalent Softkeys
Stimulus > Frequency Offset > Port n > Stop
283
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).OFFSet.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the frequency offset feature.
Allowable Values
True: frequency offset feature is ON
False: frequency offset feature is OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.SENSe(Ch).OFFSet.STATe
app.SCPI.SENSe(Ch).OFFSet.STATe = True
Equivalent Softkeys
Stimulus > Frequency Offset > Frequency Offset
284
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).ROSCillator.SOURce
Object Type
Property (read/write)
Data Type
String
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Selects the internal or external source of the reference frequency of 10 MHz.
"INTernal" : Internal source of the reference frequency
Range
"EXTernal" : External source of the reference frequency
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
An error occurs. Error code: 224.
Preset Value
"INT"
Dim Param As String
Syntax
Param = app.SCPI.SENSe(Ch).ROSCillator.SOURce
app.SCPI.SENSe(Ch).ROSCillator.SOURce = "EXT"
Equivalent Softkeys
System > Misc Setup > Ref Source
285
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).SEGMent.DATA
Object Type
Property (read/write)
Data Type
Variant (array of double)
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
The array of the segment sweep table.
The array has the following format:
{ <Buf>, <Flag1>, <Flag2>, <Flag3>, <Flag4>, <Flag5>, <N>,
<Start(1)>, <Stop(1)>, <NOP(1)> [,<IFBW(1)>] [,<Pow(1)>] [,<Del(1)>] [,<Time(1)>],
<Start(2)>, <Stop(2)>, <NOP(2)> [,<IFBW(2)>] [,<Pow(2)>] [,<Del(2)>] [,<Time(2)>],
…
<Start(N)>, <Stop(N)>, <NOP(N)> [,<IFBW(N)>] [,<Pow(N)>] [,<Del(N)>] [,<Time(N)>] }
Description
<Buf>
<Flag1>
<Flag2>
<Flag3>
<Flag4>
<Flag5>
<N>
<Start n>
<Stop n>
<NOP n>
<IFBW n>
<Pow n>
<Del n>
: Always 5,
: Stimulus start setting (0 – start/stop, 1 – center/span),
: Setting of the <IFBW> field (0 – disabled, 1 – enabled),
: Setting of the <Pow> field (0 – disabled, 1 – enabled),
: Setting of the <Del> field (0 – disabled, 1 – enabled),
: Setting of the <Time> field (0 – disabled, 1 – enabled),
: Number of segments,
: Start value of the n–th segment,
: Stop value of the n–th segment,
: Number of points of the n–th segment,
: IF bandwidth of the n–th segment (if enabled),
: Power of the n–th segment (if enabled),
: Measurement delay of the n–th segment (if enabled),
<Time n>
: Reserved for future use (if enabled).
Dim Data As Variant
Syntax
Data = app.SCPI.SENSe(Ch).SEGMent.DATA
app.SCPI.SENSe(Ch).SEGMent.DATA = Data
Equivalent Softkeys
Stimulus / Segment Table
286
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).SWEep.POINt.TIME
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The value of the delay before measurement in each measurement point.
Range
from 0 to 0.3
Resolution
5E-6
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).SWEep.POINt.TIME
app.SCPI.SENSe(Ch).SWEep.POINt.TIME = 5Е-6
Equivalent Softkeys
Stimulus > Meas Delay
287
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).SWEep.POINts
Object Type
Property (read/write)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The number of measurement points.
Range
2 to 500001 (PLANAR 804/1)
2 to 200001 (PLANAR 304/1)
2 to 200001 (S5048)
2 to 200001 (S7530)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
201
Dim Value As Long
Syntax
Value = app.SCPI.SENSe(Ch).SWEep.POINts
app.SCPI.SENSe(Ch).SWEep.POINts = 1001
Equivalent Softkeys
Stimulus > Points
288
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SENSe(Ch).SWEep.TYPE
Object Type
Property (read/write)
Data Type
String
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
Sets the sweep type.
"LINear"
: Linear frequency sweep
"LOGarithmic" : Logarithmic frequency sweep
Range
"SEGMent"
: Segment frequency sweep
"POWer "
: Power sweep
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
An error occurs. Error code: 206.
Preset Value
"LIN"
Dim Param As String
Syntax
Param = app.SCPI.SENSe(Ch).SWEep.TYPE
app.SCPI.SENSe(Ch).SWEep.TYPE = "LOG"
Equivalent Softkeys
Stimulus > Sweep Type
289
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SERVice.CHANnel.ACTive
Object Type
Property (read only)
Data Type
Long
Target
Instrument
Description
The number of the active channel.
Syntax
Dim Value As Long
Value = app.SCPI.SERVice.CHANnel.ACTive
Equivalent Softkeys
None
SCPI.SERVice.CHANnel.COUNt
Object Type
Property (read only)
Data Type
Long
Target
Instrument
Description
The maximum number of the channels.
Syntax
Dim Value As Long
Value = app.SCPI.SERVice.CHANnel.COUNt
Equivalent Softkeys
None
290
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SERVice.CHANnel(Ch).TRACe.ACTive
Object Type
Property (read only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The active trace number of the channel.
Syntax
Dim Value As Long
Value = app.SCPI.SERVice.CHANnel(Ch).TRACe.ACTive
Equivalent Softkeys
None
SCPI.SERVice.CHANnel.TRACe.COUNt
Object Type
Property (read only)
Data Type
Long
Target
Instrument
Description
The maximum number of the traces in the channel.
Syntax
Dim Value As Long
Value = app.SCPI.SERVice.CHANnel.TRACe.COUNt
Equivalent Softkeys
None
291
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SERVice.PORT.COUNt
Object Type
Property (read only)
Data Type
Long
Target
Instrument
Description
The number of the ports.
Syntax
Dim Value As Long
Value = app.SCPI.SERVice.PORT.COUNt
Equivalent Softkeys
None
SCPI.SERVice.SWEep.FREQency.MAXimum
Object Type
Property (read only)
Data Type
Double
Target
Instrument
Description
The upper limit of the measurement frequency.
Syntax
Dim Value As Double
Value = app.SCPI.SERVice.SWEep.FREQency.MAXimum
Equivalent Softkeys
None
292
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SERVice.SWEep.FREQency.MINimum
Object Type
Property (read only)
Data Type
Double
Target
Instrument
Description
The lower limit of the measurement frequency.
Syntax
Dim Value As Double
Value = app.SCPI.SERVice.SWEep.FREQency.MINimum
Equivalent Softkeys
None
SCPI.SERVice.SWEep.POINts
Object Type
Property (read only)
Data Type
Double
Target
Instrument
Description
The maximum number of the measurement points.
Syntax
Dim Value As Long
Value = app.SCPI.SERVice.SWEep.POINts
Equivalent Softkeys
None
293
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.CENTer
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The center value of the power sweep range.
Range
from –55 to 10 (PLANAR 304/1 and PLANAR 804/1)
from –50 to 5 (S5048 and S7530)
Resolution
0.025
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
–22.5
Unit
dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.SOURce(Ch).POWer.CENTer
app.SCPI.SOURce(Ch).POWer.CENTer = 5
Equivalent Softkeys
Stimulus > Center
294
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.LEVel.IMMediate.AMPLitude
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The power level for the frequency sweep.
Range
from –55 to 10 (PLANAR 304/1 and PLANAR 804/1)
from –50 to 5 (S5048 and S7530)
Resolution
0.05
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.SOURce(Ch).POWer.LEVel.IMMediate.AMPLitude
app.SCPI.SOURce(Ch).POWer.LEVel.IMMediate.AMPLitude = 10
Equivalent Softkeys
Stimulus > Power > Power
295
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.LEVel.SLOPe.DATA
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The power slope value for the frequency sweep.
Range
from –2 to 2
Resolution
0.1
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
dB/GHz (decibel/gigahertz)
Dim Value As Double
Syntax
Value = app.SCPI.SOURce(Ch).POWer.LEVel.SLOPe.DATA
app.SCPI.SOURce(Ch).POWer.LEVel.SLOPe.DATA = 0.2
Equivalent Softkeys
Stimulus > Power > Slope [dB/GHz]
296
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.LEVel.SLOPe.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The ON/OFF state of the power slope for the frequency sweep.
Allowable Values
True: Power slope ON
False: Power slope OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.SOURce(Ch).POWer.LEVel.SLOPe.STATe
app.SCPI.SOURce(Ch).POWer.LEVel.SLOPe.STATe = True
Equivalent Softkeys
Stimulus > Power > Slope [ON/OFF]
297
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.COLLect.ACQuire
Object Type
Method
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
Measures the power calibration data for the specified port using the power
meter controlled via USB or USB/GPIB. Calculates calibration coefficients on
completion of the measurement, and turns ON the power correction for the
port.
Syntax
app.SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.COLLect.ACQuire
Equivalent Softkeys
Calibration > Power Calibration > Calibrate
298
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.COLLect.TABLe.
LOSS.DATA
Object Type
Property (read/write)
Data Type
Variant (array of double)
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
The loss compensation table used during the power calibration.
The array size is 1+2N, where N is the number of the table rows.
For the n–th point, where n from 1 to N:
Description
Data(0)
the number of the table rows N integer from 0 to 100;
Data(2n–1)
the frequency of the n–th row of the table from 300 kHz
to 8.0 GHz;
Data(2n)
the loss value of the n-th table row in dB.
Dim Data As Variant
Data =
Syntax
app.SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.COLLect.TABLe.LOSS.DATA
app.SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.COLLect.TABLe.LOSS.DATA
= Data
Related Commands
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.COLLect.ACQuire
Equivalent Softkeys
Calibration > Power Calibration > Loss Compen
299
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.COLLect.TABLe.
LOSS.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
The ON/OFF state of the loss compensation function used during the power
calibration.
Allowable Values
True: Loss compensation ON
False: Loss compensation OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status
=
app.SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.COLLect.TABLe.
LOSS.STATe
app.SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.COLLect.TABLe.LOSS.STATe
= True
Related Commands
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.COLLect.TABLe.LOSS.DATA
Equivalent Softkeys
Calibration > Power Calibration > Loss Compen > Compensation
300
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.DATA
Object Type
Property (read/write)
Data Type
Variant (array of double)
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
The power correction table.
The array size is 1+2N, where N is the number of the table rows.
For the n–th point, where n from 1 to N:
Description
Data(0)
the number of the table rows N integer from 0 to 10001;
Data(2n–1)
the frequency of the n–th row of the table from 300 kHz
to 8.0 GHz;
Data(2n)
power correction value of the n–th row of the table from
–10 to +10 dB.
Dim Data As Variant
Syntax
Data = app.SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.DATA
app.SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.DATA = Data
Related Commands
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.COLLect.ACQuire
Equivalent Softkeys
None
301
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Port Pt of channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Pt:
port number 1–2 (see Table 2 on page 24)
Description
The ON/OFF state of the power correction function.
Allowable Values
True: Power correction ON
False: Power correction OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.STATe
app.SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.STATe = True
Related Commands
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.COLLect.ACQuire
Equivalent Softkeys
Calibration > Power Calibration > Correction
302
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.SPAN
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The power span for the power sweep.
Range
from 0 to 65
Resolution
0.05
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
65
Unit
dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.SOURce(Ch).POWer.SPAN
app.SCPI.SOURce(Ch).POWer.SPAN = 50
Equivalent Softkeys
Stimulus > Span
303
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.STARt
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The power sweep start for the power sweep.
Range
from –55 to 10 (PLANAR 304/1 and PLANAR 804/1)
from –50 to 5 (S5048 and S7530)
Resolution
0.05
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
–55 (PLANAR 304/1 and PLANAR 804/1)
–50 (S5048 and S7530)
Unit
dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.SOURce(Ch).POWer.STARt
app.SCPI.SOURce(Ch).POWer.STARt = 5
Equivalent Softkeys
Stimulus > Start
304
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.STOP
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Description
The power sweep stop for the power sweep.
Range
from –55 to 10 (PLANAR 304/1 and PLANAR 804/1)
from –50 to 5 (S5048 and S7530)
Resolution
0.05
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
10 (PLANAR 304/1 and PLANAR 804/1)
5 (S5048 and S7530)
Unit
dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.SOURce(Ch).POWer.STOP
app.SCPI.SOURce(Ch).POWer.STOP = 5
Equivalent Softkeys
Stimulus > Stop
305
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.OPERation.CONDition
Object Type
Property (read only)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Operation Status Condition Register.
Dim Value As Long
Syntax
Value = app.SCPI.STATus.OPERation.CONDition
Equivalent Softkeys
None
SCPI.STATus.OPERation.ENABle
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Operation Status Enable Register.
Range
from 0 to 65535
Preset Value
0
Dim Value As Long
Syntax
Value = app.SCPI.STATus.OPERation.ENABle
app.SCPI.STATus.OPERation.ENABle = Value
Equivalent Softkeys
None
306
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.OPERation.EVENt
Object Type
Property (read only)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Operation Status Event Register.
Dim Value As Long
Syntax
Value = app.SCPI.STATus.OPERation.EVENt
Equivalent Softkeys
None
SCPI.STATus.OPERation.NTRansition
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Negative Transition Filter of the Operation Status Register.
Range
from 0 to 65535
Preset Value
0
Dim Value As Long
Syntax
Value = app.SCPI.STATus.OPERation.NTRansition
app.SCPI.STATus.OPERation.NTRansition = Value
Equivalent Softkeys
None
307
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.OPERation.PTRansition
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Positive Transition Filter of the Operation Status Register.
Range
from 0 to 65535
Preset Value
65535
Dim Value As Long
Syntax
Value = app.SCPI.STATus.OPERation.PTRansition
app.SCPI.STATus.OPERation.PTRansition = Value
Equivalent Softkeys
None
SCPI.STATus.PRESet
Object Type
Method
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Initialization of all registers.
Syntax
app.SCPI.STATus.PRESet
Equivalent Softkeys
None
308
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.QUEStionable.CONDition
Object Type
Property (read only)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Questionable Status Condition Register.
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.CONDition
Equivalent Softkeys
None
SCPI.STATus.QUEStionable.ENABle
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Questionable Status Enable Register.
Range
from 0 to 65535
Preset Value
0
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.ENABle
app.SCPI.STATus.QUEStionable.ENABle = Value
Equivalent Softkeys
None
309
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.QUEStionable.EVENt
Object Type
Property (read only)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Questionable Status Event Register.
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.EVENt
Equivalent Softkeys
None
SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).CONDition
Object Type
Property (read only)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Questionable Limit Channel Status Condition Register for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).CONDition
Equivalent Softkeys
None
310
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).ENABle
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Questionable Limit Channel Status Enable Register for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Range
from 0 to 65535
Preset Value
0
Dim Value As Long
Value = app.SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).ENABle
Syntax
app.SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).ENABle = Value
Equivalent Softkeys
None
SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).EVENt
Object Type
Property (read only)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Questionable Limit Channel Status Event Register for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Dim Value As Long
Syntax
Value = app. SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).EVENt
Equivalent Softkeys
None
311
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).NTRansition
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Negative Transition Filter of the Questionable Limit Channel Status Register
for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Range
from 0 to 65535
Preset Value
0
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).NTRansition
app.SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).NTRansition = Value
Equivalent Softkeys
None
312
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).PTRansition
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Positive Transition Filter of the Questionable Limit Channel Status Register for
channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Range
from 0 to 65535
Preset Value
65535
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).PTRansition
app.SCPI.STATus.QUEStionable.LIMit.CHANnel(Ch).PTRansition = Value
Equivalent Softkeys
None
SCPI.STATus.QUEStionable.LIMit.CONDition
Object Type
Property (read only)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1).
Description
Questionable Limit Status Condition Register.
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.LIMit.CONDition
Equivalent Softkeys
None
313
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.QUEStionable.LIMit.ENABle
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Questionable Limit Status Enable Register.
Range
from 0 to 65535
Preset Value
0
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.LIMit.ENABle
app.SCPI.STATus.QUEStionable.LIMit.ENABle = Value
Equivalent Softkeys
None
SCPI.STATus.QUEStionable.LIMit.EVENt
Object Type
Property (read only)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Questionable Limit Status Event Register.
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.LIMit.EVENt
Equivalent Softkeys
None
314
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.QUEStionable.LIMit.NTRansition
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Negative Transition Filter of the Questionable Limit Status Register.
Range
from 0 to 65535
Preset Value
0
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.LIMit.NTRansition
app.SCPI.STATus.QUEStionable.LIMit.NTRansition = Value
Equivalent Softkeys
None
315
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.QUEStionable.LIMit.PTRansition
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Positive Transition Filter of the Questionable Limit Status Register.
Range
from 0 to 65535
Preset Value
65535
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.LIMit.PTRansition
app.SCPI.STATus.QUEStionable.LIMit.PTRansition = Value
Equivalent Softkeys
None
316
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.QUEStionable.NTRansition
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Negative Transition Filter of the Questionable Status Register.
Range
from 0 to 65535
Preset Value
0
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.NTRansition
app.SCPI.STATus.QUEStionable.NTRansition = Value
Equivalent Softkeys
None
317
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.QUEStionable.PTRansition
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Positive Transition Filter of the Questionable Status Register.
Range
from 0 to 65535
Preset Value
65535
Dim Value As Long
Value = app.SCPI.STATus.QUEStionable.PTRansition
Syntax
app.SCPI.STATus.QUEStionable.PTRansition = Value
Equivalent Softkeys
None
SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).CONDition
Object Type
Property (read only)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Questionable Ripple Limit Channel Status Condition Register for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).CONDition
Equivalent Softkeys
None
318
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).ENABle
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Questionable Ripple Limit Channel Status Enable Register for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Range
from 0 to 65535
Preset Value
0
Dim Value As Long
Value = app.SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).ENABle
Syntax
app.SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).ENABle = Value
Equivalent Softkeys
None
SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).EVENt
Object Type
Property (read only)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Questionable Ripple Limit Channel Status Event Register for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).EVENt
Equivalent Softkeys
None
319
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).NTRansition
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Negative Transition Filter of the Questionable Ripple Limit Channel Status
Register for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Range
from 0 to 65535
Preset Value
0
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).NTRansition
app.SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).NTRansition = Value
Equivalent Softkeys
None
320
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).PTRansition
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Positive Transition Filter of the Questionable Ripple Limit Channel Status
Register for channel Ch,
Ch:
channel number 1–16 (see Table 1 on page 24)
Range
from 0 to 65535
Preset Value
65535
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).PTRansition
app.SCPI.STATus.QUEStionable.RLIMit.CHANnel(Ch).PTRansition = Value
Equivalent Softkeys
None
SCPI.STATus.QUEStionable.RLIMit.CONDition
Object Type
Property (read only)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Questionable Ripple Limit Status Condition Register.
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.RLIMit.CONDition
Equivalent Softkeys
None
321
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.QUEStionable.RLIMit.ENABle
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Questionable Ripple Limit Status Enable Register.
Range
from 0 to 65535
Preset Value
0
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.RLIMit.ENABle
app.SCPI.STATus.QUEStionable.RLIMit.ENABle = Value
Equivalent Softkeys
None
SCPI.STATus.QUEStionable.RLIMit.EVENt
Object Type
Property (read only)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Questionable Ripple Limit Status Event Register.
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.RLIMit.EVENt
Equivalent Softkeys
None
322
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.QUEStionable.RLIMit.NTRansition
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Negative Transition Filter of the Questionable Ripple Limit Status Register.
Range
from 0 to 65535
Preset Value
0
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.RLIMit.NTRansition
app.SCPI.STATus.QUEStionable.RLIMit.NTRansition = Value
Equivalent Softkeys
None
323
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.STATus.QUEStionable.RLIMit.PTRansition
Object Type
Property (read/write)
Data Type
Long
Target
IEE488.2 Status Reporting System (Appendix 1)
Description
Positive Transition Filter of the Questionable Ripple Limit Status Register.
Range
from 0 to 65535
Preset Value
65535
Dim Value As Long
Syntax
Value = app.SCPI.STATus.QUEStionable.RLIMit.PTRansition
app.SCPI.STATus.QUEStionable.RLIMit.PTRansition = Value
Equivalent Softkeys
None
324
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SYSTem.BEEPer.COMPlete.IMMediate
Object Type
Method
Target
Instrument
Description
Generates a beep to notify of the completion of the operation.
Syntax
app.SCPI.SYSTem.BEEPer.COMPlete.IMMediate
Equivalent Softkeys
System > Misc Setup > Beeper > Test Beep Complete
SCPI.SYSTem.BEEPer.COMPlete.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Instrument
Description
The ON/OFF state of the beeper notifying of the completion of the operation.
Allowable Values
True: Completion beeper ON
False: Completion beeper OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI.SYSTem.BEEPer.COMPlete.STATe
app.SCPI.SYSTem.BEEPer.COMPlete.STATe = False
Equivalent Softkeys
System > Misc Setup > Beeper > Beep complete
325
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SYSTem.BEEPer.WARNing.IMMediate
Object Type
Method
Target
Instrument
Description
Generates a beep to notify of warning.
Syntax
app.SCPI.SYSTem.BEEPer.WARNing.IMMediate
Equivalent Softkeys
System > Misc Setup > Beeper > Test Beep Warning
SCPI.SYSTem.BEEPer.WARNing.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Instrument
Description
The ON/OFF state of the beeper notifying of warning.
Allowable Values
True: Warning beeper ON
False: Warning beeper OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI.SYSTem.BEEPer.WARNing.STATe
app.SCPI.SYSTem.BEEPer.WARNing.STATe = False
Equivalent Softkeys
System > Misc Setup > Beeper > Beep Warning
326
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SYSTem.COMMunicate.ECAL.TEMPerature.SENSor
Object Type
Property (read only)
Data Type
Double
Target
AutoCal module
Description
Reads out the temperature of the AutoCal module connected to the instrument.
Unit
°C (Celsius)
Dim Value As Double
Syntax
Value = app.SCPI.SYSTem.COMMunicate.ECAL.TEMPerature.SENSor
Equivalent Softkeys
None
327
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SYSTem.COMMunicate.ECAL.IMPedance(Pt)
Object Type
Property (read/write)
Data Type
String
Target
AutoCal module
Description
Sets or reads out the impedance state of the specified port (Pt) of AutoCal
module connected to the instrument.
"OPEN" : OPEN impedance state
"SHORt" : SHORT impedance state
Range
"LOAD" : LOAD impedance state
"LOAD"
Preset Value
Dim Param As String
Param = app.SCPI.SYSTem.COMMunicate.ECAL.IMPedance(Pt)
Syntax
app.SCPI.SYSTem.COMMunicate.ECAL.IMPedance(Pt) = "OPEN"
Equivalent Softkeys
None
SCPI.SYSTem.COMMunicate.ECAL.THRU(Pt1, Pt2)
Object Type
Method
Target
AutoCal module
Description
Sets the thru state between the specified 2 ports (Pt1, Ptr2) of AutoCal module
connected to the instrument.
Syntax
app.SCPI.SYSTem.COMMunicate.ECAL.THRU(1, 2)
Equivalent Softkeys
None
328
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SYSTem.CORRection.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Instrument
Description
The ON/OFF state of the system error correction.
Allowable Values
True: System error correction ON
False: System error correction OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI.SYSTem.CORRection.STATe
app.SCPI.SYSTem.CORRection.STATe = False
Equivalent Softkeys
System > Misc Setup > System Correction
329
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SYSTem.DATE
Object Type
Property (read/write)
Data Type
Variant (array of long)
Target
Instrument
The current date.
The array consists of three elements:
Description
Data(0)
year from 1900 to 2100;
Data(1)
month from 1 to 12;
Data(2)
day from 1 to 31.
Dim Data As Variant
Syntax
Data = app.SCPI.SYSTem.DATE
app. app.SCPI.SYSTem.DATE = Array(2009, 9, 9)
Equivalent Softkeys
None
SCPI.SYSTem.LOCal
Object Type
Method
Target
Instrument
Description
Sets the instrument to the local operation mode, when all the keys on the front
panel, mouse and the touch screen are active.
Syntax
app.SCPI.SYSTem.LOCal
Related Commands
SCPI.SYSTem.REMote
SCPI.SYSTem.RWLock
Equivalent Softkeys
None
330
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
331
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SYSTem.PRESet
Object Type
Method
Target
Instrument
Description
Resets the instrument to the factory settings. The difference from the
SCPI.IEEE4882.RST: method is that the trigger is set to the Continuous trigger
mode.
Syntax
app.SCPI.SYSTem.PRESet
Equivalent Softkeys
System > Preset > OK
SCPI.SYSTem.REMote
Object Type
Method
Target
Instrument
Description
Sets the instrument to the remote operation mode, when all the keys on the
front panel, mouse and the touch screen are not active, except for one key
labeled Return to Local. Pushing this button will reset the instrument to the
local operation mode.
Syntax
app.SCPI.SYSTem.REMote
Related Commands
SCPI.SYSTem.LOCal
SCPI.SYSTem.RWLock
Equivalent Softkeys
None
332
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SYSTem.RWLock
Object Type
Method
Target
Instrument
Description
Sets the instrument to the remote operation mode, when all the keys on the
front panel, mouse and the touch screen are not active. Only
SCPI.SYSTem.LOCal or SCPI.SYSTem.REMote command can release this
remote operation mode.
Syntax
app.SCPI.SYSTem.RWLock
Related Commands
SCPI.SYSTem.LOCal
SCPI.SYSTem.REMote
Equivalent Softkeys
None
SCPI.SYSTem.TEMPerature.SENSor(Idx)
Object Type
Property (read only)
Data Type
Double
Target
Instrument
Description
Reads out the specified sensor (Idx) temperature inside the instrument.
Unit
°C (Celsius)
Dim Value As Double
Syntax
Value = app.SCPI.SYSTem.TEMPerature.SENSor(1)
Equivalent Softkeys
None
333
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.SYSTem.TIME
Object Type
Property (read/write)
Data Type
Variant (array of long)
Target
Instrument
The current time.
The array consists of three elements:
Description
Data(0)
hours from 0 to 23;
Data(1)
minutes from 0 to 59;
Data(2)
seconds from 0 to 59.
Dim Data As Variant
Syntax
Data = app.SCPI.SYSTem.TIME
app. app.SCPI.SYSTem.TIME = Array(15, 20, 30)
Equivalent Softkeys
None
334
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.TRIGger.SEQuence.IMMediate
Object Type
Method
Target
Instrument
Generates a trigger, independently of the trigger source setting (except for the
External). If the trigger source is set to External, an error occurs (error code
221) and the command is ignored.
Description
If the instrument is not in the waiting for a trigger state (sweep is in progress or
all the channels are set to Hold), an error occurs (error code 211) and the
command is ignored.
The method returns control before the end of the sweep.
Syntax
app.SCPI.TRIGger.SEQuence.IMMediate
Related Commands
SCPI.TRIGger.SEQuence.SOURce
SCPI.INITiate(Ch).CONTinuous
SCPI.INITiate(Ch).IMMediate
Equivalent Softkeys
None
335
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.TRIGger.SEQuence.SINGle
Object Type
Method
Target
Instrument
Generates a trigger, independently of the trigger source setting (except for the
External). If the trigger source is set to External, an error occurs (error code
221) and the command is ignored.
Description
If the instrument is not in the waiting for a trigger state (sweep is in progress or
all the channels are set to Hold), an error occurs (error code 211) and the
command is ignored.
The method does not return control before the end of the sweep (waiting for the
completion of the sweep of all the channels).
Syntax
app.SCPI.TRIGger.SEQuence.SINGle
Related Commands
SCPI.TRIGger.SEQuence.SOURce
SCPI.INITiate(Ch).CONTinuous
SCPI.INITiate(Ch).IMMediate
Equivalent Softkeys
None
336
PLANAR 304/1, 804/1, S5048, S7530
COM/DCOM Programming manual
SCPI.TRIGger.SEQuence.SOURce
Object Type
Property (read/write)
Data Type
String
Target
Instrument
Description
Selects the sweep trigger source.
"INTernal" : Internal
"EXTernal" : External
Range
"MANual" : Manual
"BUS"
: Bus
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
An error occurs. Error code: 205.
Preset Value
"INT"
Dim Param As String
Syntax
Param = app.SCPI.TRIGger.SEQuence.SOURce
app.SCPI.TRIGger.SEQuence.SOURce = "BUS"
Related Commands
SCPI.TRIGger.SEQuence.IMMediate
SCPI.TRIGger.SEQuence.SINGle
SCPI.IEEE4882.TRG
Equivalent Softkeys
Stimulus > Trigger > Trigger Source > Internal | External | Manual | Bus
337
Appendix 1. IEE488.2 Status Reporting System
Σ
...
Service Request Enable Register
*SRE
7
SRQ
5
4
3
MAV
QUES
2
1
0
Status Byte Register
*STB?
RQS
Service
Request
Generation
OPER
ESB
EAV
MSS
128
64
32
16
8
4
Σ
...
Standard Event Status Enable Register
*ESE
7
6
5
4
Comman
Execution
d
Error
Error
32
16
3
2
Device
spec.
Error
Query
Error
8
1
Operation
Complete
4
a
0
Standard Event Status Register
*ESR?
Output
Queue
1
Error/
Event
Queue
Σ
...
Operation Status Enable Register
STAT:OPER:ENAB
Operation Status Event Register
STAT:OPER?
Positive Transition Filter
STAT:OPER:PTR
15
14
13
12
11
10
9
8
7
6
5
4
Waiting
for
trigger
Meas.
32
16
3
2
1
0
Negative Transition Filter
STAT:OPER:NTR
Operation Status Condition Register
STAT:OPER:COND?
Appendix 1. IEE488.2 Status Reporting System
a
Σ
...
Questionable Status Enable Register
STAT:QUES:ENAB
Questionable Status Event Register
STAT:QUES?
Positive Transition Filter
STAT:QUES:PTR
15
14
13
12
11
10
Limit
Test
FAIL
9
Ripple
Test
FAIL
1024
b
512
c
8
7
6
5
4
3
2
1
0
Negative Transition Filter
STAT:QUES:NTR
Questionable Status Condition
Register
STAT:QUES:COND?
Appendix 1. IEE488.2 Status Reporting System
b
Σ
...
Questionable Limit Status Enable
Register
STAT:QUES:LIM:ENAB
Questionable Limit Status Event
Register
STAT:QUES:LIM?
Positive Transition Filter
STAT:QUES:LIM:PTR
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
Chan 14 Chan 13 Chan 12 Chan 11 Chan 10 Chan 9
Chan 8
Chan 7
Chan 6
Chan 5
Chan 4
Chan 3
Chan 2
Chan 1
Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
16384
d14
4
.
.
.
.
.
.
.
.
.
.
.
d
2
d
0
Negative Transition Filter
STAT:QUES:LIM:NTR
Questionable Limit Status Condition
Register
STAT:QUES:LIM:COND?
Appendix 1. IEE488.2 Status Reporting System
d
Σ
...
Questionable Limit Channel 1 Status
Enable Register
STAT:QUES:LIM:CHAN1:ENAB
Questionable Limit Channel 1 Status
Event Register
STAT:QUES:CHAN1:LIM?
Positive Transition Filter
STAT:QUES:LIM:CHAN1:PTR
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
Questionable Limit Channel 1 Status
Condition Register
STAT:QUES:LIM:CHAN1:COND?
Trace 14 Trace 13 Trace 12 Trace 11 Trace 10 Trace 9
Trace 8
Trace 7
Trace 6
Trace 5
Trace 4
Trace 3
Trace 2
Trace 1
Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
16384
8192
4096
2048
1024
512
256
128
64
32
16
8
4
Negative Transition Filter
STAT:QUES:LIM:CHAN1:NTR
2
...
d14
Σ
...
Questionable Limit Channel 14 Status
Enable Register
STAT:QUES:LIM:CHAN14:ENAB
Questionable Limit Channel 14 Status
Event Register
STAT:QUES:CHAN14:LIM:?
Positive Transition Filter
STAT:QUES:LIM:CHAN14:PTR
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
Trace 14 Trace 13 Trace 12 Trace 11 Trace 10 Trace 9
Trace 8
Trace 7
Trace 6
Trace 5
Trace 4
Trace 3
Trace 2
Trace 1
Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test Limit Test
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
FAIL
16384
8192
4096
2048
1024
512
256
128
64
32
16
8
4
2
0
Negative Transition Filter
STAT:QUES:LIM:CHAN14:NTR
Questionable Limit Channel 14 Status
Condition Register
STAT:QUES:LIM:CHAN14:COND?
Appendix 1. IEE488.2 Status Reporting System
c
Σ
...
Questionable Ripple Limit Status
Enable Register
STAT:QUES:RLIM:ENAB
Questionable Ripple Limit Status
Event Register
STAT:QUES:RLIM?
Positive Transition Filter
STAT:QUES:RLIM:PTR
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
Chan 14 Chan 13 Chan 12 Chan 11 Chan 10 Chan 9
Chan 8
Chan 7
Chan 6
Chan 5
Chan 4
Chan 3
Chan 2
Chan 1
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL
16384
e14
4
.
.
.
.
.
.
.
.
.
.
.
e
2
e
0
Negative Transition Filter
STAT:QUES:RLIM:NTR
Questionable Ripple Limit Status
Condition Register
STAT:QUES:RLIM:COND?
Appendix 1. IEE488.2 Status Reporting System
e
Σ
...
Questionable Ripple Limit Channel 1
Status Enable Register
STAT:QUES:RLIM:CHAN1:ENAB
Questionable Ripple Limit Channel 1
Status Event Register
STAT:QUES:CHAN1:RLIM?
Positive Transition Filter
STAT:QUES:RLIM:CHAN1:PTR
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
Questionable Ripple Limit Channel 1
Status Condition Register
STAT:QUES:RLIM:CHAN1:COND?
Trace 14 Trace 13 Trace 12 Trace 11 Trace 10 Trace 9
Trace 8
Trace 7
Trace 6
Trace 5
Trace 4
Trace 3
Trace 2
Trace 1
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL
16384
8192
4096
2048
1024
512
256
128
64
32
16
8
4
Negative Transition Filter
STAT:QUES:RLIM:CHAN1:NTR
2
...
e14
Σ
...
Questionable Ripple Limit Channel 14
Status Enable Register
STAT:QUES:RLIM:CHAN14:ENAB
Questionable Ripple Limit Channel 14
Status Event Register
STAT:QUES:CHAN14:RLIM:?
Positive Transition Filter
STAT:QUES:RLIM:CHAN14:PTR
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
Trace 14 Trace 13 Trace 12 Trace 11 Trace 10 Trace 9
Trace 8
Trace 7
Trace 6
Trace 5
Trace 4
Trace 3
Trace 2
Trace 1
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Ripple
Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL Test FAIL
16384
8192
4096
2048
1024
512
256
128
64
32
16
8
4
2
0
Negative Transition Filter
STAT:QUES:RLIM:CHAN14:NTR
Questionable Ripple Limit Channel 14
Status Condition Register
STAT:QUES:RLIM:CHAN14:COND?
Appendix 2. Error Codes
114
"Header suffix out of range"
200
"Execution error"
211
"Trigger ignored"
213
"Init ignored"
220
"Parameter Error"
222
"Data out of range"
224
"Illegal parameter value"
201
"Invalid channel index"
202
"Invalid trace index"
203
"Invalid marker index"
204
"Marker is not active"
205
"Invalid save type specifier"
206
"Invalid sweep type specifier"
207
"Invalid trigger source specifier"
208
"Invalid measurement parameter specifier"
209
"Invalid format specifier"
210
"Invalid data math specifier"
214
"Invalid limit data"
215
"Invalid segment data"
216
"Invalid standard type specifier"
217
"Invalid conversion specifier"
218
"Invalid gating shape specifier"
219
"Invalid gating type specifier"
300
"Device-specific error"
302
"Status reporting system error"
344
Appendix 3. Programming Examples
Example 1. Instrument Information String Readout
The following program reads out and displays on the screen the instrument information
string – the Name property of the COM object. The string contains the following fields:
Manufacturer, Model, Serial Number, Software Version/Firmware Version
For example:
CMT, Planar 304/1, 09381020, 2.46/03
Dim app As Object
Sub Example1()
Set app = CreateObject("Planar304.Application")
ID = app.Name
MsgBox ("Information string read out: " + ID)
End Sub
345
Example 2. Checking the Instrument Ready State
Normally, the user control program starts when the Planar.exe application is running, the
instrument booting is completed, and the instrument is ready for use. In some cases, it is
recommended to check if the instrument is ready for use. The instrument may be not
ready for use if it is not connected to PC via USB cable. Moreover, if the Planar.exe
application has not been started in advance, the CreateObject function will automatically
start the application and then within about 10 seconds the instrument booting will be in
progress. The instrument will not be ready for use until the booting is completed. The
Ready property is used to check if the instrument is ready for use.
The following program checks the Ready property right after a COM object has been
created. If the Planar.exe application has been started in advance and the booting is
completed, “Analyzer is ready” will be displayed. If the Ready property value is False,
10 second delay is activated for the case the Planar.exe application has been started by
the COM object creation. In 10 seconds the program rechecks the Ready property. If the
value is True, “Analyzer is ready” will be displayed, if otherwise, “Analyzer is not
ready” will be displayed, what means the instrument is not connected to LAN or it is not
connected to PC via USB cable.
Dim app As Object
Sub Example2()
Set app = CreateObject("Planar804.Application")
If app.Ready = False Then
Application.Wait (Now + TimeValue("0:00:10"))
If app.Ready = False Then
MsgBox ("Analyzer is not ready")
Exit Sub
End If
End If
MsgBox ("Analyzer is ready")
End Sub
346
Example 3. Setting the Measurement Parameters
The following program shows the setting of some measurement parameters. First, the
instrument is reset to the factory settings. Then the following parameters are set:
• Two channel windows are opened and allocated one above the other.
• The number of traces is set to 2 in the first channel window, the traces are allocated in
different graphs in the channel window.
• For the first channel the stimulus parameters are set as follows: the frequency range
from 1 MHz to 2 GHz, the number of measurement points 401.
• For the second channel the stimulus parameters are set as follows: the frequency range
from 800 MHz to 900 MHz, the number of points 51, IF bandwidth 100 Hz, stimulus
power –10 dBm.
• In the first channel window: S11 measurement is set for the trace 1, S22 measurement
is set for the trace 2. The Smith chart format is set for the both traces.
• In the second channel window: S21 measurement and logarithmic magnitude format
are set for the single trace. Then the auto scale function is called for this trace.
Dim app As Object
Public Sub Example3()
Set app = CreateObject("Planar804.Application")
app.SCPI.SYSTem.PRESet
app.SCPI.DISPlay.Split = 2
app.SCPI.Calculate(1).Parameter.Count = 2
app.SCPI.DISPlay.Window(1).Split = 3
app.SCPI.SENSe(1).Frequency.Start = 100000
app.SCPI.SENSe(1).Frequency.STOP = 2000000000
app.SCPI.SENSe(1).SWEep.Points = 401
app.SCPI.SENSe(2).Frequency.Start = 800000000
app.SCPI.SENSe(2).Frequency.STOP = 900000000
app.SCPI.SENSe(2).SWEep.Points = 51
app.SCPI.SENSe(2).BANDwidth.RESolution = 100
app.SCPI.Source(2).Power.LEVel.IMMediate.AMPlitude = -10
app.SCPI.Calculate(1).Parameter(1).DEFine = "S11"
app.SCPI.Calculate(1).Parameter(2).DEFine = "S22"
347
app.SCPI.Calculate(1).Parameter(1).Select
app.SCPI.Calculate(1).Selected.Format = "SMIT"
app.SCPI.Calculate(1).Parameter(2).Select
app.SCPI.Calculate(1).Selected.Format = "SMIT"
app.SCPI.Calculate(2).Parameter(1).DEFine = "S21"
app.SCPI.Calculate(2).Parameter(1).Select
app.SCPI.Calculate(2).Selected.Format = "MLOG"
app.SCPI.DISPlay.Window(2).TRACe(1).Y.SCALe.AUTO
End Sub
348
Example 4. Measurement Data Acquisition
The following program shows data array acquisition with further writing into a file. The
program also shows the method of a sweep triggering and waiting for the sweep
completion.
Three variables F, M, P are declared in the second string of the code. They are used for
arrays of frequency values (Hz), magnitude values (dB), and phase values (degree)
respectively.
After the instrument has been reset to the factory settings, two operators are used for the
sweep triggering and waiting for the sweep completion:
app.SCPI.TRIGger.SEQuence.Source = "bus"
app.SCPI.TRIGger.SEQuence.Single
The first operator sets the GPIB/LAN bus command or the COM/DCOM interface
command as a trigger source. It aborts the sweep and switches the instrument to waiting
for a trigger. The second operator is used for a new sweep triggering and waiting for the
sweep completion.
Note
Unlike
the
SCPI.TRIGger.SEQuence.IMMediate
and
SCPI.IEEE4882.TRG commands, which are completed
immediately
after
a
trigger
generation,
the
SCPI.TRIGger.SEQuence.Single command is not completed
until
the
end
of
the
sweep.
Using
the
SCPI.TRIGger.SEQuence.Single command is the simplest way
to set the waiting for the sweep completion.
On completion of the sweep, three arrays are read out: frequency values, magnitude
values and phase values. Before the magnitude and phase arrays are read out, the
corresponding trace format is set.
The array size of frequency F is equal to the number of measurement points, and the
array size of magnitude M and phase P is equal to the double number of measurement
points (see section 11 “Measurement Data Arrays”). In rectangular formats (for
magnitude and phase) the measurement data are real numbers located in even cells of the
array. Odd cells of the array contain 0.
On completion of the program, the frequency, magnitude and phase values for each
measurement point are written string by string into the file named TESTFILE.
349
Dim app As Object
Dim app As Object
Dim F, M, P
Public Sub Example4()
Set app = CreateObject("Planar304.Application")
app.SCPI.SYSTem.PRESet
app.SCPI.TRIGger.SEQuence.Source = "bus"
app.SCPI.TRIGger.SEQuence.Single
F = app.SCPI.SENSe.Frequency.Data
app.SCPI.Calculate.Selected.Format = "MLOG"
M = app.SCPI.Calculate.Selected.Data.FDATa
app.SCPI.Calculate.Selected.Format = "PHASe"
P = app.SCPI.Calculate.Selected.Data.FDATa
Open "TESTFILE" For Output As #1
For i = LBound(F) To UBound(F)
Print #1, F(i), M(i * 2), P(i * 2)
Next i
Close #1
End Sub
350
Example 5. Program Written in C++
The following C++ program represents an example of the measurement parameter
setting, as well as acquisition and display of the measurement data array. The program
also shows a method of the sweep triggering and waiting for the sweep completion.
//--------------------------------------------------------------------------// Simple example of using COM object of Planar804.exe application.
//
// This example is console application. GUI is not used in this example to
// simplify the program. Error proccessing is very restricted too.
//
#include "stdafx.h"
//--------------------------------------------------------------------------// Generate description of COM object of Planar804.exe application.
#import "Planar804.exe" no_namespace
//--------------------------------------------------------------------------int _tmain(int argc, _TCHAR* argv[])
{
IPlanar804Ptr pNWA;
// Pointer to COM object of Planar804.exe
CComVariant Data;
// Variable for measurement data
// Init COM subsystem
HRESULT hr = CoInitialize(NULL);
if(hr != S_OK) return -1;
// Create COM object
hr = pNWA.CreateInstance(__uuidof(Planar804));
if(hr != S_OK) return -1;
// Preset network analyzer
pNWA->SCPI->SYSTem->PRESet();
// Set frequency start to 1 GHz
pNWA->SCPI->GetSENSe(1)->FREQuency->STARt = 1e9;
// Set frequency stop to 2 GHz
pNWA->SCPI->GetSENSe(1)->FREQuency->STOP = 2e9;
// Set number of measurement points to 51
pNWA->SCPI->GetSENSe(1)->SWEep->POINts = 51;
// Set measured parameter to S21
pNWA->SCPI->GetCALCulate(1)->GetPARameter(1)->DEFine = "S21";
// Set trigger source to GPIB/LAN bus or COM interface
pNWA->SCPI->TRIGger->SEQuence->SOURce = "bus";
// Trigger measurement and wait
pNWA->SCPI->TRIGger->SEQuence->SINGle();
// Get measurement data (array of complex numbers)
Data = pNWA->SCPI->GetCALCulate(1)->SELected->DATA->FDATa;
//
//
//
//
//
//
Display measurement data.
Data is array of NOP * 2 (number of measurement points).
Where n is an integer between 0 and NOP - 1.
Data(n*2) : Primary value at the n-th measurement point.
Data(n*2+1) : Secondary value at the n-th measurement point. Always 0
when the data format is not the Smith chart or the polar.
CComSafeArray<double> mSafeArray;
if (mSafeArray.Attach(Data.parray) == S_OK)
{
for (unsigned int n = 0; n < mSafeArray.GetCount() / 2; ++n)
{
printf("%+.9E\t%+.9E\n",
351
mSafeArray.GetAt(n*2),
mSafeArray.GetAt(n*2+1));
}
mSafeArray.Detach();
}
printf("Press ENTER to exit.\n");
getc(stdin);
// Release COM object
pNWA.Release();
CoUninitialize();
return 0;
}
352
Example 6. Program Written in LabView
The following LabView program represents an example of the measurement
parameter setting, as well as acquisition and display of the measurement data array.
The program also shows a method of the sweep triggering and waiting for the
sweep completion.
Below see the block diagram of the program and the front panel of the program
with the program execution result.
The front panel contains the entry field for the instrument network name or IP
address. This field is used to control the instrument with DCOM technology from a
remote PC, on which the LabView package is executed. This example can also be
used on a single PC running the instrument program. In this case, enter the local PC
name or IP address 127.0.0.1 in this field.
The user must enter the instrument name in the entry field and click the “Run”
button. As the result of the program, the instrument information string will be
displayed and the measurement trace of transmission coefficient will be plotted.
353