Dissolved Oxygen Transmitter
MJK Oxix® Dissolved Oxygen Transmitter measures dissolved oxygen
in open tanks, in wells and in closed containers.
The measuring principle is based on fluorescence, and the sensor
does not consume oxygen like standard membrane-type sensors. It is
rugged and designed to handle tough applications.
The sensor is factory calibrated and does not require recalibration. The
sensor design reduces frequent cleaning to a minimum. For extreme
conditions a built-in air or water jet cleaning system can be used to
keep maintenance to the absolute minimum.
The transmitter has one 4-20 mA analog output, 2 digital outputs for
process control/alarms, and one digital input for resetting alarms or
activating cleaning cycles.
The Oxix® has user-friendly interface and a large graphic display for
measurement values. The mobile phone-like menu system is available
in many languages. Combined with soft key operation it is very easy to
start-up the instrument.
5.4
Pending
A large build-in data logger offers 160,000 date and time stamped
entries. Logging trend curves for improved overview can easily be
retrieved and shown in the display with a press of a single button.
Features
•
Mounting accessories (optional)
•
High performance optical sensor
•
No calibration
Application sites
•
No spare parts needed in sensor’s expected 10 year lifetime
•
Wastewater treatment plants
•
Air or water jet cleaning via built-in nozzle
•
Drinking water
•
Measurements not affected by sunlight or dry environment
•
Biologic control
•
Large, easy-to-read display with trend curve capability
•
Sludge treatment
•
Multiple languages and built-in datalogger
•
Filtering stations
•
No moving parts, oxygen consumption or risk of poisoning
•
Fishponds and farms
Interconnectivity
1 pc. Oxix Converter
with display ( 296305 )
3 pcs. Oxix without display (296308)
ll
RS485
ll
RS485
4 pcs. Oxix sensors (296510)
Max 3300 ft.
US 5.4 Oxix DO Transmitter 1202
ll
RS485
MJK’s modular design allows up to 1000 ft. between the sensor and the
converter, and the Display Unit can be mounted up to 3300 ft. from the
signal converter with ordinary twisted wires.
One Display Unit can control up to four Oxix® DO transmitters, MagFlux®
flow meters and/or SuSix® TTS transmitters for cost savings, space savings and improved overview of multiple measurement values.
Oxix® allows uploading of instrument software updates (firmware), and
the onboard data logger with 160,000 readings can be retrieved as a CSV
file on a PC. This connectivity is achieved with a common USB port and
the free MJK-Field Link software.
A number of communication modules for BlueTooth, Modbus, etc. are
available.
1
We reserve the right to continuously improve our products and make any
change in the stated specifications and dimensions without prior notice.
MJK North America, Inc.
37 Sherwood Terrace, #126
Lake Bluff, IL 60044
Tel: 847-482-8655
877 MJK LINK
Fax: 847-482-8654
[email protected]
www.mjk.com
Data Sheet
Dissolved Oxygen Transmitter
Specifications
Pending
Oxix® Sensor
Oxix® Transmitter
Dissolved oxygen/% saturated
Measurement
principle
1 voltage-free elec-mech. relay (max. 50 V DC / 1 A)
1 optically isolated relay (max. 50 VAC/VDC /120 mA)
One, max. 30 V DC, < 5 V DC = 0(low), > 10 V DC =
1(high), pulse length > 100 ms
0 - 25 mg/l or ppm
0 - 120 % SAT
Optical, near infra-red fluorescence
(lin = 475 nm, Iout = 609 nm)
Sensor Check
Automatic self-diagnostics
Pressure
Max. 100 psi (6 bar)
Materials
PVC, polyurethane, silicone, epoxy, stainless steel 316
Communication
Modbus® RTU, 9600 baud, 2-wire RS-485, slave
Cleaning system
Interface
RS-485 for connection to Display Unit or PLC
Power supply
10 - 30 V DC, or 24 V AC, 50 / 60 Hz ± 10 % or
115 V / 230 V AC, 50 / 60 Hz ± 10 %
Consumption
Max. 10 W
Cabinet material
Polycarbonate, glass reinforced
Enclosure rating
IP 67, NEMA 4x
Temp. range
-4 ... 140 ºF (- 20 … 60 °C)
Weight
2.4 lbs (1.1 kg)
CE approvals
EN 61000-6-4:2001, EN 61000-6-2:2001
Accuracy
+/- 0.1% of reading
Measuring input
RS-485
Analog output
1 active 4 - 20 mA, galvanically isolated (max. 800 W)
Digital outputs
Digital inputs
Display Unit
Built-in air or water jet connection and nozzle, air/water
supplied separately
Max. water supply pressure: 35 - 50 psi (2.3 - 3.4 Bar)
Max. air supply pressure: 40 - 60 psi (2,7 - 4.0 Bar)
Cable
4 x 0.326 mm2, shielded, dia Ø5.0 mm (22 AWG), PUR
Cable length
33 ft. (10 m); can be extended up to 1000 ft.
Response time
t90 less than 1 second
Flow velocity
No flow required
Power supply
12 V DC (10 -16 V DC), approx. 10 mA (from converter)
Output
RS-485, 9600 baud, 2-wire
Accuracy
Better than 0.1% of actual reading (0.02 mg/l)
Resolution
Better than 0.02 mg/l and 0.02 ppm
Enclosure rating
Dust and waterproof IP 67, NEMA 6 (on converter)
Temp. range
32 - 140 ºF (0 - 60 ºC)
Housing material
Polycarbonate, glass reinforced
Temp. comp.
32 - 122 ºF (0 - 50 ºC)
Protection lid
Transparent polycarbonate or optional white opaque
polycarbonate
Enclosure rating
IP 68 to IEC 529 (10 m), NEMA 6x
Dimensions
2 x 5.1 in. (50 x 130 mm) (diameter x length)
Mounting
1.5 inch NPT thread
0.25 inch OD jet spray tube compression fitting
Display
White, backlit LCD (64 x 128 pixels) with softkeys
Indication
Indication measurement, configuration and graph
Clock
Real-time clock with built-in battery backup
Weight
1.7 lbs (0.77 kg)
Communication
Modbus® RTU, 9600 baud, 2-wire RS-485, master
Approvals
EN 61000-6-4:2001, EN 61000-6-2:2001
Interface
RS-485
Memory
2 Mb Flash memory, 160.000 entries with date, time, value
Interface
USB 1,1 type mini B, Female
Temp. range
-4 ... 140 ºF (- 20 … 60 °C)
Accessories for Oxix® Transmitter
205546
Modbus® and RS-485 communication module
205547
PROFIBUS® PA communication module
205550
Bluetooth® communication module
Order Numbers
Oxix® Sensor
296510
Oxix® Converters
Oxix® Sensor
Accessories for Oxix® Sensors
155244
Oxix® Sensor Adapter for Immersion (see data sheet
US 5.6 Immersion Adapters)
155251
Ball Float for Oxix® sensor-NPT thread (see data sheet
US 5.5 Oxix Ball Float)
206304
Oxix® Converter with display, 10 - 30 VDC
296305
Oxix® Converter with display, 115 VAC
206306
Oxix® Converter with display, 24 VAC
206307
Oxix® Converter without display, 10 - 30 VDC
296308
Oxix® Converter without display, 115 VAC
206309
Oxix® Converter without display, 24 VAC
Mechanical Dimensions
7.3”
¼” tube
3.5”
¼” tube fitting
3.3”
Cable to Oxix
Converter
1½” NPT
7.2”
6.4”
6.1”
3.4”
3.1”
130 mm (5”)
Ø0.8
5.9”
5.7”
112 mm (4.2”)
57 mm (1.9”)
2
MJK offices:
Distributed by:
Measuring element
Denmark
Sweden
www.mjk.dk
[email protected]
+45 45 56 06 56
www.mjk.se
[email protected]
+46 53 31 77 50
Norway
www.mjk.no
[email protected]
+47 69 20 60 70
Ireland
Australia
www.mjk.com
[email protected]
+353 87 953 5625
www.mjk.com
[email protected]
+61 3 9755 1529
The Netherlands
North America
Singapore
www.mjk.com
[email protected]
+31 251 672171
www.mjk.com
[email protected]
+1 847 482 8655
www.mjk.sg
[email protected]
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertisement