Cross Technologies 2016-0510 Downconverter, 70 ± 18 MHz 500 to 1000 MHz in 1 MHz steps Owner's Manual
Advertisement
Advertisement
Instruction Manual
Model 2016-0510
Downconverter
July 2013, Rev. 0
F=1000 G=+25.0
MENU
EXECUTE
MODEL 2016
DOWNCONVERTER
C
ROSS
T
ECHNOLOGIES INC.
Data, drawings, and other material contained herein are proprietary to Cross Technologies, Inc., but may be reproduced or duplicated without the prior permission of Cross Technologies, Inc. for purposes of operating the equipment.
When ordering parts from Cross Technologies, Inc., be sure to include the equipment model number, equipment serial number, and a description of the part.
C
ROSS
T
ECHNOLOGIES, INC.
6170 Shiloh Road
Alpharetta, Georgia 30005
(770) 886-8005
FAX (770) 886-7964
Toll Free 888-900-5588
WEB www.crosstechnologies.com
E-MAIL [email protected]
INSTRUCTION MANUAL
MODEL 2016-0510 Downconverter
TABLE OF CONTENTS
Warranty
1.0 General
1.1 Equipment Description
1.2 Technical Characteristics
1.3 Monitor & Control Interface
2.0 Installation
2.1 Mechanical
2.2 Rear I/O’s
2.3 Front Panel Controls, Indicators
2.4 Operation
2.5 Menu Settings
3.0 Environmental Use Information
PAGE
8
9
10
11
12
2
3
3
4
5
8
17
WARRANTY - The following warranty applies to all Cross Technologies, Inc. products.
All Cross Technologies, Inc. products are warranted against defective materials and workmanship for a period of one year after shipment to customer. Cross Technologies,
Inc.’s obligation under this warranty is limited to repairing or, at Cross Technologies, Inc.’s option, replacing parts, subassemblies, or entire assemblies. Cross Technologies, Inc. shall not be liable for any special, indirect, or consequential damages. This warranty does not cover parts or equipment which have been subject to misuse, negligence, or accident by the customer during use. All shipping costs for warranty repairs will be prepaid by the customer. There are not other warranties, express or implied, except as stated herein.
2016-0510 Manual, Rev. 0
C
ROSS
T
ECHNOLOGIES, INC.
6170 Shiloh Road
Alpharetta, Georgia 30005
(770) 886-8005
FAX (770) 886-7964
Toll Free 888-900-5588
WEB www.crosstechnologies.com
E-MAIL [email protected]
Page 2 07/31/13
MODEL 2016-0510 Downconverter
1.0 General
1.1 Equipment Description
The 2016-0510 Downconverter converts 500 to 1000 MHz to 70 ± 18 MHz in 1 MHz steps (125 kHz to 1 kHz step options available). Synthesized local oscillators (LO) provide frequency selection. Push button switches select the input frequency, gain, and other parameters. Front panel LEDs provide indication of DC power
(green), PLL alarm (red), and Remote operation (yellow). Gain is adjustable manually (MGC) over a 0 to 50 dB range. Remote operation allows selection of frequency and gain. Parameter selection and frequency and gain settings appear on the LCD display. Connectors are BNC female for IF and the optional external reference input and output, and Type F female for the RF input. The external 10 MHz option E includes a 10 MHz output connector which contains either the internal or external 10 MHz reference signal. A high stability Option H
(±0.01 ppm) is also available. The unit is powered by a 100-240 ± 10% VAC power supply, and housed in a
1 3/4” X 19” X 16” rack mount chassis.
AC
F=1000 G=+25.0
MENU
EXECUTE
FRONT
MODEL 2016
DOWNCONVERTER
C
ROSS
T
ECHNOLOGIES INC.
GND RF IN LNB FUSE
F1
J2 DS8
VDC on RF
IF OUT 10 MHZ REF
OUTPUT
J1 J18
REAR
MONITOR
AND
CONTROL
5 4
9 8
3
7
2
6
1
J10
10 MHZ
EXT REF
INPUT
J3 J20
FIGURE 1.1 Model 2016-0510 Front and Rear Panels
0.50
to
1.00
GHz
IN
0.40 to 1.2 GHz
HP/LP var atten
2.25
to
2.75
GHz
1750 MHz
BP
1680 or
1820
MHz
120 MHz
LP
70
MHz
OUT
EXT
10MHz*
F=1000 G=+25
CONTROLLER
2016-0510 DOWNCONVERTER BLOCK DIAGRAM
INT
10MHz
* = OPTIONAL
FIGURE 1.2 Model 2016-0510 Downconverter Block Diagram
2016-0510 Manual, Rev. 0 Page 3 07/31/13
1.2 Technical Characteristics
TABLE 1.1
2016-0510 Downconverter Specifications*
Input Characteristics (IF)
Impedance / Return Loss
Frequency
Input Level
Output Characteristics (RF)
Impedance / Return Loss
Frequency
Output Level
Output 1 dB Compression
Channel Characteristics
Gain Range (adjustable)
Image Rejection
75 Ω / 10 dB
500-1000 MHz
-70 to -20 dBm
75 Ω / 18 dB
70 ±18 MHz
-20 to -10 dBm
-5 dBm at maximum gain
Frequency Response
Spurious Response
Group Delay, max.
Frequency Sense
Synthesizer Characteristics
Frequency Accuracy
Frequency Step
10 MHz Level (In or Out)
Phase Noise @ F (Hz) >
Standard 1MHz steps: dBC/Hz
X1001-1kHz steps: dBC/Hz
Controls, Indicators
±1.5 dB, 500-1000 MHz; ±0.5 dB, 36 MHz BW; ±1.0 dB, 40 MHz BW
< -50 dBC, in band
0.015 ns/MHz 2 parabolic; 0.05ns/MHz linear; 1 ns ripple
Inverting or Non-inverting, Selectable
± 1.0 ppm maximum over temperature (±0.01 ppm, Option H)
1.0 MHz (125 kHz to 1 kHz step options available)
3 dBm, ± 3 dB, 75 ohms (Option E)
10 100 1K
-55 -70 -70
10K
-80
100K
-90
-60 -75 -85 -90 -105
Frequency / Gain Selection
Power, Alarm, Remote, Mute
Remote
Other
RF, IF Connectors
Direct readout LCD; manual or remote selection
Green LED, Red LED, Yellow LED, Yellow LED
RS232C, 9600 baud (RS485, Ethernet Optional)
IF, 10 MHz Connectors
Alarm/Remote Connector
Size
Power
Type F (female), BNC (female)
BNC (female), 75 Ω , works with 50 or 75 ohms (Option E)
DB9 (female) - NO or NC contact closure on Alarm
19 inch, 1RU Standard Chassis 1.75” high X 16.0” deep
100-240 ±10% VAC, 47-63 Hz, 25 watts max. (24, 48 VDC Optional)
Available Options
0.0 to +50.0 dB
> 50 dB, minimum
1M
-110
-120
Connectors/Impedance
B - 75 Ω BNC (RF), 75 Ω BNC (IF)
E
H
X or X1
External 10 MHz reference input & output, RF Insertion
High Stability (± 0.01 ppm) internal reference
125 kHz or 100 kHz step size
X 1001 1 kHz step, includes Option H
Communication Interface/Standard RS232
Q RS485 Remote Interface
C - 50 Ω BNC (RF), 75 Ω BNC (IF)
D - 50 Ω BNC (RF), 50 Ω BNC (IF)
N - 50 Ω N-Type (RF), 75 Ω BNC (IF)
M - 50 Ω N-Type (RF), 50 Ω BNC (IF)
Contact Cross Technologies for other available options.
W8
W18
Ethernet, with Web Browser
Ethernet, with Web Browser and SNMP
W28 Ethernet, with TCP/IP, Telnet®
*10˚C to 40˚C; Specifications subject to change without notice.
2016-0510 Manual, Rev. 0 Page 4 07/31/13
1.3 Monitor and Control Interface
A) Remote serial interface
Protocol: RS-232C, 9600 baud rate, no parity, 8 data bits, 1 start bit, and 1 stop bit.
(RS-232C, RS-422, or RS-485 - option -Q)
5
6
7
8
9
1
2
3
4
M&C Cable Diagram - Cross Technologies Frequency Converters
Female DB-9
PC Com Port
Male DB-9
2015/16/17 M&C Port
RX
TX
DTR
SG
DSR
RTS
CTS
RX
TX
SG
7
8
5
6
9
3
4
1
2
Connector: Rear panel, DB-9 male
6
7
8
9
3
4
5
Pin
1
2
Function
Rx-
Rx+ (RS-232C)
Tx+ (RS-232C)
Tx-
GND
Alarm Relay: Common
Alarm Relay: Normally Open
Not Used
Alarm Relay: Normally Closed
2016-0510 Manual, Rev. 0 Page 5 07/31/13
B) Status Requests -
Table 1.2 lists the status requests for the 2016-0510 and briefly describes them.
* PLEASE NOTE: The two character {aa}(00-31) prefix, in the table below,
should be used ONLY when RS-485, (OPTION-Q), is selected.
Command
Command Status
10MHz Ref Status
(option -E only)
Syntax*
{aaS1}
{aaS2}
Description
Returns {S1bbbbcccIA} where:
• bbbb = Rx frequency
• cc = Rx gain
• I = 1 - spectrum invert enabled
• A = summary alarm; 1=alarmed, 0=normal
Returns {S2E} where:
• E = 1 - external 10 MHz switched in
2016-0510 Manual, Rev. 0 Page 6 07/31/13
C) Commands - Table 1.1 lists the commands for the 2016-0510 and briefly describes them. After a command is sent the 2016-0510 sends a return “>” indicating the command has been received and executed.
General Command Format - The general command format is {aaCND...}, where:
{ = start byte aa = address (RS-485 only - option -Q)
C = 1 character, either C (command) or S (status)
N = 1 character command or status request
D = 1 character or more of data (depends on command)
} = stop byte
*
PLEASE NOTE: The two character {aa}(00-31) prefix, in the table below,
should be used ONLY when RS-485, (OPTION-Q), is selected.
Command
Set Receiver Frequency
Set Receiver Gain
Enable Spectrum Invert
Enable External 10MHz IN
(option -E only)
Enable Remote
Disable Remote
Syntax* Description
{aaC2xxxx} where:
• xxxx = 4 characters
• Range: 500 to 1000 MHz, in 1 MHz steps
{aaC4xx} where:
• xxx = 3 characters
• Range: 000 to 500 (0 dB to +50 dB, in 1 dB steps)
{aaC7x}
{aaCEx} where x =:
• 0 to disable spectrum invert
• 1 to enable spectrum invert where x =:
• 0 to disable External 10MHz ref signal
• 1 to enable External 10MHz ref signal
#
{aaCR0}
Just # sign
{CR and zero}
2016-0510 Manual, Rev. 0 Page 7 07/31/13
2.0 Installation
2.1 Mechanical - The 2016-0510 consists of one RF/Controller PCB housed in a 1 RU (1 3/4 inch high) by 16 inch deep chassis. The unit is powered by a 100-240 ± 10% VAC power supply, and housed in a 1 3/4” X 19”
X 16” rack mount chassis. The 2016-0510 can be secured to a rack using the 4 holes on the front panel. Figure
2.1 shows how the 2016-0510 is assembled.
POWER
SUPPLY
FIGURE 2.1 2016-0510 Mechanical Assembly
2016-0510 Manual, Rev. 0 Page 8 07/31/13
2.2 Rear Panel Input/Output Signals - Figure 2.2 shows the input and output connectors on the rear panel.
AC1 - POWER IN
AC input for switching power supply. 100-240 ±10% VAC,
47-60 Hz.
J2 - RF IN
500-1000 MHz input
-70 to -20 dBm see Table 2.2.
AC
GND
F1 - LNB FUSE (option -L)
0.5A, Fast Blo, 1/4” Fuse; installing fuse places +24 VDC, 0.4 Amps, max on the RF IN (J2 center pin).
RF IN LNB FUSE
F1
J2 DS8
VDC on RF
IF OUT 10 MHZ REF
OUTPUT
J1 J18
J1 - IF OUT
70 MHz output
-20 to -10 dBm see Table 2.2.
MONITOR
AND
CONTROL
5 4
9 8
3
7
2
6
1
J10
J10 - MONITOR AND CONTROL
DB9 female connector see Table 2.1
10 MHZ
EXT REF
INPUT
J3 J20
DS8 - LNB ALARM LED (option -L)
Lights yellow when +24 VDC LNB voltage is present on RF IN, J2, center pin.
J18 - 10 MHz REF OUTPUT (option -E)
10 MHz external reference input
3 ± 3 dBm, 75 Ω , BNC female connector.
J20 - Ethernet Connection
RJ45 Ethernet Connector
J3 - 10 MHz EXT REF INPUT (Option -E)
10 MHz external reference input
3 ± 3 dBm, 75 Ω , BNC female connector.
FIGURE 2.2 2016-0510 Rear Panel I/Os
6
7
8
9
3
4
5
Pin
1
2
Function
Not Used
Rx+ (RS-232C)
Tx+ (RS-232C)
Not Used
GND
Alarm Relay: Common
Alarm Relay: Normally Closed
Not Used
Alarm Relay: Normally Open
*Remote Serial Interface
Interface: DB-9 Male
Protocol: RS-232C, 9600 baud rate, no parity, 8 data bits, 1 start bit, 1 stop bit
Option
STD
-B
-C
-D
-M
-N
IF
BNC, 75Ω
BNC, 75Ω
BNC, 75Ω
BNC, 50Ω
BNC, 50Ω
BNC, 75Ω
RF
F-Type, 75Ω
BNC, 75Ω
BNC, 50Ω
BNC, 50Ω
Type N, 50Ω
Type N, 50Ω
2016-0510 Manual, Rev. 0 Page 9 07/31/13
2.3 Front Panel Controls and Indicators - The following are the front panel controls and indicators.
DS3 - DOWN ALARM LED
Red LED indicates downconverter alarm.
DS1 - REMOTE LED
Yellow LED indicates remote operation.
LCD DISPLAY
Display shows Downconverter frequency in MHz and Gain in dB
S1 - MENU/EXECUTE BUTTON
Press this to get into Program mode and to execute any changes.
DS6 - POWER LED
Green LED indicates presence of DC power.
F=1000 G=+25.0
MENU
EXECUTE
S2 - VERT. TOGGLE
Vertical toggle switch that controls values in the Menu items when in program mode. Does not function in the normal display mode
S3 - HORIZ. TOGGLE
Horizontal toggle switch that controls which values are being adjusted. Does not function in the normal display mode
FIGURE 2.3 2016-0510 Front Panel Controls and Indicators
2016-0510 Manual, Rev. 0 Page 10 07/31/13
2.4 Operation
2.4.1 Installing and Operating the 2016-0510 Downconverter
1. Connect a -70 dBm to -20 dBm signal to RF IN, J2 (Figure 2.2)
2. Connect the IF OUT, J1, to the external equipment
3. Connect 100-240 ±10% VAC, 47 - 63 Hz, 25 watts max. to AC on the back panel.
4. Set the input frequency (See Section 2.5 Menu Settings).
5. Set the gain for 0.0 to +50.0 dB (See Section 2.5 Menu Settings).
6. Be sure DS6 (green, DC Power) is on and DS3 (red, Alarm) is off (Figure 2.3).
7. AC Fuse - The fuse is a 5 mm X 20 mm, 2 amp slow blow (Type T) and is inserted in the far slot in the drawer below the AC input as shown in Figure 2.4. There is a spare fuse in the near slot.
If a fuse continues to open, the power supply is most likely defective.
SPARE FUSE
FUSE DRAWER
AC Fuse - 2 amp slow blow (Type T),
5 mm X 20 mm
INPUT
~
100-240± 10%VAC
47-63 Hz
2A MAX
FUSE
TYPE T 2A GDC
250 VOLT
FOR 100 - 240 V~
FIGURE 2.4 Fuse Location and Spare Fuse
2016-0510 Manual, Rev. 0 Page 11 07/31/13
2.5 Menu Settings
2.5.1 Functions - This section describes operation of the front panel controls. There are three operator switches, the LCD display and alarm indicator LEDs. All functions for the equipment are controlled by these components. The functions are (see Figure 2.5):
Power Up
Normal Display
Menu 1 Frequency in MHz
Menu 2 Gain (0 to +50, 1 dB steps)
Menu 3 Set Unit to Remote Operation
Menu 4 Select External 10 MHz Reference (Option -E)
Menu 5 Set Remote mode (Option -Q)
Menu 6 Set RS-485 address (Option -Q)
Save Menu When go to end
Alarm indications appear on the LEDs (See figure 2.3).
All program changes must start with the operation of the Menu/Execute switch and must also end with the operation of the Menu/Execute switch verified by the “Save Settings?” Menu. If this sequence is not followed, none of the changes will take effect. If programming is initiated and no operator action takes place for approximately 12 seconds (before the final press of the Menu/Execute switch) the display will revert to its previous status and you will need to start over.
2016-0510 Manual, Rev. 0 Page 12 07/31/13
2.5.2 Power On Settings
NOTE: THE LAST STATUS OF A UNIT IS RETAINED EVEN WHEN POWER IS REMOVED.
WHEN POWER IS RESTORED, THE UNIT WILL RETURN TO IT'S PREVIOUS SETTINGS.
When power is first applied, the LCD display goes through three steps.
1. The LCD goes black to show all segments are functioning.
2. The software version will be displayed.
R E V 1 . 0 0
3. The present frequency and gain of the downconverter is shown.
F=1000 G=+25.0
The unit is now operational and ready for any changes the operator may desire.
2.5.3 Control Switches
1. Menu/Execute - Any change to the programming of the unit must be initiated by pressing the
Menu/Execute switch and completed by pressing the Menu/Execute switch.
2. Horizontal Switch - This switch is mounted so its movement is horizontal and moves the cursor left or right.
3. Vertical Switch - This switch is mounted so its movement is vertical and has two functions: a. During frequency , gain changes, the vertical movement will raise or lower the number in the direction of the arrows.
b. For other functions such Mute on/off, the vertical switch will alternately turn the function on or off regardless of the direction operated.
2016-0510 Manual, Rev. 0 Page 13 07/31/13
2.5.4 Frequency Changes
At any time during the modification process, if you have made a mistake and do not wish to save the changes you have made, do not press the Menu/Execute switch; simply do nothing for approximately 12 seconds, and the system will return to the normal operating mode or scroll to “R” and push the menu/Execute switch and select “NO” in the “SAVE SETTINGS?” window.
To change the FREQUENCY:
Operate the Menu/Execute switch until you get to the menu item you want to change, see Figure 2.5 for the sequence of menu options. The following display is for changing the downconverter frequency:
F = 1000 R
Pressing the Up/Down switch down will toggle the display to:
F = 975 R
By using the horizontal rocker switch the cursor can be moved left or right .
F = 975 R
NOTE: CHANGES DO NOT TAKE PLACE ON FREQUENCY UNTIL YOU GO TO THE SAVE
MENU AND INDICATE YOU WANT TO SAVE THE CHANGES. THE CARRIER IS MUTED WHEN
FREQUENCY IS CHANGED.
When the display indicates the value desired you can push the Menu/Execute switch to get to the next item:
G = +25.0
R
OR you can scroll to “R”, push the Menu/Execute switch to get to:
SAVE SETTINGS? Y N
Selecting Y will save the new settings. Selecting N will revert to the previous settings.
Pushing the Menu/Execute switch then takes you to this:
F=1000 G=+25.0
Figure 2.5 gives the menu items and how to make changes.
2016-0510 Manual, Rev. 0 Page 14 07/31/13
2.5.5 Gain Changes
When you get to this menu note that the gain changes will be made as you make them but if you do not wish to save the changes you have made, scroll to “R” and push the Menu/Execute switch and select “NO” in the
“SAVE SETTINGS?” window or do not press the Menu/Execute switch; simply do nothing for approximately 12 seconds, and the system will return to the normal operating mode.
To change the DOWNCONVERTER GAIN:
Push the Menu/Execute switch to get to the gain setting (See Figure 2.5 for the sequence of menu options):
G = +25.0
R
Pressing the Up/Down switch will change the gain in 1 or 10 dB steps depending on the cursor location:
G = +35.0
R
By using the horizontal rocker switch the cursor can be moved left or right. Pressing the Up/Down switch will toggle the display digit selected until you have the desired gain.
G = +35.0
R
NOTE: THE GAIN WILL BE CHANGED AS YOU ADJUST THE NUMBERS. HOWEVER, THE
VALUE WILL NOT BE STORED UNTIL YOU INDICATE YES IN THE SAVE SETTINGS
WINDOW.
SAVE SETTINGS? Y N
Selecting Y will save the new settings. Selecting N will revert to the previous settings.
Pushing the Menu/Execute switch then takes you to this:
F=1000 G=+25.0
Figure 2.5 gives the menu items and how to make changes.
2016-0510 Manual, Rev. 0 Page 15 07/31/13
2.5.5 Alarm Indications
An alarm condition for will occur if any local oscillator phase lock loop (PLL) comes out of lock. The Mute
LED will light if you select Mute and the Remote LED will light when you select the Remote mode.
ON POWER UP
REV 1.00
Power Up
NORMAL DISPLAY
F = 1000 G = +25.0
Normal Display
Menu 1 Frequency
Menu 2 Gain (0.0 to +50.0)
Menu 3 Set Unit to Remote
Operation
Menu 4 Select External 10 MHz
Reference (Option -E)
Menu 6 Set Remote Mode
(Option -Q)
Menu 7 Set RS-485 Address
(Option -Q)
Save? When “R” is selected or when you get to the end
PUSHING MENU/EXECUTE SEQUENCE
F = 1000
G = +25.0
REMOTE OFF
EXT REF OFF
RS 485 OFF
ADDRESS = 00
SAVE SETTINGS? Y N
R
R
R
R
R
R
SCROLL <>
SCROLL
SCROLL <>
SCROLL
SCROLL <>
SCROLL
SCROLL <>
SCROLL
SCROLL <>
SCROLL
SCROLL <>
SCROLL
SCROLL <>
PUSH BUTTON
PUSH BUTTON
PUSH BUTTON
PUSH BUTTON
PUSH BUTTON
PUSH BUTTON
PUSH BUTTON
PUSH BUTTON
FIGURE 2.5 Menu Display and Sequence
2016-0510 Manual, Rev. 0 Page 16 07/31/13
3.0 Environmental Use Information
A. Rack-Mounting - To mount this equipment in a rack, please refer to the installation instructions located in the user manual furnished by the manufacturer of your equipment rack.
B. Mechanical loading - Mounting of equipment in a rack should be such that a hazardous condition does not exist due to uneven weight distribution.
C. Elevated operating ambient temperature - If installed in a closed or multi-unit rack assembly, the operating ambient temperature of the rack may be greater than room ambient temperature.
Therefore, consideration should be given to Tmra.
D. Reduced air flow - Installation of the equipment in a rack should be such that the amount of air flow required for safe operation of the equipment is not compromised. Additional space between units may be required.
E. Circuit Overloading - Consideration should be given to the connection of the equipment to the supply circuit and the effect that overloading of circuits could have on over current protection and supply wiring.
Appropriate consideration of equipment name plate rating should be used, when addressing this concern.
F. Reliable Earthing - Reliable earthing of rack-mounted equipment should be maintained. Particular attention should be given to supply connections other than direct connection to the Branch
(use of power strips).
G. Top Cover - There are no serviceable parts inside the product so, the Top Cover should not be removed.
If the Top Cover is removed the ground strap and associated screw MUST BE REINSTALLED prior to Top Cover screw replacement. FAILURE TO DO this may cause INGRESS and/or EGRESS emission problems.
2016-0510 Manual, Rev. 0 Page 17 07/31/13
C
ROSS
T
ECHNOLOGIES, INC.
6170 Shiloh Road
Alpharetta, Georgia 30005
(770) 886-8005
FAX (770) 886-7964
Toll Free 888-900-5588
WEB www.crosstechnologies.com
E-MAIL [email protected]
Printed in USA
2016-0510 Manual, Rev. 0 Page 18 07/31/13