AND UA-767 S Instruction manual
The AND UA-767 S is a digital blood pressure monitor that offers accurate and easy blood pressure measurements at home. It features a large, easy-to-read display, and can store up to 60 measurements in its memory. The device also has an irregular heartbeat detector and a cuff fit error indicator to ensure accurate readings. Additionally, it comes with a pressure bar indicator and a WHO classification indicator to help you understand your blood pressure readings in context.
Advertisement
Advertisement
Digital Blood Pressure Monitor
Model UA-767S
Instruction Manual
Original
Manuel d’instructions
Traduction
Manual de Instrucciones
Traducción
Manuale di Istruzioni
Traduzione
使用手冊
翻譯
1WMPD4002781G
Contents
Dear Customers.................................................................... 2
Preliminary Remarks ............................................................. 2
Precautions........................................................................... 2
Parts Identification................................................................ 4
Symbols................................................................................ 5
Operation Mode.................................................................... 7
Using the Monitor................................................................. 8
Installing / Changing the Batteries ................................... 8
Connecting the Air Hose .................................................. 9
Connecting the AC Adapter.............................................. 9
Selecting the Correct Cuff Size ....................................... 10
Applying the Arm Cuff ................................................... 10
How to Take Accurate Measurements............................. 12
Measurement................................................................. 12
After Measurement ........................................................ 12
Measurements .................................................................... 13
Normal Measurement..................................................... 13
Measurement with the Desired Systolic Pressure ............ 14
Notes for Accurate Measurement ................................... 14
Recalling the Memory Data ................................................. 15
What is an Irregular Heartbeat............................................. 16
% IHB .................................................................................. 16
Pressure Bar Indicator ......................................................... 17
WHO Classification Indicator ............................................... 17
About Blood Pressure.......................................................... 17
What is Blood Pressure? ................................................. 17
What is Hypertension and How is it Controlled?.............. 18
Why Measure Blood Pressure at Home?........................... 18
WHO Blood Pressure Classification ................................. 18
Blood Pressure Variations............................................... 18
Troubleshooting ................................................................. 19
Maintenance ....................................................................... 20
Technical Data.................................................................... 20
English 1
Dear Customers
Congratulations on purchasing a state-of-the-art A&D blood pressure monitor, one of the most advanced monitors available today. Designed for ease of use and accuracy, this device will facilitate your daily blood pressure regimen.
We recommend that you read through this manual carefully before using the device for the first time.
Preliminary Remarks
This device conforms to the European Directive 93/42 EEC for Medical
Products. This is made evident by the mark of conformity. (0123:
The reference number to the involved notified body)
The device is designed for use on adults only, not newborns or infants.
Environment for use. The device is for use in the home healthcare environment.
This device is designed to measure blood pressure and pulse rate of people for diagnosis.
Precautions
Precision components are used in the construction of this device.
Extremes in temperature, humidity, direct sunlight, shock or dust should be avoided.
Clean the device and cuff with a dry, soft cloth or a cloth dampened with water and a neutral detergent. Never use alcohol, benzene, thinner or other harsh chemicals to clean the device or cuff.
Avoid tightly folding the cuff or storing the hose tightly twisted for long periods, as such treatment may shorten the life of the components.
Take care to avoid accidental strangulation of babies or infants with the hose.
Do not twist the air hose during measurement. This may cause injury due to continuous cuff pressure.
The device and cuff are not water resistant. Prevent rain, sweat and water from soiling the device and cuff.
Measurements may be distorted if the device is used close to televisions, microwave ovens, X-ray or other devices with strong electrical fields.
Wireless communication devices, such as home networking devices, mobile phones, cordless phones and their base stations, walkie-talkies can affect this blood pressure monitor. Therefore, a minimum distance of 3.3 meters (11 feet) should be kept from such devices.
When reusing the device, confirm that the device is clean.
Used equipment, parts and batteries are not treated as ordinary household waste, and must be disposed of according to the applicable local regulations.
When the AC adapter is used, make sure that the AC adapter can be readily removed from the electrical outlet when necessary.
Do not modify the device. It may cause accidents or damage to the device.
To measure blood pressure, the arm must be squeezed by the cuff hard enough to temporarily stop blood flow through the artery. This may cause pain, numbness or a temporary red mark to the arm. This condition will appear especially when measurement is repeated successively. Any pain, numbness, or red marks will disappear with time.
Measuring blood pressure too frequently may cause harm due to blood flow interference. Check that the operation of the device does not result in prolonged impairment of blood circulation, when using the device repeatedly.
If you have had a mastectomy, please consult a doctor before using the device.
Do not let children use the device by themselves and do not use the device in a place within the reach of infants.
There are small parts that may cause a choking hazard if swallowed by mistake by infants.
Contraindications
The following are precautions for proper use of the device.
Do not apply the cuff on an arm with another medical electrical equipment attached. The equipment may not function properly.
People who have a severe circulatory deficit in the arm must consult a doctor before using the device, to avoid medical problems.
Do not self-diagnose the measurement results and start treatment by yourself. Always consult your doctor for evaluation of the results and treatment.
Do not apply the cuff on an arm with an unhealed wound.
Do not apply the cuff on an arm receiving an intravenous drip or blood transfusion. It may cause injury or accidents.
Do not use the device where flammable gases such as anesthetic gases are present. It may cause an explosion.
Do not use the device in highly concentrated oxygen environments, such as a high-pressure oxygen chamber or an oxygen tent. It may cause a fire or explosion.
Parts Identification
Air Connector Plug
Air Socket
Proper Fit Range
Index Mark
Display
Battery Compartment
Display
MEMORY
I.H.B. Symbol
(Irregular heartbeat symbol)
%IHB
Cuff Fit Error Symbol
Movement Error Symbol
Heart Mark
Average
AC adapter Plug
DC Jack
START Button
Air Hose
Artery Position Mark
Arm Cuff
Battery Cover
1.5V Batteries
(R6P, LR6 or AA)
Systolic
Pressure
Diastolic
Pressure
WHO Classification Indicator and Pressure Bar Indicator
Battery Indicator
Pulse Rate
Symbols
Symbols that are printed on the device case
Standby and Turn the device on.
SYS Systolic blood pressure in mmHg
DIA Diastolic blood pressure in mmHg
PUL Pulse per minute
IP
Battery installation guide
Direct current
Type BF: Device, cuff and tubing are designed to provide special protection against electrical shocks.
EC directive medical device label
EU-representative
Manufacturer
Date of manufacture
International protection symbol
SN
WEEE label
Serial number
Refer to instruction manual/booklet
Polarity of DC jack
Symbols that appear on the display
Symbols Function / Meaning
Appears while measurement is in progress. It blinks when the pulse is detected.
Irregular Heartbeat symbol (I.H.B.)
Appears when an irregular heartbeat is detected.
It may light when a very slight vibration like shivering or shaking is detected.
Recommended Action
Measurement is in progress. Remain as still as possible.
Appears when a body or arm movement is detected.
The reading may yield an incorrect value.
Try the measurement again. Remain still during measurement.
Symbols
Symbols Function / Meaning
Appears during measurement when the cuff is attached loosely
The cuff is not applied correctly.
Recommended Action
The reading may yield an incorrect value. Apply the cuff correctly and try the measurement again.
Detected rate of IHB in memory
%IHB =
Number of detected
IHBs in memory
Total number x 100 [%]
Previous measurements stored in
MEMORY.
Average data
FULL BATTERY
The battery power indicator during measurement.
LOW BATTERY
The battery is low when it blinks.
Unstable blood pressure due to movement during measurement.
Replace all batteries with new ones when the mark blinks.
Try the measurement again.
Remain very still during measurement.
The systolic and diastolic values are within 10 mmHg of each other.
The pressure value did not increase during inflation.
Apply the cuff correctly, and try the measurement again.
PUL DISPLAY ERROR
The pulse is not detected correctly.
Blood pressure monitor internal error
Remove the batteries and press the START button, and then install the batteries again. If the error still appears, contact the dealer.
Operation Mode
1. Normal Measurement
Press START button. Blood pressure is measured and the data is stored in memory. This device can store the last 60 measurements in memory.
2. Recalling the Data
When nothing is displayed, press and hold the START button.
In standby, press and hold the
START button
Release the button when displaying the average data.
The data number and stored data are automatically displayed in order from the last measurement.
For details on recalling the data, Refer to
“Recalling the Memory Data”.
3. Deleting all Data Stored in Memory
When turning off the device, press and hold the
START
Select " button until the " " is displayed.
" to clear the data.
START
The data is cleared when the mark blinks.
The device turns off automatically.
4. Measurement with the Desired Systolic Pressure
Refer to page 14 for measurement with the desired systolic pressure.
Using the Monitor
Installing / Changing the Batteries
1. Remove the battery cover.
2. Remove the used batteries and insert new batteries into the battery compartment as shown, taking care that the polarities (+ and -) are correct.
Step 3
Step 1
Step 2
Step 1
Use only R6P, LR6 or AA batteries.
3. Attach the battery cover.
CAUTION
Insert the batteries as shown in the battery compartment. If installed incorrectly, the device will not work.
When (LOW BATTERY mark) blinks in the display, replace all batteries with new ones. Do not mix old and new batteries. It may shorten the battery life, or cause the device to malfunction.
Replace the batteries two seconds or more after the device turns off.
If (LOW BATTERY mark) appears even after the batteries are replaced, make a blood pressure measurement. The device may then recognize the new batteries.
(LOW BATTERY mark) does not appear when the batteries are drained.
The battery life varies with the ambient temperature and may be shorter at low temperatures. Generally, four new LR6 batteries will last approximately for one year when used twice for measurement each day.
Use the specified batteries only. The batteries provided with the device are for testing monitor performance and may have a limited life.
Remove the batteries if the device is not to be used for a long time.
The batteries may leak and cause a malfunction.
Using the Monitor
Connecting the Air Hose
Insert the air connector plug into the air socket firmly.
Air socket
Air connector plug
Connecting the AC Adapter
Insert the AC adapter plug into the
DC jack.
Next, connect the AC adapter to an electrical outlet.
Use the specified AC adapter.
(Refer to page 21.)
DC jack
AC adapter plug
When disconnecting the AC adapter from the electrical outlet, grasp and pull the AC adapter body out of the outlet.
When disconnecting the AC adapter plug from the blood pressure monitor, grasp and pull the AC adapter plug out of the monitor.
Using the Monitor
Selecting the Correct Cuff Size
Using the correct cuff size is important for an accurate reading. If the cuff is not the proper size, the reading may yield an incorrect blood pressure value.
The arm size is printed on each cuff.
The index and proper fit range, on the cuff, tell you if you are applying the correct cuff. (Refer to "Symbols that are printed on the cuff" on the next page.)
If the index points outside of the range, contact your local dealer to purchase a replacement cuff.
The arm cuff is a consumable. If it becomes worn, purchase a new one.
Arm Size
31 cm to 45 cm
22 cm to 42 cm
22 cm to 32 cm
Recommended Cuff Size
Large Adult Cuff
Wide Range Cuff
Adult Cuff
Catalog Number
CUF-F-LA
CUF-I
CUF-F-A
Arm size: The circumference at the biceps.
Note: The UA-767S is not designed for using a small cuff.
Applying the Arm Cuff
1. Wrap the cuff around the upper arm, about 1-2 cm above the inside of the elbow, as shown.
Place the cuff directly against the skin, as clothing may cause a faint
Do not roll up shirtsleeve tightly.
Cuff
1-2 cm pulse, and result in a measurement error.
2. Constriction of the upper arm, caused by tightly rolling up a shirtsleeve, may prevent accurate readings.
3. Confirm that the index points within the proper fit range.
Air hose
Artery position mark
Secure the fabric fastener firmly.
Index and proper fit range
Using the Monitor
Symbols that are printed on the cuff
Symbols Function / Meaning
Artery position mark
Recommended Action
Set the mark on the artery of the upper arm or in line with the ring finger on the inside of the arm.
▲ Index
REF Catalog number
Proper fit range for the adult cuff.
It's printed on the adult cuff.
Over range printed on the adult cuff and wide range cuff.
Use the large adult cuff instead of the adult cuff or wide range cuff.
Proper fit range for the wide range cuff.
It's printed on the wide range cuff.
Proper fit range for the large adult cuff.
It's printed on the large adult cuff.
Under range printed on the adult cuff and wide range cuff.
Under range on the large adult cuff.
Use the adult cuff instead of the large adult cuff.
Lot number
Proper fit range Large adult cuff
A
Wide range cuff
S
Adult cuff
S
L
W
A
L
L
Using the Monitor
How to Take Accurate Measurements
For the most accurate blood pressure measurement:
Sit comfortably on a chair. Rest your arm on the table. Do not cross your legs. Keep your feet flat on the floor and straighten your back.
Relax for about five to ten minutes before measurement.
Place the center of the cuff at the same level as your heart.
Remain still and keep quiet during measurement.
Do not measure right after physical exercise or a bath. Rest for twenty or thirty minutes before taking the measurement.
Try to measure your blood pressure at the same time every day.
Measurement
During measurement, it is normal for the cuff to feel very tight. (Do not be alarmed.)
After Measurement
After measurement, press the START button to turn the device off.
Remove the cuff and record your data.
Note: The device has an automatic power shut-off function, which turns the device off approximately one minute after measurement.
Allow at least three minutes between measurements on the same person.
Measurements
Before measurement, refer to “Notes for Accurate Measurement” on the next page.
Normal Measurement
1. Place the cuff on the arm (preferably the left arm).
Sit quietly during measurement.
2. Press START button.
At heart level
All of the display segments are displayed.
Zero (0) is displayed blinking briefly.
The display changes, as indicated in the figure at the right, as the measurement
Press begins. The cuff starts to inflate. It is normal for the cuff to feel very tight. A pressure bar indicator is displayed, on the left edge of the display, during inflation.
Note: If you wish to stop inflation at any time, press the START button again.
3. When inflation is complete, deflation starts automatically and (heart mark) blinks, indicating that the measurement is in
All of the display segments displayed
Zero display
Starts inflation
Pressurizing progress. Once the pulse is detected, the mark blinks with each pulse beat.
Note: device starts to inflate again automatically.
To avoid re-inflation, refer to “Measurement with the Desired Systolic Pressure” on the next page.
4. When the measurement is complete, the
START systolic and diastolic pressure readings and pulse rate are displayed.
The cuff exhausts the remaining air and deflates completely.
5. Press START button to turn the device off.
Note: Allow at least three minutes between measurements on the same person.
Measurement in progress
Systolic pressure
Diastolic pressure
WHO classification
Exhausts remaining air automatically
Measurements
Measurement with the Desired Systolic Pressure
The UA-767S is designed to detect the pulse and to inflate the cuff to a systolic pressure level automatically.
Use this method when re-inflation occurs repeatedly or when the results are not displayed even if the pressure decreases to 20 mmHg or less.
1. Place the cuff on the arm at heart level
(preferably the left arm).
At heart level
2. Press START button.
3. When the zero blinks, press and hold
Press the button
START the START button until a number about 30 to 40 mmHg higher than your expected systolic pressure
Release the button at the desired systolic pressure appears.
4. When the desired number is reached, release the START button to start
Press and hold the button
START measurement. Continue to measure your blood pressure as described on the previous page.
Notes for Accurate Measurement
Refer to the previous page for measurement
Sit down in a comfortable position. Place your arm on a table with your palm facing upward and the cuff at the same level as your heart.
Relax for about five to ten minutes before taking a measurement. If you are excited or depressed by emotional stress, the measurement will reflect this stress as a higher (or lower) than normal blood pressure reading and the pulse reading will usually be faster than normal.
An individual's blood pressure varies constantly, depending on what you are doing and what you have eaten. What you drink can have a very strong and rapid effect on your blood pressure.
This device bases its measurements on the heartbeat. If you have a very weak or irregular heartbeat, the device may have difficulty determining your blood pressure.
Should the device detect a condition that is abnormal, it will stop the measurement and display an error symbol. Refer to page 6 for the description of symbols.
This blood pressure monitor is intended for use by adults only. Consult with your physician before using this device on a child. A child should not use this device unattended.
Recalling the Memory Data
Note: This device stores the last 60 measurements in memory.
1. When nothing is displayed, press and hold the _ START _ button to recall the stored data.
Press and hold
START button
Average systolic
Average diastolic
2. Release the button when displaying the average data.
Average of all the data
Average pulse
3. The data number and stored data are automatically displayed in order from the last measurement.
4. The display will turn off automatically after all data are displayed.
Note: If you press the _ START _ button while recalling data, the device turns off.
Systolic
Diastolic
Pulse
Most recent data
Systolic
Diastolic
Pulse
Last data (Oldest)
What is an Irregular Heartbeat
The UA-767S blood pressure monitor provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeats during the blood pressure measurement. It is important that you are relaxed, remain still and do not talk during measurements.
Note: We recommend contacting your physician if you see this symbol frequently.
%
IHB
%IHB is displayed as frequency of IHB detected.
IHB can detect not only noises such as physical movement but also an irregular heartbeat. Therefore, we recommend contacting your physician if %IHB level is high.
%IHB =
Number of detected
IHBs in memory x 100 [%]
Total number
Display of %IHB: %IHB is displayed when displaying average values.
(Refer to “2. Recalling the Data” in “Operation Mode”.)
%IHB is not displayed when the memory number is six or less.
Average value display
Level 0
%IHB=0-24
Level 1
%IHB=25-49
Level 2
%IHB=50-74
Level 3
%IHB=75-100
Not displayed
Pressure Bar Indicator
The indicator monitors the progress of pressure during measurement.
Pressurizing
Inflation in progress
Releasing air
Measurement in progress
Inflation complete
WHO Classification Indicator
Each segment of the bar indicator corresponds to the WHO blood pressure
Example:
classification described on the next page.
WHO Classification Indicator
Severe hypertension
Moderate hypertension
Moderate hypertension
Mild hypertension
High normal
Normal
Optimal
Mild hypertension
: The indicator displays a segment, based on the current data, corresponding to the
WHO classification.
High normal
About Blood Pressure
What is Blood Pressure?
Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg). One's natural blood pressure is represented by the fundamental pressure, which is measured first thing in the morning while one is still at rest and before eating.
What is Hypertension and How is it Controlled?
Hypertension, an abnormally high arterial blood pressure, if left unattended, can cause many health problems including stroke and heart attack. Hypertension can be controlled by altering lifestyle, avoiding stress, and with medication under a doctor’s supervision.
To prevent hypertension or keep it under control:
Do not smoke Exercise regularly
Reduce salt and fat intake
Maintain proper weight
Have regular physical checkups
Why Measure Blood Pressure at Home?
Blood pressure measured at a clinic or doctor's office may cause apprehension and can produce an elevated reading, 25 to 30 mmHg higher than that measured at home. Home measurement reduces the effects of outside influences on blood pressure readings, supplements the doctor's readings and provides a more accurate, complete blood pressure history.
WHO Blood Pressure Classification
Standards to assess high blood pressure, without regard to age, have been established by the World
Health Organization (WHO), as shown in the chart.
Blood Pressure Variations
pressure varies greatly on a daily and seasonal basis. It may vary by 30 to 50 mmHg due to various conditions during the day. In hypertensive individuals, variations are even more pronounced. Normally, the blood pressure rises while at work or play and falls to its lowest levels during sleep. So, do not be overly concerned by the results of one measurement.
Take measurements at the same time every day using the procedure described in this manual to get to know your normal blood pressure. Regular readings give a more comprehensive blood pressure history. Be sure to note the date and time when recording your blood pressure.
Consult your doctor to interpret your blood pressure data.
Troubleshooting
Nothing appears in the display, even when the device is turned on.
Batteries are drained.
Battery terminals are not in the correct position.
Replace all batteries with new ones.
Reinstall the batteries with negative and positive terminals matching those indicated on the battery compartment.
The cuff does not inflate.
Battery voltage is too low.
(LOW BATTERY mark) blinks. If the batteries are drained completely, the mark does not appear.
The cuff is not applied properly.
You moved your arm or body during measurement.
Replace all batteries with new ones.
Apply the cuff correctly.
The device does not measure.
Readings are too high or too low.
The cuff position is not correct.
Make sure you remain very still and quiet during measurement.
Sit comfortably and still.
Place your arm on a table with your palm facing upward and the cuff at the same level as your heart.
If you have a very weak or irregular heartbeat, the device may have difficulty in determining your blood pressure.
The value is different from that measured at a clinic or doctor’s office.
Refer to “Why Measure Blood
Pressure at Home?”.
Other
Remove the batteries. Place them back properly and try the measurement again.
Note: If the actions described above do not solve the problem, contact the dealer. Do not attempt to open or repair this product, as any attempt to do so will make your warranty invalid.
Maintenance
Do not open the device. It uses delicate electrical components and an intricate air unit that could be damaged. If you cannot fix the problem using the troubleshooting instructions, contact the authorized dealer in your area or our customer service department. The A&D customer service will provide technical information, spare parts and units to authorized dealers.
The device was designed and manufactured for a long service life.
However it is generally recommended to have the device inspected every
2 years, to ensure proper functioning and accuracy. Please contact the authorized dealer in your area or A&D for maintenance.
Technical Data
Type UA-767S
Measurement method
Measurement range
Oscillometric measurement
Pressure: 0 - 299 mmHg
Diastolic pressure: 40 - 200 mmHg
Pulse: 40 - 180 beats / minute
Measurement accuracy Pressure: ±3 mmHg
Power supply 4 x 1.5V batteries (R6P, LR6 or AA) or
AC adapter (TB-233) (Not included)
Number of measurements
Approx. 700 times LR6 (alkaline batteries)
Approx. 200 times R6P (manganese batteries)
With pressure value of 180 mmHg at room temperature of 23 °C.
Clinical test
EMC batteries) / Class II (Supplied by adapter)
Continuous operation mode
According to ANSI / AAMI SP-10 1992
IEC 60601-1-2: 2007
Operating conditions +10 to +40 °C / 15 to 85 %RH / 800 to 1060 hPa
Transport / Storage conditions -20°C to +60°C / 10%RH to 95%RH
Dimensions Approx. 140 [W] x 60 [H] x 105 [D] mm
Weight
Ingress protection
Approx. 245 g, excluding the batteries
Device: IP21
Applied part Cuff Type BF
Useful life Device: 5 years (when used six times a day)
Cuff: 2 years (when used six times a day)
Accessory AC adapter
The adapter is to connect the device to a power source at home. Please contact your local A&D dealer for purchasing. The AC adapter is required to be inspected or replaced periodically.
TB-233
Output: 6 V 500 mA
139°C 2 A
Output: 6 V 500 mA
139°C 2 A
Accessories sold separately
Cuff Catalog Number
CUF-F-LA
Cuff Size
Large adult cuff
CUF-I
CUF-F-A
Wide range cuff
Adult cuff
Arm Size
31 cm to 45 cm
22 cm to 42 cm
22 cm to 32 cm
Arm size: The circumference at the biceps.
AC adapter Catalog Number Plug (Outlet type)
Note: Specifications are subject to change without prior notice.
1-243 Asahi , Kitamoto-shi, Saitama 364-8585, JAPAN
Telephone: [81] (48) 593-1111 Fax: [81] (48) 593-1119
A&D INSTRUMENTS LIMITED
Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire OX14 1DY United Kingdom
Telephone: [44] (1235) 550420 Fax: [44] (1235) 550485
A&D ENGINEERING, INC.
1756 Automation Parkway, San Jose, California 95131, U.S.A.
Telephone: [1] (408) 263-5333 Fax: [1] (408)263-0119
A&D AUSTRALASIA PTY LTD
32 Dew Street, Thebarton, South Australia 5031, AUSTRALIA
Telephone: [61] (8) 8301-8100 Fax: [61] (8) 8352-7409
Advertisement