9520 Series Pulse Generator

Operating Manual

Version 5.6

Quantum Composers, Inc.

212 Discovery Drive

Bozeman, MT 59718

Phone: (406)582-0227

Fax: (406)582-0237 www.quantumcomposers.com

Your distributor:

1-1

1 Table of Contents

1 Table of Contents ....................................................................................... 1-2

2 Introduction ................................................................................................. 2-6

Technical Support ........................................................................................................ 2-6

Warranty ...................................................................................................................... 2-6

Package Contents ......................................................................................................... 2-6

3 Safety Issues .............................................................................................. 3-7

4 Front Panel Overview ................................................................................. 4-8

9520 Front Panel .......................................................................................................... 4-8

Display Layout and Indicators ................................................................................. 4-8

Description of Front Panel Area .............................................................................. 4-8

5 Pulse Concepts and Pulse Generator Operations .................................... 5-10

Counter Architecture Overview ................................................................................. 5-10

System Timer Functions ............................................................................................ 5-10

Channel Timer Functions ........................................................................................... 5-11

Digital Output Multiplexer ........................................................................................ 5-11

Dependent & Independent Timing Events ................................................................. 5-11

Navigating the 9520 Front Panel ............................................................................... 5-12

Selecting Menus ..................................................................................................... 5-12

Selecting Menu Items ............................................................................................ 5-12

Numeric Input Mode .............................................................................................. 5-12

Entering Non-Numeric Parameters ........................................................................ 5-12

Alphanumeric Input Mode ..................................................................................... 5-13

Enabling System Output ........................................................................................ 5-13

Enable/Disable Channel Output ............................................................................. 5-13

Rearming the Channel Timers ............................................................................... 5-13

Setting Pulse Timing Parameters ........................................................................... 5-14

Setting Pulse Output Parameters ............................................................................ 5-14

Using the Output Multiplexer ................................................................................ 5-14

Setting System Internal Rate Parameters ............................................................... 5-14

6 9520 Menu Structure ................................................................................ 6-16

System Mode Menus (FUNC + MODE key) ............................................................ 6-16

Setting System Mode of Operation ........................................................................ 6-16

Channel Menus (A, B, C, D, E, F, G, or H key) ........................................................ 6-16

Timing Menu ......................................................................................................... 6-16

Output Configuration Menu ................................................................................... 6-16

Mode Menu ............................................................................................................ 6-17

Wait Menu ............................................................................................................. 6-17

Enabling Channel Output ....................................................................................... 6-17

Setting the Channel Timing Parameters ................................................................ 6-17

Setting Pulse Configuration Output Type .............................................................. 6-17

Setting Channel Mode of Operation ...................................................................... 6-17

Delaying the Start of Channel Output .................................................................... 6-18

Advanced Channel Menus (FUNC + A, B, C, D, E, F, G or H key) ......................... 6-18

Multiplexer Menu .................................................................................................. 6-18

1-2

Configuring the Channel Multiplexer .................................................................... 6-18

Channel Gate Menu ............................................................................................... 6-18

Setting the Sync Source ......................................................................................... 6-18

Setting Channel Gate Control ................................................................................ 6-18

Clocks/Rate Menus (FUNC + RATE key) ................................................................ 6-19

Internal Reference Menu........................................................................................ 6-19

Setting the Internal Reference Source and Rate .................................................... 6-19

Reference Out Menu .............................................................................................. 6-20

Setting the Output Reference ................................................................................. 6-20

Trigger Menus (TRIG key) ........................................................................................ 6-20

Enabling System Trigger ....................................................................................... 6-20

Gate Menus (GATE key) ........................................................................................... 6-20

Standard Gate Menu .............................................................................................. 6-20

Enabling System Gate ............................................................................................ 6-21

System Configuration Menus (FUNC + SYSTEM key) ........................................... 6-21

Communication Interface Menu ............................................................................ 6-21

Setting System Communication Parameters .......................................................... 6-21

User Options Menu 1 ............................................................................................. 6-21

Setting Keypad Parameters .................................................................................... 6-21

User Options Menu 2 ............................................................................................. 6-22

Setting the Auto Start Mode .................................................................................. 6-22

Setting the Display Decimal Mark ......................................................................... 6-22

Setting the Display Brightness ............................................................................... 6-22

Store Menu (FUNC + STORE key) ........................................................................... 6-22

Storing a Configuration.......................................................................................... 6-22

Recall Menu (FUNC + RECALL key) ...................................................................... 6-23

Recalling System Configurations .......................................................................... 6-23

Counter Menu (FUNC + AUX1 key) ........................................................................ 6-23

Using the Counter Function ................................................................................... 6-23

Information Menus (FUNC + 0 key) ......................................................................... 6-23

Information Menu A .............................................................................................. 6-23

Information Menu B .............................................................................................. 6-23

7 Operating the 9520 ................................................................................... 7-24

Quick Start - Normal Internal Rate Generator Operation .......................................... 7-24

Quick Start - Normal External Trigger Operation ..................................................... 7-24

System Timer Overview ............................................................................................ 7-25

To Use Continuous Mode ...................................................................................... 7-25

To Use Single Shot Mode ...................................................................................... 7-25

To Use System Burst Mode Function .................................................................... 7-25

To Use System Duty Cycle Function..................................................................... 7-26

Channel Timer Overview ........................................................................................... 7-26

To Use Channel Normal Mode Function ............................................................... 7-26

To Use Channel Single Shot Function ................................................................... 7-27

To Use Channel Burst Mode Function .................................................................. 7-27

To Use the Channel Duty Cycle Function ............................................................. 7-27

To Use the Channel Gating Function..................................................................... 7-28

1-3

External Input Overview ............................................................................................ 7-28

To Generate a Pulse on Every Trigger Input ......................................................... 7-28

To Generate a Burst of Pulses on Every Trigger Input .......................................... 7-29

To Start a Continuous Stream of Pulses Using the External Trigger .................... 7-29

To use the External Gate to Control the System .................................................... 7-30

8 Programming the 9520 ............................................................................. 8-31

Personal Computer to Pulse Generator Communication ........................................... 8-31

RS232 Interface Overview ......................................................................................... 8-31

USB Interface Overview ............................................................................................ 8-32

GPIB Interface Overview .......................................................................................... 8-32

Ethernet Interface Overview ...................................................................................... 8-32

Programming Command Types and Format .............................................................. 8-33

Line Termination ....................................................................................................... 8-33

IEEE 488.2 Common Command Format ................................................................... 8-33

SCPI Command Format ............................................................................................. 8-34

SCPI Command Keywords .................................................................................... 8-34

SCPI Command Format ......................................................................................... 8-34

SCPI Keyword Separator ....................................................................................... 8-34

SCPI Optional Keywords ....................................................................................... 8-35

SCPI Specific and Implied Channel ...................................................................... 8-35

SCPI Parameter Types ........................................................................................... 8-35

Error Codes ............................................................................................................ 8-35

Programming Examples ......................................................................................... 8-36

9520 INSTrument Commands (SCPI Command Summary) ................................. 8-37

9520 System PULSe[0] Commands (SCPI Command Summary) ........................ 8-38

9520 Channel PULSe[n] Commands (SCPI Command Summary) ...................... 8-39

9520 SYSTem Commands (SCPI Command Summary) ...................................... 8-40

9520 DISPlay Commands (SCPI Command Summary) ....................................... 8-41

IEEE 488.2 Common Commands .......................................................................... 8-41

INTERNAL RATE GENERATOR ............................................................................ A-1

PROGRAMMABLE TIMING GENERATOR .......................................................... A-1

SYSTEM EXTERNAL TRIGGER/GATE INPUT(S) ............................................... A-1

MODULE SPECIFICATIONS ................................................................................... A-2

TTL/ADJUSTABLE DUAL CHANNEL OUTPUT MODULE (Standard) .............. A-2

TRIGGER INPUT .................................................................................................. A-2

GATE INPUT ......................................................................................................... A-3

OPTICAL OUTPUTS ................................................................................................. A-3

OPTICAL INPUTS ..................................................................................................... A-3

EXTERNAL CLOCK IN/OUT .................................................................................. A-3

GENERAL .................................................................................................................. A-3

OUTPUT MODULES ................................................................................................ A-4

Standard .................................................................................................................. A-4

Optional................................................................................................................... A-4

INPUT MODULES .................................................................................................... A-4

Standard .................................................................................................................. A-4

Optional................................................................................................................... A-4

1-4

System Options ....................................................................................................... A-4

Safety Marking Symbols............................................................................................. B-5

COM Overview ........................................................................................................... C-8

GPIB Interface ............................................................................................................ C-8

Ethernet Interface ........................................................................................................ C-8

IP Address and Raw TCP/IP Connection ............................................................... C-8

Determining IP Address .......................................................................................... C-8

DT15 Overview ........................................................................................................ D-10

DT15 Menus ............................................................................................................. D-10

Trigger 1 Menu (TRIG key) ................................................................................. D-10

Enabling System Trigger ...................................................................................... D-10

Trigger 2 Menu (GATE key) * ............................................................................. D-10

DT15 SCPI Command Summary.............................................................................. D-11

DT15 IEEE 488.2 Command Summary ................................................................... D-11

AT35 Description....................................................................................................... E-12

AT35 Specifications................................................................................................... E-12

TZ50 Overview .......................................................................................................... F-13

TZ50 Specifications ................................................................................................... F-13

TTL/CMOS Mode ................................................................................................. F-13

Adjustable Mode .................................................................................................... F-13

Increment Overview.................................................................................................. G-14

Increment Menus ...................................................................................................... G-14

System Mode Menu 1 Extensions (FUNC + MODE key) ................................... G-14

System Mode Menu 2 Extensions (FUNC + MODE key) ................................... G-14

Setting System Increment Modes of Operation .................................................... G-14

Advanced Channel Menu Increment Extension (FUNC + channel key) .............. G-15

Setting the Channel Increment Parameters ........................................................... G-15

Increment SCPI Command Summary ....................................................................... G-15

Increment Initialization and Reset (FUNC + CLR) .................................................. G-15

Increment Specifications ........................................................................................... G-15

AT45 Option Overview ............................................................................................ H-17

AT45 Protection Error Messages .............................................................................. H-17

Module Errors ....................................................................................................... H-17

System Limit Error ............................................................................................... H-17

Voltage Change Timing ........................................................................................ H-17

AT45 Channel Menus ............................................................................................... H-18

Channel Output Configuration Menu ................................................................... H-18

AT45 SCPI Command Extension Summary............................................................. H-18

AT45 Specifications.................................................................................................. H-18

1-5

2 Introduction

This manual is a reference designed to familiarize you with the Quantum

Composers 9520 series pulse generator and is arranged so that you can easily find the information you’re looking for. Generally, each topic has its own section and no section assumes that you’ve read anything else in the manual.

Technical Support

For questions or comments about operating the 9520 -- our technical staff can be reached via one of the following methods:

Phone:

Fax:

Online:

(406) 582-0227

(406) 582-0237 www.quantumcomposers.com

Warranty

In addition to a 30-day money back guarantee, the 9520 has a two-year limited warranty from the date of delivery. This warranty covers defects in materials and workmanship. Quantum Composers will repair or replace any defective unit.

Contact us for information on obtaining warranty service.

Package Contents

The box you receive should contain the following:

9520 Pulse Generator

AC Power Cord

Disc that includes o Operating Manual o Software Drivers o Communication Software

Contact Quantum Composers (406) 582-0227 if any parts are missing.

2-6

3 Safety Issues

Normal use of test equipment presents a certain amount of danger from electrical shock because testing must be performed where exposed voltage is present. An electrical shock causing 10 milliamps of current to pass through the heart will stop most human heartbeats. Voltage as low as 35 V (DC or RMS AC) should be considered dangerous and hazardous since it can produce a lethal current under certain conditions. Higher voltages pose an even greater threat because such voltage can easily produce a lethal current. Your normal work habits should include all accepted practices that will prevent contact with exposed high voltage, and steer current away from your heart in case of accidental contact with a high voltage. You will significantly reduce the risk factor if you know and observe the following safety precautions:

If possible, familiarize yourself with the equipment being tested and the location of its high voltage points. However, remember that high voltage may appear at unexpected points in defective equipment.

Do not expose high voltage needlessly. Remove housing and covers only when necessary. Turn off equipment while making test connections in high voltage circuits. Discharge high voltage capacitors after shutting down power.

When testing AC powered equipment, remember that AC line voltage is usually present on power input circuits, such as the on-off switch, fuses, power transformer, etc.

Use an insulated floor material or a large, insulated floor mat to stand on, and an insulated work surface on which to place equipment. Make certain such surfaces are not damp or wet.

Use the time proven “one hand in the pocket” technique while handling an instrument probe. Be particularly careful to avoid contact with metal objects that could provide a good ground return path.

Never work alone. Someone should always be nearby to render aid if necessary. Training in CPR first aid is highly recommended.

3-7

4 Front Panel Overview

9520 Front Panel

Display Layout and Indicators

A 4 line x 20 character vacuum fluorescent display module displays parameters and status information. The status information is located in the upper-left corner of the display, between the two brackets. There are four enunciators:

A vertical arrow indicates there are additional pages to the current menu.

An alternating hollow and then solid circle indicates the unit is actively generating pulses, or armed and waiting for an external trigger.

A musical note indicates the function key has been pressed.

?

In external oscillator operation, a question mark indicates the internal pll is not yet locked with the external clock signal.

The upper-right side of the display contains the title of the currently displayed menu. The rest of the display is used for system parameters. The display brightness may be adjusted, allowing the instrument to be used under various lighting conditions.

Description of Front Panel Area

Keypads

Three keypad areas provide fast access to various menus and easy editing of system parameters.

Channel Keypad Provides one touch access to the menus for setting up the channel parameters. Pressing the appropriate letter will display the parameters for the corresponding channel. Example: Pressing the A key

Arrow Keypad will access the Channel A menus.

The up (UP) and down (DOWN) arrow keys are used to increment/decrement the current parameter

(indicated by the blinking cursor). The position of the cursor controls the step size for each increment. The

4-8

right (RIGHT) and left (LEFT) arrow keys moves the cursor to different positions within the current parameter. The NEXT key selects the next parameter in the currently displayed menu.

Numeric Keypad Allows numbers and alphanumeric values to be entered. When entering alphanumeric values, pressing a key will display the first letter shown on the key. The yellow FUNC key allows the keys to select the yellow functions.

Repeated key presses will toggle through all the letters, both upper and lower case, shown on the keycap. To enter two letters which appear on the same keycap, select the first character, then use the right arrow to shift to the next position and enter the next letter. When data entry is complete the ENTER key must be pressed.

Rotary Adjustment Knob

As an alternative to the Arrow Keypad, the Rotary Adjustment Knob may be used to adjust the current parameter. The step size is controlled by the position of the cursor; however turning the knob faster will increase the step size. Pushing the knob will perform functions similar to the NEXT key and switch to the next parameter in the currently displayed menu.

Second Level Menus (Function Key)

The second level menus (indicated in yellow above certain keys) are accessed through the use of the yellow function (FUNC) key. Pressing the FUNC key once and then pressing the desired menu key will display the specified second level menu. Pressing the FUNC key twice in succession will put the unit into “Function

Lock” mode, where the second level menus can be accessed without repeatedly pressing the FUNC key. Pressing the FUNC key a third time will exit “Function

Lock” mode.

4-9

5 Pulse Concepts and Pulse Generator

Operations

Counter Architecture Overview

*Start source is: RUN/STOP key in Internal Modes

External input in External Trigger modes

*TRG command via Serial/GPIB access

**Channels are armed by the RUN button. In single shot and burst modes channels may be rearmed by pressing the RUN button.

System Timer Functions

The System Timer functions as a non-retriggerable, multi-vibrator pulse generator. This means that once started, depending on the mode, the timer will produce pulses continuously. Before pulses can be generated, the timer must be armed and then receive a start pulse. Arming the counter is done by pressing the RUN/STOP key. With external trigger disabled, the RUN/STOP key also generates the start command for the counter. With external trigger enabled, the external trigger provides the start pulse. In either case, once started, the counter operation is determined by the System Mode Generator. Standard modes include:

Continuous

Single Shot

Burst

Duty Cycle

Once started T

0

pulses are generated continuously.

One T

0

pulse is generated for each start command.

‘n’ T

0

pulses are generated for each start command.

Once started T

0

pulses cycle on and off continuously.

5-10

The T

0

pulses are distributed to all of the start inputs of the Channel Timers and

Mode Generators

Channel Timer Functions

The Channel Timer functions as a non-retriggerable, delayed, one shot pulse generator. This means that the timer will only generate one delayed pulse for every start pulse received. Once the channel timer has started counting, additional start pulses will be ignored until the pulse has been completed

(non-retriggerable). The start pulse for each channel is provided by the internal

T

0

pulse generated by the Internal System Timer. Whether or not a pulse is generated for each T

0

pulse is determined by the Channel Mode Generator.

Standard modes include:

Normal

Single Shot

A pulse is generated for each T

0

pulse.

One pulse is generated at the first T

0

pulse, after which output is inhibited.

Burst

Duty Cycle

A pulse is generated for each T

0

pulse, 'n' times, after which output is inhibited.

‘n’ pulses are generated for each T

0

pulse after which the output is inhibi ted for ‘m’ times. The cycle is then repeated.

Different modes may be selected for each output, allowing a wide variety of output combinations. Each output may also be independently disabled or gated

(using the external gate input).

Digital Output Multiplexer

The outputs of the Channel Timers are routed to a set of multiplexers. This allows routing of any or all Channel Timers to any or all of the unit outputs. In the normal mode of operation, the output of the T n

Channel Timer is routed to the T n output connector. As an example, if a double pulse is required on Channel A output, one can multiplex the Channel A timer with the Channel B timer adjusting each timer to provide the necessary pulses.

Dependent & Independent Timing Events

The 9520 allows the user to control the relationship between the Channel Timers by setting the sync source for each timer. Independent events are all timed relative to the internal T

0

start pulse. Dependent events may be linked together by setting the sync source to the controlling event. This allows the instrument to match the timed events and adjustments can be made in one event without detuning the timing between it and a dependent event.

5-11

Navigating the 9520 Front Panel

Selecting Menus

Parameters are grouped in menus, selectable using menu keys. To select the output channel parameters press the letter key corresponding to the desired channel. To select second level menus press the FUNC key and then the key corresponding to the desired function. To select advanced channel menus press the FUNC key followed by the desired channel key. Menus may include a number of different pages with each page containing up to four parameters. The status block in the upper-left corner of the display shows a vertical arrow if the current menu contains additional pages. To select the next page, press the channel button again or select the same second level menu by pressing the

FUNC key and the channel/menu key again.

Selecting Menu Items

Within a menu, the blinking cursor indicates the current menu item for editing.

The NEXT key or pressing the adjustment knob will select a different menu item.

Numeric Input Mode

When the current item is numeric, the system enters the Numeric Input Mode. In this mode data may be edited in one of three ways. Using the arrow keypad, the left (LEFT) and right (RIGHT) arrow keys are used to select a digit to edit. The selected digit blinks to identify itself as the active digit. The UP and DOWN arrow keys are then used to increment or decrement this digit. Alternately, after using the LEFT and RIGHT arrow keys to select an active digit, the adjustment knob may be used to increment and decrement this digit. The adjustment knob features speed dependent resolution. Slow rotation will increment or decrement the active digit by one. As you increase the speed of rotation, the parameter will be 10 to 1000 times faster depending on the speed.

An additional entry mode is using the numeric keypad. Enter the number, including decimal point using the numeric keypad. Complete the number using the ENTER key. To clear number entry and/or start over press the clear key

(CLR). Pressing the CLR key a second time will exit the numeric keypad mode and restore the original number.

The last entry mode is a modified form of scientific notation. The FUNC button acts as 10

-

in this case. Type in the value followed by FUNC then the number button that represents the power desired. For example 64us is entered as 6, 4,

FUNC, and then 6.

Entering Non-Numeric Parameters

When the current item is non-numeric, the UP and DOWN arrow keys are used to select among different options for the parameter. The adjustment knob may also be used to change the selection. If the item is an on-off toggle, the UP arrow

5-12

(CW adjustment knob) enables the item and the DOWN arrow (CCW adjustment knob) disables the item.

Alphanumeric Input Mode

When the current item is alphanumeric, the system enters the Alphanumeric

Input Mode. In this mode, data is entered using the alphanumeric keypad.

Pressing a key will display the first letter shown on the keypad. Repeated key presses will toggle through all the letters, both upper and lower case, shown on the key cap. To enter two letters which appear on the same key cap, select the first character, then use the right arrow to shift to the next position and enter the next letter. The Left and Right arrow keys may be used to position the cursor to edit any character. When data entry is complete, the ENTER key must be pressed. The keys contain the following characters:

1 1 2 3 4 5 6 7 8 9 0

2 A B C a b c 2

3 D E F d e f 3

4 G H I g h i 4

5 J K L j k l 5

6 M N O m n o 6

7 P Q R S p q r s 7

8 T U V t u v 8

9 W X Y Z w x y z 9

0 0 1 2 3 4 5 6 7 8 9

.

. , # $ % & ? • -- + * / space

Enabling System Output

The RUN/STOP key is used to arm the system. With external trigger disabled, the key will arm and start pulse output. With external trigger enabled, the key will arm the pulse generator. Pulse output then starts after the first valid trigger input.

Pressing the RUN/STOP key a second time disables the pulse generator.

Enable/Disable Channel Output

At the top of each channel menu page is a parameter to enable or disable the output of the channel. Each channel may be individually enabled or disabled. An illuminated channel key indicates that the channel is enabled.

Rearming the Channel Timers

In the channel single shot mode and burst mode, the Channel Timers may be rearmed after completing the initial output by pressing the FUNC key followed by the RUN/STOP key. If there are channels currently running in normal mode, single shot and burst channels can be re-armed without affecting the timing on normal mode channels by pressing function RUN/STOP button.

5-13

Setting Pulse Timing Parameters

Pulses are defined by a delay, from their sync or start pulse to the active edge, and a width.

Wid:

Dly:

Sets the width of the active portion of the pulse.

Sets the delay from the sync source to the start of the pulse.

NOTE: If Wid + Dly + 75ns (hardware reset time) > T

0

Period, the correct pulse

width will be generated but at a slower rate.

Setting Pulse Output Parameters

There are three basic types of outputs available on the 9520: (a) TTL/CMOS compatible outputs; (b) adjustable amplitude outputs; (c) optical outputs.

Out: Selects between TTL/CMOS mode and Adjustable mode when both are available on a single output.

Pol: Sets the voltage polarity of the pulse, active high or active low. Note: All outputs are positive - negative voltages are not supported.

Ampl: In adjustable mode, it sets the unloaded output voltage. The actual output voltage will depend on the load impedance. For example: If the load is 50 ohms, the output will be 50% of the stated voltage.

Using the Output Multiplexer

Each output channel includes a multiplexer which allows routing any or all of the timer outputs to the physical output. This allows double pulses and other complex pulse trains to be generated. Only timing parameters are multiplexed together, not amplitudes.

Mux:

-HGFEDCBA-

-00000101-

The multiplexer is represented by an

“n” bit binary number as shown above. “n” is the number of channels. Each bit represents a channel timer, which is enabled by setting the bit to one. In the above example, timers A and C are combined on the current output.

Setting System Internal Rate Parameters

The internal T

0

period controls the fundamental output frequency of the system.

Each channel may operate at submultiples of the fundamental frequency using their duty cycle mode.

Source: Sets the reference source for the internal T Period.

Per: Sets the internal T Period.

5-14

To set the system Internal Rate press the yellow FUNC key, then press the

RATE key, and then use the dial or number pad to specify the T

0

Period.

5-15

6 9520 Menu Structure

System Mode Menus (FUNC + MODE key)

MODE:

Continuous

MODE: Single

Shot

MODE: Burst

#/Burst

MODE: Duty

Cycle

#/On

#/Off

Setting System Mode of Operation

The MODE menu sets the T

0

system timer mode. The menu will show the extra set parameters (Burst, On & Off) only when they are appropriate.

Mode:

Burst:

Selects the T

0

mode: Continuous, Single Shot, Burst or Duty

Cycle mode.

Sets the number of pulses to be generated when in Burst mode.

On: Sets the number of pulses to be generated during each on cycle.

Off: Sets the number of pulses to skip each during off cycle when in the Duty Cycle mode.

*NOTE: Any mode may be started by either the RUN/STOP key in the internal trigger mode or armed by the RUN/STOP key and started by an external trigger in the external trigger mode. In the single shot and burst modes, (internally triggered) the unit disarms itself at the end of the pulse train. Pressing the

RUN/STOP key after the unit has been disarmed will generate a new pulse train.

Channel Menus (A, B, C, D, E, F, G, or H key)

Timing Menu

Channel Enable

Sync Source

Pulse Width

Delay

Output Configuration Menu

Channel Enable Channel Enable Channel Enable Channel Enable Channel Enable

Output Type:

TTL/CMOS

Output Type:

Adjustable

Output Type:

Optical

Output Type:

High Z

Output Type:

Low Z

Polarity Polarity

Output Level

Polarity Polarity

Output Level Output Level

6-16

Mode Menu

Channel Enable Channel Enable Channel Enable Channel Enable

Mode: Normal Mode: Single

Shot

Mode: Burst Mode:

Duty Cycle

#/Burst Pulses

#/On Pulses

#/Off Pulses

Wait Menu

Channel Enable

#/Wait Pulses

Enabling Channel Output

At the top of each of the channel menu pages is a parameter to enable or disable the channel. Each channel may be individually controlled. When enabled, the channel key will illuminate.

Setting the Channel Timing Parameters

To define a pulse requires two parameters: the delay to the active edge and the width of the pulse.

Wid:

Dly:

Sets the channel pulse width.

Sets the channel delay until the active edge.

Setting Pulse Configuration Output Type

The 9520 supports two types of outputs: a high speed TTL/CMOS compatible output and for applications which require different voltage levels or higher current, an adjustable voltage output. The pulses can also be defined to be active high or active low.

Out: Selects the output mode; TTL/CMOS, Adjustable, Optical,

High Impedance (Hi Z), or Low Impedance (Lo Z)

Pol:

Ampl:

Sets the pulse polarity, active high or active low.

Sets the output voltage level when in the Adjustable mode.

Setting Channel Mode of Operation

Each channel may be set independently to operate in one of four modes: normal, single shot, burst, or duty cycle (within the CHANNEL menus):

Mode: Selects the mode for the current channel. Additional

Brst: parameters are provided for the burst mode and the duty cycle mode.

Sets the number of pulses in the burst mode to gener-ate before inhibiting output.

6-17

On:

Off:

Sets the number of pulses to generate before inhibiting output in Duty Cycle Mode.

Sets the number of pulses to inhibit before repeating the On

Cycle in Duty Cycle Mode.

Delaying the Start of Channel Output

Within any channel mode, the output of the channel can be delayed using the wait parameter (within the CHANNEL menu):

Wait: Sets the number of T channel output.

0

pulses to wait until enabling the

Advanced Channel Menus (FUNC + A, B, C, D, E, F, G or H key)

Multiplexer Menu

-HGFE DCBA-

Multiplexed

Channels

Configuring the Channel Multiplexer

To define which channels are fed into the channel multiplexer, the corresponding bit for the desired channel to add should be set to 1. All desired omitted channels should have the corresponding bit set to 0.

Mux: Enable/disable bit field.

Channel Gate Menu

Channel Enable Channel Enable Channel Enable

Ch Gate Mode:

(Gate Menu)

Ch Gate Mode:

Pulse Inhibit

Logic Level

Ch Gate Mode:

Output Inhibit

Logic Level

Setting the Sync Source

Although each channel receives its start pulse from the internal T

0

pulse, the start pulse can be assigned such that the delay entered is relative to the T

0

pulse or any other channel pulse. This allows dependent events to link. The unit will not allow a circular chain of sync sources that would result in a channel triggering itself. The delay entered is relative to the selected sync source.

Sync Source: Selects the channel sync source.

Setting Channel Gate Control

When the global gate is set (Chan Menu), the channel can then use the gate input with independent behavior from other channels.

6-18

Gate:

Logic:

Enables the GATE input for the channel by setting the method of output control used with the gating function.

Sets the logic level used with the gating function, either active high or active low.

“Pulse Inhibit” method

T he gate prevents the channel from being triggered by the channel’s trigger source pulse. If a pulse has already started when the gate disables the channel, the pulse will continue normal output but will not restart on the next trigger pulse.

“Output Inhibit” method

The gate leaves the base triggering alone and enables/ disables the output directly.

Clocks/Rate Menus (FUNC + RATE key)

Internal Reference Menu

Source:

System Osc

T

0

Period

Source:

50MHz Ext

T

0

Period

Source:

10MHz Ext

T

0

Period

Source:

80MHz Ext

T

0

Period

Source:

20MHz Ext

T

0

Period

Source:

100MHz Ext

Source:

25MHz Ext

T

0

Period

Source:

40MHz Ext

T

0

Period

Setting the Internal Reference Source and Rate

Source:

To:

Selects the internal or external clock source from which the unit will operate.

Sets the T

0

period which determines the fundamental output frequency of the unit.

6-19

Reference Out Menu

Ref Out: T

0

Pulse

T

0

Period

Ref Out: 20MHz Ref Out:

T

0

Period

16.67Mhz

T

0

Period

Ref Out: 10MHz

T

0

Period

Ref Out:

100MHz

T

0

Period

Ref Out: 50MHz Ref Out:

33.33MHz

T

0

Period T

0

Period

Ref Out:

14.2857MHz

T

0

Period

Ref Out:

12.5MHz

T

0

Period

Ref Out: 25MHz

T

0

Period

Ref Out:

11.11MHz

T

0

Period

Setting the Output Reference

Ref Out: Selects the frequency of the output reference for synchronizing with external system components.

Trigger Menus (TRIG key)

Mode: Disabled Mode:

Triggered

Threshold Level

Trigger Edge

Enabling System Trigger

Enable the use of the TRIG input by the system timer as a trigger source.

Mode:

Level:

Selects between disabling/enabling the trigger mode.

Sets the trigger threshold. Edge: Selects between rising and falling edges as the trigger source when a trigger mode is enabled.

Gate Menus (GATE key)

Standard Gate Menu

Mode: Disabled Mode: Pulse

Inhibit

Threshold

Level

Logic Level

Mode: Output

Inhibit

Threshold

Level

Logic Level

Mode: (Chan

Menu) *

Threshold Level

* Gate Mode is controlled on a per channel basis from the Advanced Channel

Gate Menu

6-20

Enabling System Gate

Enables the use of the GATE input as a trigger inhibit or output control for all channels simultaneously, or on a per channel basis.

Mode:

Level:

Logic:

Selects between disabling the GATE inputs and method of output control

Sets the gating threshold.

Sets the active logic level.

System Configuration Menus (FUNC + SYSTEM key)

Communication Interface Menu

Interface:

RS232

Baud Rate

Echo

Interface: USB Interface: GPIB

*

Address

Interface:

Ethernet *

* Instrument must be equipped with the Extended Communications Option.

(Visit www.quantumcomposers.com

for more details).

Setting System Communication Parameters

The 9520 comes with a standard RS232 serial port and USB port. The unit will not respond to computer commands unless these ports are properly configured.

Interface: RS232, USB, GPIB (optional), Ethernet (optional)

Baud Rate:

Echo:

Selects the baud rate for the selected interface.

Selects whether to echo characters back to the host computer or not. Address: Sets the GPIB address.

User Options Menu 1

Key Rate

Key Volume

Knob Volume

Setting Keypad Parameters

The rate at which a key will repeat itself when held down may be set. This can be used to provide a controlled rate at which a parameter is incremented. In addition, the volume of the beep can be controlled for both the keypad and the adjustable knob.

Key Rate:

Key Vol:

Knob Vol:

Sets the rate at which the keys will repeat when held down.

Sets the beep volume for the keypad.

Sets the beep volume for the Rotary Knob.

6-21

User Options Menu 2

Auto Start

Mode

Decimal Mark

LCD Brightness

Setting the Auto Start Mode

The unit may be configured to automatically start generating pulses after power up.

Setting the Display Decimal Mark

Mark:

Selects the format of the decimal mark, “.” or “,”.

Setting the Display Brightness

LCD: Adjusts display brightness.

Store Menu (FUNC + STORE key)

Configuration #

Name

Help Line

Storing a Configuration

Use the following procedure to store a complete system configuration:

Set all parameters to the desired value.

Select a configuration number.

*NOTE: You cannot store to the zero location, as that contains the factory default values.

Label the configuration as desired.

From the Store menu, press the store button sequence (FUNC + STORE).

6-22

Recall Menu (FUNC + RECALL key)

Configuration #

Name

Help Line

Recalling System Configurations

Use the following procedure to recall a stored or default system configuration:

Enter the Recall Menu (FUNC + RECALL).

Select a configuration number.

From the Recall Menu, press the recall key sequence (function + recall).

*Note: Configuration 0 is the factory default setting.

Counter Menu (FUNC + AUX1 key)

Counter Enable Counter Enable

T Counts T Counts

Help Line

G Counts *

Help Line

* Enabled when the Trigger Mode is set to Dual Trigger operation

(Instrument must be equipped with the Dual Trigger Option. Visit www.quantumcomposers.com

for more details).

Using the Counter Function

The Counter function counts the number of T

0

pulses output by the system clock.

When the unit is operated in system single shot mode, the T

0

count reflects the number of incoming trigger pulses.

Information Menus (FUNC + 0 key)

Information Menu A

Model Number

Serial Number

Firmware Ver.

FPGA Ver.

Information Menu B

FW ID #

GA ID #

Module IDs

Instr. Options

The Information Menus provide all of the pertinent version numbers and serial numbers for the unit. This information should be readily available when contacting customer service for troubleshooting help.

6-23

7 Operating the 9520

Quick Start - Normal Internal Rate Generator Operation

The 9520 has a powerful set of functions providing a number of modes of opera tion for the internal or “System” rate generator (T

0

). Most of these functions can be ignored if a simple continuous stream of pulses is required. Starting from the default settings, which can be restored by recalling configuration 0, the following parameters need to be set:

Pulse Width, Delay Enter the Channel menus by pressing the letter key.

Enter the required pulse width and delay. Repeat for each output channel.

T

0

Period Enter the Rate menu by pressing the FUNC key and then the RATE key. Set the desired pulse period.

Note that in general, the pulse delay plus the pulse width, plus a 75ns hardware reset constant, for any channel must be less than the T

0

period.

Start

Stop

Press the RUN/STOP key to start generating pulses.

Press the RUN/STOP key a second time to stop gen-erating pulses.

Quick Start - Normal External Trigger Operation

To generate a single pulse for every external trigger event, based on the default configuration 0, the following parameters need to be set:

System Mode Enter the System Mode menu by pressing the FUNC key and then the MODE key. Select Single Shot mode.

Trig Enter the Trigger menu by pressing the TRIG key.

Select Triggered.

Level Press the NEXT key until the Level parameter is highlighted. Set the trigger threshold voltage to approximately 50% of the trigger signal amplitude.

Edge Press the NEXT key until the Edge parameter is high-lighted. Set the instrument to trigger off the rising edge or falling edge as desired.

7-24

Pulse Width, Delay Enter the Channel menus by pressing the letter key.

Enter the required pulse width and delay. Repeat for each output channel.

Start Press the RUN/STOP key to start/arm the instrument.

The 9520 will now generate a pulse for every valid trigger.

Stop Press the RUN/STOP key a second time to stop/disarm the instrument (i.e. to stop generating pulses).

System Timer Overview

For internal operation, the 9520 contains a timer and mode generator which generates an internal T

0

clock that is used to trigger all the channel timers.

System modes are controlled via the MODE menu.

To Use Continuous Mode

The RUN/STOP button starts and stops a continuous pulse stream at the rate specified by the Rate menu. This corresponds to the normal output mode for most pulse generators.

To generate a continuous stream of pulses:

- within the system Mode menu

Mode Select Continuous for the system mode.

- within the Rate menu

Source Select the system oscillator or the external clock in

Period frequency.

Set the desired period.

Pressing the RUN/STOP key will now generate a stream of T

0

pulses at a rate specified by the period parameter.

To Use Single Shot Mode

To generate a single pulse with every press of the RUN/STOP key:

- within the system Mode menu

Mode Select Single Shot for the system mode.

Pressing the RUN/STOP key will now generate a single pulse.

To Use System Burst Mode Function

The RUN/STOP button generat es a stream of “n” T

0 pulses, where “n” is speci-fied by the Burst parameter. The rate is specified in the Rate menu.

7-25

Pressing the RUN/STOP button while the burst is in process will stop the output.

After the burst has been completed, pressing the RUN/STOP button will generate another burst. T

0

generate a burst of pulses:

- within the system Mode menu

Mode:

Burst:

Select the Burst mode.

Set the number of pulses to produce in the burst.

To Use System Duty Cycle Function

The RUN/STOP button starts a continuous pulse stream which oscillates on for the “n” pulses and off for “m” pulses, where “n” and “m” are specified by the On and Off parameters, respectively. The rate is specified in the Rate Menu. To generate a stream of pulses which oscillates on fo r ‘n’ pulses and off for ‘m’ pulses:

- within the system Mode menu

Mode: Select the Duty Cycle mode.

On: Set the number of pulses to produce during the on

Off: cycle.

Set the number of pulses to skip during the off cycle.

- within the Rate menu

Source: Select the system oscillator or the external clock in frequency.

Period: Set desired Period.

Channel Timer Overview

The output of each channel is controlled by two timers to generate the delay timing and the pulse width. All channels are simultaneously triggered, depending on the system mode, by the internal T

0

pulse, the external trigger, or a trigger provided by the operating software. A given channel may or may not generate a pulse depending on its own channel mode as described below.

When one channel is generating a continuous stream of pulses, a user can trigger a single shot or burst of pulses on another channel without interrupting the continuous stream by pressing the FUNC and the

RUN/STOP key.

To Use Channel Normal Mode Function

The Normal mode generates a continuous stream of pulses at a rate determined by the system timer:

- within the Channel menus

Enable: Select Enable to enable channel output.

Dly:

Wid:

Set the desired delay.

Set the desired pulse width.

7-26

Mode: Select the Normal mode.

Pressing the RUN/STOP key will now generate a continuous stream of pulses.

To Use Channel Single Shot Function

The Single Shot mode generates a single pulse every time the RUN/STOP key is pressed. If the unit is in the active state, ( i.e. channels which are set to the

Normal mode are producing pulses), pressing the FUNC key and RUN/STOP key will reset the Single Shot counters and generate one pulse in sync with the other channels running in the Normal mode. To use the Single Shot mode:

- within the Channel menus

Enable:

Delay:

Select "Enabled" to enable channel output.

Set the desired delay.

Width:

Mode:

Set the desired pulse width.

Select the Single shot mode.

To Use Channel Burst Mode Function

The Burst mode generates a burst of pulses every time the RUN/STOP key is pressed. If the unit is in the active state, (i.e. channels which are set to the

Normal mode are producing pulses), pressing the FUNC - RUN/STOP key sequence will reset the Burst counters and generate a new set of pulses in sync with the other channels running in the Normal mode. FUNC - RUN/STOP will not affect T

0

pulse status. To use the Burst mode:

- within the Channel menus

Enable: Select

“Enabled” to enable channel output.

Delay: Set the desired delay.

Width:

Mode:

#/Burst:

Set the desired pulse width.

Select the Burst mode.

Set the number of pulses to produce in the burst.

To Use the Channel Duty Cycle Function

To generate a stream of pulses which oscillates on for ‘n’ pulses and off for ‘m’ pulses:

- within the Channel menus

Enable: Select

“Enabled” to enable channel output.

Delay:

Width:

Mode:

On Cycle:

Set the desired delay.

Set the desired pulse width.

Select the Duty Cycle mode.

Set the number of pulses to produce during the on

Off Cycle: cycle.

Set the number of pulses to skip during the off cycle.

*Note: Older Quantum Composers pulse generators had a divide-by-n function.

The duty cycle mode is a more general case. To reproduce the divide-by-n

7-27

function, set the on cycle to 1 and set the off cycle to (n-

1), where “n” is the divide-by-n factor.

To Use the Channel Gating Function

Each channel may use the external input to gate or control its output. The gate controls the triggering of the channel. To use the gate, set the following parameters

- within the Channel menu

Mode:

Channel Gate:

Logic:

In Gate Menu must be set to Channel Menu.

Select

“Pulse Inh” or “Output Inh”.

Select active high or active low.

In the

“Pulse inhibit” method, the gate prevents the channel from being triggered by the channel’s trigger source pulse. If a pulse has already started when the gate disables the channel, the pulse will continue normal output but will not restart on the next pulse. In the

“Output inhibit” method, the gate leaves the base triggering alone and enables/disables the output directly. Output pulses will immediately cease when the gate signal is removed

External Input Overview

The external inputs may be used to trigger the unit or to gate the system or channel timers. When using a trigger input, the external input acts as a system start pulse. Depending on the system mode, the result of a trigger input can be either a single pulse, a burst of pulses or the start of a stream of pulses.

To Generate a Pulse on Every Trigger Input

To generate a pulse on every external trigger received, set the following param-eters:

- within the Mode menu

Mode: Select the Single Shot mode.

- within the Trigger menu

Mode: Select Triggered mode.

Level:

Edge:

Set the trigger threshold level.

Select which edge, rising or falling, to trigger on.

Pressing the RUN/STOP key will arm the unit. Once the unit is armed, it will generate a T

0

pulse for every external trigger received. Pressing the

RUN/STOP key will disarm the unit. This mode corresponds to the normal external trigger mode found on most other pulse generators.

7-28

To Generate a Burst of Pulses on Every Trigger Input

To generate a burst of pulses for every external trigger received set the following parameters:

- within the Mode menu

Mode: Select the Burst mode.

Burst: Set the number of pulses to generate in each burst.

- within the Rate menu

To: Set the period between pulses.

- within the Rate menu

Source: Select the system oscillator or an external clock in

To: frequency.

Set the period between pulses.

- within the Trigger menu

Mode:

Level:

Edge:

Select Triggered mode.

Set the trigger threshold level.

Select which edge, rising or falling, to trigger on.

Pressing the RUN/STOP key will arm the unit. Once the unit is armed, it will generate a set of pulses for every external trigger received. The unit is reset at the end of a burst and will generate another set of pulses upon receiving a new trigger. Triggers that occur in the middle of a burst are ignored. Pressing the RUN/STOP key will disarm the unit.

To Start a Continuous Stream of Pulses Using the External Trigger

The external trigger may be used to start the unit generating pulses:

- within the Mode menu

Mode: Select the Continuous mode.

- within the Rate menu

Source: Select the system oscillator or an external clock in

To: frequency.

Set the period between pulses.

- within the Trigger menu

Mode: Select the Trigger mode.

Level:

Edge:

Set the trigger threshold level.

Select which edge, rising or falling, to trigger on.

Pressing the RUN/STOP key will arm the unit. Once the unit is armed, it will begin generating pulses after an external trigger is received. Triggers

7-29

that occur after the pulses start are ignored. Pressing the RUN/STOP key a second time will disarm the unit.

To use the External Gate to Control the System

The external gate may be used to control the output of the unit. To gate the system timer:

- within the Mode menu

Mode: Select the desired mode.

- within the Rate menu

Source: Select the system oscillator or an external clock in

Per: frequency.

Set the period between pulses.

- within the Gate menu

Mode: Select

“Pulse Inh” or “Output Inh”.

Level:

Logic:

Set the gate threshold level.

Select active high or active low.

Pressing the RUN/STOP key will arm the unit. Once the unit is armed, it will begin generating pulses whenever the external gate input is in the active state. Pressing the RUN/STOP key a second time will disarm the unit.

7-30

8 Programming the 9520

Personal Computer to Pulse Generator Communication

The 9520 ships standard with an RS232 serial and USB interface. Ethernet and

GPIB interfaces are available as an option. All menu settings can be set and retrieved over the computer interface using a simple command language. The command set is structured to be consistent with the Standard Commands for

Programmable Instruments. Although due to the high number of special features found in the 9520, many of the commands are not included in the specification.

The syntax is the same for all interfaces. The amount of time required to receive, process, and respond to a command at a Baud rate of 115200 is approximately

10 ms. Sending commands faster than 10 ms may cause the unit to not respond properly. It is advised to wait until a response from the previous command is received before sending the next command.

WARNING:

When communicating with the unit (USB, Serial or Ethernet), avoid sending any commands that include the “*” character as the unit is booting up.

This may result in an undesired lockup of the instrument.

RS232 Interface Overview

The serial port is located on the back of the 9520 and uses a 9-pin D-type connector with the following pinout (as viewed from the back of the unit):

1

2

No Connection

Tx - Transmit (to computer)

3

4

5

Rx - Receive (from computer)

DTR - Connected to pin 6

Ground

6

7

8

9

DSR - Connected to pin 4

RTS - Connected to pin 8

CTS - Connected to pin 7

No Connection

The serial port parameters should be set as follows:

Baud Rate 4800, 9600 19200, 38400, 57600, 115200*

Data Bits 8

Parity None

Stop Bits 1

*The default baud rate for the RS232 is 115200.

8-31

USB Interface Overview

The USB interface is standard on the 9520. Before this type of communication can be used, the appropriate drivers must be installed on the personal computer

(pc). These drivers are included on the CD that was shipped with your unit.

Please contact Quantum Composers or visit www.quantumcomposers.com for updated installation files and instructions.

USB communication is achieved by using a mapped (virtual) COM port on the

PC. The driver installation executable will obtain an unused COM port number, install the USB drivers, and make that COM port number available for typical

RS232 communication to the pulse generator. HyperTerminal or other common software may be used.

When communicating through the mapped COM port over USB, the baud rate for the communication port used by the USB chip must match the baud rate for the

COM port on the PC. Access to the USB port baud rate is done using the SCPI command “:SYSTem:COMMunicate:SERial:USB <baud rate>” command. This parameter can be accessed via any communication method. The default baud rate for USB is 38400.

USB communication notes:

The correct drivers must be installed on the personal computer before communication can be accomplished via USB.

The BAUD rates on the PC and on the pulse generator must match for successful communication.

The USB port’s BAUD rate on the pulse generator can be set using the

SCPI command “:SYSTem:COMMunicate:SERial:USB <baud rate>” where <baud rate> can be: o 4800 o 9600 o 19200 o 38400 (default)

USB 1.0 specification is used. The USB cable can be removed without

“unplugging” the device in the operating system environment.

Echo functionality is not available on the USB port.

GPIB Interface Overview

A GPIB interface is optional on the 9520. Refer to Appendix C for more

information.

Ethernet Interface Overview

An Ethernet interface is optional on the 9520. Refer to Appendix C for more

information.

8-32

Programming Command Types and Format

The 9520 Pulse Generator uses two types of programming commands: IEEE

488.2 Common Commands and Standard Commands for Programmable

Instruments (SCPI). The format is the same for all interfaces. HyperTerminal (in

Windows) or any other generic terminal program may be used to interactively test the commands using the RS232 interface. The format of each type is described in the following paragraphs.

Line Termination

The pulse generator uses text-style line terminations. When a command is sent to the unit, the firmware is programmed to read characters from a communication port until it reads the line termination sequence.

The command string is parsed and executed after reading these characters.

These characters are the “carriage return” and “linefeed”. They are ASCII char-acter set values 13 and 10 respectively (hex 0x0D and 0x0A). All command strings need to have these characters appended.

When the pulse generator responds to a command, whether it is a query or a parameter change, it also appends its return strings with these characters.

Coded applications could use this behavior to know when to stop reading from the unit. However, if the “echo” parameter is enabled, there will be two sets of line terminators, one following the echoed command string, and one following the pulse generator’s response.

Note: The pulse generator will echo commands on the DB9 serial port only.

The pulse generator responds to every communication string. If the communica-tion string is a query, the unit responds with the queried response (or error code) followed by the line terminators. If the communication string is a parameter change, the response is “ok” (or error code) followed by the line terminators. For this reason, it is not recommended that multiple commands be stacked together into single strings as is common with some other types of instruments. It is recommended that the coded application send a single command in a string and follow immediately by reading the response from the unit. Repeat this sequence for multiple commands.

IEEE 488.2 Common Command Format

The IEEE 488.2 Common Commands control and manage generic system functions such as reset, configuration storage and identification. Common commands always begin with the asterisk (*) character and may include parameters. The parameters are separated from the command pneumonic by a space character. For Example:

*RST<cr><lf>

*RCL 1<cr><lf>

8-33

*IDN?<cr><lf>

SCPI Command Format

SCPI Command Keywords

The commands are shown as a mixture of upper and lower case letters. The upper case letters indicate the abbreviated spelling for the command. You may send either the abbreviated version or the entire keyword. Upper and/or lower case characters are acceptable.

For example, if the command keyword is given as POLarity, then POL and

POLARITY are both acceptable forms; truncated forms such as POLAR will generate an error; polarity, pol, and PolAriTy are all acceptable as the pulse generator is not case sensitive.

SCPI Command Format

SCPI commands control and set instrument specific functions such as setting the pulse width, delay and period. SCPI commands have a hierarchical structure composed of functional elements that include a header or keywords separated with a colon, data parameters, and terminators. For example:

SCPI Format

:PULSE1:STATE ON<cr><lf>

:PULSe1:WIDTh 0.000120<cr><lf>

:PULSe:POL NORMal<cr><lf>

Any parameter may be queried by sending the command with a question mark appended. For example:

- QUERY FORMAT

:PULSE1:STATE?<cr><lf>

Will return: 1<cr><lf>

:PULSE1:WIDT?<cr><lf>

Will return: 0.000120000<cr><lf>

:PULSE1:POL?<cr><lf>

Will return: NORM<cr><lf>

SCPI Keyword Separator

A colon (:) must always separate one keyword from the next lower-level keyword.

A space must be used to separate the keyword header from the first parameter.

If more than one parameter is used, you must separate subsequent parameters with a comma.

8-34

SCPI Optional Keywords

Optional keywords and/or parameters appear in square brackets ( [ ] ) in the command syntax. Note that the brackets are not part of the command and should not be sent to the pulse generator. When sending a second level key-word without the optional keyword, the pulse generator assumes that you intend to use the optional keyword and responds as if it had been sent.

SCPI Specific and Implied Channel

Some commands, such as PULSe, allow specifying a channel with an optional numeric keyword suffix. The suffix will be shown in square brackets [1 / 2]. The brackets are not part of command and are not to be sent to the pulse generator.

The numeric parameters correspond to the following channels: 0 = T

0

, 1 = ChA,

2 = ChB, etc. Only one channel may be specified at a time.

If you do not specify the channel number, the implied channel is specified by the

:INSTrument:SELect command or the last referenced channel.

After power-up or reset (*RST) the instrument default is channel #1.

SCPI Parameter Types

The following parameter types are used:

<numeric value> Accepts all commonly used decimal representation of numbers including optional signs, decimal points, and scientific notation: 123, 123e2, -123, -1.23e2, .123,

1.23e-2, 1.2300E-01.

<boolean value> Represents a single binary condition that is either true

<identifier> or false. True is represented by a 1 or ON; false is rep-resented by a 0 or OFF. Queries return 1 or 0.

Selects from a finite number of predefined strings.

Error Codes

The unit responds to all commands with either: ok<cr><lf> or ?n<cr><lf>

Where "n" is one of the following error codes:

1

2

Incorrect prefix, i.e. no colon or * to start command.

Missing command keyword.

3

4

5

6

7

8

Invalid command keyword.

Missing parameter.

Invalid parameter.

Query only, command needs a question mark.

Invalid query, command does not have a query form.

Command unavailable in current system state.

8-35

Programming Examples

Example 1)

20 ms pulse width, 2.3 ms delay, 10 Hz, internal trigger, continuous operation.

:PULSE1:STATE ON<cr><lf>

:PULSE1:POL NORM<cr><lf>

:PULSE:WIDT 0.020<cr><lf>

:PULSE1:DELAY 0.0023<cr><lf> enables channel A sets polarity to active high sets pulse width to 20 ms sets delay to 2.3 ms

:PULSE0:MODE NORM<cr><lf>

:PULSE0:PER 0.1<cr><lf> sets system mode to continuous sets period to 100 ms (10 Hz)

:PULSE0:TRIG:MODE DIS<cr><lf> disables the external trigger

To start the pulses use either of the following commands:

:PULSE0:STATE ON<cr><lf>

:INST:STATE ON<cr><lf> starts the pulses alternate form to start pulses

Example 2)

25µs pulse width, 0 delay, external trigger, one pulse for every trigger.

:PULSE1:STATE ON<cr><lf>

:PULSE1:POL NORM<cr><lf>

:PULSE:WIDT 0.000025<cr><lf>

:PULSE1:DELAY 0<cr><lf>

:PULSE0:MODE SING<cr><lf> enables channel A sets polarity to active high sets pulse width to 25µs sets delay to 0 sets system mode to single shot

:PULSE:TRIG:MODE TRIG<cr><lf> sets system to external trigger

:PULS:TRIG:LEV 2.5<cr><lf> sets trigger level to 2.5 v

:PULS:TRIG:EDGE RIS<cr><lf> set to trigger on rising edge

To arm the instrument in external gate mode, use either of the following commands:

:PULSE0:STATE ON<cr><lf>

:INST:STATE ON<cr><lf> arms the instrument alternate form if T

0

is currently selected

A software generated external trigger can be generated by using the following command:

*TRG<cr><lf> generates a software external trigger

8-36

9520 INSTrument Commands (SCPI Command Summary)

Keyword

:INSTrument

Parameter

:CATalog?

:FULL?

:COMMands?

:NSELect

:SELect

:STATe

0-8

T0,CHA,CHB,CHC,CHD,

CHE,CHF,CHG,CHH

0/1 or ON/OFF

Comments

Subsystem. Supports treating each channel as a logical instrument.

Query only. Returns a comma-separated list of the names of all channels. A two channel instrument would return:

T0,CHA,CHB.

Query only. Returns a comma-separated list of the names of all channels and their associated number. A two channel instrument would return: T

0

, 0, CHA, 1,

CHB, 2.

Query only. Returns an indented list of all

SCPI commands.

Selects a channel using the channel's numeric value. All channel specific commands will refer to the selected channel.

Selects a channel using the channel's identifier string. All subsequent channel specific commands will refer to the selected channel.

Enables/Disables the selected channel output. If T

0

is selected all output is affected. Enabling T

0

is the same as pressing the RUN button.

8-37

9520 System PULSe[0] Commands (SCPI Command Summary)

Keyword

:PULSe[0]

:COUNter

:STATe

:CLear

:COUNts

:STATe

:PERiod

:MODe

:BCOunter

:PCOunter

:OCOunter

:ICLock

:OCLock

:GATe

:MODe

:LOGic

:LEVel

:TRIGger

:MODe

:LOGic

:LEVel

Parameter Comments

Subsystem. Contains commands to control the output pulse generation. Commands without suffix refer to the currently selected logical instrument. See INSTrument subsystem.

Subsystem. Contains commands to define the Counter function.

0/1 or ON/OFF

Enables/Disables the counter function.

TCNTS/GCNTS Clears the designated counter. Standard units only have the Trigger counter.

TCNTS/GCNTS Queries the number of counts for the specified input. Standard units only have the

Trigger counter.

0/1 or ON/OFF

Enables / Disables the output for all channels. Command is the same as pressing the RUN/STOP button.

100ns-5000s

NORMal /

SINGle / BURSt /

DCYCle

1-9,999,999

Sets the T

0

period.

Sets the T

0

mode.

1-9,999,999

Burst Counter. Number of pulses to generate in the Burst mode.

Pulse Counter. Number of pulses to

1-9,999,999

SYS / EXT10 /

EXT20 / EXT25 /

EXT40 / EXT50 /

EXT80 / EXT100

T0 / 10 / 11 / 12 /

14 / 16 / 20 / 25 /

33 / 50 / 100 generate during on cycle of the Duty Cycle mode.

Off Counter. Number of pulses to inhibit output during the off cycle of the Duty Cycle mode.

Sets Source for the internal rate generator.

System Clock or External Source ranging from 10MHz to 100MHz.

Sets external clock output. T

0

Pulse or 50% duty cycle TTL output from 10MHz to

100MHz.

Subsystem. Contains the commands to define the Gate function.

Sets Global Gate Mode. Disable, pulse inhibit, output inhibit, channel.

DISabled /

PULSe / OUTPut

/ CHANnel

LOW / HIGH Sets Channel Gate logic level. Active low or active high.

.20V - 15V Sets the gate threshold. Value is in volts with a range of .20 to 15 Volts.

Subsystem. Contains the commands to define the Trigger function.

Sets Trigger Mode. Disable or TRIG (enable). DISabled /

TRIGgered

RISing / FALLing Selects which edge (rising or falling) to use as the trigger signal.

.20V - 15V Sets the Trigger Threshold. Value is in volts, with a range of .20 to 15 Volts.

8-38

9520 Channel PULSe[n] Commands (SCPI Command Summary)

Keyword

:PULSe [1 / 2 / n]

:STATe

:WIDTh

:DELay

:SYNC

:MUX

:POLarity

:OUTPut

:MODe

:AMPLitude

:CMODe

:BCOunter

Parameter

0/1 or ON/OFF

10ns -

999.99999999975s

-999.99999999975s -

999.99999999975s

TO, CHA, CHB,

CHC, CHD, etc.

0-255

NORMal /

COMPlement /

INVerted

TTL/

ADJustable/

2.0V to 20V

NORMal /

SINGle /

BURSt /

DCYCle

1-9,999,999

Comments

Subsystem. Contains commands to control the output pulse generation. Valid suffix range depends on the number of channels

(ChA = 1, ChB = 2, etc). Command without suffix refers to the currently selected logical instrument. See INSTrument subsystem.

Enables/Disables the output pulse for selected channel.

Sets the width or duration of the output pulse.

Sets the time from the start of the T

0

period to the first edge of the pulse.

Selects the Sync source.

Selects which timers are enabled as output for the current channel.

Sets the polarity of the pulse. For NORMal operation the second nominal state is more positive than the first. COMPlement and

INVerted are aliases. For both, the second state is more negative than the first.

Subsystem. Contains command to control output mode.

Selects output Amplitude mode: TTL/CMOS,

ADJustable.

Sets adjustable output level.

Channel Mode. Sets the channel pulse series output mode.

:PCOunter

:OCOunter

:WCOunter

:CGATe

:CLOGic

1-9,999,999

1-9,999,999

0-9,999,999

DIS / PULS /

OUTP

LOW / HIGH

Burst Counter. Sets the number of pulses to generate when channel is in the BURST mode.

Pulse Counter. Sets the number of pulses to generate during the on cycle of the Duty

Cycle Mode.

Off Counter. Number of pulses to inhibit output during the off cycle of the Duty Cycle mode.

Sets the number of T

0

pulses to delay until enabling output.

Sets Channel Gate Mode. Disable, pulse inhibit, output inhibit. (Global Gate Mode must be set to CHAN for this command to be available).

Sets Channel Gate Logic level. Active low or active high. (Global Gate Mode must be set to CHAN for this command to be available.

8-39

9520 SYSTem Commands (SCPI Command Summary)

Keyword

:SYSTem

:STATe?

:BEEPer

:STATe

:VOLume

:COMMunicate

:GPIB

:ADDRess

:SERial

:BAUD

:USB

:ECHo

:KLOCk

:AUTorun

:VERSion?

:SERN?

:INFOrmation?

:NSID?

:CAPS

Parameter Comments

0 - 100 and/or generating pulses or "0" if the machine has been disarmed.

Subsystem. Controls the audible beeper.

0/1 or ON/OFF Enables/disables the beeper.

Range is 0 to 100, Sets the volume of the beeper where 0 is off and 100 is maximum volume.

Query only. Returns the state of the machine: returns "1" if the machine is armed

1-15

Subsystem. Controls the RS232 and GPIB interfaces.

Subsystem. Controls the physical configuration of the GPIB port.

Sets the GPIB of the instrument.

Subsystem. Controls the physical configuration of the RS232 port.

4800 / 9600 /

19200 / 38400 /

57600 / 115200 /

4800 / 9600 /

19200 / 38400

0/1 or ON/OFF

Sets the baud rate for both receiving and transmitting using the DB9 RS232 port.

Sets the baud rate for communication when using mapped comports for USB communication. Default value is 38400.

Enables/Disables transmission of characters received on the DB9 serial port.

0/1 or ON/OFF Locks the keypad.

0/1 or ON/OFF

After power-up, unit will start generating pulses .automatically

Query only. Returns SCPI version number in the form: YYYY.V ex. 1999.0

Query only. Returns the serial number.

0/1 or ON/OFF

Query only. Returns model, serial number, firmware version, and FPGA version numbers.

Query only. Returns firmware and FPGA identification numbers.

Forces unit to recognize commands only sent in capital letters. 1 turns on the feature -

0 disables the feature.

8-40

9520 DISPlay Commands (SCPI Command Summary)

Keyword Parameter

:DISPlay

:MODe

:UPDate?

:BRIGhtness

:ENABle

0/1 or ON/OFF

0-4

0/1 or ON/OFF

Comments

Subsystem. Contains commands to control the display.

Enables/Disables automatic display update.

When true, front panel display is updated with serial command parameter changes.

Setting to false decreases response time.

Query only. Forces update of display. Use when mode is false.

Controls intensity of display. Range is 0 to 4, where 0 is off and 4 is full intensity.

Enables/Disables the display and front panel lights. When Disabled the keylock is enabled to prevent parameter changes from the front panel.

IEEE 488.2 Common Commands

Keyword Parameter

*IDN?

*RCL

*RST

*SAV

*TRG

*LBL

*ARM

Identification

Query

0-12

Reset Command

1-12

Trigger

Setup Label

Channel Trigger

Reset

Comments

Queries the Pulse Generator Identification.

The ID will be in the following format: manufacturer,model#,serial#,version#

Restores the state of the Pulse Generator from a copy stored in local nonvolatile memory (0 through 12 are valid memory blocks).

Resets the Pulse Generator to the default state.

Stores the current state of the Pulse

Generator in local nonvolatile memory (1 through 12 are valid memory blocks).

Generates a software trigger pulse.

Operation is the same as receiving an external trigger pulse.

Query Form returns the label of the last saved or recalled configuration.

Command Form sets the label string for the next "*SAV" command. String must be in double quotes, 14 characters max.

Resets channel triggers when channels are set to single shot or burst mode. Functions like pressing the function then run/stop button.

8-41

Appendix A - 9520 Specifications

INTERNAL RATE GENERATOR

RATE (T

0

period)

RESOLUTION

ACCURACY

JITTER

SETTLING

BURST MODE

TIMEBASE

OSCILLATOR

SYSTEM OUTPUT MODES

PULSE CONTROL MODES

0.0002 Hz to 20.000 MHz

10ns

Same as timebase

< 50 ps

1 period

1 to 9,999,999 pulses

100 MHz, low jitter PLL

50 MHz, 25ppm

Single pulse, burst, duty cycle, external gate/trigger

Internal rate generator, external trigger/gate

PROGRAMMABLE TIMING GENERATOR

CHANNEL OUTPUT MODES Single shot, burst, duty cycle, normal

CONTROL MODES Internally triggered, externally triggered

OUTPUT MULTIPLEXER and external gate. Each channel may be independently set to any of the modes.

Timing of any/all channels may be

WAIT FUNCTION

TIMEBASE

WIDTHS

RANGE multiplexed to any/all outputs.

0 to 9,999,999 pulses

Same as internal rate generator

ACCURACY

RESOLUTION

DELAYS

RANGE

ACCURACY

10 ns-999.99999999975 s

1 ns + 0.0001 x width

250 ps

0-999.99999999975 s

1 ns + 0.0001 x delay

250 ps RESOLUTION

SYSTEM EXTERNAL TRIGGER/GATE INPUT(S)

TRIGGER INPUT

FUNCTION

RATE

SLOPE

Generate individual pulses, start a burst or continuous stream

DC to 1/ (200 ns + longest active pulse).

Maximum of 5 MHz

Rising or Falling

A-1

GATE INPUT

MODE

POLARITY

MODULE SPECIFICATIONS

Pulse inhibit or output inhibit

Active high/active low

TTL/ADJUSTABLE DUAL CHANNEL OUTPUT MODULE

(Standard)

OUTPUT IMPEDANCE

TTL/CMOS MODE

50 ohm

OUTPUT LEVEL

RISE TIME

SLEW RATE

JITTER

ADJUSTABLE MODE

OUTPUT LEVEL

4.0 V typ into 1 kohm

3 ns typ (10% - 90%)

> 0.5 V/ns

50 ps RMS channel to channel

OUTPUT RESOLUTION

CURRENT

RISE TIME

SLEW RATE

OVERSHOOT

2.0 to 20 VDC into 1 k ohm

1.0 to 10.0 VDC into 50 ohm

10 mV

200 mA typical, 400 mA (short pulses)

15 ns typ @ 20V (high imp)

25 ns typ @ 10V (50 ohms)

(10% - 90%)

>0.1 V/ns

<100 mV + 10% of pulse amplitude

TRIGGER/GATE DUAL INPUT MODULE (Standard)

Standard dual channel input module, providing one trigger input and one gate input. May be used with the dual trigger firmware option to provide two independent trigger sources.

THRESHOLD

RESOLUTION

MAXIMUM INPUT VOLT.

IMPEDANCE

0.2 to 15 VDC

60 V Peak

1.2K ohm

10 mV

TRIGGER INPUT

SLOPE Rising or Falling

A-2

JITTER

INSERTION DELAY

MINIMUM PULSE WIDTH

GATE INPUT

POLARITY

FUNCTION

CHANNEL BEHAVIOR

PULSE INHIBIT DELAY

OUTPUT INHIBIT DELAY

OPTICAL OUTPUTS

WAVELENGTH

MAXIMUM SIGNAL RATE

MAXIMUM LINK DIST.

CONNECTOR TYPE

800 ps RMS

<160 ns

2 ns

Active High/Active Low

Pulse Inhibit or Output Inhibit

Global w/Individual Channel

120 ns

50 ns

GENERAL

COMMUNICATIONS

STORAGE

DIMENSIONS

WEIGHT

POWER

FUSE

820 nm or 1300 nm

5 MBd

1.5 km

ST

OPTICAL INPUTS

WAVELENGTH

MAXIMUM SIGNAL RATE

MAXIMUM LINK DIST.

CONNECTOR TYPE

INSERTION DELAY

JITTER

820 nm or 1300 nm

5 MBd

1.5 km

ST

<300 ns

<1.4 ns RMS

EXTERNAL CLOCK IN/OUT

CLOCK IN FREQUENCIES 10 MHz

– 100 MHz user selectable in discrete values

2.3 V THRESHOLD

MAX INPUT VOLTAGE 5.5 V

DUTY-CYCLE 50% (Recommended)

FREQUENCY JITTER <10%

CLOCK OUT FREQUENCIES To or Ref out (10 MHz

– 100 Mhz) user selectable in discrete values

USB/RS232

12 storage bins

10.5” x 8.25” x 5.5”

8 lbs

100 - 240 VAC

50/60 Hz, <3 A

(Qty 2) 3.15 A, 250 V Time-lag

A-3

OUTPUT MODULES

Standard

AT20

Optional

L82

L130

AT35

AT45

TZ50

TZ35

Dual channel, TTL/CMOS & Adjustable output module

Dual channel, 820 nm optical output module

Dual channel, 1300 nm optical output module

Dual channel, TTL/35 V high voltage output module

Dual channel, 45 V high and low impedance voltage output module (limited to 4 channels)

Dual channel, high current TTL/CMOS (for driving 50 ohm loads) & adjustable output module

Dual channel, high current TTL/CMOS (for driving 50 ohm loads) & 35 V high voltage output module

INPUT MODULES

Standard

IA15 Dual channel, 1 trigger / 1 gate input module

Optional

IL82

IL130

Dual channel, 820 nm optical input module

Dual channel, 1300 nm optical input module

System Options

I Incrementing (provides automatic high speed incrementing /

DT15

COM

SRM

DRM decrementing of delay and/or pulse width for each channel)

Dual Trigger Logic

– provides additional trigger via gate input

Extended Communications

Single Rackmount

– Adds Ethernet & GPIB

Dual Rackmount

*Other custom modules (LED drivers, higher voltages, current sources, and alternative input circuits) available, call with your request.

A-4

Appendix B - Safety Symbols

Safety Marking Symbols

Technical specifications including electrical ratings and weight are included within the manual. See the Table of Contents to locate the specifications and other product information. The following classifications are standard across all QC products:

Indoor use only

Ordinary Protection: This product is NOT protected against the harmful ingress of moisture.

Class 1 Equipment (grounded type)

Main supply voltage fluctuations are not to exceed +/-10% of the nominal supply voltage.

Pollution Degree 2

Installation (overvoltage) Category II for transient overvoltages

Maximum Relative Humidity: <80% RH, non-condensing

Operating temperature range of 0 o

C to 40 o

C

Storage and transportation temperature of -40 o

C to 70 o

C

Maximum altitude: 3000 m (9843 ft.)

This equipment is suitable for continuous operation.

This section provides a description of the safety marking symbols that appear on the instrument. These symbols provide information about potentially danger-ous situations which can result in death, injury, or damage to the instrument and other components.

Symbols Publications; Descriptions & Comments

IEC 417, No. 5031

Direct current - VDC may be used on rating labels.

IEC 417, No. 5032

Alternating current - For rating labels, the symbol is typically replaced by V and Hz as in 230V, 50Hz. DO NOT USE VAC.

IEC 417, No. 5033

Both direct and alternating current.

IEC 617-2 No. 02-02-06

Three-phase alternating current .

B-5

I

IEC 417, No. 5017

Earth (ground) terminal - Primarily used for functional earth terminals which are generally associated with test and measurement circuits. These terminals are not for safety earthing purposes but provide an earth reference point.

IEC 417, No. 5019

Protective conductor terminal - This symbol is specifically reserved for the protective conductor terminal and no other. It is placed at the equipment earthing point and is mandatory for all grounded equipment .

IEC 417, No. 5020

Frame or chassis terminal - Used for points other than protective conductor and functional earth terminals where there is a connection to accessible conductive terminals to advise the user of a chassis connection.

IEC 417, No. 5021

Equipotentiality - Used in applications where it is important to indicate to the operator that two or more accessible functional earth terminals or points are equipotential - More for functional rather than for safety purposes.

IEC 417, No. 5007

On (Supply) - Note that this symbol is a bar, normally applied in the vertical orientation. It is not the number 1.

IEC 417, No. 5008

Off (Supply) - Note that this symbol is a true circle. It is not the number 0 or the letter O.

IEC 417, No. 5172

Equipment protected by double insulation or reinforced insulation

(equivalent to Class II if IEC 60536).

B-6

ISO 3864, No. B.3.6

Background colour - yellow; symbol and outline

– black

Caution, risk of electric shock - Generally used only for voltages in excess of 1000 V. It is permissible to use it to indicate lower voltages if an explanation is provided in the manual. Colour requirements do not apply to markings on equipment if the symbol is moulded or engraved to a depth or raised height of 0.5 mm, or that the symbol and outline are contrasting in colour with the background.

IEC 417, No. 5041

Background colour - yellow; symbol and outline

– black

Caution, hot surface - Colour requirements do not apply to markings on equipment if the symbol is moulded or engraved to a depth or raised height of 0.5 mm, or that the symbol and outline are contrasting in colour with the background.

ISO 3864, No. B.3.1

Background colour - yellow; symbol and outline

– black

Caution - (refer to accompanying documents) used to direct the user to the instruction manual where it is necessary to follow certain specified instructions where safety is involved. Colour requirements do not apply to markings on equipment if the symbol is moulded or engraved to a depth or raised height of 0.5 mm, or that the symbol and outline are contrasting in colour with the background.

IEC 417, No. 5268-a

In-position of bistable push control.

IEC 417, No. 5269-a

Out-position of bistable push control.

B-7

Appendix C - COM Option

COM Overview

The COM Option for the 9520 includes a GPIB and an Ethernet interface in addition to the RS232 and USB interfaces included with the standard product.

The command set is the same for the RS-232, USB, GPIB, and Ethernet.

Different interfaces may be used at the same time. Responses will be made to the most recently used interface.

GPIB Interface

Also known as IEEE-488, a GPIB computer interface is optional on the 9520.

Before using this interface, the address must be set using the GPIB address menu item.

Ethernet Interface

The Ethernet module used is a “Digi Connect ME” module supplied by Digi

Connectware, Inc. There are several ways to successfully communicate with the pulse generator over Ethernet. The two most popular methods are raw TCP/IP

(such as Labview or programming with VISA libraries) and by mapping a PC

COM port using the Digi Connectware’s “Realport Drivers”.

IP Address and Raw TCP/IP Connection

This document describes one of the most popular methods of setting up Ethernet communication for the Quantum Composers pulse generators. The method discussed is Raw TCP/IP communication.

The Ethernet module used in Quantum Composers pulse generators is a “Digi

Connect ME” device manufactured by Digi International, Inc. It supports virtually all practical Ethernet communication methods. A set of utilities and documentation by Digi is included on the CD shipped with the pulse generator.

This discussion assumes that the Digi utilities included with your pulse generator and National Instruments VISA (version 3.3 in this procedure, see National

Instruments’ website) are installed. The procedures discussed have been prepared using Windows XP service pack 2.

Determining IP Address

The Digi module has been reset to factory defaults before it left the manufactur-ing facility. In this mode, it is ready to be assigned an IP address by the local DHCP server. If a crossover cable is being used, the Ethernet device will as-sume a default IP address.

The Digi utility “Digi Device Discovery” can be used to determine the IP address that is currently assigned to the Ethernet module. Hit “Start, All Programs, Digi

C-8

Connect, Digi Device Discovery”. When the utility opens, it scans the LAN looking for Digi Ethernet modules. It may take a minute after plugging in or powering the Ethernet module before the LAN negotiates the connection with the

Digi module. Hit “Refresh View” in the left column after a minute or so if the utility fails to see the unit when you start it. In some situations it is possible that the

Windows Firewall will block the Digi Device Discovery from being able to see the unit. It is advisable to turn the Windows Firewall off while performing these tasks.

When the utility sees the Digi device, it will display it in the list (Figure 1).

Figure 1: Digi Device Discovery utility displaying Digi module discovered on the

LAN.

From this point, a web interface can be opened, allowing access to configura-tion options for the Digi module. If you are required to enter a username and password, they are as follows:

Username: “root”

Password: “dbps”

If a static IP address is desired, this change can be made from the web interface.

Please note, however, that if the IP address is changed such that it is incompatible with the LAN, all communication includ ing access to the module’s settings (including the IP address!) will no longer be possible over the LAN. If this happens, a crossover cable must be used to access the Digi module’s settings

(again using Digi Device Discovery).

Temporarily set the pc’s IP ad-dress to be compatible with the Digi module’s IP address to get the pc and pulse generator to

‘see’ each other over the crossover cable.

C-9

Appendix D - DT15 (Dual Trigger Option)

DT15 Overview

This module option allows the “Gate” input to function as a second trigger input.

For consistency, the enabling menu for this option is located under the “Trig” menu structure. Once the dual trigger mode is enabled, both the “Gate” and

“Trig” inputs can act as trigger inputs.

Adjustments for the “Gate” trigger input are located under the “Gate” menu structure. The voltage threshold level and trigger edge for the “Gate” input can be adjusted from this menu. The “Gate” trigger edge choice is only available when in dual trigger mode.

Once dual trigger functionality is enabled on the unit, each channel can be as-signed to either of the trigger inputs. The default trigger source for each channel is the “Trig” input. The trigger source selection is accessed in the secondary channel menus. To access this menu, first press the yello w “Func” button fol­lowed by the channel of interest. Continue to press “Func” then the channel button until the menu page with “Ch Gate:” and “TrigSrc:” appears. Use the “Next” button to place the cursor on the “TrigSrc” line and use the up/down arrows to change to the desired trigger source.

DT15 Menus

Trigger 1 Menu (TRIG key)

Mode: Disabled Mode:

Triggered

Mode: Dual Trig

*

Threshold Level Threshold

Level

Trigger Edge Trigger Edge

Enabling System Trigger

Enable the use of the TRIG input by the system timer as a trigger source.

Mode: Selects between disabling/enabling the trigger mode(s).

Level: Sets the trigger threshold. Edge: Selects between rising and falling edges as the trigger source when a trigger mode is enabled.

Trigger 2 Menu (GATE key) *

Mode: Disabled Mode:

Triggered

Mode: Dual Trig

D-10

Threshold Level Threshold

Level

Trigger Edge Trigger Edge

*Only enabled when TRIG

mode is set to “Dual Trig”. Functions as

Standard

Gate Menu wh

en not in “Dual Trig” mode.

DT15 SCPI Command Summary

Keyword

:PULSe [0]

:TRIGger

:MODe

:PULSe[1/2/n]

:CTRIGger

Parameter

DUAL

GATE/TRIG

Comments

Subsystem. Contains commands to control the output pulse generation. Commands without suffix refer to the currently selected logical instrument.

See INSTrument subsystem.

Subsystem. Contains commands to define the

Trigger function.

Sets the unit into dual trigger mode.

Valid suffix range depends on the number of channels (ChA = 1, ChB = 2, etc).

Sets which input is assigned to the channel trigger.

DT15 IEEE 488.2 Command Summary

Keyword

*TTG

*GTG

Command Name Comments

Trigger

Generates a software trigger pulse for the TRIG

Trigger Input input only. Operation is the same as receiving an external trigger pulse on the Trigger input.

Trigger

– Gate

Generates a software trigger pulse for the GATE

Input input only. Operation is the same as receiving an external trigger pulse on the Gate input.

D-11

Appendix E - AT35 (35V Output Option)

AT35 Description

When the Adjustable Mode is enabled for this module, the outputs will provide an adjustable output from 5 volts to 35 volts. The pulse width can be set over the standard range of the unit, but the 35 volt output will self limit to approximately 4

µs with some droop. There is no change to TTL output mode functionality with this module.

To maintain the highest possible rise time, care must be taken with cabling and termination. Low capacitance cable and 50 ohm termination will provide the fastest rise times without overshoot. Faster rise times can be achieved by increasing the termination resistance, but some overshoot is likely to occur.

While the 35 volt output provides a fast, controlled rising edge, the pulse width and falling edge are not tightly controlled. Also, when using the 35V mode, the option will only function if the ‘Polarity’ is set for “Active High”.

AT35 Specifications

Through a 50 ohm load at 200 Hz

Output

Setpoint Resolution

Rise Time

5 V

– 35 V

10 mV

< 30 ns

Accuracy 500 mV

Max. Frequency (Internal & External) 4000 Hz

E-12

Appendix F - TZ50 (Impedance Matching

Output Module)

TZ50 Overview

This module option allows a user to have a 50 Ohm load on the output while maintaining output amplitude of at least 4 Volts while in the TTL/CMOS mode. All other functionality of the module is the same as the AT20 modules, including output while using the Adjustable Mode Function of the channels.

TZ50 Specifications

TTL/CMOS Mode

Output Level

Rise Time

Slew Rate

Jitter - Channel to Channel

4.0 V typ into 50 Ohms

3 ns

0.5 V/ns

50 ps RMS

Adjustable Mode

Output Resolution

Current

Slew Rate

10 mV

100 mA typ, 400 mA max (short pulses)

0.1 V/ns

*Note: The TZ50 module has significant overshoot and ringing through high impedance (see figure below).

F-13

Appendix G - Increment Modes Option

Increment Overview

The System Increment modes are a pair of special modes, which allow the delay and width of each channel to be incremented at the end of a burst of pulses.

Each channel is independent and each may be set with different initial values and different values for the step size for both the delay and the pulse width.

There are two incrementing modes, Increment and DC Increment. In the

Incre-ment mode, each start command or external trigger produces a burst of pulses. At the end of the burst the appropriate delays and pulse widths are incremented and the instrument is armed for the next start command. In the DC

Increment (Duty Cycle) mode the output is starting as with the normal duty cycle mode. At the end of each cycle the delays and pulse widths are incremented.

This contin-ues for the number of cycles defined by the Cycles parameter. The modes are selected from the system mode menu. The step sizes are specified in the channel menus.

Increment Menus

System Mode Menu 1 Extensions (FUNC + MODE key)

MODE:

BurstIncr

#/Burst

MODE:

DCIncrement

#/On

#/Off

System Mode Menu 2 Extensions (FUNC + MODE key)

MODE:

BurstIncr

T

0

Period

MODE:

DCIncrement

Cycles

T

0

Period

Setting System Increment Modes of Operation

The MODE menu sets the T

0

system timer mode. The menu will show the extra set parameters (Burst, On & Off) only when they are appropriate.

Mode:

Burst:

On:

Selects the T

0

mode: Continuous, Single Shot, Burst or Duty

Cycle mode.

Sets the number of pulses to be generated when in Burst mode.

Sets the number of pulses to be generated during each on cycle.

G-14

Off:

Cycles:

To:

Sets the number of pulses to skip each during off cycle when in the Duty Cycle mode.

Sets the number o DC Increment Cycles to generate before completion.

Sets the T

0

period which determines the fundamental output frequency of the unit.

Advanced Channel Menu Increment Extension (FUNC + channel key)

Channel Enable

Increment Wid *

Increment Dly *

Setting the Channel Increment Parameters

To define a pulse requires two parameters: the delay to the active edge and the width of the pulse.

IncW:

IncD:

Sets the channel pulse width.

Sets the channel delay until the active edge.

Increment SCPI Command Summary

Keyword

:PULSe[0]

:MODe

:CYCLe

:IRESet

:PULSe[1/2/n]

:IWIDth

:lDELay

Parameter

BINCRement /

<numeric value>

1

Comments

Subsystem. Contains commands to control the output pulse generation. Commands without suffix refer to the currently selected logical instrument. See INSTrument subsystem.

Sets the T

0

mode. Added parameters for Burst

Increment and Duty Cycle Increment mode.

Sets the number of cycles to generate in Duty

Cycle Increment mode.

Resets the width and delay increment parameters on all channels.

Subsystem. Contains commands to control the output pulse generation. Valid suffix range depends on the number of channels (ChA-1,

ChB-2, etc . . .). Command without suffix refers to the currently selected logical instrument. See

INSTrument

’s subsystem.

<numeric value> Sets the pulse width increment step size.

<numeric value> Sets the delay increment step size.

Increment Initialization and Reset (FUNC + CLR)

Pressing the FUNC key then CLR initializes the increment parameters and resets the delays and pulse widths to their initial conditions. This must be done after setting all the step parameters, but before generating any pulses.

Increment Specifications

Width Step Size

Width Minimum Step

-1.00 s to 1.00 s

10 ns (-10 ns)

G-15

Width Step Resolution

Width Incremented Range

250 ps

10,000 s

Delay Step Size

Delay Minimum Step

Delay Step Resolution

Delay Incremented Range

-100 ms to 100 ms

10 ns (-10ns)

250 ps

10,000 s

*Note: Any increment value between -10 ns and 10 ns will disable the increment function for that parameter.

Update Rate 10 µs + 30 µs per active channel (1 Ch

@ 25 kHz to 8 Ch @ 4 kHz)

G-16

Appendix H - AT45 Option (45 Volt Output)

AT45 Option Overview

For channels with AT45 output option, the maximum frequency is limited to 100 kHz. The pulse width can be set over the standard range of the unit with both active high and low outputs when set to high impedance mode. In low impedance mode, the pulse width is limited to a maximum of 10s and the active high output is no longer allowed. To maintain the highest possible rise time, care must be taken with cabling and termination. Low capacitance cable and 50 ohm termination will provide the fastest rise times without overshoot. The channel menu structure for the AT45 module changes are described in the table below

(changes from standard outputs are in bold italics).

AT45 Protection Error Messages

When an AT45 module is present, the system performs self-checks to insure the module is not damaged when attempting to over-drive, however the overdrive protection is NOT SHORT CIRCUIT PROTECTED, and caution must be taken to prevent damage to the board when driving into a short!

Module Errors

If a channel on any AT45 module is over-driven, the channel will disable itself and the system will indicate an error on the module. The error will not clear until the user presses FUNC - PERIOD key sequence to clear the error, or power cycles the instrument. The overdriving protection is limited to low resistance errors not to direct shorts to ground. Damage to the output will occur when driving into short circuits. Module errors may occur due to any of the following:

- Over current.

- Over temperature.

- Internal hardware.

System Limit Error

The system will not allow the Lo Impedance enabled AT45 channels to exceed

150V total amplit ude. If this situation occurs, the “Over-Driving Unit” error is displayed and the currently adjusting amplitude is reduced to the 150V enabled channel limit.

Voltage Change Timing

The channels adjustable voltage changes very quickly when adjusting from a lower voltage to a higher voltage but changes slowly when changing from a higher voltage to a lower voltage. It takes approximately 30s to change from 45V to 4.0V so caution must be taken when adjusting the voltage to a lower voltage tolerant circuit.

H-17

AT45 Channel Menus

Channel Output Configuration Menu

Channel Enable Channel Enable

Output Type:

High Z

Output Type:

Low Z

Polarity

Output Level Output Level

AT45 SCPI Command Extension Summary

Keyword

:PULSe [1 / 2 / n]

:OUTPut

:MODe

:AMP

:MERRor

Parameter

HIZ / LOZ

4 V

– 45 V

1

Comments

Subsystem. Contains commands to control the output pulse generation. Valid suffix range depends on the number of channels (ChA = 1,

ChB = 2, etc.). Command without suffix refers to the currently selected logical instrument. See

INSTrument subsystem.

Subsystem. Contains command to control output mode.

Selects output Amplitude mode: High

Impedance or Low Impedance

Sets adjustable output level.

Command clears the last module error to allow the unit to generate pulses again. Query returns the last displayed error.

AT45 Specifications

Amplitude

Resolution

Accuracy

Rise Time

Fall Time

Frequency (Internal & External)

Overshoot

Polarity - High Z (>10k)

Polarity - Low Z (50 Ohms)

Pulse Width - High Z (>10k)

Pulse Width - Low Z (50 Ohms)

Current (maximum)

4 V

– 45 V

20 mV

+/-1.5%

< 2 ns Typical 10%-90% (Low Z)

< 9 ns Typical 10%-90% (High Z)

< 9 ns Typical 90%-10% (Low Z)

< 7 ns Typical 90%-10% (High Z)

DC

– 100 kHz

<35% Typical Allowed for Fast

Rise Time

Active High or Active Low

Active High Only

10 ns to DC

10 ns to 10 s

35 mA (High Z @10 ms width)

900 mA (Low Z @ 10 ms width)

H-18

Figure 1 (High Z 5V)

Figure 2 (High Z 45V)

H-19

Figure 3 (Low Z 5V)

Figure 4 (Low Z 45V)

H-20

Rise time and overshoot are tuned for best response at low impedance (low Z)

Figure 5 (Rise Time versus Output Voltage)

Figure 6 (Fall Time versus Output Voltage)

H-21

Figure 7 (Overshoot Versus Output Voltage)

H-22