True RMS
AUTORANGING MULTIMETER
AMM-1032
User’s Manual
http://www.tmatlantic.com
SAFETY INFORMATION
The following safety information must be observed to insure maximum personal safety during the operation at this meter:
Do not use the meter if the meter or test leads look damaged, or if you suspect that the meter is not operating properly.
Never ground yourself when taking electrical measurements. Do not touch exposed metal pipes, outlets, fixtures, etc., which might be at ground potential. Keep your body isolated from ground by using dry clothing, rubber shoes, rubber mats, or any approved insulating material.
Turn off power to the circuit under test before cutting, unsoldering, or breaking the circuit.
Small amounts of current can be dangerous.
Use caution when working above 60V dc or 30V ac rms. such voltages pose a shock hazard.
When using the probes, keep your fingers behind the finger guards on the probes.
Measuring voltage which exceeds the limits of the multimeter may damage the meter and expose the operator to a shock hazard. Always recognize the meter voltage limits as stated on the front of the meter.
Never apply voltage or current to the meter that exceeds the specified maximum:
Function
V DC or V AC
A DC/AC
Input Limits
Maximum Input
1000VDC/VAC
10A DC/AC (30 seconds max every
15 minutes)
600V DC/AC Frequency,
Resistance,
Capacitance, Cycle,
Diode test,
Continuity
Temperature 600V DC/AC
SAFETY SYMBOLS
This symbol adjacent to another symbol, terminal or operating device indicates that the operator must refer to an explanation in the Operating
Instructions to avoid personal injury or damage to the meter.
WARNING This WARNING symbol indicates a potentially hazardous situation, which if not avoided, could result in death or serious injury.
CAUTION
MAX
This CAUTION symbol indicates a potentially hazardous situation, which if not avoided, may result damage to the product.
This symbol advises the user that the terminal(s) so marked must not be connected to a circuit point at which the voltage with respect to earth ground exceeds (in this case) 500 VAC
or VDC. SYMBOLS AND ANNUNCIATORS
Continuity
This symbol adjacent to one or more terminals identifies them as being associated with ranges that may, in
BAT Low Battery
Diode normal use, be subjected to particularly hazardous voltages. For
HOLD Data Hold maximum safety, the meter and its test leads should not be handled when these terminals are energized.
AUTO
AC
DC
AutoRanging
Alternating Current or Voltage
Direct Current or Voltage
1.
CONTROLS AND JACKS
2.
3.
1
7
8
10
11
SPECIFICATIONS
The instrument complies with: EN61010-1.
Insulation: Class2, Double insulation.
Overvoltage category: CATIII 600V, CATII 1000V.
Display: 6000 counts LCD display with function indication.
Polarity: Automatic, (-) negative polarity indication.
Overrange: “OL” mark indication.
AC Response: True RMS (50 Hz or 60 Hz)
6
3
2
4
MAX
5
9
Ω
10A
MAX every
15 min.
COM
V
F
C
True RMS: The term stands for
“Root-Mean-Square,” which represents the method of calculation of the voltage or current value. Average responding multimeters are calibrated to read correctly only on sine waves and they will read inaccurately on non-sine wave or distorted signals.
True rms meters read accurately on either type of signal.
6000 count Liquid Crystal Display with symbolic signs
Function switch
Low battery indication: The “ ” is displayed when the battery voltage drops below the operating level.
Measurement rate: 2 times per second, nominal.
10A (positive) input jack for 10A DC or AC Auto power off: Meter automatically shuts down measurements after approx. 15 minutes of inactivity.
4.
5.
6.
7.
8.
9.
10.
11.
COM (negative) input jack
Positive input jack
HOLD & Backlight pushbutton
Max/Min pushbutton
Range/ .Diode pushbutton
HZ/% pushbutton
AC Voltage Detector Sensor
AC Voltage Detector indicator light
Operating environment: 0 oC to 50 oC (32 oF to 122 oF) at < 70 % relative humidity.
Storage temperature: -20 oC to 60 oC (-4 oF to 140 oF) at < 80 % relative humidity.
For inside use, max height: 2000m
Pollution degree: 2
Power: One 9V battery , NEDA 1604, IEC 6F22.
Dimensions: 150 (H) x 70 (W) x 48 (D) mm
Weight: Approx.:255g.
Accuracy is given at 18 oC to 28 oC (65 oF to 83 oF), less than 70 % RH
DC Voltage
600.0mV 0.1mV +0.5% of rdg + 2 dgts
6.000V 1mV
60.00V 10mV
600.0V 100mV
+1.2% of rdg + 2 dgts
1000V 1V +1.5% of rdg + 2 dgts
Input Impedance: 7.8M
.
Maximum Input: 1000V dc or 1000V ac rms.
AC Voltage
Range Resolution Accuracy
6.000V 1mV +1.5% of rdg + 10 dgts
60.00V 10mV
600.0V 100mV
+1.5% of rdg + 10 dgts
1000V 1V +2.0% of rdg + 10 dgts
Input Impedance: 7.8M
.
Frequency Range:50 to 60Hz
Maximum Input: 1000V dc or 1000V ac rms.
DC Current
6A 1mA +2.5% of rdg + 5 dgts
10A 10mA
Overload Protection: 10A / 250V Fuse.
Maximum Input: 10A dc or ac rms on 10A DC range.
AC Current
6A 1mA +3.0% of rdg + 5 dgts
10A 10mA
Overload Protection: 10A / 250V Fuse.
Frequency Range: 50 to 60 Hz
Maximum Input: 10A dc or ac rms on 10A AC range.
Resistance
600.0
0.1
6.000k
1
60.00k
10
600.0k
100
6.000M
1k
60.00M
10k
+1.2% of rdg + 4 dgts
+1.0% of rdg + 2 dgts
+1.2% of rdg + 2 dgts
+2.0% of rdg + 2 dgts
+5.0% of rdg + 10 dgts
Maximum Input: 600V dc or 600V ac rms.
Capacitance (Auto-ranging)
Range Resolution Accuracy
40.00nF 10pF
400.0nF 0.1nF
4.000uF 1nF
40.00uF 10nF
+5.0% of rdg + 50 dgts
+3.0% of rdg + 5 dgts
400.0uF 0.1uF
4000uF 1uF
+5.0% of rdg + 5 dgts
+5.0% of rdg + 5 dgts
Maximum Input:
6
00V dc or
6
00V ac rms.
Frequency (Auto-ranging)
9.999Hz 0.001Hz
99.99Hz 0.01Hz +1.5% of rdg + 5 dgts
999.9Hz 0.1Hz
9.999kHz 1Hz
99.99kHz 10Hz +1.2% of rdg + 3 dgts
999.9kHz 100Hz
10MHz 1kHz +1.5% of rdg + 4 dgts
Sensitivity: >0.5V RMS while ≤ 1MHz ;
Sensitivity: >3V RMS while >1MHz ;
Maximum Input: 600V dc or 600V ac rms.
Temperature
Range
-20oC~+760oC
Resolution Accuracy
1 oC
+3% of rdg
-4 oF~+1400 oF 1oF
+5 oC/9 oF
Sensor: Type K Thermocouple
Overload protection: 600V dc or ac rms..
Diode Test
Test current Resolution Accuracy
0.3mA typical
1 mV +10% of rdg + 5 dgts
Open circuit voltage: 1.5V dc typical
Overload protection: 600V dc or ac rms.
Audible continuity
Audible threshold: Less than 100 Test current:
<0.3mA
Overload protection: 600V dc or ac rms.
OPERATION
WARNING: Risk of electrocution. High-voltage circuits, both AC and DC, are very dangerous and should be measured with great care.
1. ALWAYS turn the function switch to the OFF position when the meter is not in use. This meter
has Auto OFF that automatically shuts the meter
2.
OFF if 15 minutes elapse between uses.
If “OL” appears in the display during a measurement, the value exceeds the range you have selected. Change to a higher range.
NOTE: On some low AC and DC voltage ranges, with the test leads not connected to a device, the display may show a random, changing reading. This is normal and is caused by the high-input sensitivity. The reading will stabilize and give a proper measurement when connected to a circuit.
AUTORANGING/MANUAL RANGE SELECTION
When the meter is first turned on, it automatically goes into Autoranging. This automatically selects the best range for the measurements being made and is generally the best mode for most measurements. For measurement situations requiring that a range be manually selected, perform the following:
1. Press the RANGE key. The “AUTO” display indicator will turn off or select diode /beeper, AC/DC Clamp
2. Press the RANGE key to step through the available ranges until you select the range you want.
3. To exit the Manual Ranging mode and return to
Autoranging, press and hold the RANGE key for 2 seconds.
Note: Manual ranging does not apply for the Frequency functions.
“min” occurs.
3. To exit MAX/MIN mode press and hold the
MAX/MIN key for 2 seconds
DISPLAY BACKLIGHT
Press and hold the HOLD key for >1 second to turn on or off the display backlight function. The backlight will automatically turn off after 10 seconds.
HOLD
The hold function freezes the reading in the display. Press the HOLD key momentarily to activate or to exit the
HOLD function.
NON-CONTACT VOLTAGE (NCV)
The NCV function works on any rotary switch position.
1.Test the detector on a known live circuit before use.
2.Hold the top of the meter very close to the voltage source as shown.
3.If voltage is present, the rim of the LCD display will flash a bright red and a motor shaking will happen.
Hz/duty
1. Switch to Hz range.
2. Press the Hz button to show the reading in the display and the "Hz" indicator will appear on the display.
DC VOLTAGE MEASUREMENTS
CAUTION: Do not measure DC voltages if a motor on the circuit is being switched ON or OFF. Large voltage surges may occur that can damage the meter.
MAX/MIN
Note: When using the MAX/MIN function in
Autoranging mode, the meter will “lock” into the range that is displayed on the LCD when
MAX/MIN is activated. If a MAX/Min reading exceeds that range, an “OL” will be displayed.
Select the desired range BEFORE entering
MAX/MIN mode.
1. Press the MAX/MIN key to activate the MAX/MIN recording mode. The display icon "MAX" will appear.
The meter will display and hold the maximum reading and will update only when a new “max” occurs.
2. Press the MAX/MIN key again and the display icon
"MIN" will appear. The meter will display and hold the minimum reading and will update only when a new
1. Set the function switch to the V DC position .
2. Insert the black test lead banana plug into the negative
(COM) jack and the red test lead banana plug into the positive (V) jack.
3. Touch the test probe tips to the circuit under test. Be sure
4. to observe the correct polarity (red lead to positive, black lead to negative).
Read the voltage in the display. The display will indicate the proper decimal point and value. If the polarity is reversed, the display will show (-) minus before the value.
AC VOLTAGE MEASUREMENTS
WARNING: Risk of Electrocution. The probe tips may not be long enough to contact the live parts inside some 240V outlets for appliances because the contacts are recessed deep in the outlets. As a result, the reading may show 0 volts when the outlet actually has voltage on it. Make sure the probe tips are touching the metal contacts inside the outlet before assuming that no voltage is present.
CAUTION: Do not measure AC voltages if a motor on the circuit is being switched ON or OFF. Large voltage surges may occur that can damage the meter.
1. Set the function switch to the V AC position.
2. Insert the black test lead banana plug into the negative (COM) jack and the red test lead banana plug into the positive (V) jack.
3. Touch the test probe tips to the circuit under test.
4. Read the voltage in the display. The display will indicate the proper decimal point, value and symbol (AC, V, etc.).
DC CURRENT MEASUREMENTS
CAUTION: Do not make current measurements on the 10A scale for longer than 30 seconds. Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1. Insert the black test lead banana plug into the negative
(COM) jack.
2. For DC current measurements, set the function switch to the DC 10A position and insert the red test lead banana plug into the 10A jack.
3. Remove power from the circuit under test, then open up the circuit at the point where you wish to measure current.
4. Touch the black test probe tip to the negative side of the circuit. Touch the red test probe tip to the positive side of the circuit.
5. Apply power to the circuit.
6. Read the current in the display. The display will indicate the proper decimal point, value and symbol.
AC CURRENT MEASUREMENTS
WARNING: To avoid electric shock, do not measure
AC current on any circuit whose voltage exceeds 250V
AC.
CAUTION: Do not make current measurements on the 10A scale for longer than 30 seconds. Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1. Insert the black test lead banana plug into the negative
(COM) jack.
2. For ACcurrent measurements , set the function switch to the AC 10A position and insert the red test lead banana plug into the 10A jack.
3. Remove power from the circuit under test, then open up the circuit at the point where you wish to measure current.
4. Touch the black test probe tip to the negative side of the circuit. And touch the red test probe tip to the positive side of the circuit.
5. Apply power to the circuit.
6. Read the current in the display. The display will indicate the proper decimal point, value and symbol.
RESISTANCE MEASUREMENTS
WARNING: To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any resistance measurements. Remove the batteries and unplug the line cords.
1. Set the function switch to the position.
2. Insert the black test lead banana plug into the negative
(COM) jack and the red test lead banana plug into the positive jack.
3. Touch the test probe tips across the circuit or part under test. It is best to disconnect one side of the part under test so the rest of the circuit will not interfere with the resistance reading.
4. Read the resistance in the display. The display will indicate the proper decimal point, value and symbol.
CONTINUITY CHECK
WARNING: To avoid electric shock, never measure continuity on circuits or wires that have voltage on them.
1. Set the function switch to the position.
2. Insert the black lead banana plug into the negative (-) jack (COM) and the red test lead banana plug into the positive (+) jack ( ).
3. Press the button until the symbol appears in the display.
4. Touch the test probe tips to the circuit or wire you wish to check.
5. If the resistance is less than approximately 100 , the audible signal will sound. The display will also show the actual resistance.
DIODE TEST
WARNING: To avoid electric shock, do not test any diode that has voltage on it.
1. Set the function switch to position.
2. Press the button until the symbol appears in the display.
3. Insert the black test lead banana plug into the negative (-) jack (COM) and the red test lead banana plug into the positive (+) jack ( ).
4. Touch the test probe tips to the diode or semiconductor junction you wish to test. Note the meter reading
5. Reverse the probe polarity by switching probe position.
Note this reading.
6. The diode or junction can be evaluated as follows:
A. If one reading shows a value and the other reading shows OL, the diode is good.
B. If both readings show OL, the device is open.
C. If both readings are very small or 0, the device is shorted.
NOTE: The value indicated in the display during the diode check is the forward voltage.
FREQUENCY MEASUREMENT
1. Set the function switch to the FREQ position.
2. Insert the black test lead banana plug into the negative (-) jack (COM) and the red test lead banana plug into the positive (+) jack (F).
3. Touch the test probe tips to the circuit under test.
4. Read the frequency in the display. The digital reading will indicate the proper decimal point, symbols (Hz, kHz) and value.
CAPACITANCE MEASUREMENTS
WARNING: To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any capacitance measurements. Remove the batteries and unplug the line cords.
1. Set the function switch to the CAP position. (“nF” and a small value will appear in the display).
2. Insert the black test lead banana plug into the negative (-) jack (COM) and the red test lead banana plug into the positive (+) jack (CAP).
Touch the test leads to the capacitor to be tested. The display will indicate the proper decimal point, value and symbol
TEMPERATURE MEASUREMENTS
WARNING: To avoid electric shock, disconnect both test probes from any source of voltage before making a temperature measurement.
1. If you wish to measure temperature in o F, set the function switch to the o F range. If you wish to measure temperature in o C, set the function switch to the o C range.
2. Insert the type K thermocouple probe black test lead banana plug into the negative COM jack and the red test lead banana plug into the positive Temp jack..
3. Touch the Temperature Probe head to the part whose temperature you wish to measure. Keep the probe touching the part under test until the reading stabilizes
(about 30 seconds).
4. Read the temperature in the display. The digital reading will indicate the proper decimal point and value.
WARNING: To avoid electric shock, be sure the thermocouple has been removed before changing to another measurement function.
REPLACING THE BATTERY
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the battery door.
1. When the batteries become exhausted or drop below the operating voltage, “BAT” will appear in the right-hand side of the LCD display. The battery should be replaced.
2. Follow instructions for installing battery. See the
Battery Installation section of this manual.
3. Dispose of the old battery properly.
WARNING: To avoid electric shock, do not operate your meter until the battery door is in place and fastened securely.
2. Open the battery door by loosening the screw using a
Phillips head screwdriver.
3. Insert the battery into battery holder, observing the correct polarity.
4. Put the battery door back in place. Secure with the two screws.
WARNING: To avoid electric shock, do not operate the meter until the battery door is in place and fastened securely.
NOTE: If your meter does not work properly, check the fuses and battery to make sure that they are still good and that they are properly inserted.
REPLACING THE FUSES
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the fuse door.
1. Disconnect the test leads from the meter and any item under test.
2. Open the fuse door by loosening the screw on the door using a Phillips head screwdriver.
3. Remove the old fuse from its holder by gently pulling it out.
3. Install the new fuse into the holder.
4. Always use a fuse of the proper size and value
(10A/250V fast blow for the 10A range).
5. Put the fuse door back in place. Insert the screw and tighten it securely.
WARNING: To avoid electric shock, do not operate your meter until the fuse door is in place and fastened securely.
BATTERY INSTALLATION
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the battery door.
1. Disconnect the test leads from the meter.