GREISINGER GCO 100 Carbon Monoxide Measuring Instrument Operating Manual
Below you will find brief information for Carbon Monoxide Measuring Instrument GCO 100. The GCO 100 is a high value CO-Measuring instrument for the measuring of up to 1000ppm (0,1Vol %) CO in ambient air. Via the adjustable alarm function the instrument warns the user optically (display) and acoustically (integrated horn) when dangerous concentrations are present. You can monitor air quality, detect CO in the breath of smokers, and detect CO-toxication of victims of fires. It comes with a calibration protocol and can be connected to a computer for data transfer.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
H65.0.01.6C-07
Operating Manual
– Carbon Monoxide Measuring Instrument
as of version 1.3
GCO 100
Seite 1
GREISINGER electronic GmbH
D - 93128 Regenstauf, Hans-Sachs-Straße 26
Tel.: 09402 / 9383-0, Fax: 09402 / 9383-33, eMail: [email protected]
WEEE-Reg.-Nr. DE 93889386
H65.0.01.6C-07
Content
1
2
Operating Manual GCO 100 page 2 of 10
General advice ...................................................................................................... 2
Intended use .......................................................................................................... 3
3
4
5
6
7
Safety instructions ................................................................................................ 3
Operating and maintenance advice ..................................................................... 4
Disposal instructions ............................................................................................ 4
Operation ............................................................................................................... 4
6.1 Display elements ............................................................................................. 4
6.2 Pushbuttons .................................................................................................... 4
6.3 Connections .................................................................................................... 4
Generals of CO-measuring ................................................................................... 5
8
9
Configuration of the instrument .......................................................................... 6
Alarm functions ..................................................................................................... 7
10 Adjustment ............................................................................................................ 7
10.1 Adjustment menue ........................................................................................ 7
10.2 Automatic zero point adjusting .................................................................... 7
10.3 Adjustment with reference gas .................................................................... 8
11 System messages ................................................................................................. 9
12 The serial interface ............................................................................................... 9
13 Specification .......................................................................................................... 10
1 General advice
Read through this document attentively and make yourself familiar to the operation of the device before you use it. Keep this document in a ready-to-hand way in order to be able to look up in the case of doubt.
H65.0.01.6C-07 Operating Manual GCO 100 page 3 of 10
2 Intended use
High value CO-Measuring instrument for the measuring of up to 1000ppm (0,1Vol %) CO in ambient air.
Via the adjustable alarm function the instrument warns the user optically (display) and acoustically
(integrated horn) when dangerous concentrations are present.
The device must not used as a monitoring device for the personal safety !
The measuring inlet is at the opening of the front plate.
Applications
control and maintenance of heating systems
monitoring air quality (signalling exceeding of the maximum allowed working place concentration MAK)
detection of CO in the breath of smokers
detection of CO-toxication of victims of fires (Fire brigades)
and much more...
Calibration protocol within scope of supply
3 Safety instructions
This device has been designed and tested in accordance to the safety regulations for electronic devices. However, its trouble-free operation and reliability cannot be guaranteed unless standard safety measures and special safety advises given in this manual will be adhered to when using it.
1. Trouble-free operation and reliability of the device can only be guaranteed if it is not subjected to any other climatic conditions than those stated under “Specification”.
2. Transporting the device from a cold to a warm environment condensation may result in a failure of the function. In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start-up.
3. The circuitry has to be designed most carefully if the device should be connected to other devices.
Internal connection in third party devices (e.g. connection GND and earth) may result in notpermissible voltages impairing or destroying the device or another device connected.
4. Whenever there may be a risk whatsoever involved in running it, the device has to be switched off immediately and to be marked accordingly to avoid re-starting. Operator safety may be a risk if:
- there is visible damage to the device or the device is not working as specified.
- the device has been stored under unsuitable conditions for a longer time.
In case of doubt, please return device to manufacturer for repair or maintenance.
5. Warning: Do not use this product as safety or emergency stop device or in other applications where failure of the product could result in personal injury or material damage. Failure to comply with these instructions could result in death or serious injury and material damage.
6. This device only serves as supervision by the monitoring of essential or other for the customer important systems.
It must not be used instead of compulsory approval monitoring devices and it is not designed for that purpose. If this device is used for the monitoring of such systems on its own, the manufacturer will not assume liability for damages whatsoever.
7. Caution, acid! The sensor contains small amounts of sulphuric acid. This can cause severe chemical burns. If leaking, avoid contact!
If there was contact:
- to skin: Flush contacted area with large amounts of water for several minutes.
- to clothing: remove contaminated clothing.
- to eyes: Flush with large amounts of water for several minutes, obtain medical
treatment.
After swallowing:
- give large volumes of water. DO NOT induce vomiting! Obtain
medical treatment.
H65.0.01.6C-07 Operating Manual GCO 100 page 4 of 10
4 Operating and maintenance advice
If the symbol "BAT" is displayed at the left side of display, the battery is weak, measuring can be continued for a short period. If „bAt“ is displayed in the display the battery is finally used up and has to be replaced. Measuring is no more possible.
The battery has to be removed, when storing device above 50°C.
Hint: We recommend removing the battery if device is not used for a longer period of time!
Risk of Leakage!
Treat device and probes carefully. Use only in accordance with above specification. (do not throw, hit against etc.).
Protect from soiling, especially in the area of the gas inlet at the front plate.
Avoid condensation at the sensor, if there was condensation, let the device dry sufficiently
Attention: Do no use solvents or silicone containing substances around the sensor opening, otherwise the measuring might be disturbed or the sensor might even be destroyed!
5 Disposal instructions
Dispense exhausted batteries at destined gathering places.
This device and the sensor must not be disposed as „residual waste‟.
According to the ElektroG (law for bringing into market, the return and the environmentally friendly disposal of electronic equipment) we accept the return of this device and/or the sensor, please send it directly to us (adequately stamped). We will dispose it appropriately and environmentally friendly.
6 Operation
6.1 Display elements
4 3 2
1
1: Main display Display for the current, frozen or max. CO-value
2: unit arrows display in ppm, mg/m
3
or %COHb
3: MAK-warning Flashes, if the allowed value for maximum working place concentration isexceeded.
(maximum exposition value for daily 8 hours work = MAK)
4: MAX-arrow The max. value since power-on will be displayed.
BAT
HLD
Press >2s: max. value will be reset
Indicates low battery
Measured value is ‚frozen„ (key 3)
6.2 Pushbuttons
Left key: on/off-button, for power-off press longer
Middle key: Max: display of the maximum measured value
(MAX-arrow in display)
Right key: Hold: hold (frozen) of the current measure
value ('HLD' in display)
6.3 Connections
1. Interface: Connect to optically isolated interface adapter
(accessory: GRS 3100, USB3100, ...)
2. Sensor opening
The mains socket is located at the left side of the instrument
H65.0.01.6C-07 Operating Manual GCO 100 page 5 of 10
7 Generals of CO-measuring
Carbon monoxide (CO) is combustible and highly toxic. It is invisible, taste- and odourless, the relative density is 0,97 (a bit lighter than air).
Already smallest concentrations can be dangerous for humans (depending of physical constitution)::
30ppm Maximum allowed working place concentration (MAK) for 8-hours of work per day
70..100ppm Flue like symptoms after some hours: sniffing nose, headaches, excoriated eyes, short breath
150...300ppm medium exposure: dizziness, sleepiness and nausea, even vomiting
400ppm
800ppm extreme exposure: headache extreme exposure: headache, dizziness, nausea and limb twitching after
45min,
1600ppm unconsciousness within 2h headache, dizziness, nausea within 45min, death after 2 hours
CO is produced when burning carbon containing substances (Wood, coal, oil, petrol, gas, cigarettes...), especially when not enough oxygen is available. In fresh uncontaminated air the COconcentration is near zero. Transportation, heating and industry exhausts are increasing CO concentrations in urban areas.
There could be measured CO also in the breath of smokers:
None smokers <6ppm
Light smoker
Smoker
6...10ppm
10...20ppm
Strong smoker >20ppm
Like with alcohol concentration in blood, also the CO is decomposed in the blood: around every 5h the CO content is cut by half. Via the CO concentration of exhaled air, the saturation of the CO in blood haemoglobin can be derived.
The carboxy-haemoglobin in % can be displayed directly by the instrument: %COHb (p.r.t.
Configuration of the instrument). This value is an estimation of the carboxy-haemoglobin of blood via the exhaled breath.
The calculation above 5ppm is done according to: M.J. Jarvis, M. Belcher, C. Vesey and D.C.S.
Hutchison, Low cost carbon monoxide monitors in smoking assessment. Thorax 41 (1986), pp.
886-887
Procedure for the measuring of exhaled gas
We suggest to use the ESA-100 adapter and the T-piece (picture at right) in combination with a suitable mouthpiece (accessories).
switch on instrument, respectively reset the max-value by pressing Max key for 2 seconds
inhale deeply and hold your breath 20 seconds, if possible (now CO migrates from the blood to the air)
exhale slowly and complete via the T-piece (preferably with mouthpiece)
read the Maximum value by pressing the Max key shortly
If no T-piece is available, an estimation can be done, by exhaling/blowing in the direction of the sensor-opening at a distance of fewer than 5 cm.
Wait sufficiently between measurings to allow the sensor to dry and to degrade CO of the previous measuring, if there was some.
H65.0.01.6C-07 Operating Manual GCO 100 page 6 of 10
8 Configuration of the instrument
To configure the instruments according to Your needs proceed like follows::
Switch off instrument.
Switch on and press max-key during the segment test ( ), until the display shows the first parameter ‚P_oF„
If a parameter should be edited, press key up or down ( ) , the setting of the parameter will be shown and can be changed via up/down keys. confirm value with
Jump to the next parameter with key
Parameter value
keys
meaning
key
Auto Power-Off
Auto Power-Off in minutes.
If there won‟t be pressed any key and no interface communication takes place for the time of the power off time setting, the device will be switched off automatically to save battery power. automatic power-off is deactivated (continuous operating)
Unit of the CO-display
Selection of the display unit “ppm” (factory setting)
Selection of the display unit “mg/m 3 “
Selection of the display unit “ %COHb“
Alarm function
Alerting over display and buzzer (factory setting)
Alerting only via display
No alarm
Alarm boundary (not available if alarm function is deactivated: AL = oFF)
Alarm boundary in the selected display unit, factory setting: 10ppm
Warning boundary
for maximum working place concentration
Value for warning boundary in the selected display unit.
Setting for Germany: 30ppm, value may vary for different countries
.
Warning for maximum working place concentration f is deactivated
Base address of the interface
Base address (refer to „the serial interface“)
Reset settings to Ex-Works
Settings are kept
Settings of configuration and adjustment menue are reset to ex-works settings
Pressing again stores the settings, the instruments restarts (segment test)
Please note: If there is no key pressed within the menu mode within 120 seconds, the configuration will be cancelled, the entered settings are lost!
H65.0.01.6C-07 Operating Manual GCO 100 page 7 of 10
9 Alarm functions
There are 3 possible settings for the alarm function: off (AL oFF), on with horn (AL on) and on without horn (AL no.So)
The alarm will be issued under the following circumstances:
- measuring value higher or equalling upper alarm limit ‟AL.Hi‟
- sensor error
- measuring values exceeding/falling below measuring ranges
- battery voltage too low or error in device (‟Err.7‟)
The alarm is supported by the interface, thus, it can be monitored by a connected computer.
The acoustic alarm is depending of the value. It is an interrupting tone, increasing in intensity with the value, above 300ppm the tone is permanent.
Visual alarming: „AL.Hi“ flashes in the display
Additionally the exceeding of the MAK value is signalled with the MAK-arrow independently from the alarm settings.
Because the maximum working place concentration differ from country to country, this value is adjustable.
10 Adjustment
The adjustment of the GCO100 is stable within 5% within in a year.
The measuring precision can be checked by means of suitable test-gas mixtures and the suitable testing accessories. Alternatively the instrument can be sent to the manufacturer for test and recalibration.
Attention! If the adjustment is altered, the included Calibration protocol is no more valid!
10.1 Adjustment menu
To adjust a measuring offset and scale proceed like follows:
Switch off instrument.
Switch on and press
“Hold”-key during the segment test ( ), until the display shows the first parameter “OFFS”
Press “up” or “down” key, the currently selected offset adjustment appears.
Choose the desired value by pressing “up” or “down” key. (max. selectable values: ±20ppm)
Enter by pressing On/Off-key: OFFS appears in the display again
Selsect the next parameter by pressing On/Off-key: SCAL appears in the display
Press “up” or “down” key, the currently selected scale appears
Choose the desired value by pressing “up” or “down” key. (28.00 ... 62.00 nA/ppm).
Store the values by pressing „On/Off‟ key. The instrument will restart (segment test).
Please note: If during the changing of the offset adjust no key is pressed within 120 seconds, the input will be aborted. Eventually made changes won’t be stored!
10.2 Automatic zero point adjusting
To compensate a deviation of the zero point proceed like following:
Switch on instrument and bring it to a CO-free atmosphere (fresh air or good vented room without any CO sources)
If there was measured at higher CO levels before, please wait at least 3 minutes.
press Hold-key 5 seconds
, until the display shows “nuLL”
The zero point adjusting is now performed automatically and will be stored in the instrument. The instrument will automatically show the measured value after successful adjustment.
If the adjusting could not be performed (display shows nuLL permanently), there is a higher COvalue in the ambient air or the sensor is defect. Turn device off and on to return to measuring
–
Note: zero point adjusting was not stored.
The zero point can be viewed in the adjustment menu at parameter “OFFS”.
H65.0.01.6C-07 Operating Manual GCO 100 page 8 of 10
10.3 Adjustment with reference gas
The instrument can be adjusted to a reference gas by means of a test gas cap and the suitable test gas bottle with gas valve. Depending on the application two concentration are offered as standard:
"GZ-02"
"GZ-03"
12l disposable test gas bottle cont. 12l of 30ppm CO
12l disposable test gas bottle cont. 12l of 300ppm CO
Ca. 10 calibrations can be done per bottle.
Additionally needed:
"GZ-04" Gas Valve unit MiniFlo
"GZ-10" test gas cap for GCO100
Preparation of device:
Perform automatic zero point adjusting first (pr.t. chapter 10.2).
The Instrument as well as the gas bottle should have adjusted to room temperature.
Set the instruments display to the unit [ppm], if another one is displayed (p.r.t. chapter 4)
Preparation of the test gas and accessories
Screw the closed gas valve unit "GZ-04 MiniFlo" on the gas bottle
Screw the test gas cap "GZ-10" to the instrument
Connect the test gas cap "GZ-10" to the valve unit via the hose
The Gas bottle has to stand upright, for the flow-display of the gas valve unit is able to work properly
Scale Adjustment
Keep bottle upright, open valve slowly, until the ball in the flow indicator shows 0.5 l/min.
(lowest marking of indicator).
Wait until the measurement display at has stabilised (ca. 1min), and write down the displayed value. Close the valve accurately an
0.5 l/min disconnect the bottle from the instrument.
Correction of scale value
If the displayed value differs from the nominal test gas concentration, the scale has to be adjusted.
For this r ead the “old” scale value ("Scal") from the adjustment menu (p.r.t. chapter 5.1)
Calculate the new scale like follows:
Scal_new = Scal_old * gas concentration / display at gas concentration
Enter Scal_new to the adjustment menu "Scal", and close the menu.
The instrument is now adjusted.
Attention:
The referring calibration protocol is then no more valid!
The ex works settings can be recalled with the "Init-YES" function of the configuration menu (p.r.t chapter 4).
H65.0.01.6C-07 Operating Manual GCO 100 page 9 of 10
11 System messages
Er. 1 =
Er. 2 =
-- =
Er. 7 = measuring range has been exceeded, measured value is to high measuring range is undershot, measured value is to low
Sensor error: Value could not be calculated
System fault - the device has detected a system fault (defective or far outside allowable ambient temperature range)
If "BAT" is displayed at the left side of display, the battery is weak, measuring can be continued for a short period.
If
“bAt” is displayed in the main display the battery is used up and needs to be replaced.
Measuring is no more possible.
12 The serial interface
By means of the serial interface and a suitable electrically isolated interface adapter (GRS3100,
GRS3105 or USB3100) the device can be connected to a computer for data transfer.
With the GRS3105 up to 5 devices of the GMH3xxx- series can be connected to one interface (see also manual of GRS3105).
To avoid transmission errors, there are several security checks implemented e.g. CRC.
The following standard software packages are available:
EBS9M: 9-channel software to display the measuring values
In case you want to develop your own software we offer a GMH3000-development package including:
a universally applicable Windows functions library ('GMH3000.DLL') with documentation
Programming examples: Visual Basic™, Delphi 1.0™, Testpoint™, EXCEL™ VBA
The device has 2 channels: Channel 1: CO[ppm], [mg/m³] or [%COHb] concentration
Channel 2: temperature
(nearly)
[°C]
Note: The measuring-/ alarm- and display range values read back from the interface are always in the selected measurement unit!
Supported functions:
Code name/function code name/function code name/function
0 Read measurement value 178 Read measuring range unit
3 Read system state
7 Read max. measured value
12 Read ID number
23 Read max alarm limit
176 Read min. measuring range
177 Read max. measuring range
202 Read display unit
204 Read display decimal point
179 Read meas. range decimal point
180 Read kind of measuring or sensor
208 Read number of channels
222 Read power-off time
199 Read kind of measuring or display
223 Set power-off time
200 Read min. display range 240 Reset
201 Read max. display range 254 Read Program version
Note for operation with GAM3000: Keep in mind that the instrument just support max-, not min-alarm ! Therefore the control-output of the GAM3000 could not be used, just the alarm output!
H65.0.01.6C-07 Operating Manual GCO 100 page 10 of 10
13 Specification
Measuring principle electrochemical CO measuring cell
Measuring range
0 ... 1000 ppm CO-concentration
Display ranges
Resolution
Reaction time
0 ... 1000 ppm CO-concentration
0 ... 1250 mg/m
3
CO- concentration conditions)
(conversion of the ppm value with factor at norm-
0 ... 60.0 % COHb (estimation of the carboxy-haemoglobin of blood via exhaled breath gas)
1 ppm, 1 mg/m
3
or 0.1 % COHb in ambient: diffusion < 50 sec
T90 at 0,5l/min gas flow: < 20sec
Life time meas. cell >5 years at proper usage at air suggested test interval: every 6months (depending on precision requirements)
Accuracy
Interference
linearity <±5 % of measured value, repeatability <±5 % of measured value (at range 0 ... 500 ppm). Calibration protocol in scope of supply
gas
hydrogen sulphide sulphur dioxide nitrogen dioxide nitric oxide chlorine hydrogen
Carbon dioxide ammonia ethanol iso-propanol acetone acetylene concentration residence time
(ppm) (minutes)
25
50
5
600
50
50
2
100
900
5
5
5
5000
100
2000
200
1000
40
5
5
30
120
5
5 display
(ppm CO)
0
<1
-1
8
0
20
0
0
80
0
0
5
Nominal temperature 25°C
Ambient condition temperature -10 ... +50°C , for short time -20 ... +50°C humiditiy 15 ... 90 %RH (non-condensing), for short time 0 ... 90 %RH
Storage temperature -10 ... +50°C
Housing impact-resistant ABS, membrane keyboard, transparent panel,
Front side IP65, integrated pop-up clip for table top or suspended use.
Dimensions
Wight
142 x 71 x 26 mm (L x W x D) approx. 155 g
Power supply
9V-battery, type IEC 6F22 (in scope of supply) as well as additional d.c. connector (diameter of internal pin 1.9 mm) for external 10.5-12V direct voltage supply. (suitable power supply: GNG10/3000)
Power consumption < 0,25mA (with standard battery > 1000 operating hours)
Display approx. 11 mm high, 4½-digit LC-display with additional elements
Pushbuttons
Alerting
3 membrane keys for on/off, menu handling, MAX/Hold-function, etc.
Hold-/Maxfunction by keypress the current measuring will be “frozen” (HLD), max. measured value will be stored
(MAX) adjustable alarm rail, value depending alarm sound, display alarm: flashing „AL.Hi“ additionally signalling of exceeding of the MAKvalue („>MAK“-arrow)
Power-Off-function the device will be automatically switched of if no key is pressed or no interface communication takes place for the time of the power-off delay.
The power-off delay can be set to values between 1 and 120 minutes or can be completely deactivated
EMC The device corresponds to the essential protection ratings established in the Regulations of the
Council for the Approximation of Legislation for the member countries regarding electromagnetic compatibility (2004/108/EG). EN61326 +A1 +A2 (appendix B, class B),
Additional fault: <1% FS.
advertisement
Key Features
- CO measuring up to 1000ppm
- Adjustable alarm function
- Optical and acoustic warning
- Display units: ppm, mg/m3, %COHb
- Data transfer via serial interface
- Automatic zero point adjusting
- Calibration protocol included