Below you will find brief information for Carbon Monoxide Measuring Instrument GCO 100. The GCO 100 is a high value CO-Measuring instrument for the measuring of up to 1000ppm (0,1Vol %) CO in ambient air. Via the adjustable alarm function the instrument warns the user optically (display) and acoustically (integrated horn) when dangerous concentrations are present. You can monitor air quality, detect CO in the breath of smokers, and detect CO-toxication of victims of fires. It comes with a calibration protocol and can be connected to a computer for data transfer.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

GCO 100 Operating Manual | Manualzz

H65.0.01.6C-07

Operating Manual

– Carbon Monoxide Measuring Instrument

as of version 1.3

GCO 100

Seite 1

GREISINGER electronic GmbH

D - 93128 Regenstauf, Hans-Sachs-Straße 26

Tel.: 09402 / 9383-0, Fax: 09402 / 9383-33, eMail: [email protected]

WEEE-Reg.-Nr. DE 93889386

H65.0.01.6C-07

Content

1

2

Operating Manual GCO 100 page 2 of 10

General advice ...................................................................................................... 2

Intended use .......................................................................................................... 3

3

4

5

6

7

Safety instructions ................................................................................................ 3

Operating and maintenance advice ..................................................................... 4

Disposal instructions ............................................................................................ 4

Operation ............................................................................................................... 4

6.1 Display elements ............................................................................................. 4

6.2 Pushbuttons .................................................................................................... 4

6.3 Connections .................................................................................................... 4

Generals of CO-measuring ................................................................................... 5

8

9

Configuration of the instrument .......................................................................... 6

Alarm functions ..................................................................................................... 7

10 Adjustment ............................................................................................................ 7

10.1 Adjustment menue ........................................................................................ 7

10.2 Automatic zero point adjusting .................................................................... 7

10.3 Adjustment with reference gas .................................................................... 8

11 System messages ................................................................................................. 9

12 The serial interface ............................................................................................... 9

13 Specification .......................................................................................................... 10

1 General advice

Read through this document attentively and make yourself familiar to the operation of the device before you use it. Keep this document in a ready-to-hand way in order to be able to look up in the case of doubt.

H65.0.01.6C-07 Operating Manual GCO 100 page 3 of 10

2 Intended use

High value CO-Measuring instrument for the measuring of up to 1000ppm (0,1Vol %) CO in ambient air.

Via the adjustable alarm function the instrument warns the user optically (display) and acoustically

(integrated horn) when dangerous concentrations are present.

The device must not used as a monitoring device for the personal safety !

The measuring inlet is at the opening of the front plate.

Applications

control and maintenance of heating systems

monitoring air quality (signalling exceeding of the maximum allowed working place concentration MAK)

detection of CO in the breath of smokers

detection of CO-toxication of victims of fires (Fire brigades)

and much more...

Calibration protocol within scope of supply

3 Safety instructions

This device has been designed and tested in accordance to the safety regulations for electronic devices. However, its trouble-free operation and reliability cannot be guaranteed unless standard safety measures and special safety advises given in this manual will be adhered to when using it.

1. Trouble-free operation and reliability of the device can only be guaranteed if it is not subjected to any other climatic conditions than those stated under “Specification”.

2. Transporting the device from a cold to a warm environment condensation may result in a failure of the function. In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start-up.

3. The circuitry has to be designed most carefully if the device should be connected to other devices.

Internal connection in third party devices (e.g. connection GND and earth) may result in notpermissible voltages impairing or destroying the device or another device connected.

4. Whenever there may be a risk whatsoever involved in running it, the device has to be switched off immediately and to be marked accordingly to avoid re-starting. Operator safety may be a risk if:

- there is visible damage to the device or the device is not working as specified.

- the device has been stored under unsuitable conditions for a longer time.

In case of doubt, please return device to manufacturer for repair or maintenance.

5. Warning: Do not use this product as safety or emergency stop device or in other applications where failure of the product could result in personal injury or material damage. Failure to comply with these instructions could result in death or serious injury and material damage.

6. This device only serves as supervision by the monitoring of essential or other for the customer important systems.

It must not be used instead of compulsory approval monitoring devices and it is not designed for that purpose. If this device is used for the monitoring of such systems on its own, the manufacturer will not assume liability for damages whatsoever.

7. Caution, acid! The sensor contains small amounts of sulphuric acid. This can cause severe chemical burns. If leaking, avoid contact!

If there was contact:

- to skin: Flush contacted area with large amounts of water for several minutes.

- to clothing: remove contaminated clothing.

- to eyes: Flush with large amounts of water for several minutes, obtain medical

treatment.

After swallowing:

- give large volumes of water. DO NOT induce vomiting! Obtain

medical treatment.

H65.0.01.6C-07 Operating Manual GCO 100 page 4 of 10

4 Operating and maintenance advice

If the symbol "BAT" is displayed at the left side of display, the battery is weak, measuring can be continued for a short period. If „bAt“ is displayed in the display the battery is finally used up and has to be replaced. Measuring is no more possible.

The battery has to be removed, when storing device above 50°C.

Hint: We recommend removing the battery if device is not used for a longer period of time!

Risk of Leakage!

Treat device and probes carefully. Use only in accordance with above specification. (do not throw, hit against etc.).

Protect from soiling, especially in the area of the gas inlet at the front plate.

Avoid condensation at the sensor, if there was condensation, let the device dry sufficiently

Attention: Do no use solvents or silicone containing substances around the sensor opening, otherwise the measuring might be disturbed or the sensor might even be destroyed!

5 Disposal instructions

Dispense exhausted batteries at destined gathering places.

This device and the sensor must not be disposed as „residual waste‟.

According to the ElektroG (law for bringing into market, the return and the environmentally friendly disposal of electronic equipment) we accept the return of this device and/or the sensor, please send it directly to us (adequately stamped). We will dispose it appropriately and environmentally friendly.

6 Operation

6.1 Display elements

4 3 2

1

1: Main display Display for the current, frozen or max. CO-value

2: unit arrows display in ppm, mg/m

3

or %COHb

3: MAK-warning Flashes, if the allowed value for maximum working place concentration isexceeded.

(maximum exposition value for daily 8 hours work = MAK)

4: MAX-arrow The max. value since power-on will be displayed.

BAT

HLD

Press >2s: max. value will be reset

Indicates low battery

Measured value is ‚frozen„ (key 3)

6.2 Pushbuttons

Left key: on/off-button, for power-off press longer

Middle key: Max: display of the maximum measured value

(MAX-arrow in display)

Right key: Hold: hold (frozen) of the current measure

value ('HLD' in display)

6.3 Connections

1. Interface: Connect to optically isolated interface adapter

(accessory: GRS 3100, USB3100, ...)

2. Sensor opening

The mains socket is located at the left side of the instrument

H65.0.01.6C-07 Operating Manual GCO 100 page 5 of 10

7 Generals of CO-measuring

Carbon monoxide (CO) is combustible and highly toxic. It is invisible, taste- and odourless, the relative density is 0,97 (a bit lighter than air).

Already smallest concentrations can be dangerous for humans (depending of physical constitution)::

30ppm Maximum allowed working place concentration (MAK) for 8-hours of work per day

70..100ppm Flue like symptoms after some hours: sniffing nose, headaches, excoriated eyes, short breath

150...300ppm medium exposure: dizziness, sleepiness and nausea, even vomiting

400ppm

800ppm extreme exposure: headache extreme exposure: headache, dizziness, nausea and limb twitching after

45min,

1600ppm unconsciousness within 2h headache, dizziness, nausea within 45min, death after 2 hours

CO is produced when burning carbon containing substances (Wood, coal, oil, petrol, gas, cigarettes...), especially when not enough oxygen is available. In fresh uncontaminated air the COconcentration is near zero. Transportation, heating and industry exhausts are increasing CO concentrations in urban areas.

There could be measured CO also in the breath of smokers:

None smokers <6ppm

Light smoker

Smoker

6...10ppm

10...20ppm

Strong smoker >20ppm

Like with alcohol concentration in blood, also the CO is decomposed in the blood: around every 5h the CO content is cut by half. Via the CO concentration of exhaled air, the saturation of the CO in blood haemoglobin can be derived.

The carboxy-haemoglobin in % can be displayed directly by the instrument: %COHb (p.r.t.

Configuration of the instrument). This value is an estimation of the carboxy-haemoglobin of blood via the exhaled breath.

The calculation above 5ppm is done according to: M.J. Jarvis, M. Belcher, C. Vesey and D.C.S.

Hutchison, Low cost carbon monoxide monitors in smoking assessment. Thorax 41 (1986), pp.

886-887

Procedure for the measuring of exhaled gas

We suggest to use the ESA-100 adapter and the T-piece (picture at right) in combination with a suitable mouthpiece (accessories).

switch on instrument, respectively reset the max-value by pressing Max key for 2 seconds

inhale deeply and hold your breath 20 seconds, if possible (now CO migrates from the blood to the air)

exhale slowly and complete via the T-piece (preferably with mouthpiece)

read the Maximum value by pressing the Max key shortly

If no T-piece is available, an estimation can be done, by exhaling/blowing in the direction of the sensor-opening at a distance of fewer than 5 cm.

Wait sufficiently between measurings to allow the sensor to dry and to degrade CO of the previous measuring, if there was some.

H65.0.01.6C-07 Operating Manual GCO 100 page 6 of 10

8 Configuration of the instrument

To configure the instruments according to Your needs proceed like follows::

Switch off instrument.

Switch on and press max-key during the segment test ( ), until the display shows the first parameter ‚P_oF„

If a parameter should be edited, press key up or down ( ) , the setting of the parameter will be shown and can be changed via up/down keys. confirm value with

Jump to the next parameter with key

Parameter value

keys

meaning

key

Auto Power-Off

Auto Power-Off in minutes.

If there won‟t be pressed any key and no interface communication takes place for the time of the power off time setting, the device will be switched off automatically to save battery power. automatic power-off is deactivated (continuous operating)

Unit of the CO-display

Selection of the display unit “ppm” (factory setting)

Selection of the display unit “mg/m 3 “

Selection of the display unit “ %COHb“

Alarm function

Alerting over display and buzzer (factory setting)

Alerting only via display

No alarm

Alarm boundary (not available if alarm function is deactivated: AL = oFF)

Alarm boundary in the selected display unit, factory setting: 10ppm

Warning boundary

for maximum working place concentration

Value for warning boundary in the selected display unit.

Setting for Germany: 30ppm, value may vary for different countries

.

Warning for maximum working place concentration f is deactivated

Base address of the interface

Base address (refer to „the serial interface“)

Reset settings to Ex-Works

Settings are kept

Settings of configuration and adjustment menue are reset to ex-works settings

Pressing again stores the settings, the instruments restarts (segment test)

Please note: If there is no key pressed within the menu mode within 120 seconds, the configuration will be cancelled, the entered settings are lost!

H65.0.01.6C-07 Operating Manual GCO 100 page 7 of 10

9 Alarm functions

There are 3 possible settings for the alarm function: off (AL oFF), on with horn (AL on) and on without horn (AL no.So)

The alarm will be issued under the following circumstances:

- measuring value higher or equalling upper alarm limit ‟AL.Hi‟

- sensor error

- measuring values exceeding/falling below measuring ranges

- battery voltage too low or error in device (‟Err.7‟)

The alarm is supported by the interface, thus, it can be monitored by a connected computer.

The acoustic alarm is depending of the value. It is an interrupting tone, increasing in intensity with the value, above 300ppm the tone is permanent.

Visual alarming: „AL.Hi“ flashes in the display

Additionally the exceeding of the MAK value is signalled with the MAK-arrow independently from the alarm settings.

Because the maximum working place concentration differ from country to country, this value is adjustable.

10 Adjustment

The adjustment of the GCO100 is stable within 5% within in a year.

The measuring precision can be checked by means of suitable test-gas mixtures and the suitable testing accessories. Alternatively the instrument can be sent to the manufacturer for test and recalibration.

Attention! If the adjustment is altered, the included Calibration protocol is no more valid!

10.1 Adjustment menu

To adjust a measuring offset and scale proceed like follows:

Switch off instrument.

Switch on and press

“Hold”-key during the segment test ( ), until the display shows the first parameter “OFFS”

Press “up” or “down” key, the currently selected offset adjustment appears.

Choose the desired value by pressing “up” or “down” key. (max. selectable values: ±20ppm)

Enter by pressing On/Off-key: OFFS appears in the display again

Selsect the next parameter by pressing On/Off-key: SCAL appears in the display

Press “up” or “down” key, the currently selected scale appears

Choose the desired value by pressing “up” or “down” key. (28.00 ... 62.00 nA/ppm).

Store the values by pressing „On/Off‟ key. The instrument will restart (segment test).

Please note: If during the changing of the offset adjust no key is pressed within 120 seconds, the input will be aborted. Eventually made changes won’t be stored!

10.2 Automatic zero point adjusting

To compensate a deviation of the zero point proceed like following:

Switch on instrument and bring it to a CO-free atmosphere (fresh air or good vented room without any CO sources)

If there was measured at higher CO levels before, please wait at least 3 minutes.

press Hold-key 5 seconds

, until the display shows “nuLL

The zero point adjusting is now performed automatically and will be stored in the instrument. The instrument will automatically show the measured value after successful adjustment.

If the adjusting could not be performed (display shows nuLL permanently), there is a higher COvalue in the ambient air or the sensor is defect. Turn device off and on to return to measuring

Note: zero point adjusting was not stored.

The zero point can be viewed in the adjustment menu at parameter “OFFS”.

H65.0.01.6C-07 Operating Manual GCO 100 page 8 of 10

10.3 Adjustment with reference gas

The instrument can be adjusted to a reference gas by means of a test gas cap and the suitable test gas bottle with gas valve. Depending on the application two concentration are offered as standard:

"GZ-02"

"GZ-03"

12l disposable test gas bottle cont. 12l of 30ppm CO

12l disposable test gas bottle cont. 12l of 300ppm CO

Ca. 10 calibrations can be done per bottle.

Additionally needed:

"GZ-04" Gas Valve unit MiniFlo

"GZ-10" test gas cap for GCO100

Preparation of device:

Perform automatic zero point adjusting first (pr.t. chapter 10.2).

The Instrument as well as the gas bottle should have adjusted to room temperature.

Set the instruments display to the unit [ppm], if another one is displayed (p.r.t. chapter 4)

Preparation of the test gas and accessories

Screw the closed gas valve unit "GZ-04 MiniFlo" on the gas bottle

Screw the test gas cap "GZ-10" to the instrument

Connect the test gas cap "GZ-10" to the valve unit via the hose

The Gas bottle has to stand upright, for the flow-display of the gas valve unit is able to work properly

Scale Adjustment

Keep bottle upright, open valve slowly, until the ball in the flow indicator shows 0.5 l/min.

(lowest marking of indicator).

Wait until the measurement display at has stabilised (ca. 1min), and write down the displayed value. Close the valve accurately an

0.5 l/min disconnect the bottle from the instrument.

Correction of scale value

If the displayed value differs from the nominal test gas concentration, the scale has to be adjusted.

For this r ead the “old” scale value ("Scal") from the adjustment menu (p.r.t. chapter 5.1)

Calculate the new scale like follows:

Scal_new = Scal_old * gas concentration / display at gas concentration

Enter Scal_new to the adjustment menu "Scal", and close the menu.

The instrument is now adjusted.

Attention:

The referring calibration protocol is then no more valid!

The ex works settings can be recalled with the "Init-YES" function of the configuration menu (p.r.t chapter 4).

H65.0.01.6C-07 Operating Manual GCO 100 page 9 of 10

11 System messages

Er. 1 =

Er. 2 =

-- =

Er. 7 = measuring range has been exceeded, measured value is to high measuring range is undershot, measured value is to low

Sensor error: Value could not be calculated

System fault - the device has detected a system fault (defective or far outside allowable ambient temperature range)

If "BAT" is displayed at the left side of display, the battery is weak, measuring can be continued for a short period.

If

“bAt” is displayed in the main display the battery is used up and needs to be replaced.

Measuring is no more possible.

12 The serial interface

By means of the serial interface and a suitable electrically isolated interface adapter (GRS3100,

GRS3105 or USB3100) the device can be connected to a computer for data transfer.

With the GRS3105 up to 5 devices of the GMH3xxx- series can be connected to one interface (see also manual of GRS3105).

To avoid transmission errors, there are several security checks implemented e.g. CRC.

The following standard software packages are available:

EBS9M: 9-channel software to display the measuring values

In case you want to develop your own software we offer a GMH3000-development package including:

a universally applicable Windows functions library ('GMH3000.DLL') with documentation

Programming examples: Visual Basic™, Delphi 1.0™, Testpoint™, EXCEL™ VBA

The device has 2 channels: Channel 1: CO[ppm], [mg/m³] or [%COHb] concentration

Channel 2: temperature

(nearly)

[°C]

Note: The measuring-/ alarm- and display range values read back from the interface are always in the selected measurement unit!

Supported functions:

Code name/function code name/function code name/function

0 Read measurement value 178 Read measuring range unit

3 Read system state

7 Read max. measured value

12 Read ID number

23 Read max alarm limit

176 Read min. measuring range

177 Read max. measuring range

202 Read display unit

204 Read display decimal point

179 Read meas. range decimal point

180 Read kind of measuring or sensor

208 Read number of channels

222 Read power-off time

199 Read kind of measuring or display

223 Set power-off time

200 Read min. display range 240 Reset

201 Read max. display range 254 Read Program version

Note for operation with GAM3000: Keep in mind that the instrument just support max-, not min-alarm ! Therefore the control-output of the GAM3000 could not be used, just the alarm output!

H65.0.01.6C-07 Operating Manual GCO 100 page 10 of 10

13 Specification

Measuring principle electrochemical CO measuring cell

Measuring range

0 ... 1000 ppm CO-concentration

Display ranges

Resolution

Reaction time

0 ... 1000 ppm CO-concentration

0 ... 1250 mg/m

3

CO- concentration conditions)

(conversion of the ppm value with factor at norm-

0 ... 60.0 % COHb (estimation of the carboxy-haemoglobin of blood via exhaled breath gas)

1 ppm, 1 mg/m

3

or 0.1 % COHb in ambient: diffusion < 50 sec

T90 at 0,5l/min gas flow: < 20sec

Life time meas. cell >5 years at proper usage at air suggested test interval: every 6months (depending on precision requirements)

Accuracy

Interference

linearity <±5 % of measured value, repeatability <±5 % of measured value (at range 0 ... 500 ppm). Calibration protocol in scope of supply

gas

hydrogen sulphide sulphur dioxide nitrogen dioxide nitric oxide chlorine hydrogen

Carbon dioxide ammonia ethanol iso-propanol acetone acetylene concentration residence time

(ppm) (minutes)

25

50

5

600

50

50

2

100

900

5

5

5

5000

100

2000

200

1000

40

5

5

30

120

5

5 display

(ppm CO)

0

<1

-1

8

0

20

0

0

80

0

0

5

Nominal temperature 25°C

Ambient condition temperature -10 ... +50°C , for short time -20 ... +50°C humiditiy 15 ... 90 %RH (non-condensing), for short time 0 ... 90 %RH

Storage temperature -10 ... +50°C

Housing impact-resistant ABS, membrane keyboard, transparent panel,

Front side IP65, integrated pop-up clip for table top or suspended use.

Dimensions

Wight

142 x 71 x 26 mm (L x W x D) approx. 155 g

Power supply

9V-battery, type IEC 6F22 (in scope of supply) as well as additional d.c. connector (diameter of internal pin 1.9 mm) for external 10.5-12V direct voltage supply. (suitable power supply: GNG10/3000)

Power consumption < 0,25mA (with standard battery > 1000 operating hours)

Display approx. 11 mm high, 4½-digit LC-display with additional elements

Pushbuttons

Alerting

3 membrane keys for on/off, menu handling, MAX/Hold-function, etc.

Hold-/Maxfunction by keypress the current measuring will be “frozen” (HLD), max. measured value will be stored

(MAX) adjustable alarm rail, value depending alarm sound, display alarm: flashing „AL.Hi“ additionally signalling of exceeding of the MAKvalue („>MAK“-arrow)

Power-Off-function the device will be automatically switched of if no key is pressed or no interface communication takes place for the time of the power-off delay.

The power-off delay can be set to values between 1 and 120 minutes or can be completely deactivated

EMC The device corresponds to the essential protection ratings established in the Regulations of the

Council for the Approximation of Legislation for the member countries regarding electromagnetic compatibility (2004/108/EG). EN61326 +A1 +A2 (appendix B, class B),

Additional fault: <1% FS.

advertisement

Key Features

  • CO measuring up to 1000ppm
  • Adjustable alarm function
  • Optical and acoustic warning
  • Display units: ppm, mg/m3, %COHb
  • Data transfer via serial interface
  • Automatic zero point adjusting
  • Calibration protocol included

Frequently Answers and Questions

What is the measuring range of the GCO 100?
The measuring range of the GCO 100 is 0 ... 1000 ppm CO-concentration.
What units can the GCO 100 display?
The GCO 100 can display measurement units in ppm, mg/m3, or %COHb.
How do I connect the GCO 100 to a computer?
The GCO 100 can be connected to a computer via a serial interface and a suitable electrically isolated interface adapter (GRS3100, GRS3105 or USB3100).