IKA® RV 05 basic

IKA® RV 05 basic
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
31 716 00
Seite 1
IKA® WERKE
IKA® RV 05 basic
RV05b 012005
DE
2
EN
9
FR
16
INDICACIONES
ES
24
NL
26
NORMI
IT
28
SV
30
DA
32
NO
34
TURVALLISUUSOHJEET
FI
36
EL
38
INSTRUÇÕES
PT
40
INSTRUCTIONS
IKA®
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
2
2
2
4
4
5
5
6
6
7
7
8
23
2
Seite 2
RV05b 012005
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 3
3
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 4
4
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 5
RV05b 012005
5
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
6
Seite 6
RV05b 012005
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 7
°
m
A
mm
kg
4,5
RV05b 012005
VAC
VAC
Hz
W
W
1/min
1/min
%
°C
%
7
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 8
8
1
2
3
5
8
11
12
13
14
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
52
53
54
55
56
57
58
61
62
64
65
66
70
71
72
5001
5002
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 9
enclosing the delivery invoice and giving reasons for the claim. You
will be liable for freight costs.
Contents
Page
EG- Declaration of conformity
Warranty
Safety instructions
Correct use
Unpacking
Assembly
Commissioning
Eliminating malfunctions
Maintenance and cleaning
Accessories
Technical data
List of spare parts RV 05 basic
Spare parts diagram RV 05 basic
9
9
9
11
11
12
12
13
13
14
14
15
23
EG-Declaration of conformity
We declare under our sole responsibility that this product corrosponds to the regulations 89/336/EG and 73/23/EG and conforms
with the standards or standardized documents DIN EN IEC 61
010-1 and DIN EN IEC 61 326-1.
Warranty
You have purchased an original IKA laboratory machine which
meets the highest engineering and quality standards.
In accordance with IKA warranty conditions, the warranty period
is 24 months. For claims under the warranty please contact your
local dealer. You may also send the machine direct to our works,
RV05b 012005
The warranty does not cover wearing parts, nor does it apply to
faults resulting from improper use or insufficient care and
maintenance contrary to the instructions in this operating manual.
If you require servicing, return the appliance in its original packaging. Storage packaging is not sufficient. Please also use suitable
transport packaging.
Safety instructions
In order to be able to use the appliance properly and safely, every
user must first read the operating instructions and observe the
safety instructions contained therein. Take care of these operating
instructions and keep them in a place where they can be accessed
by everyone.
Only trained staff who are familiar with the appliance and who are
authorised to work in this area may operate this appliance, especially under a vacuum, using flammable liquids or explosive or
noxious substances.
Please follow the relevant safety instructions and guidelines, and
occupational health and safety regulations for use in the laboratory. When working under a vacuum in particular!
When working with the device, the user must select and wear his
personal protective equipment according to the mixing hazard
category of the medium.
When using media such as toxic liquids, gases, mist, vapours or
dusts, or, in some cases, biological or microbiological substances,
the user runs the risk of coming into contact with or inhaling the
medium.
9
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Please observe the information on the hazardous substance
categories for handling solvents and toxic or biological substances.
Tasks with the RV 05 basic must only be performed when operation is monitored.
Always handle distillation fluids in the proper manner.
Operation with excess pressure is not permitted!
Adhere to the relevant regulations when working with explosive
substances and their compounds.
The device must only be used under an all side-closed exhaust,
or a comparable protective device.
The glass equipment must always be ventilated when working
under normal pressure (e.g. open outlet at cooler) in order to
prevent a pressure build-up due to extremely volatile gases or
an unknown pressure build-up. Please note that dangerous concentrations of gases, vapours or particulate matter can escape
through the outlet at the cooler. Take appropriate action to avoid
this risk, for example, downstream cold traps, gas wash bottles
or an effective extraction system.
Ensure that the appliance is on a flat, stable, non-slip surface.
Align the feet according to the appliance instructions (see
”Assembly” section). Always check before starting up that the
cantilever is fitted to the stand in accordance with the assembly instructions for drive RV 05 basic Depending on the structure
of the device, there may be a risk to the stationary stability. You
should therefore make certain the device is safe and stable each
time before placing it in service.
The appliance is designed for operation under a vacuum of up to
10 mbar. When carrying out vacuum distillation, uncondensed
vapours must be condensed out or safely dissipated. When
carrying out vacuum distillation, equipment must be evacuated
prior to heating (see ”Starting up” section). The equipment
must only be aired again after cooling. If there is a risk that the
distillation residue could disintegrate in the presence of oxygen,
only inert gas must be admitted for stress relief.
Glass parts should be checked over carefully for damage (for
example cracks) each time before placing the unit in service.
Damaged glass parts must not be used, as they can break
during operation. Glasses must be fitted without tension. When
using glass test equipment, it is important to note that bursting
can occur because of tension resulting from incorrect assembly,
because of external mechanical impact or local temperature
peaks. Evacuated glass vessels must not be heated on one side;
the evaporating flask must rotate during the heating phase.
Prevent peroxide from forming!
Liquids which are prone to forming organic peroxides must be
examined for peroxides prior to distillation and evaporation. Any
peroxides detected must be removed. Many organic compounds are prone to the formation of peroxides e.g. dekalin, diethyl ether, dioxane, tetrahydrofuran, as well as unsaturated
hydrocarbons, such as tetralin, diene, cumene and aldehydes,
ketones and solutions of these substances.
The appliance must not be operated in explosive atmospheres
or under water.
10
Seite 10
Adapt the quantity and the type of distilland to the size of the
distillation equipment. The cooler must work properly. Monitor
the coolant flow rate at the cooler outlet (see information:
”Starting up”).
ATTENTION! Organic peroxides can build up in the distillation and waste steam residue of these liquids and solutions and disintegrate explosively! Liquids which are prone to forming organic peroxides must be kept away from light, especially
UV radiation.
RV05b 012005
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 11
ATTENTION! Avoid delayed boiling!
Never heat the evaporating flask in the heating bath without
switching on the rotary drive! If there are signs that the contents
of the flask are beginning to disintegrate in the course of a chemical reaction or distillation (for example, sudden foaming or outgases) switch the heating off immediately and lift the evaporating
flask out of the heating bath using the stand. Evacuate the danger
zone and warn those in the surrounding area!
ATTENTION! Rotating parts are a source of danger!
Never rotate the evaporating flask on the raised stand.
Always lower the stand for rotation. Ensure that parts of the body,
hair, items of clothing and jewellery cannot get caught in rotating
parts.
There may be dangerous electrostatic activity between the medium and the drive unit.
Check that the filled distillation flask is balanced when setting the
speed. Imbalances can cause uncontrolled resonance behaviour in
the appliance/equipment which could damage or destroy glass
equipment! Set the drive to a speed which ensures that none of
the medium is catapulted out by the rotating evaporating flask in
the heating bath.
Correct use
The IKA RV 05 basic is designed for evating solutions and for
rapid distillation of liquids. The medium in the operating evaporator flask is circulated by rotation. The surface the liquid is increased significantly by wetting the inner surface of the flask with the
liquids. This results in a rapid exchange of heat through a large surface.
Operating in a vacuum allows for gentle evaporation.
Unpacking
Please unpack the equipment carefully and check for any damages. It is important that any damages which may have arisen
during transport are ascertained when unpacking. If applicable a
fact report must be set immediately (post, rail or forwarder).
The delivery scope covers:
One IKA RV 05 basic, one extension, one fastening screw, one
hexagon socket screw, one ring spanner and one set of operating
instructions.
Risk of burns! The motor and glass parts may be hot during
operation and for a long time afterwards. Always wear your
personal protective equipment!
If the power fails or the motor stops during operation, you must
remove the evaporating flask from the heat source immediately by
raising the stand.
The machine may only be opened by trained specialists - even
during repairs. The machine is to be unplugged from the mains
before opening. Live parts inside the machine may still be live for
some time after unplugging from the mains.
RV05b 012005
11
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Assembly
Commissioning
Stand:
Check whether the voltage specified on the type plate matches
the mains voltage available. The power socket used must be
earthed (protective earth conductor contact). If these conditions
are met, the device is ready to operate after plugging in the
mains plug. If these procedures are not followed, safe operation
cannot be guaranteed and/or the equipment may be damaged.
Observe the ambient conditions (temperature, humidity, etc.)
listed under “Technical Data”.
The stand should be pre-mounted according to directions in the
operating instructions. It should be positioned so that the arrow
mark (on the base cross) is pointing forward at the lab table. The
heat bath should be placed on the right next to the foot standing
in front.
Drive:
First, insert the extension into the bore hole on the back of the
device and screw it securely in place with the cylinder head screws enclosed.
Using an R 271 cross sleeve, fasten the drive onto the stand. To
do this, clamp the cross sleeve in place on the stand tube about
20 cm above the operating knob of the stand.
It should be easy to tip driver counter clockwise, move it into the
cross sleeve with the extension and clamp it in place. The drive
must be aligned so that it is sitting vertically above the operating
knob of the stand.
12
Seite 12
When working with the evaporator, it has proven useful to follow certain orders while placing the unit in service. These orders
are designed to largely prevent foaming, which is highly undesirable, or to at least keep it under control. If the evaporation process is running satisfactorily, a vacuum controller can cover subsequent monitoring.
The IKAVAC VC2 vacuum controller allows for recover rates of
up to 99%. Working principle: Vacuum control by assigning a
defined pressure value.
Placing the vapour tube in the drive:
Turn the ring (R) to the left unit it
will go no further. Push the
vapour tube in and then turn the
ring (R) to the right until it will go
no further.
The disassembly follows the
direction in the opposite order.
Recommended start-up sequence
Glass sets:
Drawings with lists are included with the different glass sets. The
glass sets should be mounted on the drive according to these drawings.
•Switch on the rotary drive and gradually increase the speed.
This prevents temperature stress on the glass when you start
the heating process and set the required temperature at the
heating bath.
• Set the equipment to the target pressure using the vacuum
control before filling the vacuum flask to minimise solvent losses.
(It is of course also possible to fill the vacuum flask before applying the vacuum.)
• The distilland is now sucked into the evaporating flask via the
feed pipe and the flask is immersed in the heating bath. This
is a more environmentally friendly way of filling the flask.
RV05b 012005
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 13
Note:
- Please note that the heating process takes some time.
- Only fine tune the pressure/temperature parameters once the
bath temperature has been reached.
- Please bear in mind that the cooler load must not exceed
approx. 60%. This corresponds to condensation on approx. 2/3
of the cooling coil.
- Adjust the parameters gradually when using unknown solvents
or mixtures.
- When using mixtures of substances, first be aware of the lowboiling solvents and work at a low bath temperature initially.
The IKA RV 05 basic is maintenance-free. It is subject only to the
natural wear and tear of components and their statistical failure
rate.
When ordering spare parts, please give the manufacturing
number shown on the type plate, the machine type and the
name of the spare part.
Please send in equipment for repair only after it has been cleaned and is free from any materials which may constitute a health
hazard. Please tell us if you have used any materials which are
dangerous or which may constitute a health hazard.
Eliminating malfunctions
Use only cleansing agents which have been approved by IKA to
clean IKA devices.
The drive is equipped with an overload protection device that will
turn the device off in the event of a long-term overload (motor
overheating). If the device ie turned off during operation and the
orange glow discharge light on the front of the device lights up,
the drive is overheating.
To remove use:
If the evaporator flask is immersed in the bath, it should be removed from the heat bath immediately by raising the stand. The drive
should be turned off with the On/Off switch and must be left to
cool off for some time.
Important: The device must be turned off to cool down. It will not
turn back on automatically when the motor has cooled of. The
device will not start to run again until it has been turned on again.
If the cooling off phase was too short, the orange glow discharge
light coming on when the device is turned on again will indicate
that the motore is still too hot.
RV05b 012005
Maintenance and cleaning
Dyes
Construction materials
Cosmetics
Foodstuffs
Fuels
isopropyl alcohol
water containing tenside / isopropyl alcohol
water containing tenside / isopropyl alcohol
water containing tenside
water containing tenside
For materials which are not listed, please request information
from IKA.
Wear the proper protective gloves during cleaning of the devices.
Electrical devices may not be placed in the cleansing agent for
the purpose of cleaning.
Before using another than the recommended method for cleaning or decontamination, the user must ascertain with the manufacturer that this method does not destroy the instrument.
13
3171600_RV05basic_012005
13:47 Uhr
Seite 14
IKAVAC VC 2
VC 1.1
VC 1.3
VC 2.4
MZ-2C
HB 4 basic
RV 05.5 CN
Design voltage
or
Design frequency
Input power
Output power:
Speed range 50 Hz
60 Hz
RV 05.6 CN
RV 06.1
RV 06.2
RV 06.4
RV 06.5
RV 06.6
RV 06.7
RV 06.11
RV 06.13
RV 06.15
14
23.02.2005
Vacuum controller
Water-jet pump
Vacuum solenoid valve
Pump controller
Diaphragm pump
IKA Heating bath
Set of glassware, vertical, 300 ml
(1100 cm2 cooling surface)
IKA Heating bath
Set of glassware, diagonal (1200 cm2 cooling surface)
Set of glassware, vertical (1200 cm2 cooling surface)
Evaporating flask 1 l
Evaporating flask 2 l
Evaporating flask 100 ml
Receiving flask 1 l
Rotating vapour tube, ¢ 21,6 mm
Seal (for RV 06.11), material FKM
Seal (for RV 06.11), material PTFE
VAC
VAC
Hz
W
W
rpm
rpm
Perm. duration of operation
Perm. ambient temperature
Perm. relative humidity
Protection type to DIN 60 529
Protection class
Overvoltage category
Contamination level
Perm. inclination of the device
Operation at a terrestrial altitude
%
°C
%
Fuses on mains plate
Dimensions (W x D x H)
(without attachment)
Weight
A
mm
220 - 240+10%
100 - 115+10%
50/60
133
65
46 ... 260
55 .. .312
(infinitely adjustable)
turn dial on front side
scale 1 - 10
speed table on the side
speed controlled
asynchronous motor
Self-holding temperature
switch in the motor winding (switching tempe
rature +135 °C).
malfunction display with
orange glow discharge
light
100
+5 to +40
80
IP 21
I
II
2
+ 90
max. 2000 above
sea level
4 (Id.Nr. 25 851 00)
130x200x260
kg
4,5
Speed setting
Speed display
Drive
Overload protection
°
m
RV05b 012005
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 15
List of spare parts RV 05 basic
Item
1
2
3
5
8
11
12
13
14
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
RV05b 012005
Designation
Connection line
Anti-kink protector
Cable clip
Angel bracket
PCB-power
Plastite-oval head screw
Conveyor plate
Rocker switch
Glow lamp
Grooved ball bearing
Snap ring
Shaft seal
Close-tolerance grooved pin
Application
Lock ring
Oval head screw M2,5x5
Plate spring
Bearing tube
Threaded pin M6x12
Worm wheel
V - ring SV-35
Screwed flange
O - ring
Snap ring
O - ring
Ball bearing-spacer ring
Motor support
Plate spring
A. C. motor
Cone pulley
Friction wheel
Operating knob
Scale sign
Front plate
Snap ring
Stopper rod
47
48
49
52
53
54
55
56
57
58
61
62
64
65
66
70
71
72
5001
5002
Housing bottom part
Annular spring
Screw cap
Grooved ball bearing
Worm shaft
Snap ring
Driving plate
Coupling half
Pressure spring
Condensor
Cylinder head screw M5x12
Extension
Housing upper partl
Speed sign
Connection clamp
Seal RV 06.13
Vapour tube
Ring spanner
Fan cowl
Fan wing
15
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
16
16
16
18
18
19
19
20
20
21
21
22
23
16
Seite 16
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 17
17
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 18
18
RV05b 012005
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 19
Assemblage
RV05b 012005
19
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
20
Seite 20
RV05b 012005
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 21
Accessoires
IKAVAC VC 2
VC 1.1
VC 1.3
VC 2.4
MZ-2C
HB 4 basic
RV 05.5 CN
RV 05.6 CN
RV 06.1
RV 06.2
RV 06.4
RV 06.5
RV 06.6
RV 06.7
RV 06.11
RV 06.13
RV 06.15
VAC
VAC
Hz
W
W
220 - 240 + 10%
100 - 115 + 10%
50/60
133
65
1/min
1/min
21
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 22
22
1
2
3
5
8
11
12
13
14
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
52
53
54
55
56
57
58
61
62
64
65
66
70
71
72
5001
5002
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 23
Ersatzteilbild/ Spare parts diagram / Tableau des pièces de rechange RV 05 basic
11
1
2
3
5
20 21 22 23 24
25
26
27
28
29
30
8
58
37
38
39 40
12
41 13
31
42 43
36
44
64
62
YE/GN
BU
L BN N
46
47
57
48
56
49
54
55
RV05b 012005
53 52
2
2a
45
61
35
66
1
1a
34
33 32
Motor
Glimmlampe
glow lamp
Kondensator
capacitor
23
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
24
Seite 24
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 25
25
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
26
Seite 26
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 27
27
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
28
Seite 28
RV05b 012005
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 29
29
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
30
Seite 30
RV05b 012005
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 31
31
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
32
Kontrollér, at apparatet står jævnt og stabilt på et skridsikkert
underlag. Placér fødderne i forhold til apparatet i overensstemmelse med vejledningen (se kapitlet ”Montering”). Kontrollér før
Seite 32
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 33
33
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
34
Seite 34
RV05b 012005
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 35
35
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
36
Seite 36
RV05b 012005
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 37
RV05b 012005
37
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
38
Seite 38
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 39
39
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
40
Seite 40
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 41
RV05b 012005
41
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 42
Note
42
RV05b 012005
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 43
Note
RV05b 012005
43
3171600_RV05basic_012005
23.02.2005
13:47 Uhr
Seite 44
I K A ® -WERKE GMBH & CO.KG
I K A ® Works, Inc.
I K A ® Works, (Asia) Sdn Bhd
LABORTECHNIK
ANALYSENTECHNIK
MASCHINENBAU
Europe - Middle East - Africa
North America
Asia - Australia
IKA® Works, Inc.
2635 North Chase Pkwy SE
Wilmington, NC 28405-7419
USA
TEL. +1 800 733-3037
TEL. +1 910 452-7059
FAX +1 910 452-7693
E-mail: [email protected]
IKA® Works (Asia) Sdn Bhd
Lot 2 Jalan Indah 1/2
Taman Industri Rawang Indah
48000 Rawang
Selangor, Malaysia
TEL. +60 3 6093-3322
FAX +60 3 6093-3940
E-mail: [email protected]
I K A ® Japan Y.K.
I K A ® Works Guangzhou
LABORATORY TECHNOLOGY
ANALYZING TECHNOLOGY
PROCESSING EQUIPMENT
Japan
China
IKA® Japan Y.K.
293-1 Kobayashi-cho
Yamato Koriyama Shi
639-1026 Japan
TEL. +81 74358-4611
FAX +81 74358-4612
E-mail: [email protected]
IKA® Works Guangzhou
173-175 Friendship Road
Guangzhou Economic & Technological
Development Zone
Guangzhou 510730, P.R.CHINA
TEL. +86 20 8222-6772