Unitron Shine Rev+ RIC Spec Sheet

Show PDF Add to My manuals
2 Pages

advertisement

Shine Rev+ RIC
B/2020-01 027-6395-02

©2019Unitron. All rights reserved.

Shine
Rev+ RIC
312 receiver in canal (RIC) hearing aid

Sound performance features

Auto Mic

Sound Conductor
Noise reduction
Speech enhancement

Separate directionality settings

Sound stabilization features

AntiShock 2

Feedback manager

Convenience features

DuoLink

Telecoil

Fitting

Adaptation Manager

Frequency compression

Tinnitus masker

IntelliVent

Manual programs
4
NAL-NL2/NL1 and DSLv5

Fitting channels
12
Shine Rev+ RIC is rated IP 68

Fitting guides

250 500 1000 2000 4000 8000
-10
0
10
20
30
40
50
60
70
80
90
100
110
120
Hz

Standard receiver (xS)

250 500 1000 2000 4000 8000
-10
0
10
20
30
40
50
60
70
80
90
100
110
120
Hz

Power receiver (xP)

250 500 1000 2000 4000 8000
-10
0
10
20
30
40
50
60
70
80
90
100
110
120
Hz

Super power receiver (xSP)

Open dome

Closed dome

Power dome or

sleeve mold

A Sonova brand
Shine Rev+ Standard receiver
(xS)
Power receiver
(xP)
Super power
(xSP)
ANSI 3.22 2014/IEC 60118-0: 2015 2cc coupler technical data
OSPL90
Maximum (dB SPL) 111 124 125
HFA - OSPL90 (dB SPL) 106 119 120
Full on gain (input 50 dB SPL)
Maximum (dB) 47 57 62
HFA - FOG (dB) 40 50 56
Reference test setting (RTS)
Frequency range (Hz) <100 - 8500 <100 - 7300 <100 - 5500
Reference test gain (dB) 29 42 43
Current drain at RTS (mA) 1.15 1.25 1.2
Typical battery life (h) 160 140 150
Equivalent input noise at RTS (dB SPL) 19 18 19
Total harmonic distortion at 500 Hz/800 Hz/1600 Hz/3200 Hz (%) 1.0/1.0/1.0/1.0 1.5/1.0/0.5/0.5 0.5/0.5/0.5/0.5
Induction coil sensitivity (31.6 mA/m)
HFA SPLITS/STS-RSETS (dB SPL/dB) 89/0 102/0 103/0
Standard: mic at 70 dB SPL vs induction coil at 100 mA/m
Mic
Induction Coil
Electromagnetic compatibility
EMC immunity by ANSI c63.19-2011 EMC, omni/telecoil M4/T4 M4/T4 M4/T4
Legend Test conditions
xS receiver Battery size: 312; Source: voltage 1.3 V
xP receiver The measurements obtained with a closed configuration using an HA-1 coupler (ANSI-3.7-1995) or occluded ear simulator
xSP receiver (EN 60711, coupling arrangement according to fig. 4 in the test standard). The hearing instrument set to Unitron Truefit
test settings. LLE is applied at an approximate level of 35 dB SPL. Domes should never be fit on patients with perforated
eardrums, exposed middle ear cavities, or surgically altered ear canals. In the case of such a condition,
we recommend use of a customized earmold.
We reserve the right to change specification data without notice as improvements are introduced.
80
90
100
110
120
130
140
100 1000 10000
Pout
dBSPL
Hz
60
70
80
90
100
110
120
100 1000 10000
Pout
dBSPL
Hz
10
20
30
40
50
60
70
100 1000 10000
Gain
dB
Hz
60
70
80
90
100
110
120
100 1000 10000
Pout
dBSPL
Hz
60
70
80
90
100
110
120
100 1000 10000
Pout
dBSPL
Hz

unitron.com