The Tecnomotor 810-830 Gas Analyzer is a device used to measure pollutants in vehicle exhaust gases. It measures carbon monoxide (CO), carbon dioxide (CO2), and unburned hydrocarbons (HC) using a NDIR (Non-Dispersive Infra-Red) cell, and oxygen (O2) and nitrogen oxide (optional) using electrochemical cells. The analyzer can be used in a workshop for certified check-ups or in an MCTC-NET testing line.

advertisement

Gas Analyzer 810-830
Operating manual
Gas Analyzer 810-830 __
Operating Manual
(ENGLISH)
Rev. 1.3
Manual 810-830
3
All rights reserved. No part of this user's manual can be reproduced or
transmitted by any electronic or mechanical means, including photocopies,
recording or any other storage and tracing systems for purposes other than the
purchaser's personal use without the Manufacturer's prior written authorisation.
The Manufacturer is not liable for the consequences resulting from user's
misuse of the equipment.
The data and information contained in this manual can be subject to change or
update without prior notice by the Manufacturer.
Manual 810-830
4
CONTENTS
1 Foreword _________________________________________________7
2 Package Contents__________________________________________8
3 WARRANTY _______________________________________________9
3.1 Warranty Scope _________________________________________9
3.2 Warranty Period _________________________________________9
3.3 Warranty Application______________________________________9
3.4 Material Return _________________________________________10
3.5 Exclusions and Limitations ________________________________10
4 Tester Description ________________________________________11
4.1 Reference Standards ____________________________________11
4.2 Main Elements _________________________________________12
5 Product Installation _______________________________________13
5.1 Power Supply __________________________________________13
5.2 Connect Heat Probe and Rpm Detector ______________________13
5.3 PC Connection _________________________________________13
5.4 Gas Probe Connection ___________________________________14
5.5 Identification Label ______________________________________15
5.6 Technical Specifications __________________________________17
6 GENERAL INSTRUCTIONS _________________________________18
6.1 Operating Principle ______________________________________18
6.2 Microbench Principle System ______________________________20
6.3 Intake Circuit Diagram ___________________________________20
6.4 Gas Inlet ______________________________________________21
6.5 Gas Inlet Management Diagram____________________________21
7 MAINTENANCE SERVICE___________________________________22
7.1 Cleaning ______________________________________________23
7.2 Daily Maintenance. ______________________________________23
7.3 Weekly Maintenance ____________________________________23
7.4 Six-Month Maintenance __________________________________23
7.5 Filtering System Cleaning_________________________________24
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
  • Page 1
  • Page 2
  • Page 3
  • Page 4
  • Page 5
  • Page 6
  • Page 7
  • Page 8
  • Page 9
  • Page 10
  • Page 11
  • Page 12
  • Page 13
  • Page 14
  • Page 15
  • Page 16
  • Page 17
  • Page 18
  • Page 19
  • Page 20
  • Page 21
  • Page 22
  • Page 23
  • Page 24
  • Page 25
  • Page 26
  • Page 27
  • Page 28
  • Page 29
  • Page 30
  • Page 31
  • Page 32
  • Page 33
  • Page 34
  • Page 35
  • Page 36
  • Page 37
  • Page 38
  • Page 39
  • Page 40
  • Page 41
  • Page 42
  • Page 43
  • Page 44
  • Page 45
  • Page 46
  • Page 47
  • Page 48
  • Page 49
  • Page 50
  • Page 51
  • Page 52
/

advertisement

Key Features

  • NDIR technology for CO, CO2, and HC Measurement
  • Electrochemical cells for O2 and NOx measurement
  • Workshop and MCTC-NET testing line compatibility
  • Engine rpm reading
  • Engine oil temperature reading
  • Thermal printer for test reports
  • Stand alone and MCTC-NET rev counter options
  • Comprehensive maintenance procedures provided
  • Multiple test options: Free Page, Official Procedure, Certified Workshop Check-up

Related manuals

Frequently Answers and Questions

How do I connect the gas probe to the analyser?
Connect the gas probe to the instrument gas inlet, situated on the condensate separator, as shown in the picture under 'Gas Probe Connection' section in the manual.
What are the technical specifications of the 810-830 analyser?
Refer to the 'Technical Specifications' section for details on dimensions, power supply, operating temperature, humidity, measurement fields, resolution, and weight.
What is the warranty period for the analyser?
The warranty period is 12 months from the date of delivery. Refer to the 'Warranty' section for details on coverage and exclusions.
How do I perform the 'Seal Test' on the analyser?
The 'Seal Test' is performed to check the pneumatic circuit sealing capacity. Follow the instructions displayed on the tester. You will need to close the probe and the tester will create a vacuum inside the pneumatic circuit. It then checks that the created vacuum remains within set tolerance limits.
How often should I change the oxygen sensor?
The oxygen sensor should be changed whenever the 'CHANGE OXYGEN SENSOR' message is displayed on the monitor. It is generally recommended to replace the sensor annually.