2/Scope. Ohio Medical Pediatric ISU, ISU

2/Scope. Ohio Medical Pediatric ISU, ISU

Add to My manuals
28 Pages

The ISU and Pediatric ISU are vacuum regulators that provide either intermittent or continuous suction. They are used for NG (intermittent) or pharyngeal/tracheal (continuous) suctioning throughout the hospital. The Pediatric ISU has a maximum vacuum limit of 135 mmHg (18.0 kPa), which can be adjusted between 80 and 150 mmHg (10.7 and 20.0 kPa).

advertisement

2/Scope. Ohio Medical Pediatric ISU, ISU | Manualzz

2/Scope

This service manual contains service, maintenance and parts information on four models of the ISU.

North American - ANSI ® (2)

International - ISO ® (2)

2.1 North American(ANSI) Vacuum Regulators

Note: Part numbers given are for vacuum regulators without fittings or adapters/probes.

0

(INTER)

0

0

(OFF)

(CONT)

W 40

60

20

0

M E D

120

140

160

H

G

H

Pediatric ISU

6700-1257-900

ISU

6700-1251-900

2.2 International (ISO) Vacuum Regulators

Note: Part numbers given are for vacuum regulators without fittings or adapters/probes.

6

ISU

6700-1253-900

0

(INTER)

0

0

(OFF)

160

140

180

120

100

80

20

18

16

14 12

10

8

6

4

2 24

26

F - k

Pa

60

-m m

Hg

40

20

(CONT)

Pediatric ISU

6700-1273-900

6700-0082-000 (Rev.4) 10/2012

advertisement

Key Features

  • Intermittent or continuous suction
  • Adjustable suction level
  • Adjustable timing cycles
  • Patented Unilogic Module
  • Easy to clean and sterilize
  • Pre-use checkout procedure
  • Overflow safety trap
  • High flow suction filter
  • Pediatric model with limited vacuum

Frequently Answers and Questions

What is the difference between the ISU and the Pediatric ISU?
The Pediatric ISU has a limited vacuum setting and a low gauge for use with infants and small children.
How do I adjust the suction level?
Turn the mode selector knob to | (CONT) and rotate the suction control knob until the vacuum gauge indicates the required setting.
How do I adjust the timing cycles?
The timing cycles can be adjusted using the timing valves. The adjustment procedures are outlined in Section 8.6 Timing Cycle Adjustment of this manual.
What is the pre-use checkout procedure?
The pre-use checkout procedure must be performed before using the ISU on each patient. It involves testing the regulator for correct operation, including checking the gauge reading, suction level, and timing cycles.
How do I clean and sterilize the ISU?
The ISU can be cleaned with a solution of warm water and mild detergent. The only acceptable method of sterilization is with gas (ethylene oxide). Do not steam autoclave or liquid sterilize the ISU.