4/Operation. Ohio Medical Pediatric ISU, ISU

4/Operation. Ohio Medical Pediatric ISU, ISU

Add to My manuals
28 Pages

The ISU and Pediatric ISU are vacuum regulators that provide either intermittent or continuous suction. They are used for NG (intermittent) or pharyngeal/tracheal (continuous) suctioning throughout the hospital. The Pediatric ISU has a maximum vacuum limit of 135 mmHg (18.0 kPa), which can be adjusted between 80 and 150 mmHg (10.7 and 20.0 kPa).

advertisement

4/Operation. Ohio Medical Pediatric ISU, ISU | Manualzz

4/Operation

4.1 Equipment Set-up

w WARNING: Always connect the regulator to the vacuum source and check its operation before attaching the patient connection. w WARNING: Connection to positive pressure sources such as oxygen and medical air, even momentarily could injure the patient or operator.

w CAUTION: Connection to positive pressure sources such as oxygen and medical air, even momentarily could damage the equipment.

If the regulator is equipped with an adapter/probe for wall outlets, insert the adapter/probe into the vacuum wall outlet. If the regulator is mounted elsewhere, connect a vacuum supply hose between the regulator’s adapter/ probe port and the wall outlet. Connect the collection canister or liner’s vacuum port to the regulator’s fitting or overflow safety trap with the appropriate vacuum tubing.

Note: For proper installation of adapters/probes and fittings see appendix (Appendix A-1).

For standard ISU operation, mount the collection canister or liner above the patient.

Connect the collection canister or liner’s vacuum port to the regulator’s fitting port or overflow safety trap with the appropriate suction filter and vacuum tubing.

Hospital supplied suction tubing must be used between the catheter and the collection canister or liner. The recommended minimum inside diameter is 0.25 in (6 mm). Select tubing that is collapse and kink resistant.

An Ohio Medical high flow filter and overflow safety trap should be used between the collection canister or liner and the vacuum regulator to prevent contamination of the regulator, wall outlet and pipeline system.

ISO 10079-3 (section 5.1.2) states that “the usable volume of the collection container shall be not less than 500 mL.

High Flow Suction Filters

Hydrophilic

Pkg of 20

Pkg of 200

6730-0350-800

6730-0351-800

Hydrophobic Filter

Pkg of 3

Pkg of 10

Pkg of 100

Tubing x 1/8” NPT x tubing nipple tubing nipple

6730-0570-800 6730-0580-800

6730-0571-800 6730-0581-800

6730-0572-800 6730-0582-800

Vacuum Port

Vacuum

Regulator

Connective

Tubing

Patient

Port

Disposable

Suction

Filter

Overflow

Safety

Trap

Equipment Set-Up

6700-0082-000 (Rev.4) 10/2012

Collection

Bottle

9

advertisement

Key Features

  • Intermittent or continuous suction
  • Adjustable suction level
  • Adjustable timing cycles
  • Patented Unilogic Module
  • Easy to clean and sterilize
  • Pre-use checkout procedure
  • Overflow safety trap
  • High flow suction filter
  • Pediatric model with limited vacuum

Frequently Answers and Questions

What is the difference between the ISU and the Pediatric ISU?
The Pediatric ISU has a limited vacuum setting and a low gauge for use with infants and small children.
How do I adjust the suction level?
Turn the mode selector knob to | (CONT) and rotate the suction control knob until the vacuum gauge indicates the required setting.
How do I adjust the timing cycles?
The timing cycles can be adjusted using the timing valves. The adjustment procedures are outlined in Section 8.6 Timing Cycle Adjustment of this manual.
What is the pre-use checkout procedure?
The pre-use checkout procedure must be performed before using the ISU on each patient. It involves testing the regulator for correct operation, including checking the gauge reading, suction level, and timing cycles.
How do I clean and sterilize the ISU?
The ISU can be cleaned with a solution of warm water and mild detergent. The only acceptable method of sterilization is with gas (ethylene oxide). Do not steam autoclave or liquid sterilize the ISU.