The ISU and Pediatric ISU are vacuum regulators that provide either intermittent or continuous suction. They are used for NG (intermittent) or pharyngeal/tracheal (continuous) suctioning throughout the hospital. The Pediatric ISU has a maximum vacuum limit of 135 mmHg (18.0 kPa), which can be adjusted between 80 and 150 mmHg (10.7 and 20.0 kPa).

advertisement

10/Ordering Information. Ohio Medical Pediatric ISU, ISU | Manualzz

10.1 Illustrated Parts

Parts Multi-Paks

10/Ordering Information

Kits*

Description

1. Gauge Assy. (Includes O-rings

and Lens)

North American Std, 0-200 mmHg

North American Low, 0-160 mmHg

International Std, ccw, 0-200 mmHg

International Low, ccw, 0-160 mmHg

2. Gauge Lens

3. O-ring, 2-012 Buna-N

4. Screw

5. Retaining Washer

6. Spiral Spring Washer

7. Detent Plate

8. Ball

9. Switch Main Valve Body

10. O-ring, 2-010 Buna-N

11. Regulator Module

Standard Model

Pediatric Model

12. Screw, 4-20, Ph. Hd

13. Screw, 4-40, Ph. Hd

14. Diaphragm

15. O-ring, 2-016 Buna-N

16. O-ring, 2-004 Buna-N

17. Gauge Filter

18. Flow Control Valve

19. O-ring, 2-011 Buna-N

20. Timing Valve

21. O-ring, 2-006 Buna-N

22. O-ring, 2-014 Buna-N

23. Filter

24. Unilogic Module

25. Filter, Green

26. Filter, Beige

27. Tube

28. Backplate Assy.

North American

International

28A. Backplate Assy. with Mount

Plate

North American Standard

North American Low

International Standard CE

International Low CE

29. Cover Screw

30. Nameplate Wrench

31. Cover & Knob Assy.

Standard Model

Pediatric Model

32. Mode Selector Knob

33. Suction Control Knob

Standard Model

Pediatric Model (order item 31)

34. Mounting Plate, Std ISU

Standard Model

Pediatric Model (order item 38)

35. Gasket

Standard Model

Pediatric Model (order item 38)

36. Screw

37. Positive Pressure Relief Valves

(Pediatric ISU Only)

38. Pediatric Mounting Plate

Assembly Only

39. Service Manual

Part Number

6700-0050-200

6700-0050-202

6700-0090-200

6700-0050-205

6700-0087-500

6700-0130-500

6700-0150-400

0402-0748-300

6700-0168-400

0214-3213-300

0409-1684-300

0221-5891-300

6700-0131-500

6700-1225-800

6700-0685-850

6700-0136-500

0210-0527-300

0210-6540-300

0207-0132-300

6700-0133-500

0207-8295-300

0210-0593-300

6700-0135-500

0221-5883-300

6700-0093-200

0221-5879-300

0221-5880-300

6700-0095-400

6700-0426-850

6700-0427-850

6700-0002-700

6700-0680-850

6700-0003-700

6700-0682-850

6700-0152-400

6700-0043-400

6700-0009-700

6700-0683-850

0221-5897-300

0221-5898-300

6700-0053-400

6700-0252-400

6700-0030-500

6700-0221-500

6700-0151-400

6700-0689-850

6700-0199-700

Part Number

0205-8689-870

6700-0098-200

6700-0026-700

6700-0686-850

0207-8298-870

6700-0024-700

6700-0090-700

6700-0681-850

6700-0091-700

6700-0684-850

0221-5852-870

0221-5853-870

6700-0687-850

*Kits contain the noted quantities of items in the

corresponding rows

4

4

6

6

6

4

4

4

4

4

6

6

2

2

4

2

1

2

1

2

4

2

1

1

2

4

1

4

2

2

1

6700-0082-000 (Rev.4) 10/2012

0221-5884-870 Filter Kit 6700-0030-700 Diaphragm Rpl Kit

1

1

2

1

1

1

2

1

1

25

advertisement

Key Features

  • Intermittent or continuous suction
  • Adjustable suction level
  • Adjustable timing cycles
  • Patented Unilogic Module
  • Easy to clean and sterilize
  • Pre-use checkout procedure
  • Overflow safety trap
  • High flow suction filter
  • Pediatric model with limited vacuum

Frequently Answers and Questions

What is the difference between the ISU and the Pediatric ISU?
The Pediatric ISU has a limited vacuum setting and a low gauge for use with infants and small children.
How do I adjust the suction level?
Turn the mode selector knob to | (CONT) and rotate the suction control knob until the vacuum gauge indicates the required setting.
How do I adjust the timing cycles?
The timing cycles can be adjusted using the timing valves. The adjustment procedures are outlined in Section 8.6 Timing Cycle Adjustment of this manual.
What is the pre-use checkout procedure?
The pre-use checkout procedure must be performed before using the ISU on each patient. It involves testing the regulator for correct operation, including checking the gauge reading, suction level, and timing cycles.
How do I clean and sterilize the ISU?
The ISU can be cleaned with a solution of warm water and mild detergent. The only acceptable method of sterilization is with gas (ethylene oxide). Do not steam autoclave or liquid sterilize the ISU.