10.2  Service Kits. Ohio Medical Pediatric ISU, ISU

10.2 Service Kits. Ohio Medical Pediatric ISU, ISU

Add to My manuals
28 Pages

The ISU and Pediatric ISU are vacuum regulators that provide either intermittent or continuous suction. They are used for NG (intermittent) or pharyngeal/tracheal (continuous) suctioning throughout the hospital. The Pediatric ISU has a maximum vacuum limit of 135 mmHg (18.0 kPa), which can be adjusted between 80 and 150 mmHg (10.7 and 20.0 kPa).

advertisement

10.2  Service Kits. Ohio Medical Pediatric ISU, ISU | Manualzz

10/Ordering Information

3

Description Part Number

1. Mounting Plate, ISU ................. 6700-0053-400

2. Gasket ..................................... 6700-0030-500

3. Serial Number Label

4. Screw (4 or 5) .......................... 6700-0151-400

10. Backplate Assembly ................ 6700-0043-500

(*)These parts are available as service kits. See section 10.2

.

10 4

2

1

10.2 Service Kits

1. Regulator Module Replacement

Kit ...............................................6700-0030-700

Includes the following parts:

O-ring, Stem

O-ring

Diaphragm

Stem Screw

Cap Screw (2)

These parts are available only in kit form. If other parts are needed for replacement, the entire Module must be ordered. (6700-1225-

800)

2. Complete Filter and O-ring

Replacement Kit ..........................0221-5887-300

Includes all O-rings (16) and Filters (5) for servicing the regulator

3. Complete O-ring Replacement

Kit ................................................0210-0474-870

Includes all O-rings (16) for servicing the regulator

4. Complete Filter Replacement

Kit ................................................0221-5884-870

Includes all Filters (5) for servicing the regulator

5. Mode Selector Knob Replacement

Kit ................................................0221-5897-300

Includes the following parts:

Mode Selector Knob

Extension

Hardware

6. Suction Control Knob Replacement

Kit ................................................0221-5898-300

Includes the following parts:

Suction Control Knob

Threaded Stud

Hardware

7a. ISU Backplate Replacement Kit,

North American ...........................6700-0002-700

Includes the following parts:

Backplate Assembly

Gasket

ISU Mounting Plate

Screw (4)

7b. ISU Backplate Replacement Kit,

International ................................6700-0003-700

Includes the following parts:

Backplate Assembly

Gasket

ISU Mounting Plate

Screw (4)

26 6700-0082-000 (Rev.4) 10/2012

Appendix

Installation procedure for Adapters/

Probes and Fittings

.

All adapters/probes and fittings should be sealed and installed properly to prevent leaks and to support the equipment when mounted. Both vacuum regulator ports are 1/8-27 NPTF tapered pipe threads. It is important to note that adapters/probes and fittings seal on the thread and may have threads exposed after they have been tightened properly.

w CAUTION: Do not use any Loctite ® products to seal the threads (or products which contain methacrylate ester as an active ingredient).

The torque range for installing adapters/probes and fittings is 4.0 ft-lb (5.4 N-m) minimum to 10.0 ft-lb (13.6

N-m) maximum.

Prior to installing the adapter/probe or fitting, seal the thread with Teflon ® (PTFE) tape or one of the following lubricants:

Dow Corning

Ball Vac Kote

0091-300)

® 111 (Ohio Medical P/N 6700-0074-200)

® (27951M) (Ohio Medical P/N 0220-

Adapters/probes and fittings which are not keyed for specific orientation, should be torqued to approximately

6.0 ft-lb (8.1 N-m).

Adapters/probes and fittings that are keyed to specific orientation, must be torqued initially to 4.0 ft-lbs.

Additional torque is applied only until orientation is correct.

6700-0082-000 (Rev.4) 10/2012

North America

United States

Customer Service, Distribution Center

Technical Support, Sales and Service

Equipment Service Center

Ohio Medical Corporation

1111 Lakeside Drive

Gurnee, IL 60031 USA

P: +1 866 549 6446

P: +1 847 855 0800

F: +1 847 855 6218

Ohio Medical Corporation Authorized

Representative

EC REP

OxygenCare Ltd.

2 Holfeld Business Park

Kilmacanogue Co Wicklow

Ireland

Tel: +353 1 2769700

Fax: +353 1 2764970

© 2012 Ohio Medical Corporation.

This document contains information that is proprietary and confidential to Ohio Medical Corporation. Use of this information is under license from Ohio Medical Corporation. Any use other than that authorized by Ohio Medical Corporation is prohibited.

Ohio Medical, Ohio Medical Corporation, and the Ohio Medical Corporation Logo are registered trademarks of Ohio Medical Corporation.

ISO is a registered trademark of International Organization for Standardization.

Loctite is a registered trademark of Loctite Corporation.

Teflon is a registered trademark of E.I. duPont de Demurs and Company

NFPA is a registered trademark of National Fire Protection Association, Inc.

Freon is a registered trademark of E.I. du Pont de Nemours and Company

Dow Corning is a registered trademark of Dow Corning Corporation

Vac Kote is a registered trademark of Ball Aerospace & Technology Corporation

ANSI is a registered trademark of American National Standards Institute

28 6700-0082-000 (Rev.4) 10/2012

advertisement

Key Features

  • Intermittent or continuous suction
  • Adjustable suction level
  • Adjustable timing cycles
  • Patented Unilogic Module
  • Easy to clean and sterilize
  • Pre-use checkout procedure
  • Overflow safety trap
  • High flow suction filter
  • Pediatric model with limited vacuum

Frequently Answers and Questions

What is the difference between the ISU and the Pediatric ISU?
The Pediatric ISU has a limited vacuum setting and a low gauge for use with infants and small children.
How do I adjust the suction level?
Turn the mode selector knob to | (CONT) and rotate the suction control knob until the vacuum gauge indicates the required setting.
How do I adjust the timing cycles?
The timing cycles can be adjusted using the timing valves. The adjustment procedures are outlined in Section 8.6 Timing Cycle Adjustment of this manual.
What is the pre-use checkout procedure?
The pre-use checkout procedure must be performed before using the ISU on each patient. It involves testing the regulator for correct operation, including checking the gauge reading, suction level, and timing cycles.
How do I clean and sterilize the ISU?
The ISU can be cleaned with a solution of warm water and mild detergent. The only acceptable method of sterilization is with gas (ethylene oxide). Do not steam autoclave or liquid sterilize the ISU.

Related manuals