Step 1: Unpack and Inspect the Module. Keysight PXIe System Modules and Cable
Advertisement
Advertisement
Step 1: Unpack and Inspect the Module
Step 1: Unpack and Inspect the Module
Keysight’s PXIe Modules are shipped in materials that prevent static electricity damage. The modules should only be removed from the packaging in an anti-static area ensuring that correct anti-static precautions are taken. Store all modules in anti-static envelopes when not in use.
ESD
Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe work station.
The following figure shows an example of a static-safe work station using two types of ESD protection. Purchase acceptable ESD accessories from your local supplier.
Conductive table-mat and wrist-strap combination.
Conductive floor-mat and heel-strap combination.
Both types, when used together, provide a significant level of ESD protection. Of the two, only the table-mat and wrist-strap combination provides adequate
ESD protection when used alone.
To ensure user safety, the static-safe accessories must provide at least 1 M of isolation from ground.
Inspect for damage
After unpacking the module, carefully inspect it for any shipping damage. Report any damage to the shipping agent immediately, as such damage is not covered by the warranty.
To avoid damage when handling a module; do not touch exposed connector pins.
2
3
1 After unpacking the module, verify that all items listed on the packing list are included.
Inspect the module for shipping damage.
Save all packing material for storage or return shipment to Keysight.
Keysight PXIe System and Cable Interface Installation Guide 11
Step 1: Unpack and Inspect the Module
If you need to return the module
Should it become necessary to return a Keysight module for repair or service, follow the steps below:
1 Review the warranty information shipped with your product.
2 Contact Keysight to obtain a return authorization and return address. If you need assistance finding Keysight contact information go to www.keysight.com/find/assist (worldwide contact information for repair and service) or refer to the Support information on the product web page, for example: www.keysight.com/find/M9021A .
3
4
5
Write the following information on a tag and attach it to the malfunctioning equipment.
– Name and address of owner. A Post Office box is not acceptable as a return address.
– Product model number (for example, M9021A)
– Product serial number (for example, TWXXXXXXXX). The serial number label is located on the side of the module.
– A description of failure or service required.
Carefully pack the module in its original ESD bag and carton. If the original carton is not available, use bubble wrap or packing peanuts, place the instrument in a sealed container and mark the container “FRAGILE”.
On the shipping label, write ATTENTION REPAIR DEPARTMENT and the service order number (if known).
If any correspondence is required, refer to the product by model number and serial number.
12 Keysight PXIe System and Cable Interface Installation Guide
Step 2: Verify Shipment Contents
Step 2: Verify Shipment Contents
Your shipment should have included the following:
The Keysight M902xA module that you ordered.
Controller Expansion Slot Blocker *
A driver is required for the M9022A, M9023A, and M9024A modules. Refer to
on page 19 for driver installation information. A single driver install utility services the
M9022A, M9023A, and M9024A modules. It is available on each of the Keysight product web sites: www.keysight.com/find/M9022A www.keysight.com/find/M9023A www.keysight.com/find/M9024A
The most current version of Keysight IO Libraries is required prior to installing and running any other software. The latest version can be downloaded from: www.keysight.com/find/iosuite .
A GPIB cable is NOT supplied with the M9024A System Module.
Note that the GPIB cable is a standard micro type 2 GPIB cable with standard pin wiring and is compatible with the M9037A
Embedded Controller GP-IB cable. It is available from Keysight as an accessory cable, part number Y1260A.
* Included with the M9021A, 22A, and 23A models only.
Keysight PXIe System and Cable Interface Installation Guide 13

Public link updated
The public link to your chat has been updated.
Advertisement