6 INSTALLATION
6.1 Packaging
6.2 Unpacking & installation
Instruction manual for MH & MH+M
Each MICRO-HITE or MICRO-HITE+M instrument is delivered in packaging developed to protect it from shocks and corrosion.
Only transport the height gauge in this packaging. Any other transport using unofficial packaging is not recommended and will not be covered by TESA in case of dispute.
1. Bring the pallet as close as possible from the installation area
2. Remove from pallet, accessories and panel boxes as well as documents cardboard folder.
3. Remove power supply and cables from the box.
4. Remove the masterpiece from the box and its plastic bag. Clean its base before placing it on the granite table.
30
Instruction manual for MH & MH+M
5. Remove the probe and its support from the box. Mount the probe on the support. Do not forget to tighten the assembly with the tightening wheel.
6. Accessories are now ready to be used.
7. Remove the top two protection foam pieces from pallet
8. With a second person, remove carefully the gauge from the pallet
31
32
Instruction manual for MH & MH+M
It is highly recommended not to do this step alone. Two persons are required to avoid any damage of the instrument in case of a shock or an incautious movement. Because of the weight of the instrument, it is not recommended to lift the unit on your own.
9. Carefully place the height gauge on the measuring surface keeping it in a horizontal position.
The MICRO-HITE+M are equipped with a handle directly fixed on the carriage system.
Instruction manual for MH & MH+M
This handle is used for manual fast displacement of the probe between two measurement zones. It is very fragile and no force whatsoever must be applied on it, except for the force applied for normal displacement during a measurement.
It cannot be used as a handle when moving the instrument as it could be heavily damaged or completely broken.
10. Delicately unpack the plastic cover protecting the base of the height gauge
11. Make sure you have access to the displacement lugs.
12. Remove the grease from the instrument base surface. Use a non-aggressive solvent to do so.
33
Instruction manual for MH & MH+M
13. Install the instrument vertically on the clean granite plate (or any other support).
14. Remove the protective plastic cover
15. Remove carefully the panel arm protection
34
16. Remove carefully the protective tape from the base, the handle and the gauge top’s cap.
Instruction manual for MH & MH+M
17. Remove the two screws from the front shipping bracket
18. Pull carefully the plate
19. Remove carefully the shaft protection (where the probe holder is going to be mounted).
20. Mount the probe support and its probe on the shaft.
35
Instruction manual for MH & MH+M
21. Screw the reference half-spheres to the base of the instrument
22. Unscrew the double carriage locking screw (positioned above the probe support)
36
23. Remove the panel from its box
Instruction manual for MH & MH+M
24. Screw the control panel on the articulated support.
Any cable connections have to be established while the instrument is turned off. Make sure that the instrument is turned off each time the control panel/height gauge cable is connected or disconnected.
25. Connect the control panel to the height gauge
26. Make sure that the battery is well fixed in the height gauge.
27. Connect the height gauge to a source of power using the power supply unit for direct mains operation or subsequent use (with batteries once they are charged).
37