23.4 Sensor control
Instruction manual for MH & MH+M
This page of the software enables the visualisation of certain crucial parameters of the system in order to quickly determine the status of the instrument.
23.5 Reference mark control
In order to verify the sensor and its positioning regarding the measurement scale, make sure to activate the option “Lissajous” on the right side of the screen. You should then see a circle appear on the screen, as shown in the above picture. In order to control the sensor of your instrument, you only have to move your probe slowly up or down (avoid moving to the end stop), which will lead to the display of green points on the screen. A sensor has been correctly configured when you see a centred circle appearing.
On this page, the option “reference control” has to be activated first. Then slowly move the probe up or down. It is the same principle as the initialisation of the height gauge after turning on the instrument. The encoder has to pass in front of the control mark defined on the scale
(at a height of approximately 15 cm from the base). If the mark is detected, a beep is generated and a green point is displayed on the screen.
23.6 Software update For the following procedure it is supposed that you already have the file that corresponds to the software version that you would like to load onto your instrument. If you do not have this file, please contact your local reseller.
1. Press the key in order to go to the main page of the software.
154
Instruction manual for MH & MH+M
2. Enter the service mode by clicking on the context-based option .
3. Make sure to be on the first page of the mode and to have access to the following option at the bottom of the screen . Click on it.
4. A warning message will automatically be displayed, press on in order to continue the process or in order to cancel it. If the process is not cancelled, the software will automatically close the application and open a special service mode.
5. Make sure that the software file has correctly been copied to the root folder on the USB key that you now have to connect to your control panel.
6. Press the button “1” of the numerical keyboard or click on option “Application Update” on the screen.
The software will list all available versions on your USB key and displays them in chronological order with the most recent one at top of the list. In the case above, the USB key contains 4 different software versions.
155
Instruction manual for MH & MH+M
7. Once you have chosen your version, click on the button “Update”. The software will be installed (this can take several minutes) and then warn you that the control panel will automatically shut down.
8. Wait until the control panel is automatically turned off then manually turn it on again.
9. You can now use your height gauge.
Each software version (loaded onto your control panel) is linked to firmware versions used in the electronic cards that are part of your height gauge.
When you update your height gauge with a new software version, make sure that the electronic cards do not have to be updated as well, by contacting your local reseller.
156