XPIN -BT X-ray Detector Operation Manual Detectors

XPIN -BT X-ray Detector Operation Manual Detectors

XPIN

®

-BT X-Ray Detector

Operation Manual

Appendix

Preamp Connector Pin-outs

Figure 10 Preamp Connector Pin-outs

Pin

Power Connector Standard

Internal Temperature Control

Description

Set

1

Ready Signal Open Collector

2

High Voltage +170 VDC

Detector Power - 9 VDC

3

4

5

6

Shield

Detector Power

Temperature GND

Temp Control Power

Detector Ground

+9 VDC

GND

+5 VDC

GND

Pin

Power Connector

External Temperature Control

Description

Set

1

Temperature Out (see Appendix)

2

3

4

5

6

Shield

High Voltage

Detector Power

Detector Power

Temperature GND

TEC Control Voltage

Detector Ground

+170 VDC

- 9 VDC

+9 VDC

GND

0 to 3.5 VDC

GND

Signal Connector (BNC)

Contact

Pin shield

Description

Detector Signal

Signal GND

Preamplifier Power connector part number:

Manufacture: Lemo

Part #: EPL.1S.306.HLN

Mating connector:

Manufacture: Lemo

Part #: FFA.1S.306.CLA57

DET-MAN-MED00014, Rev C Page 11

Subject to technical change without notice

XPIN

®

-BT X-Ray Detector

Operation Manual

Thermistor Temperature Sensor

The thermistor is a temperature sensitive resistor. The output of the thermistor is buffered to create feedback voltage on Pin 1 of the power connector. Refer to Table 3 and equation below for thermistor temperature vs. Pin 1 feedback voltage.

Thermistor:

Betatherm 0.3K1CG2

Thermistor Equation:

T

=

A

+

1

B

*

LN

(

R

)

+

C

*

LN

(

R

)

3

273 .

15

Where:

T = Temperature °C

R = Thermistor

Resistance Ω

A = 0.0018590668

B = 0.0002367000

C = 0.0000007811

R is the resistance of the thermistor. The resistance of the thermistor can be calculated from the voltage on pin 1 using the following equation.

DET-MAN-MED00014, Rev C Page 12

Subject to technical change without notice

XPIN

®

-BT X-Ray Detector

Operation Manual

0.473

0.489

0.505

0.522

0.540

0.558

0.576

0.363

0.375

0.388

0.401

0.414

0.428

0.443

0.457

0.246

0.254

0.262

0.271

0.280

0.289

0.298

0.308

0.319

0.329

0.340

0.351

0.596

0.616

0.636

0.657

0.679

0.702

Volts

0.184

0.190

0.196

0.203

0.209

0.216

0.223

0.230

0.238

Table 1 Temperature vs. Thermistor Voltage (Pin 1)

21

20

19

18

17

16

15

25

24

23

22

29

28

27

26

14

13

12

11

10

9

37

36

35

34

41

40

39

38

33

32

31

30

Temp °C

50

49

48

47

46

45

44

43

42

-21

-22

-23

-24

-25

-26

-27

-13

-14

-15

-16

-17

-18

-19

-20

-28

-29

-30

-31

-32

-33

-4

-5

-6

-7

0

-1

-2

-3

-8

-9

-10

-11

-12

Temp °C

8

7

6

5

4

3

2

1

1.769

1.818

1.868

1.919

1.970

2.022

2.075

1.406

1.449

1.492

1.536

1.581

1.627

1.673

1.721

2.128

2.182

2.237

2.292

2.347

2.402

1.136

1.172

1.209

1.246

1.285

1.324

1.365

0.881

0.910

0.940

0.970

1.002

1.034

1.067

1.101

Volts

0.725

0.750

0.774

0.800

0.826

0.854

-63

-64

-65

-66

-67

-68

-69

-55

-56

-57

-58

-59

-60

-61

-62

-43

-44

-45

-46

-47

-48

-49

-50

-51

-52

-53

-54

-70

Temp °C

-34

-35

-36

-37

-38

-39

-40

-41

-42

DET-MAN-MED00014, Rev C Page 13

Subject to technical change without notice

3.971

4.012

4.051

4.090

4.128

4.165

4.200

3.607

3.656

3.704

3.750

3.797

3.842

3.886

3.929

2.911

2.967

3.023

3.079

3.134

3.189

3.244

3.298

3.351

3.403

3.455

3.507

3.557

4.235

Volts

2.458

2.515

2.571

2.628

2.684

2.741

2.798

2.854

XPIN

®

-BT X-Ray Detector

Operation Manual

Mechanical Drawing of XPIN-BT Detector

DET-MAN-MED00014, Rev C Page 14

Subject to technical change without notice

XPIN

®

-BT X-Ray Detector

Operation Manual

Mechanical Drawing of Adapter Plate

DET-MAN-MED00014, Rev C Page 15

Subject to technical change without notice

XPIN

®

-BT X-Ray Detector

Operation Manual

Mechanical Drawing of XPIN-BT and Adapter Plate

DET-MAN-MED00014, Rev C Page 16

Subject to technical change without notice

XPIN

®

-BT X-Ray Detector

Operation Manual

Internal Temperature Controller Block Diagram

DET-MAN-MED00014, Rev C Page 17

Subject to technical change without notice

XPIN

®

-BT X-Ray Detector

Operation Manual

External Temperature Controller Block Diagram

DET-MAN-MED00014, Rev C Page 18

Subject to technical change without notice

Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project