2. DESCRIPTION
Preliminary remarks : Several types of action are possible on each key of the keypad, depending on whether the user presses the key briefly (short press, < 2s, validated by a beep) or at length (long press, > 2s, validated by a beep having a tone different from that of the beep emitted for a short press).
In what follows, these different actions are symbolized as follows: for a short press on the key in question for a press > 2s on the key in question
3
2
1
4
5
6
6
1 8 WAY ROTARY SWITCH :
- OFF :
- 3P :
- 2P : instrument off earth measurement with 2 auxiliary rods
AC or earth coupling resistance measurement
- REARTH / :
- ZLOOP :
- ZLINE :
- :
- SET-UP : earth measurement in a live condition, with one auxiliary rod (selective earth
if clamp connected) loop impedance measurement with 3 wires (high or low current), between phase (L) and earth (PE) loop impedance measurement with 2 wires (high current only), between two phases or between phase and neutral current measurement configuration of the instrument
Set the switch to OFF when the instrument is not in use
2 7-KEY KEYPAD :
The functions of the various keys are as follows for all the settings of the switch EXCEPT the SET-UP position (see § 3.2).
Touche 2 nd :
+ press on other key => access the secondary function of the key in question
(written in yellow italics below each key)
2 nd
=> display the current time and date as long as the key is kept pressed
Touche TEST / SMOOTH :
TEST => start/stop a measurement (except for voltage and current measurements, which are made directly) and exit from the error mode
TEST
2 nd + TEST
SMOOTH
Touche MORE / ALARM :
MORE
=>
=>
=> compensation of the measuring cables.
smooth the measurement (SMOOTH mode) display the measurements and/or complementary calculations of a function, possibly in association with the key .
2 nd + MORE
ALARM
Touche :
=> activate/deactivate the "alarm" function.
=> display the measurements and/or complementary calculations of a function, possibly in association with the key .
7
Key
2 nd
MEM or
+ MEM
MR
2 nd or
PRINT MEM
+
2 nd +
:
=> select the memory block (OBJ) or line (TEST) for storage, retrieval on screen, or printing
.
=> switch the backlighting of the display unit on/off with the switch set to ZLOOP and/or ZLINE:
=> select the type of measurement ("tripping" or
"non-tripping" mode )
=> increment the value of the memory block (OBJ) or line (TEST) 2 nd
MEM or
+ MEM
MR
2 nd or
PRINT MEM
+
Key MEM / MR :
MEM + MEM
2 nd + MEM
MR
Key PRINT (PRINT MEM) :
2 nd +
PRINT MEM
=>
=>
=>
=>
=> decrement the value of the memory block (OBJ) or line
TEST) store a measurement and all information linked to it.
display stored measurements.
print the last measurement made print the selected part of the memory (part or all)
8
advertisement
Related manuals
advertisement
Table of contents
- 4 PRESENTATION
- 5 1.1 Environmental conditions
- 5 1.2 Compliance with standards
- 5 1.3 Power supply
- 6 DESCRIPTION
- 9 GENERAL USE
- 9 3.1 Automatic checks
- 10 3.2 Instrument configuration (SET-UP)
- 11 3.3 Compensation of the leads
- 13 3.4 Recording measurement results (MEM)
- 13 3.5 Recalling recorded values (MR)
- 14 3.6 Erasing recorded values
- 15 3.7 Printing measurement results (PRINT)
- 15 3.8 Printing recorded values (PRINT MEM)
- 16 MEASUREMENTS
- 16 4.1 Voltage measurement
- 21 4.2 2P and 3P earth measurement
- 25 4.3 Earth measurement in live condition (REARTH)
- 32 4.4 3-wire loop measurement (Z LOOP)
- 35 4.5 2-wire loop measurement (Z LINE)
- 40 MAINTENANCE
- 40 6.1 Replacing the batteries
- 40 6.2 Storage
- 40 6.3 Cleaning
- 40 6.4 Metrological verification
- 40 6.5 Warranty
- 40 6.6 Customer service
- 40 6.7 Repair
- 41 LIST OF CODED ERRORS
- 41 TO ORDER