3.3 Compensation of the leads. CHAUVIN ARNOUX C.A 6456
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
Parameter
Erase memory (totally or partially)
MEM
Presses
Reference voltage for the calculation of I k
Value of the low current I
TEST in "non-tripping" measurement
Threshold voltag U
L
Alarms :
Earth resistance threshold
Loop resistance or impedance threshold
Measured current threshold
MORE x2
MORE x3
MORE x4
2 nd + MORE
2 nd + MORE x2
2 nd + MORE x3
Values see §3.6
Default values see § 4.3.2
6.9 or 12 mA see § 4.2.2
25 or 50 V voltage measured
12 mA
50 V see § 4.2.2
see § 4.4.2
see § 4.6.2
3.2.1
RESTORING THE DEFAULT CONFIGURATION
This can be used to restore the delivery configuration.
In SET- UP position:
PUSH btn
TEST dFLt
TEST for validation and programming of the default configuration
PUSH btn
3.3
COMPENSATION OF THE LEADS
The leads must be compensated for measurements of low loop and earth resistance values, in order to make the measurements more accurate.
If a compensation already exists, the symbol is displayed. To view the compensation values, use the
MORE and 4 keys.
11
3.3.1
APPLYING A COMPENSATION
Set the switch to one the LOOP/RCD positions.
Connect one end of the three-conductor cord terminated by 3 leads to the device. Short-circuit the 3 leads.
In the case of a mains outlet, connect the two earth pins with the earth pins with leads.
Perform a long press on the TEST button.
During the measurement, the device displays "LEAd" and the symbol, and the dashes blink.
At the end of the measurement, the 4 key is used to view the values of R
∆
L
, R
∆
N
and R
∆
PE
.
3.3.2
WITHDRAW A COMPENSATION
Set the switch to one of the LOOP/RCD positions.
Connect nothing to the terminals, or leave the leads open.
Then perform a long press on the TEST button.
During the measurement, the device displays "LEAd" and the symbol and the dashes blink.
To exit from compensation of the leads, perform a second long press on the TEST key.
If the compensation of the leads is effective, the symbol is lit steadily.
If the compensation has not been performed, the symbol is not displayed and the compensation values are zero.
At the end of the measurement, the device reports that the result found is greater than 5 Ω and withdraws the compensation of the leads.
For exit from compensation of the leads, perform a second long press on the TEST key.
3.3.3
POSSIBLE ERROR MESSAGES
Display - Indication
Hz
U xy
> 2V
Remark - Possible cause
The voltage on the terminals exceeds 2 V.
Check your connections.
> 5
Ω
The measured resistance of the leads is greater than 5
Ω
.
If this is not intentional, to withdraw the compensation, check your connections.
12
advertisement
Related manuals
advertisement
Table of contents
- 4 PRESENTATION
- 5 1.1 Environmental conditions
- 5 1.2 Compliance with standards
- 5 1.3 Power supply
- 6 DESCRIPTION
- 9 GENERAL USE
- 9 3.1 Automatic checks
- 10 3.2 Instrument configuration (SET-UP)
- 11 3.3 Compensation of the leads
- 13 3.4 Recording measurement results (MEM)
- 13 3.5 Recalling recorded values (MR)
- 14 3.6 Erasing recorded values
- 15 3.7 Printing measurement results (PRINT)
- 15 3.8 Printing recorded values (PRINT MEM)
- 16 MEASUREMENTS
- 16 4.1 Voltage measurement
- 21 4.2 2P and 3P earth measurement
- 25 4.3 Earth measurement in live condition (REARTH)
- 32 4.4 3-wire loop measurement (Z LOOP)
- 35 4.5 2-wire loop measurement (Z LINE)
- 40 MAINTENANCE
- 40 6.1 Replacing the batteries
- 40 6.2 Storage
- 40 6.3 Cleaning
- 40 6.4 Metrological verification
- 40 6.5 Warranty
- 40 6.6 Customer service
- 40 6.7 Repair
- 41 LIST OF CODED ERRORS
- 41 TO ORDER