3.4
RECORDING MEASUREMENT RESULTS (MEM)
IMPORTANT - Each measurement stored in the instrument is identified by 2 indices: an OBJ no. and a
TEST no.; a given object (OBJ) generally contains several TEST nos.
For example: an OBJ no. can be used to locate an installation, and the TEST nos. identify the various measurements made on this installation.
At any time, the user can store the result of a measurement and all of the parameters associated with the measurement: date, time, type of measurement, measurement parameters, etc.
The location proposed by default is the first free memory location.
measurement to be stored
MEM
03 06
OBJ TEST
FrEE (1)
selection of storage address using the following keys :
and mesure mémorisée storage at the selected address (2)
1)
"FrEE": the selected memory location is free / "OCC": the selected memory location is occupied
(2)
whether the location selected is occupied or not (previously recorded values are overwritten)
Note : Up to 100 measurements can be stored (e.g. 10 objects each having 10 tests, or any other combination).
3.5
RECALLING RECORDED VALUES (MR)
2 nd + MEM
MR
The group of measurements (OBJ) and the measurement (TEST) to be retrieved on the display unit are selected using the and keys.
13
advertisement
Related manuals
advertisement
Table of contents
- 4 PRESENTATION
- 5 1.1 Environmental conditions
- 5 1.2 Compliance with standards
- 5 1.3 Power supply
- 6 DESCRIPTION
- 9 GENERAL USE
- 9 3.1 Automatic checks
- 10 3.2 Instrument configuration (SET-UP)
- 11 3.3 Compensation of the leads
- 13 3.4 Recording measurement results (MEM)
- 13 3.5 Recalling recorded values (MR)
- 14 3.6 Erasing recorded values
- 15 3.7 Printing measurement results (PRINT)
- 15 3.8 Printing recorded values (PRINT MEM)
- 16 MEASUREMENTS
- 16 4.1 Voltage measurement
- 21 4.2 2P and 3P earth measurement
- 25 4.3 Earth measurement in live condition (REARTH)
- 32 4.4 3-wire loop measurement (Z LOOP)
- 35 4.5 2-wire loop measurement (Z LINE)
- 40 MAINTENANCE
- 40 6.1 Replacing the batteries
- 40 6.2 Storage
- 40 6.3 Cleaning
- 40 6.4 Metrological verification
- 40 6.5 Warranty
- 40 6.6 Customer service
- 40 6.7 Repair
- 41 LIST OF CODED ERRORS
- 41 TO ORDER