Log files, history files, and service files
November 2022
16 Log files, history files, and service files
16.1 Generate history files
ProLink III
Web browser
Device Tools → Configuration Transfer → Download Historical Files
Service Tools → Download Device Files
The transmitter automatically saves historical data of several types, including process and diagnostic variables, Smart Meter Verification test results, and totalizer values. To access the historical data, you can generate a log file, then view it on your PC.
WARNING
If the transmitter is in a hazardous area, do not remove the housing cover while the transmitter is powered up. Failure to follow these instructions can cause an explosion resulting in injury or death.
Prerequisites
If you want to generate a totalizer log, you must have previously configured the transmitter to record totalizer data. However, there is a totalizer history that is logged automatically.
Procedure
1. Select the type of log file that you want to generate.
2. If you selected historian data (process and diagnostic variables): a) Set the date and time for the first entry in the historian log file.
b) Set the number of days that the log file will include.
c) Select the record type.
Option
1 Second Raw Data
5 Min Average Data
Description
The current values of process and diagnostic variables, recorded at 1-second intervals.
The minimum and maximum values of the 1-second raw data over the last
5 minutes, plus the average and the standard deviation, recorded at 5-minute intervals.
The system provides an estimated file size or transfer time.
The log file is written to the specified location. File names are assigned as follows:
• Historian files: The file name is based on the transmitter tag, the starting date of the log contents, and the record type. The record type is shown as F or S:
— F=Fast, for 1-second raw data
— S=Slow, for 5-minute average data
• SMV files:
— SmvLast20Data.csv
Configuration and Use Manual 161
Log files, history files, and service files
November 2022
Configuration and Use Manual
00809-0100-1600
— SmvLongTermData.csv
• Totalizer history files: TotLog.txt
16.1.1 Historian data and log
The 1600 transmitter offers a licensable Data Historian option. If enabled, the transmitter automatically saves information about specific process and diagnostic variables. You can generate a log from this data that tracks totalizers and inventories with 21 days worth of data. The historian log is an ASCII file in .csv
format.
Contents of the historian log
There are two types of historian records:
1-second raw data The current values of process and diagnostic variables, recorded at 1-second intervals.
5-minute average data The minimum and maximum values of the 1-second raw data, plus the average and the standard deviation, calculated and recorded at 5-minute intervals.
When you generate the log, you can specify which type of record you want to see.
The historian in the transmitter internal memory contains a minimum of 7 days of 1-second raw data and
30 days of 5-minute average data.
Each record contains data for the following process and diagnostic variables:
• Timestamp
— Format: Military time
— Time and time zone: Transmitter clock
• Mass flow rate (kg/sec)
• Volume flow rate (l/sec) or GSV flow rate
• Density (g/cm ³ )
• Line temperature (°C)
• External temperature (if available)
• Pressure (if available)
• If concentration measurement is enabled:
— Standard volume flow rate
— Net mass flow rate
— Net volume flow rate
— Referred density
— Concentration
• If API Referral is enabled:
— CTPL or CTL
— Corrected density
— Corrected volume flow rate
162 Micro Motion 1600 with Ethernet Transmitters
Configuration and Use Manual
00809-0100-1600
Log files, history files, and service files
November 2022
• Alert status registers (hexadecimal format)
• Live zero (kg/sec)
• Tube frequency (Hz)
• Drive gain (%)
• Left pickoff (filtered) (V)
• Right pickoff (filtered (V)
• Left pickoff (raw) (V)
• Delta T
• Case temperature (°C)
• Voltage applied to the core processor (V)
• Temperature of the core processor board (°C)
• Temperature of the transmitter electronics (°C)
Historian data and power-cycles
Historian data is maintained across transmitter reboots and power-cycles.
Historian data and configuration files
If you restore the factory configuration or upload a configuration file, existing historian data is not affected.
Example: Historian log, 5-minute average data
MassFlow MassFlow MassFlow MassFlow S TAG:SUPPLY UID:22729F1F
SW:000000045 800:000000402
DST ON:Mountain GMT-7.0 SM:T075
SN:000000000
8/25/2020 9:58
8/25/2020 10:03
8/25/2020 10:08
8/25/2020 10:13
8/25/2020 10:18
8/25/2020 10:23
8/25/2020 10:28
8/25/2020 10:33
8/25/2020 10:38
8/25/2020 10:43
8/25/2020 10:48
8/25/2020 10:53
… kg/s Max
0.0082359
0.001018
0.00099489
0.0010835
0.0011767
0.0010243
0.0010903
0.0010319
0.0011232
0.0010841
0.0010999
0.0011523
kg/s Min
0
0.00084441
0.00086279
0.00080879
0.00084206
0.00086888
0.00084823
0.00085327
0.00088614
0.00081306
0.00086106
0.00085537
kg/s Avg
0.00091223
0.00091756
0.00092519
0.00093774
0.00094224
0.00094534
0.00094747
0.00095123
0.00095222
0.00095126
0.00095333
0.00095528
kg/s Std
9.76E-05
1.61E-05
1.44E-05
2.01E-05
2.11E-05
1.85E-05
1.81E-05
1.67E-05
1.59E-05
1.99E-05
1.93E-05
2.01E-05
…
…
…
…
…
…
…
…
…
…
…
…
…
…
Configuration and Use Manual 163
Log files, history files, and service files
November 2022
Configuration and Use Manual
00809-0100-1600
Note
The historian log displays only in English.
16.1.2 Smart Meter Verification history and log
The transmitter automatically saves test data for all Smart Meter Verification tests. You can generate a log containing data for the 20 most recent tests or for all Smart Meter Verification tests. The log is an ASCII file in .csv
format.
Contents of Smart Meter Verification log
Each record in the Smart Meter Verification log represents a Smart Meter Verification test. Each record contains the following information:
• Date and time of test
• Data collected during the test
• The abort code (15=test completed normally)
• A pass/fail result for the left pickoff (0=Pass, 1=fail)
• A pass/fail result for the right pickoff (0=Pass, 1=fail)
• The sensor type code
• The sensor serial number
Smart Meter Verification history and power-cycles
If the transmitter is rebooted or power-cycled, Smart Meter Verification history is not affected.
Smart Meter Verification history and configuration files
If you restore the factory configuration or upload a configuration file, Smart Meter Verification history is not affected.
Example: Smart Meter Verification log
Device UID: 577937183
Device Tag: SUPPLY
Time Zone: GMT -7.00
Date Time
8/13/2020 19:27
8/14/2020 7:27
8/14/2020 19:27
8/15/2020 7:27
8/18/2020 16:27
8/19/2020 19:27
8/20/2020 16:27
8/21/2020 17:10
8/22/2020 10:40
LPO Stiff
0.285876
-0.06137
0.204754
-0.15382
0.251067
-0.13654
-0.20837
-0.11062
-0.15852
RPO Stiff
0.289738
-0.05808
0.20932
-0.15216
0.251782
-0.14112
-0.20671
-0.11566
-0.16036
LPO Mass
0.155294
0.154748
0.155185
0.154612
0.155217
0.154602
0.154502
0.154641
0.154512
RPO Mass
0.158114
0.157556
0.158004
0.157416
0.158031
0.157396
0.157304
0.157435
0.157308
Damping
4.41E-05
4.02E-05
4.35E-05
3.93E-05
4.34E-05
3.89E-05
3.85E-05
3.84E-05
3.86E-05
Drv mA
1.301
1.304
1.308
1.307
1.308
1.287
1.291
1.288
1.284
…
…
…
…
…
…
…
…
…
…
164 Micro Motion 1600 with Ethernet Transmitters
Configuration and Use Manual
00809-0100-1600
Log files, history files, and service files
November 2022
8/25/2020 15:40
8/27/2020 23:16
8/28/2020 11:16
…
-0.00172
0.132787
0.04456
0.002301
0.13684
0.046158
0.154788
0.155034
0.154845
0.157599
0.15785
0.157653
4E-05
4.08E-05
3.99E-05
1.295
1.275
1.277
Note
The Smart Meter Verification log displays only in English.
16.1.3 Totalizer log
The totalizer log can track four configurable totals. The period is configurable; you can configure the transmitter to save totalizer and inventory values at a user-specified interval and then generate a totalizer log.
The totalizer log is an ASCII file.
Contents of totalizer log
The totalizer log contains one record for each logged totalizer or inventory value. Each record contains the following information:
• Default totalizer or inventory name (user-specified names are not used)
• Value and measurement unit
• Timestamp
— Format: Military time
— Time and time zone: Transmitter clock
The totalizer log also contains a line item for each totalizer or inventory reset.
Note
• The historian totalizer log is configurable, but defaults to mass and volume totalizers logged at one hour intervals.
• The entire log saves 240 lines before rolling the last lines out, so 5 days of totalizer data are saved by default.
…
…
…
Totalizer logs and power cycles
If the transmitter is rebooted or power-cycled, the totalizer log is not affected.
Totalizer logs and configuration files
If you restore the factory configuration or upload a configuration file, the totalizer log is not affected.
Example: Totalizer log
================================================================================
Device UID: 22729F1F Device Tag: SUPPLY
Name Value Units Time Zone: GMT-7.00
================================================================================
Mass Fwd Total 61.74707
grams 9/12/2020 20:00
Configuration and Use Manual 165
Log files, history files, and service files
November 2022
Configuration and Use Manual
00809-0100-1600
Mass Fwd Inv
Mass Fwd Total
Mass Fwd Inv
Mass Fwd Total
Mass Fwd Inv
Mass Fwd Total
Mass Fwd Inv
Mass Fwd Total
Mass Fwd Inv
…
61.74705
61.74707
61.74705
61.74707
61.74705
61.74707
61.74705
61.74707
61.74705
Note
The totalizer history displays only in English.
grams grams grams grams grams grams grams grams grams
9/12/2020 20:00
9/12/2020 21:00
9/12/2020 21:00
9/12/2020 22:00
9/12/2020 22:00
9/12/2020 23:00
9/12/2020 23:00
9/13/2020 0:00
9/13/2020 0:00
Read contract totals
You can read contract totals for the current 24-hour contract period and for the previous 24-hour contract period. Depending on the configuration of the totalizer log, you may be able to read contract totals for earlier periods.
The contract totals are derived from existing inventories. However, they are reset automatically at the beginning of each contract period. Therefore, the values shown will probably not match the values shown for the inventories.
Important
You can reset inventories manually, and you can stop and start inventories manually. However, if you do this, data for the current contract period will not reflect the entire 24-hour period. Data for earlier contract periods is not affected.
• The contract totals for the current contract period are stored in the Today's Total [1-4] parameters.
• The contract totals for the previous contract period are stored in the Yesterday's Total [1-4] parameters.
• The contract totals from earlier contract periods can be read in the totalizer log.
Configure contract totals into the totalizer log
The transmitter can be configured to store contract totals to the totalizer log. This allows you to access totals from earlier contract periods. Otherwise, the transmitter maintains data for only the current contract period
(today) and the immediately preceding contract period (yesterday).
Procedure
1. Navigate to the Totalizer Log .
Option
Display
Description
Menu → Configuration → Totalizer Log
166 Micro Motion 1600 with Ethernet Transmitters